ACE Easy is a digital semiconductor alcohol testing device, which is easy to handle and shows reliable test results. This way you receive results that are as accurate as possible. It provides three different ways to perform a breathalyzer test: by means of using a mouthpiece (best accuracy), by means of using a straw (good accuracy), or (passive mode) just blow your breath upon the sensor opening (sufficient accuracy). To maintain the accuracy of the breathalyzer, ACE recommends a calibration after 1,000 tests or as soon as you see "CAL" on the display.
advertisement
ACE Easy
Breathalyzer
Operating Manual
www.ace-instruments.net
ACE Easy Breathalyzer
Operating Manual
Thank you for purchasing our ACE Easy Breathalyzer. It measures alcohol concentration in human breath using a semiconductor sensor and calculates the blood alcohol level. The device must only be used pursuant to the instructions mentioned in this manual.
In case of any questions you may contact the distributor (ACE) by email to [email protected], www.ace-instruments.net.
Short description:
The ACE Easy Breathalyzer is a digital semiconductor alcohol testing device, which can be operated by only one button and shows reliable test results in % BAC (default setting).
The legal limit in the USA is
0.08% BAC (status: Oct 2016).
You can interprete the test result 1:1 (e.g. the display shows 0.03 = 0.03% BAC).
Scope of delivery:
Digital ACE Easy Breathalyzer
(3 digits, 2 decimals) with semiconductor sensor
5 mouthpieces (Figurative image of mouthpiece on the pictures in this manual)
2 pieces AA batteries (1.5V, LR6,
Alkaline), already inside the device
Quick Start Guide
Operating Manual
To maintain the measurement accuracy of this breathalyzer it should be calibrated regularely. For a calibration service you can send the device to ACE Instruments
(see www.ace-instruments.net).
For each breath test a new mouthpiece is required (to get a most accurate test result).
Label of individual components:
Mouthpiece
Sensor opening
Operation button
Digital LCD display
(3 digits, 2 decimals)
Air outlet
(always keep this part unblocked)
Battery compartment cover
This way you receive results that are as accurate as possible: a) Wait at least 15 minutes after your last sip of alcohol. Do not eat, drink or smoke during that time.
b) Do not open the casing.
c) Do not keep the breathalyzer in cold, hot, moist or badly ventilated places. d) Keep the breathalyzer out of the reach of children.
e) Conduct three inlet tests after long storage before conducting the actual alcohol test.
f) Do not drop the breathalyzer.
g) Change the battery if it is depleted. Diminished battery capacity leads to incorrect test results.
h) Do not conduct tests in loud
(sound waves disrupt the sensor), drafty or poorly ventilated places.
i) After a test result of 0.100% BAC
(% BAC = basic setting) or higher the sensor needs rest time.
In this case wait at least three minutes before the next test.
j) Do not conduct tests in very cold, very hot or an unventilated, smoky ambience.
k) Store the breathalyzer in a way to avoid penetration by dust or liquids.
l) Blowing too hard may cause damage to the sensor as saliva flows into it.
Preparation for the test:
1. If you use the breathalyzer for the first time, then push the battery compartment cover on the back downwards. Remove the small paper inside the battery compartment (this paper prevents the 2 inserted batteries from discharging). Close the battery compartment cover again. If you change the batteries please mind their polarity (+/- pole).
2. a) Put a mouthpiece or a straw on the mouthpiece connection, which is on the side of the device. Make sure it is attached properly. Use a different mouthpiece or straw for each test person. Avoid blowing cigarette smoke into the tester! b) If you want to test yourself without a mouthpiece or straw push the power button and as soon as you see “0” on the display, simply blow directly towards the mouthpiece connection (blow contantently like you would whistle).
ACE INSTRUMENTS
RECOMMENDS:
ALWAYS USE A NEW MOUTH-
PIECE FOR EACH PERSON AND
TEST THAT IS CONDUCTED. ACE
INSTRUMENTS ASSUMES NO
LIABILITY FOR ANY DAMAGES
WHICH ARISE WHEN SEVERAL
PEOPLE USE THE SAME
MOUTHPIECE FOR MULTIPLE
TESTS.
ONLY USE THE PROVIDED
SINGLE PACKED AND
HYGIENIC MOUTHPIECES -
MORE ARE AVAILABLE AT
WWW.ACE-INSTRUMENTS.NET.
DURING EACH TEST THE
BREATHALYZER´S INTERNAL
AND EXTERNAL PARTS COME
INTO CONTACT WITH HUMAN
SECRETION (SALIVA, BLOOD,
SWEAT, ETC.)!
USE SAFTEY GLOVES WHEN
PULLING OUT A MOUTHPIECE
USED BY ANOTHER PERSON!
