MicroCal generali UK 3.4
MICROCAL
Palm-top
Multifunction
Calibrators
Bulletin n. 08-30.1 (E)
TEST &
MEASUREMENT
Year 2000 compliant
•
•
•
•
•
Simultaneous
Measurement &
Generation
MicroCal 1+ MAV
MicroCal 1+
MicroCal 1 I.S.
MicroCal 2+
MicroCal 10+
mA
mV, V
°C, °F, ½
Ni100, Ni120
Tc, Pt100
Eurotron - One of the world leaders in high technology equipment : portable flue gas analyzers • high accuracy calibrators • non-contact IR thermometers
TEST &
MEASUREMENT
MICROCAL
To keep your test
and process
equipment accurately
calibrated for ISO
9000 compliance
Performance
The palm-top indicator-simulator series
MicroCal are multifunction instruments
designed to check and calibrate your test
and process equipment. MicroCal calibrators meet, in a modern and practical
way, the everyday needs of Quality and
Maintenance instrumentation engineers,
both in laboratory and fleld work.
Accurate, compact, rugged, easy to use;
the ideal solution for measuring and simulating: millivolts, volts, milliamperes,
ohms, thermocouples and resistance
thermometers.
Eurotron palm-top calibrators have been
developed using the most advanced microprocessor technology to grant high
accuracy on extended ranges and a powerful operating flexibility. The modular
firmware includes the algorithms for thermocouples and resistance thermometers in accordance with IEC, DIN and JIS
standards. Both IPTS68 and ITS90 temperature scale are memory stored and
can be selected using the keyboard. A
unique internal automatic Rj compensation system providing the accurate
input and output readings in a temperature range from -5°C to +50°C.
Further, external compensation is available with temperature
programmable from
-50°C to +100°C.
The simulationmeasurement of
resistance and
temperature with
resistance thermometers uses
a special proprietary active
circuit
(Eurotron
Patent
206327).
The calibrators are able
2
to source a single step value; to program up to 7 simulation programs using
up to 60 values; and to source programmable loops (start, step, end
values and time), using menu-driven
procedures.
The instruments can mathematically filter
unstable input signals, convert linear signals in technical units and linearize quadratic signals.
Selection of the operating function is
made using a rugged membrane keyboard which assures up to one million
operations per key. MicroCal use two
special thick film membrane “slidewires”
to set the simulated signal value.
Measured or simulated value is indicated
on a high quality alphanumeric LCD dot
matrix display which provides good contrast even in poor lighting conditions. The
instrument is powered by four Ni-MH
rechargeable batteries; an external battery
charger is supplied as standard accessory.
Quality System
Research, development, production, inspection and certification activities are
defined by methods and procedures of
the Eurotron Quality System. Eurotron
Quality System has been inspected and
certified for compliance to lSO 9001 by
Gastec, a Dutch notified body.
Five models
MicroCal palm-top series include 5 different calibrators to select the instrument more suitable for each specificy
laboratory and/or field applications.
High accuracy and stability are the key
and important technical specifications.
• MicroCal 1+ MAV:
One and two channels mA (active
and passive loop) and up to 10V
process signal calibrator
• MicroCal 1+:
13 thermocouples, 3 Rtds, mV up to
1 Volt, mA and ohms universal calibrator
• MicroCal 1 IS:
EEx ia IIC T6 Intrisically safe multifunction calibrator
• MicroCal 2+:
Two channel multifunction calibrator
for simultaneous measurement and
simulation for TRX calibration.
• MicroCal 10+:
0.01% accuracy, multifunction calibrator.
Display & Keyboard
The high contrast alphanumeric LCD
display with dot matrix (7x5 dots per
character - 16 characters) allows easy
reading even in poor light conditions.
Simultaneous indication of the measured or simulated value, engineering
unit, type of sensor or signal, temperature scale and Rj mode. MicroCal 2+,
the 2 channels calibrator, has a 2 line
back lighted LCD display for simultaneous reading of input and output signals.
A metal-click tactile polycarbonate
membrane keyboard ( MicroCal 1+
MAV, MicroCal 2+ and MicroCal 10+),
with a working life of one million operations per key, seal the internal electronics from the surrounding environment. Contact closure of the membrane keys is acknowledged, as a
coded signal, directly by the microprocessor. Two thick film membrane
“slidewires” (patent pending) allows
the quick and an easy setting of the
simulation value.
Source & Measure thermoelectric
signals
Each instrument, through a menu-driven
procedure, allows measurement or simulation of mV, mA, or any normalised
IEC, DIN and JIS thermoelectric sensor
TEST &
MEASUREMENT
MICROCAL
One & Two channels
Palm-top Multifunction
Calibrators
Features …
… Benefits
J, K, T, R, S, B, C, U, L, G, D, N, E, F, Pt100,
Ni100, Ni120, ½. The microcontroller
performs automatic polynomial linearization and cold junction compensation to assure high accuracy.
Cold junction compensation
Accurate and fast response cold junction compensation through a special
proprietary design of a low thermal capacity binding posts incorporating a high
accuracy thin film Pt100.
Simulation programs
Menu-driven simulation program set
up to generate:
• a continuous or step ramp output
where the total time, the start point,
the end point and the size of the
step are requested by the set-up
procedure to run the program;
• a continuous programmable cycle
(rise, soak, fall);
• a manual repeat increment through
keyboard;
• 60 memory stored values (20
groups of 3 memories) for 7 different simulation programs.
EMC Conformity
Each instrument has been developed
and tested for compliance with the directive 89/336/CEE Electromagnetic
Compatibility (CE mark).
Report of Calibration
Each MicroCal is factory calibrated and
certified against 1st Line Standards
mantained in the Eurotron Metrology
Laboratory. These Reference Standards
are periodically certified by an
Internationally recognised Laboratory
to ensure traceability, and shipped with
a Report of Calibration stating the nominal and actual values and the deviation errors.
3
• Accuracy up to 0.01% of reading
•
•
•
•
•
•
•
•
granted for one year with operative
conditions from 18°C to 28°C.
Source and measure of mV, mA, or
any normalised IEC, DIN and JIS
thermoelectric sensor J, K, T, R, S,
B, C, U, L, G, D, N, E, F, Pt100,
Ni100, Ni120, ½. Flexibility save
money: one instrument performs
a wide range of functions.
IPTS 68 and ITS90 linearization
mA active and passive loop (direct
TRX loop power supply)
Two wires TRX simulation
Resistance thermometers measurement and active simulation with a
propietary circuit
Electronic fuse protection on critical connection to avoid instrument
damage of misused. Don’t require
spare fuses.
Cold junction compensation.
Accurate and fast RJ compensation
with propretary binding post.
Hardware and firmware use a selflearning technique. No trimmer adjustment if a reallignement if required.
PT100
End
step
Startt
Star
Rise
Soack
Fall
Time
• Simulation programs: 60 memory
•
•
•
•
•
•
•
•
stored value for 7 different simulation programs to source continuos
or step ramps, cycles and automatic
programmable sequences.
Scale factor function: to read the
input or output signal in engeneering unit directly.
Square root function: to linearize
quadratic signals.
Programmable averaging: makes it
practical to measure unstable, noisy
input signals.
RS232 interface & software: Allow
communication to and from a PC.
CalpMan 2000 Calibration Procedure
Software for Windows 95/98/NT
Two thick film membrane “slidewires”:
allow an easy and fast setting of the
simulation value.
Shock resistant ABS case and tactile polycarbonate membrane
keypad. Rugged design for field
applications.
Traceable report of Calibration as
standard
Note:
see details inside for the specifications of the different models.
TEST &
MEASUREMENT
MICROCAL
Five models
for all process
quantity
MicroCal 1+ MAV
MicroCal 1+
MicroCal 1 IS
MicroCal 1+ MAV is a voltage and current process calibrator designed to
meet, in a practical way, the needs
of instrumentation engeneers, both
in laboratory and field work.
Accurate, compact, rugged, easy to
use it is the ideal solution to measure
and simulate millivolts, volts (up to
10V) and mA on active and passive
loop.
The two channels model allows simultaneous measurement and simulation for voltage (up to 30V) and
current on active and passive loop.
An accurate (0.03% of rdg) and powerful multi function system for testing,
measuring and calibrating into a single
compact portable instrument.
MicroCal 1+ is the ideal solution for
measuring and simulating: milliVolts
(up to 1V), milliAmperes, ohms, and
14 type thermocouples.
MicroCal 1+ Rtd allows to measure
and simulate resistance up to 400½
and five type of resistance thermometers.
MicroCal 1 I.S. is the multifunction indicator-simulator with the specifications of MicroCal 1+ plus the compliance of Intrisic Safety (I.S.) standards.
The Microcal 1 IS indicator-simulator
has successfully met the verification
and test requirements prescribed by
EN 50.014 and EN 50.020 standards
with a classification
4
EEx ia IlC T6
CESI, an approved certifying authority
of the EEC, has ensured the design
compliance to the above mentioned
intrinsic safety standards.
TEST &
MEASUREMENT
MICROCAL
To document
the inspection
and calibration
activities
MicroCal 2+
MicroCal 10+
Microcal 2+ is a high accuracy (0.02
% of rdg.) two channels palmtop calibrator. It is the ideal solution in all
check outs, measuring and simulation for the simultaneous function of:
- Channel 1 (In/Out): millivolt, volt
(up to 10V), milliampere (active
and passive loop), ½, thermocouple
and resistance thermometer;
- Channel 2 (In): volt (up to 30V)
and milliampere (active and
passive loop).
The calibrator has a 2 lines backlighted display for simultaneous
reading of input and output signals
The ideal tool for signal conditionig
calibration.
An highly accurate (0.02 % of rdg.)
and powerfull one channel multifunction calibrator designed for
testing, measuring and calibrating into
a single compact portable instrument.
MicroCal 10+ is the ideal solution for
measuring and simulating: milliVolts,
Volts (up to 10V), milliAmperes,
ohms, thermocouples and resistance
thermometers.
5
CalpMan 2000
Easy to use and Low
Cost PC software
Standard Agencies and Quality
Auditors require, to keep the compliance with ISO 9000, collection organization and analysis of traceability
documents. CalpMan (Calibration
Procedure Manager) Windows®
95/98/NT based supporting software, designed to document the calibration/inspection activities using all
Microcal calibrators. The status and
measured values of every instrument
“before” and “after” its calibration can
be memory stored, recalled or stored
into the PC data bank.
The benefits of the periodical equipment inspections achieve the following tasks:
• To grant the traceability of every test
and process equipment.
• To grant the appropriate operative
condition of each section of the
process to obtain the required level
of quality of the finished product.
• To optimise the Predictive
Maintenance period through an
analysis of test and calibration documents.
• To printout of the Report of
Calibration relevant with each TAG.
TEST &
MEASUREMENT
MICROCAL
High technology
for daily use
in field and
laboratory
Specification
Both IPTS 68 and ITS 90
IPTS 68 (MicroCal 1 I.S. only)
International Temperature Scale:
Reference Junction
Compensation (Rj):
Internal automatic from -5°C to +50°C (+23°F to +122°F)
Reference Junction accuracy:
± 0.15°C/°C for temperature exceed the +18°C to +28°C (64°F to 82°F) band
Common Mode Rejection:
>130 dB at 50/60 Hz ±1 Hz
Normal Mode Rejection:
>65 dB at 50/60 Hz ±1 Hz
External programmable from -50°C to +100°C (-58°F to +212°F)
Span: ±0.003% of rdg/°C
±0.005% of rdg/°C (Microcal 1 I.S. only)
±0.015% of rdg/°C (Channel 2 only)
Temperature stability:
for temperture exceeding
the +18°C to +28°C
(64°F to 82°F) band
Zero: ±0.2 µV/°C
±0.5 mV/°C or ±0.4µA/°C (Channel 2 only)
Rtd and ½ measurement
excitation current:
Rtd and ½ simulation
excitation current:
Rtd cable compensation error:
0.25 mA
From 0.2 to 3 mA
±0.005°C/½ of total cable resistance
max 100 ½ each wire
Rtd cable compensation limit:
38 ½
20 ½ (Microcal 1 I.S. only)
Shunt resistance (mA range):
60 ½ (Channel 2 only)
mA measurement and generation:
Active and Passive loop on MicroCal 1+ MAV and MicroCal 2+
Input impedance:
10 M½ (1 M½ for 10 V range and Channel 2 only)
Output Impedance (mV and Tc):
<0.5 ½ with a maximun current of 0.5 mA
1000 ½ at 20 mA
Maximum resistance load:
100 ½ at 20 mA Microcal 1 I.S. only
Maximum input overvoltage:
50 V (5 V Microcal 1 I.S. and Rtd ranges only)
Maximum input overcurrent:
100 mA (6 mA Rtd ranges only)
Source resistance effect:
1 µV/1000 ½
Engineering unit indication:
4 characters directly on the display
n. 4 rechargeable Ni-MH battery (sealed cartridge for Microcal 1 I.S. only)
Power supply:
Mains operation and battery charge 100, 115, 230 V 50/60 Hz
Recharging time:
Max 12 h at 90%
Battery life:
12 h on measuring mode - 4 h on 20 mA passive loop
Display:
High contrast alphanumeric LCD ( Backlight on MicroCal 2+ )
Operative ambient temperature:
From -10°C to 50°C (15°F to 122°F)
Storage temperature:
From -30°C to 60°C (-22°F to 140°F)
Case:
ABS 120x60x230 mm
Weight:
Net 1 kg
6
Gross 1.5 kg
TEST &
MEASUREMENT
MICROCAL
Easy to use
Easy to order
Ordering code
Table A Model
MicroCal 1+ MAV, MicroCal 1+,
MicroCal 2+, MicroCal 10+
Standard packing include :
-MicroCal
-Charger
-Vinyl Case
-Instruction Manual
-Report of Calibration
3440-1 MicroCal 1+ MAV
measurement and generation of mV, 10V, mA.
3440-2 MicroCal 1+ MAV
measurement and generation of mV, 10V, mA with 2nd channel
for 30V, mA measurement.
3906-A MicroCal 1+
measurement and generation of mV, 1V, mA, Tc
3906-Rtd MicroCal 1+ Rtd
measurement and generation of mV, 1V, mA, Tc, Rtd, ½
3920
MicroCal 2+
measurement and generation of mV, 10V, mA, Tc, Rtd, ½
with 2nd channel for 30V, mA measurement.
3910
MicroCal 10+
measurement and generation of mV, 10V, mA, Tc, Rtd, ½.
Table B Battery charger
1
115 V with USA plug
2
230 V with Schuko plug
3
230 V with UK plug
4
230 V with Europea plug
5
100 V with USA/Japan plug
Table C Options
Example :
MicroCal 1+ Rtd with US plug charger
Report of Calibration, Electrical test leads.
Ordering code is
MicroCal 1+ 3906-Rtd-1-14
1
Report of Calibration
2
Leather Case with shoulder strap (instead vinyl case)
3
ABS case for instrument+charger+kits (instead vinyl case)
4
Electrical leads kit
5
Thermocouple connection cables kit (Tc J,T,K and S )
9
Special
Table A Model
MicroCal 1+ IS
Standard packing include :
- MicroCal
- Charger
- Leather carrying case
- Instruction Manual
- I.S. Certificate
- Report of Calibration
3901-IS-A Microcal 1 IS
Measurement and generation of mV, 1V, mA e Tc
3901-IS-Rtd Microcal 1 IS Rtd
Measurement and generation of mV, 1V, mA, Tc, ½ and Rtd
Table B Battery charger
1
115 V with USA plug
2
230 V with Schuko plug
3
230 V with UK plug
4
230 V with Europea plug
5
100 V with USA/Japan plug
Table C Options
7
1
Report of Calibration
9
Special
Table IN-OUT range
Sensor or
parameter
Tc type J
Tc type K
Tc type T
Tc type R
Tc type S
Tc type B
Tc type C
Tc type G
Tc type D
Tc type U
Tc type L
Tc type N
Tc type E
Tc type F
Pt 100 IEC
Pt 100 JIS
Pt 100 US
Ni100
Ni120
mV
Total range
Accuracy Range
Resolution
MicroCal 1+ MAV
MicroCal 1+
MicroCal 2+
MicroCal 10+
MicroCal 1 IS
Accuracy (%of rdg)
Accuracy (%of rdg)
Accuracy (%of rdg)
Accuracy (%of rdg)
Accuracy (%of rdg)
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
-210 to +1200
-346 to +2192
-270 to +1370
-454 to +2498
-270 to +400
-454 to +752
0 to +1760
+32 to +3200
0 to +1760
+32 to +3200
+200 to +1820
+392 to +3308
0 to +2300
+32 to +4172
0 to +2300
+32 to +4172
0 to +2300
+32 to +4172
-200 to +400
-328 to +752
-200 to +760
-328 to +1400
0 to +1300
+32 to +2372
-270 to +1000
-454 to +1832
0 to +1400
+32 to +2552
-170 to 1200
-274 to 2192
-120 to 1300
-184 to 2372
-120 to +400
-184 to 752
+500 to +1700
+932 to +3092
+800 to +1760
+1472 to +3200
+1000 to +1820
+1832 to +3308
0 to +2300
+32 to +4172
200 to +2300
+392 to +4172
0 to +2200
+32 to +3992
-120 to +400
-184 to +752
-180 to +760
-292 to +1400
0 to +1300
+32 to +2372
-150 to +1000
-238 to +1832
0 to +1400
+32 to +2552
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
(2)
(2)
(1)
(1)
(1)
(1)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0. 02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
(2)
(2)
(1)
(1)
(1)
(1)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
(2)
(2)
(1)
(1)
(1)
(1)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
(2)
(2)
(1)
(1)
(1)
(1)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(1)
°C
°F
°C
°F
°C
°F
-200 to +850
-328 to +1562
-200 to +850
-328 to +1562
-200 to +850
-328 to +1562
-200 to +600
-328 to +1112
-200 to +600
-328 to +1112
-200 to +600
-328 to +1112
0.1
0.1
0.1
0.1
0.1
0.1
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
±0.03%
(2)
(2)
(2)
(2)
(2)
(2)
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
(2)
(2)
(2)
(2)
(2)
(2)
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
±0.02%
(2)
(2)
(2)
(2)
(2)
(2)
±0.03%
±0.03%
(2)
(2)
°C
°F
°C
°F
-60 to +180
-76 to +356
0 to +150
+32 to +302
-60 to +180
-76 to +356
0 to +150
+32 to +302
0.1
0.1
0.1
0.1
±0.03%
±0.03%
±0.03%
±0.03%
(1)
(1)
(1)
(1)
±0.02%
±0.02%
±0.02%
±0.02%
(1)
(1)
(1)
(1)
±0.02%
±0.02%
±0.02%
±0.02%
(1)
(1)
(1)
(1)
±0.03%
±0.03%
(1)
(1)
-18 to +21
0 to +100
mV
mV
0 to 1000
V
0 to 10
mA
0 to 21
½ (IN)
0 to 400
½ (OUT)
0 to 400
CH2 measurement only
mA
0 to +22
V
0 to 30
-18 to +21
0 to 21
21 to 53
53 to 100
0 to 1000
0 to 10
0 to 21
0 to 400
0 to 400
1µV
1µV
10µV
10µV
100µV
1mV
1µA
10m½
10m½
±(0.02%+3µV)
±(0.02%+3µV)
±(0.02%+3µV)
±(0.02%+6µV)
±(0.02%+60µV)
±(0.02%+60mV)
±(0.02%+0.5µA)
0 to +22
0 to 30
1µA
1 mV
±(0.02%+1µA)
±(0.02%+1.2mV)
Note:
• The relative accuracies shown are stated for 360 days and
operative conditions from +18°C to +28°C (+64 °F to +84°F).
• Tc’s relative accuracies are stated with 0°C external Rj sets.
• Typical 90 days relative accuracy can be estimated by dividing the “% of rdg” specifications by 1.8.
•
•
•
±(0.02%+3µV)
±(0.02%+3µV)
±(0.02%+3µV)
±(0.02%+6µV)
±(0.02%+60µV)
±(0.02%+0.5µA)
±(0.02%+38m½)
±(0.03%+78m½)
±(0.01%+3µV)
±(0.01%+3µV)
±(0.01%+3µV)
±(0.01%+6µV)
±(0.01%+40µV)
±(0.02%+0.4mV)
±(0.02%+0.5µA)
±(0.02%+38m½)
±(0.03%+78m½)
±(0.02%+1µA)
±(0.02%+1.2mV)
Typical 2 year relative accuracy can be estimated by multiplying the “% of rdg” specifications by 1.5.
All input ranges: additional error ±1 digit.
Eurotron traceability chart and uncertainty can be supplied
on request
Eurotron Italiana srl
Viale F.lli Casiraghi 409/413
I 20099 Sesto S. Giovanni (MI)
Tel.: +39 02 248820.1
Fax:+39 02 2440286
±(0.01%+3µV)
±(0.01%+3µV)
±(0.01%+3µV)
±(0.01%+6µV)
±(0.01%+40µV)
±(0.02%+0.4mV)
±(0.02%+0.5µA)
±(0.02%+38m½)
±(0.03%+78m½)
(1)Zero error ± 0.1 °C / 0.18 °F
(2)Zero error ± 0.15 °C / 0.28 °F
(3)Zero error ± 0.5 °C / 0.9 °F
Distributed by:
G.&P. Grafica & Pubblicità snc - Gazzaniga - BG • e - m a i l : G & P @ m a i l . c s g - n e t . i t
MICROCAL
Accurate
and Reliable
365 days a year
±(0.03%+3µV)
±(0.03%+3µV)
±(0.03%+6µV)
±(0.03%+60µV)
±(0.03%+0.5µA)
±(0.03%+38m½)
±(0.04%+78m½)
Specifications shown may change without notice