INSTRUCTION MANUAL

DM-500 IHB

AUTOMATIC UPPER ARM

BLOOD PRESSURE AND PULSE MONITOR

EN

0197

5127

CONTENTS

1. INTRODUCTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

1.1 Device features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

1.2 Important information on self-monitoring . . . . . . . . . . . . . . . . . . 1

2. IMPORTANT INFORMATION ON BLOOD PRESSURE AND ITS

MEASUREMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

2.1 How is hypertension / hypotension developed? . . . . . . . . . . . . . . 2

2.2 What is the correct pressure value?. . . . . . . . . . . . . . . . . . . . . . . 2

3. PRESSURE MONITOR DESIGN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

4. STARTING THE DEVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

4.1 Installing battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

4.2 Battery life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

4.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

4.4 User selection and setting date and time . . . . . . . . . . . . . . . . . . . 6

5. TAKING MEASUREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

5.1 Before measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

5.2 Most frequent errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

5.3 Wrapping the cuff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

5.4 Body posture during measurement . . . . . . . . . . . . . . . . . . . . . . . 9

5.5 Measurement procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

5.6 Finishing the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

6. MEMORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

6.1 Full memory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

6.2 Deleting all measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

7. ERROR MESSAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

8. MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

9. WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

10. SAFETY AND DISPOSAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

11. SYMBOLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

12.TECHNICAL DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Thank you for buying the blood pressure and pulse monitor DM-500 IHB. The model can be used with irregular pulse. If the device detects irregular pulse, the symbol   appears on the display. In such a case, it is advisable to consult your physician.

Please carefully read this user manual before the first use of the device.

Please keep the user manual. The information contained herein may be needed in the future.

1. INTRODUCTION

1.1. Device features

It DM-500 IHB is a fully automatic digital pressure measuring device on his arm, which allows you to perform quick and reliable measurement of systolic and diastolic pressure and pulse rate, using the method oscylometrycznej.

The device provides a very high accuracy of measurement and was designed to be as user-friendly as possible.

The device is intended for taking blood pressure measurements at home.

For more information on blood pressure and its measurement, please contact your doctor.

1.2 important information on self-measurement

PAGE  Using a cuff other than the recommended one may result in measuring error.

Do not use the device for measuring blood pressure in infants.

Do not use the device in pregnant patients in pre-eclampsia.

Pay attention not to entangle the tubing because this may result in a serious injury of the patient or disturbances in blood pressure measurement.

Too frequent measurements may cause trauma to the patient due to impaired blood flow.

Wrapping the cuff on a wound may lead to a deterioration of its condition.

Application of the cuff on the treated arm may lead to injury as a result of temporary obstruction of blood flow during pressure increase.

Do not put on and inflate the cuff, on the side where the mastectomy procedure has been performed.

Inflation of the cuff may cause temporary stoppage of equipment monitoring vital functions used on the same arm.

Pressure measurement using the automatic device for measuring blood pressure does not cause long-term impairment of the patient's circulation.

PAGE 1

The device is not suitable for simultaneous use with high frequency

(HF) electrosurgical equipment

Self-measurement means control and not diagnosis and treatment.

Unusual values should always be consulted with your doctor.

You should under no circumstances change the doses of medications prescribed by your doctor.

The displayed pulse rate is not suitable for controlling the operating

 frequency of a pacemaker!

In the case of arrhythmias, the measurement made using the device should be consulted with a doctor.

Electromagnetic interference

The device contains sensitive electronic components, therefore, one should avoid strong electrical or electromagnetic fields (e.g., nearby cellular phones, microwave ovens). Otherwise, there may be a temporary deterioration in the accuracy measurements

2. IMPORTANT INFORMATION ON BLOOD PRESSURE AND ITS

MEASUREMENT

2.1 How is hypertension / hypotension developed?

The level of blood pressure is regulated in the brain, in the circulatory center and adapted to the current conditions based on feedback involving the nervous system. To adjust the blood pressure, the frequency and the strength of heart contractions and the diameter of blood vessels (the degree of contraction of smooth muscle of blood vessel walls). The level of blood pressure changes periodically in the cardiac cycle: during the contraction the value is the highest (systolic) and at the end of the diastole the value is the lowest (diastolic pressure). In order to prevent the development of dangerous diseases, the blood pressure values should be correct.

2.2 What is the correct pressure value?

The value of blood pressure is too high if the diastolic pressure at rest is above

90 mmHg or the systolic pressure is over 160 mmHg. In such a case, you should immediately consult your doctor. Long-term persistence of pressure on such a level endangers human health due to the increased damage to bloodvessels.

If systolic pressure is within the range of 140 to 160 mmHg or the diastolic pressure is between 90 to 100 mmHg, consult your doctor. Subsequently, regular self-measurement will be necessary.

In the case of values that are too low, that is the systolic pressure is below

PAGE 2

100 mmHg or the diastolic pressure falls below 60 mmHg, you should also consult your doctor. Even in the case of pressure values in the normal range, it is recommended to perform regular blood pressure self-measurements.

This allows for detecting any changes in the value of blood pressure at an early stage and respond accordingly. If the patient is undergoing treatment for hypertension/hypotension, regular measurements should be taken at a specific time of day and the results recorded, and then presented to the doctor.

Never use the obtained results to individually change dosage of medications prescribed by your doctor.

Table of blood pressure value classification (unit: mmHg) according to the

World Health Organization (WHO):

Range Systolic pressure

Diastolic pressure

Remedial measures

Hypotension below 100 below 60 Consult your doctor

Optimal blood pressure from

100 to

120 from

60 to

80 Self-measurement

Normal blood pressure from

120 to

130 from

80 to

85 Self-measurement

Slightly elevated blood pressure from

130 to

140 from

85 to

90 Consult your doctor

Too high blood pressure from

140 to

160 from

90 to

100 Imperative contact your doctor

Significantly elevated from

160 to

180 from

100 to

110 Imperative contact blood pressure your doctor

Dangerously high blood pressure

Above 180 Above 110 Immediately contact your doctor!

 l If the values of your blood pressure at rest are usually normal, but elevated during stress, you may suffer from labile (latent) hypertension. If you suspect that this might be possible, contact your doctor.

Correctly measured diastolic pressure above 120 mmHg requires immediate medical treatment.

PAGE 3

3. PRESSURE

MONITOR DESIGN

LCD display cuff air tube connector user change button air tube socket button

START/STOP

CUFF

TIME button

MEMORY button

(Applied part type BF)

Arm circumference range:

22-42 cm

LCD DISPLAY

users

/ groups blood pressure classification indicator according to the World

Health Organization (WHO) number of the stored measurement average diastolic pressure exhausted battery symbol

PAGE 4 irregular heart rate symbol air tube connecting plug with the blood pressure monitor (insert in the air tube connector) air tube metal buckle date / time diastolic pressure measurement value symbo pulse

4. STARTING THE DEVICE

4.1 Installing battery

1. Remove the battery cover.

2. Insert 4 standard AA alkaline batteries.

Use batteries of the same brand.

Note that all the batteries are properly installed, observing polarity.

3. Reinstall the battery cover.

4. If the battery icon   is displayed on the screen, it means that there is 20% power left until the battery is drained completely

5. If the battery icon   is displayed on the screen, it indicats low batteries. Batteries must be replaced, otherwise the device may fail to

 operate properly.

Do not mix old and new batteries.

After replacing batteries, you must reset the time and date.

After the battery warning icon is displayed, the device will not turn

 on until the battery is replaced.

Use batteries of AA Long-Life type or alkaline 1.5 V. It is not recommended to use 1.2 V rechargeable batteries.

If the pressure monitor is left without use for an extended period of

.

time, you should remove the batteries

4.2 BATTERY LIFE

Four new LR6 (AA) batteries last for approximately 200 measurements (1 per day, at room temperature 23°C), battery life varies depending on the temperature in which they are used, and may be shorter at lower temperatures.

You can check the battery status in the lower left corner of the screen. If the low battery symbol   , is displayed, they should be replaced with new ones.

4.3 Power adapter (Optional)

1. Connect the plug of the power cord into the power supply connector.

2. Plug the power adapter unit into electrical outlet.

Use power adapter suitable for local mains voltage.

Power adapter specification: 100~240 V, 50/60Hz; output: 6V, min.

600mA

We recommend using only the power adapter supplied by the manufacturer, model Diagnostic ZID 6-1 (100~240 V,

50/60Hz, 6 V, 1000 mA (1 A))

If the device is defective, unplug the power supply or the power cord.

PAGE 5

Do not touch the power adapter with a wet hand.

Do not tangle the wires during usage.

The power adapter is added to the set optionally

(charged additionally) power supply

4.4 User selection and setting date and time

User selection: The blood pressure monitor allows you to track blood pressure readings of 2 users.

a) Before starting the measurement, make sure that the appropriate user is set. The device can track the results of up to 2 users (user 1, user 2).

b) Hold down the TIME button for at least 3 seconds. The screen will display a blinking user icon. Change the user by pressing the memory button (M). To confirm user selection, press the START/STOP button.

c) We recommend that the first person who takes measurement is user 1.

QUICK USER SELECTION:

After pressing the user button, the symbol  . will flash. Press again to change the user.

TIME AND DATE SETTINGS:

The device has an integrated clock and displays the date. This permits saving not only the result of blood pressure measurement, but also the exact date and time of taking the reading. After inserting the new batteries, the CLOCK will be set to 12:00 and the DATE to 1-01. You must then set the correct time and date. For this purpose, please do the following.

1. Hold down the TIME button for at least 3 seconds. The user icon starts blinking. Next, press the TIME button again to display the year (4 characters flashing).

2. Enter the year by pressing the MEMORY button.

3. Press the TIME button again. Now the date with the flashing month icon appears on the screen.

4. Set the month using the MEMORY button.

5. Press the TIME button again. Now the last two characters will flash

(day).

6. Set the day using the MEMORY button.

7. Press the TIME button again. Now the system switches to time settings; the hours character will flash.

8. Set the hour using the MEMORY button.

9. Press the TIME button again. Now the last two characters will flash

(minutes).

PAGE 6

10. Set the exact time, i.e. minutes, using the MEMORY button.

11. Press the TIME button: the measurement unit will start flashing.

12. Press the MEMORY button to set the unit of measurement (mmHg or kPa)

13. After completing the settings, press TIME. Now the settings are confirmed and the timer starts running.

14. After completing all the settings, press the TIME button once again.

The device will briefly display the date and time. Now the entered settings are confirmed and the clock starts running.

After pressing the TIME and MEMORY buttons, data is entered (e.g. switching from hours to minutes or changing the value by +1). After pressing and holding the button, the switching is much quicker.

5. TAKING MEASUREMENTS

5.1 Prior to measurement

Directly prior to measurement one should not: eat, smoke and avoid physical effort because all these activities have an impact on

 measurement results. Prior to measurement you should relax, sitting on a chair in a quiet environment for approximately 10 minutes.

Measurements should always be taken on the same arm (normally left).

Take measurements on a regular basis, every day at the same time, because blood pressure varies throughout the day.

5.2 Most frequent errors

For blood pressure measurements to be comparable, the same measurement conditions are necessary! (these conditions always include peaceful surroundings).

All the patient's efforts to support the arm may result in increased blood pressure. Select a comfortable and relaxed position. During

 the measurement, do not stretch any muscles of the arm on which the cuff is wrapped. If necessary, use a pillow as a support.

The operation of the pressure monitor may be disturbed by extreme temperatures, humidity and taking measurements at high altitudes.

Pay attention not to pinch or bend the tubes.

A loosely fitting cuff will cause incorrect measurement results

PAGE 7

In the case of repeated measurements there is a build-up of blood in the arm, leading to incorrect results. For this reason, the correct blood pressure measurement should be carried out after a 5 minute break or after lifting the arm, to facilitate the outflow of the accumulated blood (after at least 3 minutes).

5.3 Wrapping the cuff

Insert the tip of the air duct firmly in the opening on the left side of the unit (air tube socket).

2. Insert the end of the cuff under the metal buckle, with the velcro facing out.

3. Wrap the cuff approximately 2-3 cm above the elbow. For best results, wrap the cuff on bare skin, at heart level.

4. The compression of arm caused by tucked up sleeve may prevent accurate reading.

2-3 cm

The air hose should be placed in the center of the arm.

Palm facing up

5. The cuff should be wrapped easily on the shoulder and the Velcro should fasten easily.

6. After wrapping the cuff, make sure that there is sufficient space under the cuff to fit a finger.

7. If the cuff does not fit on the arm, the accuracy of measurements may be incorrect.

8. Your feet should not be crossed and rest flat on the floor; shoulders and arms should be supported.

Do not fold the cuff or the air tube.

To disconnect the cuff, remove the air tube plug from the device.

Measurement can be started only after wrapping the cuff properly.

The cuff must be replaced if there is a leak or when the cuff is not operating properly.

In order to ensure the accuracy of readings, you should only use the cuff supplied by the manufacturer.

PAGE 8

5.4 Body posture during measurement

Relax, rest the elbow on the table with palm facing up; the cuff should be at heart level. Accuracy of readings may be reduced if the cuff is not wrapped properly. The arm should be at the same height as the heart. If the arm is too low, the reading results will be too high. If the arm is too high, the reading results will be too low.

The arm should be positioned at the same height as the heart.

5.5 Measurement procedure

After wrapping the cuff properly, you can start taking the measurement.

a) Press the START/STOP button, all the elements will be displayed on the screen (FIG. 1), the cuff will start inflating. The increasing cuff pressure is displayed continuously. (FIG. 2) b) When the appropriate pressure is reached, it starts to slowly decrease.

When the pulse is detected, the heart icon will start blinking on the screen (FIG. 3) c) After completing the measurement, the values of systolic and diastolic pressure and the pulse rate appear on the screen (FIG. 4)

Example (FIG. 4): Systolic pressure 118, diastolic pressure 73, Pulse rate 75

The results of the measurement will be displayed until the device is turned off. If no button is pressed within 3 minutes, the unit will automatically turn off to save battery power.

FIG.1

FIG.2

FiG.3

FIG. 4

5.6 Finishing the measuremen

To interrupt blood pressure measurement (e.g. if the patient is feeling unwell), you can press the START/STOP button at any time. The device will automatically reduce the pressure of the cuff.

PAGE 9

6 MEMORY

Internal memory stores up to 2x120 of measurement results.

1) Recalling results from memory

To access memory, press the MEMORY button.

The device will display the average score of 3 most recent measurements, the symbol   without the measurement number

 will appear (FIG. 5).

After pressing the MEMORY button, the user can view the data from the newest to the oldest, the symbol   with measurement number will appear (FIG. 6, FIG. 7).

JIf the   , sign is displayed together with the data stored in the memory, it indicates that irregular pulse rate was detected during the measurement.

FIG. 5 FIG. 6 FIG. 7

6.1 Full memory

Care should be taken not to exceed the available memory. When the memory is full, the old values will automatically be deleted. When the memory capacity has been exhausted, the "memory full" message will display for 1 second.

6.2 Deleting all measurements

Before deleting all the results stored in the memory, make sure that they will not be needed in the future. It is prudent to conduct a measurement log, which allows to provide more information during a visit to the doctor's office. To remove all stored results, press and hold the MEMORY button for at least 5 seconds. Release the button when the screen displays "CL". To permanently delete the entire memory, press the MEMORY button while the "CL" is flashing.

PAGE 10

7. ERROR MESSAGES

If an error occurs during the measurement, the reading will be interrupted and an error code displayed.

Error code Possible cause ERR

ERR 1 No pulse detected.

ERR 2

ERR 3

Measurement results affected by interference.

Cause: there was an arm movement during measurement.

Inflation of the cuff has taken too long. The cuff has not been wrapped properly.

ERR 5 Measurement has indicated unacceptable difference between the systolic and diastolic pressure values. Perform another measurement carefully following the instructions. If unusual results persist, contact your doctor.

Further information. Blood pressure varies even in healthy people, that is why it is important to always take measurements under the same conditions (peaceful environment). If, despite following these principles, the fluctuations will be higher than 15 mmHg and irregular pulse rate occurs repeatedly, consult your doctor. In the event of problems, you should consult with Diagnosis S.A.

Free infoline 800 11 30 70.

YOU SHOULD NEVER ATTEMPT TO REPAIR THE DEVICE YOURSELF! ALL

UNAUTHORIZED ATTEMPTS AT OPENING THE DEVICE WILL VOID THE

WARRANTY!

If, during the use of the device, a problem occurs, please check the following items and undertake the listed remedial measures.

FAULT

The screen remains dark despite turning off the device and inserting new batteries.

REMEDIAL MEASURES

1. Check if batteries are arranged correctly (polarity) and, if necessary, correct their positioning.

2. If the display is incorrect, reinstall the batteries or replace them

The device is frequently unable to measure the pressure or measurement results are too low

(or too high).

The results of each measurement are different, despite the fact that the device is working correctly, and the values are also displayed correctly.

1. Check positioning of the cuff.

2. Take another blood pressure measurement in a quiet and peaceful environment, following the instructions for use.

1. Read the following information and the information included in

"Most frequent errors". Repeat the measurement. Please remember:

Blood pressure varies constantly, which is why subsequent measurements will be characterized by some variability.

The result of blood pressure measurement is different from the one that has been taken by the doctor.

1. Take daily notes of measurement results and consult them with your doctor. Please remember: during a visit to the doctor some people feel nervous, which can raise blood pressure

(relative to the readings taken at home).

PAGE 11

8. MAINTENANCE

Do not expose the device to extreme temperatures, humidity, dust or direct sunlight.

When wrapping the cuff, be careful and avoid deformation by twisting or bending.

Clean the device with a soft and dry cloth.. Do not use gasoline, thinners or similar solvents. Stains on the cuff should be removed with care using a damp cloth and suds. Do not wash the cuff!

Be careful not to drop the device and handle it with care. Avoid strong vibrations.

Do not open the device.

Periodic inspections

The measuring device requires regular inspections.

For that reason, we recommend to carry out periodic inspections of the pressure monitor every 2 years. More information will be provided by Diagnosis S.A. Free infoline 800 11 30 70.

9. WARRANTY

The DM-500 IHB blood pressure monitor is covered by a 2-year warranty starting from the purchase date. The warranty does not cover damage due to improper handling, accidents, non-compliance with the user manual, or changes made in the device by third parties.

The warranty is only valid on presentation of the warranty card.

10. SAFETY AND DISPOSAL

The device may be used only for its intended purpose as described in the user manual. The manufacturer is not liable for damage caused by incorrect use of the device.

The device has sensitive elements and must be handled with care. It is necessary to follow the conditions of storage and use (technical data).

Protect the device from water and moisture, extreme temperatures, impact, dropping, dust, direct sunlight, heat and cold.

Inflate the cuff only after it has been properly wrapped

Do not use the device if it is damaged.

If the device is not used for an extended period of time, remove the batteries.

PAGE 12

Take care to prevent children from using the device without supervision; some parts of the device are small and may be swallowed.

Use only original elements supplied by the manufacturer. The use of other elements may reduce the level of safety.

11. SYMBOLS

SYMBOL FUNCTION/MEANING

0197

Indication of battery polarity

Symbol attesting compliance with the European

Union Directive 93/42/ECC for medical devices

Application part type BF

Product catalog number

Systolic blood pressure in mmHg

Diastolic blood pressure in mmHg

Pulse. Number of beats per minute.

Irregular pulse symbol

Symbol of pulse detected during reading

SN

Rev.

IPX0

Protect against moisture

Keep away from sunlight

Read the user manual before use

Protection against ingress of water

Warnings

Direct curret

Serial number

Manufacturing date

Manufacturer

Date of the last revision

The worn out product should be taken to a waste collection facility. Contains components that are dangerous for the environment. The correct disposal of the device allows to preserve valuable resources and avoid negative effects on health and the environment, which may be threatened by inappropriate handling of waste. If you are in doubt where to return the used appliance, contact Diagnosis. Free infoline 800 70 30 11

PAGE 13

12. TECHNICAL DATA

Measurement Oscillometric Digital

Display method LCD display

Measurement range Pressure 30 to 280 mmHg (± 1 mmHg)

Pulse: 40 do 200 beats per minute

Measurement accuracy Pressure: ±3mmHg

Pulse: ±5% reading

Inflating Automatic pumping device

Deflating Automatically through air valve

Memory function 2 x 120 measurements with date and time

Power supply 4 x AA alkaline batteries or power adapter DC 6 V, min. 600 mA (optional)

Conditions of use 5~40°C, 15%~85% relative humidity

Transportation and storage -10~55°C, 10%~95% relative humidity

Dimensions 135×115×72±1.0 mm

Weight 540 g±5g

Protection against electric shock Internally powered equipment

Safety classification Type BF

Operation Continuous mode

Protection against ingress of water IPX0

Accessories Cuff sized M/L (22-42 cm), 4x batteries AA, user manual, pouch

PAGE 14

Guidelines and manufacturer's declaration - electromagnetic emissions

The devices are intended for use in the electromagnetic environment as described below.

The customer or the user of the device should assure that the device is used in such an environment

Emission test

Fulfillment of requirements

Guidelines regarding electromagnetic environment

The emission of radio frequency waves; CISPR standard

Group 1 The device uses radio-frequency energy only for its internal functions. Therefore, these emissions are very low and should not cause interference in nearby electronic equipment

The emission of radio frequency waves; CISPR standard

Harmonic emissions

IEC 61000-3-2

Group B non applicable

The device can be used in all buildings, including residential buildings, and those that are directly connected to the public low-voltage network, supplying power to buildings intended for residential purposes.

Voltage fluctuations/flicker emissions IEC 61000-3-3

RF - frequency of the electromagnetic spectrum section, which is between the low range of long-wave radio frequencies and the infrared range; frequency useful for radio transmission. 9 kHz and 3 000 GHz are generally accepted as limits

Guidelines and manufacturer's declaration regarding electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the DEVICE should assure that it is used in such an environment.

Immunity test Test level, IEC 60601 standard

Compatibility Electromagnetic environment - guidelines

Electrostatic discharge

(ESD) IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wooden, concrete or made of ceramic tiles. If floors are covered with synthetic materials, the relative humidity should be at least 30%. If ESD interferes with the device, you should consider the use of compensatory elements i.e. wrist strap, grounding.

Fast transient/burst IEC

61000-4-4

± 2 kV for power supply lines

± 1 kV for

Non applicable The quality of power supply should be adequate for typical commercial installation or hospital environment.

Surges IEC 61000-4-5 ± 1 kV differential mode

± 2 kV common mode

Non applicable The quality of power supply should be adequate for typical commercial installation or hospital environment.

Voltage dips, short interruptions and voltage changes on power supply inlets

IEC 61000-4-11

<5 % UT (>95 % clip in

UT) for 0,5 cycle

40 % UT (60 % clip in UT) for 5 cycle

70 % UT (30 % clip in UT) for 25 cycle

<5 % UT (>95 % dip de

UT) dla 5 s

3 A/m

Non applicable The quality of power supply should be adequate for typical commercial installation or hospital environment. If the user [of the device or system] requires continuous use even during power interruptions, it is recommended to connect the device or system to emergency power supply.

Magnetic field of the power supply frequency

(50/60 Hz) IEC 61000-4-8

3 A/m The level of magnetic fields of power sources should be within the limits applicable for typical commercial installations or hospital environment.

Note UT is the alternating voltage (AC) of the power grid prior to the application of the test level.

RF - frequency of the electromagnetic spectrum section, which is between the low range of long-wave radio frequencies and the infrared range; frequency useful for radio transmission. 9 kHz and 3 000 GHz are generally accepted as limits

PAGE 15

GUIDELINES AND MANUFACTURER'S DECLARATION REGARDING ELECTROMAGNETIC IMMUNITY

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the

DEVICE should assure that it is used in such an environment.

Immunity test

Conducted radiofrequency signal IEC

61000-4-6

Emitted radio-frequency signal IEC 61000-4-3

Test level, IEC 60601 standard

3 Vrms

150 kHz do 80 MHz

Compatibility level

3V

3 V/m

80 MHz do 2,5 GHz

3V/m

Electromagnetic environment - guidelines

Portable and mobile radio communication measures should be used at a distance from any of the elements [of the DEVICE or system], including cables, which is not lower than the recommended distance calculated from the transmitter frequency equation.

Recommended distance d = 1.2

d = 1.2 80 MHz to 800 MHz d = 2.3 800 MHz to 2.5 GHz where P is the maximum power rating of the transmitter in watts (W) as specified by the manufacturer, and (d) is the recommended distance in meters (m).

Field strengths from fixed RF transmitters, as determined in field measurements of electromagnetic fields, should be lower than the compatibility level for each frequency range.

Interference may occur in the vicinity of equipment marked with the following symbol:

Recommended distance d = 1.2

Note 1: For 80 MHz and 800 MHz, the higher frequency range is assumed.

Note 2: The provided information does not apply in every situation. The propagation of electromagnetic waves is affected by the absorption and reflection from the surfaces, objects, and people.

(a) Field power from certain transmitters, such as mobile communication base stations, radio transmitters, amateur radio, AM and FM radio transmission and TV transmission cannot be predicted theoretically with accuracy. To assess the electromagnetic environment, tests of local conditions should be considered. If the measured field strength in the location where the DEVICE operates exceeds the appropriate level of compliance, normal operation of the DEVICE should be checked. If improper operation is observed, it may be necessary to take appropriate preventive steps such as moving or relocating the DEVICE.

(b) For frequencies outside the range of 150 kHz to 80 MHz, the field strength should not be higher than 3 V/m.

RF - frequency of the electromagnetic spectrum section, which is between the low range of long-wave radio frequencies and the infrared range; frequency useful for radio transmission. 9 kHz and 3 000 GHz are generally accepted as limits

PAGE 16

Recommended distance between portable and mobile radio communication equipment and the DEVICE

The [DEVICE or SYSTEM] is intended for use in the electromagnetic environment in which the interference caused by the emission of radio waves is controlled. The buyer or the user of the [DEVICE or SYSTEM] can help prevent electromagnetic interference by keeping a minimum distance between portable and mobile radio communication equipment (transmitters) and the [DEVICE or SYSTEM], as recommended below, according to the maximum output power of the communication equipment.

Maximum rated power of the transmitter

W

0,01

0,1

1

150 kHz to 80 MHz d = 1,16

0,12

0,38

1,2

Distance according to frequency of the transmitter m

80 MHz to 800 MHz d = 1,16

0,12

0,38

1,2

800 MHz to 2.5 GHz d = 2,33

0,23

0,73

2,3

10 3,8 3,8 7,3

100 12 12 23

For transmitters assessed at the maximum output power not listed below, the recommended distance d in meters (m) can be estimated using the equation corresponding to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer.

NOTE 1 at 80 MHz and 800 MHz, the distance for the higher frequency range applies.

NOTE 2: these guidelines do not apply in all situations. The propagation of electromagnetic waves is affected by the absorption and reflection from the buildings, objects and people.

Rev. 2016.09.13

przygotowana na podstawie polskiej wersji językowej z dnia 2015.07.17

an

PAGE 17

EN

Diagnosis S.A.

ul. Gen. W. Andersa 38A, 15-113 Białystok, Poland tel./fax 85 732 46 22, 732 40 99 www.diagnosis.pl

store stamp and signature of salesperson

WARRANTY CARD

DEVICE NAME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

MODEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

SERIAL NUMBER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

DATE OF SALE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

WARRANTY TERMS

1. Diagnosis S.A. grants a warranty:

24 months for DIAGNOSTIC blood pressure monitors and cuffs (excluding pump assembly)

Hardware defects revealed during the warranty period will be rectified free of charge within

21 days. The term runs from the date of delivery of the equipment to the service center.

2. The purchaser shall be entitled to replace the equipment for a new one, free of defects, when:

 the repair has not been made within the time limit set in item 1

 an authorized service center found an irreparable manufacturing defect during the warranty period, 4 repairs were effected, and the equipment still shows defects that prevent its use in accordance with its intended purpose.

The concept of repair shall not include operations related to equipment check and cleaning.

3. The warranty shall not cover: batteries, products with illegible or damaged serial number, damage due to the operation and storage inconsistent with the user manual, ingress of liquids or foreign bodies, overvoltage of mains, repairs by unauthorized persons and random events.

4. Faulty equipment should be delivered by the buyer to the address of the main service center or one of the Authorized Service Centers (listed in the appendix).

5. The warranty for the sold consumer goods shall not exclude, restrict, or suspend the powers of the buyer resulting from non-conformity of the goods with the contract.

6. The only basis for the warranty rights shall be the warranty card with the date of sale, stamp and signature of the salesperson. If the card is not completed, filled in wrongly, with traces of corrections and entries made by unauthorized persons, illegible as a result of damage - it shall be invalid.