MicroCal 20 DPC
Multifunction Documenting
Process Calibrator
MULTIFUNCTION
CALIBRATORS
Accuracy up to ±0.006% rdg
Light, Rugged, and Ergonomic
for Field Use
Push & Lock, TC and 4 mm
Industrial Plug Connection
Dual Channels High Accuracy
Thermometer
Internal / External Pressure Modules
Programmable Tasks
In-Line Digit Setting
Large Graphic Backlighted
Display
Simultaneous Measure and
Simulation for TRX Calibration
Real-Time Clock with Memory
for In-Field Calibration Procedures
("as found" + "as left")
Built-in Environmental
Condition Module for Ambient
T, rH% and Barometric
Pressure
CalpMan 2007
Software for Automatic
Calibration and Documenting in
Compliance with ISO9001
n
tro
Instrum
e
dQ
ste
m
Eu
ro
fie
s
nt
Q
rti
Ce
Bulletin 08-40.3 E - ed. 04-2007
ATEX Compliant Model
ATEX Ex II 1G EEx ia IIC T4
(-20°C Tamb +50°C) X
uality
Sy
www.eurotron.com
All descriptions are related to full options instrument. See latest page for the different configurations.
MULTIFUNCTION
CALIBRATORS
MicroCal 20 DPC
Introduction
Programmable generator
l
Autoramp and Autostep capability with
Start, End, and Step programmable
parameters
l
Single and continuous cycle with Start,
End, Rises, Soaks, and Falls
programmable parameters
l
the signal value setting uses a unique
in-line single-digit setting mode or a
direct numeric entry
l
direct keypad access to n.10
programmable memory stored values
The MicroCal 20 DPC (Documenting
Process Calibrator) Series are rugged
hand tools for calibration maintenance
and trouble shooting of virtually all the
control process instrumentation.
The MicroCal 20 DPC series features:
l
Calibrate temperature, pressure, current
dc, voltage dc, frequency and
resistance
l
Dual channel display for simultaneous
measure/source
l
Measure and source 14 type of
thermocouple and 10 RTD's
l
24 Vdc loop power supply (12 Vdc on IS
models)
l
Connection for internal and external
pressure module up to 700 bar
l
Pressure switch test and leak test
l
Hold, zero, scale, Minimun, Maximum
and Average
l
Automatic Ramp/Step with
programmable Time, Step and Soak
l
HART ability for Smart Transmitters
l
Dual channel input and extended
accuracy on XP models
l
Supports customized PRT's curve for
enhanced temperature measurement
l
Documenting capabilities using
CalPMan 2007 software Package
l
Scalable 4-20 mA measure/source into
effective engineering unit
End
Increment
Start
Rise
Break
Fall
Transmitter simulation program
The instrument can be used as a
temporary signal converter replacement.
Any input signal (electric or pressure) can
be converted into a 4-20 mA output. The
galvanic insulation between the input and
output channels allow also to use of this
feature on the process.
Built-in calculator
A special calculator function is integrated
in MicroCal 20 DPC. You can read the
value from the input channel, operate on
it, and then write the result on the output
channel. All standard math functions are
included.
Built-in environmental module
To compliant the ISO9000, ambient
temperature, Relative Humidity, and
Barometric pressure sensors can measure
environmental condition (EC) parameters
to be included in calibration report.
The module is not available on IS models.
Multichannel Data logging
The calibrator can be used as a
multichannel datalogger for electrical
and pressure signals. The graphic mode
allows you to display the trend; the Replay
function allows you to generate the
electrical signal using the data stored. The
LogMan PC software allows the data
storage in the hard-disk.
both the channels and the display
(including input and output values too). In
this way the work on field is simpler and
quicker.
Pressure modules (INT & EXT)
Single or dual range internal pressure
modules can be configured to provide a
lot of combinations for gauge, absolute
and differential pressure measurements.
External interchangeable pressure
module can be connected to extend the
pressure range up to 700 bar. The
calibrator includes 23 selectable pressure
units.
Transmitter Calibration
The MicroCal 20 DPC can be configured
to easily manage the check and the
calibration of any pressure and
temperature transmitter. The wide display
lets you simultaneously display the input
and output values and to select the right
units for the transmitter under test.
The current or voltage reading can be
scaled/converted in % of span or in the
engenieristic unit to simplify the
verification operations. The measuring
circuit is also able to power the loop for a
direct connection with the transmitter
under test.
