Below you will find brief information for OT lamp Mach M3 stand lamp, OT lamp Mach M3 emergency stand lamp, OT lamp Mach M3 wall lamp, OT lamp Mach M3 ceiling lamp. The MachVISION optical system provides an illuminated field with improved contrast and excellent homogeneity. Combined with the Duo-Focus feature this offers a unique possibility for adaptation to the wound area. With this lighting system you profit from a whole range of new developments, based on 50 years of experience in the production of operating and doctor’s lights.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

OT lamp Mach M3 stand lamp, OT lamp Mach M3 emergency stand lamp, OT lamp Mach M3 wall lamp, OT lamp Mach M3 ceiling lamp Operating instructions | Manualzz

Mach M3

Operating instructions

MACH M3

Dr. Mach

Lamps and Engineering

Mach M3 stand lamp, F- / F/H / DF- / DF/H-model

Mach M3 emergency stand lamp, F- / F/H /DF- / DF/H-model

Mach M3, wall lamp, F- / F/H / DF- / DF/H-model

Mach M3 ceiling lamp for room heights up to 2,8m, F- / F/H / DF- / DF/H-model

Mach M3 ceiling lamp for room heights between 2,8m – 3m, F- / F/H / DF- / DF/H-model

Dr. Mach GmbH u. Co., Floßmannstrasse 28, D-85560 Ebersberg

Tel.: +49 (0)8092 2093 0, Fax +49 (0)8092 2093 50

Internet: www.dr-mach.com

, E-mail: [email protected]

59200001 Edition 06 28.11.2002 / Bak page 1/22

Mach M3

Dr. Mach

Lamps and Engineering

List of contents

1. Safety instructions ....................................................................................page 4

2. Operating the lamp MACH M3..................................................................page 5

2.1 Brief description of the lamp ...............................................................page 5

2.1.1 Merging of light fields (F-model)..................................................page 5

2.1.2 Focusing (DF-model) ..................................................................page 5

2.1.3 Light intensity control (/H-model).................................................page 5

2.2 Adjusting the illuminated area by........................................................page 6 merging of light fields (for F and F/H-models)

2.3 Adjusting the illuminated area by........................................................page 6 merging of light fields and focusing –

Duo Focus system (for DF and DF/H-models)

2.4 Overview of the lamp functions ..........................................................page 7

2.5 Adjusting the light intensity .................................................................page 8

2.6 Positioning..........................................................................................page 9

3. Cleaning ...................................................................................................page 10

3.1 Sterilisable handle ..............................................................................page 10

3.2 Lamp housing, dispersing lens and support system...........................page 12

4. Maintenance .............................................................................................page 12

4.1 Adjustments at the lamp head ............................................................page 12

4.2 Removing the upper housing part ......................................................page 13

4.3 Changing of spare parts .....................................................................page 14

4.3.1 Changing the halogen bulbs........................................................page 14

4.3.2 Changing the dispersing lens ......................................................page 15

5. Data..........................................................................................................page 16

5.1 Electrical data.....................................................................................page 16

5.2 Environmental conditions ...................................................................page 16

6. CE-mark ...................................................................................................page 16

7. Disposal....................................................................................................page 16

8. Spare parts ...............................................................................................page 17

9. Spare parts list..........................................................................................page 19

10. Appendix: Setting the primary voltage at the block transformer

300VA (extra) ........................................................................page 20

59200001 Edition 06 28.11.2002 / Bak page 2/22

Mach M3

Dr. Mach

Lamps and Engineering

Dear customer!

Congratulations for achieving our new OT-lamp MACH M3.

The MachVISION optical system provides an illuminated field with improved contrast and excellent homogeneity. Combined with the Duo-Focus feature this offers a unique possibility for adaptation to the wound area.

With this lighting system you profit from a whole range of new developments, based on 50 years of experience in the production of operating and doctor’s lights.

The lighting system is characterised by a previously inconceivable general colour rendition value of Ra = 96. In other words, the colours are reproduced naturally and in high

contrast. The wound area is shown in a comfortable light.

The different reds of a wound area can now be recognised very precisely. For the doctor or surgeon this means a considerable improvement in the recognition of details in the wound

area.

The R96 lighting system uses computer-optimised cold-light filters that reduce both the unwanted build-up of heat in the head area and the heat radiated on the illuminated

wound area to a minimum.

Pay attention to the special mounting instructions for ceiling and wall lamps.

All information quoted here relates only to the illuminants. Details of ceiling or wall installation can be found in the mounting instructions.

59200001 Edition 06 28.11.2002 / Bak page 3/22

Mach M3

Dr. Mach

Lamps and Engineering

1. Safety instructions

Pay attention to the instructions for use when handling the lamp.

WARNING:

This device has not been designed for use in potentially explosive areas.

According to the Medical Device Regulation the lamp is classified under class I.

Store the OT-lamp in its package for at least 24 hours in the respective room before mounting, in order to equal temperature differences.

Please read the instructions for use carefully to make the most of your lighting system and to avoid any damages to the device.

The lamps may only be repaired and special assembly work may only be carried out on the reflector or sockets by ourselves or a company that has been expressly authorised by us.

The manufacturer can only be made responsible for the safety of the lamp if repairs and alterations are carried out by the manufacturer himself or a company that guarantees to observe the safety regulations.

The manufacturer cannot be made liable for personal or material damages if the lamp is operated inexpediently or incorrectly or used for purposes other than those for which it is intended.

The lamp is to be dismantled from the spring arm in reverse order to its assembly.

This may only be carried out after the assembly-locking device has been inserted and secured since the arm is under spring tension.

Make sure that the lamp is in perfect working order before every use.

Attention, external transformer!

The lamp works only with an external transformer 300VA.

The external transformer (Order No. 6701 0208) must be tested and validated according to IEC 60601-1.

If the external transformer is not tested and validated according to IEC 60601-1, it is not allowed to use it with Dr. Mach OT-lamps.

59200001 Edition 06 28.11.2002 / Bak page 4/22

Mach M3

Dr. Mach

Lamps and Engineering

2. Operating the lamp MACH M3

2.1 Brief description of the lamp

The OT-lamp MACH M3 is available in following versions:

-

-

Standard OT-lamp (F) with merging of light fields

Duo-Focus OT-lamp (DF) with merging of light fields and focusing

- F- or DF- lamps with light intensity control (/H)

2.1.1 Merging of light fields

(F- model)

The F-model offers light field adjustment by

merging the individual light fields of each reflector by swivelling them.

NOTE: This feature is activated by turning the sterilisable handle.

2.1.2 Focusing (DF- model)

The DF-model offers a second additional facility to further adjust the illuminated field to an extra small or extra large size, depending on the specific requirements, by focusing

through moving the bulbs up or down.

NOTE: This feature is activated by turning the ring at the top of the sterilisable handle.

2.1.3 Light intensity control (/H- model)

The OT-lamp models offer an optional feature, the light intensity control.

You can adjust the light intensity between 50% and 100%, according to your requirements.

59200001 Edition 06 28.11.2002 / Bak page 5/22

Mach M3

Dr. Mach

Lamps and Engineering

2.2 Adjusting the illuminated area by merging the light fields

(for F and F/H-models)

All F and F/H lamp models offer the feature of

“merging the light fields”. This feature is activated by turning the sterilisable handle 1.

Remark:

The ring 2 has no function here.

4

5

2

1

6

2.3 Adjusting the illuminated area by merging the light fields and focusing – Duo-Focus system

(for DF and DF/H models)

DF- and DF/H-models also have a “focusing” function (Duo-Focus).

The combination of merging of light fields and focusing (Duo-Focus-system) enables the ideal adjustment of the light field to the wound area in a range from 8 cm to 35 cm diameter. Such a large range of adjustment is not possible with conventional technique.

To activate the focusing function turn the adjusting ring 2 on the handle or, -if the lamp is equipped with a keypad at the cardan bow (for DF/H-models) – press touch buttons 5,6.

Press touch button 5 to expand the focus and touch button 6 to bundle the light on a minimum area. The current focused diameter is shown in the LED-display

4.

59200001 Edition 06 28.11.2002 / Bak page 6/22

Mach M3

2.4 Overview of the lamp functions

Merging of light fields and focusing

MERGING OF LIGHT FIELDS

Dr. Mach

Lamps and Engineering

Merging the individual light fields of each reflector is done by swivelling the reflectors. The light fields overlap.

This feature is activated by turning the sterilisable

handle (1).

FOCUSING

Ring (2)

Handle (1)

!

Focusing is done by moving the bulbs inside the reflectors up or down. In this way you can further adjust the illuminated field to an extra small or extra large size, depending on the specific requirements.

This feature is activated by turning the ring (2) at the

top of the sterilisable handle.

Always activate merging of light fields first by turning the handle (1), then activate

focusing when necessary by turning the ring (2)

59200001 Edition 06 28.11.2002 / Bak page 7/22

!

Mach M3

Key pad

1

2

3

Dr. Mach

Lamps and Engineering

2.5 Adjusting the light intensity

(for F/H and DF/H models only)

Select the desired light intensity before the operation.

At these models the adjustment is done by the keypad at the cardan bow. The intensity can be infinitely adjusted between 50% and 100%.

Press push button 2 to increase the light intensity and press push button 3 to reduce the light intensity.

The light intensity set is shown by LED-display 1.

For other light models we recommend a dimmer in the supply line.

59200001 Edition 06 28.11.2002 / Bak page 8/22

Mach M3

2.6 Positioning

1

59200001

1

Use the sterile handle 2 or the handle rails 1 to position the lamps.

Use the handle rail to position the lights before the operation.

Use the sterile handle for positioning during the operation. This handle can be removed for sterilization.

2

1,4m

(1,5m)

1m

0,7m

(0,6m)

0

The working distance of the F-lamps is 0,7m to

1,4m.

The working distance of the DF-lamps is 0,6m to

1,5m.

The handle is supplied with a detectable mechanical catching. When the handle catches in this position, the lamp is optimally set for a working distance of 1 meter.

Two or more lamps can be used for an intensive illumination of large operation fields.

It is also possible to illuminate two operation fields simultaneously.

Edition 06

Dr. Mach

Lamps and Engineering

28.11.2002 / Bak page 9/22

Mach M3

Dr. Mach

Lamps and Engineering

3. Cleaning

3.1 Sterilisable handle

1

V

At delivery the lamp is equipped with the handle

sleeve 1. The handle sleeve is removable and sterilisable. Before using the first time and before every use the handle sleeve must be cleaned, disinfected and sterilised.

The handle sleeve must be removed for sterilisation:

• To remove press the lock V and pull off the sterilisable handle sleeve 1 while keeping the lock pressed.

• To attach, push on and slightly twist the handle until the lock V engages securely.

Handles often become unsterile during an OP; therefore always keep additional handles available for exchange.

Cleaning / disinfection and sterilisation

Basics

Efficient cleaning / disinfection is an essential requirement for effective sterilisation of the handle.

Within the scope of responsibility for the sterility of the products it should be noted that only sufficiently validated equipment and product specific processes are used for cleaning / disinfection and that the validated parameters are complied with in every cycle.

In addition, the hospital / clinic hygiene regulations must be observed.

Cleaning / disinfection

Cleaning and disinfection must be carried out immediately after use.

A mechanised process (disinfector) should be used for cleaning / disinfection. The efficiency of the process used must be recognised and validated in principle (e.g. listed under disinfectants and disinfection procedures tested and recognised by Robert-Koch-Institute / DGHM).

When using other procedures (e.g. a manual procedure), proof and process efficiency in principle must be provided within the scope of validation.

Proof in principle of the suitability of the handles for efficient cleaning / disinfection was provided using a cyclic cleaning system (Netsch-Bellmed T-600-IUDT/AN, programme 2 for small parts; code B).

It is not allowed to use agents / disinfectants, which contain the following substances, as these may cause changes in the material:

- High-concentration organic and inorganic acids

-

-

Chlorinated hydrocarbons

2-ethoxyethanol

59200001 Edition 06 28.11.2002 / Bak page 10/22

Mach M3

Dr. Mach

Lamps and Engineering

When cleaning / disinfecting, the following procedures must be followed:

Process

Zone 1 Pre-rinse, external, cold, 10 – 15°C

Washing, acidic, external 35°C

Re-rinse, external approx. 80°C

Draining

Re-rinse, external approx. 80°C

Draining

Zone 2 Washing, alkaline, external, 93°C

Draining

Re-rinse, external, acidic, 90°C

Draining

Re-rinse, external 90°C time

Zone 3 Drying, external 100 – 120°C

Zone 4 Drying, external 100 – 120°C

Door open / close & transport discharge)

Cycle time overall ca.

45

120

10

*10

*15

*15

15

135

10

10

15

15

15

200

200

60

290

≈ 5 minutes

* When occupying the disinfection zone (washing zone 2), the re-rinse and draining times will depend on the respective objects being washed therein!

Sterilisation

Only previously cleaned and disinfected handles may be sterilised.

The handles are placed in a suitable sterilisation pack (one-way sterilisation pack, e.g. foil / paper sterilisation bags, single or double pack) in accordance with DIN EN 868 / ISO 11607 for steam sterilisation and then sterilised.

Use only the sterilisation procedure listed below for sterilisation. Other sterilisation procedures (e.g. ethylene oxide, formaldehyde and low-temperature plasma sterilisation) are not permissible.

Steam sterilisation procedure

Validated in accordance with DIN EN 554/ISO 11134

Maximum sterilisation temperature 134°C

Proof in principle of the handles’ suitability for effective sterilisation was provided using a fractional vacuum process (Euroselectomat 666 by MMM Münchner Medizin Mechanik GmbH, sterilising temperature 134°C, holding time 7 min.)

Inspection / durability

The handles should be inspected for damage and changed before re-use, if required.

The handles may be cleaned / disinfected, sterilised and re-used for a maximum of 1000 times. If the handles are re-used more than 1000 times, then this will be the responsibility of the hospital / clinic.

59200001 Edition 06 28.11.2002 / Bak page 11/22

Mach M3

Dr. Mach

Lamps and Engineering

Alc. ≤ 20 %

2

3.2 Lamp housing, dispersing lens and support system

The OT-lamp MACH M3 has a high-quality surface, which can be cleaned with conventional cleaning agents.

The lens system 2 is made of a high-quality plastic.

Pay attention to the following during cleaning:

-

-

Never wipe over the lens system 2 with a dry cloth (always clean with a wet cloth).

Only use disinfectants with less than 20%

alcohol.

Wipe the lens system 2 after cleaning with an antistatic, non-fluffy cloth.

4. Maintenance

MACH lamps are supplied with brakes on the suspension fixture and on the lamp housing. Adjust these brakes as necessary after installation.

If the lamp is difficult to move or if it does not stay in position, the brake forces should be adjusted.

The lamps and suspension fixtures have been designed and built, so that regular maintenance intervals are not necessary.

In order to keep the system easy running throughout its life span, we recommend that the hinges should be greased once a year with acid-free grease.

When adjusting the brakes or the hinges at the ceiling attachment please consult the mounting instructions “Ceiling attachment with heavy central axis” or “Ceiling / wall attachment”.

3

Upper housing part

4.1 Adjustments at the lamp head

The fine adjustment for handling the lamps can be carried out at points 2 and 3.

To adjust the brakes use a screw driver with appropriate size.

If you cannot adjust the brakes at points 2 and 3 anymore as tight as necessary, you will have to replace them.

Remark:

In order to adjust brake 2, the upper housing

part must be removed.

4

59200001 Edition 06 28.11.2002 / Bak page 12/22

Mach M3

59200001

Dr. Mach

Lamps and Engineering

5

6

7

8

4.2 Removing the upper housing part

The upper housing part can be removed by turning the turn-lock fastener 8.

The upper housing part has two markings, 6

(Lock) and 5 (Up). The notch 7 at the turn-lock fastener 8 shows the open or close status of the upper housing part.

5

8

Turn the turn-lock fastener 8 to the left over the marking 5 until stop and hold it in this position.

9

10

9

Lift the upper housing part with the other hand at the cardanic suspension 9 and remove it uniformly.

Be careful when removing the upper housing part, do not damage the sealing profile 10!

Mount the upper housing part in reverse order:

Turn the turn-lock fastener 8 again to the left over the marking 5 until stop and hold it in this position.

Put the upper housing part flush on the bottom housing part.

Be careful when mounting the upper housing part, do not damage the sealing profile 10!

Turn the turn-lock fastener 8 under slight pressure on the upper housing part to the right until the marking 6 (Lock).

Check by eye-control, if the upper housing part is mounted uniformly on the bottom housing part of the lamp.

Edition 06 28.11.2002 / Bak page 13/22

Mach M3

4.3 Changing of spare parts

1

3

2

4

Dr. Mach

Lamps and Engineering

4.3.1 Changing the halogen bulbs

Dr. Mach uses special halogen bulbs as illuminants.

Only original Dr. Mach replacement bulbs may be used.

The use of other bulbs can lead to a considerable reduction of the light power and increase in the thermal load.

The halogen bulbs have a service life of approx.

1200 hours. We recommend changing all halogen bulbs in a lamp at the same time to keep maintenance work to a minimum.

Turn off the lamp.

• Remove the upper housing part of the lamp as described in chapter 4.2.

ATTENTION: Parts of the housing and the halogen bulb may be very hot immediately after use.

• Loosen the red knurled screw 1 and remove the lamp socket 2 without removing the electrical connection 3.

• Remove the faulty halogen bulb 4 from the socket.

ATTENTION: Never touch new halogen bulbs with your fingers. Always use the original packaging or a clean cloth.

• Carefully insert the new halogen bulb

(22.8V/50W).

• Re-assemble the lamp in reverse order

59200001 Edition 06 28.11.2002 / Bak page 14/22

1

2

4

2

3

Mach M3

1

1

Dr. Mach

Lamps and Engineering

4.3.2 Changing the dispersing lens

The dispersing lens is made of a high-quality plastic.

In case the dispersing lens looses its optical characteristics, this can reduce the luminous intensity and the light quality, which can be provided.

It may be necessary to exchange the dispersing lens.

To change the dispersion lens proceed as follows:

Always place the dispersing lens on a soft, nonscratching base to preserve its surface.

Turn off the lamp.

• Remove the upper housing part of the lamp as described in chapter 4.2.

ATTENTION: Parts of the housing may be very hot immediately after use.

• Loosen the six screws 1 at the bottom housing part 2 with a cross screw driver.

ATTENTION: Hold the dispersing lens 3 and the retaining ring 4 while loosening the screws 1, so the lens and the ring cannot fall down.

• Remove the dispersing lens 3 and the retaining ring 4 downwards.

• Replace the dispersing lens.

• Fix the dispersing lens and the retaining ring at the bottom part of the lamp housing.

ATTENTION: Secure the screws 1 against loosening with safety lake and adhesive at the screw head.

59200001 Edition 06 28.11.2002 / Bak page 15/22

Mach M3

Dr. Mach

Lamps and Engineering

5.1 Electrical data

Power consumption

Voltage AC/DC

Current

Frequency

Degree of protection

Class of protection

Standards IEC 601-1

5.2 Environmental conditions

5. Data

Mach M3

150 W to max. 162 W, depending on design

22,8 V

6,5 A to max. 7,1 A

50 / 60 Hz

Type B

I.

Relative atmospheric humidity

Air pressure

Operation

Min. Max.

30 %

700 hPa

75 %

1060 hPa

Transport / storage

Min. Max.

Relative atmospheric humidity

Air pressure

20 %

700 hPa

90 %

1060 hPa

6. CE-mark

The products Mach M3 comply with the standards 93/42/EEC for medical products of the

European Community’s Council.

Dr. Mach GmbH is certified according to DIN EN 46001:1996 and DIN EN ISO 9001:1994 by TÜV München.

7. Disposal

The OT-lamp doesn’t contain any danger goods.

The components of the OT-lamp should be properly disposed at the end of its shelf-life.

Make sure, that the materials are carefully separated.

The electrical conducting boards should be submitted to an appropriate recycling proceeding.

The rest of the components should be disposed according to the contained materials.

59200001 Edition 06 28.11.2002 / Bak page 16/22

Mach M3

8. Spare parts

Dr. Mach

Lamps and Engineering

59200001 Edition 06 28.11.2002 / Bak page 17/22

Mach M3

8. Spare parts - continuation

Dr. Mach

Lamps and Engineering

11 12

Bowden control

25

Detail A

38 37 36

Detail A

35

59200001

39

40

Edition 06 28.11.2002 / Bak page 18/22

Mach M3

Dr. Mach

Lamps and Engineering

9. Spare parts list

Item Qty.

1

2

3

1

1

1

Name

Lower housing part for lamp without camera

Lower housing part for lamp with camera

Upper housing part

6

7

1

1

8

9

1

1

10 1

12

1

1

1

Locking sheet metal

Adjusting sheet metal 0,5

Adjusting sheet metal 1,0

Adjusting sheet metal 1,5

Adjusting sheet metal 2,0

Coupling DF for sterilisable handle

Coupling F for sterilisable handle

Coupling DF-M for sterilisable handle

17 1

18 3

19 3

Sealing cord RAL 5018

Reflector unit DF

Reflector unit F

20 12 Fixing screw M3x16 DIN 912 with washer

21 3 Halogen bulb 22,8V 50W

22 3 Socket for halogen bulb

EDVNO Remarks

20012204

20012205

20011201

20088209

20088201

20089208

20088205 alternative

20088206 alternative

20088207 alternative

20088208 alternative

21150002

20220001 DF-model

20220002 F-model

20220003 DF-M-model

DF/H-model

21115003

20088211

20060001 DF-model

20060002

65052012

24 2

25 1

26 1

Fuse 0,63 AT 5x20

Dimmer module OPDIM V3.0

PCB board OPDX-M3 V0.10

31 18 Fixing screw M3x10 DIN 912 with lock washer

32 18 Washer

67100201

67320007

28060213

67370003

24080001 Check version and number

67250101 Check version and number

67030105

23250201

20086001

65052078

65512017

21200201

67610101 Accessory 34 1 Acrylic glass box for transformer 300VA

Bowden control

35 1 Primary shaft

36 1 Screw M3x6:Ni ULF

37 1 Screw M3x6 DIN 912, turned-to-size head

38 1 Tension bolt

39 6 Cable holder

40 1 Round stranded rope

20080201

65052002

28082203

20080206

67900006

20080202

59200001 Edition 06 28.11.2002 / Bak page 19/22

Mach M3

10. Appendix

Setting the primary voltage at the block transformer 300VA (extra)

Clamps on the transformer

Dr. Mach

Lamps and Engineering

3

Pos.1 earthing connection at the transformer

Pos.2 primary

1

2

• Connect the transformer to earth. Use the provided clamps pos.1 at the transformer, respectively at the transformer plate.

See also the explanations on page 19/46 of the mounting instructions “Ceiling and wall attachment” no. 59500001 / edition 05.

Overview of the possible settings of the primary voltage

Setting for115V

Setting for 132V

59200001 Edition 06 28.11.2002 / Bak page 20/22

Mach M3

Dr. Mach

Lamps and Engineering

Setting for 230V

Setting for 247V

59200001 Edition 06 28.11.2002 / Bak page 21/22

Mach M3

Dr. Mach

Lamps and Engineering

59200001 Edition 06 28.11.2002 / Bak page 22/22

advertisement

Key Features

  • Duo-Focus feature
  • Improved contrast and homogeneity
  • Light intensity control
  • Sterilisable handle
  • High-quality surface
  • Dispersing lens made of high-quality plastic

Frequently Answers and Questions

How do I adjust the illuminated area of the Mach M3 lamp?
You can adjust the illuminated area by merging the light fields or by using the Duo-Focus system, which combines merging of light fields and focusing. Refer to the instructions on pages 6 and 7 for detailed steps.
How do I change the halogen bulbs in the Mach M3?
Follow the instructions on page 14 for changing the halogen bulbs. Remember to use only original Dr. Mach replacement bulbs.
How do I clean and sterilize the Mach M3 handle?
The handle sleeve is removable and sterilisable. Before using for the first time and before every use, the handle sleeve must be cleaned, disinfected and sterilised. Follow the instructions on page 10 for details.
What is the working distance of the Mach M3 lamp?
The working distance of the F-lamps is 0.7m to 1.4m. The working distance of the DF-lamps is 0.6m to 1.5m.
How often should I grease the hinges of the Mach M3 lamp?
We recommend greasing the hinges with acid-free grease once a year to keep the system easy running throughout its life span.