- Industrial & lab equipment
- Measuring, testing & control
- Honeywell
- Satellite XT FTT
- User Guide
- 75 Pages
Honeywell Satellite XT FTT, FTT/R, FTT/C Gas Monitoring Instrument Guide to Operation
Below you will find brief information for Gas Monitoring Instrument Satellite XT FTT, Gas Monitoring Instrument Satellite XT FTT/R, Gas Monitoring Instrument Satellite XT FTT/C. Satellite XT is a gas monitoring instrument designed for the continuous surveillance of hazardous gas concentrations. It interfaces with LONWORKS™ free topology systems. The instrument is designed to interface with LONWORKS™ free topology systems. A free topology architecture allows the user to wire gas monitoring instruments and control devices with virtually no topology restrictions. The device can be installed with various sensor types, allowing it to monitor a wide range of gases.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Guide to Operation
Satellite XT FTT Versions
Satellite XT 9602-0400 FTT
Satellite XT 9602-0405 FTT/R
Satellite XT 9602-0450 FTT/C
Including Options:
Extractive Module XT
Pyrolyzer Module XT
Table of Contents
Section Operation page.
Principle
1.2
1.2.1
1.2.2
1.2.3 Warning
1.2.4
1.3
1.3.1
1.3.2
1.3.3
1.4
Menu 1-3
Configuration Menu
1.4.2
1.5
Elements
1-3
1-4
Keypad
Display
Technical Specifications 1-5
Section 2 Installation and Start-up
2.1
2.2
2.3
Network
Relay
2.6
2-5
Section 3 Main Menu
3.1
3.2
Main
3.4
3.5
Main
Menu
Main Menu - Configuration 3-5
Section 4 Maintenance
4.1
4.2
4.3
4.4 Device
4.5
Reset
4.7
Section 5 Calibration
5.1
5.1.1
5.1.2
5.2
Satellite XT FTT Versions
Section 6 Configuration
6.1 Settings
page.
6-3
Alarm
6.1.2
6.2 Language
6.3
6.4
6.5
6.6 Password
6.7
6.8
6.9
6.10
Section 7 Troubleshooting
7.1
Warnings 7-1
7.1.2
Section 8 Reference Information
8.1
8.2 K-Factors for Satellite XT C-Versions
8.3
Order
Accessories
8-2
8-3
Network Components: Technical Specifications
8.4
8.5 Module
Supervisor
8-5
Module
Input
8-6
Module
Input
8-7
Module
Modules
8-8
8-9
Section 9 Extractive Module XT - Option
9.1
9.2
9.3
9.4 Instrument
9.5
Tubing
9.7
9.8
Replacement
Technical Specifications 9-6
Section 10 Pyrolyzer Module XT - Option
10.1
10.2
10.3
Instrument
10.5
10.6
Data
10.8
10.9
Tubing
Connection
Technical specifications 10-5
Satellite XT FTT Versions
Basic Operation – Section 1
1.1
This section provides basic understanding of the instrument and its operation.
Principle of Operation
The Satellite XT is a gas monitoring instrument especially designed for the continuous surveillance of hazardous gas concentrations. The instrument is designed to interface with
LONWORKS™ free topology systems. A free topology architecture allows the user to wire gas monitoring instruments and control devices with virtually no topology restrictions
Power is supplied by a local 12 to 24 VDC power supply.
Target gas and measuring range depend on the type of sensor chosen. The sensor comes with the specific data in its internal data memory. When a sensor is inserted, these data are loaded into the instrument’s internal memory.
Satellite XT versions FTT and FTT/R use electrochemical sensors for the surveillance of toxic and corrosive gases at TLV levels (TLV = Threshhold Limit Value). The sensor output signal is amplified electronically, digitised and the resulting concentration value is transmitted to the communication network.
1.2
1.2.1
The Satellite XT FTT/C is used for monitoring combustible gases and vapours. This instrument version uses catalytic sensors and is factory calibrated for the detection of methane in air mixtures with concentrations up to 5 %-vol. (100 % LEL Lower Explosion
Limit). The sensor output signal is amplified electronically, digitised and the resulting concentration value is transmitted to the communication network. To allow the surveillance of a variety of other combustible gases, a correction factor called “K–Factor” can be
entered.
The Satellite XT FTT/R includes the relay option and provides 3 single-pole single-throw relays for activation of external alarm devices. When the actual gas concentration exceeds the alarm levels, the instrument will activate the appropriate alarm relay and display an according message. A relay will also be activated in case of an instrument fault.
Operating States
There are four different operating states for the Satellite XT, Monitoring Mode,
Maintenance Mode, Warning Condition, and Diagnostics. Depending on the selected operating state, the green status LED is either on, off, or flashing.
Monitoring Mode
The Monitoring Mode is the instrument's standard operating mode.
The green LED located above the <set> key is on.
In the Monitoring Mode, the instrument continuously monitors for hazardous gas concentrations and checks for alarm levels and instrument faults.
The instrument's self-diagnostics provide on-line preventive maintenance information identifying electronics or sensor problems. For example a sensor selftest is performed every 24†hours. The sensor selftest is not available with instruments using oxygen sensors or catalytic sensors.
Satellite XT FTT Versions 1-1
Basic Operation – Section 1
1.2.1.1 Alarm Level Settings
The settings for Alarm Level 1 and Alarm Level 2 are automatically loaded when installing the sensor. Standard settings are 1x and 2x TLV or 20 and 40 % LEL for the target gas.
The Alarm Setup function in the Configuration Menu allows the user to change the alarm levels as required.
Whenever the concentration of the target gas exceeds the factory or user programmed alarm levels, the instrument will document this concentration alarm condition on its LCD display and transmit an according message to the network
Instruments with relay option will additionally trigger the associated alarm relays.
1.2.1.2 Reset Alarms Function
Ex factory, Alarm 1 and Alarm 2 are preprogrammed "enabled" and "latching". In the event of a concentration alarm, the alarm indication with latching alarms will continue until the alarm is manually acknowledged by actuating the <set> key. If the password protection is enabled, the password must be entered (see Configuration Menu / Security Function).
Alarm Reset is not possible, if the alarm condition is still present.
1.2.1.3 Exiting the Monitoring Mode
Exiting the Monitoring Mode and entering the Maintenance Mode may be password protected. To prevent the instrument from manipulations by unauthorized personnel, it is recommended to enable this function (see Configuration Menu / Security Function).
- Password enabled:
Actuate the <esc> key to exit the Monitoring Mode and enter the password. The screen to enter the password is shown for one minute. During this period the Monitoring Mode remains active in the background. When the correct password has been entered, the instrument is in the Maintenance Mode, it is not monitoring, and the green LED is off.
1.2.2
- Password disabled:
Actuate the <esc> key to exit the Monitoring Mode. The instrument is now in the
Maintenance Mode, the instrument is not monitoring, and the green LED is off.
Maintenance Mode
The Maintenance Mode indicates a complete absence of monitoring capability. The green status LED is off. A maintenance message is transmitted to the communication network.
Instruments with relay option additionally trigger the fault relay, depending on the configuration.
1.2.3 Warning Condition
A warning condition indicates that the instrument requires some attention, but is still able to monitor and operate as programmed.
The following will occur when the instrument detects a warning condition:
- the green status LED is flashing
- a warning message is transmitted to the communication network; depending on the configuration, this function can be enabled or disabled.
- depending on the configuration, the fault relay will be activated (only instruments with relay option)
1.2.3.1 Reset Warning Condition
Press the <set> key to acknowledge and reset a Warning Condition.
Refer to Section 7, Troubleshooting for specific warnings.
Satellite XT FTT Versions 1-2
Basic Operation – Section 1
1.3.1
1.3.2
1.3.3
1.3
1.2.4 Diagnostics
Instrument faults refer to a problem which prevents the Satellite XT from operating properly and interferes with the ability to monitor or document concentration alarms.
When the instrument has detected an instrument fault, a fault message is transmitted to the communication network and the fault relay is activated (only instruments with relay option). The green status LED is off. The LCD display is flashing while indicating the specific fault message.
The Satellite XT lists specific instrument responses to instrument faults. For information on the various fault messages and instructions on how to correct a fault condition, please refer to section 7, Troubleshooting.
Menu structure
Besides the Monitoring Mode, there is also menu operation for the Satellite XT. Menu operation comprises three groups of functions, maintenance, calibration, and configuration. A password (if enabled) must be entered to exit the Monitoring Mode and to obtain access to the Main Menu.
Maintenance Menu
This function covers the realtime service procedures required for a regular maintenance of the instrument (e.g. sensor replacement).
Calibration Menu
Use the calibration function to dynamically calibrate the Satellite XT with a known concentration of the target gas or manually by entering a new correction K-Factor.
Configuration Menu
Use this function to configure the Satellite XT and to program instrument and monitoring parameters to suit individual requirements.
! Note:
If the instrument was in the Monitoring Mode when last powered down, it will automatically return to the Monitoring Mode when power is restored.
Satellite XT FTT Versions 1-3
Basic Operation – Section 1
1.4 Operational Elements
Located at the front of the Satellite XT is
- the graphic display,
- the keypad consisting of six keys,
- the green status LED above the Set Key.
S a t e l l i t e X T
A s H 3
0.00
ppm
1.4.1 Keypad Functions
o Escape Key <esc> o Set Key <set> o 4 Cursor Keys marked with an arrow,
<up>, <down>, <left>, <right>.
Satellite XT esc
In the Maintenance, Calibration, and Configuration Menu, these keys provide the following functions:
To move forwards and backwards in the menu, use the cursor keys <down> and <up>. Select a function by actuating the <set> key.
To enter text or figures, use the cursor keys <left> and <right> to reach the desired cursor position. Use the cursor keys <up> and <down> to select the letter, figure, or sign you wish to enter. Bring the cursor to the next position, select the next letter, figure, or sign. After entering the complete text, use the <set> key to confirm the changes. If you do not want to save the changes, abort by actuating the <esc> key.
set
1.4.2 Graphic Display
Depending on the operating mode the structure of the graphic display is different. Three basic versions are illustrated and explained below. A detailed description is given in the according sections.
Display Structure for Monitoring Mode
Line 1
Line 2
Line 3
S a t e l l i t e X T
A s H 3
0 . 00 p p m
Line 1 offers the possibility to enter a user specific description, e.g. monitoring location
(max. 13 spaces; the name of the instrument is entered here when delivered). Line 2 shows the name of the gas monitored. Line 3 indicates the actual gas concentration and the concentration unit. Gas name and concentration unit are data stored in the memory of the sensor installed.
Display Structure for Menu
M E N U
MA I N T ENANCE s e l e c t : 2 o f 4
Line 1
Line 2
Line 3
M A I N T E N A N C E
S ENSOR S ERV I CE s e l e c t : 1 o f 5
Line 1 shows the actual position in the menu. In the left example the actual position is
Main Menu, in the right example it is the submenu Maintenance. Line 2 shows one of the selection possibilities. Line 3 shows the number of the selection possibility presently displayed and indicates also the total of selection possibilities available in this menu.
! Note:
The number of selection possibilities is subject to the instrument version and may be different than indicated in this guide to operation.
Satellite XT FTT Versions 1-4
Basic Operation – Section 1
1.5 Technical Specifications
Satellite XT Version
Part Number
FTT
9602-0400
FTT/R
9602-0405
FTT/C
9602-0450
Sensor Type used
electrochemical
catalytic
Power Requirements
voltage consumption
Network
data transmission
wiring topologies
Wiring
Network
4-wire cable 2x2x1.0mm2, shielded / 17 AWG
(Instrument delivered with 2m of cable attached)
relay contacts
6-wire cable 6x2x0.25mm2, shielded / 23 AWG
(Instrument delivered with 3m of cable attached)
Relay Outputs
contacts 3 x SPST (Single-Pole Single-Throw) max. ratings 250 VAC / 30 VDC, 2A
Graphic Display
Status Indication
Keypad
Physical Dimensions
size (LxWxH)
X
12 ... 24 VDC (18 VDC minimum when using the
Extractive Module) max. 0.6 W max. 1,4 W max. 0,9 W
Standardized LonTalkTM protocol
78 kBit per second free, e.g. bus, star, loop, or mixed
X
X
X
X
X
X
X
122 x 32 dots with backlight
LED green
6 touch-sensitive membrane function keys weight
Mounting
Housing Protection Class
RFI / EMC
Operating Conditions
temperature pressure humidity
145 x 95 x 50 mm
5.7 x 3.7 x 2.0 inches
480 g
17 oz
DIN-rail mounting
650 g
23 oz
IP 52 Option: IP 65
EN 50270
520 g
18 oz
-20 °C ... +40 °C -4 °F ... +104 °F
700 ... 1300 hPa
20 ... 90 % r.h.
This section explains installation and initial start-up procedures.
Satellite XT FTT Versions 1-5
Installation and Start-Up - Section 2
2.1 General Guidelines
Please consider the aspects indicated below before locating the instrument.
The property of the target gas (lighter or heavier than air) must be considered. The instrument should be mounted as close to the monitoring location as possible and should be easily accessible for operation and maintenance. For narrow monitoring locations or duct mounting a sensor extension with a standard length of 2m (6 feet) is available as an
option.
