Megohmmeter AEMC® Model 6240

Add to My manuals
11 Pages

advertisement

Megohmmeter AEMC® Model 6240 | Manualzz

SERVICE/CALIBRATION MANUAL

Megohmmeter

AEMC

®

Model 6240

(Chauvin Arnoux Model C.A 6240)

v1 – 01/09

This document is confidential property of Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

Instruments and may not be reproduced without permission.

AEMC

®

Instruments retains the right to make changes to specifications herein at any time, without notice.

MODEL 6240 SERVICE/CALIBRATION MANUAL

TABLE OF CONTENTS

Cat. #2129.80

1. SPECIFICATIONS .................................................................................................................................... 3

ELECTRICAL SPECIFICATIONS .................................................................................................... 3

Mechanical ....................................................................................................................................... 4

Environmental .................................................................................................................................. 5

Safety ............................................................................................................................................... 5

2.

CALIBRATION PROCEDURE ............................................................................................................... 6

Equipment Needed .......................................................................................................................... 6

Repair and Calibration ................................................................................................................................ 9

Technical and Sales Assistance ................................................................................................................ 9

Limited Warranty ....................................................................................................................................... 10

Warranty Repairs ...................................................................................................................................... 10

Rev Description

1 Created Document January 9, 2009 JPM v1 – 01/09

- 2 -

MODEL 6240

1. SPECIFICATIONS

ELECTRICAL SPECIFICATIONS

SERVICE/CALIBRATION MANUAL Cat. #2129.80

Specifications are given for an ambient temperature of 23°C ± 3°C, relative humidity of 45 to 55%, supply voltage of 6V ± 0.2V and magnetic field of <40A/m.

INPUT VOLTAGE CHECK

Measurement Range

Resolution

Input Impedance

Frequency Range

10.0 to 39.99V 40.0 to 399.99V 400 to 499V

0.01V 0.1V 1V

800k

DC and 15.3 to 450Hz

RESISTANCE

Measurement Range

Resolution

Accuracy

Measurement current

5.0 to

3999µ



4.0 to

39.99m

10µ



40.0 to

399.9m

100µ



400 to

3999m

1m

± 0.25% ± 2ct



1.02A

± 2%

102mA

± 2%

10.2A

± 2%

(1)

Open voltage

4 to 6V

(1) With nominal value of 10.2A, the measurement current is at least 10A whatever the charge condition of the battery.

(2) The current is 10mA only up to 300

. If the battery is low, it can fall to as low as 8mA.

VOLTAGE MEASUREMENT INDICATION

Measurement Range

0.01 to

3.999mV

4.0 to

39.99mV

Resolution

1µV

CURRENT MEASUREMENT INDICATION

10µV

40.0 to

399.9mV

100µV

0.40 to

3.999V

1mV

4.0 to

39.99

10m



Measurement Range

Resolution

5.0 to

39.99mA

10µA

40.0 to

399.9mA

100µA

0.40 to

3.999A

1mA

40.0 to

399.9

100m



10.2mA

± 2%

(2)

4.0 to

4.70V

10mV

4.0 to

11.00A

10mA v1 – 01/09

- 3 -

MODEL 6240 SERVICE/CALIBRATION MANUAL Cat. #2129.80

Influences on the Resistance Measurement

Quantities of Influence

Temperature

Relative Humidity

Supply Voltage

Series Mode Rejection,

50/60Hz*

Range of use

-10 to +55°C

10 to 85% RH @ 45°C

5 to 7V

U (AC) = (R measured x I measurement)

Variation of the measurement

Typical Maximum

0.1% / 10°C

0.1 %

2cts

0.5% / 10 °C + 2cts

0.5% + 2cts

0.2% / V + 2cts

< 0.2% 2% + 1ct

Common Mode Rejection,

50/60Hz AC

0 to 50 V AC > 80 dB > 60 dB

*Example: If the measured resistance is 1m

and the measurement current is 10A, an alternating voltage of 1mV RMS in series with the resistance to

be measured will induce an error of not more than 2%.

Power Supply: 6V, 8.5Ah rechargeable NiMH battery pack

Battery Life: Range dependent

Range Number of measurements*

10A 850

1A 3500

100mA 4500

10mA 5000

In Standby or Off battery life 4 to 6 months

*Established for measurements lasting 5s, every 25s.

Mechanical

Dimensions:

10.70 x 9.76 x 7.17" (272 x 248 x 182mm)

Weight:

10 lbs (4.5kg approx)

Case Protection:

ABS plastic polycarbonate: watertight to IP54 (cover closed), water resistant to IP53 (cover open).

Color:

Safety yellow case with gray faceplate v1 – 01/09

- 4 -

MODEL 6240

Environmental

SERVICE/CALIBRATION MANUAL

Operating Temperature:

14° to 131°F (-10° to 55°C), 10 to 85% RH

Storage Temperature:

-40° to 158°F (-40° to 70°C); 10 to 90% RH

Altitude: <2000m

Cat. #2129.80

NOTE: For long-term storage (>1 year) with the battery installed, the storage range is -4 to 86°F (-20 to

30°C); 85% RH, otherwise the battery will deteriorate. For short-term storage (1 month) the temperature can reach up to 122°F (50°C).

Safety

Electrical safety as per EN 61010-1 (Ed. 2 of 2001), EN 61557 (Ed. 97) parts 1 and 4.

Degree of pollution: 2

Protection: Measurement Cat. III, 50V with respect to earth, 500V differential between terminals, and 300V Cat. II on the charger input

Electromagnetic Compatibility:

The instrument satisfies the CEM and DBT directives required for the CE marking and product standard EN 61326-1 (Ed. 97) + A1 (Ed. 98)

Emissions in residential environment

Immunity in industrial environment

*All specifications are subject to change without notice

v1 – 01/09

- 5 -

MODEL 6240 SERVICE/CALIBRATION MANUAL Cat. #2129.80

Equipment Needed

 PC : one free serial port

 a serial communication optical cable software

 Stables AC/DC voltage sources Fluke 5500 or equivalent

 DC voltages : 30, 300 and 3000mVDC ±10% whose values known better than 0.05%

 AC voltages : 30V 50 Hz ±10% whose values known better than 0.1%

 the operator will be asked to enter , in dialog boxes the values of these voltages .

Resistors/shunts:

 They will have to be large enough not to vary significantly during calibration : see the needed wattages below . Use low drift resistors.

 these values need to be known exactly but can be ± 10% from the nominal values

30

 with 10mA so 0.003 W

3

 with 100mA so 0.03 W

30m

 with 1A so 0.03 W

3m

 with 10A so 0.3 W

 the operator will be asked to enter, in dialog boxes, the values of these resistors:

If you have ordered our calibration boxes, you will have 7 others resistors for calibration check included:

0.6m

22m

 10

1.25m

 10

A

330m

 100

2.7

 100

21

 10

341

 10

_________________________________________________________________________________

The CA6240 can work using two different modes :

 the "User" mode is the normal mode with the display showing measurement values.

 the "Production" (or "factory") mode is the one used in factory or calibration lab to adjust the instrument.

When in Production mode, the instrument display shows "Prod CAL" flashing.

Install the software to the PC

It will be installed in a directory CA6240-V10

The program's toolbar includes two buttons allowing to switch an instrument from a mode to the other one.

To do a calibration, it is necessary to use the "Pr" named one to switch your instrument to the "Production" mode .

Files list after HDD installation

---------------------------------

Directory "CA6240"

"Service_CA6240_V10.exe" : executable program file V1.0

"Service_CA6240_V10.INI" : program configuration file V1.0

"CA6240_V10.uir" : main panel V1.0

--------Directory "aide"

"Service_Aide.txt" help file

- 6 -

v14 - 09/06

MODEL 6240 SERVICE/CALIBRATION MANUAL Cat. #2129.80

--------Directory "buttons"

"go.ico" : toolbar icon files

"ident.ico"

"print.ico"

"stop.ico"

"prod.ico"

"utilisateur.ico"

--------Directory "fiches"

"Service_CA6240_Ind00.dat" : ( service mode ) calibration file at Indice 00

--------Directory "images"

"Afficheur_ca6240.pcx" : instrument's pictures

"CA6240_1.pcx"

Setting the serial port communication number

From Explore/CA6240 select the service CA6240_V10.INI file

--------------------------------------------

 Set the line "5" parameter ( "5" located before "|" symbol ) to the number you want ( assuming 1 to 9 only

) .

E.g; to set the port 3 :

"

#Port Serie pour l'Appareil en test : 1 pour COM1 .......9 pour COM9 ( 1 à 9 possible )

5|3

"

 Close the file and from the Start>Programs/CA6240/Service CA6240_V10.INI launch the calibration program.

Before Calibration:

The instrument has to be at the reference temperature: 23º C ± 3 C for at least 2 Hours

Avoid to calibrate it just after a battery charge ( which increases the inside temperature significantly)

After a battery charge, you can open the case to stabilize the temperature faster.

Do not calibrate when charging.

Following the calibration steps is a step to verify:

- the battery internal resistance

- the battery temperature

- the internal 10A shunt temperature

These last tests are only indicative and can be void if. It can be an indication of a defective battery or defective temperature sensor.

Serial Number :

---------------

When you recalibrate an instrument, the serial number has to be kept without any modification; don’t execute

the last step called "Serial Number - Switch to User Mode" by using the dedicated button and quit the program.

In case of a board failure you may have to reset a serial number.

That can be done only if all the previous steps were corrects.

After the step "Serial Number - Switch to User Mode".

