The Transposal Ultra Cart UL-QD2800 and UL-DU500 are designed to collect surgical fluids from one patient at a time in a reusable suction device that can then be emptied, cleaned, and rinsed. The carts are also designed to enhance or replace main wall vacuum supplies in operating room suites. Once fluids are collected, the cart is connected to an Ultra Evacuation Unit to process the reservoirs. Two fluid collection capacity options are available. The Ultra Quad (4 reservoirs) collects up to 52-liters of fluid while the Ultra Duo (2 reservoirs) collects up to 33-liters.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Dornoch Transposal Ultra Cart Instructions for Use | Manualzz

Transposal Ultra Cart

Instructions for Use

Model No.: UL-QD2800 w/IV Pole (optional)

Model No.: UL-DU500

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

Effective Date: 07.23.2014

CN No. 14289

Contents

Introduction ------------------------------------------------------------------------------------------- 1

Important Information ------------------------------------------------------------------------------------------------- 2

Intended Use ------------------------------------------------------------------------------------------------------------ 2

Equipment Description ------------------------------------------------------------------------------------------------- 2

User/Patient Safety ----------------------------------------------------------------------------------------------------- 3

Technical Description -------------------------------------------------------------------------------- 4

Specifications ------------------------------------------------------------------------------------------------------------ 4

Instructions for Use ---------------------------------------------------------------------------------- 6

Models Using Single Use TP-DL2800 Lids --------------------------------------------------------------------------- 6

Fluid Collection ------------------------------------------------------------------------------------------------------------------------------------- 6

Models Using Single Use UL-CL500 Manifolds ---------------------------------------------------------------------- 7

Fluid Collection ------------------------------------------------------------------------------------------------------------------------------------- 7

Preparing for the Next Procedure --------------------------------------------------------------------------------------------------------------- 9

Wall Suction Operation------------------------------------------------------------------------------------------------- 9

All Models -------------------------------------------------------------------------------------------------------------- 10

Transporting the Cart ---------------------------------------------------------------------------------------------------------------------------- 10

High Flow Mode (Optional Feature) ------------------------------------------------------------------------------------------------------------ 10

General Usage Information ----------------------------------------------------------------------------------------- 11

Single Use Lids/Manifolds and Disposable Suction Tubing --------------------------------------------------------------------------------- 11

Reservoir Use -------------------------------------------------------------------------------------------------------------------------------------- 12

Early Warning and Overflow Protection System --------------------------------------------------------------------------------------------- 12

Alternate Vacuum Source ------------------------------------------------------------------------------------------------------------------------ 12

Rapid Full Vacuum and Specimen Collection ------------------------------------------------------------------------------------------------- 12

TP-DL2800 Application Information and Precautions --------------------------------------------------------------------------------------- 13

Unit Information and Error Messages ---------------------------------------------------------------------------------------------------------- 13

Periodic Maintenance ----------------------------------------------------------------------------------------------------------------------------- 13

Troubleshooting Guide --------------------------------------------------------------------------------------------------------------------------- 14

DMS Technical Service - Information ---------------------------------------------------------------------------------------------------------- 14

Symbols & Labels --------------------------------------------------------------------------------------------------------------------------------- 15

Consumables and Accessories ------------------------------------------------------------------------------------------------------------------ 17

Company Information ----------------------------------------------------------------------------- 18

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

Effective Date: 07.23.2014

CN No. 14289

List of Tables

Table 1 – Equipment Specifications ...................................................................................................... 4

Table 2 – Approved Wipes and Solutions ............................................................................................ 10

List of Figures

Figure 1 – Ultra Duo Cart Dimensions and Features (Dim. = inch (cm)).................................................. 5

Figure 2 – Ultra Quad Cart Dimensions and Features (Dim. = inch (cm)) ................................................ 5

Figure 3 – Vacuum Startup Warning Screen .......................................................................................... 6

Figure 4 – Power Cord and Switch ....................................................................................................... 6

Figure 5 – Vacuum Adapter and Lid ..................................................................................................... 6

Figure 6 – DU2800 Display .................................................................................................................. 6

Figure 7 – QD2800 Display .................................................................................................................. 6

Figure 8 – CL500 Manifold Assembly .................................................................................................... 7

Figure 9 – Vacuum Adapter, Manifold and Cover ................................................................................... 7

Figure 10 – Vacuum Startup Warning Screen ........................................................................................ 8

Figure 11 – DU500 Display .................................................................................................................. 8

Figure 12 – QD500 Display .................................................................................................................. 8

Figure 13 – Used Manifold Removal ..................................................................................................... 9

Figure 14 – Wall Suction Only .............................................................................................................. 9

Figure 15 – High Flow Mode Indicator ................................................................................................ 10

Figure 16 – Port Cap Under Lid .......................................................................................................... 13

Figure 17 – DO NOT Insert Objects .................................................................................................... 13

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

Effective Date: 07.23.2014

CN No. 14289

Introduction

The Transposal Ultra Fluid Cart (cart) is designed to collect surgical fluids from one patient at a time in a reusable suction device that can then be emptied, cleaned, and rinsed. The cart is also designed to enhance or replace main wall vacuum supplies in operating room suites. Two fluid collection capacity options are available. The Ultra Quad (4 reservoirs) collects up to 52-liters of fluid while the Ultra Duo (2 reservoirs) collects up to 33-liters. Once fluids are collected, the cart is connected to an Ultra Evacuation Unit (evac) to process the reservoirs.

Two manuals are provided with this equipment and are shipped with the equipment during the initial installation. Copies of these manuals are available to any Dornoch Medical Systems (DMS) customer upon request.

The two different manuals are as follows:

Instructions for Use Manual – This manual is designed to instruct users in the correct operation of the cart as well as safety considerations associated with the unit.

Technical Service Manual – This manual includes operation instructions, installation & disconnection instructions, technical specifications, and preventative maintenance requirements for the unit.

This is the Instructions for Use Manual, and it is broken into four sections including:

Introduction

Technical Description

Instructions for Use

Company Information

If required, DMS service personnel can be contacted at 1-888-466-6633. Technicians are available 8 am – 5 pm weekdays. An answering service is available 24-hours a day, 7 days a week and a technician will return the call the next business day.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

1 Effective Date: 07.23.2014

CN No. 14289

Important Information

Please read this manual and follow all instructions. The words WARNING, CAUTION and NOTE have special meanings and should be reviewed.

WARNING: Disregarding WARNING information may compromise the safety of the patient and/or health care staff and may result in injury or death.

CAUTION: Disregarding CAUTION instructions may compromise product reliability and may result in damage.

NOTE: NOTE information supplements and/or clarifies procedural information.

A triangle with an exclamation point alerts the health care professional to read and understand the accompanying instructions, especially the operating, maintenance and safety information.

Intended Use

The Dornoch Transposal Ultra System is self-powered suction / vacuum pump intended to collect and dispose of liquid waste within Hospital Operating Rooms, Pathology Labs, Surgical Outpatient Centers, and

Doctor's Offices.

Equipment Description

Carts collect up to 52-liters of surgical fluids without the need to tandem multiple suction canisters. In addition, the cart’s on-board suction assist pump improves or replaces facility wall vacuum when inadequate suction is otherwise present. The cart significantly reduces employee exposure to potentially infectious body fluids, while eliminating up to 70% of Operating Room red bag waste.

The cart contains either two 16.5 Liter reservoirs or four 13 Liter reservoirs which are processed using an evac. The multiple reservoir design allows separate suction level measurement. However, the carts contain a single vacuum pump. A suction level decrease in any reservoir, due to an open port or line, may cause the other reservoir to decrease. Once the system is installed, operators use the carts to collect fluid during surgical cases and then move them to the evac where they are processed. Once the evac cycle is started, the reservoirs are automatically emptied, cleaned, and rinsed. The carts are then wiped down and, with the addition of a new single use lid or manifold, are ready for another surgical procedure.

New carts come in two model options with the main difference being the single use disposable. The two models are the Ultra Fluid Cart (UL-DU2800 – Duo with Canister Lid and UL-QD2800 – Quad with

Canister Lid) and the Ultra fleX Fluid Cart (UL-DU500 – Duo with Cart Manifold and UL-QD500 – Quad with Cart Manifold). The 2800 configuration is designed for single patient use in each reservoir, as well as high fluid generating procedures. The standard configuration Ultra Fluid Carts use the Transposal Single Use

Lid #TP-DL2800 to provide the suction to the surgical field. The other configuration option is the Ultra fleX

Fluid Cart designed for the busy Operating Room demanding multi-patient fluid collection and high vacuum flow. The special Ultra fleX Fluid Cart Single Use Manifold #UL-CL500 supports these needs with a convenient, easy-to-use remove and seal design.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

2 Effective Date: 07.23.2014

CN No. 14289

User/Patient Safety

WARNINGS:

• DO NOT apply High Flow suction or allow extended exposure of suction to the tissue associated with procedures that require either no suction, low vacuum or low flow suction, for example, passive chest drainage.

ALWAYS consider the type of tissue associated with the surgical procedure BEFORE using this system. Failure to comply may result in severe injury or death.

• Before using this system, read and understand the information in this manual. Suction equipment should only be used by people that have had adequate training on the use of this type of equipment. Pay special attention to WARNING information.

Become familiar with the system components prior to use.

• The health care professional performing any procedure is responsible for determining the appropriateness of this equipment and the specific technique used for each patient.

Dornoch Medical Systems, Inc., as a manufacturer, does not recommend surgical procedure or technique.

• DO NOT use this system outside the scope of the defined indications for use.

• DO NOT use this system for applications that require a constant vacuum level.

• The reservoir scale and fluid volume display are not diagnostic tools. DO NOT use the scale or volume display to accurately calculate the amount of fluid loss from the patient.

• Upon initial receipt and before each use, operate the equipment and inspect each component for damage. DO NOT use any component if damage is apparent.

• ALWAYS close all unused ports and remove all unused tubing to maintain optimal suction levels. The suction levels of each canister are interdependent and linked to a common vacuum source. Failure to comply may result in the unexpected reduction of suction and patient injury.

• The suction level of this product relative to its vacuum limit setting may fluctuate significantly but will not exceed its limit.

DO NOT use this system if vacuum fluctuation may cause patient injury. ALWAYS consider the type of surgical procedure before using this system.

• ALWAYS follow current local regulations governing procedurespecific suction levels to remove fluid waste safely from a surgical site.

• Reservoirs are for surgical and bodily fluid collection only; do not place any items into the reservoir for disposal. Caustic or other harmful chemicals may harm the equipment and will void the warranty.

• Handling biohazard waste is potentially dangerous. ALWAYS follow current local regulations governing biohazard waste to safely handle and dispose of surgical fluid waste.

• Manifolds and suction tubing may contain surgical waste after use. ALWAYS handle these disposable accessories as

“potentially infectious materials” after use. ALWAYS wear gloves and protective eye wear when removing and disposing of these disposable accessories.

• The Blood-borne Pathogens Standards, provided by the

Occupational Safety and Health Association (OSHA), requires that all workers, having exposure to “potentially infectious materials”, should wear the correct personal protection equipment.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

3

• To avoid the risk of electrical shock, this equipment must be connected to electrical outlet with a protective earth ground.

• ALWAYS follow reservoir overfill alarms to prevent overfill.

• Verify all access doors are securely in place before operating this unit.

• Use only Dornoch Medical Systems, Inc. approved accessories.

Do not connect items to this system that are not designed for or specified for use with this system.

• ALWAYS have more than one person unpack and lift the equipment off the shipping pallet.

• Perform recommended maintenance as indicated in these instructions. Only trained and experienced health care professionals should maintain this equipment.

• No user serviceable parts inside the unit. Contact Dornoch

Medical Systems, Inc. customer service if an issue arises. Only authorized service personnel should open any of the access covers on this equipment. User operation does not require access to these areas.

• The unit is for use within the hospital and/or surgery center.

The unit is moveable but is not intended to be transportable or used in the outside environment. To be used within the healthcare facility ONLY.

• DO NOT use the system with patients that are being treated with radioisotopes or hazardous chemotherapy agents.

• ALWAYS use the handle to move the cart. DO NOT push or pull the cart by grasping any other part of the outer surface.

NEVER hang any heavy object from the cart handles.

• No incline planes of operation.

• DO NOT pump into the reservoirs. The lids could become dislodged, the reservoirs could overfill and surgical fluid could spill from the unit.

• DO NOT use the system if leakage of surgical fluid waste occurs. Disconnect power immediately and call Dornoch

Medical Systems, Inc. customer service.

• DO NOT allow fluid of any kind to spill directly onto the exterior surface of the electrically-powered cart.

• DO NOT use the cart until it has been tested properly and the

Evac Unit has been installed and tested properly.

• This unit uses both bleach and enzyme in its operation. When replacing bottles, always wear the appropriate Personal

Protective Equipment. Use only Dornoch Medical Systems, Inc. approved Bleach and Enzyme to avoid damage to the system components.

• This equipment is not suitable for use in the presence of flammable anesthetic mixtures with air, oxygen or nitrous oxide.

Is not intended for use with AP or APG equipment.

• Hot Water Temperatures higher than 120 0 F can cause damage to the unit.

• There are no known significant risks of reciprocal interference posed by the presence of this equipment or its operation in either the operating room suites or other areas when used during specific investigations and/or treatments.

• There are no known potential electromagnetic or other interferences between this unit and other devices located and/or operated within the area of the operating room suites.

Effective Date: 07.23.2014

CN No. 14289

Technical Description

Specifications

Table 1 – Equipment Specifications

Item Ultra Quad (#UL-QD2800/#UL-QD500) Ultra Duo (#UL-DU2800/#UL-DU500)

Certifications ETL Listing - Class 1 Medical Device. Complies with the following:

 IEC 60601-1. General requirements for safety of medical equipment devices.

 IEC 60601-1-1. Collateral Standards for Medical

Equipment

 IEC 60601-1-2. EMC & RFI compliance testing standard.

 IEC 60529 – Degrees of protection provided by enclosures.

 FDA 510K Certification: K081047

 ISO 10079-1. Medical Suction Equipment – electrically powered suction equipment. (FDA standard for hospital suction systems.)

 CAN/CSA Standard C22.2 No. 601.1

FDA Product Code JCX (carts) CGY (lids)

FDA Manufacturer # 1954182

Patents Patent Pending

Installation Mobile

Size (inch (cm)) 33(84)W x 23(58)D x 55(140)H

Weight (lbs (kg)) 225(102) empty; 340(154) full

Utilities

ETL Listing - Class 1 Medical Device. Complies with the following:

 IEC 60601-1. General requirements for safety of medical equipment devices.

 IEC 60601-1-1. Collateral Standards for Medical

Equipment

 IEC 60601-1-2. EMC & RFI compliance testing standard.

 IEC 60529 – Degrees of protection provided by enclosures.

 FDA 510K Certification: K081047

 ISO 10079-1. Medical Suction Equipment – electrically powered suction equipment. (FDA standard for hospital suction systems.)

 CAN/CSA Standard C22.2 No. 601.1

JCX (carts) CGY (lids)

1954182

Patent Pending

Mobile

24(61)W x 24(61)D x 55(140)H

195(88) empty; 268(122) full

Power 115-120VAC, 60Hz, 15amp (non-dedicated)

Additional Cooling None

Flammable Rating Not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or with nitrous oxide.

Operation

Method Connected to the facilities main vacuum system or as a standalone device.

Mode Continuous

Performance High Vacuum/High Flow, maximum vacuum level obtainable is 0-600mmHg. (Measured with all ports closed) If connected to hospitable suction, obtainable vacuum can be higher than 700mmHg.

Fluid Capacity 52,000mL

Manual IV Pole (Optional) 6,000mL maximum fluid capacity

Powered IV Pole (Optional) 12,000mL maximum fluid capacity

Cleaning Dornoch enzyme solution and Clorox bleach are used to clean the collection cylinders of the Ultra Fluid Cart.

115-120VAC, 60Hz, 15amp (non-dedicated)

None

Not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or with nitrous oxide.

Connected to the facilities main vacuum system or as a standalone device.

Continuous

High Vacuum/High Flow, maximum vacuum level obtainable is 0-600mmHg. (Measured with all ports closed) If connected to hospitable suction, obtainable vacuum can be higher than 700mmHg.

33,000mL

6,000mL maximum fluid capacity

12,000mL maximum fluid capacity

Dornoch enzyme solution and Clorox bleach are used to clean the collection cylinders of the Ultra Fluid Cart.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

4 Effective Date: 07.23.2014

CN No. 14289

Touch

Control

Panel

Optional

IV Pole

Single Use

Lids

Evac

Coupler

Receiver

Fluid

Reservoirs

Wall

Vacuum

Outlet

Power

Cord

Receptacle

24 (61) 24 (61)

Figure 1 – Ultra Duo Cart Dimensions and Features (Dim. = inch (cm))

Optional

IV Pole

Single Use

Lids

Touch

Control

Panel

Evac

Coupler

Receiver

55

(140)

Wall

Vacuum

Outlet

Power Cord

Receptacle

Fluid

Reservoirs

23 (58) 33 (84)

Figure 2 – Ultra Quad Cart Dimensions and Features (Dim. = inch (cm))

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

5 Effective Date: 07.23.2014

CN No. 14289

55

(140)

Instructions for Use

Models Using Single Use TP-DL2800 Lids

Fluid Collection

Step 1 – Verify fluid reservoirs are empty and clean. Ensure the unit

is equipped with new single use lids. (Figure 5)

NOTE: The volume of the pre-fill fluid in the reservoir is accounted for in the displayed volume on the cart’s touch screen.

Step 2 – To prevent inadvertent movement of the cart, lock the caster. Plug power cord into a wall outlet and verify that it is also plugged into the cart. Turn the power switch to the ON position.

(Figure 4)

Figure 4 Power Cord and Switch

Step 3 – Attach cart vacuum line adapter to the “Vacuum Supply”

port on the single use lid. ( Figure 5 )

Step 4 (Duo Cart) – Turn reservoir vacuum ON or OFF by pressing the desired reservoir icon on the cart’s touch screen. ( Figure

5 )

Step 4 (Quad Cart) – Turn reservoir vacuum ON or OFF by pressing the desired reservoir icon under the “Light” buttons on the

cart’s touch screen. ( Figure 7 )

Vacuum Startup Warning – Any reservoirs powered on for the first time since Evac processing will start with a suction level of zero.

Suction can be increased by using the up arrow next to the

appropriate reservoir icon (Figure 6 and Figure 7). Any time

following the initial powering up of a reservoir, the vacuum startup

warning screen will appear. (Figure 3)

Figure 5 – Vacuum Adapter and Lid

Figure 6 – DU2800 Display

Figure 3 – Vacuum Startup Warning Screen

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

6

Figure 7 – QD2800 Display

Effective Date: 07.23.2014

CN No. 14289

Vacuum Safety Warning Continued – When the vacuum startup warning screen appears, carefully consider which is best for patient safety and select either “Start At 0 mmHg” or “Previous Setting”.

“Previous Setting” will start suction at the last suction setting used.

Warning: DO NOT apply High Flow suction or allow extended exposure of suction to the tissue associated with procedures that require either no suction, low vacuum or low flow suction. Failure to comply may result in severe injury or death.

Step 6 – Verify suction is present at a manifold port.

Step 7 – Load any required irrigation fluid onto the optional IV pole. Do not load more than 6000mL of fluid on the manual IV pole or more than 12,000mL of fluid on the powered IV pole.

Step 8 – Attach patient tubing to lid port(s).

Step 9 – Regulate suction by pressing the up or down arrow on the vacuum scale next to the appropriate

reservoir image. (Figure 6 and Figure 7)

High Flow Mode (Optional Feature)

If increased suction is required during the procedure, adjust the regulated vacuum to full and then press the

up arrow one additional time to activate high flow mode. (Figure 15) To deactivate high flow mode press

the down arrow to return to regulated suction.

Models Using Single Use UL-CL500 Manifolds

Fluid Collection

Step 1 – Verify fluid reservoirs are empty and clean. Ensure the unit is

equipped with new single use manifold. (Figure 8)

NOTE: The volume of the pre-fill fluid in the reservoir is accounted for in the displayed volume on the cart’s touch screen.

Step 2 – To prevent inadvertent movement of the cart, lock the caster.

Plug power cord into a wall outlet and verify that it is also plugged into the cart. Turn the power switch to the ON position.

Step 3 – Attach cart vacuum line adapter to the “Vacuum Supply” port

on the single use manifold. ( Figure 9 )

Step 4 (Duo Cart) – Turn reservoir vacuum ON or OFF by pressing the desired

reservoir icon on the cart’s touch screen. ( Figure 11 )

Step 4 (Quad Cart) – Turn reservoir vacuum ON or OFF by pressing the desired reservoir icon under the “Light” buttons on the cart’s touch

screen. ( Figure 12 )

Figure 8 – CL500 Manifold Assembly

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

7

Figure 9 – Vacuum Adapter, Manifold and

Cover

Effective Date: 07.23.2014

CN No. 14289

Vacuum Startup Warning – Any reservoirs powered on for the first time since Evac processing will start with a suction level of zero. Suction can be increased by using the up arrow next to the appropriate reservoir

icon (Figure 11 and Figure 12). Any time following the initial powering up of a reservoir, the vacuum startup warning screen will appear. (Figure

10)

Figure 11 – DU500 Display

Figure 10 – Vacuum Startup Warning Screen

When the vacuum startup warning screen appears, carefully consider which is best for patient safety and select either “Start At 0 mmHg” or

“Previous Setting”. “Previous Setting” will start suction at the last suction setting used.

Warning: DO NOT apply High Flow suction or allow extended exposure of suction to the tissue associated with procedures that require either no suction, low vacuum or low flow suction. Failure to comply may result in severe injury or death.

Step 6 – Verify suction is present at a manifold port.

Step 7 – Load any required irrigation fluid onto the optional IV pole.

Do not load more than 6000mL of fluid on the manual IV pole or more than 12,000mL of fluid on the powered IV pole.

Step 8 – Attach patient tubing to manifold port(s).

Step 9 – Regulate suction by pressing the up or down arrow on the

vacuum scale next to the appropriate reservoir image. (Figure 11 and

Figure 12)

High Flow Mode (Optional Feature)

If increased suction is required during the procedure, adjust the regulated vacuum to full and then press the up arrow one additional time to

activate high flow mode. (Figure 15) To deactivate high flow mode press

the down arrow to return to regulated suction.

Figure 12 – QD500 Display

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

8 Effective Date: 07.23.2014

CN No. 14289

Preparing for the Next Procedure

To use the same reservoir for the next patient, use the following steps. Otherwise, use an empty reservoir or transport the cart to the evac for processing.

Warning: Follow the current local regulations governing biohazard waste to safely handle and dispose of surgical fluid waste.

Step 1 – Cap all of the patient ports but leave the “Vacuum Only” port uncapped. Remove the manifold from the housing and tip it forward so the ports face slightly down. Take the cap off of the back

of the housing and place it on the bottom of the manifold. (Figure 13)

Figure 13 – Used Manifold Removal

Step 2 – Cap the “Vacuum Only” port and dispose of the manifold per facility red bag waste policy.

Step 3 – Reset the reservoir fluid volume to zero by pressing the digital value below the reservoir icon and following the instructions.

(Number of resets limited to four total resets between cart cleanings.)

Cylinder vacuum level will start at zero following cylinder reservoir fluid reset.

Figure 14 – Wall Suction Only

Step 4 – Insert a new manifold and proceed to Step 3 in the “Fluid

Collection” section on the previous section.

Wall Suction Operation

Dornoch suction carts may be run with the use of hospital wall suction. This feature may be used in the event of a failure of the vacuum pump inside the cart. It also may be used if very quiet equipment operation is desired.

Step 1 – Connect suction hose from the wall suction port to the suction inlet port on the left side of the Duo or back side of the Quad suction cart.

Step 2 – Press the set up button in the bottom left hand corner of the

main OR screen. (Figure 11 and Figure 12)

Step 3 – Press the “Wall Suction Only” button. (Figure 14) (Button

will turn green). All cart functions will remain active, with the exception of the vacuum pump.

Step 4 – If vacuum pump operation is desired, press “Wall Suction

Only” button to turn this feature off (button will turn gray).

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

9 Effective Date: 07.23.2014

CN No. 14289

All Models

Transporting the Cart

After the surgical procedure is complete, the cart needs to be transported to the evac to be emptied, cleaned, and rinsed.

Warning: Always remove the irrigation accessories and lower the IV pole prior to relocating the cart. Always use the handle to retain control of the cart during relocation.

Step 1 – Turn off the suction in the cart by pressing the cylinder icon on all activated reservoirs on the carts touch screen.

Step 2 – Remove patient tubing and cart vacuum lines from the green single use lid(s)/manifold(s) and seal all ports with the attached caps.

Step 3 – Detach wall vacuum line and power cord from the wall and wind onto the cart’s storage bracket.

Unlock the caster to allow the cart to be transported and move it to the evac for processing.

Step 4 – Once the cart is processed and used single use lids/manifolds are discarded, wipe down the cart with

an approved bleach wipe or solution (Table 2) and place clean single use lids/manifolds onto the cart.

Dornoch also recommends performing a full soak cycle once per week to assure that the cylinders remain clean.

Table 2 – Approved Wipes and Solutions

Name

Sani-Cloth Bleach Germicidal Wipes (P54072)

Clorox Healthcare Bleach Germicidal Wipes (UPC 30577 )

Clorox Household Bleach (< 1200 ppm)

3oz (80mL) Clorox in 1 Gal (4L) of Water

Manufacturer

PDI

Clorox

Clorox

CAUTION: Avoid pooling of excess disinfection liquid around seams or edges. Dry area with a paper towel if necessary to avoid prolonged exposure to disinfectants. Non-approved wipes may cause surface damage to the equipment.

NOTE: Liquid will remain in the reservoir after processing is complete. This liquid contains a small amount of enzyme for starting the breakdown of fluid waste collected during the next case

High Flow Mode (Optional Feature)

High Flow Mode is an optional feature available on some carts. It allows a boost in suction to a customer selected reservoir when more suction is needed. Suction boost is still limited to number of lines hooked to the reservoir and if another reservoir is in use. If increased suction is required during the procedure, adjust the regulated vacuum to full and then press the up arrow one additional

time to activate high flow mode. (Figure 15) To deactivate high flow

mode press the down arrow to return to regulated suction.

Figure 15 – High Flow Mode Indicator

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

10 Effective Date: 07.23.2014

CN No. 14289

General Usage Information

Single Use Lids/Manifolds and Disposable Suction Tubing

Single Use Lid: (TP-DL2800)

The single use lid has three suction ports. Any one of these ports (and up to three) can be used to collect fluids. The suction ports are labeled:

1. “Patient” this port is used for routine collection of fluids.

2. “Tandem” this port is used for routine collection of fluids.

CAUTION: The use of this port to tandem reservoirs together is not allowed on carts as this can cause damage to the equipment.

3. “Ortho” this port is used for routine collection of fluids or used when larger suctioned particulate may occlude the Patient or Tandem ports.

Single Use Cart Manifold: (UL-CL500)

The single use manifold has three suction ports. Any one of these ports (and up to three) can be used to collect fluids. The suction ports are not labeled and are all the size of a standard “Ortho” port.

Single Use Lids and Manifolds:

WARNING: Both the TP-DL2800 lid and the UL-CL500 manifold are designed to be used for a single patient. Failure to change the item after each patient may result in the following:

1. Loss of available suction. Contaminated and/or wet filters will affect vacuum flow.

2. Cross contamination. Improper usage of single use items poses a health risk to patients and health care providers.

Disposable Suction Tubing: (Provided by the Hospital)

Any surgical suction tubing of variable diameters (sterile or unsterile) can be used to attach to single use lids/manifolds. Most suction tubing has flexible boots on the ends to accommodate different diameter suction ports or surgical devices. For high flow applications Dornoch strongly recommends 9/32” ID tubing as it flows noticeably better than smaller 3/16” and 1/4" sizes.

CAUTION: Do not connect items that are not designed for suction systems as a safety hazard may result.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

11 Effective Date: 07.23.2014

CN No. 14289

Reservoir Use

General Information for Reservoir Usage

Cart reservoirs, consisting of a reusable fluid collector and single use lid, shall be used in place of traditional disposable suction canisters/liners in surgery. Cart reservoirs are considered non-critical and non-sterile medical devices and should be handled accordingly. The fluid collection ranges for each reservoir is 0 to

13,000mL for an Ultra Quad and 0 to 16,500mL for an Ultra Duo. The accuracy of the digital readouts on the Ultra Duo and Ultra Quad is +/- 150mL. The graduations on the exterior of the reservoirs are approximations only. Record fluid volume as needed from the digital readouts on the touch screen display.

Note: Clots or solid masses in the collection cylinders can affect fluid readout accuracy.

CAUTION: Reservoirs are not meant to be used in tandem as this could damage the equipment. Fluid must not be pumped directly into the reservoirs as this could cause the single use lid to become dislodged or damage the equipment.

CAUTION: DO NOT use solidifier in the cart reservoirs. It will make the cart inoperable until it can be serviced to clean out the reservoir and may cause damage to the equipment.

Early Warning and Overflow Protection System

The cart will alert the user when the reservoir is a 1000mL from full capacity, and will do the following to avoid an overfill situation when the level reaches the reservoirs maximum capacity.

- Reservoir will no longer be active and the vacuum will automatically shut off.

- Reservoir will automatically vent to prevent backflow.

- The hydrophobic filter in the bottom on the single use lid or single use cart manifold is a secondary means (mechanical) of safety to prevent liquid from getting into the vacuum line.

Alternate Vacuum Source

Equipment Malfunction or Power Loss Situation

An alternate source of vacuum should be readily available in the event the cart can no longer provide suction to the reservoir. The cart can still be used with an external vacuum source hooked directly to the “Vacuum

Port” of the single use lid. Vacuum regulation and volume readout will not be available with the cart setup in this configuration.

Rapid Full Vacuum and Specimen Collection

Application that Requires Full Suction in Ten Seconds or Less:

On the back of the unit there is a place for a canister ring holder. Place a canister ring (TP-RH100) into the holder and then either an 1800cc (TP-RC1800 (reusable)/TP-DCL1800 (disposable w/lid)) or 2800cc (TP-

RC2800 (reusable)/TP-DCL2800 (disposable w/lid)) canister into the ring. Hook the black vacuum adapter from one of the reservoirs to the vacuum port on the canister lid and turn on the vacuum for that reservoir.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

12 Effective Date: 07.23.2014

CN No. 14289

TP-DL2800 Application Information and Precautions

- To place a TP-DL2800 lid onto a reservoir, press it onto the lid ring with two hands working around the outside edge toward the back of the cart. Another option is to cap all of the ports and use suction to pull the lid onto the lid ring.

- When positioning the TP-DL2800 single use lid onto the cart,

check that port caps are not caught under the lid. (Figure 16) Cap

all ports prior to transporting the cart.

Figure 16 – Port Cap Under Lid

-

CAUTION: Do NOT insert objects into the reservoir. (Figure

17) Reservoir contents are to be collected with suction only.

Depositing gloves, trash, or other miscellaneous items will CLOG the reservoir making it unable to drain and rendering it inoperable.

Figure 17 – DO NOT Insert Objects

Unit Information and Error Messages

The cart provides the operator feedback for common operating errors via the display panel. The most common messages are described below. Please call DMS at 1-888-466-6633 for help with any error message.

Warning or Error

Tandem Error: This message indicates fluid has been detected entering a reservoir that does not have vacuum applied.

Temperature Errors: This message provides notice that the temperature monitoring system in the unit has detected an elevated temperature.

Processing Required: This message notifies the user that the cart has not been processed for 16 hours since it was used to collect fluid.

Periodic Maintenance

Action Required

Verify fluid lines are only hooked to reservoirs that are turned on. Do not tandem reservoirs or pump directly into reservoirs. Call service if necessary.

Finish the current surgical procedure and then remove the cart from use. Contact service once the cart is available for repair.

Finish the current surgical procedure and then process the cart with the evac.

Interval Activity

Prior to each use Inspect the collection reservoirs for cracks and the two casters for failed lock and steer and after each cleaning levers. Ensure there are no cuts in the power cord and no bent pins in the power cord receptacle. DO NOT use if the equipment if damage is present.

As required Check the level of the enzyme and bleach in the bottles. Replace the bottles as required.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

13 Effective Date: 07.23.2014

CN No. 14289

Troubleshooting Guide

Problem: Unit Will Not Operate

Possible Solutions:

1. Check that the machine is plugged into an operational outlet.

2. Check that the switch on the rear of the unit is turned on and that the power cord is all the way into the receptacle.

3. Call maintenance to replace the internal fuses.

Problem: Display is Powered, Cannot Get Vacuum Pump to Turn On

Possible Solutions:

1. Go to the setup screen and verify wall suction only button is not green.

2. Try to run the unit with wall vacuum.

3. Call DMS at 1-888-466-6633.

Problem: Volume Not Displayed on Screen

Possible Solutions:

1. Level Sensor not calibrated or defective.

2. Call DMS at 1-888-466-6633.

Problem: No Suction or Inadequate Suction

Possible Solution:

1. Verify that new single use lids have been installed. The single use lids/manifolds are each equipped with a hydrophobic filter that will close when it becomes wet. If the unit was processed and the lids have not been replaced this will block suction at the patient ports.

2. If using the cart not attached to wall suction, verify that the suction pump has been turned “ON”.

3. Call DMS at 1-888-466-6633.

DMS Technical Service - Information

DMS Technical Service line is answered 24 hours a day, 7 days a week. Technicians are available in the office

Monday – Friday from 8AM until 5PM central standard time. For all other calls please leave information with the answering service at 1-888-466-6633 and a service technician will get back to you on the next business day. Please have the information below available for the technician. If possible, call with the equipment available for troubleshooting.

• Equipment Type (Ultra Duo, Ultra Quad, Evac or Safety Station)

• Equipment Serial Number

• Description of the problem

• Location of the equipment

• Contact information for a service technician coming to the account

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

14 Effective Date: 07.23.2014

CN No. 14289

Symbols & Labels

The explanation and location of all Labels, Safety Signs, Symbols and Displays used on the equipment is presented below.

Protective Earth Ground Symbol – This label is used to symbolize a protective earth ground location on the unit. There is one located on the ground from the main power cord and another located on the ground for the 24VDC power supply.

Functional Earth Ground Symbol – This label is used to symbolize a functional earth ground. They are located near all functional grounds.

Single Use Symbol – This label is used to symbolize that the product is single use only. This label is located on either of the single use lids.

Consult Instructions Symbol – When displayed on a device, it refers the user to accompanying instructions and identifies safety and precautionary information.

Electrical Requirements, Serial Number, and Operating Instructions Label (See Below) - Located on the back of the unit under the electrical cabinet access panel. This label is used to instruct the user to the manuals. Information contained on this label: DMS contact information, serial number, ETL compliance information, electrical information, patent information and power outlet requirements.

Dornoch Medical Systems, Inc.

200 NW Parkway Riverside, MO

64150 www.dornoch.com

(888) 466-6633

Model Number

Serial Number

This equipment complies with and is Listed under the IEC-60601-1

(2 nd Edition) Standard for a Class 1 Medical Device.

This equipment complies with and is approved for use under:

IEC-60601-1: General Requirements for Safety

IEC-60601-1-1: Safety Requirements for Medical Electrical Systems

IEC-60601-1-2: Electromagnetic Compatibility - Requirements and

Tests

ISO-10079-1: Electrically Powered Suction Equipment

IEC-60529 rating certification of IP41

115 – 120 VAC 60 Hz

Nominal: 0.2 – 4.5 A

Momentary Peak: 6.5 A

Replace fuses with T10AL250V

250 V 10 A Time Delayed Fuse

Patent Pending

Power connection to be made with a Hospital Grade power cord only and must be connected to a Hospital Grade power receptacle only.

Suction: High Vacuum/High Flow

Maximum Level 700mmHG

Warning: Do Not Touch

Internal Pins or Connectors

Suction Rating Label – This label is used to inform the user of the maximum level of vacuum that can be developed by the system. This label is located on the collar of the monitor pole.

Receiver Warning Label – This label is used to warn the user of the electrical pins and connectors inside the receiver of the cart. This label is located under the receiver doors.

Do not connect any item here except evac coupler.

Ne touchez ni les connecteurs ni les broches internes.

Connect to Hospital Suction Only

Vacuum Draw 85 lpm @ 300 mmHg

Receiver Warning Label – French version

Hospital Suction Information Label – This label is used to inform the users of the hospital suction supply requirements.

This label is located under the wall vacuum connector.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

15 Effective Date: 07.23.2014

CN No. 14289

Bleach Wipe Label – This label is used to inform the users to use only approved bleach wipes when wiping down the carts.

This label is located on the center of the top plate of the unit.

Air Filter: Filter to be Changed Once per

Year. See the “Preventative Maintenance

Section” in the Technical Manual

Air Filter Label – This label is used to inform the users of the air filter replacement requirement. This label is located on the air filter housing.

HEPA Filter: Filter to be Changed

Once per Year. See the

“Preventative Maintenance

Section” in the Technical Manual

BRAKE

STEER

HEPA Filter Label – This label is used to inform the users of the HEPA filter replacement requirement. This label is located on the air filter housing.

BRAKE Label– This label is used to inform the users of the

Brake Caster and is located on the lower cabinet side panel above the brake caster.

STEER Label– This label is used to inform the users of the

Steer Caster and is located on the lower cabinet side panel above the steer caster.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

16 Effective Date: 07.23.2014

CN No. 14289

Consumables and Accessories

Warning: Use only Dornoch-approved components and enzyme. DO NOT modify any component or accessory.

The following is a list of consumables and accessories associated with the cart.

• TP-DL2800 – Single Use Canister Lid

• UL-CL500 – Single Use Cart Manifold

• UL-IV100 – Manual IV Pole Assembly – Mounts to the back panel of the unit and can hold up to

6000mL of fluid.

• UL-PP100 – Powered IV Pole Assembly – Mounts to the back of the unit and can hold up to

12000mL of fluid.

• UL-SF100 – Ultra Smoke Filter – Placed between the vacuum port on the lid and the vacuum hose to protect the cart vacuum supply.

• UL-ST100 – Ultra Specimen Trap – Specimen collection device designed for the cart. Placed on the port on the lid and then suction tubing connected to the top of the trap.

• UL-CB100 – Ultra Clot Buster – Specially formulated chemical to either prevent clotting during a rich blood procedure or to break up any clots found during processing of the cart.

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

17 Effective Date: 07.23.2014

CN No. 14289

Company Information

DMS has been helping healthcare facilities responsibly manage infectious fluid waste since 1997. Starting out as the premier manufacturer for fluid waste management systems in the country, we now have installations in leading healthcare facilities nation-wide.

DMS was founded in 1995. Products are manufactured and shipped from a centralized facility and supported by a nation-wide sales force. The company’s principals include individuals with extensive backgrounds in product development, management, production, and sales of medical/surgical products. Jim Dunn, a successful inventor who served as an operating room nurse for 20 years, helped design DMS products.

Contact Information

Dornoch Medical Systems

S h i p p i n g Ad d r es s :

200 NW Parkway

Riverside, MO 64150

O f f i c e C o r r es p o n d en c e :

P.O. Box 681656

Riverside, MO 64168

Toll free: 1-888-466-6633

Phone: 1-816-505-2226

Fax: 1-816-505-1050

Web: www.zimmer.com

Ultra Cart Instructions for Use Manual

GD-99022 Rev. H

18 Effective Date: 07.23.2014

CN No. 14289

advertisement

Key Features

  • Reusable fluid collection reservoirs
  • Enhances or replaces wall vacuum
  • Up to 52-liter (Quad) or 33-liter (Duo) capacity
  • Single-use lids/manifolds
  • Integrated vacuum pump
  • Touchscreen control panel
  • Optional IV pole
  • Wall vacuum port
  • High flow suction mode
  • Early warning and overflow protection system

Frequently Answers and Questions

What is the intended use of the Transposal Ultra Cart?
The Transposal Ultra Cart is intended to collect and dispose of liquid waste within Hospital Operating Rooms, Pathology Labs, Surgical Outpatient Centers, and Doctor's Offices.
What are the different models of the Transposal Ultra Cart?
The Transposal Ultra Cart is available in two models: the Ultra Quad (UL-QD2800 / UL-QD500) and the Ultra Duo (UL-DU2800 / UL-DU500). The Ultra Quad has four reservoirs and a capacity of 52-liters, while the Ultra Duo has two reservoirs and a capacity of 33-liters.
How do I use the Transposal Ultra Cart?
The Transposal Ultra Cart is designed to be used in conjunction with an Ultra Evacuation Unit. Once the cart is connected to the evac, the reservoirs can be emptied, cleaned, and rinsed automatically. Then, the cart can be wiped down and placed back into use.
What kind of fluids can be collected in the Transposal Ultra Cart?
The Transposal Ultra Cart is designed to collect surgical fluids, but it can also be used to collect other types of liquid waste.
What is the maximum vacuum level obtainable with the Transposal Ultra Cart?
The maximum vacuum level obtainable with the Transposal Ultra Cart is 0-600mmHg. If the cart is connected to a hospital suction system, the obtainable vacuum can be higher than 700mmHg.
How often does the Transposal Ultra Cart need to be processed?
The Transposal Ultra Cart should be processed with the evac once every 16 hours.
How do I clean the Transposal Ultra Cart?
The Transposal Ultra Cart should be wiped down with an approved bleach wipe or solution after each use. A full soak cycle should be performed once per week to ensure that the cylinders remain clean.
What is the difference between the Ultra Quad and Ultra Duo models?
The primary difference between the Ultra Quad and Ultra Duo models is the number of reservoirs. The Ultra Quad has four reservoirs, while the Ultra Duo has two reservoirs.
What is the purpose of the high flow suction mode?
The high flow suction mode can be activated to provide a boost in suction to the selected reservoir when more suction is needed.
How do I use the wall vacuum port on the Transposal Ultra Cart?
The wall vacuum port can be used to connect the cart to an external vacuum source in the event that the cart's internal vacuum pump fails or if a very quiet operation is desired.