Scientific Industries SI-0400, SI-0401, SI-0402, SI-0403, SI-0404, SI-0405, SI-0406 Microplate Mixer Operating Instructions

Scientific Industries SI-0400, SI-0401, SI-0402, SI-0403, SI-0404, SI-0405, SI-0406 Microplate Mixer Operating Instructions

The MicroPlate Genie® SI-0400, SI-0401, SI-0402, SI-0403, SI-0404, SI-0405, SI-0406 is a laboratory mixer designed for consistent mixing within all microplate wells. It's designed for use with almost any microplate, regardless of type or material. The mixer features both timed and continuous modes for unattended operation, allowing you to set the desired run time (1 to 60 minutes) or let it run indefinitely. The unit also includes a speed control switch, allowing you to adjust the mixing speed to your specific needs.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Microplate Genie SI-0400-SI-0406 Operating Instructions | Manualzz

The MicroPlate Genie ® has been optimized for use with almost any microplate, regardless of type or material of construction. The orbit of motion and the speed have been changed to achieve consistent mixing within all microplate wells. VORTEX-GENIE mixers are the "Gold Standard" for laboratory mixers and the quality and reliability that you have come to know so well are characteristics that are by design, an integral part of all GENIE ® products.

OPERATING INSTRUCTIONS

We recommend you retain the original packaging for 90 days in case you need to return the product for any reason to your distributor or Scientific Industries.

1.0 - Plug the line cord into a properly grounded electrical outlet.

(120 VAC for the SI-0400 model;

230 VAC for the SI-0401 through

SI-0404 and SI-0406 models; 100

VAC for the SI-0405 model)

2.0 - Place a microplate into the

MicroPlate Genie. Tip the microplate slightly down to the left and back, and bank the corner into the two rubber bumpers. It is not necessary to tilt the plate excessively. Push the plate the rest of the way down onto the top of the MicroPlate Genie and release. The right side of the microplate will be held in place by the beveled edge on the right side.

3.0 - The MicroPlate Genie mixer has the following modes of operation:

3.1 - TIMED MODE- (1 to 60 minutes)

The Timed Mode allows for unattended operation that is timed from 1 to

60 minutes. To initiate timed operating mode, with the microplate in place, set the 1 to 60 minute timer (gray knob and gray scale) to the desired run time. The MicroPlate Genie is OFF until the push-button is depressed and then released. This will start the mixing action of the

MicroPlate Genie. The mixer will run until the set time has elapsed. If for any reason you wish to stop the mixing action before the set time has elapsed, simply depress the push-button for a second time and release. The mixer will stop. Depressing the push-button a third time will initiate a completely new timed cycle.

3.2 - CONTINUOUS MODE

The continuous mode allows for unattended operation that will continue indefinitely until you stop the run by depressing the push-button switch.

To initiate Continuous Mode, turn the gray timer knob fully counterclockwise until you feel the "click" and the knob is in the detent "TIMER

OFF" (continuous mode) position. Next, depress the push-button switch and release. The MicroPlate Genie

4.0 - With prolonged use, the MicroPlate Genie mixer may become warm to the touch. It should not get HOT to the touch. The motor will radiate a certain amount of heat that will be dissipated by the fan in the bottom of the unit.

® mixer will start and continue to run, unattended, until you once again depress the push-button switch and release. If you wish to restart the mixer once again, simply depress the push-button switch for a moment and release.

Note: The speed control switch should be set between positions 3 and 10 to enable the mixing action to start. After the action has started, you may, if desired, reduce the speed below position 3 for the duration of the run. Before re-starting the mixer, re-set the speed control to a setting higher than position 3.

Heat will not be transferred to your microplates. The fan (located in the bottom of the unit) directs air downward and out of the unit. As a result, room air is constantly drawn into the unit underneath the tray that holds the microplates, insuring that samples will not be heated.

advertisement

Key Features

  • Microplate mixing
  • Timed and Continuous Mode
  • Speed Control
  • Unattended Operation
  • Compatible with various microplates
  • Consistent Mixing
  • Easy to use

Frequently Answers and Questions

How do I operate the MicroPlate Genie in timed mode?
To initiate timed operating mode, set the 1 to 60 minute timer to the desired run time. Press the push-button switch to start the mixing action. The mixer will run until the set time has elapsed. You can stop the mixing action before the set time by pressing the push-button for a second time.
How do I operate the MicroPlate Genie in continuous mode?
To initiate Continuous Mode, turn the gray timer knob fully counterclockwise until you feel the “click” and the knob is in the detent “TIMER OFF” position. Press the push-button switch to start the mixer. It will run continuously until you press the push-button again to stop it.
What is the operating voltage of the different models?
The SI-0400 model operates on 120 VAC. Models SI-0401 through SI-0404 and SI-0406 operate on 230 VAC. The SI-0405 model operates on 100 VAC.
Will the MicroPlate Genie heat up my samples?
The MicroPlate Genie mixer has a fan that directs air downwards and out of the unit. This ensures that room air is constantly drawn into the unit, preventing your samples from getting heated
Where can I find replacement parts for the MicroPlate Genie®?
Replacement parts are available through most laboratory equipment distributors or directly from Scientific Industries.

Related manuals

Download PDF

advertisement