METTLER TOLEDO IND690 Sum weighing terminal Operating instructions

METTLER TOLEDO IND690 Sum weighing terminal Operating instructions
Add to My manuals

Below you will find brief information for weighing terminal IND690 Sum. This weighing terminal is designed for totalizing functions, allowing you to form three different sums: article total, customer sum, and total sum. The article total combines all items with the same article number, while the customer sum aggregates article totals for a specific customer. The total sum provides a comprehensive record of all items totalized, including both individual article sums and sums across all identical database fields. You can also manage inventory by switching between Plus and Minus modes to track receipts and removals. The device allows working with multiple weighing platforms and provides a total record printout with detailed information about each article.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Weighing terminal IND690 Sum Operating Instructions | Manualzz

Operating instructions

METTLER TOLEDO MultiRange

Application software IND690-Sum www.mt.com/support

Congratulations on choosing the quality and precision of METTLER TOLEDO. Proper use according to these instructions and regular calibration and maintenance by our factory-trained service team ensure dependable and accurate operation, protecting your investment. Contact us about a ServiceXXL agreement tailored to your needs and budget.

We invite you to register your product at www.mt.com/productregistration so we can contact you about enhancements, updates and important notifications concerning your product.

IND690-Sum

Contents

5

6

3

4

1.7

1.8

1.9

1.10

1.11

1.12

1.13

1

1.1

1.2

1.3

1.4

1.5

1.6

2

2.1

2.2

Contents

Page

Totalizing functions .....................................................................

4

Documentation .............................................................................

4

Introduction..................................................................................

4

Entering identifications ..................................................................

5

Totalizing items with same article number .......................................

6

Entering known weight values manually ..........................................

7

Cancelling item ............................................................................

7

Ending totalizing and clearing customer sum ...................................

8

Managing inventory ......................................................................

8

Working with several weighing platforms.........................................

9

Recalling application-specific information ........................................

9

Printing total record....................................................................... 10

Recalling total record from computer ............................................... 11

Printing total record on external printer ............................................ 12

Settings in the master mode ........................................................ 13

Overview of the PAC master mode block.......................................... 13

Settings in the PAC master mode block............................................ 14

Application blocks ....................................................................... 17

What to do if …? ......................................................................... 20

Technical data ............................................................................. 22

Index .......................................................................................... 23

Operating instructions 22012831B 08/08

3

Totalizing functions

IND690-Sum

4

1 Totalizing functions

1.1

Documentation

The weighing terminal IND690-… comes supplied with a CD containing all the documentation on the weighing system IND690.

These operating instructions describe the operation and configuration of the application software IND690-Sum.

The basic information for working with the weighing terminal IND690-... can be found in the operating instructions IND690-Base.

PLUS

Start automatic totalizing or totalize items manually

1.2

Introduction

With the IND690-Sum you can form 3 different sums:

• Article total, marked with #

• Customer sum, marked with *

• Total sum, per article marked with @, for all articles marked with §

The article total combines all items with the same article number.

Each article total can be added to a customer sum.

The total sum is formed independently of the article total and customer sum. It stores all items that have been totalized with the IND690-Sum in a database. Both the sums in the database fields of each individual article (@) and the individual sums for all identical database fields (§) are stored. Similar to application blocks, these database fields can be added to the printout configuration, see operating instructions for the

IND690-Base weighing terminal.

A pre-configured total record of the total sum lists all items by article numbers.

Function keys

The function keys of the IND690-Sum are assigned as follows:

A TOT C SUM TOTAL

Print article total, set article total buffer to zero, transmit article total to customer sum buffer

Print customer sum, set customer sum buffer to zero

Print sum report of total sum, save or delete list entries

MAN CANC

Enter weight values or quantities manually

Cancel items and print cancelled values

➜ Select the function by pressing the function key.

Example

➜ Press the MAN key.

Then you can enter a weight value manually with the keypad.