With the optional communication HART
module it is also possible to verify and
calibrate smart transmitters.
All the MicroCal 20 capabilities let the
calibrator useful for all the checks and
calibration activities.
Pressure
mbar
bar
Pa
psi
mmH20
Press enter chP
100.05
ch1
P1
in
X mbar
mbar
100.000
TASK
The MicroCal 20 DPC can store and
recall up to 10 complete instrument
configurations. By pressing 2 keys only you
can store or recall the configuration of
14
All descriptions are related to full options instrument. See latest page for the different configurations.
Multifunction Calibrator
for Documenting the Verification
and the Certification activity
"Excellence in measurements"
Advanced Features
Software
Switch test
CalpMan 2007
Temperature, signal and pressure switches
can be tested using this advanced
procedure. The calibrator will hold the
display reading when the contact
changes status.
For over 20 years, Eurotron Instruments has
combined its own ISO 9001, electric,
thermoelectric and pressure measurements,
and development of software application
experiences. CalpMan 2007 has been
designed to be used industries, where there
are both laboratory and maintenance on
field needs. The CalPMan 2007 is a
WindowsTM 2000/XP software designed to
plan, manage and document all the
calibrations and the certifications of the
process instrumentation. The software can
manage the automatic procedures of the
Eurotron calibrators.
Two software versions are available: the
CalpMan 2007 and the CalpMan 2007
Advanced.
The main features are:
¡
Easy to use;
¡
Compliant with ISO9001 requirements;
¡
Standards and instruments expired date
management;
¡
Automatic procedure run
¡
Built-in communication with documenting
calibrators
¡
Traceable and editable calibration
documents
¡
Database backup/Restore ability
Leak Test
This procedure allows you to measure the
pressure fall in a programmable time
interval.
MULTIFUNCTION
CALIBRATORS
LogMan
WindowsTM software to download logged
data from internal memory to PC. Data
can be saved on disks, loaded from disks,
exported in Excel format file.
LinMan
WindowsTM software to setup the
instrument with TC, RTD special
linearization. The program allows highly
accurate temperature measurement with
a calibrated Pt100/TC loading the
coefficients of the Calibration Report.
The CalpMan 2007 Advanced has the
following advanced features:
¾
Up to 3 instruments of simultaneous
management;
¾
Advanced control of the test point;
¾
Multi-parameter procedures run;
¾
Simultaneous certification of more than
one temperature sensors;
¾
Password manager and advanced data
security ability.
Module for MicroCal T Series
FIVE STEPS FOR IN-FIELD DOCUMENTING CALIBRATION
The MicroCal 20 is able to directly control
a MicroCal T Series calibrator with this
module. It is also possible execute fully
automatic "off-line" procedures.
the procedure
2 Transfer
into the calibrator
1
Prepare the procedure
using the CalPMan
Software
OFF_Pt100
1/5
Ref:100.0°C
Ref:In 100.00 °C
Act:In 101.50 °C
Limit:
+1.00 °C
Error:
+1.50 °C
Print
the report
5
3
Select the instrument
(TAG) and run the procedure
4
Transfer the results
on a PC
The MicroCal 20 is a documenting calibrator
able to load test strategies using the
CalPMan 2007 software. It can run them and
store the results automatically, calculate the
errors, highlight the out of range points, and
download data on a PC to store them and
to generate different reports. The calibrator
internal memory can store a whole week
procedures.
15
All descriptions are related to full options instrument. See latest page for the different configurations.
MULTIFUNCTION
CALIBRATORS
MicroCal 20 DPC
LCD Graphic Display
External Interchangeable
Pressure Modules
Connection for remote "SMART" pressure
modules. Calibration matrix and range
are stored on the sensor.
Gauge, Absolute and Differential
models available.
Accuracy ±0.025% F.S.
Keypad
Large display with text and graphic
capabilities.
The rugged LCD is protected by a
polycarbonate window from scratches
and impacts.
19 key sealed rubber keypad, for
direct access to the main functions
of the instrument.
Direct Up/Down Keys
5 dedicated keys for directly
increase/decrease the value of the output
signal.
Two Internal
Pressure Sensors
Single or Dual AISI316 built-in pressure
sensors (up to 20bar).
Gauge, Absolute and Differential
models available. Barometric
reference sensor capability.
Accuracy ±0.025% F.S.
