Vacuum Gauge VACUU-VIEW (extended) - Manual

Vacuum Gauge VACUU-VIEW (extended) - Manual
Technology for Vacuum Systems
Vacuum gauge
VACUU·VIEW
VACUU·VIEW extended
Instructions for use
Original instructions
N°: 999293_EN
Imprint
Original instructions
Keep for further use!
This manual is only to be used and distributed in its complete and original form. It is strictly
the users’ responsibility to check carefully the validity of this manual with respect to his product.
Manufacturer:
VACUUBRAND GMBH + CO KG
Alfred‑Zippe‑Str. 4
97877 Wertheim
GERMANY
Phone:
ƒƒ Head office +49 9342 808‑0
ƒƒ Sales
+49 9342 808‑5550
ƒƒ Service
+49 9342 808‑5660
Fax:
E‑Mail:
Web:
+49 9342 808‑5555
[email protected]
www.vacuubrand.com
Thank you for purchasing this product from VACUUBRAND GMBH + CO KG. You have chosen a modern and technically high quality product.
2
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Content
TABLE OF CONTENT
1 Introduction
5
1.1 User information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 About this document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2.1
1.2.2
1.2.3
1.2.4
Display conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Handling instructions (action steps) . . . . . . . . . . . . . . . 7
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Term definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Safety instructions
8
2.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Improper use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.1 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.2 Personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.3 ATEX category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4 Proper disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3 Product description
12
3.1 VACUU·VIEW vacuum gauge . . . . . . . . . . . . . . . . . . . . . . 13
3.1.1 Designs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.2 Device view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2 Application example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Connection and operation
16
4.1 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1.1 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1.2 Vacuum connection . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1.3 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.1.4 DCP 3000, CVC 3000, VACUU·BUS® . . . . . . . . . . . 20
4.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.2.1 Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.2.2 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.3 Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5 Operation
29
5.1 Vacuum measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.2 Pressure reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
3
Content
5.3 Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3.1 Sensor adjustment, in general . . . . . . . . . . . . . . . . . . 30
5.3.2 Adjustment at atmospheric pressure . . . . . . . . . . . . . 31
5.3.3 Adjustment to reference pressure . . . . . . . . . . . . . . . 32
5.3.4 Adjustment under vacuum . . . . . . . . . . . . . . . . . . . . . 33
6 Service menues
34
6.1 Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.1.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.1.2 Update gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.2 Factory Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.3 Submenu Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7 Resolving problems
40
7.1 Error display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2 Fault – Cause – Remedy . . . . . . . . . . . . . . . . . . . . . . . . . . 41
8 Cleaning
44
8.1 Housing surface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
8.2 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
9 Appendix
45
9.1 Technical information . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
9.2 9.3 9.4 9.5 9.6 9.7 4
9.1.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
9.1.2 Wetted materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
9.1.3 Rating plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Overview menu structure Vacuu·View (extended) . . . . . . 55
EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . 56
Declaration of Conformity – China RoHS 2 . . . . . . . . . . . 57
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Introduction
1 Introduction
This manual is part of your product.
1.1 User information
Safety
Instructions for use
and
safety
ƒƒ Read
this manual thoroughly and completely before using the
produkt.
ƒƒ Keep
this manual in an easily accessible location.
ƒƒ Proper use of the product is essential for safe operation. Comply
with all safety instructions provided!
ƒƒ In
addition to this manual, adhere to any relevant local accident
prevention regulations and comply with industrial safety regula‑
tions.
General
General
information
of the term VACUU·VIEW mostly the term Gauge or
Vacuum gauge is used in this manual, in order to make the text
more readable.
ƒƒ Instead
ƒƒ The
illustrations in this manual are provided as examples in
order for a better understanding.
ƒƒ They
are intended to aid in your understanding of the proper
use of the product.
Contact
Contact us
ƒƒ Please
ask for replacement in case of an incom‑
plete manual or download the manual on our website:
www.vacuubrand.com
ƒƒ Contact
us regarding any questions about this product, if you
need further information, or to provide us with feedback.
ƒƒ When
contacting our Customer Service Department, please be
sure to have the correct type and serial number of your product
at call  see Rating plate on the product.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
5
Introduction
Copyright
Copyright ©
The content of this manual is protected by copyright. Only copies
for internal use are allowed, e. g., for professional training.
© VACUUBRAND GMBH + CO KG
1.2 About this document
1.2.1 Display conventions
Warning levels
CAUTION
Indicates a potentially hazardous situation.
Disregarding the situation could result in slight or minor
injury or damage to property.
>>
Take
appropriate action to avoid dangerous situation!
NOTICE
Notice for a potentially harmful situation.
Disregarding the notice could lead to material damage.
Additional notes
IMPORTANT!
>>Information or specific use recommendation, which must be
observed.
>>Important information for proper operation.
>>Helpful tips and tricks
>>Additional information
6
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Introduction
1.2.2 Handling instructions (action steps)
Presentation
convention operating
steps
Action step (single step)
>>Do the described step.
55 Result of action
Handling instructions (multiple steps)
1. first step
2. next step
55 Result of action
Follow steps in the described order.
1.2.3 Abbreviations
Used
Abbreviations
abs.
ATM
(di)
di
DN
Gr.
hPa
KF
max
mbar
min
RMA-N°
Torr
VAC
absolute
Atmospheric pressure
Interior diameter
Nominal diameter
Size
Pressure unit, Hectopascal (1 hPa = 1 mbar =
0.75 Torr)
Small flange
Maximum value
Pressure unit, millibar (1 mbar = 1 hPa = 0.75 Torr)
Minimum value
Return Merchandise Authorization number
Pressure unit (1 Torr = 1.33 mbar = 1.33 hPa)
Vacuum
1.2.4 Term definition
Product specific
terms
VACUU·BUS®
CVC 3000
DCP 3000
Bus system by VACUUBRAND
Vacuum controller, controller
Vacuum gauge
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
7
Safety instructions
2 Safety instructions
The complete information of this chapter must be observed by all
persons working with the herein described product.
Use the product only when it is in proper working condition.
2.1 Intended use
Intended use
VACUU·VIEW is a laboratory instrument for the measure‑
ment of absolute pressure in the range of rough vacuum or as
VACUU·VIEW extended version for both measuring rough and
fine vacuum.
The gauge may only be used in non-explosive areas.
Any other use is considered to be improper use.
2.2 Improper use
Improper use
Improper use includes:
ƒƒ Using
the product contrary to its intended use.
ƒƒ Operation
with obvious malfunctions.
ƒƒ Operation
at inadmissible operating conditions.
ƒƒ Unauthorized modifications or repairs provided by the customer.
8
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Safety instructions
2.3 General safety instructions
2.3.1 Safety precautions
Safety precautions
>>Use the gauge only if you have understood its function and this
manual.
>>Please note that adhering process media can pose danger to
humans and the environment.
>>When handling with contaminated parts, follow the relevant
regulations and safety precautions.
>>Repairs are only allowed by the Service Department or your
local supplier.