YOU CAN BUY APPROPRIATE
GLOVES E.G. AT ACE TECHNIK
(WWW.ACE-TECHNIK.DE).
Conducting the measurement:
1. Press the power button to turn on the device * .
2. The device warms up, a countdown starts, wait until you can see “0” on the display, then you can start the test which can be done in three different ways: a) picture 2. a) - by means of using a mouthpiece (best accuracy) b) picture 2. b) - by means of using a straw (good accuracy) c) picture 2. c) - (passive mode) just blow your breath upon the sensor opening (sufficient accuracy)
3. Blow steadily for 4 seconds until a double beep tone concludes.
4. If the breath sample is insufficient the ACE Easy shows “Flo”.
Turn off the device and repeat the test. Please blow steadily and without a break into the mouthpiece (like you would whistle).
5. After a sufficient breath sample you will see the digit “0” moving left to right in the display as the device analyzes the breath sample. After a few seconds the display shows the alcohol level in the format #.##% BAC (% BAC
= default setting). In the USA the legal limit for drivers over
21 years is 0.08% BAC, below 21 years it is 0.00% BAC (status:
Oct. 2016).
6. When showing the test result the display will blink ca. 10 times. Then you hear a beep tone with each further blink and after five blinks the device turns off.
7. For turning off manually keep the power button pressed. The display shows “Out” (off). After a few seconds the ACE Easy turns off.
* If you constantely keep pushing the power button, you can see the number of all tests conducted with the device (starting with zero after a calibration) and also the last measurement result.
IMPORTANT NOTICE: WHEN
YOU HAVE EATEN, DRUNKEN
ALCOHOL OR SMOKED A
CIGARETTE, WAIT AT LEAST 15
MINUTES BEFORE STARTING
THE MEASUREMENT. ANY TEST
CONDUCTED PRIOR TO THE EXPI-
RATION OF THE RECOMMENDED
WAITING PERIOD WILL RESULT
IN INACCURATE RESULTS.IF THE
TEST RESULT IS BELOW 0.005%
BAC THEN THE DEVICE SHOWS
0.000% BAC ON THE DISPLAY.
THIS HAPPENS BECAUSE AT
A LOW RANGE OF ALCOHOL
CONCENTRATION IT IS POSSIBLE
TO MEASURE OTHER MATERIALS
IN YOUR MOUTH.
Only the usage of original
ACE mouthpieces ...
Guarantees test result accuracy
Guarantees 100% hygenity
Guarantees full warranty of this device
ACE INSTRUMENTS ADVISES:
EVEN IF YOU GET A RESULT
MEANING „SAFE RANGE“, YOU
SHOULD NEVER DRIVE AFTER
CONSUMING ALCOHOL OR DO
ANY OTHER ACTIVITIES THAT
ARE SEVERELY IMPAIRED BY
ALCOHOL CONSUMPTION AND
MIGHT ENDANGER YOUR HEALTH
AND SAFETY OR THE HEALTH
AND SAFETY OF OTHERS.
Display indications:
FLO
BAT
CAL
SEN
OUT
The breath sample was interrupted or not strong enough for a measurement. After the sensor warmed up itself automatically, please blow steadily and without a break into the mouthpiece (like you would whistle).
The battery power is low. Please replace the 2 batteries inside the device with 2 new ones (type AA, 1.5V, LR6, Alkaline batteries). If you change depleted batteries against new ones please mind their polarity (+/- pole). Do not use rechargeable batteries!
If you test over 1.000 tests, ACE Easy will display “CAL” sign for
3 seconds when warming up for test. This “CAL” sign informs about the appropriate time for calibration. To maintain the measurement accuracy of this breathalyzer it should be calibrated.
For a calibration service you can send the device to ACE Instruments
(see www.ace-instruments.net).
The sensor is defective, please contact ACE: [email protected], www.ace-instruments.net
Notice: The exchange of the sensor is excluded from warranty and liable to costs.
The device turns off (either by pressing the power button constantely for 4 seconds or 10 seconds after displaying the test result).
2. a) 2. b) 2. b) 2. c)
Calibration:
To maintain the accuracy of the breathalyzer, ACE recommends a calibration after 1.000 tests or as soon as you see “CAL” on the display. For a calibration service you can send the device to ACE
Instruments
(see www.ace-instruments.net).
The calibration is not mandatory, but the accuracy of the breathalyzer will decrease steadily without it.
We recommend sending it as a parcel, so that in cases of loss an inquiry or proof of delivery is possible.
depleted batteries to electrical appliance stores, who collect them for free.
Battery exchange:
To replace depleted batteries against new ones, first push the battery cover down. Then replace the depleted batteries with two new ones. Be aware of the polarity when inserting the new batteries
(+/- pole). Put the battery compartment cover on the rear panel and push it gently down.
All depleted batteries will be recycled. This way valuable resources like iron, zinc and nickel can be retrieved and are reusable then. Battery recycling is the easiest lesson in environmental protection - so please support it!
Due to changes in the product, the procedures stated can change. This can take place without prior notice or adaptation of the operating manual and is no cause for complaint, return or price reduction. The user is responsible for recognizing discrepancies between the operating manual and the actual handling of the breathalyzer and must react accordingly.
Battery Law notice (BattG):
The device associated with this operating manual contains two
AA Alkaline batteries (1.5V) in delivery condition. These batteries are not made for recharging. The use of rechargeable batteries is not possible with this device and can damage it. If the included batteries are depleted, they must not be thrown into the household waste. Used batteries may contain harmful substances that can cause environmental or health damage. Please return all
The shown garbage bins mean that batteries must not be thrown into the household waste.
The signs below the garbage bins are acronyms for the chemical substances inside the battery:
Hg: Battery contains Mercury
Cd: Battery contains Cadmium
Pb: Battery contains Lead
Technical data:
Statement of the BAC
Accuracy
Warm-up time
Sensor
Response time
Energy supply
Dimensions of the device
Weight
Ambient conditions
Calibration
0.00 – 0.40% BAC (the default setting of your device is % BAC)
+/- 0.01% BAC (Blood Alcohol Concentration) with reference to 0.05% BAC max. 30 seconds
Semiconductor sensor
Within 10 seconds
2 pieces AA batteries (1.5V, LR6, Alkaline)
46 x 110 x 20 mm
89 g (including batteries)
Operation: 0 – 40 °C
Storage: 0 – 50 °C
After 1.000 tests or as soon as you see “CAL” on the display. For a calibration service you can send the device to ACE Instruments
(see www.ace-instruments.net).
WARRANTY AND LIABILITY
WARRANTY
ACE Instruments (ACE) provides a warranty of two (2) years after purchase for material and processing errors for the product
(see attached warranty and liability declaration) in all versions and designations; and a warranty of one (1) year for the sensor. This warranty includes the electronics as well as the casing or sensor without sampling components.
Components that are inevitably destroyed upon effectiveness
(such as fuses) are not part of this warranty. The warranty is valid from the date of delivery or from the date of first activation (always the shorter period), if not stated otherwise in the accompanying
ACE Handels- und Entwicklungs
GmbH product literature.
Misuse:
OTHER APPLICATIONS OF,
OR MODIFICATIONS TO, THIS
BREATHALYZER CONSTITUTE
A MISUSE AND ENTAIL A RISK
OF PERSONAL INJURY AND/OR
DAMAGE TO THE DEVICE. ACE
SHALL ASSUME NO LIABILITY
FOR INJURY/DAMAGE CAUSED
BY MISUSE OF THE DEVICE. THE
DEVICE IS NOT INTENDED FOR
COMMERCIAL USE.
ACE SHALL NOT BE HELD LIABLE
FOR ANY DAMAGES THAT OCCUR
AS A RESULT OF ANY UNAUTHOR-
IZED MODIFICATIONS, REENGI-
NEERINGS OR AMENDMENTS
MADE TO THIS BREATHALYZER
OR ITS IMPROPER USE.
LIMITATION OF WARRANTY AND
LIABILITY
1. Product description:
ACE REJECTS ANY FURTHER
EXPRESSLY STATED OR IMPLIED
WARRANTY, INCLUDING THE
MARKETABILITY AND SUITABILITY
FOR A DETERMINED PURPOSE.
THIS WARRANTY ESPECIALLY
EXCLUDES CALIBRATION.
AFTER CHANGING THE SENSOR,
CALIBRATION IS ABSOLUTELY
NECESSARY.
IF AN EXCHANGE OF THE
SENSOR OR A CALIBRATION IS
REQUIRED FOR A PRODUCT,
THE USER MAY RECEIVE A COST
ESTIMATE. SHOULD THE USER
REJECT THE ESTIMATE, ACE WILL
SEND THE PRODUCT BACK UNRE-
PAIRED / UNCALIBRATED AND THE
PRODUCT WILL BE LABELLED
ACCORDINGLY. SHOULD THE
PRODUCT NOT FULFIL THE WAR-
RANTY, THE CLAIM OF THE BUY-
ER AND THE OBLIGATIONS OF
ACE ARE LIMITED EXCLUSIVELY
AND AT THE DISCRETION OF ACE
TO AN EXCHANGE OR REPAIR
OF THE AFFECTED GOODS OR
TO A REIMBURSEMENT OF THE
PURCHASING PRICE FOR THE
FAULTY GOODS.
2. Application restriction:
ACE IS ON NO ACCOUNT LIABLE
FOR ADDITIONAL DAMAGES,
INCIDENTAL DAMAGES OR CON-
SEQUENTIAL DAMAGES SUCH
AS, AMONG OTHERS BUT NOT
LIMITED TO, LOST PROFIT AND
DOWNTIME THAT RESULT FROM
SELLING, PRODUCING OR USING
PRODUCTS, WHICH ARE PART OF
THIS APPLICATION RESTRICTION.
THIS IS INDEPENDENT OF ANY
CLAIMS, SUCH AS FROM CON-
TRACTS, NEGLIGENCE OR OTHER
CULPABLE ACTIONS.
THE PRODUCT, WHICH IS AC-
COMPANIED BY THIS WARRANTY
AND LIABILITY DECLARATION, NO
MATTER IN WHICH VERSION OR
DESIGNATION, CAN ESPECIALLY
NOT BE USED LEGALLY BINDING
TO DETERMINE CAPABILITY TO
OPERATE VEHICLES OR TO EXE-
CUTE ACTION THAT COULD LEAD
TO DAMAGE TO ONESELF OR
OTHERS UNDER THE INFLUENCE
OF ALCOHOL.
THE THOROUGH CHECK FOR
DAMAGE OF ALL PRODUCTS
UPON RECEIPT, A CALIBRATION
SUITABLE FOR THE GIVEN
USE BY THE BUYER AND THE
PROPER USE, MAINTENANCE
AND SERVICE ACCORDING TO
THE INSTRUCTIONS OF THE
ACCOMPANYING PRODUCT
LITERATURE BY ACE ARE MANDA-
TORY PREREQUISITES FOR THIS
WARRANTY. MAINTENANCE AND
SERVICE WORK BY NON-QUAL-
IFIED PERSONS, AS WELL AS
USING NON-APPROVED CON-
SUMABLES OR SPARE PARTS ARE
A MISUSE, AND ALL WARRANTY
RIGHT SHALL SEIZE IMMEDIATELY.
AS FOR ALL TECHNICALLY
CHALLENGING PRODUCTS, IT IS
A FUNDAMENTAL PREREQUISITE
AND MANDATORY FOR THE WAR-
RANTY BY ACE THAT ALL USERS
OF THIS PRODUCT ARE FAMILIAR
WITH THE PROPERTIES AND
RESTRICTIONS OF THE PRODUCT
AS STATED IN THE RESPECTIVE
LITERATURE.
DISCLAIMER
TO THE EXTENT PERMITTED BY
LAW, THIS WARRANTY AND THE
REMEDIES SET FORTH ABOVE
ARE EXCLUSIVE AND IN LIEU OF
ALL OTHER WARRANTIES, REME-
DIES AND CONDITIONS, WHETH-
ER ORAL OR WRITTEN, STATU-
TORY, EXPRESS OR IMPLIED. AS
PERMITTED BY APPLICABLE LAW,
ACE SPECIFICALLY DISCLAIMS
ANY AND ALL STAUTORY OR IM-
PLIED WARRANTIES, INCLUDING,
WITHOUT LIMITATION, WARRAN-
TIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR
PURPOSE AND WARRANTIES
AGAINST HIDDEN OR LATENT
DEFECTS. IF ACE CANNOT
LAWFULLY DISCLAIM STATUTORY
OR IMPLIED WARRANTIES THEN
TO THE EXTENT PERMITTED BY
LAW, ALL SUCH WARRANTIES BE
LIMITED IN DURATION TO THE
DURATION OF THE EXPRESS
WARRANTY AND TO THE REPAIR
OR REPLACEMENT SERVICE
AS STATED ABOVE. EXCEPT AS
PROVIDED IN THIS WARRANTY
AND TO THE MAXIMUM EXTENT
PERMITTED BY LAW, ACE IS NOT
RESPONSIBLE FOR DIRECT,
SPECIAL, INCIDENTAL OR CON-
SEQUENTIAL DAMAGES UNDER
ANY LEGAL THEORY.
the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights, and you may also have other rights that vary by country, state or province. This Limited Warranty is governed by and construed under the laws of the country in which the product purchase took place.
3. Advice restriction:
The buyer confirms that he has checked the intended use and suitability of purchased goods for any purpose in sole responsibility. Between the two parties it is expressly agreed upon that all technical or other advice by
ACE regarding use of the goods or services is free of charge and was made with to the exclusive risk of the buyer. Therefore no obligation or liability can arise to
ACE from any recommendation or their result.
Some countries, states and provinces do not allow the exclusion or limitation of incidental or consequential damages or allow limitations on how long an implied warranty or condition may last, so
Imported and distributed by
ACE Handels- und Entwicklungs GmbH
Staufenstr. 1, 83395 Freilassing, Germany www.ace-instruments.net
General Support [email protected]
Calibration Service see www.ace-instruments.net
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement