advertisement
Manufacturer
EU-representative
Asia Pacific HQ
Production facility
Check following components!
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS www.omron-healthcare.com
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967 www.omronhealthcare-ap.com
OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD.
Binh Duong Province, VIETNAM
Made in Vietnam
Automatic Blood Pressure Monitor
Model HEM-7320
Instruction Manual
English
5342277-7A
Automatic Blood Pressure Monitor
Model HEM-7320
Instruction Manual
English
HEM-7320-AU_A_M04_130726.pdf
Important Safety Information
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis.
Please read this section carefully before using the unit.
Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
(General Usage)
• Do not use the device on the injured arm or the arm under medical treatment.
• Do not wrap the arm cuff on the arm while being on an intravenous drip or blood transfusion.
(AC Adapter (optional) Usage)
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off the power and unplug the power cord immediately.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a multiple-tap.
• Never plug in or unplug the power cord from the electric outlet with wet hands.
EN
Caution: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a doctor before using the unit, as cuff inflation can cause internal bleeding.
• If there are any abnormalities during the measurement, remove the arm cuff.
• Do not use this device on infants or persons who cannot express their intentions.
• Do not inflate the arm cuff more than necessary.
• Do not use the unit for any purpose other than measuring blood pressure.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect measurement results.
• Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
• Do not disassemble the unit or arm cuff.
1
Important Safety Information
2
(AC Adapter (optional) Usage)
• Fully insert the power plug.
• When disconnecting the power plug, do not pull the power cord. Be sure to hold the power plug.
• When handling the power cord, observe the following:
Do not damage. Do not break it.
Do not tamper with it.
Do not twist.
Do not pinch.
Do not forcibly bend or pull.
Do not bundle during use.
Do not place under heavy objects.
• Wipe the dust off from the power plug.
• Disconnect the power plug if the product will not be used for a long period of time.
• Disconnect the power plug before starting maintenance.
• Use only the original AC adapter designed for this unit. Use of unsupported adapters may damage and/or may be hazardous to the unit.
(Battery Usage)
• Do not insert the batteries with their polarities incorrectly aligned.
• Use only four “AA” alkaline or manganese batteries with this unit. Do not use other types of batteries. Do not use new and used batteries together.
• Remove the batteries if the unit will not be used for 3 months or more.
General Precautions
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• To unplug the air plug, pull on the air plug at the connection with the main unit, not the tube itself.
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not inflate the arm cuff when it is not wrapped around your arm.
Section.
used accessories or optional parts.
1. Overview
Open the rear cover page to read following:
The alphabet in the rear cover page correspond to those in the body page.
Main unit
A
B
C
D
E
F
G
H
Display
Cuff wrapping guide lamp
Memory button
START/STOP button
Up/Down ( ) buttons
Date/Time setting ( ) button
Blood pressure colour indicator
Air jack
I
J
Battery compartment
AC adapter jack (for optional AC adapter)
Arm cuff
K
L
M
Arm cuff (Arm circumference
22-42cm)
Air plug
Air tube
Display
N
O
P
Q
R
S
T
U
V
W
X
Y
Z
Memory symbol
Average value symbol
Systolic blood pressure
Diastolic blood pressure
Pulse display and Memory number
Heartbeat symbol
(Flashes during measurement)
Date/Time display
Movement error symbol
Irregular heartbeat symbol
Blood pressure level indicator
Cuff wrapping guide
Low battery symbol
Deflation symbol
EN
3
4
2. Preparation
2.1 Installing/Replacing the
Batteries
1.
Remove the battery cover.
2.
Insert 4 “AA” batteries as indicated in the battery compartment and then replace the battery cover.
Notes:
• If the low battery symbol ( ) appears on the display, turn the monitor off then replace all batteries at the same time.
• The measurement values continue to be stored in memory even after the batteries are replaced.
• The supplied batteries may have a shorter life.
Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries.
2. Preparation
2.2 Setting the Date and Time
1.
Press the button.
2.
Set the monitor to the correct date and time before taking a measurement for the first time.
1)
Push or buttons to change.
Year
2)
Push the button to confirm.
Next setting appears.
Month
Day
Hour
3.
Press the START/STOP button to store the setting.
Notes:
• If the batteries have been replaced, the date and time setting will need to be reset.
• If the date and time are not set, “-/ - -:--” appears during or after measurement.
Minute
EN
5
6
3. Using the Unit
Open both the front and rear covers to read following:
The alphabet and number in the cover page correspond to those in the body page.
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your upper arm.
Do not place the cuff over thick clothes.
1.
Insert
L
the air plug into
H
the air jack securely.
2.
Apply the arm cuff to your upper arm.
The bottom edge of the arm cuff should be
a
1 to 2 cm above the elbow.
M
Air tube is centred on the middle of your inner arm.
3.
Close the fabric fastener FIRMLY.
Notes:
• When you take a measurement on the right arm, air tube will be at the side of your elbow. Be careful not to rest your arm on the air tube. ---
b
• The blood pressure can differ between the right arm and the left arm, and therefore also the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your physician which arm to use for your measurement.
3. Using the Unit
3.2 How to Sit Correctly
To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. No bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight. ---
c
• The arm cuff should be at the same level as your heart. ---
d
EN
7
8
3. Using the Unit
3.3 Taking a Reading
Notes:
• To cancel a measurement, press the START/STOP button to turn the unit off and to release the air in the arm cuff.
• Remain still while taking a measurement.
1.
Press the START/STOP button.
The arm cuff will start to inflate automatically.
START INFLATE DEFLATE COMPLETE
If your systolic pressure is more than 210 mmHg
After the cuff starts to inflate, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
Cuff wrapping guide lamp
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
3. Using the Unit
Cuff Wrapping Guide Lamp
The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around the arm. Even when lights in orange, a blood pressure reading will be taken.
Note:
This reading is NOT reliable due to the incorrect wrapping of the cuff. Please wrap the cuff again, taking care to wrap it correctly and take the measurement again.
When lights in green, the arm cuff is correctly wrapped tightly enough on the arm and the reading is accurate and reliable.
2.
Remove the arm cuff.
Note:
Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.
Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
EN
3.
Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after 2 minutes.
9
10
3. Using the Unit
Important:
• If your systolic or diastolic pressure is outside the standard range, the blood pressure colour indicator will light in orange when the measurement result is displayed. If they are within the standard range, the blood pressure colour indicator will light in green.
Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
This criteria is for home blood pressure measurement.
For professional office blood pressure measurement criteria, please refer to Chapter 7.
• Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats can influence the results of the measurement. The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated. If the measurement results are affected by irregular heartbeats but the result is valid, the result is shown together with the irregular heartbeat symbol ( ). If the irregular heartbeats cause the measurement to be invalid, no result is shown. If the irregular heartbeat symbol ( is shown after you have taken a measurement, repeat the measurement. If the irregular
) heartbeat symbol ( ) is shown frequently, please make your doctor aware of it.
• If you move during measurement, the movement error symbol ( ) will appear on the display. Keep still and repeat the measurement.
3. Using the Unit
Using the Guest Mode
The monitor stores measurement values for one user in the memory. The guest mode can be used to take a single measurement for another user. No measurement values are stored in the memory when the guest mode is selected.
1.
Press and hold the START/STOP button for more than
3 seconds.
The date/time display will disappear.
EN
2.
Release the START/STOP button when the date/time display turns off.
The arm cuff will start to inflate automatically.
11
3. Using the Unit
3.4 Using the Memory Function
The monitor automatically stores the result up to 90 sets. It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes. If there are only two readings in memory for that period, the average will be based on two readings. If there is one reading in memory for that period, the average will be based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time, “-/ - -:--” is displayed instead of the date and time.
12
To View the Readings Stored in Memory
1.
Press the button.
The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered “1”.
Note:
The blood pressure colour indicator lights and the cuff wrapping guide result appears on the display with the measurement values.
2.
Press the or button to view the readings stored in memory.
: To the older readings
: To the more recent readings
3. Using the Unit
To View the Average Value
1.
Press the button for more than
3 seconds.
Notes:
• If the previous reading was taken without setting the date and time, the average value is not calculated.
• If there are no measurements results stored in the memory, the screen to the right is displayed.
EN
To Delete All the Values Stored in Memory
When the memory symbol ( ) appears, first press the Memory button. Then while holding it down, press the START/STOP button simultaneously for more than 3 seconds.
Note:
You cannot partially delete values stored in the memory.
13
14
4. Troubleshooting and Maintenance
4.1 The Icons and Error Messages
Error Display Cause
Irregular heartbeats are detected.
Remedy
Remove the arm cuff. Wait 2-3 minutes and then take another measurement.
Repeat the steps in section 3.3. If this error
continues to appear, contact your doctor.
Movement during measurement.
Carefully read and repeat the steps in section
Arm cuff is not applied correctly.
The batteries are low.
The batteries are exhausted.
Apply the arm cuff correctly.
