Fluke 1750
Three-Phase Power Recorder
Never miss capturing a disturbance—with the exclusive threshold-free measurement system, it’s automatic.
Capture every measurement, every event, on every cycle, all the time with the Fluke
1750 Power Recorder. Outstanding accuracy and resolution provide complete visibility into your installation or distribution system.
• Power quality that meets the standard:
All measurements comply with IEC61000-4-30 standards for correct evaluation of all measured values including voltage, current, power, harmonics, fl icker etc.
• Quick and reliable confi guration: PDA wireless “front panel interface” provides the ability to verify setup without a laptop along with a window into what the instrument is recording, even in awkward test locations.
•
Threshold-free setup: Apply thresholds after data is collected with Fluke Power Analyze
Software so there is no need to worry about missed information due to incorrect set-up.
•
Captures everything: Cross-channel and current triggering capture every measurement, on every channel, every time.
• Intuitive PC software: Easily analyze data and generate reports. Automated EN50160 reporting and compliance.
•
Plug and play: Set up in minutes with selfidentifying current probes and single-lead voltage connections.
•
No need to reconnect wires: Swap channels internally with the wireless PDA or PC when connections are not correct.
• Measure every parameter: voltage and current on three phases, neutral, and ground.
•
5 MHz, 8000 Vpk waveform capture: Get a detailed picture of even the shortest events.
• Quickly retrieve data: With included SD memory card or via the 100BaseT high-speed Ethernet connection.
Applications
Long-term analysis: Uncover hard-to-fi nd or intermittent issues; monitor critical equipment, capturing power quality events to correlate with equipment malfunction
Power quality surveys: Quantify power quality throughout a facility, documenting results with professional reports
Quality of service compliance: Validate incoming power quality at the service entrance
Equipment Installation/Commissioning: Benchmark power system prior to install to insure quality of service
Easy to use
The recorder automatically detects, scales, and powers current probes without needing batteries.
Requiring only single-lead voltage connections enables safe and quick setups. Once power is applied the instrument automatically begins recording and LEDs give you assurance that the recorder is powered up and signals are within range – no more uncertainty that data is being recorded. The
Fluke 1750 has an exclusive capture algorithm which makes certain all events are captured without the tedious setups and blind spots associated with threshold driven equipment.
The PDA wirelessly interfaces with the recorder, allowing quick setup and verifi cation with waveform displays, meter screens, and phasor diagrams.
The built-in wireless technology allows you to control multiple instruments from a distance easily, without the need for a laptop computer (laptops can also be used when desired).
View measurements real-time with wireless PDA interface.
Configurations are simple with wiring diagrams to guide you.
2 Fluke Corporation Fluke 1750 Three-Phase Power Recorder
No need to reconnect wires—swap channels internally using the PDA interface.
Comprehensive power system data
Every cycle on every channel is simultaneously sampled providing complete analysis of power quality, harmonics, power and energy. Crosschannel current and voltage triggering captures event data on all input channels simultaneously, no matter which channel triggered the event. When periodic analysis of the waveform may be needed, the Snapshot mode captures waveforms at userdefi ned intervals. Even dc signals can be recorded!
Events can be displayed against a variety of tolerance curves.
Full FFT on each channel to the 50th harmonic.
Display multiple parameters simultaneously on one graph.
Full-featured power meter display for each channel and totals.
Thresholds for the EN50160 report are easily customized.
Powerful data management
Data is automatically stored on the internal fl ash memory that can store records for over a month without compression or overwriting. Data is retrieved via one of two quick and easy options: download the data onto the included SD memory
EN50160 report delivers an easy to read pass or fail chart along with drill down data.
card without using your laptop, or transfer directly to your computer via Ethernet, using TCP/IP protocol.
Voice, data, and picture annotations can be made via the PDA to fl ag important points in the data stream. Imagine being able to fl ag in your recorded data when a transfer switch was operated, or load changes were made.
3 Fluke Corporation Fluke 1750 Three-Phase Power Recorder
View data the way you want
The new Fluke Power Analyze software revolutionizes your ability to analyze data. No need to worry about thresholds—with Fluke Power Analyze, thresholds can be modifi ed after the data is recorded! And the easy-to-use user interface will display voltage as a phase-to-phase or phase-toneutral reading.
Apply thresholds to data after collection using a variety of standard or customized templates.
Simplified report writer feature included in Fluke Power
Analyze.
All the latest power quality standards are built-in
IEC 61000-4-30 compliant measurement systems provide the confi dence that all parameters are measured and calculated consistently with international standards.
Automated EN50160 compliance reporting for rapid Pass/
Fail testing.
600 V CAT IV and
1000 V CAT III safety rating
Designed to help protect you and your equipment, the Fluke 1750 Three-Phase Power Recorder and accessories are all certifi ed to meet the stringent safety standards for use in 600 V CAT IV and 1000 V CAT III environments. They are the fi rst tools of their kind to carry the
CAT IV rating and, therefore, can be used for most power connections and for all outlets in a low-voltage power distribution system.
4 Fluke Corporation Fluke 1750 Three-Phase Power Recorder
Optional Accessories
Fluke power quality current transformers and fl exible current probes are specially designed to work seamlessly with Fluke power monitors (1750 and 1650).
All clamp-on and fl exi-CT’s are matched to take full advantage of your instrument’s ability to set scale factors for accurate readings.
Cases
1750/CASE Transit Case
• Rugged water tight transit case with rollers
Miscellaneous
1750/MC
•
Additional SD Memory Card for Fluke 1750
1750/SEAT-L
•
Fluke Power Analyze –
Additional SEAT LICENSE
• One additional license for installation on one additional
PC
1750/SITE-L
•
Fluke Power Analyze -
Additional SITE LICENSE
• Site license for installation on unlimited PCs
4006 Security Cable
• 1.8 m (6 ft) looped lockable steel cable for securing monitor
Model No.
CT Type
3005-PR Clamp-On
3014-PR
3140R
Clamp-On
Clamp-On
Model No.
CT Type
3110-PR Flexi-CT
3210-PR
3310-PR
3312-PR
Flexi-CT
Flexi-CT
Flexi-CT
Current Range
0.01 A to 5 A
0.1 A to 40 A
2 A to 400 A
Accuracy
Frequency
Response
1 % RDG ± 0.5 % FS 5 kHz
Jaw Opening
2 cm (.78 in) dia.
1 % RDG ± 0.1 % FS 5 kHz
2 % RDG ± 0.04 A 20 kHz
2 cm (.78 in) dia.
3.2 cm (1.25 in) dia.
Current Range
2 A to 100 A
20 A to 1000 A
Accuracy
1 % rdg ± 0.5 % FS
1 % rdg ± 0.15 % FS
Frequency
Response
7 kHz
7 kHz
100 A to 5000 A 1 % rdg ± 0.15 % FS 7 kHz
Probe Length
60 cm (24 in)
60 cm (24 in)
60 cm (24 in)
100 A to 5000 A 1 % rdg ± 0.15 % FS 7 kHz 120 cm (48 in)
5 Fluke Corporation Fluke 1750 Three-Phase Power Recorder
Specifi cations for the System:
Recorder and Power Analyze Software
General
Power quality measurement standards
Conformance
Clock/calendar
Real-time clock accuracy
Internal memory capacity for data
Maximum recording period
Measurement time control
Maximum number of events
Power requirements
Operating time during interruptions
(internal UPS operation)
Dimensions
Mass (weight)
IEC 61999-1-4 Class 1, IEC 61000-4-30, IEEE519, IEEE1159, IEEE1459 and EN50160
Leap years, 24-hour clock
Not more than ± 1 s/day
At least 2 GB
At least 31 days
Automatic
Limited only by the size of the internal memory
100 to 240 V rms ± 10 %, 47-63 Hz, 40 W
5 minutes per interruption, 60 minutes total operating time without recharging
215 mm x 310 mm x 35 mm (8.5 in x 12.2 in x 3.5 in)
6.3 kg (14 lb)
Input
Measurement types
Input channels
Voltage channels
Current input characteristics
Measuring method
One Phase Plus Neutral, One Phase IT No Neutral, One Phase Split Phase, Three Phase Wye,
Three Phase Delta, Three Phase IT, Three Phase High Leg, Three Phase Open Leg, 2 Element Delta,
2 1 /
2
Element Wye
Voltage: 4 channels, ac/dc
Current: 5 channels
Input resistance: 2 MΩ
Input capacitance: < 20 pF
2 V rms = full scale, 1 MΩ Input Impedance for ferro CTs, low impedance for Flexi-CTs
Simultaneous digital sampling of voltage and current. Digital PLL synchronized sampling, internal frequency reference used during voltage drops.
Synchronization and sampling
PLL-synchronization source
PLL lock range
Sampling frequency
A/D resolution
The PLL synchronizes to the A-N voltage for wye power types, and to the A-B voltage for delta power types. All listed power types can be characterized as either wye or delta.
42.5 to 69 Hz
Voltage and current: 256 samples/cycle
Inter-harmonics per IEC 61000-4-7: 2560 points/10 cycles (50 Hz),
3072 points/12 cycles (60 Hz)
Transient Voltage: 5 MHz
Voltage and current: 24 bits
Transient voltage: 14 bits
Voltage and current measurements
Voltage measurement range
Voltage crest factor
Current measurement range
Current crest factor
AC voltage: 1000 V rms ± 10 % over range
DC voltage: ± 1000 V + 10 % over range
3 or less
Depends on current probe used
4 or less
Voltage and current measurement accuracy
RMS voltage
Measurement type
Measurement uncertainty
True rms calculated continuously: every cycle, every 1/2 cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by IEC 61000-4-30.
AC: ± 0.2 % reading ± 0.1 % full scale, above 50 V rms
DC: ± 0.5 % reading ± 0.2 % full scale, above 50 V dc
RMS current
Measurement type
Measurement uncertainty
True rms calculated continuously: every cycle, every 1/2 cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by standards
Ferromagnetic Clamps: ± (0.1 % full scale + 0.2 % reading + current sensor accuracy), valid for 5 % to 100 % of current sensor range
Flexible Current Probes: ± (0.1 % full scale + 0.5 % reading + current sensor accuracy), valid for 5 % to 100 % of current sensor range
6 Fluke Corporation Fluke 1750 Three-Phase Power Recorder
Transient voltage (impulse)
Measurement type
Full scale
Sample resolution
Measurement uncertainty
Waveshape sampling
8000 V pk
200 nS
± 5 % reading ± 20 V (test parameters: 1000 V dc, 1000 V rms, 100 kHz)
Dip (Sag) and Swell Measurements
Voltage swell (rms swell)
Measurement type
Displayed data
Measurement uncertainty
Voltage dip (rms sag)
Measurement type
True rms (one cycle calculation by overlapping each half cycle - voltage between lines is measured for 3P3W lines and phase voltage is measured for 3P4W lines)
Amplitude and duration of swell
Same as rms voltage
Displayed data
Measurement uncertainty
Voltage dropout (interruption)
Measurement type
True rms (one cycle calculation by overlapping each half cycle - voltage between lines is measured for 3P3W lines and phase voltage is measured for 3P4W lines)
Amplitude and duration of dip or interruption
Same as rms voltage
Same as voltage dip
Power Measurements
Calculated per IEEE1459 for best performance when distortion is present
Measurement type
Measurement accuracy
True rms calculated continuously: every cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by standards
+/- (voltage uncertainty + current uncertainty + current probe uncertainty)
Frequency
Measurement range
Measurement source
Measurement accuracy
Power factor
Measurement range
Measurement accuracy
Displacement power factor
42.5 to 69 Hz
Same as PLL synchronization source
± 10 mHz (10 to 110 % of range, with sine wave)
0.000 to 1.000
± 1 digit from the calculation of each measured value (±3 digits for total)
Measurement method
Measurement range
Measurement accuracy
Calculated from the phase difference between voltage fundamental and current fundamental
- 1.000 (leading) to + 1.000 (lagging)
± 0.5 % reading ± 2 % full scale ± 1 digit
Voltage unbalance and phase sequence
Measurement method Positive sequence voltage divided by negative sequence voltage, per IEC 61000-4-30
Harmonic voltage and current
Analysis window
Analysis order
Measurement accuracy rectangular
1st to 50th order
Measurement method
Voltage / Current: 1st to 20th orders: ± 0.5 % reading ± 0.2 % full scale, 21st to 50th orders: ± 1 % reading ± 0.3 % full scale (current sensor accuracy must be included for current and power)
IEC 61000-4-7
Inter-harmonic voltage and current (intermediate harmonics)
Analysis window
Analysis orders
Measurement method
Voltage fl icker
Measurement method rectangular
1.5 to 49.5th order
IEC 61000-4-7 as per EN 61000-4-15:2003: 10 min (Pst), 2 h (Plt)
7 Fluke Corporation Fluke 1750 Three-Phase Power Recorder
External Interface Specifi cations
LAN interface
Connector
Speed and type
Communications protocol
Wireless controller interface
Connection
Speed
Communications protocol
RJ-45
10/100 Base-T, auto MDIX
TCP/IP over Ethernet wireless (2.4 GHz radio) up to 700 kbit/second
Bluetooth SPP
Environmental and safety specifi cations
Operating environment
Storage temperature and humidity
Operating temperature and humidity
Maximum rated working voltage
Voltage terminals
Voltage durability
Indoors or in covered area outdoors, up to 2000 m altitude (for compliance to IEC61010 standard)
-20 °C to 50 °C, 80 % RH max, non-condensing
0 °C to 40 °C, 80 % RH max, non-condensing
1100 V rms
5550 V rms ac for 1 minute, between voltage input terminals, voltage input terminals and current probes, and voltage input terminals and case (50/60 Hz, 1 mA sense current)
IP30 (per EN 60529) Enclosure protection
Standards conformance
EMC
Safety
EN 61326-1:1997+A1:1998 Class A
EN 61000-3-2:1995+A1:1998+A2:1998
EN 61000-3-3:1995
EN 61010-1 2nd Edition; 2000
Voltage input unit: Contamination Level 2, Overvoltage Category 1000 V CAT III, 600 V CAT IV
(anticipated overvoltage: 8000 V)
Ordering information
Fluke 1750 Three-Phase Power
Recorder
Includes:
•
1750 acquisition unit
• PDA wireless “front panel interface” and charger power plug adapters
•
4 - 400 A current probes (3140R)
•
5 test leads and clips
•
SD memory card
•
Fluke Power View and Fluke Power
Analyze software
• Power cord with international plug set
•
Ethernet cable
•
Color localization set
• Printed Getting Started manual
• Product CD with software and users manual PDF
•
Soft carrying case
Fluke 1750/B Three-Phase Basic Power Recorder
Includes items above with the exception of the
4 - 400 A current probes (3140R)
8 Fluke Corporation Fluke 1750 Three-Phase Power Recorder
Fluke offers a full suite of power quality test tools to locate, predict, prevent and troubleshoot power problems.
•
Handheld troubleshooters for instant analysis
• Ready-to-use power loggers
• Comprehensive power quality recorders
•
Power calibrators and standards backed by Fluke metrology expertise
The Fluke power quality set of products offers the highest level of performance and maintains the Fluke promise of rugged and reliable test tools.
Fluke. Keeping your world up and running.
®
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006-2009 Fluke Corporation. Windows ® is a registered trademark of Microsoft Corporation.
Specifi cations subject to change without notice.
Printed in U.S.A. 10/2009 2428201F D-EN-N
Modification of this document is not permitted without written permission from Fluke Corporation.
Pub-ID 11578-eng