Dako Autostainer User Guide

Dako Autostainer User Guide
Add to My manuals

Below you will find brief information for Autostainer. The Dako Autostainer is an automated slide processing system compatible with currently available reagents for the staining of paraffin-embedded and frozen tissue sections, cytospins, cell smears and fine needle aspirates. This system is designed to automate manual staining methods routinely used in immunohistochemistry and cytochemistry, enabling the transfer of established protocols from the bench to the Autostainer. Alteration of standard reagents or experimental conditions may not be required.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Dako Autostainer User Guide | Manualzz

Dako Autostainer |

User Guide

Document Number 0003106

Revision E

March 2007

For In Vitro Diagnostic Use

UL and cUL

Dako Colorado, Inc.

4850 Innovation Drive

Ft. Collins, CO 80525 USA

970-226-2200

Dako Autostainer Information

Dako Autostainer Model:

_______________________

Computer Configuration:

_______________________

Monitor: _______________________

Dako Autostainer Software Product License

IMPORTANT: Please read the license carefully before using this Software. The right to use this Software is granted only if the Customer agrees to the terms of this license. If you do not agree to the terms of this license, you may return the unused Software for a refund.

HOWEVER, USE OF THIS SOFTWARE INDICATES YOUR ACCEPTANCE OF THESE

TERMS AND CONDITIONS.

In return for the payment of the applicable one time license fee included with the purchase price of the Dako Autostainer, Dako grants the user a license to use one copy of the

Software subject to the following terms and conditions. No title or ownership of the Software is conferred with this license.

1. The Software may be used on one Dako Autostainer.

2. The Software may not be copied.

3. The Software may not be disassembled, decompiled, decrypted or reverse engineered.

4. Dako reserves the right to terminate this license upon breach. In the event of termination at Dako’s election, all copies of the Software must be returned to Dako.

5. No title or ownership of the Software is conferred with this license.

Copyright © 2007 Dako Colorado, Inc. All rights reserved.

This document may not be copied in whole or in part or reproduced in any other media without the express written permission of Dako. Please note that under copyright law, copying includes translation into another language.

2 Dako Autostainer | User Guide

Corporate

Headquarters

Denmark

Tel. 45 44 85 95 00

Fax 44 85 95 95

www.dako.com

Distributors in more than 50 countries

Australia

Tel. 2 9316 4633

Fax 2 9316 4773

Austria

Tel. 0800 0800 7153

Fax 0800 0800 7154

Belgium

Tel. 016 38 72 20

Fax 016 38 72 21

Canada

Tel. 905 858 8510

Fax 905 858 8801

Czech Republic

Tel. 420 541 423 710

Fax 420 541 423 711

Denmark

Corporate

Headquarters

Tel. 44 85 95 00

Fax 44 85 95 95

Sales

Tel. 44 85 97 56

Fax 44 85 84 29

C o n t a c t D a k o

France

Tel. 1 30 50 00 50

Fax 1 30 50 00 11

Germany

Tel. 040 69 69 470

Fax 040 69 52 741

Ireland

Tel. 353 91 768150

Fax 353 91 763674

Italy

Tel. 02 58 078 1

Fax 02 58 078 292

Japan

Tel. 075 211 3655

Fax 075 211 1755

The Netherlands

Tel. 020 42 11 100

Fax 020 42 11 101

Norway

Tel. 23 14 05 40

Fax 23 14 05 42

Poland

Tel. 058-661 1879

Fax 058-661 3390

Spain

Tel. 93 499 05 06

Fax 93 499 02 08

Sweden

Tel. 08 556 20 600

Fax 08 556 20 619

Switzerland

Tel. 041 760 11 66

Fax 041 760 11 77

United Kingdom

Tel. (0)1 353 66 99 11

Fax (0)1 353 66 89 89

Technical Support

Tel. (0) 1 353 66 99 65

United States of

America

Carpentaria,

California

Tel. 805 566 6655

Fax 805 566 6688

Technical Support

Tel. 800 424 0021

Customer Care

Tel. 800 235 5763

United States of

America

Fort Collins, Colorado

Flow Instrumentation

Tel. 800 822 9902

Fax 970 226 0107

Dako Autostainer | User Guide 3

This page intentionally blank.

4 Dako Autostainer | User Guide

Table of Contents

Section 1 | Introduction ........................................................................................................ 9

Dako Autostainer Features................................................................................................ 10

Universal System........................................................................................................... 10

Quick Start-up................................................................................................................ 10

Flexible Programming.................................................................................................... 10

Economic Use................................................................................................................ 10

Data Management and Report Generation.................................................................... 10

Slide and Reagent Labels.............................................................................................. 10

Section 2 | System Specifications..................................................................................... 11

Hardware Specifications.................................................................................................... 13

Software Specifications ..................................................................................................... 14

Accessories Included with the Dako Autostainer .............................................................. 14

Section 3 | Installation Requirements ............................................................................... 15

Requirements .................................................................................................................... 15

Recommendations ............................................................................................................ 15

Section 4 | Typical Operation............................................................................................. 17

Section 5 | Programming Overview................................................................................... 19

Menu Bar Options ............................................................................................................. 20

Slides Function .............................................................................................................. 20

File Menu ....................................................................................................................... 20

Edit Lists Menu .............................................................................................................. 21

Copy Function................................................................................................................ 21

Auto Function................................................................................................................. 22

Command Buttons............................................................................................................. 22

Section 6 | Accessing the Dako Autostainer Software.................................................... 23

Section 7 | Initializing the Dako Autostainer .................................................................... 25

Entering Institution Information.......................................................................................... 26

Editing Institution Information ............................................................................................ 27

Entering Staff Information.................................................................................................. 27

Adding a New User........................................................................................................ 27

Editing User Information ................................................................................................ 28

Deleting User Information .............................................................................................. 29

Adding a New Doctor..................................................................................................... 29

Deleting Doctor Information ........................................................................................... 30

Entering Tissue Information .............................................................................................. 30

Adding a New Tissue..................................................................................................... 30

Deleting a Tissue ........................................................................................................... 30

Entering Default Reagent Information ............................................................................... 31

Assigning Default Reagent Dispense Volumes ............................................................. 31

Assigning Reagent Drop Zone....................................................................................... 31

Cleaning Schedule......................................................................................................... 32

Label Printing Initialization................................................................................................. 32

Slide Labels ................................................................................................................... 32

Reagent Labels.............................................................................................................. 33

Label Adjustment ........................................................................................................... 34

Communication Port ...................................................................................................... 34

Printer Selection ............................................................................................................ 35

The Options Screen .......................................................................................................... 35

Program Grid ................................................................................................................. 35

IHC Report..................................................................................................................... 36

Custom Options ............................................................................................................. 36

Section 8 | Entering Slide Information .............................................................................. 39

Slides Function.................................................................................................................. 39

Dako Autostainer | User Guide 5

Adding Slides................................................................................................................. 39

Deleting a Slide.............................................................................................................. 40

The Slide Information Button............................................................................................. 40

Entering Slide Information Through the SLIDE INFORMATION Button ........................ 41

Deleting Slide ID Information through the SLIDE INFORMATION Button ..................... 42

Deleting a Slide Identification ........................................................................................ 42

Deleting a Case # .......................................................................................................... 43

Deleting a Block ID ........................................................................................................ 43

Deleting a Specific Slide ................................................................................................ 44

Slide Rearrangement ........................................................................................................ 44

Rearranging slide positions on the Programming Grid .................................................. 44

Section 9 | Designing a Protocol ....................................................................................... 45

Protocol Elements ............................................................................................................. 46

Rinse Buffer ................................................................................................................... 46

Rinse Water ................................................................................................................... 46

Standard Rinse Replacement........................................................................................ 46

Substrate-Batch ............................................................................................................. 47

Switch (*Swtch).............................................................................................................. 47

Creating and Editing a Protocol Template......................................................................... 47

Saving a Protocol Template........................................................................................... 48

Deleting a Protocol Template ........................................................................................ 49

Using a New Protocol Template Without Saving It ........................................................ 50

Using a Saved Protocol Template ................................................................................. 50

Using a Saved Protocol Template for a Specific Auto Program .................................... 50

Selecting and Editing the Reagent Dispense Volume....................................................... 51

Selecting and Editing the Reagent Dispense Volume for All Steps in a Protocol.......... 51

Selecting and Editing the Reagent Dispense Volume for a Specific Step in a Protocol 52

Section 10 | Programming Slides ...................................................................................... 53

Defining Slide Specific Reagents ...................................................................................... 53

Defining Detection Reagents (Secondary, Tertiary, Substrate, etc.) ............................. 53

Defining Primary Antibodies .......................................................................................... 54

Assigning Positive and Negative Control Reagents ...................................................... 55

Defining Specific Rinses ................................................................................................ 55

Auto Programming ............................................................................................................ 56

Creating and Saving an Auto Program .......................................................................... 56

Deleting an Auto Programming Item.............................................................................. 57

Using an Auto Programming Item.................................................................................. 58

Getting Help with Auto Programming ............................................................................ 58

Copy/Paste........................................................................................................................ 58

Copy/Paste Individual Reagents.................................................................................... 58

Copy/Paste Entire Rows................................................................................................ 59

Editing a Protocol Step...................................................................................................... 59

Changing a reagent for a step: ...................................................................................... 59

Replacing a reagent with skip step (no reagent added to slide): ................................... 60

Editing a reagent in a staining step for one run only:..................................................... 60

Printing Options................................................................................................................. 61

Slide Labels ................................................................................................................... 61

Reagent Labels.............................................................................................................. 61

Printing reagent labels not from a program. .................................................................. 62

The Main Grid Report .................................................................................................... 62

Immunohistochemical Report (IHC Report)................................................................... 63

Run Log Printout............................................................................................................ 63

Viewing Programmed Slides.......................................................................................... 64

Assigning Reagent Dispense Locations............................................................................ 65

Assigning Reagent Dispense Volumes ............................................................................. 66

6 Dako Autostainer | User Guide

ALL SLIDES Tool .............................................................................................................. 67

Printing a Slide Layout Map .............................................................................................. 67

Section 11 | Reagent Management.................................................................................... 69

Adding New Reagents to Reagent Lists............................................................................ 69

Adding Detection Reagents (Secondary, Tertiary, Substrate, etc.) ............................... 69

Adding Primary Antibodies ............................................................................................ 71

Editing Reagent Lists..................................................................................................... 73

Deleting Reagents from Reagent Lists .......................................................................... 73

Updating a Reagent Lot Number and Expiration Date .................................................. 74

Printing Reagent Labels .................................................................................................... 75

Printing reagent labels from a programmed run. ........................................................... 75

Printing reagent labels not in a program........................................................................ 75

Compatibility Check........................................................................................................... 76

Section 12 | Loading Reagents.......................................................................................... 79

Reagent Layout Map Screen............................................................................................. 79

Reagent List Screen.......................................................................................................... 80

Missing Reagent Error....................................................................................................... 81

Paused to Add Reagent................................................................................................. 81

Slides Skipped ............................................................................................................... 81

Partial Reagent Application ........................................................................................... 82

Section 13 | Loading Slides ............................................................................................... 83

Loading Microscope Slides ............................................................................................... 83

Section 14 | Starting a Run ................................................................................................ 85

Slides................................................................................................................................. 85

Racks ................................................................................................................................ 85

Reagents ........................................................................................................................... 85

Buffer................................................................................................................................. 85

Water................................................................................................................................. 85

Waste ................................................................................................................................ 85

Starting a Run without Delay............................................................................................. 86

Delaying the Start of a Run ............................................................................................... 88

Delaying the Start of a Run using Substrate-Batch ....................................................... 89

Section 15 | Completing a Staining Run ........................................................................... 91

Printing a Run Log............................................................................................................. 92

Printing a Run Log Without Saving................................................................................ 92

Printing and Saving a Run Log ...................................................................................... 92

Printing a Temporary Run Log....................................................................................... 93

Emergency Stop................................................................................................................ 93

Immunohistochemical Report............................................................................................ 94

Shutting Down the Autostainer.......................................................................................... 94

Section 16 | Programming During Staining Run.............................................................. 95

Reviewing the Current Program during a Staining Run .................................................... 96

Section 17 | System Maintenance ..................................................................................... 97

Automated Cleaning Cycle................................................................................................ 97

Cleaning Log ..................................................................................................................... 98

Reservoir Cleaning............................................................................................................ 98

Wash Buffer and Water Container Cleaning ..................................................................... 98

Slide Rack Cleaning .......................................................................................................... 99

Reagent Vial Cleaning....................................................................................................... 99

Computer Maintenance ..................................................................................................... 99

Backing Up Files ............................................................................................................... 99

Annual Preventative Maintenance..................................................................................... 99

Section 18 | Reagent Tracking......................................................................................... 101

Printing the Reagent Tracking Report ............................................................................. 101

Printing the entire Reagent Tracking Report ............................................................... 101

Dako Autostainer | User Guide 7

Printing only selected reagents from the Reagent Tracking Report: ........................... 101

Remove Reagent from the Reagent Tracking Log ...................................................... 102

Remove all Reagents from the Reagent Tracking Log................................................ 102

Remove Selected Reagents from the Reagent Tracking Log ..................................... 102

Section 19 | Wash Buffers................................................................................................ 103

Dako Autostainer Wash Buffer ........................................................................................ 103

Intended Use ............................................................................................................... 103

Materials Required....................................................................................................... 103

Reagent Preparation.................................................................................................... 104

Shelf Life and Storage ................................................................................................. 105

Tris-Buffered Saline......................................................................................................... 105

Intended Use ............................................................................................................... 105

Materials Required....................................................................................................... 105

Reagent Preparation.................................................................................................... 105

Shelf Life and Storage ................................................................................................. 106

Dako Phosphate Buffered Saline .................................................................................... 107

Intended Use ............................................................................................................... 107

Materials Required....................................................................................................... 107

Reagent Preparation.................................................................................................... 107

Shelf Life and Storage ................................................................................................. 107

Phosphate Buffered Saline.............................................................................................. 108

Intended Use ............................................................................................................... 108

Materials Required....................................................................................................... 108

Reagent Preparation.................................................................................................... 108

Shelf Life and Storage ................................................................................................. 109

Section 20 | Troubleshooting........................................................................................... 111

Solving Autostainer Problems ......................................................................................... 111

Solving Software Problems ............................................................................................. 115

Solving Computer Problems............................................................................................ 116

Solving Monitor Problems ............................................................................................... 117

Solving Printer Problems................................................................................................. 119

Solving Seymour Label Printer Problems........................................................................ 121

Section 21 | Warnings and Symbols ............................................................................... 123

Symbol Definitions........................................................................................................... 124

Section 22 | General Limitations ..................................................................................... 125

8 Dako Autostainer | User Guide

Section 1 |

Introduction

The Dako Autostainer System is an automated slide processing system compatible with currently available reagents for the staining of paraffin-embedded and frozen tissue sections, cytospins, cell smears and fine needle aspirates. This system is designed to automate manual staining methods routinely used in immunohistochemistry and cytochemistry, enabling the transfer of established protocols from the bench to the Autostainer. Alteration of standard reagents or experimental conditions may not be required.

This document is a comprehensive User Guide for the Dako Autostainer System.

Dako Autostainer | User Guide 9

Dako Autostainer Features

Universal System

The Dako Autostainer is a universal system enabling the direct transfer of routine staining protocols from the laboratory bench.

Quick Start-up

Dako’s Autostainer Support Team installs and tests the system to ensure proper operation.

The factory-installed software includes standard reagent protocols and commonly used Dako reagents such as Dako LSAB2, Dako EnVision, and HercepTest™ visualization systems. The complete line of ready-to-use Dako N-series primary antibodies and their recommended protocols are also listed.

Flexible Programming

The Dako Autostainer enables the use of slide-specific programmed or user defined protocols.

The system provides intuitive programming routines, reagent and slide loading maps and start/finish run time displays.

You can build and save an unlimited number of protocols containing up to 35 steps (including rinse and blow steps between reagent incubations) and 64 different reagents. A staining run can process 1 to 48 microscope slides. Individual slides can be programmed to receive different reagents, reagent dispense volumes and reagent dispense locations.

Economic Use

The Dako Autostainer provides the flexibility to program reagent volumes of 100, 150, 200, 400 and 600

µL for any step in a staining protocol. Each step can be programmed with a different reagent dispense volume. In addition, waste can be segregated into hazardous and nonhazardous collection containers, reducing disposal costs.

Data Management and Report Generation

The Dako Autostainer is designed to track a variety of data. It can generate patient, reagent and real-time operation data reports, as well as track reagent usage and log instrument maintenance.

Slide and Reagent Labels

The Dako Autostainer software provides a simple and efficient means to print slide labels through one-time programming of information for the staining run. The software enables items from the program (slide ID, slide #, case #, block ID, technician, antibody(s), pretreatment, date, doctor, tissue, run name, and static text) to be printed directly on the slide labels. You can also create reagent labels that include information stored in the reagent list (reagent name, lot number, expiration date, incubation time, and date.) The printed label text is protected by a heat and chemical resistant laminate, allowing you to apply the labels to the slides at the start of the staining procedure.

10 Dako Autostainer | User Guide

Section 2 |

System Specifications

The Dako Autostainer is a combination of hardware and software integrated with the appropriate reagents to automate immunohistochemical and immunocytochemical staining methods. This section describes the hardware and software specifications and the accessories provided with the system.

Dako Autostainer | User Guide 11

Slide Rack

Glass Slide

Reagent Racks

1 and 2

Reagent Probe

Wash Station

Sink

Y axis mechanism

X axis mechanism

Z head Assembly

is a slide carrier holding 12 microscope slides in a horizontal position. The instrument holds a maximum of 4 racks. These racks are removable and interchangeable for slide loading and unloading. is a standard microscope slide to which specimen(s) is mounted for processing on the instrument. Slide dimensions are 25mm x

75mm x 1mm. hold the reagent vials. There are 2 reagent racks: reagent rack

1(white) sits in the front right side of the sink; reagent rack 2

(blue) sits behind reagent rack 1. The racks are identical in shape and size. Each rack holds up to 32 reagent vials and is marked with letters A-D and numbers 1-8 to designate each reagent vial position. The computer always assigns reagents to the front white rack first. If additional reagents are required for a procedure or for running the cleaning protocol the computer will assign reagents to the blue rack. is a white, plastic assembly located in the rear right corner of the instrument and is used as a cleaning station for the

Teflon

™-coated reagent probe. The probe is washed in the wash station at the beginning of every run, prior to checking reagent volumes, and application of each reagent. is the Teflon-coated tray that sits underneath the slide racks and channels the buffer/DI water washes and reagents into the waste reservoir. is the robotic arm extending from the back of the instrument towards the front carrying with it the Z head assembly. The Y arm moves from right to left across the instrument. When the

Autostainer is not in operation the Y arm rests on the left side of the machine. When the instrument is in operation and slides are incubating, the arm is located on the right side. enables the Y arm to move from left to right across the instrument. The X axis is stationary across the back of the instrument. is the dome-shaped assembly that sits over the Y arm. The Z head consists of three components: the probe tip, the wash head and the blow head. These components move up and down along the Z axis while the instrument is in operation. The

Z head moves back and forth along the Y axis mechanism, allowing access to all slides and reagents.

12 Dako Autostainer | User Guide

Hardware Specifications

Dako Autostainer

Dimensions

Weight

Electrical requirements

Normal operating temperature

Slide rack capacity

Total slide capacity

Reagent capacity

1.02m W x 0.69m D x 0.61m H

40" W x 27" D x 24" H

140 lb. (63.5 kg)

120V 110/120V (

±10%) 60Hz (±2Hz)

220V 220/240V (

±10%) 50Hz (±2Hz)

18

°C-26°C (64°F-79°F)

1 - 12 glass slides

1 - 48 glass slides

64 different reagents (15 mL/reagent vial)

100, 150, 200, 400 and 600

µL

100

µL minimum, 1200µL maximum

Reagent dispense volumes

Reagent probe volume capacity

Computer Controller

These are the minimum features provided. Dako reserves the right to change the computer controller specifications

at any time.

Monitor

Printer

Surge Protector

90MHz Intel Pentium

®

processor

16 MB RAM

840 MB hard drive

Quad-speed CD ROM

14.4 K bps fax/modem

Windows

®

compatible mouse

14" Color (.28 dot pitch)

LaserJet or Inkjet printer

120 VAC, 15 Amps, 50/60 Hz

Dako Autostainer | User Guide 13

Software Specifications

Operating System

Operating Logic

Protocol Logic

Idle Rinse

Rinse Default Settings

Incubation Time

Windows

®

95/98, or equivalent.

Designed to calculate the most time-efficient sequence to complete a programmed staining run.

Flexible selection of steps including rinses.

Maximum of 35 protocol steps (including rinse steps).

Buffer is applied to slides during a staining run to keep specimens wet when no reaction is occurring.

Buffer rinse every 30 minutes prior to the start of a run, when delayed start is selected.

Buffer rinse every 30 minutes during a staining run.

Water rinse every 60 minutes after completion of run.

Default settings for time and the amount of buffer dispensed can be changed by a Dako representative upon request.

Minimum time is 2 minutes.

Accessories Included with the Dako Autostainer

Slide Racks

Reagent Racks

Reagent Vials

Buffer Container

Deionized Water

Container

Waste Containers

User Guide

4 non-disposable slide racks with a 12 slide capacity/rack. The

slide racks are suitable for standard glass slides; they are

removable and interchangeable.

2 removable reagent racks. Each reagent rack holds a

maximum of 32 reagent vials.

100 reagent vials and caps. Each vial holds a maximum of 15

mL of reagent.

1 ten liter plastic container.

1 ten liter plastic container (not available with the “01” hardware version of the Dako Autostainer).

2 ten liter plastic containers.

1 User Guide with detailed operating instructions.

14 Dako Autostainer | User Guide

Section 3 |

Installation Requirements

For installation of the Dako Autostainer the following requirements should be met to ensure that the instrument functions properly.

Requirements

Surface

Size

Environment

Prior to unpacking the Autostainer, ensure that the area to be used for instrument is a solid, level surface that can safely support the weight of the system (approximately 240 lbs. for instrument plus computer.)

Minimum dimensions for the Autostainer work area are 2.00m wide x 0.8m deep x 1.0m height (78” w x 30” d x 37” h) clearance.

The work area must also be an environment with an ambient temperature between 18

°C-26°C (64°F-79°F), not facing direct sunlight and not affected by a hood or vent.

Recommendations

Power source

Dako recommends a dedicated power source for the Autostainer to prevent interference from other instruments or equipment.

Dako Autostainer | User Guide 15

This page intentionally blank.

16 Dako Autostainer | User Guide

Section 4 |

Typical Operation

A few “off instrument” preparations will make the staining run more time-efficient. Prior to starting a run you should have the following items:

• A list of slide cases to be run with requested protocols.

• Deparaffinized and rehydrated slides soaking in buffer. This buffer should contain 0.05%

Tween 20. A minimum of 5 minutes soaking time is recommended.

• The reagents required for the staining run brought to room temperature.

• Sufficient buffer in the buffer reservoir and deionized water in the deionized water reservoir.

The buffer should contain 0.05% Tween 20. (Deionized water is not available with the “01” hardware version of the Dako Autostainer).

• Containers for hazardous and non-hazardous waste with enough capacity to accommodate the waste from the staining run.

A typical set up for a staining run follows this order:

1. Start the software by double clicking on the Dako Autostainer icon.

2. Enter your user name and password. (See Accessing the Dako Autostainer Software,

Section 6)

3. Select the PROGRAM button on the Main Menu screen.

4. Select the SLIDE INFO button from the Programming Grid, or Slides from the menu bar.

When the SLIDE INFO button is selected the SLIDE INFORMATION screen is displayed.

When the Slides menu item is selected the SLIDE COUNT window is displayed.

5. Enter slide information or slide count for the staining run. (See Entering Slide Information,

Section 8)

6. Select the FINISH ENTRY button when entering slide information or select OK when using SLIDE COUNT window. The Programming Grid is displayed with the slide information.

7. Select the PROTOCOL TEMPLATE button. The PROTOCOL TEMPLATE DESIGN screen is displayed.

8. Select the protocol template for the run. (See Designing a Protocol, Section 9)

9. Select the USE TEMPLATE button. The Programming Grid is displayed with the selected protocol.

NOTE: Steps 7 through 9 can be skipped if the default template is used.

10. Assign specific reagents to all slides. (See Programming Slides, Section 10)

11. Select the NEXT button. The SLIDE LAYOUT MAP:PROGRAM SLIDES screen is displayed. Select the PRINT button to print the Slide Layout Map.

12. Select the NEXT button the RUN TIME screen is displayed. The Autostainer calculates the run time and amount of buffer and DI water needed to complete the run.

13. Select the OK button. The user is prompted to “Save program on disk”.

Selecting YES displays the SAVE PROGRAM ON DISK window. Enter the program’s name and select OK. The program is saved and the REAGENT LAYOUT MAP screen is displayed.

Selecting NO displays the REAGENT LAYOUT MAP screen and the program is not saved.

Dako Autostainer | User Guide 17

14. Select the PRINT button to print the Reagent Layout Map.

15. Load reagents using the Reagent Layout Map.

16. Select the NEXT button. The SLIDE LAYOUT MAP:LOAD SLIDES screen appears.

17. Load slides using the Slide Layout Map. It is recommended that each slide be flooded gently with buffer after it is placed on the instrument. This buffer should contain 0.05%

Tween 20. (See Loading Slides, Section 13) At no time should the slides dry out.

18. Select the NEXT button. The SET START TIME screen is displayed. (See Starting a Run,

Section 14)

19. Select the PRIME PUMP (WATER) button. The RUN PROGRAM NOW dialog box appears. (Deionized water is not available with the “01” hardware version of the Dako

Autostainer). The user is prompted: “Is the pathway clear for the arm to move.”

20. Select the YES button and the pump is activated. Deionized water is flushed through the water intake line.

21. Select the PRIME PUMP (BUFFER) button and the pump is activated, flushing buffer through the buffer intake line. For the “01” hardware version of the Dako Autostainer the user is prompted: “Is the pathway clear for the arm to move.”

22. Set Reagent Levels to Check volumes or Do not check volumes. (See Starting a Run,

Section 14)

23. Select the START RUN button and the run starts.

18 Dako Autostainer | User Guide

Section 5 |

Programming Overview

The Dako Autostainer is a computer controlled system. The PROGRAMMING GRID is the main screen used to define slide information, primary antibodies, staining reagents and staining protocols.

Menu Bar

Grid Header

Programming

Grid

Program

Name

Software

Version

Date/Time

User Name

(security level)

Institution

Name

Autostainer serial number

Command

Buttons

1. Five menus are accessible from the menu bar. The Slides menu item is used to enter the number of slides To be run on the Autostainer. The File menu is used for general information such as opening, saving and printing previously saved runs. The Edit Lists menu is used to establish lists of reagents utilized in programming a staining protocol.

The Copy and Auto menu items are used during programming.

2. The grid header lists the steps and dispense volumes of the selected protocol template for the current staining run. Each column represents one step of the protocol template.

3. The programming grid is divided into rows and columns. Each row represents one slide and displays the complete slide identification (e.g. Slide ID, Case #, Block DI, Doctor

Name, and Tissue) and all reagents applied. The grid will accommodate a maximum of 48 slides per staining run.

4. The 6 command buttons, located at the bottom of the screen, are used to enter slide information, develop staining protocols (templates) and print reports.

5. Below the command buttons is a line of text that is displays the software version, date/time, user name, security level, institution name, and Autostainer serial number.

This information is edited from the INITIALIZE screen (see Initializing the Dako

Autostainer, section 7). This line is on the bottom of most screens of the Autostainer software.

Dako Autostainer | User Guide 19

Menu Bar Options

Slides Function

The Slides option may be used as an alternative to expedite programming by reducing the amount of information required to program a staining run. The Slides option may be used to enter the total number of slides for the current staining run.

The Slides function can be accessed by moving the mouse pointer to Slides on the menu bar and pressing the left mouse button. This selection displays the SLIDE COUNT window for entering the number of slides for the current staining run. This function is used to bypass the SLIDE

INFORMATION screen when programming a staining run. Enter the desired number of slides in the text box or use the scroll bar to select the desired number of slides.

Once a run has been set up using the SLIDE COUNT window, slide ID, case number, doctor, block ID or tissue information cannot be added to the run. Selecting the SLIDE INFO button displays the SLIDE COUNT window.

NOTE: Slide Identification, Case #, Block ID, Tissue or Doctor Name will not be

associated/displayed in the Programming Grid if the Slides function is used.

File Menu

The File menu can be accessed by moving the mouse pointer to File on the menu bar and pressing the left mouse button. The File menu commands include:

New

Open

Save

displays a blank grid to program a new staining run. displays the Load Program From Disk window with a list of all saved staining runs available to be opened. saves the current staining run program.

Save As

Print

saves an existing staining run program under a new name while the original program remains under its saved name. prints the current Program Grid, IHC Report, Slide Labels, Reagent Labels,

Reagent List, or any Run Log.

NOTE: The list at the bottom of the file menu displays the last five programs accessed in the computer. This list is constantly updated.

20 Dako Autostainer | User Guide

Edit Lists Menu

The Edit Lists menu can be accessed by moving the mouse pointer to Edit lists on the menu bar and pressing the left mouse button. For instruction on the features of the Edit List menu item, see

Reagent Management, section 10. The Edit Lists menu commands include:

End. Enz.

Block

displays the Edit Reagent List screen for endogenous enzyme block, allowing programming and editing of endogenous enzyme block reagents.

Protein Block

displays the Edit Reagent List screen for protein block, allowing programming and editing of protein block reagents.

Primary

Antibody

displays the Edit Reagent List screen for primary antibodies, allowing programming and editing of primary antibody reagents. The

PRETREATMENT list is also accessible from this screen.

Pretreatment

displays the Edit Reagent List screen for primary antibody pretreatment, allowing programming and editing of pretreatment reagents.

Secondary

Reagent

Tertiary

Reagent

Labeled

Polymer

Substrate

displays the Edit Reagent List screen for secondary reagents (e.g., biotinylated secondary antibodies or link), allowing programming and editing of secondary reagents. displays the Edit Reagent List screen for tertiary reagents (e.g., enzymelabeled streptavidin or label), allowing programming and editing of tertiary reagents. displays the Edit Reagent List screen for labeled polymer reagents used in the Dako EnVision visualization systems, allowing programming and editing of labeled polymer reagents. displays the Edit Reagent List screen for substrates, allowing programming and editing of substrate reagents.

Auxiliary

displays the EDIT Reagent List screen for auxiliary reagents (e.g., counterstain), allowing programming and editing of auxiliary reagents.

Copy Function

The Copy function can be accessed by moving the mouse pointer to Copy on the menu bar and pressing the left mouse button. This function activates the Select and Paste commands which allow selection of programmed tiles in the Programming Grid. The selected tiles can be copied over (pasted) on either unprogrammed or programmed tiles. (See the Copy section of

Programming Slides, section 10)

Dako Autostainer | User Guide 21

Auto Function

The Auto function can be accessed by moving the mouse pointer to Auto on the menu bar and pressing the left mouse button. This function is used to save all or any portion of a protocol to be used in other programs. (See the Auto Programming section of Programming Slides, section 10)

The Help function is also accessible from this menu.

Command Buttons

The command buttons on the PROGRAMMING GRID and their respective functions are:

Slide Info

displays the SLIDE INFORMATION screen to enter new or edit existing slide information (e.g. Slide ID, Case #, Block ID, # Slides, Tissue and Doctor

Name).

Protocol

Template

Next

displays the Protocol Template Design screen where a saved protocol template can be retrieved or a new one created. This template is applied to the current staining run. displays the SLIDE LAYOUT MAP:PROGRAM SLIDES screen.

Print

Exit

Help

prints the Slide labels, Reagent Labels, Main Grid, IHC Report, Reagent

List, or Run logs. returns to the MAIN MENU screen without saving the program. displays information to aid in the understanding and utilization of features in the Programming Grid.

22 Dako Autostainer | User Guide

Section 6 |

Accessing the Dako Autostainer

Software

The SIGN IN screen is the first screen that appears after the Dako Autostainer icon is double clicked.

The SIGN IN screen allows you to enter your assigned user name (up to 28 alpha-numeric characters) and password (up to 8 alpha-numeric characters). Your user security status (programmed in the INITIALIZE screen) determines the functions you are able to access. If the sign-in process is executed correctly (i.e. valid user name and password entered), the system displays the MAIN MENU screen.

To add a new user or change a password of an existing user, see Initializing the Dako Autostainer,

Section 7.

1. The cursor is located in the name box. Type your name and press the ENTER key. The cursor moves to the password box.

2. Type your password and press ENTER. Only asterisks (*) will be displayed in the entry box during keystroke entry. The MAIN MENU screen is displayed.

If an unrecognizable name and/or password is entered, the software does not allow you to continue to the next entry. For example, if an unrecognized name is entered and the ENTER key is pressed the cursor remains in the name box and does not move to the password box.

If EXIT is selected a dialog box prompts, “Quit using Autostainer?”. Select YES to return to the main Windows screen. Select NO to return to the SIGN IN screen.

The MAIN MENU screen appears when the sign-in process is properly executed. This screen allows high security level users access to all the software features. The INITIALIZE button is displayed for only high security level users. A user’s security status is entered from the INITIALIZE screen (see

Initializing the Autostainer, section 7).

If the accessing operator is assigned a middle or low security level, the INITIALIZE button is replaced with the CHANGE PASSWORD button. A middle or low security level does not allow access to set or change the parameters for slide labels, Main Grid and IHC Report printing formats, cleaning schedule or signing on new users and doctors.

Dako Autostainer | User Guide 23

NOTE: In the example below, the INITIALIZE button is displayed, indicating this user is set at high security level.

NOTE: The software version number, current date and time, user name, security level (indicated by asterisks), institution name, and Autostainer serial number are displayed across the bottom of this screen.

NOTE: A cleaning run is recommended after 150 slides have been processed on the Autostainer.

Once the cleaning cycle is set up (see System Maintenance, section 16) this number and the number of slides processed since the last cleaning appear in red letters. Once the instrument has processed

150 slides, a message will prompt the user to clean the instrument.

The software allows up to 200 slides to be run between cleaning cycles. The additional 50 slides allows for flexibility in the laboratory work flow to perform an additional run before cleaning the instrument. If more than 200 slides have been run the instrument must be cleaned. The PROGRAM button becomes unavailable and the software does not allow additional programming until the instrument has gone through a cleaning cycle.

PROGRAM

is used during routine operation of the Autostainer and allows access to the

PROGRAMMING GRID.

INITIALIZE

CHANGE

PASSWORD

CLEAN

REAGENT

TRACKING

displays the INITIALIZE screen used to establish and update default information for institution, staff, the number of slides per routine cleaning cycle, the default reagent volume and drop zone for reagent dispense. This button is only displayed for high security level users. displays the CHANGE PASSWORD screen used to edit the signed on user’s password. This button replaces the Initialize button for users with middle and low security access. allows you to initiate a cleaning run. It gives a status report on the number of slides since the last cleaning cycle and the total number of runs and generates a

Cleaning Log. allows users to monitor the total volume of each reagent that has been used on the Autostainer.

PRIME PUMP

displays the PRIME PUMP screen used to flush the intake tubing with buffer or DI water without initiating an actual run.

HELP

SIGN OFF

displays information to aid in understanding and utilizing the features on the MAIN

MENU screen. exits the Autostainer program

24 Dako Autostainer | User Guide

Section 7 |

Initializing the Dako Autostainer

The INITIALIZE screen is used to establish and update default information for the system. This includes the institution information, Autostainer serial number, user names, the doctors requesting

IHC tests, a library of tissues that may be used during staining, the number of slides allowed between routine cleaning runs, the default volume and the drop zone for reagent dispense. Slide and reagent label formatting, and user-defined printing formats for the Main Grid and IHC Reports, and custom options can also be configured from this screen.

This screen is used to enter information for: Institution, Autostainer serial number, staff, default reagent volume and drop zone, slides per cleaning cycle, and to access the Slide and Reagent

Design Label and Options screens.

Institution

Information

The Institution Information (name, address, department, phone, serial number) is stored and incorporated into reports.

Staff

Information

Tissue

Information

Default

Reagent

Information

The Staff Information includes lists of the Users and Doctors who frequently request IHC tests. The list of Doctors is accessed during the programming of the slide information.

The Tissue Information list is used to save a list of tissues that are frequently used in IHC tests. This list of tissues is accessed during the programming of slide information.

The Default Reagent Volume allows the selection of the default reagent volume to be used for all steps in any staining protocol. Reagent volume choices are: 100, 150, 200, 400, and 600

µL. The Default Reagent Drop

Zone is also selected, allowing application of reagent to single or multiple locations on each slide.

Cleaning

The Slides/Cleaning permits you to program the number of slides stained.

Schedule

150 slides. Once the programmed number is exceeded, the MAIN MENU screen displays the number of slides since the last cleaning. The maximum slides allowed between cleaning cycles is 200. Once this number is exceeded, the PROGRAM button is not available.

Dako Autostainer | User Guide 25

The buttons on the INITIALIZE screen and their respective functions are:

SLIDE

LABELS

displays the DESIGN SLIDE LABEL screen used to define slide label printing of slide labels.

REAGENT

LABELS

displays the DESIGN REAGENT LABEL screen used to define the printing of reagent labels.

OPTIONS

OK

CANCEL

displays the OPTIONS screen used to define the format for the Program

Grid, Slide Layout Map, and printing format for IHC Reports. saves the entered information and returns to the MAIN MENU screen. cancels any entries made and returns to the MAIN MENU screen.

PRINT

HELP

prints a report listing all of the institution , staff, tissue, default volume, default dispense zones, and cleaning schedule information. displays information to aid in understanding and utilizing the Initialize screen.

Entering Institution Information

INITIALIZE on the MAIN MENU screen. The INITIALIZE screen is displayed with the cursor in the Name box.

2. Type in the institution’s name and press ENTER or TAB.

The cursor moves to the

Address box.

NOTE: Information entry is limited to the space displayed on the screen. When the end of an entry box is reached, an audible tone is made by the computer and no additional information can be added in that box.

3. Type in the institution’s address and press ENTER or TAB. The cursor moves to the

Department/Lab box.

4. Type in the laboratory or department where the Dako Autostainer will be operating and press ENTER or TAB. The cursor moves to the Contact Phone/Ext. box.

5. Type in the laboratory’s phone number and press ENTER or TAB. The cursor moves to the Serial Number box.

6. Type in the Autostainer serial number located on the right side of the instrument. The number should start with “DC3400-“and followed by 6 additional numbers. Press ENTER or TAB. The cursor moves to the Users box.

NOTE: Once entered, the instrument serial number will be located in the lower right corner of many of the screens in the Autostainer software. the displayed.

Select the CANCEL button to escape without saving the entered information. The

MAIN MENU screen is displayed.

26 Dako Autostainer | User Guide

Editing Institution Information

the cursor in the Name box and the previously entered information listed.

2. Select the information to be changed and enter the new information. the displayed.

Select the CANCEL button to escape without saving the changes. The MAIN MENU screen is displayed.

Entering Staff Information

Adding a New User

INITIALIZE on the MAIN MENU screen. The INITIALIZE screen is displayed with the cursor in the Name box and the previously entered information is listed.

3. Type a user’s name and press ENTER. The

PASSWORD AND SECURITY LEVEL screen appears.

4. Type in a password for the new user in the Enter box and press ENTER. The cursor moves to the Reenter box.

NOTE: Passwords are not case sensitive; capital and lower case letters may be interchanged.

Dako Autostainer | User Guide 27

5. Type in the same password again and press ENTER. The High security level option becomes highlighted. If a different password is entered the window “TRY AGAIN:

Password must be identical” appears.

6. Using the mouse, select the desired security level. A mark appears in the raised diamond. Selecting the High Level option allows full access to the software features.

Select OK and the INITIALIZE screen appears.

NOTE: Each security level is displayed at the bottom of each screen with asterisks in parenthesis “()”. (See below for details).

The security levels are defined as follows:

High Level (**): Users have full access to all features of the software.

Middle Level (*): Users do not have access to the INITIALIZE button in the MAIN MENU

SCREEN. This button is replaced with a CHANGE PASSWORD button. Middle level users cannot set or change the parameters for Institution info, default slide settings, slide/reagent labels, Main

Grid and IHC printing formats, cleaning schedule, entering in new users and doctors, or adding new tissues.

Low Level: Users do not have access to the INITIALIZE button in the MAIN MENU screen. This button is replaced with a CHANGE PASSWORD button. Users with low level security are also restricted from the same features as users with middle level security and the following: no access to Reagent Tracking, the Edit List function, setting up Auto Program, creating/editing Protocol

Templates. is added to the current list of users and MAIN MENU screen appears.

Select the CANCEL button to escape without saving the entered information. The MAIN

MENU screen appears.

Editing User Information

INITIALIZE on the MAIN MENU screen. The INITIALIZE screen is displayed with the cursor in the Name box and the previously entered information listed.

3. Select the user name from the list and press ENTER. The PASSWORD AND SECURITY

LEVEL screen appears.

28

NOTE: The Users list can be displayed by positioning and clicking the mouse pointer in the Users box and pressing the DOWN arrow key on your keyboard until the desired user’s name is highlighted in the Users box. The scroll arrows in the drop down list can also be used.

4. The Password boxes are empty. Enter the new password information.

5. To change a user’s security status, move the mouse pointer over the desired security level and press the left mouse button. A mark will appear next the selected security level.

Select the OK button to save the changes.

Selecting the CANCEL button displays the INITIALIZE screen without saving the newly entered information.

Dako Autostainer | User Guide

Deleting User Information

INITIALIZE on the MAIN MENU screen. The INITIALIZE screen is displayed with the cursor in the Name box and the previously entered information listed.

3. Select the user name from the list and press ENTER. The PASSWORD AND SECURITY

LEVEL screen appears. The DELETE USER button located to the right of the entry boxes becomes available. user (by name) appears. the screen appears.

Select the NO button and the PASSWORDS AND SECURITY LEVEL screen is displayed without deleting the selected user. Select CANCEL to return to the INITIALIZE screen.

NOTE: A user cannot be deleted from the current user list if his/her name was used to sign in for the current session. The DELETE USER button is grayed out and not available.

Adding a New Doctor

INITIALIZE on the MAIN MENU screen. The INITIALIZE screen is displayed with the cursor in the Name box and the previously entered information listed.

3. Type in the Doctors name and press ENTER. A dialog box appears asking if you want to add the current doctor to the list. the name is added to the list.

Select the NO button. The INITIALIZE screen is displayed and the information is not saved.

5. To set the displayed doctor as the Default Doctor, move the mouse pointer over the

Default Doctor check box and press the left mouse button. A check mark is placed in the box and the doctor’s name is displayed in the SLIDE INFORMATION screen as the default doctor. This name also appears in the Doctor column in the PROGRAMMING

GRID.

To change the Default Doctor, select the name of the new default and select the Default

Doctor check box. A check is placed in the box. The initial doctor is no longer set as the default.

NOTE: To edit a doctor’s name, first delete the name, then enter the new information.

Dako Autostainer | User Guide 29

Deleting Doctor Information

INITIALIZE on the MAIN MENU screen. The INITIALIZE screen is displayed with the cursor in the Name box and the previously entered information listed.

3. Select a doctor from the current list of doctors. the

INITIALIZE screen appears.

Select the NO button and the INITIALIZE screen is displayed and the highlighted doctor is not removed from the list.

Entering Tissue Information

Adding a New Tissue

INITIALIZE on the MAIN MENU screen. The INITIALIZE screen is displayed with the cursor in the Name box and the previously entered information listed.

3. Type in the tissue and press ENTER. A dialog box appears asking if you want to add the current tissue to the list. the added to the list.

Select the NO button. The INITIALIZE screen is displayed and the information is not saved.

NOTE: To edit a tissue, first delete the tissue, then enter the new information.

Deleting a Tissue

INITIALIZE on the MAIN MENU screen. The INITIALIZE screen is displayed with the cursor in the Name box and the previously entered information listed.

3. Select a tissue from the current list of tissues. the

INITIALIZE screen appears.

Select the NO button and the INITIALIZE screen is displayed and the highlighted tissue is not removed from the list.

30 Dako Autostainer | User Guide

Entering Default Reagent Information

Assigning Default Reagent Dispense Volumes

2. Select a default reagent volume by clicking on a raised diamond next to the appropriate volume. A smaller black diamond appears next to the selected volume.

The MAIN MENU screen is displayed.

NOTE: The dispense volume selected on this screen is the default volume assigned to all reagents in any programmed staining run. The volume for all reagents as well as for specific reagents applied to all slides in a particular staining run can be changed in the

Protocol Template Design screen. In addition, the dispense volume can also be changed on an individual slide in the PROGRAM SLIDES screen. (See Programming Slides,

Section 9)

5. To change the Default Reagent Dispense volume, click on the raised diamond next to the new volume. A smaller black diamond appears next to the selected volume.

Assigning Reagent Drop Zone

The area(s) on the microscope slide where the reagent is dispensed is defined as the Drop

Zone(s). The default drop zone information is used when a new Protocol Template is created.

1. Move the mouse pointer over the top, middle or bottom portion of the microscope slide image and press the left mouse button. A yellow bar appears on the slide in the selected position. During the staining run, the selected Default Reagent Volume is dispensed on this zone of the slides for all steps of the protocol.

2. Multiple drop zones can be selected by clicking on multiple (2 or 3) areas while holding down the CONTROL (CTRL) key on the keyboard. Multiple zones are colored yellow.

During the staining run the selected Default Reagent Volume is dispensed on each of the selected zones.

Select the CANCEL button to escape without changing the Drop Zone. The MAIN MENU screen is displayed.

NOTE: The drop zones(s) are selected to correspond to the location of tissues on the slide.

Dako Autostainer | User Guide 31

Cleaning Schedule

This selection determines the number of slides the instrument will stain prior to prompting the user to start a cleaning cycle. Dako recommends setting the cleaning schedule to 150 slides. The

MAIN MENU screen always displays the total number of slides processed since the last cleaning cycle in red.

1. Move and click the mouse pointer on the Slides/Cleaning box and type in the number of slides to stain between maintenance cleanings. Select the OK button and the MAIN

MENU screen is displayed.

Select the CANCEL button to escape without saving. The MAIN MENU screen is displayed.

NOTE: This number can be changed by highlighting the entered number and typing the new number of slides. Select the OK button and the MAIN MENU screen is displayed.

NOTE: The maximum number of slides allowed per cleaning cycle is 200 slides. If a number greater than 200 is placed in this field, then you will be prompted to reduce your slide number to 200 or less. Once more than 200 slides have been run on the

Autostainer, the PROGRAM button will not be available.

Label Printing Initialization

Slide Labels

The SLIDE LABELS button is displayed on the INITIALIZE screen and it is used to access the

DESIGN SLIDE LABEL screen for configuring and adjusting slide label printing.

NOTE: Slide labels are printed from the print button on the PROGRAMMING GRID. Do not print labels while the Autostainer is operating.

1. Select the SLIDE LABELS button. The DESIGN SLIDE LABEL screen appears with six fields displayed representing the six lines of information to be printed on each slide label.

Select number of

characters per line.

32

NOTE: Each field offers 11 options: Slide ID, Case #, User, Primary 1, Primary 2,

Pretreatment, Today’s Date, Run Name, Slide #, Doctor and Tissue. The information printed

Dako Autostainer | User Guide

for each field is the data entered in the SLIDE INFORMATION screen and appearance of the

Programming Grid.

NOTE: If Slide ID, Doctor or Tissue is selected to print on the label, then these items must be selected from the OPTIONS button in the INITIALIZE screen. If these Program Grid items are not selected from the OPTIONS button in the INITIALIZE screen, then these fields will not be present in the SLIDE INFORMATION screen.

2. Move the mouse pointer over the down arrow to the right of the information box and press the left mouse button. A list box with the 11 options is displayed.

3. Select the desired option for the field. The list box disappears and the selected option is displayed in the field.

NOTE: Static information can be entered into the field. This information is printed on every label. For example, the institution name can be printed on every label. To enter static information, highlight the text in the desired field and type the text in the text box.

Press ENTER.

4. Select the number of characters for each line (10, 12, 14, or 16) by selecting a number from the LENGTH list. (Selecting 10 characters will display bold text.)

NOTE: The greater the number of characters, the smaller the text will appear on the labels.

Reagent Labels

The REAGENT LABELS button is displayed on the INITIALIZE screen and it is used to access the DESIGN REAGENT LABEL screen for configuring and adjusting reagent label printing.

NOTE: Reagent labels are printed from the print button in the PROGRAMMING GRID. Do not print labels while the Autostainer is in operation. the with six fields displayed representing the six lines of information to be printed on the reagent label.

NOTE: Each field offers 8 options: Reagent Name, Short Name, Lot, Exp. Date,

Incubation Time, Today’s Date, Reagent Name 1/2, and Reagent Name 2/2. The information printed for each field is the data entered from the EDIT REAGENT LIST screens for each reagent.

NOTE: Reagent Name 1/2 and Reagent Name 2/2 are used together on adjacent lines to split a long reagent name into 2 separate lines. Reagent Name 1/2 will contain the first half of the reagent name, and Reagent Name 2/2 will contain the second half of the reagent name. The length of each line is set up using the

LENGTH tool. (See below.)

2. Move the mouse pointer over the down arrow to the right of the information box and press the left mouse button. A list box with the 8 options is displayed.

Dako Autostainer | User Guide 33

3. Select the desired option for the field. The list box disappears and the selected option is displayed in the field.

NOTE: Static information can be entered into the field. This information is printed on every label. For example, the institution name can be printed on every label. To enter static information, highlight the text in the desired field and type the text in the text box. Press

ENTER.

4. Select the number of characters for each line (10, 12, 14, or 16) by selecting a number from the LENGTH list. (Selecting 10 characters will display bold text.)

NOTE: The greater the number of characters, the smaller the text will appear on the labels.

Label Adjustment

To check the position of the text on the slide or reagent label and to generate a test label, select the ADJUST LABEL button from the DESIGN SLIDE LABEL or DESIGN REAGENT LABEL screen. The ADJUST LABEL screen is displayed.

2. Adjust the printing on the label using the scroll bar on the right side of the displayed information field. The displayed image moves on the screen in relation to the image on the label.

3. Check the adjusted image position by selecting the TEST PRINT button. Continue to move the image until the text is located in the center of the label. the

DESIGN REAGENT LABEL screen is displayed. the displayed. the and the changes are not saved.

NOTE: The horizontal adjustment is automatically positioned by the Seymour Label

Printer. If the text does not begin along the left edge of the label (text is too far over to the right), the Seymour Label Printer may need to be reset.

To reset the Seymour Label Printer, turn off the printer. Hold down the green label feed button on top of the printer and turn the printer on again. Continue holding the button for 3 seconds. The labels advance forward and printer will print. Release the feed button. Press and release the feed button one more time to advance one label. Perform a TEST PRINT. The print will be lined up along the left slide of the label.

Communication Port

1. To select the communication port on the DESIGN SLIDE LABELS screen, move the mouse pointer over the down arrow to the right of the port box and press the left mouse button. Scroll through the list using the keyboard arrows or use the scroll bar. Select the communication port where the printer is attached and press the left mouse button or TAB.

34 Dako Autostainer | User Guide

NOTE: The communication port for the printer is identified using the Port box on the

DESIGN SLIDE LABEL screen. The number displayed in the box must match the computer serial port where the label printer is connected.

Select the CANCEL button to escape without saving. The INITIALIZE screen is displayed.

Printer Selection

1. To select the proper Seymour printer type on the DESIGN SLIDE LABEL screen, move the mouse pointer over the down arrow to the right of the Printer box and press the left mouse button. Scroll though the list using the keyboard arrows or the scroll bar. Select the printer model and press the left mouse button or TAB.

NOTE: If the incorrect printer is selected, then error messages will appear on the screen when slide or reagent labels are printed. To find the model number of your Seymour

Glass Label printer, check the cover of the printer for a number (ie.TPL2634). The last 4 digits represent the model number (i.e. 2642).

Select the CANCEL button to escape without saving. The INITIALIZE screen is displayed.

The Options Screen

The OPTIONS button is displayed on the INITIALIZE screen. This screen enables the user to select format options for the Programming Grid, IHC Report; and customer options for rinsing, Slide Layout

Maps, and blow steps.

Program Grid

This function enables the user to select format options for the PROGRAMMING GRID.

Dako Autostainer | User Guide 35

2. Select the items to be included on the Program Grid. Any or all of the listed options can be marked. Only checked items are displayed on the SLIDE INFO screen and the

Programming Grid.

NOTE: Selecting the Comment item creates a column for data entry on the Programming

Grid printout. This column is not displayed on the PROGRAMMING GRID.

NOTE: Slide ID, Block ID, Doctor, and Tissue items were selected to create this

Programming Grid example.

IHC Report

This function enables the user to select printing options for the IHC Report.

1. Select the IHC Report option by clicking on a raised diamond. Only one option can be selected.

NOTE: The One case/page option prints one case on each page of the report.

The One slide ID/page option prints each Slide ID on a separate page.

The No page break option prints all slides for a run in a continuous print out.

Select the CANCEL button to escape without saving. The INITIALIZE screen is displayed and the changes are not saved.

Select the CANCEL button to escape without saving. The MAIN MENU screen is displayed.

Custom Options

This function enables the user to select rinse and Slide Layout Map options.

1. Select any items to be made active by clicking the check box next to each option. If a check is present then the option is activated.

36 Dako Autostainer | User Guide

Delayed Start Water Rinse: For a delayed start, idle rinses may be performed with water or buffer. Select the Delayed Start water rinse option to use water. If this option is not checked then idle rinse will be with buffer. (Deionized water is not available with the “01” hardware version of the Dako Autostainer.)

NOTE: The Delayed Start Water Rinse options cannot be selected while the No Pre-Rinse option is selected.

Add Rinse to Protocol Steps: When creating a new protocol template, a buffer rinse can be added automatically, following each step. Select the Add Rinse to Protocol Steps option to automatically add buffer rinses. If this option is not checked, the rinse step must be manually selected.

Show only Programmed Slides: To display only the programmed slides on the Slide Layout

Maps, select the Show only Programmed Slides option. If this option is not checked, the Slide

Layout Maps shows all slide positions, displaying programmed and unprogrammed slides.

No Pre-Rinse: When creating a new protocol template, the default pre-rinse step may be removed. Select the No Pre-rinse option to remove the default pre-rinse. If this option is not checked, then in the Protocol Template there will be a default buffer rinse step.

NOTE: The No Pre-Rinse option cannot be selected while the Delayed Start Water Rinse option is selected.

Flex-Time Blow Option: When a program is running, the timing of blow steps can be adjusted by this feature. Check this option to minimize run time. The instrument will make minor adjustments to the incubation times of reagents that precede a blow step. Leave this option unchecked to fix incubation times. Be aware that fixed incubation times (option left unchecked) may increase the run time. the displayed and the changes are not saved. displayed.

Dako Autostainer | User Guide 37

This page intentionally blank.

38 Dako Autostainer | User Guide

Section 8 |

Entering Slide Information

Slides can be programmed from the PROGRAMMING GRID either through the Slides function for quick programming of total slide count, or through the SLIDE INFORMATION button for entering specific slide identification information. This section describes both options for programming slides.

Slides Function

Adding Slides

The Slides option expedites programming by reducing the amount of information required to program a staining run.

Slides function can be accessed from the PROGRAMMING GRID by moving the mouse pointer to Slides on the menu bar and pressing the left mouse button.

This selection displays the SLIDE COUNT window for entering the total number of slides for the current staining run. This function is used to bypass the SLIDE INFORMATION screen when programming a staining run.

2. Enter the desired number of slides in the text box or use the scroll bar to select the desired number of slides. the displayed and ready for programming.

Select the CANCEL button to escape without saving the changes. The PROGRAMMING

GRID is displayed.

NOTE: Once a run has been set up using the SLIDE COUNT window, slide, tissue or doctor information cannot be added to the run. Selecting the SLIDE INFO button displays the SLIDE COUNT window.

NOTE: Slide Identification, Case #, Block ID, Tissue or Doctor Name will not be associated/displayed in the Programming Grid when the Slides function is used.

Dako Autostainer | User Guide 39

Deleting a Slide

When using the Slides option to program a staining run, a slide can be deleted if there is at least one entirely blank (unprogrammed) row.

2. Adjust the number of slides and click OK.

NOTE: If all rows have a programmed tile the slide must be deleted from the

PROGRAMMING GRID (see Deleting a Specific Slide, later in this section.)

The Slide Information Button

The SLIDE INFO button enables the entry of specific information for every slide to be stained. Slide identification information is entered on the SLIDE INFORMATION screen. The data entry fields may include: Slide ID, Case #, Doctor, Block ID, Tissues, and # Slides. The specific combination of information fields is determined by an authorized user during initialization. (See Initializing the Dako

Autostainer, Section 7.)

A summary line indicates the number of slide identifications, cases and slides entered for the current staining run. This status line is updated as new slide information is added, providing a useful point of reference during programming.

NOTE: Slide ID, Doctor, Block ID, and Tissue fields were selected through the OPTIONS button on the INITIALIZE screen during software initialization. Case # and slides/case fields are required and are always present.

40

Summary Line

Dako Autostainer | User Guide

The buttons on the SLIDE INFORMATION screen and their functions are:

FINISH ENTRY

saves the entered information and returns to the Programming Grid.

DELETE

CANCEL

deletes selected programmed cases for the selected Slide Identification or

Case #. cancels any entries made and returns to the Programming Grid.

HELP

displays information to aid in understanding and utilizing the features of the

SLIDE INFORMATION screen.

Entering Slide Information Through the SLIDE INFORMATION Button

the displayed.

INFORMATION screen is displayed. The information to be entered on this screen is determined by a user, with high security level, during initialization (see Initializing the

Dako Autostainer, Section 6.) The minimum entry required is a Case # and the # slides/case.

NOTE: The following describes the steps required if all options are selected during initialization. Information entry is required in all activated fields except Doctor, Block ID, and/or Tissue.

NOTE: An entry is required to proceed from one field to the next. Each field requires at least one key stroke. Type the specified information or press the SPACE BAR before moving to the next field. The Slide/case box requires entry of the actual number of slides.

3. Type in the slide identification’s last name and press TAB or ENTER. Boxes for the first name and middle initial appear with the cursor in the First name box.

4. Type in the first name or press TAB or ENTER. The cursor moves to the middle initial box.

5. Type in the middle initial or press TAB or ENTER. The cursor moves to the Case # box.

NOTE: The software allows the entry of as many characters as can be displayed in each entry box. Once the box is filled an audible signal is made and no additional characters can be entered. A maximum of 33 characters (last name 28 characters, first name 5 characters and middle initial) will be displayed on the programming grid. The software shortens name entries to fit available space on the PROGRAMMING GRID.

NOTE: Steps 2-4 can be condensed by entering slide identification’s last name, (comma) first name, (comma), and middle initial. Press ENTER to move to the Case # box.

6. Type in the case # and press TAB or ENTER. The cursor moves to the Doctor box.

7. If a default doctor was selected during initialization the name appears in the Doctor box.

To maintain the defaulted doctor, press TAB or ENTER. The cursor moves to the Block

ID box.

To select an alternative doctor press the DOWN arrow key on the keyboard until the name appears highlighted in the Doctor box and press ENTER. If the doctor’s name is not listed it can be typed in the box followed by pressing the ENTER key. The cursor moves to the Block ID box.

8. Type in the Block ID information for the case and press TAB or ENTER. The cursor moves to the Tissue box.

9. Type in the tissue information for the slide identification and press TAB or ENTER. The cursor moves to the # Slides box.

Dako Autostainer | User Guide 41

To select a Tissue that was entered at the INITIALIZE screen, press the DOWN arrow key on the keyboard until the tissue appears highlighted in the Tissue box and press

ENTER. If the tissue is not listed, it can be typed in the box followed by pressing the

ENTER key. The cursor moves to the # Slides box.

10. Type in the number of slides for the case and press TAB or ENTER. The cursor returns to the Block ID box. Enter next block if necessary.

NOTE: A maximum number of 48 slides can be entered for each staining run. If the total number of slides programmed exceeds 48, a warning message displays the number of excess slides.

11. Press ENTER to return to the Case # box and enter additional cases.

12. Press ENTER to return to the Slide ID box and enter additional slide identifications.

13. Select the FINISH ENTRY button or press ENTER until the FINISH ENTRY button is highlighted to complete the slide identification entry process. The PROGRAMMING GRID appears.

Deleting Slide ID Information through the SLIDE INFORMATION

Button

There are three methods for deleting slide identification information: deleting the slide identification, case # or a specific slide. Deleting a slide identification enables the removal of a large group of slides associated with a specific ID. Deleting a case # removes only the slides associated with the specified case for the specified ID and allows other slides from the same ID with a different case # to remain on the grid. Deleting a slide removes a specific slide from the grid.

Deleting a Slide Identification

This method deletes all slides associated with the selected slide identification.

42 the

INFORMATION screen is displayed with the cursor in the Slide ID box. the

Slide ID box.

ENTER and the selected slide identification appears in the Slide ID box. The cursor moves to the Case # box. the identification. the from the system and the cursor is positioned in the Slide ID box.

Select the NO button to return to the SLIDE INFORMATION screen without deleting the selected Slide ID.

Dako Autostainer | User Guide

displayed.

Select the CANCEL button to undo any changes made to the slide identification record and the PROGRAMMING GRID is displayed.

Deleting a Case #

This method deletes all slides associated with the selected slide identification’s case.

INFORMATION screen is displayed.

2. If the cursor is in the Slide ID box, press the DOWN arrow key until the desired slide identification appears highlighted in the Slide ID box. Press ENTER and the selected slide identification appears in the Slide ID box. The cursor moves to the Case # box.

3. When the cursor is in the Case # box, press the DOWN arrow key on your keyboard until the desired case appears highlighted in the Case # box. The number of slides for this record is displayed in the # Slides/case box.

#. the deleted and the cursor is positioned in the Slide ID or Case # box.

Select the NO button to return to the SLIDE INFORMATION screen without deleting the selected Case #. the displayed.

Select the CANCEL button to undo any changes made to the slide identification record and the PROGRAMMING GRID is displayed.

Deleting a Block ID

This method deletes all Block IDs associated with the selected case number. the

INFORMATION screen is displayed.

2. Is the cursor is in the Slide ID box, press the DOWN arrow key until the desired slide identification appears highlighted in the Slide ID box. Press ENTER and the selected slide identification appears in the Slide ID box. The cursor moves to the Case # box.

3. When the cursor is in the Case # box, press the DOWN arrow key on your keyboard until the desired case appears highlighted in the Case # box. the Doctors box. The cursor moves to the Block ID box. the in the Block ID box.

The number of slides for this record is displayed in the # Slides (per case) box. the deleted and the cursor is positioned in the Slide ID or Case # box.

Select the NO button to return to the SLIDE INFORMATION screen without deleting the selected Block ID.

Dako Autostainer | User Guide 43

displayed.

Select the CANCEL button to undo any changes made to the slide identification record and the PROGRAMMING GRID is displayed.

Deleting a Specific Slide

This method deletes one slide at a time.

44

A dialog box asks if you want to delete the selected slide.

Slide # the subsequent rows in the programming grid shift up one line.

Select the NO button to return to the PROGRAMMING GRID.

Slide Rearrangement

Slides can be rearranged into a user defined sequence. This feature is used to move a slide to a different position on the grid.

NOTE: Adding slides to the grid is done using the Slides function in the menu bar or the SLIDE INFO screen. Added slides are inserted at the bottom of the Programming Grid as blank rows of tiles.

Rearranging slide positions on the Programming Grid

1. Right click on the Slide # located on the PROGRAMMING GRID for the slide to be moved.

NOTE: When the mouse pointer is positioned on the Slide # and a right click is made the slide number, Slide ID, Case #, and Block ID are displayed in a highlighted box. The reagent tiles for the selected slide disappear from the grid.

2. Move the mouse pointer to the position where the slide is to be inserted and press the left mouse button. The slide information is inserted in the new position.

NOTE: As the cursor is moved the highlighted box moves.

Dako Autostainer | User Guide

Section 9 |

Designing a Protocol

Protocol templates are defined as a sequential series of steps required to stain a microscope slide.

Each template is built by selecting the individual protocol steps listed under the Protocol Elements column in the correct order to complete a staining protocol. Protocol templates can be saved for future use.

The PROTOCOL TEMPLATE DESIGN screen enables you to create a staining protocol template. An unlimited number of protocol templates can be saved for future use.

The buttons on the PROTOCOL TEMPLATE DESIGN screen are:

DELETE

STEP

deletes a highlighted step from the protocol template displayed in the

Protocol Outline column. displays a blank Protocol Outline to build a new template.

NEW

TEMPLATE

GET

TEMPLATE

SAVE

USE

TEMPLATE

CANCEL

HELP

retrieves an existing template. saves a newly created or modified protocol template. installs the template displayed in the Protocol Outline column and returns to the Programming Grid. cancels any entries made and returns to the previous Programming Grid .

REAGENT

VOLUME

displays information to aid in understanding and utilizing the features of the

Protocol Template Design screen. assigns a reagent dispense volume to a highlighted step (or to all steps if a specific step has not been selected) in the Protocol Outline. The volume for

Rinse steps cannot be changed.

Dako Autostainer | User Guide 45

Protocol Elements

The Protocol Elements are categories of reagents used to build the Protocol Outline in the

PROTOCOL TEMPLATE DESIGN screen. Reagents can be added, deleted and edited by choosing

Protocol Element categories from the Edit Lists menu item on the PROGRAMMING GRID.

Standard Protocol Elements include:

• Endogenous

Block

• Primary

• Pretreatment

• Secondary

• Tertiary

Polymer

• Substrate

• Auxiliary

Other Protocol Elements may be defined as steps in the template:

Rinse Buffer

A buffer rinse step consists of a buffer wash and air blow cycle and should be programmed between each protocol template step. Each buffer rinse is displayed as a blue droplet in the programming grid with a “droplet” icon in each programmed row. The buffer icon is printed out on the Main Grid printout as a filled in droplet.

Rinse Water

A water rinse step consists of a water wash and air blow cycle and is an option that can be selected in the Protocol Template. Each water rinse is displayed as a white droplet in the programming grid with a “droplet” icon in each programmed row. The water icon is printed out on the Main Grid printout as a white droplet.

NOTE: The rinse water option is not available with the “01” hardware configuration of the Dako

Autostainer.

Standard Rinse Replacement

The standard rinse cycle can be replaced by a blow only step (reagent is blown off the slide but no buffer is applied) by clicking on the droplet icon in the programming grid using the right mouse button. The droplet icon is replaced by a “blow” icon. The blow icon is printed out on the

Main Grid printout as a wind symbol.

46 Dako Autostainer | User Guide

Substrate-Batch

This step splits the staining run into two distinct operations. The first includes all steps prior to the application of the substrate and the second batches all steps starting with the substrate application. This feature allows you to use unstable substrates that need to be prepared immediately prior to application. The Autostainer stops after the steps in the first operation (non-

Substrate protocol steps) are completed. An audible signal indicates when the substrate should be loaded and the screen displays where it should be placed (typically in position D8 of reagent rack

#1).

NOTE: The Substrate-Batch step can only be used once in a Protocol Template.

Switch (*Swtch)

This step programs the switching of waste disposal from one container to another for the separation of hazardous from non-hazardous waste. The Switch step is displayed in the

Programming Grid as a “skull and cross bones” icon in each programmed row. The Switch step can be used multiple times in any Protocol Template. When programming successive switch steps the “skull and cross bones” icon will be alternated with a “flower” icon to indicate switching from hazardous to non-hazardous waste. The switch icons are printed out on the Main Grid printout as

“skull and cross bones” or flower symbols.

NOTE: The switch step must be preceded by a rinse step in the protocol template.

NOTE: The first switch step in any protocol template will be a hazardous waste switch.

Creating and Editing a Protocol Template

NOTE: Low level security does not allow access to creating or editing Protocol Templates. the

PROTOCOL TEMPLATE DESIGN screen appears with the default protocol template displayed in the Protocol Outline column. default first step (Rinse Buffer) is displayed.

NOTE: If No Pre-Rinse is selected at the OPTIONS screen, then the first Rinse Buffer is not displayed. (See Initializing the Dako Autostainer, section 7)

3. Build the new protocol template from the Protocol Elements list. To select a reagent move the mouse pointer over the protocol element and press the left mouse button. The item is copied to the Protocol Outline column and sequentially numbered.

NOTE: If Add Rinse to Protocol Steps is selected at the OPTIONS screen (see

Initializing the Autostainer, section 7), then rinses will be automatically added after each step, except following a Substrate step.

NOTE: Buffer/Water rinses and hazardous/non-hazardous switches are represented in the protocol outline with the appropriate icon for easy differentiation between the different selections.

4. To delete a step, select it in the Protocol Outline box. The name is highlighted. Select the

DELETE STEP button. The step is removed.

Dako Autostainer | User Guide 47

5. Additional steps can be inserted in the Protocol Outline by highlighting a step in this column and then clicking on the desired step from the Protocol Elements. The selected step will be inserted above the highlighted area in the Protocol Outline.

NOTE: The Dako Autostainer can accommodate protocol templates containing as many as 35 steps (including rinse steps between reagent incubations).

Saving a Protocol Template

NOTE: Low level security does not allow access to saving a Protocol Template.

Protocol templates can be used with or without saving. A saved protocol template may be accessed for designing a future run.

1. To save a newly created template select the SAVE button on the PROTOCOL

TEMPLATE DESIGN screen. The SAVE TEMPLATE window appears with the cursor in the file name box.

2. To save an edited template select the USE TEMPLATE button from the PROTOCOL

TEMPLATE DESIGN screen. The TEMPLATE CHANGED dialog box appears. Select the YES button to save and the SAVE TEMPLATE window appears with the cursor in the file name box. Select the NO button and the PROGRAMMING GRID appears.

3. Type the name of the new Protocol Template in the File Name box. The name cannot exceed 8 alpha-numeric characters; symbols and spaces cannot be used.

NOTE: If the new protocol will be used frequently, click on the Default Template button. A check mark appears. The default protocol template automatically appears as the header on the PROGRAMMING GRID and the new template is saved.

NOTE: A default template is a template that appears each time a new staining run is programmed. the programmed protocol template.

48 Dako Autostainer | User Guide

NOTE: When assigning a new template to an existing programmed staining run, a dialog box appears if the new template lacks protocol elements present in the previous template.

Select the YES button and the columns of missing protocol elements disappear from the grid while newly programmed columns appear. Select the NO button and the

PROGRAMMING GRID appears unchanged.

Deleting a Protocol Template

NOTE: Low level security does not allow access to deleting Protocol Templates.

1. Select the PROTOCOL TEMPLATE button on the PROGRAMMING GRID. The

PROTOCOL TEMPLATE DESIGN screen appears.

2. Select the GET TEMPLATE button. The GET TEMPLATE screen is displayed with the cursor flashing in the File Name box.

3. Select the template to be deleted by either scrolling and clicking on the file name or using the DOWN arrow key. The name appears in the file name box. the selected protocol template.

NOTE: If the Default Template is deleted, a new template must be selected from the

GET TEMPLATE window and saved as the default template from the SAVE TEMPLATE window. Deleting the Default Template will result in an error message when attempting to program a new run.

TEMPLATE screen is displayed.

Select the NO button to return to the GET TEMPLATE screen without deleting.

Dako Autostainer | User Guide 49

Using a New Protocol Template Without Saving It

1. After a Protocol Template has been created or modified, click on the USE TEMPLATE button in the PROTOCOL TEMPLATE DESIGN screen. A dialog box asking “Save on disk now?” is displayed. the template in the header row.

NOTE: If a staining run had already been programmed and the new protocol template omits previously programmed steps, a warning message is displayed stating that programmed steps will be discarded if the new protocol template is applied.

Using a Saved Protocol Template

the

PROTOCOL TEMPLATE DESIGN screen appears with the default protocol template displayed in the Protocol Outline column. the cursor in the File Name box.

3. Select the template by clicking on it. The name appears in the file name box. appears with the selected protocol template displayed in the Protocol Outline column.

5. Click on the USE TEMPLATE button. The PROGRAMMING GRID appears with the selected protocol template displayed in the grid header.

Using a Saved Protocol Template for a Specific Auto Program

NOTE: Auto Programming is a feature for making slide programming more efficient. (See

Programming Slides section 10) the

PROTOCOL TEMPLATE DESIGN screen appears with the default protocol template displayed in the Protocol Outline column. cursor in the File Name box.

4. Select the desired Auto Program by clicking on it. The name appears in the file name box.

50 Dako Autostainer | User Guide

appears with the selected Auto Program protocol template displayed in the Protocol

Outline column.

6. Click on the USE TEMPLATE button. The PROGRAMMING GRID appears with the selected protocol template displayed in the grid header.

Selecting and Editing the Reagent Dispense Volume

NOTE: Low security level does not allow access to changing the reagent volumes for a Protocol

Template. Low level security users can change the volume for all slides using the All Slides control on the SLIDE LAYOUT:PROGRAM SLIDES screen. (See Programming Slides, section 10)

Selecting and Editing the Reagent Dispense Volume for All Steps in a

Protocol

the

TEMPLATE DESIGN screen. The PROTOCOL STANDARD VOLUME dialog box appears.

2. Select the desired reagent dispense volume by clicking on the raised diamond next to the appropriate volume. The diamond becomes recessed and a smaller black diamond appears. The selected volume is applied to all steps in the Protocol Template. the asking “Save on Disk now”. Select the YES button and the SAVE TEMPLATE dialog box is displayed. Enter a template name and select OK. Template names must be 8 alphanumeric characters or less; no symbols or spaces can be used. Select the NO button to use the template without saving the change. The newly selected dispense volume is displayed in the grid header of the PROGRAMMING GRID.

Dako Autostainer | User Guide 51

Selecting and Editing the Reagent Dispense Volume for a Specific

Step in a Protocol

1. Select a Protocol Template Step in the Protocol Outline column to assign a new reagent dispense volume. The selected step will be highlighted. the

TEMPLATE DESIGN screen. A dialog box with the dispense volume choices for the selected reagent step appears.

3. Select the diamond next to the desired reagent dispense volume by clicking on the raised diamond next to the appropriate volume. The diamond becomes recessed and a smaller black diamond appears. The selected volume is applied to the specific step in the

Protocol Template. reagent dispense volume next to the reagent step in the Protocol Outline. Select the USE

TEMPLATE button. The TEMPLATE CHANGED dialog box is displayed asking “Save on

Disk now”. Select the YES button and the SAVE TEMPLATE dialog box is displayed.

Enter a template name and select OK. Template names must be 8 alpha-numeric characters or less; no symbols or spaces can be used. Select the NO button to use the template without saving the change. The newly selected dispense volume is displayed in the grid header of the PROGRAMMING GRID.

52 Dako Autostainer | User Guide

Section 10 |

Programming Slides

The Autostainer can process a maximum of 48 microscope slides in a single staining run. Each slide can be stained using an independently designed protocol. After the slide information is entered and protocol template is selected, specific reagents are assigned to each slide. This is accomplished from the PROGRAMMING GRID. The following section describes how to program the specific reagents for a staining run.

Defining Slide Specific Reagents

Defining Detection Reagents (Secondary, Tertiary, Substrate, etc.)

1. Enter slide information and select a protocol template as described previously. (See sections 8 and 9). The PROGRAMMING GRID is displayed listing the slides identified during Slide Count or Slide Information entry. The first programmable field is highlighted

(blinking) and a list of available reagents for the step is displayed.

Each row represents one slide and its information.

List of reagents available for the highlighted step.

Dako Autostainer | User Guide 53

2. Select the desired reagent from the list using the UP and DOWN arrow keys on the keyboard, the mouse pointer, the reagent list scroll bar or by typing the name of the reagent. A dialog box prompts “Assign to the following unprogrammed slides?”.

Select YES to assign the selected reagent to the remaining slides in the grid and the next programmable step is highlighted.

Select NO and the next programmable step is highlighted without assigning the reagent to the remaining slides. the programmed. The SLIDE LAYOUT MAP:PROGRAM SLIDES screen is displayed. selecting there is a rinse missing between two reagents, a message will ask if you want to continue.

Select YES to continue on the SLIDE LAYOUT MAP:PROGRAM SLIDES screen.

Select NO to return to the PROGRAMMING GRID and the missing rinse will appear on the screen.

Defining Primary Antibodies

Each primary antibody reagent is programmed individually and will appear in the PROGRAMMING

GRID with its programmed pretreatment. (see Reagent Management, section 11) Primary antibodies and/or pretreatments can be manual steps and are noted on the PROGRAMMING

GRID and printouts. If a reagent is a manual step, then that step is preformed before the slides are loaded on the Autostainer (i.e. overnight incubations, Target Retrieval, etc.)

NOTE: A primary antibody’s pretreatment will automatically appear with the primary antibody, only if the protocol template contains a pretreatment protocol element (see Designing a Protocol, section 9).

1. Enter slide information and select a protocol template as described previously. (See sections 8 and 9). The PROGRAMMING GRID is displayed listing the slides identified during Slide Count or Slide Information entry. Add any detection reagents as described above. When the first programmable Primary Antibody is highlighted (blinking), a list of available Primary Antibodies for the step is displayed.

2. Select the desired antibody from the list using the UP and DOWN arrow keys on the keyboard, the mouse pointer, the reagent list scroll bar or typing the name of the antibody and press ENTER. The selected antibody and its pretreatment will appear on the

PROGRAMMING GRID.

NOTE: Pretreatments can be made a manual step (for more information on manual steps, see Reagent Management, section 11) by positioning the mouse pointer over a pretreatment, and pressing the right mouse button. To change the pretreatment back to a step preformed on the Autostainer, position the mouse pointer over the pretreatment to be changed and press the left mouse button. The list of available pretreatments is displayed. Select the desired pretreatment from the list.

3. Negative controls are listed at the end of the primary antibody list and can be accessed on the keyboard by pressing the SHIFT key and typing “_NC”. For example, for selecting the negative control reagent for Vimentin, press the SHIFT key and type “_NCVimentin.

Once the correct negative control is selected, press ENTER and the selected negative reagent and pretreatment will appear on the PROGRAMMING GRID. selecting there is a rinse missing between two reagents, a message will ask if you want to continue.

Select YES to continue on the SLIDE LAYOUT MAP:PROGRAM SLIDES screen.

Select NO to return to the PROGRAMMING GRID and the missing rinse will appear on the screen.

54 Dako Autostainer | User Guide

Assigning Positive and Negative Control Reagents

Positive and negative control reagents can be automatically added for each primary antibody that is added to the Programming Grid. The controls will be added as new slides and will contain the matching detection reagents of the test specimen.

1. Enter Slide Information and select a protocol template. A minimum of one slide must be programmed through SLIDE COUNT or the SLIDE INFORMATION screen. Press OK or the FINISH ENTRY button. The PROGRAMMING GRID is displayed.

2. Fill in the PROGRAMMING GRID columns with specific reagents for one slide. After the primary antibody is applied, NEG.CTL and POS.CTL buttons will appear in the menu bar.

Continue entering the detection reagents from that slide.

3. To insert a matching positive control, click on the POS.CTL button. One slide will be added to the PROGRAMMING GRID with matching reagents to the last programmed slide. In the slide number column, PC will be displayed signifying that this is a positive control reagent.

4. To insert a negative control, click on the NEG.CTL button. One slide, with the negative control reagent, will be added to the PROGRAMMING GRID with matching detection reagents to the last programmed slide (i.e. “NCVimentin”.) In the Slide # column, “NC” will be displayed signifying that this a negative control reagent.

NOTE: The NEG.CTL tool always searches for negative control reagents from the primary antibody list designated “_NC” (the “matching” negative control). If there are no negative controls of this designation, or if the matching negative control is missing from the reagent list, then an error will appear stating that there is not matching negative control reagent. To add a matching negative control, see Reagent Management, section

11.

NOTE: After the NEG.CTL or POS.CTL feature is used, the button will disappear until the next primary is programmed.

NOTE: If you do not wish to use this feature noted above, then continue programming as usual. This feature is not active when using Auto Program.

Defining Specific Rinses

Rinses are automatically added following every programmed reagent (except for Protein Block; a blow step is added following a Protein Block) when assigned as described above. The type of rinse is selected in the Protocol Template (see Designing a Protocol, section 9).

The type of rinse can also be changed from the PROGRAMMING GRID.

1. In a rinse column, position the mouse pointer over an activated tile and click the left mouse button. Each time the button is pressed, this will change the rinse to a different selection: rinse buffer, rinse (DI) water, blow, or none (for details of each option, see

Designing a Protocol, section 9).

Dako Autostainer | User Guide 55

2. In a rinse column, position the mouse pointer over a tile with a (blue) rinse buffer icon and click the right mouse button. This will change the buffer rinse to a blow step. To change the blow step back to a buffer rinse, position the mouse pointer over the blow icon and left click the mouse button.

3. In a rinse column, position the mouse pointer over a tile with a (blue) rinse buffer icon and hold down the left mouse button and click the right mouse button. Release both buttons.

This will change the buffer rinse to a DI water rinse step. To change the water rinse back to a buffer rinse, position the mouse pointer of the water rinse icon and left-click the mouse button.

NOTE: Rinses that precedes a switch (see Designing a Protocol, section 9) cannot be changed from the PROGRAMMING GRID. The rinse at the end of a run cannot be changed into a blow step.

Auto Programming

Auto Programming is a feature to make slide programming more efficient. This function is used to save all, or any portion, of a protocol to be used in future programming.

Creating and Saving an Auto Program

NOTE: Low level security does not allow access to creating and saving Auto Programs.

1. Enter Slide Information and select a protocol template. A minimum of one slide must be programmed through SLIDE COUNT or the SLIDE INFORMATION screen. Press OK or the FINISH ENTRY button. The PROGRAMMING GRID is displayed.

2. Fill in the programming grid columns with specific reagents, except primary antibody for one of the programmed slides.

NOTE: This will create an Auto Program for the detection reagents only. To create an

Auto Program for a protocol that includes the primary antibody, fill in the programming grid columns with the specific primary antibody and detection reagents.

56 Dako Autostainer | User Guide

5. Select one or more rows of programmed tiles by clicking on the first row. Move the mouse pointer over the desired rows while holding down the left mouse button. Each selected row is high-lighted. Release the mouse button and the Save Auto Program window is displayed.

NOTE: An Auto Program can consist of one or more rows of programming.

6. Enter a file name for the saved protocol. The Auto Program file name must be 8 alphanumeric characters or less; symbols and spaces cannot be used. Select the OK button.

The PROGRAMMING GRID appears.

Setup from the menu bar. Setup changes to Auto and exits the Auto program function.

NOTE: To use a saved “Auto program,” the current template being programmed must contain the same protocol elements listed in the same order as the saved Auto Program.

To find the correct protocol template, use the Auto file function. (See Using a Saved

Protocol Template for a specific Auto Program, earlier in this section.) The dispense volumes do not have to be the same.

Deleting an Auto Programming Item

3. Select any activated tile in the PROGRAMMING GRID to display the SAVE AUTO

PROGRAM window.

4. Use the mouse to scroll through the list of available Auto Programs. Select the Auto

Program to be deleted by clicking on it with the mouse.

Dako Autostainer | User Guide 57

5. Click on the DELETE PROGRAM button.

Select the YES button to delete the Auto Program and return to the SAVE AUTO

PROGRAM window.

Select the NO button to cancel changes and return to the SAVE AUTO PROGRAM window.

Using an Auto Programming Item

3. Select a tile in the row where the Auto program is to be inserted. A list displays the saved

Auto Program files that can be assigned for this slide.

4. Select the desired Auto program file name. The Auto program is inserted where indicated in step 3 above.

5. If the Auto Program does not contain a primary antibody a message box appears prompting: “Assign to all following slides”.

Click YES to assign this Auto Program to all unprogrammed slides.

Click NO to assign Auto Program to only this slide.

NOTE: (Skip) is the first item in the Auto Program list. To leave a slide unprogrammed, select (Skip) from the Auto Program list. there is a rinse missing between two reagents, a message will ask if you want to continue.

Select YES to continue on the SLIDE LAYOUT MAP:PROGRAM SLIDES screen.

Select NO to return to the PROGRAMMING GRID and the missing rinse will appear on the screen.

Getting Help with Auto Programming

Auto from the menu bar in the PROGRAMMING GRIDPROGRAM STAINING

RUN screen.

Copy/Paste

The Copy feature copies reagents from one slide to another during slide programming.

Copy/Paste Individual Reagents

Copy from the menu bar on the PROGRAMMING GRID. The word Copy changes to Select.

2. Move the mouse pointer over the tiles to be copied. As the pointer is moved, hold the left mouse button down, multiple items may be selected and highlighted. Entire rows and/or parts of rows and columns can be copies at a time.

NOTE: When copying individual reagents, be sure to select the rinse following each step.

3. Release the left mouse button after all the desired items are highlighted. The word Select changes to Paste.

58 Dako Autostainer | User Guide

4. Move the mouse pointer to the slide(s) where the items are to be pasted and press the left mouse button. The copied reagents are inserted for the selected slides.

NOTE: Copied sections can only be pasted under the same heading and cannot be pasted over areas that are currently highlighted.

Copy/Paste Entire Rows

Copy from the menu bar on the PROGRAMMING GRID. The word Copy changes to Select.

2. Move the mouse pointer over the slide number to be copied. As the pointer is moved, hold the left mouse button down, multiple rows may be selected and highlighted.

3. Release the left mouse button after all the desired rows are highlighted. The word Select changes to Paste.

4. Move the mouse pointer to the slide(s) where the items are to be pasted and press the left mouse button. The copied rows of reagent are inserted for the selected slides.

NOTE: Copied rows cannot be pasted over areas that are currently highlighted.

Editing a Protocol Step

The Autostainer software provides multiple methods for editing the reagents in a defined staining run from the PROGRAMMING GRID.

Changing a reagent for a step:

1. Move the mouse pointer over the reagent step to be changed and press the left mouse button. The list of available reagents is displayed.

2. Select the desired reagent from the list. The new reagent is displayed for the step.

Dako Autostainer | User Guide 59

Replacing a reagent with skip step (no reagent added to slide):

1. Move the mouse pointer over the reagent step to be changed and press the left mouse button. The list of available reagents is displayed.

2. Select (none). (none) is displayed in the tile.

Editing a reagent in a staining step for one run only:

1. To edit a reagent in a staining step or to use a reagent not listed in the reagent list, click on the step and select the (edit slide) item from the reagent list. The EDIT INDIVIDUAL

SLIDE REAGENT box displays the information associated with the reagent.

NOTE: Lot and date cannot be edited here and are grayed out. Only the reagent long name, short name, compatibility, and incubation time can be edited.

2. Add a new reagent by typing in the appropriate information. Select the OK button. The new reagent is added for the current staining run only and will not be added to the available list of reagents for this step. A dialog box appears asking if the new reagent is to be assigned to all following unprogrammed fields of the reagent column.

NOTE: If there are no remaining unprogrammed fields of the reagent column, this dialog box will not appear.

NOTE: If the newly added reagent is a primary antibody, the dialog box “Assign to all following unprogrammed slides” does not appear. the are filled with the edited reagent.

Select the NO button and the PROGRAMMING GRID appears with only one slide edited.

NOTE: Do not use symbols when entering reagent information. Doing so may cause errors during the run.

60 Dako Autostainer | User Guide

Printing Options

The Autostainer software allows the user to print out various reports including information on each program, run, and reagents stored in the Autostainer. Slide labels with program related information and reagent labels with information stored in the software can be created using the software.

Slide Labels

screen appears.

3. To print all slide labels for the program, select the ALL LABELS button. All slide labels for the current program will be printed. The format of each slide label is determined by the SLIDE LABEL DESIGN screen that is accessed from the INITIALIZE screen. (See

Initializing the Dako Autostainer, section 6)

To print a range of labels, type the range of labels with a hyphen between the starting and ending numbers in the range (i.e. 5-12). To print individual labels, type the label number with commas between each number (i.e. 1,14,22).

These two formats can be used together to print a combination of a range of labels with individual labels (i.e. 5-12,14,22).

NOTE: When entering individual or a range of labels, do not use a space between the numbers.

Reagent Labels

Printing reagent labels from a programmed run.

1. Program or retrieve a previously saved run on the PROGRAMMING GRID, select the

PRINT button. The WHICH REPORT? screen appears. displaying all reagents from the programmed run on the PROGRAMMING GRID.

Dako Autostainer | User Guide 61

3. To print all reagent labels, select the PRINT ALL button.

To print only selected reagents, select desired reagents by positioning the mouse pointer over a reagent and clicking the left mouse button. To select more than one reagent, repeat this step.

To de-select a selected reagent, position the mouse pointer over the highlighted reagent and click with the left mouse button. The reagent will become un-highlighted, or deselected.

Once all desired reagents are highlighted, select the OK button.

Printing reagent labels not from a program.

the screen appears. the appears displaying all reagents available from the reagent list.

To print all reagent labels, select the PRINT ALL button.

To print only selected reagents, select desired reagents by positioning the mouse pointer over a reagent and clicking the left mouse button. To select more than one

reagent, repeat this step.

To de-select a selected reagent, position the mouse pointer over the highlighted reagent and click with the left mouse button. The reagent will become un-highlighted, or deselected.

4. Once all desired reagents are highlighted, select the OK button.

The Main Grid Report

the appears.

The Program Grid Report is a printout of the Programming Grid. All slide information, protocol headings, and specific reagents are displayed in this report. All rinses and switch(es) are represented by corresponding icons for easy identification of each step.

Each Program Grid Report also contains institution information (institution name, address, and department), user, date printed, number of slide ID’s, number of case #’s, and the

Autostainer serial number.

62 Dako Autostainer | User Guide

Immunohistochemical Report (IHC Report)

the appears.

The IHC Report can also be printed from the RUN LOG screen after the completion of a run.

3. After a run is completed, a PRINT IHC button appears at the bottom of the screen.

4. Select the PRINT IHC button.

The IHC Report format is selected during Autostainer initialization. This format can be changed when necessary by selecting the OPTIONS button from the INITIALIZE screen. (see Initializing the

Dako Autostainer, Section 6)

The IHC Report can be configured to print in one of the following formats:

• The One case/page option prints one case on each page of the report.

• The One slide ID/page option prints each Slide ID on a separate page.

• The No page break option prints all slides for a run in a continuous print out.

The IHC Report groups all slides by case number and prints the following information for every slide:

#

• primary antibody information (Full name, Species reactivity, Catalog #, Lot # (if entered),

Expiration date, and Incubation time)

• full name and incubation time for all reagents used

Run Log Printout

During a staining run a Run Log is created containing each step performed during the run. The run log can be printed at the end of the run after saving the Run Log (see Completing a Run, section 14). The Run Log can also be printed from the Programming Grid: the appears.

Dako Autostainer | User Guide 63

Viewing Programmed Slides

After the completion of reagent programming, the first SLIDE LAYOUT MAP is displayed. The

SLIDE LAYOUT MAP screen is a representation of all 48 slides as they would be viewed facing the front of the Autostainer. The information presented for each slide includes the slide number, the primary antibody, the case number and reagent dispense volume. The dispense volume and drop zone(s) for each slide can be changed on this screen. The location for reagent dispense is displayed by a yellow highlight.

NOTE: If the Show Only Programmed Slide option is selected at the Options screen, then only the sides that are programmed will be displayed on this screen. (See Initializing the Dako Autostainer, section 7)

The buttons and their functions are:

NEXT

BACK TO

displays the RUN TIME CALCULATION window to calculate the optimal run time for the current staining run. returns to the Programming Grid.

GRID

PRINT

generates a printout of the slide layout map.

HELP

displays information to aid in understanding and utilizing the features of the

SLIDE LAYOUT MAP screen.

64 Dako Autostainer | User Guide

Each cell on the screen represents a microscope slide. Cells contain the position number, case number and primary antibody short name. If case number and primary antibody name have not been assigned, the cells contain only the position number.

Slide Number

Position

Abbreviated Primary

Antibody Name

(Short name)

Primary Antibody Dispense

Volume per highlighted drop zone

P or N signifies Positive or

Negative control

Case Number (if text is too long it wraps to the next line)

Drop Zone

(yellow highlight)

Slide Number

Position

Primary Antibody

Dispense Volume

N/P Notation

is located in the upper left corner and identifies the slide rack position. is located in the upper middle section and displays the volume of primary antibody to be dispensed on each highlighted section of every slide. is located in the upper right corner and displays a P or N if this slide is programmed as a positive or negative control with the

Pos.Ctl or Neg.Ctl tool. (See Programming Slides, section 10) is located at the bottom and displays the primary antibody name.

Abbreviated Primary

Antibody Name

(Short Name)

Case Number

is located in the middle and lists the Case # assigned to the slide.

Yellow Highlight

(Drop Zone)

indicates the reagent dispense location. This bar appears on all programmed slides.

Assigning Reagent Dispense Locations

The non-frosted area of each slide is equally divided into three horizontal sections or dispense locations. Each slide can have one to three dispense locations. The maximum dispense volume of a slide cannot exceed 800

µL (controlled by the software). The default dispense location is programmed during initialization. Any modifications to the dispense volume or dispense location of a slide in the SLIDE LAYOUT MAP:PROGRAM SLIDES screen will change the text color of that slide from black to red.

1. Selecting a single dispense location:

Select a new reagent dispense location on each slide in the SLIDE LAYOUT MAP:

PROGRAM SLIDES screen by clicking on the desired dispense location using the mouse pointer. The yellow highlight appears in the selected location on each programmed slide.

Dako Autostainer | User Guide 65

2. Selecting multiple dispense locations:

To assign multiple dispense locations on a slide, press the CTRL key while clicking on the desired dispense locations with the mouse pointer. Multiple yellow highlights appear on the slide identifying the selected dispense locations.

NOTE: Each of the selected dispense locations will receive the volume of primary antibody programmed. The maximum total dispense volume per slide is 800 µL.

Assigning Reagent Dispense Volumes

The default reagent dispense volume will be applied for the primary antibody unless the dispense volume is changed.

1. Move the mouse pointer over the dispense volume in the SLIDE LAYOUT

MAP:PROGRAM SLIDES screen and press the left mouse button. A reagent dialog box displays the dispense volume choices.

2. Select the dispense volume by clicking in the diamond of the desired dispense volume. A smaller black diamond appears. the and the new reagent dispense volume and all slide information appear in red letters on the slide.

Select CANCEL and the reagent volume is not changed.

NOTE: If the primary antibody volume is changed in the SLIDE LAYOUT

MAP:PROGRAM SLIDES screen, all other reagent dispense volumes for the specific slide are adjusted in proportion. For example, the default volume is 200 µL of reagent for each step. If the primary antibody dispense volume is changed to 100 µL, the dispense volume of all other reagents will also be reduced to 100 µL.

66 Dako Autostainer | User Guide

ALL SLIDES Tool

At the upper left corner of the SLIDE LAYOUT MAP:PROGRAM SLIDES screen is an ALL SLIDES tool. This tool allows users to change the dispense pattern and volume for all slides with one click.

1. Select a new reagent dispense location on the slide in the ALL SLIDES tool by clicking on the desired dispense location using the mouse pointer. The yellow highlight appears in the selected location on each programmed slide.

To assign multiple dispense locations on the ALL SLIDES tool, press (and hold) the

CTRL key while clicking on the desired dispense locations with the mouse pointer.

Multiple yellow highlights appear on the slide identifying the selected dispense locations.

2. Move the mouse pointer over the dispense volume in the ALL SLIDES tool and press the left mouse button. A reagent dialog box displays the dispense volume choices.

3. Select the dispense volume by clicking in the diamond of the desired dispense volume. A smaller black diamond appears. and the new reagent dispense volume and all slide information appear in red letters on the slide.

Select CANCEL and the reagent volume is not changed.

NOTE: If the reagent dispense volume is changed for all slides, the heading of the

Programming Grid will be changed to reflect that all the volumes for this program have been changed to the new volume.

Printing a Slide Layout Map

The Slide Layout Map is useful for ensuring that slides are loaded in the proper location on the

Autostainer and provides a record of reagent dispense volume and dispense location.

1. Select the NEXT button on the PROGRAMMING GRID. The SLIDE LAYOUT

MAP:PROGRAM SLIDES screen appears.

NOTE: The Slide Layout Map can also be printed for the SLIDE LAYOUT MAP:LOAD

SLIDES screen.

Dako Autostainer | User Guide 67

This page intentionally blank.

68 Dako Autostainer | User Guide

Section 11 |

Reagent Management

Reagent lists are defined using the commands on the Edit lists menu on the PROGRAMMING GRID.

A new reagent can be created and existing reagents can be modified or deleted from the reagent list using this function. Changes can be made to all reagent-related information (reagent name, abbreviation, compatibility code, lot #, expiration date, and incubation time). The following procedures are applicable to all lists except Primary Antibody (see Notes below).

Adding New Reagents to Reagent Lists

Adding Detection Reagents (Secondary, Tertiary, Substrate, etc.)

the categories are displayed.

2. Select the reagent category. The corresponding EDIT REAGENT LIST screen is displayed with the cursor in the Reagent Name box.

NOTE: The Reagent Type box displays the name of the selected list.

All symbol characters such as (*), (:), (/) and (-), etc. are incompatible with the

Autostainer software. Do not use any symbol characters. The use of symbol characters may interrupt the Autostainer during operation.

Dako Autostainer | User Guide 69

3. Type the full name of the reagent you want to add to the current list and press TAB or

ENTER. The software allows the entry of 33 alpha-numeric characters for detection reagents. No symbols can be used in the reagent long name. Once the box is filled an audible signal alerts that no additional characters can be entered. The cursor moves to the Short Name box.

4. Enter an abbreviated name for the reagent in the short name box and press TAB or

ENTER. The abbreviated name cannot exceed 10 alpha-numeric characters and should be unique to distinguish it from other reagents. No symbols can be used in the reagent short name. The cursor moves to the Compatibility box.

5. Enter the correct compatibility code and press TAB or ENTER. Assigning compatibility information to reagents is optional. (See Compatibility Check at the end of this section).

The cursor moves to the Lot box.

6. Enter the reagent lot number and press TAB or ENTER. The lot # cannot exceed 8 alphanumeric characters. Assigning lot number information to reagents is optional. The cursor moves to the Exp. Date box.

7. Enter the reagent expiration date and press TAB or ENTER. The reagent expiration date entry is optional. To enter, use XXYY where XX is a two-digit numeric character symbolizing a month and YY signifies a year. A maximum of 5 characters can be entered.

The entry of expired reagents is not accepted by the software. The cursor moves to the

Time box.

NOTE: All expiration dates entered are limited to a maximum of 10 years from the date of entry.

8. Enter the reagent incubation time and press TAB or ENTER. The cursor moves the

Reagent Name box to allow addition of the next reagent within the same reagent category. Follow steps 3-8 until all reagents have been added.

NOTE: A minimum of 2 minutes incubation time is required for proper operating conditions. the prompts “Save changes on disk?”. Select the YES button and the new reagent is saved to the appropriate list. Select the NO button and the new reagent is not saved. The

PROGRAMMING GRID is displayed.

Select the CANCEL button to escape without saving. The PROGRAMMING GRID is displayed.

70 Dako Autostainer | User Guide

Adding Primary Antibodies

A primary antibody can be added to the reagent list with a specific pretreatment reagent. This will facilitate automatic selection of the pretreatment whenever a primary antibody is selected at the

PROGRAMMING GRID. the categories are displayed.

2. Select the Primary Antibody category. The Primary Antibody EDIT REAGENT LIST screen is displayed with the cursor in the text box in the Antibody section.

NOTE: All symbol characters such as (*), (:), (/) and (-), etc. are incompatible with the

Autostainer software. Do not use any symbol characters. The use of symbol characters may interrupt the Autostainer during operation.

3. Type the full name of the antibody you want to add to the current list and press TAB or

ENTER. The software allows the entry of 32 alpha-numeric characters for primary antibodies. No symbols can be used in the reagent long name. Once the box is filled an audible signal alerts that no additional characters can be entered. The cursor moves to the Short Name box.

NOTE: The Dako programmed list of primary antibodies includes matched negative controls for all primary antibodies listed. The name of a negative control is that of its corresponding primary antibody with the prefix “_NC”. For example, the negative control for p53 is _Ncp53.

NOTE: To add a matching negative control be sure to use the exact same name as the positive control and add the prefix “_NC” to the beginning of the reagent name. Matching negative controls are necessary for the use of the Neg.Ctl tool. (See Programming Slide, section 10)

4. Enter an abbreviated name for the reagent in the short name box and press TAB or

ENTER. The abbreviated name cannot exceed 10 alpha-numeric characters and should be unique to distinguish it from other reagents. No symbols can be used in the reagent

short name. The cursor moves to the Manual Step check box.

Dako Autostainer | User Guide 71

5. If the primary antibody is a step that is preformed off the Autostainer (i.e. overnight incubation or requires heating) then check the Manual Step check box by pressing the

SPACE BAR. When this reagent is selected at the PROGRAMMING GRID, “(Manual

Step)” is displayed in the reagent tile, signifying that this step is preformed before the slides are loaded on the Autostainer. Press TAB or ENTER. The cursor moves to the

Compatibility box.

6. Enter the correct compatibility code and press TAB or ENTER. Assigning compatibility information to reagents is optional. (See Compatibility Check at the end of this section).

The cursor moves to the Lot box.

7. Enter the reagent lot number and press TAB or ENTER. The lot # cannot exceed 8 alphanumeric characters. Assigning lot number information for reagents is optional. The cursor moves to the Exp. Date box.

8. Enter the reagent expiration date and press TAB or ENTER. The reagent expiration date entry is optional. To enter, use XXYY where XX is a two-digit numeric character symbolizing a month and YY signifies a year. A maximum of 5 characters can be entered.

The entry of expired reagents is not accepted by the software. The cursor moves to the

Time box.

NOTE: All expiration dates entered are limited to a maximum of 10 years from the date of entry.

9. Enter the reagent incubation time and press TAB or ENTER. The cursor moves the

Pretreatment section.

10. Select the Pretreatment from the box by using the UP and DOWN arrows. When the correct pretreatment is listed in the text box, press TAB or ENTER. The pretreatment short name, manual step, and incubation time will appear in the proper text boxes. The cursor moves to the Short Name box.

NOTE: To add a new pretreatment, type the name of the pretreatment name, and press

TAB or ENTER. The cursor moves to the Short Name box.

11. Make any necessary changes to the pretreatment Short Name, and press TAB or

ENTER. The cursor moves to the Manual Step Check box.

12. Make any necessary changes to the Manual Step check box by pressing the SPACE

BAR. If the pretreatment is a step that is preformed off the Autostainer (i.e. Target

Retrieval) then select the Manual Step check box. When this reagent is assigned at the

PROGRAMMING GRID, “(Manual Step)” is displayed in the reagent tile, signifying that this step is preformed before the slides are loaded on the Autostainer. Press TAB or

ENTER. The cursor moves to the Time box.

NOTE: Manual Step setting can also be changes from the Programming Grid, see

Programming Slides, section 10.

NOTE: A minimum of 2 minutes incubation time is required for proper operating conditions.

72 Dako Autostainer | User Guide

13. Enter the reagent incubation time and press TAB or ENTER.

NOTE: If Manual Step was selected, the incubation time will be “0”. Do not change this time.

The cursor moves to the first text box to allow addition of the next primary antibody Follow steps 3-13 until all antibodies have been added.

NOTE: If a new pretreatment was used for this antibody, message appears asking if this pretreatment should be added to the Pretreatment list. Select Yes to add reagent to the pretreatment list.

14. Select the OK button to add the new antibodies to the reagent list. The system prompts

“Save changes on disk?”. Select the YES button and the new antibodies are saved to the

Primary Antibody list. Select the NO button and the new antibodies are not saved. The

PROGRAMMING GRID is displayed.

Select the CANCEL button to escape without saving. The PROGRAMMING GRID is displayed.

Editing Reagent Lists

commands are displayed.

2. Select the reagent category. The corresponding EDIT REAGENT LIST screen is displayed with the cursor in the Reagent Name box.

3. Display the specific reagent by selecting the scroll bar in the Reagent Name box or press the UP or DOWN key on the keyboard until the target reagent appears highlighted. Press

ENTER or click the left mouse button and reagent information is displayed. Negative controls are listed at the end of the primary antibody list. Use the list scroll bar or type

“_NC” and software moves the highlight to the negative controls.

4. Edit fields by highlighting the information and typing over it. When changing the reagent name, the software saves the change as a new reagent.

NOTE: No symbols should be used in the addition or editing of reagents.

5. Select disk?”

Select the NO button and the reagent is not edited. The PROGRAMMING GRID is displayed.

NOTE: Changes saved in a reagent category using Edit Reagents Lists will be reflected in the appropriate reagent list displayed during subsequent programming sessions.

Deleting Reagents from Reagent Lists

the categories are displayed.

2. Select the reagent category. The corresponding EDIT REAGENT LIST screen is displayed and the cursor will be flashing in the Reagent Name box.

Dako Autostainer | User Guide 73

3. Display the specific reagent by selecting the scroll bar in the Reagent Name box or press the UP or DOWN arrow key on the keyboard until the target reagent appears highlighted.

Press ENTER and the reagent information is displayed. the deleted. Selecting the NO button displays the EDIT REAGENT LIST screen and the reagent is not deleted. the reagent file. Select the YES button to save changes. Selecting the NO button deletes the reagent. Select the CANCEL button to discard changes. The PROGRAMMING GRID is displayed.

Updating a Reagent Lot Number and Expiration Date

the commands are displayed.

2. Select the reagent category. The corresponding EDIT REAGENT LIST screen is displayed with the cursor in the Reagent Name box.

3. Display the specific reagent by selecting the scroll bar in the Reagent Name box or press the UP or DOWN key on the keyboard until the target reagent appears highlighted. Press

ENTER or select the left mouse button and reagent information is displayed. Negative controls are listed at the end of the primary antibody list. Use the list scroll bar or type

“_NC” and software moves the highlight to the negative control antibodies.

4. Highlight the current lot number displayed in the lot box. Type the new lot number and the CHANGE button appears. the and type the new expiration date.

7. Select disk?”

Select the NO button and the reagent is not edited. The PROGRAMMING GRID is displayed.

74 Dako Autostainer | User Guide

Printing Reagent Labels

Printing reagent labels from a programmed run.

1. Program or retrieve a previously saved run on the PROGRAMMING GRID, select the

PRINT button. The WHICH REPORT? screen appears.

REAGENT LABELS button. The PRINT REAGENT LABELS screen appears displaying all reagents from the programmed run on the PROGRAMMING GRID.

3. To print all reagent labels, select the PRINT ALL button.

To print only selected reagents, select desired reagents by positioning the mouse pointer over a reagent and clicking the left mouse button. To select more than one reagent, repeat this step.

To de-select a selected reagent, position the mouse pointer over the highlighted reagent and click with the left mouse button. The reagent will become un-highlighted, or deselected.

Once all desired reagents are highlighted, select the OK button.

Printing reagent labels not in a program.

screen appears. the appears displaying all reagents available from the reagent list.

4. To print all reagent labels, select the PRINT ALL button.

To print only selected reagents, select desired reagents by positioning the mouse pointer over a reagent and clicking the left mouse button. To select more than one

reagent, repeat this step.

To de-select a selected reagent, position the mouse pointer over the highlighted reagent and click with the left mouse button. The reagent will become unhighlighted, or de-selected.

5. Once all desired reagents are highlighted, select the OK button.

Dako Autostainer | User Guide 75

Compatibility Check

A compatibility check has been built into the Dako Autostainer software to minimize programming errors resulting from assigning incompatible reagents (such as alkaline phosphatase-labeled streptavidin and DAB substrate) to a slide protocol. This feature has two check levels. The first check is the species reactivity compatibility of the primary antibody and the secondary reagent used. The second addresses the compatibility rules dictated by the enzyme used in the detection system. The following tables summarize the compatibility rules and codes assigned to each reagent type.

A. Species Reactivity

Reagent Type

(Protocol Element)

Primary Antibody

Primary Antibody

Secondary Reagent

Secondary Reagent

Description Compatibility

Code

All monoclonal primary antibodies raised in mouse (e.g., mouse anti-human)

A

B All polyclonal primary antibodies raised in rabbit

(e.g., rabbit anti-human)

All secondary reagents (antibodies) compatible with monoclonal antibodies raised in mouse

(e.g., biotinylated anti-mouse IgG)

A

All secondary reagents (antibodies) reacting with polyclonal antibodies raised in rabbit (e.g., biotinylated anti-rabbit IgG)

B

B. Enzyme System Compatibility

Reagent Type

(Protocol Element)

Endogenous Enzyme

Endogenous Enzyme

Tertiary Reagent

Description Compatibility

Code

Block, HRP-compatible

Block, AP-compatible

HRP-labeled streptavidin

X

Y

X

Y Tertiary Reagent AP-labeled streptavidin

Substrate HRP-compatible

Substrate AP-compatible substrates (i.e. Fast Red, New

Fuchsin)

X

Y

76 Dako Autostainer | User Guide

Every reagent in the reagent categories lists (Protocol Element) can be programmed with an appropriate compatibility code or combination of codes. For example a linking reagent that reacts with mouse and rabbit can be marked with compatibility codes AB. The software verifies that these compatibility rules are not violated during the programming of a staining run. If an incompatible reagent is assigned to a tile in the programming grid, an INCOMPATIBLE REAGENTS dialog box displays the two reagents which are incompatible.

Select the NO button to replace an incompatible reagent. The incompatible reagent is removed from the last programmed tile and the reagent list corresponding to that tile is displayed.

Select the YES button to override the warning message. The next available tile for programming flashes and the corresponding reagent list is displayed.

Dako Autostainer | User Guide 77

This page intentionally blank.

78 Dako Autostainer | User Guide

Section 12 |

Loading Reagents

After programming is completed the Autostainer calculates the time required for the staining run, the amount of wash buffer/deionized water required and the most efficient placement for the reagent vials. the the calculation is complete lists the run time, number of probe washes and the volume of buffer and deionized water required. (The deionized water feature is not available with the

“01” hardware version of the Dako Autostainer.) the for the program and select the OK button. The name cannot exceed 8 alpha-numeric characters; symbols and spaces cannot be used. The REAGENT LAYOUT MAP screen is displayed.

Selecting the NO or CANCEL button displays the REAGENT LAYOUT MAP:LOAD

REAGENTS screen and the program is not saved.

Reagent Layout Map Screen

The REAGENT LAYOUT MAP documents the rack positions of the reagents including the fill volume needed for completing the staining run.

The REAGENT LAYOUT MAP displays a maximum of 32 reagent vials per rack. The abbreviations for reagents are displayed in their appropriate alpha- numeric rack positions.

Each reagent position is color coded to correspond to Dako reagent colors and the colors from the

PROGRAMMING GRID.

Dako Autostainer | User Guide 79

If more than 32 reagent vials are required by the run, a SECOND RACK button becomes available at the bottom of the screen.

The buttons and their functions are:

REAGENT LIST

displays a detailed listing of the reagents used in the current staining run.

NEXT

displays the SLIDE LAYOUT MAP:LOAD SLIDES screen.

BACK TO GRID

returns to the Programming Grid. displays the SLIDE LAYOUT MAP:Review Slides screen.

SLIDE MAP

FIRST/ SECOND

RACK

PRINT

displays the alternate reagent rack when more than 32 reagents are used in a run. If more than 32 reagents are required for a run, this button is labeled SECOND RACK and a second rack can be accessed from this button. The background for the second rack is blue and the vial positions appear with a “b” prefix (for example, A2b indicates position A2 in the second/back rack). When the second rack is not available, this button is labeled FIRST Rack and is grayed out and inactive. prints the reagent layout map.

1. The reagent volume displayed in the REAGENT LAYOUT MAP screen includes a dead volume of at least 200 µL.

2. Load the reagent vials according to the reagent map.

NOTE: Using the Reagent Layout map information, make sure that enough reagent is in the vials to complete the programmed run and that the reagent rack is properly seated in its correct location and orientation.

NOTE: Check that the reagent vials have no bubbles on top of the solution. Remove any bubbles.

Reagent List Screen

To view a detailed listing of the reagents used in the current run, click on the REAGENT LIST button in the REAGENT LAYOUT MAP screen.

Select the PRINT button to generate a detailed reagent list.

Select the OK button to return to the REAGENT LAYOUT MAP screen.

80 Dako Autostainer | User Guide

Missing Reagent Error

The Dako Autostainer prompts the user when an insufficient volume of reagent is measured for a run.

The Autostainer can be paused to allow the user to add more reagent to the specified reagent vial. If sufficient volume of a reagent is not available, the Autostainer will provide a list on the Run Log of skipped or partially applied slides.

To program the Autostainer to check for insufficient volumes at the start of the run, select the Check

volumes option on the SET START TIME screen (See Starting a Run, section 14).

Paused to Add Reagent

Reagent vials are loaded on the Autostainer and a run has been started.

1. The reagent probe senses insufficient or a missing reagent. The MORE REAGENT window is displayed on the screen. The screen states the reagent vial position, amount needed and reagent that is missing. A 60 second count begins to count down and an alarm sounds.

2. If the user notices the MORE REAGENT alarm and clicks the OK button, the ADD

REAGENT dialog box is displayed. This box contains a message notifying the user how much reagent needs to be added.

3. The specified reagent volume is added and the user clicks the OK button.

4. The probe returns to the reagent vial and picks up the reagent. The Run Log records the pause to add reagent.

Slides Skipped

Reagent vials are loaded on the Autostainer and a run has been started.

1. The reagent probe senses insufficient or a missing reagent. The MORE REAGENT window is displayed on the screen. The screen states the reagent vial position, amount needed and reagent that is missing.

2. A 60 second count begins to count down and an alarm sounds.

3. If the user does not notice the MORE REAGENT alarm, after 60 seconds the RUN LOG screen appears and the pause in the run is recorded.

4. The Autostainer tries again to pump reagent into the probe. If no reagent is found, this reagent is skipped and the Autostainer continues with the next reagent. The Run Log records no reagent was found and reports the skipped slides.

5. When the run is completed, the user is prompted with a list of slide(s) that have been skipped. This list can be printed by selecting PRINT.

Dako Autostainer | User Guide 81

Partial Reagent Application

Reagent vials are loaded on the Autostainer and a run has been started.

1. The reagent probe senses insufficient or a missing reagent. The MORE REAGENT window is displayed on the screen. The screen states the reagent vial position, amount needed and reagent that is missing.

2. A 60 second count begins to count down and an alarm sounds.

3. If the user does not notice the MORE REAGENT alarm, after 60 seconds the RUN LOG screen appears and the pause in the run is recorded.

4. The Autostainer pumps and dispenses all the reagent available. A Low Reagent button appears at the right of the screen. Selecting this button exposes the ADD REAGENT window (reagent layout map) with the low reagent location and volume displayed.

5. Once sufficient reagent has been added to the vial, select the BACK button or click on the reagent circle in the ADD REAGENT window. The run will continue.

6. If the initial MORE REAGENT error screen is not noticed, a second MORE REAGENT screen is displayed. This screen informs the user how much additional reagent is needed for the run. This message is displayed for 30 seconds.

7. If the user does not notice the MORE REAGENT alarm, the RUN LOG screen appears and the pause in the run is recorded. Also listed on the RUN LOG screen is:

ƒ The amount of reagent not present.

ƒ The number of slides the Autostainer was able to process with the amount of reagent present.

ƒ Which slides have a partial amount of reagent and how much was applied to each slide.

ƒ Any slides that were skipped.

8. When the run is completed, the user is prompted with a list of slides that have a partial application of reagent and which slides were skipped. This list can be printed by selecting

PRINT.

82 Dako Autostainer | User Guide

Section 13 |

Loading Slides

After programming is completed and the reagents have been loaded, you are ready to load the slides on the slide racks. Print the Slide Layout Map by selecting the PRINT button in the SLIDE LAYOUT

MAP:LOAD SLIDES screen. This map provides a guide for loading slides and can be saved as part of your staining record.

Loading Microscope Slides

NOTE: The slides should soak in a buffer solution containing 0.05% Tween 20 for a minimum of 5 minutes prior to loading on the Autostainer.

1. Tilt the slide rack at a 30

°

angle away from you, allowing the back of the rack handles to rest in the loading slots.

2. Hold the slide by placing your thumb and forefinger on each side of the slide’s unfrosted end.

3. Load the slide into the slide rack by gently pushing the frosted end into position. The slide is securely held to the rack. (see photo)

4. Secure the slide rack in the instrument by aligning the holes in the handles over the pins in the sink frame. Gently flood the slides with buffer containing 0.05% Tween 20 using a squirt bottle to prevent the slides from drying while loading additional slides.

Slide Rack

Microscope slide

Frosted end of slide

Dako Autostainer | User Guide 83

This page intentionally blank.

84 Dako Autostainer | User Guide

Section 14 |

Starting a Run

After the specimen-containing slides and reagent vials have been loaded, it is advisable to go through the following checklist:

Slides

• should be flooded with buffer containing 0.05% Tween 20 using a squirt bottle to prevent slides from drying while loading additional slides.

• should be seated securely in slide rack.

Racks

• should be properly seated in the Autostainer.

Reagents

• should be at room temperature.

• should be placed at proper locations in the reagent racks.

• should contain no air bubbles.

• should contain at least the reagent volume in each vial as specified in the

Reagent Layout Map.

Buffer

• reservoir should contain at least the volume of buffer indicated in the Reagent Layout Map plus any additional volume listed for the completion of a delayed run.

Water

• reservoir should contain at least the volume of water indicated in the Reagent Layout Map.

Waste

• containers should have enough room to accommodate the volume of buffer and water used during the staining run.

After the above checks have been made the Start Run sequence can be initiated.

Dako Autostainer | User Guide 85

Starting a Run without Delay

Starting a staining run immediately. the loading. The SET START TIME window is displayed.

The Start Time is listed as +0:00 and the current time.

The Finish Time is listed as the +HH:MM required to complete the run. The calculated finish time is also displayed. the clear for the arm to move.

Select the YES button. The Autostainer calibrates the reagent dispensing mechanism and primes the water dispensing pump.

Select the NO button and the SET START TIME window is displayed.

NOTE: The “01” hardware version of the Dako Autostainer does not have an additional reservoir for deionized water. If your Autostainer does not have this option the software does not prompt to prime the water pump.

NOTE: If this step is omitted prior to start of the run, the pump will be automatically primed upon starting the run.

4. Select either the Check Volumes or Do Not Check Volumes options located near the bottom of the screen.

NOTE: If the Check Volumes option is selected, the Autostainer will check each required reagent vial to verify that the minimum amount of reagent is available at the start of the run. Extra buffer will be required for this function to clean the probe after checking each reagent. When this option is selected, the amount of required buffer will increase and you will be prompted to add extra buffer at the start of the run.

If the reagent volume is insufficient to complete the run, the Autostainer will stop and prompt the user with a message indicating that adequate reagent was not found. The

Autostainer will sound an alarm and will not proceed until the user returns and selects

OK. The instrument will then resume checking volumes.

If the Do Not Check Volumes option is selected, the Autostainer will not check reagent volumes prior to the start of the run.

86 Dako Autostainer | User Guide

visible throughout the run and displays the steps performed by the Dako Autostainer.

The Elapsed, Remaining, Total, Start, Current, and Finish times are listed on the RUN LOG screen.

The current step, total and current time are always displayed in blue at the bottom of the Run

Log.

NOTE: If the current time is not accurate do not change it during the run. Adjust the time at the completion of the run. To reset time refer to the Windows 95 documentation.

NOTE: After the completion of the run, the slides will be rinsed with deionized water every 60 minutes until the slides are removed from the Autostainer. (The deionized water feature is not available for the “01” hardware version of the Dako Autostainer.)

Dako Autostainer | User Guide 87

Delaying the Start of a Run

Starting a staining run at a later time.

1. Select the gray scroll bar and hold the left mouse button down. Move the scroll bar to the right until the desired Start Time and Finish Time is displayed. Release the mouse button.

The staining run will start and finish at the times indicated.

Gray scroll bar used to adjust start time in large increments.

NOTE: The total amount of buffer is calculated and displayed below the scroll bar.

Scroll arrow used to adjust start time in small increments.

Gray scroll bar

Total buffer calculation clear for the arm to move.

Select the YES button. The Autostainer calibrates the reagent dispensing mechanism and primes the water dispensing pump.

Select the NO button and the SET START TIME window is displayed.

NOTE: The “01” hardware version of the Dako Autostainer does not have an additional reservoir for deionized water. If your Autostainer does not have this option the software does not prompt to prime the water pump.

88 Dako Autostainer | User Guide

visible throughout the run and displays the steps performed by the Autostainer .

The elapsed, remaining and total times are listed on the RUN LOG screen.

The Start, Current, and Finish times are also listed. The current step, total and current time are always displayed in blue at the bottom of the Run Log.

NOTE: If the current time is not accurate do not change it during the run. Adjust the time at the completion of the run. To reset time refer to the Windows 95/98 documentation.

NOTE: The slides are rinsed with buffer every 30 minutes until the staining run begins. After the completion of the run, the slides will be rinsed with DI water every

60 minutes until the slides are removed from the Autostainer. (The deionized water feature is not available for the “01” hardware version of the Dako Autostainer.

Therefore, following completion of the run, the slides will be rinsed with buffer every

60 minutes until the slides are removed from the instrument.)

Delaying the Start of a Run using Substrate-Batch

Batch Time indicates when to load the freshly prepared substrate.

1. Select the gray scroll bar and hold the left mouse button down. Move the scroll bar to the right until the desired Start Time and Finish Time is displayed. The Batch Time indicates the time when the substrate should be loaded.

NOTE: The total amount of buffer is calculated and displayed below the scroll bar.

Dako Autostainer | User Guide 89

The elapsed, remaining and total times are listed on the RUN LOG screen.

The Start, Current, Finish, and Batch times are also listed. The current step, total and current time are always displayed in blue at the bottom of the Run Log.

NOTE: If the current time is not accurate do not change it during the run. Adjust the time at the completion of the run. To reset time refer to the Windows 95/98 documentation.

If Substrate-Batch is used, the Batch Time is listed as the +HH:MM required to apply all reagents except for the substrate. The calculated batch time is also displayed.

NOTE: The slides are rinsed with buffer every 30 minutes until the staining run begins. After the completion of the run, the slides will be rinsed with DI water every

60 minutes until the slides are removed from the Autostainer.

90 Dako Autostainer | User Guide

Section 15 |

Completing a Staining Run

During a staining run the RUN LOG screen is displayed and updated with each step performed. The current step is highlighted in blue and displayed at the bottom of the screen. Previously performed steps can be reviewed using the scroll bars.

NOTE: To view previous steps, use the scroll bar to get to the desired view field, then release the mouse button. Do not hold the button depressed for an extended period. Holding the mouse button down too long may cause the Autostainer to pause.

EMERGENCY

STOP

PRINT LOG

PRINT IHC

can be used to stop the run at any time. Selecting this button pauses the run.

This button disappears once the staining run is complete. prints the Run Log for the current run. Selecting this button during a staining run prints a partial Run Log, showing the last event completed by the Autostainer. This button changes to PRINT IHC once the run has ended. is displayed only after the run has ended. Selecting this button prints an

IHC Report as defined during Initialization. (see Initializing the Dako

Autostainer, Section 6)

NEW

PROGRAM

REVIEW

is used to program a new run while the current run is being processed.

Selecting this button displays the Programming Grid. This button disappears once the staining run is complete. is used to return to the Programming Grid to review the run in progress. No changes

PROGRAM

can be made at this screen. This button disappears once the staining run is complete.

At the end of a staining run an audible tone sounds every few seconds and the RUN LOG screen displays the message “End Program Run”.

Dako Autostainer | User Guide 91

NOTE: If the Print IHC button is not pressed, then a Run Log Report will not be printed.

To print out an IHC report at a later time, see Programming Slides, section 10. the instrument) and the waste pump activates to drain the waste reservoir. A dialog box asks if the Run Log should be saved. For instructions on how to save and print a run log see

Saving and Printing a Run Log below.

Select YES to save the Run Log. Select NO to return to the MAIN MENU screen.

3. Position the slide rack for unloading slides by picking up the slide rack, tilting it at a 30

° angle away from you and mounting the posts on the rack handles into the slots in the sink frame.

4. Unload slides by placing your thumb and forefinger on each slide of the slide (away from the frosted end) and pull the slide straight out.

NOTE: If the run is not ended immediately after the program is finished, then the instrument will idle rinse the slides with deionized water every hour until the run is ended

(idle rinses will be with buffer every hour until the run is ended with “01” hardware versions of the Dako Autostainer.)

Printing a Run Log

Printing a Run Log Without Saving

1. At the completion of the staining run select the OK button on the RUN LOG screen. A dialog box appears: “Save Run Log?” the screen.

Printing and Saving a Run Log

1. At the completion of the staining run select the OK button on the RUN LOG screen. A dialog box appears: “Save Run Log?” box and a suggested file name identical to the saved program file name is displayed. the cannot exceed 8 alpha-numeric characters; symbols and spaces cannot be used. A dialog box prompts “Print Run Log?”

92 Dako Autostainer | User Guide

Printing a Temporary Run Log

If the Autostainer comes to an unexpected stop (i.e. power failure), a temporary copy of the Run

Log is maintained on the computer. This record can be retrieved and the current step of each slide can be identified so that a run can be restarted.

1. Restart the computer.

NOTE: If the Autostainer was not shut down properly, then a Scan Disk may occur when you start up the computer. Follow the instructions on the screen if a Scan Disk occurs.

Select Programs and then select Windows Explorer. THE EXPLORING – (C:) window will appear. the highlight this folder. All the contents of this folder will appear on the right side of the window. At the top of this list are the headings of each column: Name, Size, Type, and

Modified.

4. Position the mouse pointer over the Name heading and press the mouse button. All the files will be listed in alphabetical order. (If pressing this button sorts all files in reverse alphabetical order, then click again on Name to sort again.) At the top of the list, there will be files that are preceded by the following symbol: ~.

5. Double click on the file named: ~runlog. This will launch the temporary Run Log.

NOTE: If the Run Log does not automatically launch, an OPEN WITH window will be displayed. Scroll through the list and select Note Pad. Check the check box near the bottom of this screen to always use this application to open the temporary Run Log.

Select OK.

6. Once the Run Log is displayed the screen, this report can be printed by selecting File, and selecting Print. Reviews the Run Log to find what step the slides on the Autostainer are currently in.

7. Compare steps that have already been treated with the original program. Create a new program by removing steps already performed and restart the new run. Re-check vial map as this may cause vials to be reassigned to new locations.

Emergency Stop

The EMERGENCY STOP button on the RUN LOG screen can be used to stop the run at any time. the displayed. the asks if you want to save the run log. The PROGRAMMING GRID appears.

Select the NO button if you do not want to abort the run. The Autostainer resumes the staining run and the paused time is recorded in the Run Log.

NOTE: When a run is aborted, the Autostainer does not stop immediately. It will first complete the function it was performing when the EMERGENCY STOP button was selected.

Dako Autostainer | User Guide 93

NOTE: If the Autostainer is stopped when there is reagent in the probe, it will ask you if you would like to clean the probe. Select yes if it is safe to allow the arm to move. If there is as object in the pathway of the arm or if the arm has been moved out of calibration during the run, select No.

NOTE: If the ESC key is pressed during Autostainer operation, this will emergency stop the run. Select the YES button to stop the run. Select the NO button if you do not want to abort the run.

Immunohistochemical Report

An Immunohistochemical Report is generated for every staining run and can be printed either by selecting the PRINT IHC button on the RUN LOG screen, or by selecting the PRINT button on the

PROGRAMMING GRID and choosing IHC option.

The IHC Report format is selected during Autostainer initialization. This format can be changed when necessary by selecting the OPTIONS button from the INITIALIZE screen. (see Initializing the Dako

Autostainer, Section 6)

The IHC Report can be configured to print in one of the following formats:

• The One case/page option prints one case on each page of the report.

• The One slide ID/page option prints each Slide ID on a separate page.

• The No page break option prints all slides for a run in a continuous print out.

The IHC Report groups all slides by case number and prints the following information for every slide:

#

• primary antibody information (Full name, Species reactivity, Catalog #, Lot # (if entered),

Expiration date, and Incubation time)

• full name and incubation time for all reagents used

The report includes a Staining Signal/Noise and a Comments section for grading the individual slides at the microscope. A Case Interpretation section is also provided at the end of a case’s slides to record the pathologist’s comments. Each IHC Report is uniquely identified by the header information which includes the date of the report, the institution information, the name of the technician and the slide identification, case #, block ID, tissue and doctor for every case included in the current run.

Shutting Down the Autostainer

2. A dialog box appears asking: “Quit using Autostainer?” Select the YES button.

3. Move the mouse pointer to the lower left corner until the Windows “Start” icon appears.

Press the left mouse button once. A menu of items is displayed.

5. Select the “Shut down the computer” option. the computer displays “It’s now safe to turn off your computer”.

7. Turn off the power on the surge protector.

94 Dako Autostainer | User Guide

Section 16 |

Programming During Staining Run

This feature allows the user to program subsequent runs while the Autostainer is in operation. All programming screens are accessible, including the PROGRAMMING GRID, SLIDE LAYOUT MAP screen, RUN TIME CALCULATION screen and REAGENT LAYOUT MAP screen. A print out of all screens can be generated after the current run is completed.

The Run Log screen displays the features described below. These features can be accessed even while a current run or cleaning routine is in process.

While the current run or cleaning routine is in process, and the RUN LOG screen is displayed: the

GRID appears with the current run displayed.

NOTE: To return to the currently generated run log at any time while programming a subsequent run, select the EXIT button in the PROGRAMMING GRID. A prompt is displayed asking to save any changes to the newly created program. the for the new run. (See Entering Slide Information, Section 8)

3. Set up the protocol template for the next run or select an existing template. (See

Designing a Protocol, Section 9)

4. Program slides and reagents. (See Programming Slides and Reagent Management,

Sections 10 and 11)

NOTE: Use the UP or DOWN arrow key on your keyboard to select reagents from the reagent lists while the Autostainer is in operation. (See Warnings, Section 21) the appears. Select NEXT again and the RUN TIME CALCULATION screen appears. Select the OK button and the “Save program on disk” dialog box appears. Select YES to save the program and enter a file name. The program is saved and the LOAD REAGENTS screen is displayed. Select NEXT again and the RUN LOG screen of the current run is displayed.

NOTE: The file name cannot exceed 8 alpha-numeric characters; symbols and spaces cannot be used.

Open on the File menu to recall the newly programmed run after the current run is completed. (See Completing a Staining Run, Section 15) The “Load program from disk” dialog is displayed. Select the assigned file name and press the OK button. The

PROGRAMMING GRID appears with the desired program grid. Select NEXT to get through all appropriate screens.

7. Start the run. (See Starting a Run, Section 14)

Dako Autostainer | User Guide 95

Reviewing the Current Program during a Staining Run

This feature allows the user to review the PROGRAMMING GRID for the current run while the

Autostainer is in operation. No changes can be made using this function.

While the current run is in process and the RUN LOG screen is displayed:

GRID appears with the programming grid of the current run displayed.

2. To return to the RUN LOG screen select the NEXT or EXIT button.

96 Dako Autostainer | User Guide

Section 17 |

System Maintenance

It is recommended that the Dako Autostainer be cleaned after 150 slides have been processed on the instrument. The cleaning procedure can be divided into a manual phase and an automated phase.

Manual cleaning consists of wiping down any areas exposed to reagents and buffer using deionized

(DI) water and a detergent. Organic solutions are not recommended for cleaning. The slide racks should be removed after completion of the automated cleaning cycle and rinsed with deionized water.

The automated cleaning cycle entails running the cleaning protocol. The Dako Autostainer keeps track of the number of staining runs performed. The instrument displays a maintenance message listing the number of slides since the last cleaning cycle.

During the 20-minute, automated cleaning cycle the reagent probe will be incubated with cleaning solutions to remove any build-up of chromogen. In addition, the wash head dispenses DI water on each of the 48 slide positions to remove any buffer deposits or reagent residue remaining after the staining run.

Automated Cleaning Cycle

the number of slides completed since the last cleaning cycle. the window appears with instructions on how to prepare the instrument for a maintenance run.

NOTE: The following reagents are required to perform a cleaning run:

ƒ

DI

ƒ DAB-Away 1 solution

ƒ DAB-Away 2 solution

ƒ Decolorizing solution

3. Prepare the cleaning reagents as instructed in the CLEANING INSTRUCTIONS window and load into the indicated rack positions. (Positions “A8b” and “A7b” refer to positions A8 and A7 in the back rack.) Pour 100 mL of deionized water into sink and provide 3.5 L deionized water in a water reservoir (The “01” hardware version of the Dako Autostainer

does not have a reservoir for deionized water). Select the OK button. The Autostainer starts the cleaning run and the Run Log is displayed. When the run is completed the

MAINTENANCE screen appears. Select OK to return to the OPTIONS screen.

NOTE: Do not use bleach in the Dako Autostainer. Bleach may react with other chemicals and create toxic fumes.

Dako Autostainer | User Guide 97

NOTE: If running protocols that require more frequent cleaning (i.e. using hematoxylin counterstain or high salt buffers), Dako recommends performing a cleaning run a the end of each day the instrument is used.

Cleaning Log

Each time the Autostainer is cleaned, this is logged in a Cleaning Log so that a record of each cleaning is maintained. Each time a cleaning is reset without performing an actual cleaning run, this is recorded in the Cleaning Log with the notation “(Reset)”. the

CLEANING LOG screen is displayed.

This screen displays every cleaning performed on the Autostainer according to date and time it was cleaned, the user who performed the cleaning, slides since the last clean, and total number of runs on the instrument at time of cleaning.

2. To print the entire report, select the PRINT ALL button. To print a section of the report, use the scroll bar to scroll to the section to be printed and select the PRINT SCREEN button.

NOTE: A scroll bar will only appear when the Cleaning Log becomes too long to be displayed on one screen.

Reservoir Cleaning

1. Pour 100 mL of deionized water into the instrument sink before starting an automated cleaning cycle.

2. Wipe sink dry with a non-abrasive cloth or paper towel after cleaning cycle is completed.

NOTE: Ensure no particles remain in the sink after cleaning.

NOTE: Do not use bleach in the Dako Autostainer. Bleach may react with other chemicals and create toxic fumes.

Warning: Residues of hazardous reagents may be present in the sink basin and tubing when rinsing or cleaning the Dako Autostainer. Wear appropriate personal protective equipment to prevent exposure. Hazardous waste must be disposed of according to local, state or federal regulations. Refer to the reagent-specific Material Safety Data Sheet

(MSDS) for additional information.

Wash Buffer and Water Container Cleaning

1. Dako Autostainer Wash Buffer and deionized water do not contain any preservatives. The wash buffer and deionized water container should be washed according to good laboratory practices each time before they are filled with fresh buffer or deionized water.

2. Also, the end piece of the in-line buffer tubing should be cleaned to avoid clogging.

NOTE: Do not use bleach in the Dako Autostainer. Bleach may react with other chemicals and create toxic fumes.

98 Dako Autostainer | User Guide

Slide Rack Cleaning

Dako recommends additional cleaning for slide racks.

1. Soak the slide racks overnight in a container of deionized water.

2. After soaking the racks, rinse with deionized water and brush off any salt residue before beginning the next run.

Reagent Vial Cleaning

The cleaning and re-use of reagent vials is not recommended by Dako. Additional vials can be purchased from Dako (Code No. S3425).

Computer Maintenance

Follow instructions as defined by the computer operator’s manual.

Backing Up Files

Files can be backed up onto a floppy disk from Windows Explorer. To backup flies, first exit

the Autostainer software.

1. Insert a floppy disk into the A: drive. the

Select Programs and then select Windows Explorer. THE EXPLORING – (C:) window will appear.

3. Under the Folders section on the left side of this window, click on the Autostain folder to highlight this folder. All the contents of this folder will appear on the right side of the window.

4. Highlight all files to be backed up onto disk by clicking on the files with the left-mouse button. To highlight more than one file, hold down the Ctrl key and click with the leftmouse button.

5. After all files to be copied are highlighted, position the mouse pointer over one of the highlighted items and press the right-mouse button. Select Send To and then select 3½

Floppy A. Highlighted files will be copied to the disk.

NOTE: If the floppy disk becomes full with data, you may need to repeat this process with another disk.

Annual Preventative Maintenance

Annual maintenance is essential for reliable operation of the Dako Autostainer. Without preventative maintenance, the reliability and the life of the Dako Autostainer may be compromised.

Annual maintenance must be performed by a trained technician. Contact your local Dako representative for further information.

Dako Autostainer | User Guide 99

This page intentionally blank.

100 Dako Autostainer | User Guide

Section 18 |

Reagent Tracking

Reagent Tracking allows you to monitor the total volume of each reagent that has been used on the

Autostainer. A Dako Autostainer Tracking Report can be generated to display various information about all the reagents used on the instrument.

After each Autostainer run, the usage of each reagent is tracked on the Reagent Tracking screen. The tracking system can maintain the following information: Reagent Name, Lot number, Expiration date, total volume of reagent used, and total number of slides treated with this volume.

Printing the Reagent Tracking Report

Printing the entire Reagent Tracking Report

1. Select the PRINT ALL button.

2. All reagents are printed on the Reagent Tracking Report.

Printing only selected reagents from the Reagent Tracking Report:

1. Select the desired reagent(s) to print onto the report by positioning the mouse pointer over a reagent and clicking the left mouse button.

2. To select more than one reagent, repeat step one.

3. To de-select a selected reagent, position the mouse pointer over the highlighted reagent and click with the left mouse button. The reagent will become un-highlighted, or deselected.

4. Once all desired reagents are highlighted, select the PRINT SELECTED button.

5. Only selected reagents are printed on the Reagent Tracking Report.

Dako Autostainer | User Guide 101

Remove Reagent from the Reagent Tracking Log

Reagents that are outdated or not critical to tracking can be removed from the Reagent Tracking log.

Remove all Reagents from the Reagent Tracking Log

1. Select the RESET ALL button. A RESETTING REAGENT window appears and asks if you are sure you want to reset all reagents.

2. Select YES to remove all reagents from the Reagent Tracking Log.

Select NO to cancel this action and return to the REAGENT TRACKING screen.

Remove Selected Reagents from the Reagent Tracking Log

1. Select the desired reagent(s) to be removed from the Reagent Tracking Log by positioning the mouse pointer over a reagent and clicking the left mouse button.

2. To select more than one reagent, repeat step one.

3. To de-select a selected reagent, position the mouse pointer over the highlighted reagent and click with the left mouse button. The reagent will become un-highlighted, or deselected.

4. Once all desired reagents are highlighted, select the RESET SELECTED button. A

RESETTING REAGENTS window appears asking if you are sure you want to reset selected reagents.

5. Select YES to remove selected reagents from the Reagent Tracking Report.

Select NO to cancel this action and return to the REAGENT TRACKING screen.

102 Dako Autostainer | User Guide

Section 19 |

Wash Buffers

Wash buffers are required for all immunohistochemical staining procedures. They function to eliminate background by washing away the nonspecifically bound proteins from the specimen. Commonly used buffers for immunohistochemical staining are Tris-buffered saline (TBS) or phosphate buffered saline

(PBS). These buffers are recommended by Dako for use on the Dako Autostainer. The following pages describe detailed instructions for preparation of TBS and PBS buffer solutions. Buffers used with the Dako Autostainer should contain a final concentration of 0.05% Tween 20.

Dako Autostainer Wash Buffer

The Dako Autostainer Wash Buffer is a Tris-buffered saline solution (TBS), pH 7.6, containing 0.05%

Tween 20. It is recommended that Dako prepackaged buffer solutions are used for consistent performance.

Intended Use

Dako Tris-Buffered Saline Solution containing 0.05% Tween 20 is used as the standard wash buffer for the Dako Autostainer.

Materials Required

1. Clean carboy, flask or beaker graduated to 1 Liter and 5 Liters

3. Magnetic stir plate with stir bar

4. Prepackaged Dako reagents

Code No.

S3006

S3001

S1968

S1966

Contents

One Liter of Wash Buffer 10X. Each bottle makes 10 liters of 50mM Tris-

HCl, 150mM NaCl, 0.05% Tris-Buffered Saline containing Tween 20 and a preservative, pH 7.6 or

Six packets of TBS containing Tris base, Tris-HCI and NaCl. Each packet makes 1 liter of 50mM Tris-HCl, 150mM NaCl Tris-Buffered Saline, pH 7.6 or

Two packets of TBS containing Tris base, Tris-HCl and NaCl. Each packet makes 5 liters of 50mM Tris-HCl, 150mM NaCl Tris-Buffered Saline, pH 7.6 or

Tween 20, 100 mL.

Dako Autostainer | User Guide 103

Reagent Preparation

One Liter of Tris-Buffered Saline Solution (S3001)

1. Add 1L (1000 mL) of deionized water to a 1L carboy, flask or beaker.

2. Empty entire contents of one packet. Use a magnetic stir plate to mix and dissolve the

TBS granules.

3. Add 0.5 mL of Dako Tween 20 (Code No. S1966) to the Tris-buffered saline solution and mix using a magnetic stir plate.

4. The final solution is 50mM Tris-HCl, 150mM NaCl buffered saline, pH 7.6 with 0.05%

Tween 20.

Five Liters of Tris-Buffered Saline Solution (S1968)

1. Add 5 L (5000 mL) of deionized water to a 5 L carboy, flask or beaker.

2. Empty entire contents of one packet. Use a magnetic stir plate to mix and dissolve the

TBS granules.

3. Add 2.5 mL of Dako Tween 20 (Code No. S1966) to the Tris-buffered saline solution and mix using a magnetic stir plate.

4. The final solution is 50mM Tris-HCl, 150mM NaCl buffered saline, pH 7.6 with 0.05%

Tween 20.

Ten Liters of Tris-Buffered Saline Solution (S3006)

1. Prior to dilution, allow the Wash Buffer 10X solution to reach room temperature (20-

25

°C).

2. Add 1L (1000mL) of concentrated Wash Buffer 10X solution to a 10L carboy, flask or beaker.

3. Add 9L (9000mL) of deionized water to the same container. Use a magnetic stir plate to stir gently for at least 15 minutes.

4. The final solution is 50mM Tris-HCl, 150mM NaCl buffered saline, pH 7.6 with 0.05%

Tween and a preservative.

104 Dako Autostainer | User Guide

Shelf Life and Storage

TBS solutions should be stored at 2-8

°C to inhibit bacterial growth. TBS solutions that become turbid should be discarded.

If Tris-buffered saline is stored at 4

°C, the expected shelf life is 7 days.

If Tris-buffered saline is stored at room temperature, the expected shelf life is 4 days.

NOTE: Dako Autostainer Wash Buffer does not contain any preservatives. It is recommended that the wash buffer carboy is washed according to good laboratory practices each time before it is filled with fresh buffer. Also, it is recommended that the end piece of the in-line buffer tubing is wiped clean periodically to avoid clogging.

Tris-Buffered Saline

The following is a procedure for the formulation of 50mM Tris, 150mM NaCl, buffered saline,

pH 7.6 with 0.05% Tween 20 from common laboratory chemicals.

Intended Use

Dako recommends a Tris-buffered saline solution (TBS), pH 7.6, containing 0.05% Tween 20 as wash buffer for IHC procedures performed on the Autostainer.

Materials Required

1. Clean carboy, flask or beaker graduated to 1L

3. Magnetic stir plate with stir bar

4. Common laboratory chemicals

Chemical

Trizma Base

Trizma HCl

Manufacturer and Product No.

Sigma, T-1503

Sigma, T-3253

Tween 20 Dako, S1966

Hydrochloric Acid 12N EM, HX0603-3

Reagent Preparation

One Liter of 10x Tris-Buffered Saline containing 0.5% Tween 20

1. Add approximately 500 mL of deionized water to a 1L (1000 mL) carboy, flask or beaker.

2. Add 13.9 g of Trizma base, 60.6 g of Trizma HCl and 87.66 g of NaCl and dissolve using a magnetic stir plate.

3. After all granules are dissolved adjust the solution to pH 7.6 with 12 N HCl. Allow the solution to continue to mix on the stir plate. Add the 12N HCl dropwise, observing the pH constantly.

4. Once pH 7.6 is reached, add 5 mL of Tween 20 to the Tris-buffered saline solution and mix using a magnetic stir plate.

5. Adjust the final volume of buffer to 1 liter with deionized water.

Dako Autostainer | User Guide 105

6. The final solution is 0.5M Tris-HCl, 1.5M NaCl, buffered saline, pH 7.6 with

0.5% Tween 20.

One Liter of Tris-Buffered Saline, containing 0.05% Tween 20 from 10x stock

1. Add 900 mL of deionized water to a 1L (1000 mL) carboy, flask or beaker.

2. Add 100 mL of 10X Tris-buffered saline solution and mix using a magnetic stir plate.

3. The final solution is 50mM Tris-HCl, 150mM NaCl buffered saline, pH 7.6 with 0.05%

Tween 20.

One Liter Tris-Buffered Saline containing 0.05% Tween 20

1. Add approximately 500 mL of deionized water to a 1L (1000 mL) carboy, flask or beaker.

2. Dissolve 1.39 g of Trizma base, 6.06 g of Trizma HCl and 8.77 g of NaCl and dissolve using a magnetic stir plate.

3. After all granules are dissolved adjust the solution to pH 7.6 with 12N HCl. Allow the solution to continue to mix on the stir plate. Add the 12N HCl dropwise, observing the pH constantly.

4. Once pH 7.6 is reached, add 0.5 mL of Tween 20 to the Tris-buffered saline solution and mix using a magnetic stir plate.

5. Adjust the final volume of buffer to 1 liter with deionized water.

6. The final solution is 50mM Tris-HCl, 150mM NaC, buffered saline, pH 7.6 with 0.05%

Tween 20.

Shelf Life and Storage

TBS solutions should be stored at 2-8

°C to inhibit bacterial growth. TBS solutions that become turbid should be discarded.

If Tris-buffered saline is stored at 4

°C, the expected shelf life is 7 days.

If Tris-buffered saline is stored at room temperature, the expected shelf life is 4 days.

NOTE: The above formulation for Tris-buffered saline does not contain any preservatives. It is recommended that the wash buffer carboy is washed according to good laboratory practices each time before it is filled with fresh buffer. Also, it is recommended that the end piece of the in-line buffer tubing is wiped clean periodically to avoid clogging.

106 Dako Autostainer | User Guide

Dako Phosphate Buffered Saline

Dako Phosphate Buffered Saline Solution can be used for Autostainer reagent preparation and general laboratory use.

Intended Use

Dako recommends phosphate buffered saline (PBS) in the preparation of many immunohisto- chemical and in situ hybridization procedures.

Materials Required

1. Clean carboy, flask or beaker graduated to 1 Liter

3. Magnetic stir plate with stir bar

4. Prepackaged Dako reagents

Code No.

S3024

Contents

Six packets of PBS containing sodium phosphate dibasic, sodium phosphate monobasic, and sodium chloride. Each packet makes 1 liter of 20mM sodium phosphate, 150mM NaCl, pH 7.0.

Reagent Preparation

One Liter of Phosphate Buffered Saline Solution (S3024)

1. Add l L (1000 mL) deionized water a 1L carboy, flask or beaker.

2. Empty entire contents of one packet. Use a magnetic stir plate to mix and dissolve the

PBS granules.

3. Add 0.5 mL Dako Tween 20 (Code No. 1966) to the phosphate buffered saline solution and mix using a magnetic stir plate.

4. The final solution is 20mM sodium phosphate, 150mM NaCl, phosphate buffered saline, pH 7.0 with 0.05% Tween 20.

Shelf Life and Storage

PBS solutions should be stored at 2-8° C to inhibit bacterial growth. PBS solutions which becomes turbid should be discarded.

If phosphate buffered saline is stored at 4° C, the expected shelf life is 7 days.

If phosphate buffered saline is stored at room temperature, the expected shelf life is 4 days.

NOTE: Dako PBS does not contain any preservatives. It is recommended that the wash buffer carboy is washed according to good laboratory practices each time before it is filled with fresh buffer. Also, it is recommended that the end piece of the in-line buffer tubing is wiped clean periodically to avoid clogging.

Dako Autostainer | User Guide 107

Phosphate Buffered Saline

The following is a procedure for the formulation of 20mM Phosphate, 150mM NaCl, buffered saline, pH 7.0 with 0.05% Tween 20 from common laboratory chemicals.

Intended Use

A phosphate buffered saline solution, pH 7.0 can be used on the Dako Autostainer.

Materials Required

1. Clean carboy, flask or beaker graduated to 1L

3. Magnetic stir plate with stir bar

Common laboratory chemicals

Chemical

Sodium Phosphate, dibasic (anhydrous)

Sodium Phosphate, monobasic (monohydrate)

Tween 20

Sodium Hydroxide (NaOH), 50% w/w solution

Manufacturer and Product No.

Sigma, S-087

Fisher, S369-1

Dako, S1966

Baker, 3727-03

Reagent Preparation

One Liter of Phosphate-Buffered Saline containing 0.05% Tween 20

1. Add approximately 500 mL of deionized water to a 1L (1000 mL) carboy, flask or beaker.

2. Add 0.92 g of Sodium Phosphate, monobasic (monohydrate), 1.92 g of Sodium

Phosphate, dibasic and 8.77 g of NaCl and dissolve using a magnetic stir plate.

3. After all granules are dissolved adjust the solution to pH 7.0 with NaOH (50% w/w solution). Allow the solution to continue to mix on the stir plate. Add the NaOH dropwise, observing the pH constantly.

4. Once pH 7.0 is reached, add 0.5 mL of Tween 20 and mix using a magnetic stir plate.

5. Adjust the final volume of buffer to 1 liter with deionized water.

6. The final solution is 20mM Phosphate, 150mM NaCl, buffered saline, pH 7.0 with 0.05%

Tween 20.

108 Dako Autostainer | User Guide

Shelf Life and Storage

PBS solutions should be stored at 2-8

°C to inhibit bacterial growth. PBS solutions that become turbid should be discarded. If phosphate buffered saline is stored at 4

°C, the expected shelf life is

7 days.

If phosphate buffered saline is stored at room temperature, the expected shelf life is 4 days.

NOTE: The above formulation for phosphate buffered saline does not contain any preservatives. It is recommended that the wash buffer carboy is washed according to good laboratory practices each time before it is filled with fresh buffer. Also, it is recommended that the end piece of the inline buffer tubing is wiped clean periodically to avoid clogging.

Dako Autostainer | User Guide 109

This page intentionally blank.

110 Dako Autostainer | User Guide

Section 20 |

Troubleshooting

Solving Autostainer Problems

Symptom

Autostainer will not turn on. (top

green LED not on) A Cannot

Continue box displays

“Instrument Not On” when a run is started.

The Autostainer does not respond when you initiate a prime pump or a run. (bottom

LED not on) A Cannot

Continue box displays

“Instrument Not On” when a run is started.

Buffer or DI Water does not flow out during a run or a prime.

Possible Cause

The Autostainer is unplugged at the power source.

The cable between the

Autostainer and the computer is not connected properly.

Autostainer was not primed properly.

Not enough buffer or DI water in the appropriate reservoir.

Bent or pinched tubing.

Tubing is disconnected at the inlet fitting(s).

Solutions

Check the power cord connections and verify the surge protector is plugged in and turned on.

Shut down the system and turn the power off. Disconnect and reconnect the cable between the

Autostainer and the computer.

** Please take great care in this action. Pins can be bent and cause damage to the instrument during operation.

Prime pump, see Starting a

Run, section 14. Pump may need to be activated several times in order to activate a

prime

Fill the reservoir with the proper level of fluid. Then prime the pump, see Starting a Run, section 14.

Straighten out the tubing and prime the pump, see Starting a

Run, section 14.

Check the tubing to verify that the fittings are secured properly.

Dako Autostainer | User Guide 111

Symptom

Buffer or DI Water does not flow out during a run or a prime.

(Cont.)

Possible Cause

Inlet filters are clogged and are restricting fluid flow.

All waste collects into one waste container.

Waste backs up in the sink.

Reagent probe is not dispensing reagent.

Tubing not completely inserted into carboy.

Faulty buffer / DI water pump.

“Switch” is not programmed in the Protocol Template.

Waste tubing not placed in the proper waste carboy.

Faulty waste pump.

Clogged sink filter.

Insufficient reagent in reagent vial.

Solutions

Check the inlet filter at the end of the tubing in the carboy - clean if necessary. Dako recommends using a tube and cap assembly (S3472) to prevent foreign objects from entering the carboys and possibly clogging the filter.

Insure that buffer and DI water tubing is fully inserted into the proper carboys.

Contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

Program a “Switch” in the

Protocol Template.

Insure that all waste tubes lead to proper carboy.

Contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

Clean the filter when the instrument is not in use.

Using a small brush and some water, remove the material causing the

blockage.

The Autostainer is equipped with a liquid level sensor. If the sensor detects insufficient reagent to complete the current step, a message appears on the screen prompting you to add missing reagent to the correct reagent vial. You have one minute to add reagent after which some slides may be skipped.

112 Dako Autostainer | User Guide

Symptom

Reagent probe is not dispensing reagent. (Cont.)

Possible Cause

Clogged reagent probe.

Bubbles on the surface of the reagent in the reagent vial.

The yellow 3- way “stopcock” valve on the syringe has been moved.

Z-head assembly is hitting slide racks during priming or during run.

Arm is out of alignment.

Blow head/wash head is bent.

The probe and wash/blow head are not going to the correct locations

The arm has been knocked out of calibration.

Bent probe tip.

Solutions

Contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

Remove bubbles.

Verify that the “OFF” position of the valve is in the down direction. The valve should be in a “T” configuration.

Emergency stop the run and restart the run. Remove steps already performed. Re-check

Reagent Layout Map as this may cause vials to be reassigned to new locations.

Contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

Abort run and save run log.

Reprogram the run deleting out the steps from the Protocol

Design screen that have already been completed. Restart run.

Verify that all reagent vial caps have been removed and slide racks are firmly locked in place in the down position. Contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

Dako Autostainer | User Guide 113

Symptom

No Staining

Possible Cause

Insufficient reagent in reagent vial.

Reagents vials were not loaded in the correct locations in the reagent racks.

Slides were not loaded in the correct locations.

Slides were loaded with tissue side down.

Programming error.

Slides drying out.

Solutions

The Autostainer is equipped with a liquid level sensor. If the sensor detects insufficient reagent to complete the current step, a message appears on the screen prompting you to add missing reagent to the correct reagent vial. You have one minute to add reagent after which some slides may be skipped. Check Run Log for

“Insufficient Reagent **Skipped

Slides**” message.

Check the Reagent Map to verify proper locations. Check

Run Log for “Unable to Locate

Reagent **Skipped Slides**” message.

Check the Load Slide printout to verify proper locations.

Orient slides properly.

Check the programming grid to ensure that the staining run was programmed correctly.

Check that adequate volume is programmed to dispense on each slide that is suitable for desired incubation times.

Ensure that no extreme environmental conditions are contributing to the drying of the slides (i.e. direct sunlight, heating/air conditioning vents, hoods, etc.)

114 Dako Autostainer | User Guide

Solving Software Problems

Symptom

User is unable to access many of the functions in the Initialize section of the program.

Possible Cause

User has been set up as a

“Non-Authorized” user.

Autostainer software will not accept your log-in name and password.

Illegal or Fatal exception errors.

Name and password not entered properly.

Windows error.

EWP/DWP messages on Run

Log.

Enable waste pump/disable waste pump.

EWL/PFR messages on Run

Log.

Faulty waste removal/probe failed to retract.

Solutions

Contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

Ensure that you are using the proper spelling.

Shut down the system and turn the power off at the surge protector. Wait 30 seconds and then power up the system. If this problem persists, contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

This is a normally occurring message. However, if multiple messages occur during a run,

Contact your local Dako representative.

Contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

Dako Autostainer | User Guide 115

Solving Computer Problems

Symptom

The computer will not turn on.

Possible Cause

The computer is unplugged at the power source.

Wall outlet is not working.

Surge protector is not working.

The computer turns on but

“Safe Mode” is displayed on all four corners of the screen

The computer has been improperly shut down.

Solutions

Check the power cord connections and verify the surge protector is plugged in and turned on.

Ensure the wall outlet is operational. Check also for a blown fuse, or try using a different wall outlet for the surge protector.

Check to see if the circuit breaker on the surge protector is popped out. If so, then reset by pushing it back in.

Shut down the computer by using the Windows “Start” button located at the bottom left corner of the screen and selecting Shut Down. Note: This button may be hidden requiring you to drag the mouse pointer off the bottom of the screen for it to appear. Always shut down the computer using this method.

Perform Scan Disk through

Windows

Explorer/Accessories/System

Tools/Scan Disk. For assistance, contact your local

Dako representative.

(See Dako Companies and

Distributors, Section 23)

116 Dako Autostainer | User Guide

Solving Monitor Problems

Symptom

Monitor will not turn on or goes blank intermittently. (power

LED is not on)

The monitor screen is not displaying normally.

Monitor screen went blank.

(power LED is green)

Monitor screen went blank.

(power LED is orange)

Words on your monitor screen are too dim.

Monitor screen locks-up during a run, but the Autostainer is still performing the run.

Image on the desktop is grainy and the color is incorrect.

Possible Cause

The monitor is unplugged at the power source.

The cable between the monitor and the computer is not connected properly.

Screen saver utility installed.

The cable between the monitor and the computer is not connected properly.

The brightness control is not set properly.

Windows error.

Windows started in “Safe

Mode.”

Solutions

Check the power cord connections and verify the surge protector is plugged in and turned on.

Disconnect and reconnect the cable between the monitor and the computer.

Move the mouse. Current screen will reappear. Remove screen saver as this may interfere with the Autostainer operation.

Disconnect and reconnect the cable between the monitor and the computer.

Adjust the brightness control.

Allow the Autostainer to finish the run. Re-boot the system before the next run.

Shut down the computer by using the Windows “Start” button located at the bottom left corner of the screen and selecting Shut Down. Note: This button may be hidden requiring you to drag the mouse pointer off the bottom of the screen for it to appear. Always shut down the computer using this method.

Perform Scan Disk through

Windows Explorer/

Accessories/System Tools/Scan

Disk.

For assistance, contact your local Dako representative.

(See Dako Companies and

Distributors, Section 23)

Dako Autostainer | User Guide 117

Symptom

Image on the desktop is grainy and the color is incorrect.

(Cont.)

Possible Cause

Color depth has been adjusted too low.

Solutions

At the Windows Desktop, use the mouse to right click anywhere other than an icon, a drop down menu will appear.

Select Properties. The

WINDOWS DISPLAY

PROPERTIES window should appear. Select the Settings tab.

In the box marked colors verify that “High Color (16 bit)” is displayed. If it is not, select it and save changes.

118 Dako Autostainer | User Guide

Solving Printer Problems

Symptom Possible Cause

The printer does not respond to print command.

The printer may be in Manual

Feed mode.

The parallel cable between the printer and the computer is not connected properly.

The parallel cable is defective.

The orange LED remains on continuously.

All the printer LEDs are off and the printer does not respond to print command.

The power cord is not connected to the printer properly.

The printer is out of paper.

The printer door is open.

The toner cartridge is not installed properly.

Paper Jam.

The printer is unplugged from the power source.

The printer has malfunctioned.

Solutions

Briefly press and release the

Front Panel Button on the printer.

Disconnect and reconnect the parallel cable between the printer and the computer.

Check the cable on another system with a print job that you know works, or connect another parallel cable.

Check the power cord connections and verify the surge protector is plugged in and turned on.

Add paper.

Close the door securely.

Open the door, reinstall the toner cartridge, then close the door securely.

Clear the paper out of the paper feed path.

Check the power cord connections and verify the surge protector is plugged in and turned on.

Contact your local Dako representative

Dako Autostainer | User Guide 119

Symptom

All the printer LEDs are on steadily (not blinking).

The printer is not feeding properly or jams frequently.

Possible Cause

The printer has a hardware error.

Paper was added to the Paper

Input Bin incorrectly.

The paper Input Bin is too full.

Paper was poorly cut by the paper manufacturer and is sticking together.

The output capacity was exceeded in the Paper Output

Bin or the Front Output Slot.

Solutions

Disconnect the printer from the surge protector and wait 1 minute before reconnecting. If all three LEDs come back on, contact your local Dako representative.

Remove the paper from the

Paper Input. Tap the sides of the stack together and reinsert into the bin.

Remove some of the paper. The

Paper Input Bin can hold up to

100 sheets of regular weight paper.

Although fanning the paper is not generally recommended, it may be an effective way to decrease multi-feeds if the paper was poorly cut and sticking together.

You may also try turning the paper around to feed the opposite end first.

Do not allow more than 100 sheets of

20 lb. Bond weight paper (less for heavier media) to stack in the Paper Output Bin.

120 Dako Autostainer | User Guide

Solving Seymour Label Printer Problems

Symptom Possible Cause

The text of the label does not begin along the left edge of the label and the text is getting cut off.

The Seymour printer may need to be reset.

Error occurs during printing. Labels/ribbon has run out.

Incorrect COM port selected.

Incorrect printer is selected.

Solutions

To reset the Seymour Label

Printer, turn off the printer. Hold down the green label feed button on the top of the printer.

Turn on the printer while holding down the green button and continue holding the button for 3 seconds. The labels advance forward and printer will print.

Release the feed button. Press and release the feed button one more time to advance one label.

Perform a TEST PRINT. The print will be lined up along the left slide of the label.

Replace labels/ribbon in the printer and the printer will restart. For instructions, refer to the Eltron printer User’s Manual.

Select the correct COM port from the SLIDE LABEL

DESIGN screen, see Initializing the Autostainer, section 7.

Select correct printer model from SLIDE LABEL DESIGN screen, see Initializing the

Autostainer, section 7.

Dako Autostainer | User Guide 121

This page intentionally blank.

122 Dako Autostainer | User Guide

Section 21 |

Warnings and Symbols

Keep the cover closed during operation. The robotic arm will move unexpectedly during the operation - stay clear.

Do not pour liquid down the Dako Autostainer sink. The Dako Autostainer is not equipped to drain large liquid volumes poured at high speed.

Do not use bleach in the Dako Autostainer. Bleach may react with other chemicals and create toxic fumes.

Do not impair the movement of the Dako Autostainer robotic arm in any way.

Remove caps from reagent vials before starting a run on the Dako Autostainer.

Place the slide racks in the down position before starting a run on the Dako Autostainer.

Place the reagent racks firmly in their seated position before starting a run on the Dako

Autostainer.

Hazardous reagent wastes must be disposed of according to local, state and federal regulations. Wear appropriate Personal Protective Equipment to prevent exposure.

Do not use the scroll bars to select reagents from the reagent lists when setting up

subsequent runs while the Autostainer is in operation. Using the scroll bars during operation will cause the Dako Autostainer to pause temporarily.

Do not use symbols when programming in reagents or protocols. Doing so may cause errors during the run.

Do not multitask during operation. Multitasking is defined as running more than one software application at a time (this includes the CD player and screen savers). This may lock up the Dako

Autostainer.

Do not print slide labels while the Autostainer is in operation.

Do not install third party software or hardware products. Installing third party products may lock up the Dako Autostainer and may void the warranty.

Do not install screen saver. Installing screen saver may lock up the Dako Autostainer.

Do not make any hardware or software changes before consulting your local Dako representative.

(See Dako Companies and Distributors, Section 23)

Do not attempt to service the Dako Autostainer unless instructed to do so by an authorized Dako representative. Doing so will void the warranty.

Do not relocate the Dako Autostainer System within your facility before contacting your local

Dako representative for vital information that may affect your warranty. (See Dako Companies and

Distributors, Section 23)

Dako Autostainer | User Guide 123

Contact your local Dako representative prior to using non-Dako reagents or solutions on your

Autostainer. Some solvents, acids, and other solutions may cause damage to internal components of the Autostainer and affect your instrument's performance and warrantee.

Symbol Definitions

Instrument Serial Number

Dako Model Number

Date of Manufacture

Alternating Current Input

In-vitro Diagnostic Medical Device

Caution, Consult Accompanying Documents

124

Consult Instructions for Use

Identification of Authorized Representative in the European Community

Identification of Manufacturer

Correct Disposal of this Product

(according to Directive 2002/96/EC on Waste Electrical and Electronic

Equipment [WEEE] applicable in the European Union and other

European countries with separate collection systems).

Contact a Dako representative for disposal of the equipment at the end of its working life. This product should not be mixed with other commercial waste for disposal.

Dako Autostainer | User Guide

Section 22 |

General Limitations

Immunohistochemistry is a multi-step diagnostic process that requires specialized training in the selection of the appropriate reagents, tissue selection, fixation, and processing, preparation of the IHC slide, and interpretation of the staining results.

Tissue staining is dependent on the proper handling and processing of tissues prior to staining.

Improper fixation, freezing, thawing, washing, drying, heating, sectioning or contamination with other tissues or fluids may produce artifacts, antibody trapping, or false-negative results. Inconsistent results may be due to variations in fixation and embedding methods, or to inherent irregularities within the tissue.

Excessive or incomplete counterstaining may compromise proper results.

Use of old or unbuffered fixatives, or exposure of tissues to excessive heat (greater than 60

°C) during processing may result in decreased staining sensitivity.

Normal/non-immune sera from the same animal source as the secondary antisera used in blocking steps may cause false-negative or false-positive results due to auto-antibodies or natural antibodies.

False-positive results may be seen due to non-immunologic binding of reagents to tissue sections. In some case the application of an alternate blocking reagent prior to incubation with the primary antibody may be useful for reducing background. A recommended blocking reagent is Dako

®

Protein

Block Serum-Free (Code No. X0909).

Unexpected negative reactions in poorly differentiated neoplasms may be due to loss or marked decrease of antigen expression or nonsense mutation in the gene(s) coding for the antigen.

Unexpected positive staining in tumors may be from expression of an antigen not usually expressed in morphologically similar normal cells, or from persistence or acquisition of an antigen in a neoplasm that develops morphologic and immunohistochemical features associated with another cell lineage

(divergent differentiation). Histopathologic classification of tumors is not an exact science and some literature reports of unexpected staining may be controversial.

The clinical interpretation of any positive staining or its absence should be complemented by morphological and histological studies with proper controls. Evaluations should be made within the context of the patient’s clinical history and other diagnostic tests. It is the responsibility of a qualified pathologist who is familiar with the antibodies, reagents and methods used to interpret the stained preparation. Staining must be performed in a certified licensed laboratory under the supervision of a pathologist who is responsible for reviewing the stained slides and assuring the adequacy of positive and negative controls.

Reagents may demonstrate unexpected reactions in previously untested tissues. The possibility of unexpected reactions in tested tissue groups cannot be completely eliminated due to biological variability of antigen expression in neoplasms, or other pathological tissues. Contact your local Dako representative with documented unexpected reactions.

Tissues from persons infected with hepatitis B virus and containing hepatitis B surface antigen

(HBsAg) may exhibit nonspecific staining with horseradish peroxidase.

Dako Autostainer | User Guide 125

advertisement

Key Features

  • Universal system
  • Quick start-up
  • Flexible programming
  • Economic use
  • Data Management and Report Generation
  • Slide and Reagent Labels

Frequently Answers and Questions

What is Dako Autostainer's purpose
Dako Autostainer is an automated slide processing system compatible with currently available reagents for the staining of paraffin-embedded and frozen tissue sections, cytospins, cell smears and fine needle aspirates.
How many slides can Dako Autostainer process at once?
The Dako Autostainer can process 1 to 48 microscope slides in a single staining run.
What kind of reports can Dako Autostainer generate?
Dako Autostainer can generate patient, reagent and real-time operation data reports, as well as track reagent usage and log instrument maintenance.