BALLUFF BTL7 A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _ linear transducer User's Guide

BALLUFF BTL7 A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _ linear transducer User's Guide

The BTL7 A/E501-M_ _ _ -A/B/Y/Z(8)-S32/S115/KA _ is a linear transducer designed to measure displacement, velocity, and differential position. It utilizes a magnetostrictive effect to generate a torsional wave that travels along a waveguide, and the travel time of the wave is used to calculate the position.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

BALLUFF BTL7 A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _ User's Guide | Manualzz
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
User's Guide
english
www.balluff.com
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
1
Notes to the user
5
1.1
1.2
1.3
1.4
5
5
5
5
2
Safety
6
2.1
2.2
2.3
6
6
6
3
Construction and Function
7
3.1
3.2
3.3
7
8
8
4
Installation
5
Startup
11
5.1
11
11
11
12
12
12
4.1
4.2
4.3
5.2
5.3
5.4
Intended use
General safety notes for the transducer system
Meaning of the warnings
Construction
Function
LED indicator
9
Installation variants
Preparing for installation
Installing the transducer
4.3.1 Installation recommendation for hydraulic cylinders
Electrical connection
5.1.1 Connector version S32/Cable version KA_ _
5.1.2 Connector type S115
Shielding and cable routing
Starting up the system
Operating notes
9
9
10
10
6
Configuring using the Micropulse Configuration Tool
13
6.1
6.2
6.3
13
13
13
7
Setting using the calibration device
14
7.1
7.2
7.3
Calibration device
Calibration procedure notes
Calibration procedure overview
7.3.1 Teach-in
7.3.2 Adjusting
7.3.3 Reset
14
14
15
15
16
16
Teach-in using the calibration device
17
Adjusting using the calibration device
18
Resetting all values using the calibration device (Reset)
20
8
9
10
www.balluff.com
Validity
Symbols and conventions
Scope of delivery
Approvals and markings
Micropulse Configuration Tool
Connecting the USB communication box
Configuration possibilities
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
11
Technical Data
21
11.1
11.2
11.3
11.4 11.5
11.6
21
21
21
21
21
21
12
Accessories
22
12.1
12.2
12.3
12.4
22
22
23
23
13
14
english
Accuracy
Ambient conditions
Supply voltage (external)
Outputs
Communication lines La, Lb
Dimensions, weights
Magnets
Mounting nuts
Connectors
USB communication box
Ordering code
24
Appendix
25
14.1 Converting units of length
14.2 Part label
25
25
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
1
Notes to the user
1.1
Validity
This guide describes the construction, function and
setting possibilities for the BTL7 Micropulse Transducer
with analog interface. It applies to models BTL7-A/E501M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _ (see part numbering code on page 24).
The guide is intended for qualified technical personnel.
Read this guide before installing the transducer and
placing it in operation.
1.2
1.4
UL authorization
File No.
E227256
US Patent 5 923 164
The US patent was awarded in connection with this product.
The CE Mark verifies that our products
meet the requirements of EC Directive
2004/108/EG (EMC Directive).
Symbols and conventions
Enumerations are shown in list form with bullet points
Action instructions are indicated by a preceding triangle.
The result of an action is indicated by an arrow.
► Action instruction 1.
⇒ Action result.
Cross-references indicate where additional information
on the topic can be found (see "Technical data" on page
21).
The transducer meets the requirements of the following
Generic Standards:
– EN 61000-6-2 (noise immunity)
– EN 61000-6-4 (Emission)
Emission tests:
–
Note, tip
This symbol indicates general notes.
Œ ‚
1.3
–
–
–
Approvals and markings
RF emission
EN 55011 Group 1, Class A and B
Noise immunity tests:
These symbols indicate the buttons on the
calibration device.
–
Static electricity (ESD)
EN 61000-4-2
Severity Level 3
Symbols of this type indicate the LED indicators.
–
Electromagnetic fields (RFI)
EN 61000-4-3
Severity Level 3
–
Electrical fast transients (Burst)
EN 61000-4-4
Severity Level 3
–
Surge EN 61000-4-5
Severity Level 2
–
Conducted interference induced
by high-frequency fields
EN 61000-4-6
Severity Level 3
Magnetic fields
EN 61000-4-8
Severity Level 4
Scope of delivery
BTL7 transducer,
calibration device,
condensed guide.
–
The magnets are available in various form
factors and must be ordered separately.
www.balluff.com
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
2
2.1
Safety
Intended use
The BTL7 Micropulse Transducer together with a machine
controller (PLC) comprises a displacement measurement
system. It is fitted into a machine or piece of equipment for
its application. Flawless function in accordance with the
specifications in the technical data is ensured only when
using original BALLUFF accessories, and use of any other
components will void the warranty.
Opening of the transducer or non-approved use are not
permitted and will result in loss of warranty and void any
liability claims against the manufacturer.
2.2
General safety notes for the transducer system
Installation and startup are to be performed by trained
technical personnel only.
The operator is responsible for ensuring that local safety
regulations are observed.
In particular, the operator must take measures ensure that
a defect in the displacement measurement system will not
result in hazards to persons or equipment.
When defects and non-clearable faults in the transducer
occur, take it out of service and secure against unauthorized use.
english
2.3
Meaning of the warnings
Note!
The pictogram used with the word "Caution"
warns of a possible hazardous situation affecting the health of persons or equipment damage.
Ignoring these warnings can result in injury or
equipment damage.
► Always observe the described measures for
preventing this danger.
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
3
Construction and Function
Output signal rising:
Error signal
100 %
0%
Null point
End point
B: 30-1 mm
Z: 2" -0.04" Nominal stroke =
measuring area
BTL7…KA_ _
BTL7…S32
Magnet ‚
Thread size
B: M18×1.5
Z: 3/4"-16UNF
Type
D1
G
...-A/B/Y/Z-... 10.2 mm Thread
M4 × 4/6 deep
...-A/B/Y/Z8-... 8 mm
No thread
Mounting surface
Calibration device
Ø D1
BTL7…S115
Damping
zone 

G
Mounting surface
A: 30-1 mm
Y: 2" -0.04"

‚
Non-usable area
 Not included
‚ Thread size
A: M18×1.5
Y: 3/4"-16UNF
Fig. 3-1: BTL7 Transducer… construction and function
3.1
Construction
Electrical connection: Depending on the version the
electrical connection is made via a cable or a connector
(see part numbering code on page 24).
BTL housing : Aluminum housing containing the processing electronics.
Mounting thread: It is recommended that the transducer
be installed using its mounting threads.
– BTL7-…-A/B: M18×1.5,
– BTL7-…-Y/Z: 3/4"-16UNF.
The transducers having Ø 10.2 mm include an additional
thread at the rod end to be used for support with long
stroke lengths.
Magnet: Defines the position to be measured on the
waveguide. Magnets are available in various form factors
and must be ordered separately (see accessories on
page 22).
Nominal stroke length: Defines the travel/length area
available. Rods with various nominal stroke lengths are
available depending on the model:
– Ø 10.2 mm: Nominal stroke length from 25 mm to
7600 mm;
– Ø 8 mm: Nominal stroke length from 25 mm to
1016 mm.
Damping zone: Unusable zone at the end of the rod
which may not be entered without generating a spurious
signal.
Calibration device: Additional device for calibrating the
transducer.
www.balluff.com
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
3
3.2
Construction and function (cont.)
Function
3.3
The BTL7 transducer contains the waveguide which is
protected by an outer stainless steel tube (rod). A magnet
is moved along the waveguide. This magnet is attached to
the moving member of the equipment whose position is
being determined.
LED indicator
LED 1
LED 2
The magnet defines the position to be measured on the
waveguide.
An internally generated INIT pulse together with the magnetic field from the magnet generates a torsional wave in
the waveguide through the magnetostrictive effect which
propagates along the waveguide at ultrasonic velocity.
The component of the torsional wave which arrives at the
end of the waveguide is absorbed in the damping zone to
prevent reflection. The component of the torsional wave
which arrives at the beginning of the waveguide is converted by a coil into an electrical signal. The travel time of
the wave is used to calculate the position. Depending on
the BTL7 version, this information is made available as a
voltage or current with rising or falling gradient.
The following functions can be selected for the output
values:
– Position
– Velocity (with or without sign)
– Differential position
2 outputs are provided which can be allocated independently of each other. 2 magnets may be used.
The entire range of functions can only be
configured using the "Micropulse Configuration
Tool" PC software. For this the USB communication box must be connected (see Accessories on page 23). The transducer can be calibrated without PC software by using the
calibration device (see page 14 ff).
english
Fig. 3-2: BTL7 LED indicators
In normal operation the LEDs indicate the
operating states of the transducer. LED 1 is
assigned to Output 1, LED 2 is assigned to
Output 2.
In calibration mode the LEDs indicate additional
information (see page 17 ff).
LED 1/LED 2
Operating status
green
Normal function (magnet is within the
measuring range)
red flashing
Measuring range left (magnet is outside
the measuring range)
red
Error (no magnet or magnet outside the
limits)
Tab. 3-1: LED indicators in normal operation
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
4
4.1
Installation
Installation variants
4.2
Non-magnetizable material
Preparing for installation
Installation variant: For holding the transducer and
magnet we recommend non-magnetizable material.
min. Ø D2
Non-magnetizable
material
Magnet
min. Ø D2: Minimum hole diameter
Fig. 4-1: Installation variants in non-magnetizable material
Magnetizable material
When using magnetizable material, the transducer must be
protected from magnetic disturbances by suitable means
(e.g. spacer ring made of non-magnetizable material,
sufficient distanced from strong external magnetic fields).
Horizontal mounting: For horizontal mounting with
nominal stroke lengths > 500 mm we recommend fastening the tube at its end (only possible for Ø 10.2 mm tube)
or supporting it.
Hydraulic cylinders: When installing in a hydraulic cylinder, be sure to keep the minimum value for the hole diameter of the piston (see Tab. 4-1).
Threaded hole: The transducer comes with an M18×1.5
(ISO) or 3/4"-16UNF (SAE) mounting thread. Depending on
the model, the threaded hole must be prepared before
installation.
Spacer made of non-magnetizable material
Magnet
min. Ø D2
Fig. 4-3: Threaded hole M18×1.5 per ISO 6149, O-Ring 15.4×2.1
Spacer made of non-magnetizable material
Magnet
min. Ø D2: Minimum hole diameter
Fig. 4-2: Installation variants in magnetizable material
Tube diameter
Hold diameter D2
10.2 mm
minimum 13 mm
8 mm
minimum 11 mm
Tab. 4-1: Hole diameter for installation in a hydraulic cylinder
www.balluff.com
Fig. 4-4: Threaded hole 3/4"-16UNF per SAE J475, O-ring 15.3×2.4
Magnets: Various magnets are available for the BTL7
transducer (see Accessories on page 22).
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
4
4.3
Installation (continued)
Installing the transducer
Note!
Improper installation can compromise the
function of the transducer and result in
increased wear.
► The flange surface of the transducer must
make full contact with the mounting surface.
► The hole must be perfectly sealed (O-ring/
flange gasket).
► Construct threaded hole (and, if needed, recess for the
O-ring) as per Fig. 4-3 and Fig. 4-4.
► Thread the transducer into the threaded hold (max.
tightening torque 100 Nm).
► Install magnet (accessory).
► Nominal stroke length 500 mm or longer: Fasten rod at
its end (only possible for 10.2 mm rod) or support it.
Appropriate nuts for the mounting thread are
available as accessories (see page 22).
4.3.1 Installation recommendation for hydraulic
cylinders
When sealing the hole using a flat gasket, the max. operating pressure is reduced based on the larger surface subjected to pressure
The slide element may be threaded or glued on.
► Secure the screw from loosening or becoming
detached.
► Use a suitable glue.
Slide surface
Flow gap
Piston rod
BTL outer rod
Slide element
Fig. 4-6: Detail view and top view of slide element
There must be a sufficient gap between the slide element
and the piston hole to allow for the flow of hydraulic fluid.
Methods of attaching the magnet:
– Screws
– Threaded ring
– Press-in
– Indentation (center punching)
The hole in the spacer must be matched to the slide
element for optimal guiding of the protective tube.
Magnet
Magnet attachment
For horizontal installation in hydraulic cylinders (nominal
stroke lengths > 500 mm) we recommend using a slide
element.
Detailed dimensioning is the responsibility of the
cylinder manufacturer.
The slide element material must be matched to the load
situation, the medium used and the relevant temperatures.
Possible materials include Torlon, Teflon or bronze.
Spacer
Slide element
Magnet
Fig. 4-7: Magnet attachment
Fig. 4-5: Installation in a hydraulic cylinder
10
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
5
5.1
Startup
Electrical connection
Depending on the model, the electrical connection is made
using a cable (KA_ _) or a connector (S32 or S115).
The wire assignments or pinout depends on the respective
model (see Tab. 5-1, 5-2).
5.1.1 Connector version S32/Cable version KA_ _
S32
KA
Interface BTL7-…
PIN
Cable color
-A501
-E501
1
YE yellow
not used
4…20 mA2)
2
GY gray
3
PK pink
4
RD red
5
GN green
6
BU blue
GND3)
7
BN brown
+24 V
8
WH white
Lb (communication line)
1)
0V
10…0 V
20…4 mA
2)
2)
Fig. 5-1: Pinout for S32 connector
(topview of connector on transducer)
La (communication line)
0…10 V2)
not used1)
Tab. 5-1: Pinouts connector S32/KA_ _
5.1.2 Connector type S115
PIN
Interface BTL7-…
-A501
-E501
1
0 V (Pin 3)
2
3
0 V (Pin 5)
10…0 V
4
5
20…4 mA
2)
2)
Fig. 5-2: Pinout for S115 connector
(topview of connector on transducer)
La (communication line)
0…10 V2)
4…20 mA2)
6
GND3)
7
+24 V
8
Lb (communication line)
Tab. 5-2: Pinout for S115 connector
1) Unused wires may be connected to GND on the controller side, but not to the shield.
2) Factory settings, freely configurable using PC software.
3) Reference potential for supply voltage and EMC GND.
www.balluff.com
english
11
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
5
Startup (cont.)
5.2
Shielding and cable routing
5.3
Defined ground!
The transducer and the control cabinet must be
at the same ground potential.
Shielding
To ensure electromagnetic compatibility (EMC) observe the
following:
– Connect transducer and controller using a shielded
cable.
Shield: Braided copper shield with minimum 80%
coverage.
– Connector version: Area connection of shield in connector with the connector body.
– Cable version: Connect cable shield to housing on
transducer side.
Ground cable shield on controller side (connect to
protection ground).
Magnetic Fields
The transducer system is a magnetostrictive system.
Ensure that there is sufficient distance between the transducer and between the holding cylinder and strong, external magnetic fields.
Note!
When starting up and if the displacement
measurement system is part of a closed loop
system whose parameters are not yet set, the
system may exhibit uncontrolled movements.
This could result in personal injury and equipment damage.
► Startup should be performed only by trained
technical personnel.
► Observe safety instructions of the equipment or system manufacturer.
1. Check connections for tightness and correct polarity.
Replace damaged connections.
2. Turn on the system.
. Check measured values.
Especially after replacing the transducer or after repair
by the manufacturer, check for correct null point and
end point values. If necessary recalibrate the transducer (see page 13 ff).
5.4
–
Cable routing
Do not route the cable between the transducer, controller
and power supply near high voltage cables (inductive stray
noise is possible).
Bending radius for fixed cable
The bending radius for fixed cable must be at least twice
the cable diameter.
Cable length
Linear
Transducer
Maximum
cable length1)
BTL7-A
30 m
BTL7-E
100 m
Cable diameter
6 - 8 mm
Tab. 5-3: Cable lengths
1) Prerequisite: Construction, shielding and routing preclude the effect of
any external noise fields.
12
english
Starting up the system
–
Operating notes
Check the function of the transducer and all associated
components on a regular basis.
Take the displacement measurement system out of
operation whenever there is a malfunction. Secure the
system against unauthorized use.
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
6
6.1
Configuring using the Micropulse Configuration Tool
Micropulse Configuration Tool
6.3
The Micropulse Configuration Tool allows Balluff transducers type BTL7-A/E501... to be quickly and simply configured on the PC.
The most significant features are:
– Online display of the current position of the magnet
– Graphical support for setting the functions and curves
– Display of information about the connected transducer
– Selectable number formats and units for display
– Resetting to factory settings is possible
– Calibration device can be locked
– Demo mode without having a transducer connected
Note!
The PC software and associated manual can
be obtained in the Internet at
www.balluff.com/downloads-btl7.
6.2
Connecting the USB communication box
For transducer models BTL7-A/E501-M…-S32 and BTL7A/E501-M…-S115 (S32 or S115 connector) the communication box must be inserted between the transducer and
controller. The communication box is connected to the PC
using a USB cable.
USB cable
Connection cable
approx. 0.3 m with
S32 or S115
connector (male)
8-pin
6-pin
Controller
Connection cable
approx. 0.3 m with
S32 or S115
connector (female)
Fig. 6-1: Connecting the communication box using S32 or S115
connector
8-pin
USB cable
Prerequisites
– USB communication box connected to transducer and
PC.
– Software correctly installed.
– Transducer connected to power supply.
– Magnet on transducer.
Magnets and outputs
– 2 Magnet selectable
– 2 Outputs independently assignable
Output functions
Position: Position in the measuring range.
Velocity: Velocity of the magnet, with the sign indicating
the direction of motion. Movement from the start point to
the end point is output with a plus sign, and movement
from the end point to the start point with a minus sign.
Velocity (no sign): Velocity of the magnet, but the direction of motion cannot be read.
Differential position: Distance between two magnets.
This selection is only allowed if two magnets are selected.
Curve freely configurable
– The start and end points can be read (teach-in) or
defined by using the mouse.
– The curve can be scaled by adjusting the output
values.
– The curve can be inverted or copied from the other
output.
– The limits can be adjusted to the measuring range.
– The error value can be set according to the limits.
Disabling the calibration device
The function for manually setting the transducer using the
calibration device can be locked, so that configuration is
only possible using the PC software.
Note!
The transducers may be connected to the
communication box only for configuration. After
configuration the communication box must be
removed.
Controller
For a BTL7-A/E501-M…-KA_ _ transducer the communication lines La and Lb are connected to the USB communication box.
Configuration possibilities
Connection cable approx.
0.6 m with luster terminal
Fig. 6-2: Connecting the communication box with cable connection
www.balluff.com
english
13
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
7
7.1
Setting using the calibration device
Calibration device
Automatic deactivation!
If the buttons on the calibration device are not
pressed for approx. 10 min., the Programming
mode is automatically quit.
The calibration device is an ancillary unit for setting the
transducer when it is not possible to configure it with PC
and software.
Requirements
– Button function is unlocked by the PC software (button
lock).
– The calibration device is in place.
– The transducer is connected to the system controller.
– Voltage or current values from the transducer can be
read (using a multimeter or the system controller).
Values for null and end point
– Any desired position of the magnet can be used as the
null or end point. The null and end points may not however be reversed.
– The absolute null and end points must lie within the
minimum or maximum limits of what can be output.
– The distance between the null point and end point
must be at least 4 mm.
If the button function is locked, the lock must
be lifted using the PC software.
The last set value are always saved, regardless
of whether the setting was made using the
buttons or automatically after 10 min. have
expired.
Settings
The transducer has the following factory settings which
can be adjusted using the calibration device:
– One magnet,
– "Position" function on both outputs, Output 1 with
rising curve and Output 2 with falling curve,
– Identical measuring range for both outputs.
During setting the start and end points for both outputs are
set the same. No other function may be selected and no
additional magnet used.
If the button function is locked, the lock can only be lifted if
these conditions are met.
7.2
Calibration procedure notes
Button 1 (blue)
► Before calibrating: Place the calibration device on the
connection side of the transducer.
► When finished: Remove the calibration device to
prevent changes.
► Keep the calibration device for later use.
Button 2 (gray)
Fig. 7-1: Calibration device in place
End point
Null point
Output gradient
Linear Transducer
Unit
Min. value
Null value
End value
Max. value
Error value
Rising
BTL7-A…
V
-0.5
0
+10.0
+10.5
+10.5
BTL7-E…
mA
3.6
4.0
20.0
20.4
3.6
BTL7-A…
V
+10.5
+10.0
0
-0.5
-0.5
BTL7-E…
mA
20.4
20.0
4.0
3.6
3.6
Falling
Tab. 7-1: Value table for factory settings
14
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
7
7.3
Setting using the calibration device (cont.)
Calibration procedure overview
► Move magnet to the new end position.
► Read end point by pressing the buttons.
⇒ The present null point is retained.
7.3.1 Teach-in
The factory set null and end point is replaced by a new null
and end point. The null and end points can be set independently of each other, and the output gradient changes.
The detailed procedure for teach-in is described
on page 17.
New end point
before
Steps:
► Move magnet to the new null position.
► Read null point by pressing the buttons.
⇒ The present end point is retained.
after
New measuring length 100 %
Fig. 7-3: Reading the new end point
► The curves can be inverted by pressing the buttons.
⇒ When inverting, the curves for both outputs are
inverted. For example, the rising curve of Output 1
becomes a falling curve and the falling curve of
Output 2 becomes a rising curve.
New null point
Fa
llin
g
before
g
isin
R
after
Fig. 7-2: Reading the new null point
www.balluff.com
english
15
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
7
Setting using the calibration device (cont.)
7.3.2 Adjusting
7.3.3 Reset
The factory set null and end point can be replaced by a
new start and end point and the associated output values
can be adjusted. The start and end values can be adjusted
as desired up to the limits.
Restoring the transducer to its factory settings.
The detailed procedure for adjustment is
described starting page 18.
Steps:
► Move magnet to the new start position.
► Read new start point by pressing the buttons.
► Set new start value by pressing the buttons.
before
after
New start value
Start point
Fig. 7-4: Adjusting the new start value
► Move magnet to the new end position.
► Read end point by pressing the buttons.
► Set new end value by pressing the buttons.
New end value
before
after
End point
Fig. 7-5: Adjusting the new end value
16
english
The detailed procedure for the reset is
described on page 20.
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
8
Teach-in using the calibration device
Note!
Teach-in while the system is running may result
in malfunctions.
► Stop the system before performing teach-in.
LED indicator Displayed values (example)
Output:
– Transducer with magnet within measuring range
Activate teach-in
► Œ Hold down for at least 4 s.
>4s
1
Set null point
► Bring magnet to the new null point.
►
Œ Hold down for at least 2 s.
> 2 s
Set end point
► Bring magnet to the new end point.
‚ Hold down for at least 2 s.
at 4…20 mA
5.39 V
9.15 mA
5.39 V
9.15 mA
1.04 V
4.82 mA
0.00 V
4.00 mA
9.89 V
19.13 mA
10.00 V
20.00 mA
1
⇒ After releasing, the new null point is set.
►
LED1 LED2 at 0…10 V
2
> 2 s
⇒ After releasing, the new end point is set.
Invert curves
► Hold down Œ and ‚ together for at least 4 s, until
both LEDs come on red.
1
2
1
2
> 4 s
⇒ After releasing, the curves for both outputs are
inverted.
Quit Teach-in
► Press Œ and ‚ briefly (< 1 s).
⇒ After releasing, the present position value is displayed.
< 1 s
The individual steps for the settings can be
selected as desired. Teach-in can be quit at any
desired point in time.
LED legend:
LED not on
LED green solid on
LED red
www.balluff.com
LED 1 and LED 2 flashing green-green in
alternation
english
17
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
9
Adjusting using the calibration device
Note!
Adjustment while the system is running may
result in malfunctions.
► Stop the system before performing adjustment.
LED indicator Displayed values (example)
LED1 LED2 at 0…10 V
Output:
– Transducer with magnet within measuring range
Activate adjusting
► ‚ Hold down for at least 4 s.
> 4 s 2
Set start point
► Bring magnet to the new start point.
►
Œ Hold down for at least 2 s.
at 4…20 mA
5.39 V
9.15 mA
5.39 V
9.15 mA
1.04 V
4.82 mA
0.00 V
4.00 mA
0.00 V
4.00 mA
2.00 V
7.20 mA
2.00 V
7.20 mA
> 2 s 1
⇒ The new start point is set with the last valid start
value.
Adjust start value
► The start value can be changed using Œ and ‚ 1).
The curve gradient changes (see page 16).
1
1
2
2
► Quit calibration procedure: Briefly press Œ and ‚ at
the same time (< 1 s).
⇒ Set position value is saved.
< 1 s 1
2
Set end point, adjust end value and quit adjusting, see
page 19.
The individual steps for the settings can be
selected as desired. Adjusting can be quit at
any desired point in time.
1) Press button briefly: Present value is increased or
decreased by approx. 1 mV or 1 mA .
If a button is held down longer than 1 s, the step
interval is increased.
LED legend:
LED not on
LED green solid on
LED 1 and LED 2 flashing green-red in alternation
LED 1 and LED 2 flashing red-red in alternation
18
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
9
Adjusting using the calibration device (cont.)
Note!
Adjustment while the system is running may
result in malfunctions.
► Stop the system before performing adjustment.
LED indicator Displayed values (example)
Set end point
► Bring magnet to the new end point.
►
‚ Hold down for at least 2 s.
> 2 s
at 4…20 mA
9.89 V
19.13 mA
10.00 V
20.00 mA
10.00 V
20.00 mA
8.00 V
16.80 mA
8.00 V
16.80 mA
2
⇒ The new end point is set with the last valid end
value.
Adjust end value
► The end value can be changed using Œ and ‚ 1).
The curve gradient changes (see page 16).
LED1 LED2 at 0…10 V
1
1
2
2
► Quit calibration procedure: Briefly press Œ and ‚ at
the same time (< 1 s).
⇒ Set position value is saved.
< 1 s 1
2
< 1 s 1
2
Quit adjusting
► Press Œ and ‚ briefly at the same time (< 1 s).
⇒ After releasing, the present position value is displayed.
1) Press button briefly: Present value is increased or
decreased by approx. 1 mV or 1 mA .
If a button is held down longer than 1 s, the step
interval is increased.
LED legend:
LED not on
LED green solid on
LED 1 and LED 2 flashing red-green in alternation
LED 1 and LED 2 flashing red-red in alternation
www.balluff.com
english
19
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
10
Resetting all values using the calibration device (Reset)
Note!
Resetting the values while the system is running
may result in malfunctions.
► Stop the system before performing the
reset.
The Reset function can be used to restore all the settings
to the factory settings. For a Reset the magnet may also
be located outside the measuring range.
LED indicator
LED1 LED2
1. Activate Reset
► Hold down Œ and ‚ simultaneously for at least 4 s.
► Release buttons.
>4s
1
2
> 4 s
1
2
⇒ Reset is activated.
2. Reset
► Hold down Œ and ‚ for at least 4 s.
► Release buttons.
⇒ All values are reset.
⇒ Current position value is displayed.
⇒ Reset is deactivated.
LED legend:
LED not on
LED green-red flashing
20
english
LED 1 and LED 2 flashing green-red together
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
11
Technical Data
11.1 Accuracy
11.3 Supply voltage (external)
Typical values at 24 V DC and room temperature. BTL7A501-… and BTL7-E501-… Nominal stroke length BTL =
500 mm.
Ready immediately, full accuracy after warmup phase.
Together with magnet BTL-P-1013-4R,
BTL-P-1013-4S, BTL-P-1014-2R, BTL-P-1012-4R or
BTL-P-0814-GR-PAF.
Voltage, stabilized
10 to 30 V DC
Ripple
≤ 0.5 Vpp
Current draw (at 24 V DC)
BTL7-A501-…
BTL7-E501-…
≤ 150 mA
≤ 180 mA
Inrush
≤ 500 mA/10 ms
Polarity reversal protected
to 36 V
Overvoltage protection
to 36 V
Dielectric strength GND to
housing
500 V AC
For special versions, other technical data may
apply.
Special versions are indicated by the suffix -SA
on the part label.
Repeatability
Voltage, typical
Current, typical
±10 µm
±5 µm
Sampling rate
With one magnet:
Depends on nominal stroke length
For nominal stroke length = 500 mm
With two magnets:
Depends on nominal stroke length
For nominal stroke length = 500 mm
Non-linearity at
nominal stroke length ≤ 500 mm
nominal stroke length > 500 to
≤ 5500 mm
nominal stroke length > 5500 mm
Temperature coefficient
(nominal stroke length = 500 mm,
magnet in middle of measuring range)
250 µs1) … 5.4 ms
500 µs
375 µs1) … 8.1 ms
750 µs
±50 µm
±0.01 % FS
±0.02 % FS
11.4 Outputs
BTL7-A501
Output voltage
max. configurable3)
Factory settings
Load current
-10…10 V / 10…-10 V
0…10 V / 10…0 V
≤ 5 mA
BTL7-E501
Output current
max. configurable3)
Factory settings
Load resistance
0…20 mA / 20…0 mA
4…20 mA / 20…4 mA
≤ 500 Ohms
Short circuit protected
Signal line to 36 V
Signal line to GND
11.5 Communication lines La, Lb
≤ 30 ppm/K
Short circuit protected
Signal line to 36 V
Signal line to GND
11.2 Ambient conditions
Operating temperature
-40 °C to +85 °C
Storage temperature
-40 °C to +90 °C
Relative humidity
< 90 %,
non-condensing
Pressure rating (when installed in
hydraulic cylinders),
for Ø 8 mm
for Ø 10.2 mm
Shock rating per IEC 60068-2-27
2)
Diameter of outer rod
8 mm or 10.2 mm
nominal stroke length for
Ø 8 mm
Ø 10.2 mm
≤ 1016 mm
≤ 7600 mm
≤ 250 bar
≤ 600 bar
Weight (depends on length)
approx. 2 kg per m
Housing material
Anodized aluminum
150 g/6 ms
Outer rod material
Stainless steel 1.4571
Outer rod wall thickness
for Ø 8 mm
for Ø 10.2 mm
0.9 mm
2 mm
Young's modulus
approx. 200 kN/mm2
Mounting thread
M18×1.5 or
3/4"-16UNF
Continuous shock per IEC 60068-2-292) 150 g/2 ms
Vibration per IEC 60068-2-62)
(note resonant frequency of the outer
rod)
Degree of protection per IEC 60529
Connector S32/S115 (when
attached)
Cable KA_ _
11.6 Dimensions, weights
20 g, 10 – 2000 Hz
IP67
IP682)
1) Position output only. For velocity output, next higher sampling rate
(500 ms or 750 ms).
2) Individually determined according to Balluff Factory Standard
3) Only using PC software (Configuration Tool)
www.balluff.com
english
21
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
12
Accessories
Accessories not included in scope of delivery and must be
ordered separately.
12.1 Magnets
BTL-P-1013-4R
Magnet
BTL-P-1013-4R, BTL-P-1013-4S, BTL-P-1012-4R,
BTL-P-1014-2R:
Weight:
approx. 10 g Housing:
Anodized aluminum
Operating temperature: -40 °C to +85 °C
BTL-P-0814-GR-PAF:
Weight:
approx. 2 g
Housing:
Ferrite bound in PA
Operating temperature: -40 °C to +85 °C
BTL-P-1013-4S
Included with BTL-P-1013-4R, BTL-P-1013-4S, BTLP-1012-4R magnet:
Spacer:
8 mm, Polyoxymethylene (POM)
Magnet BTL5-P-4500-1:
(electromagnet)
BTL-P-1012-4R
Weight:
approx. 90 g
Housing:
Plastic
Operating temperature: -40 °C to +60 °C
12.2 Mounting nuts
–
Mounting nut M18×1.5:
BTL-A-FK01-E-M18×1.5
–
Mounting nut 3/4“-16UNF:
BTL-A-FK01-E-3/4“-16UNF
BTL-P-1014-2R
BTL-P-0814-GR-PAF
Fig. 12-1: Magnet installation dimensions
22
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
12
Accessories (cont.)
12.3 Connectors
BKS-S32M-00:
– Straight connector, for user assembly
– BKS-S32M-00: M16 per IEC 130-9, 8-pin
BKS-S116-PU-_ _:
– Right-angle connector, molded-on cable, preassembled
– M12, 8-pin
– Various cable lengths can be ordered,
e.g. BKS-S116-PU-05: Cable length 5 m
approx. 62
�����
��
Fig. 12-2: BKS-S32M-00 connector
BKS-S33M-00:
– Right-angle connector, for user assembly
– BKS-S33M-00: M16 per IEC 130-9, 8-pin
approx. 54
Fig. 12-5: Connector BKS-S116-PU-_ _
12.4 USB communication box
Fig. 12-3: BKS-S33M-00 connector
BKS-S115-PU-_ _:
–
–
–
Straight connector, molded-on cable, preassembled
M12, 8-pin
Various cable lengths can be ordered,
e.g. BKS-S115-PU-05: Cable length 5 m
Fig. 12-4: Connector BKS-S115-PU-_ _
www.balluff.com
BTL7-A-CB01-USB-S32:
– For BTL7-A/E501… with S32 connector.
– Included in scope of delivery:
USB communication box, USB cable, 2 adapter cables
0.3 m long each, condensed guide
BTL7-A-CB01-USB-S115:
– For BTL7-A/E501… with S115 connector.
– Included in scope of delivery:
USB communication box, USB cable, 2 adapter cables
0.3 m long each, condensed guide
BTL7-A-CB01-USB-KA:
– For BTL7-A/E501… with cable connection.
– Included in scope of delivery:
USB communication box, USB cable, 1 adapter cable
approx. 0.6 m long, condensed guide
english
23
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
13
Ordering code
Example:
BTL7 - A 5 0 1 – M0500 – B – S32
Linear Transducer
Interface
A = Analog interface, voltage output 0…10 V (factory settings)
E = Analog interface, current output 4…20 mA (factory settings))
Supply voltage
5 = 10 to 30 V DC
outputs
01 = 2 outputs, configurable
Nominal stroke
M0500 = Metric specification in mm, nominal stroke length 500 mm
Construction
A = Metric mounting thread M18×1.5, rod diameter 10.2 mm
B = Metric mounting thread M18×1.5, O-ring, rod diameter 10.2 mm
Y = Inch thread 3/4"-16UNF, rod diameter 10.2 mm
Z = Inch thread 3/4"-16UNF, O-ring, rod diameter 10.2 mm
A8 = Metric mounting thread M18×1.5, rod diameter 8 mm
B8 = Metric mounting thread M18×1.5, O-ring, rod diameter 8 mm
Y8 = Inch thread 3/4"-16UNF, rod diameter 8 mm
Z8 = Inch thread 3/4"-16UNF, O-ring, rod diameter 8 mm
Electrical connection
S32 = Connector S32, M16 per IEC 130-9, 8-pin
S115 = Connector S115, M12, 8-pin
KA05 = Cable, 5 m
24
english
BTL7-A/E501-M_ _ _ _-A/B/Y/Z(8)-S32/S115/KA_ _
Micropulse Linear Transducer - Rod-Style
14
Appendix
14.1 Converting units of length
14.2 Part label
1 mm = 0.0393700787 inch
Conversion table:
mm
inch
1
0.03937008
2
0.07874016
3
0.11811024
4
0.15748031
5
0.19685039
6
0.23622047
7
0.27559055
8
0.31496063
9
0.35433071
10
0.393700787
USB-Configurable
BTL7-A501-M0500-B-S32
BTL_ _ _ _
07112200054321 DE
Type
Order code
Serial number
Tab. 14-1: Examples of conversion for mm-inch
1 inch = 25.4 mm
Fig. 14-1: Part label BTL7
Conversion table:
inch
mm
1
25.4
2
50.8
3
76.2
4
101.6
5
127
6
152.4
7
177.8
8
203.2
9
228.6
10
254
Tab. 14-2: Example for converting inch-mm
www.balluff.com
english
25
World Headquarters
Germany
Balluff GmbH
Schurwaldstrasse 9
73765 Neuhausen a.d.F.
Phone +49 7158 173-0
Fax +49 7158 5010
Service hotline
+49 7158 173-370
[email protected]
Subsidiaries
Australia
Balluff-Leuze Pty. Ltd.
Sensors for Automation
2 Rocco Drive
Scoresby VIC 3179
Melbourne
Phone 03/97 64 23 66
Fax 03/97 53 32 62
Austria
Gebhard Balluff
VertriebsgesmbH
Industriestrasse B16
2345 Brunn am Gebirge
Phone 0 22 36/3 25 21-0
Fax 0 22 36/3 25 21 46
Brazil
Balluff Controles
Elétricos Ltda.
Rua Francisco Foga, Nr. 25
Distrito Industrial
de Vinhedo
cep: 13280.000
Sao Paulo
Phone 0 19/3 8 76 56 11
Fax 0 19/3 8 76 27 63
Canada
Balluff Canada Inc.
2840 Argentia RD Unit 2
Mississauga On
L5N 8G4
Phone 9 05/ 8 16-14 94
1-8 00-9 27-96 54
Fax 9 05/8 16-14 11
Czech Republic
Balluff CZ, s.r.o
Pelušková 1400
198 00 Praha 9
Phone/Fax
02/81 94 00 99
Fax 02/81 94 00 66
France
Balluff SAS
ZI Nord De Torcy-Bat 3
Rue des Tanneurs - BP 48
77201 Marne La Vallée
Cedex 1
Phone 01/64 11 19 90
Fax 01/64 11 19 91
Great Britain
Balluff Ltd.
The Automation Centre
Finney Lane, Cheadle,
Cheshire SK8 3DF
Phone 0161 437 1234,
Fax 0161 436 1435
Poland
Balluff Sp. z o.o.
ul. Smolenskiego 4-7
01-698 Warszawa
Phone 22/8 33 15 64,
8 33 15 82, 8 33 58 60
Fax 22/8 33 09 69
Hungary
Balluff Elektronika Kft
Pápai u. 55
8200 Veszprém
Phone 0 88/42 18 08
Fax 0 88/42 34 39
Singapore
Balluff Asia Pte. Ltd.
BLK 1004 Toa Payoh
Ind. Park
Lorong 8, #03-1489
Singapore 319076
Phone 62 52 43 84
Fax 62 52 90 60
Balluff Elektronika Kft
Marketing Iroda
Vihar u. 22/4
1221 Budapest XXII
Phone/Fax 1/2 29 20 95
Italy
Balluff Automation s.r.l.
Corso Giambone 63
10134 Torino
Phone 0 11/3 15 07 11
Fax 0 11/3 17 96 65,
3 17 01 40
Japan
Balluff Co. Ltd.
Ishikawa Bldg. 2nd Fl.
1-5-5 Yanagibashi,
Taito-Ku
Tokio 111-0052
Phone +81 03 5833-5440
Fax +81 03 5833-5441
Switzerland
Balluff Sensortechnik AG
Riedstrasse 6
8953 Dietikon
Phone 043/322 32 40
Fax 043/322 32 41
USA
Balluff Inc.
8125 Holton Drive
Florence,
Kentucky 41042-0937
Phone 8 59/7 27-22 00,
1-8 00-5 43-83 90
Fax 8 59/7 27-48 23
[email protected]
No. 858412-726 E . 00.000000 . Edition 0803; Subject to modification.
www.balluff.com

advertisement

Key Features

  • Analog interface
  • 2 outputs independently assignable
  • USB-configurable
  • High accuracy and repeatability
  • Wide range of nominal stroke lengths
  • IP67/IP68 protection
  • Various mounting options

Frequently Answers and Questions

What is the accuracy of the BTL7 transducer?
The typical accuracy is ±10 µm for voltage output and ±5 µm for current output, with a repeatability of ±10 µm for voltage and ±5 µm for current.
What is the maximum operating temperature of the BTL7 transducer?
The maximum operating temperature is +85 °C.
What are the available mounting options for the BTL7 transducer?
The BTL7 transducer is available in metric and inch mounting threads (M18×1.5 or 3/4-16UNF) and with or without an O-ring. It can be installed in both magnetizable and non-magnetizable materials.
Can I configure the BTL7 transducer without a PC?
Yes, you can configure the transducer using the calibration device, but you can only set the null and end points, and the curve gradients. For full configuration options, you need to use the Micropulse Configuration Tool software.
What is the purpose of the damping zone at the end of the transducer rod?
The damping zone absorbs the torsional wave that arrives at the end of the waveguide, preventing reflections that could affect the accuracy of the measurement.

Related manuals

Download PDF

advertisement