OPERATION MANUAL

ISM-M1000

Metallurgical Microscope

Congratulations and Thank You for your purchase of

The Microscope

This precision instrument has been designed to require a minimum of optical and mechanical maintenance. Its excellent design assures years of high quality, reliable service. We recommend you read this entire manual carefully before beginning to use the instrument.

WARNING

PLEASE DISCONNECT PLUG FROM MAINS SOCKET BEFORE REMOVING

THE BOTTOM PLATE, OPENING THE LAMP REPLACEMENT DOOR, OR

REMOVING THE LAMP HOUSE

WARNING

THE POWER CORD PROVIDED WITH THE EQUIPMENT HAS A GROUNDED

PLUG. ALWAYS USE THE POWER CORD WITH A PROPERLY GROUNDED

WALL OUTLET.

DO NOT EXPOSE THE INSTRUMENT TO HIGH TEMPERATURES OR

HUMIDITY. AVOID USING THE INSTRUMENT IN EXTREMELY DUSTY

LOCATIONS.

OPERATING TEMPERATURE 5°C TO 35°C

OPERATING HUMIDITY: 20% TO 80% @ 25°C

CAUTION: NEVER IMMERSE THE INSTRUMENT IN WATER OR SOLVENT

CAUTION: DO NOT PUT ANY FOREIGN OBJECTS IN THE FRAME OR INTO

ANY MOVING MECHANICAL PARTS

1

Contents

Getting Started

Components

Technical Specifications

Setup Instructions

Basic Operation

Changing the Lamp

Changing the Fuse

Preventative Maintenance

3

4

5

5

6

10

11

12

2

Getting Started

This guide is written on the assumption that it will be used by advanced students and experienced microscopists. It covers the essential adjustments and controls of microscopes. No attempt has been made to include fundamentals of microscopy. Read this guide prior to unpacking the microscope. This guide is designed to illustrate a sequential method for assembly and operation. We recommend the instructions be followed in a sequential process. After assembling the instrument, re-read the guide and become familiar with all adjustments and functions before using your new microscope.

3

Components

21 20

19

18

17

16

15

1

2

3

4

5

6

22 23

7 8 9 10 11 12 13 14

24

25

26

Figure 1

27 28

1.Eyepiece 2.Switch Handspike For Observe/Photograph 3.

Thumbscrew 4.Trinocular 5. Up Stop 6. Table-board Longitudinal

Adjustment Knob 7. Brightness Control Knob 8. Table-board

Transverse Adjustment Knob 9. Power Switch 10. Field Diaphragm

Centering Screw 11. Focusing Lens Adjustment Lever 12. Light Bulb

Vertical Adjustment knob 13. Light Bulb Transverse Adjustment knob

14. Lamp House 15. Filter Wheel 16. Aperture Diaphragm Adjustment

Lever 17. Field Diaphragm Adjustment Lever 18.Objective 19.Stage

20.Stretchy Clip 21.Binocular 22.Polarizer

(Optional) 23 Lamp

House Set Screw.24.Focus Tensional Adjustment Knob 25.Coarse

Focus Control Knob 26. Fine Focus Control Knob 27. Power Outlet with Fuse Seat 28. Power Inlet

4

Technical Specifications

Trinocular: Inclined 30°.

Objectives:

Type(no cover glass)

Plan achromatic

Magnification

10X

20X

40X

100X(Spring,Oil)

N.A.

0.25

0.40

0.60

1.25

W.D.(mm)

8.8

8.6

3.73

0.33

Nosepiece: Quadruple nosepiece ball-bearing reversed nosepiece with positive click stops and smooth operation.

Eyepieces: 10X wide field eyepiece, focal length 25mm, field Ø 18mm

.

Stage: Double layer mechanical,

Size: 180mmX150mm,

Moving range: 15mmX15mm.

Filters: Ground glass, blue filter, green filter and yellow filter.

Illumination: 6V/20W halogen lamp, adjustable brightness.

Power supply: 220V(50/60HZ) or 110V(50/60HZ).

Anti-fungus: Yes.

Setup Instructions

1. Remove all parts from their packing materials and retain the

packaging in the event you need to transport the product.

2. Connect the power cord to a suitable power supply.

Basic Operation

1. Illumination controls

1) The power switch to the illuminator and the brightness control is located on the base.

The electrical system is fuse protected and the fuse holder located on the socket.

2) Turn on the light with the power switch, see figure 2. If the light does not appear to be ON, check the brightness control to see if it's on a sufficiently low setting. Then adjust the brightness control until image can be observed comfortably.

Note: Using the light at brightest setting reduces life span of lamp.

2. Focusing Controls

1) Focusing adjustment is accomplished by using the l a r g e c o a r s e a d j u s t m e n t knobs located comfortably on each side of the frame. Fine adjustment is accomplished using the smaller knobs located on the same focus shaft. This coaxial arrangement allows for easy, precise a d j u s t m e n t w i t h o u t d r i f t or discomfort.

6

1

Figure 2

2

1. Brightness Control Knob

2. Power Switch

1 2 3

Figure 3

1. Focus Tension Adjustment

K nob

2. Up Stop

3. Coarse Focus Control Knob

4. Fine Focus Control Knob

4

5

will raise or lower the stage. The smallest graduation on the fine adjust knob index scale is 2μm of vertical.

adjustable and preset at the factory for ease of use. If you wish to adjust the coarse focus tension, first locate the tension adjustment ring, It is located between the frame and coarse adjustment knob.

Turning the ring toward the rear of the microscope increases the tension, and toward the front of the microscope loosens it. Tension is too high if you experience physical discomfort.

4)

Pre-focusing or Focus Stop Control Use of this feature will insure that the shorter working distance objectives don't contact the stage or specimen when using the microscope. Its use also simplifies focusing. After focusing on the specimen with the coarse adjustment by low magnification eyepiece, rotation of the lever toward the handler will set and upper limit on the coarse adjustment movement. After changing specimens or objectives, focusing is easily accomplished by rotating the coarse adjustment knob to reach the pre-focused position, then making fine adjustments with the fine adjustment knob. Focusing movement with the fine adjustment isn't affected by using the pre-focusing lever.

3. Diopter and Interpupillary Adjustments

Diopter Adjustments Ring

1) D i o p t e r A d j u s t m e n t P r o p e r correction for individual vision is accomplished via the diopter adjustment located at the left eyepiece, see figure 4. Using the

40X magnification objective, bring

An image into focus with your right eye only. Once the image is well focused, observe with left eye, make fine adjustments with the diopter Adjustment ring to correct for your Vision.

Figure 4

7

2) Proper interpupillary distance, Or the distance between eyepieces, is crucial to the comfort of the user. Adjusting the interpupillary distance is accomplished through a “folding” action of the optical head, at figure 4, allowing for quick and easy adjustment.

4. Mechanical Stage Controls, see figure 5, the coaxial transverse/ longitudinal knob allow for easy adjustment.

1. Longitudinal Adjustment

Konb

2. Transverse Adjustment

Knob

3. Stage

1 2

Figure 5

3

5. Illumination Alignment, See Figure 6.

1 2 3 4 5 6 7

Figure 6

1. Field Diaphragm Centering Screw 2. Field Diaphragm Adjustment Lever

3. Aperture Diaphragm Adjustment Lever 4. Filter wheel(Polarizer(optional))

5. Focusing Lens Adjustment Lever 6. Light Bulb Transverse Adjustment Knob

7 Light Bulb Vertical Adjustment Knob

8

1) The aperture diaphragm near the illuminator house in the vertical illuminator tube may be adjusted slightly to change contrast.

2) To center the incident or vertical lamp.

a) Put a piece of paper and place it on the stage.

Take out one objective form nosepiece and rotate the nosepiece until this objective opening hole is in position over the paper.

Filament Image b) Turn the power on and adjust the brightness c o n t r o l t o e s t a b l i s h sufficient light.

Overpass Hole

Figure 7 c) This paper will allow for focusing of the lamp filament on it. If the filament image is not centered in the overpass hole, reduce the aperture diaphragm, adjust the light bulb transverse and vertical adjustment to obtain a uniform image. Adjust the focusing lens lever until a sharp image of the lamp filament is attained, refer to figure 7 for this step.

3) Reinstall the objective again.

4) Adjust the focusing control until the image clear. If brightness of field not equality, you may move the lamp adjustment and focusing lens adjustment lever slightly to make the brightness of field equality.

5) Reduce the field diaphragm, If it is not centered, move it to the center using the centering screw. Adjustment can be done by opening the field diaphragm.

6) The green, yellow, blue filter and ground glass can be inserted in the socket to obtain an image of good quality; also the polarizer

(optional) can be inserted in the socket to make the polarizing analyzer on work.

9

6.

To Observe the Facility Polarizer

(Optional)

Turn the Analyzer and Polarizer to the light way (Figure 8), so that the analyzer used for transmitted light work., rotate the polarizer (Figure 6)while watching the brightness in the field and let the polarizer and analyzer orthogonally.

7.

The white push rod to the upper left of the trinocular head will send 100% of the image light to the binocular eyepieces or to the top port. It satisfies high-grape photography.

1 2

Figure 8

1. Analyzer (Optional)

2. Polarizer (Optional)

Changing the Lamp

1.

Turn off and unplug the

Microscope.

2.

Loosen the set screw, see figure

9, remove the entire cover board from the lamp house.

3.

The halogen lamp can be removed by grasping the lamp and pulling it firmly from its fixture. Take care not to touch the lamp with bare fingers, as the lamp will be hot.

4.

Insert the new halogen lamp into the same fixture. When installing the new lamp, be careful not to touch the glass with your fingers.

1 2 3 4

5

Figure 9

1. Set Screw

2. Cover Board

3. Lamp

4. Lamp House

5. Lamp Socket

10

The new lamp should be supplied in a plastic protective envelope. If not, use a tissue or other medium to grasp the lamp. This will prevent contamination from your hand from reducing the lamp's intensity and life.

5.

Reinstall the cover board to the lamp house and retighten the screw.

6.

If necessary, you may need to readjust the lamp as previously described.

Changing the Fuse

Draw the Power inlet and remove the fuse, reinstall the new one and insert the fuse seat and power line (Figure 10).

1

Preventative Maintenance

1.

Cleaning frame and stage

Disconnect the plug from mains socket before cleaning. Clean the frame and stage with a soft cloth moistened with a mild detergent solution. Be sure the instrument is dry before using.

2.

Cleaning optical parts

Microscope eyepieces and objectives are coated.

They should not be wiped while dry as dirt or dust may scratch the coating. It is best to remove parts from the frame prior to cleaning.

Always blow loose dust away first. Use cotton swabs or lens tissue moistened with a lens cleaner or a small amount of alcohol, then wipe the surface clean with a good quality lens tissue. Solvents such as Xylene should NOT be used as cleaner.

3.

Cleaning 100X oil immersion lens

The immersion oil should be removed from the lens at the end of each workday using cotton swabs or lens tissue moistened with a lens cleaner or a small amount of alcohol.

DO NOT DISASSEMBLE OBJECTIVE LENSES.

Figure 10

1. Fuse seat

2. Power inlet

2

11 12