Below you will find brief information for Compact balance FFN-N. This document describes how to use, maintain, and troubleshoot your new balance. It includes information on unpacking, setting up, adjusting, and weighing with the balance, as well as instructions on how to use the menu and understand error messages.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Compact balance FFN-N Operating instructions | Manualzz

KERN & Sohn GmbH

Ziegelei 1 Phone: +49-[0]7433- 9933-0

D-72336 Balingen email: [email protected]

Fax: +49-[0]7433-9933-149

Internet: www.kern-sohn.com

Operating instructions

Compact balance

KERN FFN-N

Version 2.0

05/2011

GB

FFN-N-BA-e-1120

GB

KERN FFN-N

Version 2.0 05/2011

Operating instructions

Compact balance

Table of Contents

1 Technical Data ............................................................................................... 4

1.1

Dimensions................................................................................................................................. 5

2 Appliance Overview....................................................................................... 6

2.1

Overview of displays................................................................................................................... 6

2.2

Keyboard Overview .................................................................................................................... 7

3 Basic Information........................................................................................... 8

3.1

Proper use .................................................................................................................................. 8

3.2

Improper use .............................................................................................................................. 8

3.3

Warranty ..................................................................................................................................... 8

3.4

Monitoring of Test Resources .................................................................................................... 8

4 Basic Safety Precautions .............................................................................. 9

4.1

Pay attention to the instructions oft the Operation Manual ........................................................ 9

4.2

Personnel training....................................................................................................................... 9

5 Transportation & Storage.............................................................................. 9

5.1

Testing upon acceptance ........................................................................................................... 9

5.2

Packaging / return transport ....................................................................................................... 9

6 Unpacking, Setup and Commissioning ..................................................... 10

6.1

Installation Site, Location of Use .............................................................................................. 10

6.2

Unpacking and implantation ..................................................................................................... 11

6.2.1

Scope of supply ........................................................................................................................ 13

6.3

Battery power supply ................................................................................................................ 13

6.3.1

Mains connection during rechargeable battery operation ........................................................ 15

6.4

Initial Commissioning................................................................................................................ 16

6.5

Protection type IP-67 ................................................................................................................ 16

7 Adjustment ................................................................................................... 16

7.1

Adjustment................................................................................................................................ 16

8 Operation...................................................................................................... 18

8.1

Weighing................................................................................................................................... 18

8.2

Taring........................................................................................................................................ 18

8.3

Weighing units switch-over....................................................................................................... 19

9 The menu...................................................................................................... 20

9.1

Navigation in the menu............................................................................................................. 20

9.2

Menü-Übersicht ........................................................................................................................ 21

9.3

Set auto-OFF function .............................................................................................................. 22

FFN-N-BA-e-1120 2

10 Error messages............................................................................................ 23

11 Instant Help .................................................................................................. 24

12 Service, maintenance, disposal.................................................................. 25

12.1

Cleaning ................................................................................................................................... 25

12.2

Service, maintenance ............................................................................................................... 25

12.3

Disposal .................................................................................................................................... 25

1 Technical Data

KERN

Weighing range (max)

Readability (d)

Reproducibility

Linearity

Recommended adjustment weight, not added (class)

Stabilization time (typical)

FFN

3K0.5IPN

3 kg

0,5 g

0,5 g

1 g

3 kg

(M3)

FFN

6K1IPN

6 kg

1 g

1 g

1 g

6 kg

(M3)

2 sec.

FFN

15K2IPN

15 kg

2 g

2 g

4 g

15 kg

(M3)

FFN

25K5IPN

25 kg

5 g

5 g

10 g

25 kg

(M3)

Warm-up time

Electric Supply

Battery power supply

Auto-off (Rechargeable battery)

Display mode

Operating temperature

Humidity of air

Weighing plate dimensions

(stainless steel) (mm)

Casing dimensions

Total dimensions mounted

(mm)

Weight kg (net)

IP protection

30 min.

Supply voltage: 110V-230V AC

Mains adapter: 12 V, 0,8 A

Operating period: Background lighting on: 30 h

Background lighting off: 50 h

Loading time: 12 h

15 min., 5 min., 3 min., off

LCD, digit size 25 mm

0° C …. + 40° C

25 % - 95 % (non-condensing)

230 x 190

230 x 300 x 130

230 x 300 x 130

3,2

IP67

FFN-N-BA-e-1120 4

1.1 Dimensions

2 Appliance Overview

2.1 Overview of displays

Display

Designation

Zeroing display

Description

Should the balance not display exactly zero

Stability display despite empty scale pan, press the button.

Will be reset to zero after short waiting time.

Scales are in a steady state

Net weight will be displayed Net weight display

Capacity display

Rechargeable battery

Rechargeable battery almost empty, please connect power pack

Power supply connected Icon is glowing for power supply via power pack

Display weighing unit kg Displayed weight in kg

Display weighing unit lb Displayed weight in lb

FFN-N-BA-e-1120 6

2.2 Keyboard Overview

Button Designation Function

TARE button Tare balance

Zero setting key Balance set at zero

+

TARE + ZERO-key

Change units

in the menu

ƒ Call up menu

ƒ Change to the next menu item /parameter

ƒ Select menu item / parameter

3 Basic Information

3.1 Proper use

The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.

3.2 Improper use

Do not use balance for dynamic weighing. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a container on the balance.)

Do not leave permanent load on the weighing plate. This may damage the measuring system.

Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damaged by this.

Never operate balance in explosive environment. The serial version is not explosion protected.

The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance.

The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.

3.3 Warranty

Warranty claims shall be voided in case

ƒ Our conditions in the operation manual are ignored

ƒ The appliance is used outside the described uses

ƒ The appliance is modified or opened

ƒ Mechanical damage or damage by media, liquids, natural wear and tear

ƒ The appliance is improperly set up or incorrectly electrically connected

ƒ The measuring system is overloaded

3.4 Monitoring of Test Resources

In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test.

Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In

KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.

FFN-N-BA-e-1120 8

4 Basic Safety Precautions

4.1 Pay attention to the instructions oft the Operation Manual

Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.

4.2 Personnel training

The appliance may only be operated and maintained by trained personnel.

5 Transportation & Storage

5.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.

5.2 Packaging / return transport

Ö Keep all parts of the original packaging for a possibly required return.

Ö Only use original packaging for returning.

Ö Prior to dispatch disconnect all cables and remove loose/mobile parts.

Ö Reattach possibly supplied transport securing devices.

Ö Secure all parts such as glass wind screen, weighing platform, power unit etc. against shifting and damage.

6 Unpacking, Setup and Commissioning

6.1 Installation Site, Location of Use

The balances are designed in a way that reliable weighing results are achieved in common conditions of use.

You will work accurately and fast, if you select the right location for your balance.

Therefore, observe the following for the installation site:

ƒ Place scales on a stable, even surface;

ƒ Avoid extreme heat as well as temperature fluctuation caused by installing next to a radiator or in the direct sunlight;

ƒ Protect the balance against direct draughts due to open windows and doors;

ƒ Avoid jarring during weighing;

ƒ Protect the balance against high humidity, vapors and dust;

ƒ Do not expose the device to extreme dampness for longer periods of time.

Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.

ƒ Avoid static charge of goods to be weighed or weighing container.

Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.

FFN-N-BA-e-1120 10

6.2 Unpacking and implantation

Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.

1

3

2

7

5

4

8

6

1 Weighing plate

2 Locking screw

3 Carrier weighing plate

4 Display

5 Bubble level

6 Keyboard

7 Footscrews

8 Rechargeable battery compartment cover

9 Screws for battery compartment cover

9

Level balance with foot screws until the air bubble of the water balance is in the prescribed circle.

Pressure compensation screw

Mains connection

Battery compartment

Pressure compensation screw:

[1]

On the lower side of the balance there is a pressure compensation screw [1] which has to be opened for the weighing process. Fasten the screw before cleaning the balance.

FFN-N-BA-e-1120 12

Membrane for degree of protection IP67

6.2.1 Scope of supply

Serial accessories:

ƒ Balance

6.3 Battery power supply

The rechargeable battery is charged via the delivered power supply.

The operating time of the rechargeable battery with background illumination is 30h, without that it is 50 hrs. Charging time until complete recharging 12h.

In the menu you can activate the AUTO-OFF function

[time off], see chap. 9.3.

According to menu settings, the balance switches automatically off in order to spare the battery.

Insert rechargeable battery:

Ö Remove any adherent moisture from the balance

Ö Move both levers 90° to the left

Ö Remove the rechargeable battery cover

Ö Connecting the rechargeable battery

Watch the colours:

Red on red !

Black on black!

Ö Insert the rechargeable battery in a manner that it cannot slip (fasten with a rubber mat)

Ensure that the cables are not

squeezed.

FFN-N-BA-e-1120 14

Ö Close the rechargeable battery cover

Ö Move both levers 90° to the right

If in the display the triangle appears above the capacity display , the rechargeable battery is nearly exhausted. Connect power pack, the rechargeable battery is loaded.

Avoid excess pressure on the balance in order to avoid damaging them, particularly as they are supported by the weighing plate.

6.3.1 Mains connection during rechargeable battery operation

During the rechargeable battery operation ensure that the mains connection is covered with a rubber cap.

Only by this the type of protection IP67 is ensured.

6.4 Initial Commissioning

In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1).

The weighing scale must be connected to the power supply (batteries) during this warm-up time.

The accuracy of the balance depends on the local acceleration of gravity.

Strictly observe hints in chapter Adjustment.

6.5 Protection type IP-67

Dust and spray water protection:

The KERN FFN is in compliance with the requirements set for protection type IP67.

Suitable for short-term use in wet areas Dustproof.

7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location

(only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.

7.1 Adjustment

Carry out adjustment as near as possible to the maximum load of the scale (See chpt. 1 “Techn. data”). The accuracy of the weight used for adjustment has to correspond approximately to readability d of the scale, or rather closer than that.

Info about test weights can be found on the Internet at: http://www.kern-sohn.com

Procedure when adjusting:

Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization.

FFN-N-BA-e-1120 16

Display Operation

CAL

Ö Start balance by pressing

Ö Whilst the weighing scale is carrying out a self-test,

(00…- 99…) press

CAL“ appears.

and hold until the message “F1

UnLQD

Ö Press

, in the display appears „UnLod“.

(Example)

Ö Press

again the weight value will be displayed.

Ö Enter value for adjustment weight (See chpt 1) by

Ö Position weight

Ö Press

Remove weight whilst scale is carrying out self-test

Weighing scale changes to zero display.

This completes the adjustment process.

0.0

8 Operation

8.1 Weighing

U1.04

Ø

0.0

Ö Start balance by pressing

Program version is displayed.

Afterwards the balance will carry out a self-test.

The balance will be ready for operation as soon as weight reading “0.0” and the triangular icon above the stability display appear.

ƒ The time.

key can be used to set the balance to zero at any

Ö Turn off scale by pressing

.

The “0.0” display disappears and the weighing scale is now switched off.

8.2 Taring

The tare weight of any preloads can be deducted by pressing a button so that the actual weight of the baby is displayed in subsequent weighings.

200.0

Ø

Ö Put on weighing receptacles and press

.

The zero display appears as well as the triangle above the zero setting icon as well as the stability icon and the net weight icon .

The weight of the container is now internally saved.

0.0

1000.0

Ö Place goods to be weighed in the weighing container.

The net weight of the goods to be weighed is displayed.

-200.0

The weight of the weighing container will be displayed as a minus number after removing the weighing container (= gross weight).

FFN-N-BA-e-1120 18

0.0

The tare weight is saved until it is deleted. Remove the load from the balance and press . The zero display appears and the triangle above the net weight icon disappears.

-

1200.0

Gross weight:

Ö Press

as long as the weighing tray and the load are present on the weighing platform.

Ö Remove weighing goods and receptacle

Gross weight will be shown as a negative value.

8.3 Weighing units switch-over

Ö Press

and at the same time.

The weighing scale switches between the kg and lb units.

9 The menu

9.1 Navigation in the menu

Ø

Ö Start balance by pressing

Ö While the balance carries out a self test, press keep pressed until „F1 CAL“ appears

and

Ö Use

to change to the next menu item

Ö Select menu item using

Ö Use

to change between the individual parameters

Ö Select the desired parameter using

FFN-N-BA-e-1120 20

9.2 Menü-Übersicht

Menu item Function

Calibration not documented not documented not documented

Auto-OFF function, adjustable between off, 3, 5 and 15 minutes not documented

Return to weighing mode by

9.3 Set auto-OFF function

Ø

Ö Start balance by pressing

Ö While the balance carries out a self test, press keep pressed until „F1 CAL“ appears

Ø

Ø

Ø

Ø

Ø

and

Ö Press

repeatedly until „F5 oFF“ appears

Ö Select function using

Ö Use

to change between the individual parameters

Ö Select the desired parameter using

“F5 oFF“ appears, now the function is enabled

Ö Press

repeatedly until “bACK“ appears

Ö Press

The balance carries out a restart and changes to weighing mode

FFN-N-BA-e-1120 22

10 Error messages

Display Description

Zero range exceeded

ERR4

Electronic error

ERR6

11 Instant Help

In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning.

The displayed weight does not glow.

ƒ The balance is not switched on.

ƒ Batteries are inserted incorrectly or empty

ƒ No batteries inserted.

The displayed weight is permanently changing

ƒ Weighing plate has contact with other objects.

ƒ Electromagnetic fields / static charging

(choose different location/switch off interfering device if possible)

The weighing value is obviously

ƒ The display of the balance is not at zero

ƒ Adjustment is no longer correct. wrong

ƒ The balance is on an uneven surface.

ƒ Electromagnetic fields / static charging

(choose different location/switch off interfering device if possible)

Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.

FFN-N-BA-e-1120 24

12 Service, maintenance, disposal

12.1 Cleaning

Remove batteries from instrument before cleaning.

Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth.

Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.

Spilled weighing goods must be removed immediately.

12.2 Service, maintenance

The appliance may only be opened by trained service technicians who are authorized by KERN.

Before opening, disconnect from power supply.

12.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.

advertisement

Key Features

  • IP67 protection
  • Rechargeable battery
  • Auto-off function
  • Tare function
  • kg/lb weighing units
  • LCD display
  • Menu navigation
  • Error messages

Frequently Answers and Questions

How do I adjust my balance?
Adjusting the balance is essential for obtaining accurate weighing results. It involves calibrating the balance to the local acceleration due to gravity. The process is outlined in the "Adjustment" section of the manual. It involves placing a known weight onto the balance and following the on-screen prompts to calibrate the balance.
What are the different weighing units available?
The balance offers two weighing units: kg (kilograms) and lb (pounds). You can switch between these units by pressing the "kg/lb" button on the keyboard. The current unit is indicated by the icon next to the displayed weight.
How do I use the tare function?
The tare function allows you to subtract the weight of a container or other preload from the overall weight. This is useful for weighing goods inside a container. To use the tare function, press the "TARE" button while the container is on the weighing plate. The balance will then display the weight of the goods only.
How do I clean my balance?
Keep your balance clean and free of dust and debris to ensure accurate and reliable weighing results. The balance should be cleaned with a soft, damp cloth and mild soap suds, avoiding harsh chemicals or solvents. Ensure no liquid penetrates into the balance. You can remove loose residue with a brush or a manual vacuum cleaner. It is crucial to keep the balance clean to maintain its accuracy and longevity.
What are the error messages and what do they mean?
The balance displays error messages to alert you of potential problems or issues. The ERR4 message indicates that the zero range has been exceeded, while the ERR6 message indicates an electronic error. The manual provides descriptions of each error message and possible solutions. If you encounter an error message that persists, contact the manufacturer for assistance.