Environmental
Conditions Module (EC)
*
You can include in documenting report the
ambient temperature, pressure and the
relative humidity.
Powerful Documenting
Capability
Automatic
Backlight Sensor
Only ONE KEY to enter in Calibration
Procedure Mode.
Select the TAG and run the calibration
procedure. All procedure data are loaded
from the PC and the Calibration report can
be downloaded to the PC with the
CalpMan software.
The photoelectric sensor will detect
dark condition switching on backlight
when necessary. Manual operation is
also available.
16
*
Battery Charger
Modular for a
Total Flexibility
MULTIFUNCTION
CALIBRATORS
"Excellence in measurements"
2 Channels
Dual simultaneous IN/OUT channels.
mV, V, mA (active and passive loop),
TCs, 3/4w RTDs, Frequency, Pulse.
High accuracy, high repeatability and
low drift.
Mini-DIN
TC Connector
Integrated HART© Ability
The MicroCal 20 DPC has a built-in
option for HART calibration and maintenance.
No external adapter is required. It supports a
wide instruction set:
Four operative mode: measure,
simulate, measure/simulate and
measure/measure. (XP model only)
Universal Commands available for all process
instruments like the model and the
manufacturer reading, the primary variable (PV)
and the output current reading.
Isothermic binding post for
TC's with Rj compensation.
Advanced Commands available on many (but
not for all) instruments like the multiple variables
and the damping reading, and the run of the
Loop Test.
Specific Commands available on few
instruments (verify the updated list on the
website) like the sensor trim. The transmitter
upgrade can be run in a simple and quick way
using the PC software and the RS232/USB serial
cable.
Capabilities
Basic
Plus
XP
•
•
•
•
•
•
±0.02%
±0.01%
±0.006%
•
•
•
•
•
•
•
•
•
•
•
mA, V, T/C's, RTDs, Hz, W
Measure / Source
Basic Accuracy (rdg)
Dual Channel Display
•
•
Dual Input
INT pressure module
EXT pressure module
Loop P.S.
HART ability
Documenting capability
EC Module
RS232 interface
Remote Rj input
Digital HART remote input
Switch test external contact input
*
*
*
•
•
•
•
•
•
•
•
Intrinsecally Safe
The Intrinsecally Safe for zone 0 with ATEX
certification, class II 1G EEx ia IIC T4
(-20°C TAmb + 50°C) X is available for
IS model.
* Not available on IS model
17
All descriptions are related to full options instrument. See latest page for the different configurations.
MULTIFUNCTION
CALIBRATORS
MicroCal 20 DPC
Specifications
Measure or Source Voltage
Input impedance:
>10 MW
for ranges up to 2000 mV f.s.
>500 kW
for ranges up to 20 V f.s.
Output impedance (emf output):
less than 0.5W
with a maximum current of 0.5
mA
Output noise (at 300 Hz):
<2 µVpp for ranges up to 200 mV f.s.,
<10 µVpp for ranges up to 2000 mV f.s.
<80 µVpp for ranges up to 20 V f.s.
Report of Calibration
Each MicroCal is factory calibrated and
certified against Eurotron Standards,
which are periodically certified by an
Internationally recognised Laboratory to
ensure traceability, and shipped with a
Report of Calibration stating the nominal
and actual values and the deviation
errors.
Firmware
The firmware is stored on a flash memory
and allows a fast and easy upgrade of
the instrument using the USB cable and
the STFlash software.
Measure or Source Current
Input impedance: <20 W
at 1 mA
Maximum load resistance:
1000W
at 20 mA
600W
at 21 mA (IS model)
Loop Supply: 24V ± 5 %
12V ± 5 % (IS model)
Over-Voltage protection
The unit is equipped with an advanced
system including thermal fuse (auto repair,
do not need replacement), high voltage
suppressor and resistor-diode voltage
limiter.
Measure or Source Resistance and RTDs
Connections: 2, 3 and 4 wires
Source resistance effects: ±1 µV error for
1000W
source resistance
Rtd and W
simulation excitation current: from
0.100 to 4 mA without incremental error
Rtd and W
measurement excitation current:
0.2 mA
Rtd cable compensation: up to 100 mW
(for
each wire)
Rtd cable compensation error (Pt100):
±0.005°C/ W
of total wire
Maximum load resistance: 1000 W
at 20 mA
600 W
at 20 mA (IS model only)
EMC Conformity
The instrument fulfils the prevision of the
directive 89/336/CEE Electromagnetic
Compatibility.
