1788 0-32v/0-6a single output dc programmable power supply

1788 0-32v/0-6a single output dc programmable power supply
1788 0-32V/0-6A
SINGLE OUTPUT DC PROGRAMMABLE POWER SUPPLY
TYPE: SINGLE OUTPUT DC PROGRAMMABLE POWER SUPPLY.
1788 0-32V/0-6A
SINGLE OUTPUT DC PROGRAMMABLE POWER SUPPLY
TYPE: SINGLE OUTPUT DC
PROGRAMMABLE
POWER SUPPLY.
Ratings :
PARAMETER
Output ratings,(0°C-40°C)
Load regulation ±(% of
output+ offset)
Line regulation ±(% of
output+ offset)
Programming regulation
Readback regulation
VOLTAGE
0~32V
<̲0.02% + 5mV
CURRENT
0~6A
<̲0.1% + 2mV
<̲0.1% + 3mV
<̲0.1% + 2mV
10mV
0.05% +
120mV(>̲20V)
10mA
<̲0.1% +
15mA
Programming accuracy, 12
months,
(@25°C ± 5°C) ±(% of
output+ offset)
Readback accuracy, 12
months,
(@25°C ± 5°C) ±(% of
output+ offset)
<̲0.05% + 10mV
<0.2% +
10mA
<0.05% +
15mV(<20V)
<0.05% +
120mV(>̲20V)
<̲0.1% +
15mA
1788 0-32V/0-6A
SINGLE OUTPUT DC PROGRAMMABLE POWER SUPPLY
TYPE: SINGLE OUTPUT DC
PARAMETER
PROGRAMMABLE
Ripple (20Hz ~ 20MHz) <̲1mVrms/3mVp-p
POWER SUPPLY.
Ratings :
VOLTAGE
CURRENT
5mArms
Temperature
<0.02% + 5mV
<0.1% + 5mA
coefficient,( 0°C-40°C)
±(% of output+ offset)
Readback
<0.02% + 15mV(<20V) <0.1% + 15mA
Temperature
<0.02% + 120mV(<20V)
coefficient,
±(% of output+ offset)
1788 0-32V/0-6A
SINGLE OUTPUT DC PROGRAMMABLE POWER SUPPLY
TYPE: SINGLE OUTPUT DC PROGRAMMABLE POWER SUPPLY.
Specifications :
AC Input Ratings
OP1: 220V AC ± 10%,47 to 63 Hz
OP2: 110V AC ± 10%,47 to 63 Hz
Maximum input power
350VA
Cooling
Fan cooled
Operating temperature
32 to 104°F (0 to 40°C) for full rated output
Storage temperature
-68 to 158°F (-20 to 70°C) for storage environment
Weight
N.W
14.8Lbs.
6.7kg
G.W
16.5Lbs.
7.5kg
Dimensions W×H×D
10”×4.16”×15”(255.7mm × 105.7mm × 382.7mm)
Relative humidity
95%
Altitude range
2000 meters
APPLICATIONS: IT OFFERS NEW LEVELS OF EASE OF USE AND PROGRAMMABLE FUNCTIONALITY.
CODE: DSE0PS000017880019.
MANUAL NO.: 19.
9000 SERIES DEAD WEIGHT TESTER
TYPE: Pressure calibration.
9000 SERIES DEAD WEIGHT TESTER
TYPE: Pressure calibration.
Specifications
RANGE
Higher accuracy
1 to 2000bar/15 to 30,000 psi
(kg/cm² also available)
0.015% of reading (optional)
Standard accuracy
0.025% of reading
Traceability
NPL,NAMAS
Weights
Stainless steel, non-magnetic
35-10k(dependent on pressure range)
Direct via ram screw
Priming
Standard corrections
weight
Dimensions
Gravity
: 9.80665m/s²
Temperature : 20°C
Air density : 1.2 kg/m³
Tester
: 18 kg.
Model 9190 : 30 kg.
455×340×478mm max.
9000 SERIES DEAD WEIGHT TESTER
PRESSURE RANGE
TYPE: Pressure calibration.
ORDER
CODE
NO.
SINGLE
PISTON
/4
Psi
Weights
Per set
/5
Kgf/cm²
Weights
Per set
/6
bar
Weights
Per set
9120
10-500
12
1-35
13
1-35
13
9130
20-1000
12
2-75
14
2-75
14
9150
50-5000
12
5-350
13
5-350
13
9170
100-10000
12
10-700
13
10-700
13
9180
100-16000
15
10-1100
17
10-1100
17
9190
500-30000
19
35-2000
23
35-2000
23
9230
10-5000
12
1-350
13
1-350
13
9250
10-10000
12
1-700
13
1-700
13
9260
10-16000
15
1-1100
17
1-1100
17
DUAL
PISTON
Ordering Codes:
9000 SERIES DEAD WEIGHT TESTER
TYPE: Pressure calibration.
APPLICATIONS: used to calibrate accurate pressure.
Code: DSE00P000090000006
MANUAL NO. : 06
500000 COUNT DUAL DISPLAY MULTIMETER
MODEL NO.: 5491A
TYPE: MULTIMETER.
500000 COUNT DUAL DISPLAY MULTIMETER
MODEL NO.: 5491A
TYPE: MULTIMETER.
PARAMETERS
Maximum Range
Best Accuracy
Best Resolution
DC Voltage
1000V
+0.02%+4 digits
10µV
AC Voltage (True RMS)
Freq. 50Hz to 10KHz
DC Current
750V
+0.5%+15 digits
10µV
10A
+0.05 + 4 digits
1µA
AC Current (True RMS,
AC Coupled) Freq. 50Hz
to 2KHz
Resistance
10A
+0.5 + 25 digits
10µA
50MΩ
+0.1 + 3 digits
10mΩ
Frequency
500KHz
+0.01 + 3 digits
10mHz
500000 COUNT DUAL DISPLAY MULTIMETER
MODEL NO.: 5491A
TYPE: MULTIMETER.
GENERAL SPECIFICATIONS
Warm up time
At least 30 minutes
Temperature Coefficient
Add 0.15 x (the applicable accuracy)/°C at 0°C to 18°C and 28°C to 50°C
Operating Temperature
0°C to 50°C (32°F to 122°F)
Storage Temperature
-20°C to 60°C
Altitude
Up to 2000 M
Pollution Degree
II
Over-voltage category
CAT III- 600V and CAT II-1000V
Relative Humidity
Up to 80% (60% for 50MΩ range or resistance measurement)
Common Mode Voltage
1000V dc or peak ac rms maximum between any input and earth ground
Dimension
Approx. 255(w) x 105(h) x 305(d) mm (with holsters)
Weight
<3.0kgs
Line Voltage
100V / 120V / 220V / 240V ac ± 10%, 50/60Hz, 16VA maximum
500000 COUNT DUAL DISPLAY MULTIMETER
MODEL NO.: 5491A
TYPE: MULTIMETER.
APPLICATIONS: Resistance measurements up to 50MΩ, frequency to 500 KHz, true RMS (AC, AC+DC).
CODE: DSE0MM0005491A0017.
MANUAL NO.: 17.
AMTEK TEMPERATURE CALIBRATOR-650SE
TYPE: TEMPERATURE CALIBRATOR
AMTEK TEMPERATURE CALIBRATOR-650SE
TYPE: TEMPERATURE CALIBRATOR
Parameter
Value
Range
50℃/122F to 650℃/1202F
Display Resolution
0.1℃/0.1F
Stability
±0.05℃/0.1F
Accuracy
±0.6℃/1.1F
Heating time, Min to Max
24min
Setting time
5min
Cooling time, Max to Min(50℃)
50min
Storage(100℃)
33min.
Supply
115/230VAC±10%,50/60Hz±10%
Power consumption
1150VA
Weight
7kg
Dimension-H/W/L
320/135/370mm
AMTEK TEMPERATURE CALIBRATOR-650SE
TYPE: TEMPERATURE CALIBRATOR
INPUT CURRENT
Input
0-24mA
Accuracy
±(0.005%+0.01)mA
Internal supply
24V(min12V)max30mA
ANALOGUE OUTPUT
Output
0-6.5V,10mV/℃
Accuracy
±(0.03%+0.003)V
INSERTION TUBE
Tube length
Hole depth
Hole diameter(
std)
Hole
diameter(min)
Hole
diameter(max)
120mm
110mm
φ6.6mm
φ3mm
φ20mm
AMTEK TEMPERATURE CALIBRATOR-650SE
TYPE: TEMPERATURE CALIBRATOR
APPLICATIONS: Used For Temperature Calibration of Thermocouples, RTDs, Thermo
switches, Thermistors and other common temperature sensing devices
CODE: DSE00T000650SE0013
MANUAL NO.: 13
COMPACT PRESSURE CALIBRATOR
CPC SERIES
TYPE: PRESSURE CALIBRATOR.
COMPACT PRESSURE CALIBRATOR
CPC SERIES
TYPE: PRESSURE CALIBRATOR.
PRESSURE
TECHNICAL SPECIFICATIONS
RANGE
-0.80 to 2BAR
RESOLUTION
Minimum 5 digits for full scale
ACCURACY
0.75% of full scale +1 digit
WORKING TEMPERATURE WITH TEMPERATURE
COMPENSATION
OVERLOAD
0°C to50°C/320°F to 1220°F
OVERLOAD ALARM
+20% of full scale, indicated by visual and acoustic alarm ,with
ZERO
Automatic by activation, if pressure is below 1% of full scale
PEAK DETECTION
Freezes the min. and max. pressure in display.
