Operating and Installation Instructions
Diaphragm Vacuum
Pumps and Compressors
Type
Range:
UN035ANP
UN035.3ANP
UN035.1.2SNP
UN035TTP
UN035ATP
UN035.3ATP
UN035.1.2STP
UN035.1.2TTP
UN035AVP
UN035.3AVP
UN035.1.2SVP
UN035.2TTP
UN035.1.2ANP
UN035SNP
UN035.3SNP
UN035.3TTP
UN035.1.2ATP
UN035STP
UN035.3STP
UN035.1.2AVP
UN035SVP
UN035.3SVP
Contents
KNF Neuberger, Inc
2 Black Forest Rd
Trenton, NJ 08691-1810
Phone: 609-890-8600
Fax:
609-890-8323
www.knf.com/usa.htm
Ident#121228-210800
Revision (6/12)
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
Page
About this document................................................................ 3
Use.......................................................................................... 4
Safety ...................................................................................... 5
Technical Data ........................................................................ 7
Design and Function ............................................................... 9
Installation and connection .................................................... 11
Operation .............................................................................. 12
Servicing ............................................................................... 14
Troubleshooting .................................................................... 21
Spare parts and accessories ................................................. 23
Product Return ...................................................................... 24
2
Diaphragm Vacuum Pump UN035_P
About this document
1.
About this document
1.1. Using the Operating Instructions
The Operating Instructions are part of the pump.
Carefully study the Operating Instructions before using a pump.
Always keep the Operating Instructions handy in the work area.
Pass on the Operating Instructions to the next owner.
Project pumps
Customer-specific project pumps (pump models which begin with
“PU” or “MPU”) may differ from the Operating Instructions.
For project pumps, also observe the agreed upon specifications.
1.2. Symbols and Markings
Warning
A danger warning is located here.
Possible consequences of a failure to observe the
warning are specified here. The signal word, e.g.
WARNING Warning, indicates the danger level.
Measures for avoiding the danger and its consequences are specified here.
Danger levels
Signal word Meaning
DANGER
warns of immediate danger
WARNING
CAUTION
Consequences if not observed
Death or serious injuries
and/or serious damage are the
consequence.
warns of possible Death or serious injuries
danger
and/or serious damage are
possible.
warns of a possibly Minor injuries or damage are
dangerous situapossible.
tion
Tab. 1
Other information and symbols
An activity to be carried out (a step) is specified here.
1. The first step of an activity to be carried out is specified here.
Additional, consecutively numbered steps follow.
This symbol refers to important information.
3
Use
Diaphragm Vacuum Pump UN035_P
2.
Use
2.1. Proper use
The pumps are exclusively intended for transferring gases and
vapors.
Owner's responsibility
Operating parameters and
conditions
Only install and operate the pumps under the operating parameters
and conditions described in Chapter 4, Technical data.
Make sure that the installation location is dry and the pump is
protected against rain, splash, hose and drip water.
Requirements for
transferred medium
Before using a medium, check whether the medium can be
transferred danger-free in the specific application case.
Before using a medium, check the compatibility of the materials of
the pump head, diaphragm and valves with the medium.
Only transfer gases which remain stable under the pressures and
temperatures occurring in the pump.
2.2. Improper use
The pump may not be operated in an explosive atmosphere.
The pump is not suitable for transferring dusts.
The pump is not suitable for transferring liquids.
Pumps designed to create either a vacuum or an overpressure
must not be used for these two purposes simultaneously.
An overpressure must not be applied to the suction side of the
pump.
4
Diaphragm Vacuum Pump UN035_P
3.
Safety
Safety
Note the safety precautions in sections 6. Installation and
connection, and 7. Operation.
The pumps are built according to the generally recognized rules of
technology and in accordance with the occupational safety and
accident prevention regulations. Nevertheless, dangers can result
during their use which leads to injuries to the user or others, or to
damage to the pump or other property.
Only use the pumps when they are in a good technical and proper
working order, in accordance with their intended use, observing the
safety advice within the operating instructions, at all times.
Personnel
Make sure that only trained and instructed personnel or specially
trained personnel work on the pumps. This especially applies to
assembly, connection and servicing work.
Make sure that the personnel has read and understood the
operating instructions, and in particular the "Safety" chapter.
Working in a safetyconscious manner
Observe the accident prevention and safety regulations when
performing any work on the pump and during operation.
Do not expose any part of your body to the vacuum.
Open housing parts with notice sticker (see fig. 1) only after
separating mains plug from power source.
Fig. 1: Notice sticker
Handling dangerous media
When transferring dangerous media, be sure to observe the safety
regulations when handling these media.
Handling flammable media
Be aware that the pumps are not designed to be explosion-proof.
Make sure the temperature of the medium is always sufficiently
below the ignition temperature of the medium, to avoid ignition or
explosion. This also applies for unusual operational situations.
Note that the temperature of the medium increases when the pump
compresses the medium.
Hence, make sure the temperature of the medium is sufficiently
below the ignition temperature of the medium, even when it is
compressed to the maximum permissible operating pressure of the
pump. The maximum permissible operating pressure of the pump
is stated in the technical specifications (see section 4).
If necessary, consider any external sources of energy, such as
radiation, that may add heat to the medium.
In case of doubt, consult the KNF customer service.
Environmental protection
Store all replacement parts in a protected manner and dispose of
them properly in accordance with the applicable environmental
protection regulations. Observe the respective national and
international regulations. This especially applies to parts
contaminated with toxic substances.
5
Safety
Diaphragm Vacuum Pump UN035_P
Standards
The pumps conform to the safety regulations of the EC Directive
2004/108/EG concerning Electromagnetic Compatibility and the
EC Directive 2006/42/EG concerning Machinery.
The following harmonized standards have been used:
IP20
DIN EN 55014-1/2
DIN EN 61000-3-2/3
DIN EN 60335-1
IP44
DIN EN 55014-1/2
DIN EN 61000-3-2/3
DIN EN 60204-1
Tab. 2
Customer service and
repairs
Only have repairs to the pump carried out by the KNF Customer
Service responsible.
Only authorized personnel should open those parts of the housing
that contain live electrical parts.
Use only genuine parts from KNF for servicing work.
6
Diaphragm Vacuum Pump UN035_P
4.
Technical Data
Technical Data
Pump materials
Pump type
UN035ANP
UN035.1.2ANP
UN035.3ANP
UN035ATP
UN035.1.2ATP
UN035.3ATP
UN035AVP
UN035.1.2AVP
UN035.3AVP
UN035TTP
UN035.1TTP
UN035.1.2TTP
UN035.2TTP
UN035.3TTP
UN035SNP
UN035.1.2SNP
UN035.3SNP
UN035STP
UN035.1.2STP
UN035.3STP
UN035SVP
UN035.1.2SVP
UN035.3SVP
Material*
Pump head
Diaphragm
Valve
Gasket
Aluminium
CR
Stainless steel
CR
Aluminium
PTFE-coated
Stainless steel
FPM
Aluminium
FPM
Stainless steel
FPM
Aluminum with
PTFE-coated
PVDF plating
Stainless steel
FPM
Stainless steel CR
CR
-
Stainless steel PTFE-coated
PTFE
-
Stainless steel FPM
FPM
-
Tab. 3
*According to DIN ISO 1629 und 1043.1
7
Technical Data
Diaphragm Vacuum Pump UN035_P
Pneumatic Values
Pump type
UN035ANP
UN035SNP
UN035ATP
UN035TTP
UN035STP
UN035AVP
UN035SVP
UN035.1.2ANP
UN035.1.2SNP
UN035.1.2ATP
UN035.1.2TTP
UN035.1.2STP
UN035.1.2AVP
UN035.1.2SVP
UN035.3ANP
UN035.3SNP
UN035.3ATP
UN035.3TTP
UN035.3STP
UN035.3AVP
UN035.3SVP
Delivery rate* at
Maximal operating
atm. pressure (l/min) pressure (bar g)
30
4
27
30
2
100
55
4
50
55
2
30
-
13
27
-
20
30
-
13
Tab. 4
8
Ultimate vacuum
(mbar abs.)
*Liters in standard state (1013 bar)
Diaphragm Vacuum Pump UN035_P
5.
Design and Function
Design and Function
Design UN035A_P
1
2
3
4
Pneumatic pump outlet
Pneumatic pump inlet
On/Off Switch
Handle
Fig. 2: Diaphragm Pump U 035ANP
Design UN035.1.2A_P
1
2
3
4
5
6
Pneumatic pump outlet
Pneumatic pump inlet
On/Off Switch
Pneumatic head connection 1
Handle
Pneumatic head connection 2
Fig. 3: Diaphragm Pump UN035.1.2ANP
9
Design and Function
Diaphragm Vacuum Pump UN035_P
Design UN035.3A_P
1
2
3
4
5
6
Pneumatic pump outlet
Pneumatic pump inlet
On/Off Switch
Handle
Pneumatic head connection
Silencer/filter (accessory)
Fig. 4: Diaphragm pump U 035.3ANP
Function diaphragm pump
1
2
3
4
5
6
7
Outlet valve
Inlet valve
Transfer chamber
Diaphragm
Eccentric
Connecting rod
Pump housing
Fig. 5: Pump head
The pump transfers, compresses (depending on pump version)
and evacuates gases and vapors.
The elastic diaphragm (4) is moved up and down by the eccentric
(5) and the connecting rod (6). In the downward stroke it aspirates
the gas to be transferred via the inlet valve (2). In the upward
stroke, the diaphragm presses the medium out of the pump head
via the outlet valve (1). The transfer chamber (3) is hermetically
separated from the pump drive (7) by the diaphragm.
10
Diaphragm Vacuum Pump UN035_P
Installation and connection
6.
Installation and connection
Only install and operate the pumps under the operating parameters
and conditions described in chapter 4, Technical data.
Observe the safety precautions (see chapter 3).
6.1. Installation
Before installation, store the pump at the installation location to
bring it up to room temperature.
Dimensions
Cooling air supply
Installation location
See chapter 4, Technical data, for the dimensions of pump.
Install the pump so that the motor fan can intake sufficient cooling
air.
Make sure that the installation location is dry and the pump is
protected against rain, splash, hose and drip water.
Choose a safe location (flat surface) for the pump.
Protect the pump from dust.
Protect the pump from vibrations and jolts.
6.2. Connection
Connected components
Only connect components to the pump which are designed for the
pneumatic data of the pump (see section 4).
Pump exhaust
If the pump is used as a vacuum pump, safely discharge the pump
exhaust at the pump’s pneumatic outlet.
Connection
A marking on the pump head shows the direction of flow.
1. Remove the protective plugs from the pneumatic connectors of
the pump.
2. Mount accessory silencer/filter (where applicable):
If the pump is used as a vacuum pump, mount the silencer at
the pressure side if necessary. If the pump is used as a compressor (not permitted with series UN 035.3) mount the filter at
the suction side if necessary.
−
Unscrew corresponding hose connector from pump head.
−
Screw silencer/filter into pump head.
3. Connect the suction line and pressure line (UN035._A _P: tube
inner diameter 9 mm; UN035._A _P: tube inner diameter 10
mm).
4. Lay the suction and pressure line at a downward angle to
prevent condensate from running into the pump.
5. Insert the power cable’s plug into a properly installed
shockproof socket.
11
Operation
Diaphragm Vacuum Pump UN035_P
7.
Operation
7.1. Preparing for Start-up
Before switching on the pump, observe the following points:
Pump
Operational requirements
All hoses attached properly
Fan openings not blocked
Specifications of the power supply correspond with the
data on the pump’s type plate.
The pump outlet is not closed or constricted.
Tab. 9
7.2. Starting
Only operate the pump under the operating parameters and conditions described in chapter 4, Technical data.
Make sure the pump is used properly (see section 2.1).
Make sure the pump is not used improperly (see section 2.2).
Observe the safety precautions (see chapter 3).
Hazard of the pump head bursting due to excessive
pressure increase
Do not exceed max. permissible operating pressure
WARNING (see section 4).
Monitor pressure during operation.
If the pressure exceeds the maximum permissible
operating pressure, immediately shut down pump
and eliminate fault (see chapter 9. Troubleshooting).
Only throttle or regulate the air or gas quantity in the
suction line to prevent the maximum permissible
operating pressure from being exceeded.
If the air or gas quantity in the pressure line is throttled or regulated, make sure that the maximum
permissible operating pressure of the pump is not
exceeded.
Excessive pressure (with all of the related hazards) can be
prevented by placing a bypass line with a pressure-relief valve
between the pressure and suction side of the pump. For further
information, contact our technical adviser.
Pump standstill
12
With the pump at a standstill, open pressure and suction lines to
normal atmospheric pressure.
Diaphragm Vacuum Pump UN035_P
Operation
Automatic starting can cause personal injury and
pump damage
When the operation of the pump is interrupted by the
WARNING thermal switch, the pump will restart automatically
after cooling down.
After triggering of the thermal protection or in the
event of power failure, remove the pump’s mains
plug from the socket so that the pump cannot start
uncontrollably.
Attempt work on the pump only if the pump is
separated from mains power.
7.3. Switching pump on and off
Switching pump on
The pump may not start up against pressure or vacuum during
switch-on. This also applies in operation following a brief power
failure. If a pump starts against pressure, it may block. This
activates the thermal switch, and the pump switches off.
Make sure that no pressure or vacuum is present in the lines
during switch-on.
Switch on pump with mains switch (see fig. 2, 3 or 4, position 3).
Switching off the pump/removing from operation
When transferring aggressive media, flush the pump prior to
switch-off to increase the service life of the diaphragm (see section
8.2.1).
Switch off pump with mains switch (see fig. 2, 3 or 4, position 3).
Open pressure and suction lines to normal atmospheric pressure.
13
Servicing
Diaphragm Vacuum Pump UN035_P
8.
Servicing
8.1. Servicing Schedule
Component
Pump
Diaphragm, reed valves
(valve plate)
Silencer/filter (accessory)
Servicing interval
Regular inspection for external
damage or leaks
Replace at the latest, when pump
output decreases
Change if it is dirty
Tab. 10
8.2. Cleaning
When cleaning, make sure that no liquids enter the inside of
the housing.
8.2.1. Flushing Pump
Before switching off the pump, flush it with air (if neccesary for
safety reasons: with an inert gas) for about five minutes under
atmospheric conditions (ambient pressure).
8.2.2. Cleaning Pump
Only use solvents for cleaning if the head materials cannot be
attacked (check the resistance of the material!).
If compressed air is available, blow out the components.
14
Diaphragm Vacuum Pump UN035_P
Servicing
8.3. Changing Diaphragm and Valves
8.3.1. Pumps with aluminium head
Conditions
Spare parts
UN035ANP
UN035.1.2ANP
UN035.3ANP
UN035ATP
UN035.1.2ATP
UN035.3ATP
UN035TTP
UN035.1.2TTP
UN035.3TTP
UN035AVP
UN035.1.2AVP
UN035.3AVP
ƒ
Pump is switched off and mains plug is removed from the
socket
ƒ
Pump is clean and free of hazardous materials
ƒ
Tubes removed from pump’s pneumatic inlet and outlet
Spare part*
Position**
Diaphragm
Countersunk
screw
Reed valve
Gasket
(F)
(D)
Quantity per
pump head
1
1
(M,P)
(V)
2
1
Tab. 11
Tools
Quantity
1
1
1
1
1
1
1
*According to Spare parts list, chapter
10
**According to Fig. 6.
Tools/Material
Allen key 3 mm
Allen key 4 mm
Allen key 5 mm
Screwdriver blade width 6.5
Screwdriver blade width 4.0
Fork wrench 16 mm (only for two-headed pumps)
Pencil
Tab. 12
Information on procedure
With multi-head pumps, parts of the individual pump heads can be
confused.
Replace the diaphragm and reed valves of the individual pump
heads consecutively.
Health hazard due to dangerous substances
in the pump!
Depending on the substance transferred, caustic
WARNING burns or poisoning are possible.
Wear protective clothing if necessary, e.g. protective
gloves.
Flush pump before replacing the diaphragm and
redd valves (see section 8.2.1).
15
Servicing
Diaphragm Vacuum Pump UN035_P
Fig. 6: Pump parts for versions with aluminium head
1. For pumps UN035.1.2A_P and UN035.1.2TTP:
On both pneumatic head connections, open the head connection union nut on one pump head and pull the hose off.
2. For pumps UN035.3A_P and UN035.3TTP:
At one pump head open the union nut of pneumatic connection
and pull off the tube.
3. Mark the position of the diaphragm head C in relation of the
housing A with a pencil.
4. Loosen the four allen screws B and remove the diaphragm
head C.
5. Unscrew the countersunk screw D, remove the retainer plate E
and the diaphragm F.
6. Loosen the four screws G and remove the cover plate H.
7. Turn the counterweight J so that the connection rod K is in the
mid-position; fit the new diaphragm F.
8. Place the retainer plate E on the diaphragm F and carefully but
formly tighten the new countersunk screw D (torque: 5.0 Nm).
16
Diaphragm Vacuum Pump UN035_P
Servicing
The self-locking screw D can only be used once.
9. Change lower reed valve:
−
Undo the cheese head screw N and exchange the reed
valve M.
10. Change upper reed valve:
−
Loosen the allen screws S, remove the cover plate T and
the gasket V.
−
Undo the cheese head screw U and exchange the reed
valve P; tighten the cheese head screw U.
−
Replace the cover plate T with a new gasket V and tighten
the allen screws S.
11. Place the diaphragm head C according to the marks made
previously and tighten the screws B uniformly and diagonally
(torque: 10.0 Nm).
12. Turn the counterweight J to check that the pump run freely.
13. For two-headed pumps:
Carry out steps 3 to 12 for the second pump head.
14. Replace the cover plate H and secure it with the four screws G.
15. For two-headed pumps:
Reattach the tube of pneumatic head connection onto the hose
connector and tighten the union nut.
For UN035.1.2A_P and UN035.1.2TTP pumps, this applies
to both pneumatic head connections.
17
Servicing
Diaphragm Vacuum Pump UN035_P
8.3.2. Pumps with stainless steel head
Conditions
Spare parts
UN035SNP
UN035.1.2SNP
UN035.3SNP
UN035STP
UN035.1.2STP
UN035.3STP
UN035SVP
UN035.1.2SVP
UN035.3SVP
ƒ
Motor disconnected from mains and de-energized
ƒ
Pump is clean and free of hazardous materials
Spare part*
Position**
Diaphragm
Countersunk
screw
Valve plate
(F)
(D)
Quantity per
pump head
1
1
(Z)
1
Tab. 13
Tools
Quantity
1
1
1
1
*According to Spare parts list, chapter
10
**According to Fig. 7
Tools/Material
Allen key 4 mm
Allen key 5 mm
Screwdriver blade width 6.5
Pencil
Tab. 14
Information on procedure
With multi-head pumps, parts of the individual pump heads can be
confused.
Replace the diaphragm and valve plate of the individual pump
heads consecutively.
Health hazard due to dangerous substances
in the pump!
Depending on the substance transferred, caustic
WARNING burns or poisoning are possible.
Wear protective clothing if necessary, e.g. protective
gloves.
Flush pump before replacing the diaphragm and the
valve plate (see section 8.2.1).
18
Diaphragm Vacuum Pump UN035_P
Servicing
Fig. 7: Pump parts for versions with stainless steel head
1. For pumps UN035.1.2S_P:
On the pneumatic connection between the pump heads on the
suction side, pull the hose off one pump head.
On the pneumatic connection on the pressure side, loosen the
hose clip on one pump head and pull the hose off.
2. For pumps UN035.3S_P:
Pull the pneumatic head connection hose off one pump head.
3. Mark the position of the head plate W and intermediate plate X
in relation of the housing A with a pencil.
4. Loosen the allen screws Y and remove the head plate W, valve
plate Z and intermediate plate X.
5. Unscrew the countersunk screw D, remove the retainer plate E
and the diaphragm F.
6. Loosen the four screws G and remove the cover plate H.
7. Turn the counterweight J so that the connection rod K is in the
mid-position; fit the new diaphragm F.
8. Place the retainer plate E on the diaphragm F and carefully but
formly tighten the new countersunk screw D (torque: 5.0 Nm).
19
Servicing
Diaphragm Vacuum Pump UN035_P
The self-locking screw D can only be used once.
9. Place the intermediate plate X on the top of the diaphragm F so
that it corresponds to the marks on the housing.
10. Place the new valve plate Z on the intermediate plate X.
11. Place the head plate W according to the marks made previously and tighten the screws Y uniformly and diagonally (torque:
10.0 Nm).
12. Turn the counterweight J to check that the pump run freely.
13. For two-headed pumps:
Carry out steps 3 to 12 for the second pump head.
14. Replace the cover plate H and secure it with the four screws G.
15. For two-headed pumps:
Pull the pneumatic head connection hose (or hoses in the case
of UN035.1.2S_P pump models) back onto the hose connector.
For UN035.1.2S_P pump models: Retighten the hose clip on
the pneumatic head connection on the pressure side.
20
Diaphragm Vacuum Pump UN035_P
Troubleshooting
9.
Troubleshooting
Extreme danger from electrical shock!
Disconnect the pump power supply before working on
the pump.
DANGER Make sure the pump is de-energized and secure.
Check the pump (see Tab. 15 to 18).
Pump does not transfer
Cause
No voltage in the power source
Fault remedy
Check room fuse and switch on if necessary.
Thermal switch has operated
following to over-heating.
Disconnect pump from mains.
Allow pump to cool.
Trace cause of over-heating and eliminate it.
Check connections and lines.
Remove blockage.
Check external valves and filters.
Connections or lines blocked.
External valve is closed or filter
is clogged.
Condensate has collected in
pump head.
Diaphragm or reed valves
(valve plate) are worn.
Detach the condensate source from the pump.
Flush pump (see Section 8.2.1).
Replace diaphragm and reed valves (valve plate), (see Section
8.3).
Tab. 15
Flow rate, pressure or vacuum too low
The pump does not achieve the output specified in the Technical data or the data sheet.
Cause
Fault remedy
Condensate has collected in
Detach the condensate source from the pump.
pump head.
Flush pump (see Section 8.2.1).
There is gauge pressure on
Change the pressure conditions.
pressure side and at the same
time vacuum or a pressure
above atmospheric pressure on
suction side.
Disconnect pump from system to determine output values.
Pneumatic lines or connection
parts have an insufficient cross Eliminate throttling (e.g. valve) if necessary.
section.
Use lines or connection parts with larger cross section if necessary.
Leaks occur on connections,
Check that tubes sit correctly on hose nozzles.
lines or pump head.
Replace leaky tubes.
Eliminate leaks.
Connections or lines completely Check connections and lines.
or partially jammed.
Remove the jamming parts and particles.
Head parts are soiled.
Clean head components.
Diaphragm or reed valves
(valve plate) are worn.
Replace diaphragm and reed valves (valve plate),
(see Section 8.3).
21
Troubleshooting
Diaphragm Vacuum Pump UN035_P
Flow rate, pressure or vacuum too low
The pump does not achieve the output specified in the Technical data or the data sheet.
Cause
Fault remedy
Check the tubing for leaks.
After diaphragm and reed
valves (valve plate) have been
Pumps UN035_A_P and UN035_TTP:
replaced
Possibly carefully tighten the screws (B) and (S) (see fig. 6)
crosswise.
Pumps UN035_S_P:
Possibly carefully tighten the screws (Y) (see fig. 7) crosswise.
Tab. 16
Pump is switched on, but does not run, the on/off-switch on the pump is not lit
Cause
Pump is not connected with the
power source.
No voltage in the power source
Fault remedy
Connect pump to mains power.
Check room fuse and switch on if necessary.
Tab. 17
Pump is switched on, but does not run, the on/off-switch on the pump is lit
Cause
The thermal switch has opened
due to overheating
Cause
Remove pump´s mains plug from the socket.
Allow pump to cool.
Trace cause of over-heating and elimate it.
Tab. 18
Fault cannot be rectified
If you are unable to determine any of the specified causes, send
the pump to KNF Customer Service (see first page for the address).
1. Flush the pump to free the pump head of dangerous or aggressive gases (see Section 8.2.1).
2. Clean the pump (see Section 8.2.2).
3. Send the pump to KNF with a filled out decontamination declaration (see chapter 11) and specification of the medium transferred.
22
Diaphragm Vacuum Pump UN035_P
Spare parts and accessories
10. Spare parts and accessories
Spare parts
UN035ANP
UN035.1.2ANP
Spare part
Diaphragm
Countersunk screw
Reed valve
Gasket
Position*
(F)
(D)
(M, P)
(V)
UN035.3ANP
Order No.
001312
110711
001328
001326
Tab. 19
According to Fig. 6
UN035ATP
UN035.1.2ATP
UN035.3ATP
UN035TTP
UN035.1.2TTP
UN035.3TTP
Spare part
Diaphragm
Countersunk screw
Reed valve
Gasket
Position*
(F)
(D)
(M, P)
(V)
Order No.
001406
110711
001328
011796
Tab. 20
UN035AVP
According to Fig. 6
UN035.1.2AVP
Spare part
Diaphragm
Countersunk screw
Reed valve
Gasket
Position*
(F)
(D)
(M, P)
(V)
UN035.3AVP
Order No.
001405
110711
001328
011796
Tab. 21
UN035SNP
According to Fig. 6
UN035.1.2SNP
Spare part
Diaphragm
Countersunk screw
Valve plate
Position*
(F)
(D)
(Z)
Tab. 22
UN035STP
UN035.3SNP
Order No.
001312
110711
001308
*According to Fig. 7
UN035.1.2STP
Spare part
Diaphragm
Countersunk screw
Valve plate
Position*
(F)
(D)
(Z)
UN035.3STP
Order No.
001406
110711
001407
Tab. 23
UN035SVP
*According to Fig.7
UN035.1.2SVP
Spare part
Diaphragm
Countersunk screw
Valve plate
Tab. 24
Position*
(F)
(D)
(Z)
UN035.3SVP
Order No.
001405
110711
001404
*According to Fig. 7
23
Product Return
Diaphragm Vacuum Pump UN035_P
11. Product Return
Î KNF provides warranty and non-warranty repair services for all
products.
Î A Return Material Authorization (RMA) number is required for
all product returns.
• To receive an RMA number, submit a completed
Decontamination Declaration form to [email protected]
Î The Decontamination Declaration form can obtained from our
website or by contacting KNF Technical Services.
KNF Neuberger, Inc
2 Black Forest Rd
Trenton, NJ 08691-1810
Phone 609-890-8600
Fax
609-890-8323
www.knf.com/usa.htm
24
•
http://www.knf.com/pdfs/decontamdec.doc
•
Phone: 609-890-8600
Î Product return instructions will be provided when the RMA is
issued.
Was this manual useful for you? yes no