1byone TMB-1585 User manual
Add to my manuals14 Pages
1byone TMB-1585 is a digital blood pressure monitor intended for use in measuring blood pressure and heartbeat rate with an arm circumference ranging from about 22 cm to 42 cm. It features a 100mm×68 mm Digital LCD display, can store up to 60 records per user, and utilizes advanced 3rd technology for measuring during inflation. It is suitable for adult indoor use only.
advertisement
User Manual
Arm Blood Pressure Meter
Model : TMB-1585
Arm Type
Distributed by:
1byone Products Inc.
2313 E PHILADELPHIA ST STE M, CA, US, , 917618048 www.1byone.com
Thank you very much for selecting 1byone Arm Blood Pressure Meter.
Please do read the user manual carefully and thoroughtly so as to ensure the safe usage of this product, and keep the manual well for further reference in case you have problems.
CATALOGUE CATALOGUE
Table of Contents
INTRODUCTION...................................................................................................................2
General Description
Indications for Use
Safety Information
LCD Display Signal
Monitor Components
List
BEFORE YOU START...........................................................................................................7
The Choice of Power Supply
Installing and Replacing the Batteries
Measurement Principle
Setting Date, Time and Measurement Unit
Select the User
MEASUREMENT.................................................................................................................12
Tie the Cuff
Start the Measurement
DATA MANAGEMENT.........................................................................................................14
Recall the Records
Delete the Records
INFORMATION FOR USER................................................................................................17
Tips for measurement
Maintenances
ABOUT BLOOD PRESSURE..............................................................................................19
What are systolic pressure and diastolic pressure?
What is the standard blood pressure classification?
Irregular Heartbeat Detector
Why does my blood pressure fluctuate throughout the day?
Why do I get a different blood pressure at home compared to the hospital?
Is the result the same if measuring on the right arm?
TROUBLESHOOTING........................................................................................................21
SPECIFICATIONS..............................................................................................................22
AUTHORIZED COMPONENT ...........................................................................................23
COMPLIED STANDARDS LIST..........................................................................................24
FCC STATEMENT...............................................................................................................25
EMC GUIDANCE................................................................................................................25
1
INTRODUCTION
General Description
This manual contains important safety and care information, and provides step by step instructions for using the product.
Read the manual thoroughly before using the product.
Features:
100mm×68 mm Digital LCD display
Maximum 60 records per each user
3rd technonoly: Measuring during inflation
(The updated technology in the world)
Indications for Use
The 1by one Arm Blood Pressure Meter is digital monitors intended for use in measuring blood pressure and heartbeat rate with arm circumference ranging from about 22 cm to 42 cm ( about 8¾˝-16½˝ ).
It is intended for adult indoor use only.
2
INTRODUCTION
Safety Information
The signs below might be in the user manual, labeling or other component.
They are the requirement of standard and using.
SN
Symbol for “THE OPERATION
GUIDE MUST BE READ”
Symbol for “SERIAL NUMBER”
Symbol for “DIRECT CURRENT”
Symbol for “MANUFACTURE
DATE”
Symbol for “TYPE BF APPLIED
PARTS”
Symbol for “ENVIRONMENT
PROTECTION - Electrical waste products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local authority or retailer for recycling advice”
Caution: These notes must be observed to prevent any damage to the device.
CAUTION
The device is not suitable for measuring the blood pressure of children. Ask your doctor before using it on older children.
The device is not suitable for use on pregnant women, patients with implanted, electrocical devices, patients with pre-elcampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy. Please consult your doctor prior to using the unit if you suffer from illnesses
Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure.
Please keep the unit out of reach of infants, children or pets, since inhalation or swallowing of small parts is dangerous or even fatal.
The device has been evaluated clinically using manual cuff/stethoscope auscultation as the reference.
Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultatory method, within the limits prescribed by the American National Standard, Manual, electronic, or automated sphygmomanometers.”
3
INTRODUCTION
CAUTION
This device is intended only for adult use in homes.
This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than arm or for functions other than obtaining a blood pressure measurement.
If you experience discomfort during a measurement, such as pain in the arm or other complaints, press the START/STOP button to release the air immediately from the cuff. Loosen the cuff and remove it from your arm.
On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately. Prolonged high pressure (cuff pressure >300mmHg or constant pressure
>15mmHg for more than 3 minutes) applied to the arm may lead to an ecchymosis.
Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.
Do not wrap the cuff on the same arm which other monitoring ME EQUIPMENT is applied simultaneously, because this could cause temporary loss of function of those simultaneously-used monitoring ME EQUIPMENT.
Don’t kink the connection tube during use, otherwise, the cuff pressure may continuously increase which can prevent blood flow and result in harmful injury to the PATIENT.
The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with oxygen or nitrous oxide.
This unit is not suitable for continuous monitoring during medical emergencies or operations.
This device cannot be used with HF surgical equipment at the same time.
This device is not intended for patient transports outside a healthcare facility.
To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated interference signal or electrical fast transient/burst signal.
The operator shall not touch output of batteries/adapter and the patient simultaneously.
Manufacturer will make available on request circuit diagrams, component parts list etc.
The materials of the cuff have been tested and found to comply with requirements of ISO
10993-5:2009 and ISO 10993-10:2010. It will not cause any potential sensization or irritation reaction.Never apply the cuff over hurt skin.
Do not use the unit in case of existing polyester resp. synthetic allergies.
Be careful to strangulation due to cables and hoses, particularly due to excessive length.
Do not connect the air hose to other medical equipment, as this could cause air to be pumped into intravascular systems or high pressure, what could lead to dangerous injuries.
Before use, make sure the device functions safely and is in proper working condition.
Please use the device under the environment which was provided in the user manual. Otherwise, the performance and lifetime of the device will be impacted and reduced.
Please use ACCESSORIES and detachable partes specified/ authorised by MANUFACTURE.
Otherwise, it may cause damage to the unit or danger to the user/patients.
Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the local guidelines.
Please do not attempt to repair the unit yourself in the event of malfunctions. Only have repairs carried out by authorized service centers.
Please report to Manufacturer if any unexpected operation or events occur.
The device doesn’t need to be calibrated in two years of reliable service.
Please use the soft cloth to clean the whole unit. Don’t use any abrasive or volatile cleaners.
4
LCD display signal
INTRODUCTION
SYMBOL DESCRIPTION EXPLANATION
Systolic blood pressure High pressure result
Diastolic blood pressure Low pressure result
Pulse per minute Beats per minute, BPM
Deflating
Memory kPa mmHg
CUFF air is exhausting of deflating
The displayed measurement values is from the memory.
Measurement Unit of the blood pressure
(1kPa=7.5mmHg)
Measurement Unit of the blood pressure
(1mmHg=0.133kPa)
Low battery
Irregular heartbeat
Grade
Batteries are low and need to be replaced
Irregular heartbeat Detection
The grade of the blood pressure
Current Time
Heartbeat
User A
User B
Inflation
Year/Month/Day, Hour/Minute
Heartbeat detection during the measurement
Start measurement and save the measuring results for User A
Start measurement and save the measuring results for User B
The air is inflating into the cuff.
5
INTRODUCTION
Monitor Components
CUFF
AIR HOSE
AIR CONNECTOR PLUG
Component list of pressure measuring system
1 Cuff
2 Air pipe
3 PCBA
4 Pump
5 Valve
LCD DISPLAY
DC POWER SOCKET
SET BUTTON
START/STOP BUTTON
MEM BUTTON
6
BATTERY COMPARTMENT
List
1.Arm Blood Pressure Meter
2.Cuff (Type BF applied part)
(about 22cm~42cm)
(Please use 1byone authorized cuff. The size of the actual cuff please refer to the label on the attached cuff.)
3. Storage bag 4.User manual
BEFORE YOU START
The Choice of Power Supply
1.Battery powered mode:
6VDC 4×AAA batteries(Not Included)
2.AC adaptor powered mode:
6V 1A
(Please only use the recommended AC adaptor model).(Not Included)
Please unplug the adaptor to depart from the using utility power.
AC adaptor
CAUTION
In order to get the best effect and protect your Arm Blood Pressure Meter, please use the right battery and special power adaptor which complies with U.S. safety standard.
Installing and Replacing the Batteries
• .Open the battery cover.
• .Install the batteries by matching the correct polarity, as shown.
• .Replace the cover.
Replace the batteries whenever the below happen
The shows
The display dims
The display does not light up
CAUTION
Remove batteries if the device is not likely to be used for some time.
The old batteries are harmful to the environment, do not dispose with other daily trash.
Remove the old batteries from the device and follow your local recycling guidelines.
Do not dispose of batteries in fire. Batteries may explode or leak.
7
BEFORE YOU START
Measurement Principle
This product uses the Oscillometric Measuring method to detect blood pressure. Before every measurement, the unit establishes a “zero pressure” equivalent to the air pressure.
Then it starts inflating the arm cuff, meanwhile, the unit detects pressure oscillations generated by beat-to-beat pulsatile, which is used to determine the systolic and diastolic pressure, and also pulse rate.
The device also compares the longest and the shortest time intervals of detected pulse waves to mean time interval then calculates standard deviation. The device will display a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.
Setting Date, Time and Measurement Unit
It is important to set the clock before using your Arm Blood Pressure
Meter, so that a time stamp can be assigned to each record that is stored in the memory. (The setting range of the year :2014—2054 time format:12H/24H)
1.When the Arm Blood Pressure
Meter is off, hold pressing “S” button about 3 seconds to enter the mode for year setting.
2.Press “M” button to change the [YEAR]. Each press will increase the numeral by one in a cycling manner.
8
3.When you get the right year, press “S” button to set down and turn to next step. You can set the [DATE FORMAT] between D/M and M/D by pressing 'M' button
4.Repeat steps 2 and 3 to set the [MONTH] and [DAY].
BEFORE YOU START
5
.Repeat steps 2 and 3 to set the [TIME FORMAT] between 12H and 24H.
9
BEFORE YOU START
6
.Repeat steps 2 and 3 to set the [HOUR] and [MINUTE].
7.Repeat steps 2 and 3 to set the [UNIT].
8.After the unit is set,the LCD will display “done” first, then display all the settings you have done and then turn off.
10
BEFORE YOU START
Select the User
1.When the Arm Blood Pressure Meter is off , Press and hold
'M' button for 2 seconds to enter user setting mode. The user ID will blink.
2.Then press “M” button again, select the user ID between user A and user B.
3. After selecting the suitable user ID, press “S” button to confirm. Then the LCD will turn off.
11
MEASUREMENT
12
Tie the cuff
1.Tie the cuff on your upper arm, then position the tube off-center toward the inner side of arm in line with the little finger.
Or position the artery mark
Φ
over the main artery (on the inside of your arm).Note: Locate the main artery by pressing with 2 fingers approximately 2 cm above the bend of your elbow on the inside of your left arm. Identify where the pulse can be felt the strongest.
This is your main artery.
2.The cuff should be snug but not too tight. You should be able to
3.Sit comfortably with your tested arm resting on a flat surface.
4
insert one finger between the cuff and your arm.
.Patients with Hypertension:
The middle of the cuff should be at the level of the right atrium of the heart;
Before starting measurement, please sit comfortably with legs uncrossed, feet flat on the floor, back and arm supported.
Rest for 5 minutes before measuring.
Wait at least 3 minutes between measurements. This allows your blood circulation to recover.
For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, position of upper arm, or as directed by a physician.
2~3cm
MEASUREMENT
Start the Measurement
1.When the Arm Blood Pressure Meter is off, press the “START/STOP” button to turn on it, and it will finish the whole measurement.
(Take user A for example.)
LCD display Adjust to zero.
Inflating and measuring.
Display and save the measurement results.
2.Press the “START/STOP” button to power off, otherwise it will turn off within 1 minute.
Tips: Maximum 60 records are both for User A and User B.
13
DATA MANAGEMENT
Recall the Records
1. When the Arm Blood Pressure
Meter is off, please press “M” button to show the average value of the latest three records.
(Note: If the records are less than
3 groups, the LCD will display the recent record instead. Take user A for example.)
2. Press “M” button or “S” button to get the record you want.
The order, date and time of the record will be shown alternately.
The current No. is No. 1.
The corresponding date is May 25 .
The corresponding time is P.M. 9:37.
14
DATA MANAGEMENT
3. If you want to check another user’s records, press “START/STOP” button to turn off the Arm Blood Pressure Meter when it is in the memory inquiry mode. Press and hold 'M' button for 2 seconds to enter into the selecting user
ID mode, press “M” again to select the user ID between user A and user B, press “S” button to confirm the user ID , then press “M” button to check the selected user’s measurement records.
CAUTION
The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record (60) is dropped from the list.
15
DATA MANAGEMENT
Delete the Records
If you did not get the correct measurement, you can delete all results by following steps below. (Take User A for example.)
1.Hold pressing “M” button for 3 seconds when the
Arm Blood Pressure Meter is in the memory recall mode , the flash display “ User ID+ dEL ALL” will show.
2.Press “S” button to confirm deleting and the Arm Blood Pressure
Meter will display “dEL dOnE” and then turn off.
16
Note: To exit out of delete mode without deleting any records, press START/STOP button before pressing "S" button to confirm any delete commands.
3. If there is no record. the following display will show.
INFORMATION FOR USER
Tips for Measurement
Measurements may be inaccurate if taken in the following circumstances.
Within 1 hour after dinner or drinking
Immediate measurement after tea, coffee, smoking
Within 20 minutes after taking a bath
When talking or moving your fingers
In a very cold environment
When you want to discharge urine
17
INFORMATION FOR USER
18
Maintenance
In order to get the best performance, please follow the instructions below.
Put in a dry place and avoid the sunshine
Avoid touching water, clean it with a dry cloth in case.
Avoid intense shaking and collisions
Avoid dusty and unstable temperature environment
Using wet cloths to remove dirt
Do not attempt to clean the reusable cuff with water and never immerse the cuff in water.
ABOUT BLOOD PRESSURE
What are systolic pressure and diastolic pressure?
When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
Systolic
blood discharging artery press
Diastolic
blood entering vein relax
What is the standard blood pressure classification?
The chart on the right is the standard blood pressure classification published by
American Heart Association
(AHA).
This chart reflects blood pressure categories defined by American Heart Association.
Blood Pressure Category
Normal
Prehypertension
Systolic mmHg (upper#) less than 120
120-139 and or
Diastolic mmHg (lower#) less than 80
80-89
AHA Home Guideline for
Upper Limit of Normal BP
SYS 135 mm Hg
DIA 85 mm Hg
High Blood Pressure
(Hypertension) Stage 1
High Blood Pressure
(Hypertension) Stage 2
Hypertensive Crisis
(Emergency care needed)
140-159
160 or higher
Higher than 180 or or or
90-99
100 or higher
Higher than 110
CAUTION
Only a physician can tell your normal BP range. Please contact a physician if your measuring result falls out of the range. Please note that only a physician can tell whether your blood pressure value has reached a dangerous point.
Irregular Heartbeat Detector
An irregular heartbeat is detected when a heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure.During each measurement, this equipment records the heartbeat intervals and works out the standard deviation.
If the calculated value is larger than or equal to 15,the irregular heartbeat symbol appears on the symbol when the measurement results are displayed.
CAUTION
The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heart-beat was detected during measurement. Usually this is NOT a cause for concern.
However, if the symbol appears often, we recommend you seek medical advice. Please note that the device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.
19
ABOUT BLOOD PRESSURE
Why does my blood pressure
fluctuate throughout the day?
1. Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.
2.If the person takes medicine, the pressure will vary more.
3.Wait at least 3 minutes for another measurement.
Why do I get a different blood pressure at home compared to the hospital?
The blood pressure is different even throughout the day due to weather, emotion, exercise etc, Also, there is the
“white coat” effect, which means blood pressure usually increases in clinical settings.
What you need to pay attention to when you measure your blood pressure at home:
If the cuff is tied properly.
If the cuff is too tight or too loose.
If the cuff is tied on the upper arm.
If you feel anxious.
Taking 2-3 deep breaths before beginning will be better for measuring.
Advice: Relax yourself for 4-5 minutes until you calm down.
Is the result the same if measuring on the right arm?
It is ok for both arms, but there will be some different results for different people. We suggest you measure the same arm every time.
20
TROUBLESHOOTING
This section includes a list of error messages and frequently asked questions for problems you may encounter with your Arm
Blood Pressure Meter . If the products not operating as you think it should, check here before arranging for servicing.
