GW Instek GOS 6112 100MHz Readout Oscilloscope User Manual
Below you will find brief information for 100MHz Readout Oscilloscope GOS 6112. This 100MHz dual-sweep oscilloscope provides two channels for general purpose use. Its microprocessor-based operating system controls most of the instrument functions, including cursor readout and digitized panel settings. This device is equipped with a high intensity 6-inch rectangular type cathode-ray tube with red internal graticule for clear readability of traces even at high sweep speeds. The vertical deflection system has two input channels, each with 11 basic deflection factors from 2mV to 5V per division. The horizontal deflection system provides single, dual or delayed sweeps from 0.5s to 50ns per division (delayed sweep, 50ms to 50ns per division). This oscilloscope also features a 20MHz bandwidth limit to eliminate undesired high-frequency signals. It features a temperature compensation circuit to reduce the drift of the baseline and DC balance.
Advertisement
Advertisement
EN 12687 в/м 134 65 5 100MHz Readout Oscilloscope GOS-6112 User Manual 150-9001 CERTIFIED MANUFACTURER {Ed insTex - All Rights Reserved This manual contains proprietary information, which is protected by copyrights. Al rights are reserved, No part of this manual mey be photocopied. reproduced or translated to another language witionn prior written consent of Gecd Will compar The informatloa In ibis manual was conect at the time of printing However, Good Will continues to improve producis and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice. Good Will instrument Co, Etd. Na. 953-11, Pao-Chung Road, Hsin-Tien City, Taipei Hsien , Taiwan (08-6112 OSCILLOSCOPE USER MANUAL ny CONTENTS PAGE I. PRODUCT INTRODUCTION ‚ 1 1-1.Description...........................iemene sesso 1 1-2.Feature..................... “eee 2 2, TECHNICAL SPECIFICATION......eeccnovorerorar e 4 3. PRECAUTIONS BEFORE OPERATION,..................... 7 3-1.Unpacking the instrument........ e... 7 3-2.Checking the Line Voltage... 7 3-3.Environment... . 8 3-4.Equipment Installation and Operation... avorsasssaeeees В 3-5.CRT Intensity...............reree-r inver neon reorecarióós 8 3-6.Withstanding Voltage of Input Terminals 8 4. PANEL INTRODUCTION........... esoncentacuarooonramnecas Y 1 rrereareaas H 4-2.Rear Panel.............. er... neerenenenirarnecaere nenes 30 5. OPERATION METHOD..........reececerenero reee aa 32 5-1.Readout Display... erenoaartmareroecene sor aro oo rene race 32 52.Connecting Input Signals. eanmarvanaracon meamenaa oe Eme ENs 34 5-3.Adjustment and Checks... "eoceevinarcamanacne ree 39 54.Function Check...............e......erreriaeaoanooeame carre 37 3-5.Basic Operation... x... cree 39 5-6.Measurement Application... .. 47 A anar aa eeorerareróós Si 6-I.Fuse Replacement.,...............e20aeer ara eraeaaan 51 6-2,Line Voltage... ..........nee.enineeo oia rannerre neos SÍ ET an raane cena er cnaee 52 N cena rence 53 GO5-6112 OSCILLOSCOPE USER MANUAL SAFETY TERMS AND SYMBOLS These terms may appear in this manual or on the product: 0 WARNING. Waming statements identify condition or practices that could result in injury or loss of life. practices that could result in damage to this product or | CAUTION. Caution statements identify conditions or other property. The following symbols may appear In this manual or on the product: A A © + DANGER ATTENTION Protective Earth{ground) High Voltage referto Manual Conductor Terminal Terminal GO5-6112 OSCILLOSCOPE USER MANUAL bai ——— FOR UNITED KINGDOM ONLY NOTE: This lead/appliance must only be wired by competent persons WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following cade: Green/ Yellow: Earth Blue: Neutral Brown: Live (Phase) As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows: The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth symbol © or coloured Green or Green & Yellow. The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black, The wire which is coloured Brown must be connected to the terminal marked with the letter L ar P or coloured Brown ar Red. If in doubt, consult the instructions provided with the equipment or contact the supplier, GOS-6112 OSCILLOSCOPE USER MANUAL "EC A This cable/appliance should be protected by a snitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm* should be protected Бу a ЗА ог 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used. Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label. G0OS-6112 OSCILLOSCOPE USER MANUAL ee ey EC Declaration of Conformity We GOOD WILL INSTRUMENT CO, LTD. (1) No. 95-11, Pao-Chung Rd, Hsin-Tien City, Taipei Hsien, Taïwan 2) Plot 522, Lorong Perusahaan Baru 3, Prai Industrial Estate, 13600 Prai, Penang, Malaysia declare, that the below mentioned product GOS-6112 is herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Clectromagnetic Compatibility (89/366/EEC) and Low Voltage Equipment Directive (73/23/EEC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Equipment Directive, the following standards were applied: 61000-3-2 (1995) Voltage Fluctuation N 61000 3-3 (199) _ E toéiromagtetic Capa Genie nit § чат UT be Part 1- Residentiäl Commercial and Light Industry =: EU ВАО | lectrostatic Discharge EN 61000-4-2: 1994 Radiated [Immunity NV 50140: 1993 lectrica] Fast Transients IN 61000-4-4: 1993 urge Immunity 61000-4-5: 1995 aro Dips / Interrupts НМ 61010-1: 1995 GOS-6112 OSCILLOSCOPE USER MANUAL "E e — 1.PRODUCT INTRODUCTION 1-1. Description The GOS-6112 is a 100MHz, two-channel, dual-sweep, portabie oscilloscope for general purpose use. A microprocessor-based operating system controls most of the functions of the instrument, including cursor readout and digitized panel setting. On-screen alphanumeric readout and cursor function for voltage, time, frequency and phase measurement provide extraordinary operational convenience. The vertical deflection system has two input channels. Each channel has 11 basic deflection factors from 2mY to 5V per division. The honzontal deflection system provides single, dual or delayed swecps from 0.5s to 50ns per division (delayved sweep, 50ms to 50ns per division). The trigger system provides stable triggering over the full bandwidth of the vertical deflection system. GO5-6112 OSCILLOSCOPE USER MANUAL I ne —_ 1-2. Features Additionally, the oscilioscope offers several other features: 1) High intensity and internal graticule CRT The oscilloscope employs a high intensity 6-inch retangular type cathode-ray tube with red internal graticule. Tt displays clear readable traces cven at high sweep speeds. Internal graticule lines eliminate parallax-viewing crror between the trace and the graticule line, 2) Temperature compensation The oscilloscope uses a temperature compensation circuit to reduce the drift of base line and DC balance. 3) 20MI1z bandwidth limit When it is hard to observe or trigger a signal because a high-frequency component is superimposed on the signal, use the 20MHz BWL function to reduce the bandwidth of the vertical deflection system and trigger system te 20MHz. 4) TV triggering Exciusive TV sync separator circuit technology provides stable TV signal measurements on fields, frames and lines. 3) Z-axis intensity modulation For applying a blanking signal from an external source, The trace displayed on the screen may be intensity-modulated where pulse signal or time-scale marks are required. {(r08-6112 OSCILLOSCOPE USER MANUAL == zZ. 6) Trigger signai output The signal selected by the TRIGGER SOURCE is available. This output may be used to connect to a frequency counter or other Instrument. 7) Panel setups lock To avold unintentional touch of the setting, the feature is extremely uscful for long term and repetitive measurements that used to be performed under the same test condition of the oscilloscope setting, 8) LED indicator and buzzer alarm The LEDs located in the front panel assist operation and indicated additional information. Incorrect operation and ithe electrical end position of control knobs are indicated by a warning beep. 9) SMD manufacturing technology The instrument 1s built by using the most advanced SMD technology so as 1» reduce the number of internal wiring and shorten the fol route on the pc board. This will also greatly increase the high frequency performance and the reliability of the product. GOS-6112 OSCILLOSCOPE USER MANUAL E— E ——————]];[;—l;—L—L я 2. TECHNICAL SPECIFICATIONS Sensitivity 2mYy-5Y/DIY,I1 step in 1-2-5 sequence Sensitivity Accuracy — 3% (5 DIV at the center display ) Vernier Vertical Sensitivity Continuously variable to 172,5 or less than panel- | indicated value Frequency Bandwidth(-3d8 } De - 100MEZ (2m Y/DiV:DC — 20MHz) Rise Time 3.505 (2mY/DIY:17 55) VERTICAL [Signal Delay Leading edge can be monitored Maximum Input Voltage 400Y (DC+AC) at 1kHz or less DEFLECTION (Input Coupling AC, DC, GND Input Impedance IM (3=2% #/ 25pF approx. SYSTEM Vertical Modes CHI1,CH2,DUAL(CHOP/ALT), ADD(DIFF mode can be established when the CH2 is in the INV mode) CHOP Repetition Frequency | Approx. 250kHz. Polarity (INV) CH? only Bandwidth Limited 20MHz Common-mode Rejection Ration 50:1 or better at 50kHz Dynamic Range 5 div at 60MHz, 5 div at 100MHz. Trigger Modes AUTO, NORM, TV Trigeer Source CHI. CH, LINE, EXT Trigger Coupling AC, DC, HFR, LFR Trigger Slope +/- polarity or TV sync polarity Mode Frequency INT EXT AUTO 10Hz-20MHz D.3SDIY| SOmvpp TRIGGER В 20MHz—100MHz 1.$DIV| 150mVpp Trigger Sensitivity NORM | DC-20MHz 0.35DIV|__50mVp SYSTEM 20MHz7-100MHz 1 4DIV 150mVp TV Sync signal IDIV|_ 200mVp Trigger Level Range INT :+4 DIY or mare > EXT :+0d4 Y or more TV Syne TV-V TV-H Max. External Input Voltage|300Ÿ (DC + AC peak) at | kHz [External Input Impedance — (IMEI 5% //25pF approx. GOS8-6112 OSCILLOSCOPE USER MANTAL HORIZONTAL DEFLECTION Harizontai Modes MAIN(A), ALT, DELAY(B) À (main) Swcep Time 50n5~0.35/D1V, continuously variable (UNCAL) B (delay) Sweep Time SOnS~50mS/DIV Accuracy + 3% (+ 5% at * 10 MAG) Sweep Magnification x 10 (maximum sweep time $nS/DIV) Hold Off time Variable SYSTEM Delay Time lu$-55 Delay Jitter 1720000 or less Alternate Separation Vartable X-axis, Y-axis selectable Mee X-axis: CH1, CH?> 2mY-5Y/DIY £ 3% X.Y Sensitivity Accuracy EXT > 0.1V/DIV = 5% Y-axis: CH1,.CH2> 2mY-3 Y/DIVY + 3% OPERATION — |X-axis Bandwidth DC-500kHz (-3dB) Phase Error 3° or less at DC~50kHz Cursor Measurement Function] A V. AVY AVIB, AT VAT ATH A В. Cursor Resolution ¿100 DIV Effective Cursor Range Vertical: + 3 DIV; horizontal: + 4 DIV CURSOR Panel setting Vertical: V/DIV (CHI, CH2),UNCAL, ADD, EFADOUT INV. PLO, AC/DC/GND. FUNCTION Horizontal: S/DIV (MTB, DTB), UNCAL x 10MAG, Delay time, Hold-off. Trigger: Source, Coupling, Slope, Level, TV-V/TV-H. Others: X-¥, LOCK. é-inch rectangular type with internal graticule Type 0%, 10%, 90% and 100% markers. CRI 8 x 10 DIV (1 DIV = 1 em} Phosphor P31 Accelerating Potentral 12KY approx. External mtensity modulation Coupling DC L-AXIS Voltage 5V or more INPLT Maximum Input Voltage 30Y (DC+AC peak) at LkHz or less Bandwidth DC-3MHz {05-6112 OSCILTOSCOPE. USER MANLAL TRIGGER Voltage 23mV/DIV approx. in 50¢] termination SIGNAL Frequency Response DC—10MHz OUTPLT Output Impedance 3052 approx. , e Waveform IkHz + 55%, squaré wave OUTPUE TOR Voltage 2Vpp = 2% Impedance 2k) approx. SPECIAL | FUNCTION Panel Setups Lock Provided F Г + 0 = LINE POWER — Voltage AC 100V_120V, 230V + 10% selectable REQUIREMENT [ETÉGREnCy ЗОН ог 6012 Power Consumption Approx. SOYA. 70 Wimax.) Indoor use Altitude up 10 2000 m Ambient temperature OPERATING To salisty specifications : 10% to 33 (309 to 95%) ENVIRONMENT | Maximum operating ranges: 0° to 40°C ( 32 "to L104"F) Relative humidity 85% RH(max.) non condensing Installation Category 11 Pollution degree 2 STORAGE TEMPERATURE [-10°10 707, T0%RH{maximum) & HUMIDITY MECHANICAL ¡Dimensions 310 W = 150 H x 455 D (mm) SPECIFICATION |Weight Approx. 9kgs (19.8 lbs) Power cord... 1 ACCESSORIES [Instruction manual............... 1 Prehc (X 17X 10... 2 (308-6112 OSCILLOSCOPE USER MANUAL — o vs 3.PRECAUTIONS BEFORE OPERATION 3-1.Unpacking the Oscilloscope The product has been fully inspected and tested before shipping from the factory. Upon receiving the instrument, please unpack and inspect it to check if there is any damages caused during transportation. If any sign of damage is found, notify the bearer and/or the dealer immediately. 3-2.Checking the Line Voltage The oscilloscope can be applied any kind of line voltage shown in the table below. Before connecting the power plug to an AC line outlet, make sure the voltage selecior of th. va? panel 15 set to the correct position corresponding to the line voltage. It might he damaged the instrument 1f connected to the wrong AC line voltage. WARNING. To avoid electrical shock the power cord protective grounding conductor must be connected to ground. When line voltages are changed, replace the required fuses shown as below: Line voltage| Range Fuse Line voltage | Range Fuse 100V 90-110V ov 108-132v | T1A2S0V | 230V |207-250V | T0.4A 250V WARNING. To avoid personal injury, disconnect the power cord before removing the fuse holder. GOS-6112 OSCILLOSCOPE USER MANUAL 3-3. Environment The normal ambient temperature range of this instrument is from 0° to 40°C {32° to 104°F). To operate the instrument over this specific temperature range may cause damage to the circuits. Do not use the instrument in a place where strong magnetic or electric field exists as it may disturb the measurement, 3-4. Equipment Installation, and Operation Ensure there is praper ventilation for the vents in the oscilloscope case. If this equipment is used not according to the specification, the protection provided by the equipment may be impaired. 3-5.CRT Intensity To prevent permanent damage to the CRT phosphor, do not make the CRT trace brighten excessively or long time. leave the spot stay for an unreasonably 3-6.Withstanding Voltages of Input Terminals The withstanding voltages of the instrument Input terminals and probe Input terminals are shown in the following table. Do not apply voltages higher than these limits. Input terminal Maximum input voltage CHI, CH2, inputs 400\ (DC + AC peak) EXT TRIG input 400V (DC + AC peak) Probe inputs Z AXIS input 600V (DC + AC peak) 30V (DC + AC peak) | maximum input voltage of 1 kHz the frequency to the instrument } CAUTION. To avoid instrument damage, de not apply exceeding GOS-6112 OSCILLOSCOPE USER MANUAL ZOO а 4. PANEL INTRODUCTION After the instrument is switched on, all the important settings are displayed in the readout. The LEI}’s located on the front pang] assist operation and indicate additional information. Incorrect operation and the electrical end positions of control knobs are indicated by a warning beep. Except the Power pushbutton (POWER), the Focus control (FOCUS), and the Trace Rotation control, all other controls are electronically selected, and their functions and settings can therefore be stored. The front panel is subdivided into six sections: ® Display controls Vertical controls Horizontal controls Trigger controls Measurement control Input connectors GOS-6112 OSCILLOSCOPE USER MANUAL д NT ea " OSCALISCOPE О ФОБИИ Ea) Y = ñ ™ Front panel of GOS-6112 GO5-6112 OSCILLOSCOPE USER MANUAL 4-1.Front Panel Display controls The display controls adjust the on-screen appearance of the waveform and provide a probe compensation signal source, ho ; I — 1 re ... . .. : — | : а - HOLE [AL HTA RACE E ; 7 wh | РМГ SEAL — | H y Г > - — | В \ { ) L ща a 7 | CHR mt | 5. A и + AT Fa > 7 (5) ` 4 - J A i" vy (1) POWER - Pushbutton and symbols for ON([) and OFF(0). When switch on the oscilloscope to have all LEDs lighted and the software version will be displaved on the screen, After the Internal test is completed successfully, the normal operation mode is present. Then the last settings become activated and the LED indicates the ON condition. (2).INTEN — TRACE/READOUT - Control knob with associated pushbutton and readout display. The control knob is used for adjusting the traces and readout intensity. Turning the knob clockwise to increase the intensity and turning it counterclockwise to decrease the intensity. he TRACE/READOUT pushbutten is for selecting the intensity function and indicates the letter “TRACE INTEN™ or “READOUT INTEN” in the readout. Press the pushbutton briefly for the following sequences: “TRACE INTEN” — “READOUT INTEO" — “TRACE INTEN” GOS-6112 OSCILLOSCOPE USER MANUAL (3)TRACE ROTATION The TRACE ROTATION is for aligning the horizontal trace in parallel with graticule lines. This potentiometer can be adjusted with a small screwdriver. (4) FOCUS The control knob effects both the trace and the readout sharply. (5). CAL The terminal provides a reference signal of 2Vp-p at 1kHz for probe adjustm Vertical controls The vertical controls select the displaved signals and control the amplitude characteristics. — (9) (9) — VE=TICAL a 4 a д РУ Y oi fa” (3 (а CO | ay | ` “4 | mu 2 thin Col Ca 1; и тои — A O) O « Ju . 7 (3 Av; Me -D 7 чб “ОТ > — — {1 à Lo ! , — E Ww Jen y Zeit x O) SD 6) NY wD ME HD OO CD T pt A $ Y ©) GOS-6112 OSCILLOSCOPE USER MANUAL - - -_-_—_-_-_-_ -_ _ —_ — —_— _—_ _ _—_—_—————_m_—", (7).20MHz BWL — Pushbutton with indicator LED. Briefly pressing the pushbutton, the bandwidth is reduced to approx. 20MHz, and the measurement 18 made by climinating undesired high frequency signal from the wave orm Also the high frequency component over 20MHz 1s eliminat. fro nthe tngger ste '' (8) CURSOR FOS - AVLI/2 — Pushbutton with double function nd associated indicator LED. The function of Cursor Position or CH1/CH2 Position can be selected only after the pushbutton cf Cursor Function is pressed to appear enabling their cursor measurement. Press the pushbutton once briefly to have the related LED lighted, the CH1/CH2 POSITION control knob is then operated as CURSOR CURSOR 2 POSITION control. AY E The function is required and available only in DUAL mode in combination with AY (Voltage) measurement. Pressing and holding the pushbutton, then switch between CHI and CHZ, the measured result will be displayed by the readout with “AVI...” or “AV2..." providing the defection coefficient is calibrated. The settings of the cursors must be related to the signal of the selected channel. (5).CH1 POSITION — C1 — Control knob with double function. The vertical trace position of channel 1 can be set with the control knob, which is also operated as CURSOR 1 position control in cursor measurement mode. (10)CH2 POSITION — C2 — Control knob has several functions. The vertical trace position of channel 2 can be set with the control knob, which is also operated as CURSCR 2 position control in cursor measurement mode. In alternate time base mode, this control knob can be used to scparatc the DELAY time basc trace from the MIAN time base trace, Please note TRACE SEP (11). — 13 — GO5-5112 OSCILLOSCOPE USER MANUAL Te _ (11YTRACE SEP The instrument contains a trace separate function which is required in the alternate time base mode to separate the DELAY time base trace(s) from the MAIN time base in vertical direction. Consequently this function is only available in alternate time base mode. Press the pushbutton once to have the related LED lighted, the CH1 POSITION control knob 1s then operated as vertical position control for the trace(s) of the DELAY time base. (12)ALT/CHOP/ADD-INV The pushbutton has several functions, which arc required and available only when both channels are active. ALT- Displays in the readout, indicates alternate channel switching. After each time base sweeps the instrument internally, switches over from channel ] and channel 2 and vice versa. CHOP Indicates chopper mode. The channel switching occurs constantly between channel and channel 2 during each sweep. ADD- Displays in the readout, indicates additional mode. Whether the algebraic sum (addition) or the difference (subtraction) of both input signals is displayed, depends on the phase relationship and the INV setting. As a result, both signals are displayed as one signal. For correct measurements, the deflection coefficients for both channels must be equal. INV—Pressing and holding the pushbutton to set the channel 2 invert function on or off. The invert on condition is indicated with a horizontal bar above “CH2” in the readout. The invert function causes the signal display of channel 2 to be inverted by 180°. GOS-6112 OSCILEOSCOTE USER MANUAL J (13)CH1 VOLTS/DIV. (14)CH2 VOLTS/DIV- Control knob for channel l/channel 2 has double function. Turning the knob clockwise to increase the sensitivity in 1-2-5 sequence and turning it in the opposite direction (CCW) to decrease. The available range is from 2mVidiv up to 5V/div. The knob is automatically switched inactive if the related channel is switched oft. The deflection coefficients and additional information regarding the active channels are displayed in the readout. ie. “CH I=deflection coefficient, input coupling”. The “=" symbolizes calibrated measuring conditions and is replaced by the “>” symbol in uncalibrated conditions. (15)CH1-VAR. (16\CH2-VAR — Pushbution with double function. CHECH2 Pressing briefly the CH1(CH2) button to set the cannel 1 (channel 2) of the instrument on, the deflection coefficient will be displayed in the readout indicating the current conditions (“C1I1...7/ “CH2...”). VAR Pressing and holding the pushbutton to select the VOLTS/DIV function between attenuator and vernier (variable). The current setting is displayed by the “>” symbol in the readout. After switching on the VAR, turn the VOLTS/DIV control knob counterclockwise to reduce the signal height. and the deflection coefficient becomes uncalibrated. GOS-6112 OSCILLOSCOPE USER MANUAL "VA epee (17)CHI AC/DC. (18)CH2 AC/DC Pressing the pushbutton briefly to switch over from AC (~ symbol) ta DC {= symbol) input coupling. The seiting 15 displayed in the readout with the deflection coefficient. (19)CHI GND- Px 10 (20CH2 GND — PX 10 -Pushbutton has two functions. GND Each time when the pushbutton is pressed bricfiy, the input. of the vertical amplifier is grounded, It is displayed in the readout as an earth (ground) symboi “ 74”, PX10 Pressing and holding the pushbutton to select the indicated deflection coefficient of the channel displayed in the readout between 1:1 and 10:1. The probe factor of 10:1 is displayed in the readout with the probe symbol in front of channel indication (e.g. “P10”, CHI) When proceed cursor voltage measurement, the probe factor will be automatically included. The symbol must not he activated unless a 10:1 attenuator реб are dag, LE. ——];—;——;—;—;—;]—]]][];]—;]]];]]];]]]—————;——;;;;];Ú;]]] ]o];];];” ON DD A E oa a (08-6112 OSCILLOSCOPE USER MANUAL "AAO, Horizontal controls: The horizontal controls select the time base operation mode and adjust the horizontal scale, position and magnification of the signal. 29 HOT SON AL Eh чё СОЗ № fo Т MAN" L. ; El E ; ЧА) Dm ED =D Ge FINE Thy “ua PER 2 . E ur { Je a — u. wet (ZDTIME/DIV- Control knob with double function. Tuming the knob clockwise to reduce the deflection coefficient in a 1-2-5 sequence and turning it in the opposite direction (CCW) to increase, The time coefficient(s) will be displayed in the readout. In MAIN time base (MTB) mode, time deflection coefficients between 0.5s/div and 50ns/div can be chosen in 1-2-5 sequence, 1f the 10 MAG [function is not activated. During alternate (ALT) and DELAY time base (DTB) operation, the control knob changes the DELAY time base setting in 1-2-5 sequence. The available deflection coefficient range is from 50ms/div up to SOns/div (without =10 MAG), but the availability depends on the MAIN time base setting. The internal control of the oscilloscope prevents the DELAY time deflection coefficient from becoming higher than the MAIN deflection coefficient. as such an eperation condition wanld mate no sense. GOS-6112 OSCILLOSCOPE USER MANUAL COC ZZ > (22)\MAFN/ALT/DELAY—X-Y — Pushbutton for time base mode selection. The instrument contains two-time base designated MAIN and DELAY. With the aid of the DELAY time base, signal parts displayed by the MAIN time basc can be expanded in X-direction. The expansion ratio depends on the time deflection coefficient ratio of both time bases (ie. “MTB=0.1ms”, “DTB=1 ¿2 57=100). With higher expansion ratio the DELAY time base trace intensity reduces. Each time when press the pushbutton briefly, the time base mode changes in the sequence of MAIN-ALT-DELAY-MAIN. The actual setting is displaved in the readont. MAIN The TIME/DIV control knob is operated only under the MAIN time base mode. The readout then displays the main time coefficient alone. The time base setting for this condition will be stored if the time base mode is changed. ALT If the alternate time base mode is selected, the TIME/DIV knob only controls the DELAY time base switch. The alternate time base mode 15 a sub-function of the DELAY time base mode and both time base traces can be displayed simultaneousty. Consequently the readout can display both time deflection coefficient. A window sector which indicates part of signal is also visible on the MAIN trace and is displayed by the DELAY time base. The window segment can be shifted horizontally by the DELAY TIME coniral continuously. The difference between the beginning of both the MAIN time base trace and the window sector shows the delay time. The information is also displaved in the readout with an approximate value (ee. “DLY=0.125ns") related to the calibrated MAIN time o GOS-61 £2 OSCILLOSCOPE [USER MANUAL EE CT a EE coefficient (uncalibrated ie. “DLY>0.125ms”) The width of the window segment decreases when the DELAY time coefficient is set to a lower value (higher time deflection speed). For better reading, the vertical position of the DELAY time base trace position can be shifted (please note TRACE SEP (11)). DELAY In the DELAY time base mode, the display of the MAIN traces, the window sector and the MAIN time coefficient will disappear from the readout. As the trace separation is no longer required under the circumstances, the function would be switched off too. Consequently, only the DELAY timc coefficient is displayed by the readout. X-Y Switch on or off the X-Y mode by pressing and holding the button. ln the X-Y mode, the deflection coefficient is displayed in the readout The Y axis input is selected by selling the vertical mode pushbutton to the CHI, CH2 and both modes, and the X axis input is selected by setting the TRIGGER SOURCE pushbutton to the CHI, CH2 and EXT. (23)H POSITION The control knab enables a horizontal position shift of the signals. In combination with x10 MAG the function makes it possible to shift anv part of the signal on the screen. (24)x*10 MAG—SETUPS LOCK— Fushbutton has double function and associated MAC LED, Each time when this pushbutton is pressed, the MAG LED located above will be switch on or off, If the MAG LED is lighted, the signal display in all time base modes will be expanded 10 folds and consequently only a tenth part of the signal curve 1s visible. The interesting part of the signal can be made visible with the aid of the H. POSITION control. GOS-6112 OSCILT OSCOPE USER MANUAL ET Es SETUPS LOCK Pressing and holding the pushbutton, then switch the panel setups lock function en or off. To avoid unintentional touch of the setting, the feature is extremely useful for long term and repetitive measurements that need to bc performed under the same test condition of the oscilloscope setting. (25) YAR Pressing and holding the pushbutton to select the TIME/DIV (21) control knob function between time base switch and vernier (variable). The variable function is activated in the MAIN time base only. After switching on the VAR, the time deflection coefficient is still calibrated until further adjustments are made. Turn the TIME/DIV (21) control knob counter clockwise to increase the time deflection coefficient (reduce the deflection speed) and the deflection coefficient becomes uncalibrated. Instead of “A=10 g 5”, the readout then displays “A>10 zs” indicating the uncalibrated condition. This setting is stored if the instrument is switched to ALT or DELAY time base mode. Switch off the VAR by pressing and holding the pushbutton of time base mode again, then set the time deflection coefficient back into the calibrated condition. GOS-6112 OSCILLOSCOPE USER MANUAL EE —]];—]Ñ];——TDD ня Trigger controls The trigger controls determine the sweep start timing for both signal and dual trace operation. ¿E RSE E LLL A) ROLE EO.RE COJPUSC Car] (o) (a) Cir] Cf Cal Lm] + + wv oe, 25 — tr) 10 (3; Cu] — во 7 > CEE "ME —— (y ZA ho Uy SFE -. Yu - Y-H (25)MODE — Pushbutton and indicator LEDs, Pressing the pushbutton to select the trigger mode. The actual setting is indicated by a LED. Each time when the MODE pushbutton is pressed the trigger mode changes in the seguence: ATO—NML—TV—ATO ATO (Auto) Select the automatical mode, the sweep frec-runs will display a baseline trace when there is no trigger signal or the frequency is below 10Hz. The setting of triggering level changed only when the TRIGGER LEVEL control is adjusted to a new level setting. GOS-6112 OSCILLOSCOPE USER MANUAL re ey NML (Normal) Select the normal mode, the input signal will trigger the sweep when the TRIGGER LEYEL control is set within the peak-to-peak limits of an adequate trigger signal. When the sweep is not triggered, no baseline trace will be displayed. it to the triggering circuit. The horizontal or vertical SYNC signals are selected by TV-V/TV-H pushbutton. Please refer to the TV-V/TV-H (31). 27)LEYEL—-Control knob Turning the control knob causes a different trigger input setting (voltage), and set to a suitable position for the starting of triggered sweep of the waveform. An approximate tri gger level setting (voltage) value will be displayed in the readout. When rotate clockwise the control knob, the trigger point moves toward the positive peak of the trigger signal and rotate it counterclockwise to move the trigger point toward the negative peak of the trigger signal. When the setting (voltage) value is out of the changing portion of the observation waveform, the synchronization sweep stops. Sometimes a “7” will be displayed on the left of the valued display, that indicates that direct reading is impossible if AC, HFR, LFR coupling or VAR of | TY Separate the video sync signal from the composite waveform and direct vertical deflection is set, [05-6112 OSCILLOSCOPE USER MANUAL ——P — — _ — ___® (28)COUPLING —Pushbutton and indicator LEDs. Pressing the pushbutton to select the trigger coupling. The aclual setting is indicated by a LED and by the readout (“source, slope, AC”). Each time when the COUPLING pushbutton is pressed the trigger coupling changes in the sequence: AC—DC—HFR—LFR AC Attenuates trigger signal frequency components below 10Hz and blocks the DC component of the signal. AC coupling is useful for triggering on AC waveforms that have a large DC offset. DC Couples DC and ali frequency components of a triggering signal to the trigger circuitry. DC coupling 1s useful for most signals, especially for providing a stable display af low-frequency or low-repetition-rate signals. HFR (High Frequency Reject) Attenuates high-frequency triggering signal components above 40kHz. HFR coupling is useful [or providing a stable display of low-frequency components of complex waveforms and eliminates high-frequency interference from the trigger signal. LFR (Low Frequency Reject) Attenuates low-frequency triggering signal components below 40kHz and blocks the DC component of the trigger signal. LFR coupling 1s uscful for producing stable triggering on the high- frequency components of complex waveforms and rejecting low- frequency interference or power supply hum from the trigger signal. GOS-6112 OSCILLOSCOPE LISER MANUAL р (29)SOURCE—Pushbutton and associated LEDs. Pressing the pushbutton to select the trigger signal source or the X signal for an X-Y operation. The actual setting is indicated in a LED and by the readout (* SOURCE”, slope, coupling). CHI The signal applied to the channel 1 input connector is the source of the trigger signal. CH2 The signal applied to the channel 2 input connector is the source of the trigger signal. , (Line) The triggering signal is obtained from a sample of the AC power source waveform. The trigger source is useful when the displayed waveform frequency is time related to the AC power source frequency. EXT The external signal applied through the EXT input connector 1s used for the external triggering source signal. When in the dual X-Y operation, the X-axis operates with the external signal. GOS-61 [2 OSCILLOSCOPE USER MANUAL —Bgo oOoOo:sqOAO3IE $ — E [— (30)DELAY-HO— Control knob with a double function and associated LED. The control knob has two different functions depending on the time base mode. HO (Hold-off time) In MAIN time base mode, the control knob applies to the hold off time setting, the HO-LED associated with the knob is dark, the hold off time is sel to minimum. Switch on the LED by turning the control knob clockwise and extend the hold off time until the maximum is reached. An approximate hold off time value will be displayed in the readout (HO: — %”). The hold off time is automatically set to minimum (LED is dark), if the MAIN time base setting is changed. The hold off time setting is stored and deactivaled if ALT (MAIN and DELAY) or DELAY time base mode is selected, DELAY TIME In ALT (MAIN and DELAY) and DELAY time mode, the knob controls the delay time setting. Under the ALT time base mode, the delay time is visible on the main trace, beginning at the trace start and ending at the start of the window sector. An approximate delay time value will be displayed in the readout (“"DLY="). If only select DELAY time base, the delay time can also be varied, but there would be no window sector as the main trace is not visible. GOS5-6112 OSCILLOSCOPE USER MANUAL — ooo —————————;———;— (31)TY-V/TY-H-— Pushbutton for video sync signal selection. TV-V Start the main trace at the beginning of a video signal field. SLOPE polarity must match the composite sync polarity (i.e, “== for negative sync) to obtain TV field triggering on the vertical sync pulse. TV-H Start the main trace at the beginning of a video signal line. SLOPE polarity must match the composite sync polarity to obtain TV line triggering on the horizontal sync pulse. The current setting is displayed in the readout under item “Source, video polarity, TV-H*, (32)SLOPE ( y 7 }—Pushbution for the triggering slope or video polarity selection. If in the AUTO or NML trigger mode, bricfly pressing the pushbutton to select the slope of the signal which is used for triggering the time base generator. Each time when the pushbutton is briefly pressed, the slope direction will switch from falling edge to rising edge, and vice YETSA. The current setting is displayed in the readout under item “source, SLOPE, coupling”, If in the TV trigger mode, bricfly pressing the pushbutton to select the video polarity, which will be displayed in the readout with a “—” symbol of positive video signal and a “— symbol of negative video signal. GOS-6117 OSCILLOSCOPE USER MANUAL —————————]———]].[D)[ E ——— ОА Инин Measurement Control The measurement section controls the cursor measurement. и a = . . -— . << Clk: JHCTHR DE (33)CURSOR FUNCTION-ON/OFF-—Pushbution with two functions. ON/OFF Pressing and holding either pushbutton to switch both cursor lines on or off. As the cursor lines are part of the readout, they are visible only when the readout is switched on. CURSOR FUNCTION Each time when the pushbutton is briefly pressed the seven measurement functions will be selected in the sequence as below: AY Voltage difference measurement. ANS : Voltage difference percentage measurement (3div=100% reference) AYdB Voltage gain measurement. {Sdiv=odB reference, /AVdB=20 log, div/5div). AT : Time différence measurement. 1/AT Frequency measurement. AT% Time difference percentage measurement, (3div=100% reference). Аб Phase measurement. (ádiv=350" reference). GOS-6112 OSCILLOSCOPE USER MANUAL Input connectors The input section is where the input signals are commonly connected to the oscilloscope. CHI 19 RÉ Cha NA EXT kag oY car II yl T SEs t ave GH, VA, 37 Lo (35)CHI—Input BNC socket This BNC socket is the signal input for channel 1. In X-Y mode, signals at this input are used for the Y or X deflection. The outer (ground) connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line/mains plug. (36)CH2—Input BNC socket This BNC socket is the signal input for channel 2. In X-Y mode, signals at this input are used for the X or Y deflection. The outer (ground) connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line/mains plug. (37)Ground socket—Banana Socket galvanically connected to safety earth, This socket can be used a reference potential connection for DC and low frequency signal measurement purposes. (05-6112 OSCILLOSCOPE USER MANUAL д (38)EXT—This BNC socket is the external trigger signal input. In dual X-Y mode, signals at this input are used for the X deflection. Pressing the TRIG. SOURCE (29) pushbutton until the information of “EXT, slope, coupling” is shown up in the readout and the TRIG. SOURCE “EXT” LED is lighted, switches the input on. The outer (ground) connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line/mains plug. The maximum input voltages of the instrument input terminals and probe input terminals are listed in the section of 3-6.Withstanding voltage of [nput terminals”. Do not apply voltage higher than the limit. 505-6112 OSCILLOSCOPE USER MANUAL 4-2.Rear Panel The rear panel provides input power and additional signal connections. or ot % ña” a o a" iif + E) o Ta aa: ao so HER mee Ce PT еб. & вен e. o ko аи RE NF TE "WE J PETE CR BEE " CPT CIO CE CA - nio WE TEN PETE NFL JT BM PRT WN ME, CE E FLO 00 CM EF ODO EE 1 La E E ica M1 _ tr | 0 ye é ne i A A ® E „т, | ® £ 87 ee MANCO: ala. ue „” —_— na MAT "A Tim E OHM E (CIAO HO "Pr MOON TO ad 1% mf WR Wok Le Hal; Fei Te E Л, _ ‘5 (+ Te xl) Ur — A ' — - Гы = ме ' | 1 EA Tia ; 1 — = — Capp ws NE ! I I ! 1 PACA gal GD ITS, 1 | 7 > 5 I И Tn MT E NA > E — | SHE THE N > dé eat) = (39)Line voltage selector and input fuse holder —Select power source and contain the primary power fuse The fuse rating 15 shown in the section of 3-2 Checking the line voltage. (40)AC power input connector Connect the AC power cord to the power supply of instrument, the power cord protective-ground connection is connected to the exposed metal part of the instrument. The power cord must be connected to a proper grounded source for electrical-shock protection. (4 TRIGGER SIGNAL Output—BNC socket The signal selected by the TRIG. SOURCE (29) is available. This output may be used to connect to a frequency counter or other instrument. GO5-6112 OSCILLOSCOPE USER MANUAL yr re Em (42)Z-Axis Input —BNC socket Connect external signals to the Z-axis amplifier for intensity modulating the CRT display. This terminal is DC-coupled. The intensity is lowered by a positive signal, while it is increased by a negative signal. GOS-6112 OSCILLOSCOPE USER MANUAL —— A re >. OPERATION METHOD This section contains basic operation information and techniques that should be considered before proceeding any measurement. As for the location and function of instrument controls, connectors, and indicators, refer to the “Instruction of Front Panel and Rear Panel” of this manual, 5-1.Readout Display The CRT readout display indicates how to set up the instrument controls. No physical marking shown on the rotating switches indicates the control setting. A key to the location and type of readout information displayed is Hlustrated in figure 5-1: GOS-6112 OSCILLOSCOPE USER MANUAL DELAY TIME BASE Ss Dv MAIN TIME BASE — DELAY TIME —— © HOLD OFF TIME - - si CH YALUE VALUE % MEASUREMENT PANEL SETJFS | ХО МАС FUNCTION AND LOCK SYMBOL | VALLE I A | чо 100 ---- = = = | = = = [|-> - o -- a0 1 — VERT MODE È Eb 1 4d Ll LLE TU | АСТ 41 + | | H | | } F TT Ti | Ti I mi 1 14+1-1 CHOP TI ADD — TRIG LEVEL 10— - UNKNOWN SYMBOL 7? 0% - PU Ce | INYERTOR РАО ik 4 A 1 & A | CHA CH2 TRIG SOURCE TRIG LEYEL PROBE X1C PROZE X10 CH! VALJE | CHZ ! or Cki — : CHE — EXT CAL = CAL - Ro | Ft CH1 — сна — 1 | VOLTS/DIV VOLTS/D V | YE ARITY CH1 . A HZ — | TRIG COUPLING — _ Ih PLT COUPLING INPUT COUPLING AC AS: fy AC. Ny DC DE == DZ. = HE= GND: nr GND: gy LFR or TY SYNC Ty Figure 3-1 TV-H 08-6112 OSCILLOSCOPE USER MANUAL -— —_ —_ "—_—_—_—_—_“A——#"., 5-2.Connecting Input Signals Grounding The most reliable signal measurements are made when the oscilloscope and the unit under test are connected by a common reference (ground lead) in addition to the signal lead or probe. The ground lead of the probe provides the best grounding method for signal interconnection and ensures the maximum amount of signal-lead shielding in the probe cable. A separate ground lead (with a banana plug) can also be connected from the unit under test to the oscilloscope ground jack on the front panel. Probes A probe provides the most convenient way to connect an input signal to the oscilloscope. The standard x1/<10 probes supplied to the oscillascope are shielded agamst electromagnetic interference and have a high input impedance for low circuit Joading. CAUTION. To get the best waveform precisely, keep probe ground and signal leads as short as possible. Misadjust probe compensation can cause measurement error. Check and adjust probe compensation whenever a probe is moved to a different channel or oscilloscope. As for the probe compensation adjustment procedure, refer to the “Probe Compensation”. Coaxial Cables Signal input cable can greatly affect the accuracy of a displayed waveform, To maintain original frequency characteristics of the input signal. use only high-quality, low-loss coaxial cables. Coaxial cables must be terminated at both ends in their characteristic impedance fo prevent signal reflections within the cable. Use suitable impedance-matching devices. (008-6112 OSCILLOSCOPE: USER MANUAL "ECO, 5-3. Adjustments and checks Trace Rotation Adjustment Normally, when the trace is in parallel with the center horizontal graticule line, there will be no need to adjust the TRACE ROTATION. If necessary, adjust the TRACE ROTATION to make the baseline trace parallel to the center horizontal graticule line by using a small straight-blade screwdriver or alignment tool. Probe Compensation To minimize the distortion of measured waveforms, check the compensation of your probes before using them. The probe compensation should be checked pericdically whenever the probes are moved to different input channels, 1. Install the probes onto the oscilloscope (Press the BNC connector onto the channel input and rotate the connector to lock it into place). 2. Set the probe slide switches to the x10 position. 3. Briefly pressing the CH1/CH2 button to set the oscilloscope to channel | and channel 2. 4. Pressing and holding the P-10 button to set the indicated deflection coefficient of the channel displayed in the readout as a symbol “P10”, 5. Attach the probe tips to the CAL connection in the front of the oscilloscope. 6. Set the oscilloscope controls to display both channels: VERTICAL: VOLTS/DIV 1V COUPLING DC ALT/CHOP/ADD CHOP HORIZONTAL: MODE MTB TIME/DIV 0.3ms TRIGGER: MODE ATO SOURCE CHI or СН2 COUPLING AC SLOPE LL —GOS-5117 OSCILLOSCOPE USER MANUAL "HI _ -— ———_—-—_—_m———— 7. Observe the displayed waveform and compare them with the waveforms shown in figure 5-2. If either probe needs to be adjusted, proceed the step 8. If either probe does not need to be adjusted, proceed the “Function Check”. Cver compensated | | | | Under compensated | | | | Correctly compensated Figure 5-2 Typical Compensation Waveform 8.Adjust the probe by using a smal! insulated screwdriver. Slowly rotate the adjustment control until the probe is properly compensated. GOS-6112 OSCILLOSCOPE USER MANUAL TE АО ОЕ ДОН оО ОНО АО НИЕ 5-4. Function Check When vou start to check the operation of your oscilloscope, proceed the following instruction: |. Install the x10 probes onto CHI and CHZ inputs. 2. Connect the probe tips to the CAL test point of the oscilloscope. 3. Set the oscilloscope controls to display both channels: VERTICAL: VOLTS/DIV iv COUPLING DC ALT/CHOP/ADD CHOP HORIZONTAL: MODE MTB TIME/DIV 0.5ms TRIGGER: MODE ATO SOURCE CHI or CH? COUPLING AC SLOPE LL The figure 5-3 below illustrates a satisfactory display. The waveform should be approximately 2Vp-p at a frequency of 1kHz that confirms the vertical and horizontal deflection function of the oscilloscope. — - + + = ALT CHI FAC Figure 5-3 Display 1 — 37 — (105-6112 OSCILLOSCOPE USER MANUAL "WE ZZ en: 4. Set both CH1 and CH2 COUPLING to GND. 3.Use the CHI and CH2 POSITION controls to align both traces on the center graticule. 6. Open the CH2 INV by pressing and holding the pushbutton. 7.5¢t to the ADD mode by pressing the ALT/CHOP/ADD pushbutton briefly. 8. Set both CITI and CH2 COUPLING to ОС. 9. The figure 5-4 below shows a satisfactory display. The display will show a flat trace located on the center gracticule that confirms the channel balance and ADD offset function. NEN | | CHI=1Y ее ALT CHI fat Figure 5-4 Display 2 10. Set to the CHOP mode by pressing the ALT/CHOP/ADD pushbutton briefly. I'l. Turn off the CH2 INV by pressing and holding the pushbutton, GOS8-6112 OSCILLOSCOPE USER MANUAL ————— zz U. 5-5.Basic Operation Displaying CH1 or CH2 To display the signal from a signal channel, pressing briefly the CHI or CH2 pushbutton to set the oscilloscope to channel ! or channel 2. Displaying CHI and CHZ To display both signals ar the same time, proceed the following steps: 1 Set the CH! and CH2 on. The figure 5-5 below shows two synchronous waveforms in the both modes. 2.Adjust the CHI or CH? POSITION control to position the two waveforms. 3.Set the ALT/CHOP/ADD button to CHOP mode if the waveforms are flickering. J AC Figure 5-5 Both typical waveforms GOS-6112 OSCILLOSCOPE USER MANUAL Ee ye en Displaying the sum or difference of CHI and CH? To display the algebraic sum or difference of CHI and CH?2, proceed the following steps: 1.Set the ALT?CHOP/ADD button to ADD mode. The figure 5-6 below shows the sum of the waveforms from figure 5-5. 2.Set the CH2 INV on by pressing and holding the button, if necessary, to display the different waveform, 3. Pressing and holding one of the VAR buttons to set the VOLT/DIV control knob to vernier (variable). Then adjust one channel to the other in the event of gain difference. AC Figure 5-6 Typical ADD waveform GOS-6112 OSCILLOSCOPE USER MANUAL "EE 2222 EE Comparing Frequency and phase (Single X-Y Operation) To compare the frequency and phase between two signals by using the X-Y mode. The X-Y waveform displays different amplitude, frequency, and phase. The figure 5-7 shows a typical waveform made up of two signals that are of the same frequency and amplitude, but approximate 45° out of phase. To use the oscilloscope in the signal of X-Y mode, proceed the following steps: |. Connect the horizontal or X-axis signal to the CHI input. 2 Connect the vertical or Y-axis signal to the CH2 input. 3.Set the CHI off, and set the CH2 on. 4,5et the X-Y mode on by pressing and holding the button. 5.Set the TR1G SORCE button to CHI. Use the HORIZONTAL POSITION control to adjust the X-axis. Note: When high frequency signals are displayed in the X-Y operation, note the frequency bandwidths and phase difference between X and Y axis. Refer to “2. SPECIFICATION" section for details. Figure 5-7 Typical single X-Y display. — 41 — GOS-60117 OSCILLOSCOPE USER MANUAL. es Setting up Dual X-Y Operation To use the oscilloscope in the dual X-Y mode, proceed the following steps: [. Connect the horizontal or X-axis signal to the EXT (X) input. 2.Connect one of the vertical or Y-axis signal to the CHI (Y E) input. 3.Connect either of the vertical or Y -axis stgnal to the CH? (Y?) input. 4.5ct the CHI and CIR on. 5, Set the ALT/CHOP/ADD button to CHOP mode. 6.Set the X-Y mode on by pressing and holding the button. The figure 5-8 shows two X-Y waveforms in the dual X-Y mode. Figure 5-8 Typical dual X-Y display Setting Up delayed-sweep Operation A delayed sweep is used to magnify any portion of a complex waveform En the horizontal direction. lo display the delayed sweep operation, proceed the following steps: l. Briefly pressing the MAIN/ALT/DELAY pushbutton ta set to MAIN time base of the horizontal mode. Effect triggering by main sweep and set MAIN TIME/DIV control as desired. GOS-6112 OSCILLOSCOPE USER MANUAL Е О_О 2.Set the MAIN/ALT?DELAY button to ALT mode, and sct the time range of the DELAY TIME/DIVE control to be magnified. The figure 5-9 below shows the man and deiaved sweeps appear simultancously an the screen, and the window sector will appear. 3.Adjust the DELAY TIME control to move continuously the window sector. Bring the window sector to the position te be magnified. Then, the waveform between the window sector is magnified to occupy the full area of the screen, 4.1{ necessary, press the TRACE SEP pushbutton to sct the LED on. The delayed sweep trace can be shilted verticalty abont +3 divisions with respect lo the main sweep trace for the convenience of observation by the POSITION control. Figure 5-9 ALT Mode Figure 5-10 Delay Mode 5 Set the DELAY mode to measure the magnified waveform only by pressing the MAIN/ALT/DELAY button. The magnilied waveform shown in figure 3-10. GOS-6112 OSCILLOSCOPE USER MANUAL ree ene Magnifying Waveform Events Use the x10 MAG pushbutton to view small portions of a waveform as which is too far back from the starting point to view by using the TIME/DIV control. To use the x10 MAG button, proceed the following steps: 1. Adjust the TIME/DIV to the fastest sweep that displays the event. 2.Rotate the HORIZONTAL POSITION control to move the event 19 display on the center of screen. 3. Press the x10 MAG button to switch the MAG LED on. When above procedures have been done, the displayed waveform will be expanded 10 times to the right and left from the center of screen as center of expansion. Operating HO (Hold off time) Control When the measured signal is a complex waveform with two or more repetition frequencies (period), triggering with the LEVEL control alone may not be sufficient to attain a stable waveform display. In such a case, the sweep can be stable synchronized to the measured signal waveform by adjusting the HO (Hold off) time of the sweep waveform. Figure 5-11(a) shows several different waveforms which overlapped on the scrcen, marking the signal observation unsuccessful when the hold off is set to minimum (the HO-LED is dark). Figure 5-11(b) shows the undesirable portion of the signal is held off. The same waveforms are displayed on the screen without overlapping, GOS-6112 OSCILLOSCOPT. USER MANUAL EE (gai signal cyde 4 | Highlighted parts ane displayed и L — E Rd or Figure 5-11(a) Adpusting The : \ HOLD ОРЕ «та | \ Figure 5-11(b) Triggering of Video signal In the work concerned with TV, complex signals and containing video signal, blanking pedestal signal, and synchronizing signal are often measured. Press the TRIG MODE pushbutton to set the TV position. The built-in active TV-Sync-separator provides the separation of frame or line sync pulses from the video signal. To trigger the oscilloscope at the vertical (frame) rate. press the TV-V/TY-H pushbutlon to set TV-V coupling. To trigger the oscilloscope at the horizontal (line), press the TV-Y/TV-H pushbutton to set TV-H coupling. The figure 5-12(a) shows vertical signal of TV-V and Figure 3-12(b} shows horizontal signal of TY-l1. — 45 — GOS-6112 OSCILLOSCOPE USER MANUAL Figure 5-12(a) TV-V Figure 5-12(b) TV-H The polarity of the synchronization pulse is critical for the slope selection. The figure 5-13(a) and 3-13(b) shows the examples of TV polarity synchronization signals. A | x ; iden ; signal E Syxchronimng y Aina Figure 5-13(a) (-) Sync signal. Figure 5-13(b) (+) Sync signal GOS-5112 OSCILLOSCOPE USER MANUAL ee — — ]]—]]—»» -——]]]]Ó 3-6. Measurement Application The oscilloscope has a cursor measurement svstem for making accurate, direct-readout voltage, time, frequency and phase measurements. The measurements described in this section arc examples of typical applicalions using this measurement systèm. After becoming familiar with the controis, indicators, and capabilities of the instrument, vou can develop convenient methods to make the special measurement for Y CUT own applications. Froceed a measurement by using the cursor according to the following steps: 1. о Pressing and holding the CURSOR FUNCTION-ON/OFF pushbutton to turn on the cursor and measurement readout. . Briefly pressing the pushbutton to select (he seven measurement function in the sequence as below: AV AVR —AVIB —AT — VAT — АТИ —A BG AV . If the associated indicator CURSOR POS-LED is hghted, rotate the CI-PGSITION control to position the cursor | and rotate the С2- POSITION control to position the cursor 2. - Read the measurement value on the screen. Typical measurement readouts and applications are shown in Figure 5-14. The measurement values arc automatically controlled bv the VOLTS/DIV and TIME/DIV control seilings, G0S5-6112 OSCILLOSCOPE USER MANUAL "EZ Figure 5-14: Cursor Measuremeg (a). Typical AV (Voltage difference) for AC voltage, When both CHI and CH2 are turned on, the measurement value of CHICA YI) or CH2(/ Y7) can be displayed by pressing and holding the AV 1/2 pushbutton. (b).Typical A Y%iVoltagc percentage) cursor measurement for overshoot of square waveform, À vollage percentage measurement is done by first sctting a reference for the full scale (amplitude) waveform: 5div=10{0%. (c). Typical A VdR(voltage gain) cursor measurement for -3dD bandwidth application. The percentage reference is: 5div=Cdb. The measurement values calculate the voltage gain from the formula: ES Y dB=20 log( À Vdiv/3div) (d).Typical 2 T(Time difference) cursor measurement for rise time. Proceed rise-time or fail-time measurement requiring some additional signal scaling by using the graticale rise-time measurement aids. Number 0%, 10, 90 and 100 arc etched near the left vertical gratical line. Use the following steps as a guideline to in making rise-time measurement: — 48 — GOS-6112 OSCILLOSCOPE USER MANUAL —s gs a ЗОО 1. Set the VOLTS/DIV and VAR controls to provide an exact five- division vertical display. FF _ Use the vertical POSITION control to control the negative amplitude of the signal on the 0% reference line and the positive amplitude on the 100% reference line. 3. Increase the TIME/DIV setting to stretch out the rising edge of the waveform as much as possible to improve the cursor placement accuracy. 4. Use the C1-POSITION control to align the cursor 1 lo the rising edge at the point where it crosses the 10% reference graticule Ine. Then use the C2-POISITION control to align the cursor 2 to the point where the rising edge crosses the 90% graticule line and read the rise time displayed in the CRT readout. (e). Typical 1/AT cursor function for frequency measurement. When the two cursors are superimposed at two edge points of the one period waveform by the C1-POSITION and C2-POSITION controls, the measurement value is displayed in frequency units on the upper side of the SCrçèmt. ({.Typical A T%(Time difference percentage) cursor function for duty-cycle measurement of square waveform. A time difference percentage measurement is done by first setting a reference for the full cycle of waveform period: 3div=100%. — 49 — GOS-61 12 OSCILLOSCOPE USER MANUAL (2).Typical A 9 cursor function for phase measurement, A phase measurement is done by first setting a reference for the full 360° waveform period: Idiv-366, NOTE. When the VOLTS/DIV or the TIME/DIY controls are in uncalibrated setting, the AY and AT measurement values will be displayed with divisions. When the vertical mode is set to the ADD mode, and the CHI and CH? VOLTS/DIV controls are set to different scales, the AV measurement values will be displayed with divisions. G0OS-6112 OSCILLOSCOPE USER MANUAL 6.MAINTENENCE The following instructions are executed by qualified personnel only. To avoid electrical shock, do not perform any servicing other than the operating instructions unless you are qualified to da so. 6-1.Fuse Replacement I the fuse blows, the power lamp indicators will not light and the oscilloscope will not start. The fuse should not normally open unless a probiem has developed in the unit. Try lo determine and correct the cause of the blown fuse and replace only with a fuse of the correct rating and type on the rear panel. WARNING. For continued fire protection. Replace fuse only with 250% fuse of the specified type and rating, and disconnect power cord before replacing fuse. 6-2.Line Voltage Conversion The primary winding of the power transformer is tanped to permit operation from 100, 120 or 230VAC 50/6011z line voltage. Conversion from one line voltage to another is done by changing the line voltage selector switch as shown in page 7. The rear panel identifics the line voltage to which the unit was factory set. To convert to a different line voltage, perform the following procedure: (1).Make sure the power cord is unplugged. (2). Adjust the line voltage selecior switch to the desired line voltage position, (3).A change in line voltage may also require a corresponding change of fuse value. Install the correct fuse value as listed on rear panel. — 5] — GOS-6112 OSCILLOSCOPE USER MANUAL "EC tee ep 6-3.Cleaning le clean the oscilloscope. use a soft cloth dampened in a solution of mild detergent and water. Do not spray cleaner directly onto the oscilloscope because it may leak into the cabinet and cause damage. Do not use chemicals containing benzine, benzene, toluene, Xylene, acetone, or similar solvents, Do not use abrasive cleaners on any portion of the oscilloscope. GOS-6112 OSCILLOSCOPE USER MANUAL 7.Block Diagram EH Was ' LN] LH AE INN e Ha HAMEC HOL 3135 TR LW LEA FLAN НО dar LALA TNA dg 194170 TELA be Boyd Ol — — i oo Zo odo A MOLY dE DILE AL a TOMAS 473A HOLY"! HG aL неа” ва dE ni HOLT = ATI Tod Ih 0 , AM = | ans MOL ет dez IMT £ [i T3N%d TOWNE : ALA IA LEA ЭлНО | NT AE WYHOYIA YO0019 2112-500 ЭМП AID 115 241 OMA —m—— diy Jal т 2195 SAL дано Le | LAN LA} CHO Liv = & - ina (X) VIS ">

Download
Advertisement
Key features
- 100MHz bandwidth
- Two channels
- Dual sweep
- Cursor readout
- Microprocessor-based operating system
- Temperature compensation
- 20MHz bandwidth limit
- TV triggering
- Z-axis intensity modulation
- Trigger signal output