+20% of full scale
INPUT SPECIFICATION
mA:
SWITCH:
RANGE
0 to 25Ma
RESOLUTION
0.01mA
ACCURACY
0.1% of full scale
IMPEDANCE
Less than 6 Ohm
CONNECTION
2mm test sockets
FUNCTION
Open/closed
Meas. current
Approx. 1mA
IMPEDANCE
Larger than 100kOhm
CONNECTION
2mm test sockets
COMPACT PRESSURE CALIBRATOR
CPC SERIES
TYPE: PRESSURE CALIBRATOR.
SPECIFICATIONS
AMBIENT OPERATING TEMPERATURE RANGE
15-40°C
STORAGE TEMPERATURE
10-50°C
HUMIDITY RANGE
40-75% RH
BATTERY
,
BATTERY LIFETIME
1*9V alkaline
SIZE
,
WEIGHT
348mm/180mm/202mm , Approx. 4.5kg
, Approx. 15 hours
APPLICATIONS: Used to test pressure measuring devices connected to the manifold only.
CODE: DSE00P00000CPC0007
MANUAL NO.: 07
CURRENT CALIBRATOR 72-6694
TYPE: CURRENT CALIBRATOR
CURRENT CALIBRATOR 72-6694
TYPE: CURRENT CALIBRATOR
DETAILS:
DISPLAY
GENERAL SPECIFICATIONS
FUNCTION RANGE & RESOLUTION
OVER INPUT
INDICATION
OPERATING TEMPERATURE
13mm (0.5”) digit size. Super large LCD display.
Current source: 2 ranges :
0-19.99mA × 0.01mA
0-24mA × 0.1mA
Indication of “1”
0°C to 50°C (32°F to 122°F)
OPERATING HUMIDITY
Less than 80% RH.
POWER SUPPLY
DIMENSION
066P DC 9V, MN 1604/PP3 battery or equivalent.
Alkaline or heavy duty type.
Current source (under 10mA signal output): approx.
DC 33mA.
185 ×78×38 mm(7.3×3.0×1.5 inch ).
WEIGHT
265 g/0.58 LB (including battery).
POWER CONSUMPTION
CURRENT CALIBRATOR 72-6694
ELECTRICAL SPECIFICATIONS
CURRENT SOURCE
RANGE
DISPLAY RESOLUTION
ACCURACY
0-19.99Ma
0.01mA
±(0.25% FS + 1d)
0-24mA
0.1mA
±(0.5% FS + 1d)
OUTPUT
*Output 0-20mA current for loads up to 500 ohms.
Output >20mA current for loads up to 400 ohms.
*FS: full scale
REMARK (FOR ELECTRICAL SPECIFICATION): The above specifications are tested under the environment RF Field Strength
less than 3V/M & frequency less than the 30 MHz only.
APPLICATIONS: Used for current calibration.
CODE: DSE00C0072-66940004.
MANUAL NO : 04
DC MILLIVOLT REFERENCE SOURCE E100 RC
TYPE: MILLIVOLT REFERENCE SOURCE
DC MILLIVOLT REFERENCE SOURCE E100 RC
TYPE: MILLIVOLT REFERENCE SOURCE
PARAMETER
RANGE1
RANGE2
OUTPUT
±10 V dc+ 10% over ranging
±100mV dc+ 10% over ranging
RESOLUTION
100V(10 ppm) 5decades
1V(10 ppm) 5decades
ACCURACY
±0.005% of setting+50V
±0.01% of setting+5V
RIPPLE AND NOISE, rms(BAND
PASS 0.8 HZ TO 100KHZ)
CURRENT
100V rms
1V rms
75mA with short circuit protection
IMPEDANCE
50 milliohms
Not applicable. (output voltage is
within stated accuracy if load
impedance is greater than 1MΩ)
20 ohms(constant)
DC MILLIVOLT REFERENCE SOURCE E100 RC
TYPE: MILLIVOLT REFERENCE SOURCE
WARM UP TIME
30 Seconds(min); 15 minutes(max)
Temperatures
Calibration:
Ambient:
Operating limit:
Storage:
23°C ±1V°C
20°C to 30°C
-10°C to 50°C
-40°C to 85°C
Temperature coefficient
Power requirements:
Ambient: ±0.0005%/°C
Operating limit: ±0.001%/°C
10W, 50-400Hz
Dimensions: bench series
3½×16¾×7 inches
Weight
10lbs; 4.53 kg
105Vac to 125Vac
220Vac to 240Vac
DC MILLIVOLT REFERENCE SOURCE E100 RC
TYPE: MILLIVOLT REFERENCE SOURCE
STABILITY
load regulation no load to full load
1hour ±0.0005%
8hours ±0.0010%
1year ±0.0025% of range+5V
±0.0005%
Thermal EMF mV, V ranges
1V
Vernier control range
±100V
APPLICATIONS: USED FOR CALIBRATION OF DIGITAL VOLTMETERS, ANALOG METERS,
SEMICONDUCTOR ANALYZING SYSTEMS, ANALOG REFERENCES FOR A/D CONVERTERS,
TELEMETRY AND DATA ACQUISITION SYSTEMS AND AS A POTENTIOMETER FOR MEASURING DC
VOLTAGES.
CODE: DSE0VS0E100RC0030.
MANUAL NO.: 0030
DRUCK DPI 601
TYPE: PORTABLE PRESSURE INDICATOR/CALIBRATOR
DRUCK DPI 601
TYPE: PORTABLE PRESSURE INDICATOR/CALIBRATOR
Specifications:
ACCURACY
RESOLUTION
70mbar to 35bar range
±0.05%FS
35bar to 700bar range
±0.1%FS
Maximum working pressure
150%FS
Instrument<100mbarFS
0.01%FS
Instrument>100mbarFS
0.005%FS
DIMENSIONS
320mm×195mm×125mm
WEIGHT
4kg nominal.
POWER
SUPPLY
Internal: 4×size D( IEC R20) ZINC- CARBON/ALKALINE/NICKEL-CADMIUM BATTERIES.
TEMPERATURE
OPERATING
External:3.5to14volts at 100Ma
9voltsDC at 500mA.
STORAGE
CALIBRATED
-10 to +50°C.
0 TO + 40° C
-20 TO +60°C
DRUCK DPI 601
TYPE: PORTABLE PRESSURE INDICATOR/CALIBRATOR
VOLTAGE RANGE: maximum input and read output………………+50V DC
CURRENT RANGE: maximum input and read output……………….30mA.
POWER OUTPUT: maximum combined load………………………..30mA.
OUTPUT
+24V OUTPUT
+10V OUTPUT
ACCURACY
±5%
±0.1%
TEMPERATURE STABILITY
±0.02%/°C
±0.005%/°C
ANALOGUE OUTPUT
OUTPUT
2V full scale.
Bandwidth
2kHz(-3db)
MAXIMUM LOAD
1000Ω.
APPLICATIONS: used for pressure calibration.
CODE: DSE00P000006010001.
MANUAL NO.: 01.
DRUCK DPI 802
TYPE: PRESSURE LOOP CALIBRATOR
DRUCK DPI 802
TYPE: PRESSURE LOOP CALIBRATOR
GENERAL SPECIFICATIONS
OPERATING TEMPERATURE
14…122°F(-10…50°C)
STORAGE TEMPERATURE
-4…158°F(-20…70°C)
HUMIDITY
SHOCK/VIBRATION
0 to 90% without condensation
(Def Stan 66-31,8.6 cat III)
BS EN 61010:2001; Def Stan 66-3,8.4 cat III
EMC
BS EN 61326-1:1998+A2:2001
SAFETY
SIZE(L:W:H)
Electrical- BS EN 61010:2001;
Pressure equipment directive- class:
Sound Engineering Practice(SEP);CE Marked
Maximum: 7.5×3.3×2.0 in (190×85×50mm)
WEIGHT
19 oz (530g)
POWER SUPPLY
3×AA alkaline batteries
DURATION(MEASURE)
P1:≈50 hours
P1+mA:≈5 hours (24 V Source at 12mA)
DRUCK DPI 802
TYPE: PRESSURE LOOP CALIBRATOR
SPECIFICATIONS - ELECTRICAL CONNECTORS (A2):
RANGE(MEASURE)
0 to ±55mA
ACCURACY
0.02% of reading + 3 counts
TEMPERATURE COEFFICIENT
14…50°F,86…122°F
(-10…10°C,30…50°C)
SWITCH DETECTION
0.0011%FS/°F
(0.002%FS/°C)
Open and closed,2mA current
LOOP POWER OUTPUT
24V±10%
HART RESISTOR
250Ω(menu selection)
CONNECTORS(A2)
Three 0.16 in (4mm)sockets
APPLICATIONS: Used as a pressure loop calibrator.
CODE: DSEOOP000008000009.
MANUAL NO.: 09.
DRUCK DPI 821
TYPE: THERMOCOUPLE CALIBRATOR
DRUCK DPI 821
TYPE: THERMOCOUPLE CALIBRATOR
THERMOCOUPLE STANDARD
TYPE
R
IEC 584
RANGE°F
RANGE°C
ACCURACY °F*
-58...32
-50...0
5.4
ACCURACY
°C*
3.0
R
IEC 584
32...572
0...300
3.6
2.0
R
IEC 584
572...3216
300...1769
1.8
1.0
S
IEC 584
-58...32
-50...0
4.5
2.5
S
IEC 584
32...212
0...100
3.4
1.9
S
IEC 584
212...3216
100...1769
2.5
1.4
E
IEC 584
-454...-238
-270...-150
1.6
0.9
E
IEC 584
-238...1832
-150...1000
0.7
0.4
N
IEC 584
-454...-4
-270...-20
1.8
1.0
*COLD JUNCTION ERROR(MAX) : RANGE 50°...86°F(10....30°C) =0.4°F(0.2°C)
Add 0.01° CJ Error/°ambient temperature for ranges 14....50° F, 86....122°F (-10....10°C,30....50°C)
DRUCK DPI 821
TYPE: THERMOCOUPLE CALIBRATOR
THERMOCOUPLE
TYPE
N
STANDARD
RANGE°F
RANGE°C
ACCURACY °F*
IEC 584
-4...2372
-20...1300
1.1
ACCURACY
°C*
0.6
L
DIN 43710
-328...1652
-200...900
0.6
0.3
U
DIN 43710
-328...212
-200...100
0.9
0.5
U
DIN 43710
212...1112
100...600
0.6
0.3
C
32...2732
0...1500
1.8
1.0
C
2732...3632
1500...2000
2.5
1.4
C
3632...4208
2000...2320
3.4
1.9
D
32...3092
0...1700
1.8
1.0
D
3092...3992
1700...2200
2.9
1.6
D
3992...4523
2200...2495
6.5
3.6
*COLD JUNCTION ERROR(MAX) : RANGE 50°...86°F(10....30°C) =0.4°F(0.2°C)
Add 0.01° CJ Error/°ambient temperature for ranges 14....50° F, 86....122°F (-10....10°C,30....50°C)
DRUCK DPI 821
TYPE: THERMOCOUPLE CALIBRATOR
THERMOCOUPLE STANDARD
TYPE
K
IEC 584
RANGE°F
RANGE°C
ACCURACY °F*
-454...-328
-270...-200
3.6
ACCURACY
°C*
2.0
K
IEC 584
-328...2498
-200...1370
1.1
0.6
J
IEC 584
-346...2192
-210...1200
0.9
0.5
T
IEC 584
-454...292
-270...-180
2.5
1.4
T
IEC 584
-292...-94
-180...-70
0.9
0.5
T
IEC 584
-94...752
-70...400
0.6
0.3
B
IEC 584
122...932
50...500
7.2
4.0
B
IEC 584
932...2192
500...1200
3.6
2.0
IEC 584
2192...3308
1200...1820
1.8
1.0
B
*COLD JUNCTION ERROR(MAX) : RANGE 50°...86°F(10....30°C) =0.4°F(0.2°C)
Add 0.01° CJ Error/°ambient temperature for ranges 14....50° F, 86....122°F (-10....10°C,30....50°C)
DRUCK DPI 821
TYPE: THERMOCOUPLE CALIBRATOR
SPECIFICATIONS:
Range (mV)
Impedance
-10...100
<0.2Ω
OPERATING TEMPERATURE
14.....122°F(-10....50°C)
STORAGE TEMPERATURE
-4......158°F(-20....70°C)
HUMIDITY
SIZE(L:W:H)
0 TO 90% Without condensation (Def stan 6631,8.6 cat lll)
7.1×3.3×2.0 in ( 180×85×50)mm
WEIGHT
14oz(400gm)
POWER SUPPLY
3×AA alkaline batteries
SHOCK/VIBRATION
BS EN 61010:2001; Def Stan 66-31,8.4 cat lll
EMC
BS EN 61326-1:1998 + A2:2001
SAFETY
ELECTRICAL- BS EN 61010:2001; CE MARKED
DRUCK DPI 821
TYPE: THERMOCOUPLE CALIBRATOR
APPLICATIONS: USED TO MEASURE AND SIMULATE TEMPERATURE.
CODE: DSE00T000008210003
MANUAL NO. 03
EUROTRON MICROCAL 200
TYPE: MULTI FUNCTION CALIBRATOR
EUROTRON MICROCAL 200
TYPE: MULTI FUNCTION CALIBRATOR
Parameter
Resolution
Accuracy
Ni 100, Ni 120
0.1°C
±(0.02% +0.05°C)
mV(L)
1 V
±(0.02% +2 V)
mV(H)
10 V
±(0.02% +10 V)
V
0.1mV
±(0.02%+0.08mV)
mA(In), (Out)
0.1A
±(0.02% +0.4A)
Ω In
1mΩ
±(0.02% +12mΩ)
Ω Out
1mΩ
±(0.02% +20mΩ)
Frequency
0.001,0.01,
0.1Hz
±(0.005%+0.001
Hz)
APPLICATIONS: Used to simulate and measure voltage, current, resistance, thermocouple, resistance
thermometers, frequency and pulse.
CODE: DSEMFC0MCAL2000029.
MANUAL NO.: 0029.
EUROTRON MICROCAL 200
TYPE: MULTI FUNCTION CALIBRATOR
Dimensions
264×96×172mm
Weight
Nett: 4kg
Line operation
100-120-230Vac through external battery charger
Line transformer insulation
2500Vac
Operating temperature
-10°C to +55°C
Storage temperature
-30°C to +60°C
Power supply
External charger and rechargeable Ni-Cd battery
Common mode rejection
>140dB at AC operation
Normal mode rejection
>70dB a 50 or 60Hz
Channel 1- channel 2 insulation
50V dc
In/out data memory
20 data with manual or automatic recall
Measurement sampling time
250ms
Temperature stability( For temperature
exceeding the band +18°C to + 28°C)
Max load resistance
Full scale : ±8ppm/°C
Zero
: ±0.2V/°C
1000Ω@ 20mA
Output impedance(emf output)
<0.5Ω with 0.5mA max current
Input impedance
mV, V and Tc ranges:
mA ranges:
>10MΩ
<140Ω@1mA
EUROTRON MICROCAL 200
TYPE: MULTI FUNCTION CALIBRATOR
Reference junction compensation
Reference junction drift
Reference junction error
Internal automatic: -10°C to
+55°C
External adjustable: -50°C to
+100°C
Remote with external Pt-100: 10°C to +100°C
±0.015°C/°C (da-10°C a+55°C)
Internal: ±0.15°C
Remote: ±0.3°C
RTD and Ω
simulation
excitation current
0.01 to 5mA
RTD and Ω
measurement
excitation current
[email protected]Ω
[email protected]Ω
RTD connection
RTD cable
compensation
2,3 and 4 wires
Up to 100Ω(each
wire)
RTD cable
compensation error
(Pt-100)
±0.005°C/Ω of
total wire
EUROTRON MICROCAL 200
TYPE: MULTI FUNCTION CALIBRATOR
Parameter
Resolution
Accuracy
Tc J,K,T,U,L,N,E,F
0.1°C
±(0.02% +0.1°C)
Tc R,S,C
0.1°C
±(0.02% +0.1°C)
Tc B,G,D
0.1°C
±(0.02% +0.3°C)
Pt100 IEC
OIML, a3946
Pt100 a3902
Pt100 JIS SAMA
Pt 200
0.01°C
0.01°C
0.01°C
0.1°C
±(0.02% +0.5°C)
±(0.02% +0.5°C)
±(0.02% +0.5°C)
±(0.02% +0.15°C)
Pt 500
0.1°C
±(0.02% +0.1°C)
Pt1000 IEC OIML
0.01°C
±(0.02% +0.1°C)
CU 10
0.1°C
±(0.02% +0.4°C)
CU 100
0.1°C
±(0.02% +0.05°C)
FLUKE 85
TYPE: Multimeter
FLUKE 85
TYPE: Multimeter
SPECIFICATIONS
FUNCTION
RANGE
RESOLUTION
ACCURACY
CAPACITANCE
5.00nF
0.0500μF
0.500μF
5.00μF
3.000V
0.01nF
0.0001μF
0.001μF
0.01μF
0.001V
±(1% + 3)
±(1% + 3)
±(1% + 3)
±(1% + 3)
±(2% + 1)
DIODE TEST
FUNCTION
mA
A ͠
(45Hz to 2kHz)
mA --͞A
μ͠A
(45Hz to 2kHz)
-͞-μA
RANGE
RESOLUTION
ACCURACY
40.00mA
400.0mA
4000mA
10.00A³
40.00mA
400.0mA
4000mA
10.00A³
400.0 mA
400.0 μA
400.0 μA
400.0 μA
0.01mA
0.1mA
1mA
0.01A
0.01mA
0.1mA
1mA
0.01A
0.1μA
1μA
0.1μA
1μA
±(0.6% +2)
±(0.6% +2)
±(0.6% +2)
±(0.6% +2)
±(0.2%+2)
±(0.2%+2)
±(0.2%+2)
±(0.2%+2)
±(0.6%+2)
±(0.6%+2)
±(0.2%+3)
±(0.2%+2)
BURDEN
VOLTAGE
TYPICAL
1.6mV/mA
1.6mV/mA
0.03V/A
0.03V/A
1.6mV/mA
1.6mV/mA
0.03V/A
0.03V/A
100μV/μA
100μV/μA
100μV/μA
100μV/μA
FLUKE 85
TYPE: Multimeter
SPECIFICATIONS
FUNCTION
RANGE
RESOLUTION
ACCURACY
Frequency
(0.5 Hz to 200kHz,
Pulse width >2μs)
199.99
199.99
199.99kHz
199.99kHz
>200kHz
0.01Hz
0.1Hz
0.001kHz
0.01kHz
0.1kHz
±(0.005%+1)
±(0.005%+1)
±(0.005%+1)
±(0.005%+1)
Unspecified
FUSE PROTECTION
mA or μA
A
1A 600V FAST fuse
15A 600V FAST fuse
MAXIMUM VOLTAGE BETWEEN ANY
TERMINAL AND EARTH GROUND
1000 Volts
FLUKE 85
TYPE: Multimeter
SPECIFICATIONS
OPERATING TEMPERATURE
-20°C to 55°C
STORAGE TEMPERATURE
-40°C to 60°C
TEMPERATURE COEFFICIENT
0.05×(specified accuracy)/°C(<18°C or>28°C)
ELECTROMAGNETIC COMATIBILITY
BATTERY TYPE
In an RF field of 1V/m on all ranges and functions:
Total Accuracy=Specified Accuracy+0.5% of range.