IMPORTANT!
For all service works hazardous substances need to be
excluded.
>>Fill in the form Health and Safety Clearance thoroughly and
completely and confirm with your signature.
2.3.2 Personnel
IMPORTANT!
It is the owner's responsibility to observe the proper use of the
device.
>>Always be conscious of safety, and work in a safe manner.
>>Observe the owners‘ directives at work, the national accident
prevention regulations and occupational safety provisions.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
9
Safety instructions
2.3.3 ATEX category
Installation and explosive environment
Installation and operation in areas where explosive atmo‑
sphere can occur is not allowed.
ATEX equipment labeling
ATEX category
VACUUBRAND products which are labeled with ` have an
ATEX approval 3 G; i. e. flammable substances as mixture with
air: Pumped gas or vapor.
>Use
>
the VACUUBRAND product only when it is in proper
working condition.
ATEX category
and peripherals
ATEX approval1 is only valid for the internal, wetted parts
area (vacuum sensor), not for the environment of the gauge.
The ATEX category of the gauge is dependent on the con‑
nected peripheral devices. Peripherals and connect‑
ed devices to the gauge need to have the same or higher
ATEX approval. Without concordant categorization of pe‑
ripherals, the specified category of the VACUUBRAND
equipment loses its validity.
Prevent ignition
source
The use of gas ballast or the operation of venting valves is only
permitted if thereby explosive atmospheres normally do not occur
in the interior of the equipment or, if they do occur, are likely to do
so only infrequently and for a short period.
>If
> necessary vent with inert gas.
For more and detailed information about ATEX approval visit our
website: www.vacuubrand.com/Information-ATEX
1 compare with rating plate
10
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Safety instructions
2.4 Proper disposal
NOTICE
Risk of environmental damage due to incorrect disposal of
the product.
>>
Do
not dispose your product in household waste!
Electronic components are subject to hazardous waste
treatment and must only be disposed of by certified
specialists.
>>
Observe the national regulations for safe disposal and envi‑
ronmental protection.
>>
Receive detailed information for respective regulations from
your competent administrative authority.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
11
Product description
3 Product description
Goods arrival
Goods arrival
Check the shipment for transport damage and completeness.
>>Report any transit damage immediately to the supplier.
NOTICE
Condensate could damage the gauge.
A large difference in temperature between storage location and
installation location can cause condensation.
>>
Let
the product acclimatise for 3-4 hours before using it.
Included materials
Included materials
Gauge
VACUU·VIEW with 2 m connection cable
or
VACUU·VIEW extended with 2 m connection cable
Hose nozzle 10/6 G1/4” with O ring
Wall power supply plug* 30W 24V; with adapters and
2 m connection cable
Instructions for use
Safety Information for Vacuum Equipment
Original packaging
683220
683210
642474
612090
999293
999254
------
* not required when connected to a VACUU-BUS® compatible gauge or
controller.
12
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Product description
3.1 VACUU·VIEW vacuum gauge
Gauge description
and designs
The gauge as stand-alone version will be supplied with wall power
supply plug. The gauge includes a vacuum sensor and is equi‑
ped with an illuminated display for pressure display. The gauge is
highly, chemically resistant.
VACUU·VIEW completes the VACUU·BUS® accessories program.
For more demanding tasks the gauge can be used as an external
vacuum sensor when connected to a controller CVC 3000 or the
gauge DCP 3000 .
When operating with a DCP 3000 measured values can be stored
(data logger) and graphically displayed. Via the RS 232 interface
the pressure can be read by an external Computer.
3.1.1 Designs
VACUU·VIEW
With chemically resistant ceramic diaphragm sensor for precise
measurings in the range of rough vacuum. VACUU·VIEW pro‑
vides gas-independent pressure indication with precise capacitive
readout.
The material of the connection flange of VACUU·VIEW consists
of black PP and is therefore easy to distinguish from the extended
design.
VACUU·VIEW extended
The heavy duty combination of ceramic diaphragm sensor and
ceramic jacketed Pirani sensor ensures reliable readings in the
wide range from atmosphere down to rough until fine vacuum.
The gauge reliably measures in the complete measuring range.
The connection flange of VACUU·VIEW extended consists of
aluminium and has at the interior a PPS coating.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
13
Product description
3.1.2 Device view
Front and rear side
1
2
3
5
4
Side view
6
1 VACUU·VIEW
2 Display
3 Operating elements
4 Small flange KF DN16, inside thread G1/4”
`` Material: PP black  VACUU·VIEW
`` Material: Aluminium + PPS  VACUU·VIEW extended
5 Rating plate (here VACUU·VIEW)
6 Connection cable, 2 m
`` for connection to wall power supply plug
or
`` for VACUU·BUS® connection to CVC 3000 or DCP 3000
14
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Product description
3.2 Application example
 Example
VACUU·VIEW direct
installation
4
6
1
3
7
5
2
Wall power supply, wall power supply plug
Plug connector, connection cable (each 2 m)
Vacuum chamber, tank, apparatus
VACUU·VIEW vacuum gauge
(Stand-alone version including wall power supply plug)
5 Vacuum valve
6 Vacuum hose
7 Diaphragm pump, vacuum pump
1
2
3
4
Please observe the following points to get an optimal
measuring result:
>>Connect the gauge as close as possible to the apparatus.
>>If possible use the small flange for connection.
>>Connect the vacuum line with a cross-section as wide as
possible.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
15
Connection and operation
4 Connection and operation
4.1 Connection
4.1.1 Installation
IMPORTANT!
Never install the measuring instrument too close to hot equip‑
ment, like on top of a rotary vane pump, in order to avoid incor‑
rect measurements or damages to the sensor.
The gauge is intended for assembly directly to the apparatus
(application).
>>Observe all specifications for installation, connection and
operation according to technical data,
 see chapter 9.1 Technical information.
>>Also observe rating plate data.
>>Compare the permitted limits which are described in this
manual, with your actual application regarding operating
media, pressures, forces, moments, temperatures and volt‑
age.
Installation conditions
ƒƒ The
gauge has acclimatized.
ƒƒ Ambient
of use.
Consider installation
conditions
16
conditions are observed and are within the limitation
Limitation of use
Ambient temperature
Altitude, max.
(US)
50–104 °F
10–40 °C
3000 m
9840 ft
above sea level
above sea level
Relative humidity
30–85 %, non condensing
Protection type
IP 54
Avoid condensation or contamination by dust or liquids.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Connection and operation
4.1.2 Vacuum connection
IMPORTANT!
>>Maximum admissable pressure at vacuum sensor: 1.5 bar/
750 Torr (absolute).
>>Pollution and damages, especially at the flange, could affect
the measurement.
>>If possible assemble the vacuum gauge vertically; this pre‑
vents condensate accumulation.
Connection via small flange
Required connection material: Clamping ring, centering or cen‑
tering ring for KF DN16.
1. Remove dust cap.
2. Put the gauge with the centering onto the connection of the
apparatus  small flange KF DN16.
3. Fix the vacuum gauge with a clamping ring.
Connection via hose nozzle
Required connection material: Hose nozzle DN 6/10 mm G1/4”
with O ring and compatible hose clip (option).
1. Remove dust cap.
2. Screw the hose nozzle plus the o ring hand-tight into the
inside thread.
3. Use the hose nozzle to assemble the gauge to a vacuum
hose or directly at the apparatus.
4. Fix the vacuum hose, e. g., with a hose nozzle.
5. Fix the vacuum gauge.
IMPORTANT!
>>Use a stable vacuum hose that is suitable for the required
vacuum range. For fine vacuum range, flexible rubber hoses
for example are not useful because of possible gas emis‑
sion.
>>Connect hose tubes as short as possible.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
17
Connection and operation
4.1.3 Electrical connection
NOTICE
No wall power supply plug is required when connected
as VACUU·BUS® component, to a controller CVC 3000 or
gauge DCP 3000.
>>
Power supply of the gauge via VACUU·BUS®.
Wall power supply plug*
Wall power supply,
wall power supply
plug
* short-circuit-proofed
multi-voltage power
supply with integrated overload protection and changeable
mains plugs.
Prepare wall power supply plug
Prepare connection
1. Take the wall power supply kit out of the packaging.
2. Select the mains plug that fits to your mains socket.
3. Connect the mains plug to the metal contacts of the wall pow‑
er supply plug.
4. Slide the mains plug until it locks.
Remove mains plug
Remove mains plug
1. Press the locking knob on top of the wall power supply plug.
2. Remove the mains plug.
55 Another mains plug can be fixed.
18
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Connection and operation
Connect to mains
IMPORTANT!
>>Please install the power supply line in such a way, that no
damage can cause to the cable due to sharp edges, chemi‑
cals or hot surfaces.
1. Connect the mains connector from the gauge to the female
plug of the wall power supply.
2. Plug the wall power supply into the mains socket.
55 Green LED at wall power supply plug glows.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
19
Connection and operation
4.1.4 DCP 3000, CVC 3000, VACUU·BUS®
Meaning
VACUU·BUS® is a communication system for peripheral acces‑
sories that are connected to VACUUBRAND controller or gauge.
VACUU·BUS® components will be automatically detected by
CVC 3000/DCP 3000. By uniform connectors and Y adapter the
bus-system can be extended with up to 32 peripherals.
Connenct VACUU·VIEW as VACUU·BUS® component (client)
>>Plug the mains connector of VACUU·BUS® into the port on
the rear side of DCP 3000 or CVC 3000.
55 Mains supply via CVC 3000 or DCP 3000.
Plug connectors of the newest series have a guide tongue
for proper connection. Slide the connection into the guiding
groove on the rear side of CVC 3000 or DCP 3000.
Feature
VACUU·BUS
Address assignment
IMPORTANT!
When using the gauge as VACUU·BUS® component, e. g., when
connected to a CVC 3000, the controller detects the gauge auto‑
matically as vacuum sensor.
First perform address assignment at CVC or DCP, when working
with several VACUU·VIEW gauges of the same type.
For detailed descriptions about address assignment
 see online manual 999151 (CVC 3000).
Address assignment
Vacuubus component
Address-N°
VACUU·VIEW
VACUU·VIEW extended
Referenzsensor
20
1–4
1–4
1–4
Abbreviations
in CVC/DCP
VSK _
VSP _
Ref._
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Connection and operation
Connection example VACUU·BUS® mit VACUU·VIEW
 Example
VACUU·VIEW
as VACUU·BUS®
component (client)
2
3
1
1
2
3
CVC 3000* (or DCP 3000*)
Connection cable (power supply via VACUU·BUS)
VACUU·VIEW – Rough vacuum measurement
or
VACUU·VIEW extended – Rough- until fine vacuum measure‑
ment
* Options with CVC 3000 or DCP 3000:
`` using the RS 232 interface with an external application for
reading the measuring data,
`` with connection of several VACUU·VIEWs, address assign‑
ment and differential pressure measurement.
Y adapter
Up to 4
sensors VACUU·VIEW type and
4 sensors VACUU·VIEW extended type and
4 sensors VACUU·VIEW type -> reference sensor for
differential pressure
measurement
... can be connected to CVC 3000 or DCP 3000 by
VACUU·BUS®-Y adapters.
 see also: Address assignment on page 20
Please regard the maximum total cable length of 30 m.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
21
Connection and operation
4.2 Operation
The operation is limited to the pre-settings about pressure unit
and display settings, such as the sensitivity of the touch panel.
Aside from that, version information can be displayed and factory
settings or updates can be loaded. Operation of the gauge is un‑
necessary during the running measurement.
4.2.1 Display elements
Icons on the display show which key needs to be pressed to ac‑
cess a menu or submenu.
Meaning display and symbols
 Examples
Icons and
illustrations on the
display
1
2
5
3
4
6
7
1
2
3
4
5
6
7
22
Title bar
Marking = active selection
. . . = Refers to submenu
Press key arrow down
Adapt value with up or down
PressEnter
PressEsc
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Connection and operation
4.2.2 Operating elements
The front side of the gauge consists of a glass with a touch-sensi‑
tive control pad including 4 keys.
Operating elements
touch keys
Key
Meaning
Arrow up
`` Menu selection
`` Navigate upwards
`` Increase value
Arrow down
`` Menu selection
`` Navigate downwards
`` Decrease value
Enter
`` Confirm entry
`` Call up menu
`` Confirm value
Escape
`` Abort action or exit menu
`` Return to previous menu
`` Return to pressure display
NOTICE
Damage of the glass surface.
Pointed or edged objects could damage the glass surface of
the gauge.
>>
Operate
finger.
the vacuum gauge only by a suitable touch stylus or
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
23
Connection and operation
4.2.3 Menu structure
The display with text Menu in the footer can be called up by any
touch key. The menu language is English.
VACUU·VIEWMenus
Menu structure
1
or press any key
2
3
Pressure reading
2 Call-up selection menus
3 Menus
`` Settings
`` Info
`` Service
1
Without any action, the display returns automatically to
pressure display.
``Menu  return to basic screen after approximately
5 seconds.
``Submenu  return to basic screen after approximately
20 seconds.
``Submenu Diagnostics  return to basic screen after
approximately 60 seconds.
24
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Connection and operation
Menu Settings
Adjustment option
for Settings
Menu Settings
Unit
Adjustment
Default settings pressure unit
`` mbar
`` hPa
`` Torr
Sensor adjustment
`` ATM (700-1080 mbar)
`` Vacuum (0 mbar)
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
25
Connection and operation
Menu Settings
Touch
Display
Adjust sensivity of the touch panel
`` High
`` Media
`` Low
Adjust brightness and contrast
`` Backlight: 10% – 100 %
`` Contrast: 10% – 100 %
Menu Info
Call-up version
information
Menu Info
Info
Gauge type
`` VACUU·VIEW
`` VACUU·VIEW extended
Gauge version (display example)
`` Hardware: VB.xx
`` Firmware: V1.xx
26
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Connection and operation
Service menu
Service menus
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
27
Connection and operation
Service menu
Service menus
Update
To perform software updates
`` Start update
Factory
Settings
Reset to factory settings
`` reset to Factory Settings
Diagnostics*
from software
version V1.04
Call-up diagnostics menu
`` Adjustment
`` Measurements
`` Raw Values
from software
version V1.05
`` Other Values
* Menu for our Service department for error determination.
The display of diagnostics sub-menus for VACUU·VIEW
and VACUU·VIEW extended do vary.
28
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Operation
5 Operation
5.1 Vacuum measurement
Directly after mains connection the gauge VACUU·VIEW
(extended) starts measuring and displaying pressure reading.
Also when conneted to a VACUU·BUS® system, pressure read‑
ing starts immediately. The measurement starts immediately after
switching on CVC 3000 or DCP 3000 .
The gauge is intended for continuous operation.
Warm-up (warm-up times)
IMPORTANT!
Warm-up times
>>Observe the warm-up times until full measurement precision
is reached.
VACUU·VIEW
Rough vacuum
`` 3 minutes
VACUU·VIEW extended
Fine vacuum
`` 15 minutes
5.2 Pressure reading
Display with pressure reading
Pressure reading,
basic display
1
2
1
Pressure reading
VACUU·VIEW
(min./max.)
`` 0.1–1100 [mbar/hPa]
`` 0.1–825 [ Torr
VACUU·VIEW extended
`` 0.001–1100 [mbar/hPa]
`` 0.001–825 [Torr]
2 Pressure unit (mbar, hPa, Torr)
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
29
Operation
5.3 Adjustment
NOTICE
For pressure < 5 mbar the VACUU·VIEW extended
measures the pressure-dependent thermal conductivity of
gas.
The gauge is factory adjusted for air. For gases with higher
mass, this may (<5 mbar) result in incorrect measurements.
· light gases: Pressure reading increased
· heavy gases: Pressure reading decreased
>>
If
required perform the adjustment with the gas that is to
measured. H2, He, Ar, CO2.
The adjustment is only possible after the warm-up time of
the gauge has been completed.
>>
Perform
adjustment only when the gauge is ready for use.
5.3.1 Sensor adjustment, in general
Adjustment is not part of the everyday operation. Perform
adjustment only when the measured values differ from reference
normal or when irregularities in pressure reading emerge.
IMPORTANT!
Only perform sensor adjustment during operation with a fully
warmed-up sensor (consider warm-up time 15–20 minutes).
Precondition for a proper sensor adjustment is the connection
to a reliable source of vacuum, e. g., adjustment of a
VACUU·VIEW extended with a high vacuum pump and a precise
reference gauge.
In case of a polluted vacuum system, e. g., with oil, particles
or humidity could contaminate the sensor and cause
mismeasurements and/or falsify sensor adjustment.
>>Clean polluted sensors before adjustment  see chapter: 8
Cleaning.
30
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Operation
5.3.2 Adjustment at atmospheric pressure
Perform sensor adjustment at atmospheric pressure
Sensor adjustment
at atmospheric
pressure
1. Call up the menu Settings/Adjustment.
55
2. If necessary, correct the value to the precise
atmospheric pressure in accordance with the reference
gauge by pressing the arrow keys.
3. Press Enter to confirm entry.
4. Subsequently confirm safety prompt by pressing Enter.
55 The pressure value will be stored automatically after con‑
firming the safety prompt.
55 VACUU·VIEW (extended) adjusted to atmospheric pres‑
sure.
VACUU·VIEW already displays the actual pressure.
Normally, only corrections in the range of ± 5 are required.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
31
Operation
5.3.3 Adjustment to reference pressure
Adjustment VACUU·VIEW to reference pressure
Adjustment at
reference pressure
1. Connect the gauge VACUU·VIEW to a vacuum pump which
pumps to a precise vacuum, e. g., down to 2 mbar.
IMPORTANT!
>>Check the accuracy of the ultimate vacuum with a calibrated
reference vacuum gauge.
2. Call up the menu Settings/Adjustment. The value on the dis‑
play should be 2 referring to the
reached vacuum of the vacuum pump.
3. If necessary, correct the value for reference pressure 2 by
pressing the arrow keys.
55
4. Press Enter to confirm entry.
5. Subsequently confirm safety prompt by pressing Enter.
55 VACUU·VIEW adjusted to reference pressure.
The adjustment to a reference pressure should only be
carried out when this pressure is accurately and reliably
known.
We recommend the adjustment to 0 mbar by using a high
vacuum pump (end vacuum < 0,1 mbar)
 see also 5.3.4 Adjustment under vacuum
32
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Operation
5.3.4 Adjustment under vacuum
NOTICE
The adjustment at vacuum for VACUU·VIEW extended
always occurs to the final measured value with 0 mbar.
An adjustment to a reference vacuum value is not possible.
>>
Pump
down to an ultimate vacuum as low as possible.
VACUU·VIEW (extended) adjusted under vacuum.
Adjustment under
vacuum
1. Connect the gauge VACUU·VIEW extended to a high vacu‑
um pump which pumps to a precise end vacuum lower than
< 10-3 mbar
or
connect the gauge VACUU·VIEW to a high vacuum pump
which pumps to a precise vacuum lower than < 0,1 mbar.
>>Check the accuracy of the ultimate vacuum with a calibrated
reference vacuum gauge.
IMPORTANT!
2. Please wait until the high vacuum pump has reached ultimate
vacuum and until the gauge has completed warm-up.
3. Call up the menu Settings/Adjustment. The value on the dis‑
play should be 0 .
55
0
4. Press Enter to confirm entry.
5. Subsequently confirm safety prompt by pressing Enter.
55 VACUU·VIEW (extended) adjusted under vacuum.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
33
Service menues
6 Service menues
6.1 Update
NOTICE
Damage to the gauge due to incorrect execution of
updates.
Gauges may be damaged by incorrect or unauthorized proce‑
dure.
>>
Please
note that you must have the necessary authorizations
and basic knowledge for an update.
>>
Always connect only one gauge to the VACUU·BUS® Update
kit and run updates individually.
6.1.1 Preparation
Connect VACUU·BUS®-Update-Kit1
Connect Update kit
VACUU·BUS® Update kit
1. Disconnect VACUU·BUS® connector from gauge and wall
power supply plug.
2. Connect wall power supply plug, gauge and VACUU·BUS®
Update kit one after the other to the Y adapter.
3. Connect the VACUU·BUS® Update kit to a PC or laptop (=
end device).
55 With the first connection of the Update kit the operating
system displays an information prompt, e. g., Hardware
Wizard displays Found New Hardware.
4. Plug the wall power supply into the mains socket.
1
34
 see chapter 9.2 Ordering information
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Service menues
Download and start update software2
Download update
file
1. Download the ZIP file with updates for your device from
VACUUBRAND homepage onto your computer.
2. Unzip the ZIP file:
Right-click on the ZIP file  Extract to... .