You should replace them with new ones ahead of time.
You should replace them with new ones at once.
4. Troubleshooting and Maintenance
Error Display Cause
Air plug disconnected.
Arm cuff not applied correctly.
Air is leaking from the arm cuff.
Movement during measurement and the arm cuff has not been inflated sufficiently.
The arm cuff was inflated above 299 mmHg when inflating the cuff manually.
Movement during measurement.
Clothing is interfering with the arm cuff.
Remedy
Insert the plug securely.
Apply the arm cuff correctly.
Replace the cuff with the new one.
Repeat measurement. Remain still and do not talk during measurement.
If “E2” appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result.
Do not inflate the cuff above 299 mmHg.
Repeat measurement. Remain still and do not talk during measurement.
Remove any clothing interfering with the arm cuff.
EN
Device error.
Contact your local OMRON representative.
15
4. Troubleshooting and Maintenance
16
4.2 Troubleshooting
Problem
The reading is extremely low (or high).
Arm cuff pressure does not rise.
Arm cuff deflates too soon.
Cannot measure or readings are too low or too high.
The arm cuff has not been inflated sufficiently.
Cause
Arm cuff not applied correctly.
Movement or talking during measurement.
Clothing is interfering with the arm cuff.
The air tube is not securely connected into the main unit.
Air is leaking from the arm cuff.
The arm cuff is loose.
Remedy
Apply the arm cuff correctly.
Remain still and do not talk during measurement.
Remove any clothing interfering with the arm cuff.
Make sure that the air tube is connected securely.
Replace the arm cuff with a new one.
Apply the cuff correctly so that it is firmly wrapped around the arm.
Inflate the cuff so that it is 30 to
40 mmHg above your previous measurement result.
4. Troubleshooting and Maintenance
Problem
Nothing happens when you press the buttons.
Other problems.
Cause Remedy
The batteries are empty.
Replace the batteries with new ones.
The batteries have been inserted incorrectly.
Insert the batteries with the correct
(+/ -) polarity.
• Press the START/STOP button and repeat measurement.
• If the problem continues, try replacing the batteries with new ones.
If this still does not solve the problem, contact your local OMRON representative.
EN
17
4. Troubleshooting and Maintenance
4.3 Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and the cuff to extreme temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations (for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean the arm cuff.
• Do not carry out repairs of any kind yourself. If a defect occurs, consult your local OMRON representative.
18
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and neutral soap to clean the arm cuff.
Note:
Read and follow the "Correct Disposal of This Product" in "6.Technical Data" when disposing of the
device and any used accessories or optional parts.
Calibration and Service
• The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected every 2 years to ensure correct functioning and accuracy. Please consult your local OMRON representative.
4. Troubleshooting and Maintenance
4.4 Storage
Keep the unit in its storage case when not in use.
1.
Unplug the air plug from the air jack.
2.
Gently fold the air tube into the arm cuff.
Note:
Do not bend the air tube excessively.
3.
Place the arm cuff and main unit in the storage case.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an angle.
EN
19
5. Optional Parts
Arm cuff
Arm circumference
22 - 42 cm
AC adapter S
20
Note:
Please check with your local OMRON representatives for the appropriate optional part models.
Using the Optional AC Adapter
1.
Insert the AC adapter plug into the AC adapter jack on the rear side of the main unit.
2.
Plug the AC adapter into an electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the main unit.
6. Technical Data
Product Description
Model
Display
Measurement Method
Measurement Range
Accuracy
Inflation
Deflation
Memory
Rating
Power Source
Battery Life
Automatic Blood Pressure Monitor
OMRON HEM-7320
LCD Digital Display
Oscillometric method
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/ min.
Pressure:
3 mmHg
Pulse:
5% of display reading
Fuzzy-logic controlled by electric pump
Automatic pressure release valve
90 Measurements with date and time
DC6V 4W
4 “AA” batteries 1.5V or AC adapter (optional, INPUT AC100-240V 50/60Hz 0.12A)
Approximately 1000 measurements (Using new alkaline batteries)
Applied Part
Protection Against Electric Shock
= Type BF
Internally powered ME equipment (When using only the batteries)
= Class II ME equipment (Optional AC adapter)
Operating temperature/ Humidity
Storage temperature/ Humidity/ Air pressure
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Cuff/ Tube Material
Package Contents
+10°C to +40°C / 30 to 85% RH
-20°C to +60°C / 10 to 95% RH / 700-1060hPa
Approximately 380g without batteries
Approximately 210g
Approximately 124 (w) mm
90 (h) mm 161 (l) mm
Approximately 152 mm
600 mm (Cuff: arm circumference 22 to 42 cm)
Nylon, polyester, polyvinyl chloride
Main unit, arm cuff, battery set, storage case, instruction manual
Note:
Subject to technical modification without prior notice.
EN
21
22
6. Technical Data
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive).
• This blood pressure monitor is designed according to the European Standard EN1060, Noninvasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.
• This OMRON product is produced under the strict quality system of OMRON HEALTHCARE
Co., Ltd., Japan. The core component for OMRON blood pressure monitors, which is the
Pressure Sensor, is produced in Japan.
6. Technical Data
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
EN
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.
This product does not contain any hazardous substances.
23
24
7. Some Useful Information about Blood Pressure
What is Blood Pressure?
Blood pressure is a measure of the force of blood flowing against the walls of the arteries.
Arterial blood pressure is constantly changing during the course of the heart’s cycle.
The highest pressure in the cycle is called the Systolic Blood Pressure; the lowest is the
Diastolic Blood Pressure.
Both pressure readings, the Systolic and Diastolic, are necessary to enable a doctor to evaluate the status of a patient’s blood pressure.
What is Irregular Heartbeat?
An irregular heartbeat is a heartbeat rhythm that varies by more than 25% from the average heartbeat rhythm detected while the unit is measuring the systolic and diastolic blood pressure.
Pulse
Blood pressure
Normal Heartbeat
Irregular Heartbeat
Short Long
If such an irregular rhythm is detected more than twice during measurement, the irregular heartbeat symbol ( appears on the display when the measurement results are
) displayed.
Pulse
Blood pressure
What is Arrhythmia?
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in the bioelectrical system that drives the heartbeat. Typical symptoms are skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) or slow (bradycardia) pulse.
7. Some Useful Information about Blood Pressure
Why is it a Good Thing to measure Blood Pressure at Home?
Many factors such as physical activity, anxiety, or the time of day, can influence your blood pressure. A single measurement may not be sufficient for an accurate diagnosis. Thus it is best to try and measure your blood pressure at the same time each day, to get an accurate indication of any changes in blood pressure. Blood pressure is typically low in the morning and increases from afternoon to evening. It is lower in the summer and higher in the winter.
200
150
100
50
6 12
Upper curve: systolic blood pressure
Lower curve: diastolic blood pressure
Time of day
18
Example: fluctuation within a day (male, 35 years old)
24
EN
25
26
7. Some Useful Information about Blood Pressure
Classification of Blood Pressure by the World Health Organization
The World Health Organization (WHO) and the
International Society of Hypertension (ISH) developed the Blood Pressure Classification shown in this figure.
This classification is based on the blood pressure values measured on people in a sitting position in outpatient departments of hospitals.
Note:
There is no universally accepted definition of hypotension.
However, those having the systolic pressure below
100 mmHg are assumed as hypotensive.
Systolic blood pressure
(mmHg)
Severe hypertension
180
Moderate hypertension
160
Mild hypertension
140
Normal systolic value
130
Normal blood pressure
120
Optimal blood pressure
(target value)
80 85 90 100 110
Diastolic blood pressure
(mmHg)
c d
1
H
L
2 a
Contents
Thank you for purchasing the OMRON HEM-7320 Automatic Blood Pressure Monitor.
The OMRON HEM-7320 is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly.
For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the device uses its advanced “IntelliSense” technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of people within the range of the designated arm cuff, following the instructions in this instruction manual. It is mainly designed for general household use. Please read the Important Safety Information in this instruction manual before using the unit.
Important Safety Information........................1
1. Overview ...................................................3
2. Preparation ...............................................4
2.1 Installing/Replacing the Batteries...........4
2.2 Setting the Date and Time .....................5
3. Using the Unit ...........................................6
3.1 Applying the Arm Cuff ............................6
3.2 How to Sit Correctly ...............................7
3.3 Taking a Reading...................................8
3.4 Using the Memory Function .................12
4. Troubleshooting and Maintenance.......14
4.1 The Icons and Error Messages ........... 14
4.2 Troubleshooting................................... 16
4.3 Maintenance........................................ 18
4.4 Storage................................................ 19
5. Optional Parts.........................................20
6. Technical Data ........................................21
7. Some Useful Information about Blood
Pressure ..................................................24
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
H
G
F
K
A
M
L
N
O
P
B
C
D
E
Q
R
S
GI
HJ
T
U
V
W
HX
HY
Z
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project