When the function keys are otherwise allocated

➜ Press the cursor keys < or > repeatedly until the function key assignment shown above appears.

Operating instructions 22012831B 08/08

IND690-Sum

Totalizing functions

1.3

Entering identifications

To mark the items identifications must be entered prior to totalizing, e. g. the article number, see "Additional functions" chapter in the operating instructions for the

IND690-Base weighing terminal.

Depending on the setting of the OPERATOR GUIDE in the master mode the

IND690-Sum requests Code F to Code A in sequence.

The identification data buffers have the following standard assignment:

• Code A: ARTICLE NO

• Code B: DESTINATION

• Code C: IDENT C

• Code D: IDENT D

• Code E: CUSTOMER

• Code F: OPERATOR

If the name of the identification data buffers is to be changed in the master mode, make sure that the sum memories list all weighings by the article number or by the identification Code A.

When entering the identifications, no fixed text memories can be recalled with the

IND690-Sum.

Input

➜ Enter identifications and confirm each with ENTER.

After entering Code B the header is automatically printed.

Totalizing is started after entering Code A.

Notes

• The values marked with * only appear when the master mode block OPERATOR

GUIDE is set to TARG. VALUE A+B+C(+D+E+F).

• The first character of the article number (Code A) must be a number, as otherwise the message ENTER NUMBER appears.

All other identifications can be confirmed without entry with ENTER.

• The input of Code A or Code B can be cancelled by pressing the keys CODE A or

CODE B again.

• For article numbers already present in the article database, it is sufficient when all characters (numbers and separating characters) are entered up to the first letter.

Example: The existing article number "1004 Screw" can be called up with "1004 "

(with blank space).

To handle article numbers, see DATABASE master mode block on page 14.

Operating instructions 22012831B 08/08

5

Totalizing functions

IND690-Sum

1.4

Totalizing items with same article number

Totalizing manually

If AUTO TOTALIZING OFF is set in the master mode, the weight values of the items must be copied with the PLUS key.

1. Press PLUS key and enter identifications, see section 1.3.

2. Lay item on weighing platform.

3. Press PLUS key.

The display shows the article number, the serial number and the net value # NET of the article total. The current article total is automatically printed out.

4. Relieve weighing platform.

5. To totalize additional items, repeat steps 2 to 4.

Totalizing automatically

If AUTO TOTALIZING ON is set in the master mode, the weight values of the items are automatically totalized.

If the items are in tare containers of an equal weight, the weighing platform can be tared with one of these containers prior to totalizing. The stored tare weight is then taken into account for all subsequent weighings.

1. Press PLUS key and enter identifications, see section 1.3.

2. Lay item on weighing platform.

The display shows the article number, the serial number and the net value # NET of the article total. The current article total is automatically printed out.

3. Relieve weighing platform.

4. To totalize additional items, repeat steps 2 and 3.

Note

The PLUS key need only be pressed after the 1st item to start automatic totalizing.

Transmitting current article total

➜ Press A TOT key.

The current article total is transmitted to the customer sum buffer and printed out.

Then the article total buffer is set to zero.

The display shows the last article number.

Continuing totalizing

➜ Copy displayed article number with ENTER.

– or –

Enter new article number and confirm with ENTER.

Cancelling totalizing

➜ Cancel entry of article number with the CODE A key.

6

Operating instructions 22012831B 08/08

IND690-Sum

Totalizing functions

Notes

• If the article total or the customer sum memory overflows, the sum saved last is automatically printed and then deleted before copying the current item to the memory. The display shows *SUM OVERFLOW or # SUM OVERFLOW.

• The printout can be formatted as desired, see operating instructions for IND690-

Base weighing terminal.

1.5

Entering known weight values manually

Known net weight values can be copied without weighing or quantities entered.

Manual entries in a running weighing series are stored under the current article number. However, the manual entries can also be stored under a separate article number.

