OPERATING INSTRUCTION
Model MTP-3095
AC True RMS CLAMP METER
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates that the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present
Double insulation
SAFETY NOTES
Do not exceed the maximum allowable input range of any function
Do not apply voltage to meter when resistance function is selected.
Set the function switch to OFF when the meter is not in use.
WARNINGS
Set function switch to the appropriate position before measuring.
When measuring volts do not switch to current/resistance modes.
When changing ranges using the selector switch
, always disconnect the test leads from the circuit under test.
Do not exceed the maximum rated input limits.
2
CAUTIONS
Improper use of this meter can cause damage, shock, injury or death. Read and understand this user manual before operating the meter.
Always remove the test leads before replacing the battery.
Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard.
Remove the battery if the meter is to be stored for long periods.
Always discharge capacitors and remove power from the device under test before performing Diode, Resistance or
Continuity tests.
Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not "live".
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Input Limits
A AC
V DC, V AC
1000A
1000V DC/AC
3
Frequency, Resistance, Diode,
Continuity , Capacitance Test
Temperature (°C/°F)
Meter Description
1000V DC/AC
1000V DC/ AC
1. Current clamp
2. Non-contact AC voltage indicator light
3. Clamp trigger
4. Data Hold and Backlight button
5. Rotary Function swith
6. Hz% Hold button
7. RANGE select button
8. Relative button
9. MIN/MAX Hlod button
10. MODE select button
11. LCD display
12. COM input jack
13. V ΩCAP TEMPHz jack
14. Battery Cover
1. AC DC AC(alternating current) and DC (direct currrent)
2. Minus sign
3. 8.8.8.8 6000 count (0 to 5999) measurement reading with 61segments
Analog Bargraph
4. AUTO AutoRange mode
5. REL Relative mode
6. Diode test mode
7. •))) Audible Continuity
4
8. HOLD Data Hold mode
9. °C,°F,
,m,V,A,K,M,Ω, Units of measure list
10. MAX MIN MAX/MIN mode
11. Hz % Frequency/duty cycle mode
Specifications
Function
Range &
Resolution
Accuracy (% of reading)
AC Current
(50/60Hz)
AC Current
(400Hz)
DC Voltage
AC Voltage
(50-400Hz)
Resistance
600.0 AAC
1000 AAC
600.0 AAC
1000 AAC
600.0 mVDC
6.000 VDC
60.00 VDC
600.0 VDC
6.000 VAC
60.00 VAC
600.0 VAC
600.0 Ω
6.000KΩ
60.00KΩ
600.0KΩ
6.000MΩ
± (2.8 % + 8 digits)
± (3.0 % + 8 digits)
± (5.0 % + 10 digits)
± (5.0 % + 10 digits)
± (0.8% + 2 digits)
± (1.5% + 2 digits)
± (1.8% + 8 digits)
± (1.0% + 4 digits)
± (1.5% + 2 digits)
± (2.5% + 3 digits)
± (3.5% + 5 digits)
±(4.0% reading + 20 digits)
Capacitance
60.00MΩ
40.00nF
400.0nF
4.000μF
40.00μF
400.0μF
4000μF
±(3% reading + 5 digits)
±(4.0% reading + 10 digits)
±(5.0% reading + 10 digits)
Frequency
Sensitivity:15Vrms
Temp
10-10kHz ±(1.5% reading + 2 digits)
-20.0 to 760.0
C ±(3%rdg+5C)
5
(type-K)
(probe accuracy not included)
-4.0 to1400.0
F ±(3%rdg+9F)
Clamp size Opening 1.5" (40mm) approx
Diode Test Test current of 0.3mA typical;
Open circuit voltage 1.5V DC
Continuity Check typical.
Threshold <40
;
Test current < 0.5mA
Low Battery Indication “ ” is displayed
Over range Indication
“OL” is displayed
Measurements Rate 2 per second, nominal
Input Impedance
Display
AC Current
10M
(VDC and VAC)
6000 counts LCD
50-400Hz (AAC)
AC Voltage bandwidth
50-400Hz (VAC)
Operating Temperature
41 to 104 o
F (5 to 40 o
Storage Temperature -4 to 140 o
C)
F (-20 to 60 o
C)
Operating Humidity
Max 80% up to 87
F (31C) decreasing linearly to 50% at
104
F(40C)
Storage Humidity
<80%
Operating Altitude
Over voltage
7000ft.(2000meters)maximum.
Category 600V
15
Dimensions/Weight 232x77x39mm/271g
Safety For indoor use and in accordance with Overvoltage
Category II, Pollution Degree
2. Category II includes local
6
level, appliance, portable equipment, etc., with transient overvoltages less than
Overvoltage Cat. III
Operation
NOTICES: Read and understand all warning and
precaution statements listed in the safety section of this operation manual prior to using this meter. Set the function select switch to the
OFF position when the meter is not in use.
AC Current Measurements
WARNING: Ensure that the test leads are disconnected from the meter before making current clamp measurements.
1. Set the Function switch to the 1000 or 600A range.
2. If the range of the measured current is not known, select the higher range first then move to the lower range if necessary.
3. Press the trigger to open jaw.
Fully enclose one conductor to be measured.
Read the current measurement on LCD display.
DC/AC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the red test lead into the positive V terminal.
2. Set the function switch to the V position.
3. Select AC or DC with the MODE button.
7
4. Connect the test leads in parallel to the circuit under test.
5. Read the voltage measurement on the LCD display.
Resistance Measurements
1.
Insert the black test lead into the negative COM terminal and the red test lead into the positive terminal.
2.
Set the function switch to the
position.
3.
4.
Touch the test probe tips across the circuit or component under test. It is best to disconnect one side of the device under test so the rest of the circuit will not interfere with the resistance reading.
For Resistance tests, read the resistance on the LCD display.
Diode and Continuity Measurements
1. Insert the black test lead banana plug into the negative
COM jack and the red test lead banana plug into the positive diode jack.
2. Turn the rotary switch to the •))) position.
3. Press the MODE button until “ ” appears in the display.
4. Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to 0.7V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0mV and an open device will indicate “OL” in both polarities.
Red Black Black Red
Probe Probe Probe Probe
Forward test Reverse test
For Continuity tests, if the resistance is < 40
, a tone will sound.
8
Capacitance Measurements
WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any capacitance measurements. Remove the batteries and unplug the line cords.
1. Set the rotary function switch to the cap position.
2. Insert the black test lead banana plug into the negative
(COM) jack. Insert the red test lead banana plug into the positive (V) jack.
3. Touch the test leads to the capacitor to be tested.
4. Read the capacitance value in the display
Frequency/duty cycle Measurements (electronic)
1. Set the rotary function switch to “V” position.
2. Select with
3. Press the Hz/% button to indicate “Hz” in the display.
4. Insert the black lead banana plug into the negative COM jack and the red test lead banana plug into the positive
Hz jack.
5. Touch the test probe tips to the circuit under test.
6. Read the frequency on the display.
7. Press Hz/% button again to indicate “%” on the display. Read the % of duty cycle on the display.
Temperature Measurements
WARNING: To avoid electric shock, disconnect both test probes from any source of voltage before making a temperature measurement.
1. Set the function switch to TEMP.
2. Insert the Temperature Probe into the negative (COM) and the V jacks, making sure to observe the correct polarity.
C or F with the MODE button.
4. Touch the Temperature Probe head to the part whose temperature you wish to measure. Keep the probe touching the part under test until the reading stabilizes
9
(about 30 seconds).
5. Read the temperature on the display. The digital reading will indicate the proper decimal point and value.
WARNING: To avoid electric shock, be sure the thermocouple has been removed before changing to another measurement function
Non-Contact AC Voltage Measurements
WARNING: Risk of Electrocution. Before use, always test the Voltage Detector on a known live circuit to verify proper operation
1. Touch the probe tip to the hot conductor or insert into the hot side of the electrical outlet.
2. If AC voltage is present, the detector light will illuminate.
NOTE: The conductors in electrical cord sets are often twisted. For best results, rub the probe tip along a length of the cord to assure placing the tip in close proximity to the live conductor.
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy may randomly trip the sensor. This is normal operation
MODE BUTTON
To select DC/ACV,OHM/ Diode/Continuity/CAP ,
C/F
MIN & MAX Recording Button
Press MIN/MAX to enter MIN, MAX Recording mode
(manual range only ). Select the proper range before selecting MIN MAX to ensure that the MIN/MAX reading will not exceed the testing range. Press once to select MIN.
Press again to select MAX. and press again release
MIN/MAX recording function.
10
REL BUTTON
For Offset adjustment.
Data Hold Button
To freeze the LCD meter reading, press the data hold button. The data hold button is located on the left side of the meter (top button). While data hold is active, the HOLD display icon appears on the LCD. Press the data hold button again to return to normal operation.
Note: The HOLD feature will activate when the backlight is turned on. Press the HOLD key again to exit Hold.
RANGE BUTTON
When the meter is first turned on, it automatically goes into
Auto Ranging. This automatically selects the best range for the measurements being made and is generally the best mode for most measurements. For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE button. The “Auto Range” display indicator will turn off, The “Manual Range” display indicator will turn on
2. Press the RANGE button to step through the available ranges until you select the range you want.
3. Press and hold the RANGE button for 2 seconds to exit the Manual Ranging mode and return to Auto Ranging.
Backlight Button
The backlight function illuminates the display and is used when the ambient light to too low to permit viewing of the displayed readings. Press the button for two seconds to
11
turn the backlight on and press the button a second time to turn the backlight off.
Battery Replacement
1. Remove the one rear Phillips head screw
2. Open the battery compartment
3. Replace the Requires One 9V battery
4. Re-assemble the meter
MTP Warranty
MTP Instruments warrants this instrument to be free of
defects in parts and workmanship for one (1) year from date of shipment. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. MTP
Instruments specifically disclaims any implied warranties
or merchantability or fitness for a specific purpose and will not be held liable for any direct, indirect, incidental or consequential damages. MTP Instruments total liability is limited to repair or replacement of the product.
12
MTP
Instruments
4409, Charleroi Street
Montreal-North, Que
Canada
H1H 1T6
Tel: (514) 326-7167 Fax:(514) 326-7835
Toll Free Number: 1 888 326-7167
WebSite: www.mtpinc.com