LM-280 USER'S MANUAL

LM-280 USER'S MANUAL
LM-280
ΑUΤΟ
USER’S
LΕΝSΜΕΤΕR
MANUAL
Introduction
This manual mainly introduces how to correctly use LM-280 Auto
Lensmeter . In order to operate it more safely and effectively
, please
read this manual carefully before you use this machine.
Surely , if you find any mistake or problem in this manual please do
not hesitate to contact Shanghai Yanke Equipment Co., Ltd.
LM-280 Auto Lensmeter is applied to measure the top focus degree
and prism degree of eye-glass lens and cornea Screen lens, orient the axial
direction of prism, locate and mark the center and axis of the rough lens. It
can also be used to check whether the glass is installed properly and test
the penetration ratio of purple light.
LM-280 Auto Lensmeter can be used for not only monochromatic
lens, but also twin focus lens, three focuses lens, progressive multi focus
lens and contact lens.
LM-280 Auto Lensmeter is digital-show focus degree instrument
made according to the theory of auto-focus. Its resolving power is 0.01
m-1.
LM-280 Auto Lensmeter is far more favorable in the function and
Test Step contrasted to manual focus instrument.
LM-280 Auto Lensmeter is the first-degree focus apparatus, based on
the Enterprise Standard of NWLK01-2004.
1
Contents
Chapter 1
Safety Measures and Cautions
Chapter 2
Label Instruction
Chapter 3
Using The Instrument
3.1 Names of out elements and parts
3.2 Installation and operation
3.3 Measuring
3.4Checking Lens
A. Checking single beam lens
B. Checking lens with framework
C. Measuring two or three beams lens
D. Checking the progressive multi-focus lens
E. Checking progressive lens with framework
F. Checking for high refraction ratio lens
Chapter 4
Other Instructions
A. Revise
B. Maintenance
C. General faults treatment
2
Chapter 1
Safety Measures and Cautions
The safety contents are defined as different signs in these instructions as
follows:
! Danger: indicate the emergent situations, if not avoid, death or
grievous harm will happen.
! Warning: indicate the potential dangerous situations, if not avoid,
death or grievous harm may happen.
!
Caution: indicate the potential dangerous situations, if not avoid,
light harm or farther harm may happen or the property of instrument
may be influenced.
A.
Operation Cautions:
! Caution:
z Not open and touch the inside without permission, otherwise
electric shock or instrument exception will happen
z Please apply the stated power supply voltage. The voltage can’ t be
too low or high, otherwise apparatus can not run normally, even
damage the instrument and cause to electric shock.
z Power supply lines under the instrument is not permitted,
otherwise the surface layer of the line will be damaged and the
followed accident will be caused: short circuit , fire and electric
shock.
If the metal core of the Power supply line is exposed, it must be
changed in order to avoid electric shock and fire.
3
B.
Deposit Cautions:
Hints:
z Don’t locate the instrument near the poisonous, erosive liquid and
gas.
z
It is not permitted to expose the instrument under the sunlight and
keep proper temperature and humidity.
Move Cautions:
! Warning:
z Pulling the instrument lines and cables forcibly is forbidden,
otherwise instrument damage and personal harm may happen;
z
Don’t move the whole instrument by catching the framework, the
permissive method is to support the base with hands in the back
when moved, otherwise it may cause man hurt or instrument
damage.
C. Installation Cautions:
! Warning:
z Don’t locate the instrument near wet place. If water enters it will
cause electric shock and instrument faults.
z
Please install the instrument firm and flatly. Body harm and
instrument damage will happen because of instrument fall when
improperly fixed.
Hints:
z Don’t locate the instrument under the sunlight or use it beside the
lamp; The surface such as mirror, glass show-case or polished table
which reflect the light intensively is not suitable to fix the
instrument, otherwise the instrument may work improperly.
z Don’t use the instrument in the place exposed in the wind of the
4
fans and air conditioner, otherwise dust will enter into the base and
the Step will be influenced.
z
Don’t use the instrument under the circumstances of web, heat and
dust. They will be bad for the instrument.
Connection Cautions
! Warning:
z Don’t plug excessive pin into one socket, otherwise fire will be
caused.
D.
Cautions After Use
! Warning:
z Please turn off the power supply when finished and put on the
dust-proof cover. If power is on for over long time the fire may be
caused because of overheat. The dust may influence the Test Step
if exposed over long time without cover.
z
Please pull out the power line against fire if the instrument is free
for long time.
E. Maintain Cautions
! Warning:
z
When adding ink, please pull out the ink box lightly and put it
flatly on the table lest the print center hurt your fingers.
Hints:
z If your instrument is in exception and need to open for repair,
please contact with our agent. We will be derelict of duty if it open
by yourself.
z
Please do mind not to scratch the surface protect lens under the
base. The nick will reduce the Step and reliability of Test.
5
F. Contingency Cautions:
! Warning:
z In case of LCD screen destroyed, please smash it into fragments
and clear off the Liquid Crystal with iso-propyl alcohol and burn
them off.
z
When contact the Liquid Crystal left, please water off as soon as
possible.
6
Chapter 2
Label Instruction
To guarantee safe operation, this instrument is with a label on the back.
Please use the power supply according to the instruction and replace fuse.
A: Signs on the instrument
!
Read carefully before use。
O
Signs “ on/off” of power supply
Power off
Fuse sign and standard value
Warning: Fire! Please replace
the fuse with standard
value..
100—120V~
T—2A 250V
200—240V~
T—1A 250V
Fuse Parameters is on the lower label of the back
B: Main Technology Parameters
1. Test scope
Ball lens:
0~±20D
(Reading Increment 0.01D, 0.12D, 0.25D)
Column lens: 0~±9.99D (Reading Increment 0.01D, 0.12D, 0.25D)
Ball lens axial position: 0~180°(Reading Increment 1°)
Prism: 0~9.99△ (Reading Increment 0.01△,0.12△,0.25△)
(ADD): 0~9.99D (Reading Increment 0.01D, 0.12D, 0.25D)
2. Astigmatic Test Mode: MIX /+/—
3. Prism Test Mode:
7
X—Y: Cartesian Coordinates
P—B:
Polar Coordinates △
mm: mm expression
4. Reading resolving power: 0.01m-1
5. Multi-beam type: double-beam, progressive multi focus.
6. High Refraction ratio Lens Compensation: ABBE value 30-60
7. Display: LCD
8. Diameter of Compatible Lens: 10mm-90mm
9. Power Supply: AC 220V
50Hz or AC 110V 60Hz
10. Max Power: 55W
11. Weight: 7KG
12. Outline Size: 200(W)x 290(D)x 450(H)mm
13. Optional Accessories:
P. Thermal Printer
8
Chapter 3
Using The Instrument
3.1 Names of out elements and parts
Power indicator light
Slice handle
Monitor contrast
Adjust hole
Nose bridge block
Center print
handle
Lens base
Baffle move
handle
Power on/off
Fig. 1. LM-280 Auto Lensmeter Outline
LM-280 Auto Lensmeter Outline showed as Fig. 1, Names of the elements
and parts as follows:
3.2
INSTALLATION AND OPERATION
1.Preparation
a)To install the instrument on the table with no sunlight and intensive
reflection, avoiding heat, wet and dust.
b) To get rid of the sub of center print ink box.
c) To pull up the cushion handle to get rid of the dust-proof cushion.
pull the handle and the framework is settled on the top, repeat the
operation and the framework set down on the bottom.
d) To collect and check all accessories, documents and put them to right
place of the tool box.
e) To check the power supply and make it satisfied with the instrument,
then plug the two ends into the instrument and socket respectively.
9
f) To turn on the power
“Initial ” Screen (Fig.2) is showed; soon it will convert to Screen 1 (Fig.
3)
Attention: if put the lens on the bracket before opening the power, the instrument will not
convert to Test Screen after the “open” Screen but to display the hint “ INITAI ERROR” under
the Screen, please take away with the lens and reopen the power.
2.Screen and Menu
A. Initial Screen
Turn on the AUTO LENSMETER, the “Initial” Screen is showed on the
display
B. Test Screen
The Test Screen 1 is displayed automatically after the “Initial” Screen
(Fig. 3).
Fig. 3 Test Screen
The Test can be made both under the state of Screen 1 and Screen 2.
● Operation key is under the control Screen of display. Don’t press the function
indicator on the LCD, otherwise the display will be damaged.
C. Menu Screen
The choices of Test, object, method and unit parameters etc. are made
through the Menu Screen. When press the “:Menu” key, the target Screen
is displayed (Fig.4).
10
Once the move-up (move-down) key is pressed, the selected
parameter will move up (or down) one line. Now press the move-up or
move-down key, the favorable option is available. The selected option on
the Screen is marked with white beam and up-towards arrow. When the
selection is finished, return to the Test Screen.
Fig.4 Menu Screen
Each parameter option is given as (table 1) shows:
Table 1.
Parameters options
Parameters
Options
Step
0.01D, 0.12D, 0.25D
Pole lens
Mixture, +, -
Len
Normal, Contact Lens
Prism
Not display, X-Y, P-B
Save
Close, 5 min, 10 min, 30 min, 60 min,
Auto-memory
Close, Near center, Aim at center
ABBE Constant
60, 50, 40, 30,
Time
2005.06.23
Language
05:24
Chinese,English
11
3.3
Measuring
Caution: Please read the introduction closely, be familiar with function, operation of every
element and corresponding instructions before use.
A. Test single beam lens
1. Stop the cushion framework on the top and put the convexity of the
lens on the bracket upward, hold the lens by hand and raise the
cushion slowly to the top to fix the lens.
2. Move lens and when the lens is close to the drone, the “0” type drone
will be show on the display. When optical center is parallel with drone,
the horizontal line will become longer and the cursor will display the
hint “MARKING OK” .Then you can get the data from it.
3. Press “MEMORY” if you want to save the data ,then “S”
showed with contrast color.
will be
If printing is necessary, please press
“PRINT”, otherwise press “CLEAR”.
4. Identify left lens or right lens
Press the left lens key or right, then “L” appears on the top of left side
or “R” on the top of right side of Test Screen.
If it is not requested to identify left and right lens, then “S” will appear
on the right top of the Screen.
B. Put the lens with eyeglass framework
1) ensure the lens left or right as the same way.
2) put lens
put the lens on the bracket with the convexity of the lens upward and
the top
of the framework towards the operator.
3) adjust lens baffle
make the baffle close to bottom of the framework by moving the baffle
handle.
12
4) Fix lens
press the lens by cushion framework
Attention: the state will be kept when the instrument be reopened after Screen check and
option.
C. Check lens
Make options of Step, prism, auto-memory before measure lens. The Step
of 0.01D, 0.12D, 0.25D represents different resolving capacity. Four
options of prism represent respectively as follows:
Close…… not display
X-Y……
Cartesian Coordinates
P-B …… Polar Coordinates
Three items of auto-memory option will be explain later.
A. Check single beam lens
Move lens and when the lens is close to the drone, the “0” type
drone will be show on the display. When optical center is parallel with
drone, the horizontal line will become longer and the cursor will
display the hint “MARKING OK”. At the same time the sound “beep”
will be heard and all parameters checked will be displayed and the
“L” “R” “S” will be showed with contrast color.
To auto-memory, the hint sound “close to center” is easy to
operate but the Step is low; while “ in center” is with high Step but
require more Caution and care. Operator can choose mode as he likes
Attention: if data is saved, the lens must be taken away before the next Test, press
clear key. The lens can be put on the bracket for new Test only after the data is
cleared
.
If the option “ Auto-memory” is closed, the cross line is drawn
longer but the hint sound will not be heard when the lens optical
center is close to the drone. It has the same Step with optical center
focus. Press the auto-memory key and then the data can be latched by
13
manual.
Fig. 7 and Fig. 8 show the second Screen during the Test, and Fig.
9 is Screen after the operation is
finished.
Fig.5Start Screen of Test
Fig6.opticalcenter is close to the Screen
Fig.7. Final Screen of Test
B.
Check lens with framework
Lens fix method is same with A,1, while measure method is same
with B,2 for single beam lens, here the lens position can be changed
by moving framework.
C.
Measuring two beams lens
1) Put the lens on the bracket with the convexity side to the
far-sighted part. (Fig.8)
2) Measure the parameters of far-sighted part with the same
14
method for single beam lens .
3) Press the memory key under the bracket to confirm and record
the measured parameter of far-sighted part
4) Press “ADD” function key as displayed on Fig. 11
5) Draw the lens outward and move properly to make the myopic
part on the center of the bracket (Fig. 12), the result of ADD
will be showed on the displayed. (Fig. 13)
6) Repress the memory key to save the data.
Hints: fine focus is not required when test ADD value. But the sub-lens must
be in scope of penetration hole of lens bracket.
Fig. 8 installation of the far-sighted
Fig. 9 the display after the ADD part for double
beam lens
key is pressed for double beam lens
Fig. 10 installation of the myopic
part for double beam lens
Fig. 11 the display of measure results
for double beam lens
15
Fig. 12 the mark of progressive
multi-focus lens
D.Check the uncut progressive multi-focus lens
Both of the far-sighted and myopic parts of uncut progressive
multi-focus lens are marked (Fig.12). Measure the parts according to
the marks
1. Check by the Test Screen.
a) press “ADD” key to enter into Progessive Screen.(Fig.13)
b) Put the convexity upward and myopic parts towards inside, keep the
datum line horizontal and put the lens on the bracket, then fix it.
Make the far-sighted optical center focused on the optical enter of
the instrument, record and save the value
If the hint voice items of the automatic options are selected, then
instrument will emit “beep” when far-sighted optical center focus, the
parameters is confirmed and recorded, memory key is unnecessary.
c)
Draw the lens outward to make the myopic part close to the optical
center of the instrument. The display is showed as Fig. 14. When the
optical focus “+” is in the same position with the center line of the lens,
while the optical center “+” is close to the lower horizontal line of the
lens, two focuses are made together. The hint sound emits and the
parameter value ADD of myopic part is displayed. (Fig. 15)
Attention: if the optical center “+” line of the myopic part is in superposition with the
upper line of the lens, the optical center “+” line is lower than the bottom horizontal line of
the upper lens, the instrument will automatically latch and emit hint voice,. The ADD value
is displayed, but the checked point is still in the progressive zone and ADD value is
small.
16
Press the progressive key to return to the Test Screen.
Fig. 13. Progressive Screen
Fig. 14 Test finished for far-sighted part of progressive
Test is finished
17
Fig. 15 Progressive view zone when the
E. Check progressive multi-focus lens with framework
Check with progressive Screen. Operation must be careful because of
no mark on the lens.
a) Put the lens with framework in the same way .
b) Measure parameters of far-sighted part first. The center of far-sighted
part is in the scope close to the framework center.
Do the same way .
c) Measure parameters of myopic part.
Move the lens outward by handle and baffle. Move the framework
between left and right along the direction near the framework center
and myopic part optical center to look for the optical center of myopic
part. The optical center “+” of myopic part may be not able to cross
with the mark center of lens, so the max value should be recorded and
then press the memory key.
F. Check for high refraction ratio lens
When high refraction ratio lens is measured, please select ABBE
value from the menu, which is correspondent to value of the lens. Then
enter into the Screen to measure. The operation is finished the displayed
parameters have been made some compensation for ABBE values.
1.Test of eyeglass optical center.
a) Select the “memorized” from the auto-memory menu, return to the
Test Screen, press the right (or left) eyeglass lens key.
b) Make the eyeglass framework face to the operator and convexity
upward, put the nose bridge block lock into the concave position,
adjust the lens baffle, make the framework bottom against the baffle
and put the right (or left) lens on the bracket.
c) Make the framework always close to the baffle through nose bridge
block when move left to right, forward or back by parallel. When the
optical centers of lens and instrument in the superposition, the hint
sound “beep” emits. The screen display the parameters of right (or left)
lens, at the same time, the right (or left)
d) part horizontal distance value RPD (LPD) from the optical center is
showed under the left side (or left) part.
e) Move the nose bridge and framework together right (or left), put the
left (or right) eyeglass lens on the bracket, look for the lens optical
center following c), when the hint sound stop, the right (or left) part
horizontal distance value RPD (LPD) will be displayed under the left
side (or left) part, the optical center horizontal distance PD value is
showed too.
Fig. 16 display of the lens
Fig. 17 display of the optical center horizontal distance
2. Print
When print is necessary after the check is finished, press the memory
key first, then press the print key the instrument will send out the results
from the print on the right side. When do this the option “On” must be
held.
18
Chapter 4
A.
Other Instructions
Revise
AUTO LENSMETER is afforded with the function for revising the
error. When used for long time, error between the check value and real
value is intolerable. It can be amended with the standard lens. It is
complicate and professional man is required.
B. Maintenance
a) Please prepare a dust-proof cover and cover the instrument when it is
free in order to reduce the influence of dust.
b) Please set screen protection when the instrument
is in use, if it is free of test the screen can shut
automatically.
lens
bracket
protecting
lens
c) Clean the protecting lens regularly. Take
down the bracket first and clear the dust
or oil with soft brush or lens paper.
Warning:Don’t scratch the protecting
lens, otherwise the precision will be
reduced.
d) Check the filter regularly. Clear it lightly
with soft brush or lens paper to get rid
of the deposited dust.
filter
lens
Warning:Don’t scratch the filter.
e) Add printer oil
when the sign is dim the printer oil should be added. Draw out the ink
20
box to the table, open the cover, infuse the ink on the sponge. When
work is finished recover the box and get it back.
f) Replace print paper
Resume the paper according to the instruction.
g) Replace fuse
shut down the power before replacement, pull off the power line and
take out the outlet under the power socket and replace it. Pay Caution
to the fuse type: the required voltage and current.
h) Surface clear
Wipe the surface with soft and dry rag. If the oil is heavy, the neutral
wash can be used with the rag. Then dry it with cloth. Warning: organic
solvent and gas are forbidden.
C. General faults treatment
General faults of LM-280 AUTO LENSMETER can be treated as
follows:
SN
1
Fault character
Power indicator lamp
doesn’t flash
Reason
Solution
Fuse broken
Replace fuse
2
Display is dim
Lightness is too low
Adjust the lightness
3
Display is blurry
Contrast is too weak
Adjust the contrast
4
Display disappears
suddenly
Press the memory key or
The screen protection works
There
is
lens
on
Start the screen again
the
bracket,”INITAI ERROR”is
displayed under the Screen
5
Take out the lens and
restart the instrument.
Stick at Start Screen
No lens but still display
Clear the bracket and
“INITAI ERROR” too much restart the instrument
dust
Too
6
intensive No paper or paper is installed Install or reinstall the
circumstance light
incorrectly.
21
paper correctly.
ADD of ball, column The “PRINT” function is off in Open
and prism go beyond
7
8
the menu.
Mark on lens is not
clear or invisible
“PRINT”
function in the menu.
No paper is out when Paper run our or not install
press the print key
the
Install the paper again
Printer is not open
Restart the printer
Print oil is dry
Add oil
LM280 Series coke-meter calibration data that
In view of the use of this product in the process of
measuring data error problem, a special new features
calibration measurements. This feature is based on
measurement of national production house vertex power
of the lens in the standard spherical 0 to ± 20D in
lens as the standard for this product data calibration
operation.
Note: This product is calibrated before the factory
to see the machine code, code to the factory V at the
beginning of the machine that has the calibration
function.
Important
In the calibration, the machine should be shut down under the
lens measuring bearing down, the optical-glass wipe clean,
well-being re-measurement of lens support, and then opened
machine calibration (the picture shows off the lens measuring
sticks Block of optical-glass cleaning)
Light-gl
ass
Measuring
support
22
By the following steps calibration operation:
1. to ensure that light-cleaning glass, open the
machine. (The picture shows the normal boot
interface)
2.under the light of a map icon, holding down the key
blank (about 3 seconds) until the hearing, Di. ...
Di continuous sound, the hand can be let go. Into the
machine calibration, the screen will show if the
interface is shown in Figure 2 into the machine to
calibrate the model, when the space bar to do the
calibration button to the top of the screen
"calibration model" prompt.
Space
Diagram 1
23
Diagram 1
Calibra
tion key
Diagram 2
3.and then take the standard lens, lens normal
measurement, when the target at the center into a rough
sha
pe
of
the
cro
sslin
e
cro
ss
(X),
suc
h
as
24
that shown in the diagram under the
4.and then click the button calibration, the machine
can automatically complete the lens of the current
degree of automatic calibration, the calibration
process that follows the screen appear as shown in
Figure suggest that the current standards that mirror
the degree of calibration is in progress.
25
5.the current standard lens calibration is
complete, the screen appears as follows map "has
been completed!" Prompt.
Then remove the current standard lenses, said the
resumption of the screen interface shown in Figure
2.
26
6.and then on the need to calibrate the degree lens
can be calibrated in order.
7.calibration operation is completed, the machine
re-boot, after the calibration of the machine can be
used to normal.
Prompts the question:
●
calibrated machine, follow the
carefully calibrated operation.
●
If the calibration process to operate properly,
there might be death. At this point can only
restart the machine.
●
steps
After resuming the original calibration has
been the lens is no need to re-calibrate again。
When you press the button calibration, the calibration
into the state, and do not want to calibration, restart
your machine, measuring return to normal status.
27