attoCFM II
low temperature confocal microscope, highly stable and compact
Technical Specifications
General Specifications
type of instrument
fiber optics based confocal microscope for maximum stability
sensor head specifics
fiber-coupled low temperature compatible aspheric objectives for ultimate stability
Confocal Unit
configuration
one single-mode fiber for excitation and detection;
50/50 fiber coupler integrated into laser/detector module LDM600 (for more
information, see SPM fu
key benefits
ultimate longterm stability & ease-of-use
pinhole configuration
pinhole provided by single mode fiber core
pinhole size
dependent on fibers, typically 3 .. 9 µm mode field diameter
LT- compatible objective
LT-IWDO, LT-LWDO (see accessory section for more information)
Illumination
excitation wavelength range
limited to wavelength range of single mode fiber
default: 650 nm (others on request)
Detection
detection mode
e.g. reflection, luminescence, flourescence
detection wavelength range
limited to wavelength range of single mode fiber
default: 650 nm (others on request)
Sample Positioning
total travel range
5 x 5 x 5 mm³ (open loop)
step size
0.05..3 µm @ 300 K, 10..500 nm @ 4 K
fine scan range
50 x 50 µm² @ 300 K, 30 x 30 µm² @ 4 k (optional, open loop)
sample holder
ASH/QE/0 quick exchange sample holder and integrated heater with calibrated
temperature sensor
Suitable Operating Conditions
temperature range
1.5 K..300 K (dependent on cryostat); mK compatible setup available on request
magnetic field range
0..15 T+ (dependent on magnet)
operating pressure
designed for He exchange gas (vacuum compatible version down to 1E-6 mbar on
request)
Suitable Cooling Systems
titanium housing diameter
48 mm
bore size requirement
designed for a 2" (50.8 mm) cryostat/magnet bore
compatible cryostats
attoDRY1000/1100/2100
attoLIQUID1000/2000/3000
(attoLIQUID5000 on request)
Electronics
scan controller and software
ASC400 (optional; for detailed specifications please see attoCONTROL section)
laser
LDM600 laser/detector module (for detailed specifications please see attoCONTROL
section)
Options
closed loop scanning & global sample
coordinates
Interferometric encoders for scan linearization and closed loop sample navigation
ultra-large scan range upgrade
80 x 80 ?m2 @ 300 K
125 x 125 ?m2 @ 4 K
closed loop upgrade for coarse
positioners
resistive encoder, range 5 mm, sensor resolution approx. 200 nm, repeatability 1-2
µm
sample holder upgrade
ASH/QE/4CX quick-exchange sample holder with 8 electrical contacts and integrated
heater with calibrated temperature sensor
* Resolution may vary depending on applied tip, sample, and cryostat
All rights, including rights created by patent grant or registration of a utility model or design as well as rights of technical modifications are reserved. Delivery