HAZ-DUST IV Personal Monitor Cat. Nos. 770-4004 and

HAZ-DUST IV Personal Monitor Cat. Nos. 770-4004 and
• Immediate display of particle concentration in mg/m3
– TWA, STEL, Max, and Min
– Data is stored for downloading
• Interchangeable impactors
– Inhalable impactor adapts to IOM Sampler
– Thoracic impactor (included)
– Respirable impactor adapts to GS-3 Cyclone
• Operate without an impactor for total dust measurement by NIOSH Method 0500
• Internal adjustable sample pump — 1.0 to 3.3 L/min
• In-line cassette directly behind sensor for concurrent filter sampling to
correct data for local dust
• Audible alarm with adjustable threshold
• Small and lightweight (1.5 lbs) with miniature sensor
• DustComm Pro Software for downloading, basic trend analysis,
and comprehensive graphical reporting
• Run time over eight hours with rechargeable NiMH battery
• Highly accurate, NIST-traceable calibration
– ± 10% to filter gravimetric SAE fine test dust
The HAZ-DUST IV, with new infrared detector, displays immediate breathing zone measurements and comprehensive graph reporting
of aerosol and dust for industrial hygiene and environmental air investigations. Operate without an impactor for total dust measurement
(NIOSH 0500) or target specific particle fractions by installing an inhalable, thoracic, or respirable impactor. An in-line 37-mm cassette behind
the sensor may be used to collect a concurrent filter sample for gravimetric or chemical analysis.
Easy-to-use HAZ-DUST IV
Four simple keys and menu-driven instructions displayed on an easy-to-read LCD make HAZ-DUST IV operation easy. Simply select from
the menus the desired particle type and connect the appropriate impactor to the sensor. Place a filter in the filter cassette behind the sensor,
zero the unit, and clip the unit and sensor to a worker’s clothing. Use the four-button keypad to set the threshold alarm and begin monitoring.
SKC Inc. 724-941-9701
SKC-West 714-992-2780
SKC Gulf Coast 281-859-8050
www.skcinc.com
SKC South 434-352-7149
Publication 1685 Rev 1702
Personal Monitor with Concurrent Filter Sample
HAZ-DUST IV Impactors
Performance Profile
• Inhalable — SKC IOM Sampler
Accuracy:
(Cat. No. 770-115A)
Sensing Range:
Resolution:
Particle Size
Range:
Precision:
Calibration:
• Thoracic — impactor included
with the HAZ-DUST IV
• Respirable — SKC GS-3 Cyclone
(Cat. No. 225-103) with adapter
(Cat. No. 770-313)
HAZ-DUST IV with GS-3 Cyclone for real-time
personal respirable silica monitoring
• The only real-time personal dust monitor with both a sensor and a
cyclone/filter cassette in the breathing zone
• Concentrations over user-defined target levels can be further
± 10% to filter gravimetric
SAE fine test dust
0.01 to 200 mg/m3
0.01 mg/m3
0.1 to 100 μm
± 0.02 mg/m3
NIOSH 0600 gravimetric reference NIST-traceable SAE fine test dust
Sample Flow Rate: 1.0 to 3.3 L/min
Power:
Rechargeable NiMH battery
Operating Time: ≥ 8 hrs
Charging Time: 10 to 12 hrs
Data Storage:
21,500 data points
Digital Output: RS-232
Dimensions:
Case: 5.5 x 3.25 x 2.75 in (14 x 8.3 x 7 cm)
Sensor: 1.75 x 1.5 in (4.4 x 3.8 cm)
Weight:
1.5 lbs (0.68 kg)
characterized and quantified by laboratory analysis of the filter.
• Real-time detection can also be used to determine and then
mitigate peak exposures, which will lower TWA levels as well.
Principle of Operation
HAZ-DUST IV operates on the principle of near-forward light
scattering of infrared radiation. It immediately and continuously
measures the concentration of airborne dust particles and displays the
result in mg/m3.
What is near-forward light scattering? This measurement technique
uses an infrared light source positioned at a 90-degree angle from a
photodetector. As the airborne particles enter the infrared beam, they
scatter the light. The amount of light received by the photodetector is directly
proportional to the aerosol concentration. Unique signal processing
internally compensates for noise and drift.
Applications
• Determining levels of worker respirator protection for
OSHA compliance
• Reviewing compliance programs
• Evaluating work practices and controls of any
dust-generating practice
• Performing safety audits
• Monitoring welding fume exposures
• Performing air quality studies in occupational health
and industrial hygiene
Description
HAZ-DUST IV Monitor includes monitor, thoracic impactor,
in-line filter cassette, carry case, computer cable, battery
charger, HAZ-DUST Media CD with instruction manual, and
DustComm Pro Software
110-240 V
HAZ-DUST IV Monitor Kit includes all items listed above
plus GS Cyclone and adapter, IOM and adapter, calibration
jar, and standard
110-240 V
Impactors
Inhalable Impactor and Adapter, IOM Sampler, mounts on
HAZ-DUST IV sensor
Thoracic Impactor, mounts on inlet or HAZ-DUST IV
sensor
Respirable Impactor, GS-3 Cyclone, mounts on HAZDUST IV sensor
Adapter for Respirable Impactor, required when using
GS-3 Cyclone
Accessories
Calibration Standard for Monitor, for verifying span and
optical sensor performance
Calibration Jar, for calibrating and setting pump flow when
using respirable impactor
Zeroing Accessory, for clean-air zeroing of HAZ-DUST IV
sensor when using inhalable impactor
Zeroing Filter, for use with respirable and thoracic
impactors
Replacement Battery Pack, NiMH
Charger, 110-240 V
Cat. No.
770-4004
770-4004K
770-115A
770-4107
225-103
770-313
770-140
225-112
770-4202A
770-4102
770-4105
770-319
Requires calibration with equipment sold separately, see accessories
• Performing hazmat air quality investigations and
monitoring waste site remediation
• All air monitoring applications involving lung-damaging particulates
• Measuring total dust according to NIOSH 0500
SKC Inc. 724-941-9701
Ordering Information
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return
Policy, which provides SKC’s sole liability and the buyer’s exclusive
remedy. To view the complete SKC Limited Warranty and Return
Policy, go to http://www.skcinc.com/warranty.asp.
SKC-West 714-992-2780
SKC Gulf Coast 281-859-8050
www.skcinc.com