The EGT-AMP is a universal high-quality K-type thermocouple amplifier with Cold Junction Compensation. It is designed for nominal 12V supply and it's output gives 0-4,994V voltage signal in 0-1000°C (32-2012°F) sensor temperatures. Because of differential output stage EGT-AMP gives accurate readings also in situations where grounding potential errors are present. EGT-AMP is protected from environmental and electrical hazards normally present on automotive environment.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Thermocouple Amplifier EGT-AMP User Manual - VPV -Motorracing Electronics | Manualzz

-Motorracing Electronics

EGT-AMP USER MANUAL

EGT-AMP User Manual v1.0 11/2011

1/9

-Motorracing Electronics

Table Of Contents

1.

GENERAL ...................................... 3

1.1.

2.

2.1.

SPECIFICATION...............................3

ASSEMBLY .................................... 4

MECHANICAL ASSEMBLY .................4

2.2.

3.

ELECTRICAL ASSEMBLY ...................4

INSTRUCTIONS FOR USE .............. 6

3.1.

THEORY OF THERMOCOUPLE

MEASUREMENT ................................6

3.2.

4.

5.

ELECTRICAL OPERATION .................7

WARRANTY................................... 8

PROOF OF CALIBRATION .............. 9

EGT-AMP User Manual v1.0 11/2011

2/9

-Motorracing Electronics

1. GENERAL

EGT-AMP is a universal high-quality K-type thermocouple amplifier with Cold

Junction Compensation. It is designed for nominal 12V supply and it’s output gives 0-4,994V voltage signal in 0-1000°C (32-2012°F) sensor temperatures.

Because of differential output stage EGT-AMP gives accurate readings also in situations where grounding potential errors are present. EGT-AMP is protected from environmental and electrical hazards normally present on automotive environment.

1.1. SPECIFICATION

Type of thermocouple: K

Thermocouple connector: Miniature, K-type

Gain: 121 x

Output voltage function: 0°C=0V, 1000°C=4,994V

(32°F = 0V, 2012°F=4,994V)

Operational supply voltage: 8.5-16Vdc

Polarity protection: -16Vdc continuous, -45V t<20ms

Supply voltage transient: +40V t<20ms

Supply voltage transient, power: 600W max, 10/1000us

Internal fuse protection: 2 x F2A

Operational temperature range: -40 - +85°C (-40 - +185°F)

Cold junction compensation: -30 - +85°C (-22 - +185°F)

Measurement accuracy: ±3°C (±5,5°F)

Cold junction compensation accuracy: ±5°C (±9°F)

Dimensions (W x D x H): 73.5 x 70 x 21.5mm

Ingress protection: IP57 (not applicable to TC-connector)

Weight: 130g

Mating connector, body: TYCO 794805-1

Mating connector, socket: TYCO 1-770988-0

Mating connector, cable seal: TYCO 794758-1

Mating connector, interface seal: TYCO 794772-4

EGT-AMP User Manual v1.0 11/2011

3/9

-Motorracing Electronics

2. ASSEMBLY

Due to high quality parts used in EGT-AMP it can be assembled in many different conditions; wet, dry, cold or hot. Naturally best possible accuracy and reliability can only be achieved in dry, room temperature atmosphere.

Mechanical fixing is easiest to be done via three (3) ready-made vibration dampened fixing points. Maximum fixing screw diameter is 4mm, M4 bolt or

3.9mm thin sheet screws are suitable. Head of the screw must be at least 9mm

(11/32”) in diameter; otherwise the screw may slide thru rubber grommet. In some cases suitable washers should be used.

Local heat, cold and wind sources should be avoided in the assembly location.

Strong sources of heat or cold may deteriorate the operation of the Cold

Junction Compensation.

EGT-AMP needs a stable +12V feed for operation, +12V feed ( red cable) should be free of excessive disturbances. Feed point near starter, alternator of ignition coil is not recommended; high voltage spikes may interfere EGT-AMP operation or in worst case harm it. Grounding is good when ground cable

(black cable) is short and equipped with a proper grounding connector.

Output of the EGT-AMP is differential and it can adapt to a small ground potential differences (±1,5V) between EGT-AMP and the unit to be controlled.

This potential means the voltage measured between black and blue wire.

Negative Output- cable should always be connected; not doing so will give falsely readings.

Best possible accuracy can only be achieved when negative Output- cable

( blue cable) is connected very near the grounding pin of the controlled unit like seen on Picture 1. With this arrangement a voltage drop in controlled unit grounding cable don’t have effect on measurement. If controlled unit has a separate signal ground, then negative Output- cable should be connected to that. Especially this is important when controlling ECU’s; these have strong ground currents because of injector and ignition control, for example 0.1V voltage drop in ground cable can cause error of 20°C (36°F).

Positive Output+ cable ( yellow cable) should be connected directly to the positive input of the controlled unit.

EGT-AMP output stage is designed to control high impedance inputs; excessive loading of the output can cause measurement errors or in extreme case can cause permanent failure of the EGT-AMP.

Maximum load of the output of EGT-AMP: 5 k Ω

EGT-AMP User Manual v1.0 11/2011

4/9

-Motorracing Electronics

Direct capacitive loading of the EGT-AMP output is forbidden! Especially small loading capacitances may cause oscillations in the output signal; this causes no harm for the EGT-AMP, but measurement accuracy will be lost. If there is a direct capacitive loading on the input of the controlled unit, then a suitable series resistor should be added between the Output+ of the EGT-AMP and the input of the controlled unit, 4.7k

Ω should be ok. Measurement accuracy should be studied case by case after this modification

Never connect Output+ and Output- signals to any high voltage circuits, voltages exceeding -2V - +8V may cause permanent damage to the EGT-AMP!

Pin Signal Cable color

1 +12V Red

2 Ground / - Black

Table 1  Pinning of the EGT‐AMP 

- +

Battery

Picture 1 

EGT-AMP

- +

Sensor

EGT‐AMP electrical connection 

ECU/Datalogger

Analog IN

EGT-AMP User Manual v1.0 11/2011

5/9

-Motorracing Electronics

3. INSTRUCTIONS FOR USE

3.1. THEORY OF THERMOCOUPLE

MEASUREMENT

Thermocouple temperature measurement is based on voltage generation over dissimilar metal junction by the temperature. Voltage is dependant on metals used and the temperature of the junction, voltage generation is nearly linear over usable temperature range of the materials used.

Real measurement system always has two measuring points; the actual measuring point called ‘hot junction’ and the reference junction called ‘cold junction’. Measuring device sees the difference of the voltage produced by these two junctions and to get the actual hot point temperature the measuring system needs to know the absolute temperature of the cold junction. Below is a picture of the measuring principle. t1

Sensor

U = F(t1-t2)

Compensation

cable

U = combined voltage t1 = hot junction temperature t2 = cold junction temperature

Connector t2

Cu

U

Cu

Electronics

Measuring device

Picture 2  Principle of the thermocouple measurement 

Temperature

°C

Temperature

°F

Thermovoltage mV

-50 -58 -1.889

0.000

100 212 4.096

200 392 8.138

300 572 12.209

400 752 16.397

500 932 20.644

29.129

33.275

900 1652

1000 1832 41.276

1100 2012 45.119

Table 2  K‐type sensor output voltage (t2 =0°C (32 °F)) 

EGT-AMP User Manual v1.0 11/2011

6/9

-Motorracing Electronics

-50

0

100

200

300

400

500

600

EGT-AMP is basically an analog amplifier and it has linear input-to-output voltage conversion. Voltage output of the K-type sensor is non-linear over full operational temperature range; a table type conversion from voltage to temperature is needed for best accuracy. In low-end ECUs etc there are only linear type conversions available, this causes some measurement errors that are not induced by EGT-AMP.

For temperature conversion a following table can be used (Table 3) or in simplified case this can be done by function:

0,000V = 0°C, 4,994V = 1000°C (0,000V = 32°F, 4,994V = 2012°F)

Temperature

°C

Temperature

°F

Nominal output voltage

(V)

700

800

900

1000

1100

-58

32

212

392

572

752

932

1112

1292

1472

1652

1832

2012

-0.229

0.000

0.496

0.985

1.477

1.984

2.498

3.014

3.525

4.026

4.516

4.994

5.459

Table 3 

NOTE! Due to the technical reason the output voltage of the EGT-AMP is not limited from 0V to 5V and in extreme temperatures (< 0 °C (<32°F) or >1000 °C (>1832°F)) or in special operational cases (start or sensor failure) the output voltage can be from –2V to +8V. Before connecting

EGT-AMP to any system, please make sure the controlled unit can handle this voltage range without problems. Resistor connected between Output+ of the EGT-AMP and the input of the controlled unit can solve this problem in most cases. Measurement accuracy should be studied case by case after this modification. If there are any questions, please ask from the manufacturer of the devices.

EGT-AMP User Manual v1.0 11/2011

7/9

-Motorracing Electronics

4. WARRANTY

EGT-AMP has a full one (1) year warranty from the date of the purchase.

Warranty includes component failures and workmanship. Not included are natural wear, usage against specification and Force Majeure -type failures.

Warranty does not include failures in other systems connected to EGT-AMP, it is assumed that the end-user of EGT-AMP has skills necessary to assemble and analyse the product so it causes no risk to any other systems even in cases of malfunction and failure. Warranty is valid also in competition use.

EGT-AMP User Manual v1.0 11/2011

8/9

-Motorracing Electronics

5. PROOF OF CALIBRATION

Temperature

(°C)

-50

0

100

200

300

400

500

600

700

800

900

1000

1100

Output Voltage

(V)

Nominal Output

Voltage (V)

Error mV

(5mV=1°C)

-0.229

0.000

0.496

0.985

1.477

1.984

2.498

3.014

3.525

4.026

4.516

4.994

5.459

Table 4  Calibration table 

DATE: _________________ OPERATOR: ____________________________

List of calibration equipment:

Fluke 714 thermocouple calibrator, serial number: 8660083

Fluke 87 multimeter, serial number: 69470043

Hewlett Packard 6632B power supply, serial number: US37476309

Operational voltage: ____________________________V

Test room temperature: _________________°C

Serial number sticker

EGT-AMP User Manual v1.0 11/2011

9/9

advertisement

Key Features

  • K-type thermocouple amplifier
  • Cold Junction Compensation
  • 0-4,994V voltage signal
  • 0-1000°C (32-2012°F) sensor temperatures
  • Differential output stage
  • Protected from environmental hazards
  • IP57 ingress protection
  • Dimensions: 73.5 x 70 x 21.5mm

Frequently Answers and Questions

What type of thermocouple does the EGT-AMP use?
The EGT-AMP uses a K-type thermocouple.
What is the output voltage function of the EGT-AMP?
The EGT-AMP has an output voltage function of 0°C=0V, 1000°C=4,994V (32°F = 0V, 2012°F=4,994V).
What is the operational supply voltage of the EGT-AMP?
The EGT-AMP's operational supply voltage is 8.5-16Vdc.
What is the measurement accuracy of the EGT-AMP?
The EGT-AMP has a measurement accuracy of ±3°C (±5,5°F).
What is the cold junction compensation accuracy of the EGT-AMP?
The EGT-AMP has a cold junction compensation accuracy of ±5°C (±9°F).