Battery Analyzer BA5100
Bat t ery Anal yzer BA5100
CHARGE
DISCHARGE
EIS
CYCLING
WITH
EIS
Features
-
Programmable battery charger of 5V/100A
-
Programmable load 100A/400W continuous and peak pulse current up to
300A/1s for hybrid pulse power characteristic test
-
Impedance response analyzer for EIS measurements; 0.05Hz to 10KHz
-
Battery cycling test for estimating SoC and SoH
-
Impedance measurements; Nyquist plot, terminal resistance, charge
transfer resistance, diffusion capacitance and cell's inductance
-
Simultaneously measurement of voltage, current, temperature and
impedance
-
Precision coulomb counter for battery pack gas gauge calibration
-
EIS Analog Interface for external high power charger and load for stack
measurements
-
Interface for external DC contactor in order to isolate the battery under test
-
Ultra-capacitor measurement
-
Standard Ethernet, USB and RS-232 interface for application software
Specif icat ions
CHARGE
Programming Voltage
Range
2 ~ 5V
EIS (Electrochemical Impedance Spectroscope)
Impedance
Frequency Range
0.05 ~ 10KHz
Resolution
1mV
Resolution
0.01Hz
Accuracy
± 0.05% +0.1%
Amplitude
Programming Current
Range
Galvano 0.1% to 20%
of the current range
0 ~ 100A
Accuracy
0.2% , 0.1° with
auto-sampling interval
Resolution
10mA
Connection
4-terminals
Accuracy
± 0.05% +0.1%
Cutoff Condition
Time, CC to CV+Time,
Temp, Capacity
GENERAL
User Interface
Display
20x2 characters OLED
Input
16Keys and rotary
Remote Interface
Ethernet, USB, RS232
DISCHARGE
Programming Current
Range
0 ~ 100A,
Pulse 0 ~ 300A/ 1s
Resolution
10mA
Mechanical and Environmental
Dimension
44H x 445W x 568L mm
Accuracy
± 0.05% +0.1%
Weight
10KG
Cutoff Condition
Time, Vfinal, Capacity
Operating Temperature
0 ~ 40?
Mode
Continuous, List, Pulse
Storage Temperature
-30 ~ 70?
Power Requirement
Voltage
90 ~ 250VAC
MEASUREMENT
Voltage
Range
0 ~ 5.5V
Frequency
50/ 60Hz
Resolution
1mV
Watt
700W @ 100A charging
Accuracy
± 0.05% +0.1%
Current
Range
-300A ~ +100A
SOFTWARE
Charge/ Discharge
Test Mode
Resolution
10mA
Accuracy
± 0.05% +0.1%
Temperature
Range
-40 ~ 100?
Resolution
0.1?
Accuracy
± 1?
Coulomb
Range
-200 ~ +200Ah
Resolution
0.001Ah
Accuracy
±0.2%
Parameter Setting
EIS
Display Mode
I-V, Cycle life,
Charge/ discharge rate
CV charge,
CC charge,
CC discharge,
Pulse current discharge,
List mode for
charge/ discharge,
Cutoff condition
Nyquist, Meter readout
of voltage, current,
Z(real) and Z(imaginary)
Design and specifications subject to change without notice
Biol ogic Technol ogy Inc.
5F., No. 93, Xinhu First Rd., Neihu Dist., Taipei City,
Taiwan, 114
www.biologictek.com.tw
+886-2-2792-6976
[email protected]