The Brannan 13/483/0 Humidity & Temperature Meter is a compact, versatile device designed for measuring and monitoring both humidity and temperature in various environments. It features a wide measuring range, providing accurate and reliable readings in a variety of industrial, commercial, and residential settings. This manual will guide you through the operation of the meter, highlighting its key features and functionalities.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Brannan 13/483/0 Humidity & Temperature Meter User's Manual | Manualzz

User's Manual

Mini Humidity & Temperature Meter 13/483/0

Please read this user manual thoroughly before use

and keep it safe for future reference.

1. General Description

This Temperature and Humidity Meter is designed to provide simple, accurate measurements for Health and Safety Engineers, Industrial Safety Officers in a variety of industrial environments such as factories, schools, kitchens, garages and construction sites.

2. Features

Provides continuous readings of temperature and humidity.

Stores Maximum and Minimum readings

Hold function

3. Basic User Maintenance

This device contains no user serviceable parts and repairs should only be performed by qualified personnel. Clean by wiping the unit with a soft dry cloth, do not use abrasives or solvents on this instruments. The battery will need to be replaced periodically, see section 10.

4. Safety

Read the following safety information carefully before attempting to operate this meter.

Use the meter only as specified in this manual; otherwise, the readings provided by the meter may be impaired.

Suitable conditions for use:

Altitude up to 2000 metres

Storage and operating conditions: -20 to 60 ℃, RH≤90%(Non-Condensing)

EMC compliant (IEC61672-1 Class 2)

5.

Specifications

Measuring

Range

Accuracy

Humidity

Temperature

0%~100%RH

-30℃~70℃

-22℉~158℉

Response time

Sampling rate

Working Temp.

Power Resource

1 second

2 times per second

-30 ℃to 60℃

9V battery

Dimension(L*W*H) 165* 55* 38mm

Weight 88g

20%~40%RH : ±3.0%

40%~60%RH: ±2.5%

60%~80%RH: ±3.0%

0%~20%RH

80%~100%RH: ±3.5%

-30 ℃~-10℃&F:

±1.0

℃&F

-10

℃~70℃&F:

±0.5

℃&F

Resolution

0.1%RH

0.1℃

6. Meter Description

① Temperature & Humidity sensor

② LCD display

③ Power ON/OFF button

④ MAX/MIN button

⑤ Tripod nut

⑥ Battery cover automatically turn off after approx. 30 seconds of inactivity.

Hold & MAX/MIN Function

HOLD: After turning the meter on, momentarily press " which indicates the data is frozen. To release the held reading, press the button again.

MAX/MIN: Press and hold "

" button "HOLD" icon will appear on the LCD,

" button for at least 3 seconds until "MAX" icon appears on the LCD to view the current MAX reading; press the button again, "MIN" icon will appear to view the current MIN reading.

7. LCD Display Description

① Low battery indication

② MAX/MIN icon

③ Humidity icon

④ Humidity reading

⑤ ℃/℉ temperature units

⑥ Temperature reading

⑦ Auto power off icon

⑧ Hold icon

8. Operating Instructions

Backlit & Power ON/OFF Function

Power ON:Press and release the power button" " and the LCD screen will be activated.

To exit the MAX/MIN mode, press and hold the button "

/

temperature units

While the meter is switched off, continuously press the

" for 3 seconds.

button and then also continuously press the

button. Keep holding both buttons until the LCD displays the current readings and the temperature units will change from °C to °F. Release both buttons before the meter switches off.

Notes

① This meter has already been calibrated before delivery. Do not revise the calibration parameters without professional personnel and device. The recommended calibration period is one year.

② Repairs or servicing not covered in this manual should only be performed by qualified personnel.

③ Wipe the unit with a dry soft cloth. Do not use abrasives or solvents on this instrument.

④ Remove the battery when the meter is to be stored for long periods of time to avoid battery leakage.

Accessories

① User's manual

② 9V battery

Power OFF: Press and hold the power button" " for approximately 3 secs until the LCD switches off.

Brannan Thermometers & Instrumentation

Backlit:When the meter is on, press the power button " ", again and the backlight will be switch on.

Cleator Moor

Pressing the power button again will switch it off, but it will automatically turn off after approximately 30 secs Cumbria, UK

CA25 5QE of inactivity. Auto power off icon, the meter will automatically turn off after approx. 15 minutes of inactivity.

Backlit: After turning the meter on, momentarily press " " button, the backlit will be on/off, it will www.brannan.co.uk

BRAN.2015.Rev.0

advertisement

Key Features

  • Continuous readings of humidity and temperature
  • Stores maximum and minimum readings
  • Hold function
  • Backlit LCD display
  • Automatic power off
  • ℃/℉ switchable units
  • Tripod nut for mounting
  • 9V battery powered
  • EMC compliant (IEC61672-1 Class 2)
  • Wide measuring range

Frequently Answers and Questions

What is the measuring range of the humidity sensor?
The humidity sensor has a measuring range of 0% to 100%RH.
What is the accuracy of the humidity readings?
The accuracy varies depending on the humidity level, ranging from ±3.0% at 20%-40%RH to ±3.5% at 80%-100%RH.
How often does the meter sample readings?
The meter samples readings 2 times per second.
What is the purpose of the hold function?
The hold function freezes the current readings on the LCD display, allowing you to record or note them without the readings changing.
How do I change the temperature units from Celsius to Fahrenheit?
While the meter is switched off, continuously press and hold both the power button and the button until the temperature units change from °C to °F. Release both buttons before the meter switches off.