1. With separate article number: Press A TOT key and enter article number for manual entry.

2. Press MAN key.

3. Weight value: Enter known net weight value and confirm with ENTER.

– or –

Pieces: Press the cursor keys < or > and enter the number of pieces in the unit PCS.

4. The display shows the net value of the current article total.

The manual entry is automatically printed and the display shows the net weight value of the article total.

Notes

• The weight unit for entering known weight values can be selected with the cursor keys < or >.

• Weight values can only be entered in the displayed unit.

• The printout can be formatted as desired, operating instructions for IND690-Base weighing terminal.

1.6

Cancelling item

An item of the current article total can be cancelled by weighing back. Manually entered weight values cannot be cancelled.

➜ Leave the item to be cancelled on the weighing platform or place it on the platform again and press the CANC key.

Net, gross and tare weight values of the current item are negatively booked in the article total and customer sum memory.

The cancelled values are automatically printed.

Operating instructions 22012831B 08/08

7

Totalizing functions

IND690-Sum

Notes

• If a negative weight value is formed in the article total or customer sum memory during cancelling, the item is not cancelled and the display shows NEGATIVE SUM.

• The serial number of the item is increased by 1 during cancelling, and the item counter in the article total or customer sum memory is decreased by 1.

• The printout can be formatted as desired, see operating instructions and installation information for printer GA46.

1.7

Ending totalizing and clearing customer sum

After completing a shipment or an order the customer sum can be printed. If new identifications are then entered, a new totalizing process is started automatically.

1. Press C SUM key.

The customer sum is printed and the customer sum buffer set to zero.

2. Enter identifications, see section 1.3.

Notes

• Article data are only listed individually when ENLARGED *SUM ON is set in the master mode.

• The printout can be formatted as desired, see operating instructions and installation information for printer GA46.

• The entries of the customer sum are available in the application blocks

323 … 328, see section 3.

8

1.8

Managing inventory

The default setting for the IND690-Sum is the PLUS MODE, in which it adds together all items with the same article number and the item counter increases.

If inventory control is to be carried out with the IND690-Sum, it is possible to switch back and forth between the PLUS MODE and the MINUS MODE to detect inventory receipts and removals.

Weight values adopted in the MINUS MODE are subtracted from the sum and the item counter is reduced.

Prerequisite

ITEMS ACTIVE is set in the master mode.

Inventory removals

1. To switch over to the MINUS MODE, press and hold the PLUS key until MINUS

MODE appears in the display.

2. Totalize the inventory removals, see section 1.4.

IND690-Sum operates in the MINUS MODE until it is switched over to the PLUS

MODE again.

Operating instructions 22012831B 08/08

IND690-Sum

Totalizing functions

Inventory receipts

1. To switch over to the PLUS MODE, press and hold the PLUS key until PLUS MODE appears in the display.

2. Totalizing the inventory receipts, see section 1.4.

IND690-Sum operates in the PLUS MODE until it is switched over to the MINUS

MODE again.

Note

If the inventory level is dropped below, the display shows INVENTORY 0 and the weight value of the current item is not subtracted.

1.9

Working with several weighing platforms

When working with several weighing platforms, the weighing platform on which the individual items are weighed can be selected as desired.

Prior to taring the weighing platform or placing the item on the scale, select the number of the weighing platform, see "Basic functions" chapter in the operating instructions for the IND690-Base weighing terminal.

The number of the weighing platform is indicated on the printouts of the article total and the customer sum.

Note

The weighing sums of all weighing platforms are rounded off with the roughest increments according to the weighing platform.

1.10

Recalling application-specific information

Information on totalizing can be recalled with the following key combinations:

INFO, PLUS

INFO, A TOT

INFO, C SUM

INFO, TOTAL

INFO, MAN

INFO, CANC

Display setting of AUTO TOTALIZING master mode block.

Display current net value of the article total.

Display current net value of the customer sum.