0% to 90%(0°C to 35°C)
0% to 70%(35°C to 55°C)
9V, NEDA 1604 or 6F22 or 006P
BATTERY LIFE
500 hrs typical with alkaline
SHOCK, VIBRATION
Per MIL-T-28800 for a class 2 instrument
SIZE(H×W×L)
1.25 in × 3.41 in × 7.35 in (3.1cm × 8.6cm ×
18.6cm)
12.5 oz (355g)
RELATIVE HUMIDITY
WEIGHT
APPLICATIONS: Used for different measurements.
CODE: DSE0MM0FLUKE850012.
MANUAL NO.: 12.
FLUKE 725
MULTIFUNCTION PROCESS CALIBRATOR
TYPE: PROCESS CALIBRATOR
FLUKE 725
MULTIFUNCTION PROCESS CALIBRATOR
TYPE: PROCESS CALIBRATOR
GENERAL SPECIFICATIONS
Operating temperature
-10°C to 55°C
Storage temperature
-20°C to 71°C
Relative humidity(% RH operating without
condensation)
Vibration
90% (10 to 30°C)
75% (30 to 40°C)
45% (40 to 50°C)
35% (50 to 55°C)
Uncontrolled<10°C
Random,2g, 5 to 500Hz
Power requirements
4AA alkaline batteries
Size
96 × 200 × 47mm.
Weight
650 gm (1lb, 7 oz)
Ohms range
(3.75 × 7.9 × 1.86 in)
ACCURACY±Ω
OHMS
MEASUREMENT
4-wire
0 to 400Ω
400Ωto1.5kΩ
1.5 to 3.2kΩ
2- and 3-wire
0.1
0.1
0.5
1
0.15
1.0
1.5
FLUKE 725
MULTIFUNCTION PROCESS CALIBRATOR
TYPE: PROCESS CALIBRATOR
PARAMETER
DC voltage
measurement
RANGE
DC voltage
source
DC mA measurement and source
30V
20V
90V
100mV
10mV
24mA
Millivolt measurement and source
-10 to 75mV
FREQUENCY MEASUREMENT
FREQUENCY SOURCE
2 to 1000CPM
1 to 1000
1 to 10kHz
2 to 1000CPM
1 to 1000
1 to 10kHz
RESOLUTION
0.001V
0.001V
0.01mV
0.01mV
0.001mV
0.001mA
ACCURACY
0.02%+2
0.02%+2
0.02%+2
0.02%+2
0.02%+2
0.02%+2
0.01mV
±(0.025%+1count)
0.1CPM
1Hz
0.01kHz
0.1CPM
1Hz
0.01kHz
±(0.05%+1)
±(0.05%+1)
±(0.05%+1)
±0.05%
±0.05%
±0.25%
APPLICATIONS: Used to measure electrical and physical parameters.
CODE: DSEMPC000007250016.
MANUAL NO.: 16.
FLUKE 1953A COUNTER TIMER
TYPE: COUNTER/ TIMER
GWINSTEK GFG-8216A FUNCTION GENERATOR
TYPE: FUNCTION GENERATOR
GWINSTEK GFG-8216A FUNCTION GENERATOR
TYPE: FUNCTION GENERATOR
MAIN OUTPUT
Frequency Range:
0.5Hz ~ 5MHz in 7 ranges
Amplitude:
> 10Vpp (into 50W load)
Impedance
/ Attenuator:
50W ±10%
/
-20dB ±1dB x 2
DC Offset:
<-5V ~ >+5V (into 50W load)
Duty Control:
80% : 20% : 80% to 1MHz continually variable
Display:
6 digit LED display
SINE WAVE
Distortion:
£1%, 0.5Hz ~ 100kHz THD £ -30dB fundamental in all
ranges (specification applied from MAX to 1/10 level)
Flatness:
£0.3dB, below 500kHz
GWINSTEK GFG-8216A FUNCTION GENERATOR
TYPE: FUNCTION GENERATOR
TRIANGLE WAVE
Linearity:
³98% 0.5Hz to 100kHz
³95% 100kHz to 5MHz
SQUARE WAVE
Symmetry:
±2% 1Hz ~ 100kHz
Rise/Fall Time:
£50ns at maximum output (into 50W load)
CMOS OUTPUT
Level:
4Vpp ±1Vpp ~ 14.5Vpp ±0.5Vpp adjustable
Rise/Fall Time:
£120ns
GWINSTEK GFG-8216A FUNCTION GENERATOR
TYPE: FUNCTION GENERATOR
TTL OUTPUT
Level:
³3Vpp
Fan Out:
20 TTL Loads
Rise/Fall Time:
£25ns
VCF
Input Voltage:
0V ~ 10V ±1V (100:1)
Input Impedance:
10kW ±10%
FREQUENCY COUNTER
INT/EXT:
Switch selectable
Range:
0.5Hz to 5MHz (EXT 5Hz ~ 150MHz)
Accuracy:
Time base accuracy ±1 count
Time Base:
±20ppm (23°C ±5°C) after 30 minutes warm up
Resolution:
Maximum resolution is 100nHz for 1Hz and 1Hz for
100MHz
Input Impedance:
1MW // 150Pf
GWINSTEK GFG-8216A FUNCTION GENERATOR
TYPE: FUNCTION GENERATOR
Sensitivity:
£35mVrms (5Hz ~ 100MHz)
£45mVrms (100MHz ~ 150MHz)
OTHER DETAILS
Power Source:
AC 115V/230V ±15%, 50/60Hz
Accessories:
Power cord x 1
Dimensions
/
weight:
251mm (w) x 91mm (h) x 291mm (d) /
approx
APPLICATIONS: Used to generate different signals.
CODE: DSE0FGGFG8216A0023.
MANUAL NO.:23
Instruction manual x 1
2.4kg
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
DETAILS:
GENERAL SPECIFICATIONS
Operating temperature
0°C to 50°C
Storage temperature
-40°C to 175°C
Warm-up time
1 hr to meet all specifications
Humidity range
95% relative humidity for temperatures between 0°C and 40°C
Power requirements
100/120/240 volts ± 10% , 48Hz to 66Hz. 30VA maximum
Dimensions
Height
Width
Depth
425mm
292mm
Net weight
89mm(with feet
removed)
5.05kgm
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
DETAILS:
GENERAL SPECIFICATIONS
Parameter
Range
Max reading
DC voltage
30mV
300mV
3V
30V
300V
300A
3mA
30mA
300mA
1A
30.03000mV
303.0000mV
3.030000V
30.30000V
303.0000V
303.0000A
3.030000mA
30.30000mA
303.0000mA
1.000000A
DC current
Best 6½ digit accuracy³ ± (% rdg +
cnts)
% of reading
Count error
0.0040
0.0025
0.0017
0.0035
0.0050
0.02
0.02
0.02
0.07
0.07
365
39
6
19
6
104
104
104
204
604
Input
resistance
10GΩ
10GΩ
10GΩ
10MΩ
10MΩ
1000Ω
100Ω
10Ω
1Ω
0.1Ω
Best 6½ digit accuracy³ ± (% rdg + cnts)
Parameter
Range
Max reading
Resistance
30Ω
300Ω
3kΩ
30kΩ
300kΩ
3MΩ
30MΩ
300MΩ³
3.0G Ω³
30.30000Ω
303.0000Ω
3.030000kΩ
30.30000kΩ
303.0000kΩ
3.030000MΩ
30.30000MΩ
303.0000MΩ
3.030000GΩ
% of reading
Count error
0.0065
0.0045
0.0035
0.0035
0.0040
0.0055
0.0250
1.6
16.0
315
34
6
6
7
12
80
1000
1000
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
DETAILS:
GENERAL SPECIFICATIONS
Operating temperature
0°C to 50°C
Storage temperature
-40°C to 175°C
Warm-up time
1 hr to meet all specifications
Humidity range
95% relative humidity for temperatures between 0°C and 40°C
Power requirements
100/120/240 volts ± 10% , 48Hz to 66Hz. 30VA maximum
Dimensions
Height
Width
Depth
425mm
292mm
Net weight
89mm(with feet
removed)
5.05kgm
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
DETAILS:
GENERAL SPECIFICATIONS
Parameter
Range
Max reading
DC voltage
30mV
300mV
3V
30V
300V
300A
3mA
30mA
300mA
1A
30.03000mV
303.0000mV
3.030000V
30.30000V
303.0000V
303.0000A
3.030000mA
30.30000mA
303.0000mA
1.000000A
DC current
Best 6½ digit accuracy³ ± (% rdg +
cnts)
% of reading
Count error
0.0040
0.0025
0.0017
0.0035
0.0050
0.02
0.02
0.02
0.07
0.07
365
39
6
19
6
104
104
104
204
604
Input
resistance
10GΩ
10GΩ
10GΩ
10MΩ
10MΩ
1000Ω
100Ω
10Ω
1Ω
0.1Ω
Best 6½ digit accuracy³ ± (% rdg + cnts)
Parameter
Range
Max reading
Resistance
30Ω
300Ω
3kΩ
30kΩ
300kΩ
3MΩ
30MΩ
300MΩ³
3.0G Ω³
30.30000Ω
303.0000Ω
3.030000kΩ
30.30000kΩ
303.0000kΩ
3.030000MΩ
30.30000MΩ
303.0000MΩ
3.030000GΩ
% of reading
Count error
0.0065
0.0045
0.0035
0.0035
0.0040
0.0055
0.0250
1.6
16.0
315
34
6
6
7
12
80
1000
1000
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
Max reading
Rates(DCV,
DCI
Resistance
up to
30kOhm)²
Reading
rates(ACV
And ACI)
Power line
cycles³
Readings per second-60Hz(50Hz)
Max # of
digits
.0005
.005
.1
1.0
10
100
3½
4½
5½
6½
7½
7½
.0005
.005
.1
1.0
10
100
3½
4½
5½
6½
6½
6½
Auto zero
ON
300
280
140(128)
26(22)
2.5(2.0)
.25(0.2)
Auto zero OFF
NMR
1350
1250
360(312)
53(45)
4.8(4.0)
0.5(0.4)
Readings per second-60Hz(50Hz)
0
0
0
60dB
80dB
90dB
1
1
1(1)
1(1)
0.7(0.65)
0.2(0.17)
9.5
9.5
9.25(9.2)
7.25(6.9)
2.0(1.7)
0.25(0.2)
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
PARAMETER
TRUE RMS ACV and (AC+DC)V
TRUE RMS ACI and (AC+DC)I
Band width
20 Hz TO 1MHz
20 Hz TO 100 KHz
Crest factor
3.5 TO 1 at full scale
3.5 to full scale
Common mode rejection ratio
>76dB , DC TO 60Hz
---------
Accuracy
(90day) accuracy specified for
sine wave inputs,>10%of range.
dc component<10% of ac
component after 1hr warm-up and
within one week of autocal. AC
band set to<400hz. dc coupled
mode requires 2 hour warm-up.
10PLC(POWER LINE CYCLES)
(90day) accuracy specified for
sine wave inputs,>10%of range.
dc component<10% of ac
component after 1hr warm-up and
within one week of autocal. AC
band set to<400hz. dc coupled
mode requires 2 hour warm-up.
10PLC(POWER LINE CYCLES)
Integration Time
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
TRUE RMS ACV and (AC+DC) V
Range
Max reading
(100Hz to 20 kHz) best 5½ digit accuracy ± (% Rdg + Cnts)
AC coupled
30mV
300mV
3.0V
30.0V
300V
32.50000Mv
325.0000mV
3.250000V
32.50000V
303.0000V
DC coupled
% of reading count error
% of reading
count error
0.13
0.13
0.13
0.13
0.19
0.17
0.17
0.17
0.17
0.23
364
364
364
364
364
116
116
116
116
116
Input
impedanc
e
1M
Ohm±1%
shunted
by <90pf
TRUE RMS ACI and (AC+DC) I
Range
Max reading
(100Hz to 20 kHz) best 5½ digit accuracy ± (% Rdg + Cnts)
AC coupled
30mA
300mA
1.0A
32.50000mA
325.0000mA
1.000000A
DC coupled
% of reading
count error
% of reading
count error
0.25
0.25
0.35
290
290
290
0.3
0.3
0.4
1600
1600
1600
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
APPLICATION: Used to measure different parameters.
CODE: DSE0MM0003457A0028.
MANUAL NO.: 0028.
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
Max reading
Rates(DCV,
DCI
Resistance
up to
30kOhm)²
Reading
rates(ACV
And ACI)
Power line
cycles³
Readings per second-60Hz(50Hz)
Max # of
digits
.0005
.005
.1
1.0
10
100
3½
4½
5½
6½
7½
7½
.0005
.005
.1
1.0
10
100
3½
4½
5½
6½
6½
6½
Auto zero
ON
300
280
140(128)
26(22)
2.5(2.0)
.25(0.2)
Auto zero OFF
NMR
1350
1250
360(312)
53(45)
4.8(4.0)
0.5(0.4)
Readings per second-60Hz(50Hz)
0
0
0
60dB
80dB
90dB
1
1
1(1)
1(1)
0.7(0.65)
0.2(0.17)
9.5
9.5
9.25(9.2)
7.25(6.9)
2.0(1.7)
0.25(0.2)
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
PARAMETER
TRUE RMS ACV and (AC+DC)V
TRUE RMS ACI and (AC+DC)I
Band width
20 Hz TO 1MHz
20 Hz TO 100 KHz
Crest factor
3.5 TO 1 at full scale
3.5 to full scale
Common mode rejection ratio
>76dB , DC TO 60Hz
---------
Accuracy
(90day) accuracy specified for
sine wave inputs,>10%of range.
dc component<10% of ac
component after 1hr warm-up and
within one week of autocal. AC
band set to<400hz. dc coupled
mode requires 2 hour warm-up.
10PLC(POWER LINE CYCLES)
(90day) accuracy specified for
sine wave inputs,>10%of range.
dc component<10% of ac
component after 1hr warm-up and
within one week of autocal. AC
band set to<400hz. dc coupled
mode requires 2 hour warm-up.
10PLC(POWER LINE CYCLES)
Integration Time
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
TRUE RMS ACV and (AC+DC) V
Range
Max reading
(100Hz to 20 kHz) best 5½ digit accuracy ± (% Rdg + Cnts)
AC coupled
30mV
300mV
3.0V
30.0V
300V
32.50000Mv
325.0000mV
3.250000V
32.50000V
303.0000V
DC coupled
% of reading count error
% of reading
count error
0.13
0.13
0.13
0.13
0.19
0.17
0.17
0.17
0.17
0.23
364
364
364
364
364
116
116
116
116
116
Input
impedanc
e
1M
Ohm±1%
shunted
by <90pf
TRUE RMS ACI and (AC+DC) I
Range
Max reading
(100Hz to 20 kHz) best 5½ digit accuracy ± (% Rdg + Cnts)
AC coupled
30mA
300mA
1.0A
32.50000mA
325.0000mA
1.000000A
DC coupled
% of reading
count error
% of reading
count error
0.25
0.25
0.35
290
290
290
0.3
0.3
0.4
1600
1600
1600
HP 3457A DIGITAL MULTIMETER
TYPE : DIGITAL MULTIMETER
APPLICATION: Used to measure different parameters.
CODE: DSE0MM0003457A0028.
MANUAL NO.: 0028.
FLUKE 1953A COUNTER TIMER
TYPE: COUNTER/ TIMER
DETAILS:
PARAMETER
RANGE
ACCURACY
FREQUENCY
MEASURMENTS
0-125MHz(dc coupled)
5Hz to 125MHz(ac
coupled)
0 to 25MHz(dc
coupled) 5Hz to
25MHz(ac coupled)
0.1 μs to 10⁷s
Time base accuracy ±1 0.1Hz at 10s gate time to
count
10KHz at 0.1ms gate time
KHz or MHz automatically displayed with
decimal point
Time base accuracy ±1
count + trigger error of
signal on input A
±1 count + time base
accuracy + trigger error
±1 count of signal on
input A+ trigger error of
signal on input B.
1ps at 10⁵ periods
to0.1μs at 1period
ms or μs automatically displayed with
decimal point
0.1μs to 10ms in 6
decade steps
DISPLAY
ms or s automatically displayed with
decimal point
Decimal point without unit annunciation
PERIOD
MEASURMENTS
TIME INTERVAL
MEASURMENT
RATIO
MEASURMENT
f1: 0 to 120MHz (dc
coupled) 5Hz
to120MHz (ac)
f2: 0 to 25MHz (dc)
5Hz to25MHz(ac)
RESOLUTION
f1/f2,where f1 and f2 are
applied at the two input
channels, A and B,
respectively
READOUT
FLUKE 1953A COUNTER TIMER
TYPE: COUNTER/ TIMER
Frequency required
10MHz
Impedance
1MΩ, 20pF
Sensitivity
250mVrms
Dynamic range without limiting
5V peak-to-peak
Operating temperature
0°C to +50°C
Storage temperature
-40°C to + 75°C
Dimensions (W×D×H)
36.2×34.29×8.76 cm
Weight
4.32kg(9.5 lbs)
FLUKE 1953A COUNTER TIMER
TYPE: COUNTER/ TIM
INPUT IMPEDANCE
Channel A or B
1mΩ in parallel with 30pf max
Channel C (pre scaled)
50Ω nominal, VSWR 2:1 max when not
limiting
-3.5 to +3.5V(channel A and B)
1Vrms (channel C)
120KΩ in parallel with 75pf (channel A and
B).VSWR less than 3:1(channel C)
Dynamic range without limiting
Impedance in limiting condition
APPLICATIONS: Capable of measuring frequency, frequency ratio, time interval, period and total events.
