Hanna Instruments GroPro HI9814 pH/EC/TDS/Temperature Meter Instruction manual

Hanna Instruments GroPro HI9814 pH/EC/TDS/Temperature Meter Instruction manual

Below you will find brief information for pH/EC/TDS/Temperature Meter HI9814. The HI9814 is designed to offer a combination of pH, conductivity, total dissolved solids, and temperature measurements. This instrument is easy to use with automatic temperature compensation and calibration. The rugged waterproof housing is rated for IP67 conditions. This instrument is ideal for use in agriculture, hydroponics, and other applications where pH, EC, TDS, and temperature measurements are required.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

HI9814 Instruction Manual | Manualzz
TM
HI9814
pH / EC / TDS / Temperature
Meter with Quick Calibration
LCD Description
Thank You
Specifications
Range*
0.00 to 14.00 pH
Resolution
0.01 pH
Accuracy
±0.01 pH
Calibration
For technical support, contact your local Hanna
office or email us at [email protected]
automatic, one or two-point
calibration (using pH 4.01,
7.01, 10.01 buffers); one-point
calibration using quick calibration
solution
Temperature
Compensation
automatic
Find your local Hanna office on www.hannainst.com
Range
0.00 to 6.00 mS/cm
Resolution
0.01 mS/cm
Accuracy
±2% F.S.
Calibration
automatic, one-point at
1.41 mS/cm or 5.00 mS/cm;
one-point calibration using
quick calibration solution
Temperature
Compensation
automatic, with β = 1.9%/°C
Range
0 to 3000 ppm (500 CF);
0 to 3999 ppm (700 CF)
Resolution
10 ppm (mg/L)
Accuracy
±2% F.S.
Conversion
Factor (CF)**
0.5 (500 ppm) or 0.7 (700 ppm)
Range*
0.0 to 60.0°C/32.0 to 140.0°F
Resolution
0.1°C/0.1°F
Accuracy
±0.5°C/±1°F
Typical EMC
Deviation
±0.02 pH; ±0.2°C or ±0.4°F
Probe
(included)
HI1285-7 pH/EC/TDS/
temperature with DIN connector
and 1 m (3.3’) cable
Battery Type/
Life
1.5V AAA (3) /approximately 500
hours of continuous use
Auto-Off
after 8 minutes, 60 minutes, or
disabled
Environment
0 to 50°C (32 to 122°F);
RH max. 100%
Dimensions
152 x 58 x 30 mm (6.0 x 2.3 x 1.2”)
Weight
205 g (7.2 oz)
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before
using the instrument.
For more information about Hanna and our
products, visit www.hannainst.com or e-mail
us at [email protected].
pH
EC
Preliminary Examination
Remove the instrument from the packing material
and examine it carefully to make sure that no damage
has occurred during shipment. If noticeable damage
is evident, contact your local Hanna office.
1. Automatic Temperature Compensation indicator
Note: Save all packing material until you are
sure that the instrument functions correctly. All
defective items must be returned in the original
packing together with the supplied accessories.
2. Battery percentage (visible at power up)
3. Secondary display
4.Low battery indicator
6.Calibration indicator
7. Selectable temperature unit
TDS
8.Primary display
9.Measuring unit for primary display
5. Stability indicator
Temperature
General Description
HI9814 is designed to offer you the combination of pH,
conductivity, total dissolved solids, and temperature
measurements. All operations and settings, including
calibration buffers and temperature scale selections,
are made through only two buttons. The housing
is waterproof and rated for IP67 conditions.
Maintenance:
The supplied HI1285-7 multiparameter probe measures
pH, EC/TDS, and temperature in one convenient, rugged
probe. A solid-state preamplifier is integrated into the
probe to protect the pH measurement from transient
electrical noise. Sources of electrical noise include
ballasts used in lighting and pumps to circulate water
and nutrient solutions. Other user-selectable features
include selectable TDS factors of 0.5 and 0.7 as well as
auto-off after 8 minutes, 60 minutes, or disabled.
• After use, rinse the electrode with water and replace the
protective cap filled with a few drops of HI70300 storage solution. If HI70300 is not available, use pH buffer.
• While unpacking, the appearance of salt deposits
around the protective cap is normal. The salt deposits
will dissolve when rinsed with water.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR
STORAGE PURPOSES.
Warranty
Each meter is supplied with:
• HI1285-7 pH/EC/TDS probe with built-in temperature
sensor, DIN connector and 1 m (3.3’) cable
• HI9814P Quick Calibration solution sachets (3)
• HI700661P Electrode cleaning solution sachets
for agriculture (3)
• 1.5V AAA batteries (3)
• Instruction manual
Hanna Instruments reserves the right to modify the design,
construction, or appearance of its products without advance notice.
Additional
Specifications
* The sensor is rated 0 to 12 pH and -5 to 30°C
**1000 µS/cm = 500 ppm with 0.5 CF
ISTHI9814GROPRO 1.0 07/15
INSTRUCTION MANUAL
The meter is warranted for a period of two years against defects in
workmanship and materials when used for their intended purpose
and maintained according to instructions. The electrode is warranted
for a period of six months. This warranty is limited to repair or
replacement free of charge. Damage due to accidents, misuse,
tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna office. If under warranty, report
the model number, date of purchase, serial number and the nature of
the problem. If the repair is not covered by the warranty, you will be
notified of the charges incurred. If the instrument is to be returned
to Hanna Instruments, first obtain a Returned Goods Authorization
(RGA) number from the Technical Service department and then
send it with shipping costs prepaid. When shipping any instrument,
make sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the
written consent of the copyright owner.
Accessories
Code
Description
HI1285-7
pH/conductivity probe with built-in temperature
sensor, DIN connector and 1 m (3.3’) cable
HI9814P
quick calibration solution, 20 mL sachets (25)
HI70004P
pH 4.01 buffer solution, 20 mL sachets (25)
HI70007P
pH 7.01 buffer solution, 20 mL sachets (25)
HI70010P
pH 10.01 buffer solution, 20 mL sachets (25)
HI70031P
1413 µS/cm (1.41 mS/cm) solution, 20 mL sachets (25)
HI70039P
5000 µS/cm (5.00 mS/cm) solution, 20 mL sachets (25)
HI70300M
electrode storage solution, 230 mL bottle
HI700661P
general purpose cleaning solution for agriculture,
20 mL sachets (25)
HI710025
shockproof boot (green)
Operational Guide
Meter Setup
Before using the instrument for the first time,
open the battery compartment and insert batteries,
observing the polarity.
While in measurement mode, press and hold the ON/OFF/
MODE button until “UNIT“ appears on the secondary
display. Pressing the ON/OFF/MODE button will now cycle
through the various units and features below which can
then be modified with the SET/HOLD button.
To connect the probe
With the meter turned off, connect the HI1285-7 probe to
the DIN socket on the bottom of the meter by aligning the
pins and pushing in the plug. Tighten the nut to ensure a
good connection. Remove the protective cap from the probe
before taking any measurements.
To turn the meter ON and check
the battery status
Press the ON/OFF/MODE button to turn the meter on. At
start-up, all the LCD segments are displayed for 1 second,
then the percent indication of the remaining battery life is
displayed for another second. The meter then enters the
normal measuring mode.
To select EC or TDS
To select TDS measurement with 0.5 TDS factor press
SET/HOLD until “500 UNIT” is displayed. To select TDS
measurement with 0.7 TDS factor press SET/HOLD until
“700 UNIT” is displayed. By pressing SET/HOLD again
“EC UNIT” unit is selected.
To select calibration type
Press the ON/OFF/MODE until “CAL“ appears on primary
display. Press the SET/HOLD button to choose from “CAL
STD“ (standard calibration) or “CAL QUIK“ (one-point quick
calibration).
To select the temperature unit (°C/°F)
• To reset the default values and clear a previous calibration,
press the SET/HOLD button after entering the calibration mode, before the first point is accepted. The LCD displays
“----CLR“ for 1 second, the meter resets to the default calibration and the “Calibrated” tag on the LCD disappears.
pH Measurement and Calibration†
• Make sure the meter has been calibrated before use.
• If the probe is dry, soak it in HI70300 storage solution
for 30 minutes to reactivate it.
• Submerge the probe in the sample to be tested while
stirring it gently. Wait until the “Not Stable“ tag on the LCD disappears.
• The LCD displays the pH value (automatically compensated for temperature) on the primary LCD,
while the secondary LCD displays the sample temperature.
•
If measurements are taken in different samples
successively, rinse the probe tip thoroughly† to eliminate cross-contamination. After cleaning, rinse
the probe tip with some deionized water and some of
the sample to be measured.
Note: Keeping the ON button pressed while turning the
meter on will display all LCD segments as long as the button
is pressed.
Press ON/OFF/MODE button until “TEMP” and the selected
temperature unit “°C“ or “°F“ is displayed. Press SET/HOLD
button to select temperature unit.
pH calibration
To select the measurement range
To select the Auto-Off time
• Enter calibration mode while in pH measurement mode.
Press ON/OFF/MODE button until “AOFF“ appears on
the secondary display and selected time “8,“ “60“ or “----“
(disabled) is displayed on the primary display. Press SET/
HOLD button to select Auto-Off interval.
• The meter will enter the calibration mode, displaying
“pH 7.01 USE.”
While in normal measurement mode, press the SET/HOLD
button quickly to toggle between pH and EC or TDS
reading on the primary LCD, while temperature will be
simultaneously displayed on the secondary LCD.
To freeze the display
While in measurement mode, press and hold the SET/HOLD
button until “HOLD“ appears on the secondary display and
the reading will be frozen on the LCD. Press any button to
return to normal mode.
To turn the meter OFF
While in normal measurement mode, press the ON/OFF/
MODE button. “OFF“ will appear on the secondary display.
Note: When the meter detects the absence of probe at its
input, the message “Probe not connected“ appears on LCD
and “----“ blinks on LCD lines. When a probe is connected,
the “Probe not connected“ tag is turned off, and the
readings are displayed on the LCD.
To enter calibration mode
Press and hold down the ON/OFF/MODE button until “OFF“
is replaced by “CAL.“ Release the button.
To enter setup mode
Press and hold ON/OFF/MODE button until “CAL“
is replaced by “UNIT“ on the secondary display.
Release the button.
To return to measurement mode
Press ON/OFF/MODE button.
Quick Calibration†
Select calibration type “CAL QUIK“ from meter setup.
• Enter calibration mode.
• Immerse the probe in the HI9814-0 calibration solution.
• When the standard value is recognized and stability is
reached the meter automatically accepts the calibration.
• The LCD will display “OK” for 1 second and return to normal measurement mode.
•
If the standard is not recognized or the slope is out
of accepted range “---- WRNG“ is displayed. Change the
calibration solution, clean the electrode or press any key
to exit calibration.
• When the calibration procedure is completed, the
“Calibrated“ tag is turned on.
To exit calibration and reset default values
•
After entering the calibration mode and before the point is
accepted, it is possible to quit the procedure and return
to the last calibration data by pressing the ON/OFF/MODE
button. The LCD displays “---- ESC“ for 1 second and the
meter returns to measurement mode.
Select calibration type “CAL STD“ from meter setup.
• Place the sensor into the first calibration buffer. If performing a two-point calibration, use pH 7.01 buffer first.
Follow directions for single and two-point calibration below:
Single-point calibration
1. Place the probe in any buffer from the selected buffer set.
The meter will automatically recognize the buffer value.
2. If the buffer is not recognized or the calibration offset
is out of the accepted range “---- WRNG“ is displayed.
3. If the buffer is recognized “REC“ is displayed until the
reading is stable and the calibration is accepted.
• If using pH 7.01, after acceptance of the buffer press any
key to exit. “OK1“ message is displayed and meter returns
to pH measurement mode.
• If using 4.01 or 10.01 buffer the “OK1“ message is
displayed and meter returns to pH measurement mode.
Two-point calibration
Proceed with steps 1 through 3 under single-point calibration
using 7.01 pH buffer first. Then follow steps below:
• The “pH 4.01 USE“ message is then displayed.
•
Place the probe in the second calibration buffer
(pH 4.01 or 10.01). When the second buffer is accepted,
the LCD will display “OK2” for 1 second and the meter will return to the normal measurement mode.
•
If the buffer is not recognized or the slope is out of
accepted range “---- WRNG“ is displayed. Change
the buffer, clean the electrode or press any key to
exit calibration.
Note: When the calibration procedure is completed,
the “Calibrated“ tag is turned on.
It is always recommended to carry out a two-point
calibration for better accuracy.
EC Measurement and Calibration†
• Place the probe in the sample to be tested. Use plastic
beakers or containers to minimize any electromagnetic
interference.
• Tap the probe lightly on the bottom of the container to remove air bubbles that may be trapped inside the tip.
• Wait for a few minutes for the temperature sensor
to reach thermal equilibrium, when the “Not Stable“
tag disappears.
• The LCD displays the EC or TDS value (automatically
compensated for temperature) on the primary LCD,
while the secondary LCD displays the sample temperature.
EC calibration
Select calibration type “CAL STD“ from meter setup.
• Enter calibration mode while in EC measurement mode.
• The meter enters the calibration mode and “USE”
is displayed. Immerse the probe in 1.41 mS/cm or
5.00 mS/cm calibration solution.
• If the standard value is recognized ”REC” is displayed
until the reading is stable and calibration is accepted.
• The LCD will display “OK” for 1 second and return to
normal measurement mode.
•
If the standard is not recognized or the slope is out
of accepted range “---- WRNG“ is display. Change the
buffer, clean the electrode or press any key
to exit calibration.
• When the calibration procedure is completed, the
“Calibrated“ tag is turned on.
Battery Replacement
The meter is supplied with batteries.
The meter displays the remaining battery percentage
when turned on. When the level is below 5%, the
symbol on the LCD blinks to indicate a low battery condition.
If the battery level is low enough to cause erroneous
readings, the Battery Error Prevention System (BEPS) turns
the meter off. It is recommended to replace the batteries as
soon as the display flashes the battery symbol. To replace
the batteries:
• Open the battery compartment cap (on the bottom of
the instrument).
• Remove old batteries.
• Replace new batteries, observing the polarity on the
rear of the instrument.
• Close the battery compartment cap.
† The probe tip should be rinsed with purified water (reverse
osmosis, distilled, or deionized) before and after placing in
any solution (buffer, storage or sample).

advertisement

Key Features

  • pH, EC, TDS, and temperature measurement
  • Automatic temperature compensation
  • One-point quick calibration
  • Rugged waterproof housing (IP67)
  • Multiparameter probe (HI1285-7)
  • Selectable TDS factors (0.5 and 0.7)
  • Auto-off function
  • Battery percentage indicator

Frequently Answers and Questions

What is the purpose of the HI9814P Quick Calibration solution sachets (3) included with the meter?
The HI9814P Quick Calibration solution sachets are used for one-point quick calibration of the meter. They provide a convenient and fast way to calibrate the meter without using multiple buffer solutions.
What is the HI700661P Electrode cleaning solution sachets (3) for agriculture used for?
The HI700661P Electrode cleaning solution sachets are specifically designed for cleaning the probe after use in agricultural applications. They help to remove any residues that might be present on the probe, ensuring accurate measurements.
What is the function of the Auto-Off feature?
The Auto-Off feature automatically turns off the meter after a set period of inactivity to conserve battery life. You can choose between 8 minutes, 60 minutes, or disable the Auto-Off feature completely.

Related manuals

Download PDF

advertisement