The optimum position to mount the instrument is in an upright position on a flat surface. If the instrument is used for ambient air monitoring, ensure that the sensor is facing downwards. For personal monitoring applications the instrument should be mounted at head level.
Ensure that the area immediately surrounding the instrument is free from objects that might hinder free flow of air. The instrument should be installed away from any possible liquid sources, extreme dust and dirt sources and should be protected against rain and sunlight. An instrument version protection class IP 65 is available for outdoor applications.
A junction box is recommended to connect the instrument to external control systems.
A distance of 2 m (6 feet) must not be exceeded.
Regarding the power requirements and operating conditions, please follow the specifications indicated in Section 1, Technical Specifications, for wiring instructions refer to Section 2, Wiring.
Sensors should be stored in a cool, dry place when not in use.
2.2 Standard Mounting
The instrument is designed for DIN–rail mounting. The mounting rail standard shipped with each instrument is used for wall mounting. Mount the DIN–rail to a wall and fasten the instrument with the sensor downwards and the display facing the front.
Attach the wires to the junction box according to the wiring diagram provided in the course of this section..
Mounting rail standard (P/N 9602.0050.10.03)
90,0 7,5
Standard mounting rail (P/N 9602.0050.10.03)
Satellite XT FTT Versions 2-1
Installation and Start-Up - Section 2
For applications where wall mounting is not possible, an L-shaped mounting plate with DIN–rail is available as an option.
L-shaped mounting plate with DIN–rail (P/N 9602.0051.10.02)
! Note:
Do not apply power until system is ready for start-up!
Satellite XT FTT Versions 2-2
2.3
2.3.1
Installation and Start-Up - Section 2
Duct Mounting
Observe the guidelines for installation of the instrument. For assembly and installation of the sensor extension see the schematics below and overleaf.
Connecting the Sensor Extension to the Instrument
1. Connect the Socket (A) to the plug in the sensor compartment located at the bottom of the Satellite XT. The groove of the Socket must face to the front.
2. Screw in the Internal Cap (B) and fasten it hand-tight.
3. Insert the Sensor Housing (C) with the nose of the Bayonet Catch facing to the front.
Lock the Bayonet Catch with a quarter-turn right.
A
B
1
C
2
3
D
A Socket with Groove
B Internal Cap
C Sensor Housing with Bayonet Catch
D Extension Cable to the Sensor
Satellite XT FTT Versions 2-3
Installation and Start-Up - Section 2
2.3.2
Duct
Mounting the Sensor Extension to the Duct
5,2cm
3,3cm
*) Duct Mounting Saddle Assembly various types available, see spare parts list, Section 8, Reference
Information
2,6cm 2,6cm
Duct
99
02
.00
80
.40
Du
Se ali ng
.00
S ilic ct
Mo un tin on g S ad dle
*)
9902.0005.40.01
T hre ad ed
Jo int
Sensor d
**) m n S tandar
R
O-Rin g 24x1.5m
1
Mi lle d N ut
9902.0004.40.0
9602
.0085
**)
F or
.10
Sa
9602
.00
Adap
XT
C tel lite
.0084
ter
.10.00
so
-V ers
Adap r E xtensio
Sen ion s: ter
Sensor
E xt.
C om bu sti ble
Satellite XT
Sensor Extension
Satellite XT
Satellite XT FTT Versions 2-4
Installation and Start-Up - Section 2
2.4 Network Wiring
For network wiring the 4-wire shielded cable extending from the bottom of the instrument is used. This cable has a length of 2 meters.
The 4-wire shielded cable is required to power the instrument and to interface with additional control devices. The open end of the cable should be terminated in a junction box.
To interface with additional control devices, a 4-wire shielded cable, 2x2x1.0 mm2 / 17 AWG, is required.
Junction Box Terminals
N+ white Network +
N– brown
P+ green
Network –
Power +
P– yellow
S shield
Power – cable shield
109,5mm
Junction Box 2 nodes, standard
Satellite XT FTT Versions 2-5
Installation and Start-Up - Section 2
2.5 Relay Wiring
Instruments with relay option have an additional
6–wire shielded cable extending from the bottom of the instrument. This cable has a length of 3 meters and is also permanently attached to the instrument.
The 6-wire shielded cable is used for relay wiring and provides a pair of connections for each of the 3 internal relays for activation of visual or audible external alarm devices.
Relay Contacts
1 white
1 brown
2 green
2 yellow
3 grey
3 pink
S shield
Alarm 1 Relay
Alarm 1 Relay
Alarm 2 Relay
Alarm 2 Relay
Fault Relay
Fault Relay cable shield
165
120
P+ P+ PPN+ N+ NNS S D D 1/A1 1/A1 2/A2 2/A2 3 / F 3 / F
P+ P+ PPN+ N+ NNS S D D 1/A1 1/A1 2/A2 2/A2 3 / F 3 / F
180
Junction Box 1 node, R-version for instruments with relay option.
! Note:
All wiring for the Satellite XT and its related equipment must be in compliance with any local electrical and fire codes. The Satellite XT wiring must be kept away from any high power lines. All shielding for the wiring must be connected together and grounded at only one point in the system.
Satellite XT FTT Versions 2-6
Installation and Start-Up - Section 2
Satellite XT FTT
Only Satellite XT with relay option
Junction Box
6-wire shielded cable
3 m provided with instrument
4-wire shielded cable
2 m provided wíth instrument
Junction Box
230 / 115 VAC
Power Supply
24 VDC
2-wire shielded cable
4-wire shielded cable to additional control devices
Wiring Diagram Satellite XT FTT and FTT/R - Basic Configuration
Satellite XT FTT/R
Service Center
Touch
Screen
Information
Panel
Satellite XT FTT
Modem
Power
Supply
PROFI / L2
MOD, CAN
Visualization
Software
Ethernet
Junction Box
Relay
Output
Module
Wiring Diagram Satellite XT - Typical Configuration
Network
Router /
Repeater to further segments
Satellite XT FTT Versions 2-7
Installation and Start-Up - Section 2
2.6 Initial Start-Up
When wiring is completed, a sensor must be assigned to each instrument. Every sensor is gas calibrated and these sensor specific calibration parameters are stored in its integrated data memory. Make sure to use only sensors designed for use with the Satellite XT.
Ex factory the instruments are shipped either unconfigured or already preconfigured according to the customer's specifications.
If the instruments are preconfigured, a certain sensor, which can be identified by its serial number printed on the packaging and the sensor label, is assigned and documented to one particular instrument. These data are stated in the test certificate provided with the
shipment.
Switch on power supply. If no sensor is fitted, the display of the instrument shows the below message.
- - - - F AU L T - - - -
NO S ENSOR !
For applications without sensor extension open the bayonet catch of the sensor housing located at bottom of the instrument with a quarter-turn left and remove the sensor housing.
For applications with sensor extension hold the plastic fitting at the duct side with two fingers and unscrew the metal fitting.
Take the assigned sensor out of the packaging. Remove shorting strap or wire link placed in the sensor plug before inserting it into the instrument. Not all sensors types require this strap or link.
Caution: Failure to remove the shorting strap or wire link may result in damage to the Satellite.
For applications without sensor extension insert the sensor into the instrument's sensor compartment with the arrow printed on the sensor label facing upwards and to the front. Fit sensor housing by fastening the bayonet catch with a quarter-turn right.
For applications with sensor extension insert the sensor into the socket inside the metal fitting, properly aligning notch, groove, and arrow printed on the sensor label.
A sensor warm-up is performed and the graphic display shows the according screen until the displayed value is zero. This does not apply for oxygen sensors.
The warm-up time required depends on the type of sensor installed. The instrument is in the Maintenance Mode and the green LED is off.
S ENSOR WARM - UP !
1 . 23 p p m
S a t e l l i t e X T
A s H 3
0 . 00 p p m
As soon as the displayed value reads zero, the instrument automatically switches to the
Monitoring Mode. The graphic display shows the according screen, the green status LED is on, the instrument is monitoring.
! Note:
If the instrument does not perform as described above and a fault message is shown, please refer to Section 7, Troubleshooting.
Instruments with catalytic sensors:
Zero Adjustment must be performed prior to start–up. It is recommended to repeat
Zero Adjustment every 4 to 6 weeks. For details refer to Section 5, Calibration.
Instruments with oxygen sensors:
Span Adjustment must be performed prior to start-up. It is recommended to repeat
Span Adjustment every 4 to 6 weeks. For details refer to Section 5, Calibration.
Satellite XT FTT Versions 2-8
3.1
Main Menu - Section 3
This section explains the different monitoring conditions and the submenus available under the Main Menu.
The Main Menu Screens
When the <esc> key is actuated, the instrument will leave the Monitoring Mode and go to the Main Menu. The instrument is now in the Maintenance Mode and the green LED is off.
Actuate the cursor keys <up> or <down> to move forwards and backwards in the Main
Menu. Actuate the <set> key to select a submenu. Actuate the <esc> key to go back to the
Main Menu.
To return to the Monitoring Mode, go to the menu 'MONITORING' and actuate <set> to select. The display shows the monitoring screen again and the illuminated green LED indicates, that the instrument is in the Monitoring Mode.
M E N U
MON I TOR I NG s e l e c t : 1 o f 4
M E N U
MA I N T ENANCE s e l e c t : 2 o f 4
M E N U
CA L I BRA T I ON s e l e c t : 3 o f 4
M E N U
CON F I GURA T I ON s e l e c t : 4 o f 4
Satellite XT FTT Versions
Monitoring Mode
Submenus
Submenus
Submenus
3-1
M E N U
MON I TOR I NG s e l e c t : 1 o f 4
Main Menu - Section 3
Main Menu - Monitoring
The instrument is in the operating mode
Maintenance and the green LED is off.
Actuate the cursor keys <up> or <down> to move forwards and backwards in the Main Menu.
Actuate the <set> key to enter the Monitoring Mode.
S a t e l l i t e X T
A s H 3
0 . 00 p p m
<Normal Condition>
This screen and the illuminated green LED indicate that the instrument is in the Monitoring Mode and is operating properly.
S a t e l l i t e X T
A s H 3
0 . 07
A L 1 p p m
<Alarm 1 Condition>
This screen shows that Alarm Level 1 has been exceeded and that there is an actual gas concentration of 0.07 ppm AsH3.
(f.i. Alarm 1 set at 0.05 ppm AsH3)
S a t e l l i t e X T
A s H 3
0 . 12
A L 2
A L 1 p p m
<Alarm 2 Condition>
This screen shows that Alarm Level 2 has been exceeded and that there is an actual gas concentration of 0.12 ppm AsH3.
(f.i. Alarm 2 set at 0.10 ppm AsH3)
3 - 2
Satellite XT FTT Versions 3-2
M E N U
MA I N T ENANCE s e l e c t : 2 o f 4
M A I N T E N A N C E
S ENSOR S ERV I CE s e l e c t : 1 o f 7
Main Menu - Section 3
Main Menu - Maintenance
The instrument is in the Maintenance Mode and the green LED is off.
Actuate the cursor keys <up> or <down> to move forwards and backwards in the Main Menu.
To select a submenu, actuate the <set> key to enter.
Submenus - Maintenance
Sensor Service screen 1 of 7
M A I N T E N A N C E
A L ARM RE S E T s e l e c t : 2 o f 7
M A I N T E N A N C E
A L ARM / WARN T E S T s e l e c t : 3 o f 7
M A I N T E N A N C E
DE V I CE I N FOS s e l e c t : 4 o f 7
M A I N T E N A N C E
S ENSOR I N FOS s e l e c t : 5 o f 7
Alarm Reset screen 2 of 7
Alarm / Warn Test screen 3 of 7
Device Infos screen 4 of 7
Sensor Infos screen 5 of 7
M A I N T E N A N C E
RE S E T DE V I CE s e l e c t : 6 o f 7
M A I N T E N A N C E
S ERV I CE s e l e c t : 7 o f 7
Satellite XT FTT Versions
Reset Device screen 6 of 7
Service screen 7 of 7
3-3
Main Menu - Section 3
M E N U
CA L I BRA T I ON s e l e c t : 3 o f 4
C A L I B R A T I O N
GA S CA L I BRA T I ON s e l e c t : 1 o f 2
Main Menu - Calibration
The instrument is in the Maintenance Mode and the green LED is off.
Actuate the cursor keys <up> or <down> to move forwards and backwards in the Main Menu.
To select a submenu, actuate the <set> key to enter.
Submenus - Calibration
Gas Calibration screen 1 of 2
C A L I B R A T I O N
MANUA L K - F AC TOR s e l e c t : 2 o f 2
Manual K-Factor screen 2 of 2
3 - 4
Satellite XT FTT Versions 3-4
M E N U
CON F I GURA T I ON s e l e c t : 4 o f 4
C O N F I G U R A T I O N
A L ARM S E T T I NGS s e l e c t : 1 o f 1 0
Main Menu - Section 3
Main Menu - Configuration
The instrument is in the Maintenance Mode and the green LED is off.
Actuate the cursor keys <up> or <down> to move forwards and backwards in the Main Menu.
To select a submenu, actuate the <set> key to enter.
Submenus - Configuration
Alarm Settings screen 1 of 10
C O N F I G U R A T I O N
L ANGUAGE s e l e c t : 2 o f 1 0
Language screen 2 of 10
C O N F I G U R A T I O N
DA T E FORMA T s e l e c t : 3 o f 1 0
Date Format screen 3 of 10
C O N F I G U R A T I O N
AU TO S E L F T E S T s e l e c t : 4 o f 1 0
C O N F I G U R A T I O N
S ECUR I T Y s e l e c t : 5 o f 1 0
Auto Selftest screen 4 of 10
! This function is not available with instruments using oxygen sensors or catalytic sensors.
Security screen 5 of 10
C O N F I G U R A T I O N
P A S SWORD s e l e c t : 6 o f 1 0
Password screen 6 of 10
continued
Satellite XT FTT Versions 3-5
continued
Main Menu - Section 3
Main Menu - Configuration
Submenus - Configuration
C O N F I G U R A T I O N
L OCA T I ON s e l e c t : 7 o f 1 0
C O N F I G U R A T I O N
NEW S ENSOR T Y P E s e l e c t : 8 o f 1 0
C O N F I G U R A T I O N
GA S NAME s e l e c t : 9 o f 1 0
Location screen 7 of 10
New Sensor Type screen 8 of 10
Gas Name screen 9 of 10
C O N F I G U R A T I O N
RE L A Y S s e l e c t : 1 0 o f 1 0
Relays screen 10 of 10
! This submenu is only available for instruments with relay option.
Satellite XT FTT Versions 3-6
Maintenance - Section 4
M E N U
This section explains routine maintenance procedures including sensor replacement and specific information on sensor and instrument.
MA I N T ENANCE s e l e c t : 2 o f 4
Main Menu - Maintenance
The instrument is in the Maintenance Mode and the green LED is off.
Actuate the cursor keys <up> or <down> to move forwards and backwards in the Main Menu.
To select a submenu, actuate the <set> key to enter.
Submenus - Maintenance
M A I N T E N A N C E
S ENSOR S ERV I CE s e l e c t : 1 o f 7
Sensor Service screen 1 of 7
M A I N T E N A N C E
A L ARM RE S E T s e l e c t : 2 o f 7
M A I N T E N A N C E
A L ARM / WARN T E S T s e l e c t : 3 o f 7
M A I N T E N A N C E
DE V I CE I N FOS s e l e c t : 4 o f 7
M A I N T E N A N C E
S ENSOR I N FOS s e l e c t : 5 o f 7
Alarm Reset screen 2 of 7
Alarm / Warn Test screen 3 of 7
Device Infos screen 4 of 7
Sensor Infos screen 5 of 7
continued
Satellite XT FTT Versions 4-1
continued
M A I N T E N A N C E
RE S E T DE V I CE s e l e c t : 6 o f 7
M A I N T E N A N C E
S ERV I CE s e l e c t : 7 o f 7
Maintenance - Section 4
Main Menu - Maintenance
Submenus - Maintenance
Reset Device screen 6 of 7
Service screen 7 of 7
Satellite XT FTT Versions 4-2
Maintenance - Section 4
Sensor Service
To perform a sensor replacement with a new sensor, which must have the same part number. See section 8 for sensor information.
Use the <set> key to enter and follow the dialogue displayed at the bottom line of the screen.
When the instrument has finished the sensor service procedure, it will automatically enter the Monitoring
Mode.
Remove Sensor
Remove the sensor presently installed and actuate the
<set> key to proceed.
Replacement
Insert the new sensor and actuate the <set> key to continue.
Load New Data ?
If you want the data of the new sensor to be loaded, confirm this by actuating the <set> key.
Loading Data !
The new data are now loaded from the sensor into the unit's internal memory.
<Sensor Warm-Up Condition>
A sensor warm-up is performed and this screen is shown until the displayed value is zero.
This does not apply for oxygen sensors.
The warm-up time required depends on the type of sensor.
<Normal Monitoring Condition>
The instrument will automatically switch to the
Monitoring Mode when the sensor service procedure is finished.
Only instruments with catalytic sensors:
If the displayed value is not zero, a zero adjustment must be performed; refer to section 5, Calibration.
!Note:
When a new sensor is installed the K-factor will automatically revert to the default value 1.00.
If individual settings are required, they must be entered again; refer to section 5, Calibration.
Catalytic sensors are calibrated for methane. When monitoring for other combustible gases, a K-factor must be entered.
Satellite XT FTT Versions 4-3
M A I N T E N A N C E
A L ARM RE S E T s e l e c t : 2 o f 7
A L A R M R E S E T
A L ARMS OF F ?
< s e t > t o c o n f i r m
M A I N T E N A N C E
A L ARM / WARN T E S T s e l e c t : 3 o f 7
Maintenance - Section 4
Alarm Reset
Alarm Reset allows the user to reset the alarm indication of latched alarms. An according message is sent to external alarm devices.
Use the <set> key to enter and follow the dialogue displayed at the bottom of the screen. Actuate the
<esc> key to exit.
Alarms Off ?
Actuate the <set> key to cofirm the reset of alarm indication with latching alarm.
For non-latching alarms, alarm indication will be reset automatically..
Alarm/Warn Test
Used to verify the alarm function of any associated external device by simulating an Alarm 1, Alarm 2, and warning condition.
Use the <set> key to select or use the cursor keys
<up> and <down> to continue in the menu.
A L A R M / WA R N T E S T
S E T A L ARM 1 ?
s e l e c t : 1 o f 3
Set Alarm 1 ?
Allows the user to simulate an Alarm 1 condition.
Actuate the <set> key to enter.
S E T A L A R M 1 ?
A l a r m 1 s e t !
< e s c > t o e x i t !
Alarm 1 set !
An Alarm 1 message is transmitted to the network.
Instruments with relay option: the according relay is activated. Use the <esc> key to exit.
A L A R M / WA R N T E S T
S E T A L ARM 2 ?
s e l e c t : 2 o f 3
Set Alarm 2 ?
Allows the user to simulate an Alarm 2 condition.
Actuate the <set> key to enter.
S E T A L A R M 2 ?
A l a r m 2 s e t !
< e s c > t o e x i t !
continued
Satellite XT FTT Versions
Alarm 2 set !
An Alarm 2 message is transmitted to the network.
Instruments with relay option: the according relay is activated. Use the <esc> key to exit.
4-4
Maintenance - Section 4
Alarm/Warn Test
continued
A L A R M / WA R N T E S T
S E T WARN I NG ?
s e l e c t : 3 o f 3
Set Warning ?
Allows the user to simulate a warning condition.
Actuate the <set> key to enter.
S E T WA R N I N G ?
Wa r n i n g s e t !
< e s c > t o e x i t !
Warning set !
A warning message is transmitted to the network.
Instruments with relay option: the according relay is activated. Use the <esc> key to exit.
M A I N T E N A N C E
DE V I CE I N FOS s e l e c t : 4 o f 7
Device Infos
Used to obtain instrument specific information, i.e. software version and ID number.
In general these informations are required for service purposes. Actuate the <set> key to select and move with the cursor keys <up> and <down>. Actuate the
<esc> key to exit.
D E V I C E I N F O S
SW : S X T
_
X x . x x s h ow i n g : 1 o f 2
<Actual Software Version>
The version of the software presently installed is shown.
D E V I C E I N F O S
I D : 0 0 0 1 2 3 4 5 6 7 8 9 s h ow i n g : 2 o f 2
<Specific Identification Number>
The instrument’s specific ID identification number is shown.
Satellite XT FTT Versions 4-5
M A I N T E N A N C E
S ENSOR I N FOS s e l e c t : 5 o f 7
Maintenance - Section 4
Sensor Infos
Provides specific information about the sensor presently installed, i.e. part number, serial number, date of first calibration, sensitivity, and revision number.
These data are stored in the sensor’s memory.
Actuate the <set> key to select or use the cursor keys
<up> and <down> to continue in the menu.
S E N S O R I N F O S
P AR T NUMB ER s e l e c t : 1 o f 5
Part Number
Actuate the <set> key to display the part number of the sensor presently installed.
P A R T N U M B E R
9 6 0 2 - 6 0 0 0
< e s c > t o e x i t !
<Actual Sensor Part Number>
Use this information to order replacement sensors.
Additional order information is provided in Section 8,
Sensor Order Information.
Actuate <esc> to leave.
S E N S O R I N F O S
S ER I A L NUMB ER s e l e c t : 2 o f 5
Serial Number
Actuate the <set> key to display the serial number of the sensor presently installed.
S E R I A L N U M B E R
3 1
< e s c > t o e x i t !
<Actual Sensor Serial Number>
This information may be required for service purposes.
Actuate the <esc> key to leave.
S E N S O R I N F O S
1 s t CA L I BRA T I ON s e l e c t : 3 o f 5
First Calibration
Use the <set> key to display the date when the installed sensor was calibrated for the first time.
1 s t C A L I B R A T I O N
2 7 . 0 9 . 2 0 0 6
< e s c > t o e x i t !
continued
Satellite XT FTT Versions
<Date of First Calibration>
This information may be required for service purposes and to check the age of the sensor.
Actuate <esc> to leave.
4-6
Maintenance - Section 4
Sensor Infos
continued
S E N S O R I N F O S
S ENS I T I V I T Y s e l e c t : 4 o f 5
Sensitivity
Use the <set> key to display the sensor sensitivity determined during the first calibration.
S E N S I T I V I T Y
1 2 3 n A / p pm
< e s c > t o e x i t !
<Actual Sensitivity>
This information may be required for service purposes.
Actuate the <esc> key to leave.
S E N S O R I N F O S
RE V I S I ON NUMB ER s e l e c t : 5 o f 5
Revision Number
Use the <set> key to display the revision number of the gas related information stored in the sensor’s memory.
R E V I S I O N N U M B E R
0
< e s c > t o e x i t !
<Actual Revision Number>
This information may be required for service purposes.
Actuate the <esc> key to leave.
Satellite XT FTT Versions 4-7
M A I N T E N A N C E
RE S E T DE V I CE s e l e c t : 6 o f 7
Maintenance - Section 4
Reset Device
Offers the possibility to perform a “warm start” of the instrument. Actuate the <set> key to select.
R E S E T D E V I C E
ARE YOU SURE ?
< e s c > t o e x i t !
Are You Sure ?
This screen is to confirm that a software reset should be performed. Actuate the <set> key to confirm or use the <esc> key to exit.
S ENSOR WARM - UP !
1 . 2 3 p p m
<Sensor Warm-Up Condition>
A sensor warm-up is performed and this screen is shown until the displayed value is zero, except for
Oxygen sensors. The warm-up time depends on the type of sensor.
S a t e l l i t e X T
A s H 3
0 . 0 0 p p m
<Normal Monitoring Condition>
The instrument will automatically switch to the
Monitoring Mode when the sensor service procedure is finished.
M A I N T E N A N C E
S ERV I CE s e l e c t : 7 o f 7
Service
This submenu is used exclusively by trained Service
Personnel. The functions are password protected.
S E R V I C E
P A S SWORD : *** p l e a s e e n t e r !
Satellite XT FTT Versions 4-8
Calibration - Section 5
This section describes calibration procedures for the Satellite XT.
Calibration can be performed either automatically by a dynamic gas calibration or manually by entering a calculated correction factor called K-factor.
M E N U
CA L I BRA T I ON s e l e c t : 3 o f 4
Main Menu - Calibration
The instrument is in the Maintenance Mode and the green LED is off.
Actuate the cursor keys <up> or <down> to move forwards and backwards in the Main Menu.
To select a submenu, actuate the <set> key to enter.
G A S C A L I B R A T I O N
GA S CA L I BRA T I ON s e l e c t : 1 o f 2
Submenus - Calibration
Gas Calibration screen 1 of 2
G A S C A L I B R A T I O N
MANUA L K - F AC TOR s e l e c t : 2 o f 2
Manual K-Factor screen 2 of 2
!Note:
If a high degree of accuracy in monitoring is required, monthly calibration with calibration gas of a known concentration is recommended.
When performing a dynamic calibration, a new correction factor (K-Factor) is calculated automatically. The actual value can be displayed in the submenu Manual K-Factor.
Use appropriate safety precautions when handling toxic or corrosive gases and properly vent, if possible.
Instruments with CATALYTIC sensors:
Zero Adjustment must be performed prior to start–up.
It is recommended to repeat Zero Adjustment every 4 to 6 weeks.
Catalytic sensors must be protected from silicone vapours, which may permanently reduce the sensor’s sensitivity.
Instruments with OXYGEN sensors:
Span Adjustment must be performed prior to start-up.
It is recommended to repeat Span Adjustment every 4 to 6 weeks.
Satellite XT FTT Versions 5-1
Calibration - Section 5
C A L I B R A T I O N
GA S CA L I BRA T I ON s e l e c t : 1 o f 2
G A S C A L I B R A T I O N
Z ERO AD J US T s e l e c t : 1 o f 2
G A S C A L I B R A T I O N
A P P L Y A I R !
u s e c l e a n a i r !
Gas Calibration
The gas calibration submenu is used for dynamic calibration. For zero adjustment use clean or synthetic air free from measuring gas or any other interfering gases. The sensor must be warmed-up before calibration.
For span adjustment it is recommended to use test gas of a known concentration at or slightly above TLV for the target gas.
Always keep to the correct sequence, first perform zero adjustment, then Span
Adjustment.
Zero Adjust
Zero adjustment applies for all sensors except oxygen sensors.
Use the <set> key to enter and follow the dialogue displayed.
Apply Air !
Apply clean or synthetic air to the sensor and actuate the <set> key to continue.
G A S C A L I B R A T I O N
0 . 0 1 p pm
V a l u e s t a b l e ?
Allow the zero reading to stabilize.
<Actual Zero Point>
If the reading is already zero, zero adjustment is not necessary. Actuate the <esc> key to exit.
If the reading is not zero, actuate the <set> key to re-adjust the zero point and to continue the procedure.
G A S C A L I B R A T I O N
0 . 0 0 p pm
V a l u e o . k . ?
<Re-adjusted Zero Point>
The reading is zero. Actuate the <set> key to proceed.
G A S C A L I B R A T I O N
S A V E ?
< s e t > t o c o n f i r m
Save ?
Actuate the <set> key to save the new zero point and store it in the unit's internal memory. continued
Satellite XT FTT Versions 5-2
Calibration - Section 5
Gas Calibration
continued
G A S C A L I B R A T I O N
S P AN AD J US T s e l e c t : 2 o f 2
Span Adjust
Span adjustment applies for all sensors, including oxygen sensors.
Use the <set> key to enter and follow the dialogue displayed.
S P A N A D J U S T
A s H 3 i s c a l i b r a t i o n g a s !
<Actual Calibration Gas>
The instrument shows which calibration gas is required for the actual sensor type.
S P A N A D J U S T
0 . 0 5 p pm e n t e r c o n c e n t r a t i o n
S P A N A D J U S T
A P P L Y GA S !
u s e c a l i b r a t i o n g a s !
S P A N A D J U S T
0 . 0 4 p pm
V a l u e s t a b l e ?
<Actual Concentration for Calibration>
The recommended concentration for span adjustment is displayed. Default value is TLV concentration for the target gas, 20 % LEL for catalytic sensors. When using calibration gas of a different concentration, change the value accordingly. After the new value has been entered, actuate the <set> key to continue.
Apply Gas !
Place the calibration cap over the sensor and start the calibration gas flow (18 l/h or 300 ccm/min).
Actuate the <set> key to continue.
<Actual Span Value>
Allow the reading to stabilize. If the reading matches the concentration of the calibration gas, span adjustment is not necessary. Actuate the <esc> key to exit.
If the reading does not match the concentration of the calibration gas, actuate the <set> key to re-adjust the span value and to continue the procedure.
S P A N A D J U S T
0 . 0 5 p pm
V a l u e o . k . ?
<Re-adjusted Span Value>
The reading matches the concentration of the calibration gas. Actuate the <set> key to proceed.
S P A N A D J U S T
S A V E ?
< s e t > t o c o n f i r m
Save ?
Actuate the <set> key to save the new span value and store it in the unit's internal memory.
Remove Gas !
Stop the calibration gas flow. Remove gas and calibration cap. Allow several minutes for the sensor to clear.
S P A N A D J U S T
REMOV E GA S !
V a l u e s a v e d !
Satellite XT FTT Versions 5-3
C A L I B R A T I O N
MANUA L K - F AC TOR s e l e c t : 2 o f 3
M A N U A L K - F A C T O R
1 . 0 0 p l e a s e e n t e r !
Calibration - Section 5
Manual K-Factor
This submenu is used to manually calibrate the instrument by entering a new K-Factor. This factor is a multiplication or correction value used to calibrate the response of the instrument to a specific gas concentration.
The factory default K-Factor for all gases is 1.00.
Use the <set> key to enter. The instrument will display the actual K-Factor.
<Actual K-Factor>
Use the cursor keys <left> or <right> to reach the desired position. Use the cursor keys <up> and
<down> to select the figures you wish to enter. The K-
Factor must be in the range of 0.20 and 5.00.
After a new value has been entered, actuate the <set> key to confirm or <esc> to exit.
!Note:
Individual settings for the Manual K-Factor will automatically revert to the default value 1.00, whenever the sensor is replaced by a new sensor.
Only instruments using catalytic sensors:
Catalytic sensors are calibrated for methane. When monitoring for other combustible gases, a
K-factor must be entered.
Satellite XT FTT Versions 5-4
Configuration - Section 6
M E N U
This section describes the instrument's default configuration and procedures how to adapt it to individual requirements.
CON F I GURA T I ON s e l e c t : 4 o f 4
Main Menu - Configuration
The instrument is in the Maintenance Mode and the green LED is off.
Actuate the cursor keys <up> or <down> to move forwards and backwards in the Main Menu.
To select a submenu, actuate the <set> key to enter.
Submenus - Configuration
C O N F I G U R A T I O N
A L ARM S E T T I NGS s e l e c t : 1 o f 1 0
Alarm Settings screen 1 of 10
C O N F I G U R A T I O N
L ANGUAGE s e l e c t : 2 o f 1 0
Language screen 2 of 10
C O N F I G U R A T I O N
DA T E FORMA T s e l e c t : 3 o f 1 0
Date Format screen 3 of 10
C O N F I G U R A T I O N
AU TO S E L F T E S T s e l e c t : 4 o f 1 0
Auto Selftest screen 4 of 10
! This function is not available with instruments using oxygen sensors or catalytic sensors.
C O N F I G U R A T I O N
S ECUR I T Y s e l e c t : 5 o f 1 0
Security screen 5 of 10
continued
Satellite XT FTT Versions 6-1
continued
Configuration - Section 6
Main Menu - Configuration
Submenus - Configuration
Password screen 6 of 10
C O N F I G U R A T I O N
P A S SWORD s e l e c t : 6 o f 1 0
C O N F I G U R A T I O N
L OCA T I ON s e l e c t : 7 o f 1 0
Location screen 7 of 10
C O N F I G U R A T I O N
NEW S ENSOR T Y P E s e l e c t : 8 o f 1 0
New Sensor Type screen 8 of 10
C O N F I G U R A T I O N
GA S NAME s e l e c t : 9 o f 1 0
Gas Name screen 9 of 10
C O N F I G U R A T I O N
RE L A Y S s e l e c t : 1 0 o f 1 0
Relays screen 10 of 10
! This submenu is only available for instruments with relay option.
Satellite XT FTT Versions 6-2
C O N F I G U R A T I O N
A L ARM S E T T I NGS s e l e c t : 1 o f 1 0
Configuration - Section 6
Alarm Settings
Allows the user to enter or change various settings for
Alarm 1 and Alarm 2 activation.
Use the <set> key to enter.
A L A R M S E T T I N G S
A L ARM 1 s e l e c t : 1 o f 2
Alarm 1
For Alarm 1 settings, use the <set> key to enter or continue in the menu with the cursor keys <up> or
<down>.
A L A R M S E T T I N G S
A L ARM 2 s e l e c t : 2 o f 2
Alarm 2
For Alarm 2 settings, use the <set> key to enter or continue in the menu with the cursor keys <up> or
<down>.
Satellite XT FTT Versions 6-3
A L A R M S E T T I N G S
A L ARM 1 s e l e c t : 1 o f 2
Configuration - Section 6
Alarm 1
Allows the user to configure all parameters for Alarm 1
(lower alarm level) activation.
A L A R M 1
A L ARM 1 S T A T E s e l e c t : 1 o f 4
Alarm 1 State
Used to enable / disable the Alarm 1 activation.
A L A R M 1
A L ARM 1 L E V E L s e l e c t : 2 o f 4
Alarm 1 Level
Used to set individual alarm levels for Alarm 1 activation.
A L A R M 1
A L ARM 1 T R I GGER s e l e c t : 3 o f 4
Alarm 1 Trigger
Used to define the trigger condition for Alarm 1 activation.
A L A R M 1
A L ARM 1 L A T CH s e l e c t : 4 o f 4
Alarm 1 Latch
Used to define the Alarm 1 activation to be latching or non-latching.
Satellite XT FTT Versions 6-4
A L A R M 1
A L ARM 1 S T A T E s e l e c t : 1 o f 4
Configuration - Section 6
Alarm 1 State
Used to enable / disable the Alarm 1 activation. Default setting is enabled. Use the <set> key to enter. The instrument will show the actual setting.
A L A R M 1 S T A T E
D I S A B L ED s e l e c t : 1 o f 2
Disabled
Allows the user to turn off the Alarm 1 activation.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
A L A R M 1 S T A T E
ENA B L ED s e l e c t : 2 o f 2
Enabled
Allows the user to turn on the Alarm 1 activation.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
A L A R M 1
A L ARM 1 L E V E L s e l e c t : 2 o f 4
Alarm 1 Level
Used to set individual alarm levels for Alarm 1 activation. Default setting is 1 x TLV respectively 20 %
LEL for the target gas.
Actuate the <set> key to enter and show the actual setting.
A L A R M 1 L E V E L
0 . 0 5 p pm p l e a s e e n t e r !
<Actual Alarm 1 Level>
To change the value, use the cursor keys <left> or
<right> to reach the desired position. Use the cursor keys <up> and <down> to select the figures you wish to enter. After the new value has been entered, actuate the
<set> key to confirm or <esc> to exit.
Satellite XT FTT Versions 6-5
A L A R M 1
A L ARM 1 T R I GGER s e l e c t : 3 o f 4
Configuration - Section 6
Alarm 1 Trigger
Used to define the trigger condition for Alarm 1 activation. Default setting is Greater Than. Use the
<set> key to enter. The instrument will show the actual setting.
A L A R M 1 T R I G G E R
GRE A T ER T HAN s e l e c t : 1 o f 2
Greater Than
Defines that an Alarm 1 condition will be indicated when the actual gas concentration exceeds the preset level for Alarm 1.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
A L A R M 1 T R I G G E R
L E S S T HAN s e l e c t : 2 o f 2
Less Than
Defines that an Alarm 1 condition will be indicated when the actual gas concentration falls below the preset level for Alarm 1.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit..
A L A R M 1
A L ARM 1 L A T CH s e l e c t : 4 o f 4
A L A R M 1 L A T C H
D I S A B L ED s e l e c t : 1 o f 2
Alarm 1 Latch
Used to define the Alarm 1 activation to be either latching or non-latching.
Default setting is enabled. Use the <set> key to enter.
The instrument will show the actual setting.
When latching is enabled, alarm reset must be done manually. Nonlatching alarms will be reset automatically, upon correction of the alarm condition.
Disabled
Defines the Alarm 1 activation to be non-latching.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
A L A R M 1 L A T C H
ENA B L ED s e l e c t : 2 o f 2
Enabled
Defines the Alarm 1 activation to be latching.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Satellite XT FTT Versions 6-6
A L A R M S E T T I N G S
A L ARM 2 s e l e c t : 2 o f 2
Configuration - Section 6
Alarm 2
Allows the user to configure all parameters for Alarm 2
(upper alarm level) activation.
For all subsequent Alarm 2 settings, please follow the procedures described in the corresponding section for
Alarm 1 settings.
A L A R M 2
A L ARM 2 S T A T E s e l e c t : 1 o f 4
Alarm 2 State
Used to enable / disable the Alarm 2 activation.
A L A R M 2
A L ARM 2 L E V E L s e l e c t : 2 o f 4
Alarm 2 Level
Used to set individual alarm levels for Alarm 2 activation.
A L A R M 2
A L ARM 2 T R I GGER s e l e c t : 3 o f 4
Alarm 2 Trigger
Used to define the trigger condition for Alarm 2 activation.
A L A R M 2
A L ARM 2 L A T CH s e l e c t : 4 o f 4
Alarm 2 Latch
Used to define the Alarm 2 activation to be latching or non-latching.
Satellite XT FTT Versions 6-7
C O N F I G U R A T I O N
L ANGUAGE s e l e c t : 2 o f 1 0
Configuration - Section 6
Language
Used to select the desired language for menu operation. Default setting is English.
Use the <set> key to enter. The instrument will display the actual language.
L A N G U A G E
GERMAN s e l e c t : 1 o f 2
German
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
L A N G U A G E
ENG L I SH s e l e c t : 2 o f 2
English
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
C O N F I G U R A T I O N
DA T E FORMA T s e l e c t : 3 o f 1 0
Date Format
Used to define the format how to display the date (only used in the submenu Sensor Infos, Date of First
Calibration). Default setting is International.
Use the <set> key to enter. The actual date format is displayed.
D A T E F O R M A T
I N T ERNA T I ONA L s e l e c t : 1 o f 2
International
International date format is DD.MM.YYYY
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
D A T E F O R M A T
US A s e l e c t : 2 o f 2
USA
USA date format is MM-DD-YYYY
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Satellite XT FTT Versions 6-8
Configuration - Section 6
C O N F I G U R A T I O N
AU TO S E L F T E S T s e l e c t : 4 o f 1 0
A U T O S E L F T E S T
D I S A B L ED s e l e c t : 1 o f 2
Auto Selftest
The instrument's self-diagnostics provides an on-line preventive sensor selftest performed automatically every 24 hours, which may be enabled or disabled.
Default setting is enabled. The Auto Selftest function is not applicable for Oxygen sensors. Use the <set> key to enter. The display will show the actual setting.
! The function Auto Selftest is not available with instruments using oxygen sensors or catalytic sensors.
Disabled
Allows the user to turn off the Auto Selftest.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
A U T O S E L F T E S T
ENA B L ED s e l e c t : 2 o f 2
Enabled
Allows the user to turn on the Auto Selftest.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
C O N F I G U R A T I O N
S ECUR I T Y s e l e c t : 5 o f 1 0
Security
Exiting the Monitoring Mode and entering the
Maintenance Mode, may be password protected, to prevent the instrument from manipulations by unauthorized personnel. Ex factory the password protection is disabled.
Use the <set> key to enter. The instruments shows the actual setting.
S E C U R I T Y
P A S SWORD ON s e l e c t : 1 o f 2
Password On
Allows the user to enable the password protection.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
S E C U R I T Y
P A S SWORD OF F s e l e c t : 2 o f 2
Password Off
Allows the user to disable the password protection.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit..
Satellite XT FTT Versions 6-9
C O N F I G U R A T I O N
P A S SWORD s e l e c t : 6 o f 1 0
Configuration - Section 6
Password
Allows the user to enter or change the password.
Ex factory the password is <000>.
Use the <set> key to enter. The instrument will display the actual password.
P A S SWO R D
0 0 0 p l e a s e e n t e r !
<Actual Password>
To change the password, use the cursor keys <left> or
<right> to reach the desired position. Use the cursor keys <up> and <down> to select the figures you wish to enter. After the new password has been entered, actuate the<set> key to confirm or <esc> to exit.
C O N F I G U R A T I O N
L OCA T I ON s e l e c t : 7 o f 1 0
L O C A T I O N
G a s C a b i n e t p l e a s e e n t e r !
Location
Allows the user to enter a desciption defining the monitoring point. Ex factory it reads <Satellite XT>. Up to 13 alphanumeric characters can be entered.
Use the <set> key to enter. The instrument will display the actual location..
<Actual Location>
Use the cursor keys <left> or <right> to reach the desired position. Use the cursor keys <up> and <down> to select the characters you wish to enter. When the new location has been entered completely, actuate the
<set> key to confirm or <esc> to exit.
Satellite XT FTT Versions 6-10
C O N F I G U R A T I O N
NEW S ENSOR T Y P E s e l e c t : 8 o f 1 0
Configuration - Section 6
New Sensor Type
Allows the user to configure the instrument for a new sensor type with a part number, which is different from the one presently used.
See Section 8 for Sensor Information.
Actuate the <set> key to enter and follow the dialogue displayed.
Remove Sensor
Remove the sensor presently installed and actuate the
<set> key to proceed.
N EW S E N S O R T Y P E
REMOV E S ENSOR
< s e t > wh e n r e a d y
N EW S E N S O R T Y P E
NEW S ENSOR p l e a s e i n s e r t !
New Sensor
Insert the new sensor and actuate the <set> key to proceed.
N EW S E N S O R T Y P E
L OAD NEW DA T A ?
< s e t > t o c o n f i r m
Load New Data ?
If you want the data of the new sensor to be loaded, confirm this by pressing the <set> key..
N EW S E N S O R T Y P E
L OAD I NG DA T A !
p l e a s e wa i t !
Loading Data !
The new data are now loaded from the sensor into the instrument's internal memory.
C O N F I G U R A T I O N
GA S NAME s e l e c t : 9 o f 1 0
G A S N A M E
A s H 3 p l e a s e e n t e r !
Gas Name
Allows the user to enter a different gas name than the one stored in the sensor's data memory. Up to 7 alpha-numeric characters can be entered.
Use the <set> key to enter. The instrument will display the actual gas name.
<Actual Gas Name>
Use the cursor keys <left> or <right> to reach the desired position. Use the cursor keys <up> and
<down> to select the characters you wish to enter.
When the new gas name has been entered completely, actuate the <set> key to confirm or <esc> to exit.
Satellite XT FTT Versions 6-11
C O N F I G U R A T I O N
RE L A Y S s e l e c t : 1 0 o f 1 0
Configuration - Section 6
Relays
Allows the user to enter or change various settings for the internal relay contacts for Alarm 1, Alarm 2, and
Fault. Use the <set> key to enter.
!The RELAYS menu and all subsequent menu items are only available for instruments with relay option.
R E L A Y S
RE L A Y S T A T E s e l e c t : 1 o f 3
Relay State
Used to individually define the relays to be either deenergized (normally open contact) or energized
(normally closed contact).
Use the <set> key to enter or continue in the menu with the cursor keys <up> or <down>.
R E L A Y S
A L ARM DE L A Y s e l e c t : 2 o f 3
R E L A Y S
F AU L T T R I GGER s e l e c t : 3 o f 3
Alarm Delay
Used to set a delay for the relay activation in case of concentration alarms.
Use the <set> key to enter or continue in the menu with the cursor keys <up> or <down>.
Fault Trigger
Used to define the trigger conditions for activation of the Fault Relay.
Use the <set> key to enter or continue in the menu with the cursor keys <up> or <down>.
Satellite XT FTT Versions 6-12
R E L A Y S
RE L A Y S T A T E s e l e c t : 1 o f 3
Configuration - Section 6
Relay State
Used to individually define the Alarm 1, Alarm 2, and
Fault Relay to be either deenergized (normally open contact) or energized (normally closed contact).
Ex factory the relays are deenergized (normally open contacts).
Use the <set> key to enter.
R E L A Y S T A T E
A L ARM 1 RE L A Y s e l e c t : 1 o f 3
Allows to set the Alarm 1 Relay state.
Alarm 1 Relay
R E L A Y S T A T E
A L ARM 2 RE L A Y s e l e c t : 2 o f 3
Allows to set the Alarm 2 Relay state.
Alarm 2 Relay
Fault Relay
R E L A Y S T A T E
F AU L T RE L A Y s e l e c t : 3 o f 3
Allows to set the Fault Relay state.
R E L A Y S T A T E
A L ARM 1 RE L A Y s e l e c t : 1 o f 3
Alarm 1 Relay
Allows the user to set the Alarm 1 Relay state.
Use the <set> key to enter.
The instrument will show the actual setting.
A L A R M 1 R E L A Y
DE ENERG I Z ED / NO s e l e c t : 1 o f 2
Deenergized / NO
Defines the Alarm 1 Relay to be deenergized (normally open contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Energized / NC
A L A R M 1 R E L A Y
ENERG I Z ED / NC s e l e c t : 2 o f 2
Satellite XT FTT Versions 6-13
R E L A Y S
RE L A Y S T A T E s e l e c t : 1 o f 3
Configuration - Section 6
Defines the Alarm 1 Relay to be energized (normally closed contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
R E L A Y S T A T E
A L ARM 1 RE L A Y s e l e c t : 1 o f 3
Allows to set the Fault Relay state.
Fault Relay
R E L A Y S T A T E
A L ARM 2 RE L A Y s e l e c t : 2 o f 3
Alarm 1 Relay
Allows the user to set the Alarm 1 Relay state.
Use the <set> key to enter.
The instrument will show the actual setting.
R E L A Y S T A T E
F AU L T RE L A Y s e l e c t : 3 o f 3
Deenergized / NO
Defines the Alarm 1 Relay to be deenergized (normally open contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
R E L A Y S T A T E
A L ARM 1 RE L A Y s e l e c t : 1 o f 3
A L A R M 1 R E L A Y
DE ENERG I Z ED / NO s e l e c t : 1 o f 2
Energized / NC
Defines the Alarm 1 Relay to be energized (normally closed contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Alarm 2 Relay
Allows the user to set the Alarm 2 Relay state.
Use the <set> key to enter.
The instrument will show the actual setting.
A L A R M 1 R E L A Y
ENERG I Z ED / NC s e l e c t : 2 o f 2
Deenergized / NO
Defines the Alarm 2 Relay to be deenergized (normally open contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit..
Satellite XT FTT Versions 6-14
R E L A Y S T A T E
A L ARM 2 RE L A Y s e l e c t : 2 o f 3
Configuration - Section 6
Alarm 2 Relay
Allows the user to set the Alarm 2 Relay state.
Use the <set> key to enter.
The instrument will show the actual setting.
A L A R M 2 R E L A Y
DE ENERG I Z ED / NO s e l e c t : 1 o f 2
Deenergized / NO
Defines the Alarm 2 Relay to be deenergized (normally open contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit..
A L A R M 2 R E L A Y
ENERG I Z ED / NC s e l e c t : 2 o f 2
Energized / NC
Defines the Alarm 2 Relay to be energized (normally closed contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
R E L A Y S T A T E
F AU L T RE L A Y s e l e c t : 3 o f 3
Allows the user to set the Fault Relay state.
Use the <set> key to enter.
The instrument will show the actual setting.
Fault Relay
F A U L T R E L A Y
DE ENERG I Z ED / NO s e l e c t : 1 o f 2
Deenergized / NO
Defines the Fault Relay to be deenergized (normally open contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
F A U L T R E L A Y
ENERG I Z ED / NC s e l e c t : 2 o f 2
Energized / NC
Defines the Fault Relay to be energized (normally closed contact).
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Satellite XT FTT Versions 6-15
R E L A Y S
A L ARM DE L A Y s e l e c t : 2 o f 3
A L A R M D E L A Y
3 . 0 s e c .
p l e a s e e n t e r !
Configuration - Section 6
Alarm Delay
Used to define an alarm delay for concentration alarms.
Minimum value is 0, maximum setting is 99.9 seconds.
Default setting is 3.0 seconds.
Actuate the <set> key to enter and show the actual setting.
<Actual Alarm Delay>
To change the value, use the cursor keys <left> or
<right> to reach the desired position.
Use the cursor keys <up> and <down> to select the figures you wish to enter.
After the new value has been entered, actuate the <set> key to confirm or <esc> to exit.
R E L A Y S
F AU L T T R I GGER s e l e c t : 3 o f 3
Fault Trigger
Allows the user to define, which fault conditions will activate the Fault Relay.
Ex factory the setting is Fault Only.
Use the <set> key to enter.
The instrument will show the actual setting..
F A U L T T R I G G E R
F AU L T ON L Y s e l e c t : 1 o f 4
Fault Only
Used to configure the Fault Relay to be activated in case of a fault condition only.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
F A U L T T R I G G E R
F AU L T +WARN I NG s e l e c t : 2 o f 4
Fault + Warning
Used to configure the Fault Relay to be activated in case of a fault or warning condition only.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
F A U L T T R I G G E R
F AU L T +MA I N T s e l e c t : 3 o f 4
Fault + Maintenance
Used to configure the Fault Relay to be activated in case of a fault or maintenance condition only.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
F A U L T T R I G G E R
A L L s e l e c t : 4 o f 4
Used to configure the Fault Relay to be activated in case of a fault, warning, or maintenance condition.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
All
Satellite XT FTT Versions 6-16
Troubleshooting - Section 7
7.1
This section will help to determine the source of a fault orwarning condition and will provide a corrective action.
Warning and Fault Messages
Warning and Fault are two different kinds of messages, which require different attention.
For any message not explained in this section or any other assistance required, please do not hesitate to contact our service department.
7.1.1 Warnings
A warning condition indicates that the instrument requires some attention, but the instrument is still able to monitor and operate as programmed.
The following will occur when the instrument detects a warning condition:
- the green status LED is flashing
- a warning message is sent to the communication network
- the fault relay is activated, relays and their appropriate configuration provided.
Actuate the <set> key to acknowledge and reset a Warning Condition.
L OW Z E R O !
A s H 3
0 . 00 p p m
Low Zero !
This message indicates that the zero point of the sensor is out of the optimum range.
To correct this problem, check the zero adjustment and recalibrate, if necessary.
C H E C K S E N S O R !
A s H 3
0 . 00 p p m
Check Sensor !
This message indicates that the sensor sensitivity does not comply with the specifications. The sensor must soon be replaced.
Whilst most sensors will work for another 2-6 weeks, there are circumstances that may cause an earlier failure of the sensor cell. Honeywell therefore suggests that customers with more critical needs carry out the replacement of the sensor cell within 3 days (72 hours) of receiving the "check sensor" message.
To receive this message, the function Auto Selftest must be configured enabled; refer to Section 6, Configuration.
! This function is not available for instruments with oxygen sensor or catalytic sensor.
Satellite XT FTT Versions 7-1
Troubleshooting - Section 7
7.1.2 Faults
Instrument faults refer to a problem which prevents the instrument from operating properly and interferes with the ability to monitor or document concentration alarms.
When the instrument has detected an instrument fault, a fault message for each different fault condition is sent to the communication network. The green status LED is off and the fault relay is activated. The LCD display is flashing while indicating the specific fault
message.
For instruments with relay option the fault relay is activated additionally.
- - - - F AU L T - - - -
ADC NOT RE ADY !
Fault - ADC Not Ready !
This indicates that there is a problem with the internal electronics. The AD Converter has failed.
Switch power supply for this unit off and on again. If the instrument still shows this message, switch off power supply for this unit and contact our service department for further instructions.
- - - - F AU L T - - - -
- - -
WR
NO S ENSOR !
-
ON G
F A U L
S E
T
NS
- -
OR
- -
!
Fault - No Sensor !
This message indicates that no sensor is or a sensor type not suitable for this instrument is fitted.
To eliminate the problem, insert an appropriate sensor.
If the instrument is operated with a sensor extension, this fault message may also be caused by a bad connection.
Make sure all connections are firmly attached.
Fault - Wrong Sensor !
This message indicates that the sensor installed is not the sensor assigned to this particular instrument.