PROBLEM SYMPTOM CHECK THIS
No power
Low batteries
REMEDY
Display will not light up.
Display is dim or show
E 1 shows
Batteries are exhausted.
Replace with new batteries
Batteries are inserted incorrectly.
AC adaptor is inserted incorrectly.
Insert the batteries correctly
Insert the AC adaptor tightly
Batteries are low.
The cuff is not secure.
Replace with new batteries
Refasten the cuff and then
Error message
E 2 shows
E 3 shows
E10 or E11
shows
E20 shows
E21 shows
EExx,shows on the display.
The cuff is very tight Readjust the cuff ,not too loose or too tight and then measure again.
The pressure of the cuff is excess.
Relax for a moment and then measure again.
The Arm Blood Pressure
Meter detected motion, talking or the pluse is too poor while measuring.
moment and then measure again.
The measurement process does not detect the pulse signal.
Loosen the clothing on the arm and then measure again
The treatment of the measurement failed.
A calibration error occurred.
Relax for a moment and then measure again.
Retake the measurement.
If the problem persists, contact the retailer or our customer service department for further assistance.Refer to the warranty for contact information and return instructions.
21
22
SPECIFICATIONS
Power supply
Display mode
Measurement mode
Measurement range
Accuracy
Battery powered mode:
6VDC 4×AAA batteries (Not Included)
AC adaptor powered mode:
6V 1A
(Please only use the recommended AC adaptor model).(Not Included)
Digital LCD V.A.100mm×68mm
Oscillographic testing mode
Rated cuff pressure:
0mmHg~300mmHg(0kPa ~ 40kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Pressure:
5℃-40℃within±3mmHg(0.4kPa) pulse value:±5%
Normal working condition
Storage & transportation condition
Measurement perimeter of the upper arm
Net Weight
External dimensions
Attachment
Mode of operation
Degree of protection
Protection against ingress of water
Software Version
Temperature:5℃ to 40℃
Relative humidity: ≤85%RH
Atmospheric pressure: 86kPa to 106kPa
Temperature:-20℃-60℃
Relative Humidity: 10%RH-93%RH
Atmospheric Pressure: 50kPa-106 kPa
About 22cm~42cm
Approx.282g(Excluding the dry cells and cuff)
Approx.154mm×106mm×57.1mm
1×storage bag,user manual
Continuous operation
Type BF applied part
IP21
V01
WARNING: No modification of this equipment is allowed.
AUTHORIZED COMPONENT
Authorized Component
1.
please use the 1byone authorized adapter. (Not Included)
Adapter
SKU 703US-0003
(MODEL: KH0601000UW)
Input:
100~240V~
50/60Hz, 0.3A Max
Output:
6V 1000mA
23
COMPLIED STANDARDS LIST
Complied Standards List
Risk management
ISO/EN 14971:2012 Medical devices — Application of risk management to medical devices
Labeling
User manual
ISO/EN 15223-1:2012 Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. General requirements
EN 1041: 2008 Medical equipment manufacturers to provide information
General Requirements for Safety
IEC 60601-1: 2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-11 Medical electrical equipment - Part
1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
Electromagnetic compatibility
Performance requirements and
Clinical investigation
Software life-cycle processes
IEC/EN 60601-1-2:2007 Medical electrical equipment -
Part 1-2: General requirements for basic safety and essential performance - Collateral standard:Electromagnetic compatibility - Requirements and tests
IEC 80601-2-30:2009 Medical electrical equipment-
Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
ANSI/AAMI SP10:2002/A2: 2008 Manual, electronic, or automated sphygmomanometers
IEC/EN 62304:2006+AC: 2008 Medical device software - Software life cycle processes
Usability
IEC 62366 Medical devices - Application of usability engineering to medical devices (IEC 62366:2007)
IEC 60601-1-6 Medical electrical equipment - Part 1
-6 : General requirements for basic safety and essential performance - collateral standard : Usability
24
EMC GUIDANCE
FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
EMC Guidance
1) This equipment needs to be installed and put into service in accordance with the information provided in the user manual;
2) Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d=3,3m away from the equipment.
(Note: As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical cell phone with a maximum output power of 2 W yields d=3, 3m at an IMMUNITY LEVEL of 3V/m)
25
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project