Display current net value of the total sum.

Display last manually entered weight value.

Display last cleared weight value.

Note

If several pieces of information are recalled with one key, the display changes automatically after the set DISPLAY DURATION. It is also possible to switch back and forth between these pieces of information with the CLEAR key.

Operating instructions 22012831B 08/08

9

Totalizing functions

IND690-Sum

1.11

Printing total record

For complete detection of all items, a total record should be printed at regular intervals, e. g. after one day, one week or after completing an entire shipment.

A database entry is created for each article number in the customer sum memory. A database entry consists of the following values:

• Article number

• Gross weight of all items (Gross)

• Net weight of all items (Net)

• Tare weight of all items (Tare)

• Quantities of the manual entries (Hand)

• Item counter (Items)

When the total record is printed out you can select whether you print the total record for an individual database entry or for all database entries.

The total record lists the database entries and the corresponding total values in the total sum memory.

The total record can be formatted. The article data can be called up as database fields.

Following printing you can store or clear the database entries. The setting of the

DATABASE master mode block determines which data are cleared.

1.11.1

Printing individual database entry

1. Press the TOTAL, CODE A key sequence.

The display shows the article number last entered.

2. To adopt the article number, press the ENTER key.

– or –

Enter a new article number and confirm with ENTER.

The total record for this database entry is automatically printed and the display shows TOTAL RECORD and then SAVE ?

3. To save the database entry, press the ENTER key.

– or –

To delete the database entry, press the CLEAR key.

Note

If the entered article number is not present in the total sum memory, the display shows NO. NOT FOUND.

10

Operating instructions 22012831B 08/08

IND690-Sum

Totalizing functions

1.11.2

Printing all database entries

1. Press the TOTAL, ENTER key sequence.

The total record is automatically printed and the display briefly shows TOTAL

RECORD and then SAVE?.

2. To save all database entries, press the ENTER key.

– or –

To delete all database entries, press the CLEAR key.

1.12

Recalling total record from computer

If the interface is set to DIALOG MODE in the INTERFACE master mode block, the total record of the total sum can be recalled from the computer.

The following interface commands are available for this purpose:

• Print total record for all list entries and then save:

{A[W|3[0[6|_|$[$|0[5}

(corresponds to the TOTAL, ENTER key sequence)

• Print total record for all list entries and then delete:

{A[W|3[0[6|_|$[$|1[5}

(corresponds to the TOTAL, CLEAR key sequence)

In accordance with the printout of the total sum, the database entries and the customer sums are listed in the total sum memory in sequence.

In the process, the data are transmitted in their maximum length in accordance with the data formats in the application blocks.

During the data transmission, the keypad of the IND690-Sum is locked and the display shows DAILY LOG. TO EDP.

Note

After pressing the CLEAR key or when the total sum memory overflows, the

IND690-Sum waits approximately 2 seconds before it deletes the article total memory. During this time the computer can request data transmission via the interface.

Operating instructions 22012831B 08/08

11

Totalizing functions

IND690-Sum

1.13

Printing total record on external printer

If REPORT EXTERNAL is set in the master mode, the total record of the total sum can be printed out on an external form printer.

In the PRINT MODE the printout can be formatted, see INTERFACE master mode block in the operating instructions for the IND690-… weighing terminal.

Here article data can be called up as database fields.

Example of printout on a certified printer

Total record

Date 06/06/06 Time 09-05-47

Last deletion Date 06/06/01 Time 09-25-41

ARTICLE

1000.ART01

1100.ART02

1200.ART03

1300.ART04

Total

GROSS

1000 kg

1500 kg

1000 kg

2500 kg

§ Gross 6000 kg

NET

800 kg

1200 kg

800 kg

2200 kg

§ Net 5000 kg

TARE

200 kg

300 kg

200 kg

300 kg

§ Tare 1000 kg

ITEMS

2

4

3

1

§ Items 10

12

Operating instructions 22012831B 08/08

IND690-Sum

2 Settings in the master mode

2.1

Overview of the PAC master mode block

In this block the following system settings can be carried out:

Settings in the master mode

Legend

• Blocks on a grey background are described extensively in the following.