At the end of this step, automatically :

- the instrument is softly rebooted

- the Instrument Identification's window appears

to verify :

- serial number

- calibration date

- etc ......

- 7 -

v14 - 09/06

MODEL 6240 SERVICE/CALIBRATION MANUAL Cat. #2129.80

- the Instrument is switched back to the "User" mode .

Note :

The instrument always starts in "User" mode.

To calibrate it, you have to switch it to the "Production" mode. (from the software menu)

After calibration, the PC will switch it automatically back to the "User" mode to be able to use it normally.

 Use the two right buttons on the toolbar to enter the values, use decimal point "." but not comma ","

 Take care with the unities to use when entering these values

 (µ or 0.1µ for resistances and µV or mV for voltages : that will be written in the dialog boxes ) .

 During the calibration, if an incorrect coefficient is detected, the software will switch to manual calibration

(if automatic selected before) and the screen "Coefficients" is displayed.

 The operator can then stop and exit or redo the same test: to do so, click on the "Calibration/Control" screen and then click on the "Go" button.

 At completion of the test, the number of a correct step or coefficient becomes green colored and red if the test has failed.

 The "?" button allows the instrument's identification ( serial number, firmware version , coefficients , .... ) : so it can be used to check the serial communication .

 To uninstall this program, run again the "SetUp.exe" of the CD.

Quick Start :

 Using the CD , install the program .

 Set the Serial Port Communication Number in the configuration "Service_CA6240_V10.INI" file (1 to 9 for

COM1 to COM9 ).

 Connect the instrument to the PC using an optical serial link ( DB9 or USB ).

 Start the "Service_CA6240_V10.exe" program using the navigator or "Start" button.

 Switch your instrument from the "User" mode to the "Production" mode, using the toolbar's "Pr" button.

 Check the right communication PC/Instrument using the "?" button (Instrument Identification).

 Start the calibration by clicking the "Go" button.

- 8 -

v14 - 09/06

MODEL 6240

Repair and Calibration

SERVICE/CALIBRATION MANUAL Cat. #2129.80

To ensure that your instrument meets factory specifications, we recommend that it be submitted to our factory

Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures.

For instrument repair and calibration:

You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus recorded calibration data).

Chauvin Arnoux

®

, Inc.

d.b.a. Instruments

15 Faraday Drive

Dover, NH 03820 USA

Tel: (800) 945-2362 (Ext. 360)

(603) 749-6434 (Ext. 360)

Fax: (603) 742-2346 or (603) 749-6309

[email protected]

(Or contact your authorized distributor)

Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.

NOTE: All customers must obtain a CSA# before returning any instrument.

Technical and Sales Assistance

If you are experiencing any technical problems, or require any assistance with the proper operation or application of your instrument, please call, mail, fax or e-mail our technical support hotline:

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

Instruments

200 Foxborough Boulevard

Foxborough, MA 02035, USA

[email protected]

www.aemc.com

NOTE: Do not ship Instruments to our Foxborough, MA address

- 9 -

v14 - 09/06

MODEL 6240

Limited Warranty

SERVICE/CALIBRATION MANUAL Cat. #2129.80

The Model 6240 is warranted to the owner for a period of one year from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC

®

Instruments, not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the defect is related to service not performed by AEMC

®

Instruments.

For full and detailed warranty coverage, please read the Warranty Coverage Information, which is attached to the

Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the Warranty Coverage

Information with your records.

What AEMC

®

Instruments will do:

If a malfunction occurs within the one-year period, you may return the instrument to us for repair, provided we have your warranty registration information on file or a proof of purchase. AEMC

®

Instruments will, at its option, repair or replace the faulty material.

Warranty Repairs

What you must do to return an Instrument for Warranty Repair:

First, request a Customer Service Authorization Number (CSA#) by phone or by fax from our Service Department

(see address below), then return the instrument along with the signed CSA Form. Please write the CSA# on the outside of the shipping container. Return the instrument, postage or shipment pre-paid to:

Arnoux

®

, Inc. d.b.a. AEMC

®

Instruments

15 Faraday Drive • Dover, NH 03820 USA

Tel:

Fax:

(800) 945-2362 (Ext. 360)

(603) 749-6434 (Ext. 360)

(603) 742-2346 or (603) 749-6309

[email protected]

Caution: To protect yourself against in-transit loss, we recommend you insure your returned material.

NOTE: All customers must obtain a CSA# before returning any instrument.

- 10 -

v14 - 09/06

MODEL 6240 SERVICE/CALIBRATION MANUAL Cat. #2129.80

Chauvin Arnoux

®

. Inc. d.b.a. AEMC

®

Instruments

15 Faraday Drive • Dover, NH 03820 USA

Phone: (603) 749-6434 • (800) 945-2362 • Fax: (603) 742-2346

www.aemc.com