GL240 Data Logger - Fieldworks Direct

GL240 Data Logger - Fieldworks Direct
GL840-M and GL840-WV Data Loggers
99 20 analog input channels, expandable to 200
99 Programmable per channel
99 ±20 mV to ±100 V over 12 ranges
99 Supports direct-connected thermocouples of any type
99 Supports direct-connected PT100/PT1000 RTDs
99 Full isolation per channel (60 V GL840-M; 300 V GL840-WV)
99 4 discrete input channels
99 Programmable as a group as logic or pulse inputs
99 Pulse inputs support counter or frequency inputs
99 4 discrete alarm outputs
99 Optional WiFi wireless operation
99 Flexible triggering
99 Built-in, 7-inch color display
99 Built-in Web server operation for remote operations
99 Removable SD memory support up to 32 GB capacity
99 Operates either stand-alone or
PC-connected.
99 PC-side software included
GL840 Description
Model GL840 Series is a third-generation data logger product with exceptional price/performance. It’s a 20 analog channel device expandable to 200 channels, augmented by four discrete inputs and outputs. Its discrete inputs can be configured as a group to be either logic inputs or pulse inputs. When configured for pulse, each of the four channels can be
configured to measure frequency or to count. The four discrete outputs are alarms that can be triggered by a variety of easily-defined analog and pulse/discrete input channel conditions. The 20 GL840 analog input channels may each be configured to measure a direct connected voltage in the range of 20 mV to 100 V full scale across 12 ranges, a direct-connected
thermocouple of any type, or a 3-wire PT100 or PT1000 RTD. Each of the GL840’s analog input channels is electrically
isolated from other channels and from power ground allowing off-ground measurements using shunts, as well as powered
or grounded thermocouples. Further, two GL840 versions are available to tolerate up to a 60 or 300 Vp-p common mode
voltage. Model GL840-M supports up to 60 Vp-p common mode. Model GL840-WV supports up to 300 Vp-p common
mode to address more aggressive applications, like stacked battery cell measurements for both temperature and voltage.
The most powerful GL840 feature is its triggering flexibility. Data recording can be independently started or stopped as a
function of analog and pulse/discrete signal level (single or windowed), alarm, date and time, and day-of-the-week. Triggers can also be configured to operate only once, or to automatically repeat. The ability of the GL840 to adapt to virtually
any desired trigger condition allows the instrument to operate unattended for long periods of time with complete autonomy.
The GL840 operates either connected to a PC or entirely stand-alone. Connection to a PC may be over a standard Ethernet, USB, or optional wireless connection. Each connection approach may take advantage of supplied PC-side software
to configure, acquire, display, and record digitized information for storage directly to the PC’s HDD. Acquired data may
be retrieved for review and analysis after recording, including the ability to export to Microsoft Excel. Using the GL840’s
built-in Ethernet port or wireless option enables the instrument’s networking features, allowing it to be remotely configured and managed using the standard Web browser of any computer or smart phone. Automatic backup of data acquired to
a user-accessible SD memory card is also supported via its FTP facility.
Finally, the measurement reach of the GL840 is further extended by use of GS Sensor accessories that can be either connected directly to the GL840 or wirelessly accessed with the B-568 wireless option using the GL100-WL. Extended measurements of temperature/humidity, acceleration, CO2, illuminance, ac current, and more are available using GS Sensors.
GL840 Close Up
330-668-14442
www.dataq.com
GL840 External Dimensions
Dimension: mm
Precision: ±5 mm
www.dataq.com3
330-668-1444
GL840 Analog Input Circuit and Measurement Ranges
Each of the twenty GL840 analog input channels offers isolation between them and ground. That means that a potential
difference in the ground of one or more channels relative to each other, or relative to the power ground of the GL840 have
little or no effect on measurements when used within spec. The isolation feature provides a tremendous advantage in terms
of noise immunity while making typical measurements, and extends the reach of the instrument to include those that can
only be made with an isolated configuration, like current shunts and powered and grounded thermocouples.
Analog Input Channel Circuit Diagram (typical per channel)
Analog Connections and Measurement Range (typical per channel)
Item
Description
Input configuration
Isolated input, scanning
Measurement range
20, 50, 100, 200, 500 mV/F.S.; 1, 2, 5, 10, 20, 50, 100 V/F.S.; 1-5V
Thermocouples
K, J, E, T, R, S, B, N, W (WRe 5-26)
RTDs
PT100/PT1000 (3-wire)
A/D resolution
16-bit (Effective resolution: Approx. 1/40,000 of the +/- range)
Filter
Off, 2, 5, 10, 20, 40
Filter operation is on a moving average basis.
The average value of the set sampling count is used.
If the sample interval exceeds 5 seconds, the average value of data
obtained in a sub-sample (5 seconds) is used.
RTD input
Thermocouple input
DC voltage input
4-20 mA process current input
330-668-14444
www.dataq.com
GL840 Analog Channel Expansion
The included base analog channel count of 20 channels can be expanded in 20-channel increments to a total of 200, and
the GL840 supports various options to accomplish this. At a minimum you’ll need one of two types of Terminal Units, an
Extension Terminal Base Unit, and at least one of two cables. The expansion process begins by removing the GL840s included Terminal Unit. Next, the removed Terminal Unit is attached to a Extension Terminal Base Unit and paired with one
or more other assemblies depending upon the desired analog channel count. The assemblies are cabled back to the GL840
to complete the expansion.
Item
Description
Model B-564 Standard Terminal Unit. This is the 20-channel Terminal Unit that ships with model GL840-M. It supports a
maximum common mode voltage of 60 Vp-p, and a typical voltage and temperature accuracy of ±0.1% FS and ±1.55°C
respectively.*
Model B-565 High CMV Terminal Unit. This is the 20-channel Terminal Unit that ships with model GL840-WV. It supports a
maximum common mode voltage of 300 Vp-p, and a typical voltage and temperature accuracy of ±(0.05% FS + 10 µV) and
±1.1°C respectively.*
Model B-566 Extension Terminal Base Unit. Provided with a joiner plate and screws to mechanically attach it to others, the
Extension Terminal Base Unit is required to externally mount either the B-564 or B-565 Terminal Units.
Model B-567-05 or B-567-20 Connection Cable. This connection cable is offered in either 50 cm (-05) or 2 m (-20) lengths
and connects the B-566 to the GL840 mainframe or to other B-566 units.
* The standard and high CMV terminals may be combined, but the CMV specification of the standard terminal is applied to all channels.
GL840 Expansion Steps
1. Remove exisiting Terminal Unit
2. Attach Terminal to Base unit
3. Attach Connection Cable to Base unit
4. Join more expanders (10 units max.)
5. Connect to the GL840
6. Add cables as desired (20m total max.)
www.dataq.com
5
330-668-1444
GL840 Discrete I/O and Pulse Inputs
Depending upon the application, discrete I/O and pulse inputs can play a crucial data logging role. The GL840 supports
four discrete alarm outputs that can signal alarm or event states that are a function of virtually any combinations of analog
and pulse or discrete input values. These alarm outputs may be used to handshake with a PLC or other devices to start or
stop processes or simply signal the beginning or end of events. The GL840 also offers four discrete input ports, which may
be configured as simple binary true/false input flags, or for pulse and counter inputs. Pulse inputs can be used to acquire
frequency data such as rpm or flow, or reconfigured to acquire count data to derive volume from flow or simply count the
number of iterations from a process. Pulse data is neatly folded into acquired analog data so that all measured parameters
can be evaluated in the same timeframe during analysis to easily identify cause and effect. A final discrete input is reserved for externally triggering the GL840’s A-D conversion to allow the instrument to synchronize to external processes.
Access to all discrete I/O requires the B-513 cable option.
Logic/Pulse Input Specifications
Item
Description
Number of input channels
4 (switch between logic and pulse)
Input voltage range
0 to +24V max. (single-ended ground input)
Threshold level
Approx. +2.5V
Hysteresis
Approx. 0.5 V (+2.5 to +3 V)
Trigger Input/External Sampling Input Specifications
Item
Description
Number of input channels
1
Input voltage range
0 to +24V max. (single-ended ground input)
Threshold level
Approx. +2.5V
Hysteresis
Approx. 0.5 V (+2.5 to +3 V)
Alarm Output Specifications
Item
Description
Number of output channels
4
Output format
Open collector output
+5 V, 10 KΩ pull-up resistance
Discrete I/O Instrument-side Equivalent Circuits
330-668-14446
www.dataq.com
SD Memory Card and Wireless Option Access
The GL840 supports two SD memory slots. SD Card 1 is provided with a 4 GB SD memory for data recording. Any
SDHC memory card can be used to a 32 GB capacity. SD Card 2 can be populated with a secondary SD memory card,
or with the B-568 wireless option. Model B-568 enables the networking features of the GL840, and its installation in a
secondary SD slot without affecting the primary SD slot does not compromise the GL840’s storage capacity. The wireless option supports IEEE 802.11b/g/n with the following security protocols: WEP64, WEP128, WPA-PSK/WPA2-PSK,
AKIP/AES. The B-568 wireless option may be configured as either an access point to allow peer-to-peer communication
(such as with the GL100-WL data logger option), or as a router-managed device on a LAN.
SD Memory and B-568 Wireless Access
Inserting the SD memory into SD Card1 Slot
SD Card2 Slot and Wireless Installation
Programmable Sampling Interval Speed versus Measurement
Interval
10mS
20mS
50mS
100mS
200mS
500mS
1S
2S
>2S
Number of channels
1
2
5
10
20
50
100
200
200
Measurement
Voltage
Temperature
Yes
No
Yes
(Chart applies when the captured data format is binary. Limited sampling speed when GL100-WL sensors are in use. )
www.dataq.com
7
330-668-1444
B-568 Wireless Option Networking Modes
Use the B-568’s access point mode to network the GL840 with the optional GL100-WL (up to 5 units) to expand measurement flexibility, or to provide direct peer-to-peer access to a PC and even a smart phone. Alternatively, the B-568’s router
mode neatly folds the GL840 into an existing LAN. The B-568 wireless option or hard-wired Ethernet enables an entire
upper level of GL840 performance in terms of FTP backup, Web server operation, and email notification. Web server
mode supports all popular Web browsers and allows remote operation of the GL840 and real time screen monitoring.
As a Wireless Access Point
As a Router-managed Device
E-mail Configurations
Selection item
E-mail address
Notification
Description
TO:
Set the e-mail address of the e-mail destination. (Up to 63 characters)
CC1: to CC3:
Up to three e-mail addresses can be set as CC (carbon copy). (Up to 63 characters)
Subject:
The e-mail subject. (Up to 63 characters)
Alarm
When it is set to On, the occurrence of alarm is notified.
Low battery
When it is set to On, the low battery information is notified.
Low communication strength
When it is set to On, the low communication strength information is notified.
SD memory card free space
When it is set to On, the SD memory card free space information is notified.
Periodic notification
Set the time to send the notification setting information with the e-mail to any address.
330-668-14448
www.dataq.com
Connecting the GL840 Directly To A PC
Don’t need a wireless or hard-wired LAN? No problem. The GL840’s integral USB port allows it to connect directly to a
PC on to which is typically installed the included Graphtec APS software for real time data acquisition.
Ethernet Connection
Use a crossover cable when connecting the GL840 directly to a PC. Otherwise, a standard Ethernet cable can be used.
Internal Battery Pack Option B-569
The GL840 supports one or two battery packs, option (B-569), that allows operation independent of ac power. A fully
charged battery allows an operating time of 5 hours (one battery pack) to 10 hours (two battery packs) depending upon
data logging configuration. The battery can also operate as a UPS (uninterruptable power supply.) In the event that ac
power is lost, the GL840 will seamlessly continue operation on battery power without data logging disruption.
www.dataq.com9
330-668-1444
GL840 Display Close Up
The GL840’s 7-inch display shows acquired analog and discrete/pulse data in real time as it is acquired. Values can be
scaled into meaningful engineering units, and one of three selectable modes can displayed with a simple button push.
Waveform and digital screen
Expanded waveform screen
Digital display screen
330-668-144410www.dataq.com
GL840 Analog Measurement Modes
The GL840 in tandem with the optional GL100-WL wireless add-on or its GS Sensor accessories is capable of a remarkable range of measurements. Measurements that the GL840 can make on its own are voltage, 4-20 mA process current
(with optional R250 shunt resistor), thermocouple, RTD, and humidity (with optional B-530 sensor.) When a GL100-WL
is added to the mix (requires the optional B-568 wireless interface), the GL840 has wireless access to an additional range
of measurements. Without the wireless option or the GL100-WL, the GL840 directly accepts GS Series add-on modules.
Direct GL840 Measurements
Function
Analog
Discrete
Channels
20 (expandable
to 200)
4
Measurements
Comments
Voltage
±20 mV to ±100 V
Thermocouple
Types K, J, E, T, R, S, B, N, W (WRe 5-26)
4-20 mA
Requires R250, 250-Ohm current shunt
Humidity
Requires B-530 humidity sensor option
RTD
3-wire PT100, PT1000
Logic
True/false
Pulse
Count, instantaneous count, revolution
GL100-WL Add-on Modules (refer to the GL100 Accessories datasheet for details)
Model
Channels
Measurements
Alarms
GL100-WL
-
-
GS-TH
4
Ambient temperature and RH
1*
-
GS-3AT
4
Temp + 3-axis acceleration
-
GS-4VT
4
Thermocouple/voltage
-
GS-4TSR
4
Thermistor
-
GS-LXUV
4
Illuminance / Ultraviolet
-
GS-CO2
1
CO2
-
GS-DPA-AC
4
AC current sensor (1 or 3 phase)
-
GS-DPA
-
Dual branch adator
-
8
GS-TH + GS-LXUV
-
5
GS-TH + GS-CO2
-
5
GS-CO2 + GS-LXUV
-
Comments
Provides wireless access to the GL240
Programmable per channel. Supports K and J TCs
Used to combine (pair) the indicated modules
* GS modules support a level alarm function that is detected by the GL100-WL and communicated to a wireless-equipped GL840.
Use GS Series Add-ons Wirelessly with a GL100-WL, or Directly Connected
www.dataq.com11
330-668-1444
GL840 Global Device Measurement Settings
The GL840 allows an array of settings that define how all of its channel information is acquired. The following table provides a overview of the major setting categories and selections within them:
Measurements
Comments
Sampling
10, 20, 50, 100, 125, 200, 250, 500ms, 1, 2, 5, 10, 20, 30s, 1, 2, 5, 10, 20, 30min, 1h; External
Capture destination
SD CARD 1, SD CARD 2
File Name
Ring/Relay capture
Ring capture
AC Line Filter
Backup
Name of the recorded data file
Off, Ring, Relay
Number of recording points
Off, On
Backup Interval
Off, 1, 2, 6, 12, 24 hours
Backup Destination
SD CARD 1 (SD1), SD CARD 2 (SD2), FTP
Save Folder
Folder name
Calc. Settings 1
Off, Average, Max, Min, Peak, RMS
Calc. Settings 2
Off, Average, Max, Min, Peak, RMS
“Ring Capture” and “Relay Capture” Explained
In addition to continuous data logging, the GL840 supports two special-purpose recording modes: Ring and Relay Capture.
The Ring Capture feature allows the GL840 to acquire a definable number of data values to consecutive data files while
automatically deleting the oldest file. For example, if 1,000 data points are specified, data is acquired to File 1 until 1,000
data values have been recorded. Recording then seamlessly continues to File 2 for another 1,000 values. Before data
recording continues to File 3 for another 1,000 data values, File 1 is deleted. When File 3 is full, File 2 is deleted and recording continues to File 4. This process continues until recording stops. In this manner Ring Capture allows data logging
to continue indefinitely without concern for filling the target memory. Further, since the number of data values recorded
to each file and sampling interval are definable and constant, the timeframe before data is erased is precisely known in
advance. Thus, all critical data leading up to, during, and after an event can be captured for analysis. Maximum file size is
2 GB, but SD memory sizes as large as 32 GB are supported.
The Relay Capture feature of the GL840 is almost identical to Ring Capture, except that data files are never deleted. The
feature essentially exchanges unlimited record time for an entire history of recorded data. Like Ring Capture, maximum
file size is 2 GB, but SD memory sizes as large as 32 GB are supported.
“External Sampling” Explained
Sometimes asynchronous sampling rates just won’t do. If you need to acquire data at a precise moment that’s coincident
with an event, and you can generate a trigger signal for that occurrence, External sampling can be used. The following diagram describes the timing relationship between the various components that define an externally triggered application.
330-668-144412www.dataq.com
Start or Stop Recording on Any Trigger Condition
The GL840’s range of stop and start trigger conditions is massive and unrivaled. Select from single or windowed levels to
the day and time of the week with everything in between per pulse or analog channel. Want to start or stop acquiring data
when the signal level on channel 1 is above 200 psi, but only on Saturday at 12 noon? No problem. Finally, select Boolean
AND/OR operators to tie any variety of trigger conditions together.
Finally, trigger conditions can be independently set for the GL840’s alarm output ports as a function of virtually any combination of analog or pulse input values. The following table describes the array of GL840 trigger conditions to stop, start,
or alarm the instrument:
Setting
Start Side Source Setting
[Level]
[Alarm]
[Date]
[Weekly]
Off, Level, Alarm, External Input, Date, Weekly, Time
Mode
Analog: Off, H, L, Window In, Window Out
Logic: Off, H, L
Pulse: Off, H, L, Window In, Window Out
Combination
Level OR, Level AND, Edge OR, Edge AND
Level
Set numeric value
Alarm port number
1, 2, 3, 4, WL1
Date
From 2005.1.1 to 2035.12.31
Time
From 0:0:0 to 23:59:59
Day of week
Off or On setting for each of Sunday through Saturday
Time
From 0:0:0 to 23:59:59
[Time]
From 0:0:1 to 9999:59:59
Stop Side Source Setting
[Level]
[Alarm]
[Date]
[Weekly]
Off, Level, Alarm, External Input, Date, Weekly, Time
Mode
Analog: Off, H, L, Window In, Window Out
Logic: Off, H, L
Pulse: Off, H, L, Window In, Window Out
Combination
Level OR, Level AND, Edge OR, Edge AND
Level
Set numeric value
Alarm port number
1, 2, 3, 4, WL1
Date
From 2005.1.1 to 2035.12.31
Time
From 0:0:0 to 23:59:59
Day of week
Off or On setting for each of Sunday through Saturday
Time
From 0:0:0 to 23:59:59
[Time]
Repeated Capturing
Alarm Level
Settings
Selections available
From 0:0:1 to 9999:59:59
Off, On
Mode
Analog: Off, H, L, Window In, Window Out
Logic: Off, H, L
Pulse: Off, H, L, Window In, Window Out
Level
Set numeric value
Output
1, 2, 3, 4, WL1
Detection Method
Level, Edge
Alarm Hold
Held or Not held
Send Burnout Alarm
Sent or not sent
www.dataq.com13
330-668-1444
Trigger Operations Close Up
Trigger and Alarm Operations
Rising
Falling
Window-in
Window-out
General Record Time
The following record time table assumes analog channels only with Logic/Pulse inputs disabled. Figures are approximate.
File size of captured data is 2GB in GBD or CSV file format. Sampling interval is limited by the number of channels in
use.
Sampling Interval (10 acquired channels)
Storage Format
10mS
50mS
100mS
200mS
500mS
1S
10S
Binary
31 days
77 days
95 days
108 days
270 days
>365 days
>365 days
CSV
3 days
11 days
16 days
21 days
54 days
109 days
>365 days
330-668-144414www.dataq.com
GL840 Specifications
Overall Specifications
Number of analog inputs:
Analog terminal unit type:
GS sensor and terminal/
adapter connection
terminal:
Data backup functions:
Clock accuracy:
(23°C environment)
Operating environment:
Withstand voltage:
Power supply:
Power Consumption:
Maximum 200ch available for 20ch/1 terminal or
extension unit
Standard terminal or High CMV, High-precision
terminal unit (B-565)
One module
Optional (GS sensor and terminal / adapter ) connections only
Setup conditions: EEPROM; Clock: Lithium
battery
±0.002% (accurate within about 50 seconds per
month)
0 to 45°C, 5 to 85% RH
(0 to 40°C when operated in batteries/15 to 35°C
when battery is charging)
Standard terminal: Between each input ch and
GND terminal: 350Vp-p 1 minute; Between each
input terminals: 350Vp-p 1 minute
High CMV, High-precision terminal unit (B565): Between each input ch and GND terminal:
2300VACrms 1 minute; Between each input
terminals: 600Vp-p
AC adapter : 100 to 240 VAC, 50 to 60 Hz
DC input : 8.5 to 24 VDC (26.4 V max.)
Battery pack (option) : 7.2 VDC (2900 mAh), two
packs can be mounted
AC Power consumption (when AC adapter is
used)
Condition
LCD on
Screensaver on
Normal
Consumption
24 VA
19 VA
Consumption during
battery recharge
38 VA
33 VA
DC Power consumption
DC
Voltage
+24V
+24V
+12V
+12V
+8.5V
+8.5V
Condition
LCD on
Screensaver on
LCD on
Screensaver on
LCD on
Screensaver on
Normal
Consumption during
Consumption
battery recharge
0.36 A
0.65 A
0.27 A
0.56 A
0.70 A
Can’t Recharge
0.50 A
Can’t Recharge
1.00 A
Can’t Recharge
0.70 A
Can’t Recharge
*Set the LCD to “Bright” as normal condition.
External Dimensions: Standard terminal: 240×158×52.5mm*
High CMV, High-precision terminal unit (B-565):
240×166×52.5mm*
*Not including protruding parts
Weight: Standard terminal: 1010g*
High CMV, High-precision terminal unit (B-565):
1035g*
*AC adapter and battery are not included, but one
terminal unit is included
Vibration-tested condi- Equivalent to automobile parts Type 1 Category A
tions: classification
Memory devices
Memory capacity: SD CARD Slot: 2 (Compatible with SDHC, up to
approx. 32GB memory available)
• Approx. 4GB SD memory card included
• Possible to save up to 2GB for one file
Memory contents: Setup conditions, Measured data, Screen copy
PC I/F
Interface types: Ethernet (10BASE-T/100BASE-TX); USB 2.0;
Wireless LAN (Option)
Functions: Data transfer to the PC (realtime, SD memory card
data); PC control of the GL840; Control of wireless sensor GL100GL100-WL, Data capture (when
connected to the wireless LAN: up to 5 units)
330-668-1444
15
Ethernet functions:
(10BASE-T/100BASE-TX)
Web server functions: Displays the screen images
FTP server function: Transfer and delete the captured data in the SD memory card.
FTP client function: Back up the captured data to
the FTP server.
NTP client function: Time-synchronize the NTP
server.
DHCP client function: IP address automatic
acquisition
DHCP client function : Automatically retrieves the
IP address
E-mail function: Send and receive the e-mail
USB functions: USB drive mode: Transfer and delete the captured
data in the SD memory card.
Realtime data transfer 10 ms/1 ch maximum (dependant on number of
speed: channels).
Monitor
Display:
Displayed languages:
Backlight life:
Backlight:
7-inch TFT color LCD (WVGA: 800 × 480 dots)
Japanese, English, French, German, Chinese,
Korean, Russian, Spanish
50,000 hrs (until the brightness is reduced to
50%), varies with operating environment.
Screen saver function provided (10, 30 sec., 1, 2,
5, 10, 30, 60 min.)
Input Unit Specifications for GL840-M (with B-564)
Number of input channels:
Input terminal type:
Input method:
Scan speed:
Measurement ranges:
Measurement accuracy:
23°C ±5°C; When 30 minutes or more have elapsed
after power was switched
on; Sampling 1 s/10 ch;
Filter ON (10); GND connected
20ch (200ch available when used with the extension terminal base). Possible to direct-connect or
connect with the extension terminal connection
cable (sold separately) between the GL840 and
terminal unit, or between terminal units.
M3 screw type terminals (Rectangular flat washer)
Photo MOS relay scanning system
All channels isolated, balanced input
Terminal b to be used to connect the resistance
temperature detector is shorted within all channels.
10 ms/1 ch maximum
Voltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20,
50, 100 V; 1-5 V F.S.
Thermocouples : K, J, E, T, R, S, B, N, W (WRe5-26)
Resistance temperature detector: Pt100, JPt100,
Pt1000 (IEC751)
Temperature range : 100°C, 500°C, 2000°C
(150°F, 750°F, 3000°F)
Humidity: 0 to 100% (voltage 0 to 1 V scaling
conversion) fixed (requires B-530)
Voltage: 0.1% of F.S.
Temperature Thermocouple:
Measurement Temperature
Range (°C)
0 ≤ TS ≤ 100
100 < TS ≤ 300
R/S
R : 300 < TS ≤ 1600
S : 300 < TS ≤ 1760
400 ≤ TS ≤ 600
B
600 < TS ≤ 1820
-200 ≤ TS ≤ -100
K
-100 < TS ≤ 1370
-200 ≤ TS ≤ -100
E
-100 < TS ≤ 800
-200 ≤ TS ≤ -100
T
-100 < TS ≤ 400
-200 ≤ TS ≤ -100
J
-100 < TS ≤ 100
100 < TS ≤ 1100
-200 ≤ TS < 0
N
0 ≤ TS ≤ 1300
W
0 ≤ TS ≤ 2000
Reference contact compensation
accuracy
TC
Measurement
Accuracy (°C)
±5.2
±3.0
± (0.05% of rdg +2.0)
± (0.05% of rdg +2.0)
±3.5
± (0.05% of rdg +2.0)
± (0.05% of rdg +2.0)
± (0.05% of rdg +1.0)
± (0.05% of rdg +2.0)
± (0.05% of rdg +1.0)
± (0.1% o f rdg +1.5)
± (0.1% o f rdg +0.5)
±2.7
±1.7
± (0.05% of rdg +1.0)
± (0.1% o f rdg +2.0)
± (0.1% o f rdg +1.0)
± (0.1% o f rdg +1.5)
±0.5
* Thermocouple diameters T, K: 0.32 φ, others: 0.65 φ
www.dataq.com
GL840 Specifications
Measurement Accuracy
(cont.):
Resistance temperature detector:
Measurement Temperature Range (°C)
Pt100 -200 to 850 (FS=1050)
JPt100 -200 to 500 (FS=700)
Pt1000 -200 to 500 (FS=700)
Type
Applied
Current
1mA
1mA
0.3mA
Measurement
Accuracy (°C)
±1.0
±0.8
±0.8
(cont.)
Scan speed:
Measurement ranges:
*3-wire system
Temperature Range:
Type
R/S
B
Temperature
Resolution
Range (°C F.S.)
100
0.01
500
0.05
2000
0.1
500
2000
100
0.05
0.1
0.01
500
0.05
2000
0.1
100
500
2000
100
500
0.01
0.05
0.1
0.01
0.05
2000
0.1
K/E/T/J/N
W
Pt
Measurement
Range (°C)
0 to 100
0 to 500
R: 0 to 1600
S: 0 to 1760
400 to 500
500 to 1820
-100 to 100
K/E/J/N: -200 to 500
T : -200 to 400
K : -200 to 1370
E : -200 to 800
T : -200 to 400
J : -200 to 1100
N : -200 to 2000
0 to 100
0 to 500
0 to 2000
-100 to 100
-200 to 500
Pt100: -200 to 850
JPt100/Pt1000:
-200 to 500
*Measurement accuracy does not change due to the
temperature range
Reference contact compensation accuracy:
A/D converter:
Temperature coefficient:
Input resistance:
Allowable signal source
resistance:
Maximum permissible
input voltage:
Withstand voltage:
Insulation resistance:
Common mode rejection
ratio:
Noise:
Filter:
Internal/External switching
Method:ΔΣ method; Resolution:16-bit (Effective
resolution: About 1/40,000 of the +/- range)
Gain : 0.01% of F.S./°C; Zero : 0.02% of F.S./°C
(Occurs when sampling speed is 10, 20, or 50 ms.)
1 MΩ ±5%
Within 300Ω
Between +/– input terminals :20mV to 2V range
(60Vp-p); 5V to 100V range (110Vp-p)
Between input terminal/input terminal :60 Vp-p
Between input terminal/GND :60 Vp-p
Between input terminal/input terminal : 350 Vp-p
1 minute
Between input terminal/GND : 350 Vp-p 1 minute
Between input terminal/GND : 50MΩ or more (at
500 VDC)
90 dB or more (50/60 Hz; signal source 300Ω or
less)
48 dB or more (with +/– terminals shorted)
Off, 2, 5, 10, 20, 40
Filter operation is on a moving average basis.
The average value of the number of set samples is
used. If the sample interval exceeds 30 seconds,
the average value of data obtained in a sub-sample
(30 seconds) is used.
Input Unit Specifications for GL840-WV (with B-565)
Number of input channels:
Measurement accuracy:
23°C ±5°C; When 30 minutes or more have elapsed
after power was switched
on; Sampling 1 s/10 ch;
Filter ON (10); GND connected
10 ms/1 ch maximum
Voltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20,
50, 100 V; 1-5 V F.S.
Thermocouples : K, J, E, T, R, S, B, N, W (WRe5-26)
Resistance temperature detector: Pt100, JPt100,
Pt1000 (IEC751)
Temperature range : 100°C, 500°C, 2000°C
(150°F, 750°F, 3000°F)
Humidity: 0 to 100% (voltage 0 to 1 V scaling
conversion) fixed (requires B-530)
Voltage: ± (0.05% of F.S. + 10μV)
Temperature Thermocouple:
Measurement Temperature
Range (°C)
0 ≤ TS ≤ 100
100 < TS ≤ 300
R/S
R : 300 < TS ≤ 1600
S : 300 < TS ≤ 1760
400 ≤ TS ≤ 600
B
600 < TS ≤ 1820
-200 ≤ TS ≤ -100
K
-100 < TS ≤ 1370
-200 ≤ TS ≤ -100
E
-100 < TS ≤ 800
-200 ≤ TS ≤ -100
T
-100 < TS ≤ 400
-200 ≤ TS ≤ -100
J
-100 < TS ≤ 100
100 < TS ≤ 1100
-200 ≤ TS < 0
N
0 ≤ TS ≤ 1300
W
0 ≤ TS ≤ 2000
Reference contact compensation
accuracy
Measurement
Accuracy (°C)
±4.5
±3.0
±2.2
±2.2
±3.5
±2.5
±1.5
±0.8
±1.0
±0.8
±1.5
±0.6
±1.0
±0.8
±0.6
±2.2
±1.0
±1.8
TC
±0.3
* Thermocouple diameters T, K: 0.32 φ, others: 0.65 φ
Resistance temperature detector:
Measurement Temperature Range (°C)
-200 ≤ TS ≤ 100
Pt100
100 < TS ≤ 500
500 < TS ≤ 850
-200 ≤ TS ≤ 100
JPt100
100 < TS ≤ 500
-200 ≤ TS ≤ 100
Pt1000
100 < TS ≤ 500
Type
Applied
Current
1mA
1mA
0.3mA
Measurement
Accuracy (°C)
±0.6
±0.8
±1.0
±0.6
±0.8
±0.6
±0.8
*3-wire system
Temperature Range:
Type
R/S
B
Temperature
Resolution
Range (°C F.S.)
100
0.01
500
0.05
2000
0.1
500
2000
100
0.05
0.1
0.01
500
0.05
2000
0.1
100
500
2000
100
500
0.01
0.05
0.1
0.01
0.05
K/E/T/J/N
W
Measurement
Range (°C)
0 to 100
0 to 500
R: 0 to 1600
S: 0 to 1760
400 to 500
500 to 1820
-100 to 100
K/E/J/N: -200
to 500
T : -200 to 400
K : -200 to 1370
E : -200 to 800
T : -200 to 400
J : -200 to 1100
N : -200 to 2000
0 to 100
0 to 500
0 to 2000
-100 to 100
-200 to 500
Pt100: -200
to 850
JPt100/Pt1000:
-200 to 500
20ch (200ch available when used with the extension terminal base). Possible to direct-connect or
Pt
connect with the extension terminal connection
2000
0.1
cable (sold separately) between the GL840 and
terminal unit, or between terminal units.
*Measurement accuracy does not change due to the
Input terminal type: M3 screw type terminals (Rectangular flat washer)
temperature range
Input method: Photo MOS relay scanning system
Reference contact com- Internal/External switching
All channels isolated, balanced input
pensation accuracy:
Terminal b to be used to connect the resistance
temperature detector is shorted within all channels.
330-668-144416www.dataq.com
GL840 Specifications
A/D converter:
Temperature coefficient:
Input resistance:
Allowable signal source
resistance:
Maximum permissible
input voltage:
Withstand voltage:
Insulation resistance:
Common mode rejection
ratio:
Noise:
Filter:
Method:ΔΣ method; Resolution:16-bit (Effective
resolution: About 1/40,000 of the +/- range)
Gain : 0.01% of F.S./°C; Zero : 0.02% of F.S./°C
(Occurs when sampling speed is 10, 20, or 50 ms.)
1 MΩ ±5%
Within 100Ω
Between +/– input terminals :20mV to 2V range
(60Vp-p); 5V to 100V range (110Vp-p)
Between input terminal/input terminal :600 Vp-p
Between input terminal/GND :300 Vp-p
Between input terminal/input terminal : 600 Vp-p
1 minute; Between input terminal/GND : 2300
VACrms 1 minute
Between input terminal/GND : 50MΩ or more (at
500 VDC)
90 dB or more (50/60 Hz; signal source 300Ω or
less)
48 dB or more (with +/– terminals shorted)
Off, 2, 5, 10, 20, 40
Filter operation is on a moving average basis.
The average value of the number of set samples is
used. If the sample interval exceeds 30 seconds,
the average value of data obtained in a sub-sample
(30 seconds) is used.
Function Specifications
Display screen: Waveform + Digital screen, All Waveform screen,
Digital + Calculation Display
screen, Expanded digital screen
* Can be switched using the dedicated key (toggle
operation)
* For the Expanded Digital screen, the number of
channels and the display channel
must be specified
* The waveform is not rewritten due to the change
of the TIME / DIV.
Sampling interval: 10 ms/1 ch maximum (GBD/CSV-formatted)
10, 20, 50, 100, 125, 200, 250, 500 ms; 1, 2, 5,
10, 20, 30 sec.; 1, 2, 5, 10, 20, 30 min.; 1 hour;
External
* The settings of 125 ms or below can be used
depending on the input settings and the measuring
channel.
EU (scaling function): 4 points can be set for each channel
The temperature range scaling function is available.
Functions during cap- Confirmation of the captured data; Saving of data beture: tween cursors; Replacement of the SD memory card
* When the wireless sensor (GL100-WL) is connected,
the sample interval among 10, 20, and 50ms cannot be
replaced during recording.
Data save function: Capture destination: SD memory card (Available both
slot 1 and 2)
Captured data: Settings, Screen data, Measurement data
Capture function: Function: OFF, Ring recording, Relay recording
Ring recording: Number of recording points: 1000 to 2000000
When ring capture is ON, the memory space that can be
used for capture is one-third of the free space.
Relay recording: The data is continuously recorded in 2GB-separated files
without missing data.
Replaying data: GBD/CSV-formatted data file (only data captured in this
GL840)
Calculation between Calculation type: Four arithmetic operations (+, -, ×, ÷)
channels: Target input: Analog CH1 to CH200
GS sensor and terminal/adapter: CH1 to CH8
Wireless sensor: CH1-1 to CH5-8
330-668-1444
17
(cont.)
Statistical calculation: Statistical calculation type: Average value, peak value,
maximum value, minimum value, root mean square
value; Number of calculations: Two arithmetic operations can be set to each channel; Calculation method:
Real-time calculation and specified between cursors
(during replay)
* Real-time calculation results are displayed on the
Digital screen + Calculation Display screen.
Search functions: Function : Search the captured data for the required
number of points
Search type : Channel Pulse, Logic, Level, Alarm search
Annotation input Function : A comment can be entered for each channel
function: Input table characters : Alphanumerics
Number of characters : 31
(The number of characters can be displayed on the
screen is up to eight characters.)
Navigation function: Easy capture measurement, easy trigger setting, wireless
LAN setting functions
Trigger/Alarm Functions
Repeat Trigger: Off, On
Trigger types: Start: Data capture starts when a trigger is generated; Stop: Data capture stops when a trigger is
generated
Trigger conditions: Start: Off, Level, Alarm, External, Time
Stop: Off, Level, Alarm, External, Time
Trigger judgment modes: Combination: Level OR, Level AND, Edge OR,
Edge AND; Analog channel judgment mode: H
(↑), L (↓), Window In, Window Out; Logic channel
judgment mode: H (↑), L (↓); Pulse channel judgment mode: H (↑), L (↓), Window In, Window Out
Alarm judgment modes: Combination: Analog, Logic or “AND” / “OR”
of pulse; Analog judgment: H (↑), L (↓), Window
In, Window Out; Logic judgment: Pattern; Pulse
judgment: H (↑), L (↓), Window In, Window Out
External Input/Output Functions
Input/output types:
Input specifications:
Alarm output specifications:
Pulse input:
Trigger input (1 ch) or External sampling input
(1 ch)
Logic input (4 ch) or Pulse input (4 ch)
Alarm output (4 ch)
* Switch between Logic and Pulse
* Switch between Trigger and External sampling.
* The Input/output cable for GL B-513 (option) is
required to use the external output function.
Input voltage range : 0 to +24 V (single-ended
ground input); Input signal : No-voltage contact (acontact, b-contact, NO, NC), Open collector, Voltage
input; Input threshold voltage : Approx. +2.5 V;
Hysteresis : Approx. 0.5 V (+2.5 to + 3 V)
Output format: Open collector output (5 V, pull-up
resistance 10KΩ))
<Maximum ratings of output transistor>
• Collector-GND voltage : 30 V
• Collector current : 0.5 A
• Collector dissipation : 0.2 W
Output conditions: Level judgment, window judgment, logic pattern judgment, pulse judgment
Revolutions mode (engines, etc.): Counts the number of pulses per sampling interval, and converts
them to RPM. Set the number of pulses per revolution during revolution. Spans : 50, 500, 5000, 50
k, 500 k, 5 M, 50 M, 500 M PRM/F.S.
Counts mode (electric meters, etc.): Counts the
number of pulses for each sampling interval from
the start of measurement. Spans : 50, 500, 5000,
50 k, 500 k, 5 M, 50 M, 500 M C/F.S.
www.dataq.com
Pulse input (continued):
GL840 Specifications
Inst. mode: Counts the number of pulses for each
sampling interval. Resets the count value after
each sampling interval. Spans : 50, 500, 5000, 50
k, 500 k, 5 M, 50 M, 500 M C/F.S.
Maximum input frequency : 50kHz
Maximum number of count : 50kC/sampling (16bit counter)
Connectable sensor, terminal, Number of GS sensors and terminals/adapter conadapter: nection terminals: 1
(Generic name: GS sensor GS-TH: Temperature and humidity sensor
and terminal/adapter) GS-4TSR: 4CH thermistor terminal
GS-LXUV: Illuminance / ultraviolet sensor
GS-DPA: Branch adapter for GS
GS-DPA-AC: Adapter for AC current sensor
GS-CO2: CO2 sensor
GS-3AT: 3-axis acceleration / temperature sensor
GS-4VT: 4CH voltage / temperature terminal
* The items above are optional accessories.
Sampling: 500 ms to 1 hour
* Specific setting is not required. It is synchronized to the setting of the GL840.
(cont.)
Control Software
Compatible OS:
Function:
Number of groups:
Number of CHs per group:
Max number of channels:
Settings:
Statistic/History:
Report function:
Windows8.1/Windows8/Windows7/Windows Vista
Main unit control, realtime data capture, data conversion
4 groups MAX
Up to number of connected module
1000 ch maximun
AMP settings, capture settings, trigger/alarm settings, report settings, others
Realtime data (CSV, GBD Binary)
Data in SD memory card (CSV, GBD binary)
Analog waveforms, logic waveforms, pulse waveforms, digital values
Y-T View, Digital View, X-Y View between Cursors (only during replay)
Between cursors, All data
Alarm monitor enables sending of email to the
specified address
Displays max, min and average values
Enables creation of daily or monthly files
E-mail function:
E-mail sent to specified address on alarm
Captured data:
Display:
Display modes:
File conversion:
Monitor functions:
See ordering information on next page
330-668-144418www.dataq.com
Ordering Guide
Description
Order No.
GL840-M
Compact, lightweight, multi-channel data logger with 20 analog measurement channels (expandable to 200),
20mV to 100V Full Scale measurement range, 4 discrete input channels, and 4 alarm outputs. Includes GL840
data logger, 4 GB flash memory card, AC adapter, software on CD, and an NIST-traceable calibration certificate.
GL840-M
GL840-WV
Compact, lightweight, multi-channel data logger with 20 high common mode voltage analog measurement
channels (expandable to 200), 20mV to 100V Full Scale measurement range, 4 discrete input channels, and 4
alarm outputs. Includes GL840 data logger, 4 GB flash memory card, AC adapter, software on CD, and an NISTtraceable calibration certificate.
GL840-WV
Accessories
Description
Order No.
Standard input terminal
20 channel input terminal.
Extension terminal base
Base unit, connection plate, and screws.
Battery pack
7.2V/2900mAh lithium battery pack.
DC Power Cable
2-meter DC power cable, bare tips.
B-536-US-840
Carrying case.
R250
4-20mA shunt resistor.
GL100-WL
Wireless GS series sensor coupler.
GS-4TSR
GL100 terminal for thermistor temperature.
GS-103AT-4P
GL100 3m thermistor sensor (-40 to 105°C).
GS-103JT-4P
GL100 3m thermistor sensor (-40 to 120°C).
GS-CO2
GL100 Sensor for CO2
GS-LXUV
GL100 Sensor illuminance/UV.
GS-TH
GL100 sensor for ambient temperature/RH.
GS-DPA
GL100 adapter for two sensors.
B-564
B-566
B-569
B-514
B-536US-240
R250
GL100-WL
GS-4TSR
GS-103AT-4P
GS-103JT-4P
GS-CO2
GS-LXUV
GS-TH
GS-DPA
Description
Order No.
High CMV input terminal
B-565
20 channel input terminal for high CMV.
Extension terminal cable
B-567-05
50cm extension connection cable.
Extension terminal cable
B-567-20
2m extension connection cable.
Humidity Sensor
B-530
3-meter with dedicated power connector.
Logic/Alarm Cable
B-513
2-meter logic/alarm cable, bare tips.
Wireless Option
B-568
Wireless communication option. 802.11/b/g/n
Power supply
GLACP
Spare AC power supply.
GS-4VT
GS-4VT
GL100 Voltage/thermocouple terminal.
GS-3AT
GS-3AT
GL100 sensor for acceleration and temperature.
GS-DPA-AC
GL100 adapter for AC power measurements.
GS-DPA-AC
GS-AC50A
GS-AC50A
Dedicated 50A current transformer.
GS-AC100A
Dedicated 100A current transformer.
GS-AC200A
Dedicated 200A current transformer.
GS-AC100A
GS-AC200A
GS-EXC
1.5m extension cable for GL100 sensors.
DATAQ Instruments, Inc.
241 Springside Drive
Akron, Ohio 44333
Phone: 330-668-1444
Fax: 330-666-5434
Data Acquisition Product Links
(click on text to jump to page)
Data Acquisition | Data Logger
DATAQ, the DATAQ logo and WinDaq are registered trademarks of DATAQ Instruments, Inc. All rights reserved. Copyright © DATAQ Instruments, Inc.
The information on this data sheet is subject to change without notice.