• Factory settings are printed in bold type.

Operating instructions 22012831B 08/08

13

Settings in the master mode

IND690-Sum

AUTO TOTALIZING

2.2

Settings in the PAC master mode block

Switch automatic totalizing on or off

Factory setting: AUTO TOTALIZING OFF

ENLARGED *SUM Also divide up customer sum printout by article numbers

Factory setting: ENLARGED *SUM OFF

DATABASE

EDIT

DATABASE

PRINT DATABASE

CLEAR DATABASE

CLEAR TOTAL

CLEAR ALL

Edit article data

Edit database entries, see page 16.

Print a list of all database entries.

Clear only selected data when deleting a database entry in the total sum memory.

Clear only weight sums and item counters; the article entry with the article number is retained (factory setting).

Clear article number, all other identification data, weight sums and item counters.

ITEMS

ITEMS ACTIVE

ITEMS PASSIVE

Comment

Switch item counter for inventory control on or off

For inventory receipts the item counter is increased with the PLUS key, for inventory removals it is decreased (factory setting).

Switch off item counter; the display shows the article total and a serial number.

The item counter is available in application block 317, see section 3.

REPORT

REPORT GA46

REPORT EXTERNAL

Select printer for total record of total sum

Print total record on printer GA46; the printout cannot be formatted (factory setting).

Print total record on a form printer; the printout can be formatted as desired.

14

Operating instructions 22012831B 08/08

IND690-Sum

Settings in the master mode

OPERATOR GUIDE

TARG. VALUE A+B

TARG. VALUE A+B+C

Interrogation of the identification data memory prior to totalizing

Code B and Code A are always requested (factory setting).

Code C, Code B and Code A are always requested.

TARG. VALUE A+B+C+D Code D, Code C, Code B and Code A are always requested.

TARG. VALUE

A+B+C+D+E

Code E, Code D, Code C, Code B and Code A are always requested.

TARG. VALUE

A+B+C+D+E+F

Code F, Code E, Code D, Code C, Code B and Code A are always requested.

MIN. CONTROL Set minimum load control

If the current weight value is less than the value of the MIN. CONTROL, the display shows INSUFFICIENT WEIGHT.

Possible values: 1 … 99 digit (factory setting: 10)

DISPLAY

STANDARD

EXTENDED

TOTAL REPORT

Select display presentation

Article number can only be displayed with 14 characters.

If entered, the article number is displayed immediately after entry in a larger font on the middle line. Up to 20 characters can be displayed.

Switch the total report on/off

If a total is not required, the total report can be deactivated.

If the total report is deactivated, no total sum is added up and the TOTAL key is not displayed. The error message EL is displayed at the AW_306 ... interface command.

Factory setting: TOTAL REPORT ON

Operating instructions 22012831B 08/08

15

Settings in the master mode

IND690-Sum

RESET PAC Reset all functions to the factory settings

Block

AUTO TOTALIZING

ENLARGED *SUM

CLEAR DATABASE

Factory setting

off off clear total

ITEMS

OPERATOR GUIDE

MIN. CONTROL

TOTAL REPORT active enter A+B

10 digit on

Editing database entries

Function keys

The function keys in the block EDIT DATABASE are assigned as follows for editing database entries:

<

Display previous database entry

>

Display next database entry

F

u

NEW

Select function of function key

F5:

NEW,

SEARCH, EDIT

SEARCH

EDIT

DEL

The database entry can be edited with function key F5:

Accept changes and return to higher level block

NEW Create new database entry. Here the first character of the article number

(Code A) must be a number, as otherwise the message ENTER

NUMBER appears.

SEARCH Search for an existing database entry.