To correct the problem, insert the sensor assigned and documented to this instrument. The data are stated in the test certificate provided with the shipment.
- - - - F AU L T - - - -
RE P L ACE S ENSOR !
Fault - Replace Sensor !
This message indicates that the sensor presently installed is depleted (the warning CHECK SENSOR ! has been ignored).
To correct the problem, replace the old sensor immediately.
! This function is not available for instruments with oxygen sensor or catalytic sensor.
Satellite XT FTT Versions 7-2
- - -
E
-
X T
F A
RA
U L
C T
T
I V E
- -
!
- -
Troubleshooting - Section 7
Fault - Extractive !
Only for instruments with Extractive Module.
This message indicates that there is a problem with the
Extractive Module. Check power supply, sample tubing, sample tubing connections, and sample line filter. Perform any corrective action required.
If the message is still shown, please contact our service department for further instructions.
- - -
P
- F A
YRO L
U L T
Y Z ER
- - - -
!
Fault - Pyrolyzer !
Only for instruments with Pyrolyzer Module.
This message indicates that there is a problem with the
Pyrolyzer Module.
Please contact our service department for further instructions.
Satellite XT FTT Versions 7-3
Reference Information - Section 8
8.1 Sensor Order Information
Substance / Sensor
AsH3 Arsine (3 El.)
AsH3 Arsine (2 El.)
AsH3 Arsine (2 El.)
B2H6 Diborane
Br2 Bromine
C4F6
C5F8
Hexafluorobutadiene
Octafluorocyclopentene
CH3F Methyl Fluoride
CH4 Methane
Cl2:
ClF3
Chlorine
Chlorine Trifluoride
ClO2 Chlorine Dioxide
CO Carbon Monoxide
COCl2 Phosgene
DCE
1,2
Di-chloro-ethylene 1,2
C2H4O Ethylene Oxide
F2
F2
Fluorine
Fluorine
GeH4 Germane
H2 Hydrogen (1%)
H2
H2S
H2S
H2Se
HBr
HCl
Hydrogen (4 %)
Hydrogen Sulfide
Hydrogen Sulfide (org.)
Hydrogen Selenide
Hydrogen Bromide
Hydrogen Chloride
HCN
HF
Hydrogen Cyanide
Hydrogen Fluoride
HMDS Hexamethyldisilazane
HMDS Hexamethyldisilazane
N2H4 Hydrazine
NF3 Nitrogen Trifluoride
NH3
NH3
NO
NO2
O2
O3
O3
Ammonia (100 ppm)
Ammonia (1000 ppm)
Nitric Oxide
Nitrogen Dioxide
Oxygen
Ozone
Ozone
PH3
PH3
SF6
SiH4
Phosphine (3 El.)
Phosphine (2 El.)
Sulfur Hexafluoride
Silane
SO2 Sulfur Dioxide
TEOS Tetraethyl Orthosilicate
Part No
Sensor
9602-9600
9602-8000
9602-6400
9602-6401
9602-6902
9602-5100
9602-5101
9602-5200
9602-5201
9602-5601
9602-7000
9602-5800
9602-5700
9602-6500
9602-6714
9602-6715
9602-7600
9602-9700
9602-6704
9602-6705
9602-7200
9602-7300
9602-5500
9602-7100
9602-7101
9602-6102
9602-6100
9602-9710
9602-6301
9602-5900
9602-7500
9602-6004
9602-6000
9602-6002
9602-6202
9602-6800
9602-9732
9602-9730
9602-9720
9602-9900
9602-5300
9602-7410
9602-7400
9602-5400
9602-6600
Unit
ppm ppm ppm ppm ppm
% vol ppm ppm ppm ppm ppm ppm
% vol ppm ppm ppm ppm
% vol ppm ppm ppm ppm ppm ppm
% vol
% vol ppm ppm ppm ppm ppm
% vol
% LEL ppm ppm ppm ppm ppm ppm ppm ppm ppm ppm ppm ppm
Nominal
Range
0 ... 1000
0 ... 20.0
0 ... 5.00
0 ... 30
0 ... 5.0
0 ... 1.000
0 ... 4.00
0 ... 100
0 ... 30.0
0 ... 5.00
0 ... 30.0
0 ... 30.0
0 ... 30.0
0 ... 10.0
0 ... 500
0 ... 0.500
0 ... 1.00
0 ... 50.0
0 ... 100
0 ... 1000
0 ... 250
0 ... 25.0
0 ... 25.0
0 ... 1.00
0 ... 1.00
0 ... 1.00
0 ... 1.00
0 ... 0.500
0 ... 50.0
0 ... 25.0
0 ... 100
0 ... 1.00
0 ... 1.00
0 ... 10.0
0 ... 1.00
0 ... 5.00
0 ... 50.0
0 ... 20.0
0 ... 0.500
0 ... 100
0 ... 5.00
0 ... 1.00
0 ... 1.00
0 ... 500
0 ... 1.00
2)
1)
3)
1)
3)
2)
1)
Note
3)
4)
1)
1)
3)
3)
3)
Dust Filter allowed
no data no data
No
No
Yes
Yes
No no data
No
No
Yes
Yes
Yes
No
No
Yes
Yes no data
Yes
Yes
No
Yes
Yes
Yes
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
No no data no data no data no data
Yes
No
No
Yes
Yes
Satellite XT FTT Versions 8-1
Substance / Sensor
TMB Trimethyl borate
TMP Trimethyl phosphite
Part No
Sensor
Nominal
Range
9602-7510
9602-7800
0 ... 500
0 ... 30.0
Notes: 1) Special application
2) Special range
Unit
ppm ppm
3) Pyrolyzer Module required for detection
4) Only for use with Satellite XT C-versions
Further gases and ranges on request
Reference Information - Section 8
Note Dust Filter allowed
No
Yes
8.2 K-Factors for Satellite XT C-Versions
Catalytic sensors are calibrated for methane. When monitoring for other combustible gases, a correction factor must be entered; consult our local representative.
Satellite XT FTT Versions 8-2
Reference Information - Section 8
9630-0441
9630-0443
9660-0110
9660-0120
9660-0220
9660-0330
9660-0440
9675-0440
9902-4000
9902-4010
9902-4020
9902-4030
9902-4040
9902-4100
9902-4110
9902-4120
9902-4130
9902-4200
8.3 Spares and Accessories
Part Number Description
9602.0050.10.03
Mounting Rail Standard
9602.0050.10.02
Mounting Plate DIN-Rail, Option
9602.0051.10.02
Mounting Plate L-shaped with DIN-Rail, Option
9602.0090.00.01
Sensor Extension, 2 meters
9602.0091.00.01
Sensor Extension, 3 meters
9602.0092.00.01
Sensor Extension, 1 meter
9602.0093.00.01
Sensor Extension Combustible Gases, 2 meters
9602.0094.00.01
Sensor Extension Combustible Gases 3 meters
9602.0095.40.00
Calibration Cap
9630-0146
9650-0540
PC Network Adapter PCLTA-21 TP/FT-10
PC LonTalk Adapter PCC-10 TP/FT 10 for notebook use
9630-0423
9630-0424
9630-0505
9630-0431
Junction Box for max. 2 nodes
Junction Box for max. 6 nodes
Junction Box 1 node /R-Version
4-wire shielded cable, 2x2x1.0 mm2, 100 m coil
Network Terminator for FTT-10
Network Terminator for FTT-10 Bus Topology
Relay Output Module for DIN-Rail Mounting
Supervisor Module for DIN-Rail Mounting
Digital Input Module for DIN-Rail Mounting
Analog Input Module for DIN-Rail Mounting
Programmable Logic Module
Router FTT-10 FTT-10, 24 VDC for DIN-Rail Mounting
Duct Mounting Saddle Assy, 4 inches
Duct Mounting Saddle Assy, 6 inches
Duct Mounting Saddle Assy, 8 inches
Duct Mounting Saddle Assy, 10 inches
Duct Mounting Saddle Assy, 12 inches
Duct Mounting Saddle Assy, 1,5 inches
Duct Mounting Saddle Assy, 2 inches
Duct Mounting Saddle Assy, 2,5 inches
Duct Mounting Saddle Assy, 3 inches
Duct Mounting Saddle (Assy), flat
Satellite XT FTT Versions 8-3
8.4
Reference Information - Section 8
Network Terminators
Network Terminators are designed to provide electrical termination for twisted pair channels. They are passive devices and do not require electrical power.
In a free topology FTT-10 segment, one Network Terminator type 9630-0441 is required and may be located anywhere on the segment.
In a bus topology FTT-10 channel, two Network Terminators type 9630-0443 are required, one located at each end of the bus.
Technical Specifications
Network
wiring topologies
Physical Dimensions
size
weight
Mounting
Standardized LonTalk™ protocol free, e.g. Bus, Star, Loop, or mixed
18 x 58 x 60 mm
0.7” x 2.3” x 2.4”
40 g
1.4 oz
DIN rail
Housing Protection Class
IP 20
Operating Conditions
temperature
humidity
Part Number
-20 °C ... +50 °C
-4°F ... +122 °F
0 ... 95 % r.h.
FTT-10 9630-0441
FTT-10 Bus Topology 9630-0443
(W x H x D)
(W x H x D)
Satellite XT FTT Versions 8-4
Reference Information - Section 8
8.5 Relay Output Module
The Relay Output Module is a generic unit interfacing with LONWORKS™ systems, that use a free topology wiring scheme supporting star, loop, and/or bus wiring.
The Relay Output Module contains four relays with SPDT contacts and one dedicated power loss relay with SPST contact for activation of external alarm devices.
The relays can be operated in either energized or deenergized condition. All relay states are indicated by LEDs.
Network configuration data is customer specific and is programmed into nonvolatile
memory.
Technical Specifications
Power Requirements
voltage consumption
Network
data transmission
12 ... 24 VDC max. 1.9 W
Standardized LonTalk™ protocol wiring topologies
Wiring
Relay Outputs
contacts
78 kBit per second free, e.g. Bus, Star, Loop, or mixed
4-wire shielded cable 2x2x1.0 mm2 / 17 AWG
max. ratings
Status Indication
4 x SPDT (single-pole-double-throw)
1 x SPST (single-pole-single-throw)
250 VAC / 30 VDC, 8 A
4 red relay state LEDs
1 green status LED
Physical Dimensions
size
weight
1 yellow service LED
45 x 80 x 105 mm
1.8” x 3.1” x 4.1”
220 g
Mounting
7.8 oz
DIN rail
Housing Protection Class
IP 20
RFI / EMC
EN 55022
EN 61000-6-2
(W x H x D)
(W x H x D)
Operating Conditions
temperature
humidity
Part Number
0 °C ... +40 °C
+32 °F ... +104 °F
20 ... 90 % r.h.
9660-0110
Out 1
Out 2
NO
1
NO
7
C
2
C
8
NC
3
NC
9
C
4 not used
10
Power
P+
5
P–
6
P+
11
ROM
P–
12
Out 3
Out 4
NO
13
NO
19
C
14
C
20
NC
15
NC
21
16
NO
22
Network
N+
17
N–
18
N+
23
N–
24
Power Requirements:
Relay Ratings:
12 ... 24 VDC, max. 2 W
250 VAC / 30 VDC, 8 A
Satellite XT FTT Versions 8-5
Reference Information - Section 8
8.6 Supervisor Module
The Supervisor Module is designed to identify a fault condition within a subnet (max. 127 nodes) of a twisted pair LONWORKS™ network. Meaning one or more nodes on the subnet lose the ability to communicate, due to a bad node hardware or a broken wire.
In the latter case, the Supervisor Module automatically restores communication by closing the relay bus switch and thus reconnecting the both sides of the open circuit. Either fault condition is indicated by individual fault relays and reported to an optional monitoring
station.
The Supervisor Module contains four relays with SPDT contacts and one dedicated power loss relay with SPST contact for activation of external alarm devices.
The relays can be operated in either energized or deenergized condition. All relay states are indicated by LEDs. Network configuration data is customer specific and is programmed into nonvolatile memory.
Technical Specifications
Power Requirements
voltage consumption
Network
data transmission wiring topologies
Wiring
Relay Outputs
contacts
12 ... 24 VDC max. 1.9 W
Standardized LonTalk™ protocol
78 kBit per second free, e.g. Bus, Star, Loop, or mixed
4-wire shielded cable 2x2x1.0 mm2 / 17 AWG
max. ratings
Status Indication
2 x SPDT: Bus Switch
1 x SPDT: Node Failure
1 x SPDT: Broken Wire
1 x SPST: Power Loss
(SPDT = single-pole-double-throw)
(SPST = single-pole-single-throw)
250 VAC / 30 VDC, 8 A
4 red relay state LEDs
1 green status LED
1 yellow service LED
Physical Dimensions
size
weight
45 x 80 x 105 mm
1.8” x 3.1” x 4.1”
220 g
Mounting
7.8 oz
DIN rail
Housing Protection Class
IP 20
RFI / EMC
EN 55022
EN 61000-6-2
Operating Conditions
temperature
humidity
Part Number
0 °C ... +40 °C
+32 °F ... +104 °F
20 ... 90 % r.h.
9660-0120
Switch
N+
Switch
N–
Bus
Fault
Node
Fault
NO
1
NO
7
NO
13
NO
19
C
14
C
20
C
8
C
2
NC
3
(W x H x D)
(W x H x D)
Power
Loss
C
4
SBS
Power
P+ P–
5 6
NC
9
NC
15
NC
21
10 not used
16
NO
22
P+
11
P–
12
Network
N+ N–
17 18
N+
23
N–
24
Power Requirements:
Relay Ratings:
12 ... 24 VDC, max. 2 W
250 VAC / 30 VDC, 8 A
Satellite XT FTT Versions 8-6
Reference Information - Section 8
8.7 Digital Input Module
The Digital Input Module is a generic unit interfacing with LONWORKS™ systems, that use a free topology wiring scheme supporting star, loop, and/or bus wiring.
The Digital Input Module contains two groups of four optocoupler isolated digital inputs, each group with separate common signal ground. It is used for processing binary and digital values, e.g. switch signals.
All input states are indicated by LEDs. The module provides one dedicated power loss relay with SPST contact for activation of external alarm devices.
Network configuration data is customer specific and is programmed into nonvolatile
memory.
Technical Specifications
Power Requirements
voltage consumption
Network
data transmission
12 ... 24 VDC max. 0.8 W
Standardized LonTalk™ protocol
78 kBit per second wiring topologies
Wiring
Digital Inputs
channels
input voltage free, e.g. Bus, Star, Loop, or mixed
4-wire shielded cable 2x2x1.0 mm2 / 17 AWG
8 (2 groups of 4) optocoupler isolated
High Level 12 ... 24 VDC
Low Level 0 ... 2 VDC
Status Indication
8 red input state LEDs
1 green status LED
1 yellow service LED
Relay Output
contact
max. ratings
Physical Dimensions
size
weight
190 g
6.7 oz
Mounting
DIN rail
Housing Protection Class
IP 20
RFI / EMC
EN 55022
EN 61000-6-2
Operating Conditions
temperature
humidity
Part Number
1 x SPST (single-pole-single-throw)
250 VAC / 30 VDC, 8 A
45 x 80 x 105 mm
1.8” x 3.1” x 4.1”
0 °C ... +40 °C
+32 °F ... +104 °F
20 ... 90 % r.h.
9660-0220
In1
1
In3
7
In5
13
In7
19
In2 C1...4
2
In4
3 not used
8 9
In6
14 not used
15
In8 C5...8
20 21
(W x H x D)
(W x H x D)
Power
Loss
C
4 not used
10 not used
16
NO
22
Power
P+
5
P–
6
P+
11
P–
12
Network
N+ N–
17
N+
23
18
N–
24
DIM
Power Requirements:
Digital Input Voltage:
Relay Ratings:
12 ... 24 VDC, max. 0.8 W
High Level: 12 ... 24 VDC
Low Level: 0 ... 2 VDC
250 VAC / 30 VDC, 8 A
Satellite XT FTT Versions 8-7
Reference Information - Section 8
8.8 Analog Input Module
The Analog Input Module is a generic unit interfacing with LONWORKS™ systems, that use a free topology wiring scheme supporting star, loop, and/or bus wiring.
The Analog Input Module contains two isolated groups of each two analog input channels, for processing ± 0-10 VDC or ± 0-25 mA current signals. The analog inputs can be logically connected via a network to different input/output modules and data visualization software.
For each channel the module offers two user-settable alarm levels and input scaling.
All states are indicated by LEDs. The module provides one dedicated power loss relay with
SPST contact for activation of external alarm devices.
Network configuration data is customer specific and is programmed into nonvolatile
memory.
Technical Specifications
Power Requirements
voltage consumption
Network
data transmission wiring topologies
12 ... 24 VDC typically 2 W
Standardized LonTalk™ protocol
78 kBit per second free, e.g. Bus, Star, Loop, or mixed
Wiring
Analog Inputs
channels
signal range
resolution
Status Indication
4-wire shielded cable 2x2x1.0 mm2 / 17 AWG
2 isolated groups of each 2 inputs with common signal ground
± 0 - 10 VDC or ± 0 - 25 mA
14-bit, 10 samples per second
8 red input state LEDs
1 green status LED
1 yellow service LED
Relay Output
contact
max. ratings
Physical Dimensions
size
1 x SPST (single-pole-single-throw)
250 VAC / 30 VDC, 8 A
45 x 80 x 105 mm
1.8” x 3.1” x 4.1”
weight
200 g
7 oz
Mounting
DIN rail
Housing Protection Class
IP 20
RFI / EMC
EN 55022
EN 61000-6-2
Ch 1
U+ Com– I+
1 2 3
Operating Conditions
temperature
humidity
Part Number
0 °C ... +40 °C
+32 °F ... +104 °F
20 ... 90 % r.h.
9660-0330
Ch 2
Ch 3
Ch 4
U+ Com– I+
7 8 9
U+
13
U+
19
Com–
14
Com–
20
I+
15
I+
21
(W x H x D)
(W x H x D)
Power
Loss
C
4
10 not used
16
NO
22
Power
P+
5
P–
6
P+
11
Network
N+ N–
17 18
N+
23
AIM
P–
12
N–
24
Power Requirements: 12 ... 24 VDC, typ. 2 W
Input Voltage Range: ± 0 ... 10 VDC
Input Current Range: ± 0 ... 25 mADC
Relay Ratings: 250 VAC / 30 VDC, 8 A
Satellite XT FTT Versions 8-8
Reference Information - Section 8
8.9 Router Modules
Router Modules connect two communication channels and route LonTalk™ messages between them. These modules interface two different twisted-pair channels, e.g. high speed backbone and a free topology channel. Additionally the routers are used to manage network traffic, increase the total number of nodes, or extend the maximum channel
length.
All states are indicated by LEDs. The module provides one dedicated power loss relay with
SPST contact for activation of external alarm devices.
Network configuration data is customer specific and is programmed into nonvolatile
memory.
Technical Specifications
Power Requirements
voltage consumption
Network
Wiring
Transceiver Types
channels
Status Indication
FTT-10 FTT-10
12 ... 24 VDC max. 1.2 W
Standardized LonTalk™ protocol shielded 4-wire cable 2x2x1.0 mm2 /17 AWG
FTT-10A (78 kBit per second)
TP/XF-1250 (1.25 MBit per second)
1 red network traffic LED
1 green status LED
2 yellow service LEDs
Relay Output
contact
max. ratings
Operating Conditions
temperature
humidity
Part Number
1 x SPST (single-pole-single-throw)
250 VAC / 30 VDC, 8 A
Physical Dimensions
size
weight
Mounting
45 x 80 x 105 mm
1.8” x 3.1” x 4.1”
220 g
7.8 oz
DIN rail
Housing Protection Class
IP 20
RFI / EMC
EN 55022
EN 61000-6-2
0 °C ... +40 °C
+32 °F ... +104 °F
20 ... 90 % r.h.
9675-0440
Net
Side A
N+
1
N+
7
N–
2
N–
8
S
3
S
9
(W x H x D)
(W x H x D)
Power
C
4 not used
10
RTM
Power
P+ P–
5
P+
11
6
P–
12
Net
Side B
N+
13
N–
14
N+ N–
19 20
S
15
S
21 not used
16
NO
22 not used
17 not used
23 not used
18 not used
24
Power Requirements:
Relay Ratings:
12 ... 24 VDC, max. 1.2 W
250 VAC / 30 VDC, 8 A
Satellite XT FTT Versions 8-9
Extractive Module XT - Section 9
9.1 Principle of Operation
The Extractive Module XT is an add-on module for use in combination with the
Satellite XT. It enables for keeping hazardous, difficult to reach, or otherwise inaccessible areas under continuous surveillance. It is capable of sampling areas up to 50 m away.
Power is supplied via the Satellite XT. Status information is provided by the Satellite XT that provides the digital interface as well.
9.2
The module contains a diagnosis for electronics and pump. The correct gas flow is factory set and saved in the instrument. If there is a problem with the Extractive Module XT, the display of the Satellite XT shows the fault message FAULT EXTRACTIVE. For information on fault messages and instructions on how to correct a fault condition, please refer to
Section 7, Troubleshooting.
General Instructions
Listed hereafter several aspects that should be reflected when locating monitoring point and instrument. When placing the sampling point , the property (lighter or heavier than air) of the target gas must be considered. The instrument should be mounted as close to the monitoring location as possible to minimize sample transport time. The area immediately surrounding the sample inlet at the monitoring point must be free from objects that might hinder free flow of air. The instrument should be installed away from any possible liquid sources, extreme dust and dirt sources and should be protected against rain and sunlight.
The sample line should be run as direct as possible to improve transport time. Avoid running the sample line through areas of great temperature extremes. The sample outlet must be connected to an exhaust line and must be properly vented. Both, sample and exhaust tubing must not be crimped or placed in an area where weight could collapse the
tubing.
Dust may be a result of construction as well as manufacturing activities. If unusual quantities of dust are expected at the monitoring location, an appropriate filter for the sample line must be installed. The use of a dust filter depends on the target gas, refer to the table Sensor Order Information in Section 8, Reference Information, or contact our service department.
As excess amounts of dirt in the filters will reduce the sample flow and will thus affect con centration readings of the instrument, make sure to regularly change sample line filters
(every 1 to 6 months, depending on the ambient conditions). For order information refer to
Section 8, Reference Information, Spares and Accessories.
9.3
Moisture may occur as rain entering a line at an outdoor sampling location or as condensation from outdoor-to-indoor temperature fluctuations. In case of unusual quantities of moisture, appropriate steps must be taken to protect the instrument; consult us for respective assistance.
Safety Instructions
Never use compressed air to clean the tubing while it is connected to the instrument.
Use isopropanol and a soft cloth to clean the housing. Do not use any aggressive
detergents.
! Note:
To extend an existing Satellite XT to an extractive system, please contact our service department. This modification makes it necessary to open the housing of the Satellite XT, which is only allowed for authorized persons.
Satellite XT FTT Versions 9-1
Extractive Module XT - Section 9
9.4 Instrument Design
Front View showing Extractive Module XT connected to Satellite XT.
A B
Exhaust Line - Outlet
Sensor Holder
MST Extractive esc
MST Satellite XT set
Satellite XT FTT Versions 9-2
Side View (left)
A Exhaust Line - Outlet
B Sensor Holder
C Sample Line - Gas Inlet
C
B
A
Extractive Module XT - Section 9
Side View (right)
B Sensor Holder
D Sensor Connection to Satellite XT
E Connection Extractive Module XT and Satellite XT
B D
E
D
Top View (only Extractive Module XT)
A Exhaust Line - Outlet
B Sensor Holder
C Sample Line - Gas Inlet
D Sensor Connection to Satellite XT
Satellite XT FTT Versions
C
A
Gas
Inlet
Outlet
B
9-3
Extractive Module XT - Section 9
9.5 Mounting
The Extractive Module XT is prepared for DIN-rail mounting. An appropriate mounting rail is delivered with the instrument, which allows mounting the Extractive Module XT and the
Satellite XT side by side.
Mount this rail to a wall with the appropriate fasteners and slide both modules onto it.
Plug in male and female connector of Satellite XT and Extractive Module XT.
1
MST Extractive esc
MST Satellite XT set
Module Connections
Module Connections
sample tubing.
sensor connection between Extractive Module XT and
Satellite XT.
9.6 Tubing Connections
Tubing material is PTFE or PFA with an outer diameter of 1/4” and an inner diameter of
3/16”. Length and inner diameter of the sample line will affect the response time. Length of the sample lines should be kept as short as possible. With a sample line ID of 3/16”
(4.7 mm) and a length of 10 m, the delay in response does not exceed 30 seconds.
Sample Line Inlet
The fitting for connecting the sample line inlet is located on top of the housing and marked
“Gas Inlet”. Improper installation of the sample inlet tube may result in dilution or even complete loss of the sample.
Exhaust Line Outlet
The fitting for connection of the exhaust line outlet is located on top of the instrument and marked “Outlet”. Since hazardous gas concentrations may still be present at the exhaust line outlet, it is recommended to connect the exhaust line to an exhaust duct.
Note:
If the pump is operated at the maximum sample line distance of 50 m, the maximum exhaust line length is 25 m to avoid flow failure due to back pressure in the pump.
Note:
Do not apply power until the system is ready for start-up.
Satellite XT FTT Versions 9-4
Extractive Module XT - Section 9
9.7 Sensor Replacement
This procedure is used to replace a depleted sensor by a fresh sensor. The replacement sensor must have the same part number as the installed sensor. Press the <esc> key on the Satellite XT keypad and enter the password to exit the Monitoring Mode. The green status LED of the Satellite XT is off, the instrument is not monitoring. A maintenance message is transmitted to the communication network.
Go to the Maintenance Menu and press the <set> key to enter.
Go to the submenu Sensor Service and press the <set> key to enter.
Follow the dialogue displayed at the bottom line of the screen - Figure 1.
Figure 1
M A I N T E N A N C E
S ENSOR S ERV I CE s e l e c t : 1 o f 7
<REMOVE SENSOR>
Unscrew the gland nut (F) to remove the sensor holder (B) from the flow adapter of the Extractive Module XT - Figure 2.
Remove the installed sensor (G) -
Figure 3.
Press the <set> key to continue.
Figure 2
B
S E N S O R S E R V I C E
REMOV E S ENSOR
< s e t > w h e n r e a d y
S E N S O R S E R V I C E
RE P L ACEMEN T p l e a s e i n s e r t !
S E N S O R S E R V I C E
L OAD NEW DA T A ?
< s e t > t o c o n f i r m
S E N S O R S E R V I C E
L OAD I NG DA T A !
p l e a s e w a i t !
S E N SOR W A RM - U P !
1 . 23
S a t e l l i t e X T
A s H 3
0 . 00 p p m p p m remove shorting device from the sensor plug, where necessary.
<REPLACEMENT>
Insert the new sensor (G) into the socket inside the sensor holder, aligning nose of the socket, groove of the sensor, and the arrow printed on the sensor label -
Figure 4.
Insert the sensor holder (B) with the replacement sensor installed into the flow adapter of the instrument and fasten the gland nut (F) - Figure 5.
Press the <set> key to continue.
<LOAD NEW DATA ?>
be loaded, confirm this message by pressing the <set> key.
<LOADING DATA !>
The instrument is now loading the new data from the sensor into the unit’s internal memory.
<SENSOR WARM-UP>
A sensor warm-up is performed and the display of the Satellite XT shows the according message, until the displayed value is zero. The warm-up time required depends on the type of sensor.
After the sensor warm-up, the instrument will automatically switch to the Monitoring
Mode.
The Sensor Service procedure is also described in Section 4, Maintenance.
Figure 3
F
Figure 4
F
Figure 5
B
G
F
G
F
Satellite XT FTT Versions 9-5
Extractive Module XT - Section 9
9.8 Technical Specifications
Power Requirements
Voltage
Consumption
Physical Dimensions
Size
Weight
Housing Protection Class
RFI / EMC
Operating Conditions
Sample tubing
Response time
Temperature
Pressure
Humidity power is provided by the Satellite XT, power supply range 18 - 24 VDC max. 2.4 W
78 x 95,5 x 50 mm
3,1" x 3,8" x 2,0"
(W x H x D)
(W x H x D)
500 g
17.6 oz
IP 30
0 °C ... +40 °C
+32 °F ... +104 °F
700 ... 1300 hPа
20 ... 90 % r.h.
EN 50270
OD 1/4” ID 3/16” PTFE or PFA
< 30 sec. with a tubing length of 10m
Maximum negative pressure (at pump)
-150mbar
Maximum sample line length
Recommended sample line length
Part Number
version for toxic/corrosive gases
(except C5F8 and SF6) version for combustible gases version for C5F8 and SF6
50 m (see note in section 9.6)
10 m
20404-0200
20404-0250
20404-0300
Warning:
Extractive Module 20404-0200 must only be used with Pyrolyzer
Modules 20408-0110, 20408-0112, 20408-0116 and 20408-0122.
Extractive Module 20404-0300 must only be used with Pyrolyzer
Modules 20408-0114 and 20408-0120.
Satellite XT FTT Versions 9-6
Pyrolyzer Module XT - Section 10
10.1 Principle of Operation
The Pyrolyzer Module XT is an add-on module for monitoring gases requiring pyrolytic sample preparation for detectability. It requires both, a Satellite XT and an Extractive Module XT.
The connected Satellite XT displays the module’s status of operation and provides the digital interface.
Each instrument is only configured for monitoring the substance specified on the identification label. The instrument is delivered with a certain sensor installed. Reference for assignment of instrument and sensor is the serial number. These data are also documented in the Certificate of Quality Control provided with the shipment. Instrument and sensor have been gas calibrated and these specific calibration parameters are stored in the integrated data memory of sensor and instrument.
Make sure to use only sensors for the substance specified on the module’s identification label.
Use only sensors designed for use with the Pyrolyzer Module XT.
The Pyrolyzer Module XT is powered by a wide range power supply. The module contains a diagnosis for electronics and filament. The filament voltage is factory set and saved in the instrument. If there is a problem with the Pyrolyzer Module XT, the display of the Satellite XT reads FAULT PYROLYZER. Information on fault messages and instructions on how to correct a fault condition, are provided in Section 7, Troubleshooting.
10.2 General Instructions
Listed hereafter are several aspects that should be reflected when locating monitoring point and instrument. When placing the sampling point , the property (lighter or heavier than air) of the target gas must be considered. The instrument should be mounted as close to the monitoring location as possible to minimize sample transport time. The area immediately surrounding the sample inlet at the monitoring point must be free from objects that might hinder free flow of air. The instrument should be installed away from any possible liquid sources, extreme dust and dirt sources and should be protected against rain and sunlight.
The instrument must be mounted in an upright position. The ventilation apertures on top and rear of the the housing must not be covered. The instrument must be easily accessable for operation and there must be sufficient space above the instrument to allow sensor replacement or to open the housing for maintenance purposes.
The sample line should be run as direct as possible to improve transport time. Avoid running the sample line through areas of great temperature extremes. The sample outlet must be connected to an exhaust line and must be properly vented. Both, sample and exhaust tubing must not be crimped or placed in an area where weight could collapse the tubing.
10.3 Safety Instructions
Always disconnect the Pyrolyzer Module XT from mains power before you open the housing.
The vents on top and rear of the instrument must not be covered. Do not enter any tapered or sharp items into the vents.
Use isopropanol and a soft cloth to clean the housing. Do not use any aggressive detergents.
Never use compressed air to clean the tubing when connected to the instrument. This will destroy the internal pressure switch.
WARNING: The Pyrolyzer Module XT will not respond to the target gas in conditions of less than 20 % r.h. Observe the specified operating conditions for humidity.
Satellite XT FTT Versions 10-1
Pyrolyzer Module XT - Section 10
10.4 Instrument Design
The Pyrolyzer Module XT can only be operated in combination with the Satellite XT and the
Extractive Module XT. All three modules are prepared for DIN-rail assembly.
Figure 2: Front view of the complete assembly
Pyrolyzer Module, Extractive Module, and Satellite XT.
Figure 1: Side view of the Pyrolyzer
Module only.
F G
H
M J
MST Extractive esc
MST Satellite XT set
L
K
Figure 3: Top View of the complete assembly
G Pyrolyzer Module XT
Inlet
Gas
Outlet
Gas
Inlet
Outlet
B Sensor Holder
F Sample Line - Inlet
G Vent
B
H Sample Line - connection to Extractive Module XT
J DIN-rail
Extractive Module XT
K Data Link - Pyrolyzer Module XT to Extractive Module XT
L Power Fitting
M Adapter for DIN-rail
Satellite XT FTT Versions
Satellite XT
10-2
Pyrolyzer Module XT - Section 10
10.5 Mounting
The Pyrolyzer Module XT is prepared for DIN-rail assembly. An appropriate mounting rail is delivered with the instrument. Mount this rail to a flat surface with the appropriate fasteners.
When the rail is mounted, follow the steps below:
1. Slide the Pyrolyzer Module XT onto the rail mounted to the wall, see Figure 4.
Figure 4
Figure 4
1
1
2. Slide the Extractive Module XT onto the rail at the front of the Pyrolyzer Module XT, see
Figure 5.
3. Slide the Satellite XT onto the rail at the front of the Pyrolyzer Module XT, as illustrated in
Figure 5.
4. Plug in male and female connector of Satellite XT and Extractive Module XT.
Figure 5
2
2
MST Extractive
MST Extractive
4
3
esc
4
3
Satellite XT
MST Satellite XT set set
Satellite XT FTT Versions 10-3
Pyrolyzer Module XT - Section 10
10.6 Tubing Connections
Tubing material is PTFE or PFA with an outer diameter of 1/4” and an inner diameter of 3/16”.
Length and inner diameter of the sample line will affect the response time. Length of the sample lines should be kept as short as possible. With a sample line ID of 3/16” (4.7 mm) and a length of 10 m, the delay in response does not exceed 30 seconds.
Moisture may occur either as rain entering a line at an outdoor sampling location or as condensation from outdoor-to-indoor temperature fluctuations. In case of unusual quantities of moisture, appropriate steps must be taken to protect the instrument; consult us for respective assistance.
Sample Line Inlet
The fitting for connecting the sample line inlet is located on top of the housing and marked
“Inlet”. Improper installation of the sample inlet tube may result in dilution or even complete loss of sample.
Sample Line Filter
Dust may be a result of construction as well as manufacturing activities. If unusual quantities of dust are expected at the monitoring location, an appropriate filter for the sample line must be installed. The use of a dust filter depends on the target gas, refer to the table Sensor Order
Information in Section 8, Reference Information, or contact our service department. As excess amounts of dirt in the filters will reduce the sample flow and will thus affect concentration readings of the instrument, make sure to regularly change sample line filters (every 1 to 6 months, depending on the ambient conditions). For order information refer to Section 8,
Reference Information, Spares and Accessories.
Module Connections
sample tubing.
Module Connections
sensor connection between Extractive
Module XT and Satellite XT.
data link connecting Pyrolyzer Module sensor connection between Extractive
XT.
data link connecting Pyrolyzer Module XT and Extractive Module XT / Satellite XT. bottom plate of the Pyrolyzer Module into the connector in the bottom of the
Extractive Module XT.
10.8 Power Connection
Use a standard power cable (obey any local regulations) and bring it through the power cable fitting into the module. Open the module from the top and connect
90 ...
240 VAC
Satellite XT FTT Versions 10-4
Pyrolyzer Module XT - Section 10
Inlet
Outlet
Inlet
Outlet
Extractive Modul XT
30 l/h 0,5 l/min
Flowmeter
Satellite XT
Pyrolyzer XT with bypass for C5F8
When the Pyrolyzer is used with C5F8, the flow needs to be adjusted at the outlet of the Extractive
Module XT to 0.5 l/min or 30 l/h.
On the underside of the Extractive Module XT is the interconnection to the Pyrolyzer Module XT, and next to it, two small holes which gives access to two pushbuttons. The left hand button will increase the flow, and the right hand button will decrease it.
The adjustment is required as the internal flow resistance is different to the other types of the Pyrolyzer
Module XT and this resistance would damage the Extractive Module XT.
Satellite XT FTT Versions 10-5
Pyrolyzer Module XT - Section 10
10.9
Technical Specifications
Power requirements
Voltage
Consumption
Physical dimensions
Size:
Weight:
Housing Protection Class
RFI / EMC
Operating Conditions
Sample tubing
Response time
Temperature
Pressure
Humidity
90 ... 240 VAC approx. 22 W / 2.13 A
223 х 97 х 100 mm (W х H х D)
8,8" x 3,8" x 3,9" (W х H х D)
1800 g
4 lbs
IP 30
EN 61010-1
EN 50270
OD 1/4” ID 3/16”
< 30 sec. with a tubing length of 10m
0 °C ... +40 °C
+32 °F ... +104 °F
700 ... 1300 hPа
20 ... 90 % r.h.
Note: For applications where the humidity is less than 20 % r.h., consider using the MST Sample Conditioning System - Humidifier, part number 20408-4010.
Minimum Oxygen content of sample
Part Numbers
Pyrolyzer Module XT NF3
Pyrolyzer Module XT 1,2-DCE
Pyrolyzer Module XT SF6
Pyrolyzer Module XT CH3F
Pyrolyzer Module XT C5F8
Pyrolyzer Module XT C4F6
3 %v/v
20408-0110
20408-0112
20408-0114
20408-0116
20408-0120
20408-0122
Warning:
Pyrolyzer Modules 20408-0110, 20408-0112, 20408-0116 and 20408-0122 must only be used with Extractive Module 20404-0200.
Pyrolyzer Modules 20408-0114 and 20408-0120 must only be used with
Extractive Module 20404-0300.
Satellite XT FTT Versions 10-6
Find out more
www.honeywellanalytics.com
Contact Honeywell Analytics:
Europe, Middle East, Africa, India
Life Safety Distribution AG
Javastrasse 2
8604 Hegnau
Switzerland
Tel: +41 (0)44 943 4300
Fax: +41 (0)44 943 4398
India Tel: +91 124 4752700 [email protected]
Americas
Honeywell Analytics Inc.
405 Barclay Blvd.
Lincolnshire, IL 60069
USA
Tel: +1 847 955 8200
Toll free: +1 800 538 0363
Fax: +1 847 955 8210 [email protected]
Asia Pacific
Honeywell Analytics Asia Pacific
#701 Kolon Science Valley (1)
43 Digital-Ro 34-Gil, Guro-Gu
Seoul 152-729
Korea
Tel: +82 (0)2 6909 0300
Fax: +82 (0)2 2025 0388 [email protected]
Technical Services
EMEAI: [email protected]
www.honeywell.com
Please Note:
While every effort has been made to ensure accuracy in this publication, no responsibility can be accepted for errors or omissions.
Data may change, as well as legislation, and you are strongly advised to obtain copies of the most recently issued regulations, standards, and guidelines.This publication is not intended to form the basis of a contract.
Issue 5 08/2013
H_MAN0859_EMEAI
PN 2401M3005 ECO A03534
© 2013 Honeywell Analytics
advertisement
Key Features
- Continuous gas monitoring
- LONWORKS™ free topology system compatibility
- Various sensor types support
- Alarm level settings
- Relay outputs for external alarm devices
- Self-diagnostic capabilities
- Menu-driven operation
- DIN-rail mounting
- IP52 protection class
- Sensor service procedures