CODE: DSE0CT0001953A0026.
MANUAL NO.: 26.
HP 3468A MULTIMETER
TYPE: MULTIMETER
GENERAL INFORMATION
HP 3468A MULTIMETER
TYPE: MULTIMETER
GENERAL INFORMATION
Operating
temperature
Humidity Range
0 to 55°C
Storage
Temperature
Warm-up Time
-40°C to 75°C
Expect for battery option, -40°C to +65°C
1hr to meet all specifications
Integration Time
95% R.H., 0 to 40°C
Number of digits
Line frequency
50Hz
60Hz
Power
5½
4½
3½
AC line 48-440Hz; 86-250V
Battery (option
001)
Maximum power
Rechargeable lead-acid; minimum continuous operation for 5 hours at 25°C;
Recharge time is 16 hours with 3468A off and 36 hours with 3468A on.
<13 VA
Size
98.4mm H × 238.1mm W × 276.2mm D
(3.88 in H × 9.38 in W × 10,88 in D)
2.1kg(4.63 lbs)
With option 001- 3.1 kg (6.83 lbs)
Weight
200ms
20ms
2ms
166.7ms
16.67ms
2ms
HP 3468A MULTIMETER
TYPE: MULTIMETER
PARAMETERS
DC Voltage
Resistance
AC Voltage
DC Current
GENERAL INFORMATION
MEASUREMENT
ACCURACY
±(% of reading +
number of counts)
Auto-zero ON
±(% of reading +
number of counts)
Auto-zero ON. 4-wire Ω
±(% of reading +
number of counts)
Auto-zero ON. 5½ digit
display.
Accuracy is specified for
sine-wave
inputs only,>10%of FS
±(% of reading +
number of counts)
Auto-zero ON. 5½ digit
display.
RANGE
0.3V.30V
3V,300V
TEMPERATURE
COEFFICIENT
0.0008 + .5
0.0007 + .05
3000Ω
3k-30kΩ
3MΩ
30MΩ
.3V
3V
30V
300V
0.0009 + 0.5
0.0009 + 0.05
0.0021 + 0.05
0.021 + 0.05
0°C to(Cal. Temp. -5°C),
( Cal. Temp.
+5°C)to55°C, 5½ digit
display.
Autozero ON
.3A
3A
0°C to(Cal. Temp. -5°C),
( Cal. Temp.
+5°C)to55°C, 5½ digit
display. Auto-zero ON.
±(0.012 of reading + 0.5
counts )/°C
HP 3468A MULTIMETER
TYPE: MULTIMETER
GENERAL INFORMATION
APPLICATIONS: Used For Different Measurements.
CODE: DSE0MM0003468A0015.
MANUAL NO.:15
HP6181C DC CURRENT SOURCE
TYPE: POWER SUPPLY
HP6181C DC CURRENT SOURCE
TYPE: POWER SUPPLY
Specifications
OUTPUT VOLTAGE
ACCURACY
RANGE
0.025%
RANGE A
RANGE B
RANGE C
1% full-scale current
10% full-scale current
100% full-scale current
LOAD REGULATION
0.0025% +(5× 10⁻⁶× range full-scale current)
LINE REGULATION
0.0025% +(5× 10⁻⁶× range full-scale current)
RIPPLE AND NOISE
Rms : 0.00032 × range full-scale current
peak-to-peak: 0.008 × range full-scale current
DC OUTPUT:
full-scale current
max. voltage compliance
AC INPUT
0 to 250mA
0 to 100mV
115 ±10%Vac, 48 to 63 Hz(for 230 Vac)
REGULATORY
COMPLIANCE
SIZE(H×W×D)
Listed to UL 1244, IEC-348
WEIGHT
Net 4.5 kg(10 lb); shipping 5.9kg(13 lb)
87.3mm×197.6mm×298.5mm
HP6181C DC CURRENT SOURCE
TYPE: POWER SUPPLY
APPLICATIONS: Used for resistivity measurements, semiconductor characterization and
applications in electrochemistry and electromagnetic.
CODE: DSE0PS0006181C0027
MANUAL NO.: 27
IPM-100 INDUSTRIAL PROCESS CONTROL LOOP CALIBRATOR
TYPE: CONTROL LOOP CALIBRATOR
INSTRUMENT
IPM-100 INDUSTRIAL PROCESS CONTROL LOOP CALIBRATOR
TYPE: CONTROL LOOP CALIBRATOR
PARAMETER
VALUE
Ranges
0.00 mA to 24.00 mA and
–25.0% to 125.0% of 4mA to 20mA
Accuracy
± (0.08% of 24 mA span+0.01 mA)
=0.03 mA
Temperature effect
Warm-up time
0.01% per °C between –2 °C and45°C
10 seconds to specified accuracy;
2 minutes to maximum accuracy
Overload protection
120 V AC/DC for 30 seconds
Operating conditions
–20 °C to 55 °C (–4 °F to 131 °F), 10% to 90% relative
humidity,
Without condensation, for 24 hours
Indoor use only
Storage conditions
–20 °C to 55 °C (–4 °F to 131 °F), 10% to 90% relative
humidity, without
Condensation, for 24 hours
Remove battery
Battery
Two 9-volt batteries (NEDA 1604, JIS 006P OR IEC 6F22)
IPM-100 INDUSTRIAL PROCESS CONTROL LOOP CALIBRATOR
Source
Typical Drive Capability
Compliance Voltage
Read
Voltage Burden
700 Ω @ 20.00 mA with fresh batteries
300 Ω @ 20.00 mA with low batteries
Nominal 14 VDC @ 20 mA with fresh batteries
0.9 VDC @ 4 mA, 1.2 [email protected] 20mA, 1.9 VDC @ 24
mA excitation current
Automatic Power-Off: After 30 minutes of inactivity
APPLICATIONS:
• Used to calibrate, check and measure all current signal instruments in a
4mA to 20mA dc loop.
• Used to simulate a 2-wire transmitter.
CODE: DSE00C00IPM-1000010
MANUAL NO. :10
NAGMAN TEMPERATURE CALIBRATOR (BATH) ETC -600
TYPE : TEMPERATURE CALIBRATOR
NAGMAN TEMPERATURE CALIBRATOR (BATH) ETC -600
TYPE : TEMPERATURE CALIBRATOR
Specifications :
Dimension (W x L x H )
270 x 124 x 240 mm
Instrument Weight
7 Kgs
Well - Immersion Depth
120 mm
Well Diameter
25 mm
Range
50°C to 600°C
Accuracy
±0.1% F.S. + 1 digit
Temperature Readout
°C/°F Switchable
Heating Time (Typical Ambient to 600°C)
18 Minutes (approx.)
Resolution
1°C
Power supply
230V ac/50Hz/1200Watts (maximum)
Cooling Time (600°C to 100°C)
60 Minutes (approx.)
Stability
±0.05°C
Analog Output (F.S.)
Digital Display
10 mV / °C
3½Digit LED
NAGMAN TEMPERATURE CALIBRATOR (BATH) ETC -600
TYPE : TEMPERATURE CALIBRATOR
APPLICATIONS: Used To Calibrate Temperature.
CODE: DSE00T00ETC6000011
MANUAL NO.: 11
OMEGA OMNICAL-8A-110
TYPE: TEMPERATURE CALIBRATOR
OMEGA OMNICAL-8A-110
TYPE: TEMPERATURE CALIBRATOR
BATTERY MODE
BATTERY LIFE
RECHARGE MODE
AND
RECAHRGE SOURCE
OPERATE POWER
8hours continuous operation minimum for NiCad battery
pack
Minimum 1000charge/discharge cycles
14 to 16 hours for full charge. Instrument can be
operated continuously and indefinitely while on recharge.
110v ac or 220v ac, ±20%,47 to 400Hz. Includes line
cord for specified power source.
1.5watts(Battery mode)
PRIMARY/SECONDARY TRANSFORMER ISOLATION
WEIGHT
1400V ac.
9.5 lb (4.3kg)
DIMENSIONS
8̎(203mm)×12.5̎(317mm)×6̎(152mm)
OPERATING TEMPERATURE
STORAGE TEMPERATURE
RELATIVE HUMIDITY
14° to 104°F(-10° to +40°C)
-40° to +140°F(-40° to +60°C)
0-90% non-condensing.
OMEGA OMNICAL-8A-110
TYPE: TEMPERATURE CALIBRATOR
THERMOCOUPLE MEASUREMENT
THERMOCOUPLE
TYPE(metals)
SCALE
TEMPERATURE RANGE
RESOLUTIO
N
CONFORMIT
Y
J(Fe/CuNi)
°F
°C
°F
°C
°F
°C
-346 to +2192
-210 to +1200
-328to +2501
-200 to +1372
-337 to +752
-405 to -337
-205 to +400
-243 to +205
--389 to +1832
-422 to -389
-234 to +1000
-252 to -234
-58 to +3214
-50 to +1768
-58 to +3214
-50 to +1768
+32 to +2192
+2192 to+3812
+3812 to +4200
0 to+1200
+1200 to+2100
+2100 to+2315
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
1
1
1
1
1
1
1
1
1
1
±0.2
±0.1
±0.2
±0.1
±0.2
±9.2
±0.1
±5.1
±0.2
±5.0
±0.1
±2.8
±0.2
±0.1
±0.2
±0.1
±0.2
±0.2
±0.2
±0.1
±0.1
±0.1
K(NiCr/NiAl)
T(Cu/CuNi)
E(NiCr /CuNi)
°F
°C
R(Pt13Rh/Pt)
°F
°C
°F
°C
°F
°C
S(Pt10Rh/Pt)
C(W5Re /W26Re)
RSS
CALIBRATION
ACCURACY
±0.29
±0.17
±0.33
±0.19
±0.25
±9.11
±0.15
±5.10
±0.27
±4.81
±0.16
±2.71
±0.76
±0.65
±0.61
±0.54
±0.68
±0.84
±0.97
±0.56
±0.62
±0.69
OVERALL ACCURACY
1YEAR
±1
±0.6
±1
±0.6
±1
±10
±0.6
±5.6
±1
±5
±0.6
±2.8
±2
±1.1
±2
±1.1
±1
±2
±3
±0.6
±1.1
±1.7
OMEGA OMNICAL-8A-110
TYPE: TEMPERATURE CALIBRATOR
mV MEASUREMENT
Voltage range
Resolution
RSS calibration
Overall accuracy 1year
-20.000mV to +20.000mV
0.001mV
-101.10mV to -20.00mV;
+20.00mV to +101.10mV
0.01mV
±(0.005% of reading
+2digits )
±(0.01% of reading +1digit
)
±(0.003% of reading
+5digits )
±(0.003% of reading
+2digits )
APPLICATIONS: Used For Temperature Calibrating And mV Measurement.
CODE: DSE00T0008A1100008
MANUAL NO.:08
PASCOLO DIGITAL PRESSURE GAUGE
CH-8404
INSTRUMENT
PASCOLO DIGITAL PRESSURE GAUGE
CH-8404
Specifications
Ranges
-1…2 bar
Resolution
1 mbar
Total Accuracy (0…50°C)
typ. 2 mbar max.5 mbar
Accuracy
0.1% FS type
Operating Temperature
-10…80°C
Storage Temperature
-30…80°C
Compensated Temperature Range
0…50°C
Stability of Transducer (6 months)
0.1 %FS type / 0.2 %FS max.
Supply
Integrated 3 V Battery Type CR 2430
Battery Life
360 hours
Pressure Connection
7/16”-20 UNF (Adapter G 1/4” included)
Protection, CEI 529
IP67
Diameter x Height x Depth
59 x 95 x 32 mm
Weight
125 g
Hose
1.5m, to withstand pressures up to 1000 bar
Comes with a 7/16”-20 UNF male thread
PASCOLO DIGITAL PRESSURE GAUGE
CH-8404
TYPE: PRESSURE MEASURING INSTRUMENT
APPLICATIONS: Used as a low power signal processor , transmits the signals of the pressure
and temperature sensor to the microprocessor.
CODE: DSE00P00CH-84040018.
MANUAL NO.: 18.
PRECISION PRESSURE CALIBRAOR
700 BAR/10000PSI
TYPE: PRECISON PRESSURE CALIBRATOR
PRECISION PRESSURE CALIBRAOR
700 BAR/10000PSI
TYPE: PRECISON PRESSURE CALIBRATOR
FUNCTIONAL SPECIFICATIONS
Accuracy
+0.05% of full scale.
Range
700 BAR, 10000PSI
Overload Maximum
+50%F.S.
Overload Alarm
Activates at +20% Full Scale
Resolution
Minimum 5 digits
Display
2 Line, 16-character LCD
Display Update
Twice per second
mA Input
4-20mA or 10-50mA
Voltage Input
0-30V
Pressure Switch
Supply maximum 9Vdc
PRECISION PRESSURE CALIBRAOR
700 BAR/10000PSI
TYPE: PRECISON PRESSURE CALIBRATOR
PHYSICAL SPECIFICATIONS
Temperature Compensation
0°C to 50°C
Battery
9Vdc Alkaline
Approximately 15-hours battery life
Low Battery Indicator at 7 volts
Weight
400g
Dimensions (L x W x H)
160 × 90 × 45 mm
APPLICATIONS: USED FOR PRESSURE CALIBRATION.
CODE: DSEPPC00700BAR0021
MANUAL NO.: 21
PRESSURE CALIBRATOR- MPCE
TYPE: PRESSURE CALIBRATOR.
PRESSURE CALIBRATOR- MPCE
TYPE: PRESSURE CALIBRATOR.
SPECIFICATIONS
Ranges
0 to 70 / 100 / 200 / 350
Resolution
Minimum 4½ digits
Accuracy
±0.1% F.S.
Display
2 line, 16 characters LCD (Alpha-Numeric)
mA Input Instrument
0 to 25 mA DC
Engineering Units
Pressure Switch Test
Mpa, At m, Bar, Kg/cm², mH2O, PSI, in Hg, ftH2O,
Kpa, inH2O, mmHg, cmH2O
Bar Maximum 5 VDC
Over Pressure Alarm
105% FS (Audio / Visual)
Pressure Generation
Battery Type & Life
Thro‟ matching Hydraulic Hand Pump with Fine
Adjustment Vernier & Release Valve
9 Volts ; Approx. 12 Hours
Low Battery Indication
Provided
Dimension(L x D x H)
480 x 260 x 150 mm
Weight
7 Kg.
PRESSURE CALIBRATOR- MPCE
TYPE: PRESSURE CALIBRATOR.
APPLICATIONS: Used for pressure calibration.
CODE: DSE00P0000MPCE0020.
MANUAL NO.: 20.
RACAL-DANA 1991 NANOSECOND UNIVERSAL COUNTER
TYPE: UNIVERSAL COUNTER
RACAL-DANA 1991 NANOSECOND UNIVERSAL COUNTER
TYPE: UNIVERSAL COUNTER
GENERAL SPECIFICATIONS
Automatically determined by resolution selected(range 1ms-10s)
GATE TIME
(FREQUENCY,PERIOD AND
RATIO MODES)
DISPLAY
WEIGHT
DIMENSIONS
VIBRATION
SHOCK
Resolution selected
Gate time (seconds)
9+overflow
10
9
1
8
0.1
7
0.01
6,5,4,3
0.001
9 digit high brightness 14mmLED display in engineering format with
exponent digit.
Net 3.63kg(8lb)
331×212×88mm
2g
30g
PARAMETER
STORAGE
OPERATING
Temperature
-40°C to +75°C at 75% RH
0°C to +50% C
Altitude
12000meters
3050meters
RACAL-DANA 1991 NANOSECOND UNIVERSAL COUNTER
TYPE: UNIVERSAL COUNTER
FREQUENCY RANGE
SENSITIVITY
SIGNAL OPERATING RANGE
INPUT IMPEDANCE
(×1 and ×10attenuation)
MAXIMUM INPUT
(WITHOUT DAMAGE)
INPUT CHARACTERISTICS:
input a
input b
sine wave
pulse
×1 attenuation
×10 attenuation
Separation mode
Common mode
50 OHMS
1Meg OHM
(×1 attenuation)
1Meg OHM
(×10 attenuation)
DC to 160 MHz DC-coupled
10 Hz to 160 MHz AC coupled
DC to 100 MHz DC-coupled
10 Hz to 100 MHz AC coupled
25mVrms DC to 100MHz
50mVrms to 160 MHz
75mV p-p, 5ns min. width
±5.1V
±51V
50 ohms or 1 Meg Ohm//≤45 pF
50 ohms or 1 Meg Ohm//≤55 pF
5V (DC+AC rms)
260V (DC+AC rms), DC TO 2 kHz
DECREASING TO 5V RMS AT
100 kHz AND ABOVE
260V (DC+AC rms), DC TO 2
kHz
DECREASING TO 50V RMS AT
100 kHz AND ABOVE
RACAL-DANA 1991 NANOSECOND UNIVERSAL COUNTER
TYPE: UNIVERSAL COUNTER
Coupling
Low pass filter
Trigger slope
Attenuator
Dynamic range
(×1 attenuation)
Trigger level range
Manual:
X1 attenuation:
X10 attenuation:
Automatic:
Trigger level accuracy:
Manual and Automation:
X1 attenuation:
X10 attenuation:
AC or DC
50kHz nominal
+ve or –ve
×1 or ×10. In auto trigger mode attenuator selected
automatically if necessary
75mV p-p to 5V p-p to 50MHz
75mV p-p to 2.5V p-p to 100MHz
150mV p-p to 2.5V p-p to 160MHz
±5.1V in 20 mV steps
±51V in 200 mV steps
±51V
±30 mV ±1% of Trigger level reading
±300 mV ±1% of Trigger level reading
RACAL-DANA 1991 NANOSECOND UNIVERSAL COUNTER
TYPE: UNIVERSAL COUNTER
Auto Trigger:
Frequency range:
Min. Amplitude(AC):
X10 attenuator
DC and 50 Hz to 100 MHz
(typically 160 MHz)
Typically 150 mVp-p*
Automatically selected if input signal exceeds
±5.1V or 5,1Vp-p*
Trigger level outputs:
(real panel)
Range:
Accuracy(relative to true Trigger level)
X1 attenuation:
X10 attenuation:
impedance
±5.1V
±1% V output ±10mV
±1% V output ±100mV
10 Kohm nominal
POWER REQUIREMENTS
VOLTAGE
FREQUENCY
90-110 Vac
103-127 Vac
193-227 Vac
207-253Vac
45-450Hz
RATING
35 VA max
RACAL-DANA 1991 NANOSECOND UNIVERSAL COUNTER
TYPE: UNIVERSAL COUNTER
APPLICATIONS: Used for measurement of functions like frequency, period , time of interval, phase and
ratio.
CODE: DSE0UC000019910025.
MANUAL NO. : 25
TAYLOR SQUEEZE GAUGE 44S-390
TYPE: PRESSURE CALIBRATOR
TAYLOR SQUEEZE GAUGE 44S-390
TYPE: PRESSURE CALIBRATOR
Specifications
RANGE
ACCURACY
0 to 18 psig
±1.5%
PRODUCT TYPE
WEIGHT
Pressure/Vacuum Calibrators
8 oz
APPLICATIONS: Used to calibrate Pneumatic controllers, Low
transmitters, Flow transmitters
CODE: DSE00P0044S-3900014.
MANUAL NO.:14.
pressure gauges, pneumatic
TYPE: TEMPERATURE CONTROLLER
TYPE: TEMPERATURE CONTROLLER
GENERAL SPECIFICATIONS
Input power
90-125,180-250Vac 50-60Hz
Size H×W×D
Weight
88×215×280mm
(3.5̎×8.5̎×11̎)
2.9kg
Operating temperature
0°C to +50°C
Storage temperature
-40°C to +70°C
Humidity
<90% relative , non-condensing
Connectors
OUTPUT
Analog output
Analog interface card(optional)
15-pin,
D-sub(female)
Isolated BNC
15-pin,high density D-sub(male)
TYPE: TEMPERATURE CONTROLLER
DISPLA
Y
TEMPERATURE
RESISTANCE(10A)
RESISTANCE(100A)
TE Current
RANGE
-50.0 to +150.0°C
0.1KΩ to 199.9KΩ
0.01KΩ to 19.99KΩ
-5.00Amps to +5.00Amps.
RESOLUTION
0.1°C
100Ω
10 Ω
10mA
ACCURACY
±0.2°C
±200 Ω
±20 Ω
±20mA
TEMPERATURE SENSORS
SENSOR TYPES
THERMISTOR(NTC-2wire)
AD590/AD592
LM335
TEMPERATURE RESOLUTION
0.1°C
0.1°C
0.1°C
TEMPERATURE ACCURACY(25°C)
0.05°C
0.5°C
3.0°C
SENSOR BIAS
10A/100A
+12V
1mA
TYPE: TEMPERATURE CONTROLLER
OUTPUT
VALUE
Maximum current
5Amps
Compliance voltage
10 Volts
Available output power
50 Watts
Current limit Range
Accuracy
Ripple/noise(rms)
0to5A
<2mA
Stability
Short term (10to30min)
Long term (24hour period)
External fan power
0.005°C
0.01°C
12v 800mA max
APPLICATIONS: Used for providing superior temperature stability and wide range of power outputs.
CODE: DSE00T00TC-16000005
DSE00T00TC-20100005
MANUAL NO.: 05
THANDAR TM351 DIGITAL MULTIMETER
TYPE: MULTIMETER
THANDAR TM351 DIGITAL MULTIMETER
TYPE: MULTIMETER
GENERAL SPECIFICATIONS
Maximum Common Mode Voltage
1000V DC or AC peak
Power requirements
Six alkaline or zinc carbon „C‟ cells
Environmental operating range
5°C to 4°C, <80% RH
Environmental storage range
-40°C to 60°C(with alkaline batteries) 75%RH
Dimensions
250 × 150 × 50 mm(excluding handle)
Weight
800gms excluding batteries; 1200gms including
alkaline batteries.
THANDAR TM351 DIGITAL MULTIMETER
DETAILS
PARAMETER
DC Voltage
AC Voltage
DC Current
AC Current
Resistance and
diode check
RANGE
RESOLUTION
200mV
2V
20V
200V
1000V
200mV
2V
20V
200V
750V
200μA
2mA
20mA
200mA
2000mA
10A
200μA
2mA
20mA
200mA
2000mA
10A
200Ω
2KΩ
20KΩ
200KΩ
2000KΩ
20MΩ
100μV
1mV
10mV
100mV
1V
100μV
1mV
10mV
100mV
1V
100nA
1μA
10μA
100μA
1mA
10mA
100nA
1μA
10μA
100μA
1mA
10mA
100mΩ
1Ω
10Ω
100Ω
1KΩ
10KΩ
ACCURACY
±(0.1% of reading+ 1 digit)
±(5% rdg +5 d)
±(0.5% rdg +2 ±(1.5% rdg +2
d)
d)
±(0.5% rdg +2d) at 50/60 Hz
±(0.3% rdg + 1d)
±(2% rdg + 2d)
±(1% rdg + 2d)
±(2.5% rdg + 2d)
±(0.2% rdg + 1d)
±(1% rdg + 1d)
THANDAR TM351 DIGITAL MULTIMETER
TYPE: MULTIMETER
APPLICATIONS: Used for different measurements.
CODE: DSE0MM000TM3510024.
MANUAL NO.: 24.
UNIVERSAL CALIBRATOR-3001M
TYPE: MULTIFUNCTION CALIBRATOR
UNIVERSAL CALIBRATOR-3001M
TYPE: TEMPERATURE CALIBRATOR
TEMPERATURE
RANGE
RESOLUTION
ACCURACY
Ambient
0 to 50°C
0.1°C
± 0.2°C
B Type T/C
450°C to 1750°C
0.1°C
± 0.2°C
E Type T/C
-100°C to 1000°C
0.1°C
± 0.5°C
J Type T/C
0°C to 760°C
0.1°C
± 0.5°C
K Type T/C
0°C to 1370°C
0.1°C
± 0.5°C
R Type T/C
200°C to 1750°C
0.1°C
± 0.5°C
S Type T/C
0°C to 1750°C
0.1°C
± 0.5°C
T Type T/C
-200°C to 850°C
0.1°C
± 0.5°C
CU-53
-100°C to 200°C
0.1°C
± 0.5°C
PT-100 RTD
-200°C to 850°C
0.1°C
± 0.5°C
UNIVERSAL CALIBRATOR-3001M
TYPE: TEMPERATURE CALIBRATOR
VOLTAGE
RANGE
RESOLUTION
ACCURACY
20mV
-5mV to +25 mV
1μV
±0.03%FS
200 mV
-50 mV to +250 mV
10μV
±0.03%FS
2V
-0.5V to +2.5V
0.1mV
±0.03%FS
10V
-2.5V to +10V
1mV
±0.03%FS
CURRENT
RANGE
RESOLUTION
ACCURACY
20mA
0 to +25mA
1μA
±0.03%FS
60mA
0 to +60mA
10μA
±0.03%FS
RESISTANCE
2K ohm
RANGE
0 to 2K ohm
RESOLUTION
0.1 ohm
ACCURACY
±0.03%FS
UNIVERSAL CALIBRATOR-3001M
TYPE: TEMPERATURE CALIBRATOR
DIMENSIONS: 240(W)× 135(H)×185(D)mm.
POWER SUPPLY:
EXTERNAL POWER SUPPLY: 230/110V±10%,50Hz/60Hz, OR BUILT IN RECHARGEABLE BATTERY.
INTERNAL POWER SUPPLY: RECHARGEABLE BATTERY 7.2V,2.2AH POWER PACK.
COLDJUNCTION COMPENSATION: automation from 0°C to 50°C in both measurement and simulation.
CJC switch off with 0°C reference provided.
DISPLAY: 5 Digit 0.56” RED high efficiency LED
Discrete Leds to indicate selected function, battery low and charging status indication.
INPUT IMPEDANCE: more than 10Mohm(mV, V,T/C)
10 ohm in mA
OPERATING TEMPERATURE: 0 TO 55°C.
HUMIDITY RANGE: 0 TO 80%R.H. non-condensing.
TEMPERATURE COEFFICIENT: +0.002% rdg/°C.
POWER CONSUPTION: 10VA.
APPLICATIONS:
•capable of measuring the output of or simulating thermocouple types B,E,G,K,R,S,T(ANSI STANDARD)
•capable of measuring output of Pt-100 and Cu-53 RTD and 2K Resistance. Simulation of resistance and RTD by simulator
module.
CODE: DSE00T0003001M0002.
MANUAL NO.:02.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF