Directions for use

Models 5002, 5122, 5702, 6002

© 2016 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.

Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.

Software in this product is Copyright 2016 Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded.

The software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors.

For information about any Welch Allyn product, contact your local Welch Allyn representative: http:// www.welchallyn.com/en/other/contact-us.html

.

This manual applies to 901109 MASTED SCALES.

725556

DIR 80021615 Ver. A Revision date: 2016–07

Welch Allyn, Inc.

4341 State Street Road

Skaneateles Falls, NY 13153 USA www.welchallyn.com

Contents

Introduction ............................................................................................. 1

Intended use ........................................................................................................ 1

Indications for use ................................................................................................ 1

Symbols ............................................................................................................... 1

About warnings and cautions ............................................................................... 2

Controls and indicators ........................................................................................ 3

Assembly and setup ................................................................................ 7

Models 5002 and 5702 stand-on scales .............................................................. 7

Model 5122 stand-on scale .................................................................................. 8

Model 6002 wheelchair scale .............................................................................. 8

Custom setup ...................................................................................................... 8

Scale operation ...................................................................................... 11

Power options .................................................................................................... 11

Audible beeps .................................................................................................... 11

Weighing procedure ........................................................................................... 12

Wheelchair weighing procedure (model 6002) .................................................. 12

Print a weight measurement ............................................................................. 13

Storage, cleaning, and maintenance ................................................... 15

Storage ............................................................................................................... 15

Cleaning ............................................................................................................. 15

Battery replacement .......................................................................................... 15

Change the printer paper ................................................................................... 16

Maintenance ...................................................................................................... 17

Troubleshooting ................................................................................................. 17

Appendices ............................................................................................ 19

Specifications ..................................................................................................... 19

Product configurations ....................................................................................... 20

Approved accessories ........................................................................................ 21

Warranty ............................................................................................................ 22

iii

iv Contents Welch Allyn Masted Scales

Introduction

Intended use

The Welch Allyn Masted Scales are intended to be used by clinicians for weighing patients up to 660 pounds (300 kg) for model 5122, up to 880 pounds (400 kg) for models 5002 and 6002, and up to 1000 pounds (454 kg) for model 5702.

Indications for use

The Welch Allyn Masted Scales are used by clinicians to weigh patients. An optional height gauge is available on some models.

Symbols

Documentation symbols

WARNING The warning statements in this manual identify conditions or practices that could lead to illness, injury, or death. Warning symbols will appear with a grey background in a black and white document.

Consult Directions for Use

Caution The caution statements in this manual identify conditions or practices that could result in damage to the equipment or other property, or loss of data.

Shipping, storing, and environment symbols

Temperature limit

Keep dry

Fragile

Separate collection of Electrical and

Electronic Equipment. Do not dispose as unsorted municipal waste.

1

2 Introduction

Humidity limitation

This way up

Recyclable

Welch Allyn Masted Scales

Atmospheric pressure limitation

Stacking limit by number

Miscellaneous symbols

Manufacturer Reorder Number

Product Identifier For indoor use only

Serial Number

Battery

Prescription only or "For Use by or on the order of a licensed medical professional"

Power

About warnings and cautions

Caution statements can appear on the Welch Allyn Masted Scales, the packaging, the shipping container, or in this Directions for use.

The Welch Allyn Masted Scales are safe for patients and clinicians when used in accordance with the instructions and caution statements presented in this Directions for

use.

Before using the device, you must familiarize yourself with all cautions, with the steps to power up the device, and with the sections of this Directions for use that pertain to your use of the device. In addition to reviewing the general cautions presented in the next section, you must also review the more specific cautions that appear throughout the manual in conjunction with setup/startup, operation, and maintenance tasks.

• Failure to understand and observe any warning statement in this manual could lead to patient injury or illness.

• Failure to understand and observe any caution statement in this manual could lead to damage to the equipment or other property, or loss of patient data.

Directions for use Introduction 3

General warnings and cautions

WARNING Patient injury risk. Make sure the scale is properly cleaned between patients.

WARNING Patient or operator injury risk. Make sure the scale is positioned on a stable surface and in a way to avoid shock or bumping hazards.

CAUTION Make sure that you routinely perform general maintenance and equipment safety checks on your scale. Remove the scale from service when you notice damaged power cords.

CAUTION Make sure the scale is set in the proper measurement units for your facility. Make sure you record the measurement units as they are displayed.

CAUTION Do not use this scale to transport patients or items.

Controls and indicators

Scale

Models 5002 and 5702

3

4

1

2

Steering handle/hand grip

Front panel

Mast

Hand rail

4 Introduction

5

6

Wheels

Weighing platform

N/A Battery access door (located on the rear of the mast)

N/A Height gauge (optional)

Model 6002

Welch Allyn Masted Scales

1 Steering handle/hand grip

2 Control panel

3 Mast

4 Hand rail (optional)

5 Wheels

6 Weighing platform

7 Wheelchair ramp

N/A Battery access door (located on the rear of the mast)

N/A Height gauge (optional)

Directions for use

Model 5122

1 Control panel

2 Mast (battery access door located on front)

3 Weighing platform

Introduction 5

6 Introduction

Front panel buttons

Welch Allyn Masted Scales

Button

ZERO

ON

ST

RECALL

LB

Function

Press this button to reset the scale to zero.

Press this button to turn on the scale.

Press this button in conjunction with

ON

to enter the setup mode.

Press this button to recall the previously stored scale reading.

Press this button to display the weight in pounds.

Note

This button is not present if you purchased a scale with the kilogram-only option.

KG Press this button to display the weight in kilograms.

PRINT (DATA)

SEND

Press this button to print the current weight.

On models equipped with a RS-232 port, pressing this button when the scale is connected to a computer sends the current measurement to the computer.

REWEIGH Press this button to retake the patient's weight.

Assembly and setup

The scale is fully assembled at the factory and tested. In some instances, the scales are shipped assembled. However, due to size and shipping restrictions, most scales are partially disassembled. If assembly is required, it can be done with common hand tools.

When special tools are required, they are supplied with the scale. Ensure that all bolts and fasteners are securely tightened.

Use the following instructions to assemble and setup your scale for use.

Models 5002 and 5702 stand-on scales

1. Carefully remove all parts from the shipping carton and inspect the components for shipping damage.

2. Place the scale base on the floor or on a workbench.

3. Insert the mast into the base. Ensure that the readout display is facing the patient when the patient stands on the scale.

4. Fasten the mast with the bolts and washers provided.

5. Tighten the bolts securely to ensure a solid connection is made.

Weighing deck mounting (model 5702)

1. Position the weighing deck so that the two sets of holes are closest to the mast.

2. Line the weighing deck up with the mounting holes in the aluminum plate.

3. Install the four flat head screws through the weighing deck.

4. Securely tighten the supplied mounting screws with the provided hex key.

Handrail mounting

1. Attach the handrail to the top of the weighing platform using the button head screws and the provided hex key.

Three screws are required for each mounting plate (six total screws).

2. Tighten all six screws securely to ensure a safe and stable handrail.

Height gauge assembly (purchasable option)

If your scale was ordered with a height gauge, follow the installation directions packed with the height gauge.

7

8 Assembly and setup Welch Allyn Masted Scales

Model 5122 stand-on scale

The 5122 model scale is shipped assembled, so no assembly is required. It operates the same as the 5002 model. Handrails are not available for this scale.

Model 6002 wheelchair scale

1. Carefully remove all parts from the shipping carton and inspect the components for shipping damage.

2. Place the scale base on the floor or on a workbench.

3. Insert the mast into the base. Ensure that the readout display is facing the patient when the patient stands on the scale.

4. Fasten the mast with the bolts and washers provided.

5. Tighten the bolts securely to ensure a solid connection is made.

Weighing deck mounting

1. Position the weighing deck so that the two sets of holes under the flaps are closest to the mast.

2. Line the weighing deck up with the mounting holes in the aluminum plate.

3. Install the four flat head screws through the weighing deck.

4. Securely tighten the supplied mounting screws with the provided hex key.

Handrail mounting (purchasable option)

1. Align the mounting feet of the handrail to the underside of the weighing deck.

2. Install and securely tighten the two flat head screws and locking nut on each foot of the handrail.

Height gauge assembly (purchasable option)

If your scale was ordered with a height gauge, follow the installation directions packed with the height gauge.

Custom setup

You can customize the scale to best suit your needs. Configurations that can be selected include the automatic shut-off time, weighing units, weight resolution, and more.

Enter custom setup mode

1. Make sure the scale is powered down.

2. Press and hold ST while pressing ON.

3. Once the scale powers up, press and release ST five times.

The scale displays SEt-UP and produces a series of four long beeps.

Directions for use Assembly and setup 9

Set the options

Enter the custom setup as described.

Press ST to scroll through the custom setup options.

The options displayed with each press are as follows:

Press the setup mode select button to turn these options On or OFF.

Note

Options indicated with an asterisk (*) require an additional press of ST to change the value.

Option displayed Feature

SOFt This option displays the software version of your scale.

dAtE

SCALE

AutOFF*

This option displays the release date of the software. The format is MM/DD/YY.

This option displays the model number of the scale.

This displays the number of seconds before the scale turns off when operating on battery power.

The

Cont

value prevents the scale from turning off automatically.

AC con* rES

This option causes the scale to remain on when plugged in to AC power.

This option allows you to change the resolution of weight. Pressing following options:

ST

switches between the

0.1 pounds / 0.05 kilograms

0.1 pounds / 0.1 kilograms

0.2 pounds / 0.1 kilograms

0.5 pounds / 0.2 kilograms

1 pound / 0.5 kilograms

UnitS* This option allows you to change the displayed weight unit.

Note

Do not change the scale units if you have purchased the kilogram-only option.

KILOS* This option allows the weight to display in kilograms.

POUNDS* This option allows the weight to display in pounds.

bEEPEr* rECALL rS-232

OutPut*

This option determines the audible signal that occurs when a front panel button is pressed.

This option allows you to turn

ON

or

OFF

the

RECALL

button functionality.

Detailed instructions for this feature are available from Welch Allyn. Go to http:// www.welchallyn.com/en/other/contact-us.html

to find your local representative.

This option displays the current data port output option. Detailed instructions for this feature are available from Welch Allyn. Go to http://www.welchallyn.com/en/other/contact-us.html

to find your local representative.

Set the value to

OFF

to disable the optional internal printer or the RS-232 port.

PrtUnt*

This option determines whether pounds or kilograms are displayed on a printout.

If the pounds and kilograms indicators are illuminated on the control panel, the printed output is determined by whichever unit is selected on the control panel.

If the kilogram indicator is illuminated, it will only print in kilograms, regardless of the unit is selected on the control panel.

10 Assembly and setup

PrtOPT* bAud* pArity*

StPbit* rtS.ctS*

Prtbtn* rtEPrt*

PC brd*

SEt-UP

Welch Allyn Masted Scales

If the pounds indicator is illuminated, it will only print in pounds, regardless of the unit is selected on the control panel.

Press

RECALL

to choose between the following options:

PnlPrt

: The current weight is printed when

PRINT

is pressed on the control panel.

AutPrt

: The current weight is automatically printed when a weight reading occurs.

This option allows you to set the baud rate. Values are

1200

,

2400

,

4800

,

9600

, and

19200.

This option allows you to change the parity between the following options:

Off

: No parity, 8 data bits

Odd

: Odd parity, 7 data bits

EuEn

: Even parity, 7 data bits

This option allows you to change the stop bits between the following options:

1 bit

: One stop at the end of a word

2 bits

: Two stops at the end of a word

This option allows you to turn

On

for the printer or

Off

for RS232 serial transmissions.

This option allows you to turn

On

or

Off

the front panel

PRINT

button.

This option allows you to turn

On

or

Off

connecting the receive line to the ground.

the enabling of printing from a remote location by

This option displays the model number of the printed circuit board. A second press of displays the revision of the printed circuit board.

ST

This option is displayed when you have cycled through all the options. Press and hold power down the scale.

ON

to

Scale operation

Power options

The scale can be used with battery power or with AC power.

Battery power

CAUTION Use only size D disposable alkaline batteries. The use of any other battery will void the warranty.

Press ON to turn on the scale.

The scale automatically turns off after 45 seconds of non-use when operating on battery power.

The LO BATT indicator illuminates when the battery power is low.

bAttrY displays when the batteries are unable to properly power the scale.

Replace the batteries when either of these messages are displayed.

AC power

When plugged in, the scale continuously runs on AC power automatically.

The scale switches to battery power when AC power is not present.

Audible beeps

The following beeps occur when using the scale.

Sound

One beep

Meaning

Acknowledgment of a button press.

Two beeps The scale is zeroed and ready to take a measurement.

Three beeps The weight reading is obtained and displayed.

No print, four short beeps

A print was requested, but valid weight is not displayed on the scale, or the printer is in use.

PRINT

was pressed while

No print, four long beeps

There is a problem with the printer. Contact your service department.

11

12 Scale operation Welch Allyn Masted Scales

Four long beeps

Four short beeps

The battery is too weak to operate the scale.

The scale has entered setup mode.

Weighing procedure

1. Press ON to turn on the scale.

The scale sounds one beep.

2. Wait for the scale to display 0.0.

The scale sounds two beeps.

3. Place the patient on the scale and make sure that his or her weight is evenly distributed.

The scale sounds three beeps and the patient's weight is displayed.

4. If necessary, press REWEIGH to perform another weight measurement.

5. If necessary, press RECALL to display the previously stored scale measurement.

Note

A measurement is stored in the scale until a new weight is taken and zeroes out.

A stationary display of - - indicates a negative weight. Make sure that there is nothing on the scale and press ZERO to set the zero.

If O-LOAd is displayed, the capacity of the scale has been exceeded.

Wheelchair weighing procedure (model 6002)

1. Make sure that the scale ramps are down.

2. Press ON to turn on the scale.

The scale sounds one beep.

3. Wait for the scale to display 0.0.

The scale sounds two beeps.

4. Place the empty wheelchair on the scale and make sure the weight is evenly distributed.

The scale sounds three beeps and the empty wheelchair weight is displayed. Write down and save this measurement.

5. Place the patient in the wheelchair and place both on the scale. Make sure the weight is evenly distributed.

The scale sounds three beeps and the patient's weight is displayed.

6. Subtract the measurement obtained in Step 4 from the measurement in Step 5 to get the patient's weight.

7. If necessary, press REWEIGH to perform another weight measurement.

8. If necessary, press RECALL to display the previously stored scale measurement.

Note

A measurement is stored in the scale until a new weight is taken and zeroes out.

Directions for use Scale operation 13

A stationary display of - - indicates a negative weight. Make sure that there is nothing on the scale and press ZERO to set the zero.

If O-LOAd displays, the capacity of the scale has been exceeded.

If OFF SCALE PLEASE – displays, remove the patient from the scale. Wait for the scale to display 0.0 and retry the measurement.

Print a weight measurement

Printers are a purchasable option for some scales. The printer will only print a valid weight. If your printer is set up with the auto-print option, a print is automatically generated after a measurement is taken.

1. Take a weight measurement as described.

A single beep sounds.

2. Press PRINT (DATA) SEND.

A single beep sounds and the measurement is printed.

3. To print a previously stored scale measurement, press RECALL and then press

PRINT (DATA) SEND.

The following errors can occur:

• Four short beeps indicate an invalid print request. Make sure that the weight is properly displayed on the scale, or that a measurement is not already in the printing process.

• Four long beeps indicate an internal problem with the printer. See the

Troubleshooting section.

14 Scale operation Welch Allyn Masted Scales

Storage, cleaning, and maintenance

Storage

Cleaning

The scale should be stored in a convenient storage facility or closet. Make sure to not leave the scale in an area where someone could trip on it or run into it.

1. Clean the scale at least once per day, and the optional handrail after each use.

2. Use one of the following products when cleaning your scale:

70 percent isopropyl alcohol

PDI Sani-Cloths

®

Battery replacement

CAUTION Use only size D disposable alkaline batteries. The use of any other battery will void the warranty.

1. Using a screwdriver, remove the two fasteners on the access door and remove the door.

2. Install six new batteries in the battery holder. Make sure to follow the polarity instructions.

3. Re-attach the access door.

15

16 Storage, cleaning, and maintenance Welch Allyn Masted Scales

Change the printer paper

The printer is a purchasable option.

1. Use a screwdriver to loosen the screw on the battery access door and remove the printer access door.

2. Remove the spindle containing the empty printer paper roll from the printer paper well.

3. Replace the printer paper roll in the spindle. Make sure the new printer paper is advancing from the bottom of the roll.

4. Lift the green knob to lift the print head lever.

5. Insert the paper through the paper guide slot, and make sure it goes under and around the black roller bar.

6. Pull the paper under the print head lever and push the green knob down to push the print head lever down.

7. Press PAPER ADVANCE to move the paper through the printer.

Directions for use Storage, cleaning, and maintenance 17

8. Install the printer access door, and advance the new paper through the slot on the door.

9. Install the screw on the printer access door.

Maintenance

Routinely perform the following preventive maintenance to keep your scale in working order.

1. Check the calibration annually or as required.

2. Inspect the scale for cracks or loose mounting hardware. Replace or repair as necessary.

3. Visually inspect the scale enclosure for damage or loose or missing hardware.

Replace or repair as necessary.

4. If equipped, inspect the AC line cord for abrasions or other signs of wear.

5. Do not expose the scale to excessive water or moisture.

6. Do not store the scale where heavy objects can be placed on it.

7. Replace the batteries annually or as required.

8. Do not service or perform maintenance while the scale is in use with a patient.

Troubleshooting

This section presents tables of information to help you troubleshoot issues with the scale.

To use these tables, locate the specific problem with your scale in the left column of the table. The remainder of the row explains possible causes and suggests actions that can resolve the issue.

Issue Possible cause

The weight reading is not accurate There may be a mechanical obstruction on the platform

Suggested action

Make sure the platform is not touching a foreign object or something restricting its downward motion.

Connect the cable.

The cable between the platform and the instrument circuit board may not be connected

The cable on the mast may not be connected

The weight reading takes too long to display

The platform may be in motion

Connect the cable.

Wait for the platform to settle.

The scale displays

CAbLE

The scale displays

O-LOAd

There may be a cable failure

The object on the scale may be larger than the maximum value allowed

Contact your service department.

If the weight value is within the specified range, contact your service department.

The reading does not change when weight is applied to the scale

The load cell transducer, connector, or cable may be defective

Contact your service department.

18 Storage, cleaning, and maintenance Welch Allyn Masted Scales

Issue Possible cause Suggested action

The scale buttons are not responding

The buttons may be damaged Check for visible signs of damage.

Contact your service department.

Contact your service department.

The error message displayed

E-FAIL

is

The error message displayed

r-FAIL

is

The printer failed to print, and there were four short beeps

The scale's memory has failed

The scale's memory has failed.

Contact your service department.

A valid weight may not be present Make sure the patient's weight is indicated on scale before you press

PRINT

.

Contact your service department.

The printer failed to print, and there were four long beeps

The printer runs but the result is blank or is faint

The print is distorted

There may be a problem with the printer

There may be a problem with the printer paper Make sure you are using thermal paper in the printer.

There may be a problem with the printer head or the printer interface door

Contact your service department.

Replace the battery.

The low battery light flashes during printing

The battery may be low

The printed paper tears poorly or unevenly

The tear bar may be damaged Check for signs of damage such as missing teeth or cracks. If damaged, contact your service department.

Appendices

Specifications

Item

Scales

Resolution (factory default)

Weighing capacity

Readout

Power source

Platform dimensions

Specification

5002, 5122, 5702, 6002

0.1 lb / 100 gm

Models 5002 and 6002: 880 lb (400 kg)

Model 5702: 1000 lb (454 kg)

Model 5122: 660 lb (300 kg)

Large, bright light-emitting diode display.

Selectable weight displayed in pounds or kilograms.

Indicator light to show selected units.

Battery

• 6 D alkaline batteries

Power adapter

• Medical grade UL listed external power supply

• Nominal input voltage: 100-240V AC ±10%

• Nominal input frequency: 50/60 Hz

• Nominal input current: 0.3-0.15Arms @ max load

• Nominal output voltage: U out

• Nominal output current: I out

: 12V DC +5% / -5%

: 1000mA

5002: 18 in. x 19.5 in. x 1.5 in.

5702: 24 in. x 26 in. x 2 in.

6002: 24 in. x 26 in. x 2 in.

5122: 16 in. x 14 in. x 1.5 in.

19

20 Appendices Welch Allyn Masted Scales

Product configurations

The Welch Allyn Masted Scales are available in the following configurations:

Model 5002, Mobile Stand-On Scale

Model

5002-KP

5002-KX

5002-XP

5002-XX

Description

Mobile Stand-On Scale, kilogram only, printer

Mobile Stand-On Scale, kilogram only, data port

Mobile Stand-On Scale, standard weight, printer

Mobile Stand-On Scale, standard weight, data port

Model 5122, Low-Profile Stand-On Scale

Model Description

5122-K

5122-X

Low-Profile Stand-On Scale, kilogram only

Low-Profile Stand-On Scale, standard weight

Model 5702, Mobile Bariatric Stand-On Scale

Model Description

5702-KP

5702-KX

5702-XP

5702-XX

Mobile Bariatric Stand-On Scale, kilogram only, printer

Mobile Bariatric Stand-On Scale, kilogram only, data port

Mobile Bariatric Stand-On Scale, standard weight, printer

Mobile Bariatric Stand-On Scale, standard weight, data port

Model 6002, Wheelchair Scale

Model Description

6002-KP

6002-KX

6002-XP

6002-XX

Wheelchair Scale, kilogram only, printer

Wheelchair Scale, kilogram only, data port

Wheelchair Scale, standard weight, printer

Wheelchair Scale, standard weight, data port

Directions for use

Approved accessories

Model 5002

Part number

845010

845010M

845010W

50231

23709

845233

Model 5122

Description

Height gauge (in and cm)

Height gauge (metric only)

Wall-mounted height gauge (in and cm)

Printer

Thermal printer paper (box, 15 rolls)

UL/CSA power adapter with hospital-grade plug

Part number

845010W

341001

Description

Wall-mounted height gauge (in and cm)

Power adapter

Model 5702

Part number

845010B

845010W

50231

23709

845233

Description

Height gauge for 5702 (in and cm)

Wall-mounted height gauge (in and cm)

Printer

Thermal printer paper (box, 15 rolls)

UL/CSA power adapter with hospital-grade plug

Model 6002

Part number

60224

845010

845010M

845010W

50231

Description

Handrail

Height gauge (in and cm)

Height gauge (metric only)

Wall-mounted height gauge (in and cm)

Printer

Appendices 21

22 Appendices Welch Allyn Masted Scales

Part number

23709

845233

Description

Thermal printer paper (box, 15 rolls)

UL/CSA power adapter with hospital-grade plug

Warranty

Welch Allyn will warranty the weight scale to be free of defects in material and workmanship and to perform in accordance with manufacturer specifications for the period of one year from the date of retail purchase.

The warranty period shall start on the date of purchase. The date of purchase is: 1) the invoiced ship date if the device was purchased directly from Welch Allyn, 2) the date specified during product registration, 3) the date of purchase of the product from a

Welch Allyn authorized distributor as documented from a receipt from said distributor.

This warranty does NOT cover damages caused by misuse or abuse, including but not limited to:

• Failure caused by unauthorized repairs or modifications

• Damage caused by shock or dropping during transportation

• Damage caused by improper use of the power supply

• Failure caused by improper operation not consistent with the instructions stated in this Directions for use

Should this device require maintenance (or replacement at our option) under warranty, contact your local Welch Allyn representative: http://www.welchallyn.com/en/other/ contact-us.html

Material No. 725556