EDIT

CLEAR

Edit an existing database entry.

Clear displayed database entry.

16

Operating instructions 22012831B 08/08

Application blocks

IND690-Sum

3 Application blocks

In the following description, the application blocks are shown in the syntax for the

MMR command set. When used with the SICS command set, please observe the

SICS conventions, see Operating instructions for IND690-Base weighing terminal.

No.

301

302

305

306

310

311

312

313

314

315

316

Content

Pac version

Format

Response:

{A[B|_|IND690-Sum_Vx.xx_}

Program number

Keypad entry or readin barcode

Total record

Response:

{A[B|_|IP66-0-0xxx_}

Response:

{A[B|_|Entry}

Write:

{A[W|3[0[5|_|$|$|Entry}

Comment: Entry = Text_20, number or weight value

Write:

Print and save total record

{A[W|3[0[6|_|$[$|0[5}

TOTAL, ENTER key sequence

Print and delete total record

{A[W|3[0[6|_|$[$|1[5}

TOTAL, CLEAR key sequence

Consecutive number Response:

{A[B|_|Number_4}

Print article total: fixed text article total

Response:

{A[B|_|Text_20}

Article total: gross weight (#Gross)

Response:

{A[B|_|Weight value|_|Unit}

Article total: net weight (#Net)

Article total: tare weight (#Tare)

Article total: manual quantity eentries (#Hand)

Article total: item counter (#Items)

Response:

{A[B|_|Weight value|_|Unit}

Response:

Response:

Response:

{A[B|_|Weight value|_|Unit}

{A[B|_|Number_10|_|Unit}

{A[B|_|Number_6}

Unit = PCS

317

318_001

318_006

Item counter for inventory control

Identification data

Code A … Code F

Response:

{A[B|_|Number_6}

Response:

{A[B|_|Name (text_20)|_[_|Identification (text_20)}

Write:

{A[W|3[x[x|_|Name (text_20)|$|$|Identification (text_20)}

Comment: xx = 18_001 … 18_006; corresponds to the application blocks 094 … 099

Operating instructions 22012831B 08/08

17

Application blocks

18

IND690-Sum

325

326

327

No.

318

321

322

323

324

328

329

330

331

332

333

Content

Identification data

Code A … Code D

Format

Response: equal to 318_001

Write: equal to 318_001

Comment: xx = 18 … 21; corresponds to the application blocks 094_001 … 094_004

Response:

{A[B|_|Text_20 }

Total record: fixed text total record

Print customer sum: fixed text customer sum

Customer sum: gross weight

(*Gross)

Customer sum: net weight (*Net)

Customer sum: tare weight (*Tare)

Customer sum: manual quantity entries (*Hand)

Customer sum: item counter (*Items)

Total record: date/time at last deletion

Total sum, cancelled items: gross weight

(–Gross)

Total sum, cancelled items: net weight (–Net)

Response:

Response:

Response:

Response:

Response:

Response:

Total record: fixed text for cancelled items

Response:

{A[B|_|Text_20|_[_|Date|_[_|Time}

Write:

{A[W|0[1[9|_|Text_20|$[$|Date|$[$|Time}

Comment: Text: e. g. last deletion

Date: DD.MM.YY, MM.DD.YY

Time: hh.mm.ss, A/PMhh.mm.ss

Response:

{A[B|_|Text_20}

Write:

{A[W|3[1[1|_|Text_20}

Comment: e. g. cancel

Response:

{A[B|_|Weight value|_|Unit}

Response:

{A[B|_|Text_20 }

{A[B|_|Weight value|_|Unit}

{A[B|_|Weight value|_|Unit}

{A[B|_|Weight value|_|Unit}

{A[B|_|Number_10|_|Unit}

{A[B|_|Number_4}

Unit = PCS

{A[B|_|Weight value|_|Unit}

Total sum, cancelled items: tare weight (–Tare)

Response:

{A[B|_|Weight value|_|Unit}

Operating instructions 22012831B 08/08

IND690-Sum

344

345

346

347

336

337

338

339

340

No.

