LumaSense Technologies IGA 6/23 Advanced pyrometer MANUAL
Below you will find brief information for pyrometer IGA 6/23 Advanced. The IMPAC IGA 6/23 is a short wave infrared temperature measuring device. It is used for non-contact temperature measurements on metals, ceramics, graphite, etc. with a temperature range between 50 and 1800 °C. It features digital signal processing and can be used with various accessories for safe mounting and alignment to the measuring object. The device has a viewfinder or laser targeting light option, and you can adjust the emissivity, exposure time, and other settings through the InfraWin software.
Advertisement
Advertisement
IMPAC Pyrometer
IGA 6/23 Advanced
Confidential Information
The material contained herein consists of information that is the property of LumaSense
Technologies and intended solely for use by the purchaser of the equipment described in this manual. All specifications are subject to change without notice. Changes are made periodically to the information in this publication, and these changes will be incorporated in new editions.
LumaSense Technologies prohibits the duplication of any portion of this manual or the use thereof for any purpose other than the operation or maintenance of the equipment described in this manual, without the express written permission of LumaSense Technologies.
Copyright
© LumaSense Technologies 2013. All rights reserved.
Trademarks
IMPAC is a trademark of LumaSense Technologies.
All other trademarks are trademarks, registered trademarks, and/or service marks of their respective holders.
LumaSense Technologies, Inc.
North America
Sales & Service
Santa Clara, CA, USA
Ph: +1 800 631 0176
Ph: +1 408 727 1600
Fax: +1 408 727 1677
Service Centers
LumaSense Technologies GmbH
Other Than North America
Sales & Support
Frankfurt, Germany
Ph: +49 (0) 69 97373 0
Fax: +49 (0) 69 97373 167
Our Headquarters
LumaSense Technologies, Inc.
Santa Clara, CA, USA
Ph: +1 800 631 0176
Fax: +1 408 727 1677
Brazil
LumaSense, Vendas Brasil
Campinas, Brasil
Ph: +55 19 3367 6533
Fax: +55 19 3367 6533
Global and Regional Centers
Americas, Australia, & Other Asia
LumaSense Technologies, Inc.
Santa Clara, CA, USA
Ph: +1 800 631 0176
Fax: +1 408 727 1677
Europe, Middle East, Africa
LumaSense Technologies GmbH
Frankfurt, Germany
Ph: +49 (0) 69 97373 0
Fax: +49 (0) 69 97373 167
India
LumaSense Technologies, India
Mumbai, India
Ph: + 91 22 67419203
Fax: + 91 22 67419201
China
LumaSense Technologies, China
Shanghai, China
Ph: +86 133 1182 7766
Fax: +86 21 5877 2383
E-mail [email protected]
Website http://www.lumasenseinc.com
Part No 3 904 226 - EN
Revision B
November 2013
Contents
1.1.1
1.1.2
2.6.1
2.6.2
2.6.3
2.6.4
2.6.5
2.6.6
3.1.1
3.1.2
3.1.3
3.1.4
3.2.1
3.2.2
3.3.1
3.3.2
Deviation from the focused measuring distance .............................................. 20
3.3.3
IGA 6/23 Advanced Manual Contents · iii
4.5 Response Time (t
90
) .............................................................................................. 24
4.6 Clear Peak Memory (t
CLEAR
) ................................................................................... 25
4.6.1
4.6.2
5.4.1
5.4.2
5.4.3
5.4.4
5.4.5
5.9 PC sampling rate (time interval between two measurements) ........................ 33
6.2.1
6.2.2
IGA 6/23 Advanced Manual Contents · iv
1 General Information
1.1 Information about the user manual
Congratulations on choosing the high quality and highly efficient IMPAC IGA 6/23 Advanced pyrometer.
This manual provides important information about the instrument and can be used as a work of reference for installing, operating, and maintaining your IGA 6/23 Advanced pyrometer. It is important that you carefully read the information contained in this manual and follow all safety procedures before you install or operate the instrument.
To avoid handling errors, keep this manual in a location where it will be readily accessible.
1.1.1 Legend
Note: The note symbol indicates tips and useful information in this manual. All notes should be read to effectively operate the instrument.
Attention: This sign indicates special information which is necessary for a correct temperature measurement.
Warnings and Cautions: The general warnings and cautions symbol signifies the potential for bodily harm or damage to equipment.
MB Shortcut for Temperature range (in German: Messbereich)
1.1.2 Terminology
The terminology used in this manual corresponds to the VDI- / VDE-directives 3511, Part 4.
1.2 Safety
This manual provides important information on safely installing and operating the
IGA 6/23 Advanced pyrometer. Several sections of this manual provide safety warnings to avert danger. These safety warnings are specified with a warning symbol. You must read and understand the contents of this manual before operating the instrument even if you have used similar instruments or have already been trained by the manufacturer.
It is also important to continually pay attention to all labels and markings on the instrument and to keep the labels and markings in a permanent readable condition.
Warning: The pyrometer is only to be used as described in this manual. It is recommended that you only use accessories provided by the manufacturer.
In addition, signs and markings on the device is to be observed and maintained in legible condition.
1.3 Limit of liability and warranty
All general information and notes for handling, maintenance, and cleaning of this instrument are offered according to the best of our knowledge and experience.
LumaSense Technologies is not liable for any damages that arise from the use of any examples or processes mentioned in this manual or in case the content of this document should be
IGA 6/23 Advanced Manual Contents · 5
incomplete or incorrect. LumaSense Technologies reserves the right to revise this document and to make changes from time to time in the content hereof without obligation to notify any person or persons of such revisions or changes.
All instruments from LumaSense Technologies have a regionally effective warranty period.
Please check our website at http://info.lumasenseinc.com/warranty for up-to-date warranty information. This warranty covers manufacturing defects and faults which arise during operation, only if they are the result of defects caused by LumaSense Technologies.
The Windows compatible software was thoroughly tested on a wide range of Windows operating systems and in several world languages. Nevertheless, there is always a possibility that a Windows or PC configuration or some other unforeseen condition exists that would cause the software not to run smoothly. The manufacturer assumes no responsibility or liability and will not guarantee the performance of the software. Liability regarding any direct or indirect damage caused by this software is excluded.
The warranty is VOID if the instrument is disassembled, tampered with, altered, or otherwise damaged without prior written consent from LumaSense Technologies; or if considered by
LumaSense Technologies to be abused or used in abnormal conditions.
There are no user-serviceable components in the instrument:
·
No adjustments may be made to the targeting laser. It is fixed at the factory.
·
No adjustments may be made to the targeting laser's power level.
1.4 Unpacking the Instrument
Before shipment, each instrument is assembled, calibrated, and tested at the LumaSense Factory.
When unpacking and inspecting your system components, you need to do the following:
1. Check all materials in the container against the enclosed packing list.
LumaSense Technologies cannot be responsible for shortages against the packing list unless a claim is immediately filed with the carrier. Final claim and negotiations with the carrier must be completed by the customer.
2. Carefully unpack and inspect all components for visible damage. If you note any damage or suspect damage, immediately contact the carrier and LumaSense Technologies, Inc.
3. Save all packing materials, including the carrier’s identification codes, until you have inspected all components and find that there is no obvious or hidden damage.
Note: LumaSense encourages you to register your product with us to receive updates, product information, and special service offers: http://info.lumasenseinc.com/registration .
.
1.5 Service Request, Repair, or Support
Contact LumaSense Technologies Technical Support in case of a malfunction or service request.
Provide clearly stated details of the problem as well as the instrument model number and serial number. Upon receipt of this information, Technical Support will attempt to locate the fault and, if possible, solve the problem over the telephone.
If Technical Support concludes that the instrument must be returned to LumaSense Technologies for repair, they will issue a Return Material Authorization (RMA) number.
