Racine RVL Inline, RVL Wafer, RVL Tube Vortex Shedding Flow Meter, HT Flow Meter Operator's Manual
Below you will find brief information for RVL Inline, RVL Wafer, RVL Tube. The RVL series vortex-shedding flow meter is a general-purpose electronic liquid flow meter. Three outputs are available. The standard output is a two wire (loop powered) 4-20 mA current. Options for a 0 to 5 VDC or a frequency output proportional to the volumetric flow rate are also available. The choice of output must be made at the time of ordering. High accuracy is assured by individual flow testing. Since it uses no moving parts, maintenance is minimized. The RVL is available in several types of plastics allowing them to be used in a wide variety of chemical applications.
PDF
Download
Document
Advertisement
Advertisement
Operator’s Manual RVL Vortex Shedding Flow Meter 8635 Washington Avenue Racine, WI 53406-3738 USA Racine Federated Inc. Racine is a trademark of Racine Federated Inc. Tel: 262-639-6770 09-VRX-UM-00007 Toll Free: 888-572-2463 TABLE OF CONTENTS INTRODUCTION ....................................................................................................................................................... 1 DESCRIPTION ......................................................................................................................................................................... 1 OPERATING PRINCIPLE ....................................................................................................................................................... 1 FLUIDS ....................................................................................................................................................................................... 1 GENERAL INSTALLATION INFORMATION ...................................................................................................... 2 FLOW RATE AND RANGE REQUIREMENTS .................................................................................................................. 2 PIPING REQUIREMENTS ..................................................................................................................................................... 2 BACK PRESSURE .................................................................................................................................................................... 7 TEMPERATURE ....................................................................................................................................................................... 7 OUTPUTS .................................................................................................................................................................................. 8 K-FACTORS .............................................................................................................................................................................. 8 ELECTRICAL INSTALLATION POWER ...................................................................................................................................................................................... 8 WIRING ...................................................................................................................................................................................... 8 4-20 mA LOOP .................................................................................................................................................................... 8 0-5 VDC Output ................................................................................................................................................................11 Pulse Output .....................................................................................................................................................................11 3 PIN CONNECTION OPTION ..........................................................................................................................................12 RVL (INLINE) SERIES..............................................................................................................................................14 SPECIFICATIONS ..................................................................................................................................................................14 FLOW RATES..........................................................................................................................................................................14 MECHANICAL INSTALLATION ........................................................................................................................................15 Dimensions ........................................................................................................................................................................16 Temperature Limits .........................................................................................................................................................16 Pressure Drop....................................................................................................................................................................16 RVL (WAFER) SERIES .............................................................................................................................................17 SPECIFICATIONS ..................................................................................................................................................................17 FLOW RATES..........................................................................................................................................................................17 MECHANICAL INSTALLATION ........................................................................................................................................17 Dimensions ........................................................................................................................................................................20 Temperature Limits .........................................................................................................................................................20 Pressure Drop....................................................................................................................................................................21 RVL (TUBE) SERIES.................................................................................................................................................22 SPECIFICATIONS ..................................................................................................................................................................22 FLOW RATES..........................................................................................................................................................................22 MECHANICAL INSTALLATION ........................................................................................................................................22 Dimensions ........................................................................................................................................................................23 Temperature Limits .........................................................................................................................................................24 Pressure Drop....................................................................................................................................................................24 MAINTENANCE .......................................................................................................................................................25 TROUBLESHOOTING ............................................................................................................................................25 APPENDIX .................................................................................................................................................................26 INTRODUCTION DESCRIPTION The RVL series vortex-shedding flow meter is a general-purpose electronic liquid flow meter. Three outputs are available. The standard output is a two wire (loop powered) 4-20 mA current. Options for a 0 to 5 VDC or a frequency output proportional to the volumetric flow rate are also available. The choice of output must be made at the time of ordering. High accuracy is assured by individual flow testing. Since it uses no moving parts, maintenance is minimized. The RVL is available in several types of plastics allowing them to be used in a wide variety of chemical applications. OPERATING PRINCIPLE Counter An everyday example of a vortex shedding phenomenon is a flag waving in the breeze: the flag waves due to the vortices shed by air moving across the flagpole. Detector Bluff Body Detector Within the flow meter, as a fluid moves across a tiny strut or “bluff body”, vortices are also shed but FIGURE 1 on a smaller scale. The vortices form alternately, from one side to the other, causing pressure fluctuations. These are detected by the crystals in the sensor tube, and are converted to an analog signal or pulse output. The frequency of the vortices is directly proportional to the flow. This results in extremely accurate and repeatable measurements with no troublesome moving parts (see figure 1). FLUIDS Any clean liquid compatible with the plastic material of construction that does not contain significant amounts of fibers or abrasive materials can be used. Danger - Do not use with: explosive or flammable materials, food or beverages, or gaseous fluids. Viscosities above 1 cSt will raise the minimum usable flow rate (in effect reducing range-ability). This effect is linear to viscosity. No adjustments are required for viscosities up to 2.0 cSt. Liquids with higher viscosities will adversely affect the permissible amount and duration of over range flow (see table 1). Form No. 09-VRX-UM-00007 Viscosity 1 cSt 2 cSt 3 cSt 4 cSt 5 cSt 6 cSt Viscosity and Rangeability Minimum Maximum Flow Range 1 12 12:1 2 12 6:1 3 12 4:1 4 12 3:1 5 12 2.4:1 6 12 2:1 TABLE 1 1 GENERAL INSTALLATION INFORMATION Prior to installation, the following items should be considered. 1) The vortex transmitter contains electronic circuitry which can be affected by high electromagnetic or electrostatic fields. Care should be taken to locate the installation in an area away from large electrical motors, transformers, or other devices which can produce such interference. 2) Proper grounding is required to eliminate electrical noise which may be present within the fluid and piping system or in the near vicinity of the vortex transmitter. For non-conductive piping systems, an exterior grounding strap should be used to provide a path to earth ground. For conductive piping systems, a properly grounded pipe will require no additional preparation. FLOW RATE AND RANGE REQUIREMENTS Most manufacturers state flow range capabilities by publishing the maximum allowed flow rates. Then they provide a turndown ratio to determine minimum flow rate. To use the turndown ratio, simply divide the maximum rate by the ratio to determine the minimum rate. Racine vortex flow meters have a 12:1 turndown ratio at a viscosity of 1 cSt. Higher viscosities will reduce the turndown. NOTE: The ¼” NPT and ½” flare end meters have a standard turndown ratio of 8:1 PIPING REQUIREMENTS Turbulence in the pipe line can affect the accuracy of most flow meters. Sources of turbulence are pumps, valves, or changes-in-direction in the line. To avoid these potential problems, it is standard practice to place the meter a certain distance from the turbulence source. These distances are indicated in Pipe Diameters (PD). For example, 10 PD means place the flow meter ten times its inside diameter away from the source of turbulence. Downstream distances between the meter and a valve or a change-indirection must also be followed. The best accuracy is achieved with at least 20 PD upstream and 5 PD downstream for RVL vortex flow meters. If an upstream elbow is closely coupled to another elbow creating a change in plane, 27 PD is required upstream and 10 PD downstream. (see Figure 2, 3, and 4) When the diameter of the meter is smaller than the pipe line, at least 20 PD of pipe with the same diameter as the meter upstream, and 2 PD downstream is needed. Overall, 25 PD of straight run prior to the meter is required (see Figure 5). If there is a plane change in the installation, this IN OUT requirement increases to 25 PD upstream (30 overall). The downstream requirement is now 2 PD of pipe with the same diameter as the meter, and a minimum of 5 PD overall of straight run. If there is a valve downstream the usual 10 PD between the meter and a valve is still required. If the required piping parameters are not met, there will be a reduction in accuracy. NOTE: Pulsating flow will affect accuracy (pressure pulses will not). 2 Form No. 09-VRX-UM-00007 HORIZONTAL FLOW - (Sensing element in vertical orientation) Configuration 1 plane change 1 plane change w/outlet valve 2 plane changes 2 plane changes w/outlet valve Piping Requirements (pipe diameters) Inlet Outlet 5 20 10 5 27 10 20 Dia Minimum Accuracy (full scale) Repeatability (of point) ±1.00% 0.25% 5 Dia Minimum Flow 20 Dia Minimum 10 Dia Minimum Flow Two Plane Changes 27 Dia Minimum 5 Dia Minimum Flow Two Plane Changes 27 Dia Minimum 10 Dia Minimum Flow FIGURE 2 Form No. 09-VRX-UM-00007 3 HORIZONTAL FLOW - (Sensing element in horizontal orientation) Configuration 1 plane change 1 plane change w/outlet valve 2 plane changes 2 plane changes w/outlet valve Piping Requirements (pipe diameters) Inlet Outlet 5 20 10 5 27 10 20 Dia Minimum Accuracy (full scale) Repeatability (of point) ±1.50% 0.25% 5 Dia Minimum Flow 20 Dia Minimum 10 Dia Minimum Flow Two Plane Changes 27 Dia Minimum 5 Dia Minimum Flow Two Plane Changes 27 Dia Minimum 10 Dia Minimum Flow FIGURE 3 4 Form No. 09-VRX-UM-00007 VERTICAL FLOW - (upward or downward flow and sensor in any orientation) Configuration 1 plane change 1 plane change w/outlet valve 2 plane changes 2 plane changes w/outlet valve Two Plane Changes Repeatability (of point) ±1.00% 0.25% 5 Dia Minimum 20 Dia Minimum 10 Dia Minimum Flow 27 Dia Minimum Flow 10 Dia Minimum Accuracy (full scale) Two Plane Changes Flow Flow 27 Dia Minimum Piping Requirements (pipe diameters) Inlet Outlet 5 20 10 5 27 10 20 Dia Minimum 5 Dia Minimum FIGURE 4 Form No. 09-VRX-UM-00007 5 25 Dia Minimum 10 Dia Minimum Flow 20 Dia Minimum 5 Dia Minimum 2 Dia Minimum 5 Dia Minimum 25 Dia Minimum Flow 20 Dia Minimum 5 Dia Minimum 2 Dia Minimum Two Plane Changes 5 Dia Minimum 30 Dia Minimum Flow 5 Dia Minimum 25 Dia Minimum 2 Dia Minimum FIGURE 5 6 Form No. 09-VRX-UM-00007 BACK PRESSURE Back pressure (the pressure immediately downstream of the meter) must be maintained above a minimum level in order to avoid cavitation. For most applications, this may be ignored if the flow rate is less than 75% of maximum. For other liquids, use the following formula to calculate the minimum back pressure. Back Pressure = 2.75 ΔP + 1.25 PV - 14.7 Where: ΔP = Pressure drop in psi at max flow PV = Vapor pressure in psia of the liquid at operating temp. (eg. the PV of water at 100 °F is 0.42.) BP = Back pressure (downstream of meter) in psig. As an example, for water, at 100 °F (37 °C) in a ½” meter, where the maximum pressure drop is 8 psi minimum back pressure is 7.8 psig. BP = (2.75 × 8) + (1.25 × 0.42) - 14.7 BP = 22 + 0.525 - 14.7 BP = 7.825 TEMPERATURE To protect the internal crystals in each unit, temperature limitations must be adhered to. All permissible operating temperatures are identified by meter and material type. Additionally fluid temperature will affect maximum working pressures. For de-rating information see the maximum fluid operating pressures for the specific flow meter model. OUTPUTS The RVL series meters can be obtained with either an analog output or a rate frequency output. The standard analog output is a 4-20 mA current an optional 0-5 VDC is also available. The analog output can also be re-configured in the field using a PC communications cable and programming software which are both available from Racine Vortex as P.N. RVS220-954. The analog current output varies between 4 mA (0 flow) and 20 mA (maximum flow). The 0 to 5 VDC analog output is also continuously variable between 0 V (0 flow) and 5 V (maximum flow). NOTE: One of the two analog output options (4-20 mA or 0-5 VDC) are hardware selected at the factory and can not be changed in the field. The rate frequency output produces pulses whose frequency is proportional to the flow going through the meter. Each meter has a slightly different output frequency which is listed on the calibration sheet that accompanies the meter. Table 2 shows the long term average full scale output frequency for standard size meters. Form No. 09-VRX-UM-00007 Meter Size Average Full Scale Pulse Width in (mm) Frequency (Hz) (msec) ¼ (6.4) 1055 0.47 ½ (12.7) 820 0.61 ½ (12.7) 570 0.88 ¾ (19.1) 284 1.76 1 (25.4) 292 1.71 1½ (38.1) 144 3.47 2 (50.8) 148 3.38 61 8.20 3 (76.2) TABLE 2 7 The frequency output option generates a square wave with an amplitude that matches the input power level. The pulse width varies with frequency and is found by using the following formula. The result is in seconds. 1 PW = 2 x Maximum Frequency (Hz) K-FACTORS The K-factor (with regards to flow) is the number of pulses that must be accumulated to equal a particular volume of fluid. Think of each pulse as representing a small fraction of the totalizing unit. Calibration reports that accompany RVL series meters include a nominal K-factor in both gallons and liters. See the sample calibration sheet in the appendix of this manual. ELECTRICAL INSTALLATION POWER The meter requires an 8 to 28 VDC power supply. The specific connections depend upon which output is option is used. See wiring details for the specific output option . NOTE: This instrument requires clean electrical line power. Do not operate this unit on circuits with noisy components (i.e. Fluorescent lights, relays, compressors, variable frequency drives, etc.) Linear power supplies are also much preferable to switching power supplies. NOTE: The power and output connections share a common ground. WIRING 4-20 mA LOOP Connect a twisted pair wire (not provided) to the terminals of the transmitter marked +8-28 VDC and Output. If the twisted pair wire is shielded, do not connect the shield to the transmitter. The shield should be grounded at the receiver only (see figure 6). The transmitter is reverse-polarity protected. 8 Form No. 09-VRX-UM-00007 8 to 28 VDC Output +8-28 VDC POWER SUPPLY 8-28 VDC Output Gnd 4-20 mA Programming RECEIVER Load FIGURE 6 The receiving equipment must accept industry standard “true two-wire” or “loop powered” 4-20 mA process transmitter inputs. This means that the receiving equipment, such as a recorder or controller, must supply power for the transmitter along the twisted wire pair. If the receiving equipment does not provide power, a separate power supply, typically 24 VDC and at least 30 mA, must be used, as shown in figure 6. There are many brands of receivers which provide 24 VDC for this purpose. Several receivers may be connected in a series as shown in figure 7, but only one should provide power, and all should have isolated inputs. The voltage provided by the receiver must be within the limits shown in the Supply Voltage chart (see figure 8). 8 to 28 VDC Output +8-28 VDC Output Gnd 4-20 mA RECEIVER/POWER SUPPLY Programming RECEIVER RECEIVER Additional Loads FIGURE 7 Form No. 09-VRX-UM-00007 9 1100 1000 Supply Voltage - 8 VDC Loop Load (Ohm's) 900 0.02 = Maximum Loop Resistance 800 700 600 500 400 Operate in the Shaded Region 300 200 100 8 10 12 14 16 18 20 22 24 26 28 Supply Voltage (VDC) FIGURE 8 To use this figure, first add the resistance of all the receivers, indicators, etc., and the wire in the loop. If the wire resistance is unknown, use a value of 50 Ohm for a twisted wire of 1,000 feet or less with a gauge of #22 AWG or heavier. Find the total load (in ohms) on the left hand side of the chart in figure 8 and follow that value horizontally until it intersect with the shaded area. From the intersection point look straight down to where a vertical line would intersect the voltage scale. This is the minimum voltage needed for the transmitter to operate properly under the specific load conditions. Example: After checking the specification for all the loads in an application the total amounted to 800 ohms. Following the 800 ohm line in figure 9 to the right the intersection point is about ¾ of the way across the chart in figure 9. Loop Load (Ohm's) 1100 A vertical line through the intersection point crosses the voltage axis at about 24 VDC so with a load of 800 ohms a standard 24 volt power supply would be used. 1000 Supply Voltage - 8 VDC 900 0.02 = Maximum Loop Resistance 800 700 600 500 400 Operate in the Shaded Region 300 200 100 8 10 12 14 16 18 20 22 24 26 28 Supply Voltage (VDC) FIGURE 9 10 Form No. 09-VRX-UM-00007 Figure 10 and figure 11 show typical connections for the 0-5 VDC and pulse output options. 0 - 5 VDC OUTPUT Digital Display 0 - 5 VDC 8 to 28 VDC Output Ground +8-28 VDC 8-28 VDC POWER SUPPLY Output Gnd Programming FIGURE 10 PULSE OUTPUT Counter Pulse Output +8-28 VDC 8 to 8-28 VDC POWER SUPPLY OutputOutp Gnd Grou FIGURE 11 Form No. 09-VRX-UM-00007 11 3 PIN CONNECTION OPTION An optional 3 pin connection is available for when the transmitter/meter combination are to be mounted remotely from the power source/receiver. The mating connector is P.N. RF8687000. White +4-20 mA Input 8 to 28 VDC Output +8-28 VDC Output Gnd Black -4-20 mA Input Meter Electronics + + 4-20 mA Input - 8 to 28 VDC Power Supply P.N. RF8687000 Connector -4-20 mA Input + mA Ammeter FIGURE 12 Black - Supply Green Ground +8-28 VDC Output Gnd 8 to 28 VDC Output Ground White Output Meter Electronics + Green Ground Black +V - 8 to 28 VDC Power Supply P.N. RF8687000 Connector White Output Counter FIGURE 13 12 Form No. 09-VRX-UM-00007 Division of Racine Federated Inc. White + 4-20 mA Input Black - 4-20 mA Input 8635 Washington Avenue Racine, WI 53406-3738 USA [email protected] FIGURE 14 Form No. 09-VRX-UM-00007 13 RVL (INLINE) SERIES RVL (Inline) Specifications Fluid: Connection: Turndown Ratio: Accuracy: Repeatability: Materials Output Signals: Power Supply: CSA Certification: Response Time Enclosure: Liquids NPT Female or Butt (PVDF only) 12:1 (½” - 2”) 8:1 (¼”) ±1% of full scale (4-20 mA or 0-5 VDC) ±2% of full scale, frequency pulse ±0.25% of actual flow PVC standard CPVC, PVDF optional 4-20 mA standard 0-5 VDC or frequency pulse optional (Push - Pull Driver) 150 mA Sink or Source 8 to 28 VDC CSA File 215035, CSA Standard C22.2 No. O-M and No 142-M 2 seconds minimum, step change in flow. Type 4X (IP 66) RVL (Inline) Nominal Flow Rates Tube Size (inches) Minimum Flow gpm (lpm) ¼ ½ ¾ 1 1½ 2 0.6 (2.3) 1.3 (4.7) 2.1 (7.9) 4.2 (15.8) 8.3 (31.5) 16.7 (63.1) Maximum Flow gpm (lpm) 5 (18.9) 15 (56.8) 25 (94.6) 50 (189.3) 100 (378.5) 200 (757.1) Nominal Full Scale Frequency Hz 1052 570 284 292 144 142 Weight lbs (Kg) 1.5 (0.68) 1.6 (0.72) 1.7 (0.77) 1.8 (0.80) 3.1 (1.40) 2.7 (1.22) MECHANICAL INSTALLATION This meter will provide years of accurate service if good flow meter installation practices are followed. The flow meter should be installed where pipe vibration is minimal. Observe the upstream piping requirements listed under “Piping requirements”. Upstream valves should not be used to control flow rate. They should always be kept fully open. Good quality ball valves with integral unions may be connected directly to the flow meter if the valves are fully open during operation. This allows easy isolation and removal of the flow meter, should maintenance be required. Cavitation and flow rate pulsation will adversely affect flow meter performance. 14 Form No. 09-VRX-UM-00007 Diaphragm or piston pumps may not be used. Do not use Teflon tape or any kind of pipe dope when piping. The simple appearance of the flow meter may tempt an installer to handle it as an ordinary nipple. Remember, it is a precision electronic instrument. Treat it with care. Do not use excessive force. Mating fittings (FNPT) and flanges should be screwed into meter hand tight; then tighten an additional ½ to ¾ turn. Always use two wrenches when turning the flow meter into a fitting, one across the flats on the flow meter end, close to the fitting, and one on the fitting. Do not use tools inside the flow meter, as this may damage the vortex sensor, and invalidate the warranty. The flow meter may be mounted in any orientation. Three holes, tapped ¼-20 UNC-2B, .375-inch deep, on ¾” centers are provided on the ¾ inch and smaller flow meters. These holes may be used (at the user’s discretion) to provide support for the flow meter should pipe supports not be practical. CORD GRIP Cord Grip COVER F CONDUIT ADAPTOR TERMINAL STRIP ELECTRONICS MODULE I NPT/BUTT END THREE PIN CONNECTOR FLOW SENSOR BODY A B C E D FIGURE 11 Form No. 09-VRX-UM-00007 15 RVL (Inline) Dimensions PVC/CPVC Size inches (mm) A B C D E F I ¼ ½ ¾ 1 1½ 2 3.81 (97) 3.81 (97) 3.81 (97) 3.92 (100) 3.90 (99) 4.31 (109) 1.75 (45) 1.75 (45) 1.75 (45) 1.75 (45) 2.00 (51) 2.00 (51) 5.25 (133) 7.13 (181) 7.63 (194) 8.03 (204) 8.37 (213) 8.37 (213) 2.50 (64) 2.50 (64) 2.50 (64) 2.50 (64) 2.50 (64) 2.50 (64) 0.30 (8) 0.55 (14) 0.74 (19) 0.96 (24) 1.50 (38) 1.94 (49) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) 3.00 (76) 3.00 (76) 3.00 (76) 3.00 (76) 3.38 (86) 3.38 (86) RVL (Inline) Dimensions PVDF (BUTT Fusion Only) Size inches (mm) A B C D E F I ¼ ½ ¾ 1 1½ 2 5.90 (150) 5.75 (146) 5.75 (146) 5.88 (149) 6.21 (158) 6.60 (168) 0.63 (16) 0.78 (20) 0.94 (24) 1.19 (30) 1.50 (38) 1.88 (48) 4.87 (124) 4.87 (124) 4.87 (124) 5.09 (129) 6.24 (158) 6.77 (172) 1.31 (33) 1.31 (33) 1.44 (37) 2.00 (51) 2.50 (64) 3.00 (76) 0.30 (8) 0.55 (14) 0.74 (19) 0.96 (24) 1.50 (38) 1.94 (49) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) 3.00 (76) 3.00 (76) 3.00 (76) 3.00 (76) 3.38 (86) 3.38 (86) RVL (Inline) Maximum Operating Pressure PSIG (KPa) Maximum Fluid Temperature °F (°C) CPVC PVDF 203 (95) Not Recommended Consult Factory Consult Factory 150 (66) 100 (38) 70 (21) Not Recommended 93 (641) 150 (1034) 63 (434) 120 (827) 150 (1034) 130 (896) 150 (1034) 150 (1034) .5 .2 .1 .05 .3 .5 1 2 5 10 15 25 50 100 200 FLOW (GPM) . 1 in in. ¾i n. 100 1½ i 2 in n. . . 2 in 1 1½ in. in. n. ¾i ½ 2 200 ½ . 1 in n. ¼i PRESSURE DROP (PSID) 12 10 8 5 1000 750 500 350 n. 20 ¼i PRESSURE DROP (MILLIBAR) PVC 50 35 20 10 5 3.5 2 5 10 20 50 30 100 200 600 300 800 FLOW (LPM) FIGURE 12 16 Form No. 09-VRX-UM-00007 RVL (WAFER) SERIES RVL (Wafer) Specifications Fluid: Connection: Turndown Ratio: Accuracy: Repeatability: Materials Output Signals: Power Supply: CSA Certification: Response Time Enclosure: Liquids Wafer 12:1 ±1% of full scale (4-20 mA or 0-5 VDC) ±2% of full scale, frequency pulse ±0.25% of actual flow PVC standard CPVC, Polypropylene, PVDF optional 4-20 mA standard 0-5 VDC or frequency pulse optional (Push - Pull Driver) 150 mA Sink or Source 8 to 28 VDC CSA File 215035, CSA Standard C22.2 No. O-M and No 142-M 2 seconds minimum, step change in flow. Type 4X (IP 66) RVL (Wafer) Nominal Flow Rates Tube Size (inches) Minimum Flow gpm (lpm) ½ ¾ 1 1½ 2 3 1.3 (4.7) 2.1 (7.9) 4.2 (15.8) 8.3 (31.5) 16.7 (63.1) 25.0 (94.6) Maximum Flow gpm (lpm) 15 (56.8) 25 (94.6) 50 (189.3) 100 (378.5) 200 (757.1) 300 (1136) Nominal Full Scale Frequency Hz 570 284 292 144 148 61 Weight lbs (Kg) 0.8 (0.36) 0.9 (0.41) 1.1 (0.50) 1.7 (0.77) 2.6 (1.17) 4.8 (2.16) MECHANICAL INSTALLATION The RVL (Wafer) series transmitters are designed with wafer style flow bodies, which mount easily between standard ANSI style pipe flanges. Observe the upstream piping requirements listed under “Piping requirements”. Upstream valves should Qot Ee used to FoQtrol ÀoZ rate. They should always be kept fully open. Good quality ball valves with integral unions may be connected directly to the flow meter if the valves are fully open during operation. This allows easy isolation and removal of the flow meter, should maintenance be required. Cavitation and flow rate pulsation will adversely affect flow meter performance. Form No. 09-VRX-UM-00007 17 Diaphragm or piston pumps may not be used. Do not use Teflon tape or any kind of pipe dope when piping. For flanged meters, do not allow gaskets to protrude into the flow stream. The following steps will insure proper installation and operation. 1) A uniform flow profile is required to assure proper vortex shedding. This requires a non-pulsating flow along with the proper length of straight pipe run before and after the transmitter. Figures 2,3,4 and 4 shows the proper piping requirements and dimensions. FLANGE SIZE ½ -1½” 2-3” RECOMMENDED TORQUE 10-15 ft. Ibs. 20-30 ft. Ibs. 2) Flanges are to be spaced to accommodate the width of the flow body. Dimensions are listed on page 20. 3) Align the flow body centered with respect to flanges and gaskets, insert threaded rods, retaining nuts and lock washers. 4) Install all retaining nuts hand tight, and then uniformly tighten the nuts in an alternating sequence, diametrically opposed to each other. Uniform stress across the flange will prevent leakage at the gasket. Torque ratings are listed below. 5) The use of grounding rings is recommended when metal pipes are used in conjunction with this meter. See figure 13 18 Form No. 09-VRX-UM-00007 Grounding Rings FIGURE 13 Form No. 09-VRX-UM-00007 19 CORD GRIP Cord Grip E COVER CONDUIT ADAPTOR TERMINAL STRIP ELECTRONICS MODULE THREE PIN CONNECTOR A FLOW SENSOR BODY B C D FIGURE 14 RVL (Wafer) Dimensions PP/PVC/CPVC/PVDF Size inches ½ ¾ 1 1½ 2 3 A B C D E 5.85 (149) 5.90(150) 5.69 (145) 6.00 (152) 6.37 (162) 6.88 (175) 0.78 (20) 0.94 (24) 1.19 (30) 1.50 (38) 1.88 (48) 2.50 (64) 2.03 (52) 2.03 (52) 2.25 (57) 2.63 (67) 3.22 (82) 4.25. (108) 1.75 (45) 1.75 (45) 1.75 (45) 1.75 (45) 1.75 (45) 1.75 (45) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) 2.88 (73) RVL (Wafer) Standard Specifications Maximum Operating Pressure PSIG (KPa) Maximum Fluid Temperature °F (°C) PVC CPVC Polypropylene PVDF 203 (95) Not Recommended Consult Factory Not Recommended Consult Factory 150 (66) 100 (38) 70 (21) Not Recommended 100 (690) 150 (1034) 63 (434) 120 (827) 150 (1034) 90 (621) 130 (896) 150 (1034) 130 (896) 150 (1034) 150 (1034) 20 Form No. 09-VRX-UM-00007 RVL (Wafer) High Pressure Specifications Maximum Operating Pressure PSIG (KPa) Maximum Fluid Temperature °F (°C) CPVC 1 .5 .2 .1 .5 1 PVDF 2 5 10 15 25 50 100 200 300 FLOW (GPM) . 2 in n. . 3 in 100 1½ i 200 1 in 3/4 in . . . in. 2 in 1½ 3 in . 2 in. . 1 in ¾i n. in. ½ PRESSURE DROP (PSID) 12 10 8 5 1000 750 500 350 1/2 20 .05 .3 Polypropylene Not Recommended Not Recommended Not Recommended Consult Factory Consult Factory Consult Factory 90 (621) 300 (2068) Consult Factory Consult Factory 400 (2750) 130 (896) Consult Factory Consult Factory 150 (1034) 400 (2750) PRESSURE DROP (MILLIBAR) 203 (95) 150 (66) 100 (38) 70 (21) PVC 50 35 20 10 5 3.5 2 5 10 20 50 30 100 200 600 1200 300 800 FLOW (LPM) FIGURE 15 Form No. 09-VRX-UM-00007 21 RVL (TUBE) SERIES RVL (Tube) Specifications Fluid: Connection: Turndown Ratio: Accuracy: Repeatability: Materials Output Signals: Power Supply: CSA Certification: Response Time Enclosure: Liquids Tube (Flare end) 12:1 (¾”, 1”) 8:1 (½”) ±1% of full scale (4-20 mA or 0-5 VDC) ±2% of full scale, frequency pulse ±0.25% of actual flow PVC standard CPVC, Polypropylene, PVDF optional 4-20 mA standard 0-5 VDC or frequency pulse optional (Push - Pull Driver) 150 mA Sink or Source 8 to 28 VDC CSA File 215035, CSA Standard C22.2 No. O-M and No 142-M 2 seconds minimum, step change in flow. Type 4X (IP 66) RVL (Tube) Nominal Flow Rates Tube Size (inches) ½ ¾ 1 Minimum Flow Maximum Flow Weight gpm (lpm) gpm (lpm) lbs (Kg) 0.6 (2.3) 5 (18.9) 1.5 (0.68) 1.3 (4.7) 15 (56.8) 1.6 (0.72) 2.1 (7.9) 25 (94.6) 1.7 (0.77) MECHANICAL INSTALLATION This meter will provide years of accurate service if good flow meter installation practices are followed. The flow meter should be installed where pipe vibration is minimal. Observe the upstream piping requirements listed under “Piping requirements”. Upstream valves should not be used to control flow rate. They should always be kept fully open. Good quality ball valves with integral unions may be connected directly to the flow meter if the valves are fully open during operation. This allows easy isolation and removal of the flow meter, should maintenance be required. Cavitation and flow rate pulsation will adversely affect flow meter performance. Diaphragm or piston pumps may not be used. Do not use Teflon tape or any kind of pipe dope when piping. 22 Form No. 09-VRX-UM-00007 The simple appearance of the flow meter may tempt an installer to handle it as an ordinary nipple. Remember, it is a precision electronic instrument. Treat it with care. 1) To install a flare fitting, first remove any burrs from the pipe ends, then slide the flare nut onto the pipe. Push it back far enough so that it will be out of the way when you use the flaring tool. 2) Clip the pipe in the flaring tool, keeping the end flush with the face of the tool. 3) Slowly turn the handle on the tool until it bottoms out. 4) Unscrew the handle and remove the tool to check the quality of the flare. (If the flare isn’t smooth or even the first time, cut off the end with your pipe cutter, and try the technique again.) 5) Line up and tighten the nut and flared pipe to the fitting body. Make the connection as tight but not so tight as to distort the flow meter body. Always use two wrenches when turning a fitting onto the flow meter, one across the flats on the flow meter end, close to the fitting, and one on the fitting. Do not use tools inside the flow meter, as this may damage the vortex sensor, and invalidate the warranty. The flow meter may be mounted in any orientation. CORD GRIP COVER CONDUIT ADAPTOR TERMINAL STRIP ELECTRONICS MODULE THREE PIN CONNECTOR FLOW SENSOR BODY B A C FIGURE 16 RVL (Tube) Dimensions Tube Size (inches) A B) C ½ ¾ 1 1.31 (33.3) 1.31 (33.3) 1.44 (36.6) 6.25 (158.8) 6.25 (158.8) 6.59 (167.4) 4.87 (123.7) 4.66 (118.4) 5.42 (137.7) Form No. 09-VRX-UM-00007 23 RVL (Tube) Specifications Maximum Fluid Temperature °F (°C) 150 (66) 100 (38) 70 (21) Maximum Operating Pressure PSIG (KPa) PVDF 130 (896) 150 (1034) 150 (1034) ¾ 2 1 .5 .2 .1 .05 .3 .5 1 2 5 10 15 25 50 100 200 FLOW (GPM) . 1 in 200 n. 1 in in. . ½i n. PRESSURE DROP (PSID) 12 10 8 5 1000 750 500 350 ¾i 20 ½i n. PRESSURE DROP (MILLIBAR) FLOW (LPM) 100 50 35 20 10 5 3.5 2 5 10 20 50 30 200 600 100 300 800 FLOW (LPM) FIGURE 17 24 Form No. 09-VRX-UM-00007 MAINTENANCE This flow meter requires no maintenance in normal service if properly installed. If the flow tube should become clogged with debris, it will be necessary to remove the meter from service for cleaning. Significant clogging will often result in high (up to 20%) and/or erratic output. Do not stick tools into the tube, as this may permanently damage the vortex sensor. The vortex sensor can not be repaired in the field. To clean the flow tube, run hot (up to 160 °F) soapy water into the downstream end of the flow tube. Large objects jammed against the bluff body maybe dislodged by lightly tapping the upstream end of the flow tube against a firm surface. WARNING: Do not remove vortex meter during operation. Always disconnect primary power source before inspection of service. Do not tap the flow tube so hard that the threads, on threaded units, become damaged A schedule of maintenance checks should be determined based upon environmental conditions and frequency of use. Inspect the meter at least once a year. The inspection should consist of performing visual, electrical, and mechanical checks on all components. 1) Visually check for evidence of overheating by noting discoloration of wires or other components. 2) Check for damaged or worn parts, especially the bluff body, or indications of corrosion. 3) Check for tight , clean electrical connections and that the device is operating properly. TROUBLESHOOTING If difficulty is encountered, locate the symptom most likely present and follow the appropriate instructions. CURRENT LOOP SYMPTOM - NO CURRENT OUTPUT 1) Place a DC voltmeter across the two terminal block screws. With the electronics module powered there must be at least 8 VDC present. If there is less than 8 VDC, but more than 0 VDC, check the power source for sufficient voltage to drive the loop, as shown on page 10, in Figure 8. If there is 0 VDC present check for a broken wire or connector in the loop. 2) Check for the proper polarity of the current loop connections. 3) Make sure the receiving device is configured to provide (source) current to the electronics module. SYMPTOM - ZERO FLOW INDICATION (4 mA IN LOOP). 1) Check that the flow is greater than the minimum specified for the particular size flow meter in use. If the flow rate is too low, replace the flow meter with the proper size flow meter. 2) If the flow rate is sufficient, partially remove the electronic module. Check that the three pin connector that connects the electronics module to the flow transducers is positively connected (See Figure 18). If it is disconnected, align and insert the connector on to the bottom of the electronics module. Form No. 09-VRX-UM-00007 25 SYMPTOM - Erratic Flow Indication Check that there is at least 8 VDC present across the two terminal block screws. Check for material clogging the flow meter and in the upstream piping. Check for erosion of the bluff body by sighting down the meters bore. Erosion or damage to the bluff body will cause erratic readings and compromise accuracy. If the erosion continues, the flow meter will need to be periodically replaced. Check upstream piping to assure that the required piping is used. (See the piping requirements specifications starting on page 2 for this information.) Check for pipe vibration. Normal amounts of pipe vibration are easily tolerated. In addition to this, the transmitter module contains a highly effective active filter that rejects false signals caused by pipe vibration. This filter is most effective under flowing conditions. If vibration is causing the meter to indicate flow when the flow is stopped it will most likely not cause error under flowing conditions. The false flow indication may be ignored, or the pipe may be restrained by firm clamps, or the noise adjustment may be readjusted. Be sure to read the noise adjustment instructions. Check for electrical noise. Under some conditions there can be high common mode AC noise present between the fluid and the power supply ground. The flow meter is designed to reject up to 50 volts of AC common mode noise without loss of accuracy. If metal piping is used, place a ground strap on the pipe on both sides of the flow meter (the flow meter is made of non-conductive plastic) and connect them both to the one point where the loop is grounded (see wiring diagrams starting on page 9). If plastic piping is used, a grounding orifice should be used. The transmitter module contains a highly effective active filter that will reject false signals due to high common mode voltage. This filter is most effective under flowing conditions. If a false indication of flow is encountered at zero flow, it will probably not cause error under flowing conditions. In addition, the noise adjustment may be used, but it will reduce the ability of the flow meter to measure low flow rates. See the noise adjustment instructions before attempting to make this adjustment. OVER-STRESSED SENSOR If the maximum permitted flow rate of 125% of Recommended capacity (100% of HT meters) is exceeded, it is possible to over-stress the sensor. 26 Form No. 09-VRX-UM-00007 APPENDIX CALIBRATION CERTIFICATE SAMPLE Racine Vortex A Division of Racine Federated Inc. 8635 Washington Ave Racine, Wisconsin 53406 Phone: 262-639-6770 Fax: 262-639-2266 Calibration Report Unit Under Test (UUT) Information: Description: ¾” In-Line NPT End Flow Meter Model Number: RVL075-N 1 VNN Serial Number: 99999 Sensor Type: Vortex Shedding Output type: 0-5V Minimum Flow: 2.1 GPM 7.9 LPM Maximum Flow: 25 GPM 94.6 LPM Calibration Date: October 24, 2007 Calibration Interval: 12 Months Cal. Liquid: Water Ambient Temperature: 71.74 °F Ambient Humidity: 31.39 %RH Linear Points: 5 Master Meter: Std uncertainty: Traceability No: Model No: Serial No: ±0.25% 30400/31801 FT8-8N EXW-LEG-5/FT-16 NEXW-LEG-1 806890/16011903 Customer Information: Customer Name: Customer No.: Order No.: UUT Calibration Data Table In GPM: Flow Standard Actual GPM UUT Hz UUT Temp °F Visc. cSt 1 25.00 100.000 72.00 0.949 1 18.00 75.000 72.00 0.949 1 12.00 50.000 72.00 0.949 1 6.00 25.000 72.00 0.949 1 2.10 10.000 72.00 0.949 e l p m a UUT F/V Hz/cSt UUT K CYC/GAL (Hz*60)/NK *60)/NK 60 GPM Linear COEFF. Raw Err % FS Calc. 0-5V Meas. 0-5V Output Err % FS 105.406 240.00 24.57 .57 1.0174 1 1.71 5.000 5.000 0.00 79.055 250.00 50.00 18.43 0.9767 -1.71 3.600 3.680 0.40 52.703 250.00 0 12.29 2.29 0.9767 -1.14 2.400 2.420 0.10 352 26.352 250.00 6.14 0 0.9767 -0.57 1.200 1.200 0.00 10.541 285.71 2.46 0.8547 -1.43 0.420 0.420 0.00 Raw Err % FS Calc. 0-5V Meas. 0-5V Output Err % FS Nominal K (NK (NK) UUT Calibration Data Table In LPM: M: Flow Standard Actual GPM UUT Hz UUT Temp mp °F Visc. cSt 1 94.64 100.000 72.00 0.949 1 68.14 75.000 72.00 0.949 1 45.42 50.000 72.00 1 22.71 25.000 72.00 1 7.95 10.000 72.00 244.186 244 UUT F/V Hz/cSt UUT K CYC/GAL CY (Hz*60)/NK GPM Linear COEFF. 10 105.406 63.40 93.01 1.0174 1.71 5.000 5.000 0.00 79.0 79.055 66.04 69.76 0.9767 -1.71 3.600 3.680 0.40 09 0.949 52.703 66.04 46.51 0.9767 -1.14 2.400 2.420 0.10 0.949 49 26.352 66.04 23.25 0.9767 -0.57 1.200 1.200 0.00 0.949 10.541 75.48 9.30 0.8547 -1.43 0.420 0.420 0.00 S Nominal K (NK) 64.507 Status: PASS Meter Accuracy (of FS): ± 0.4 % Average Calib. Temperature : 72 F Average Calib. Specific Gravity : 1 Average Calib. Viscosity : 0.95 cSt Flow Direction : Forward Calibrated By: Ramon Benedict Certified By: Larry Perez Racine calibrations are performed using standards traceable to National Institute of Standards and Technology. The equipment and calibration procedures complies with ISO 9001 Form No. 09-VRX-UM-00007 27 A division of Racine Federated Inc. Limited Warranty and Disclaimer Racine, division of Racine Federated Inc. warrants to the end purchaser, for a period RIRQH\HDUIURPWKHGDWHRIVKLSPHQWIURPWKHIDFWRU\WKDWDOOÀRZPHWHUVHOHFWURQLF accessories and other products manufactured by it are free from defects in materials and workmanship. This warranty does not cover products that have been damaged due to misapplication, abuse, lack of maintenance, or improper installation. Racine’s obligation under this warranty is limited to the repair or replacement of a defective product, at no charge to the end purchaser, if the product is inspected by Racine and found to be defective. Repair or replacement is at Racine’s discretion. An authorization number must be obtained from Racine before any product may be returned for warranty repair or replacement. The product must be thoroughly cleaned and any process chemicals removed before it will be accepted for return. The purchaser must determine the applicability of the product for its desired use and assumes all risk in connection therewith. Racine assumes no responsibility or liability for any omissions or errors in connection with the use of its products. Racine will under no circumstances be liable for any incidental, consequential, contingent or special damages or loss to any person or property arising out of the failure of any product, component or accessory. All expressed or implied warranties, including the implied warranty of merchantDELOLW\DQGWKHLPSOLHGZDUUDQW\RI¿WQHVVIRUDSDUWLFXODUSXUSRVHRUDSSOLFDtion are expressly disclaimed and shall not apply to any products sold or services rendered by Racine. The above warranty supersedes and is in lieu of all other warranties, either expressed or implied and all other obligations or liabilities. No agent or representative has any authority to alter the terms of this warranty in any way. 28 Form No. 09-VRX-UM-00007 8635 Washington Avenue t Racine, Wisconsin 53406 Toll Free: 888-572-2463 Phone: 262.639.6770 t Fax: 262.639-2267 www.racinevortex.com · [email protected] Racine Vortex is a registered trademark of Racine Federated Inc. UL is a registered trademark of Underwriters Laboratories. © 2011 Racine Federated Inc. Printed in USA Form No. 09-VRX-UM-00007 ">
Advertisement
Key features
- Three output options
- High accuracy
- Minimal maintenance
- Available in various plastics
- General-purpose liquid flow meter
- Loop powered
Frequently asked questions
The meter operates by using the vortex shedding principle. As a fluid moves across a tiny strut (bluff body) within the device, vortices form alternately, causing pressure fluctuations. These fluctuations are detected by crystals in the sensor tube and converted into an analog signal or pulse output. The frequency of the vortices is directly proportional to the flow rate.
Any clean liquid compatible with the plastic material of construction that does not contain significant amounts of fibers or abrasive materials can be used. However, it is important to note that the meter should not be used with explosive or flammable materials, food or beverages, or gaseous fluids.
The RVL series meters have a 12:1 turndown ratio at a viscosity of 1 cSt. This means that the minimum flow rate is one-twelfth of the maximum flow rate. Higher viscosities will reduce the turndown ratio.