3. Open the extracted folder.
Start application
(exe)
4. Start the update application by double-click on the
icon
.
55
Update application
5. Select the required COM port from the drop down list to which
the VACUU·BUS® Update kit is connected: USB Serial Port.
55
 Example
If no COM port is displayed for selection, you need to
install the hardware driver for RS485/USB.
>>Ask the responsible staff member of your IT department to
install the required hardware driver (CD-ROM is included to
the Update kit).
>>
First,
please read the description in chapter
6.1.2 Update gauge carefully before proceeding.
2
Valid for both VACUU·VIEW as well as VACUU·VIEW extended.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
35
Service menues
6.1.2 Update gauge
IMPORTANT!
>>Please regard that you ever need to use the gauge and then
promptly switch to the Updater window on your monitor.
>>Also note that the time window in which the updater applica‑
tion searches a connected device, takes 20 seconds.
Perform software update
1. At your VACUU·VIEW gauge call-up the menu
Service/Update.
Perform software
update
55
2. Click on the button
.
55 Detection for connected device starts.
3. Press Enter key on the gauge within the time the progress
bar is displayed.
55 Update starts.
55 Within the update progress the display is switched off.
36
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Service menues
Perform software
update
4. Wait until the update is completely loaded.
55 The following prompt appears in the update application:
55 Update succeeded.
55 The display of the gauge is switched on again.
IMPORTANT!
>>Regard the warm-up time also after updating the gauge.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
37
Service menues
6.2 Factory Settings
Reset to factory settings
Factory settings
1. Call up menu Service/Factory Settings .
55
2. Press Enter to start Reset.
55 Factory settings loaded.
38
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Service menues
6.3 Submenu Diagnostics
From software version V1.04 a diagnostic menu is included for
VACUU·VIEW® gauges.
Call-up diagnostics menu
Diagnostics menus
>>Call up the menu Service/Diagnostics ....
55
Call-up sub-menus
1. Press arrow key to select one of the sub-menus.
2. Press Enter, to call up the selected sub-menu.
When contacting our Service Department the Diagnostics values
might help to localize an error.
IMPORTANT!
>>Please send us photos of the displayed values. As well
measured under vacuum as at atmospheric pressure.
>>Email the photos to: [email protected]
>>Details about the product type and serial number from the
rating plate are also required.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
39
Resolving problems
7 Resolving problems
CAUTION
Malfunction because of incorrect repair by the
customer.
The gauge is not intended for the repair by customer.
>>
Never
open the vacuum gauge.
>>
If the gauge is defective, please send it to our
Service Department or your local supplier.
7.1 Error display
For error indication the backlight of the display switches to red.
Example error display
Error display
(red)
1
2
1
2
40
Error display with red backlight
Error indication as clear text
`` Over Pressure – pressure overload
`` Under Range – measuring fallen below lowest range
`` Sensor Failure – sensor error
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Resolving problems
7.2 Fault – Cause – Remedy
Resolving problems
Fault
``Possible cause
3Remedy
3
Over Pressure
``Pressure too high.
33
Vent the system or appara‑
``Measuring range
tus.
33
Reduce pressure.
33
Perform sensor adjustment.
33
Call-up diagnostics menu
and inform our Service about
the status.
Under Range
``Below measuring
33
Perform sensor adjustment.
Sensor Failure
``Defective sensor.
33
Call-up diagnostics menu
exceeded.
range (negative pres‑ 33
Call-up diagnostics menu
sure reading).
and inform our Service about
the status.
and inform our Service about
the status.
33
Send in.
Front glass bro‑ ``Incorrect cleaning
agent used.
ken
``Mechanically
damaged.
33
Send in.
Readings de‑
``Sensor measures
incorrectly.
viate from the
reference stan‑ ``Reading of other gas
dard
than air.
33
Perform sensor adjustment.
Adjustment ----
33
Perform sensor adjustment
with the gas to be measured.
33
Call-up diagnostics menu
and inform our Service about
the status.
``A for sensor adjust‑
33
Wait approximately 5 –
ment inadmissible
10 minutes until the sensor is
pressure is reached
ready for use.
(no adjustment possi‑ 33
Perform adjustment at
ble in between pres‑
> 700 mbar or < 20 mbar.
sure range 20 – 700
33
For adjustment connect
mbar).
a vacuum pump with pre‑
cise vacuum and then move
pump down to the possible
pressure range.
Menu Settings/ ``Sensor warm-up not
completed.
Adjustment
warm up
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
33
Wait approximately 5 –
10 minutes until the sensor is
ready for use.
33
Subsequently perform sensor
adjustment if necessary.
41
Resolving problems
Possible error during
an update
Fault
``Possible cause
3Remedy
3
No display
``No power, wall power
33
Check power supply, connect
supply plug defective
or not connected.
wall power supply plug to a
socket.
USB serial port ``No driver installed for 33
Install the driver for the
RS485/USB interface.
RS485 / USB interface from
not detected
the supplied CD-ROM or via
``Driver for RS485/USB
Internet download.
(No COM-port
interface not enabled.
for selection
33
Enable connection at the
listed)
Device Manager.
Display
``Update running = no
error.
switched off,
no device reac‑ ``Update failed.
tion
33
Wait until the update is com‑
Update failed
33
Connection: Do not inter‑
``Update failed.
pleted.
33
Repeat Update:
1.Remove VACUU·BUS® con‑
nector from Y adapter.
2.Click on Update button in the
update application.
3.Reconnect VACUU·BUS®
connector to Y adapter within
detection for connected
device (progress bar).
33
Send in the gauge, if the
problem persists.
rupt connection: Device 
Update kit  End device.
Update kit  End
device, connection
33
Check power supply, connect
interrupted.
wall power supply plug to a
socket.
``No power, wall power
supply plug defective 33
Connect the device that
or not connected.
needs to be updated  see
also illustration in chapter
``No connected device
6.1.1 Preparation.
found.
``Connection: Device
42
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Resolving problems
Technical support
>>To identify errors and potential remedies, please refer to the
troubleshooting
Fault – Cause – Remedy table:
For technical help or in case of errors you need additional help for,
please contact your local supplier or our Service1 department.
>>For practical help in contact with our Service department also
use 6.3 Submenu Diagnostics on page 39
1 -> Phone: +49 9342 808‑5660, Fax: +49 9342 808‑5555, [email protected]
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
43
Cleaning
8 Cleaning
Clean the sensor to remove malfunctions that are caused by a
polluted sensor. We recommend to clean the sensor before ad‑
justment.
IMPORTANT!
This chapter does not contain descriptions for the decontam‑
ination of the product. This chapter describes only simple
cleaning and care measures.
8.1 Housing surface
Clean surface
Clean surface
>>Clean polluted surface with a clean, slightly wetted cloth. To
moisten the cloth we recommend water or mild soap.
8.2 Sensor
Clean sensor
Clean sensor
1. Fill a small amount of solvent via flange into the gauge, e. g.,
cleaning solvent.