335

341

342

343

348

Content

Total sum, manual entry: fixed text for printing and weight value

Total record: fixed text total

Format

Response:

{A[B|_|Text_20|_|Weight value|_|Unit}

Write:

Response:

{A[W|3[3[5|_|Text_20|Weight value|_|Unit}

{A[B|_|Text_20}

Total sum: gross weight (§Gross)

Total sum: net weight (§Net)

Total sum: tare weight (§Tare)

Total sum: manual weight entries

(§Hand weight)

Total sum: manual quantity entries (§Hand pieces)

Total sum: item counter (§Items)

Response:

{A[B|_|Weight value|_|Unit}

Response:

{A[B|_|Weight value|_|Unit}

Response:

{A[B|_|Weight value|_|Unit}

Response:

{A[B|_|Weight value|_|Unit}

Response:

{A[B|_|Number_10|_|Unit}

Unit = PCS

Response:

{A[B|_|Number_6}

Last closed article:

Date, time, name of destination, identification of destination, name of article, identification of article

Last closed article:

Gross weight

Response:

{A[B|_|Date|_|Time |

{ |_|Name of destination (Text_20) |

{ |_|Identification of destination (Text_20)|

{ |_|Name of article (Text_20) |

{ |_|Identification of article (Text_20) }

Response:

{A[B|_|Weight value|_|Unit}

Response:

{A[B|_|Weight value|_|Unit}

Last closed article:

Net weight

Last closed article:

Tare weight

Last closed article:

Manual entry of pieces

Last closed article:

Item counter

Response:

Response:

Response:

{A[B|_|Weight value|_|Unit}

{A[B|_|Number_10|_|Unit}

{A[B|_|Number_6}

Operating instructions 22012831B 08/08

Application blocks

19

What to do if …?

IND690-Sum

4 What to do if …?

20

Error / Display

ENTER NUMBER

NOT USED

NO. NOT FOUND

VALUE TOO HIGH

WRONG INPUT

WEIGHT TOO LOW

LOAD SCALE

NEGATIVE NET

NEGATIVE SUM

BUFFER FULL

OVERFLOW PIECES

OVERFLOW TOTAL

Possible causes

• No numerical entry for Code A

• Buffer contains no data

• No database entry exists for entered article number

• Place number for manual entry of pieces exceeded

Remedy

• Manual entry of pieces contains decimal places

• Minimum weight dropped below when totalizing or clearing

➜ Enter quantity without decimal places

➜ Place weight > minimum weight on scale

➜ Place weight on scale or move scale

• Minimum deflection dropped below when totalizing or clearing

• Negative weight values cannot be totalized

➜ Relieve scale and retare

• Clearing would lead to a negative subtotal or total sum

• Max. number of article numbers in total sum buffer reached

➜ Enter article number

➜ Totalize items with + key

➜ Enter existent article number

➜ Enter maximum of 999 999 PCS

➜ Place right item on scale

➜ Print total record and clear

• Manual entry of pieces leads to a sum greater than 999 999

• Totalizing leads to a value greater than

999 999 999 kg or to more than

9 places

• Itemcounter exceeds 999 999

➜ Manual entry of pieces is automatically printed and deleted

➜ Total sum is automatically printed and deleted

➜ Itemcounter starts over at 1

➜ Subtotal is automatically printed and deleted

*SUM OVERFLOW

# SUM OVERFLOW

• Totalizing leads to a value greater than

999 999 999 kg or to more than

9 places

• Itemcounter exceeds 999 999

• Totalizing leads to a value greater than

999 999 999 kg or to more than

9 places

• Itemcounter exceeds 999 999

SERIAL NO. OVERFLOW • Serial number for individual printout is greater than 999 999

➜ Itemcounter starts over at 1

➜ Total sum is automatically printed and deleted

➜ Itemcounter starts over at 1

➜ Serial number automatically starts over at 1

Operating instructions 22012831B 08/08

IND690-Sum

What to do if …?

Error / Display

OFF LIMIT

STOCK 0

Possible causes

• When DeltaTrac is activated, net value is not within set tolerances, therefore cannot be totalized and cleared

• Current item cannot be subtracted, as the stock will be dropped below

Remedy

➜ Conclude weigh-in properly

➜ Restock

Operating instructions 22012831B 08/08

21

Technical data

IND690-Sum

5 Technical data

Totalizing functions

Subtotal memory

All sum memory

Total sum memory

Article management

Serial No.

Item counter

Input routine

Clear function

Manual input

Automatic totalizing

• For gross, net and tare

• Capacity up to 999 999 999 kg or a maximum of 9 places including decimal point

• For totalizing subtotals

• Capacity up to 999 999 999 kg or a maximum of 9 places including decimal point

• For exact article detection of all sub-quantities within certain time period or for completion of a shipment

• Capacity up to 999 999 999 kg or a maximum of 9 places including decimal point

Maximum of 999 article numbers

On every sub-quantity printout, capacity up to 999 999

On every sum printout, capacity up to 999 999

Automatic input request for destination data and article numbers for exact article sum detection

Reverse entry of incorrectly totalized sub-quantities from sum memories

Manual input of known weight values (net) in sum memories or piece inputs for logging on output

Automatic totalizing and printing after placing weighing sample on scale and stabilization of weighing platform

22

Operating instructions 22012831B 08/08

IND690-Sum

6 Index

A

Article total 4

Automatic totalizing 14

C

Cancel item 7

Carry out inventory control 8

D

Database 14

Database entry 10, 14

E

Edit data string 16

Enlarged *sum 14

Enter identifications 5

Enter known weight values manually 7

Error messages 20

F

Function keys 4

M

Minimum load control 15

O

Operator guide 15

P

Print total record 12, 14

R

Recall information 9

Recall total record from computer 11

Reset Pac 16

T

Technical data 22

Total record 10

Total sum 4, 8

Totalize automatically 6

Totalizing 6, 8

Totalizing functions 4, 22

Transmit article total 6

W

What to do if …? 20

Work with several weighing platforms 9

Operating instructions 22012831B 08/08

Index

23

*22012831B*

22012831B

Subject to technical changes © Mettler-Toledo (Albstadt) GmbH 08/08 Printed in Germany

22012831B

Mettler-Toledo (Albstadt) GmbH

D-72458 Albstadt

Tel. ++49-7431-14 0, Fax ++49-7431-14 232

Internet: http://www.mt.com

advertisement

Key Features

  • Totalizing functions
  • Article total, Customer sum, Total sum
  • Inventory management
  • Multiple weighing platforms
  • Detailed total record printout

Frequently Answers and Questions

How can I totalize items with the same article number?
You can totalize items manually by pressing the PLUS key, entering identifications, placing the item on the weighing platform, and pressing PLUS again. Alternatively, if AUTO TOTALIZING ON is set in the master mode, weight values are automatically totalized when you place the item on the platform. You can then transmit the article total to the customer sum buffer by pressing the A TOT key.
How can I manage inventory with this device?
To manage inventory, you can switch between PLUS and MINUS modes. The PLUS mode adds weight values to the sum, while the MINUS mode subtracts them. This allows you to track inventory receipts and removals. To switch to MINUS mode, press and hold the PLUS key until MINUS MODE appears. To switch back to PLUS mode, follow the same procedure.
How do I print a total record?
To print a total record, press the TOTAL key followed by either CODE A or ENTER. If you press CODE A, you can then enter an article number to print a total record for that specific entry. Pressing ENTER will print a total record for all database entries. You can then choose to save or delete the database entries.

Related manuals