DENTSPLY Cavitron Plus Ultrasonic Scaler User manual
The Cavitron Plus is a dental ultrasonic scaler that provides efficient and effective removal of plaque, calculus, and stains from teeth. It features water-based solutions and offers a variety of functions, including purge mode and a water line filter to ensure optimal performance and hygiene.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
BETWEEN PATIENTS:
1. Remove ultrasonic insert used. Clean and sterilize the ultrasonic insert(s) following the procedures outlined in the Cavitron Ultrasonic Insert Infection
Control Direction for Use enclosed with every insert.
2. Hold the handpiece over a sink or drain and activate
Purge function as described in Step 6 of Start-Up
Procedures.
3. After the purge cycle is complete, turn the system
OFF, (O) position.
4. Remove the Steri-Mate handpiece. Clean and sterilize the handpiece following the procedure outlined in the booklet enclosed with your unit.
5. Disinfect the surfaces of the cabinet, Power Cord,
Handpiece Cable, Water Supply Line, Foot Control and Auxillary Cable (if applicable) by applying an approved non-immersion type disinfectant solution* carefully following the instructions provided by the disinfectant solution manufacturer. To clean System, generously spray disinfectant solution on a clean towel and wipe all surfaces. Discard used towel. Dry with a clean cloth. To disinfect system, generously spray disinfectant on a clean towel and wipe all surfaces. Allow disinfectant solution to air dry. Never spray disinfectant solution directly on the system.
6. Inspect the handpiece cable for any breaks or tears.
7. If using a closed water supply or DualSelect
Dispensing system, check for adequate fluid volume for the next patient.
8. When ready for use, place a sterilized Steri-Mate handpiece on the handpiece cable and a sterilized insert into the handpiece and adjust system controls to preferred operator positions.
Use manual. If connected to another device, please follow those directions for use, keeping in mind that a chemical flush should be performed at maximum water flow for at least
30 seconds. The system should be left undisturbed for 10 minutes but no more than 30 minutes to allow the sodium hypochlorite solution to soak in the lines. As a suggestion, it is recommended that a sign be placed on the system stating that the SYSTEM IS BEING DISINFECTED WITH A STRONG
DISINFECTANT AND SHOULD NOT BE USED. When ready, flush system with clean water for at least 30 seconds or until sodium hypochlorite odor disappears. ALL CHEMICALS
MUST BE FLUSHED FROM THE SYSTEM BEFORE IT IS READY
FOR PATIENT USE.
9.3 Water Line Filter Maintenance
When the water line filter becomes discolored, the filter should be replaced to prevent reduced water flow to the
Cavitron Plus Ultrasonic Scaler. A 10-pack of replacement filters is available by ordering Part Number 90158 from your local authorized DENTSPLY Distributor.
1. Verify system is turned OFF.
2. Disconnect the water supply hose from the water source. If a quick-disconnect connector is attached to the end of the hose, relieve the water pressure by pressing the tip of the connector in an appropriate container to drain the water.
3. Grasp the fitting on either side of the filter disk and twist counterclockwise. Remove the filter section from either side of the water hose.
4. Install the replacement filter onto the water hose fittings. The filter should be positioned to match up with the correct hose fitting.
5. Hand tighten the two hose fittings in a clockwise direction. Reconnect the water supply hose, operate the unit to bleed the air and test for leaks.
SHUT-DOWN PROCEDURES AT THE
END OF THE DAY:
Follow the “Between Patients” maintenance procedures,
Steps 1 through 6. In addition, it is recommended to close the manual shut-off valve on the dental water supply system.
SECTION 10:
Troubleshooting
*NOTE: Water-based disinfection solutions are preferred. Some alcohol-based disinfectant solutions may be harmful and may discolor plastic materials.
9.2 Weekly Maintenance
Although service and repair of the Cavitron Plus Ultrasonic
Scaler should be performed by DENTSPLY personnel, the following are some basic trouble shooting procedures that will help avoid unnecessary service calls. Generally, check all lines and connections to and from the System, a loose plug or connection will often create problems. Check the settings on the System’s controls.
It is strongly recommended that this system be disinfected by chemically flushing the waterlines with a 1:10 Sodium
Hypochlorite solution (NaOC1) at the end of each week.
This can be accomplished by connecting this device to the
Cavitron DualSelect Dispensing System or a number of other devices available from your local distributors. Where this device is connected to the Cavitron DualSelect Dispensing
System, please follow the DualSelect system’s Directions for
10.1 Troubleshooting Guide
Symptom:
System will not operate: No Power ON indicator
13
1. Check that the Main Power Switch is in the ON (l)
position, and that the detachable Power Cord is fully
seated in the receptacle on back of System.
2. Check that the system’s power cord plug is fully seated
in an approved AC wall outlet.
3. Check that the wall outlet is functional.
Symptom:
System will not operate: Power ON Indicator is illuminated
1. If the office has more than one foot control, test each to ensure that the proper foot control is being used.
With a handpiece and insert installed, depress the foot control to the first position. The system should dispense water. If none of the foot controls operate the system, continue to the next step.
2. Resynchronize one foot control to the system (see
Section 6.8 Foot Control Synchronization).
Symptom:
System operates: No water flow to insert tip
1. Assure that handpiece lavage control is properly
adjusted.
2. Check for clogged insert.
3. Check that dental office water supply valves are open.
4. If the system is connected to DualSelect Dispensing
System, check that fluid level in the selected bottle is
sufficient. Make sure valves are open when using
external water source.
5. Check that the water line filter is clean. Replace filter
if needed.
Symptom:
System operates: No insert cavitation
1. Check that the Power Level Control is not in Rinse
Mode.
2. Check the insert for damage and that it is properly
installed in the handpiece.
3. Check that the handpiece is properly installed to the
cable assembly.
4. If Steri-Mate grip is used on the handpiece, verify that
the grip is flush with the hard plastic of the insert port.
5. Turn the system’s Main Power Switch to the OFF (0)
position. Wait 5 seconds and turn the system back ON.
6. If problem still exists, replace both “AA” batteries in
foot control with new “AA” batteries (Refer to Section
6.7) or connect auxiliary foot control cable.
Symptom:
System operates: Purge Mode will not function
– icon flashing
1. Check that there is no insert in the handpiece.
2. Check that handpiece is properly installed to the cable
assembly.
Symptom:
System operates: Service Indicator blinking
• Slow Blinking (1 blink per second)
The system is not operating within factory
specifications.
1. Remove insert.
2. Turn Main Power Switch OFF, (O) position. Wait five seconds. Turn unit ON, (I) position.
3. Operate Purge function.
4. If service indicator still blinks, refer to Section 10.2
Technical Support and Repairs to have unit
serviced as soon as possible.
• Fast Blinking (3 blinks per second)
– Indicates improper set-up
1. If insert is in the handpiece, remove. Verify the handpiece is properly seated and depress the foot control for 2 seconds. If blinking stops, the system is ready for use. If blinking remains, continue to the next step.
2. Attach a NEW handpiece and depress foot control for 2 seconds. If blinking stops, the system is ready for use. Discard the old handpiece or return if within warranty. If blinking remains, continue to the next step.
3. Install and fully seat an insert into handpiece.
Depress foot control for 2 seconds. If blinking stops, unit is ready for use. If blinking remains, continue to the next step.
4. Install and fully seat a NEW insert in handpiece and depress foot control for 2 seconds. If blinking stops, system is ready for use. Discard old insert or return if within warranty. If blinking remains, refer to Section 10.2 Technical Support and
Repairs to have unit serviced as soon as possible.
Symptom:
System operates: Service Indicator illuminated
1. Ensure that the base unit has adequate ventilation and
is not near a heat source (i.e. radiator, heat lamp,
sunlight or other heat producing operatory equipment).
2. Turn Main Power Switch to the OFF (O) position. Allow
system to cool for 10 minutes and turn system ON, (I)
position. Verify light is not illuminated.
3. If light is still illuminated, refer to Section 10.2
Technical Support and Repairs to have unit serviced as
soon as possible.
10.2 Technical Support and Repairs
For technical support and repair assistance call DENTSPLY
Professional Cavitron Care SM Factory Certified Service at 1-
800-989-8826 Monday through Friday, 8:00 A.M. to 5:00
P.M. (Eastern Time). For areas outside the U.S., contact your local DENTSPLY Professional representative.
14