2. Let the solvent react for a few minutes.
3. Pour the solvent.
55 Dissolved substances or discolorations in the solvent are
possible.
4. Repeat this procedure until no more pollutants are in the sol‑
vent.
5. Air or ventilate the gauge until the internal chamber has dried.
6. Re-adjust the sensor.
44
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Appendix
9 Appendix
9.1 Technical information
Designs
Vacuum gauge – Rough vacuum
VACUU·VIEW
Vacuum gauge – Fine vacuum
VACUU·VIEW extended
9.1.1 Technical data
Technical data
Ambient conditions
Ambient temperature
Working temperature
Transport and storage tem‑
perature
Altitude, max.
10–40 °C
10–40 °C
(US)
50–104 °F
50–104 °F
-10–60 °C
14–140 °F
3000 m
above sea level
9840 ft
above sea level
Relative humidity
30–85 %, non condensing
Avoid condensation or contamination by dust, liquids or
corrosive gases.
Wall power supply plug
Input voltage
Frequency
Input current, max.
Output voltage,
short-circuit-proof
Output current, max.
Cable length, approx.
Dimension
Weight
Mains plug
Electrical data gauge
Supply voltage, max.
Capacity, max.
Protection type
Interface
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
90–264 VAC
50–60 Hz
0,8 A
(US)
90–264 VAC
50–60 Hz
0.8 A
24 VDC
24 VDC
1,25 A
1.25 A
2 m
79 in.
108 mm x 58 mm x 34 mm
4.3 in. x 2.3 in. x 1.4 in.
300 g
0.66 lb
AC, changeable: CEE/CH/UK/US/AUS/CN
24 VDC
1,3 W
IP 54
(US)
24 VDC
1.3 W
VACUU·BUS®
45
Appendix
Technical data
Vacuum data
VACUU·VIEW
Measuring range,
absolute
Accuracy of
measurement
Measuring principle
(US)
1100–0,1 mbar
825–0.1 Torr
< ±1 mbar/hPa/Torr, ±1 digit
(after adjustment, constant temperature)
Ceramic diaphragm (alumina),
capacitive, gas type independent, absolute
pressure
< ±0,07 mbar/K
< ±0.05 Torr/K
internal
internal
Temperature coefficient
Sensor
Max. admissible pres‑
1,5 bar
1125 Torr
sure, absolute
Max. admissible media temperature (gas) non-explosive atmosphere:
80 °C
176 °F
momentarily
40 °C
104 °F
Continuous operation
Max. admissible media temperature (gas) ` atmosphere:
momentarily
40 °C
104 °F
40 °C
104 °F
Continuous operation
VACUU·VIEW extended (only deviant items)
1100–0.001 mbar
Measuring range,
absolute
1100–0.001 hPa
< ±0.2 mbar/K
Temperature coefficient
825–0.001 Torr
< ±0.15 Torr/K
Resolution at pressure
ranges
0.001 hPa (0.001–0,1 hPa)
0.01 hPa (0.1– hPa)
0.1 hPa (1–10 hPa)
1 hPa (10–1100 hPa)
Accuracy of measurement ±15% of indicated value in the range from
0.01–5 mbar/hPa/Torr, ±3 mbar for > 5 mbar
Ceramic diaphragm sensor
Measuring principle
+ ceramic jacketed Pirani sensor
Connections
Cable (length)
Plug connector
Vacuum connection
46
2 m
79 in.
®
VACUU·BUS
Small flange KF DN 16
Hose nozzle DN 6/10
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Appendix
Technical data
Display
Type
Brightness control
Pressure reading
LC display (LCD)
yes
switchable: mbar, Torr, hPa
(US)
Weight and dimensions*
190 g
0.4 lb
Weight
103 mm x 62 mm x 50 mm
Dimension sheet
4 in. x 2.4 in. x 2 in.
* without wall power supply
9.1.2 Wetted materials
Wetted materials
Component
Wetted materials
VACUU·VIEW
Vacuum sensor
Metering box + small flange
Sealing ring at the sensor
Hose nozzle
O ring (KF 16)
Aluminium oxide ceramics
PP
chemically resistant fluorelastomer
PPS, glass fiber reinforced
FKM
VACUU·VIEW extended (only deviant items)
Metering box + small flange
Sealing ring at the sensor
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
PPS, glass fiber reinforced
FFPM
47
Appendix
9.1.3 Rating plate
>>In case of malfunction, please note type and serial number
on the rating plate.
>>When contacting our service department, name us product
type and serial number. With this information we can offer
selective support and advice for your product.
Rating plate VACUU·VIEW, in general
Rating plate
Manufacturer
Type/Year of contruction/Month
Serial number
Pressure measuring range
Voltage supply
compatible to VACUU·BUS
ATEX specification*
Address
* Group and category, marking G (gas), type protection, explosion group, temperature class
(additionally see : Approval for ATEX equipment).
48
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Appendix
9.2 Ordering information
Ordering information
accessories
Ordering information
spare parts
Vacuum gauge
VACUU·VIEW, ready-for-use inclusive wall power supply
plug
VACUU·VIEW extended, ready-for-use inclusive wall
power supply plug
Order N°
Accessories
Vacuum hose DN 6 mm (l = 1000 mm)
Vacuum hose DN 10 mm (l = 1000 mm)
PTFE hose KF DN 16 (l = 1000 mm)
Stainless steel tubing KF DN 16 (l = 1000 mm)
Fitting for PTFE tube 10/8 mm
DAkkS calibration with first delivery
DAkkS recalibration
VACUU·BUS®/USB Update kit
Order N°
Spare parts
VACUU·VIEW
VACUU·VIEW extended
Hose nozzle 10/6 G1/4” with O ring
Wall power supply plug 30W 24V; with adapters
Y adapter VACUU·BUS
Extension cable VACUU·BUS, 2m
Wall duct VACUU·BUS
Instructions for use
Order N°
683220
683210
686000
686002
686031
673336
900214
900215
683230
635490
635489
642474
612090
636656
612552
636153
999293
Source of supply
International
sales offices and
specialized trade
Purchase original accessories and spare parts from your spe‑
cialized distributor or through international sales offices of
VACUUBRAND GMBH + CO KG.
>>Information about the complete product range are available
in the current product catalog.
>>For orders, questions about vacuum control and
optimal accessories, please contact your specialized
distributor or an international sales office of
VACUUBRAND GMBH + CO KG.
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
49
Appendix
9.3 Service
Service offer and
service range
Take advantage of the comprehensive service range of
VACUUBRAND GMBH + CO KG.
Service in detail
ƒƒ product
ƒƒ fast
guidance and practical solutions,
delivery of spare parts and accessories,
ƒƒ professional
ƒƒ immediate
ƒƒ Service
repairs processing,
on the spot (on request),
ƒƒ Calibration
ƒƒ return,
maintenance,
(DAkkS accredited),
disposal.
>Visit
>
our website for further information:
www.vacuubrand.com.
Servicing handling
Meet the
terms of service
1. Contact your local supplier or our Service Department.
2. Request a RMA number for your order.
3. Clean the product thoroughly and if necessary decontaminate
it professionally.
4. Please fill in this form Health and Safety Clearance complete‑
ly.
Return (reshipment)
5. Return your product including:
ƒƒ RMA-N°,
ƒƒ Repair- or service order,
ƒƒ Form Health and Safety Clearance,
ƒƒ Short error description.
>>Reduce downtime, speed up the service process. Please
keep the required data and documents ready when
contacting our Service Department.
``Your order can be quickly and easily processed.
``Hazards can be excluded.
``A short description or photos may help for error location.
50
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Appendix
9.4 Index
Index
A
Accessories . . . . . . . . . . . . . . . . . . 49
Action step . . . . . . . . . . . . . . . . . . . . 7
Address assignment . . . . . . . . . . . . 20
Adjustment at atmospheric pressure . .
31
Adjustment at reference pressure . 32
ATEX category . . . . . . . . . . . . . . . . 10
ATEX equipment labeling . . . . . . . . 10
Atmospheric pressure . . . . . . . . . . . 31
B
Basic display . . . . . . . . . . . . . . . . . . 29
C
Call-up diagnostics menu . . . . . . . . 28
Cleaning . . . . . . . . . . . . . . . . . . . . . 44
Clean sensor . . . . . . . . . . . . . . . . . 44
Clean surface . . . . . . . . . . . . . . . . . 44
Connection example for VACUU·BUS® 21
Connection plug vacuum gauge . . . 19
Connect to mains . . . . . . . . . . . . . . 19
Connect Update kit . . . . . . . . . . . . . 34
Contact . . . . . . . . . . . . . . . . . . . . . . . 5
Copyright © . . . . . . . . . . . . . . . . . . . 6
country-specific Interchangeable plug .
18
CVC 3000 . . . . . . . . . . . . . . . . . . . . . 7
D
Data logger . . . . . . . . . . . . . . . . . . . 13
DCP 3000 . . . . . . . . . . . . . . . . . . . . . 7
Designs . . . . . . . . . . . . . . . . . . . . . . 13
Device view . . . . . . . . . . . . . . . . . . 14
Diagnostics . . . . . . . . . . . . . . . . . . . 39
Diagnostics menus . . . . . . . . . . . . . 39
Download update file . . . . . . . . . . . 35
E
EC Declaration of Conformity . . . . . 56
Electrical connection . . . . . . . . . . . . 18
Error display . . . . . . . . . . . . . . . . . . 40
Examples of use: . . . . . . . . . . . . . . 15
F
Factory settings . . . . . . . . . . . . . . . 38
Factory Settings . . . . . . . . . . . . . . . 38
Fault – Cause – Remedy . . . . . . . . 41
Firmware . . . . . . . . . . . . . . . . . . . . . 26
Front and rear side . . . . . . . . . . . . . 14
G
Gauge description . . . . . . . . . . . . . 13
Goods arrival . . . . . . . . . . . . . . . . . 12
H
Handling instruction . . . . . . . . . . . . . 7
Health and Safety Clearance . . . . . 50
heavy gases . . . . . . . . . . . . . . . . . . 30
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
I
Improper use . . . . . . . . . . . . . . . . . . 8
Included materials . . . . . . . . . . . . . 12
incorrect readings . . . . . . . . . . . . . . 30
Installation . . . . . . . . . . . . . . . . . . . 16
Installation conditions . . . . . . . . . . . 16
Intended use . . . . . . . . . . . . . . . . . . . 8
Interchangeable plug . . . . . . . . . . . 18
L
light gases . . . . . . . . . . . . . . . . . . . 30
Limitation of use . . . . . . . . . . . . . . . 16
M
Menu language . . . . . . . . . . . . . . . . 24
Menu structure . . . . . . . . . . . . . . . . 24
multi-voltage power supply . . . . . . . 18
O
Operating elements . . . . . . . . . . . . 23
Ordering information . . . . . . . . . . . . 49
P
Perform software update . . . . . 36, 37
Peripherals and ATEX . . . . . . . . . . 10
Personnel (staff) . . . . . . . . . . . . . . . . 9
Power supply via VACUU·BUS . . . 21
Power supply via wall power supply
plug . . . . . . . . . . . . . . . . . . . . . . . . . 18
Preparation for update . . . . . . . . . . 34
Presentation convention operating
steps . . . . . . . . . . . . . . . . . . . . . . . . . 7
Pressure reading . . . . . . . . . . . . . . 29
Prevent ignition source . . . . . . . . . . 10
Proper disposal . . . . . . . . . . . . . . . . 11
R
Rating plate . . . . . . . . . . . . . . . 14, 48
Red display . . . . . . . . . . . . . . . . . . . 40
Reset . . . . . . . . . . . . . . . . . . . . . . . 38
Resolving problems . . . . . . . . . . . . 41
Return (reshipment) . . . . . . . . . . . . 50
S
Safety . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety precautions . . . . . . . . . . . . . . 9
Sales offices . . . . . . . . . . . . . . . . . . 49
Sensor adjustment (vacuum) . . . . . 33
Service menus . . . . . . . . . . . . . 27, 28
Service range . . . . . . . . . . . . . . . . . 50
Servicing handling . . . . . . . . . . . . . 50
Side view . . . . . . . . . . . . . . . . . . . . 14
Software version V1.04 . . . . . . . . . 39
Source of supply . . . . . . . . . . . . . . . 49
Spare parts . . . . . . . . . . . . . . . . . . . 49
Start application (exe) . . . . . . . . . . . 35
T
Technical data . . . . . . . . . . 45, 46, 47
Technical information . . . . . . . . . . . 45
Touch keys . . . . . . . . . . . . . . . . . . . 23
51
Appendix
Index
U
Update failure . . . . . . . . . . . . . . . . . 42
Update gauge . . . . . . . . . . . . . . . . . 36
Update kit (VACUU·BUS/USB) . . . . 49
Update software . . . . . . . . . . . . . . . 35
User information . . . . . . . . . . . . . . . . 5
V
VACUU·BUS®/USB Update kit . . . 49
Vacuum connection . . . . . . . . . . . . 17
VACUU·VIEW . . . . . . . . . . . . . . . . . 13
VACUU·VIEW as VACUU·BUS® com‑
ponent . . . . . . . . . . . . . . . . . . . . . . . 20
VACUU·VIEW extended . . . . . . . . . 13
Version information . . . . . . . . . . . . . 26
W
Warm-up times . . . . . . . . . . . . . . . . 29
Wetted materials . . . . . . . . . . . . . . . 47
52
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Appendix
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
53
Appendix
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
9.5 Overview menu structure Vacuu·View (extended)
55
Appendix
9.6 EC Declaration of Conformity
EU-Konformitätserklärung
EC Declaration of Conformity
Déclaration CE de conformité
Hersteller / Manufacturer / Fabricant:
VACUUBRAND GMBH + CO KG · Alfred-Zippe-Str. 4 · 97877 Wertheim · Germany
Hiermit erklärt der Hersteller, dass das Gerät konform ist mit den Bestimmungen
der Richtlinien:
Hereby the manufacturer declares that the device is in conformity with the
directives:
Par la présente, le fabricant déclare, que le dispositif est conforme aux directives:
2014/30/EU (EMV-RL), 2014/35/EU (Niederspannungs-RL),
2014/34/EU (ATEX-RL), 2011/65/EU (RoHS-2)
Messgerät / Vacuum gauge / Vacuomètre
Typ / Type / Type: VACUU·VIEW / VACUU·VIEW extended
Artikelnummer / Order number / Numéro d‘article: 683210, 683220, 635490, 635489
Seriennummer / Serial number / Numéro de série: Siehe Typenschild / See rating
plate / Voir plaque signalétique
Angewandte harmonisierte Normen / Harmonized standards applied /
Normes harmonisées utilisées: DIN EN 12100:2011, DIN EN 61010-1:2010
(Ed. 3), DIN EN 61010-1:2011, DIN EN 61326‑1:2013, DIN EN 1127-1:2011,
DIN EN ISO 80079-36:2016, DIN EN 50581:2013
Bevollmächtigter für die Zusammenstellung der technischen Unterlagen / Person
authorised to compile the technical file / Personne autorisée à constituer le dossier
technique: Dr. J. Dirscherl · VACUUBRAND GMBH + CO KG · Germany
Ort, Datum / place, date / lieu, date: Wertheim, 07.07.2017
ppa.
56
(Dr. F. Gitmans)
(Dr. J. Dirscherl)
Geschäftsführer / Managing Director /
Gérant
Technischer Leiter / Technical
Director / Directeur technique
VACUUBRAND GMBH + CO KG
Alfred‑Zippe‑Str. 4
97877 Wertheim
Tel.: Fax:
E‑Mail:
Web:
+49 9342 808‑0
+49 9342 808‑5555
[email protected]
www.vacuubrand.com
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
Appendix
9.7
Declaration
of Conformity
– China
RoHS 2
DECLARATION
OF CONFORMITY
– China RoHS
2
VACUUBRAND GMBH + CO KG has made reasonable efforts to ensure that hazardous mate‑
rials and substances may not be used in its products.
In order to determine the concentration of hazardous substances in all homogeneous materials of the
subassemblies, a “Product Conformity Assessment” (PCA) procedure was performed. As defined in
GB/T 26572 the “Maximum Concentration Value” limits (MCV) apply to these restricted substances:
0.1%
 Lead (Pb):
0.1%
 Mercury (Hg):
0.01%
 Cadmium (Cd):
0.1%
 Hexavalent chromium (Cr(+VI)):
0.1%
 Polybrominated biphenlys (PBB):
0.1%
 Polybrominated diphenyl ether (PBDE):
Environmental Protection Use Period (EPUP)
EPUP defines the period in years during which the hazardous substances contained
in electrical and electronic products will not leak or mutate under normal operating
conditions. During normal use by the user such electrical and electronic products will
not result in serious environmental pollution, cause serious bodily injury or damage to
the user’s assets.
The environmental Protection Use Period for VACUUBRAND products is 40 years.
此表格是按照SJ/T 11363‑2006中规定所制定的。
This table is created according to SJ/T 11363‑2006.
MATERIAL CONTENT DECLARATION FOR VACUUBRAND PRODUCTS
部件名称
Part name
包装
Packaging
塑料外壳 / 组件
Plastic housing / parts
真空油
Vacuum oil
电池
Battery
玻璃
Glass
电子电气组件
Electrical and electronic parts
控制器 / 测量设备
Controller / measuring device
金属外壳 / 组件
Metal housing / parts
电机
Motor
配件
Accessories
铅
汞
Pb
Hg
有毒有害物质或元素
Hazardous substances
镉
六价铬 多溴联
苯
Cd
Cr(+VI)
PBB
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
X
X
X
O
O
O
X
O
X
O
O
O
X
O
O
O
O
O
X
O
O
O
O
O
X
O
O
O
O
O
Declaration of Conformity – China RoHS 2
多溴二 环保期限标
苯醚
识
PBDE
EPUP
version 01 of April 2017 / aw
Copyright 2017
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
57
Appendix
注释: 此表格适用于所有产品。以上列出的元件或组件不一定都属于所附产品的组成。
Note: Table applies to all products. Some of the components or parts listed above may not be part of
the enclosed product.
O:
O:
表示该有毒有害物质在该部件所有均质材料中的含量均在GB/T 26572规定的限量要求以下。
Indicates that the above mentioned hazardous substance contained in all homogeneous mate‑
rials of the part is below the required limit as defined in GB/T 26572.
X:
X:
表示该有毒有害物质至少在该部件某一均质材料中的含量超出GB/T 26572规定的限量要求。
Indicates that the above mentioned hazardous substance contained in at least one of the ho‑
mogeneous materials of this part is above the required limit as defined in GB/T 26572.
除上表所示信息外,还需声明的是,这些部件并非是有意用铅(Pb)、汞 (Hg)、铬(Cd)、六价铬
(Cr(+VI))、多溴联苯(PBB)或多溴二苯醚(PBDE)来制造的。
Apart from the disclosures in the above table, the subassemblies are not intentionally manufactured or
formulated with lead (Pb), mercury (Hg), cadmium (Cd), hexavalent chromium (Cr+VI), polybrominated
biphenyls (PBB), and polybrominated diphenyl ethers (PBDE).
Products manufactured by VACUUBRAND may enter into further devices (e.g., rotary evaporator) or
can be used together with other appliances (e.g., usage as booster pumps).
With these products and appliances in particular, please note the EFUP labeled on these products.
VACUUBRAND will not take responsibility for the EFUP of those products and appliances.
Place, date:
Wertheim, 04/24/2017
ppa.
(Dr. F. Gitmans)
Managing Director
(Dr. J. Dirscherl)
Technical Director
VACUUBRAND GMBH + CO KG
Alfred‑Zippe‑Str. 4
97877 Wertheim
Germany
Declaration of Conformity – China RoHS 2
Tel.:
Fax:
E‑Mail:
Web:
+49 9342 808‑0
+49 9342 808‑5555
[email protected]
www.vacuubrand.com
version 01 of April 2017 / aw
Copyright 2017
58
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
N°: 999293_EN_VACUU•VIEW (extended)/07.07.2017
59
Manufacturer:
VACUUBRAND GMBH + CO KG
Alfred‑Zippe‑Str. 4
97877 Wertheim
GERMANY
Phone:
ƒƒ Head office +49 9342 808‑0
ƒƒ Sales
+49 9342 808‑5550
ƒƒ Service
+49 9342 808‑5660
Fax:
E‑Mail:
Web:
+49 9342 808‑5555
[email protected]
www.vacuubrand.com
Version: 999293_EN_VACUU•VIEW (extended) TDD2/V1.7 (07.07.2017)
© VACUUBRAND GMBH + CO KG – 2017