Return the instrument upon receipt of the RMA number, transportation prepaid. Clearly indicate the assigned RMA number on the shipping package exterior. Refer to Section 1.6,
Shipments to LumaSense for Repair, for shipping instructions.
IGA 6/23 Advanced Manual Contents · 6
Technical Support can be contacted by telephone or email:
Santa Clara, California
·
Telephone: +1 408 727 1600 or +1 800 631 0176
·
Email: [email protected]
Frankfurt, Germany
·
Telephone: +49 (0) 69 97373 0
·
Email: [email protected]
Erstein, France
·
Telephone +33 (0)3 88 98 98 01
·
Email [email protected]
1.6 Shipments to LumaSense for Repair
All RMA shipments of LumaSense Technologies instruments are to be prepaid and insured by way of United Parcel Service (UPS) or preferred choice. For overseas customers, ship units airfreight, priority one.
The instrument must be shipped in the original packing container or its equivalent. LumaSense
Technologies is not responsible for freight damage to instruments that are improperly packed.
Contact us to obtain an RMA number (if one has not already been assigned by Technical
Support). Clearly indicate the assigned RMA number on the shipping package exterior.
Send RMA Shipments to your nearest technical service center:
Santa Clara, California
LumaSense Technologies, Inc.
3301 Leonard Court
Santa Clara, CA 95054 USA
Telephone: +1 408 727 1600
+1 800 631 0176
Email: [email protected]
Frankfurt, Germany
LumaSense Technologies GmbH
Kleyerstr. 90
60326 Frankfurt
Germany
Telephone: +49 (0)69-97373 0
Email: [email protected]
1.7 Disposal / decommissioning
Inoperable IMPAC pyrometers must be disposed of in compliance with local regulations for electro or electronic material.
IGA 6/23 Advanced Manual Contents · 7
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced Manual Contents · 8
2 Introduction
2.1 Appropriate use
The IMPAC IGA 6/23 is a short wave infrared temperature measuring device with digital signal processing. It is used for non-contact temperature measurements on metals, ceramics, graphite, etc. with a temperature range between 50 and 1800 °C.
2.2 Scope of delivery
Pyrometer, PC adjustment, and evaluation software InfraWin, works certificate, and operating instructions.
Note: A connection cable is not included with the instrument and has to be ordered separately (see Chapter
, Reference numbers).
2.3 Technical Data
Measurement
Temperature Ranges:
Sub Range:
Spectral Ranges:
Resolution:
Emissivity e
:
Transmittance t
:
Response Time t
90
:
(
Measurement Uncertainty: e
= 1, t
Repeatability: e
= 1, t
90
= 1 s, T
Umg
.
= 25 °C)
(
90
= 1 s, T
Noise Equivalent
Umg
.
= 25 °C)
Temperature Difference
(NETD):
( e
= 1, s
= 1, T
Umg
.
= 25 °C, t
90
= min)
50 to 1000 °C (MB 10)
75 to 1300 °C (MB 13)
150 to 1800 °C (MB 18)
Any range adjustable within the temperature range, minimum span
50 °C
2 to 2.6 µm (main wavelength 2.3 µm)
0.1 °C or 0.2 °F at interface; < 0.0015% of adjusted temperature range at analog output, 16 bit; 1 °C or 1 °F on display
0.050 to 1.000 in steps of 1/1000
0.050 to 1.000 in steps of 1/1000
0.5 ms (with dynamic adaptation at low signal levels) adjustable to min; 1 ms; 3 ms; 5 ms; 10 ms; 50 ms; 250 ms; 1 s; 3 s; 10 s
< 1500 °C: 0.3% of reading in °C + 2 °C
> 1500 °C: 0.6% of reading in °C
0.15% of reading in °C + 1 °C
MB 10
MB 13
Temperature / °C
60
100
150
100
200
250
NETD
0.9
0.5
0.1
1.2
0.6
0.1
MB 18
200
300
400
1.5
0.6
0.2
IGA 6/23 Advanced Manual Introduction · 9
Optics
Sighting: Built-in laser aiming light (max. power level < 1 mW, λ = 630 to 680 nm,
CDRH class II) or through-lens sighting
Optics:
Distance Ratio:
Communication
Analog Output:
Digital Interface:
Maximum Value Storage:
Manually focusable from rear cover measuring distance a = 210 to 5000 mm
MB 10: approx. 50 : 1
MB 13: approx. 100 : 1
MB 18: approx. 350 : 1
Environment
Protection Class:
Operating Position:
Ambient Temperature:
Note: During operation the instruments will warm up and might reach an intrinsic temperature of up to 58 °C
Storage Temperature:
Relative Humidity:
Weight:
Housing:
CE-Label:
Interface
Connection:
Display (in rear cover):
Parameters:
IP 65 IEC 60529 (value in mated condition)
Any
0 to 70 °C at housing
-20 to 80 °C
Non condensating conditions
0.6 kg
Stainless steel
According to EU directives about electromagnetic immunity
12-pin connector
LED, 4 digit matrix, 5 mm high temperature signal or measuring distance
Adjustable via interface: emissivity, sub range, ambient temperature compensation, wait time t w
, settings for maximum value storage, address, baud rate, transmittance, response time t
90
, 0 to 20 mA or 4 to 20 mA analog output range, °C/°F
Readable via interface: measured value, internal temperature of the unit, measuring distance
Adjustable 0 to 20 mA or 4 to 20 mA, linear (via digital interface)
RS485 addressable (half-duplex)
Baud rate: 1200 Bd to 115.2 kBd (on request RS232, not addressable)
Built-in single or double storage. Clearing with adjusted time t next measuring object, hold-function clear
(off;
10 ms; 50 ms; 250 ms; 1 s; 5 s; 25 s), via interface, automatically with the
Electrical
Power Supply:
Power Consumption:
Load (analog output):
Isolation:
24 V DC ± 25%, ripple must be less than 50 mV
Max. 3 W (incl. laser)
0 to 500 Ohm
Power supply, analog output, and digital interface are galvanically isolated from each other
IGA 6/23 Advanced Manual Introduction · 10
Note: The calibration / adjustment of the instruments was carried out in accordance with VDI/VDE directive “Temperature measurement in industry, Radiation thermometry, Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4.
For additional details on this directive, see http://info.lumasenseinc.com/calibration or order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany.
2.4 Dimensions
IGA 6/23 Advanced with the view finder
IGA 6/23 Advanced with laser sighting
2.5 Physical User Interface
IGA 6/23 Advanced with laser sighting
IGA 6/23 Advanced with the view finder
1 12-pin connector
2 Digital display
3 Optional sighting
4 Screw to adjust the focus
5 LED distance display
6 LED indicator light
7 Adjusting for the view finder
IGA 6/23 Advanced Manual Introduction · 11
Warning: If selected, the IGA 6/23 Advanced laser light option is a Class 2 and emits laser light.
To minimize the risk of eye injury, do not look directly into the beam, and do not point the laser beam of the pilot light into the eyes of another person.
2.6 Accessories (optional)
Numerous accessories guarantee easy installation of the pyrometer. The following overview shows a selection of suitable accessories. You can find the entire accessory program with all reference numbers in Chapter
, Reference numbers.
2.6.1 Mounting
Mounting brackets are available for safe mounting and alignment of the pyrometer to the measuring object.
Mounting angle
2.6.2 Cooling Jacket
The stainless steel housing with integrated water cooling air purge allows use of the pyrometer above the maximum permissible ambient temperature. The pyrometer can be operated in ambient temperatures up to 180 °C.
2.6.3 Air Purge
The air purge protects the lens from contamination of dust and moisture. It has to be supplied with dry and oil-free pressurized air and generates an air stream shaped like a cone.
Water cooling jacket with integrated air purge
Air purge
2.6.4 Vacuum Pickup
The vacuum pickup KF 16 with window allows the user to easily mount the pyrometer on vacuum systems.
Vacuum Pickup
2.6.5 Scanning Mirror
The scanning mirror unit SCA 5 allows the measured object to be scanned over a certain range. The measuring beam of the pyrometer moves straight in one line across the object and collects temperature data of this line. This is useful when used in combination with the maximum value storage (peak picker) to measure objects which move out of the target area. The scanning angle of the mirror is 0 to 12° and the scanning frequency 0 to 5 Hz. Both values are easily adjustable at the scanner.
2.6.6 Flange System
Tube support with air purge nozzle and flange
Scanning Mirror
IGA 6/23 Advanced Manual Introduction · 12
The flange system is a modular mounting system to fix the pyrometer on furnaces, vacuum chambers, etc.
Ceremaic sighting tube
600 x 24, closed
It can consist of e.g. mounting support, tube support with air purge and flange and an open or closed ceramic sighting tube. The mounting support can be equipped with a quartz window for support vacuum applicationsIt may consist of an equipment rack, flange, and an open or closed ceramic
Schematic drawing of the flange system
2.7 Transport, Packing, Storage
With faulty shipping, the instrument can be damaged or destroyed. To transport or store the instrument, please use the original box or a box padded with sufficient shock-absorbing material. For storage in humid areas or shipment overseas, the device should be placed in welded foil (ideally along with silica gel) to protect it from humidity.
The pyrometer is designed for a storage temperature of -20 to 80 °C with non-condensing conditions. Storing the instrument out of these conditions can cause damage or result in malfunction of the pyrometer.
IGA 6/23 Advanced Manual Introduction · 13
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced Manual Introduction · 14
3 Controls and Installation
3.1 Electrical Installation
The IGA 6/23 Advanced is powered by a voltage of 24 V DC ± 25% (ripple < 50 mV). It is important to ensure correct polarity when connecting the device to the power supply.
To meet the electromagnetic requirements (EMV), a shielded connecting cable must be used.
LumaSense offers connecting cables, which are not part of the standard scope of delivery. The shield of the connecting cable has to be connected only on the pyrometer’s side. If the connecting cable is extended, the shield of the extension also needs to be extended. The shield must be open on the power supply side (switch board), to avoid ground loops.
The connecting cable has wires for the power supply, interface, analog output, switch contact, and external clearing of the maximum value storage via contact and 12-pin connector (see
Chapter
, Reference numbers).
Attention: When connecting the power supply, ensure the polarity is correct.
Once the instrument has been connected to the power supply, it is immediately ready for use.
Although it does not need to be warmed up, it does need to run for approximately 15 to 30 minutes before achieving full accuracy. The instrument can be switched off by interrupting the power supply or unplugging the electrical connector.
3.1.1 Pin assignment of the connector
12-pin connector
Pin assignment
(view of the connector pins)
IGA 6/23 Advanced Manual Controls and Installation · 15
Pin Color Function
K white +24 V DC power supply
A brown
J pink
0 V DC power supply
L
B green + I
Ausg.
yellow
H gray
– I
Ausg.
analog output analog output
Targeting light activate / deactivate via external switch (bridged with K)
External clearing of max. value storage
(bridge to K) or hold function*
G
F red DGND (GND for interface) black B1 (RS485) or RxD (RS232)
C violet A1 (RS485) or TxD (RS232)
D gray-pink B2 (RS485) (bridged with F)
E red-blue A2 (RS485) (bridged with C)
M orange
Screen only for cable extension, don’t connect at the switchboard
*The connector pin J can be used for two different functions:
1. External clearing of the maximum value storage: When the pyrometer is in operating mode, pin J can be used for external clearing of maximum value storage.
When external clearing is selected from the tclear dropdown menu, pin J is connected for a short time to pin K (power supply voltage) to clear the stored maximum value. The function “external clearing” is triggered when the clear time is set to “extern”.
2. Hold function: When the hold function mode is activated, the current temperature reading is frozen as long as pin J and pin K are connected. (See section 4.6 for clear time for the maximum value storage).
3.1.2 Connecting the pyrometer to a PC
The pyrometer is equipped with an RS485 serial interface. With the RS485, long transmission distances can be realized and the transmission is, to a large extent, free of problems. The RS485 also allows several pyrometers to be connected in a bus system.
If an RS485 connection is not available at the PC, it can be accomplished using the RS485 to USB connector. When using a RS485 to USB adapter, make sure that the adapter is fast enough to receive the pyrometer’s answer to an instruction of the master. Most of the commonly used adapters are too slow for fast measuring equipment, so it is recommended to use the
RS485 ó USB converter “USB nano“ (ref. no. 3 852 600).
3.1.3 Connection to RS485
Half-duplex mode: A1 and A2 as well as B1 and B2 are bridged in the 12-pin round connector of the connecting cable, to prevent reflections due to long stubs. It also safeguards against the interruption of the RS485 bus system, should a connecting plug be pulled out. The master labels mark the connections on the RS485 converter.
The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the cable. Adjusted values are from 1200 to 115200 Bd.
IGA 6/23 Advanced Manual Controls and Installation · 16
Terminator
Pyrometer 1 e.g. Address 00
Pyrometer 2 e.g. Address 01
Pyrometer 32 e.g. Address 31
RS485 Bus System
3.1.4 Connection of Additional Units
For temperature indication of the pyrometer, LumaSense offers pure indicators (series DA 4000).
LumaSense also offers indicators with features to change pyrometer parameters (DA 6000 and
DA 6000-N) as well as a fast digital PID controller PI 6000.
Digital Display Parameterizing Indicator Digital Controller
Additional analyzing instruments, including LED digital displays only need to be connected to a power supply and the analog outputs of the pyrometer (exception: the digital display DA 6000 can also be connected with its serial interface, whereas the digital display DA 6000-N has to be connected with its serial interface).
Connection of additional units
Other instruments like an analog controller or printer can be connected to the display in a series as shown above (total load of resistance max. 500 Ohm).
IGA 6/23 Advanced Manual Controls and Installation · 17
3.2 Sighting
The IGA 6/23 Advanced can be purchased with Through-Lens Sighting (viewfinder) or with a
Laser Targeting Light. These sighting options allow you to easily align the pyrometer to the measuring object.
3.2.1 Viewfinder
The IGA 6/23 Advanced can be equipped with a viewfinder which offers through-lens sighting.
The viewfinder is true-sided and parallax-free. A circle marks the position of the measuring spot, but not the exact spot size.
The viewfinder is equipped with an adjustable eye-protection filter, which allows you to view hot objects without exposing your eye to high intensity light. When you are viewing hot objects, turn the adjustment ring on the viewfinder so the filter will let in less light.
Warning: To ensure eye protection, make sure you use the shaded filter when viewing objects with temperatures of 1500 °C or higher.
Viewfinder Adjustment Ring
Note: You can turn the adjustment ring on the viewfinder in both a clockwise and counterclockwise direction to change the filter from light to dark.
When you are viewing low temperature objects, turn the viewfinder adjustment ring so the filter will let in more light.
3.2.2 Laser Targeting Light
The IGA 6/23 Advanced can be equipped with a laser targeting light to assist with aligning the pyrometer to the measuring object. The laser targeting light is a visible red light with a wavelength between 630 and 680 nm and a maximum power of 1 mW. The laser is classified as product of laser class II.
The laser spot marks the center of the measuring spot on the target. The diameter does not correspond to the spot size. The smallest diameter of the laser spot approximately indicates the measuring distance.
Warning: The instrument is equipped with a class II laser that emits radiation.
To reduce the risk of injury to the eyes, do not look directly into the targeting laser and do not point the targeting laser into anyone's eyes.
IGA 6/23 Advanced Manual Controls and Installation · 18
Never look directly into the laser beam. The beam and spot can be watched safely from side.
Also, make sure that the beam will not be reflected into eyes of people by mirrors or shiny surfaces.
The laser targeting light can be switched on and off by pressing the button of the rear cover of the housing.
Push button for
Laser Targeting Light
Note: The laser warning signs on the pyrometer should be easily viewable at all times, even after it has been installed.
The laser targeting light can also be switched on and off by using an external contact (see section 3.1.1 Pin Assignment) or through the InfraWin software. If it is not switched off by one of the above-mentioned methods, it will be switched off automatically after approximately
2 minutes.
To prevent damage to the laser, the targeting light is also switched off automatically if the internal temperature of the pyrometer exceeds 61 °C. It can only be used again once the temperature falls below 61 °C.
3.3 Optics
3.3.1 Spot Sizes
The IGA 6/23 Advanced has a Vario optics, which can be manually adjusted at all distances between 210 mm and 5000 mm.
The table of Spot Sizes in Relation to Measuring Distance shows examples of the pyrometer’s spot size M [mm] in relation to the measuring distance a [mm] (min. 90% of the radiation intensity). Increasing or decreasing the measuring distance will change the spot size.
Note: Since these are 1-color (mono) pyrometers, they can measure objects at any distance (whether focused or non-focused). However, the object has to be bigger than, or at least as big as, the spot size of the pyrometer in the measuring distance.
IGA 6/23 Advanced Manual Controls and Installation · 19
Aperture D for all temperature ranges is 13 to 15 mm with the aperture being the effective diameter of the lens. This is dependent on the optical setting. The largest value applies to a very small measuring distance, while the minimum value applies to the largest measuring distance.
Measuring
Distance
a [mm]
210
300
500
800
1300
2000
5000
Spot Diameter
M [mm]
MB 10
4.2
6
10
16
26
40
100
Spot Diameter
M [mm]
MB 13
2.1
3
5
8
13
20
50
Examples of Spot Sizes in Relation to Measuring Distance
Spot Diameter
M [mm]
MB 18
0.6
0.9
1.5
2.3
3.7
5.8
15
Note: Effective aperture D for all temperature ranges is 13 mm (focused to longest distance) to 15 mm (focused to shortest distance).
Focused spot sizes between the listed distances can be found by linear interpolation between the listed values.
3.3.2 Deviation from the focused measuring distance
Spot sizes for non-focused distances (shorter or longer than the focused distance) may be calculated by using the formula below.
Formula for Calculating Spot Sizes
The InfraWin software also includes a Spot Size Calculator that calculates the data for the non-focused regions, if you enter the values of aperture D, focused measurement distance a, and focused measuring field diameter M as found in the above table (see section 3.3.1).
3.3.3 Finding the right measuring distance
The measuring distance can be set using the Focus Adjustment Screw on the back of the device.
The focused distance value can be adjusted at all distances between 210 mm and 5000 mm. To focus, turn the focus adjustment set screw to make the target image appear sharp and clear.
Note: The optics are manually focusable with a measuring distance of a = 210 to
5000 mm.
IGA 6/23 Advanced Manual Controls and Installation · 20
The LED Distance Indicator Light (labeled mm) will turn red and the approximate focused measuring distance in mm will automatically be shown on the Digital Display for a few seconds after making an adjustment using the Focus Adjustment Set Screw.
Note: Turning the focus adjustment screw counterclockwise will shorten the measuring distance.
Turning the focus adjustment screw clockwise will lengthen the measuring distance.
IGA 6/23 Advanced Manual Controls and Installation · 21
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced Manual Controls and Installation · 22
4 Settings / parameter descriptions
The pyrometer is equipped with a wide range of settings for optimal adaptation to the required measuring condition and to measure the temperature correctly.
All settings can be read and set only in the pyrometer parameters window of the software InfraWin. Adjusting the settings at the instrument is not possible (detailed description of the software see Chapter
Selecting the pyrometer parameters window shows the current settings of the pyrometer. Changing a value can either be done by typing a value in an input box or by selecting a preset value from the list field.
4.1 Factory Settings
Temperature display (°C / °F) = °C
Emissivity (ε) = 100%
Transmittance (W) = 100%
Exposure time (t
90
) = min
Dynamical adaptation dm = 4
Clear time (t
Clear
) = off max / min = maximum value storage preset
Analog output (out) = 0 ... 20 mA
Sub range same as basic temperature range
Ambient temperature compensation = auto
Address = 00
Baud rate = 19200 Bd
4.2 Temperature Display
The Digital Display on the back of the pyrometer will show the temperature measurement in either °C or °F. You can select which temperature scale you wish to use through the InfraWin software or by using the UPP Data Format commands.
The Measuring Value can also be viewed at any time through the InfraWin Software or by using the UPP Data Format commands.
4.3 Emissivity ε
Different materials have different emissivities ranging between 0% and 100%. The emissivity is also dependent upon on the surface condition of the material, the spectral range of the pyrometer, and the measuring temperature. The emissivity setting of the pyrometer has to be adjusted accordingly.
Emissivity settings between 5% and 100% can be established through the InfraWin software or by using the UPP Data Format commands.
IGA 6/23 Advanced Manual Settings / parameter descriptions · 23
Measuring object
“Black body furnace“
Extruded Aluminum
Brass
Brass oxidized (tarnished)
Copper
Copper, oxidized
Inconel
Inconel, oxidized
Oxidized Iron
Steel rolling scale
Emissivity [%]
(at 2.3 µm)
100
13
18
65 to 70
5
70 to 80
30
85
85 to 90
80 to 88
Measuring object
Steel, molten
Nickel
Titanium, non-oxidized
Titanium, oxidized
Molybdenum
Molybdenum, oxidized
Black Carbon
Graphite
Stoneware, glazed
Porcelain rough
Emissivity [%]
(at 2.3 µm)
20 to 30
15 to 20
50
75 to 80
10
75 to 80
95
80 to 90
60
80 to 90
Note: Emissivity e can be set from 5% to 100% in steps of 0.1%
4.4 Transmittance t
Transmittance is a parameter that can compensate for signal loss due to external windows etc.
For example, if the emissivity of the material is 0.6 and the transmittance of an additional window is 0.9, then the product would be 0.54 which is well inside the allowed range.
The product of transmittance and emissivity ( t x e
) must not be less than 5%
.
Note: Transmittance t can be set from 5% to 100% in steps of 0.1%
4.5 Response Time (t
90
)
The response time t
90
is the time interval for the analog output of the pyrometer to go from a low temperature value up to 90% of the temperature step to a high value when measuring an abrupt increase from said low to said high temperature.
The dynamic exposure time adjustment prolongs the exposure time at the lower range limit.
With the InfraWin software and the UPP Data Format commands it is possible to configure the dynamic exposure time.
Dependency of the response time on the temperature value
MB 10
MB 13
Temperature value t
90 min
Temperature value
60 °C
2 ms
MB 18 t
90 t
90 min
Temperature value min
100
°C
2 ms
200
°C
4 ms
120 °C
1 ms
150
°C
1 ms
250
°C
1 ms
150 °C
0.5 ms
200
°C
0.5 ms
300
°C
0.5 ms
Independently of this, the pyrometer performs a measurement every 25 µs and updates the analog output. Slower response times can be used to achieve a constant temperature reading for measuring objects that have rapidly fluctuating temperatures.
The response time is set using the InfraWin software or by using the UPP Data Format commands.
When the setting is set to min., the IGA 6/23 Advanced operates using a time constant of
IGA 6/23 Advanced Manual Settings / parameter descriptions · 24
approximately 0.5 ms. The response time can be extended to 1 ms; 3 ms; 5 ms; 10 ms; 50 ms;
250 ms;1 s; 3 s; 10 s.
Note: Settings for Response Time t
90 min, 1 ms; 3 ms; 5 ms; 10 ms; 50 ms; 250 ms; 1 s; 3 s; 10 s
4.6 Clear Peak Memory (t
CLEAR
)
The integrated maximum value storage is activated when the parameter t clear is set to something other than “OFF” or “HOLD”.
If the maximum value storage is switched on, the highest last temperature value will always be displayed and stored. As such, it may be beneficial to periodically clear and reset the stored maximum values in order to obtain new temperature readings.
This storage also has to be cleared at regular intervals when fluctuating object temperatures cause the display or the analog outputs to change too rapidly or when the pyrometer is not constantly viewing an object to be measured.
4.6.1 Single and Double Storage Modes
Depending upon the selected settings, the maximum value storage will either work in single storage mode or in double storage mode.
Single Storage
Mode:
Double Storage
Mode:
Settings: off
10 ms
50 ms
250ms
1 s
5 s
25 s extern auto
Single storage mode is used when you want to reset the stored value using an external impulse via one contact closure from an external relay (such as between two measured objects). The relay contact is connected directly to the pyrometer between pins J and K. This mode allows a new value to be established after each impulse from the reset signal. Single storage mode also comes into effect when the Clear Peak Memory t clear
is set to AUTO.
Double storage mode comes into effect when selecting one of the reset intervals. This mode utilizes two memories. With the first memory, the highest measured value is held and is deleted alternately in the time interval set (clear time). The other memory retains the maximum value throughout the next time interval. The disadvantages of fluctuations in the display with the clock frequency are thereby eliminated.
Note: The maximum value storage setting coincides with adjustments made to the response time.
The response time setting (working like a low-pass filter) is applied first. After that, the maximum storage is processed. So when using both, the maximum storage takes the peak of the signal that was previously smoothed by the response time filter.
IGA 6/23 Advanced Manual Settings / parameter descriptions · 25
4.6.2 Clear Time Settings
The following settings are available through the InfraWin software or by using the UPP data format commands.
OFF
10 ms…25,0 s
EXTERN
AUTO
HOLD
When set to OFF, the maximum value storage is switched off and all new temperature values are measured but not stored.
If the clear time is set between 10 ms and 25.0 s, the maximum value is held in double storage mode. After the entered time, the value will be cleared alternately from one of the storages, while the value of the other storage is shown.
With the external clearing function, the storage operates in single storage mode. The values are immediately cleared from the storage by contacting the wires connected to pins J and K, if the EXTERN mode was selected.
The AUTO mode is used for discontinuous measuring tasks. For example, when objects are being transported on a conveyer belt and pass the measuring beam of the pyrometer only for a few seconds. In this case, the maximum value for each object has to be obtained.
With the AUTO mode, the maximum value is stored until a new hot object appears in the measuring beam. The temperature, which has to be recognized as “hot” is defined by the low limit of the adjusted sub range.
The stored maximum value will be deleted once the temperature of the new hot object exceeds the low limit of the sub range by 1% (transition in positive direction) or by at least 2 °C. This is also valid if the sub range equals the basic range.
The HOLD function allows you to freeze the current temperature reading at any moment. This feature is activated using an external switch that has been connected between connector pins J and K.
The temperature reading will remain frozen as long as the contact remains closed.
4.7 Analog Output
The analog output has to be selected according to the signal input of the connected instrument (controller, PLC, etc.). If 4 to 20 mA is set, the analog output gives 3.9 mA for temperatures below lower range limit.
Einstellungen:
0 … 20 mA
4 … 20 mA
4.8 Sub Range
You have the opportunity to choose a sub range (minimum span 50 °C) within the basic measuring range of the pyrometer. This sub range corresponds to the analog output.
Example: Range 150…1800 °C, Sub Range 925…975 °C.
The sub range setting also affects the maximum value storage when the Clear Peak Memory t clear
is set to AUTO. For more information on the t clear
AUTO setting, refer to section
.
Note: Settings for Sub Range - any range adjustable within the temperature range with a minimum span of 50 °C.
IGA 6/23 Advanced Manual Settings / parameter descriptions · 26
4.9 Device Address
When connecting several pyrometers to one serial interface with RS485, it is necessary for each instrument to have its own device address for communication purposes. First, it is necessary to connect each instrument separately to give it an address. After that, all instruments can be connected and addressed individually.
Settings:
00
...
97
Note: Only via own communication program with interface command (not possible with InfraWin, because InfraWin automatically detects a connected pyrometer): If parameters should be changed simultaneously on all pyrometers, the global Address 98 can be used. This allows you to program all pyrometers at the same time, regardless of the addresses that have already been assigned. If the address of a pyrometer is unknown, it is possible to communicate with it using the global
Address 99 (connect only one pyrometer).
4.10 Focused Distance
The focused distance value can be adjusted at all distances between 210 mm and 5000 mm using the Focus Adjustment
Screw on the back of the device.
The LED Distance Indicator Light will turn red and the focused measuring distance in mm will automatically be shown on the
Digital Display within a few seconds of making an adjustment using the Focus
Adjustment Set Screw.
The focused distance can be viewed at any time through the InfraWin software or by using the UPP Data Format commands.
4.11 Baud Rate
The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the cable. A maximum cable length for 19200 Bd with RS485 is 2 km. The baud rate is reduced by
50% if the transmission distance is doubled.
Baud Rates can be set at 1200, 2400, 4800, 9600, 19200, 38400, 57600, or 115200.
4.12 Pyrometer Internal Temperature
The internal temperature of the pyrometer can be read through the PC interface using the
InfraWin software or by using the UPP Data format commands. It is a few degrees higher than the ambient temperature due to the heat generated by the electronics.
When using the Laser Targeting Light, the targeting light is switched off automatically if the internal temperature of the pyrometer exceeds
61 °C. This safety feature is used to prevent damage to the laser. It can only be used again once the temperature falls below 61 °C.
IGA 6/23 Advanced Manual Settings / parameter descriptions · 27
4.13 Ambient Temperature Compensation
The compensation of the ambient temperature can be set for temperatures within the basic measuring range. This compensation is only used for very few special applications. The standard setting of this parameter is “auto", because the temperature of the air around the pyrometer is normally the ambient temperature of the measured object. Should the measured object be instrument). After switching over to “man” the corresponding data field T the ambient temperature value can be entered.
amb
Settings: auto
...
70 °C (158 °F) placed in an area with a higher wall temperature (e.g. inside a furnace), the measurement might be falsified (probably too high temperature indication). This influence can be compensated by presetting the ambient temperature of the object (presetting within the measuring range of the
is activated so that
Attention: It has to be considered that this method only improves the results if the ambient temperature at the place of the measured object is always constant and the emissivity is well known.
IGA 6/23 Advanced Manual Settings / parameter descriptions · 28
5 Software InfraWin
The operating and analyzing InfraWin software is included with delivery of the pyrometer. In addition to allowing you to make parameter adjustments, the InfraWin software also provides temperature indication, data logging, and measurement analysis features.
This section gives an overview about the functions of the software. It also provides a description of the individual icons found in the program's help menu. Click on the F1 button after loading
InfraWin or click on the ? in the menu bar.
The following descriptions refer to software InfraWin version 5.0. The latest version is available for free as download from the homepage www.lumasenseinc.com.
5.1 Connecting the pyrometer to a PC
The program InfraWin can operate up to two devices. Two devices using RS485 may be operated simultaneously by the same interface, if two different addresses have been properly entered
(see section
Device Address for more information).
5.2 Installation
To install the InfraWin software, select setup.exe from the InfraWin-CD or from the downloaded and unpacked zip file from the internet and then follow the installation instructions.
5.3 Program start
The first time you load InfraWin 5, you will be prompted to select a default language. The
InfraWin software is available in German, English, Spanish, French, Portuguese, and Chinese.
Once installed, click Language/Languages if you would like to select another language.
On the start page, the screen shows the following start menu icons.
Opens a saved file
Storage of measured values for further processing
Starts measurement for the selected device
Setting of the parameters of the instruments
Time interval between two measurements
Listing of measured or stored values in tabular form
Processing of measured (stored) readings in graph form
Processing of measured (stored) readings in a text file
Calculation of spot sizes in various measuring distances
Only if available: controls the programmable controller PI 6000
Search I/O module
IGA 6/23 Advanced Manual Software InfraWin · 29
5.4 Basic settings
All preset values for the device can be displayed and modified, if necessary under the
Devices/Parameters window.
Changing an existing pyrometer setting can be accomplished by typing a value in an input box or by selecting a preset value from the list field.
Choose the correct settings for your application from the displayed options.
This window contains the parameter settings described in Chapter
,
Parameters.
5.4.1 Open/Save
The open and save buttons enable you to save customized configurations, which can then be opened and reused at any time.
5.4.2 1 measur…
The 1 measur… function shows the current measuring temperature in the pyrometer parameters window for approximately 1 second.
5.4.3 Print
The print icon will generate a screenshot of the parameter form and allow you to send it a printer.
5.4.4 Close
The close icon will close the pyrometer parameters window.
5.4.5 Test
Clicking on the Test icon will open a window that allows you to directly communicate with the pyrometer using the UPP (Universal
Pyrometer Protocol) Data Format commands.
After entering an interface command (00 is the adjusted address ex works, ms is the command reading temperature value) and clicking on Send, the window shown to the right will be opened.
This window already shows the answer of the pyrometer in 1/10°. The actual temperature reading is 49.0 °C.
Len indicates the length of the answered data string, incl. Carriage Return (Chr(13)).
IGA 6/23 Advanced Manual Software InfraWin · 30
In the lower part of the window, the connection with the preset baud rate can be checked. Here the command was sent 100 times with 19200 baud. It has taken 1.100 seconds without transmission errors.
Emi: AutoFind: If the true temperature of the measured object is known, you can calculate the emissivity of the measured object using the Emi: AutoFind function:
A measured temperature is displayed with the current set emissivity (in this example
100%) (here: 116 °C).
If you press Emi: Autofind a window will open which allows you to enter the "true" temperature (here 130 °C).
Once the temperature entry has been entered and confirmed with Calculate, InfraWin will then calculate the emissivity which occurs with the new temperature (here 79.0%). This is displayed immediately and can be used for further temperature measurement.
5.5 Measurement online trend
The measurement function allows you to access a number of input tabs located on the left side of the screen. The main or home tab is the Output
Screen. You can toggle the input tabs on and off by clicking them. This window displays:
· temperature as graphical diagram
· internal temperature of the instrument
· current temperature
· quantity of the measured values and file size of the current measurement
The example shows a sample reading over the period of approximately 15 seconds with a temperature range between 400 and 800 °C. The final temperature (at the end of the reading) is
431 °C.
The Mark Zone button allows you to color mark a temperature range for easier recognition.
The Threshold button allows you to set a temperature value as a baseline to prevent recording values above or below the baseline temperature. This allows you to keep the output file size small.
IGA 6/23 Advanced Manual Software InfraWin · 31
The Scaling Trend button allows you to scale temperature trend view.
Note: The measuring values of “measurement online trend” are automatically saved as "standard.i12". Should you need to edit the data later, you need to save the file as another .i12-file because old values are over-written when a new measurement is taken.
Files from older program versions (.i10-files) can be opened and saved as .i12.
5.6 Listing (analyzing)
For analyzing the measured values in this field, all measured data appears in a numeric list.
The column between time and temperature provides a time resolution of milliseconds. The value specifies the time in seconds after midnight (0:00 h). The amount of data depends on the frequency that readings were taken
(settings at 10.9 PC sampling rates).
As the amount of data increases, so does the amount of storage space required to save it. In order to save room, all .i12 data files are stored by a binary code.
5.7 Output .TXT file (analyzing)
The same file as under Output listing may be converted into a text file and can be easily opened, for example with Microsoft Excel. With the standard import settings, Excel automatically formats the columns accordingly (with tabs as separators).
5.8 Trend output (analyzing)
The graph’s curve depicts the temperature change over time within the specified temperature range.
Additionally, other information appears in this window; such as recorded time (x-axis) and temperature in degrees (y-axis) as well as the time and temperature at the vertical cursor line which can be dragged with the mouse.
Selecting the Trend output initially causes all the saved data to be displayed.
If the data exceeds an amount that can be reasonably represented, you may “Zoom“ in on a partial segment using the mouse (such as the segment represented in the example). Under
“Total” you can return to the representation of the entire curve.
IGA 6/23 Advanced Manual Software InfraWin · 32
Note: The last reading is saved in the standard.i12 file and automatically appears in this form upon opening Listing or Trend output.
Selecting file open with another file, the previous file will be overwritten and replaced by the standard.i12 file.
5.9 PC sampling rate (time interval between two measurements)
This function sets a time interval. After each interval, one measured value is stored on the PC. Longer time intervals will result in creating smaller stored file sizes. This function is mainly used for long term measurements.
5.10 Spot size calculator
The InfraWin Spot Size Calculator calculates the data for the nonfocused regions. To calculate data, enter the values of aperture D, focused measurement distance a, and focused measuring field diameter M as found in the documentation relating to the optics supplied with your specific pyrometer.
After entering the aperture and the main spot size, the input of interim values calculates spot sizes in different measuring distances of the fixed optics.
5.11 Search I/O Module
The I/O Module allows accessories to connect to the software and is used to trigger measurement externally or to send a signal
(like a relay) under certain conditions.
IGA 6/23 Advanced Manual Software InfraWin · 33
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced Manual Software InfraWin · 34
6 Maintenance
6.1 Cleaning the front window
Since the device does not contain parts that require regular maintenance, the only regular maintenance required is periodic inspection of the front window for build-up of foreign particiles. If allowed to build up, the particles can influence the energy received by the instrument.
The IGA 6/23 Advanced window is not water soluable and can be cleaned with standard lens tissue dampened with a commercially available glasses or camera lens cleaning solution. Use a soft blower/brush (available in camera stores) to remove any grit on the window before you rub the lens with lens tissue and solution.
Attention: NEVER CLEAN THE IGA 6/23 Advanced WINDOW WITH A DRY TISSUE
OF ANY KIND! The only time dry lens tissue may be used is to dry a window which has already been cleaned with wet lens tissue.
6.2 Calibration
LumaSense calibrated your IGA 6/23 Advanced pyrometer at the factory and delivered your instrument with a Works Certificate. Normally we advise against changing the factory set calibration. If you believe that the calibration may have changed, perhaps because your operating environment is severe, an approximate field (on-site) calibration is possible. You have the choice of an on-site calibration or arranging a more precise calibration at the LumaSense factory.
6.2.1 Laboratory Calibration
Contact LumaSense for information about calibration at the LumaSense factory. For most of our customers who do not have large numbers of infrared thermometers in service, we recommend that our laboratory be chosen to do calibration. When you have many infrared thermometers, you may find it most convenient and economical to have your own calibration laboratory.
LumaSense has a variety of blackbody calibration sources including very economical ones.
Among these you may find the source most suitable for your own laboratory.
6.2.2 On-Site Calibration
See Chapter 9 Troubleshooting before attempting to perform your own calibration. The instrument was calibrated at the factory to its original accuracy as stated.
Nearly all erroneous temperature readings are caused by application problems such as:
· Reflections: Unexpectedly high readings may be caused by the IGA 6/23 Advanced
“seeing” a reflection of another hotter source, especially if the target emissivity is less than 0.8.
