HM8115-2 - Electrocomponents
HM8115-2
8 kW Power Meter
HM8115-2
HZ815 Power adapter
Wide measurement range 1 mW…8 kW
Voltage range 100 mV…500 V, current range 1 mA…16 A
Frequency range DC…1 kHz
Simultaneous voltage, current and power display
RMS value
Display of apparent, effective and reactive power
Power factor display
Autoranging, simple operation
Monitor output (BNC) representing the instantaneous active
power
Active power
Suitable for measurements on frequency converters
Software for remote control and data acquisition included
Galvanically isolated USB/RS-232 Interface, optional IEEE-488
8 kW Power Meter HM8115-2
All data valid at 23 °C after 30 minute warm-up
Voltage
Ranges:
Resolution:
Accuracy:
Input impedance:
Crest factor:
Input protection:
Current
Ranges:
Resolution:
Accuracy:
Crest factor:
Input protection:
True RMS voltage measurement (AC+DC)
50 V
150 V
500 V
0.1 V
1V
1V
20 Hz…1 kHz:
± (0.4 % + 5 digit)
DC:
± (0.6 % + 5 digit)
1 MΩ II 100 pF
max. 3.5 at full scale
max. 500 Vp
True RMS current measurement (AC+DC)
160 mA
1.6 A
16 A
1 mA
1 mA
10 mA
20 Hz…1 kHz:
± (0.4 % + 5 digit)
DC:
± (0.6 % + 5 digit)
max. 4 at full scale
fuse, FF 16 A 6.3 x 32 mm (superfast)
Miscellaneous
Safety Class:
Power supply:
Power consumption:
Operating temperature:
Storage temperature:
Max. rel. humidity:
Dimensions (W x H x D):
Weight:
Safety Class I (EN 61010)
115/230 V ± 10 %, 50/60 Hz, CAT II
approx. 15 W at 50 Hz
+5°C...+40°C
-20°C...+70°C
5%...80% (non condensing)
285 x 75 x 365 mm
approx. 4 kg
Accessories supplied: Operator’s Manual, power cable, software
Optional accessories:
HZ10S/R Silicone test lead
HZ815 Socket adapter
HO880 IEEE-488 (GPIB) Interface (galvanically isolated)
Active power measurement
Ranges:
8W
24 W
80 W 240 W 800 W 2400 W 8000 W
Resolution:
1 mW 10 mW 10 mW 100 mW 100 mW
1W
1W
Accuracy:
20 Hz…1 kHz:
± (0.8% + 10 digit)
DC:
± (0.8% + 10 digit)
Display:
4-digit, 7-segment LED
Reactive power measurement
Ranges:
8 var
24 var
80 var 240/800 var
2400/8000 var
Resolution: 10 mvar 100 mvar 100 mvar
1 var
1 var
Accuracy:
20…400 Hz: ± (2.5 % + 10 digit + 0.02 x P)
P = active power
Display:
4-digit, 7-segment LED
Apparent power measurement
Ranges:
8 VA
24 VA
80 VA 240/800 VA
2400/8000 VA
Resolution:
1 mVA 10 mVA 10 mVA
100 mVA
1 VA
Accuracy:
20 Hz…1 kHz:
± (0.8 % + 5 digit)
Display:
4-digit, 7-segment LED
Power factor measurement
Display:
0.00…+1.00
Accuracy:
50 Hz…60 Hz:
± (2% + 3 digit) (sine wave)
voltage and current › 1/10 of full scale
Monitor output (analog)
Connection:
BNC connector (galvanic isolation
to test circuit and RS-232 interface)
Reference potential:
protective earth
Level:
1 VAC at full scale (2400/8000 digit)
Accuracy:
typ. 5 %
Output impedance:
approx. 10 kΩ
Bandwidth:
DC…1 kHz
Protected up to:
± 30 V
Functions and displays
Measurement functions:
Range selection:
Overrange alarm:
Display resolution
Voltage:
Current:
Power:
Power factor:
Interface
Interface:
Connection RS-232:
Protocol:
Data rate:
Functions:
voltage, current, power, power factor
automatic/manual
visual and acoustic
3-digit, 7-segment LED
4-digit, 7-segment LED
4-digit, 7-segment LED
3-digit, 7-segment LED
USB/RS-232 (HO820), IEEE-488 (option)
D-sub connector (galvanic isolation
to test circuit and monitor output)
Xon / Xoff
9600 Baud
control/data fetch
w w w. h a m e g . co m
HM8115-2E/260309/ce · Subject to changes · © HAMEG Instruments GmbH® · DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 800 0 · Fax +49 (0) 6182 800 100 · www.hameg.com · [email protected]
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement