Manual

Add to my manuals
20 Pages

advertisement

Manual | Manualzz
ENERGY AND COMFORT
Indoor Air Quality
Q-TRAK
Indoor Air Quality Meter
TM
Model 7565
Operation and Service Manual
Copyright©
TSI Incorporated / May 2007 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective July 2000)
Seller warrants the goods sold hereunder, under normal use and service as described in the
operator's manual, shall be free from defects in workmanship and material for twenty-four (24)
months, or the length of time specified in the operator's manual, from the date of shipment to the
customer. This warranty period is inclusive of any statutory warranty. This limited warranty is
subject to the following exclusions:
a.
Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of
shipment.
b.
Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment.
c.
Seller does not provide any warranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty
applies.
d.
Unless specifically authorized in a separate writing by Seller, Seller makes no warranty
with respect to, and shall have no liability in connection with, goods which are
incorporated into other products or equipment, or which are modified by any person other
than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated
herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR
BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED
ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL
BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE
PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE
GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL,
CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE
RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS
OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12
months after a cause of action has accrued. The goods returned under warranty to Seller's
factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This
LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms
waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our
customers, our service policy is designed to give prompt attention to any problems. If any
malfunction is discovered, please contact your nearest sales office or representative, or call
Customer Service department at (800) 874-2811 (USA) or (1) 651-490-2811 (International).
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............ 1
CHAPTER 2 SETTING-UP................................................................. 3
Supplying Power to the Model 7565............................................... 3
Installing the Batteries ...............................................................3
Using the AC Adapter................................................................ 3
Using The Probe.............................................................................3
Connecting the Optional Bluetooth Portable Printer....................... 3
Connecting to a Computer..............................................................4
CHAPTER 3 OPERATION ................................................................. 5
Keypad Functions ...........................................................................5
Common Terms ..............................................................................5
Menus .............................................................................................6
DISPLAY SETUP.................................................................................6
SETTINGS .........................................................................................6
DATA LOGGING .................................................................................6
Measurements...........................................................................6
Log Mode/Log Settings .............................................................7
Delete Data................................................................................7
% Memory .................................................................................7
APPLICATIONS ..................................................................................7
Printing Data Using the Portable Printer ........................................ 7
TRAKPRO™ Data Analysis Software ............................................... 8
CHAPTER 4 MAINTENANCE............................................................ 9
Recalibration...................................................................................9
Cases..............................................................................................9
Storage ...........................................................................................9
CHAPTER 5 TROUBLESHOOTING................................................11
APPENDIX A SPECIFICATIONS.....................................................13
APPENDIX B DIP SWITCH SETTINGS...........................................15
iii
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components below. If
anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
3. IAQ Probe
4. Calibration collar
5. AC Adapter
6. USB cable
7. CD-ROM with downloading software
1
Chapter 2
Setting-up
Supplying Power to the Model 7565
The Model 7565 Q-TRAK Indoor Air Quality Meter can be powered in one
of two ways: four size AA batteries or the optional AC adapter.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on the
inside of the battery compartment. The Model 7565 is designed to
operate with either alkaline or NiMH rechargeable batteries. Battery life
will be shorter if NiMH batteries are used. If NiMH batteries are used
the DIP switch will need to be changed. Refer to Appendix B, DIP
Switch Settings. Carbon-zinc batteries are not recommended because of
the danger of battery acid leakage.
Using the AC Adapter
When using the AC adapter, the batteries (if installed) will be bypassed.
Be sure to provide the correct voltage and frequency, which is marked
on the back of the AC adapter.
Using The Probe
The sensing probe relies on the diffusion of air. For best results, try to
keep the sensing probe surrounded by moving air. Do not breathe on the
probe. Humans exhale CO2 levels exceeding 10,000 ppm and it may
take time for the probe to re-stabilize. Use the probe holder to support
the probe when in continuous data logging mode.
Connecting the Optional Bluetooth Portable Printer
To connect the Bluetooth printer to the Model 7565, power on the unit and
the printer. Then select the MENU soft key. From the Menu use the S and T
keys to highlight Discover Printer and press the key. If other TSI Bluetoothprinters are in the area, turn them off before searching. The Model 7565 will
then search for and list all available Bluetooth devices. Select the device
”Handy700”.
If the Model 7565 has previously been connected to a TSI printer, then it
should automatically reconnect to that printer.
If the printer prints question marks (??????), asterisks (******), or random
characters, reset it by turning it off and then on again. If necessary, refer to
the Portable Printer Manual.
3
Connecting to a Computer
Use the Computer Interface USB Cable provided with the Model 7565 to
connect the instrument to a computer for downloading stored data or for
remote polling. Connect the end labeled “COMPUTER” to the computer
USB port and the other end to the data port of the Model 7565.
For more information on how to download stored data see Chapter 3 section
titled TRAKPRO™ Data Analysis Software.
Caution: This symbol is used to indicate that the data port
of the Model 7565 is not intended for connection to a
public telecommunications network. Connect the USB
data port only to another USB port.
4
Chapter 2
Chapter 3
Operation
Keypad Functions
ON/OFF Key
Press to turn the Model 7565 on and off. During
the power up sequence the display will show the
following: Model Number, Serial Number,
Software Revision and Last Date Calibrated.
Arrow (ST) Keys
Press to scroll through choices while setting a
parameter. Pressing the ST keys simultaneously
will lock the keypad to prevent unauthorized
adjustments to the instruments. To unlock the
keypad, press the ST keys simultaneously.
(Enter) Key
Press to accept a value or condition.
Arrow (gor h) and
Menu Soft Keys
Press arrow keys to change choices while setting a
parameter. Press the Menu soft key to select the
Menu selections, which are Display Setup,
Pressure Zero, Settings, Flow Setup, Actual/Std
Set up, Data Logging, Applications, Calibration
and Printer.
Common Terms
In this manual there are several terms that are used in different places. The
following is a brief explanation of the meanings of those terms.
Sample
Consists of all of the measurement parameters
stored at the same time.
Test ID
A group of samples. The statistics (average,
minimum, maximum, and count) are calculated for
each test ID. The maximum number of test IDs is
100.
5
Time Constant
The time constant is an averaging period. It is used
to dampen the display. If you are experiencing
fluctuating flows, a longer time constant will slow
down those fluctuations. The display will update
every second, but the displayed reading will be the
average over the last time constant period. For
example, if the time constant is 10 seconds, the
display will update every second, but the displayed
reading will be the average from the last 10
seconds. This is also referred to as a “moving
average”.
Logging Interval
The logging interval is a frequency period that the
instrument will log readings. For example, if the
logging interval is set to 30 minutes, each sample
will be the average of the last 30 minutes.
Menus
DISPLAY SETUP
Display setup menu is where you will setup the desired parameters to be
displayed on the running screen. With a parameter highlighted you can then
use the ON soft key to have it show up on the running screen or select the
OFF soft key to turn off the parameter. Use PRIMARY soft key to have a
parameter show up on the running screen in a larger display. Only one
parameter can be selected as a primary, and up to 4 secondary parameters
can be selected at one time.
SETTINGS
Settings menu is where you can set the general settings. These include
Language, Beeper, Select Units, Time Constant, Contrast, Set Time, Set
Date, Time Format, Date Format, Number Format, Backlight and Auto Off.
Use the g or h soft keys to adjust the settings for each option and use the key to accept settings.
DATA LOGGING
Measurements
Measurements to be logged are independent of measurements on the
display, and must therefore be selected under DATA LOGGING Æ
Measurements.
6
Chapter 3
Log Mode/Log Settings
You can set Log Mode to Manual, Auto-save, Cont-key, Cont-time,
Program 1 or Program 2.
•
Manual mode does not automatically save data, but instead prompts
the user to save a sample.
•
In Auto-save mode, the user manually takes samples that are
automatically logged.
•
In Cont-key mode, the user starts taking readings and logging by
pressing the key. The instrument will continue taking
measurements until the key is pressed again.
•
In Cont-time mode, the user starts taking readings by pressing the key. The instrument will continue taking samples until a set period
of time has passed.
•
Auto-save, Cont-Key and Cont-time modes have the following
additional Log Settings:
Log Settings
Mode
Auto-save
Log Interval
Cont-key
Log Interval
Cont-time
Log Interval
Test Length
•
Pressing the ST keys simultaneously will lock the keypad to
prevent unauthorized adjustments to the instruments. To unlock the
keypad, press the ST keys simultaneously.
Delete Data
Use this to delete all data, delete test or delete sample.
% Memory
This option displays the memory available. Delete All, under Delete
Data, will clear memory and reset the memory available.
APPLICATIONS
You can choose Draft Rate, Heat flow, Turbulence and % Outside Air in the
Applications menu. After choosing one of these applications, take
measurements or enter data for each line.
Printing Data Using the Portable Printer
To print logged data, first enter the DATALOGGING menu. Then, use the
CHOOSE TEST item to select the data to be printed. After the test is
selected, use the VIEW STATS and VIEW SAMPLES items to select
statistics or individual data points to view and print. After selecting VIEW
STATS or VIEW SAMPLES, press the PRINT key to print the data.
Operation
7
TRAKPRO™ Data Analysis Software
The Q-TRAK Model 7565 comes with special software called TRAKPRO™
Data Analysis Software, which is designed to provide you with maximum
flexibility and power. To install this software on your computer, follow the
instructions on the label of the TRAKPRO software.
Follow the instructions on the label of the TRAKPRO software to install the
software on your computer. TRAKPRO software contains a very
comprehensive Help Function. This utility provides all the necessary
information to guide you in all aspects of software operation. The software is
shipped on a CD-ROM.
To download data from the Model 7565, connect the supplied computer
interface USB cable to the Model 7565 and to a computer USB port. Any
USB port from can be used.
8
Chapter 3
Chapter 4
Maintenance
The Model 7565 requires very little maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model 7565 to TSI for annual recalibration.
Please contact one of TSI’s offices or your local distributor to make service
arrangements and to receive a Return Material Authorization (RMA)
number.
The Model 7565 can also be recalibrated in the field using the
CALIBRATION menu. These field adjustments are intended to make minor
changes in calibration to match a user's calibration standards. The field
adjustment is NOT intended as a complete calibration capability. For
complete, multiple-point calibration and certification, the instrument must be
returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft
cloth and isopropyl alcohol or a mild detergent. Never immerse the
Model 7565. If the enclosure of the Model 7565 or the AC adapter becomes
broken, it must be replaced immediately to prevent access to hazardous
voltage.
Storage
Remove the batteries when storing the unit for more than one month to
prevent damage due to battery leakage.
9
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions
for common problems encountered with the Model 7565. If your symptom is
not listed, or if none of the solutions solves your problem, please contact
TSI.
Table 5-1: Troubleshooting the Model 7565
Symptom
No Display
Possible Causes
Unit not turned on
Low or dead batteries
Dirty battery contacts
Humidity reading
near zero or not
believable
No response to
keypad
Probe exposed to intense
light
Instrument Error
message appears
Memory is full
Keypad locked out
Fault in instrument
Probe Error
message appears
Fault in probe
Corrective Action
Switch unit on.
Replace batteries or plug
in AC adapter.
Clean the battery
contacts.
Shade the probe while
taking samples.
Unlock keypad by
pressing ST keys
similtaneously.
Download data if desired,
then DELETE ALL
memory.
Factory service required
on instrument.
Factory service required
on probe.
WARNING!
Remove the probe from excessive temperature immediately:
excessive heat can damage the sensor. Operating temperature limits
can be found in Appendix A, Specifications.
11
Appendix A
Specifications
Specifications are subject to change without notice.
CO2:
Range:
Accuracy1:
Resolution:
Sensor type:
0 to 5000 ppm
±3% of reading or ±50 ppm, whichever is greater
1 ppm
Non-Dispersive Infrared (NDIR)
Temperature:
Range:
Accuracy:
Resolution:
Response time:
Type:
32 to 140°F (0 to 60°C)
±1.0°F (±0.6°C)
0.1°F (0.1°C)
30 seconds (90% of final value, air velocity at 400 ft/min
[2 m/s])
Thermistor
Relative Humidity:
Range:
5 to 95% RH
±3% RH (includes ±1% hysteresis.)
Accuracy2:
Resolution:
0.1% RH
Response time: 20 seconds (for 63% of final value)
Sensor type:
Thin-film capacitive
CO Sensor:
Range:
Accuracy:
Resolution:
Response time:
Sensor type:
0 to 500 ppm
±3% of reading or 3 ppm whichever is greater [add
±0.5%/°C (0.28%/°F) away from calibration temperature]
1 ppm
<60 seconds to 90% of final value.
Electro-chemical
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C)
Storage:
-4 to 146°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Pollution degree 1 in accordance with IEC 664
Transient over voltage category II
13
Data Storage Capabilities:
Range:
Logs up to 56,035 data points with key (4) measured
parameters enabled, 38.9 days at 1-minute log intervals
Logging Interval:
Intervals:
1 second to 1 hour (user selectable)
Time Constant:
Intervals:
1 sec, 5 sec, 10 sec, 20 sec, 30 sec, (user selectable)
External Meter Dimensions:
3.8 in. × 8.3 in. × 2.1 in. (9.7 cm × 21.1 cm × 5.3 cm)
Meter Probe Dimensions:
Probe length:
Probe diameter of tip:
Meter Weight:
Weight with batteries:
7.0 in. (17.8 cm)
0.75 in. (1.9 cm)
0.8 lbs (0.36 kg)
Power Requirements:
Batteries:
Four AA-size alkaline or rechargeable
or
AC adapter:
6 VDC nominal, 300 mA [Q-TRAK monitor mates with
5.5 mm OD × 2.1 mm ID plug, center pin positive(+)]
1
2
At 77°F (25°C). Add uncertainty of ±0.2%/°F (±0.36%/°C) away from calibrated temperature.
At 77°F (25°C). Add uncertainty of ±0.03% RH/°F (±0.05% RH/°C) away from calibrated
temperature.
14
Appendix A
Appendix B
DIP Switch Settings
To access the DIP switch, remove the batteries from the battery
compartment. On the inside of the battery compartment, there is a window
with a single DIP switch (see Figure B-1). The table below shows the
functions for the switch.
Caution: Make certain that power is turned off before changing the
DIP switch settings.
Switch
1
Function
NiMH
Settings
OFF: Alkaline Batteries
ON: NiMH Rechargeable Batteries
Figure B-1: DIP Switch Location
15
Thank you for reading this data sheet.
For pricing or for further information, please contact us at our UK Office, using the details
below.
UK Office
Keison Products,
P.O. Box 2124, Chelmsford, Essex, CM1 3UP, England.
Tel: +44 (0)1245 600560
Fax: +44 (0)1245 808399
Email: [email protected]
Please note - Product designs and specifications are subject to change without notice. The user is responsible for determining the
suitability of this product.