Below you will find brief information for Ultrasound Scanning System LOGIQ 200 PRO Series. The LOGIQ 200 PRO Series is a versatile ultrasound system that is designed for a wide range of applications, including abdominal, cardiac, and vascular imaging. It features a variety of advanced technologies that provide high-quality images and improved diagnostic accuracy. The system is also easy to use, with a user-friendly interface that makes it easy for clinicians to perform exams efficiently.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Ultrasound Scanning System LOGIQ 200 PRO Series Service Manual | Manualzz

Technical

Publications

2235374

Revision 6

LOGIQt200 PRO Series

Service Manual

GE Medical Systems

Copyright

E

2000, 2001 By General Electric Co.

Operating Documentation

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time of delivery. If damage is apparent, write “damage in shipment” on all copies of the freight or express bill before delivery is accepted or “signed for” by a General Electric representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if the inspection is not requested within this 14 day period.

GE Medical Systems -- Americas

USA: (1) 800--437--1171

Canada: (1) 800--668--0732

Latin & South Am: (1) 305--735--2304

GE Medical Systems -- Asia

Japan:

Korea

(81) 426--48--2594

(82)342--740--6000

GE Ultrasound -- Europe

Germany (49) 212--2802--207

GE MEDICAL SYSTEMS

REV 6

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

LIST OF EFFECTIVE PAGES

REV DATE

0 August 1. 1999

1 February 10. 2000

2

3

May 10. 2000

July 20. 2000

4 November 22. 2000

5 January 24. 2001

6 August 14. 2001

PRIMARY REASON FOR CHANGE

Inital release

System Version 3.10 release

System Version 3.15 release

System Version 3.16 release

System Version 3.16b release

Manufacturer Name Change

System Version 3.17a release

PAGE REV

Title Page . . . . . . . . . .

GE Logo Page . . . . . .

6

0

2184996

2184997

2184998

. . . . . . . . . . .

. . . . . . . . . . .

. . . . . . . . . . .

0

1

0

A . . . . . . . . . . . . . . . . . .

i to vi . . . . . . . . . . . . . .

6

0

Chapter 1

1--1 to 1--7

1--8 to 1--12

. . . . . . . . . .

. . . . . . . . .

1--13 to 1--19

1

5

. . . . . . .

1

PAGE REV PAGE

1--20 . . . . . . . . . . . . . . .

5

REV

Chapter 4

4--1 to 4--12 . . . . . . . . .

0

Chapter 2

2--1 to 2--16 . . . . . . . . .

0

Chapter 3

3--1 to 3--9

3--10

3--15

. . . . . . . . . .

. . . . . . . . . . . . . . .

3--11 to 3--14 . . . . . . . .

. . . . . . . . . . . . . . .

0

4

0

6

3--16 . . . . . . . . . . . . . . .

0

Chapter 5

5--1 to 5--8 . . . . . . . . . .

0

Chapter 6

6--1 to 6--2 . . . . . . . . . .

0

6--3

6--4

. . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . .

2

6

6--5 to 6--6

6--7

. . . . . . . . . .

. . . . . . . . . . . . . . . .

0

6

6--8 . . . . . . . . . . . . . . . .

0

PAGE REV

6--9 . . . . . . . . . . . . . . . .

6--10 . . . . . . . . . . . . . . .

6--11 . . . . . . . . . . . . . . .

6

0

6

6--12 . . . . . . . . . . . . . . .

6--13 to 6--14 . . . . . . .

0

3

6--15 to 6--16 . . . . . . .

6--17 . . . . . . . . . . . . . . .

0

3

6--18 to 6--19 . . . . . . .

6--20 . . . . . . . . . . . . . . .

6

3

6--21 to 6--44

6--45

. . . . . . .

0

. . . . . . . . . . . . . . .

6--46 to 6--100 . . . . . .

2

0

6--101 . . . . . . . . . . . . . .

6

PAGE REV

6--102 . . . . . . . . . . . . . .

6--103 . . . . . . . . . . . . . .

6--104 to 6--108 . . . . .

0

6

0

Chapter 7

7--1 to 7--14 . . . . . . . . .

0

Chapter 8

8--1 to 8--3 . . . . . . . . . .

8--4 . . . . . . . . . . . . . . . .

8--5 to 8--12 . . . . . . . . .

0

0

0

A

GE MEDICAL SYSTEMS

REV 0

SECTION

1--1

1--2

1--3

1--4

2--1

2--2

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

TABLE OF CONTENTS

TITLE PAGE

CHAPTER 1 -- INTRODUCTION

SERVICE MANUAL CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3

SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--2--1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--2--2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--4

1--4

1--17

EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--1 EMC Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--2 Notice upon Installation of Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--3 General Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--4 Countermeasures against EMC--related Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--5 Notice on Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--18

1--18

1--18

1--19

1--19

1--19

ADDRESS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--20

CHAPTER 2 -- INSTALLATION

PREINSTALLATION

2--1--1 Introduction

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--1--2 Power Line Requirements

2--1--3 Physical Specifications

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--1--4 Recommended Ultrasound Room Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--3

2--3

2--3

2--4

2--5

INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--2 Average Installation Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--3 Installation Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--4 Checking the Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--5 Unpacking LOGIQ 200 PRO Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--6 MTZ Probe Holder Installation (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--7 Transducer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--8 Moving into Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--9 Adjusting System Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--10 Product Locator Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--8

2--8

2--8

2--8

2--8

2--9

2--11

2--14

2--14

2--14

2--15 i

TABLE OF CONTENTS

GE MEDICAL SYSTEMS

REV 0

SECTION

3--1

3--2

3--3

3--4

3--5

3--6

TITLE

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

PAGE

CHAPTER 3 -- SYSTEM CONFIGURATION

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--3

DIMENSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--3

ELECTRICAL SPECIFICATIONS

3--3--1 Power Supply

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--3--2 Facility Power Receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--4

3--4

3--4

STORAGE AND OPERATION REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--4

OPTIONAL PERIPHERALS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--5--1 Peripherals/Accessories Connector Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--5--2 List of Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--5--3 Power Consumption of Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--5

3--5

3--9

3--11

TEST POINT, LED, DIP SWITCH AND RESET SWITCH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--6--1 Test Point List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--6--2 LED List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--6--3 DIP Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--6--4 Reset Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--6--5 Video Signal Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--12

3--12

3--14

3--15

3--15

3--16 ii

TABLE OF CONTENTS

GE MEDICAL SYSTEMS

REV 0

SECTION

4--1

4--2

4--3

5--1

5--2

5--3

5--4

5--5

TITLE

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

PAGE

CHAPTER 4 -- FUNCTIONAL CHECKS

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--1--1 Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--3

4--3

FUNCTIONAL CHECK PROCEDURES

4--2--1 Basic Controls

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--2--2 M--Mode Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--3

4--3

4--5

SMPS ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--3--1 SMPS Assy Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--3--2 SMPS Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--7

4--7

4--10

CHAPTER 5 -- DIAGRAMS

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--3

LOGIQ 200 PRO Series SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--3

BLOCK DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--4

WIRING DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--5

CIRCUIT BOARD DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--7 iii

TABLE OF CONTENTS

GE MEDICAL SYSTEMS

REV 0

SECTION TITLE

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

PAGE

6--1

6--2

CHAPTER 6 -- RENEWAL PARTS

RENEWAL PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--3

DISASSEMBLY/RE--ASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--1 Monitor Assy (FRU No. 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--2 Pot Set Assy (FRU No. 101)

6--2--3 CRT Filter(FRU No. 102)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--4 Monitor Cover Set (FRU No. 103)

6--2--5 Tilt Assy(FRU No. 104)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--6 Filter Clamp Set (FRU No. 105) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--7 Probe Holder (FRU No. 201) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--8 Holder Bracket Assy (FRU No. 202) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--9 Cover Set (FRU No. 203) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--10 Keyboard Assy (FRU No. 204) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--11 Trackball Assy (FRU No. 205) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--12 TGC Assy (FRU No. 206) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--13 Encoder Set (FRU No. 207) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--14 Freeze/Record Key Assy (FRU No. 208)

6--2--15 TGC Knob Set (FRU No. 209)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--16 Rubber Key Set (FRU No. 210) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--17 P.C.Board(s) (FRU No.301 through 304)

6--2--18 BPHV Assy (FRU No. 305)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--19 SCAN Assy (FRU No. 306) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--20 Shield Panel (FRU No. 307) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--21 Mother 3 Assy (FRU No. 308) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--22 Nest Box (FRU No. 309) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--23 DC Fan Assy (FRU No. 310) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--24 Swing Arm Assy (FRU No. 401) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--25 Pipe Assy (FRU No. 402) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--26 Pole Cover (FRU No. 403) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--27 Left Cover (FRU No. 404) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--28 Right Cover (FRU No. 405) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--29 Front Base Cover (FRU No. 406) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--30 Front Cover (FRU No. 407) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--31 Rear Cover (FRU No. 408) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--32 Top Cover (FRU No. 409) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--33 Printer Cover (FRU No. 410) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--34 Printer Bracket Assy(FRU No. 411)

6--2--35 Rear Handle (FRU No. 412)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--36 Neck Frame (FRU No. 413) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--37 Rear Panel Assy (FRU No. 414)

6--2--38 Power S/W Assy (FRU No. 415)

6--2--39 Front Caster (FRU No. 416)

6--2--40 Rear Caster (FRU No. 417)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--64

6--65

6--66

6--68

6--70

6--72

6--74

6--76

6--78

6--80

6--44

6--46

6--48

6--50

6--52

6--54

6--56

6--58

6--60

6--62

6--82

6--84

6--86

6--87

6--29

6--30

6--32

6--34

6--36

6--37

6--38

6--39

6--40

6--42

6--21

6--22

6--24

6--25

6--26

6--27

6--28 iv

TABLE OF CONTENTS

GE MEDICAL SYSTEMS

REV 0

SECTION

6--3

6--4

7--1

7--2

7--3

TITLE

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

PAGE

CHAPTER 6 -- RENEWAL PARTS (continued)

6--2--41 Bumper Set (FRU No. 418) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--42 EMI Cover L (FRU No. 419) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--43 PCB Guide Assy (FRU No. 420) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--44 Terminal Block 12 Assy (FRU No. 501) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--45 Terminal Block 6 Assy (FRU No. 502) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--46 SMPS Assy (FRU No. 503) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--47 Power Trans Assy (FRU No. 504)

6--2--48 Circuit Breaker Set (FRU No. 505)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--88

6--89

6--90

6--92

6--94

6--96

6--98

6--100

FUSE REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--3--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--3--2 Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--101

6--101

6--101

SMPS Assy REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--4--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--4--2 Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--104

6--104

6--104

CHAPTER 7 -- PERIODIC MAINTENANCE

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--1--1 Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3

7--3

PERIODIC MAINTENANCE PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--2--1 Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--2--2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--2--3 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--2--4 Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3

7--3

7--4

7--4

7--4

ELECTRICAL SAFETY TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--1 Outlet Test Wiring Arrangement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--2 Grounding Continuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--3 Chassis Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--4 Probe Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--5 When There’s Too Much Leakage Current...

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--5

7--5

7--6

7--7

7--10

7--14 v

TABLE OF CONTENTS

GE MEDICAL SYSTEMS

REV 0

SECTION

8--1

8--2

TITLE

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

PAGE

CHAPTER 8 -- INSTALLATION FOR OPTIONS

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--3

MOD ASSY INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--2 TOOLS REQUIRED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--3 TIME REQUIRED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--4 PARTS REQUIRED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--5 Functional Check--out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--6 Disassembling Dummy Bracket Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--7 Installing MOD Top Bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--8 Assembling MOD Assy

8--2--9 Jumper Setting

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--10 Installing MOD Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--11 Operational Check--out

8--2--12 Final Procedure

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--13 Renewal Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--5

8--6

8--7

8--8

8--8

8--10

8--11

8--12

8--4

8--4

8--4

8--4

8--4

8--4 vi

TABLE OF CONTENTS

GE MEDICAL SYSTEMS

REV 1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CHAPTER 1 -- INTRODUCTION

TABLE OF CONTENTS

TITLE SECTION

1--1

1--2

1--3

1--4

PAGE

SERVICE MANUAL CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3

SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--2--1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--2--2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--4

1--4

1--17

EMC (Electromagnetic Compatibility) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--1 EMC Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--2 Notice upon Installation of Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--3 General Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--3--4 Countermeasures against EMC--related Issues

1--3--5 Notice on Service

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--18

1--18

1--18

1--19

1--19

1--19

ADDRESS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--20

1--1

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

1--2

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1--1 SERVICE MANUAL CONTENTS

This manual provides service information on the LOGIQ 200 PRO Series Ultrasound Scanning System. It contains the following chapters:

1.

Chapter 1, Introduction: Contains a content summary and warnings.

2.

Chapter 2, Installation: Contains physical and electrical requirements that must be considered prior to installation and a complete LOGIQ 200 PRO Series installation procedure with installation checklist.

3.

Chapter 3, System Configuration: Contains system configuration and specifications.

4.

Chapter 4, Functional Checks: Contains functional checks that must be performed as part of the installation, or as required during servicing and periodic maintenance.

5.

Chapter 5, Diagrams: Contains block diagrams and functional explanations of the LOGIQ 200 PRO Series electronics.

6.

Chapter 6, Renewal Parts: Contains a complete list of replacement parts for the LOGIQ 200 PRO Series and disassembly procedures for all changeable FRU.

7.

Chapter 7, Periodic Maintenance: Provides periodic maintenance procedures for the LOGIQ 200 PRO

Series.

8.

Chapter 8, Installation for Options: is provided to keep the option installation instructions supplied with each option.

1--3

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

1--2 SAFETY

1--2--1 Warnings

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

WARNING!

CAREFULLY READ ALL THE WARNINGS LISTED BELOW!

1.

The operator manual should be fully read and understood before operating the LOGIQ 200 PRO Series and kept nearby for quick reference.

2.

Although the ultrasound energy transmitted from the LOGIQ 200 PRO Series transducer is within

AIUM/NEMA standards, unnecessary exposure should be avoided. Only trained personnel should operate the

LOGIQ 200 PRO Series.

3.

To prevent electrical shock, the LOGIQ 200 PRO Series should be connected to a properly grounded power receptacle. Do not use a three to two prong adapter. This defeats safety grounding.

4.

Do not use with Defibrillator when LOGIQ 200 PRO Series is being operated .

5.

Probes are fragile, please handle with care.

6.

Concerning Outside Markings, refer to Illustration 1--1, 1--2, 1--3, 1--4, 1--5, 1--6, 1--7, 1--8, and 1--9.

7.

For the cleaning, disinfection, and sterilization, refer to Probe section in LOGIQ 200 PRO Series Advanced

Reference Manual (2233222--100) and Caution Sheet supplied with each probe.

NOTICE

This medical equipment is approved, in terms of the prevention of radio wave interference, to be used in hospitals, clinics and other institutions which are environmentally qualified. The use of this equipment in an inappropriate environment may cause some electronic interference to radios and televisions around the equipment. Proper handling of this equipment is required in order to avoid such trouble according to the operator and service manuals.

This equipment can be used in residential areas only under the supervision of physicians or qualified technicians.

1--4

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR ALL UNITS)

ILLUSTRATION 1--1

1--5

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

OUTSIDE MARKINGS OFLOGIQ 200 PRO Series (FOR ALL UNITS)

ILLUSTRATION 1--2

Note

For further details regarding the cautions above, refer to NO TAG MOVING INTO POSITION in

Chapter 2.

1--6

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Defibrillator

Caution

Label

OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR ALL UNITS)

ILLUSTRATION 1--3

1--7

INTRODUCTION

GE MEDICAL SYSTEMS

REV 5

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

MODEL NUMBER : 2270968

Class1/Classe1

MANUFACTURED

LOCATION:

GE MEDICAL SYSTEMS KOREA

S.N.

VOLT

:

: 220--240Vac ~

AMP. LONG TERM : 2.1A

KVA : 0.5kVA

PHASE

DESC.

: 1

: LOGIQ 200PRO Series

CISPR 11 / EN 55011

CLASS : A GROUP : 1

CLASSE : A GROUPE : 1

OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR EUROPE )

ILLUSTRATION 1--4

Note

For the symbols shown in the illustration above, refer to latter pages in this chapter.

1--8

INTRODUCTION

GE MEDICAL SYSTEMS

REV 5

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

For USA

Canada,

Mexico

MADE FOR GE MEDICAL SYSTEMS

MILWAKEE, WISCONSIN BY

GE MEDICAL SYSTEMS KOREA

CLASS 1

MODEL 2270969

SERIAL.

MANUFACTURED

VOLTS : 120Vac ~ 1 PHASE

POWER

FREQUENCY

DESC,

: 500VA

: 60HZ

: LOGIQ 200PRO Series

For Brazil

MADE FOR GE MEDICAL SYSTEMS

MILWAKEE, WISCONSIN BY

GE MEDICAL SYSTEMS KOREA

MODEL

SERIAL.

CLASS 1

2270970

MANUFACTURED

VOLTS : 220--240Vac ~ 1 PHASE

POWER : 500VA

FREQUENCY

DESC,

: 50HZ

: LOGIQ 200PRO Series

OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR USA CANADA MAXICO BRAZIL)

ILLUSTRATION 1--5

1--9

INTRODUCTION

GE MEDICAL SYSTEMS

REV 5

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

MADE FOR GE MEDICAL SYSTEMS

MILWAKEE, WISCONSIN BY

GE MEDICAL SYSTEMS KOREA

MODEL

SERIAL.

MANUFACTURED

CLASS 1

2270967

VOLTS

POWER

:220--240Vac ~ 1 PHASE

: 500VA

FREQUENCY

DESC,

: 50HZ

: LOGIQ 200PRO Series

For SEA, China, Hongkong, India, Australia, New Zealand

MADE FOR GE MEDICAL SYSTEMS

MILWAKEE, WISCONSIN BY

GE MEDICAL SYSTEMS KOREA

CLASS 1

MODEL 2270966

SERIAL.

MANUFACTURED

VOLTS : 120Vac ~ 1 PHASE

POWER

FREQUENCY

DESC,

: 500VA

: 60HZ

: LOGIQ 200PRO Series

For Taiwan

OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR SEA CHINA HONGKONG INDIA AUSTRALLIA NEWZEALAND TAIWAN

PHILLIPPINES)

ILLUSTRATION 1--6

1--10

INTRODUCTION

GE MEDICAL SYSTEMS

REV 5

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

MADE FOR GE MEDICAL SYSTEMS

MILWAKEE, WISCONSIN BY

GE MEDICAL SYSTEMS KOREA

CLASS 1

MODEL 2270971

SERIAL.

MANUFACTURED

:

VOLTS : 220--240Vac~ 1 PHASE

POWER

FREQUENCY

DESC

: 500VA

: 60HZ

: LOGIQ 200PRO Series

OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR CHILE)

ILLUSTRATION 1--7

1--11

INTRODUCTION

GE MEDICAL SYSTEMS

REV 5

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR KOREA)

ILLUSTRATION 1--8

1--12

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Do not use a Defibrillator simultaneously with the ECG, as its excessive voltage will damage the signal input block of the ECG unit.

OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR UNITS WITH ECG)

ILLUSTRATION 1--9

1--13

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

1--2--1 Warnings (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

NOTE: Two labels are attached on rear of the SMPS assy box inside the rear cover and front of BPHV Board Assy inside the front base cover.

MARKINGS OF LOGIQ 200PRO Series (INSIDE COVERS)

ILLUSTRATION 1--10

1--14

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1--2--1 Warnings (Continued)

The following table describes the purpose and location of safety labels and other important information provided on the equipment.

Label/Icon

Identification and Rating Plate

Type/Class Label

IP Code

(IPX1)

Purpose/Meaning

Manufacturer’s name and address

Date of manufacture

Model and serial numbers

Electrical ratings (Volts, Amps, phase, and frequency)

Used to indicate the degree of safety or protection.

Location

Rear of console near power inlet

Indicates the degree of protection provided by the enclosure per IEC 529. IPX1 indicates drip proof.

Foot Switch

Equipment Type BF (man in the box symbol)

IEC 878-02-03 indicates B Type equipment having a floating applied part.

Probe connectors

Equipment Type CF (heart in the box symbol) IEC 878-02-05 indicate equipment having a floating applied part having a degree of protection suitable for direct cardiac contact.

ECG connector and surgical probes

Device Listing/

Certification Labels

Laboratory logo or labels denoting conformance with industry safety standards such as UL or IEC.

Rear of console

“DANGER -- Risk of explosion used in...”

The system is not designed for use with flammable anesthetic gases.

Rear of console

“CAUTION” The equilateral triangle is usually used in combination with other symbols to advise or warn the user.

Various

“ATTENTION -- Consult accompanying documents” is intended to alert the user to refer to the operator manual or other instructions when complete information cannot be provided on the label.

Various

1--15

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

1--2--1 Warnings (Continued)

Label/Icon

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Purpose/Meaning

“CAUTION -- Dangerous voltage” (the lightning flash with arrowhead in equilateral triangle) is used to indicate electric shock hazards.

Various

Location

“Mains OFF” Indicates the power off position of the mains power switch.

Front of system,

Main power switch

“Mains ON” Indicates the power on position of the mains power switch.

Front of system,

Main power switch

“Equipotentiality” Indicates the terminal to be used for connecting equipotential conductors when interconnecting (grounding) with other equipment.

CAUTION

This is only for ”FUNCTIONAL

GROUNDING”, NOT ”PROTECTIVE

EARTH”.

Rear of console

Indicates Main protective earth terminal

Various

1--16

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1--2--2 Specifications

Type of protection against electric shock: Class I EQUIPMENT (*1)

Degree of protection against electric shock: Type BF EQUIPMENT (*2) (Except ECG)

Type CF EQUIPMENT (*3) (ECG Only)

Ordinary Equipment

Continuous Operation

*1. Class I EQUIPMENT

EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but which includes an additional safety precaution in that means are provided for the connection of the

EQUIPMENT to the protective earth conductor in the fixed wiring of the installation in such a way that

ACCESSIBLE METAL PARTS cannot become LIVE in the event of a failure of the BASIC INSULATION.

*2. Type BF EQUIPMENT

TYPE B EQUIPMENT with an F--TYPE APPLIED PART

TYPE B EQUIPMENT: EQUIPMENT providing a particular degree of protection against electric shock, particularly regarding:

-- allowable LEAKAGE CURRENT;

Normal mode

Patient leakage current Less than 100

µ

A

Single failure mode

Less than 500

µ

A

*3. Type CF EQUIPMENT

EQUIPMENT providing a particular degree of protection higher than that for TYPE OF BF EQUIPMENT against electronic shock particularly regarding allowable LEAKAGE CURRENT, and having an F--TYPE

APPLIED PART.

-- allowable LEAKAGE CURRENT;

Normal mode

Patient leakage current Less than 10

µ

A

Single failure mode

Less than 50

µ

A

1--17

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1--3 EMC (Electromagnetic Compatibility)

1--3--1 EMC Performance

All types of electronic equipment may characteristically cause electromagnetic interference with other equipment, either through air or connecting cables. The term EMC (Electromagnetic Compatibility) indicates capability of the equipment, which curbs electromagnetic influence from other equipment and at the same time does not affect other equipment with similar electromagnetic radiation from itself.

This product is designated to fully comply with EN60601--1--2 (IEC 601--1--2), In Medical electronic equipment

EMC regulations.

Proper installation following this service manual is required in order to achieve the full EMC performance of the product.

The product must be installed as stipulated in 1--3--2, Notice upon Installation of Product.

In case of issues related to EMC, please follow procedures stated in 1--3--4, Countermeasures against

EMC--related Issues.

1--3--2 Notice upon Installation of Product

1) Use either power supply cords provided or designated by GEMS or SGMS. Products equipped with power source plug should be plugged into the fixed power socket which has the protective grounding conductor.

Connect a three--pole plug to a three--pole socket without using a three--pole--to--two--pole converter.

2) Locate the equipment as far as possible from other electronic equipment.

3) Be sure to use cables provided by GEMS and SGMS. Wire these cables following these installation procedures.

(Example) Wire power cables separately from signal cables.

4) Lay out the main equipment and other peripherals following the installation procedures described in

Chapter2, INSTALLATION.

1--18

INTRODUCTION

GE MEDICAL SYSTEMS

REV 1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1--3--3 General Notice

1) Designation of Peripheral Equipment Connectable to This Product

Equipment which conforms to EN60601--1--2 (IEC601--1--2), can be hooked up to the product without compromising its EMC performance.

Avoid using other equipment. Failure to comply with this instruction may result in poor EMC performance of the product.

2) Notice against User Modification

Never modify this product. Unilateral user modification may cause degradation in EMC performance.

Modification of the product includes: a) Changes in cables (length, material, wiring etc.) b) Changes in system installation/layout c) Changes in system configuration/components d) Changes in means of fixing system/parts (cover open/close, screwing cover)

3) Operate the system with all covers closed. If you open any cover for some reason, be sure to close it before starting/resuming operation.

Operating the system with any cover open may affect EMC performance.

1--3--4 Countermeasures against EMC--related Issues

Generally it is very difficult to handle issues related to EMC. It may be time consuming and costly.

General countermeasures

Electromagnetic interference with other equipment

1) Electromagnetic interference may be alleviated by positioning other equipment far from the system.

2) Electromagnetic interference may be mitigated by changing the relative location (installation angle) between the system and other equipment.

3) Electromagnetic interference may be eased by changing wiring locations of power/signal cables of other equipment.

4) Electromagnetic influence may be reduced by altering the path of power supply for other equipment.

1--3--5 Notice on Service

1) Ensure all screws are tight after servicing. Loose screws may cause degradation in EMC performance.

2) In case the high frequency gasket of this system is broken, replace it with a new one immediately.

1--19

INTRODUCTION

GE MEDICAL SYSTEMS

REV 5

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1--4 ADDRESS

This system is not repairable by the customer. If this equipment does not work as indicated in the Operator

Manual, please contact your service support center. If the service engineer needs additional information to repair this equipment, please contact the following address (The necessary information will be provided to the Service

Engineer as needed):

GE Medical Systems

Ultrasound Business Group

4855 W. Electric Ave., Milwaukee, WI 53219

USA

TEL: (1) 800--437--1171

FAX: (1) 414--647--4090

CANADA

TEL: (1) 800--668--0732

LATIN & SOUTH AMERICA

TEL: (1) 305--735--2304

GE Ultrasound Europe

GE Ultraschall GmbH & Co.KG

Beethovenstra β e 239 Postfach 11 05 60

D--42655 Solingen GERMANY

TEL: 0130--81--6370 (OLC--Europe Toll Free Number)

(49) 212--2802--207 (English/German Hotline)

(49) 212--2802--208 (English/German/French Hotline)

FAX: (49) 212--2802--431

GE Yokogawa Medical Systems

On--Line Center (OLC), Asia

Ultrasound Group

67--4, Takakura--cho, Hachioji--shi, Tokyo, 192--0033

JAPAN

TEL: (81) 426--48--2940

FAX: (81) 426--48--2950

GE MEDICAL SYSTEMS KOREA

65--1, Sangdaewon--dong, Chungwon--Ku,

Sungnam--Si, Kyunggi--do,

KOREA

TEL: (82) 31--740--6000

FAX: (82) 31--740--6436

1--20

INTRODUCTION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CHAPTER 2 -- INSTALLATION

TABLE OF CONTENTS

TITLE SECTION

2--1

2--2

PAGE

PREINSTALLATION

2--1--1 Introduction

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--1--2 Power Line Requirements

2--1--3 Physical Specifications

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--1--4 Recommended Ultrasound Room Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--3

2--3

2--3

2--4

2--5

INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--2 Average Installation Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--3 Installation Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--4 Checking the Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--5 Unpacking LOGIQ 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--6 MTZ Probe Holder Installation (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--7 Transducer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--8 Moving into Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--9 Adjusting System Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2--10 Product Locator Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--8

2--8

2--8

2--8

2--8

2--9

2--11

2--14

2--14

2--14

2--15

2--1

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

2--2

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2--1 PREINSTALLATION

2--1--1 Introduction

This section describes various general electrical, operational, and environmental considerations that must be considered before installing the LOGIQ 200 PRO Series Ultrasound unit.

2--1--2 Power Line Requirements

The following power line parameters should be monitored for one week before installation. We recommend that you use an analyzer Dranetz Model 606--3 or Dranetz Model 626:

PARAMETER

Voltage Range

Voltage Range

Line Frequency

Power Transients

:

: Korea

: USA

: Europe, LA

: Japan.

: Asia

LIMITS

: 220 VAC

±

10% (198 -- 242 VAC)

: 115 VAC

±

10% (103 -- 127 VAC)

: 220 -- 240 VAC ¦ 10%(198 --264 VAC)

: 100 VAC

±

10% (90 -- 110 VAC)

: 115 VAC

±

10% (103 -- 127 VAC)

: 220 -- 240 VAC

¦

10%(198 --264 VAC)

: All applications : MAX. 500 VA

: All applications : 50/60Hz (

±

2Hz)

: Less than 25 % of nominal peak voltage for less than 1 millisecond for any type of transient, including line frequency, synchronous, asynchronous, or aperiodic transients.

Decaying Oscillation : Less than 15 % of peak voltage for less than 1 millisecond.

2--3

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2--1--3 Physical Specifications

The LOGIQ 200 PRO Series (excluding accessories) weighs 76 Kg (168 lbs). See Chapter 3, ILLUSTRATION

3--1 for dimensions.

Operating Conditions

The LOGIQ 200 PRO Series is designed to operate within a temperature range of 10

_

C to 40

_

C (50

_

F to 104

_ F), and a relative humidity range of 5 % to 90 % (Non--condensing).

Patient Comfort

Concerning permissible operating temperature and humidity tolerances, we recommend that ambient room temperature should be maintained between 20 _ C to 26 _ C (68 _ F to 79 _ F), Humidity should be maintained between 50 % and 70 % for patient comfort during ultrasound scanning.

Electromagnetic Interference (EMI)

Ultrasound machines are susceptible to interference from the radio frequencies, magnetic fields, and transients in the air or power leads. Possible EMI sources should be identified. Electrical and electronic equipment may produce EMI unintentionally or as a result of a malfunction. These sources include medical lasers, cauterizing guns, computers, monitors, fans, gel warmers, microwave ovens, and cellular phones. The presence of broadcast station or van may also cause interference.

Carefully read the following precautions before installing the unit.

1.

Connect the power plug for any other equipment into the fixed outlet with ground wire.

2.

Securely connect any equipment with permanent ground connection to the earth ground furnished in the building.

3.

Install the unit as far from any electrical or electronic equipment as possible.

If any EMI troubles are known or suspected to be present, try to deal with the equipment suspected to have influence on the Ultrasound machine as follows:

1.

Move the ultrasound machine as far from the equipment as possible.

2.

Change the arrangement of the equipment in the room.

3.

Plug the equipment into other outlet.

4.

Move the power cable or signal cable connected with the equipment.

Securely re--tighten drive any screws for the Ultrasound machine after re--assembling for service operation.

2--4

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

2--1--4 Recommended Ultrasound Room Layout

TABLE 2-- 1 provides the requirements for an ultrasound room:

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

POWER SOURCE

CURRENT RATING

RADIATION SHIELDING

TEMPERATURE

HUMIDITY

HEAT DISSIPATION

FLOOR LOADING

FLOOR CONDITION

LOGIQ 200 PRO Series Weight

TABLE 2-- 1

ULTRASOUND ROOM REQUIREMENTS

230VAC, 50Hz, SINGLE PHASE

115V, 60Hz, SINGLE PHASE

For Europe Version

For USA Version

10A (115V, 100V) ; 5A (220--240V) CIRCUIT BREAKER

NONE REQUIRED for ULTRASOUND ENERGY

20--26

_

C (68--79

_

F) for PATIENT COMFORT

50% to 70% for PATIENT COMFORT

2000 BTU/Hr for LOGIQ 200 PRO SeriesMD

Approximately 240 -- 300 kg/m

2 without Accessories

Gradient : WITHIN 5 degrees

76 kg (168lbs) without Accessories

2--5

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

2--1--4 Recommended Ultrasound Room Layout (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Optional Desirable Ultrasound Room Facilities

These facilities are:

1.

Lab sink with hot and cold water.

2.

Emergency oxygen supply.

3.

Dimmer control for overhead lights.

4.

Film viewer.

5.

Film and linen storage.

6.

Medical equipment storage.

7.

Hospital grade equipment electrical outlet.

8.

Document storage area for operating and service manuals.

9.

Nearby waiting room, dressing room, lavatory.

10. Trash bin.

Recommended and Alternate Ultrasound Console Floor Plans

ILLUSTRATION 2--1 provides a recommended standard floor plan and a minimal floor plan for ultrasound equipment

2--6

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

2--1--4 Recommended Ultrasound Room Layout (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Dedicated power outlet

GE cabinet for software and manuals

Film

Processing

Room

Film Viewer

Counter

Top

LOGIQ

200

Stool

Suction

Line

Emergency

Oxygen

61 cm

Desk

Patient

Toilet

Facility

Scale: each square equals one square foot

Film Viewer

Film Supplies

Sink

107 cm

42 in

A 14 by 17 Recommended Floor Plan

Linens

Probes/supplies

Stool

LOGIQ

200

Examination Table

193 cm

76 cm

30 in

An 8 by 10 Minimal Floor Plan

Dedicated power outlet

GE cabinet for software and manuals

RECOMMENDED ULTRASOUND FLOOR PLAN

ILLUSTRATION 2--1

2--7

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2--2 INSTALLATION

2--2--1 Introduction

This section contains many of the procedures required to install the LOGIQ 200 PRO Series console.

2--2--2 Average Installation Time

The LOGIQ 200 PRO Series has been designed to be installed and checked out by an experienced service technician in approximately three hours. LOGIQ 200 PRO Series consoles with optional equipment may take slightly longer.

2--2--3 Installation Warnings

1.

Since the LOGIQ 200 PRO Series weighs approximately 76 kg (168 lbs) without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items.

2.

There are no operator serviceable components. To prevent shock, do not remove any covers or panels.

Should problems or malfunctions occur, unplug the power cord. Only qualified service personnel should carry out servicing and troubleshooting.

Note

For information regarding packing labels, refer to ILLUSTRATION 2--3, LABELS ON PACKAGE.

3.

After being transported, the unit may be very cold or hot. If this is the case, allow the unit to acclimate before you turn it on. It requires one hour for each 2.5

°

C increment if it’s temperature is below 10

°

C or above 40

°

C.

CAUTION

Equipment damage possibility. Turning the system on without acclimation after arriving at site may cause the system to be damaged.

TABLE 2-- 2

TIME FOR SETTLEMENT

°

C 60 55 50 45 40 35 30 25 20 15 10 5 0 --5 --10 --15 --20 --25 --30 --35 --40

°

F 140 131 122 113 104 95 86 77 68 59 50 41 32 23 14 5 --4 --13 --22 --31 --40 hrs 8 6 4 2 0 0 0 0 0 0 0 2 4 6 8 10 12 14 16 18 20

2--2--4 Checking the Components

When a new system arrives, check that any components are not damaged and or missing. If shipping damage or shortages occur, contact the address shown in Chapter 1.

2--8

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

2--2--5 Unpacking LOGIQ 200 PRO Series

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Do not lift the unit by the Keyboard or Monitor arm. Equipment damage may result.

CAUTION

The unit weighs approximately 76 kg (168 lbs). Be prepared for a sudden shift of weight as the unit is removed from its base (pallet).

Refer to ILLUSTRATION 2--2 while performing the following procedure.

1.

Cut the two BANDs.

2.

Lift the CAP up and off.

3.

Lift the TOP PAD up and off.

4.

Remove the OPTION BOX.

5.

Remove the VINYL COVER.

6.

Remove the MIDDLE PLATE ASSY.

7.

Remove the MONITOR COVER.

8.

Lift the PACKING CASE up and off.

9.

Lift the Monitor up by pressing the <UP/DOWN> button located on the Monitor Arm.

10. Remove the MONITOR PAD.

11. Return the Monitor arm to its lowest position.

12. Carefully roll the LOGIQ 200 PRO Series from the PALLET.

13. Remove the Caution Label attached on the CRT Filter and clean the CRT Filter.

Note

Check the shipping container for special instructions. Verify that the container is intact. In some cases a secondary container may be used. If so, ask the carrier for unpacking instructions.

2--9

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

2--2--5 Unpacking LOGIQ 200 PRO Series (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

MIDDLE PLATE

CAP

TOP PAD

PACKING

PLATE

OPTION BOX

PALLET

BAND

UNPACKING LOGIQ 200 PRO Series

ILLUSTRATION 2--2

MONITOR PAD

2--10

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

2--2--5 Unpacking LOGIQ 200 PRO Series (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

60

d

C

Air Pressure

700 - -1060hPa

Humidity 30- -95%

EXcluding Condensation

- -10

d

C

LABELS ON PACKAGE

ILLUSTRATION 2--3

2--2--6 MTZ Probe Holder Installation (Option)

One MTZ probe holder and two brackets are supplied with the LOGIQ 200 PRO Series console as shown in

ILLUSTRATION 2--4.

Note

Two sets of screw holes are provided at the bottom of standard probe holder and left side of keyboard to install the MTZ probe holder. You can choose the most convenient position for your customer between the two sets of screw holes.

Bracket1 (RIGHT)

MTZ Probe Holder

Bracket 2 (LEFT)

MTZ PROBE HOLDER SET

ILLUSTRATION 2--4

2--11

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2--2--6 MTZ Probe Holder Installation (Option) (Continued)

1.

Assemble the MTZ probe holder at the bottom of standard probe holder by screwing four (1 -- 4) screws as shown in ILLUSTRATION 2--5.

MTZ PROBE HOLDER INSTALLATION (1)

ILLUSTRATION 2--5

1

2

3

4

2--12

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2--2--6 MTZ Probe Holder Installation (Option) (Continued)

2.

After removing the Keyboard Assy (Refer to NO TAG, 6--2 DISASSEMBLY/RE--ASSEMBLY), Assemble the

MTZ probe holder at the left side of keyboard by screwing four (1 -- 4) screws as shown in

ILLUSTRATION 2--6.

1

2

3

4

MTZ PROBE HOLDER INSTALLATION (2)

ILLUSTRATION 2--6

2--13

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2--2--7 Transducer Connection

1.

Connect a transducer to the left side connector slot #1, transducer receptacle as follows: a. Ensure that the transducer twist lock lever points towards the 12 o’clock position.

b. Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating surface.

c. Twist the transducer twist lock lever to the 4 o’clock position to lock it in place. Twist the lever to the 12 o’clock position to disconnect the transducer.

Note

It is not necessary to turn OFF power to connect or disconnect a transducer.

2.

Connect the main power cable to a hospital grade power receptacle with the proper rated voltage checked during preinstallation. Never use a three--to--two prong adapter; this defeats the safety ground.

2--2--8 Moving into Position

CAUTION

Do not lift the unit by the Keyboard.

Do not tilt the unit more than 5 degrees to avoid tipping it over.

To avoid injury by tipping over. Set the monitor to the lowest position before moving.

CAUTION

Equipment Damage Possibility. Lifting the console by holding covers may damage the covers. Do not lift the console by holding any covers.

In general, a single adult can move the LOGIQ 200 PRO Series. (It is better to pull from the rear rather than push from the front of the unit). Before moving, store all loose parts in the unit. Wrap transducers in soft cloth or foam to prevent damage.

2--2--9 Adjusting System Clock

Set system clock for the LOGIQ 200 PRO Seriesto the local time. For procedure of adjusting the System clock, refer to “Customizing Your System” in the Users manual.

2--14

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2--2--10 Product Locator Installation Card

Fill out proper customer Information the Product Locator Installation Card. Refer to ILLUSTRATION 2--7. Mail this Installation Card “Product Locator” to the address corresponding to your pole.

Note

The Product Locator Installation Card shown in ILLUSTRATION 2--7 may not be same as the provided Product Locator card.

PRODUCT LOCATOR INSTALLATION CARD

ILLUSTRATION 2--7

2--15

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

2--16

INSTALLATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CHAPTER 3 -- SYSTEM CONFIGURATION

TABLE OF CONTENTS

TITLE SECTION

3--1

3--2

3--3

3--4

3--5

3--6

PAGE

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--3

DIMENSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--3

ELECTRICAL SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--3--1 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--3--2 Facility Power Receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--4

3--4

3--4

STORAGE AND OPERATION REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--4

OPTIONAL PERIPHERALS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--5--1 Peripherals/Accessories Connector Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--5--2 List of Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--5--3 Power Consumption of Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--5

3--5

3--9

3--11

TEST POINT, LED, DIP SWITCH AND RESET SWITCH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--6--1 Test Point List

3--6--2 LED List

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--6--3 DIP Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--6--4 Reset Switch

3--6--5 Jumper

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--12

3--12

3--14

3--15

3--15

3--15

3--1

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

3--2

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

3--1 INTRODUCTION

This chapter describes system configuration and specifications.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--2 DIMENSIONS

Regarding LOGIQ 200 PRO Series dimensions, Refer to ILLUSTRATION 3--1 for planning the position of your

LOGIQ 200 PRO Series.

WEIGHT: 76 kg

168 lbs

NOTE

LENGTH : mm (inches)

ABERRATION :

±

5%

400 (15.7)

350 (13.8)

663 (26.1)

250 (9.8)

335 (13.2)

245 (9.6)

192 (7.6)

1200 (47.2)

800 (31.5)

162 (6.4)

OVERALL DIMENSIONS

ILLUSTRATION 3--1

3--3

PRO Series

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--3 ELECTRICAL SPECIFICATIONS

Electrical conduit, junction boxes, outlets, circuit breakers, and switches should be in place before installing the

LOGIQ 200 PRO Series console.

3--3--1 Power Supply

Voltage setup is performed in the factory. Different power cables and circuit breakers are used for the 100 (115)

Vac and 220 (240) Vac versions.

3--3--2 Facility Power Receptacle

A separate power outlet with a 10 amp circuit breaker for 100 (115) Vac units, or a 5 amp circuit breaker for 220

(240) Vac units, is recommended. The specific power receptacle used depends on your country’s power line standards.

The receptacle should have International Electrotechnical Commission (IEC) approval, or equivalent.

3--4 STORAGE AND OPERATION REQUIREMENTS

Probes and peripherals are shipped in separate containers. Shipping weight is approximately 210 lbs (96 kg).

The size of the container is L82 cm x W53 cm x H140 cm (32 in. x 21 in. x 55 in). Table 3--1 provides a summary of temperature, atmospheric pressure, and humidity tolerances for shipping, installation, and operation.

TABLE 3-- 1

STORAGE AND OPERATION REQUIREMENTS

PARAMETER STORAGE OPERATION

TEMPERATURE (

_

C)

( _ F)

--10 to 60

14 to 140

10 to 40

50 to 104

ATMOSPHERIC PRESSURE

(hPa)

HUMIDITY (%)

(Non--condensing)

700 to 1060

5 to 90

700 to 1060

5 to 90

3--4

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

3--5 OPTIONAL PERIPHERALS

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--5--1 Peripherals/Accessories Connector Panel

LOGIQ 200 PRO Series peripherals and accessories can be properly connected using the rear connector panel.

Located on the panel are video input and output connectors, camera expose connector, foot switch connector, power connector and service ports.

This section indicates the pin assignment for each connector (1 -- 4 in ILLUSTRATION 3--2) at pages 3--6 through 3--7.

4

5

1

6

3

2

115V 1.7A Max including front panel

100V 2.0A Max including front panel

220--240V 0.8A Max including front panel

REAR CONNECTOR PANEL

ILLUSTRATION 3--2

3--5

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--5--1 Peripherals/Accessories Connector Panel (Continued)

Note

Each outer (case) ground line of peripheral/accessory connectors are protectively grounded.

Signal ground lines are Not Isolated.

8

9

10

11

12

13

Pin No.

1

2

3

6

7

4

5

1. Pin Assignment of RS232C for Service

Connector : Female, D--SUB, 25--pin

Signal

Frame GND

TXD

RXD

RTS

CTS

DSR

Signal GND

DCD

+ 5 V

*1

(300mA Max)

Pin No.

14

15

16

17

18

19

20

21

22

23

24

25

Signal

DTR

Ring Indicator

*1

: This voltage shall be apply to this Pin in case of MODEM usage.

Note

Output level of RS232C signals :

High

Low

+3V to +15V

--15V to 0V

14

25

1

13

3--6

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--5--1 Peripherals/Accessories Connector Panel (Continued)

2. Pin Assignment of Foot Switch

Connector : Round 5--pin connector

Pin No.

1

2

3

Output Signal

FOOT SW

FOOT SW_G

Pin No.

4

5

Output Signal

Frame GND

4

5

1

2

3

3. Pin Assignment of Mini Jack for Controlling B/W Printer

Connector :Stereo Mini Jack

Pin No.

1

Output Signal

PRINT

*1

Pin No.

2

Output Signal

Signal GND

*1

: Printer starts printing by receiving the LOW pulse more than 100ms.

Note

Output level of control signals indicated in the above tables are TTL level.

4. Pin Assignment of ECG

Connector :

Pin No.

1

2

3

Output Signal

RA

LF

RF

Pin No.

4

5

Output Signal

Signal GND

Signal GND

1

2

4

5

3

3--7

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

3--5--1 Peripherals/Accessories Connector Panel (Continued)

5. Pin Assignment of Parallel Printer for Printer--A

Connector :

8

9

10

11

12

13

Pin No.

1

2

3

6

7

4

5

Signal

STB (L)

PD0

PD1

PD2

PD3

PD4

PD5

PD6

PD7

ACK (L)

BUSY

PE

SLCT

Pin No.

14

15

16

17

18

19

20

21

22

23

24

25

Signal

AFD(L)

ERR(L)

INIT(L)

SLIN(L)

GND

GND

GND

GND

GND

GND

GND

GND

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

14

25

1

13

6. Pin Assignment of for Printer--B

Connector :

Pin No.

1

2

3

4

5

Signal

DCD

RXD

TXD

DTR

Signal GND

Pin No.

6

7

8

9

Signal

DSR

RTS

CTS

Ring Indicator

6

9

1

5

3--8

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--5--2 List of Optional Peripherals

The tables below shows the suggested optional peripherals for LOGIQ 200 PRO Series.

1. RECORDING DEVICES

TABLE 3-- 2

LIST OF RECORDING DEVICES

DEVICE

Video Cassette Recorder

Video Graphic Printer

Multi Image Camera

Magneto Optic Disk Driver

MANUFACTURER

SONY

SONY

SONY

MITSUBISHI

International Imaging

Electronics

FUJITSU

MODEL

SVO--9500MD

SVO--9500MDP

UP890

P90

IIE460

CATALOG No.

VIDEO SIGNAL

Local

Local

Local

Local

Local

MDD3064AP Local

Note

See each option installation instructions for installation and connection procedures.

NTSC

PAL

3--9

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 6

3--5--2 List of Optional Peripherals (Continued)

2. TRANSDUCER (PROBE)

LOGIQ 200 PRO Series SERVICE MANUAL

TABLE 3-- 3

LIST OF TRANSDUCERS

PROBE

NAME

CBF

CAE

MTZ

CZB

CS

ATR

LH

LE

LI

LT

LB

LD

ERB

10Lb

SY

3C

S317

MATERIAL OF

HEADSHELL

PES

PES

PES

NORYL

PES

PES

PES

PES

PES

PES

PES

NORYL

PES

AREA OF USING

Abdom.

Abdom.

Intercavity

Neonatal

Abdom.

Urology

Superficial

OB/Gyn.

Intraoperative

Intraoperative

OB/Gyn.

Intraoperative

Urology

PES

PES

PES

NORYL

Superficial

Abdom.

OB/Gyn

Cardiology

TYPE CATALOG

NO.

REQUIRED

ADAPTER

Convex H46022CB Not Required

Convex H46022CA Not Required

Convex H46022MT Not Required

Convex H45202CZ Not Required

Convex H45222CS Not Required

Convex H4061PR Not Required

Linear H46022LH Not Required

Linear H46022LE Not Required

Linear H46022LI Not Required

Linear H46022LT Not Required

Linear H46022LB Not Required

Linear H45202LD Not Required

Linear,

Convex

H45202ER Not Required

Linear H45202LM Not Required

Sector H46022SY Not Required

Linear H45202WB Not Required

Sector H45202SD Not Required

PART NO.

FOR JAPAN

P9603AA

P9603AB

P9603AL

2152402

2202315

2201223

P9601AC

P9601AB

P9601AG

P9601AJ

P9601AA

P9601AD

--

--

--

--

2144268--2

2235374

PART NO.

P9603AD

P9603AE

P9603AU

2152422

2202320

2201222

P9601AS

P9601AR

P9601AW

P9601AX

P9601AQ

2124317

--

--

--

--

--

CAUTION

Equipment damage possibility. Be sure to use the following recommended connecting cables to connect recording devices with LOGIQ 200 PRO Series console.

3. CONNECTING CABLES

NAME

Printer Install Kit

MOD Install Kit

PART NO.

2176459

2235371

TABLE 3-- 4

LIST OF CONNECTING CABLES

FIGURE

Printer Cable Assy, BNC cable,

Shutter cable

MOD Driver, Top Bracket,

Bottom Bracket, Screw

NOTE

For B/W Printer

For MOD Driver

3--10

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--5--3 Power Consumption of Optional Peripherals

The table below shows the power consumption of each optional peripheral for LOGIQ 200 PRO Series.

TABLE 3-- 5

POWER CONSUMPTION OF OPTIONAL RECORDING DEVICES

DEVICE

Video Cassette Recorder

Video Graphic Printer

Multi Image Camera

MOD Driver

MODEL

SVO--9500MD

SVO--9500MDP

UP890

IIE460

MDD3064AP

POWER CONSUMPTION

64 W

110 W

60 W

5.3 W

3--11

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--6 TEST POINT, LED, DIP SWITCH AND RESET SWITCH

3--6--1 Test Point List

The table below shows The Test Point list for LOGIQ 200 PRO Series.

LOCATION

MVP ASSY

MSTE ASSY

TBF ASSY

NAME

TP1

TP2

TP3

TP4

TP5

TP6

TP7

TP8

GND

VCC1

--5V

TP1

TP2

TP3

DGND2

AGND1

AGND2

TABLE 3-- 6

TEST POINT

DESCRIPTION

Non--interlaced HSYNC

Non--interlaced VSYNC

M--mode data clock

M_EN

SLN7

B mapping clock

B DTC

CPU clock

Ground

+5V for analog/digital

--5V for analog

+5V for digital

GND

Foot switch status

Ground for Digital

Ground for Analog

Ground for Analog

POSITION

Edge of board

Edge of board

--

--

--

--

--

--

Edge of board

Edge of board

--

Edge of board

Edge of board

--

--

--

--

3--12

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--6--1 Test Point List (continued)

The table below shows The Test Point list for LOGIQ 200 PRO Series.

LOCATION

BPHV ASSY

RBF ASSY

SCAN ASSY

MOTHER 3 ASSY

TABLE 3-- 7

TEST POINT

NAME

TP

TP

¦

¦

THV

SHV

TP HV_REF

V5AP

V5AN

V5BG

V5AD

VCC1

AGND1

AGND2

GND1

GND2

GND3

GND4

GND5

+5V

¦ SHV

MON_12P

ECG_12P

ECG_12M

GND1

GND2

GND3

TRIG

DESCRIPTION

High Voltage for transmitting

High Voltage for High Voltage switch on CONN Assy

D/A output for HV control

+5V for Analog

--5V for analog

+5V for Gain Control Block

+5V for ADC Block

+5V for Digital

Ground for Analog

Ground for Analog

Ground for Digital

Ground for Digital

Ground for Digital

Ground for Digital

Ground for Digital

+5V for digital

¦

SHV for high voltage switch

+12V for Monitor

+12V for ECG Module

--12V for ECG Module

Ground

Ground

Ground

TRIG Signal

POSITION

--

--

--

Edge of board

Edge of board

--

--

--

--

--

--

--

--

--

--

Edge of board

Edge of board

--

--

--

--

--

--

--

3--13

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

3--6--2 LED List

The table below shows the LED list for LOGIQ 200 PRO Series.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

TABLE 3--7

LED

LOCATION

MSTE ASSY

MVP ASSY

NAME

DS9

DS10

DS1~DS8

DS11

DS1

DS2

DS3

DS4

DS5

DS1, DS2

DESCRIPTION

+5V for digital

Watchdog timer out

(CPU FAULT)

Reserved

MOD Active

+5V for digital

Scan Frame Clock

POSITION

Edge of board

NORMAL

ON

OFF

--

--

ON

BLINKING

BPHV ASSY

SMPS ASSY GREEN

LED

DSP running

BPHV ASSY status

(ON:normal)

SMPS ASSY status

(ON:normal)

SMPS ASSY

ON

ON

ON

Note

Three LED (DS2~DS4) of MVP assy should be blinked when the system operated with the

Probe.

ABNORMAL

OFF

ON

--

--

OFF

FIXED

OFF

OFF

OFF

3--14

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 4

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--6--3 DIP Switch

The table below shows the DIP Switch list for LOGIQ 200 PRO Series.

LOCATION NAME Switch No.

TABLE 3--8

DIP SWITCH Setting

DESCRIPTION

ON

STATUS

OFF

-enable

-enable

-disable

-disable

MSTE ASSY S1

S2

1 ~6

7

4

5

2

3

8

1

6

7

8

Assy Version Read, I/O check

System Cold Reset

Reserved

Initialize Setup data to factory default

Reserved

Reserved

A--Mode

Emergency mode for download by MOD or PL

Reserved

Flash memory/SRAM Test & Clear

NTSC/PAL Selector

Note

To activate the dip switch S2--7 fuction, S2--1 should be set to ON.

Note

To activate the A--Mode fuction, S2--4 should be set to ON.

LOCATION NAME Switch No.

MVP ASSY S1 1 ~8

TBF ASSY

RBF ASSY

S1

S1

1 ~8

1 ~7

DESCRIPTION

Assy Version Read, I/O Check

Assy Version Read, I/O check

Assy Version Read, I/O check

PAL NTSC

POSITION

--

Edge of board

DIN connector side

3--6--4 Reset Switch

Reset switch (S3) on edge of MSTE Assy used for resetting the system.

3--15

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3--6--5 Video Signal Specification

The table below shows the video signal specification for LOGIQ 200 PRO Series.

TABLE 3--9

VIDEO SIGNAL SPECIFICATIONS

NTSC

VCR

OUT

Monitor

OUT

VCR

OUT

Total Line per Frame

Vertical Field Frequency

[Line]

[Hz]

525

60

525

60

625

50

Horizontal Scanning Frequency [KHz]

15.733

31.466

15.625

Total Horizontal Line Time

Horizontal Display

Horizontal Porch Width

Horizontal Pulse Width

Horizontal Porch Width

Total Horizontal Blanking

Vertical Front Porch Width

[

µ

s]

[

µ

s]

[

µ

s]

[

µ

s]

[

µ

s]

[

µ

s]

[H]

63.56

49.54

2.76

4.73

6.53

14.02

6.5

31.78

27.3

1.18

2.81

0.49

4.48

0

64.00

48.81

3.09

4.68

7.42

15.19

9.0

Vertical Sync Width

Vertical Back Porch Width

Total Blanking Interval

Pixel Clock

H

Display Area Size

[H]

[H]

[H]

3

22

31.50

[MHZ]

12.27

[

µ

s]

63.56

[mm]

--

4

0

0

24.54

31.87

207X157

2.5

27

38.50

14.75

64

--

PAL

Monitor

OUT

0.3

4.48

0

4

0

0

29.5

32

625

50

31.25

32.00

27.52

1.42

2.76

207X157

3--16

SYSTEM CONFIGURATION

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CHAPTER 4 -- FUNCTIONAL CHECKS

TABLE OF CONTENTS

TITLE SECTION

4--1

4--2

4--3

PAGE

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--1--1 Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--3

4--3

FUNCTIONAL CHECK PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--2--1 Basic Controls

4--2--2 M--Mode Check

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--3

4--3

4--5

SMPS ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--3--1 SMPS Assy Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--3--2 SMPS Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4--7

4--7

4--9

4--1

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

4--2

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

4--1 INTRODUCTION

This chapter provides procedures for quickly checking major functions of the LOGIQ 200 PRO Series console, and SMPS adjustments.

4--1--1 Required Equipment

To perform these tests, you’ll need a linear, or a convex transducer.

4--2 FUNCTIONAL CHECK PROCEDURES

4--2--1 Basic Controls

Step Check

1

2

Connect the convex transducer to ”Probe 1” connector.

Power On

3

4

5

6

7

8

Expected Result

Rotate B/M Gain knob

After few seconds, the B mode screen should appears as shown in ILLUSTRATION 4--1.

Image gets lighter with CW rotation and darker with

CCW.

The gray scale adjusts to each new Map selected.

Press Map key to select another gray scale Map.

Press Dyn Range Arrow up or down key.

At lower Dynamic Range settings, image speckle fades and prominent objects in the display are more pronounced from the background image.

Rotate Depth knob.

The depth of image should be increased /decreased.

Slide TGC potentiometers (pots) Image grows darker or brighter at depth equivalent to pot’s location.

Press Zoom. key.

The image should increase to X2 size.

9

Press it again to exit.

Press Frame Avg key.

10

11

12

Press Edge Enhc key.

Press Reverse key.

Press Reverse key again.

Image speckle fades and probe or wire movement is smeared.

The edges inside the focal area(s) should become lighter when you increase and darker as you decrease its value.

The image reverses the left/right orientation.

The image reverses again.

4--3

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

4--2--1 Basic Controls (Continued)

Map1

B--Mode

Cine Gauge

B--MODE DISPLAY SCREEN

ILLUSTRATION 4--1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

GE

4--4

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

4--2--2 M--Mode Check

Step Check

13

Press M key.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Roll trackball, position cursor over area you want to see in motion.

Press M key again.

14

Rotate B/M Gain knob

15

Press Dyn Range Arrow up or down key.

16

Press Sweep Speed key

Press it again to exit.

17

Press Freeze key.

Press it again to exit.

18

Press Edge Enhc key.

19

Press B key.

Expected Result

The M mode timeline should appear next to the B image as shown in ILLUSTRATION 4--2. Whether it takes half the screen or two--thirds depends on the preset.

The Mode cursor should follow trackball movement and the timeline should update for new location of focus.

The full M--mode should appear on the CRT monitor.

The M timeline should get brighter with CW rotation and darker with CCW.

Dynamic Range affects grays and the last added scan mode; to adjust the basic B, M must be off. Turn

Dynamic Range down to increase contrast, turn up to soften.

The timeline speed should increase to 4 second sweeps and decrease to 16 second sweeps.

Fast=4 Medium=8 Slow=16

The image should freeze.

The image revives acquisition.

Changes the M image

The M Mode timeline should disappear and the

B--mode image should appear as shown in

ILLUSTRATION 4--1.

4--5

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

4--2--2 M--Mode Check (Continued)

Map1

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

GE

M--Mode

MuFq

Cine Gauge

M--MODE DISPLAY SCREEN

ILLUSTRATION 4--2

Note

You can select several types of display formats by using the Setup Menu. For the Preset Menu, refer to Customizing Your System in the LOGIQ 200 PRO Series User Manual.

4--6

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

4--3 SMPS ADJUSTMENTS

This section provides SMPS adjustment procedures for the LOGIQ 200 PRO Series. Adjustments should be only made when necessary. SMPS adjustments should be made in accordance with the schedule for periodic maintenance in Chapter 7 of this manual.

Before beginning the SMPS adjustments procedure, make sure the power outlet conforms to the proper power line standards. Refer to Chapter 2, Installation.

Note

If the adjustment pot is turned to far clockwise, the SMPS output shuts down to protect the circuits against the over--voltage. In that case, power the LOGIQ 200 PRO Series OFF and turn the pot counterclockwise all the way. Wait for 5 minutes for recovering of SMPS protection circuit. Then power it ON and try to adjust the SMPS again.

The SMPS Assy is in the bottom of the LOGIQ 200 PRO Series as shown in ILLUSTRATION 4--7.

4--3--1 SMPS Assy Access

1.

Remove the four screw caps and unscrew the screws to remove the Left Cover as shown in the

ILLUSTRATION 4--3.

419 EMI Cover

404 Left Cover

1

2

3

4

LEFT COVER REMOVAL

ILLUSTRATION 4--3

4--7

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

4--3--1 SMPS Assy Access (Continued)

2.

Remove the EMI Cover L as shown in ILLUSTRATION 4--4.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

419 EMI Cover L

3

1

2

5

4

7

6

8

EMI COVER REMOVAL

ILLUSTRATION 4--4

3.

Remove the four screw caps and unscrew the screws to remove the Right Cover as shown in the

ILLUSTRATION 4--5.

RIGHT COVER REMOVAL

ILLUSTRATION 4--5

4--8

405 Right Cover

1

2

3

4

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

4--3--1 SMPS Assy Access (Continued)

4.

Remove the four screw caps and unscrew the screws to remove the Rear Cover as shown in the

ILLUSTRATION 4--6.

408 Rear Cover

1

2

3

4

REAR COVER REMOVAL

ILLUSTRATION 4--6

4--9

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

4--3--2 SMPS Adjustment Procedure

1.

Power LOGIQ 200 PRO Series ON. Wait for about 30 seconds to warm up the console.

2.

Connect a DVM to the appropriate place shown in Table 4--1.

3.

Verify that the voltages are as shown in Table 4--2.

SMPS

±

12V for ECG

+12 for Monitor

+5V for Digital

TABLE 4-- 1

SMPS MEASUREMENT LOCATION

MEASURE AT

ECG_12P, ECG_12M on

MOTHER3 BD ASSY

RETURN AT

AGND3 (Ground) on

MOTHER3 BD ASSY

MON_12P on MOTHER3 BD

ASSY

AGND3 on MOTHER3 BD

ASSY

VCC1 (+5V) on MVP BD ASSY GND1 (GND) on MVP BD

ASSY

ADJUST AT

1 on SMPS ASSY

See ILLUSTRATION 4--7

3 on SMPS ASSY

See ILLUSTRATION 4--7

2 on SMPS ASSY

See ILLUSTRATION 4--7

SMPS

--5V,

±

12V for Analog

+12 for Monitor

+5V for Digital and Analog

TABLE 4--2

SMPS MEASUREMENT TOLERANCES

MIN

--4.8 V

--12.6V

+12.6 V

+11.8 V

+4.8 V

MAX

--5.2 V

--11.4 V

+11.4 V

+12.2 V

+5.2 V

4--10

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

4--3--2 SMPS Adjustment Procedure (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1

2

3

1. Adjustment Point for --5V,

±

12V SMPS

2. Adjustment Point for +5V SMPS

3. Adjustment Point for +12V SMPS

SMPS ASSY

ILLUSTRATION 4--7

4--11

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

4--12

FUNCTIONAL CHECKS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CHAPTER 5 -- DIAGRAMS

TABLE OF CONTENTS

TITLE SECTION

5--1

5--2

5--3

5--4

5--5

PAGE

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--3

LOGIQ 200 SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--3

BLOCK DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--4

WIRING DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--5

CIRCUIT BOARD DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5--7

5--1 DIAGRAMS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

5--2 DIAGRAMS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

5--1 INTRODUCTION

LOGIQ 200 PRO Series is a compact ultrasound scanner supporting a wide range of probes. This gives the system added benefits and flexibility to meet diverse applications.

5--2 LOGIQ 200 PRO Series SYSTEM

The LOGIQ 200 PRO Series has a 48 channel beamformer that digitizes the RF signal. It uses digital delay to focus the acoustic beam. It offers parallel receive beam formation which can increase frame rate by a factor of two in time critical applications. This system also features many advanced image processing controls.

User surveys and the latest technology were used to increase console ease--of--use.

System Features

The key design goals of this system are:

High Image Quality

Increased User Productivity

Multiple Clinical Applications

Planned Upgradeability

High Mobility

Types of Applications

The system supports many clinical uses. Scan and display parameters may be user selected to default to desired values for each application. The system presets many parameters to clinically determined, optimal values.

RAD/ABDOMEN

OB/GYN

UROLOGY

SMALL PARTS

UROLOGY

CARDIOLOGY

See Illustration 5--1, the LOGIQ 200 PRO Series system can be divided into an beamforming section (from TBF to

RBF assy) and an image reconstruction processing section (from MVP to MSTE). The beamforming section is to focus a RF signal in transmiting and receiving. Image reconstrcution section converts the receiving scan data to some kinds of images that are B mode convex image, B--mode linear image, M--mode image. Another section has the microprocessor driven system control section, which controls the system based on operator commands and system status information.

5--3 DIAGRAMS

GE MEDICAL SYSTEMS

REV 0

5--3 BLOCK DIAGRAM

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Power

Supply

To Mother board

To Mother board

PIF

External Video In

External Video Out

ECG

Shutter Control

Foot Switch

Service

Printer B (RS232C)

Printer A (Parallel)

MSTE

MOD

Keyboard

SHV

BPHV

THV

TBF

RF Data

RBF

BM Data

MVP

B/W Monitor

ECHO SIGNAL

SCAN

NOVA BUS

USC BUS

LOGIQ 200 PRO Series SYSTEM BLOCK DIAGRAM

ILLUSTRATION 5--1

5--4 DIAGRAMS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

5--4 WIRING DIAGRAM

PLUG

L(Brown)

N(Light Blue)

G(Green/Yellow)

TB2

TB1

1

2

3

4

1

NFB1

3

2

4

10A/115V

5A/220V

N(Light Blue)

L(Brown)

ISOLATION TRANSFORMER

PRI

:

SEC

0

100

7

4

115

Core

9

8

7

8

5

0

100

12

0

9

6

115

10

1

100

115

0

11

10

POWER SW

TB2

1

2

3

4

5

6

L

N

AC OUTLET B/W PRT

L G

Orange

Gray

NFB3

NFB4

2A/115V

1A/220V

N

TB3

1

2

N G L

CON1

11

12

2

3

100

115

15 17

100

0

16

NFB2

3A/115V

3

4

5

6

5

6

2

3

7

8

1

4

J2

J3

J1

8

2

4

2

3

5

6

2

1

BPHV ASSY

GND

AC110 V

+SHV

7

+THV

GND

HV--REF

--5V

+5V

-- SHV

--THV

1

2

1

2

4

3

5

MON ASSY

+12V

GND

FG

MON_VID

RTN

REAR PANEL

F SW

SHTR

EXT_VID_IN

EXT_VID_OUT

PRINTER_A

SERVICE

PRINTER_B

ECG

PIOP ASSY

DC FAN SET

MOD

KEY B’D

L N

FG

HK100A--5/A

+5V G

L N FG

HKT--50--522/A

--12V +12V G1 +5V G2

L N FG

HK50A--12/A

+12V G

CON2

B/W PRINTER

PRT_VID

SHTR

1 2

+

5

VA

G

N

D

J25

1 2 3 4 5

+

5

VD

+

5

VD

+

5

VD

G

N

D

G

N

D

J12

1

--

12

VA

+

12

VA

2

J19

3 4 5

--

5

V

G

N

D

G

N

D

1 2

1

+

12

V

J6

G

N

D

+

5V

J7

G

N

D

2

+

12

V

1 2

G

N

+

12

V

G

N

+

12

V

D

G

N

D

D

J3 J23 J24

J1 J29 J4 J16 J17 J18

J5

J11

MOTHER B’D

LOGIQ 200 PRO Series SYSTEM WIRING DIAGRAM

ILLUSTRATION 5--2

5--5

J5 J6 J9 J8 J1 J2 J3 J500

J10

J4

PIF B’D

DIAGRAMS

GE MEDICAL SYSTEMS

REV 0

5--4 WIRING DIAGRAM (Continued)

Brown

Brown

Light Blue

Light Blue

Orange

Gray

Orange

Gray

6

5

4

3

2

1

10

9

8

7

TB2

12

11

115V

100V

100V

0V

115V

100V

0V

0V

115V

100V

0V

115V

Brown

Brown

Light Blue

Orange

Orange

Gray

(100 V)

TB2

12

11

10

9

8

7

6

5

4

3

2

1

115V

100V

0V

115V

100V

0V

115V

100V

0V

115V

100V

0V

AC IN

Peripheral

Out

AC IN

Peripheral

Out

LOGIQ 200 PRO Series SERVICE MANUAL

Brown

Brown

Light Blue

Light Blue

Orange

Orange

Gray

Gray

Brown

Brown

Light Blue

Orange

Orange

Gray

Gray

(115 V)

TB2

12

11

6

5

4

3

2

1

10

9

8

7

115V

100V

0V

115V

100V

0V

115V

100V

0V

115V

100V

0V

6

5

4

3

2

1

10

9

8

7

TB2

12

11

115V

100V

100V

0V

115V

100V

0V

0V

115V

100V

0V

115V

(220 -- 240 V)

(INPUT: 220 -- 240 V

OUTPUT: 115 V)

LOGIQ 200 PRO Series SYSTEM WIRING DIAGRAM (CONTINUED)

ILLUSTRATION 5--3

AC IN

2235374

Peripheral

Out

AC IN

Peripheral

Out

5--6 DIAGRAMS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

5--5 CIRCUIT BOARD DESCRIPTION

The following table lists circuit boards and their respective card cage slot assignments on the mother board of the

LOGIQ 200 PRO Series system.

CARD CAGE SLOT

3

4

1

2

BOARD NAME

TBF BD ASSY

RBF BD ASSY

MVP BD ASSY

MSTE BD ASSY

SCAN BD ASSY

BPHV BD ASSY

MOTHER 3 ASSY

TABLE 5-- 1

CIRCUIT BOARD DESCRIPTION

DESCRIPTION

Transmit Beam Former Board Assembly

Receive Beam Former Board Assembly

Multi--Function Video Processor Board Assembly

Master Enhanced Board Assembly

Scan Board Assembly

Bipolar High Voltage Board Assembly

Motherboard Assembly

NOTE

5--7 DIAGRAMS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

5--8 DIAGRAMS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CHAPTER 6 -- RENEWAL PARTS

TABLE OF CONTENTS

TITLE SECTION

6--1

6--2

PAGE

RENEWAL PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--3

DISASSEMBLY/RE--ASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--1 Monitor Assy (FRU No. 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--2 Pot Set Assy (FRU No. 101)

6--2--3 CRT Filter(FRU No. 102)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--4 Monitor Cover Set (FRU No. 103)

6--2--5 Tilt Assy(FRU No. 104)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--6 Filter Clamp Set (FRU No. 105) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--7 Probe Holder (FRU No. 201) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--8 Holder Bracket Assy (FRU No. 202) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--9 Cover Set (FRU No. 203) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--10 Keyboard Assy (FRU No. 204) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--11 Trackball Assy (FRU No. 205) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--12 TGC Assy (FRU No. 206) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--13 Encoder Set (FRU No. 207) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--14 Freeze/Record Key Assy (FRU No. 208)

6--2--15 TGC Knob Set (FRU No. 209)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--16 Rubber Key Set (FRU No. 210) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--17 P.C.Board(s) (FRU No.301 through 304)

6--2--18 BPHV Assy (FRU No. 305)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--19 SCAN Assy (FRU No. 306) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--20 Shield Panel (FRU No. 307) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--21 Mother 3 Assy (FRU No. 308) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--22 Nest Box (FRU No. 309) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--23 DC Fan Assy (FRU No. 310) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--24 Swing Arm Assy (FRU No. 401) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--25 Pipe Assy (FRU No. 402) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--26 Pole Cover (FRU No. 403) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--27 Left Cover (FRU No. 404) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--28 Right Cover (FRU No. 405) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--29 Front Base Cover (FRU No. 406) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--30 Front Cover (FRU No. 407) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--44

6--46

6--48

6--50

6--52

6--54

6--56

6--58

6--60

6--62

6--64

6--65

6--66

6--68

6--29

6--30

6--32

6--34

6--36

6--37

6--38

6--39

6--40

6--42

6--21

6--22

6--24

6--25

6--26

6--27

6--28

6--1 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

TABLE OF CONTENTS (continued)

SECTION TITLE PAGE

6--2--31 Rear Cover (FRU No. 408)

6--2--32 Top Cover (FRU No. 409)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--33 Printer Cover (FRU No. 410) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--34 Printer Bracket Assy(FRU No. 411) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--35 Rear Handle (FRU No. 412) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--36 Neck Frame (FRU No. 413) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--37 Rear Panel Assy (FRU No. 414) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--38 Power S/W Assy (FRU No. 415)

6--2--39 Front Caster (FRU No. 416)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--40 Rear Caster (FRU No. 417) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--41 Bumper Set (FRU No. 418) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--42 EMI Cover L (FRU No. 419) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--43 PCB Guide Assy (FRU No. 420) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--44 Terminal Block 12 Assy (FRU No. 501) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--45 Terminal Block 6 Assy (FRU No. 502) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--46 SMPS Assy (FRU No. 503) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--47 Power Trans Assy (FRU No. 504) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--2--48 Circuit Breaker Set (FRU No. 505) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--88

6--89

6--90

6--92

6--94

6--96

6--98

6--100

6--70

6--72

6--74

6--76

6--78

6--80

6--82

6--84

6--86

6--87

6--3

6--4

FUSE REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--3--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--3--2 Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--101

6--101

6--101

SMPS Assy REPLACEMENT

6--4--1 Introduction

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--4--2 Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6--104

6--104

6--104

6--2 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 2

6--1 RENEWAL PARTS

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

LOGIQ 200 PRO Series

MATERIAL LIST (1)

PART NAME

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

PART NO.

QUANTITY

2233819 1

2242419

2233820

2233821

2233822

2233823

2233824

2233825

2233826

1

1

1

1 1

1 1

1

1

220V, 4SW

220V, 4SW

100V, 4SW

115V, 4SW

220V, 4SW

220V, 4SW

115V, 2SW

220V, 2SW

1 220V, 2SW

DESCRIPTION

VER. 3.0

VER. 3.0

VER. 3.0

VER. 3.0

VER. 3.0

VER. 3.0

VER. 3.0

VER. 3.0

VER. 3.0

PROBE CABLE ARM 2215992 1 1 1 1 1 1 1 1 1 1 1

MOD INSTALL KIT 2235371 1 1 1 Include MOD

RIGHT & LEFT MTZ INSTALL KIT

FOOT SWITCH ASSY

CINE S/W

VINYL COVER

INSITE KIT

2242590

2162242

2242585 1

2235373

2242586

1 1 1 1

1 1 1

1 1 1

PRINTER INSTALL KIT 2242589 1 1 1 1 1 1 1 1 1 1 1

ATO S/W 2251544 1

1

ADVANCED REF. MANUAL

2233222--100

1 1 1 1 1

SERVICE MANUAL 2235374

QUICK START GUIDE

2234974--100

SONY B/W PRINTER E8310KA

1 1 1 English

1 1 1 1 1 English

1 1 1 English

1 1 1

ECG INSTALL KIT

L500 ECG CABLES

ECG CABLE SET

ECG CABLE ASSY

ECG CABLE ASSY

2242584

P9509KH

2137161

P9509KG

2137160

Clip type

Rod type

Japan only, Clip type

Japan only, Rod type

KOREA (H43132LA)

PHILIPPINES (H43132LK)

JAPAN (H43132LB)

TAIWAN (H43132LC)

CHINA, HONGKONG, INDIA (H43132LD)

SOUTHEAST ASIA, AUSTRALIA, NJ (H43132LD)

CHILE (H43132LH)

BRAZIL (H43132LG)

USA (H43132LF)

EUROPE (H43132LL)

EUROPE (H43132LE)

6--3 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 6

6--1 RENEWAL PARTS (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

PART NAME

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

CONSOLE ASSY

LOGIQ 200 PRO Series

MATERIAL LIST (2)

PART NO.

QUANTITY

2270964 1

2270972

2270965

2270966

2271424

2270967

2270968

2270969

2270970

2270971

1

1

1

1

1

1 1

1

1

220V, 4SW

220V, 4SW

100V, 4SW

115V, 4SW

220V, 4SW

220V, 4SW

220V, 4SW

115V, 2SW

220V, 2SW

1 220V, 2SW

DESCRIPTION

VER. 3.1

VER. 3.1

VER. 3.1

VER. 3.1

VER. 3.1

VER. 3.1

VER. 3.1

VER. 3.1

VER. 3.1

VER. 3.1

PROBE CABLE ARM 2215992 1 1 1 1 1 1 1 1 1 1 1

MOD INSTALL KIT 2235371 1 1 1 1 Include MOD

MTZ INSTALL KIT 2242590 RIGHT

LEFT MTZ INSTALL KIT

FOOT SWITCH ASSY

CINE S/W

2242591

2162242

2242585 1 1 1 1 1

1 1 1

1 1 1

PRINTER INSTALL KIT 2242589 1 1 1 1 1 1 1 1 1 1 1

ATO S/W

VINYL COVER

INSITE KIT

2251544

2235373

2242586

1

1

ADVANCED REF. MANUAL

SERVICE MANUAL

2233222--100

2235374

QUICK START GUIDE

2234974--100

SONY B/W PRINTER E8310KA

ECG INSTALL KIT 2242584

1 1 1 1 1 1 1 1 English

1 1 1 1 1 English

1 1 1 English

1 1 1

KOREA (H43142LA)

PHILIPPINES (H43142LK)

JAPAN (H43142LB)

TAIWAN (H43142LC)

INDIA (H43142LM)

SEA, AUS, NJ, CHINA, HONGKONG (H43142LD)

6--4

CHILE (H43142LH)

BRAZIL (H43142LG)

USA (H43142LF)

EUROPE (H43142LL)

EUROPE (H43142LE)

RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

OPERATOR CONSOLE ASSY

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

MONITOR BLOCK

FRU 100 -- 105

See ILLUSTRATION 6--2

KEYBOARD BLOCK

(FRU 201 -- 210)

See ILLUSTRATION 6--3

PC BOARDS(2)

See ILLUSTRATION 6--4

BODY BLOCK

See ILLUSTRATION 6--5

PC BOARDS(1)

See ILLUSTRATION 6--4

POWER SUPPLY BLOCK

See ILLUSTRATION 6--6

ILLUSTRATION 6--1

6--5 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

OPERATOR CONSOLE ASSY 1/7

101

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

102

105

104

103

ILLUSTRATION 6--2

6--6 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 6

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

MATERIAL LIST (1/8)

FRU

NO.

PART NAME

100 MONITOR ASSY

MONITOR ASSY

MONITOR ASSY

101 POT SET ASSY

102 CRT FILTER (PLASTIC)

CRT FILTER (GLASS)

103 MONITOR COVER SET

MONITOR COVER SET

MONITOR COVER SET

104 TILT ASSY

105 FILTER CLAMP SET

: V3.0

: V3.1

PART NO QTY

.

2235375--3 1

FRU

2235376--3

2249964--3

2148193

2251535

1

1

1

1

DESCRIPTION

1 Including Caution Label,

Glass Filter

1 Including Caution Label

, Plastic Filter

1 Including Caution Label,

For JAPAN only

2 Bright and Contrast,

Include Knob

2 Include 2 brackets

SECTION FOR

REFERENCE

6--2--1

6--2--1

6--2--1

6--2--2

6--2--3

2175438

2235385

2249415

2249966

2249414

2214397

2214398

1

1

1

1

2 Include 2 brackets

For USA, Brazil, Chile,

2 For Plastic Filter

For Glass Filter

For Japan only

2

2 For Plastic Filter

For Glass Filter

6--2--3

6--2--4

6--2--4

6--2--4

6--2--5

6--2--6

6--7 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

OPERATOR CONSOLE ASSY 2/7

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

206, 209

205

208

204

207

210

203

201

202

PG Rivet

ILLUSTRATION 6--3

6--8 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 6

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

FRU

NO.

PART NAME

201 PROBE HOLDER

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

202 HOLDER BRACKET ASSY

203 COVER SET

204 KEY BOARD ASSY

205 TRACKBALL ASSY

206 TGC ASSY

207 ENCODER SET

208 FREEZE/RECORD KEY ASSY

209 TGC KNOB SET

210 RUBBER KEY SET

: V3.0

: V3.1

MATERIAL LIST (2/8)

PART NO QTY.

FRU

2175439

2148199

2235384

2235386--2

2235387

2235388--2

2235390

2235391

2214594

2249182

1

1

1

1

1

1

1

1

1

1

2

2

2

1

2

2

2

2

2

1

DESCRIPTION

Include Probe Cup

Holder and Probe Sheet

Include PG Rivet

SECTION

FOR

REFERENCE

6--2--7

Include FRU 205--210

3 Knobs

8 Knobs

6--2--8

6--2--9

6--2--10

6--2--11

6--2--12

6--2--13

6--2--14

6--2--15

6--2--16

6--9 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

OPERATOR CONSOLE ASSY 3/7

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

308

310

309

301--304

306

305

307

ILLUSTRATION 6--4

6--10 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 6

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

MATERIAL LIST (3/8)

FRU

NO.

PART NAME PART NO QTY FRU DESCRIPTION

301 TBF ASSY

302 RBF ASSY

303 MVP ASSY

304 MSTE ASSY

305 BPHV ASSY

306 SCAN ASSY

307 SHIELD PANEL

308 MOTHER3 ASSY

309 NEST BOX

310 DC FAN ASSY

2216833--2

2216835--2

2233428--3

2266029--3

2216832--9

2266030--10 1

2240652 1

1

1

1

1

1

2217956

2240645

2220786--2

2235392

2244961

1

1

1

1

1

1 software ver 3.02 or later

1

1 software ver 3.05 or later

1 software ver 3.15 or later

1 software ver 3.06b

1 software ver 3.17a

1 Include Bracket

1

2

1 Include Bracket Probe

2

1 2 fans

6--2--18

6--2--19

6--2--20

6--2--21

6--2--22

6--2--23

: V3.0

: V3.1

SECTION

FOR

REFERENCE

6--2--17

6--11 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

OPERATOR CONSOLE ASSY 4/7

LOGIQ 200 PRO Series SERVICE MANUAL

412

408

402

401

403

409

405

420

413

419

414

415

404

417

407

406

416

418

411

410

2235374

ILLUSTRATION 6--5

6--12 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 3

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

FRU

NO.

PART NAME

401 SWING ARM ASSY

402 PIPE ASSY

403 POLE COVER

404 LEFT COVER

405 RIGHT COVER

406 FRONT BASE COVER

407 FRONT COVER

408 REAR COVER

409 TOP COVER

410 PRINTER COVER SET

411 PRINTER BRACKET ASSY

PRINTER BRACKET ASSY

PRINTER BRACKET ASSY

MATERIAL LIST (4/8)

PART NO

QTY.

FRU

2148212

2256761

2148214

1

1

1

1

1

2

DESCRIPTION

Include Cable Assy

Include Cover Pole,

Curtain Pole

2235396

2235397

2148217

2148218--2

2175441

2158086

2235395

2242780

1

1

1

1

1

1

1

1

1

2

1

2

2

1

1

1

Include Caution Label

2242781

2242782

1

1

2

2

For Korea, Taiwan,

Phillippines, USA, Chile

For Japan

For China, Hongkong,

Southeast Asia, India,

Australia, Europe, Brazil

: V3.0

: V3.1

SECTION

FOR

REFERENCE

6--2--24

6--2--25

6--2--26

6--2--27

6--2--28

6--2--29

6--2--30

6--2--31

6--2--32

6--2--33

6--2--34

6--13 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 3

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

FRU

NO.

PART NAME

412 REAR HANDLE

413 NECK FRAME

414 REAR PANEL ASSY

REAR PANEL ASSY

REAR PANEL ASSY

415 POWER S/W ASSY

416 FRONT CASTER

417 REAR CASTER

REAR CASTER

418 BUMPER SET

419 EMI COVER L

420 PCB GUIDE ASSY

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

: V3.0

: V3.1

MATERIAL LIST (4/8)(Continued)

PART NO

QTY.

FRU DESCRIPTION

2148222

2148223--2

2240733

1

1

1

2

2

1

2240734

2240735

1

1

1

1

For Korea, Taiwan,

Phillippines, USA, Chile

For Japan

For China, Hongkong,

Southeast Asia, India,

Australia, Europe, Brazil

2148226

2192111

2192112

1

2

2

1

1

1

2192113 2 2

Swivel

For Japan, Taiwan,

Phillippines, China, Hongkong,

Southeast Asia, India,

Australia, Europe

Fixed

For USA, Brazil, Chile

2148230

2169025

2169263

1

1

1

2

2

2

SECTION

FOR

REFERENCE

6--2--35

6--2--36

6--2--37

6--2--38

6--2--39

6--2--40

6--2--41

6--2--42

6--2--43

6--14 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--15 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

OPERATOR CONSOLE ASSY 5/7

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

504

503

501

505

502

ILLUSTRATION 6--6

6--16 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 3

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

MATERIAL LIST (5/8)

FRU

NO.

PART NAME

501 TERMINAL BLOCK 12 ASSY

502 TERMINAL BLOCK 6 ASSY

503 SMPS ASSY

504 POWER TRANS ASSY

(INPUT:220V--OUTPUT:115V)

POWER TRANS ASSY

(INPUT:100V--OUTPUT:100V)

POWER TRANS ASSY

(INPUT:115V--OUTPUT:115V)

POWER TRANS ASSY

(INPUT:220V--OUTPUT:220V)

505 CIRCUIT BREAKER SET

CIRCUIT BREAKER SET

CIRCUIT BREAKER SET

PART NO QTY.

FRU

2148231 1 2

2248909

2235398

2243660

1

1

1

2

1

1

2243661 1 1

DESCRIPTION

12 pins

6 pins

For Korea, Chile,

Phillippines

For Japan

2243662

2243664

2174991

2174992

2174993

1

1

1

1

1

1

1

1

1

1

For Taiwan, USA

For China, Hongkong,

India, Southeast Asia,

Australlia, Europe,

Brazil

2A(2),3A, 5A ,

For Korea, Chile,

Phillippines

2A(2), 3A, 10A

For Japan, Taiwan,

USA

1A(2), 3A, 5A

For China, Hongkong,

India, Southeast Asia,

Australia, Europe, Brazil

: V3.0

: V3.1

SECTION FOR

REFERENCE

6--2--44

6--2--45

6--2--46

6--2--47

6--2--48

6--17 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 6

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

MATERIAL LIST (6/8)

FRU

NO.

PART NAME PART NO QTY.

FRU DESCRIPTION

601 CABLES SET

602 HARDWARE SET

2235401

2235402

1

1

1

2

: V3.0

: V3.1

CABLES for Boards,

Power Switch and

Cable Tie.

Screws, Washers, Nuts,

Cable Tie, Screw Cap...

Include Washer

SECTION

FOR

REFERENCE

603 CABLE HOOK ARM SET

604 TOP COVER SHEET

605 FUSE SET

2170597

2158087

2249717--2

1

1

1

2

2

1 BPHV--1A250V,

0.5A250V

Monitor--3A 250V

6--

3

6--18 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 6

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

FRU

NO.

PART NAME

701 CBF PROBE

702 CAE PROBE

703 MTZ PROBE

704 CZB PROBE

705 LH PROBE

706 LE PROBE

707 LI PROBE

708 LT PROBE

709 LB PROBE

710 LD PROBE

711 CS PROBE

712 10Lb PROBE

713 ERB PROBE

714 3C PROBE

715 S317 PROBE

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

: V3.0

: V3.1

2152422

2152402

P9601AS

P9601AC

P9601AR

P9601AB

P9601AW

P9601AG

P9601AX

P9601AJ

P9601AQ

P9601AA

2124317

P9601AD

2202320

2202315

H45202LM

H45202ER

H45202WB

H45202SD

MATERIAL LIST (7/8)

PART NO

P9603AD

P9603AA

P9603AE

P9603AB

P9603AU

P9603AL

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

QTY.

FRU

1 1

1 1

1

1

1

1

1

1

1 1

DESCRIPTION

For Japan Only

For Japan Only

Not Include MTZ HOLDER

ASSY

Not Include MTZ HOLDER

ASSY

For Japan Only

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1 For Japan Only

For Japan Only

For Japan Only

For Japan Only

For Japan Only

For Japan Only

For Japan Only

For Japan Only

SECTION FOR

REFERENCE

6--19 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 3

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2233819 -- 2233826

2270964 -- 2270972, 2271424

LOGIQ 200 PRO Series

OPERATOR CONSOLE ASSY

MATERIAL LIST (8/8)

FRU

NO.

PART NAME

717 FOOT SWITCH ASSY

PART NO QTY.

FRU

2162242 1 2

718 RIGHT MTZ OPTION INSTALL KIT 2242590

719 B/W PRINTER INSTALL KIT 2242589

720 ECG INSTALL KIT 2242584

1

1

1

2

2

2

721 CINE S/W

722 INSITE INSTALL KIT

2242585

2242586

723 PROBE CABLE ARM INSTALL KIT 2215992

724 MOD INSTALL KIT 2235371

725 ATO S/W

726 PIOP2 ASSY

2251544

2181871

727 LEFT MTZ OPTION INSTALL KIT 2242591

728 MOD ASSY 2242409

729 VER 3.16 UPGRADE KIT 2273388

1

1

1

1

1

1

1

1

1

2

2

2

2

2

2

2

2

2

DESCRIPTION

FOR B/W PRINTER

Version between 3.10 to 3.15

: V3.0

: V3.1

REFEREN

CE

8--1

8--1

6--20 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2 DISASSEMBLY/RE--ASSEMBLY

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

WARNING!

ONLY QUALIFIED SERVICE PERSONNEL SHOULD REMOVE ANY COVERS OR PANELS.

ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE. BECOME THOROUGHLY

FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT LEVELS TO AVOID

ACCIDENTAL CONTACT

CAUTION

Do not wear the ESD wrist strap when you remove the SMPS Assy. Turn OFF power and unplug the power cord before removing any part of SMPS Assy. However be sure to turn off power and wear the strap before you remove a circuit boards.

CAUTION

Do NOT unplug the power cord before turning OFF the power switch.

6--21 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--1 Monitor Assy (FRU No. 100)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

WARNING!

PERSONAL INJURY HAZARD. VIDEO MONITOR SUPPORT ARM IS SPRING LOADED.

RELEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD

MOVEMENT. KEEP YOUR HEAD AND BODY AWAY FROM ABOVE VIDEO MONITOR

SUPPORT ARM.

Refer to ILLUSTRATION 6--7.

1.

Turn OFF the system.

2.

Lift and set the Monitor to the highest position by pushing down the Up/down Release Button located on the

Swing Arm Assy.

3.

Remove the Monitor Bottom Cover. Refer to 6--2--4.

4.

Disconnect the connector.

5.

Unscrew one screw (1).

6.

Pull the Monitor Assy upwards.

6--22 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--1 Monitor Assy (FRU No. 100) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

100 Monitor Assy connector

1

MONITOR ASSY DISASSEMBLY

ILLUSTRATION 6--7

6--23 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--2 Pot Set Assy (FRU No. 101)

Time Required

6 Minute

Tool Required

Screwdriver

Soldering Iron

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Procedure

Refer to ILLUSTRATION 6--8.

1.

Turn OFF the system.

2.

Remove the Monitor Cover Set. Refer to 6--2--4.

3.

Remove two Potentiometer Knob and unscrew two hexagonal nuts on each Pot.

4.

Unscrew four screws (1 -- 4) and remove the EMI Cover and PCB.

5.

Remove the wire from PCB with Soldering Iron.

6.

Remove the Pot Set Assy.

101 Pot Set Assy

Hexagonal Nut

Potentiometer Knob

POT SET DISASSEMBLY

ILLUSTRATION 6--8

6--24 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--3 CRT Filter(FRU No. 102)

Time Required

2 Minutes

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Not necessary

Procedure

Refer to ILLUSTRATION 6--9.

1.

Turn OFF the system.

2.

Push the two Filter Clamp Assy inward (1) as shown in ILLUSTRATION 6--9 and remove Filter Clamp Assy.

3.

Pull forward CRT Filter.

1

1

102 CRT Filter

105 Clamp Set

CRT FILTER DISASSEMBLY

ILLUSTRATION 6--9

6--25 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--4 Monitor Cover Set (FRU No. 103)

Time Required

4 Minutes

Tool Required

Screwdriver

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Procedure

Refer to ILLUSTRATION 6--10.

1.

Turn OFF the system.

2.

Lift and set the Monitor to the highest position by pushing down the Up/down Release Button located on the

Swing Arm Assy.

3.

Unscrew one screws (1) and remove the Monitor Bottom Cover.

4.

Unscrew four screws (2 -- 5) and remove the Monitor Rear Cover.

5.

Unscrew four screws (6 -- 9) and remove the Monitor Front Cover.

Monitor Rear Cover

2

4

3

5

8

Monitor Bottom Cover

9

Monitor Front Cover

6

1

7

MONITOR COVER ASSY DISASSEMBLY

ILLUSTRATION 6--10

103 Monitor Cover Set

6--26 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--5 Tilt Assy(FRU No. 104)

Time Required

8 Minutes

Tool Required

Screwdriver

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Procedure

Refer to ILLUSTRATION 6--11.