Quality system
Research, development, production,
inspection and certification activities are
defined by methods and procedures of
the Eurotron Quality System inspected for
compliance and certified ISO9001 by
GASTEC, a Dutch notified body.
Measure or Source Thermocouples
Engineering unit: °C/°F/K selectable
Resolution: 0.01°C / 0.01°F
Temperature scale: ITS90 and IPTS68
selectables
Reference junction compensation:
internal automatic from -10 °C to +55 °C
external adjustable from -50 °C to +100 °C
remote with external Pt100 from -10°C to +100 °C
(only on XP model)
Rj compensation drift: ± 0.002°C/°C (from -10
°C to +45 °C) - Class A Pt100
Input impedance (Tc ranges): >10 MW
"Push & Lock" binding posts
The multi-connection binding posts is an exclusive
project designed to connect the calibrator in a
simpler and faster way. The 3 different connection
system available are:
¾
Standard 4 mm industrial plugs
¾
Mini isothermic TC's connectors
¾
Push & Lock system for wires
Frequency
Input impedance:>500KW
Pressure
Pressure media: AISI 316 SS compatible fluids
(water, gas, and oil)
Temperature compensation: Automatic with
built-in calibration matrix.
Engineering units: mbar, bar, Pa, hPa, kPa, MPa,
kg/cm2, kg/m2, psi, mmH2O, cmH2O, mH2O, Torr,
2
atm, lb/ft , inH2O, ftH2O, mmHg, cmHg, mHg,
inHg, programmable.
Accuracy: the above accuracies are stated for
365 days and includes non linearity, histeresis,
and repeatability. The average temperature
coefficient, inside the temperature
compensated range, is ±0.002% of rdg/°C (w.t.r.
+23°C/+73°F).
18
Compensation temperature range:
+0 to +45°C (+32°F +113°F)
Internal sensors
Accuracy: ±0.025% F.S.
Ranges: see table on ordering code
Resolution: see table on ordering code
Overpressure: 125% F.S.
Port: (female) 1/8" BSP
External modules
Accuracy: ±0.025% F.S.
Ranges: see table on ordering code
Resolution: see table on ordering code
Overpressure: 125% F.S.
Port: (male) 1/4" BSP
Connection wire lenght: 2 meters
Advanced Functions
Calculation functions: hold, max, min, offset,
zero, average
In/Out data memory: 10 data with manual or
automatic recall
Convert function: displays the electrical
equivalent of the engineering unit
Scale factor: setting with zero and span
programmable within -399999 and +999999
Square root: in combination with scale factor
General Specifications
Calibration: self learning technique with
automatic procedure
Channel 1-Channel 2 insulation: 250 Vdc
Common mode rejection: 140 dB at ac
operation
Normal mode rejection:60 dB at 50/60 Hz
Display: graphic LCD display with automatic and
manual backlight device (not available on IS
model)
Measurement sampling time: 250 ms
Digital interface: full bidirectional RS232
Power supply: external charger and
rechargeable Ni-MH battery
Battery life (typical):
10 h (8 h on IS model) on Tc and mV input/output
(backlight Off)
4 h (3 h on IS model) with 20 mA simulation
(backlight Off)
Recharging time (typical): 5 h (8 h on IS model)
at 90% and 6 h (10 h on IS model) at 99% with
instrument switched off.
Battery charge indication: bar graph on the LCD
display (flashing on charge) - not available on IS
model
Line operation: 100V - 120 V - 230V - 240 Vac
with the external battery charger
Line transformer insulation: 2500 Vac
Operating environment temperature range:
from -10 °C to +55 °C (from -10°C to +50°C on IS
model)
Storage temperature range: from 0 °C to +60
°C (excluding batteries)
Humidity: max 95%RH non condensing
Case: Injection moulded policarbonate case
(injection molded ATEX approved material on IS
model)
Sealing: IP54
Weights: nett 1.4 Kg gross 2.5 Kg
Dimensions: 290x98x57 m
All descriptions are related to full options instrument. See latest page for the different configurations.
Multifunction Calibrator
for Documenting the Verification
and the Certification activity
"Excellence in measurements"
MULTIFUNCTION
CALIBRATORS
Ranges and Accuracy
Measure or
Source
Tc J
Range
-210 to 1200°C
-350 to 2200°F
Tc K
-270 to 1370°C
-454 to 2500°F
Tc T
-270 to 400°C
-454 to 760°F
Tc R
-50 to 1760°C
-60 to 3200°F
Tc S
-50 to 1760°C
-60 to 3200°F
Tc B
50 to 1820°C
140 to 3310°F
Tc C
0 to 2300°C
32 to 4170°F
Tc G
0 to 2300°C
32 to 4170°F
Tc D
0 to 2300°C
32 to 4170°F
Tc U
-200 to 400°C
-330 to 760°F
Tc L
-200 to 760°C
-330 to 1400°F
Tc N
-270 to 1300°C
-450 to 2380°F
Tc E
-270 to 1000°C
-450 to 1840°F
Tc F
0 to 1400°C
32 to 2560°F
Pt100 IEC
-200 to 850°C
OIML, a
=.3926 -330 to 1570°F
Pt100
-200 to 650°C
a
=.3902
-330 to 1210°F
Pt100
-200 to 600°C
JIS SAMA
-330 to 1120°F
Pt200
-200 to 850°C
-330 to 1570°F
Pt500
-200 to 850°C
-330 to 1570°F
Pt1000
-200 to 850°C
IEC OIML
-330 to 1570°F
Cu10
-70 to 150°C
-100 to 310°F
Cu100
-180 to 150°C
-300 to 310°F
Ni100
-60 to 180°C
-80 to 360°F
Ni120
0 to 150°C
32 to 310°F
mV
-20 to 200mV
V
-0.2 to 2V
-2 to 20V***
mA
0 to 50mA**
mA (IN CH1)
-5 to 50mA
W
(IN)
0 to 500W
0 to 5000W
W
(OUT)****
0 to 500W
0 to 5000W
Frequency
1 to 200 Hz
1 to 2 kHz
1 to 20 kHz
Pulse
0 to 106
Accuracy
Resolution
MicroCal 20 DPC basic
0.01 °C*
0.01 °F
0.01 °C*
0.01 °F
0.01°C*
0.01 °F
0.1°C
0.1°F
0.1°C
0.1°F
0.1°C
0.1°F
0.1°C
0.1°F
0.1°C
0.1°F
0.1°C
0.1°F
0.01°C
0.01°F
0.01°C
0.01°F
0.01°C
0.01°F
0.1°C
0.1°F
0.1°C
0.1°F
0.01°C
0.01 °F
0.01°C
0.01 °F
0.01°C
0.01 °F
0.1°C
0.1°F
0.1°C
0.1°F
0.01°C
0.01°F
0.1°C
0.1°F
0.1°C
0.1°F
0.1°C
0.1°F
0.1°C
0.1°F
1µV
10µV
100µV
0.1µA
0.1µA
10mW
100mW
10mW
100mW
0.001Hz
0.01Hz
0.1Hz
1 count
MicroCal 20 DPC plus
MicroCal 20 DPC XP
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.2°C)
±(0.01% rdg. + 0.2°C)
±(0.01% rdg. + 0.2°C)
±(0.02% rdg. + 0.4°F)
±(0.01% rdg. + 0.4°F)
±(0.01% rdg. + 0.4°F)
±(0.02% rdg. + 0.2°C)
±(0.01% rdg. + 0.2°C)
±(0.01% rdg. + 0.2°C)
±(0.02% rdg. + 0.4°F)
±(0.01% rdg. + 0.4°F)
±(0.01% rdg. + 0.4°F)
±(0.02% rdg. + 0.3°C)
±(0.01% rdg. + 0.3°C)
±(0.01% rdg. + 0.3°C)
±(0.02% rdg. + 0.6°F)
±(0.01% rdg. + 0.6°F)
±(0.01% rdg. + 0.6°F)
±(0.02% rdg. + 0.2°C)
±(0.01% rdg. + 0.2°C)
±(0.01% rdg. + 0.2°C)
±(0.02% rdg. + 0.4°F)
±(0.01% rdg. + 0.4°F)
±(0.01% rdg. + 0.4°F)
±(0.02% rdg. + 0.3°C)
±(0.01% rdg. + 0.3°C)
±(0.01% rdg. + 0.3°C)
±(0.02% rdg. + 0.6°F)
±(0.01% rdg. + 0.6°F)
±(0.01% rdg. + 0.6°F)
±(0.02% rdg. + 0.3°C)
±(0.01% rdg. + 0.3°C)
±(0.01% rdg. + 0.3°C)
±(0.02% rdg. + 0.6°F)
±(0.01% rdg. + 0.6°F)
±(0.01% rdg. + 0.6°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.02% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.02% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.02% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.02% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.02% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.02% rdg. + 0.15°C)