·
Spot Sizes: Ensure the IGA 6/23 Advanced is using a proper focused distance and takes into account the spot size in relation to measuring distance. The pyrometer can measure objects at any distance. However, the object has to be bigger than or at least as big as the spot size of the pyrometer in the measuring distance.
·
Obscured window: Clean the window. Refer to the cleaning window section 6.1.
IGA 6/23 Advanced Manual Maintenance · 35
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced Manual Maintenance · 36
7 Data format UPP (Universal Pyrometer
Protocol)
Software commands can be exchanged directly with the pyrometer through an interface using suitable communication software or by using the “Test” function located in the “Pyrometer
Parameters” window of the InfraWin software package.
The data exchange occurs in ASCII format with the following transmission parameters:
• The data format is: 8 data bits, 1 stop bit, even parity (8,1,e) no handshake;
• The device responds to the entry of a command with output (such as the measuring value) + CR (Carriage Return, ASCII 13), to pure entry commands with “ok” + CR, or “no”
+ CR.
• Every command starts with the 2-digit device address AA followed by two lower case command letters and finished with CR.
Example Read Command: Entry: “00em“ + <CR>
The emissivity setting (e) of the device with the address 00 is returned.
Answer: “0970“ + <CR> means Emissivity = 0.97 or 97.0%
• The ASCII parameter “X” indicates a change to be made in a parameter. When used, the command contains the new value.
Example Write Command: Entry: “00emXXXX“ + <CR>
The parameter used for the emissivity setting (e) with the address 00 is changed.
Answer: “00em0853“ + <CR> changes the Emissivity to 0.853 or 85.3%
·
A “?” after the lower case command letters answers with the limits of the respective settings (only at setting commands, not at query commands).
Example Read Limits Command: Entry: “00em?“ + ˜CR!
Answer: Could be 00501000 + <CR>, which means that E can vary between 0.050 and
1.000 (or 5% and 100%)
Description
Analog Output
Command
AAasX
Parameters
Baud rate (set)
Emissivity e
Transmittance window t
of
AAbrX
AAemXXXX
AAem
AAetXXXX
AAet
X=0 (0 to 20 mA) X=1 (4 to 20 mA)
Output: XXXXXX (hex 6-digit)
X = 0 to 6 or 8 (dec.)
0 = 1200 Baud 1 = 2400 Baud 2 = 4800 Baud
3 = 9600 Baud 4 = 19200 Baud 5 = 38400 Baud
6 = 57600 Baud 7 = is not allowed 8 = 115200 Baud
XXXX = 0050 to 1000 e
=0.050 to 1.000
Answer: DDDD 4 decimial digits 0050 to 1000
XXXX = 0050 to 1000 e
=0.050 to 1.000
Answer: DDDD 4 decimial digits 0050 to 1000
IGA 6/23 Advanced Manual Data format UPP
Ò
· 37
Description
Response time t
90
Dynamic exposure time
Internal temperature
(read)
AAdmX
Temp display °C or °F AAfhX
Device address AAgaXX
AAgt
Ambient temperature compensation
AAtm
AAutXXXX
Laser
Command
AAezX
AAlaX
AAla
Parameters
X=0 to 9
0=min
4=10 ms
X=0 to 9
0 = off, 1 = min, 9 = max
5=50 ms
8=3 s
3=5 ms
6=250 ms
9=10 s
Output: X=0 Display
X=1 Display in °F
XX=(00 to 97)
00 to 97=regular device addresses
99=global address with response
Answer: DDD 3 decimal digits (000 to 098 °C or
032 to 210 °F) gt=current temp. tm=maximum temp. (memory)
XXXX = value of ambient temperature, 4-digit, hex
XXXX e.g. FFEC corresponds to -20 degrees
-99dez = FF9Dhex means: automatic, no manual compensation
X=0 off X=1 on
Answer: 1-digit: “0” or “1”
Wait time AAtwXX XX = 00 … 99
(decimal, in bit time of the current baud rate)
Clears maximum storage Software simulation of external clearance
Clear peak memory t clear range
Sub range (set)
AAlx
AAlzX
AAmb
AAme
1. AAm1XXXXYYYY
X=0 to 9
2=50 ms
4=1 s 5=5 s
8=AUTO
6=25 s
9=HOLD
Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
XXXXYYYY=2x4 hex-digit for lower and upper sub range limit (°C or °F)
2. AAm2
Measuring AAms
Device type AAna
AAm2 confirms the change (auto reset)
Answer: QQQQQ (88880=Overflow)
5 decimal digit (in °C or °F, last digit is 1/10 °C or °F)
Output: “IGA 6/23 Advanced”
(16 ASCII-characters)
IGA 6/23 Advanced Manual Data format UPP
Ò
· 38
Description Command Parameters
Read Aapa Answer: 15 decimal digits
DD.............
: Emissivity (see em)
..D............
: Response time (see ez)
...D...........
: Clear peak memory (see lz)
....D..........
: Analog output (see as)
.....DD........
: Internal temperature (see gt)
.......DD......
: Device address (see ga)
.........D.....
: Baud rate (see br )
..........01000
: Fixed value
Serial number
Device type /
Software version
AAsn
AAve
Output: XXXXX (hex 5-digit)
Answer: VVMMJJ VV=27
MM=Month JJ = Year of software version
AAvc Communication
Module/ software version in detail
Software version in detail
AAvs tt.mm.jj XX.YY
tt = day; mm = month; jj = year;
XX.YY = Software version tt.mm.jj XX.YY
tt = day; mm = month; jj = year;
XX.YY = Software version
Note: the letter “l” means the lower case letter of “L”.
Additional instruction for the RS485 interface:
Requirements to the master system during half-duplex operation:
1. After an inquiry, the bus should be switched into a transmission time of 3 bits (some older interfaces are not fast enough for this).
2. The pyrometer's response will follow after 5 ms, at the latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be repeated.
4. After receiving the response, the master has to wait at least 1.5 ms before a new command can be entered.
IGA 6/23 Advanced Manual Data format UPP
Ò
· 39
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced Manual Data format UPP
Ò
· 40
8 Reference Numbers
8.1 Reference numbers instrument
Laser Viewfinder
50 to 1000 °C (MB 10)
75 to 1300 °C (MB 13)
3 914 210
3 914 250
3 914 220
3 914 260
150 to 1800 °C (MB 18) 3 914 290 3 914 300
Ordering note:
A connection cable is not included in scope of delivery and has to be ordered separately.
8.2 Reference numbers accessories
3 820 320 Special connection cable with angled connector and additional targeting light push button, 5 m long
3 820 330
3 820 500
3 820 510
3 820 810
3 820 820
3 820 520
3 820 340
3 820 530
3 820 540
3 820 830
3 820 840
3 820 550
Connection cable, 5 m, with straight connector
Connection cable, 10 m, with straight connector
Connection cable, 15 m, with straight connector
Connection cable, 20 m, with straight connector
Connection cable, 25 m, with straight connector
Connection cable, 30 m, with straight connector
Connection cable, 5 m, with right angle connector
Connection cable, 10 m, with right angle connector
Connection cable, 15 m, with right angle connector
Connection cable, 20 m, with right angle connector
Connection cable, 25 m, with right angle connector
Connection cable, 30 m, with right angle connector
(All connection cables include a short adapter cable with a
9-pin D-SUB connector. This connector may be used in combination with the RS485 to USB adapter.)
3 826 510 PI 6000: PID programmable controller, very fast, for digital IMPAC pyrometers
3 826 720 USB to RS485 adapter cable, 1.8 m long
3 834 210 Mounting support (adjustable)
3 835 160 Air purge unit, aluminum
3 835 590 90° mirror for Series 5 and Series 6, quartz glass window
3 837 230 Water cooling jacket (heavy duty) with integrated air purge
3 843 250 ROT5 scanning mirror attachment up to 70°
3 843 430 Scanning mirror attachment up to 12°
3 846 260 Instrument's support (Series 5 & 6)
3 846 290 Instrument's support (Series' 5 & 6) with fused silica window
3 846 590 Vacuum support KF16 with quartz glass window
3 852 290
Power supply NG DC for DIN rail mounting; 100 to 240 V AC Þ 24 V DC, 1 A
IGA 6/23 Advanced Manual Reference Numbers · 41
3 852 550
Power supply NG 2D for DIN rail mounting; 85 to 265 V AC Þ 24 V DC, 600 mA with two settable limit switches
3 890 640 DA 4000-N: LED-digital display to be built into the switchboard, 230 V AC
3 891 210 DA 4000-N: LED-digital display to be built into the switchboard, 115 V AC
3 890 650 DA 4000: like DA 4000-N, but additionally with 2 limit switches, 230 V AC
3 891 220 DA 4000: like DA 4000-N, but additionally with 2 limit switches, 115 V AC
3 890 570 DA 6000-N digital display to allow adjustment of the Pyrometer through the RS485 interface
3 890 530 DA 6000: like DA 6000-N with analog input and 2 limit switches for the RS485 interface
3 890 630 LD24-UTP; large digital indicator, 57 mm height of digits
3 834 210 Adjustable mounting support (Series 5 & 6)
IGA 6/23 Advanced Manual Reference Numbers · 42
9 Troubleshooting
No analog output even if the display shows a temperature above lower range limit
Symptom
Temperature reading differs between builtin display and external instrumentation
Temperature readings are too low
Temperature readings are too high
Probable Cause Comments
Reversed leads to IGA 6/23 Advanced The IGA 6/23 Advanced is
Reversed leads to other instruments in the current loop
Open circuit in lines connecting all instruments in current loop diode protected if power leads are reversed.
See section
Total burden higher than 500 Ohms
Insufficient supply voltage (See wiring diagram)
Pyrometer output setting differs from external instrument’s input setting
Target size too small
Target is not at correct focus distance
Incorrect alignment of the pyrometer to the object
Check the 0 to 20 mA or
4 to 20 mA on both sides
Check the sub range setting on pyrometer against the range setting on the external instrument
See section
With small target/close focus models, aiming may require X-
Y positioning table for fine adjustment during aiming
Realign pyrometer to achieve the maximum temperature signal
See Section
Dirty window or lens
Target temperature too low.
Incorrect slope adjustments
Faulty IGA 6/23 Advanced
Targets hotter than expected. (This often happens with first time infrared
Pyrometer users)
The measurement is influenced by reflections of hot machine parts
Contact LumaSense. See section
Verify target temperature with contact type method.
(Thermocouples may not accurately determine surface temperature due to “heat sink” effect)
Use mechanical accessories to avoid the influence of the interfering radiation
If target emissivity is less than 1, reflections from nearby hot objects will influence readings
Incorrect spectral response for application. IGA 6/23 Advanced may be “seeing” through target
Block reflection paths to see if readings drop to expected temperature
See Technicial Specifications
Consult LumaSense for analysis.
IGA 6/23 Advanced Manual Troubleshooting · 43
Symptom
Temperature readings are too high
Noisy readings (fast fluctuations)
Laser Targeting Light fails
Probable Cause
Faulty IGA 6/23 Advanced
Comments
Contact LumaSense. See section
See section
Electrical noise in lead wires caused by intense magnetic fields and/or improper selection of interconnecting cable and/or cable run in same conduit as AC power lines
Lens or window clouding up Increase or decrease air purge rate and/or investigate other cause.
Improper grounding of cable shield and/or IGA 6/23 Advanced housing (or cooling jacket)
IGA 6/23 Advanced mounting not secure
Target moves more than 95% out of field of view. Conveyor type applications will also cause fluctuations as targets move by the field of view
Target may be changing temperature quickly. (Thermocouples do not respond as fast as the
IGA 6/23 Advanced)
Target may be reflecting nearby objects that have fluctuating temperatures
Loose connection in wiring
Electrical noise lead wires caused by intense magnetic fields, improper selection of inter-connecting cable
Cable run into in same conduit as AC power lines
Flame or reflections of flame may be entering the field of view
Extremely dense steam, smoke, thick dust, personnel or moving machinery intermittently blocking 95% of optical path of IGA 6/23 Advanced
Non-uniform emissivity of large moving targets
Faulty IGA 6/23 Advanced
The instrument‘s internal temperature is higher than 61 °C
Application may require
Maximum to be stored. Apply t clear
settings. See section
Speed of response may be slowed down
Apply t
90
settings. See section
Slower response speed is required
Contact LumaSense. See section
Use cooling jacket
Note: The wavelength band of the IGA 6/23 reacts at low measuring temperatures
(below 130 °C) to incandescent lamps or very bright daylight (not valid for fluorescent tube). For a correct measurement strong external light to the measured object should be avoided.
IGA 6/23 Advanced Manual Troubleshooting · 44
Index
1
1 measure… 30
A
Accesories 12
Air Purge 12
Analog Output 26
Aperture 20
Appropriate use 9
B
Basic settings 30
Baud Rate 27
C
Calibration 35
Laboratory 35
On-Site 35
Cleaning the front window 35
Clear Peak Memory 25
Clear Time Settings 26
Close 30
Connecting the pyrometer to a PC 16, 29
Connection cable 9
Connection of Additional Units 17
Connection to RS485 16
Cooling Jacket 12
D
Data format UPP 37
Device Address 27
Device Settings
Clear Peak Memory tclear 25
Dimensions 11
Disposal 7
Double Storage Mode 25
E
Electrical Installation 15
Electromagnetic requirements 15
Emi: AutoFind 31
Emissivity 23
IGA 6/23 Advanced Manual
F
Focus Adjustment Screw 20, 27
Focused Distance 27
G
General Information 5
I
I/O Module 33
Installation 29
Installation, Electrical 15
L
Laboratory Calibration 35
Laser Targeting Light 18
LED Distance Indicator Light 21
Legend 5
Liability 5
Listing (analyzing) 32
M
Maintenance 35
Mark Zone 31
Measurement online trend 31
Measuring distance 20
Measuring Distance 20
Mounting 12
O
On-Site Calibration 35
Open/Save 30
Optics 19
P
Packing 13
PC sampling rate 33
Physical User Interface 11
Pin assignment 15
Print 30
Program start 29
Index · 45
Pyrometer Internal Temperature 27
Pyrometer parameters 23
R
Reference numbers
Accessories 41
Instrument 41
Reference Numbers 41
Repair 6, 7
Response Time 24
RS485 16
S
Safety 5
Scaling Trend button 32
Scope of delivery 9
Service Request 6
Settings / parameter descriptions 23
Sighting 18
Single Storage Mode 25
Software InfraWin 29
Spot size calculator 33
Spot Sizes 19, 20
Spot sizes for non-focused distances 20
Storage 13
Storage Modes 25
Sub Range 26
Support 6
T
Technical Data 9
Temperature Display 23
Test 30
Threshold button 31
Time interval 33
Transmittance 24
Transport 13
Trend output 32
Troubleshooting 43
TXT file 32
U
Unpacking the Instrument 6
UPP data format 37
V
Vacuum Pickup 12
Viewfinder 18
IGA 6/23 Advanced Manual
W
Warranty 5
Index · 46
Advertisement
Key Features
- Non-contact temperature measurement
- Temperature range between 50 and 1800 °C
- Digital signal processing
- Vario optics with manual focus adjustment
- Through-lens sighting or laser targeting light
- Adjustable emissivity, exposure time, and other settings
- RS485 interface for communication
- Analog output for integration with other systems