A-dec 300® System INSTALLATION GUIDE CONTENTS INTRODUCTION Introduction ................... 1 Install the Chair ............... 3 Install the Support Center .. 9 Install the Assistant's Instrumentation ............. 15 Install the Delivery System 22 Install the Dental Light .... 34 Install the Monitor Mount . 36 Install the Contoured Floor Box ............................ 37 Install the Remote Floor Box ............................. 39 Connect the Utilities ....... 40 Install the Cuspidor ........ 60 Install Upholstery ........... 61 Prepare and Adjust the System ........................ 64 Level the System ........... 77 Touchpad Settings .......... 80 Test the System ............. 93 Verify Proper Clearance Between the Cuspidor Bowl and the Armrest ............ 94 Install the Covers ........... 96 Appendix: Install the Air Vacuum System (AVS) .....105 Regulatory Information ...108 This document contains installation instructions for the A-dec 300 system. Information that is critical to a successful and safe installation is shaded like this note throughout the guide. NOTE When removing or replacing covers, take care not to damage any wiring. Verify that the covers are secure after replacing them. CAUTION Before you begin: • Clear the room of all debris and thoroughly clean the floors. • Check that manual air and water shut-offs are installed. • Purge any debris from air and water lines. • Check with local building and code authorities about installation requirements. They differ from state to state and internationally. Your installation may not require all components described in this document. Before you begin: 1. Assess what modules you will install. 2. Use “Installation Sequence” on page 2 to note the order of the modules that are to be installed. 86.0087.00 Rev F A-dec 300 System Installation Guide Recommended Tools Tools Needed For This Installation Drill bits: 3/8" wood, 1/4" and 1/2" masonry Hex key set Drill 5/16" hex key driver Diagonal cutters A-dec silicone lubricant Ball driver set Needle nose and standard pliers Umbilical snake Adjustable wrench Phillips head screwdriver Rubber mallet 3/4" and 9/16" socket and ratchets Roto hammer drill Sleeve pusher 1/4", 1/2", and 3/4" combination wrenches Magnetic level Voltmeter Tape measure Pliers Installation Sequence Modules for the A-dec 300 system installation are shown in Figure 1. Install the modules for your configuration in the order they are listed. Figure 1. A-dec 300 System Shipping Boxes NOTE The box for each module contains all of the parts needed to install that module. 1 4 Dental Chair Delivery System 2 5 Support Center 3 6 Assistant’s Instrumentation 2 Dental Light Monitor Mount 86.0087.00 Rev F Install the Chair INSTALL THE CHAIR If you have a post mount system, skip this section and go to “Install the Support Center” on page 9. NOTE Tools Needed For This Section 3/4" and 9/16" socket and ratchets 3/8" wood drill bit or 1/4" and 1/2" masonry drill bits Drill 3/16" hex key Roto hammer drill 3/4" combination wrench Phillips head screwdriver Figure 2. Place the Chair Position the Chair 1. Remove all items and cardboard from around the chair. NOTE When removing modules from their packaging, watch for kits and manuals included for the doctor (such as the A-dec 311 Dental Chair Instructions For Use). Set these aside during installation. Grab Points Shipping Strap 2. Remove the covers. 3. Use a 3/4" socket and ratchet to remove the bolts securing the chair to the pallet. 4. Grasp an armrest and the front of the chair frame. Lift and place the chair in position in the treatment room. Shipping Bolts Figure 3. Contoured Floor Box If the system includes a contoured floor box, failure to provide enough space between the utilities and the contoured floor box cover will prevent the installation of the power supply cover. For more information, see “Install the Contoured Floor Box Frame” on page 37. CAUTION 5. Remove the shipping strap and th42packaging from the armrests. 86.0087.00 Rev F Contoured Floor Box 3 A-dec 300 System Installation Guide Anchor the Chair Use the lag screw or masonry anchor with the cleat to anchor the chair to the floor. Select the procedure for your type of flooring structure. WARNING Anchoring the chair to the floor is required for mechanical stability. Failure to anchor properly could result in damage, serious injury, or death. Check the flooring and/or framing material where you will anchor the chair. If it is not at least 3-1/4" (82 mm) thick, contact a licensed contractor about reinforcing the floor. CAUTION Anchor to a Concrete Floor Prepare the Installation Area Figure 4. Prepare the Chair P3 1. Plug in the chair and press the Mains On/Off button on the lower right of the power supply to turn it on. Electric components on the circuit board are static sensitive and require handling precautions. CAUTION 2. Move the jumper in P3 of the chair circuit board to the Base Up position. Once the chair is raised, return the jumper to the Spare position. 3. Turn off the power. 4. Unplug the chair. 4 Mains On/Off Button 86.0087.00 Rev F Install the Chair 5. Place the cleat in the large circle in the chair base (see Figure 5). 6. Mark where to drill the hole for the masonry anchor and move the chair. 7. Use a 1/2" bit and roto hammer to drill a 4" (101 mm) deep hole; then remove any debris. Secure the Chair Figure 5. Anchor the Chair to a Concrete Floor 1. Drive the 3-1/2" masonry anchor into the hole until the washer is flush with the floor. 2. Use a 9/16" socket and ratchet to tighten the anchor until it is securely fixed in the hole then remove the screw and washer. 3. Return the chair to position. 4. From the left front of the chair, place the screw through the washer and cleat and into the hole. Cleat Masonry Anchor 5. Use a 9/16" socket and ratchet to tighten the bolt against the cleat until it firmly holds the chair to the floor. 86.0087.00 Rev F 5 A-dec 300 System Installation Guide Anchor to a Wood Floor Figure 6. Start the Lag Screw Prepare the Installation Area 1. Plug in the chair and press the Mains On/Off button on the lower right of the power supply to turn it on (see Figure 4 on page 4). Electric components on the circuit board are static sensitive and require handling precautions. CAUTION 2. Move the jumper in P3 of the chair circuit board to the Base Up position. Once the chair is raised, return the jumper to the Spare position. 3. Turn off the power and unplug the chair. 4. Place the cleat in the large circle in the chair base. Lag Screw 5. Mark where to drill the hole for the lag screw and move the chair. Cleat 3/4" Socket and Ratchet 6. Use a 3/8" bit to drill a 2" (50 mm) deep hole; then remove any debris. 7. To thread the hole, use a 3/4" socket and ratchet to drive the 2-1/2" lag screw about 1/2" (12.8 mm) into the hole; then remove the lag screw. Figure 7. Tighten the Lag Screw Secure the Chair 1. Put the chair back into position so that the hole is positioned on the right rear edge of the circle in the chair base (see Figure 6). 2. From the right rear of the chair, place the lag screw through the cleat and into the hole. Lag Screw Cleat 3. With your fingers, start the lag screw as far as you can; then use a 3/4" socket and ratchet with a short extension to tighten the lag screw until it is almost flush against the cleat. 4. Move the chair into position and use a 3/4” combination wrench to tighten the lag screw against the cleat until it firmly holds the chair to the floor. 6 3/4" Combination Wrench 86.0087.00 Rev F Install the Chair Install the Power Supply or Integrated Floor Box Cover Frames Figure 8. Cover Frames A-dec 300 systems come with either a power supply cover or an integrated floor box cover. NOTE The power supply cover frame ships attached to the power supply cover. The steps for installing their frames are the same. 1. With the chair fully raised, use a 3/16" hex key to remove the two mounting screws from the chair base. Power Supply Cover Frame Integrated Floor Box Cover Frame Figure 9. Install the Cover Frame Attach Frame to Chair Base 2. Attach the cover frame to the chair base with the two mounting screws. Attach Frame to Floor Never use a cover frame as a handle when moving the chair. CAUTION 3. Attach the cover frame to the floor. ○ If the floor is made of wood, use a Phillips head screwdriver and 1-1/4" size #10 screws. ○ If it is a concrete floor, use a 1/4" masonry drill bit to make holes where the screws fit through the frame. Insert the plastic anchors in the holes then use a Phillips head screwdriver and 1-1/4" size #10 screws. Mains On/Off Button Power Supply Cover Frame 86.0087.00 Rev F 7 A-dec 300 System Installation Guide Install the Back Support Figure 10. Install the Back Support 1. Use a 3/16" hex key to remove the 3 mounting screws from the back support. 2. Attach the back support to the chair back with the mounting screws. Back Support Mounting Screws TIP Start all three screws before tightening them. Chair Back Be sure to tighten the screws firmly to avoid the back becoming loose during use. CAUTION Install the Footswitch (Optional) Figure 11. Install the Footswitch Touchpad Or Footswitch Connection If the system contains a footswitch, complete the following steps to install it. 1. From the rear of the chair, route the cord to the chair’s circuit board and connect it to the Touchpad Or Footswitch P5 connector on the chair circuit board. 2. Use a 3/16" hex key to secure the footswitch cord in one of the side holes of the strain relief. Circuit Board Strain Relief NOTE If you are not installing any modules on to the chair, go to “Install Upholstery” on page 61. 8 86.0087.00 Rev F Install the Support Center INSTALL THE SUPPORT CENTER If the system includes a cuspidor, do not remove the positioning guide that is cable tied to the cuspidor bowl support until the support center is properly aligned. Complete the following section to ensure proper alignment of the support center. WARNING Tools Needed For This Section 5/16" hex key driver 3/16" hex key Level Diagonal cutters Adjustable wrench Sleeve pusher Figure 12. A-dec 361 Support Center and Cuspidor You can mount the support center on either side of the chair. This section describes how to install the support center and connect the utilities. Install the Chair-Base Mount Figure 13. Install the Chair-Base Mount NOTE If you have a post mount system, skip this procedure and follow the directions provided with the chair adaptor kit. Then go to “Install the Support Center Post” on the next page. Chair-Base Mount Use a 5/16" hex key driver and two 1-1/2" socket head screws to attach the chair-base mount to the side of the chair frame. Be sure to tighten the screws securely (approximately 30 ft-lb [4.15 m-kgs] of torque) so the mount does not loosen during use. CAUTION 86.0087.00 Rev F Use a 5/16" hex key driver to apply 30 ft-lb (4.15 m-kgs) of torque pressure. 9 A-dec 300 System Installation Guide Install the Support Center Post Figure 14. Install the Support Center Post Notch 1. Place the support center post in the chairbase mount with the notched bottom over the screw near the base of the mount. NOTE If you have a post mount system, the support center post installs into a chair adaptor instead of the chair-base mount and the post does not have notches in the bottom. Square Hole One Set Of Leveling Screws WARNING If you’re installing a post mount system, anchoring the chair to the floor is required for mechanical stability. Failure to anchor properly could result in damage, serious injury, or death. If the support center is installed to the left of the patient, face the square hole away from the chair. If the support center is installed to the right of the patient, face the square hole toward the chair. 2. Place a level vertically against the post and align it with the four leveling screws near the top of the chair-base mount. NOTE If you have a post mount system on an A-dec 511 chair, the chair adaptor has eight leveling screws. 10 86.0087.00 Rev F Install the Support Center 3. Use a 3/16" hex key to adjust the screws, switching from one set of leveling screws to the other until the post is level. Always align the level with the screws you are adjusting. 4. Once the post is level, evenly tighten the leveling screws to secure it. NOTE This is the initial leveling of the post. The system requires a final level after everything is installed. For final leveling instructions, see “Level the System” on page 77. Install the Support Center Figure 15. Position the Support Center 1. Wrap the support center tubing bundle around your arm and lift the support center out of the box. 2. Remove the covers from the support center. 3. Place the support center over the support center post with the water bottle connection toward the toe of the chair. Support Screw Be careful to clear all wires and tubing so they stay on the open side of the support center frame and do not get pinched or kinked. CAUTION 4. Slide the support center down the post until its frame rests on the support screws on the support center post. 86.0087.00 Rev F 11 A-dec 300 System Installation Guide 5. Align the support center depending on the configuration of the system: ○ If you have an A-dec 300 system that includes a cuspidor, align the support center so the positioning guide clears the armrest. Use the jumper and test points to move the chair up and down when testing clearance. Figure 16. Ensure the Positioning Guide Clears the Armrest Positioning guide must clear armrest. Positioning Guide WARNING The positioning guide must clear the armrest to provide the proper clearance of 1-1/8" (29 mm) between the cuspidor bowl and armrest. ○ If you have a base mount system without a cuspidor, or a post mount system, align the support center so it is parallel to the dental chair. 12 86.0087.00 Rev F Install the Support Center 6. Use a 5/16" hex key driver to tighten the button-head screw. While tightening the screw, ensure the positioning guide retains its clearance of the armrest. Figure 17. Secure the Support Center Button Head Screw To ensure that the button-head screw is properly tightened, use approximately 13 ft-lb. (1.8 m-kgs) of torque. CAUTION NOTE If you have a post mount system, you are finished installing the support center. Skip to the note after step 9 for instructions on routing tubing and wiring. 7. Use a 5/16" hex key driver to install each of the two 1-1/2" socket head screws until they are touching the back of the holes in the support center post. Use a 5/16" hex key driver to apply 13 ft-lb (1.8 m-kgs) of torque pressure to the screws. Install two 1-1/2" Socket Head Screws. 8. Alternately tighten both socket head screws until they are firmly secure (use approximately 13 ft-lb. [1.8 m-kgs] of torque). While tightening the screws, ensure the positioning guide retains its clearance of the armrest. 9. Route the support center tubing and wiring through the chair-base mount into the utilities area at the base of the chair. Figure 18. Route the Support Center Tubing Bundle NOTE If you are installing a post mount system, remove the Y-connector from the data line before routing the tubing. (The Y-connector is provided to connect to multiple data communication system devices.) Route the tubing and wiring from the bottom of the support center through the convolute into the remote floor box. Utilities Area If you are installing a post mount system on an A-dec 511 chair, Chair-Base Mount 86.0087.00 Rev F 13 A-dec 300 System Installation Guide separate the power cables and data line from the tubing group and route them from the support center, underneath the mount, and down the lift arm to the power supply. Route the rest of the tubing bundle through the convolute into the remote floor box. For information about installing the remote floor box, see “Install the Remote Floor Box” on page 39. Once you’ve completed installing the support center, verify that the end of the cuspidor ventilation tube is 1/2" (13 mm) above the support center frame. This ensures that the cuspidor drains properly. CAUTION Figure 19. Verify the Cuspidor Ventilation Tubing Position Support Center Frame 1/2" (13mm) Cuspidor Ventilation Tube Install the Moisture Separator (Optional) To install the moisture separator, use the directions included in the moisture separator kit (p/n 41.1477.00); then see “Connect the Utilities” on page 40. 14 86.0087.00 Rev F Install the Assistant's Instrumentation INSTALL THE ASSISTANT'S INSTRUMENTATION Figure 20. Assistant’s Instrumentation Assistant’s instrumentation can be mounted three ways. Radius®-Style (351) . . . . . . . . . . . . . 16 Cuspidor-Mounted (353) . . . . . . . . . . 18 Telescoping (352) . . . . . . . . . . . . . . 20 A-dec 351 Radius-Style Assistant’s Instrumentation with Touchpad A-dec 352 Telescoping Assistant’s Instrumentation with Optional Touchpad 86.0087.00 Rev F A-dec 353 Cuspidor-Mounted Assistant’s Instrumentation with Optional Touchpad 15 A-dec 300 System Installation Guide Install a Radius-Style Assistant’s Instrumentation (351) Tools Needed For This Section Hex key set Figure 21. Types of Washers Used Nylatron Washer Flat Washer Bearing Install the Assistant’s Instrumentation Arm 1. Slide the bearing and two washers from the kit onto the screw. Figure 22. Install the Assistant’s Instrumentation Arm 2. Insert the screw through the hole in the assistant’s instrumentation arm. 3. Slide the Nylatron® washer onto the screw. 4. Put Loctite® on the end of the screw. 5. Use a 1/4" hex key to securely tighten the arm to the chair. The arm should rotate smoothly and not drift. 6. Place the syringe in its holder on the assistant’s instrumentation. Nylatron Washer 7. Place the high volume evacuation (HVE) and saliva ejector in their holders; then attach their tubing to the vacuum canister. Assistant’s Instrumentation Arm Flat Washer Bearing Flat Washer Screw 16 86.0087.00 Rev F Install the Assistant's Instrumentation Route the Tubing, Cables, and Lines Figure 23. Install the Assistant’s Instrumentation Plate 1. From the back of the chair, route the vacuum line to the left of the lift cylinder and above the clear hydraulic tubing. 2. Route the tubing, power cables, and data line to the right of the lift cylinder and above the overflow bottle. Top Mount Screw Plate Notch 3. Use a 3/16" hex key to loosen the socket head screw securing the middle white mounting block. Loosen Middle Screw 4. Remove the top screw and mounting block. 5. Slide the notch in the plate between the middle mounting block and the lift arm. 6. Replace the top screw and mounting block. Make sure the screw is securely tightened. Vacuum Line Tubing and Cables 7. Tighten the middle socket head screw. Clear Hydraulic Line 8. Cable tie the tubing, power cables, and data line to the bottom two mounting blocks. Overflow Bottle Figure 24. Cable Tie the Tubing and Wires 9. Route the syringe tubing between the motor pump and the wire cover, then behind the power supply. 10.Route the power cables and data line to the chair circuit board. NOTE If the system does not include other modules, go to “Install the Contoured Floor Box” on page 37. 86.0087.00 Rev F Cable Tie to These Blocks 17 A-dec 300 System Installation Guide Install a Cuspidor-Mounted Assistant’s Instrumentation (353) 1. Use diagonal cutters to cut the cable tie and remove the positioning guide. Tools Needed For This Section Diagonal cutters Sleeve pusher Hex key set Figure 25. Routing with a Touchpad Viewed From Below NOTE Save the positioning guide and the foam plug holding it in the cuspidor bowl support. You will need it later to verify proper clearance between the cuspidor bowl and armrest. For more information, see “Verify Proper Clearance Between the Cuspidor Bowl and the Armrest” on page 94. 2. If the system includes an AVS, use a 1/8" hex key to remove the two screws connecting the toggle assembly to the bottom of the cuspidor bowl support. Data Ports WAGO Connectors Figure 26. Routing with a Touchpad Viewed From Above 3. If the assistant's instrumentation includes a touchpad, route the power cable and data line up through the cuspidor holder. For the wires to fit properly, place the power cable into the grooves first then the data line. Data Line and Power Cable Fit in Cuspidor Grooves 4. Place the assistant's instrumentation under the cuspidor holder and angle it away from the chair so that the post fits into the hole in the bottom of the cuspidor bowl support. 5. To attach the assistant’s instrumentation to the bottom of the cuspidor holder, use a 3/16" hex key to install the 1/4-20 x 1" screw in the middle and a 5/32" hex key to install the 10-32 x 5/8" screw in the large recessed hole that is off center. 6. If you disconnected the AVS toggle assembly in step 2, reconnect it now. Figure 27. Socket Head Screws (Actual Size Shown) 1/4-20 x 1" - Goes in the Middle Hole 18 10-32 x 5/8" - Goes in the Recessed Hole 86.0087.00 Rev F Install the Assistant's Instrumentation 7. If the system includes a touchpad, route the data line and power cable through the support center (see Figure 25 on page 18). Figure 28. Route the HVE, Saliva Ejector, and Syringe Tubing 8. Place the high volume evacuation (HVE) and saliva ejector in their holders; then attach their tubing to the vacuum canister. Vacuum Canister NOTE If the system includes an AVS, the HVE and saliva ejector are installed before the product is shipped and cannot be removed. 9. Place the syringe in the holder and route its tubing under the vacuum canister through the center hole. 10.Use a cable tie and a washer to create a strain relief for the syringe tubing. Make a double loop with the cable tie to firmly hold the tubing. Syringe Tubing Figure 29. Install the Strain Relief for the Assistant’s Syringe TIP When setting the strain relief, leave enough tubing to match the drape of the instrument tubing. NOTE If the system does not include other modules, go to the section for the system’s configuration: • Base Mount System - “Install the Contoured Floor Box” on page 37 • Post Mount System - “Install the Remote Floor Box” on page 39 86.0087.00 Rev F 19 A-dec 300 System Installation Guide Install an Assistant’s Instrumentation Mounted on a Telescoping Arm (352) 1. Use a 3/16" hex key and three 1" socket head screws to mount the hub for the telescoping arm on the support center. The stop on the hub goes toward the toe of the chair. Tools Needed For This Section 3/16" hex key Sleeve pusher Diagonal cutters Figure 30. Install the Telescoping Arm Hub Toe of the Chair Stop 2. Push the sheath onto the barb on the bottom of the assistant’s holder assembly. Figure 31. Install the Telescoping Arm Wave Washer 3. Set the assistant’s arm on the hub. 4. Push down one end of the wave washer and catch it on the hub ridge. Continue to work systematically around the hub to push the washer down until it is completely installed. 5. Place the HVE and saliva ejector in their holders; then attach their tubing to the vacuum canister. 6. If the system includes a touchpad: (1) Route the touchpad tubing under the vacuum canister through the hole on the right (it is the largest hole). Assistant’s Holder Assembly Sheath Figure 32. Install the Strain Relief for the Touchpad Tubing (2) Install the strain relief by placing the washer on the tubing then inserting the bushing into the tube. TIP When setting the strain relief, leave enough tubing to match the drape of the instrument tubing. 20 86.0087.00 Rev F Install the Assistant's Instrumentation (3) Route the data line and power cables through the support center. Figure 33. Route the Data Line and Power Cables WAGO Connectors 7. Place the syringe in the holder and route its tubing under the vacuum canister through the center hole (see Figure 31 on page 20). Figure 34. Route the Syringe Tubing Syringe Tubing 8. Use a cable tie and a washer to create a strain relief for the syringe tubing. Make a double loop with the cable tie to firmly hold the tubing. TIP When setting the strain relief, leave enough tubing to match the drape of the instrument tubing. Figure 35. Install the Strain Relief for the Assistant’s Syringe NOTE If the system does not include other modules, go to the section for the system’s configuration: • Base Mount System - “Install the Contoured Floor Box” on page 37 • Post Mount System - “Install the Remote Floor Box” on page 39 86.0087.00 Rev F 21 A-dec 300 System Installation Guide INSTALL THE DELIVERY SYSTEM Figure 36. Delivery Systems The A-dec 300 system can be configured with a Traditional or Continental delivery system. The installation steps are the same for both of them. Delivery systems can be Radius-style or mounted on a support center. Radius-Style (332 & 333) . . . . . . . . . . 23 Support Center Mounted (354 & 335) . 29 A-dec 332 Radius-Style Traditional Delivery System A-dec 335 Continental Delivery System mounted on a Support Center 22 86.0087.00 Rev F Install the Delivery System Install a Radius-Style Delivery System (332 & 333) Install the Rigid Arm Tools Needed For This Section Hex key set Diagonal cutters Sleeve pusher Adjustable wrench Tape measure 1. Raise the chair all the way up. 2. Turn off the power. 3. Cut the cable tie holding the clear hydraulic tubing that goes from the lift cylinder to the motor pump. Figure 37. Prepare the Chair Tower 4. Push the tubing back so that it is behind the post on the chair tower. Post Cable Tie Clear Tubing 86.0087.00 Rev F 23 A-dec 300 System Installation Guide 5. Place the notches in the front mount casting over the posts on the chair towers. Figure 38. Install the Rigid Arm Front Mount Casting Make sure all tubing and wiring are away from the posts so they do not get pinched. CAUTION 6. While holding the rigid arm in place, install the front bolt then the back bolt on one side and finger tighten. 7. Install the bolts on the other side; then use a 5/16" hex key to securely tighten all bolts. Mount Bolts Post Chair Tower 24 Notches Clear Hydraulic Tubing Positioned Between the Post and Axle for the Link Arm 86.0087.00 Rev F Install the Delivery System Install the Water Bottle 1. Use a 1/8" hex key to loosen the screw below the hole in the rigid arm. 2. Route the water bottle tubing into the hole, through the arm and out the bottom. Figure 39. Install the Water Bottle Water Bottle Housing Hole Rigid Arm 3. Slide the water bottle housing down over the mount screw and secure it to the arm. Mount Screw Install the Flexarm and Control Head 1. Cut the cable tie holding the trim ring to the delivery system tubing bundle. 2. Remove the delivery system from the box and balance it over your shoulder, with the control head in back of you, so one hand is free to route the tubing bundle. Figure 40. Install the Flexarm Flexarm Trim Ring Rigid Arm 3. Route the delivery system tubing bundle down through the rigid arm. TIP To keep Traditional delivery system handpiece tubing out of the way, wrap them around the control head before installing the flexarm. 4. Insert the flex arm into the rigid arm until it is fully seated. 86.0087.00 Rev F Tubing Bundle 25 A-dec 300 System Installation Guide 5. Use a 5/32" hex key to loosen the button-head screw that holds the control head’s position. Figure 41. Install the Control Head 6. Rotate the control head so it is positioned over the hole for the missing leveling screw. Hole for Missing Leveling Screw Install 3/8" Setscrew 7. To hold the control head in place, use a 5/32" hex key to install and tighten the 1/4-20 x 1/2" leveling screw. 8. Use a 5/32" hex key to install the 3/8" setscrew. 9. Place the handpiece tubing in their holders and insert the adjustment keys into their holes on the side of the control head. Loosen Button Head Screw Install and Tighten 1/4-20 x 1/2" Leveling Screw Figure 42. Leveling and Setscrews (Actual Size Shown) 1/4-20 x 1/2" Leveling Screw 5/16-18 x 3/8" Setscrew Figure 43. Insert Whip Hooks and Adjustment Keys NOTE If you are installing a Continental delivery system, insert the whip hooks into the whips. The top ends of the whip hooks face the front of the control head. Whip Hook Adjustment Key 26 86.0087.00 Rev F Install the Delivery System Route the Tubing and Wiring Figure 44. Route the Foot Control Tubing 1. From the back of the chair, route the foot control tubing over the strain relief and behind the power supply. 2. Measure three feet of the water bottle tubing from where it comes out of the rigid arm. Cut off the excess. 3. Strip the sheath around the water bottle tubing back to the rigid arm. TIP Mark the ends of the water bottle tubing to distinguish them from the delivery system tubing when connecting the utilities. Tubing Ridges Face Up 4. Route both the water bottle and delivery system tubing bundles through the convolute (see Figure 45). 5. Form a loop with a cable tie through the two holes in the rigid arm. Figure 45. Route the Tubing Bundles Through the Convolute Front Mount Casting 6. Push the convolute about an inch (a few centimeters) through the loop into the rigid arm. 7. Tighten the cable tie to secure the convolute. Rigid Arm Convolute Secure with Cable Tie 86.0087.00 Rev F 27 A-dec 300 System Installation Guide 8. Remove the tape from the ends of the water bottle and delivery system tubing. 9. Route the water bottle and delivery system tubing up through the front mount (be sure to go behind the black hydraulic tubing) then down to behind the power supply. NOTE Improper routing of the tubing can hinder the movement of the delivery system rigid arm. Figure 46. Route the Tubing and Wires Power Cables and Data Line Hydraulic Tubing Water Bottle and Delivery System Tubing Front Mount 10.Route the delivery system power cables, ground wire, and data line up through the front mount and down to the chair circuit board on the power supply. 11.If the system includes an assistant’s instrumentation, route its syringe tubing around the lift cylinder to behind the power supply. Cable Tie Rigid Arm Straight Out Arrange the Tubing Assistant’s Syringe Tubing Foot Control Tubing 1. Position the rigid arm so that it is straight out from the middle of the chair, in line with the front mount. 2. Neatly gather the tubing. Cable tie it to the chair tower. NOTE If the system includes a tray holder, see “Install the Tray Holder (Optional)” on page 32. If not, go to “Install the Contoured Floor Box” on page 37. 28 86.0087.00 Rev F Install the Delivery System Install a Delivery System Mounted on a Support Center (334 & 335) Install the Components 1. Lower the chair. Turn off the power. Tools Needed For This Section Hex key set Diagonal cutters Adjustable wrench Sleeve pusher Figure 47. Delivery System Post Trim Rings 2. Ensure the correct trim ring is on the delivery system post. ○ If the system includes a delivery system but no assistant’s instrumentation mounted on a telescoping arm, use the top trim ring in Figure 47. This trim ring ships on the delivery system post. ○ If the system includes a delivery system and assistant’s instrumentation mounted on a telescoping arm, use the bottom trim ring in Figure 47. Trim Ring Used With Delivery System This trim ring ships in the assistant’s instrumentation box. Trim Ring Used With Delivery System and Assistant’s Instrumentation Mounted on a Telescoping Arm 86.0087.00 Rev F 29 A-dec 300 System Installation Guide 3. Remove the delivery system from the box and balance it over your shoulder so one hand is free to route the tubing bundle. Figure 48. Carry the Delivery System The delivery system post inserts into the support center in the post hole toward the toe of the chair. 4. Route the tubing bundle into the support center then insert the delivery system post into the support center until it is fully seated. Figure 49. Install the Control Head 5. Use a 5/32" hex key to loosen the button-head screw that holds the control head’s position. 6. Rotate the control head so it is positioned over the hole for the missing leveling screw. Hole For Missing Leveling Screw Install 3/8" Setscrew 7. To hold the control head in place, use a 5/32" hex key to install and tighten the 1/4-20 x 1/2" leveling screw. 8. Use a 5/32" hex key to install the 3/8" setscrew. 30 Loosen Button Head Screw Install and Tighten 1/4-20 x 1/2" Leveling Screw Figure 50. Leveling and Setscrews (Actual Size Shown) 1/4-20 x 1/2" Leveling Screw 3/8" Setscrew 86.0087.00 Rev F Install the Delivery System 9. Place the handpiece tubings in their holders and insert the adjustment keys into their holes on the side of the control head. Figure 51. Insert Whip Hooks and Adjustment Keys NOTE If you are installing a Continental delivery system, insert the whip hooks into the whips. The top ends of the whip hooks face the front of the control head. Whip Hook Adjustment Key Route the Foot Control Tubing Figure 52. Route the Foot Control Tubing 1. Raise the chair all the way up. 2. From the back of the chair, route the foot control tubing through the chairbase mount and up into the support center. NOTE If you have a post mount system, the foot control tubing routes through a hole in the bracket in the bottom of the support center that holds the convolute. If the system includes a tray holder, see “Install the Tray Holder (Optional)” on the next page. If not, go to the section for the system configuration: • Base Mount System - “Install the Contoured Floor Box” on page 37 • Post Mount System - “Install the Remote Floor Box” on page 39 Tubing Ridges Face Up 86.0087.00 Rev F 31 A-dec 300 System Installation Guide Install the Tray Holder (Optional) To install the tray holder, select the type of delivery system and complete the steps. Figure 53. Tray Mount Position on a Continental Control Head Tray Mount Location Continental Delivery System Use a 5/32" hex key and three 10-32 x 3/4" screws to attach the mount arm to the middle of the bottom of the control head. 32 86.0087.00 Rev F Install the Delivery System Traditional Delivery System Figure 54. Install a Large Tray Holder Mount Mount Arm Tray Holder NOTE Large tray holders attach to mount arms in a different orientation than small tray holders. For proper orientation, see Figure 54 and Figure 55. 1. Use a 1/8" hex key and the 10-32 x 1/2" screws to attach the bottom of the tray holder to the mount arm. Do not over tighten the screws or you may damage the surface of the tray or break the mount. CAUTION 2. Use a 1/8" hex key and two 10-32 x 1" button head screws to attach the mount arm to the side of the bottom of the control head. Tray Mount Locations Four Screw Holes When installing a large tray holder, face the four screw holes in the mount away from the side of the control head. TIP While the tray holder can be installed on either side of the control head, it is helpful to mount it on the side opposite of the adjustment keys to provide better access to the keys. If the system does not include other modules, go to the section for the system’s configuration: Figure 55. Install a Small Tray Holder Tray Holder NOTE • Base Mount System - “Install the Contoured Floor Box” on page 37 • Post Mount System - “Install the Remote Floor Box” on page 39 Mount Arm Four Screw Holes Mount Tray Mount Locations When installing a small tray holder, face the four screw holes in the mount toward the front of the control head. 86.0087.00 Rev F 33 A-dec 300 System Installation Guide INSTALL THE DENTAL LIGHT Tools Needed For This Section Rubber mallet The procedure for installing A-dec 371 or A-dec 571 for 300 dental light components is the same regardless of the type of chair they are installed on. Routing and connecting the light cable varies depending on the system you are installing. Phillips head screwdriver Figure 56. A-dec 371 Dental Light Install the Dental Light Components 1. Slide the trim ring onto the top of the support center post then insert the intermediate post into the support center post. Orient the plugged hole in the intermediate post toward the toeboard. NOTE When the post is fully seated, there is still a gap between the white painted sections. The gap will be behind the support center covers. Figure 57. Install the Dental Light Light Post Trim Cover 2. Lightly tap the top of the intermediate post with a rubber mallet. Do not hit it too hard or it will be difficult to remove. Plastic Bearing 3. Insert the plastic bearing into the intermediate post. Connect Wires 4. Place the trim cover on the light post. 5. Route the light cable into the intermediate post. 6. Insert the light post into the intermediate post until fully seated and slip the trim cover down over the bearing and intermediate post. Intermediate Post Trim Ring Chair Circuit Board 34 Plug Chair-Base Mount 86.0087.00 Rev F Install the Dental Light Route and Connect the Dental Light Cables CAUTION Electric components on the circuit board are static sensitive and require handling precautions. 1. With the dental light components installed, finish routing the light cable through the bottom of the support center and the chair-base mount to the chair circuit board in the utilities area below the chair (see Figure 57 on the previous page). 2. Turn off the power. 3. Attach the dental light wires for the system’s type of light to their corresponding connections on the chair circuit board. ○ A-dec 371 has black and violet wires which connect to terminal strip J5 plus black/white and white wires that connect to terminal strip J6. ○ A-dec 571 for 300 has black and violet wires which connect to terminal strip J5 plus black/white, white, blue, and yellow wires that connect to terminal strip J6. Figure 58. Dental Light Wire Connections A-dec 371 Light Cable A-dec 571 for 300 Light Cable Chair Circuit Board If you have a post mount system, see the following documents that come with the kit for installing the A-dec 371 or A-dec 571 for 300 light on those chairs: NOTE • A-dec 511 Dental Chair - p/n 86.0119.00 • A-dec Cascade, Decade, Performer, or Priority Dental Chairs - p/n 86.0018.00 86.0087.00 Rev F 35 A-dec 300 System Installation Guide INSTALL THE MONITOR MOUNT Tools Needed For This Section 5/32" and 1/8" hex keys The monitor mount installs on the intermediate post. Figure 59. Monitor Mount NOTE Monitor cables are not supplied by A-dec. 1. Use a 5/32" hex key and four 3/4" socket head screws to attach the clamp for the monitor toward the top of the intermediate post. Position the mounting hole for the monitor arm toward the toe of the chair. Figure 60. Install the Monitor Mount Toe of the Chair 2. Use a 1/8" hex key to remove the ends from the monitor arm. 3. Route the power and video cables through the arm and replace the end caps. 4. Install the thrust washer on the post on the end of the mount arm then insert the post into the clamp. TIP To make it easier for the monitor arm to fit in the clamp and to avoid pinching the cables, pull the cables back into the slot of the monitor arm. Mount Arm Power and Video Cables Thrust Washer 5. Route the cables down through the square hole in the intermediate post and into the support center. Clamp 6. Insert the slotted plug into the square hole around the wires. Slotted Plug 36 86.0087.00 Rev F Install the Contoured Floor Box INSTALL THE CONTOURED FLOOR BOX Install the Contoured Floor Box Frame 1. Place the contoured floor box frame around the utilities, leaving at least 3" (76 mm) between any object that is taller than 4.5" (114 mm) and the back of the frame. Make sure that nothing extends into the gray striped area in Figure 61. Tools Needed For This Section Tape measure 1/4" masonry drill bit Phillips head screwdriver Sleeve pusher Diagonal cutters Pliers Figure 61. Contoured Floor Box Cover and Frame Contoured Floor Box Cover 3" 76 mm 4.5" 114 mm Failure to provide adequate space as described above will prevent installation of the power supply cover. CAUTION 2. Attach the frame to the floor. Nothing can extend into this zone. Contoured Floor Box Frame ○ On a wood floor, use a Phillips head screwdriver and 1-1/4" size #10 screws. ○ On a concrete floor, use a 1/4" masonry drill bit to make holes where the screws fit through the frame. Insert the plastic anchors into the holes; then use a Phillips head screwdriver and 1-1/4" size #10 screws. 86.0087.00 Rev F 37 A-dec 300 System Installation Guide Install the Convolute Figure 62. Install the Convolute and Route the Utilities 1. Unplug the chair. 2. Measure the distance between the power supply cover frame and the contoured floor box cover frame and cut the convolute to size. Cable Ties Convolute TIP The convolute should extend a few inches past each frame so you can secure it to the frames. 3. Route the convolute through the cable tie in the power supply frame until it is 2" (5 cm) past the cable tie. 4. Use pliers to tighten the cable tie so it securely holds the convolute. Trim the excess tie. 5. Route the convolute through the cable tie in the contoured floor box frame until it is 2" (5 cm) past the cable tie. Power Supply Frame Contoured Floor Box Frame 6. Use pliers to tighten the cable tie so it securely holds the convolute. Trim the excess tie. 7. Route the tubing, power cord, and vacuum line through the convolute. You are ready to connect the system. See “Connect the Utilities” on page 40 for further instructions. 38 86.0087.00 Rev F Install the Remote Floor Box INSTALL THE REMOTE FLOOR BOX 1. Remove the protective layer from the remote floor box; then position the floor box over the utilities. Tools Needed For This Section Diagonal cutters Phillips head screwdriver 5/64" hex key Sleeve pusher Figure 63. Remote Floor Box 2. Measure the distance from the floor box to the bottom of the support center and cut the convolute to match that distance. 3. Use a Phillips head screwdriver and 6 size #10 screws to secure the floor box to the floor. You are ready to connect the system. See “Connect the Utilities” on the next page for further instructions. 86.0087.00 Rev F 39 A-dec 300 System Installation Guide CONNECT THE UTILITIES See the section for the system’s configuration. Radius-Style Floor Box Utilities . . . . . 40 Radius-Style Modules . . . . . . . . . . . . 42 Radius-Style Assistant’s Instrumentation (only) . . . . . . . . . . . . . . . . . . . . . . . 46 Support Center Power and Data . . . . . 47 Support Center Floor Box Utilities . . . 53 Support Center Modules . . . . . . . . . . 56 Radius-Style Floor Box Utilities Figure 64. Radius-Style Modules with Contoured Floor Box Set Radius-style modules connect to the utilities in the contoured floor box. Use the umbilical from the kit to complete the air and water connections pictured in Figure 65. There is a blue tubing in the umbilical from the kit. Do not connect the blue tubing. CAUTION 40 86.0087.00 Rev F Connect the Utilities Figure 65. Floor Box Tubing Connections for Radius-Style Modules Umbilical from the Kit CAUTION 86.0087.00 Rev F Do not connect this blue tubing. To the Chair Base 41 A-dec 300 System Installation Guide Radius-Style Modules Figure 66. Electrical and Data Line Connections Connect Connect Black Gray Wires Wires Connect the Power and Data Plug In Data Lines Connect Ground Wire Electric components on the circuit board are static sensitive and require handling precautions. CAUTION NOTE This procedure is for both the assistant’s instrumentation and the delivery system. 1. Turn off the power. 2. Use an adjustable wrench to attach the delivery system’s ground wire to the post on the power supply. 3. Connect the ground wire to the power supply, black wire to OVAC, and gray wire to 24VAC of the chair board’s terminal strip J2. 4. Connect the two-position black connector to the two-position black 24VAC isolated lead on the power supply. Mains On/Off Button Figure 67. Properly Tied Wiring 5. Plug the data lines into any data port. NOTE The top of the 311 power supply has a diagram of the electrical connections for the chair circuit board. 6. Gather and cable tie the wiring. NOTE Make sure the power cables and data lines do not come between the power indicator light and the lens in the power supply cover. 42 Cable Ties 24VAC Isolated Lead Two-Position Connector Blue Power Light Indicator 86.0087.00 Rev F Connect the Utilities Connect the Tubing Use the umbilical from the kit to complete the air and water connections pictured in Figure 68. There is a blue tubing in the umbilical from the kit. Do not connect the blue tubing. CAUTION NOTE When installing the water manifold, be sure to connect to the barb on the end to prevent standing water in the manifold (see Figure 69 on page 45). 86.0087.00 Rev F 43 A-dec 300 System Installation Guide Figure 68. Radius-Style Modules Utility Connections Power Supply To Delivery System To Foot Control Pinch Valve Umbilical from the Kit To Floor Box To Assistant’s Syringe CAUTION Do not connect this blue tubing. With a Moisture Separator (Optional) If the system includes a moisture separator, make the connections in this gray area. To install the moisture separator, use the directions included in the moisture separator kit (p/n 41.1477.00). 44 86.0087.00 Rev F Connect the Utilities 7. If the system has an assistant’s instrumentation, install the pinch valve on the syringe yellow and red tubing. Figure 69. Install the Pinch Valve Make sure to connect to this barb. Pinch Valve 8. Use a 3/16" hex key to secure the foot control tubing to the chair base in the strain relief. Figure 70. Secure the Foot Control Tubing NOTE For proper positioning, the ridges on the foot control tubing face up to match the grooves in the strain relief. 9. Carefully remove the sticker from the top of the front mount. Tubing Ridges Face Up Figure 71. Remove the Routing Sticker Remove This Sticker 86.0087.00 Rev F 45 A-dec 300 System Installation Guide Radius-Style Assistant’s Instrumentation (only) Figure 72. Electrical and Data Line Connections Connect Connect Black Gray Wires Wires Plug In Data Lines Connect the Power and Data Electric components on the circuit board are static sensitive and require handling precautions. CAUTION 1. Turn off the power. 2. Connect the power cables to the chair circuit board. The black wire connects to 0 VAC and the gray wire to 24 VAC. 3. Plug the data lines into any data port. Connect the Tubing Figure 73. Route and Connect the Syringe Tubing 1. Connect the utilities as shown in Figure 73: ○ Use the umbilical from the kit to complete the air connection. ○ Connect the water manifold to a remote water bottle. NOTE When installing the water manifold, be sure to connect to the barb on the end to prevent standing water in the manifold. 2. Install the pinch valve on the syringe tubing. Pinch Valve Syringe Air 3. For the rest of the utility connections, see “Install the Contoured Floor Box” on page 37 and the instructions that come with the remote water bottle. Syringe Water 46 86.0087.00 Rev F Connect the Utilities Support Center Power and Data Electric components on the circuit board are static sensitive and require handling precautions. CAUTION Base Mount Systems 1. Turn off the power. 2. Use an adjustable wrench to connect the ground wire. 3. Connect the power cables to the chair circuit board terminal strip J3. The black wire connects to 0 VAC and the gray wire to 24 VAC. 4. Plug the data line into any data port. The top of the 311 power supply has a diagram of the electrical connections for the chair circuit board. NOTE Figure 74. Electrical and Data Line Connections Connect Connect Plug In Black Gray Data Wire Wire Line Connect Ground Wire Mains On/Off Button 86.0087.00 Rev F 47 A-dec 300 System Installation Guide Post Mount System on an A-dec 511 Dental Chair Figure 75. A-dec 511 Power and Data Connections Power Supply 1. Route the power cables and data line through the bottom of the support center, underneath the mount, and down the lift arm to where the power supply is located. 2. Turn off the power. 3. Use a Phillips head screwdriver to connect the ground wire to the screw on top of the power supply. 4. Connect the power cables to the chair circuit board terminal strip J2. The black wire connects to 0VAC and the gray wire to 24VAC. 5. Connect the two-position black connector to the two-position black 24VAC isolated lead on the power supply. 6. Plug the data line into a data port on the chair circuit board. 7. Route the support center tubing group through the convolute. 8. Push the convolute up several inches through the bracket in the bottom of the support center frame until it is securely attached. 9. Connect master toggle to power supply air electric switch. Ground Wire Two-Position Connector Power Cables Data Line 24VAC Isolated Lead Figure 76. Attach the Retaining Clip Inside the Remote Floor Box Inside of Floor Box 10.Route the umbilical into the floor box. 11.Use a 5/64" hex key to attach the retaining clip to the floor box and secure the convolute. Retaining Clip 12.Insert the plugs into the unused holes in the floor box. 48 86.0087.00 Rev F Connect the Utilities Post Mount System on an A-dec Cascade, Decade, or Performer Dental Chair Figure 77. Chair Circuit Board Without Data Ports Complete the procedure for the system’s type of chair circuit board. Chair Circuit Board Without Data Ports NOTE A-dec 300 touchpads cannot operate the auto dental light and chair movements when using this chair circuit board. 1. Route the tubing group and wires from the bottom of the support center through the convolute. 2. Push the convolute up several inches through the bracket in the bottom of the support center frame until it is securely attached. Figure 78. Attach the Retaining Clip Inside the Remote Floor Box 3. Route the umbilical into the floor box. Inside of Floor Box 4. Use a 5/64" hex key to attach the retaining clip to the floor box and secure the convolute. Retaining Clip Figure 79. Power Supply Connections 5. Turn off the power. 6. In the floor box, use a Phillips head screwdriver to connect the ground wire to the top of the power supply. Ground Wire 0 VAC 24 VAC 7. Connect the power cables to the power supply. The black wire connects to 0VAC and the gray wire to 24VAC. 8. Connect the two-position black connector to the two-position black 24VAC isolated lead on the power supply. 9. Insert the plugs into the unused holes in the floor box. 86.0087.00 Rev F 24VAC Isolated Lead Two-Position Connector 49 A-dec 300 System Installation Guide Chair Circuit Board With Data Ports Figure 80. Chair Circuit Board With Data Ports NOTE A-dec 300 touchpads can operate the auto dental light and chair positions when using this chair circuit board. 1. Take the extra data line from the floor box kit and plug it into a data port on the support center port board. 2. Route the data line under the support center mount and through the chair to the chair circuit board. NOTE Cable ties and adhesive-backed mounts are provided with the Performer post mount chair adaptor (p/n 77.0922.00). Use them to secure the data line under the chair adaptor. Data Ports 3. Plug the data line into a data port on the chair circuit board. 4. Route the tubing group and wires from the bottom of the support center through the convolute. 5. Push the convolute up several inches through the bracket in the bottom of the support center frame until it is securely attached. 50 86.0087.00 Rev F Connect the Utilities 6. Route the umbilical into the floor box. Figure 81. Attach the Retaining Clip Inside the Remote Floor Box 7. Use a 5/64" hex key to attach the retaining clip to the floor box and secure the convolute. Inside of Floor Box Retaining Clip 8. Turn off the power. Figure 82. Power Supply Connections 9. In the floor box, use a Phillips head screwdriver to connect the ground wire to the top of the power supply. Ground Wire 0 VAC 24 VAC 10.Connect the power cables to the power supply. The black wire connects to 0VAC and the gray wire to 24VAC. 11.Connect the two-position black connector to the two-position black 24VAC isolated lead on the power supply. 12.Insert the plugs into the unused holes in the floor box. 24VAC Isolated Lead Two-Position Connector 86.0087.00 Rev F 51 A-dec 300 System Installation Guide Post Mount System on an A-dec Priority Dental Chair Figure 83. Attach the Retaining Clip Inside the Remote Floor Box 1. Route the tubing group, power cables, and data line from the support center through the convolute. Inside of Floor Box 2. Push the convolute up several inches through the bracket in the bottom of the support center frame until it is securely attached. Retaining Clip 3. Route the umbilical into the floor box. 4. Use a 5/64" hex key to attach the retaining clip to the floor box and secure the convolute. 5. In the floor box, plug in the data line. 6. Turn off the power. 7. Connect the power cables to the power supply. The black wire connects to 0VAC and the gray wire to 24VAC. Figure 84. Power Supply Connections Ground Wire 0 VAC 24 VAC 8. Connect the two-position black connector to the two-position black 24VAC isolated lead on the power supply. 9. Insert the plugs into the unused holes in the floor box. 24VAC Isolated Lead Two-Position Connector 52 86.0087.00 Rev F Connect the Utilities Support Center Floor Box Utilities The following instructions are for both base mount and post mount systems. Support Center tubing can connect in several locations. See the section that applies to where your system’s utilities are located. • Contoured Floor Box – See Figure 85 below. • Integrated Floor Box and Remote Floor Box – See Figure 87 on page 55. NOTE There is an overall flow diagram on the inside of the support center cover. Contoured Floor Box Figure 85. Support Center Utility Connections in a Contoured Floor Box To the Support Center NOTE 86.0087.00 Rev F For moisture separator tubing connections, see Figure 86. 53 A-dec 300 System Installation Guide Figure 86. Moisture Separator Tubing Connections for Support Center Modules with a Contoured Floor Box Cover Set Power Supply From the Floor Box 54 86.0087.00 Rev F Connect the Utilities Integrated and Remote Floor Boxes Support center utilities in integrated and remote floor boxes connect in the same way. Figure 87. Support Center Utility Connections in Integrated and Remote Floor Boxes Power Supply To Support Center With a Moisture Separator (Optional) If the system includes a moisture separator, make the connections in this gray area. To install the moisture separator, use the directions included in the moisture separator kit (p/n 41.1477.00). 86.0087.00 Rev F 55 A-dec 300 System Installation Guide Support Center Modules Figure 88. Delivery System Electrical and Data Line Connections Connect the Power and Data Electric components on the circuit board are static sensitive and require handling precautions. CAUTION NOTE This procedure is for both the assistant’s instrumentation and the delivery system. Data Ports 1. Turn off the power. 2. Use an adjustable wrench to attach the delivery system’s ground wire to the post below the window in the support center frame. 3. Attach the power cables to the WAGO connectors, keeping the gray wires with gray wires (24VAC) and black wires (0VAC) with black. WAGO Connector Ground Wire 4. Connect all ground wires to the ground stud, and connect the two-position black connector from the delivery system to the two-position black connector coming from the support center umbilical. 5. Plug the data line into any data port. Connect the Tubing 1. For utility connections, see the diagram for your system’s configuration: ○ System without a cuspidor - Figure 89 on page 57 ○ System with a cuspidor - Figure 90 on page 58 NOTE When making the connections, be sure the tubing is routed through the bottom of the support center frame and you make all connections inside of the support center frame. 56 86.0087.00 Rev F Connect the Utilities Figure 89. Support Center Modules Utility Connections - No Cuspidor To Delivery System To Assistant’s Syringe To Floor Box 86.0087.00 Rev F To Foot Control 57 A-dec 300 System Installation Guide Figure 90. Support Center Modules Utility Connections - With Cuspidor To Delivery System To Assistant’s Syringe To Cuspidor To Floor Box 58 To Foot Control 86.0087.00 Rev F Connect the Utilities 2. If the system has an assistant’s instrumentation, install the pinch valve on the syringe yellow and red tubing. Figure 91. Install the Pinch Valve Pinch Valve 3. Use a 3/16" hex key to secure the foot control tubing to the chair base in the strain relief. Figure 92. Secure the Foot Control Tubing NOTE For proper positioning, the ridges on the foot control tubing face up to match the grooves in the strain relief. Tubing Ridges Face Up 86.0087.00 Rev F 59 A-dec 300 System Installation Guide INSTALL THE CUSPIDOR Tools Needed For This Section Diagonal cutters If the system has a cuspidor, install its components. Figure 93. A-dec 361 Support Center with Cuspidor Install the Cupfill Spout To install the cupfill spout, push it straight down. When installing the cupfill spout, do not rotate it or it may break. CAUTION Install the Bowl Rinse Spout To install the bowl rinse spout, push it straight in. Install the Cuspidor Bowl and Bowl Strainer Figure 94. Cuspidor Components Bowl Screen Cuspidor Bowl Cupfill Spout 1. Use diagonal cutters to cut the cable tie and remove the positioning guide. NOTE Save the positioning guide and the foam plug holding it in the cuspidor bowl support. You will need them later to verify proper clearance between the cuspidor bowl and armrest. For more information, see “Verify Proper Clearance Between the Cuspidor Bowl and the Armrest” on page 94. Bowl Rinse Spout 2. Place the cuspidor bowl in its holder with the high side away from the patient. 3. Place the bowl screen in the bottom of the bowl. 60 86.0087.00 Rev F Install Upholstery INSTALL UPHOLSTERY Tools Needed For This Section 7/64" hex key The upholstery for the A-dec 311 Dental Chair includes the seat, back, and either a patient-adjustable neck support or double-articulating headrest. Figure 95. A-dec 311 Thin-Line Back with Adjustable Neck Support Install the Seat Upholstery Figure 96. Install the Seat Upholstery Upholstery 1. Raise the chair and lower the chair back until the holes in the chair frame are accessible. 2. Move the armrests back. 3. Pull the pins out of the upholstery far enough so the seat upholstery can sit properly on the chair frame. 4. Place the seat upholstery in position and line up its holes for the pins with the holes in the chair frame. Pin 5. Push the pins back through the seat upholstery and chair frame until the rings touch the seat. 86.0087.00 Rev F 61 A-dec 300 System Installation Guide Install the Back Upholstery NOTE Back upholstery installation is the same for both back styles. Figure 97. Install the Back Upholstery Key Hole Fastener Post 1. Put the chair back up and lower the chair. 2. Place the key holes in the back upholstery over the fastener posts on the chair back. 3. Slide the back upholstery into place and push down until the back upholstery aligns with the chair back. Install the Neck Support or Headrest Figure 98. Install the Patient-Adjustable Neck Support Install a Patient-Adjustable Neck Support Snap the neck support armature into the neck support track. 62 86.0087.00 Rev F Install Upholstery Install the Double-Articulating Headrest Figure 99. Install the Double-Articulating Headrest Upholstery 1. Use a 1/8" hex key to remove the mounting screws from the headrest upholstery. Headrest Back Plate 2. Line up the holes in the headrest upholstery with those in the headrest back plate. 3. Insert and tighten the mounting screws. 4. Place the glide bar into the slot at the top of the back upholstery and push it down into place. Mounting Screws Figure 100. Install the Double-Articulating Headrest Glide Bar 86.0087.00 Rev F 63 A-dec 300 System Installation Guide PREPARE AND ADJUST THE SYSTEM Secure the Base Mount System Tubing Bundle Tools Needed For This Section 1/2" combination wrench Voltmeter Hex key set Phillips head screwdriver Figure 101. Secure the Tubing Bundle Cable Tie Around Support Center Post 1. Loosely loop a cable tie around the tubing bundle and slide the tie under the chair-base mount. 2. From the utilities area, pull the tubing bundle to take up any slack. 3. Tighten the cable tie under the mount and push the tubing bundle up into the mount so it is not visible. 4. Use a cable tie around the corner of the mount to secure the wires and tubing so nothing is hanging down under the mount. 5. Use a cable tie to secure the tubing group to the support center post so it does not push against the lower covers. Cable Tie Around Tubing Bundle Inside Mount Cable Tie Around Corner of the Mount Figure 102. Install the Water Bottle Install the Water Bottle The instructions for installing a water bottle are the same regardless of where it is located. Lubricate O-Ring 1. Following ICX instructions, place an ICX tablet (p/n 90.1065.00) in the water bottle. 2. Fill the bottle with water. 3. Put A-dec silicone lubricant on the receptacle O-ring. Use only A-dec silicone lubricant or the O-ring may be damaged. CAUTION 4. Push the water bottle up onto the water bottle receptacle and turn it to the right. 64 86.0087.00 Rev F Prepare and Adjust the System Adjustments Lock the Armrests (Optional) Figure 103. Adjust the Armrest Position Stop Screw (From Front) When the dental chair is shipped, the armrests rotate back and forth. The armrests can be locked into the upright position if preferred. To lock an armrest: 1. Use a 3/16" hex key to remove the stop screw from the front of the armrest. 2. Install the stop screw in the back of the armrest. 3. Use a 1/8" hex key to tighten the tension adjustment screw on the front of the armrest. Stop Screw (From Back) Adjust the Double-Articulating Headrest Glide Bar Tension Tension Adjustment Screw Figure 104. Adjust the Double-Articulating Headrest Glide Bar Tension A double-articulating headrest may be difficult to move or may drift downward because of the amount of tension on the glide bar. To adjust the tension, use a 1/8'' hex key and turn the tension adjustment screw clockwise to increase friction or counterclockwise to decrease friction. Tension Adjustment Screw 86.0087.00 Rev F 65 A-dec 300 System Installation Guide Adjust the Speed for Lowering the Chair To adjust the speed at which the chair lowers, raise the chair and use a 3/32" hex key to adjust the base down control valve. Turn the screw clockwise to decrease speed or counterclockwise to increase speed. TIP For an accurate estimate of Figure 105. Adjust the Base Down Control Valve Adjustment Screw Base Down Control Valve down travel speed, have someone sit in the chair when making the adjustments. 66 86.0087.00 Rev F Prepare and Adjust the System Adjust the Delivery System Spring-Assisted Flexarm Figure 106. Adjust the Spring-Assisted Flexarm End Cap If the control head drifts up or down, complete the following steps to adjust the spring-assisted flexarm. Button Head Screw NOTE After adjusting the flexarm, test for balance at its normal working position. 1. Turn off the power. 2. Load the control head for normal use, attaching handpieces and placing a tray with instruments on the tray holder. 3. Use a 1/8" hex key to loosen the screw that secures the end cap. Flexarm Spring Adjustment Screw TIP Lower the control head for easier access to the button head screw. 4. Remove the flexarm end cap. 5. Use a 3/16" hex key to adjust the flexarm spring adjustment screw until the control head maintains position at the normal working position. TIP To make it easier to adjust the spring, raise the control head to the highest point. ○ If the control head drifts up, turn the screw counterclockwise. ○ If the control head drifts down, turn the screw clockwise. 86.0087.00 Rev F 67 A-dec 300 System Installation Guide Adjust the Delivery System Flexarm Rotation Tension Do not adjust the flexarm rotation tension until the system has been leveled. For information on adjusting the flexarm rotation, see “Adjust the Delivery System Flexarm Rotation Tension” on page 79. Install the Delivery System Flexarm Rotation Stop Screw Figure 107. Install the Flexarm Rotation Stop Screw Use a 1/8" hex key to install the 10-32 x 1/2" button head screw in the bottom hole on the knuckle of the delivery system. Tighten it all the way down. Install the 10-32 x 1/2" button head stop screw (actual size) in the bottom hole. Adjust the Control Head Rotation Tension Figure 108. Adjust the Control Head Rotation Tension If the control head rotation is too tight or too loose, adjust the rotation tension by tightening or loosening the screw under the control head. Use a 5/32" hex key to turn the screw clockwise to increase tension and counterclockwise to decrease tension. 68 86.0087.00 Rev F Prepare and Adjust the System Adjust the Tray Rotation Tension If the tray holder rotation is too tight or too loose, adjust the tray rotation tension. 1. Insert a 1/8" hex key through the mounting bracket. If necessary, rotate the holder or arm until the key slides completely into the mounting bracket. 2. While holding the hex key in the bracket, rotate the tray clockwise to increase the tension and counterclockwise to decrease tension. Figure 109. Adjust the Tray Holder on a Traditional Control Head Hex Key Tray Holder Tray Arm Mounting Bracket Control Head Figure 110. Adjust the Tray Holder on a Continental Control Head For Continental-style delivery, you may need to make this adjustment in two locations: under the control head and under the tray holder. B A E D C (A) Hex Key; (B) Tray Holder Mounting Bracket; (C) Tray Holder; (D) Tray Arm; (E) Control Head Mounting Bracket 86.0087.00 Rev F 69 A-dec 300 System Installation Guide Adjust Handpieces Figure 111. Handpiece Adjustment Keys A-dec 300 delivery systems supply and regulate air and water to operate dental handpieces, syringes, and ancillary devices. To adjust handpieces, complete the following sections in their order for each handpiece. 1) Water Coolant Flow The A-dec 300 system includes a water coolant flow control for each handpiece. 1. Turn the drive air, air coolant, and water coolant all the way down. Air and water coolant adjustment keys are not intended to completely shut off flow. CAUTION Air Coolant Flow Control Water Coolant Flow Control For instructions on accessing drive air pressure controls, see “3) Drive Air Pressure”on the next page. 2. Lift a handpiece from the holder. 3. Flip the wet/dry toggle toward the blue dot on the foot control. 4. Press the foot control to activate the handpiece. 5. Use the adjustment keys on the side of the control head to adjust the water coolant flow (normally about 1 drop every 2 seconds): ○ Turn the control clockwise to decrease flow. ○ Turn the control counterclockwise to increase flow. 2) Air Coolant Flow The air coolant flow control adjusts the air coolant flow to all handpieces. 1. With the handpiece out of its holder, press the foot control to activate the handpiece. 2. Adjust the air coolant flow until the spray is a fine mist. ○ Turn the control clockwise to decrease flow. ○ Turn the control counterclockwise to increase flow. Do not continue turning the air coolant adjustment key left when the flow is no longer increasing. The stem may come out completely. CAUTION 70 86.0087.00 Rev F Prepare and Adjust the System 3) Drive Air Pressure Figure 112. Remove the Control Head Back Cover The optional deluxe touchpad allows you to easily check drive air pressure. On the touchpad, press and hold the Plus (+) and Minus (–) buttons at the same time. If an adjustment is necessary, use a 7/64" hex key to remove the control head back cover and complete the steps below for each handpiece. NOTE Use a handpiece pressure gauge attached to the handpiece tubing for exact drive air measurement or if you do not have a deluxe touchpad. One bar equals 14.5 psi. 1. Lift the handpiece from the holder. 2. Locate the drive air pressure controls inside the control head. 3. Set the wet/dry toggle to dry. 4. Press the foot control. 5. With the handpiece running, watch the deluxe touchpad readout or handpiece pressure gauge and adjust the handpiece drive air pressure to meet manufacturer’s specifications. • Turn the control clockwise to decrease flow. • Turn the control counterclockwise to increase flow. Control Head Back Cover Screws Figure 113. Adjust the Drive Air Pressure Complete these steps for each handpiece. NOTE Adjust the drive air pressure to meet the handpiece manufacturer’s drive air pressure specification. See the handpiece documentation for the drive air pressure specification. Drive Air Pressure Control 86.0087.00 Rev F 71 A-dec 300 System Installation Guide Adjust the Intraoral Light Source Voltage Figure 114. Adjust the Intraoral Light Source Voltage To adjust the intraoral light source, complete the following steps. 1. Use a 7/64" hex key to remove the control head back cover (see Figure 112 on page 71). Use Buttons to Adjust Voltage 2. Set the voltmeter to DC voltage and place its probes on the IOLS output terminals for the handpiece you're adjusting. 3. Lift the handpiece from its holder and turn the light on. 4. Use the buttons behind the terminal to adjust the voltage according to the following table. IOLS Output Terminals NOTE The following table pertains to fiber optics with 26 AWG wires, 750mA loads, and a desired bulb voltage of 3.2 VDC. For fiber optics powered with 26AWG wires and other ratings, use the equation T = (Z x 0.006 x Y) + X where: T = Terminal strip voltage (VDC) X = Desired voltage at lamp (VDC) Y = Rated lamp/load current (in Amps) Z = Length of 26AWG wire (inches) from terminal strip to lamp Length and Voltage Table Wire length in A-dec tubing 72 Voltage at terminal strip A-dec/W&H, Bien Air or other bulbs rated at 3.2V Wire length in A-dec tubing Voltage at terminal strip A-dec/W&H, Bien Air or other bulbs rated at 3.2V (in) (cm) VDC +/- .02 (in) (cm) VDC +/- .02 48 122 3.40 108 274 3.69 54 137 3.43 114 290 3.72 60 152 3.46 120 305 3.75 66 168 3.49 126 320 3.78 72 183 3.52 132 335 3.81 78 198 3.55 138 351 3.84 84 213 3.58 144 366 3.87 90 229 3.61 150 381 3.90 96 244 3.64 156 396 3.93 102 259 3.67 86.0087.00 Rev F Prepare and Adjust the System Adjust Syringes Figure 115. Adjust the Syringe Flow To adjust the water and air flow for syringes, complete the following steps. 1. Access the pinch valve: Adjust Air Flow ○ For the doctor’s syringe, use a 7/64" hex key to remove the control head back cover (see Figure 112 on page 71). ○ For the assistant’s syringe, go to the support center. 2. On the syringe, push the water button. 3. Tighten or loosen the screw on the pinch valve to adjust the water flow. 4. Push both buttons on the syringe. 5. Tighten or loosen the screw on the pinch valve to adjust the air flow until there is a fine mist. Adjust the Continental Whip Tension Doctor’s Syringe Adjust Water Flow Figure 116. Adjust the Continental Whip Tension 1. Use a 7/64" hex key to remove the control head back cover (see Figure 112 on page 71). 2. Use a 7/64" hex key to undo the two screws holding the control head front cover and raise the cover up. 3. Pull the two tabs on the top of the drip shield out; then pull down the shield. 4. Remove the spring from the front cover and the bail. Least Tension Most Tension Spring Bail Whip 5. Pull the whip forward. 6. To remove the bail from the control head, squeeze its sides until its ends clear the holes in the frame holding the bail. 7. Insert the ends of the bail into the holes for the desired whip tension; then replace the spring. The deepest hole provides the least tension. Drip Shield 8. Repeat steps 4 through 7 for each whip to be adjusted. 9. Replace the drip shield and control head covers. 86.0087.00 Rev F 73 A-dec 300 System Installation Guide Adjust the Dental Light Flexarm Counterbalance If the dental light drifts up or down, complete the following steps to adjust the flexarm counterbalance. Figure 117. Adjust the Dental Light Flexarm Counterbalance Cover Cover Screw End Cap 1. Use a Phillips head screwdriver to remove the end cap. 2. Use a 5/16" hex key to remove the two screws that secure the cover. 3. Remove the cover. 4. Use a 1/2" combination wrench to adjust the nut on the end of the spring. If the dental light drifts up, turn the nut counterclockwise. If the dental light drifts down, turn the nut clockwise. 5. Set the cover back onto the flexarm (but do not reattach it yet), and check for drift. 6. Repeat steps 3 and 4 until drift is eliminated. NOTE An optional travel stop limit kit (p/n 90.1044.00) can be installed to limit the upward and downward motion of the flexarm. 74 86.0087.00 Rev F Prepare and Adjust the System Adjust the A-dec 371 Dental Light Swivel Tension and Vertical Drift Figure 118. Horizontal Rotation Adjustments If the light head is difficult to position, moves too easily, or tends to drift out of position, complete the following steps to adjust the horizontal and vertical tension. • For horizontal rotation, use a 5/32" hex key to adjust the screw at the top of the switch housing. Turn the screw clockwise to increase tension and counterclockwise to decrease tension. Horizontal Rotation Tension Screws Figure 119. Vertical Rotation Adjustments • For vertical rotation, use a 3/16" hex key to adjust the tension. Turn the screw clockwise to increase tension and counterclockwise to decrease tension. Vertical Tension Screw 86.0087.00 Rev F 75 A-dec 300 System Installation Guide Adjust the Cuspidor Bowl Rinse Flow Settings Figure 120. Bowl Rinse Flow Pinch Valve Location Complete the following steps to adjust the bowl rinse flow, located inside the support center. 1. Remove the upper support center cover by pulling out on the lower edges of the cover. 2. With the cuspidor bowl rinse on, tighten or loosen the pinch valve to adjust the flow. 3. Rotate the bowl rinse spout to adjust the flow pattern and achieve the best rinsing action. 4. Use the touchpad buttons to set the length of cupfill and bowl rinse flow (see “Cupfill and Bowl Rinse” on page 81 for more information). Bowl Rinse Flow Pinch Valve Support Center Cover Figure 121. Adjust the Monitor Mount Tilt and Drift Monitor Handle Adjust the Monitor Tilt and Drift If you want to change the tilt of the monitor or if it tends to drift out of position, complete the following steps to make adjustments. Horizontal Tension Screw • To adjust the tilt of the monitor, turn the monitor handle to the left, adjust the angle of the monitor; then turn the handle to the right to lock it in position. • To adjust the horizontal drift of the monitor, use a 5/16" hex key to tighten or loosen the horizontal tension screw. 76 86.0087.00 Rev F Level the System LEVEL THE SYSTEM Tools Needed For This Section Complete the following sections to level the system. A level system prevents the modules from drifting and instruments from rolling around the tray. Magnetic level Hex key set NOTE To successfully level the system, it is important to complete the following sections in the order that they are listed. Figure 122. Level the Support Center 1. Level the Support Center 1. Move the control head and dental light into the normal working position for the doctor. 2. Place a magnetic level vertically on the knuckle of the delivery system’s rigid arm. Align the level with a set of leveling screws that are on opposite sides of the chair-base mount. TIP If needed, use a piece of tape to help hold the level in place. Magnetic Level 3. Use a 3/16" hex key to adjust the 4 leveling screws while moving the level’s alignment from one set of screws to the other. 4. Once the system is level, securely tighten the leveling screws. One Set of Leveling Screws 86.0087.00 Rev F 77 A-dec 300 System Installation Guide 2. Level the Delivery System 1. With the control head in the normal working position for the doctor, turn it so it faces straight out from the flex arm. Figure 124. Delivery System Ready to be Leveled Flex Arm 2. Place a level on the tray holder, if the system has one, or on top of the control head. 3. Use a 5/32" hex key to loosen the button-head screw holding the control head in place. 4. Make the following adjustments to level the control head. When making adjustments, align the level with the screws you are adjusting (see Figure 123). ○ Use a 5/32" hex key to level the control head from side to side. Control Head ○ Use a 5/32" hex key to level the control head from front to back. 5. Tighten the button-head screw to hold the control head in place. Figure 123. Level the Delivery System Button Head Screw Adjust to Level Front to Back Adjust to Level Side to Side NOTE If the system does not have an assistant’s instrumentation mounted on a telescoping arm, you have completed the leveling process and can skip to “Adjust the Delivery System Flexarm Rotation Tension” on page 79. 78 86.0087.00 Rev F Level the System 3. Level the Assistant’s Instrumentation Telescoping Arm 1. Extend the telescoping arm and align it parallel to the chair. 2. Place a level on top of the telescoping arm. Figure 125. Level the Assistant’s Instrumentation Telescoping Arm Extended Telescoping Arm Level 3. Use a 1/8" hex key to level the arm. Adjust the Delivery System Flexarm Rotation Tension NOTE This adjustment must be made after the final leveling of the system. For information on other adjustments, see “Prepare and Adjust the System” on page 64. Figure 126. Adjust the Delivery System Flexarm Rotation Tension Rotation Tension Setscrew is on Top If the delivery system arm drifts, complete the following steps to adjust the tension of the flexarm rotation. 1. Load the control head for normal use. 2. Use a 1/8" hex key to turn the setscrew clockwise to increase tension and counterclockwise to decrease tension. 86.0087.00 Rev F 79 A-dec 300 System Installation Guide TO U C H P A D SETTINGS Figure 127. Standard and Deluxe Touchpads Dental Chair and Light Settings Use the touchpad to program commonly used settings for the system. The touchpad comes with the following preset positions: • Entry/Exit • Treatment 1 NOTE Touchpad symbols are proprietary to A-dec Inc. • Treatment 2 (Standard touchpad only) • X-ray/Rinse Table 1. Chair Preset Positions Program the Chair Positions Button To program Entry/Exit, Treatment 1, and Treatment 2 buttons: 1. Move the chair into the desired position. 2. Press and release and you hear a beep indicating program mode. Position Description Entry/Exit: Automatically positions chair for entry/exit and turns off the dental light. Treatment 1: Automatically positions the chair base and back down and turns on the dental light. 3. Press the button you want to program and you hear three beeps confirming the button has been set. Treatment 2 (Standard touchpad only): Automatically positions the chair base and back and turns on the dental light. Customize the X-Ray/Rinse Button X-ray/Rinse: Automatically positions the chair for either x-ray or rinse. Toggles between the xray/rinse and the last position, and turns off the dental light. The X-ray/Rinse button functions as either x-ray/rinse or as another fully programmable preset position (Treatment 3). To change the function: 1. Press and hold buttons and same time for three seconds. NOTE The A-dec 300 has a fixed cuspidor, which requires setting both the base and back position. at the ○ One beep indicates the button has been configured as Treatment 3. ○ Three beeps indicate that the X-ray/Rinse button has been configured as the x-ray/rinse function (alternates between the x-ray/rinse and the previous position). 80 86.0087.00 Rev F Touchpad Settings 2. Program the preset position as instructed in the previous section, “Program the Chair Positions” on page 80. TIP If X-ray/Rinse is changed to a fully programmable preset position, it operates the same as Treatment buttons 1 and 2. Cupfill and Bowl Rinse Table 2. Cupfill and Bowl Rinse Presets Button Position Description The cupfill controls water flow from the water bottle into a cup. The bowl rinse provides rinse water for the cuspidor. See Table 2 for those buttons preset run times. Cupfill Button: Controls water flow from the water bottle into a cup. • Press the Cupfill button for a timed operation. The factory preset is a 2.5 second fill. To change the timing for the cupfill and bowl rinse: • Press and hold the Cupfill button for manual operation. 1. Press and release . One beep indicates the programming mode is ready. NOTE You can also enter the cuspidor programming mode by pressing and holding both the Cupfill and Bowl Rinse buttons located near the spout. One beep indicates the programming mode is ready. Bowl Rinse Button: Provides rinse water for the cuspidor bowl. • Press the Bowl Rinse button for a timed operation. The factory preset is a 30 second rinse. • Press and hold the Bowl Rinse button for manual operation. 2. Press and hold the Cupfill or Bowl Rinse button for the desired amount of time. 3. Release the button. Three beeps confirm the setting. TIP If you press the Bowl Rinse button twice in less than two seconds, the operation changes to continuous rinse mode. Press the button once to end the continuous bowl rinse mode. 86.0087.00 Rev F 81 A-dec 300 System Installation Guide Dental Light Use the Dental Light button on the touchpad to turn the dental light on or off. Press the Dental Light button to toggle between intensity and composite settings. Table 3. Dental Light Button Button Description Dental LIght Button: Press to toggle between intensity settings. Press the button to turn on the light and hold the button to turn off the light. • 371 Dental Light: Toggle between high intensity and composite. • 571 for 300 Dental Light: Toggle between either high or medium intensity and composite, depending on the selection of the intensity setting on the light. To turn off the dental light, hold down for two seconds. Auto Light Feature When you use a programmed chair position, the dental light turns on when the chair back reaches operating position. Press or and the dental light automatically turns off. To deactivate the auto light feature, press and hold and at the same time for three seconds. One beep confirms the auto light feature is off. To re-activate the auto light feature, press and hold and at the same time for three seconds. Three beeps confirm the dental light auto feature is on. TIP If X-ray/Rinse is changed to a preset position, the dental light auto feature operates the same as the Treatment buttons 1 and 2. 82 86.0087.00 Rev F Touchpad Settings Electric Handpiece Operation Table 4. Electric Motor Presets (Standard Mode) Memory Settings Preset Speed M1 2,000 RPM You can program handpieces in standard or endodontics mode. M2 10,000 RPM M3 20,000 RPM • Standard mode: Handpieces are always 100 percent torqued for electric handpieces. M4 36,000 RPM Figure 128. Standard Mode Touchpad Screen • Endodontics mode: If you have a motor controller that supports endodontics, you can program handpieces in this mode. Endodontics mode allows you to control torque and ratio settings. Program the Electric Handpieces in Standard Mode E You can program memory settings on the deluxe touchpad for specific RPMs. To program the handpiece setting: D A B C 1. Lift the handpiece from the holder. 2. To adjust the RPM, press the Minus (-) or Plus (+) buttons. You see the RPM values on the screen (see Figure 128). 3. To place the setting into memory, press . One beep sounds. 4. To toggle through memory settings (M1 through M4), press . 5. When the desired memory location is selected, press . Three beeps confirm the setting. Forward/Reverse Button Use the Forward/Reverse button to change the electric motor’s direction. The system defaults to the forward position when you return the motor to the holder or turn off the power (see Figure 128). In reverse mode, the screen icon flashes continuously. 86.0087.00 Rev F Item Description A Forward/Reverse Indicator (forward is shown) B Forward/Reverse Button C Minus and Plus Buttons D Memory Button E RPM Value 83 A-dec 300 System Installation Guide Program Electric Handpieces in Endodontics Mode The endodontics mode allows you to change settings based on the specific file and desired handpiece behavior. NOTE For more information regarding speed and torque limits for a specific file, consult the file manufacturer. To program the endodontics memory settings: Figure 129. Endodontics Mode Touchpad Screen A L K B J C H D E G F 1. Lift the handpiece from the holder. 2. If the touchpad screen does not display endodontics mode, press . 3. To change settings in endodontics mode, press the Minus (-) or Plus (+) buttons. You see a white reverse video box on the touchpad screen. 4. To move from setting to setting on the touchpad screen, press the chair positioning buttons. 5. To change the setting, press the Minus (-) or Plus (+) buttons. 6. To set the speed limit, torque limit, and ratio into memory (optional), press (one beep sounds); then press to toggle through the memory locations (M1 through M4). 7. Once the memory location is selected, press . Three beeps confirm the setting. 84 Item Description Item Description A Endodontics Mode Button G "Minus" Button B Forward/Reverse Indicator (with automode indicator inside arrow) H Memory Button C File Torque Unit Indicator J Handpiece Ratio Setting D Forward/Reverse Button K Speed Limit Indicator E "Plus" Button L Memory Setting Indicator F Program Button 86.0087.00 Rev F Touchpad Settings Table 5 lists and defines the touchpad screen icons for endodontics mode. Table 5. Endodontics Mode Settings Icon The A-dec|W&H endodontics attachments have special features due to their ball-bearing design. Their life-long efficiency factors are stable and known; therefore, the A-dec endodontics system is able to control and display file torque very accurately. All other attachments have unknown life-long efficiency factors and therefore stated torque values are approximate. CAUTION Setting Description Speed Setpoint for file speed limit. For more information, consult the file manufacturer. Torque Setpoint for file torque limit. For more information, consult the file manufacturer. Torque Units Toggles between Ncm (Newton centimeters) and gcm (Gram centimeters). Adjusting this setting for one handpiece changes it for all handpiece settings. Note: 1 Ncm = 102 gcm. Ratio Sets the handpiece ratio. For more information, consult the handpiece manufacturer. Forward/ Reverse Mode Adjusting this setting for one handpiece changes it for all handpiece settings. This icon is displayed within the Forward/Reverse indicator. Auto off Auto-off — The motor shuts off when the file speed reaches the torque limit. Auto reverse Auto-reverse — The motor stops and reverses direction when the file reaches the torque limit. Auto forward Auto-forward — When the file reaches the torque limit, the motor stops, reverses three turns, then changes back to forward again. Note: If the file is stuck, this cycle repeats three times before the motor stops. 86.0087.00 Rev F 85 A-dec 300 System Installation Guide Technician Touchpad Setup Options The deluxe touchpad allows service technician access to adjust handpiece and touchpad settings for user preferences. Navigate with Touchpad Buttons Chair button functions become navigation buttons while you are in setup mode. You will use the back up () and back down () button to navigate the setup screens. Holder Setup You can set up how handpieces are configured for each handpiece holder. To set up handpiece holders: 1. From the deluxe touchpad main screen, press and hold and at the same time for three seconds; then press to begin. 2. From the System Setup screen, press Minus (-) or Plus (+) to highlight Handpieces (see Figure 130), and press . 3. From the Handpiece Setup screen, select Holder Setup (see Figure 131), and press . 4. Lift the desired handpiece from the holder. 86 Figure 130. System Setup: Handpieces SYSTEM SETUP Handpieces Press -/+ to change < Exit Select > Figure 131. Handpiece Setup: Holder Setup Selection HANDPIECE SETUP Holder Setup Press -/+ to change <Back Select > 86.0087.00 Rev F Touchpad Settings 5. From the Handpiece Setup screen, press Minus (-) or Plus (+) to highlight the handpiece type (see Figure 132). Handpiece types include Electric, Ultrasonic, Vacuum, Other, and Turbine. Figure 132. Handpiece Setup: Option Selection HANDPIECE SETUP Electric Press -/+ to change Select > 6. Press . Three beeps confirm the setup for the handpiece is complete. 7. Return the handpiece to the holder. 8. Repeat steps 5 through 7 to set up each handpiece. 9. When you are finished setting up handpieces, press until you see the A-dec logo. Light Source Setup You can set up various intraoral light source options if a quad voltage intraoral light source (QVIOLS) is installed. The following options are available: • On When Selected: Specify whether the intraoral light source turns on or remains off when the handpiece is removed from the holder. • Auto Off Delay: Determine how long the light remains on when the foot control is released. This time is reset when you use the drive air again. • On in Endo: Specify whether the intraoral light source turns on or off when in endodontics mode. Because most enododontics handpieces do not have light pipes, it is recommended that Off is selected to reduce heat and to extend bulb life. 86.0087.00 Rev F 87 A-dec 300 System Installation Guide To set up the light source: 1. From the deluxe touchpad main screen, press and hold and at the same time for three seconds; then press to begin. 2. From the System Setup screen, press Minus (-) or Plus (+) to highlight Handpieces (see Figure 133), and press . 3. From the Handpiece Setup screen, press Minus (-) or Plus (+) to highlight Intraoral Lt Sc (see Figure 134), and press . 4. Lift the desired handpiece from the holder. 5. From the Light Source Setup screen, press Minus (-) or Plus (+) to display the desired option, such as On When Sel (see Figure 135), and press . 6. Specify options for the handpiece setting by pressing Minus (-), Plus(+), and to move through the screens. Figure 133. System Setup: Handpieces SYSTEM SETUP Handpieces Press -/+ to change < Exit Select > Figure 134. Handpiece Setup: Intraoral Lt Sc Selection HANDPIECE SETUP Intraoral Lt Sc Press -/+ to change < Back Select > Figure 135. Light Source Setup: Option Selection LIGHT SOURCE SETUP On When Sel LT SOURCE SETUP Press -/+ to change On When Sel < Back Select Press -/+>to change <Back Select> Once the setup is complete, three beeps confirm the setting. 7. Return the handpiece to the holder. 8. Repeat steps 3 through 7 to configure each handpiece. 9. When you are finished setting up the handpieces, press until you see the A-dec logo. 88 86.0087.00 Rev F Touchpad Settings Ultrasonic Setup Specify whether to turn on or off the ultrasonic colors. To set up the ultrasonic: 1. From the deluxe touchpad main screen, press and hold and at the same time for three seconds, and press to begin. 2. From the System Setup screen, press Minus (-) or Plus (+) to highlight Handpieces (see Figure 136), and press . Figure 136. System Setup: Handpieces 3. From the Handpiece Setup screen, press Minus (-) or Plus (+) to highlight Ultrasonic Setup (see Figure 137), and select . Figure 137. Handpiece Setup: Ultrasonic Setup Selection 4. Press Minus (-) or Plus (+) to display On or Off, and press . SYSTEM SETUP Handpieces Press -/+ to change < Exit Select > HANDPIECE SETUP Ultrasonic Setup HANDPIECE SETUP Press -/+ to change Setup < Back Select > Press -/+ to change Select> <Back Three beeps confirm the setting. 5. Press until you see the A-dec logo. 86.0087.00 Rev F 89 A-dec 300 System Installation Guide Set Up Electric Options You can change display information and the electric functions. The following options are available: • Torque Units: Select how to display the units, either Ncm (Newtons per centimeters) or gcm (grams per centimeters). • Endo Handpiece Auto Mode: Configure how the electric motor reacts when the torque limit is reached. ○ Auto Forward: Motor will stop, reverse three turns, and return forward again. ○ Auto Reverse: Motor will reverse. ○ Auto Stop: Motor will stop. • Auto Reverse Beep: Select whether to hear three beeps from the touchpad when auto-reversing begins. • Torque Warning: Select whether to turn on the beep sounds from the touchpad when there is a torque warning. 90 86.0087.00 Rev F Touchpad Settings To set up electric options: 1. From the deluxe touchpad main screen, press and hold and at the same time for three seconds; press to begin. 2. From the System Setup screen, press Minus (-) or Plus (+) to highlight Handpieces (see Figure 138), and press . Figure 138. System Setup: Handpieces SYSTEM SETUP Handpieces Press -/+ to change < Exit Select > 3. From the Handpiece Setup screen, press Minus (-) or Plus (+) to highlight Electric Setup (see Figure 139), and press . Figure 139. Handpiece Setup: Electric Setup Selection 4. From the Electric HPC Setup screen, press Minus (-) or Plus (+) to highlight an option, such as Torque Units (see Figure 140), and press . Figure 140. Electric Handpiece Setup: Option Selection 5. Specify options for the setting by pressing Minus (-), Plus(+), and to move through the screens. HANDPIECE SETUP Electric Setup Press -/+ to change < Back Select > ELECTRIC HPC SETUP Torque Units Press -/+ to change < Back Select > Once the setup is complete, three beeps confirm the setting. 6. Repeat steps 4 through 5 to configure electric settings for each handpiece. 7. When all settings are complete, press until you see the A-dec logo. 86.0087.00 Rev F 91 A-dec 300 System Installation Guide Touchpad Setup Use the touchpad setup to change the contrast of the touchpad display and specify whether to display help messages in the technical mode help screens. To set up touchpad options: 1. From the deluxe touchpad main screen, press and hold and at the same time for three seconds, and press to begin. 2. From the System Setup screen, press Minus (-) or Plus (+) to highlight Touchpad (see Figure 141), and press to begin. Figure 141. Touchpad Setup 3. From the Touchpad Setup screen, press Minus (-) or Plus (+) to highlight Contrast or Help Messages: Figure 142. Touchpad Setting: Contrast ○ Select Contrast to adjust the contrast on the screen. Press Minus (-) or Plus (+) to adjust the contrast (see Figure 142). ○ Select Help Messages and press Minus (-) or Plus (+) to turn on or off the technician help messages (see Figure 143). 4. Press . Three beeps confirm the setting. SYSTEM SETUP Touchpad Press -/+ to change < Exit Select > CONTRAST ADJUST 26 of 63 Press -/+ to change < Back Save > Figure 143. Touchpad Setting: Help Messages HELP MESSAGES On Press -/+ < Back to change Save > 5. When settings are complete, press until you see the A-dec logo. NOTE For a listing of touchpad help messages, see A-dec 300 Systems Instructions For Use. 92 86.0087.00 Rev F Test the System TE S T T H E S Y S T E M Assistant's Instrumentation, Cuspidor, and Cupfill • Vacuum/suction is set to the following specifications: After you have completed the installation of the A-dec 300 system, use the following checklist to test the system; then follow up with the customer. Dental Chair • Dental chair is securely anchored to the floor. • The chair functions properly, including the chair's lift, tilt, and programmable functions controlled by the touchpads and foot switch. • Headrest or neck support is functioning properly. • Armrests are fixed or positionable based on staff preference. Delivery System • The handpieces function properly and are set to doctor’s preferences: ○ Water coolant, air coolant, and spray. ○ Handpiece tubing flush. ○ Wet vacuum - 10±2 inches of Hg (34±7 KPa), 9 scfm (255 sl/min) minimum. ○ Dry/semi-dry vacuum - 4.5±1 inches of Hg (16±3.5 KPa), 12 scfm (340 sl/min) minimum. • Air and water syringe buttons operate smoothly. • Valves on HVE and saliva ejector move freely. • Cupfill timing is set properly. • Cuspidor flow pattern effectively rinses the bowl, and the bowl drains properly. • The end of the cuspidor ventilation tube is 1/2" (13mm) above the support center frame. • There is 1-1/8" (29 mm) clearance between the cuspidor bowl and armrest. Utilities • Regulated air pressure is at 80 psi. • Floor box is free of air and water leaks. • The gravity drain is functioning properly. • Tubing and connections are not kinked and are free of air and water leaks. • Excess tubing and wires are coiled and stored away from moving parts. ○ Handpiece holder valves and switches. ○ Handpiece drive air pressure. • Spring-assisted flexarm functions properly. • Delivery system flexarm brake functions properly. • Control Head is level as measured on tray. • Articulating arms do not drift. • Foot control operates properly. Dental Light • All light intensity settings work properly (two settings for the 371, three for the 571 for 300). • Dental light has a spare bulb (A-dec p/n 041.709.00). • Dental light flexarm and head tension are properly adjusted. 86.0087.00 Rev F 93 A-dec 300 System Installation Guide VERIFY PROPER C LEARANCE BETWEEN THE CUSPIDOR BOWL AND THE ARMREST NOTE If you have a Radius-style system, a base mount system without a cuspidor, or a post mount system, skip this section and go to “Install the Covers” on page 96. Tools Needed For This Section 5/16" hex key driver Figure 144. Ensure the Positioning Guide Clears the Armrest Positioning guide must clear armrest. Check For Proper Clearance 1. Remove the cuspidor bowl. 2. Place the positioning guide on the cuspidor bowl support with the tab toward the head of the chair and secure it with the foam plug in the guide’s center. Positioning Guide 3. Raise and lower the chair to check that the positioning guide clears the armrest. If the guide does not properly clear the armrest, complete the steps on the next page to ensure proper clearance. 94 86.0087.00 Rev F Verify Proper Clearance Between the Cuspidor Bowl and the Armrest Adjust the Clearance Between the Cuspidor and the Armrest Figure 145. Loosen the Support Center Screws 1. With the positioning guide in place, use a 5/16" hex key driver to loosen the button head and socket head screws. Button Head and Socket Head Screws 2. If the system includes a delivery system and light, position them over the support center. Figure 146. Secure the Support Center 3. Rotate the support center until the positioning guide properly clears the armrest. 4. If the system includes a delivery system, pull its rigid arm toward the screws until the ribs on the support center frame are flush against the support center post. 5. Use approximately 13 ft-lb. [1.8 m-kgs] of torque to tighten the button head screw. 6. Alternately tighten both socket head screws until they are firmly secure (use approximately 13 ft-lb. [1.8 m-kgs] of torque). Ribs on the Support Center Frame 7. Raise and lower the chair to check that the positioning guide clears the armrest. 8. If the guide does not properly clear the armrest, repeat steps 1 through 8. 9. Verify that the support center is level. If it needs to be adjusted, complete the steps in “1. Level the Support Center” on page 77. Pull the Rigid Arm Toward the Screws 86.0087.00 Rev F 95 A-dec 300 System Installation Guide INSTALL THE COVERS Tools Needed For This Section Diagonal cutters The A-dec 300 system includes a variety of covers for different configurations. Depending on the system's configuration, you may need to modify the covers during installation. Phillips head screwdriver Figure 147. A-dec 300 System with Support Center NOTE Be sure to adjust and test the system before installing the covers. 96 86.0087.00 Rev F Install the Covers Install the Lift Arm Covers 1. Raise the chair base all the way up. 2. Snap the lift arm cover tabs into the slots of the lower lift arm cover. Figure 148. Snap the Covers Together Lower Lift Arm Cover Lift Arm Cover 3. Set the lift arm cover assembly into the base of the chair. Snap Tabs into Slots Figure 149. Install the Lift Arm Cover Assembly Lift Arm Cover Assembly 86.0087.00 Rev F 97 A-dec 300 System Installation Guide 4. Select the appropriate upper lift arm cover depending on whether the dental chair has an assistant’s arm. NOTE The upper lift arm cover used with an assistant’s instrumentation ships in the box for the assistant’s instrumentation. 5. Raise the lift arm cover up and hold it in place. NOTE If the lift arm cover sticks when inserted into the gap between the lift arm and the link arm, use a standard screwdriver between the lift arm and link arm to widen the gap. 6. Insert the tabs of the upper lift arm into the inside slots on the back of the chair. 7. With your thumbs on the top of the upper lift arm cover, pull down and outward until the cover is snapped into place. 98 Figure 150. Upper Lift Arm Covers Upper Lift Arm Cover for Assistant’s Arm Upper Lift Arm Cover Figure 151. Install the Upper Lift Arm Cover Upper Lift Arm Cover Lift Arm Cover Inside slots on back of chair. Figure 152. Insert the Upper Lift Arm Cover Tabs 86.0087.00 Rev F Install the Covers Install the Contoured Floor Box Cover Set Figure 153. Install the Power Supply Cover Bracket Contoured Floor Box Frame 1. Use a Phillips head screwdriver to loosen the screw in the top of the power supply. 2. Position the power supply cover bracket against the power supply as shown in Figure 153. The power cord and tubing must be inside the bracket for the cover to fit properly. CAUTION Power Supply Power Supply Cover Bracket Place Bracket Between the Guides Cord and Tubing Inside of Bracket 3. Slide the slot in the bracket around the screw. Figure 154. Secure the Power Supply Cover Bracket 4. Tighten the screw to secure the bracket. Cord and Tubing Inside of Bracket 86.0087.00 Rev F 99 A-dec 300 System Installation Guide 5. Align the power supply cover with the connectors on its frame and press down. Figure 155. Covers with a Radius-Style Delivery System 6. Line up the connectors for the contoured floor box cover and press down. Notch over the Front Mount Casting Motor Pump Cover Contoured Floor Box Cover 7. Modify and install the motor pump cover according to the system’s configuration: ○ No support center or Radius-style delivery system: No modifications are needed. Slide the cover down into the grooves in the chair base plate. Power Supply Cover Figure 156. Modify the Motor Pump Cover Front End Plug Side Plug ○ With a Radius-style delivery system: Remove the plug from the front. Slide the cover over the front mount casting and down into the grooves in the chair base plate. The notch in the front goes around the front mount casting. ○ With a support center: Remove the plug from the side of the motor pump cover that is toward the chair-base mount. Slide the cover down over the chair-base mount and into the grooves in the base plate. NOTE Be sure to place any ancillaries so that the motor pump cover fits properly. 100 Figure 157. Ancillaries Bracket Location Ancillaries Mount Here Power Supply 86.0087.00 Rev F Install the Covers Install the Integrated Floor Box Cover 1. Raise the chair. 2. Modify the integrated floor box cover according to the system’s configuration: Figure 158. Install the Integrated Floor Box Cover Front End Plug Side Plug ○ Dental chair only: no modifications. ○ Dental chair with support center: remove the plug from the side of the integrated floor box cover that is toward the chair-base mount. ○ A-dec 300 system with air vacuum system: replace the plug at the front end of the integrated floor box cover with the exhaust hole cover. Post Grooves 3. Slide the cover over the utilities. Align it in the grooves of the integrated floor box cover frame and chair base plate. 4. Push the cover into the grooves and lock it over the two small posts on the frame. The cover should fit snug and secure. Make sure there are no visible cable ties, tubing, or wires. Be sure to place any ancillaries so that the integrated floor box cover fits properly. NOTE The integrated floor box cover needs to be securely replaced after removing. Verify that the cover is correctly re-attached and secured into place. WARNING 86.0087.00 Rev F Figure 159. Ancillaries Bracket Location Ancillaries Mount Here Power Supply 101 A-dec 300 System Installation Guide Install the Post Hole Caps Figure 160. Post Hole Caps If the system does not have a delivery system, dental light, or monitor, install caps to cover the post holes for those modules: • Support Center Post Holes: Use a Phillips head screwdriver and a 6-19 x 1.2" screw to attach the trim cap to the support center cover that does not have the flow diagram. • Intermediate Post: If the system has a monitor mounted on the intermediate post but does not include a dental light, push the trim cap onto the end of the intermediate post. • Assistant’s Instrumentation Hub: If the system includes an assistant’s instrumentation mounted on a telescoping arm but does not include a delivery system, push the trim cap onto the top of the hub. 102 Trim Cap for Support Center Post Holes Trim Cap for Intermediate Post and Assistant’s Instrumentation Hub 86.0087.00 Rev F Install the Covers Install the Support Center Covers Figure 161. Cover Ring Installation Options Cover Ring Slotted Side Up 1. Push the cover ring into place at the bottom of the chair-base mount until it's flush against the mount legs. If the support center is to the left of the patient, place the ring smooth side up. If the support center is to the right of the patient, place the ring slotted side up. Cover Ring Smooth Side Up TIP You may need to firmly tap the ring to get it flush against the mount legs. 2. Install the lower covers around the cover ring. The indents on the side of the covers go toward the bottom. The cover with a hole goes over the chair-base mount. If the support center is installed to the right of the patient, use diagonal cutters to widen the hole in the cover so it will fit over the mount. Push the covers' edges into the cover ring’s grooves until they snap into place. 86.0087.00 Rev F 103 A-dec 300 System Installation Guide 3. The upper support center covers attach to the support center frame. For support centers installed to the patient's left, the cover with the flow diagram goes to the outside of the support center frame, away from the chair. For support centers installed to the patient's right, this cover goes on the side toward the chair. Insert the small pins on the covers into the holes in the frame and the larger pins on the one cover into the connectors on the other cover. Figure 162. Install the Support Center Covers Trim Cap Upper Cover with Flow Diagram NOTE If the system includes a dental light or monitor mount, slide its trim ring up before installing the support center covers. The covers should fit without a gap under the trim rings and cover the top of the lower covers. ChairBase Mount Lower Cover Indents 104 86.0087.00 Rev F Appendix: Install the Air Vacuum System (AVS) APPENDIX: INSTALL THE AIR VACUUM S YSTEM ( AV S ) Tools Needed For This Section 5/64" hex key Phillips head screwdriver 1/4" combination wrench Figure 163. AVS Air-Out Tubing Connections NOTE The Air Vacuum System (AVS) is only available with A-dec 300 systems that include a integrated floor box cover or a remote floor box. A The AVS generates a vacuum using air supply at 70 PSI minimum when a central vacuum system is not available. It also separates moisture and air that flows through the system and discharges them through a drain and air outlet. B NOTE You cannot have an AVS without a cuspidor because the cuspidor is required to mount the switch that turns on the air flow through the air saliva ejector. C F Install the AVS Tank and Exhaust Pipes E B D 1. Line up the brackets with the pilot holes on the integrated floor box cover frame. 2. Use a Phillips head screwdriver to secure the brackets to each side of the frame with two 3/4" #10 screws. 3. Line up the slots in the brackets to the holes in the liquid separating tank under the air-out stub. 4. Use a 5/64" hex key and two 3/8" socket head screws to attach the tank to the brackets. Item Description Item Description A Air-Out Stub D Air Exhaust B Bracket Screwed into the Integrated Floor Box Cover Frame E Brackets Screwed into Tank C Exhaust Hole Cover F Liquid Separator Tank 5. Connect an elbow to the air-out stub. 6. Connect a splicer to the elbow. 7. Connect a second elbow. 8. Connect the 25 mm tubing to the end of the splicer and route it out of the exhaust hole. 86.0087.00 Rev F 105 A-dec 300 System Installation Guide Connect the Corresponding Tubing Figure 164. Overhead View of the AVS System 1. Connect the saliva ejector tube to the saliva ejector inlet. P 2. Connect the cuspidor drain tube to one of the ends at the top of the Y using a 20 mm adapter and a connector clip. C A D O 3. Connect a 20 mm tube to the bottom end of the Y using a 20 mm adapter and a connector clip. B C D E C D N M 4. Connect the open end of the 20 mm tube from step 3 to the drain using a 20 mm adapter and a connector clip. 5. Connect a 20 mm tube to the remaining open end at the top of the Y using a 20 mm adapter and a connector clip. F L G 6. Connect the open end of the 20 mm tube from step 5 to the liquid separator drain using a 20 mm adapter. 7. Connect the HVE/AVS exhaust tube to the HVE inlet. NOTE Connect the drain according to local plumbing codes. 8. Use a 1/4" combination wrench to replace the plug on the air filter/regulator with a 5/16" barb; then connect the 5/16" yellow tube that leads to the air vacuum generator. 9. Replace the plug in the front of the integrated floor box cover with the exhaust hole cover; then install the cover (for instructions, see “Install the Integrated Floor Box Cover” on page 101). 106 K D C J H Item Description Item Description A Cuspidor Drain Tube (5/8") J Air Exhaust Outlet B Saliva Ejector Tube (3/8") K Drain Outlet C Connector Clip (four) L HVE/AVS Inlet D 20 mm Adapter (four) M Clamp E Y Connector N 20 mm Tube Drain to Liquid Separator F Saliva Ejector Inlet O HVE/AVS Exhaust (5/8") G 20 mm Tube to Drain Outlet P H Exhaust Hole Cover Liquid Separator Drain Figure 165. Replace Air/Filter Regulator Plug with a Barb Replace either of these plugs with a barb. 86.0087.00 Rev F Appendix: Install the Air Vacuum System (AVS) Install the AVS in a Remote Floor Box 1. Insert the air-out stub through the 1.25" hole and use a 5/64" hex key and two 3/8" socket head screws to attach the tank to the floor box. 2. Attach the 90 degree elbow to the end of the air-out valve. 3. Complete the HVE, saliva ejector, and drain connections (see “Connect the Corresponding Tubing” on page 106). Figure 166. Install the AVS in a Remote Floor Box 1.25" Hole 90 Degree Elbow Two 3/16" Pilot Holes Air-Out Stub NOTE Connect the drain according to local plumbing codes. 86.0087.00 Rev F 107 A-dec 300 System Installation Guide REGULATORY INFORMATION Regulatory information is provided with A-dec equipment as mandated by agency requirements. This information is delivered in the equipment’s Instructions for Use or the separate Regulatory Information and Specifications document. If you need this information, please go to the Document Library at www.a-dec.com. A-dec® Headquarters 2601 Crestview Drive Newberg, OR 97132 USA Tel: 1.800.547.1883 Within USA/Canada Tel: 1.503.538.7478 Outside USA/Canada Fax: 1.503.538.0276 www.a-dec.com / www.a-dec.biz A-dec Inc. makes no warranty of any kind with regard to the content in this document including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. ÍvÈ.ÇÂwÈ.00ÆÎ 86.0087.00 Rev F Copyright 2012 A-dec Inc. All rights reserved.
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
advertisement