±(0.01% rdg. + 0.15°C)
±(0.01% rdg. + 0.15°C)
±(0.02% rdg. + 0.27°F)
±(0.01% rdg. + 0.27°F)
±(0.01% rdg. + 0.27°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.01% rdg. + 0.1°C)
±(0.02% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.01% rdg. + 0.2°F)
±(0.02% rdg. + 0.4°C)
±(0.01% rdg. + 0.4°C)
±(0.01% rdg. + 0.4°C)
±(0.02% rdg. + 0.7°F)
±(0.01% rdg. + 0.7°F)
±(0.01% rdg. + 0.7°F)
±(0.02% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.02% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.02% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.02% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.02% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.01% rdg. + 0.05°C)
±(0.02% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.01% rdg. + 0.09°F)
±(0.02% rdg. + 3 µV)
±(0.01% rdg. + 3 µV)
±(0.006% rdg. + 3 µV)
±(0.02% rdg. + 10 µV)
±(0.01% rdg. + 10 µV)
±(0.006% rdg. + 10 mV)
±(0.02% rdg. + 100 µV)
±(0.01% rdg. + 100 µV)
±(0.006% rdg. + 100 µV)
±(0.02% rdg. + 0.4µA)
±(0.01% rdg. + 0.4µA)
±(0.01% rdg. + 0.4µA)
±(0.02% rdg. + 0.4µA)
±(0.01% rdg. + 0.4µA)
±(0.01% rdg. + 0.4µA)
±(0.02% rdg. + 12mW
)
±(0.01% rdg. + 12mW
)
±(0.008% rdg. + 12mW
)
±(0.02% rdg. + 120mW
)
±(0.01% rdg. + 120mW
)
±(0.008% rdg. + 120mW
)
±(0.02% rdg. + 20mW
)
±(0.01% rdg. + 20mW
)
±(0.008% rdg. + 20mW
)
±(0.02% rdg. + 200mW
)
±(0.01% rdg. + 200mW
)
±(0.008% rdg. + 200mW
)
±(0.005% rdg.+0.001Hz)
±(0.005% rdg.+0.001Hz)
±(0.005% rdg.+0.001Hz)
±(0.005% rdg.+0.01Hz)
±(0.005% rdg.+0.01Hz)
±(0.005% rdg.+0.01Hz)
±(0.005% rdg.+0.1Hz)
±(0.005% rdg.+0.1Hz)
±(0.005% rdg.+0.1Hz)
NOTES:
The relative accuracies shown above are stated for 360 days and the operative conditions are
from 18 to 28°C
Typical 2 year relative accuracy can be estimated by multiplying the "% of reading"
specifications by 1.4.
All input ranges: additional error ±1 digit.
Eurotron traceability chart and uncertainty can be supplied on request.
* Resolution is 0.1°C with temperature lower than -200°C.
** 21mA max. on passive current loop, and for the IS model.
*** IS model channel 2 source Max 11 V.
**** IS model ohm generation: additional error: ±20 mW
on 500 W
range
±200 mW
on 5 kW
range
19
All descriptions are related to full options instrument. See latest page for the different configurations.
MicroCal 20 DPC
4 Models for the Process
and for the Laboratory
Ordering Codes
Accessories
Standard packing includes: calibrator, rubber holster,
charger, instruction manual and report of calibration.
EXTERNAL PRESSURE MODULES - AISI 316SS - ±0.025% F.S.
GAUGE
EE812009 from -100 to 100 mbar (1.5 PSI)
res. 0.001mbar
EE812010 from -500 to 500 mbar (7 PSI)
res. 0.01mbar
EE812011 from -0.95 to 1 bar (15 PSI)
res. 0.01mbar
EE812012 from -0.95 to 2 bar (30 PSI)
res. 0.01mbar
EE812013 from -0.95 to 7bar (100 PSI)
res. 0.1mbar
EE812014 from -0.95 to 20 bar (300 PSI)
res. 0.1mbar
EE812015 from -0.95 to 35 bar (500 PSI)
res. 1mbar
EE812016 from 0 to 70 bar (1000 PSI)
res. 1mbar
EE812017 from 0 to 150 bar (2000 PSI)
res. 1mbar
EE812018 from 0 to 350 bar (5000 PSI)
res. 10mbar
EE812019 from 0 to 700 bar (10000 PSI)
res. 10mbar
3925 basic - A - 00 - C - D
MicroCal 20 DPC basic: ±0.02% rdg
2 channel (IN - OUT) multifunction calibrator
3925 plus - A - BB - C - D
MicroCal 20 DPC plus: ±0.01% rdg
2 channel (IN - OUT) multifunction calibrator
3925 XP - A - BB - C - D
MicroCal 20 DPC XP: ±0.006% rdg
2 channel (IN - IN/OUT) multifunction calibrator
Intrinsically Safe Models
3926 basic - A - 00 - C - D
MicroCal 20 DPC IS basic: ±0.02% rdg
2 channel (IN - OUT) multifunction calibrator
3926 plus - A - BB - C - D
MicroCal 20 DPC IS plus: ±0.01% rdg
2 channel (IN - OUT) multifunction calibrator
3926 XP - A - BB - C - D
MicroCal 20 DPC IS XP: ±0.006% rdg
2 channel (IN - IN/OUT) multifunction calibrator
Table A
Basic
plus
XP
1
1
2
2
3
3
4
4
5
5
Table B
Basic
0
--------
Line charger
120V 50/60 Hz with USA plug
230V 50/60 Hz with Schuko plug
230V 50/60 Hz with UK plug
230V 50/60 Hz with European plug
100V 50/60 Hz with USA/Japan plug
Internal pressure - AISI316SS - ±0.025% FS
plus
XP
0
None
2
-100 to 100 mbar Gauge res. 0.001mbar
3
-500 to 500 mbar Gauge res. 0.01mbar
5
-0.95 to 2 bar Gauge
res. 0.01mbar
5A
2 bar Absolute
res. 0.01mbar
6
-0.95 to 7 bar Gauge
res. 0.1mbar
7
-0.95 to 20 bar Gauge
res. 0.1mbar
7A
20 bar Absolute - res. 0.1mbar
IMPORTANT:
Basic model cannot install internal pressure sensors.
plus and XP models can install up to 2 internal pressure sensors
Table C
Basic
0
1
2
3
3
Table D
Basic
1
from 0 to 2 bar (30 PSI)
from 0 to 20 bar (300 PSI)
SOFTWARE
BB530203
BB530212
BB260198
BB260199
BB260215
BB530204
RS232 cable
USB cable
LogMan-Data Logging software
LinMan-Linearization software
CalpMan 2007 - Calibration Procedure Manager
MicroCal T series communication module
EE300040
BB300122
EE300204
EE300205
BB880056
EE300284
EE480054
EE480055
Test leads kit for electric signal
TC, J, K, T, S compensated cables Kit
Mini DIN TC, J, K, T, S female connector set
Mini DIN TC, J, K, T, S male connector set
Rubber holster
Pedestral for rubber holster
1/4" BSP pressure adapter kit (1/2" M, 1/2" F, 3/8" F, 1/8" F)
1/4" NPT pressure adapter kit (1/8" F, 1/4" F, 3/8" F, 1/2" F)
res. 0.01mbar
res. 0.1mbar
BB880033
Aluminium carrying case
BB880048
Vinyl case with shoulder strap
BB880043
Compact vinyl carrying case with
shoulder strap.
F3280013 Manual pump for vacuum
and pressure up to 2 bar
EE300280 Adapter Kit for internal
pressure module
F3280019 Manual pump for vacuum
and pressure up to 40 bar
Options
plus
XP
0
1
2
ABSOLUTE
EE812020
EE812021
none
HART protocol
EC module (T + RH% + barometric
measurements) - not available on IS model
Extended memory - not available on IS
models
EE300281 Adapter Kit for internal
pressure module
EE170049 Adapter Kit for external
pressure module
F3280015 Manual pump for vacuum
and pressure up to 700 bar
Report of calibration
plus
XP
1
EE372008 High Pressure tube
for pump F3280015
Eurotron Certificate
EE170049 Adapter Kit for external
pressure module
20
All descriptions are related to full options instrument. See latest page for the different configurations.
Specifications may change without notice.
MULTIFUNCTION
CALIBRATORS
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement