Hioki 7016 Operation Manual
Hioki 7016 is an advanced signal source capable of generating a variety of signals, including constant voltage, constant current, and pulse. It also offers a SCAN function for outputting sequential values and a RAMP function for simulating linear waveforms. With its wide range of features and high accuracy, the Hioki 7016 is ideal for applications such as flow meter calibration, checking and calibrating counters, rotation counters, oscilloscopes, frequency converters, and other frequency-input systems.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
OPERATION GUIDE
7016
SIGNAL SOURCE
Contents
Contents
Introduction ............................................................... 1
Chapter 1 Source Function 3
1.1
Basic Sourcing Procedure........................... 4
1.2
Constant-Voltage Source (DCCV)............... 6
1.3
Constant-Current Source (DCCC)............... 8
1.4
Pulse Source (PULSE).............................. 10
1.5
Using Memory Function
(SCAN and RAMP).................................... 12
1.5.1 Using the SCAN Function .............. 12
1.5.2 Using the RAMP Function.............. 17
Chapter 2 Measurement Function 21
2.1
Basic Measurement Procedure
(Common Procedure)................................ 21
2.2
Measuring Voltage .................................... 23
2.3
Measuring Current..................................... 25
2.4
Measuring Frequency................................ 27
2.5
Measuring the Resistance......................... 28
2.5.1 Continuity Test ............................... 30
2.6
Diode Test ................................................. 31
2.7
Measuring Temperature ............................ 32
i
ii
Contents
Chapter 3 Special Measurement
Functions 33
3.1
Recording Function ................................... 34
3.2
Data Hold / Refresh Hold Function ........... 36
3.3
Relative (relative display) Function ........... 38
3.4
1-ms Peak Hold Function.......................... 39
3.5
Dual Display Function ............................... 41
Introduction
1
Introduction
This manual describes the basic operation of
HIOKI 7016 SIGNAL SOURCE.
❖
→Instruction manual for the 7016
Symbols
❖
Indicates the reference.
Indicates quick references for operation.
2
Introduction
Source Function
Chapter 1
3
Important!
• The yellow labels around the outer circumference of the function selector indicate signal source/generation
• functions.
• The [
SHIFT
] button must first be pressed to enable the source-related button functions.
4
1.1 Basic Sourcing Procedure
1.1 Basic Sourcing Procedure
1.
Set the slide switch to the
[
M/S
] position.
2.
Set the function selector to the following position.
(ex. Sourcing constantcurrent)
Sourcing constant-voltage
[
CV
]
Sourcing constant-current
[
CC
]
Sourcing pulse
Using memory function
[
PULSE
]
Voltage source: [
CV
]
Current source: [
CC
]
1.1 Basic Sourcing Procedure
3.
Press the [
SHIFT
] button and confirm that displayed on the LCD.
is
(Ex. Voltage Source)
5
The source-related button functions are now enabled.
4.
Press the [
MODE
] button as necessary to select the source mode or the parameter on the secondary display.
5.
Making the settings with the [
W
], [
X
], [
S
], or [
T
] button.
6.
Press the [
OUTPUT
] button and output the setting value. Verify that displayed on the LCD.
is
(Ex. Voltage Source)
7.
Press the [
OUTPUT
] button again to disable output. Verify that is no longer displayed.
6
1.2 Constant-Voltage Source (DCCV)
1.2 Constant-Voltage Source (DCCV)
Procedure
This mode outputs a constant current at the set value regardless of load.
1.
Set the slide switch.
[
M/S
]
2.
Set the function selector.
[
CV
]
3.
Press the [
SHIFT
] button.
Confirm that is displayed.
4.
Press the [
MODE
] button as necessary to select the source mode on the secondary display.
Source Mode
• Constant-voltage source 1.5 V range
• Constant-voltage source 15 V range
• Constant-voltage
SCAN source 1.5 V range
• Constant-voltage
SCAN source 15 V range
• Constant-voltage
RAMP source 1.5
V range
• Constant-voltage
RAMP source 15 V range
5.
1.2 Constant-Voltage Source (DCCV)
Select a digit to change.
With the [ W ] or [ X ] button
7
6.
Set a value.
With the [ S ] or [ T ] button
Settanble Ranges
[ +0.0000 V ]
(1.5 V range)
[ +00.000 V ]
(15 V range)
-1.5000 V to 1.5000 V
-15.000 V to 15.000 V
7.
Press the [
OUTPUT
] button to output the set voltage value.
Verify that displayed.
is
8.
Press the [
OUTPUT
] button again to disable output.
Verify that is no longer displayed.
8
1.3 Constant-Current Source (DCCC)
1.3 Constant-Current Source (DCCC)
This mode outputs a constant current at the set value regardless of load.
Procedure
1.
Set the slide switch [
M/S
]
2.
Set the function selector.
[
CC
]
3.
Press the [
SHIFT
] button.
Confirm that is displayed
4.
Press the [
MODE
] button as necessary to select the source mode on the secondary display.
Source Mode
• Constant-current source
• Constant-current
SCAN source
• Constant-current
RAMP source
5.
Select a digit to change.
6.
Set current value.
With the [ W ] or [ X ] button
With the [ S ] or [ T ] button
1.3 Constant-Current Source (DCCC)
Settable range
[ +00.000 mA ]
(25 mA range)
-25.000 mA to 25.000 mA
9
7.
Press the [
OUTPUT
] button to output the set current value.
Verify that displayed.
is
8.
Press the [
OUTPUT
] button again to disable output.
Verify that is no longer displayed.
10
1.4 Pulse Source (PULSE)
1.4 Pulse Source (PULSE)
• In addition to flow meter calibration, the pulse source can be used for checking and calibrating counters, rotation counters, oscilloscopes, frequency converters and other frequency-input systems.
• Frequency, duty cycle, pulse width and amplitude can be set.
Procedure
1.
Set the slide switch.
[
M/S
]
2.
Set the function selector.
[
PULSE
]
3.
Press the [
SHIFT
] button.
Confirm that displayed.
is
4.
Press the [
MODE
] button to select a parameter for setting.
• Pulse source frequency
• Pulse source duty cycle
• Pulse source pulse width
• Pulse source amplitude
5.
Set a value.
With the [ S ] or [ T ] button
Parameter
Frequency
Duty cycle
Pulse width
Amplitude
1.4 Pulse Source (PULSE)
11
Settable Ranges
0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60,
75, 80, 100, 120, 150, 200, 240, 300, 400,
480, 600, 1200, 1600, 2400, 4800 (Hz)
As there are 256 settable steps, each step is 0.390625%, which is displayed as 0.39% per step increments.
Example: If an factory default settings value of 50.00% is incremented by one step, the display shows
50.39%.
As there are 256 settable steps, the pulse width per step is 1/(256 * frequency).
+5V, ±5 V, +12 V, ±12 V
6.
Press the [
OUTPUT
] button to output the set pulse value.
Verify that displayed.
is
7.
Press the [
OUTPUT
] button again to disable output.
Verify that is no longer displayed.
12
1.5 Using Memory Function (SCAN and RAMP)
1.5 Using Memory Function (SCAN and RAMP)
S C A N
Function
Up to 16 steps of different source values and time intervals can be set.
R A M P
Function
A dual slope can be generated to simulate a linear waveform.
The number of divisions in each slope can be set.
❖
1.5.1 "Using the SCAN
Function"
(p.12)
❖
1.5.2 "Using the RAMP
Function"
(p.17)
1.5.1 Using the SCAN Function
Procedure
1.
Set the slide switch.
[
M/S
]
2.
Set the function selector.
Voltage source: [
CV
]
Current source: [
CC
]
3.
Press the [
SHIFT
] button.
Confirm that is displayed.
4.
Press the [
MODE
] button as necessary to select the source mode on the secondary display.
• Voltage source
(SCAN±1.5 V or
SCAN±15 V)
• Current source
SCAN±25 mA
1.5 Using Memory Function (SCAN and RAMP)
13
During SCAN function setup, is displayed.
SCAN function factory default settings
Range
±1.5000 V ±15.000 V ±25.000 mA
Time Interval
Step Output Value Output Value Output Value
13
14
15
16
9
10
11
12
7
8
5
6
3
4
1
2
-0.9000 V
-1.2000 V
-1.5000 V
+0.0000 V
+0.0000 V
+0.0000 V
-1.5000 V
+0.0000 V
+1.5000 V
+1.2000 V
+0.9000 V
+0.6000 V
+0.3000 V
+0.0000 V
-0.3000 V
-0.6000 V
-09.000 V
-12.000 V
-15.000 V
+00.000 V
+00.000 V
+00.000 V
-15.000 V
+00.000 V
+15.000 V
+12.000 V
+09.000 V
+06.000 V
+03.000 V
+00.000 V
-03.000 V
-06.000 V
+00.000 mA
+04.000 mA
+08.000 mA
+12.000 mA
+16.000 mA
+20.000 mA
+16.000 mA
+12.000 mA
+08.000 mA
+04.000 mA
+00.000 mA
+04.000 mA
+08.000 mA
+12.000 mA
+16.000 mA
+20.000 mA
0 S
0 S
0 S
0 S
2 S
2 S
2 S
0 S
2 S
2 S
2 S
2 S
2 S
2 S
2 S
2 S
Do you want to change the values from the factory default settings?
❖
"Changing Data
Settings" (p.16)
14
1.5 Using Memory Function (SCAN and RAMP)
5.
Select the SCAN source mode on the secondary display.
With the [ W ] or [ X ] button
Cont
• Output proceeds sequentially from the first step to the last, then returns to the first step and repeats continuously.
• However, if the time interval of the last step is set to 0 seconds, it is ignored, and output continues from the first step.
Example:
When the factory default settings time interval setting for Step 12 is 0 seconds, after output from Step 11, output continues from Step 1.
CyCLE
• Output proceeds sequentially from the first step to the last, then continues output at the settings of the last step.
• However, if the time interval of a step is set to 0 seconds, output continues according to the settings of the previous step.
Example:
When the factory default settings time interval setting for Step 12 is 0 seconds, output from Step 11 is maintained continuously.
StEP
Pressing the [
S
] or [
T
] button moves to the previous or next output step. The value set for each step is output until the next step is selected.
1.5 Using Memory Function (SCAN and RAMP)
Step (StEP)
15
Continuouce
(Cont)
Cycle
(CyCLE)
Time interval (second)
6.
Press the [
OUTPUT
] button to output the set voltage or current value.
Verify that displayed.
is
7.
Press the [
OUTPUT
] button again to disable output.
Verify that is no longer displayed.
• Continuous (
CONT
) and Cycle (
CyCLE
) source modes always start at Step 1.
• If the time interval of Step 1 is set to 0s, the source out put is not activated. (Except for STEP source mode)
• When Continuous (
CONT
) and Cycle
(
CyCLE
) source modes are stopped by the [
OUTPUT
] button, subsequent source output resumes at Step 1.
16
1.5 Using Memory Function (SCAN and RAMP)
Changing Data Settings
1.
With the SCAN function active, hold the [
MODE
] button for one second to enter the data setting mode for each source step.
2.
Set the time interval and source output value for each step.
Secondary display
Leftmost two digits of the primary display
Rightmost two digits of the primary display
Output value
Step number
Time interval
Press the [
MODE
] button to select the step number, time interval or output value. The parameter digit to be set blinks.
• Press the [ S ] or [ T ] button to increment or decrement the value, and press the [ W ] or [ X ] button to move to another digit of the output value.
• Time intervals can be set from 0 to 99 seconds.
3.
Press the [
OUTPUT
] button to save your settings.
4.
To clear all settings, hold the [ X ] button for one second. Time interval and output values are cleared to zero.
5.
Hold the [
MODE
] button for one second to exit the step data setting mode.
1.5 Using Memory Function (SCAN and RAMP)
1.5.2 Using the RAMP Function
Procedure
1.
Set the slide switch.
[
M/S
]
2.
Set the function selector.
17
Voltge source: [
CV
]
Current source: [
CC
]
3.
Press the [
SHIFT
] button.
Confirm that is displayed
4.
Press the [
MODE
] button as necessary to select the source mode on the secondary display.
• Voltage source
(RAMP±1.5 V or
RAMP±15 V)
• Current source
RAMP±25 mA
During RAMP function setup, is displayed.
18
1.5 Using Memory Function (SCAN and RAMP)
RAMP function factory default settings
±1.5000 V ±15.000 V ±25.000 mA
Output value
Number of devision
Output value
Number of devision
Output value
Number of devision
Start
-1.5000 V
End
+1.5000 V
015
015
-15.000 V
+15.000 V
015
015
Do you want to change the values from the factory default settings?
+25.000 mA
-25.000 mA
025
025
❖
"Changing Data
Settings" (p.20)
5.
Select the RAMP source mode on the secondary display.
With [ W ] or [ X ] button
Cont:
After the first slope is output (start to end values), the second slope is output to return to the start value, forming one repeatable cycle.
The interval of the first slope = (end output value - start output value) , divisions.
Example:
The initial interval setting with ±1.5000 V range is (1.5 + 1.5) / 15 steps = 0.2 V.
The interval of the second slope = (starting output value - end output value) / divisions.
CyCLE:
After the first slope is output (start to end values), the starting value of the second slope is output, continuing output according to the settings.
1.5 Using Memory Function (SCAN and RAMP)
6.
Press the [
OUTPUT
] button to output the set voltage or current value.
Verify that displayed.
19
is
The output value changes according to settings every
0.33 seconds.
7.
Press the [
OUTPUT
] button again to disable output.
Verify that is no longer displayed.
20
1.5 Using Memory Function (SCAN and RAMP)
Changing Data Settings
1.
With the RAMP function active, hold the [
MODE
] button for one second to enter the data setting mode for each source step.
2.
The RAMP function involves dual slope generation.
• Set the number of divisions from the start to the end position, and from the end to the start position.
• Also set the output values of the start and end positions.
Secondary display Output value
Leftmost digit of primary display
Start [
S
] or End [
E
] position
Rightmost three digits of primary display
Number of divisions
3.
Press the [ S ] or [ T ] button to select the start position [
S
] or end position [
E
]. Then press the [
MODE
] button to switch between the number of divisions and output value. The parameter digit to be set blinks.
• Press the [
S
] or [
T
] button to increment or decrement the value, and press the [ W ] or [ X ] button to move to another digit to set.
• The number of divisions can be set from 1 to 999.
4.
Press the [
OUTPUT
] button to save your settings.
5.
Hold the [
MODE
] button for one second to exit the step data setting mode.
2.1 Basic Measurement Procedure (Common Procedure)
Measurement
Function
21
Chapter 2
2.1 Basic Measurement Procedure
(Common Procedure)
1.
Set the slide switch to the[
M
] or [
M/S
] position.
• Setting [
M
] is for measurement only.
• Setting [
M/S
] is for measuring and sourcing at the same time.
22
2.1 Basic Measurement Procedure (Common Procedure)
2.
Set the function selector to the following position.
(ex. Measuring voltage)
Measuring voltage
Measuring current
Measuring frequency
Measuring resistance
Measuring diode
Measuring temperature
[
[
Voltage: [
Current: [
[
[
Ω
]
[
mV
]
]
/ V
] or [
mV
]*
/ mA
]
/
/ V
] or [
mV
]
mA
]
* Setting
Setting
[
[
mV
] is for 50 mV and 500 mV range measurement.
/ V
] is for the range over 500 mV measurement.
3.
Connect the test leads to the INPUT terminals.
Red *: (+) INPUT terminal
Black*: (-) INPUT terminal
*Serves as a temperature probe during temperature measurement
The steps up to here are common to all functions.
Follow the appropriate separate procedure from here.
2.2 Measuring Voltage
2.2 Measuring Voltage
Procedure
1.
Set the slide switch.
[
M
] or [
M/S
]
2.
Set the function selector.
[
/ V
] or [
mV
]
23
3.
Connect the test leads to the terminals.
Red : (+) INPUT terminal
Black: (-) INPUT terminal
4.
Press the [
AC/DC
] button as needed to select the appropriate voltage measurement function.
5.
To enable manual range selection, press the [
RANGE/AUTO
] button.
is not displayed on the LCD.
To re-enable auto-ranging, hold the [
RANGE/AUTO
] button for one second.
6.
Connect the test leads to the circuit to be measured, and read the displayed value.
24
2.2 Measuring Voltage
• Press the [
DUAL
] button to activate the
Dual Display mode.
• Press the [
Hz
] button to activate frequency measurement (or duty cycle or pulse width) in the Dual Display mode.
❖
Details
→ 3.5 "Dual Display Function"
(p.41)
2.3 Measuring Current
2.3 Measuring Current
Procedure
1.
Set the slide switch.
[
M
] or [
M/S
]
2.
Set the function selector.
[
/ mA
]
25
3.
Connect the test leads to the terminals.
Red : (+) INPUT terminal
Black: (-) INPUT terminal
4.
Press the [
AC/DC
] button as needed to select the appropriate current measurement function.
5.
To enable manual range selection, press the [
RANGE/AUTO
] button is not displayed on the LCD.
To re-enable auto-ranging, hold the [
RANGE/AUTO
] button for one second.
6.
Connect the test leads to the circuit to be measured, and read the displayed value.
26
2.3 Measuring Current
• Press the [
DUAL
] button to activate the
Dual Display mode.
• Press the [
Hz
] button to activate frequency measurement (or duty cycle or pulse width) in the Dual Display mode.
❖
Details
→
3.5 "Dual Display Function"
(p.41)
DC Current
Display
AC Current
Display
"4-20 mA%" or "0-20 mA%"
Display
AC+DC Current
Display
2.4 Measuring Frequency
2.4 Measuring Frequency
Procedure
1.
Display voltage or current.
27
2.2 "Measuring Voltage" (p.23) or procedure 1. to 4.
of 2.2 "Measuring
Voltage" (p.23).
2.
Press the [
Hz
] button and read the displayed voltage.
3.
ÅA
Hold the [
Hz
] button repeatedly toggles through Hz, duty cycle and pulse measurement.
Hz
Measurement
Duty Cycle
Measurement
Pulse
Measurement
Press the button for one second or more.
Selecting the Trigger Slope
4.
Hold the [
Hz
] button for one second to return to the status in Step 1.
28
2.5 Measuring the Resistance
2.5 Measuring the Resistance
Procedure
1.
Set the slide switch.
[
M
] or [
M/S
]
2.
Set the function selector.
[
Ω
]
3.
Connect the test leads to the terminals.
Red : (+) INPUT terminal
Black: (-) INPUT terminal
4.
Connect the test leads to the circuit to be measured, and read the displayed value.
5.
To enable manual range selection, press the [
RANGE/AUTO
] button.
is not displayed on the LCD.
To re-enable auto-ranging, hold the [
RANGE/AUTO
] button for one second.
6.
2.5 Measuring the Resistance
Pressing the [
AC/DC
] button toggles the continuity test function on and off.
At this time, autoranging is disabled.
The
29
mark is displayed/not displayed.
30
2.5 Measuring the Resistance
2.5.1 Continuity Test
Procedure
1.
When the resistance measuring function is active, pressing the [
AC/
DC
] button enables the continuity test function using resistance measurement.
The mark is displayed.
At this point, the 500
Ω range is selected, and if autoranging was active, it is disabled and manual ranging is enabled.
• Pressing the [
RANGE
] button selects each range.
• The beeper sounds whenever the resistance is 1000 counts or less in the selected range.
2.
Press the [
AC/DC
] button again to exit the continuity testing function.
2.6 Diode Test
31
2.6 Diode Test
Procedure
1.
Set the slide switch.
2.
Set the function selector.
[
M
] or [
M/S
]
[ ]
3.
Connect the test leads to the terminals.
Red : (+) INPUT terminal
Black: (-) INPUT terminal
4.
Connect the test leads to the circuit to be measured.
The display appears as follows:
Forward direction
0.3 to 0.8 V forward voltage drop displayed
Reverse direction
[
OL
] displayed.
Short-circuit
Approx. 0 V displayed, and beeper sounds
Open-circuit
[
OL
] displayed in forward direction
32
2.7 Measuring Temperature
2.7 Measuring Temperature
Temperature can be measured using an optional temperature probe.
Procedure
1.
Set the slide switch.
[
M
] or [
M/S
]
2.
Set the function selector.
[
mV
]
3.
Connect the temperature probe to the terminals
(+) INPUT terminal
(-) INPUT terminal
4.
Hold the [
TEMP
] button for one second to start temperature measurement.
33
Special Measurement
Functions
Chapter 3
This instrument provides the following special measurement functions:
Recording
Data Hold and Refresh Hold
Relative (relative display)
1-ms Peak Hold
Dual Display
❖
3.1 (p.34)
❖
3.2 (p.36)
❖
3.3 (p.38)
❖
3.4 (p.39)
❖
3.5 (p.41)
34
3.1 Recording Function
3.1 Recording Function
During measurement, maximum (MAX), minimum
(MIN) and average (AVG) values can be automatically recorded in memory.
In what situations is it used?
For acquiring long-term data and checking for transients.
Procedure
1.
While measuring, hold the [
HOLD/
MAX MIN
] button for one second.
The indicators appear on the display as the recording function starts.
2.
Press the [
HOLD/MAX MIN
] button to change the display as follows:
Display indication
Real-time measurement value
Current measurement value
Maximum value
Minimum value
Average value
3.1 Recording Function
35
3.
The
Hold the [
HOLD/MAX MIN
] button again for one second to stop the recording function.
indicators disappear from the display.
• The beeper sounds and the relevant value is refreshed when a new maximum or minimum value is recorded.
• When an overload value is measured, calculation of the average value stops, and
[
OL
] is displayed for the average value.
• When the recording function is started with auto-ranging enabled, the maximum
(MAX), minimum (MIN) and average
(AVG) values can be in different ranges.
36
3.2 Data Hold / Refresh Hold Function
3.2 Data Hold / Refresh Hold Function
In what situations is it used?
Use to simplify reading fluctuating measurements.
Data Hold Function
Procedure
This function holds measured values on the display.
1.
Press the [
HOLD/MAX MIN
] button to activate the Data Hold function.
appears on the display, and the displayed value is held
(retained).
2.
Press the [
HOLD/MAX MIN
] button again to measure one time and hold the measurement data.
3.
Press the [
HOLD/MAX MIN
] button for one second to cancel the Data
Hold function.
disappears from the display.
3.2 Data Hold / Refresh Hold Function
Refresh Hold Function
37
Procedure
• This function automatically refreshes the (held) displayed value when the measured value changes while Data Hold is active.
• The Data Hold function can be changed to the
Refresh Hold function in the Power-On Options.
❖
7016 instruction manual 5.4 Power-On Options
1.
Press the [
HOLD/MAX MIN
] button to activate the Refresh Hold function.
appears on the display.
1.
Press the [
HOLD/MAX MIN
] button for one second to cancel the
Refresh Hold function.
disappears from the display.
38
3.3 Relative (relative display) Function
3.3 Relative (relative display) Function
The function applies to all measurement functions, subtracting a reference value from the measured value and displaying the calculated result.
In what situations is it used?
• When you want to zero out the internal resistance of probes
• When you want to read only the difference (offset) of a measured value from another value
• When you want high-accuracy measurement in the 50 mV DC or DCA ranges
Procedure
1.
Press the [
REL
] button to activate the Relative function.
appears on the display.
At this point,
• The displayed value becomes the reference for subsequent measurements, and is redisplayed as zero.
• If measured data is at overload level, [OL] appears on the display.
2.
Press the [
REL
] button again to cancel the Relative function. disappears from the display
3.4 1-ms Peak Hold Function
39
3.4 1-ms Peak Hold Function
• Works with voltage and current measurement functions
• Shows the peak value of an input signal waveform
In what situations is it used?
Use to determine the crest factor of a waveform, where
crest factor = waveform peak value / true rms value
Procedure
1.
Hold the [
AC/DC
] button for one second to activate the 1-ms Peak
Hold function.
appears on the display.
At this point,
• The measurement displayed when the function was activated is held (retained).
• If auto-ranging was enabled, manual ranging is activated with the currently operative range retained.
40
3.4 1-ms Peak Hold Function
• Pressing the [
HOLD/MAX MIN
] button updates the displayed (held) MAX and
MIN values.
Display indication
Measurement value
Maximum value
Minimum value
• The beeper sounds when a new maximum or minimum value is measured.
• Pressing the [
DUAL
] button resets the
MAX and MIN values and restarts 1-ms
Peak Hold operation.
The maximum (MAX) and minimum (MIN) display mode are not changed.
2.
Press the [
AC/DC
] button for one second to cancel the 1-ms Peak
Hold function.
disappears from the display.
3.5 Dual Display Function
3.5 Dual Display Function
41
Procedure
This function simultaneously monitors two different measurement parameters (voltage, current, frequency, duty cycle or pulse width) of one signal.
1.
Press the [
DUAL
] or [
Hz
] button to activate dual display.
(1) Pressing the [
DUAL
] button selects the Dual
Display mode
Pressing it repeatedly changes the display as follows.
Function
Selector
Position
/ mV mA
V
Primary
Display
Secondary Display
Voltage (V) measurement
Voltage (mV) measurement
Current (mA) measurement
Frequency measurement
Source function display
Frequency measurement
Source function display
Frequency measurement
Source function display
42
3.5 Dual Display Function
(2) Press the [
Hz
] button to activate frequency measurement (or duty cycle or pulse width) in the
Dual Display mode.
Pressing it repeatedly changes the display as follows.
Function
Selector
Position
Primary
Display
Secondary Display
/ mV mA
V
Frequency
Duty cycle
Pulse width
Frequency
Duty cycle
Pulse width
Frequency
Duty cycle
Pulse width
Voltage (V) measurement
Voltage (mV) measurement
Current (mA) measurement
• When measuring frequency or pulse width, pressing the [RANGE/AUTO] button toggles between auto- and manual ranging.
• When measuring duty cycle or pulse width, holding the [AC/DC] button for one second toggles the trigger slope between and .
3.5 Dual Display Function
2.
To cancel the Dual Display mode
• When pressed [
DUAL
] button at procedure 1. :
Press [
DUAL
] button again.
• When pressed [
Hz
] button at procedure 1. :
Hold the [
Hz
] button for one second.
43
44
3.5 Dual Display Function
advertisement
Related manuals
advertisement
Table of contents
- 5 Introduction
- 8 Basic Sourcing Procedure
- 10 Constant-Voltage Source (DCCV)
- 12 Constant-Current Source (DCCC)
- 14 Pulse Source (PULSE)
- 16 (SCAN and RAMP)
- 16 1.5.1 Using the SCAN Function
- 21 1.5.2 Using the RAMP Function
- 25 (Common Procedure)
- 27 Measuring Voltage
- 29 Measuring Current
- 31 Measuring Frequency
- 32 Measuring the Resistance
- 34 2.5.1 Continuity Test
- 35 Diode Test
- 36 Measuring Temperature
- 38 Recording Function
- 40 Data Hold / Refresh Hold Function
- 42 Relative (relative display) Function
- 43 1-ms Peak Hold Function
- 45 Dual Display Function