Below you will find brief information for AED DDU-2300 Semi-Automatic External Defibrillator. The DDU-2300 AED is a Semi-Automatic External Defibrillator that is designed to be easy to use, portable, and battery powered. It has two primary user controls: the ON/OFF and SHOCK buttons. Voice prompts, text prompts, and a display screen with visual prompts provide a simple interface for the operator. The DDU-2300 AED is capable of recording event information, including ECG, audio data (optional), and SHOCK/NO-SHOCK recommendations.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

AED DDU-2300 Semi-Automatic External Defibrillator User Manual | Manualzz
Defibtech DDU-2300
Semi-Automatic
External Defibrillator
User Manual
DAC-2513EN-AB
PRESS “ON”
BUTTON
1
APPLY PADS
FOLLOW AED
INSTRUCTIONS
2
IF INSTRUCTED,
PRESS“SHOCK”
BUTTON
3
B
Notices
Defibtech shall not be liable for errors contained herein or for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.
Information in this document is subject to change without notice. Names and data used in the examples are
fictitious unless otherwise noted.
Limited Warranty
The “Limited Warranty” shipped with Defibtech AED products serves as the sole and exclusive warranty provided
by Defibtech, LLC with respect to the products contained herein.
Copyright
Copyright © 2010 Defibtech, LLC
All rights reserved. Copyright questions should be directed to Defibtech. For contact information, refer to
Chapter 12 of this manual.
!
CAUTION
Federal Law (USA) restricts this device
to sale by or on the order of a physician.
iii
DAC-2513EN-AB
Contents
1
Introduction To The DDU-2300 AED ........................................................... 1-1
Overview ............................................................................................................................ 1-1
The Defibtech DDU-2300 AED ...........................................................................................1-2
Indications...........................................................................................................................1- 4
Contraindications ................................................................................................................1- 4
Operator Training Requirements .........................................................................................1- 4
2
Dangers, Warnings, And Cautions ............................................................... 2-1
Dangers .............................................................................................................................. 2-1
Warnings ............................................................................................................................. 2-1
Cautions .............................................................................................................................. 2-2
3
Setting Up The DDU-2300 AED ................................................................... 3-1
Overview ............................................................................................................................ 3-1
Connecting The Pads .......................................................................................................... 3-1
Installing The Defibtech Data Card (DDC card) (Optional) .................................................. 3-2
Installing And Removing the Battery Pack .......................................................................... 3-3
Checking The DDU-2300 AED Status ................................................................................. 3-3
Completing The Installation ................................................................................................ 3-4
Storing The DDU-2300 AED ............................................................................................... 3-4
4
Using The DDU-2300 AED In Rescue Mode................................................. 4-1
Overview ............................................................................................................................ 4-1
Preparation.......................................................................................................................... 4-2
Heart Rhythm Analysis ....................................................................................................... 4-5
Delivering The Shock .......................................................................................................... 4-5
CPR Period.......................................................................................................................... 4-5
Post Use Procedures .......................................................................................................... 4-6
Rescue Mode Voice And Text Prompts .............................................................................. 4-6
Operational Environment .................................................................................................. 4-10
5
Maintenance And Troubleshooting ............................................................. 5-1
Routine Unit Maintenance .................................................................................................. 5-1
Self-Tests ............................................................................................................................ 5-5
Cleaning .............................................................................................................................. 5-5
Storage ............................................................................................................................... 5-5
Operator’s Checklist............................................................................................................ 5-6
Troubleshooting .................................................................................................................. 5-7
Repair.................................................................................................................................. 5-8
Click any of the section headings above to be automatically re-directed to that page.
See page references noted in green throughout this book for other interactive links.
iv
DAC-2513EN-AB
6
Maintenance Mode ...................................................................................... 6-1
Overview ............................................................................................................................ 6-1
Navigation (in Maintenance Mode) ..................................................................................... 6-1
Entering Maintenance Mode .............................................................................................. 6-2
AED Main Menu Screen ..................................................................................................... 6-2
AED Status Screen ............................................................................................................. 6-2
AED Maintenance Screen ................................................................................................... 6-3
AED Options Screen ........................................................................................................... 6-5
Rescue Options Screen ...................................................................................................... 6-7
Help Topics Screen ............................................................................................................. 6-9
7
DDU-2300 AED Accessories........................................................................ 7-1
Defibrillation Pads ............................................................................................................... 7-1
Battery Packs ...................................................................................................................... 7-1
Data Cards .......................................................................................................................... 7-1
USB Cable........................................................................................................................... 7-1
8
Event Viewing ............................................................................................. 8-1
DefibView II ........................................................................................................................ 8-1
Defibtech Data Cards (DDC cards) .................................................................................... 8-1
Downloading The Internal Data Log ................................................................................... 8-1
9
Technical Specifications.............................................................................. 9-1
Defibtech DDU-2300 AED .................................................................................................. 9-1
Battery Packs ...................................................................................................................... 9-5
Self-Adhesive Defibrillation Pads ........................................................................................ 9-5
Event Documentation ......................................................................................................... 9-6
Defibtech Event Viewer ...................................................................................................... 9-6
Recycling Information ......................................................................................................... 9-6
Notice To European Union Customers ............................................................................... 9-6
10
Electromagnetic Conformity ..................................................................... 10-1
Guidance And Manufacturer’s Declaration ........................................................................ 10-1
11
Glossary Of Symbols ................................................................................. 11-1
12
Contacts .................................................................................................... 12-1
Click any of the section headings above to be automatically re-directed to that page.
See page references noted in green throughout this book for other interactive links.
v
DAC-2513EN-AB
Introduction To The DDU-2300 AED
1. Introduction
1
This User Manual provides information to guide trained operators in the use and maintenance of the Defibtech
DDU-2300 Semi-Automatic External Defibrillator (“AED”) and its accessories. The DDU-2300 refers to all AEDs in the
DDU-2300 series of AEDs.
This chapter includes intended use, an overview of the AED, a discussion of when it should and should not be used,
and information on operator training.
Overview
The DDU-2300 AED is a Semi-Automatic External Defibrillator (“AED”) that is designed to be easy to use, portable,
and battery powered. It has two primary user controls: the ON/OFF and SHOCK buttons. Voice prompts, text
prompts, and a display screen with visual prompts provide a simple interface for the operator. The DDU-2300
AED is capable of recording event information, including ECG, audio data (optional), and SHOCK/NO-SHOCK
recommendations.
When connected to a patient who is unconscious and not breathing, the DDU-2300 AED performs the following tasks:
UÊ*Àœ“«ÌÃÊ̅iʜ«iÀ>̜À]Ê̅ÀœÕ}…Ê>Õ`ˆœ]ÊÌiÝÌ]Ê>˜`Êۈ`iœÊ«Àœ“«ÌÃ]Ê̜ʫÀi«>ÀiÊ̅iÊ«>̈i˜ÌÊvœÀÊÌÀi>̓i˜Ì°
UÊÕ̜“>̈V>ÞÊ>˜>ÞâiÃÊ̅iÊ«>̈i˜Ì½ÃÊ
°
UÊiÌiÀ“ˆ˜iÃÊ܅i̅iÀÊ>ÊŜVŽ>LiÊÀ…Þ̅“ʈÃÊ«ÀiÃi˜Ì°
UÊ
…>À}iÃÊ̅iÊ`iwLÀˆ>̈œ˜ÊV>«>VˆÌœÀÊ>˜`Ê>À“ÃÊ̅iÊ-"
ÊLÕÌ̜˜ÊˆvÊ̅iÊÊ`iÌiVÌÃÊ>ÊŜVŽ>LiÊÀ…Þ̅“°
UÊ*Àœ“«ÌÃÊ̅iʜ«iÀ>̜ÀÊ̜ʫÀiÃÃÊ̅iÊ-"
ÊLÕÌ̜˜Ê܅i˜Ê̅iÊ`iۈViʈÃÊÀi>`ÞÊ>˜`Ê>ÊŜVŽÊˆÃÊÀiVœ““i˜`i`°
UÊiˆÛiÀÃÊ>ÊŜVŽÊœ˜ViÊ̅iÊ`iۈViʅ>ÃÊ`iÌiÀ“ˆ˜i`Ê>ÊŜVŽÊˆÃÊÀiµÕˆÀi`Ê>˜`Ê̅iÊ-"
ÊLÕÌ̜˜Ê…>ÃÊLii˜Ê«ÀiÃÃi`°Ê
UÊ*ÀœÛˆ`iÃʈ˜ÃÌÀÕV̈œ˜ÃÊ̜ʫiÀvœÀ“Ê
*,°
UÊ,i«i>ÌÃÊ̅iÊ«ÀœViÃÃʈvÊ>``ˆÌˆœ˜>ÊŜVŽÃÊ>ÀiÊÀiµÕˆÀi`°
The Defibtech AED will NOT shock a patient automatically; it will only advise the operator. The SHOCK button is
enabled only when a shockable rhythm is detected and the device is charged and ready to shock. Charging occurs
automatically when the device detects a shockable rhythm. The operator must press the SHOCK button to initiate
defibrillation.
The DDU-2300 AED uses two self-adhesive defibrillation pads (also known as electrode pads, electrodes, monitoring
pads, or pads) to monitor ECG signals and, if necessary, to deliver defibrillation energy to the patient. These pads
are provided in a single-use, disposable package. The DDU-2300 AED determines proper pad-to-patient contact by
monitoring the impedance between the two pads (impedance varies with the electrical resistance of the patient’s
body).
The DDU-2300 AED user interface is clear and concise. It has two primary push-button controls and a display screen.
Easily understandable voice messages and text and video prompts guide the operator through the use of the unit.
The device communicates the status of the AED and of the patient to the operator.
Defibrillation energy is delivered as an impedance compensated biphasic truncated exponential waveform. The
device delivers 150 Joules of defibrillation energy (into a 50-ohm load) when using adult pads and 50J of defibrillation
energy (into a 50-ohm load) when using attenuated child/infant pads (also known as pediatric pads). Energy delivered
does not change significantly with patient impedance, although the duration of the generated waveform will vary.
Defibrillation and AED operating power is supplied by a replaceable (non-rechargeable) battery pack that provides
vœÀʏœ˜}ÊÃÌ>˜`LÞʏˆviÊ>˜`ʏœÜʓ>ˆ˜Ìi˜>˜Viʜ«iÀ>̈œ˜°ÊÊ>ÌÌiÀÞÊ«>VŽÃÊ>ÀiÊ>Û>ˆ>Liʈ˜ÊÃiÛiÀ>ÊVœ˜w}ÕÀ>̈œ˜Ãʜ«Ìˆ“ˆâi`Ê
for use in specific applications. Each battery pack is marked with an expiration date.
The DDU-2300 AED records event documentation internally and, optionally, on Defibtech Data Cards (“DDC cards”).
The optional DDC card plugs into a slot in the AED and enables the AED to record event documentation and, optionally,
audio data onto the card. Audio recording is selectable through configuration settings. Event documentation stored
internally can be downloaded onto a DDC card for review.
A USB port is provided to perform maintenance and data recovery. The USB interface allows connection to a personal
computer. Defibtech PC maintenance software helps support event downloading and unit maintenance operations.
1-1
DAC-2513EN-AB
1. Introduction
DAC-2513EN-AB
1-2
I. Defibtech Data Card (DDC card). This
optional plug-in card provides enhanced
storage capabilities to the DDU-2300 AED.
H. USB Port. The USB port is provided to
perform data recovery and maintenance.
Not to be used during rescue operation.
G. Softkey Buttons. Three context
sensitive softkey buttons are used to
navigate menus or select actions.
F. Active Status Indicator (ASI). The ASI
indicates the current status of the AED.
This indicator flashes green to indicate the
unit is ready for use and flashes red to
indicate unit needs attention from the
user or needs servicing.
E. Pads Connector Socket. The Pads
Connector (item O) is inserted into
this socket.
D. ON/OFF Button. This button is used to turn
the DDU-2300 AED on and off.
C. Display Screen. Color display panel
used to display text and video prompts,
messages, indicators for rescue, unit status,
and maintenance operations. The display
screen provides visual prompts, including
CPR coaching, to assist rescuers with
step-by-step instruction.
B. SHOCK Button. This button will flash
when a shock is recommended. Pressing
this button will deliver a shock when the
button is flashing. This button is disabled at
all other times.
A. Speaker. The speaker projects the voice
prompts when the DDU-2300 AED is on.
The speaker also emits a “beep” when the
unit is off and has detected a condition that
requires attention from the user or needs
servicing.
T. Battery Pack Expiration Date. The battery
pack expiration date is printed on the label on
the battery pack. Do not use the battery
pack after the printed date has past.
S. Battery Pack. The battery pack provides a
replaceable main power source for the
DDU-2300 AED.
R. Battery Pack Serial Number. The battery
pack’s serial number is located on the label on
the battery pack.
Q. Defibrillation Pads Expiration Date (back
side). The defibrillation pads expiration date
is located on the back side of the pads
package. Do not use the pads after the
printed date has passed.
P. Defibrillation Pads. The defibrillation
pads are pads that are placed on the patient.
The pads may be stored in the pad storage
area on the back of the unit.
O. Pads Connector. This connector attaches
the patient pads to the unit at the Pads
Connector Socket (item E).
N. Pad Storage Area. The pad storage area
is found on the back of the AED allowing the
pads to be stored in a pre-connected state
for rapid deployment during an emergency.
M. Battery Pack Eject Release Latch.
This release latch releases the battery pack
from the DDU-2300 AED.
L. Battery Pack Opening. This opening is
where the battery pack is inserted into the
unit.
K. Unit Serial Number. The unit’s serial
number can be found on the back of the
AED, above the battery pack opening.
J. USB and Defibtech Data Card (DDC card)
Access Door. Behind the access door is the
USB connector port and Defibtech Data Card
(DDC card) slot.
P. Defibrillation
Pads
BATTERY PACK
R. Battery Pack
Serial Number
1-3
DAC-2513EN-AB
T. Battery Pack
Expiration Date
S. Battery Pack
Q. Defibrillation Pads
Expiration Date
(on back of package)
N. Pad Storage Area
BACK OF AED
L. Battery Pack Opening
K. Unit Serial Number
J. USB and Defibtech Data Card
(DDC Card) Access Door
I. Defibtech Data Card
(DDC Card) (optional)
H. USB Port
G. Softkey Buttons
(Top, Center, Bottom)
F. Active Status Indicator (ASI)
M. Battery Pack
Eject Release Latch
DEFIBRILLATION
PADS PACKAGE
O. Pads Connector
FRONT OF AED
A. Speaker
B. SHOCK Button
C. Display Screen
D. ON/OFF Button
E. Pads Connector Socket
AW -102
The Defibtech DDU-2300 AED
1. Introduction
1. Introduction
Indications
The DDU-2300 Semi-Automatic External Defibrillator (“AED”) is indicated for use on victims of sudden cardiac arrest
(SCA) who are:
UÊ1˜Vœ˜ÃVˆœÕÃÊ>˜`Ê՘Ài뜘ÈÛi
UÊ œÌÊLÀi>̅ˆ˜}
For patients under 8 years old or less than 55 pounds (25kg), use child/infant electrode pads. Do not delay therapy
to determine exact age or weight.
The DDU-2300 AED must be used by or on the order of a physician.
Contraindications
The DDU-2300 AED should not be used if the patient shows any of the following signs:
UÊ
œ˜ÃVˆœÕÃÊ>˜`ɜÀÊÀi뜘ÈÛi
UÊÀi>̅ˆ˜}
UÊ>ÃÊ>Ê`iÌiVÌ>LiʫՏÃi
Operator Training Requirements
In order to safely and effectively operate the DDU-2300 AED, a person shall have met the following requirements:
UÊ1‡ÓÎääÊÊ>˜`ɜÀÊ`iwLÀˆ>̈œ˜ÊÌÀ>ˆ˜ˆ˜}Ê>ÃÊÀiµÕˆÀi`ÊLÞʏœV>]ÊÃÌ>Ìi]Ê«ÀœÛˆ˜Vˆ>]ʜÀʘ>̈œ˜>ÊÀi}Տ>̈œ˜Ã°Ê
UʘÞÊ>``ˆÌˆœ˜>ÊÌÀ>ˆ˜ˆ˜}Ê>ÃÊÀiµÕˆÀi`ÊLÞÊ̅iÊ>Õ̅œÀˆâˆ˜}Ê«…ÞÈVˆ>˜°
UÊ/…œÀœÕ}…ÊŽ˜œÜi`}iÊ>˜`Ê՘`iÀÃÌ>˜`ˆ˜}ʜvÊ̅iʓ>ÌiÀˆ>Ê«ÀiÃi˜Ìi`ʈ˜Ê̅ˆÃÊ1ÃiÀÊ>˜Õ>°
1-4
DAC-2513EN-AB
2
Dangers, Warnings, And Cautions
This chapter includes a list of danger, warning, and caution messages that relate to the Defibtech DDU-2300 AED and
its accessories. Many of these messages are repeated elsewhere in this User Manual and on the DDU-2300 AED or
accessories. The entire list is presented here for convenience.
!
DANGERS:
Immediate hazards that will result in serious personal injury or death.
UÊÊ>â>À`œÕÃÊiiVÌÀˆV>ÊœÕÌ«ÕÌ°Ê/…ˆÃÊiµÕˆ«“i˜ÌʈÃÊvœÀÊÕÃiʜ˜ÞÊLÞʵÕ>ˆwi`Ê«iÀܘ˜i°
UÊÊ/…iÊ1‡ÓÎääÊʈÃʘœÌÊÃՈÌ>LiÊvœÀÊÕÃiʈ˜Ê̅iÊ«ÀiÃi˜ViʜvÊ>Êy>““>LiÊ>˜iÃ̅ïVʓˆÝÌÕÀi°Ê
UÊÊ œÌÊÃՈÌ>LiÊvœÀÊÕÃiʈ˜ÊœÝÞ}i˜Êi˜ÀˆV…i`Ê>̓œÃ«…iÀi°
2. Dangers, Warnings,
And Cautions
UÊÊ/…iÊ1‡ÓÎääÊʅ>ÃʘœÌÊLii˜ÊiÛ>Õ>Ìi`ʜÀÊ>««ÀœÛi`ÊvœÀÊÕÃiʈ˜Ê…>â>À`œÕÃʏœV>̈œ˜ÃÊ>ÃÊ`iw˜i`ʈ˜Ê̅iÊ >̈œ˜>Ê
Electric Code standard. In compliance with IEC classification, the DDU-2300 AED is not to be used in the presence
of flammable substance/air mixtures.
!
WARNINGS:
Conditions, hazards, or unsafe practices that may result in serious personal injury or death.
UÊʓ«Àœ«iÀÊÕÃiÊV>˜ÊV>ÕÃiʈ˜ÕÀÞ°Ê1ÃiÊ̅iÊ1‡ÓÎääÊʜ˜ÞÊ>Ãʈ˜ÃÌÀÕVÌi`ʈ˜Ê̅iÊ1ÃiÀÊ>˜Õ>°Ê/…iÊ1‡ÓÎääÊÊ
delivers electrical energy that can potentially cause death or injury if it is used or discharged improperly.
UÊʓ«Àœ«iÀʓ>ˆ˜Ìi˜>˜ViÊV>˜ÊV>ÕÃiÊ̅iÊ1‡ÓÎääÊʘœÌÊ̜Êv՘V̈œ˜°Ê>ˆ˜Ì>ˆ˜Ê̅iÊ1‡ÓÎääÊʜ˜ÞÊ>ÃÊ`iÃVÀˆLi`Ê
in the User Manual. The AED contains no user serviceable parts – do not take the unit apart.
UÊÊʏiVÌÀˆV>Ê-…œVŽÊ>â>À`°Ê>˜}iÀœÕÃʅˆ}…ÊۜÌ>}iÃÊ>˜`ÊVÕÀÀi˜ÌÃÊ>ÀiÊ«ÀiÃi˜Ì°ÊœÊ˜œÌʜ«i˜Ê՘ˆÌ]ÊÀi“œÛiÊVœÛiÀÊ­œÀÊL>VŽ®]Ê
or attempt repair. There are no user serviceable components in the DDU-2300 AED. Refer servicing to qualified
service personnel.
UÊʈ̅ˆÕ“ÊL>ÌÌiÀÞÊ«>VŽÃÊ>ÀiʘœÌÊÀiV…>À}i>Li°Ê˜ÞÊ>ÌÌi“«ÌÊ̜ÊÀiV…>À}iÊ>ʏˆÌ…ˆÕ“ÊL>ÌÌiÀÞÊ«>VŽÊ“>ÞÊÀiÃՏÌʈ˜ÊwÀiʜÀÊiÝ«œÃˆœ˜°
UÊʜʘœÌʈ““iÀÃiÊL>ÌÌiÀÞÊ«>VŽÊˆ˜ÊÜ>ÌiÀʜÀʜ̅iÀʏˆµÕˆ`ðʓ“iÀȜ˜Êˆ˜ÊyՈ`Ãʓ>ÞÊÀiÃՏÌʈ˜ÊwÀiʜÀÊiÝ«œÃˆœ˜°
UÊʜʘœÌʏiÌÊyՈ`ÃÊ}iÌʈ˜ÌœÊ̅iÊ1‡ÓÎääÊ°Êۜˆ`Ê눏ˆ˜}ÊyՈ`Ãʜ˜Ê̅iÊʜÀʈÌÃÊ>VViÃÜÀˆiðÊ-«ˆˆ˜}ÊyՈ`Ãʈ˜ÌœÊ
̅iÊ1‡ÓÎääÊʓ>ÞÊ`>“>}iʈÌʜÀÊV>ÕÃiÊ>ÊwÀiʜÀÊŜVŽÊ…>â>À`°Ê
UÊʜʘœÌÊÃÌiÀˆˆâiÊ̅iÊ1‡ÓÎääÊʜÀʈÌÃÊ>VViÃÜÀˆið
UÊÊ1ÃiÊ œ˜ÞÊ iwLÌiV…Ê `ˆÃ«œÃ>LiÊ Ãiv‡>`…iÈÛiÊ `iwLÀˆ>̈œ˜Ê «>`Ã]Ê L>ÌÌiÀÞÊ «>VŽÃ]Ê >˜`Ê œÌ…iÀÊ >VViÃÜÀˆiÃÊ ÃÕ««ˆi`Ê LÞÊ
iwLÌiV…ÊœÀʈÌÃÊ>Õ̅œÀˆâi`Ê`ˆÃÌÀˆLÕ̜ÀðÊ-ÕLÃ̈ÌṎœ˜Êœvʘœ˜‡iwLÌiV…Ê>««ÀœÛi`Ê>VViÃÜÀˆiÃʓ>ÞÊV>ÕÃiÊ̅iÊ`iۈViÊ
to perform improperly.
UʜʘœÌʜ«i˜ÊÃi>i`Ê«>`ÃÊ«>VŽ>}iÊ՘̈Ê«>`ÃÊ>ÀiÊ̜ÊLiÊÕÃi`°Ê
UÊʜʘœÌÊ̜ÕV…Ê̅iÊ«>̈i˜ÌÊ`ÕÀˆ˜}Ê`iwLÀˆ>̈œ˜°ÊÊiwLÀˆ>̈œ˜ÊVÕÀÀi˜ÌÊV>˜ÊV>ÕÃiʜ«iÀ>̜ÀʜÀÊLÞÃÌ>˜`iÀʈ˜ÕÀÞ°ÊÊÊ
UÊʜʘœÌÊ>œÜÊ«>`ÃÊ̜Ê̜ÕV…Ê“iÌ>ÊœLiVÌÃʜÀÊiµÕˆ«“i˜Ìʈ˜ÊVœ˜Ì>VÌÊ܈̅Ê̅iÊ«>̈i˜Ì°ÊœÊ˜œÌÊ̜ÕV…ÊiµÕˆ«“i˜ÌÊVœ˜˜iVÌi`Ê
to the patient during defibrillation. Disconnect other electrical equipment from the patient before defibrillation.
UÊʜʘœÌÊŜVŽÊ܈̅Ê`iwLÀˆ>̈œ˜Ê«>`ÃÊ̜ÕV…ˆ˜}Êi>V…ʜ̅iÀ°ÊœÊ˜œÌÊŜVŽÊ܈̅Ê}iÊÃÕÀv>ViÊiÝ«œÃi`°
UÊÊœÊ ˜œÌÊ >œÜÊ `iwLÀˆ>̈œ˜Ê «>`ÃÊ ÌœÊ ÌœÕV…Ê i>V…Ê œÌ…iÀ]Ê œÀÊ ÌœÊ ÌœÕV…Ê œÌ…iÀÊ Ê iiVÌÀœ`iÃ]Ê i>`Ê ÜˆÀiÃ]Ê `ÀiÃȘ}Ã]Ê
transdermal patches, etc. Such contact can cause electrical arcing and patient skin burns during defibrillation and
may divert defibrillating energy away from the heart.
UÊÊ/…iÊ `iwLÀˆ>̈œ˜Ê «>`ÃÊ >ÀiÊ ˆ˜Ìi˜`i`Ê vœÀÊ œ˜iÊ Ìˆ“iÊ ÕÃiÊ œ˜ÞÊ >˜`Ê “ÕÃÌÊ LiÊ `ˆÃV>À`i`Ê >vÌiÀÊ ÕÃi°Ê ,iÕÃiÊ V>˜Ê i>`Ê ÌœÊ
potential cross infection, improper performance of the device, inadequate delivery of therapy and/or injury to the
patient or operator.
UÊÊۜˆ`ÊVœ˜Ì>VÌÊLiÌÜii˜Ê«>ÀÌÃʜvÊ̅iÊ«>̈i˜Ì½ÃÊLœ`ÞÊ>˜`ÊVœ˜`ÕV̈ÛiÊyՈ`ÃÊÃÕV…Ê>ÃÊÜ>ÌiÀ]Ê}i]ÊLœœ`ʜÀÊÃ>ˆ˜i]Ê>˜`ʓiÌ>Ê
objects, which may provide unwanted pathways for defibrillating current.
2-1
DAC-2513EN-AB
WARNINGS (continued)
UÊʈÃVœ˜˜iVÌÊ>Ê˜œ˜‡`iwLÀˆ>̜ÀÊ«ÀœœvÊiµÕˆ«“i˜ÌÊvÀœ“Ê̅iÊ«>̈i˜ÌÊLivœÀiÊ`iwLÀˆ>̈œ˜Ê̜ʫÀiÛi˜ÌÊiiVÌÀˆV>ÊŜVŽÊ…>âard and potential damage to that equipment.
UÊÊ}}ÀiÃÈÛiʜÀÊ«Àœœ˜}i`Ê
*,Ê̜Ê>Ê«>̈i˜ÌÊ܈̅Ê`iwLÀˆ>̈œ˜Ê«>`ÃÊ>ÌÌ>V…i`ÊV>˜ÊV>ÕÃiÊ`>“>}iÊ̜Ê̅iÊ«>`ðÊ,i«>ViÊ
the defibrillation pads if they become damaged during use.
UÊÊ*œÃÈLiÊ,>`ˆœÊvÀiµÕi˜VÞÊ­,®Êˆ˜ÌiÀviÀi˜ViÊvÀœ“Ê,Ê`iۈViÃÊÃÕV…Ê>ÃÊViÕ>ÀÊ«…œ˜iÃÊ>˜`ÊÌܜ‡Ü>ÞÊÀ>`ˆœÃÊV>˜ÊV>ÕÃiÊ
improper AED operation. Normally using a cell phone near the AED should not cause a problem; however, a distance
of 2 meters (6 feet) between RF devices and the DDU-2300 AED is recommended.
UÊ
*,Ê`ÕÀˆ˜}Ê>˜>ÞÈÃÊV>˜ÊV>ÕÃiʈ˜VœÀÀiVÌʜÀÊ`i>Þi`Ê`ˆ>}˜œÃˆÃÊLÞÊ̅iÊ«>̈i˜ÌÊ>˜>ÞÈÃÊÃÞÃÌi“°Ê
UÊÊ>˜`ˆ˜}ʜÀÊÌÀ>˜Ã«œÀ̈˜}Ê̅iÊ«>̈i˜ÌÊ`ÕÀˆ˜}Ê
Ê>˜>ÞÈÃÊV>˜ÊV>ÕÃiʈ˜VœÀÀiVÌʜÀÊ`i>Þi`Ê`ˆ>}˜œÃˆÃ]ÊiëiVˆ>ÞʈvÊÛiÀÞÊ
low amplitude or low frequency rhythms are present.
UÊʘʫ>̈i˜ÌÃÊ܈̅ÊV>À`ˆ>VÊ«>Vi“>ŽiÀÃ]Ê̅iÊ1‡ÓÎääÊʓ>Þʅ>ÛiÊÀi`ÕVi`ÊÃi˜ÃˆÌˆÛˆÌÞÊ>˜`ʘœÌÊ`iÌiVÌÊ>ÊŜVŽ>LiÊÀ…Þ̅“ðÊ
If you know the patient has an implanted pacemaker, do not place electrodes directly over an implanted device.
2. Dangers, Warnings,
And Cautions
UÊÊÕÀˆ˜}Ê `iwLÀˆ>̈œ˜]Ê >ˆÀÊ «œVŽiÌÃÊ LiÌÜii˜Ê ̅iÊ ÃŽˆ˜Ê >˜`Ê `iwLÀˆ>̈œ˜Ê «>`ÃÊ V>˜Ê V>ÕÃiÊ «>̈i˜ÌÊ ÃŽˆ˜Ê LÕÀ˜Ã°Ê /œÊ …i«Ê
prevent air pockets, make sure self-adhesive defibrillation pads completely adhere to the skin. Do not use dried out
or expired defibrillation pads.
UÊÊ1ÃiÀ‡ˆ˜ˆÌˆ>Ìi`Ê>˜`Ê>Õ̜“>̈VÊÃiv‡ÌiÃÌÃÊ>ÀiÊ`iÈ}˜i`Ê̜Ê>ÃÃiÃÃÊ̅iÊ1‡ÓÎääʽÃÊÀi>`ˆ˜iÃÃÊvœÀÊÕÃi°ÊœÜiÛiÀ]ʘœÊ
degree of testing can assure performance or detect abuse, damage, or a defect that occurred after the most recent
test is completed.
UÊÊ1ÃiʜvÊ`>“>}i`ÊiµÕˆ«“i˜ÌʜÀÊ>VViÃÜÀˆiÃʓ>ÞÊV>ÕÃiÊ̅iÊ`iۈViÊ̜ʫiÀvœÀ“ʈ“«Àœ«iÀÞÊ>˜`ɜÀÊÀiÃՏÌʈ˜Êˆ˜ÕÀÞÊ̜Ê̅iÊ
patient or operator.
!
CAUTIONS:
Conditions, hazards, or unsafe practices that may result in minor
personal injury, damage to the DDU-2300 AED, or loss of data.
UÊʜœÜÊ>ÊL>ÌÌiÀÞÊ«>VŽÊ>Liˆ˜}ʈ˜ÃÌÀÕV̈œ˜Ã°ÊœÊ˜œÌʈ˜ÃÌ>ÊL>ÌÌiÀÞÊ«>VŽÃÊ>vÌiÀÊ̅iÊiÝ«ˆÀ>̈œ˜Ê`>Ìi°
UÊʜœÜÊ >Ê `iwLÀˆ>̈œ˜Ê «>`Ê >LiÊ ˆ˜ÃÌÀÕV̈œ˜Ã°Ê 1ÃiÊ `iwLÀˆ>̈œ˜Ê «>`ÃÊ «ÀˆœÀÊ ÌœÊ Ì…iˆÀÊ iÝ«ˆÀ>̈œ˜Ê `>Ìi°Ê œÊ ˜œÌÊ Ài‡ÕÃiÊ
defibrillation pads. Discard defibrillation pads after use (in the event of suspected pad malfunction, return pads
to Defibtech for testing).
UÊÊ,iVÞViʜÀÊ`ˆÃ«œÃiʜvʏˆÌ…ˆÕ“ÊL>ÌÌiÀÞÊ«>VŽÃʈ˜Ê>VVœÀ`>˜ViÊ܈̅Êvi`iÀ>]ÊÃÌ>Ìi]Ê>˜`ɜÀʏœV>Ê>ÜðÊ/œÊ>ۜˆ`ÊwÀiÊ>˜`Ê
iÝ«œÃˆœ˜Ê…>â>À`]Ê`œÊ˜œÌÊLÕÀ˜ÊœÀʈ˜Vˆ˜iÀ>ÌiÊ̅iÊL>ÌÌiÀÞÊ«>VŽ°ÊʜʘœÌÊVÀÕÅ°ÊÊ
UÊÊ1ÃiÊ >˜`Ê Ã̜ÀiÊ Ì…iÊ 1‡ÓÎääÊ Ê œ˜ÞÊ ÜˆÌ…ˆ˜Ê ̅iÊ À>˜}iÊ œvÊ i˜ÛˆÀœ˜“i˜Ì>Ê Vœ˜`ˆÌˆœ˜ÃÊ Ã«iVˆwi`Ê ˆ˜Ê ̅iÊ ÌiV…˜ˆV>Ê
specifications.
UÊÊvÊ«œÃÈLi]Ê`ˆÃVœ˜˜iVÌÊ̅iÊ1‡ÓÎääÊÊvÀœ“Ê̅iÊ«>̈i˜ÌÊ«ÀˆœÀÊ̜ÊÕÃiʜvʜ̅iÀÊ`iwLÀˆ>̜Àð
UÊÊœÊ ˜œÌÊ Vœ˜˜iVÌÊ Ì…iÊ 1‡ÓÎääÊ ÌœÊ >Ê *
Ê œÀÊ œÌ…iÀÊ `iۈViÊ ­ÕȘ}Ê Ì…iÊ 1-Ê «œÀÌ®Ê Ü…ˆiÊ Ì…iÊ Õ˜ˆÌ½ÃÊ iiVÌÀœ`iÃÊ >ÀiÊ Ã̈Ê
connected to the patient.
UÊ1Ș}ʘœ˜‡iwLÌiV…Ê>Ì>Ê
>À`ÃÊ­
ÊV>À`îʓ>ÞÊ`>“>}iÊ̅iÊ՘ˆÌÊ>˜`Ê܈Êۜˆ`Ê̅iÊÜ>ÀÀ>˜ÌÞ°
UÊʏ̅œÕ}…Ê̅iÊ1‡ÓÎääÊʈÃÊ`iÈ}˜i`ÊvœÀÊ>Ê܈`iÊÛ>ÀˆiÌÞʜvÊwi`ÊÕÃiÊVœ˜`ˆÌˆœ˜Ã]ÊÀœÕ}…Ê…>˜`ˆ˜}ÊLiޜ˜`ÊëiVˆwV>̈œ˜ÃÊ
may result in damage to the unit.
UÊÊi`iÀ>Ê>ÜÊ­1-®ÊÀiÃÌÀˆVÌÃÊ̅ˆÃÊ`iۈViÊ̜ÊÃ>iÊLÞʜÀʜ˜Ê̅iʜÀ`iÀʜvÊ>Ê«…ÞÈVˆ>˜°
2-2
DAC-2513EN-AB
3
Setting Up The DDU-2300 AED
This chapter describes the steps required to make your Defibtech DDU-2300 AED operational. The DDU-2300 AED
is designed to be stored in a “ready” state. This chapter tells you how to make the device ready, so that if and when
you need it, few steps are required to begin using the device.
Overview
The following components and accessories are included with your DDU-2300 AED. Replacement and other
accessories are detailed in the “DDU-2300 AED Accessories” section. Before getting started, identify each
component and ensure that your package is complete.
DDU-2300 AED
Battery Pack
Defibrillation Pads
Package
Defibtech Data Card
(DDC card) (optional)
Defibtech DDU-2300
Semi-Automatic
External Defibrillator
3. Setting Up The
DDU-2300 AED
User Manual
(on Defibtech
User CD)
Operating Guide
(Booklet)
Operating Guide
Connecting The Pads
The DDU-2300 AED defibrillation pads are supplied sealed in a package with the connector and part of the cable
exposed. This allows the pads to be stored in a pre-connected state for rapid deployment during an emergency.
DO NOT open the sealed pads package until the pads
are to be used. The packaging should be opened only
immediately prior to use, otherwise the pads may dry
WARNING
out and become non-functional.
!
Note: The DDU-2300 AED is designed to be stored with the pads connector already installed. This simplifies the
procedure for deploying and operating the device in an emergency.
First, check to ensure that the pads package has not expired. The expiration date is printed on the pad pouch
and is also reported in the AED status screen. Do not use pads past expiration date. Discard expired pads.
3-1
DAC-2513EN-AB
Connecting The Pads (continued)
Insert the connector end of the defibrillation pad cable into the pads connector socket on
the top-left corner of the DDU-2300 AED as shown. Insert pads connector firmly until it is
fully seated in the unit. The connector will only fit in one way – if the connector does not fit,
rotate the connector before trying again.
The connected pads package can then be stored in the pad storage slot in the back of the
DDU-2300 AED. After connecting the pads connector to the unit, push the pads package,
rounded end first, with the pictures on the package facing out, into the pad holder
compartment on the back of the AED. When the pads package is fully inserted, press the
pad cable into the groove in the back of the unit to hold the cable in place and tuck any
excess cable behind the pads package.
The pads are intended for one time use only and must
be discarded after use or if the package has been opened
CAUTION or damaged.
!
Installing The Defibtech Data Card (DDC card) (Optional)
The Defibtech Data Card (“DDC card”) is used to store event and audio information collected by the AED. All DDU2300 AEDs will operate without DDC cards and will still store critical event information internally. DDC cards may
be reviewed with a separate Defibtech PC-based software package. (Refer to the “DefibView II” section in Chapter
8 of this manual.)
3. Setting Up The
DDU-2300 AED
Before installing the DDC card, ensure the AED is turned OFF. Locate the data card/USB port access door on the
right hand side of the unit. Open the data card/USB port access door by slightly pushing and then sliding the door
down to release the latch. The door will spring open. Insert the DDC card, into the thin slot in the side of the AED
centered above the USB port opening, notched end first, label side up until it clicks into place. The card should be
flush with the surface of the slot. If the card does not push in all the way, it may have been inserted upside down.
In that case, remove the card, flip it over, and try inserting it again.
To remove the DDC card, press the card as far as it will go and then release. Upon release, the DDC card will be
partially ejected and can be removed by pulling the DDC card the rest of the way out.
Close the data card/USB port access door by closing and then pushing the door up until the door latch engages.
Using non-Defibtech Data Cards
(DDC cards) may damage the unit
CAUTION and will void the warranty.
!
3-2
DAC-2513EN-AB
Installing And Removing The Battery Pack
The battery pack provides power to the DDU-2300 AED. Do not install the battery pack after the expiration date
printed on the label. The battery pack is non-rechargeable.
Before inserting the battery pack into the DDU-2300 AED, ensure that the
battery pack opening in the back of the AED is clean and clear of any foreign
objects. Insert the battery pack into the opening in the back of the AED.
Push the pack all the way in until the latch clicks. The battery pack will only
fit in one way – if the battery pack does not fit, rotate the battery pack before
trying again. Once fully inserted, the battery pack surface should be flush with
the back of the AED.
Within moments of insertion the DDU-2300 AED will turn on and run a
battery pack insertion self-test.* When the test is completed, the unit will
report the status of the battery pack and shut down. Afterwards, the Active
Status Indicator, adjacent to the ON/OFF button of the DDU-2300 AED, will
periodically flash. If the indicator flashes green, the AED and battery pack
are ready for use. If the indicator flashes red, is solid red, or there is no flashing
light, the AED requires service. (Refer to the “Checking The DDU-2300 AED
Status” section below for more details on the meaning of the indicator.)
*Note: The battery pack must be removed from the unit for more than 10 seconds for the battery pack self-test to
be performed automatically.
To remove the battery pack, push the battery pack eject release latch. After the battery pack is partially ejected, pull
the battery pack out.
Checking The DDU-2300 AED Status
Active Status Indicator (ASI)
3. Setting Up The
DDU-2300 AED
Once a fully functional battery pack is installed in the DDU-2300 AED, an LED indicator located to the right of the
ON/OFF button actively indicates unit status. If the unit is ready for use, the Active Status Indicator (“ASI”) will flash
green. Ready for use means that the DDU-2300 has passed the most recent self-test (scheduled or user initiated).
If the unit needs service, the ASI will flash red. Anytime the ASI flashs red, the unit will also “beep” periodically to
call attention to itself. The ASI also uses a distinct flash pattern to assist people with color blindness: green will flash
a single flash and red will flash a double flash.
The ASI is powered by the battery pack. If the battery pack has been completely discharged or is not installed in the
unit, the active status indication will be off. In this case, immediately replace the battery pack or reinsert it into the
unit to restore active status indication.
UÊ Flashing Green: DDU-2300 AED is OFF and ready for use.
UÊ Solid Green: DDU-2300 AED is ON and ready for use.
UÊ Flashing or Solid Red: DDU-2300 AED needs immediate service. Refer to the
“Troubleshooting” section in Chapter 5 of this manual or call Defibtech for service
Active
Status
Indicator
UÊ No Flashing Light: DDU-2300 AED needs immediate service. Refer to the
“Troubleshooting” section in Chapter 5 of this manual or call Defibtech for service.
3-3
DAC-2513EN-AB
Checking The DDU-2300 AED Status (continued)
AED Status Screen
To check the status of the unit when the unit is off, press the
center softkey button. The display screen will show unit status,
battery pack status, and pad status. After a short period of time,
the display screen, and the unit will turn off.
Completing The Installation
Once you have completed the previous steps to set up your DDU-2300 AED, follow this procedure:
1. Turn the unit on by pressing the power ON/OFF button.
2. Listen for the “Call for Help” voice prompt.
3. Turn unit off by pressing and holding the ON/OFF button.
4. Listen for the “Powering Off ” voice prompt.
5. Check Active Status Indicator to verify that it is flashing green.
(Refer to the ”Self-Tests” section in Chapter 5 of this manual for instructions on how to run a manually initiated
self-test.)
Storing The DDU-2300 AED
Store the DDU-2300 AED, with pads attached, in environmental conditions within range of the specifications. (Refer
to the “Environmental” section in Chapter 9 of this manual.) The unit should also be stored so that the Active Status
Indicator can be readily seen.
3. Setting Up The
DDU-2300 AED
The Active Status Indicator should periodically flash with a green light. If it flashes with a red light or does not flash at
all, the DDU-2300 AED needs servicing. (Refer to the “Checking The DDU-2300 AED Status” section in this chapter
for more information.)
Defibtech recommends storing your AED in an easily accessible location.
3-4
DAC-2513EN-AB
4. Using The DDU-2300 AED
4
Using The DDU-2300 AED In Rescue Mode
This chapter describes how to use the DDU-2300 AED in rescue mode. The DDU-2300 AED was designed for simple
operation, allowing the operator to focus on the patient. There are two primary control buttons and a display screen.
Concise and easily understandable voice messages and text and video prompts guide the operator through the use of
the unit.
The following sections describe in detail how to use the DDU-2300 AED. The basic steps for use are:
UÊ/ÕÀ˜Ê̅iÊ1‡ÓÎääÊÊ" ÊLÞÊ«ÀiÃȘ}Ê̅iÊON/OFF button.
UÊ*Õ}ʈ˜Ê«>`ÃÊVœ˜˜iV̜Àʈ˜ÌœÊ*>`Ê
œ˜˜iV̜ÀÊ-œVŽiÌʜ˜ÊʈvʘœÌÊÞiÌÊ«Õ}}i`ʈ˜°
UÊ*>ViÊ«>`Ãʜ˜Ê«>̈i˜ÌÊ­follow instructions on pads package).
UʜœÜÊۜˆViÊ>˜`Ê`ˆÃ«>ÞÊ«Àœ“«Ìð
UÊ*ÀiÃÃÊSHOCK button if instructed by the AED.
Overview
Pads Connector Socket –
Socket for pads connector
Active Status Indicator (ASI) –
Indicates the current
status of the AED
ON/OFF Button –
Turns AED on and off
Display Screen –
Displays video,
text prompts,
and information
Softkey Buttons
(Top, Center, Bottom) –
Buttons used to navigate
menus or select actions
Speaker –
Creates audio output
from device
SHOCK Button –
Enabled/disabled by
software to allow the
user to discharge a
shock to the patient
4-1
DAC-2513EN-AB
4. Using The DDU-2300 AED
Overview (continued)
Unit Display Screen (During Rescue Mode)
Battery Indicator
Main Screen
Rescue Breathing
Options Softkey Icon
Information
Softkey Icon
Text Prompts
(Sample display screen during Rescue Mode)
Battery Indicator – The Battery Indicator indicates the approximate remaining battery capacity.
Main Screen – The Main Screen displays video instructions to guide the user during a rescue.
Text Prompts – The Text Prompt Area displays text prompts to guide the user during a rescue.
Softkey Buttons (not pictured) – The Softkey Buttons are located to the right of the display screen. If a softkey
button is active, it will have a softkey icon displayed next to it. The softkey buttons are used to navigate menus or
select actions.
Rescue Breathing Options Softkey Icon – When this icon is present on the screen (during a rescue), the
user may press the corresponding softkey button to select CPR coaching with compressions only (no breathing) or
CPR coaching with compressions and breathing.
Information Softkey Icon – When this icon is present on the screen, the user may press the corresponding
softkey button for additional information with video instruction. The additional information is context dependent;
topics include preparing the patient and performing CPR. To exit, press the softkey button again.
Preparation
Checking The DDU-2300 AED Status
Visually check the Active Status Indicator (ASI). The ASI should flash green. The ASI flashes green to indicate ready
for use status. The ASI flashes red, solid red, or is not lit at all to indicate that service is required.
The ASI is powered by the battery pack. If the battery pack has been completely discharged or is not installed in the
unit, the active status indication will not be available. In this case, immediately replace the battery pack or reinsert
it into the unit to restore active status indication.
UÊ Flashing Green: DDU-2300 AED is OFF and ready for use.
UÊ Solid Green: DDU-2300 AED is ON and ready for use.
UÊ Flashing or Solid Red: DDU-2300 AED needs immediate service. Refer to the
“Troubleshooting” section in Chapter 5 of this manual or call Defibtech for service
Active
Status
Indicator
UÊ No Flashing Light: DDU-2300 AED needs immediate service. Refer to the
“Troubleshooting” section in Chapter 5 of this manual or call Defibtech for service.
4-2
DAC-2513EN-AB
4. Using The DDU-2300 AED
Turning On The DDU-2300 AED
Press the green ON/OFF button to turn the DDU-2300 AED on. The unit will emit a “beep” and the display screen will
turn on. The ASI indicator next to the ON/OFF button will illuminate green anytime the AED is on. (To turn the unit off,
press AND HOLD the ON/OFF button for approximately two seconds; the unit will emit a “beep” and power off.)
Call For Help
Call professional emergency services for help. As soon as the AED is turned on the unit will prompt the user to “Call for
help.” This is a reminder that the first step in a rescue should always be to contact professional emergency services.
If another person is available, the user should direct that person to call for help and then continue the rescue without delay.
Preparing The Patient
Prepare the patient by removing any clothing from the patient’s chest. Wipe away moisture from the chest if
necessary (the defibrillation pads will stick better on dry skin). If necessary, shave excessive chest hair, which can
prevent effective patient-electrode contact. To ensure that defibrillation pads fully contact the patient’s skin, check
that no jewelry or other objects are directly underneath where the pads will be placed.
Connecting Defibrillation Pads To The DDU-2300 AED
Connect pads to unit, if not already connected. Follow AED voice and display instructions.
The DDU-2300 AED is designed to be stored with the defibrillation pads connector attached
to the unit, while the pads themselves remain sealed in their package. This reduces the time
needed to setup and start treatment in an emergency.
The Defibtech AED should be stored with the pads connector attached to the unit. However, if
pads were damaged or not properly connected, you may need to substitute a new set of pads
during an emergency. The pads connector socket is on the top left corner of the AED.
To detach a set of pads from the unit, pull firmly on the pad connector. Do not reuse used pads. Insert the
connector for the new pads as shown above. The connector will only fit in one way – if the connector does not fit,
rotate the connector before trying again. Insert connector firmly until it is completely seated in the unit.
When this Information Softkey Icon is present on the screen, the user may
press the corresponding softkey button for additional information with video
instruction. To exit, press the softkey button again.
Opening The Pads Package
Remove the pads package from the pad storage slot at the back of the AED. Open the pads package by tearing
along the dotted line, starting at the black arrow (follow directions on the package). Check that the pads are:
UÊÀiiÊvÀœ“ÊœLۈœÕÃÊÈ}˜ÃʜvÊ`>“>}i°
UÊ
i>˜ÊœvÊiÝViÃÈÛiÊ`iLÀˆÃÊ­vœÀÊiÝ>“«i]Ê`ˆÀÌʈvÊ̅iÊ«>`ÊÜ>ÃÊ`Àœ««i`®°
UÊ œÌÊ`Àˆi`ʜÕÌ]Ê>˜`Ê̅>ÌÊ̅iÊ}iÊˆÃÊÃ̈VŽÞÊ>˜`Ê܈Ê>`…iÀiÊ̜Ê̅iÊ«>̈i˜Ì°
UÊ œÌÊiÝ«ˆÀi`°ÊœÊ˜œÌÊÕÃiÊ«>`ÃÊ>vÌiÀÊ̅iÊiÝ«ˆÀ>̈œ˜Ê`>ÌiÊ«Àˆ˜Ìi`ʜ˜Ê̅iÊ«>VŽ>}i°
If any of these conditions are found, use a new set of pads, if possible.
4-3
DAC-2513EN-AB
4. Using The DDU-2300 AED
Applying Pads To The Patient
Apply pads correctly to patient. Follow AED voice and text prompts. Correct pad placement is essential for
effective analysis of the patient’s cardiac rhythm and subsequent shock delivery (if required).
Remove the pads from the pads package by tearing the package along the dotted line near the top of the package.
Follow the directions and diagram showing proper defibrillation pad placement located on the defibrillation pads
package and on the pads.
Peel off each pad from the blue liner before placing it as shown on the picture on the pad. Peel the pad off the
blue liner only when the pad is ready to be placed on the patient.
Place the pads with the sticky side of the pad on the patient’s skin. Pad placement on infants and children under
8 years or less than 55 pounds (25kgs) is different than placement for adults and children 8 years or older or over
55 pounds (25kgs). If you are unsure of a child’s age or weight or do not have child/infant pads, do not delay
treatment.
Place the pads on the patients bare chest exactly as shown in the picture on the pad. See diagrams below:
Sample Adult Pads
Sample Child/Infant Pads
For adults and children 8 years or older or
over 55 pounds (25kgs), use adult pads:
Place one pad just below the patient’s right collar bone
as shown in the picture. Place the second pad over
the ribs on the patient’s left side below the left breast.
Use picture on pad to determine individual
pad placement.
For infants and children under 8 years or less
than 55 pounds (25kgs), use child/infant pads
(Note: Child/infant pads can be identified by
their blue connector and pads package):
Place one pad in the center of chest and one pad on
the center of the back, as shown. Use picture on
pad to determine individual pad placement.
4-4
DAC-2513EN-AB
4. Using The DDU-2300 AED
Follow DDU-2300 AED Instructions
At this point, the DDU-2300 AED will check to make sure that the pads are well connected to the patient and that
an adequate ECG signal is being received. Do not touch the patient. Eliminate any patient movement, and cease
CPR at this time.
If there is a problem with the pad connection, socket connection, patient motion, or other interference, the AED
will guide the operator with both audible and displayed instructions. Text prompts are identical to or are an abbreviated form of the audio prompts. Video prompts reinforce the audio and text prompts and aid in high ambient
noise environments.
Heart Rhythm Analysis
Once the DDU-2300 AED has determined that the pads are making a good connection to the patient, the AED
܈ÊÃÌ>ÀÌÊ̅iÊ
ÊÀ…Þ̅“Ê>˜>ÞÈðÊÊ/…iÊ՘ˆÌÊ>˜>ÞâiÃÊ̅iÊ
ÊÈ}˜>Ê>˜`Ê`iÌiÀ“ˆ˜iÃÊ܅i̅iÀÊ>ÊŜVŽ>LiʜÀʘœ˜‡
ŜVŽ>LiÊÀ…Þ̅“ʈÃÊ«ÀiÃi˜Ì°ÊÊ7…ˆiÊ>˜>Þ∘}]Ê̅iÊÊ܈ÊVœ˜Ìˆ˜ÕiÊ̜ʓœ˜ˆÌœÀÊÈ}˜>Ê>˜`Ê«>`ÊVœ˜`ˆÌˆœ˜ÃÊ>˜`Ê܈Ê
reassess analysis and prompt the user if additional user action is needed.
Delivering The Shock
If the DDU-2300 AED ECG analysis algorithm has determined that a shock is required, the unit will automatically
charge in preparation for shock delivery. While the AED charges, the unit may emit a charging tone and will continue
̜Ê>˜>ÞâiÊ̅iÊ«>̈i˜Ì½Ãʅi>ÀÌÊÀ…Þ̅“°ÊÊvÊ̅iÊ՘ˆÌÊ`iÌiVÌÃÊ̅>ÌÊ̅iʅi>ÀÌÊÀ…Þ̅“Ê…>ÃÊV…>˜}i`Ê̜ʜ˜iÊ̅>ÌÊ`œiÃʘœÌÊ
ÀiµÕˆÀiÊ>ÊŜVŽ]Ê̅iÊ՘ˆÌÊ܈Ê«Àœ“«ÌÊ̅iÊÕÃiÀÊ̜ÊLi}ˆ˜Ê
*,°ÊÊ7…ˆiÊ>˜>Þ∘}]Ê̅iÊÊ܈ÊVœ˜Ìˆ˜ÕiÊ̜ʓœ˜ˆÌœÀÊÈ}˜>Ê
and pad conditions.
If the unit has determined that a shock is required and has completed charging, the SHOCK button will flash and the
user will be instructed to press the flashing SHOCK button. The user should follow the AED instructions and press
the SHOCK button.
Important: The DDU-2300 AED will not automatically deliver a shock – the user must press the flashing SHOCK button.
The user can abort charging or shock delivery at any time by pressing and holding the ON/OFF button for approximately
two seconds to turn the unit off.
CPR Period
The operator will be prompted to begin CPR. The unit will not monitor the patient’s ECG rhythm during the CPR period.
During the CPR period, the AED will not advise the user to “stop motion” even if motion is present.
The user should follow the AED instructions during this time. Once the CPR period is complete, the unit will
continue in Heart Rhythm Analysis mode.
CPR coaching is provided through a series of voice and visual prompts and audible tones. The factory default setting
provides prompts for chest compressions only (no breathing).
However, breathing prompts can be enabled/disabled by pressing the softkey button next to the Rescue Breathing
Options icon displayed on the screen during rescue. (Refer to the “Rescue Breathing Options Softkey Icon” section
below.) Breathing prompts can also be enabled/disabled by setting the menu option in Maintenance Mode. (Refer
to the “CPR Breathing” section in Chapter 6 of this manual.)
Rescue Breathing Options Softkey Icon: During a rescue, when this icon is present on the
sreen, the user may press the corresponding softkey button to select CPR coaching with
compressions only (no breathing) or CPR coaching with compressions and breathing.
Note: Refer to the “CPR Breathing” section of Chapter 6 for instructions on how to change
the factory default setting.
When this Information Softkey Icon is present on the screen, the user may
press the corresponding softkey button for additional information with video
instruction. To exit, press the softkey button again.
4-5
DAC-2513EN-AB
4. Using The DDU-2300 AED
Post Use Procedures
After the DDU-2300 AED has been used on a patient, the unit should be cleaned following procedures in the
“Cleaning” section in Chapter 5 of this manual and prepared for the next use. The following steps should be
performed:
1. Connect a new pads package (check to make sure the package is not expired and package is not damaged).
2. Perform a self-test manually. Unit will report status at the end of the self-test. (Refer to the “Self-Tests”
section in Chapter 5 of this manual for instructions on how to run a manually initiated self-test.)
3. Turn off the unit by pressing the ON/OFF button.
4. Check to make sure that the Active Status Indicator is flashing green.
Rescue Mode Voice And Text Prompts
The following section provides brief descriptions of some of the voice and text prompts that the user will hear and
see in Rescue Mode.
General Prompts
Voice
Text
“Call for help”
Call For Help
Purpose: As soon as the DDU-2300 is turned on, the user will be prompted to “call for help.” This indicates
that the first step in a rescue should always be to contact professional emergency services. If another person is
available, the user should direct that person to call for help and then continue the rescue without delay.
“Pediatric mode”
Pediatric Mode
Purpose: This informs the user that child/infant pads are attached to the unit. Child/infant pads should only
be used if the patient is an infant or a child under the age of 8 or less than 55 pounds (25kgs.). For children 8
years or older or over 55 pounds (25kgs.) and for adults, adult pads should be used. Do not delay therapy to
determine exact age or weight.
“Training pads”
Training Pads
Purpose: This informs the user that training pads are attached to the unit. Training pads are used for training
purposes only and will not deliver a shock. In a rescue, immediately replace the training pads with defibrillation
pads.
“Powering off”
Powering Off
Purpose: This informs the user that the unit is turning off.
Pad Connection/Pad Application Related Prompts
Voice
Text
“Follow instructions to apply pads”
Follow Instructions
Purpose: This instructs the user to follow the AED prompts in order to apply the pads to the patient.
“Remove clothing from patient’s chest”
Remove Clothing
Purpose: This instructs the user to remove all clothing from patient’s chest. Pads must be applied to the
patient’s bare chest.
“Locate pads package in back of AED”
Locate Pads
Purpose: This helps the user locate the pads in the pad storage area, which is located on the back of the unit.
“Plug in pads connector”
Plug In Pads
Purpose: The DDU-2300 is unable to detect that the pads are plugged in. Check that the connector is fully
inserted into the unit. If the pads are properly plugged in, continue to follow audio and visual instructions.
“Tear open pads package”
Open Pads Package
Purpose: This instructs the user to tear open the pads package on the dotted line on the top of the package.
Once the package is open, the user will be able to remove the pads from inside the package.
4-6
DAC-2513EN-AB
4. Using The DDU-2300 AED
Pad Connection/ Pad Application Related Prompts (continued)
Voice
Text
“Peel pads from blue liner”
Peel Pads
Purpose: This instructs the user to peel each pad from the blue liner before placing the pads on the patient. Peel
the pads from the blue liner only when the pad is ready to be placed. Place the pads with the sticky side of the
pad on the patient’s bare skin.
“Apply pads to patient’s bare chest as shown”
Apply Pads to Patient
Purpose: The DDU-2300 AED has determined that the pads are not placed on the patient or not properly applied.
Place pads on the patient following instructions on the pads package. If the prompts continue, try replacing the
pads with a new set.
“Poor pad contact to patient”
“Press pads firmly”
Poor Pad Contact
Press Pads Firmly
Purpose: The pads are not making proper contact with the patient and the impedance is out of range for proper
ECG analysis and shock delivery. Check that the pads are properly placed and fully adhering to the patient and
that there are no air bubbles between the pads and the patient. If the pads are not sticking due to moisture, dry
the patient. If the pads are not sticking due to excessive hair, shave or clip excessive chest hair. If the prompts
continue, try replacing the pads with a new set.
“Check pads”
Check Pads
Purpose: The pads are making improper contact with the patient or touching each other and the impedance is out
of range for proper ECG analysis and shock delivery. Check that the pads are not touching each other and that the
patient is dry. If the prompts continue, try replacing the pads with a new set.
“Pausing for CPR”
Pausing for CPR
Purpose: If too long a period of time has passed, the user should stop attempting to resolve problems with the pads
and assess the condition of the patient. The user will be prompted to begin CPR.
“Replace pads”
Replace Pads
Purpose: The pads are making improper contact with the patient or touching each other and the impedance is
out of range for proper ECG analysis and shock delivery. If another set of pads is available, replace the pads,
otherwise check that the pads are properly placed and fully adhering to the patient. Make sure that the pads
are not touching each other. If the pads are not sticking due to moisture, dry the patient. If the pads are not
sticking due to excessive hair, shave or clip excessive chest hair. If the prompts continue, try replacing the pads
with a new set.
Motion/Interference Prompts
Voice
Text
“Stop motion”
Stop Motion
Purpose: The DDU-2300 AED has detected possible motion in the patient. Stop all patient motion, including
CPR, in response to this prompt.
“Stop interference”
Stop Interference
Purpose: The DDU-2300 AED has detected interference on the ECG signal. Eliminate any radio or electrical
sources of interference. Check the pads to make sure they are adhering properly to the patient. If the
i˜ÛˆÀœ˜“i˜ÌʈÃÊÛiÀÞÊ`ÀÞ]ʓˆ˜ˆ“ˆâiʓœÛi“i˜ÌÊ>ÀœÕ˜`Ê̅iÊ«>̈i˜ÌÊ̜ÊÀi`ÕViÊÃÌ>̈VÊ`ˆÃV…>À}iðÊ
“Pausing for CPR”
Pausing for CPR
Purpose: The user should stop attempting to resolve motion and/or interference problems and assess the
condition of the patient. The user will be prompted to begin CPR.
4-7
DAC-2513EN-AB
4. Using The DDU-2300 AED
Heart Rhythm Analysis Prompts
Voice
Text
“Analyzing heart rhythm”
“Analyzing”
Analyzing Rhythm
Analyzing
Purpose:Ê/…iÊ1‡ÓÎääÊʈÃÊ>V̈ÛiÞÊ>˜>Þ∘}Ê̅iÊ«>̈i˜Ì½ÃÊ
ÊÈ}˜>°ÊÊ/…iÊÊ܈ÊVœ˜Ìˆ˜ÕiÊ>˜>Þ∘}Ê՘̈Ê
ˆÌʅ>ÃÊ`iÌiÀ“ˆ˜i`Ê܅i̅iÀÊ>ÊÀ…Þ̅“ʈÃÊŜVŽ>LiʜÀʘœ˜‡Ã…œVŽ>LiʜÀʈvÊ>˜>Þ∘}ʈÃʈ˜ÌiÀÀÕ«Ìi`ÊvœÀÊܓiÊÀi>ܘ°ÊÊ
“Do not touch the patient”
Do Not Touch Patient
Purpose:Ê/…iÊ1‡ÓÎääÊʈÃÊÌÀވ˜}Ê̜Ê>˜>ÞâiÊ̅iÊ«>̈i˜Ì½Ãʅi>ÀÌÊÀ…Þ̅“°ÊÊ/…iʜ«iÀ>̜ÀÊŜՏ`ʘœÌÊ̜ÕV…Ê̅iÊ
patient. This prompt will be spoken at the beginning of the analysis period and also if motion or interference has
been detected.
“Analyzing interrupted”
Analyzing Interrupted
Purpose: The DDU-2300 AED has determined that accurate ECG analysis is not possible and has ceased
>˜>Þ∘}°ÊÊ/…iʜ«iÀ>̜ÀʈÃÊ«Àœ“«Ìi`Ê̜ÊÀi܏ÛiÊ̅iÊ«ÀœLi“°ÊÊ­,iviÀÊ̜Ê̅iʺMotion/Interference Prompts” and
the “Pad Connection/Pad Application Related Prompts” section in this chapter.) Once the problem is resolved,
the unit will enter analysis mode again.
“No shock advised”
No Shock Advised
Purpose: The DDU-2300 AED has determined that a shock is not required. The unit will NOT charge and the
SHOCK button will NOT be enabled. The user will be prompted to begin CPR.
“Shock advised”
Shock Advised
Purpose: The DDU-2300 AED has determined that a shock is recommended and the unit will begin charging in
anticipation of delivering a defibrillation shock.
Shock Related Prompts
Voice
Text
“Charging”
Charging
Purpose: The DDU-2300 AED has determined that a shock is recommended and is charging the unit in
anticipation of a defibrillation shock. Analysis will continue during this phase. A tone will sound to indicate
charging progress. If the unit detects a rhythm change to a non-shockable one, charging will abort and the user
will be prompted to begin CPR.
“Stand clear”
Stand Clear
Purpose: The DDU-2300 AED is charging and the operator and others should stand clear of the patient. Analysis
܈ÊVœ˜Ìˆ˜ÕiÊ`ÕÀˆ˜}Ê̅ˆÃÊ«…>ÃiÊ>˜`Ê>˜>Þ∘}Ê«Àœ“«ÌÃÊ܈ÊVœ˜Ìˆ˜ÕiÊ̜ÊLiÊ`ˆÃ«>Þi`°ÊÊÊ̜˜iÊ܈ÊÜ՘`Ê̜Ê
indicate charging progress. If the unit detects a rhythm change to a non-shockable one, charging will abort and
the user will be prompted to begin CPR.
“Press flashing SHOCK button”
Press SHOCK Button
Purpose: The DDU-2300 AED has fully charged, the heart rhythm analysis algorithm still indicates that a shock is
recommended, and the unit is ready to deliver a shock. The operator should press the SHOCK button to deliver
the shock. The SHOCK button will flash during this phase and will cancel after 30 seconds.
Important: The DDU-2300 AED will not automatically deliver a shock – the user must press the SHOCK button.
“Shock ‘x’ delivered”
Shock “x” Delivered
Purpose: The DDU-2300 AED has delivered the shock. The ‘x’ indicates the number of shocks that have been
delivered since the unit was turned on. After each shock, the AED will enter Post-Shock CPR mode.
(AHA 2005 Protocol)
“Shock cancelled”
Shock Cancelled
Purpose: The DDU-2300 AED has aborted shock mode. If while waiting for the SHOCK button to be pressed,
the unit detects a rhythm change to a non-shockable rhythm, the unit will cancel the shock. Also, if the SHOCK
button is not pressed within 30 seconds of the initial “press flashing SHOCK button” prompt, the unit will
automatically cancel the shock.
4-8
DAC-2513EN-AB
4. Using The DDU-2300 AED
Shock Related Prompts (continued)
Voice
Text
“SHOCK button not pressed”
Button Not Pressed
Purpose: After shock is advised, the DDU-2300 AED will prompt user to press the flashing shock button.
If after 30 seconds the shock button is not pressed, the DDU-2300 AED will give this prompt and immediately
go to CPR mode.
No Shock Required Prompts
Voice
Text
“No shock advised”
No Shock Advised
“It is safe to touch the patient”
OK to Touch Patient
Purpose: The DDU-2300 AED has determined that a shock is not required. The unit will not charge and the
SHOCK button will not be enabled. The user will be prompted to begin CPR.
“Check airway”
“Check breathing”
Check Airway
Check Breathing
Purpose: The user should check the condition of the patient in order to determine if it is appropriate to perform CPR.
CPR Prompts
Note: CPR breathing coaching prompts can be set through the Rescue options menu option listed on the AED
Main Menu screen. The factory default setting provides prompts for chest compressions only (no breathing).
Breathing prompts can be included either by changing the menu option (refer to the “CPR Breathing” section in
Chapter 6 of this manual) or by pressing a softkey button during rescue. (Refer to ”Rescue Breathing Options
Softkey Icon” section of this chapter.)
Voice
Text
“Begin CPR now”
Begin CPR Now
Purpose: This indicates that the user should begin performing CPR immediately. The unit will not monitor the
patient’s ECG rhythm during this CPR period.
“Give Compressions”
Give Compressions
Purpose: This indicates that the user should begin CPR compressions immediately. The unit will emit a beep at
the rate that compressions should be given.
“Continue”
“Continue for 1 minute ‘xx’ seconds”
Continue “xx” Seconds
Purpose: This indicates that the user should continue performing CPR. This phrase is spoken to let the user
know that the unit is still operating normally. The unit will not be monitoring the patient’s ECG rhythm during this
mandatory two minute CPR period. (AHA 2005 Protocol)
“Ending in 5, 4, 3, 2, 1”
Ending in “xx” Seconds
Purpose: This indicates that the user should prepare to finish performing CPR. This phrase is spoken during the last
several seconds of the CPR period to let the operator know that the unit is still operating normally and that the CPR
period is ending.
“Stop CPR”
“Stop Now”
Stop CPR
Stop Now
Purpose: This indicates that the CPR period has ended and the user should stop CPR. The unit will enter
˜>Þâiʜ`i°
“Do not touch the patient”
“Analyzing heart rhythm”
Do Not Touch Patient
Analyzing Rhythm
Purpose:Ê/…ˆÃʈ˜`ˆV>ÌiÃÊ̅iÊ՘ˆÌʅ>ÃÊi˜ÌiÀi`ʘ>Þâiʜ`iÊ>˜`ʈÃÊ«iÀvœÀ“ˆ˜}Ê>˜Ê
Ê>˜>ÞÈðÊ/…iÊÕÃiÀÊŜՏ`Ê
not touch the patient during the ECG analysis.
4-9
DAC-2513EN-AB
4. Using The DDU-2300 AED
CPR Coaching Help Prompts
Voice
Text
“Place hands”
Place Hands
Purpose: This reminds the user of the correct placement of hands for CPR.
“Press”
“Compress Chest”
Press...
Compress Chest
Purpose: This reminds the user to perform CPR compressions.
“Tilt head back”
“Pinch nose”
“Give rescue breaths”
Tilt Head Back
Pinch Nose
Give Breaths
Purpose: This guides the user to prepare the patient for rescue breaths and to give breaths.
“Breathe”
Give “x” Breaths
Purpose: This instructs the user to give rescue breaths. Each time the instruction is given, the user should give the
patient a rescue breath.
Operational Environment
The Defibtech AED is designed to operate in a wide range of environmental conditions. To ensure the reliability and
safety of the AED in a given environment, refer to the “Environmental” section in Chapter 9 of this manual for a
detailed list of approved environmental conditions.
4-10
DAC-2513EN-AB
5
Maintenance And Troubleshooting
This chapter describes the maintenance and troubleshooting procedures for the DDU-2300 AED. The self-tests that
are performed by the device are described along with the frequency and nature of routine maintenance for which the
owner/operator is responsible. A troubleshooting guide is provided to help diagnose user serviceable problems.
The DDU-2300 AED contains no user serviceable parts.
Routine Unit Maintenance
Although the DDU-2300 AED is designed to be very low maintenance, simple maintenance tasks must be performed
by the owner/operator on a regular basis to ensure the unit’s dependability.
Monthly
After Each Use
U
U
U
Check that the Active Status Indicator (ASI) is flashing green
U
U
Check the condition of the unit and accessories
U
Run manually initiated self-test
U
Replace pads
U
Action
5. Maintenance And
Troubleshooting
Daily
Check pad and battery pack expiration dates
U
Check the DDC card, if one was installed
Note: If the unit has been dropped, mishandled, or abused, a manually initiated self-test should be performed.
Checking The Active Status Indicator
The Active Status Indicator (“ASI”) is located to the right of the ON/OFF button of the DDU-2300 AED and indicates
the operational readiness state of the unit. The ASI will periodically flash green to indicate that the unit is ready for
use. Ready for use means that the DDU-2300 has passed the most recent self test (scheduled or user initiated). If
it is flashing red, solid red, or not flashing at all, the AED needs service. Anytime the ASI is flashing red, the unit will
periodically emit two “beeps” to call attention to it.
If the ASI is not flashing at all, the most likely cause is that the battery pack needs to be replaced. (Refer to the
“Installing And Removing The Battery Pack” section in Chapter 3 of this manual.) Once the battery pack has
been replaced with a fresh battery pack, the ASI should once again flash green. If it still does not flash green after
inserting a new battery pack, the DDU-2300 AED is non-operational and may need servicing. Call Defibtech for
service. (Refer to the “Contacts” section in Chapter 12 of this manual.)
If the ASI is flashing red, turn the DDU-2300 AED on. If the unit does not turn on or does not speak, the AED is nonoperational and requires servicing. If the unit does turn on, then turn the unit off and the voice prompts will indicate
the nature of the problem.
UÊ Flashing Green: DDU-2300 AED is OFF and ready for use.
UÊ Solid Green: DDU-2300 AED is ON and ready for use.
UÊ Flashing or Solid Red: DDU-2300 AED needs immediate service. Refer to the
“Troubleshooting” section in Chapter 5 of this manual or call Defibtech for service
Active
Status
Indicator
UÊ No Flashing Light: DDU-2300 AED needs immediate service. Refer to the
“Troubleshooting” section in Chapter 5 of this manual or call Defibtech for service.
5-1
DAC-2513EN-AB
Checking The AED Status Using The AED Status Screen
You may also check the status of the unit when it is off by pressing the center softkey button to enter Maintenance
Mode and display the AED Status Screen.
The AED Status Screen is used to provide a quick overview of the
DDU-2300 AED’s status and to display select information without
turning the unit on in Rescue Mode.
With the AED off, press and release the CENTER softkey button
to display the AED Status Screen. The AED Status Screen will be
displayed for a short period of time.
If the unit does not turn on, check to make sure a good battery pack
is installed. (Refer to the “Troubleshooting” section in this chapter.)
From the AED Status Screen you can enter Maintenance Mode by pressing
the softkey button to the right of the tool icon.
5. Maintenance And
Troubleshooting
Note: If the unit requires service, the AED Status Screen will present information about the problem to the user.
The user should follow the text prompts to address the condition requiring attention.
Card Application Softkey Icon: If an application is present on an inserted Defibtech
Data Card (DDC card), a card icon will also appear next to the center softkey button.
Pressing this button will load and run the application contained on the card.
Maintenance Related Voice Prompts
Voice
Text
“Power-on test failed”
“Service code ‘xxxx’ ”
Power On Test Failed
Service Code “xxxx”
Purpose: This indicates that the DDU-2300 AED has failed the power-on self-test and is non-operational
and needs servicing. The code number will indicate to service personnel the type of problem the unit is
experiencing.
“Battery pack self-test failed, service code ‘xxxx’ ”
Battery Pack Self-Test Failed, Service Code“xxxx”
Purpose: This indicates that the DDU-2300 AED’s battery pack is non-operational and needs servicing. The code
number will indicate to the service personnel the type of problem that the unit is experiencing.
“Service code ‘xxxx’ ”
Service Code “xxxx”
Purpose: The DDU-2300 AED will report this message when it powers off, indicating a service code that was
previously detected.
“Service required”
Service Required
Purpose: This indicates that the DDU-2300 AED has detected an internal error, is non-operational, and
needs servicing.
“Battery low”
Battery Low
Purpose: This indicates that the battery pack capacity is low and that the battery pack should be replaced soon.
The AED will still be able to deliver at least six defibrillation shocks the first time this message is spoken.
“Replace battery now”
Replace Battery Now
Purpose: This indicates that the battery pack is almost discharged and that the AED may not be able to deliver
defibrillation shocks. Replace the battery pack immediately.
“Pads missing”
Pads Missing
Purpose: This indicates that the unit did not detect connected pads during a self-test.
“Pads expired”
Pads Expired
Purpose: This indicates that pads have expired. Replace the pads immediately.
5-2
DAC-2513EN-AB
Checking The Condition Of The Unit And Accessories
Inspect the unit for dirt and contamination, especially in the pads connector socket and around the battery pack
opening. (Refer to the “Cleaning” section in this chapter of this manual for guidance on cleaning your AED.)
Inspect the unit display screen for damage. Look for cracks or other signs of damage on the case, especially near
the connector socket.
vÊ >˜ÞÊ VÀ>VŽÃÊ œÀÊ œÌ…iÀÊ Ãˆ}˜ÃÊ œvÊ `>“>}iÊ >ÀiÊ ÛˆÃˆLi]Ê Ài“œÛiÊ Ì…iÊ Ê vÀœ“Ê ÃiÀۈViÊ >˜`Ê Vœ˜Ì>VÌÊ >˜Ê >Õ̅œÀˆâi`Ê
service center.
Replacing Pads
The Defibtech defibrillation pads are intended for one time use only. The pads must be replaced after each use or
if the package has been damaged.
The DDU-2300 AED defibrillation pads are supplied in a sealed pouch with the connector and part of the cable
exposed. The DDU-2300 AED is designed to be stored with the electrode cable already installed. This allows the
pads to be stored in a pre-connected state for rapid deployment during an emergency.
5. Maintenance And
Troubleshooting
DO NOT remove the defibrillation pads from the sealed
package until the pads are to be used. The packaging
should be opened only immediately prior to use, otherwise
CAUTION
the pads may dry out and become non-functional.
!
STEP 1 – Inspect The Pads – First, check to ensure that the pads package has not expired. Do not use pads
past their expiration date. Discard expired pads. Next, check to ensure that the pads package has not been torn,
opened, or damaged. Dispose of the pads if the package is open or damaged. Inspect the pads cable and replace
pads if any nicks, cuts, or broken cables are found.
STEP 2 – Connect The Pads To The Unit – Insert the connector end of the defibrillation pad cable into the pads
connector socket on the top-left corner of the DDU-2300 AED as shown. Press the pads connector in firmly until
it is fully seated in the unit.
STEP 3 – Store The Pads In Back Of Unit – The pads package can then be stored in the pad storage slot in the
back of the DDU-2300 AED. After connecting the pads connector to the unit, push the pads package, rounded
end first, with the pictures on the package facing up and out, into the pad holder compartment on the back of the
AED. When the pads package is fully inserted, press the pad cable into the groove in the back of the unit to hold
it in place and tuck any excess cable behind the pads package.
!
CAUTION
The pads are intended for one time use only and must
be discarded after use or if the package has been opened.
5-3
DAC-2513EN-AB
Checking Pad And Battery Pack Expiration Dates
It is important to check the expiration dates of the pads and battery pack. The expiration date of the pads package
is printed on the outside of the sealed package. The expiration date of the battery pack is printed on the label on the
battery pack. Once an accessory is past its expiration date, it should be taken out of service and replaced as soon
as possible. Follow the instructions in the “Installing And Removing The Battery Pack” and “Connecting The Pads”
sections of Chapter 3 in this manual to replace an expired part with an un-expired part. Pads should be discarded.
Battery packs should be appropriately recycled.
You may also check the status of pads and battery pack when the unit is off by pressing the center softkey button to
display the AED Status Screen and enter Maintenance Mode. (Refer to the “AED Status Screen” section in Chapter
6 of this manual.)
Checking The DDC Card, If One Is Installed
Each time the DDU-2300 AED is used, an event file is created on the DDC card (if installed). If the unit was used to
treat a patient, the DDC card in the unit should be removed and provided to the patient’s care provider. A new DDC
card should be installed before the next use.
5. Maintenance And
Troubleshooting
To remove the DDC card, ensure the AED is OFF. Locate the data card/USB port access door on the right-hand
side of the unit. Open the data card/USB port access door by slightly pushing and then sliding the door down to
release the latch. The door will spring open. To remove the DDC card, press the card in as fas as it will go, and then
release. Upon release, the DDC card will be partially ejected and can be removed by pulling the DDC card the rest
of the way out.
To install a new DDC card, insert the DDC card into the thin slot in the side of the AED centered above the USB port
opening, notched end first, label side up, until it clicks into place. The card should be flush with the surface of the
slot. If the card does not push in all the way, it may have been inserted upside down. In that case, remove the card,
flip it over, and try inserting it again.
Close the data card/USB port access door by closing and then pushing up the door until the door latch engages.
Note: A DDC card is not required for the DDU-2300 AED to operate. Even if a DDC card is not installed, the unit
will still record basic essential information internally. The AED will still operate properly even after a “replace data
card” message.
5-4
DAC-2513EN-AB
Self-Tests
The DDU-2300 AED provides for both automatic and manually initiated self-tests. These self-tests test various
components of the AED, including the system controls, battery pack condition, charge/discharge functions, and
measurement and signal acquisition functions.
Automatic Unit Self-Tests
Every time the unit is turned on, power-on self-tests are performed to test the basic operation of the unit.
The unit also performs daily, weekly, monthly, and quarterly self-tests automatically (without any intervention from
the operator) to check the integrity of the unit’s hardware and software. The unit will also perform a battery insert
self-test when the battery pack is inserted.
Manual Self-Tests
Manually initiated self-tests may be run at any time by the user to test the DDU-2300 AED’s systems, including the
charging and shocking functions (the shock is internally dissipated; i.e., no voltage will be present at the pads).
5. Maintenance And
Troubleshooting
To run a manually initiated AED test, the unit must be put into Maintenance Mode of operation. (Refer to the “AED
Maintenance Screen” section in Chapter 6 of this manual for detailed information on performing these self-tests.)
Note: Running a manually initiated self-test will consume approximately one shock’s worth of energy from the battery.
Cleaning
After each use, clean the DDU-2300 AED of any dirt or contaminants on the case and connector socket.
The following are important guidelines to be adhered to when cleaning the device:
UÊ/…iÊL>ÌÌiÀÞÊ«>VŽÊŜՏ`ÊLiʈ˜ÃÌ>i`Ê܅i˜ÊVi>˜ˆ˜}Ê̅iÊ1‡ÓÎää°
UʜʘœÌʈ““iÀÃiÊ̅iÊ1‡ÓÎääʈ˜ÊyՈ`ÃʜÀÊ>œÜÊyՈ`ÃÊ̜Êi˜ÌiÀÊ̅iÊ՘ˆÌ°Ê
UʜʘœÌÊëÀ>ÞÊVi>˜ˆ˜}Ê܏Ṏœ˜ÃÊ`ˆÀiV̏Þʜ˜Ê̅iÊ՘ˆÌʜÀʈÌÃÊVœ˜˜iV̜Àð
UʜʘœÌÊÕÃiÊ>LÀ>ÈÛiʓ>ÌiÀˆ>ÃʜÀÊÃÌÀœ˜}Ê܏Ûi˜ÌÃÊÃÕV…Ê>ÃÊ>Vi̜˜iʜÀÊ>Vi̜˜iÊL>Ãi`ÊVi>˜ˆ˜}Ê>}i˜Ìð
UÊÊ/œÊ ܈«iÊ Ì…iÊ 1‡ÓÎääÊ V>ÃiÊ Vi>˜]Ê ÕÃiÊ >Ê ÃœvÌÊ VœÌ…Ê `>“«i˜i`Ê ÜˆÌ…Ê œ˜iÊ œvÊ Ì…iÊ vœœÜˆ˜}Ê ÀiVœ““i˜`i`Ê
cleaning agents:
– Soapy water
– Ammonia based cleaners (e.g., Windex®, Formula 409®, Fantastik®)
– Hydrogen peroxide
– Isopropyl alcohol (70 percent solution)
– Chlorine bleach (30 ml/liter water)
UÊʘÃÕÀiÊ̅>ÌÊ̅iÊVœ˜˜iV̜ÀÊÜVŽiÌʈÃÊVœ“«iÌiÞÊ`ÀÞÊLivœÀiÊÀiˆ˜ÃÌ>ˆ˜}Ê̅iÊ«>`ÃÊV>Li°ÊÊvÌiÀÊVi>˜ˆ˜}]Ê>œÜÊ
the unit to completely dry. Before returning it to service, always turn the unit on for a few seconds. If the unit
detects a problem, a service code voice prompt will be heard. Otherwise, turn the unit off.
Storage
The DDU-2300 AED should be placed in a readily accessible location in an orientation where the Active Status
Indicator next to the ON/OFF button can be easily seen. In general, the unit should be stored in clean, dry and
moderate temperature conditions. Make sure that the environmental conditions of the storage location are within
the ranges detailed in the “Environmental” section in Chapter 9 of this manual.
5-5
DAC-2513EN-AB
Operator’s Checklist
The following checklist may be used as the basis for an Operator’s Checklist. The table should be copied and
filled out as recommend by the schedule in the “Routine Maintenance” section of this chapter. As each item is
completed it should be checked off.
Defibtech DDU-2300 Operator’s Checklist
Defibtech DDU-2300 Serial Number: ___________________________________________________
Defibtech DDU-2300 Location: __________________________________________________________
5. Maintenance And
Troubleshooting
Date:
Check unit and accessories for damage, dirt, and
contamination. Clean or replace as necessary.
Check that spare battery pack and pads are available.
Check that battery pack and pads are not past expiration dates.
Check that the ASI is flashing green.
Comments:
Inspection by: (initials or signature)
5-6
DAC-2513EN-AB
Troubleshooting
The following table lists the symptoms, the possible causes, and the possible corrective actions for common
problems. Refer to the other sections of the user manual for detailed explanations on how to implement the
corrective actions. If the unit continues to be non-functional, refer the unit for servicing. (Refer to Chapter 12 of this
manual for contact information.)
Symptom
Possible Cause
Corrective Action
Insert battery pack
Battery pack depleted or
needs servicing
Replace battery pack
or call for service.
Unit needs servicing
Call for service
Battery pack depleted
Replace battery pack
Unit needs servicing
Call for service
Unit needs servicing
Go to AED Status
Screen by pressing the
CENTER softkey button
or call for service
Battery pack non-functional
Replace battery pack
Defibrillation pads are not
pre-connected to unit
Connect defibrillation
pads to unit
Defibrillation pads
or battery pack expired
Replace expired
component
Battery pack not inserted
Insert battery pack
Battery pack is low
or needs servicing
Replace battery pack
or call for service.
Unit needs servicing
Call for service
Power on self-test failed,
service code “xxxx”
Unit needs servicing
Record code number
and call for service
Battery pack self-test failed,
service code “xxxx”
Battery pack needs servicing
Record code number
and call for service
Service required
Unit needs servicing
Call for service
“Replace battery now” prompt
Battery pack capacity is
critically low
Unit may not deliver a
shock, replace battery
pack immediately
“Battery low” prompt
Battery pack capacity is
getting low
Replace battery pack as
soon as possible
Battery pack depleted
Replace battery pack
Battery pack not
inserted properly
Make sure battery pack
is oriented correctly and
fully inserted
Unit needs servicing
Call for service
Pads connector not plugged in
Plug in pads connector
Pads connector broken
Replace pads
Unit’s connector broken
Call for service
Unit will not turn on
Unit immediately turns off
ASI flashes red and/or unit makes
periodic ”beep” sound.
ASI does not flash at all
Display screen does not work
“Plug in pads connector” prompt
5. Maintenance And
Troubleshooting
Battery pack not inserted
5-7
DAC-2513EN-AB
Troubleshooting (continued)
Symptom
Possible Cause
Corrective Action
Place pads on patient
Pads not making good
connection to patient
Check pad connection
to patient
Pads or pad cable damaged
Replace pads
Dry pads
Replace pads
“Poor pad contact to patient” or
“Press pads firmly” prompt
Partial pad connection
Check that pads are
placed securely on
patient
“Check pads” prompt
Pads touching
Separate pads and place
correctly on patient
“Stop motion” prompt
Patient motion has
been detected
Stop patient motion
“Stop interference” prompt
External interference
has been detected
Stop external
interference
“Analyzing interrupted” prompt
Motion or interference detected
Stop motion or
interference
Patient’s ECG rhythm changed
No action necessary
Shock button not pressed
within 30 seconds
Press shock button
within 30 seconds
Low battery –
insufficient to charge
Replace battery pack
Bad pad to patient connection
Check that pads are
placed securely on
patient
Dry pads
Replace pads
DDC card is full
Replace DDC card with
a card that is not full
DDC card has failed
Replace DDC card
Pads not connected to unit
Make sure pads
connector is oriented
correctly and fully
inserted into unit
“Apply pads to patient’s bare
chest as shown” prompt
“Shock cancelled” prompt
“Replace data card” prompt
“Pads missing” prompt
5. Maintenance And
Troubleshooting
Pads not connected to patient
Repair
The DDU-2300 AED contains no user serviceable parts. If the unit needs servicing, call Defibtech. (Refer to
Chapter 12 of this manual for contact information.)
5-8
DAC-2513EN-AB
6
Maintenance Mode
Overview
Maintenance Mode for the Defibtech DDU-2300 permits the user to perform maintenance-related actions such as
viewing unit information, initiating unit self-tests, changing unit parameters, downloading rescue data, and upgrading
software.
Maintenance Mode is navigated through a series of screens, menus, and menu options. In Maintenance Mode, the
softkey buttons located directly to the right of the display screen are used to scroll through and select menu options.
When a softkey icon (for example, an arrow) appears on the display screen directly to the left of a softkey button, the
softkey button is functional for that screen. If a softkey icon is not displayed on the screen, then the corresponding
softkey button has no functionality for that screen.
Note: While the unit is in Maintenance Mode, it cannot perform a rescue. Maintenance Mode allows the user to go
directly to Rescue Mode by selecting the Rescue now option. The Rescue now option appears at the top of every
screen/menu when the unit is in Maintenance Mode. The user can also exit Maintenance Mode at any time and go to
Rescue Mode by pressing the ON/OFF button to turn the unit off and then immediately pressing the ON/OFF button
again to turn the unit back on.
The Display Screen (During Maintenance Mode):
Title
Top
Menu
Options
Center
Softkey
Buttons
Bottom
6. Maintenance Mode
Title – The name of the menu appears in a banner across the top of the screen.
Softkey Buttons – Softkey buttons are used to scroll through menus and to select menu options.
Softkey Button Icons – The presence of an on-screen icon located directly to the left of a softkey
button indicates that the softkey is functional.
Menu Options – The list of options available from a particular menu. Use the top or bottom
softkey buttons to scroll through the list of options. Select an option by
pressing the center softkey button once an option is highlighted (with a box).
Navigation (in Maintenance Mode)
The three softkey buttons located to the right of the display screen are used to navigate in Maintenance Mode.
Typical functions of the softkey buttons are the following:
Top softkey button: Scroll up
Center softkey button: Select highlighted option
Bottom softkey button: Scroll down
When a menu option is highlighted and then selected (typically by pressing the CENTER softkey button), either
another screen will be displayed with additional menu options or an action will be performed.
Exiting Maintenance Mode
To exit Maintenance Mode and return to Rescue Mode, scroll to and select Rescue now or simply turn the unit off
and then back on.
To exit Maintenance Mode and turn the unit off, scroll to and select Turn AED off or simply turn the unit off by
pressing the ON/OFF button.
6-1
DAC-2513EN-AB
Entering Maintenance Mode
Before You Begin: Make certain the DDU-2300 is turned off and a battery pack is installed.
STEP 1 – Press and release the CENTER softkey button.
Result – The unit will turn on and display the AED Status
Screen for short period of time.
If the unit does not turn on, check to make sure a good battery pack
is installed. (Refer to the “Troubleshooting” section in Chapter 5
of this manual.)
STEP 2 – Press the BOTTOM softkey button (to the right of the tool icon). Note: If the bottom softkey button
is not pressed within a short period of time, the unit will automatically turn off.
Result – The unit will enter Maintenance Mode and display the AED Main Menu screen.
AED Main Menu Screen
The AED Main Menu Screen allows the user to view the status of the AED, perform maintenance functions, change
AED options, and access help topics. All maintenance functions are accessed through the AED Main Menu Screen.
The user may select from the following options using the softkey buttons:
Rescue now – Puts device in Rescue Mode
AED status – Displays current AED status information
AED maintenance – Displays AED Maintenance Menu screen
AED options – Displays AED Options Menu screen
Rescue options – Displays Rescue Options Menu screen
6. Maintenance Mode
Help topics – Displays Help Topics screen
Turn AED off – Turns unit off
When the user selects “Rescue now,” the unit will exit Maintenance Mode and proceed directly to Rescue Mode.
The other menu options will perform different functions and are described in detail below.
AED Status Screen
The AED Status Screen displays unit-specific data such as current status, battery pack charge state, battery pack
expiration date, pads expiration date, unit serial number, battery pack serial number, and software version number.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to AED status:
AED Main Menu ➜ AED status
Note: When the unit is turned off, the AED Status Screen can also be accessed by pressing the center softkey button.
What it does: The unit will display the AED Status Screen. This is an informational screen only; no action is taken
by the AED.
To exit: To exit the AED Status Screen, press and release the BOTTOM softkey button.
The unit will exit the AED Status Screen and return to the AED Main Menu screen.
6-2
DAC-2513EN-AB
AED Maintenance Screen
The AED Maintenance screen allows the user to select such options as AED tests, software upgrades, data
backups, and data card functions.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to AED Maintenance:
AED Main Menu ➜ AED maintenance
What it does: The unit will display the AED Maintenance menu screen. This screen will allow the user to navigate
further to perform various maintenance tasks:
UÊ*iÀvœÀ“ÊÊÌiÃÌ
UÊ1«}À>`iÊ
UÊ/À>˜ÃviÀÊ`>Ì>Ê̜ÊV>À`
UʜÀ“>ÌÊ`>Ì>ÊV>À`
UÊ,՘Ê>««ˆV>̈œ˜ÊvÀœ“ÊV>À`
To exit: Use the TOP or BOTTOM softkey buttons to scroll to and highlight the selection “Go to main menu.”
Press the CENTER softkey button. The unit will exit the AED Maintenance screen and return to the AED Main
Menu screen.
➜ Perform AED Test
Perform AED test will initiate a system hardware and software self-test.
Note: Running manually initiated AED tests will consume approximately one shock’s worth of battery life.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Perform AED test:
AED Main Menu ➜ AED maintenance ➜ Perform AED test
6. Maintenance Mode
What it does: When the user selects “Perform AED test” selection and presses the CENTER softkey
button, the unit will begin performing the self-test procedure:
The unit says: “Performing AED test“
The unit displays: Testing AED
The unit will then prompt the user to “Press flashing shock button.” Continue to follow the directions
until the test is complete. Once the AED test is complete the unit will verbally and visually report the
status of the AED. The information will be displayed in a pop-up window. The user must press any
softkey button to confirm the test status and return to the AED Maintenance screen.
If self-test passes: The unit will speak and display: “AED OK”
If self-test fails: The unit will display an error screen with text prompts providing instructions to
address the condition.
Note: If self-test fails, the user should follow the text prompts to address the condition requiring
attention or refer to the “Troubleshooting” section in Chapter 5 of this manual.
To exit: Press any softkey button. The self-test status pop-up window will clear and the display will
return to the AED Maintenance menu screen.
➜ Upgrade AED
The Upgrade AED menu selection is a method of performing a unit upgrade and will activate the
system upgrade procedure from a Defibtech Data Card (DDC card) containing an upgrade application.
Note: Upgrades can also be performed directly from the AED Status Screen if an upgrade card is
present when the AED Status Screen mode is launched.
Before you begin: Be sure the unit is in Maintenance Mode.
6-3
DAC-2513EN-AB
To enter: Navigate to Upgrade AED:
AED Main Menu ➜ AED maintenance ➜ Upgrade AED
What it does: If an upgrade data card is present, the unit will start performing the upgrade process.
Follow any prompts and instructions that the upgrade application provides.
!
WARNING
Do not turn off the unit or remove the
battery pack or data card until the operation
is complete.
Note: If the DDC card is not inserted, the unit will speak and display “Data Card Missing.”
(Refer to the “Installing The Defibtech Data Card (DDC card)” section in Chapter 3 of this manual.)
Press any softkey button to acknowledge the message and then install a Defibtech Data Card (DDC card).
To exit: When the unit finishes performing the AED upgrade, follow the displayed and
spoken instructions.
➜ Transfer Data To Card
Transfer data to card will initiate a data transfer from the DDU-2300 AED to a Defibtech Data Card
(DDC card) inserted in the device. Internal event data and device history is written to the DDC card.
Before you begin: Be sure the unit is in Maintenance Mode. Be sure a DDC Card is installed in
the device. (Refer to the “Installing The Defibtech Data Card (DDC card)” section in Chapter 3 of
this manual.)
To enter: Navigate to Transfer data to card:
AED Main Menu ➜ AED maintenance ➜ Transfer data to card
What it does: The unit begins transferring rescue data to the card:
The unit says: “Transferring data to data card”
The unit displays: Transferring Data
!
WARNING
6. Maintenance Mode
The unit will finish transferring data to the data card and will speak and display:
“data transfer complete”
Do not turn off the unit or remove the
battery pack or data card until the operation
is complete.
Note: If a data card is not inserted, the unit will speak and display “Data Card Missing.”
(Refer to the “Installing The Defibtech Data Card (DDC card)” section in Chapter 3 of this manual.)
To exit: When the unit finishes transferring the data to the data card, it will automatically return to the
AED Maintenance menu screen.
➜ Format Data Card
Format data card is a maintenance tool used to repair corrupted cards. It is unnecessary to perform
this step on cards purchased with your DDU-2300 AED.
!
This step will erase all data on the data card!
WARNING
Before you begin: Be sure the unit is in Maintenance Mode. Be sure a Defibtech Data Card (DDC card)
is currently installed in the device. (Refer to the “IInstalling The Defibtech Data Card (DDC card)” section
in Chapter 3 of this manual.)
To enter: Navigate to Format data card:
AED Main Menu ➜ AED maintenance ➜ Format data card
6-4
DAC-2513EN-AB
What it does: The unit will format the DDC card that is inserted in the AED:
The unit says: “ Formatting data card”
The unit displays: Formatting Data Card
When unit finishes formatting the DDC card, the unit will return to the menu.
!
WARNING
Do not turn off the unit or remove the
battery pack or data card until the operation
is complete.
Note: If the data card is not inserted, the unit will speak and display “Data Card Missing.”
(Refer to the “Installing The Defibtech Data Card (DDC card)” section in Chapter 3 of this manual.)
To exit: When the unit finishes formatting the data card, it will automatically return to the AED
Maintenance menu screen.
!
CAUTION
Using non-Defibtech Data Cards
(DDC cards) may damage the unit
and will void the warranty.
➜ Run Application From Card
Run application from card will initiate a card application on the Defibtech Data Card (DDC card).
The most typical application is a software upgrade.
Before you begin: Be sure the unit is in Maintenance Mode. Be sure a DDC Card, with a card
application, is installed in the device. (Refer to the “Installing The Defibtech Data Card (DDC card)”
section in Chapter 3 of this manual.)
To enter: Navigate to Run application from card:
AED Main Menu ➜ AED maintenance ➜ Run application from card
WARNING
6. Maintenance Mode
!
Do not turn off the unit or remove the
battery pack or data card until the operation
is complete.
Note: If a data card is not inserted, the unit will speak and display “Data Card Missing.”
Refer to the “Installing The Defibtech Data Card (DDC card)” section in Chapter 3 of this manual.)
To exit: When the unit finishes performing the application, follow the displayed and
spoken instructions.
AED Options Screen
To manually configure AED options such as time, date, volume, and audio recording, select the AED options
from the AED Main Menu screen.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to AED Options:
AED Main Menu ➜ AED options
What it does: The unit will display the AED Options menu screen. This screen will allow the user to modify the
following user changeable parameters:
UÊ-ÞÃÌi“Ê̈“i
UÊ-ÞÃÌi“Ê`>Ìi
UÊ6œÕ“iʏiÛi
UÊÕ`ˆœÊÀiVœÀ`ˆ˜}
To exit: Using the TOP or BOTTOM softkey buttons, scroll to and highlight the selection
Go to main menu. Press the CENTER softkey button. The unit will exit the AED Options
menu screen and return to the AED Main Menu.
6-5
DAC-2513EN-AB
➜ System Time
The System time option allows the user to set the time of the internal AED clock.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to System time:
AED Main Menu ➜ AED options ➜ System time
What it does: The System time option allows the user to set the time of internal AED clock (using the
24 hour clock). Once the System time option is selected, press the CENTER softkey button to enter
set time mode:
The hours selection will be highlighted green:
Ê
UÊ*ÀiÃÃÊ̅iÊ/"*ʜÀÊ"//"ÊÜv̎iÞÊLÕÌ̜˜ÃÊ̜Ê>`ÕÃÌÊ̅iʅœÕÀÃÊ̜Ê̅iÊ`iÈÀi`Ê̈“i°
Ê
UÊ*ÀiÃÃÊ̅iÊ
/,ÊÜv̎iÞÊLÕÌ̜˜Ê̜Ê>VVi«ÌÊ̅iʅœÕÀÃÊÃiÌ̈˜}°
The minutes selection will be highlighted green:
Ê
UÊ*ÀiÃÃÊ̅iÊ/"*ʜÀÊ"//"ÊÜv̎iÞÊLÕÌ̜˜ÃÊ̜Ê>`ÕÃÌÊ̅iʓˆ˜ÕÌiÃÊ̜Ê̅iÊ`iÈÀi`Ê̈“i°
Ê
UÊ*ÀiÃÃÊ̅iÊ
/,ÊÜv̎iÞÊLÕÌ̜˜Ê̜Ê>VVi«ÌÊ̅iʓˆ˜ÕÌiÃÊÃiÌ̈˜}°
The seconds selection will be highlighted green:
Ê
UÊ*ÀiÃÃÊ̅iÊ/"*ʜÀÊ"//"ÊÜv̎iÞÊLÕÌ̜˜ÃÊ̜Ê>`ÕÃÌÊ̅iÊÃiVœ˜`ÃÊ̜Ê̅iÊ`iÈÀi`Ê̈“i°
Ê
UÊ*ÀiÃÃÊ̅iÊ
/,ÊÜv̎iÞÊLÕÌ̜˜Ê̜Ê>VVi«ÌÊ̅iÊÃiVœ˜`ÃÊÃiÌ̈˜}°
The time is now set and the user may use TOP or BOTTOM softkey buttons to navigate to other
choices in the menu.
Note: The factory default setting of the internal AED clock is Universal Time (GMT).
➜ System Date
The System date option allows the user to set the date of the internal AED clock.
Before you begin: Be sure the unit is in Maintenance Mode.
6. Maintenance Mode
To enter: Navigate to System date:
AED Main Menu ➜ AED options ➜ System date
What it does: The System date option allows the user to set the date of internal AED clock.
Once the System date option is selected, press the CENTER softkey button to enter set
date mode:
The year selection will be highlighted green:
Ê
UÊ*ÀiÃÃÊ̅iÊ/"*ʜÀÊ"//"ÊÜv̎iÞÊLÕÌ̜˜ÃÊ̜Ê>`ÕÃÌÊ̅iÊÞi>À°
Ê
UÊ*ÀiÃÃÊ̅iÊ
/,ÊÜv̎iÞÊLÕÌ̜˜Ê̜Ê>VVi«ÌÊ̅iÊÞi>ÀÊÃiÌ̈˜}°
The month selection will be highlighted green:
Ê
UÊ*ÀiÃÃÊ̅iÊ/"*ʜÀÊ"//"ÊÜv̎iÞÊLÕÌ̜˜ÃÊ̜Ê>`ÕÃÌÊ̅iʓœ˜Ì…°
Ê
UÊ*ÀiÃÃÊ̅iÊ
/,ÊÜv̎iÞÊLÕÌ̜˜Ê̜Ê>VVi«ÌÊ̅iʓœ˜Ì…ÊÃiÌ̈˜}°
The day selection will be highlighted green:
Ê
UÊ*ÀiÃÃÊ̅iÊ/"*ʜÀÊ"//"ÊÜv̎iÞÊLÕÌ̜˜ÃÊ̜Ê>`ÕÃÌÊ̅iÊ`>Þ°
Ê
UÊ*ÀiÃÃÊ̅iÊ
/,ÊÜv̎iÞÊLÕÌ̜˜Ê̜Ê>VVi«ÌÊ̅iÊ`>ÞÊÃiÌ̈˜}°
The date is now set and the user may use TOP or BOTTOM softkey buttons to navigate to other
choices in the menu.
Note: The factory default setting of the internal AED clock is Universal Time (GMT).
6-6
DAC-2513EN-AB
➜ Volume Level
The 6œÕ“iʏiÛi option allows the user to set AED audio output to high, medium, or low volume.
Changing the volume level will not change the volume of the Active Status Indicator “beep.”
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to 6œÕ“iʏiÛi:
AED Main Menu ➜ AED options ➜ 6œÕ“iʏiÛi
What it does: The 6œÕ“iʏiÛiÊoption allows the user to set AED audio to high, medium or low
volume. Once the 6œÕ“iʏiÛi option is selected, use the TOP and BOTTOM softkey buttons to
cycle through the different volume settings. When the desired volume selection is chosen, press
the CENTER softkey button to set that volume level. The AED will now use that volume setting for
all audio (except the volume of the Active Status Indicator “beep”). The user may use the TOP or
BOTTOM softkey buttons to navigate to other choices in the menu.
Note: The factory default setting of the volume level is “high.”
➜ Audio Recording
The Audio recording option allows the user to enable or disable recording of event audio data to a
Defibtech Data Card (DDC card).
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Audio recording:
AED Main Menu ➜ AED options ➜ Audio recording
What it does: The Audio recording option allows the user to enable/disable recording of event audio
data. Once the Audio recording option item is selected, use the TOP and BOTTOM softkey buttons
to select either enabled or disabled settings. When the desired selection is chosen, press the CENTER
softkey button to set the feature. The AED will now use that audio recording setting. The user may
use the TOP or BOTTOM softkey buttons to navigate to other choices in the menu.
6. Maintenance Mode
Note: The factory default setting of the audio recording is “disabled.”
Rescue Options Screen
To manually configure rescue options such as Rescue protocol and CPR breathing, select the Rescue options
from the AED Main Menu screen.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Rescue Options:
AED Main Menu ➜ Rescue options
What it does: The unit will display the Rescue Options menu screen. This screen will allow the user to modify
certain user changeable parameters:
UÊ
*,ÊLÀi>̅ˆ˜}
UÊ,iÃVÕiÊ«ÀœÌœVœ
– Settings
To exit: Using the TOP or BOTTOM softkey buttons, scroll to and highlight the selection Go to main menu.
Press the CENTER softkey button. The unit will exit the Rescue Options menu screen and return to the
AED Main Menu.
6-7
DAC-2513EN-AB
➜ CPR Breathing
The CPR breathing option allows the user to enable or disable the CPR breathing coaching prompts
during CPR.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to CPR breathing:
AED Main Menu ➜ Rescue options ➜ CPR breathing
What it does: CPR breathing allows the user to enable/disable CPR breathing coaching prompts.
Use the TOP and BOTTOM softkey buttons to select the desired mode. When the desired selection is
chosen, press the CENTER softkey button to set the feature. The AED will now use that coaching setting.
Note: The factory default setting of CPR breathing is set to “disabled.”
➜ Rescue Protcol
The Rescue protocol option allows the user to select a rescue protocol. Rescue protocol options
include the AHA 2005 protocol or “Custom.”
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Rescue protocol:
AED Main Menu ➜ Rescue options ➜ Rescue protocol
What it does: The rescue protocol option allows the user to select between up to two rescue protocols
that have been enabled in the unit. The factory default setting of the rescue protocol is AHA 2005.
To change the protocol, press the CENTER softkey button to highlight the protocol. The user will
be prompted to enter a password to proceed. The password may be obtained from your medical
director or from Defibtech. (For Defibtech contact information, please refer to the “Contacts” section
in Chapter 12 of this manual.) Once the password has been entered, the user may select between the
two protocols.
6. Maintenance Mode
To enter the password use the TOP softkey button to scroll through numbers. Once the correct
number appears, use the CENTER softkey button to move to the next space. Once all of the numbers
have been entered press the CENTER softkey button. The user will now be able to choose a different
rescue protocol.
➜ Settings
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Settings:
AED Main Menu ➜ Rescue options ➜ Settings
What it does: The Settings option allows the user to change the currently enabled
protocol by entering a special protocol code. This code is a special code that encodes
all the important information regarding the protocol. This code is custom generated by
Defibtech. If the code is not entered correctly the protocol will not be changed. Based
on the protocol code entered, the currently selected protocol will be changed to that
described by the special protocol code. You can obtain this code from your medical
director or from Defibtech. (For Defibtech contact information, please refer to the
“Contacts” section in Chapter 12 of this manual.) Once the code has been entered,
the settings will have been changed.
To enter the code use the TOP softkey button to scroll through numbers/letters. Once
the correct number/letter appears, use the CENTER softkey button to move to the next
space. Once all of the numbers/letters have been entered, press the CENTER softkey
button. The settings will have been changed based on the code entered.
6-8
DAC-2513EN-AB
Help Topics Screen
The Help Topics option on the AED Main Menu provides a list of available help topics.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Help topics:
AED Main Menu ➜ Help topics
What it does: The Help topics provides a list of available help topics.
The Help topics are the following:
UÊ*Ài«>Àˆ˜}Ê̅iÊ«>̈i˜Ì
Uʘ>Þ∘}Ê>˜`ÊŜVŽˆ˜}
UÊ*iÀvœÀ“ˆ˜}Ê
*,
UÊ,i«>Vˆ˜}Ê̅iÊL>ÌÌiÀÞ
UÊ,i«>Vˆ˜}Ê̅iÊ«>`Ã
UÊ
…iVŽˆ˜}Ê̅iÊÊÃÌ>ÌÕÃ
UÊ,i«>Vˆ˜}Ê̅iÊ`>Ì>ÊV>À`
Use the TOP and BOTTOM softkey buttons to scroll through the different Help topics items. When the desired
help selection is highlighted (with a box), press the CENTER softkey button to get more information.
To exit: Using the TOP or BOTTOM softkey buttons, scroll and highlight the selection Go to main menu.
Press the CENTER softkey button. The unit will exit the Help Topics menu screen and return to the
AED Main Menu.
6. Maintenance Mode
6-9
DAC-2513EN-AB
DDU-2300 AED Accessories
7. DDU-2300 AED
Accessories
7
This chapter describes the component parts and the accessories that can be used with the Defibtech DDU-2300 AED.
For contact information on obtaining replacement component parts and accessories, refer to Chapter 12 in this manual.
Defibrillation Pads
The DDU-2300 AED must be used with Defibtech self-adhesive defibrillation pads for adults or with child/infant pads
for infants and children. These pads serve two functions:
UʏœÜÊ̅iÊ՘ˆÌÊ̜ÊÀi>`Ê̅iÊ«>̈i˜Ì½ÃÊiiVÌÀœV>À`ˆœ}À>«…Ê­
®ÊÀ…Þ̅“°
UÊiˆÛiÀÊ`iwLÀˆ>̈œ˜Êi˜iÀ}ÞÊ̜Ê̅iÊ«>̈i˜Ìʈvʘii`i`°
The Defibtech self-adhesive defibrillation pad assembly comes in a “leads-out” sealed package that allows the
device to be stored with the pads connected to the AED. When the DDU-2300 AED is used, the operator needs
only to turn the device on, remove the pad package, tear open the package, remove the pads from the blue liner,
apply the pads to the patient, and administer care. The AED has a storage area in the back of the unit that allows for
storage of a single sealed adult pads package.
Battery Packs
The DDU-2300 AED uses a lithium battery pack to provide the AED with a long shelf and standby life. The battery
pack is inserted into the battery pack opening on the back of the AED and latches into place. Battery packs are
not rechargeable.
Data Cards
DDU-2300 AED is designed to optionally use Defibtech Data Cards (“DDC cards”). The AED will operate with or
without a DDC card, but if a DDC card is installed, additional event storage capacity is available.
The DDU-2300 AED accepts DDC cards capable of recording an assortment of data for a given period of time. The
DDU-2300 allows the user to enable or disable recording of audio data. (Refer to the “AED Options Screen” section
in Chapter 6 of this manual.)
The DDC card is inserted into a slot behind the data card/USB port access door found on the side of the AED. (Refer
to the “Installing The Defibtech Data Card (DDC card)” section in Chapter 3 of this manual.) A new event file is
created on the DDC card each time the AED is turned on, and the following information is recorded:
UÊ/…iÊ̈“iÊ̅iÊÊÜ>ÃÊÌÕÀ˜i`ʜ˜°
UÊÊ"̅iÀÊ `>Ì>Ê ÃÕV…Ê >Ã\Ê Ê `>Ì>]Ê Ìˆ“iÊ `>Ì>]Ê >Õ`ˆœÊ `>Ì>Ê ­>Õ`ˆœÊ i˜>Li`Ê V>À`Ê œ˜Þ®]Ê iÛi˜ÌÊ “ˆiÃ̜˜iÃÊ ÃÕV…Ê >Ã\Ê
motion detection, shock advice, shock delivery information.
Multiple events may be recorded on a single DDC card. If the DDC card becomes full, the AED will stop recording
to the card, however, the most critical event documentation for the current session will still be recorded internally.
Internally recorded event information can be downloaded for external review by inserting a blank DDC card into the
unit and following the Data Download procedure. (Refer to the “Downloading The Internal Data Log” section in
Chapter 8 of this manual.)
Using non-Defibtech Data Cards
(DDC cards) may damage the unit
CAUTION and will void the warranty.
!
USB Cable
An optional USB cable may be used with the DDU-2300 AED for connecting the AED to a personal computer running
Defibtech maintenance software. The AED has a mini-USB connector located on the right side of the unit behind
the data card/USB port access door.
!
WARNING
Do not have a USB cable attached
to the unit during a rescue.
7-1
DAC-2513EN-AB
8
Event Viewing
This chapter includes information about DefibView II, Defibtech Data Cards (DDC cards), and downloading internal data logs.
DefibView II
DefibView II is a Windows-based software application that reads data stored on a DDC card or downloaded through
the USB port and displays the data on a personal computer. DefibView II serves the following primary functions:
UÊʘ>LiÃÊi“iÀ}i˜VÞÊV>ÀiÊ«iÀܘ˜iÊ̜ÊÀiۈiÜÊ>ÊV>À`ˆ>VÊi«ˆÃœ`iÊvÀœ“Ê̅iÊ̈“iÊ̅iÊÊÜ>ÃÊÌÕÀ˜i`ʜ˜Ê>˜`Ê
connected to the patient until the unit is turned off.
UÊÊ*ÀœÛˆ`iÃʓ>ˆ˜Ìi˜>˜ViÊ«iÀܘ˜iÊ܈̅Ê>``ˆÌˆœ˜>Ê«>À>“iÌiÀʈ˜vœÀ“>̈œ˜Ê̜Ê>ÃÈÃÌʈ˜ÊÌÀœÕLiŜœÌˆ˜}Ê>Ê`iۈViÊ
suspected of malfunctioning.
DefibView II is a stand-alone software application. DefibView II cannot be used while the AED is in operation, and its
function is solely to support post-event review.
CAUTION
Not intended for clinical use. Information presented by
DefibView II should not be used for making clinical decisions.
8. Event Viewing
!
Defibtech Data Cards (DDC cards)
If a DDC card is installed in the unit, every time the DDU-2300 AED is turned on, the following information is
recorded on a new file on the card:
UÊ/…iÊ̈“iÊ̅iÊÊÜ>ÃÊÌÕÀ˜i`ʜ˜°
UÊÊ"̅iÀÊ`>Ì>ÊÃÕV…Ê>Ã\Ê
Ê`>Ì>]Ê̈“iÊ`>Ì>]Ê>Õ`ˆœÊ`>Ì>Ê­>Õ`ˆœÊi˜>Li`ÊV>À`Ãʜ˜Þ®]ÊiÛi˜ÌʓˆiÃ̜˜iÃÊÃÕV…Ê>Ã\Ê
motion detection, shock advice, shock delivery information.
This information can be reviewed using the DefibView II application.
Using non-Defibtech Data Cards
(DDC cards) may damage the unit
CAUTION and will void the warranty.
!
Downloading The Internal Data Log
Regardless of whether a DDC card is installed in the unit, select information is recorded internally within the DDU2300 AED. The information recorded is limited to:
UÊ/…iÊ̈“iÊ̅iÊÊÜ>ÃÊÌÕÀ˜i`ʜ˜°
UÊ"̅iÀÊ`>Ì>ÊÃÕV…Ê>ÃÊiÛi˜ÌʓˆiÃ̜˜iÃÊ­“œÌˆœ˜Ê`iÌiV̈œ˜]ÊŜVŽÊ>`ۈVi]ÊŜVŽÊ`iˆÛiÀÞʈ˜vœÀ“>̈œ˜]ÊiÌV°®
Uʓ«œÀÌ>˜ÌÊ
ʈ˜vœÀ“>̈œ˜°
Note: Audio data is not logged internally.
Downloading The Internal Data Log Using The DDC Card
To download the internally logged information, perform the following procedure:
UʘÃiÀÌÊ>Ê
ÊV>À`ʈ˜ÌœÊ̅iÊ՘ˆÌ°
UÊ/ÕÀ˜Ê̅iÊ՘ˆÌʜ˜Êˆ˜Ê>ˆ˜Ìi˜>˜Viʜ`iÊLÞÊ«ÀiÃȘ}Ê̅iÊVi˜ÌiÀÊÜv̎iÞÊLÕÌ̜˜°
UÊ*ÀiÃÃÊ̅iÊ̜œÊˆVœ˜Ê̜Êi˜ÌiÀÊ̅iÊÊ>ˆ˜Ìi˜>˜ViÊÃVÀii˜°
UÊÀœ“Ê̅iÊÊ>ˆ˜Ìi˜>˜ViÊÃVÀii˜ÊÃiiVÌÊ̅iÊTransfer Data To Card option.
UÊʏœÜÊ̅iÊ՘ˆÌÊ̜ÊÜÀˆÌiÊ̅iÊVœ˜Ìi˜ÌÃʜvÊ̅iʈ˜ÌiÀ˜>Êœ}Ê̜Ê̅iÊ
ÊV>À`°
The DDU-2300 AED will write the contents of the internal log onto the DDC card. This information can then be
reviewed using DefibView II software.
Downloading The Internal Data Log Using The USB Port
To download the internal data log using the USB port in the unit, connect a USB cable between the unit and a PC.
Launch the DefibView II software and follow USB download directions.
8-1
!
WARNING
Do not operate the DDU-2300 AED in Rescue
Mode while a USB cable is plugged into the unit.
DAC-2513EN-AB
9
Technical Specifications
Defibtech DDU-2300 AED
General
Category
Specification
-ˆâi
7.3 x 9.5 x 2.3 inches
18.5 x 24 x 5.8 cm
Weight
Less than 3 lbs (1.4 kg) (with battery)
Power
Battery Pack (not rechargeable)
Design standards
Meets applicable requirements of
UÊ
ÊÈäÈ䣇£
UÊ1ÊÈäÈ䣇£
UÊ
É
-Ê
ÓÓ°ÓÊ œ°È䣰£‡™ä
UÊ
ÊÈäÈ䣇£‡Ó
UÊ
ÊÈäÈ䣇Ӈ{
UÊÊnä
Patient safety
All patient connections are electrically isolated.
Defibrillator
Category
Specification
Waveform
Impedance Compensated Biphasic Truncated Exponential
Adult: 150 Joules (nominal [+/-15%] delivered into a 50 ohm load)
Energy
Child/Infant: 50 Joules (nominal [+/-15%] delivered into a 50 ohm load)
Automatic by Patient Analysis System
9. Technical Specifications
Charge control
< 4 seconds*
Charge time from shock-advised
Charge complete indication
Shock delivery
Charge time may increase at the end of battery life and for temperatures
below 10°C.
UÊ-"
ÊLÕÌ̜˜Êy>ň˜}
Uʺ*ÀiÃÃÊ-…œVŽÊLÕÌ̜˜»ÊۜˆViÊ«Àœ“«Ì
Shock is delivered by a single SHOCK button
UÊvÊ*>̈i˜Ìʘ>ÞÈÃÊ-ÞÃÌi“Ê`iVˆ`iÃÊÀ…Þ̅“ʈÃʘœÊœ˜}iÀÊŜVŽ>Li]ʜÀ
Automatic
UÊÊ7ˆÌ…ˆ˜ÊÎäÊÃiVœ˜`ÃÊ>vÌiÀÊV…>À}iÊVœ“«iÌi]ʈvʜ«iÀ>̜Àʅ>ÃʘœÌÊ«ÀiÃÃi`Ê
SHOCK button,
or
DISARM
UÊvÊ`iwLÀˆ>̈œ˜Ê«>`ÃÊ>ÀiÊÀi“œÛi`ÊvÀœ“Ê«>̈i˜ÌʜÀÊ՘«Õ}}i`ÊvÀœ“Ê՘ˆÌ°
Manual
UÊÊvʜ«iÀ>̜ÀÊ«ÀiÃÃiÃÊ>˜`ʅœ`ÃÊ̅iÊ" É"ÊLÕÌ̜˜ÊvœÀÊ>««ÀœÝˆ“>ÌiÞÊÌܜÊ
seconds, device will disarm and turn off.
*typical, new battery, at 25°C
9-1
DAC-2513EN-AB
Defibtech DDU-2300 AED (continued)
Waveform Specifications
The DDU-2300 AED delivers a 150J Biphasic Truncated Exponential
waveform to patients with impedances ranging from 25 to 180 ohms.
V
V max
V1
A
t
B
V min
V1'
The waveform is adjusted to compensate for measured patient impedance.
Nominal phase times and energy delivered are shown in the tables below.
ADULT
Phase A Duration
(ms)
Phase B Duration
(ms)
Energy Delivered
(Joules)
25
2.8
2.8
153
50
4.1
4.1
151
75
7.2
4.8
152
100
9.0
6.0
151
125
12.0
8.0
153
150
12.0
8.0
146
175
12.0
8.0
142
Patient Impedance
(Ohms)
Phase A Duration
(ms)
Phase B Duration
(ms)
Energy Delivered
(Joules)
25
4.1
4.1
35
50
5.8
3.8
47
75
5.8
3.8
51
100
7.2
4.8
53
125
7.2
4.8
52
150
9.0
6.0
53
175
9.0
6.0
51
9. Technical Specifications
Patient Impedance
(Ohms)
PEDIATRIC
If impedance is out of range for proper analysis and shock delivery, voice and/or visual prompts will inform the user.
9-2
DAC-2513EN-AB
Defibtech DDU-2300 AED (continued)
Environmental
Category
Specification
Temperature
Operating/
Maintenance
Standby/Storage/
Transport
0 – 50°C (32 – 122°F)
Humidity
5% – 95% (non-condensing)
Air Pressure
700 to 1060 hPa (21 to 31 inHg)
Temperature
0 – 50°C (32 – 122°F)
Humidity
5% – 95% (non-condensing)
Air Pressure
500 to 1060 hPa (15 to 31 inHg)
Altitude
-150 to 4500 meters (-500 to 15,000 feet) per
MIL-STD-810F 500.4 Procedure II
Shock/Drop Abuse Tolerance
MIL-STD-810F 516.5 Procedure IV 48 in, (1.2 meters), any edge,
corner, or surface, in standby mode
Crush test
1,000 lbs (450 kg)
MIL-STD-810F 514.5 Category 20 (Ground)
Vibration
RTCA/DO-160D, Section 8.8.2, Cat R, Zone 2, Curve G (Helicopter)
RTCA/DO-160D, Section 8, Cat H, Zone 2, Curves B & R (Jet Aircraft)
Sealing/Water Resistance
IEC60529 class IP55; Dust Protected, Protected against water jets
(Battery pack installed)
ESD and EMI (radiated and immunity)
Refer to Chapter 10 for details
R&TTE Directive 1999/5/EC
Radio Frequency Emissions Applicable
Directive and Standards
ETSI EN 300 220-2 V2.1.2 (2007-06)
ERC RECOMMENDATION 70-03
ETSI EN 301 489-3 V1.4.1 (2002-08)
9. Technical Specifications
Patient Analysis System
The DDU-2300 Patient Analysis System ensures that the pad/patient impedance is within the proper range and
>˜>ÞâiÃÊ Ì…iÊ «>̈i˜Ì½ÃÊ Ê À…Þ̅“Ê ÌœÊ `iÌiÀ“ˆ˜iÊ Ü…i̅iÀÊ >Ê Ã…œVŽÊ ˆÃÊ ÀiµÕˆÀi`°Ê Ê "˜Ê `iÌiV̈œ˜Ê œvÊ >Ê ˜œ˜‡Ã…œVŽ>LiÊ
rhythm, the user is prompted to perform CPR. For shockable rhythms, the AED automatically charges in preparation
for shock delivery.
The patient analysis system will detect electrical “noise” or artifact in the ECG signal that may interfere with accurate
rhythm analysis. This artifact may be caused by excessive motion to the patient or by external electrical noise.
When this artifact is present, the AED will prompt the user to “Stop motion” or “Stop Interference” until the ECG
signal is free of noise and then proceed to analysis.
9-3
DAC-2513EN-AB
Defibtech DDU-2300 AED (continued)
Shockable Rhythm Criteria
When placed on a patient meeting the indications for use criteria, the DDU-2300 AED is designed to recommend a
defibrillation shock when it detects proper pad impedance and one of the following:
Ventricular Fibrillation: Peak-to-peak amplitude at least 200 μVolts.
!
Some very low amplitude or low frequency VF rhythms may not be interpreted as shockable.
WARNING
Ventricular Tachycardia (Including ventricular flutter and polymorphic VT):
Cardiac rhythm rate of at least 180 bpm and peak-to-peak amplitude at least 200 μVolts.
!
Some very low amplitude or low frequency VT rhythms may not be interpreted as shockable.
WARNING
The DDU-2300 AED is designed to recommend no shock for all other rhythms, including Normal Sinus Rhythms,
fine Ventricular Fibrillation (<200 μVolts), and some slow Ventricular Tachycardia and Asystole.
Patient Analysis System Performance
Rhythm Class
ECG Test
Sample1 Size
Shockable Rhythm –
Ventricular Fibrillation
Non-Shockable
Rhythm – Normal
Sinus Rhythm
Non-Shockable
Rhythm – Asystole
Non-Shockable
Rhythm – All other
non-shockable
rhythms
101
213
113
248
2
Performance
>97%
99%
100%
100%
>99%
Specifications
90% Lower
Confidence Limit2
>95%
Meets the AAMI
DF80 requirement
and AHA
recommendation2 of
Sensitivity > 90%
>97%
Meets the AAMI
DF80 requirement
and AHA
recommendation2 of
Sensitivity > 75%
100%
Meets the AAMI
DF80 requirement
of Specificity
>95% and the AHA
recommendation2 of
Specificity > 99%
100%
Meets the AAMI
DF80 requirement
and the AHA
recommendation2 of
Specificity > 95%
>98%
Meets the AAMI
DF80 requirement
and the AHA
recommendation2 of
Specificity > 95%
9. Technical Specifications
Shockable Rhythm
– Ventricular
Tachycardia
227
Algorithm Performance3
1. From Defibtech ECG Rhythm Databases.
2. Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety.
American Heart Association (AHA) Task Force on Automatic External Defibrillation, Subcommittee on AED Safety and Efficacy. Circulation, 1997;95:1677-1682.
ΰÊÊ-V…˜iˆ`iÀÊ/]Ê>ÀÌi˜ÃÊ*,]Ê*>ÃV…i˜Ê]ÊiÌÊ>°ÊÊՏ̈Vi˜ÌiÀ]ÊÀ>˜`œ“ˆâi`]ÊVœ˜ÌÀœi`ÊÌÀˆ>ÊœvÊ£xäÊLˆ«…>ÈVÊŜVŽÃÊVœ“«>Ài`Ê܈̅ÊÓää‡Ê̜ÊÎÈä‡Ê“œ˜œ«…>ÈVÊŜVŽÃʈ˜Ê̅iÊÀiÃÕÃVˆÌ>̈œ˜ÊœvÊ
out-of-hospital cardiac arrest victims. Circulation 2000;102:1780-1787.
Note: Additional information available upon request.
9-4
DAC-2513EN-AB
Battery Packs
Use only Defibtech battery packs in the DDU-2300 AED.
DBP-2003 Battery Pack
Category
Specification
Model number
DBP-2003
Main battery type
12VDC, 2800 mAh, Lithium/Manganese Dioxide.
Disposable, recyclable, non-rechargeable.
Capacity
125 shocks or 8 hours of continuous operation.*
Charge time
< 4 seconds*
Charge time from the initiation of rhythm analysis to
readiness for discharge with a fresh battery and on the
charge after 6 discharges
Meets or exceeds AAMI DF80 and
IEC 60601-2-4 requirements
Charge time measured from initially switching power on
to charge ready on the charge after 6 discharges
Meets or exceeds AAMI DF80 and
IEC 60601-2-4 requirements
Standby-life (installed in unit)
4 years*
*typical, new battery, at 25°C
Self-Adhesive Defibrillation Pads
Use only Defibtech Pads with your DDU-2300 AED. Defibtech self-adhesive defibrillation pads have the following
characteristics:
DDP-2001
DDP-2002
Type
Adult
Child/Infant < 8 years
< 55 lbs. (25kgs)
Intended use
Disposable
Disposable
Adhesion
Self-adhesive
Self-adhesive
2
Active gel surface area
77 cm each (nominal)
50 cm2 each (nominal)
Cable/connector type
Integrated
Integrated
Cable length
122 cm (typical)
122 cm (typical)
Expiration date
2.5 years from date of manufacture
2.5 years from date of manufacture
9. Technical Specifications
Model number
Note: In the event of a suspected pad defect, the pads should be clearly marked “Not for Use” and returned
to Defibtech, LLC for analysis. (Refer to Chapter 12 of this manual for contact information regarding returns.)
9-5
DAC-2513EN-AB
Event Documentation
Internal Event Record
Critical ECG segments and rescue event parameters are recorded (greater than 60 minutes) and can be downloaded
to a removable data card.
Removable Storage (optional)
Up to 30 hours of ECG and event data storage (no audio option) or up to 3 hours of audio (audio option). ECG and
event storage on a removable data card. Actual length of storage is dependent on card capacity.
Defibtech Event Viewer
DefibView II is a PC-based application program that allows review of ECG data and other patient and device
performance parameters after an emergency event.
DefibView II runs on various Windows platforms including Windows XP and newer versions. Minimum system
requirements for adequate performance are as follows:
UÊ*i˜ÌˆÕ“Ê{Ê*ÀœViÃÜÀ°
UÊx£ÓÊÞÌiÊ-ÞÃÌi“Êi“œÀÞ°
UÊ£ÊÞÌiÊvÀiiÊë>Viʜ˜Ê…>À`Ê`ˆÃŽ°
UÊ1-Ê£°äÊVœ˜˜iV̈ۈÌÞ°
Recycling Information
At the end of useful life, recycle the defibrillator and its accessories.
Recycling Assistance
For recycling assistance contact your local Defibtech distributor.
Recycle in accordance with local and national regulations.
Preparation For Recycling
9. Technical Specifications
Items should be clean and contaminant-free prior to being recycled.
When recycling used disposable electrodes, follow local clinical procedures.
Packaging For Recycling
Packaging should be recycled in accordance with local and national requirements.
Notice To European Union Customers
The crossed-out wheeled bin symbol on this device indicates that this equipment has been
put on the market after 13 August 2005, and is included in the scope of the directive 2002/96/
EEC on Waste Electrical and Electronic Equipment (WEEE) and of the national decree(s), which
transpose provisions of such directive.
At the end of its lifetime, this device can only be disposed of in compliance with the provisions of the above-mentioned
European directive (and following possible revisions) as well as with the corresponding national regulation. Severe
«i˜>ÌˆiÃÊ>ÀiÊ«œÃÈLiÊvœÀÊ՘>Õ̅œÀˆâi`Ê`ˆÃ«œÃ>°Ê
iVÌÀˆV>Ê >˜`Ê iVÌÀœ˜ˆVÊ µÕˆ«“i˜ÌÊ ­®Ê “>ÞÊ Vœ˜Ì>ˆ˜Ê «œṎ˜}Ê Vœ“«œ˜i˜ÌÃÊ >˜`Ê …>â>À`œÕÃÊ ÃÕLÃÌ>˜ViÃ]Ê Ì…iÊ
accumulation of which could pose serious risk for the environment and human health. It is for this reason that local
administrations provide regulations, which encourage reuse and recycling, and prohibit the disposal of WEEE as
՘ÜÀÌi`ʓ՘ˆVˆ«>ÊÜ>ÃÌiÊ>˜`ÊÀiµÕˆÀiÊ̅iÊVœiV̈œ˜ÊœvÊÃÕV…Ê7ÊÃi«>À>ÌiÞÊ­>ÌÊëiVˆwV>ÞÊ>Õ̅œÀˆâi`ÊÌÀi>̓i˜ÌÊ
v>VˆˆÌˆiî°ÊÊ>˜Õv>VÌÕÀiÀÊ>˜`Ê>Õ̅œÀˆâi`Ê`ˆÃÌÀˆLÕ̜ÀÃÊ>ÀiÊÀiµÕˆÀi`Ê̜ÊÃÕ««Þʈ˜vœÀ“>̈œ˜Ê>LœÕÌÊ>ÊÃ>viÊÌÀi>̓i˜ÌÊ>˜`Ê
disposition of the specific device.
You may also return this equipment to your distributor when purchasing a new one. As for reuse and recycling,
notwithstanding the limits imposed by the nature and the use of this device, the manufacturer will do its best to
develop recovery processes. Please contact the local distributor for information.
9-6
DAC-2513EN-AB
10 Electromagnetic Conformity
Guidance And Manufacturer’s Declaration
The DDU-2300 is intended for use in the electromagnetic environment specified below. The customer or the user
of the DDU-2300 should assure that it is used in such an environment.
ELECTROMAGNETIC EMISSIONS
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
The DDU-2300 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
Group 1
Class B
CISPR 22
Class B
FCC Part 15
Class B
Harmonic emissions IEC 61000-3-2
Not applicable
Battery operated equipment
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable
Battery operated equipment
10. Electromagnetic
Conformity
CISPR 11
ELECTROMAGNETIC IMMUNITY
Immunity test
Electrostatic discharge (ESD)
IEC 60601-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 60601
test level
Compliance
level
Electromagnetic environment guidance
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
There are no special requirements
with respect to electrostatic discharge.
Not applicable
Battery operated equipment
Not applicable
Battery operated equipment
Not applicable Not applicable
Battery operated equipment
±2 kV for
power line
supply lines
±1 kV for
input/output
lines
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
IEC 61000-4-11
*œÜiÀÊvÀiµÕi˜VÞÊ­xäÉÈäÊâ®Ê“>}˜ïVÊ
field
3 A/m
IEC 61000-4-8
3 A/m
Power frequency magnetic fields
should not be greater than levels
characteristic of a typical location in a
commercial or hospital environment.
10-1
DAC-2513EN-AB
Electromagnetic Immunity (continued)
IEC 60601
test level
Immunity test
Compliance
level
10 V/m
näÊâÊ̜Ê
Ó°xÊâ
Radiated RF
10 V/m
IEC 61000-4-3
nä¯ÊxâÊÊ
Modulation
Electromagnetic environment guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
the DDU-2300, including cables,
than necessary. The recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter is shown
in the following table.
10. Electromagnetic
Conformity
Interference may occur in
the vicinity of equipment
marked with this symbol.
Note 1\ÊÌÊnäÊâÊ>˜`ÊnääÊâ]Ê̅iʅˆ}…iÀÊvÀiµÕi˜VÞÊÀ>˜}iÊ>««ˆið
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
/…iÊ-Ê­ˆ˜`ÕÃÌÀˆ>]ÊÃVˆi˜ÌˆwVÊ>˜`ʓi`ˆV>®ÊL>˜`ÃÊLiÌÜii˜Ê£xäʎâÊ>˜`ÊnäÊâÊ>ÀiÊÈ]ÇÈxÊâÊ̜ÊÈ]ǙxÊâÆÊ
£Î]xxÎÊâÊ̜ʣÎ]xÈÇÆÊÓÈ]™xÇÊâÊ̜ÊÓÇ]ÓnÎÊâÆÊ>˜`Ê{ä]ÈÈÊâÊ̜Ê{ä]ÇäÊâ°
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the DDU-2300 is
used exceeds the applicable RF compliance level above, the DDU-2300 should be observed to verify normal
operation. If abnormal performance is observed additional measures may be necessary, such as reorienting or
relocating the DDU-2300.
Separation Distances
The DDU-2300 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the DDU-2300 can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the DDU-2300
as recommended below, according to the maximum output of the communications equipment.
Recommended separation distances between portable and mobile RF communications equipment and the DDU-2300
Separation distance according to frequency of transmitter (m)
Rated maximum
output power of
transmitter (W)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2√P
d = 2.3√P
0.01
0.12
0.23
0.1
0.38
0.73
1
1.20
2.30
10
3.79
7.27
100
12.00
23.00
10-2
DAC-2513EN-AB
Separation Distances (continued)
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1\ÊÊÃÊnäÊâÊ>˜`ÊnääÊâ]Ê̅iÊÃi«>À>̈œ˜Ê`ˆÃÌ>˜ViÊvœÀÊ̅iʅˆ}…iÀÊvÀiµÕi˜VÞÊÀ>˜}iÊ>««ˆið
Note 2\ÊÊ/…iÊ-Ê­ˆ˜`ÕÃÌÀˆ>]ÊÃVˆi˜ÌˆwVÊ>˜`ʓi`ˆV>®ÊL>˜`ÃÊLiÌÜii˜Ê£xäʎâÊ>˜`ÊnäÊâÊ>ÀiÊÈ]ÇÈxÊâÊ̜ÊÈ]ǙxÊâÆÊ
£Î]xxÎÊâÊ̜ʣÎ]xÈÇÆÊÓÈ]™xÇÊâÊ̜ÊÓÇ]ÓnÎÊâÆÊ>˜`Ê{ä]ÈÈÊâÊ̜Ê{ä]ÇäÊâ°
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM
vÀiµÕi˜VÞÊL>˜`ÃÊLiÌÜii˜Ê£xäʎâÊ>˜`ÊnäÊâÊ>˜`ʈ˜Ê̅iÊvÀiµÕi˜VÞÊÀ>˜}iÊnäÊâÊ̜ÊÓ°xÊâÊ̜Ê`iVÀi>ÃiÊ̅iʏˆŽiˆ…œœ`Ê
that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas.
Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Regulatory Compliance
10. Electromagnetic
Conformity
Changes or modifications of this product, not expressly approved by Defibtech, may void the user’s authority to
operate the equipment.
This device complies with part 15 of the FCC Rules and Industry Canada Radio Standard RSS-210. Operation is
subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment off and on, the user
is encouraged to try to correct the interference by one or more of the following measures:
UÊ,iœÀˆi˜ÌʜÀÊÀiœV>ÌiÊ̅iÊÀiViˆÛˆ˜}Ê>˜Ìi˜˜>°
UʘVÀi>ÃiÊ̅iÊÃi«>À>̈œ˜ÊLiÌÜii˜Ê̅iÊiµÕˆ«“i˜ÌÊ>˜`ÊÀiViˆÛiÀ°
UÊ
œ˜˜iVÌÊ̅iÊiµÕˆ«“i˜Ìʈ˜ÌœÊ>˜ÊœÕ̏iÌʜ˜Ê>ÊVˆÀVՈÌÊ`ˆvviÀi˜ÌÊvÀœ“Ê̅>ÌÊ̜Ê܅ˆV…Ê̅iÊÀiViˆÛiÀʈÃÊVœ˜˜iVÌi`°
UÊ
œ˜ÃՏÌÊ̅iÊ`i>iÀʜÀÊ>˜ÊiÝ«iÀˆi˜Vi`ÊÀ>`ˆœÉ/6ÊÌiV…˜ˆVˆ>˜ÊvœÀʅi«°
CE Marking And European Union Compliance – Radio Transmitter
Defibtech, LLC declares that the DDU-2300 AED radio transmitter is in compliance with the essential requirements
and other relevant provisions of Directive 1999/5/EC. Applicable standards are listed in the “Environmental” section
in Chapter 9 of this manual.
10-3
DAC-2513EN-AB
11. Glossary Of Symbols
11
Glossary Of Symbols
Symbol
Meaning
High voltage present.
SHOCK Button – Delivers defibrillation shock to the patient when the device is ready to shock.
ON/OFF Button
UÊÊ/ÕÀ˜ÃÊ̅iÊ`iۈViÊ" Ê܅i˜ÊˆÌʈÃÊ"°
UÊÊ/ÕÀ˜ÃÊ̅iÊ`iۈViÊ"Ê܅i˜ÊˆÌʈÃÊ" °
!
Caution, consult accompanying documents.
Do not expose to high heat or open flame. Do not incinerate.
Recyclable.
Consult operating instructions.
Do not damage or crush.
Follow proper disposal procedures.
Meets the requirements of the European Medical Device Directive.
Note: XXXX represents the identification number of the notified body.
XXXX
Temperature limitation.
Use by (yyyy-mm).
Defibrillation proof - Can withstand the effects of an externally applied defibrillation shock. Internally
powered with defibrillator-proof BF-type patient applied parts (per EN 60601-1)
Manufacturer.
Date of manufacture.
YYYY
Do not reuse.
For USA users only.
11-1
DAC-2513EN-AB
11. Glossary Of Symbols
Glossary Of Symbols (continued)
Symbol
Meaning
Catalogue number.
Keep dry.
Handle with care.
Transportation and storage requirements.
See environmental requirements.
EC
REP
LATEX
Õ̅œÀˆâi`ÊÕÀœ«i>˜Ê,i«ÀiÃi˜Ì>̈Ûi°Ê
Does not contain latex.
Lot number.
IP55
Dust protected; Protected against water jets.
>ÃÈwi`ÊLÞÊ/16Ê,…iˆ˜>˜`ʜvÊ Ê܈̅ÊÀiëiVÌÊ̜ÊiiVÌÀˆVÊŜVŽ]ÊwÀi]Ê>˜`ʓiV…>˜ˆV>Ê…>â>À`ʜ˜ÞÊ
in accordance with UL 60601-1, CAN/CSA C22.2 No.601.1-M90, IEC 60601-1, and IEC 60601-2-4.
Conforms to UL Standard UL 60601-1. Certified to CAN/CSA Standard C22.2 No. 601.1-M90.
Serial number.
Lithium Manganese Dioxide Battery.
Product is not sterile.
11-2
DAC-2513EN-AB
12
Contacts
Manufacturer
Defibtech, LLC
741 Boston Post Road
Guilford, CT 06437
Tel.: (866) 333-4241 (Toll-free within North America)
(203) 453-4507
Fax: (203) 453-6657
Emails:
sales@defibtech.com
reporting@defibtech.com
service@defibtech.com
(Sales)
(Medical Device Reporting)
(Service and Repair)
European Authorized Representative
EC
REP
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands
Tel.: +31 70 345 8570
Fax: +31 70 346 7299
Patents pending.
12. Contacts
This product and its accessories are manufactured and sold under one or more of the following United States
patents: D523,393, D548,346, D551,628.
This product and its accessories are manufactured and sold under license to at least one or more of the following
United States patents: 5,591,213; 5,593,427; 5,601,612; 5,607,454; 5,611,815; 5,617,853; 5,620,470; 5,662,690;
5,735,879; 5,749,904; 5,749,905; 5,776,166; 5,800,460; 5,803,927; 5,836,978; 5,836,993; 5,879,374; 6,016,059;
6,047,212; 6,075,369; 6,438,415; 6,441,582.
12-1
DAC-2513EN-AB

advertisement

Key Features

  • Easy to use
  • Portable
  • Battery powered
  • Voice prompts
  • Text prompts
  • Display screen
  • Event information
  • SHOCK/NO-SHOCK recommendations

Frequently Answers and Questions

How do I turn on the DDU-2300 AED?
To turn the DDU-2300 AED on, press the green ON/OFF button. The unit will emit a “beep” and the display screen will turn on. The ASI indicator next to the ON/OFF button will illuminate green anytime the AED is on.
What does the Active Status Indicator (ASI) mean?
The ASI is powered by the battery pack. If the battery pack has been completely discharged or is not installed in the unit, the active status indication will not be available. In this case, immediately replace the battery pack or reinsert it into the unit to restore active status indication. Flashing Green: DDU-2300 AED is OFF and ready for use. Solid Green: DDU-2300 AED is ON and ready for use. Flashing or Solid Red: DDU-2300 AED needs immediate service. Refer to the “Troubleshooting” section in Chapter 5 of this manual or call Defibtech for service. No Flashing Light: DDU-2300 AED needs immediate service. Refer to the “Troubleshooting” section in Chapter 5 of this manual or call Defibtech for service.
What are the steps for using the DDU-2300 AED in rescue mode?
The basic steps for use are: Check the DDU-2300 AED Status. Turn On The DDU-2300 AED. Call For Help. Prepare The Patient. Connect Defibrillation Pads To The DDU-2300 AED. Analyze The Patient’s Heart Rhythm. Deliver A Shock If Instructed. Perform CPR. Follow Post Use Procedures.