1.

Turn OFF the system.

2.

Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6--2--1.

3.

Remove the Monitor Rear Cover. Refer to 6--2--4.

4.

Unscrew four screws (1 -- 4).

5.

Remove the Tilt Assy.

104 Tilt Assy

3

4

1

2

TILT ASSY DISASSEMBLY

ILLUSTRATION 6--11

6--27 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--6 Filter Clamp Set (FRU No. 105)

Time Required

6 Minutes

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--9.

1.

Turn OFF the system.

2.

Push the two Filter Clamp Assy inward (1) as shown in ILLUSTRATION 6--12 and remove Filter Clamp Assy.

1

1

105 Filter Clamp Set

FILTER CLAMP SET DISASSEMBLY

ILLUSTRATION 6--12

6--28 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--7 Probe Holder (FRU No. 201)

Time Required

2 Minutes

Tool Required

Not necessary

Procedure

Refer to ILLUSTRATION 6--13.

1.

Turn OFF the system.

2.

Pull the PG rivet located at bottom of the Probe Holder.

3.

Pull the Probe Holder out as shown in ILLUSTRATION 6--13.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

201 Probe Holder

PG Rivet

PROBE HOLDER DISASSEMBLY

ILLUSTRATION 6--13

6--29 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--8 Holder Bracket Assy (FRU No. 202)

Time Required

3 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--14.

1.

Turn OFF the system.

2.

Remove the Probe Holder (FRU No. 201). Refer to 6--2--7.

3.

Unscrew three screws (1 -- 3).

4.

Remove the Holder Bracket Assy.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

202 Holder Bracket Assy

1

2

3

HOLDER BRACKET ASSY DISASSEMBLY

ILLUSTRATION 6--14

6--30 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--31 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--9 Cover Set (FRU No. 203)

Time Required

11 Minutes

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--15.

1.

Turn OFF the system.

2.

Remove the Probe Holder (FRU No. 201). Refer to 6--2--7.

3.

Remove the Holder Bracket Assy (FRU No.202) Refer to 6--2--8.

4.

Remove MOD Assy. (FRU No.723). Refer to Chapter 8.

5.

Unscrew the two screw(1 and 2) on the front bottom of Keyboard .

6.

Unscrew the ground screw ( ) and remove the Ground Terminal.

7.

Disconnect the connector and cut the tie wraps off.

8.

Remove the Keyboard Assy (FRU No. 204).

9.

Unscrew nine screws (3 -- 11) and remove the Keyboard assy.

6--32 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--9 Cover Set (FRU No. 203) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Connector

Cover Set

Ground Terminal

1

2

3

4

5

6

11

10

9

7

8

COVER SET DISASSEMBLY

ILLUSTRATION 6--15

6--33 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--10 Keyboard Assy (FRU No. 204)

Time Required

6 Minutes

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--16.

1.

Turn OFF the system.

2.

Remove MOD Assy (FRU.723). Refer to Chapter 8.

3.

Unscrew two screws (1 and 2) on the front bottom of Keyboard.

4.

Unscrew the ground screw ( ) and remove the Ground Terminal.

5.

Disconnect the connector and cut the tie wraps off.

6.

Pull the Keyboard Assy out.

6--34 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--10 Keyboard Assy (FRU No. 204) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

204 Keyboard Assy

Connector

Ground Terminal

1

2

KEYBOARD ASSY DISASSEMBLY

ILLUSTRATION 6--16

6--35 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--11 Trackball Assy (FRU No. 205)

Time Required

8 Minutes

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--17.

1.

Turn OFF the system.

2.

Pull the Keyboard Assy out.

Refer to 6--2--10.

3.

Unscrew three screws (1 -- 3) and disconnect one connector (CN5) on Trackball Assy.

4.

Remove Trackball Assy.

3

1

2

Connector

CN5

TRACKBALL ASSY DISASSEMBLY

ILLUSTRATION 6--17

6--36 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--12 TGC Assy (FRU No. 206)

Time Required

8 Minutes

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--18.

1.

Turn OFF the system.

2.

Pull the Keyboard Assy out . Refer to 6--2--10.

3.

Pull out the TGC Knobs (5 -- 12) on Keyboard Assy.

4.

Unscrew four screws (1 -- 4) and disconnect one connector (CN12).

5.

Remove TGC Assy.

1

2

3

4

Connector

CN12

5

6

7

8

9

11

12

10

TGC Knob

TGC ASSY DISASSEMBLY

ILLUSTRATION 6--18

6--37 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--13 Encoder Set (FRU No. 207)

Time Required

14 Minutes

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Screwdriver, Hexagonal Wrench

Procedure

Refer to ILLUSTRATION 6--19

1.

Turn OFF the system.

2.

Pull the Keyboard Assy out. Refer to 6--2--10

3.

Unscrew hexagonal screws (1) on each Encoder Knob and remove three knobs.

4.

Unscrew two screws on each Encoder Bracket and disconnect three connectors (CN1 --CN3).

5.

Remove Encoder Set.

Encoder Bracket

1

CN3

CN2

CN1

ENCODER SET DISASSEMBLY

ILLUSTRATION 6--19

6--38 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--14 Freeze/Record Key Assy (FRU No. 208)

Time Required

10 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--20.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1.

Turn OFF the system.

2.

Pull the Keyboard Assy out. Refer to 6--2--10.

3.

Unscrew three screws (1 -- 3) and disconnect the connector (CN11).

4.

Remove Freeze/Record Key Assy.

1

2

3

1

2

Connector

Connector

3

CN11

FREEZE/RECORD KEY ASSY DISASSEMBLY

ILLUSTRATION 6--20

6--39 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--15 TGC Knob Set (FRU No. 209)

Time Required

5 Minutes

Tool Required

None

Procedure

Refer to ILLUSTRATION 6--21.

1.

Turn OFF the system.

2.

Pull out the TGC Knobs (1 -- 8).

1 2

3 4

5 6

7 8

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

TGC KNOB SET DISASSEMBLY

ILLUSTRATION 6--21

6--40 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--41 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--16 Rubber Key Set (FRU No. 210)

Time Required

20 Minutes

Tool Required

Screw Driver.

Procedure

Refer to ILLUSTRATION 6--22.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1.

Turn OFF the system.

2.

Remove the Keyboard assy. Refer to 6--2--10.

3.

Pull out the TGC knobs. Refer to 6--2--15.

4.

Using the hexagonial wrench, remove the three Encoder knobs. Refer to 13.

5.

Remove the Freeze/Record key. Refer to 6--2--14.

6.

Disconnect five connectors (CN1, CN2, CN3, CN5, CN11, CN12).

7.

Unscrew 20 screws (1--20) and remove the board from the Keyboard assy. Refer to

ILLUSTRATION 6--22.

8.

Carefully remove 2 Rubber key pieces (1--2) from the Keyboard assy. Refer to

ILLUSTRATION 6--22.

6--42 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--16 Rubber Key Set (FRU No. 210) (continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

2

3

4 5

1

Board

6

CN12

17

7

9

8

CN3

10

CN11

20

CN2

CN1

13

Track

Ball

19

CN5

14

15

16

11

12

18

TGC

Board

Rubber Key 1

FRU 208.

RUBBER KEY SET DISASSEMBLY

ILLUSTRATION 6--22

6--43

Rubber Key 2

Track

Ball

TGC

RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--17 P.C.Board(s) (FRU No.301 through 304)

Time Required

7 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--23.

Refer to Proprietary Service Manual (2242594) for initialization.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

An electrostatic discharge may damage a component. Turn OFF power and wear the wrist strap before you remove a circuit boards. Do not unplug the power cord to keep ground continuity.

Do not bend or flex the boards when mounting/dismounting each board. Surface mount IC boards are very susceptible to damage from flex/torque.

Make sure the Dip switch position is set correctly before changing the MSTE Assy. Refer to

.

chapter 3, section 3--6--3 DIP Switch for correct setting.

Refer to ILLUSTRATION 6--23.

1.

Turn OFF the system.

2.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

3.

Remove the PCB Guide Assy (FRU 420). Refer to 6--2--43.

4.

Unscrew four screws (1 -- 4) and Remove the Bracket Nest.

5.

Disconnect the Cable Assy(5) from Board Assy.

6.

Pull the upper and lower ejector in the direction indicated by the arrows(a). Refer to ILLUSTRATION 6--23.

7.

Pull out the board, do not bend it (b).

6--44 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 2

6--2--17 P.C.Board(s) (FRU No.301 through 304) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374 a a b

Cable Assy(5)

Bracket Nest

1

2

3

4

Card Cage No.

3

4

1

2

Board

TBF

RBF

MVP

MSTE

P.C.BOARD(S) REMOVAL

ILLUSTRATION 6--23

CAUTION

In the case of the system for India, MSTE Dip switch SW2 No.4 should be on for A--Mode

Operation. Make sure the Dip switch position is set correctly before changing the MSTE

Assy. Refer to chapter 3, section 3--6--3 DIP Switch for correct setting.

6--45 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--18 BPHV Assy (FRU No. 305)

Time Required

10 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--24.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover (FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

5.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

6.

Unscrew four screws (1 -- 4), and remove top cover.

7.

Unscrew four screws (5 -- 8) being used fixing TR.

8.

Disconnect three connectors (J1 -- J3) on the BPHV Assy.

9.

Unscrew and remove four supporters(S1 -- S4).

10.

Remove the BPHV Assy.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--46 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--18 BPHV Assy (FRU No. 305) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

J2

1

J1

S1

S3

3

S2

J3

S4

2

4

BPHV ASSY DISASSEMBLY

ILLUSTRATION 6--24

5

6

7

8

6--47 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--19 SCAN Assy (FRU No. 306)

Time Required

13 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--25.

1.

Turn OFF the system and plug out.

2.

Remove the Shield Panel (FRU 307). Refer to 6--2--20.

3.

Unscrew four screws (1 -- 5).

4.

Disconnect two connector.

5.

Pull out the Probe SCAN Assy.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

306 SCAN Assy

Two Connectors

3

SCAN ASSY DISASSEMBLY

ILLUSTRATION 6--25

1

2

4

5

6--48 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--49 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--20 Shield Panel (FRU No. 307)

Time Required

11 Minutes

Tool Required

Screwdriver

Procedure

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Do NOT unplug the power cord before OFF the power switch.

Refer to ILLUSTRATION 6--26.

1.

Turn OFF the system & plug out

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover (FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

5.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

6.

Remove the Power S/W Assy (FRU 415). Refer to 6--2--38.

7.

Unscrew eleven screws (1 --11).

8.

Remove the Shield Panel (FRU 307).

6--50 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--20 Shield Panel (FRU No. 307) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

307 Shield Panel

9

10

8

7

11

6

5

4

1

2

3

SHIELD PANEL DISASSEMBLY

ILLUSTRATION 6--26

6--51 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--21 Mother 3 Assy (FRU No. 308)

Time Required

0.5 Hours

Tool Required

Screwdriver

Hexagonal Wrench

Procedure

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

An electrostatic discharge may damage a component. Turn OFF power and wear the wrist strap before you remove a circuit boards. Do not unplug the power cord to keep ground continuity. Do not bend or flex the boards when mounting/dismounting each board.

Surface mount IC boards are very susceptible to damage from flex/torque.

Refer to ILLUSTRATION 6--27.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29

5.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

6.

Remove the EMI Cover L (FRU 419). Refer to 6--2--42.

7.

Remove the four PCB Guide Assy (FRU 420). Refer to 6--2--43.

8.

Remove the four P.C.Boards (FRU 301 to 304). Refer to 6--2--17.

9.

Remove the Shield Panel (FRU 307). Refer to 6--2--20.

10. Remove the Scan Assy (FRU306). Refer to 6--2--19.

11.

Disconnect 29 connectors connected with the mother board. (J1 -- J29)

12.

Unscrew two screws (a and b), and remove the Bracket Harness. If necessary, cut the tie wrap.

13.

Unscrew 25 screws (1 -- 25).

14.

Remove the Mother3 Assy (FRU No. 308).

6--52 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--21 Mother 3 Assy (FRU No. 308) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

308 Mother Assy

J4

1

J7

J10

8

J13

J16

J17

9 10

J20

J23

J26

19

15

J24

20

J18

2

J1

J2 J5

3

J8

J11

J14

4

11

J21

16

J27

21

22

J30

J31

5

J9

J15

12

J22

17

J28

23 a b

Bracket Harness

J3 J6

6

7

J12 J19

13

18

14

J25

24

J29

25

MOTHER3 ASSY DISASSEMBLY

ILLUSTRATION 6--27

CAUTION

Before connect the cables, Check and distinguish the labels especially for J12, J19, J3, J6.

6--53 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--22 Nest Box (FRU No. 309)

Time Required

0.6 Hour

Tool Required

Screwdriver

Hexagonal Wrench

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

An electrostatic discharge may damage a component. Turn OFF power and wear the wrist strap before you remove a circuit boards. Do not unplug the power cord to keep ground continuity.

Do not bend or flex the boards when mounting/dismounting each board. Surface mount IC boards are very susceptible to damage from flex/torque.

Procedure

Refer to ILLUSTRATION 6--28.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29

5.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

6.

Remove the Rear Cover (FRU 408). Refer to 6--2--31.

7.

Remove the EMI Cover L (FRU 419). Refer to 6--2--42.

8.

Remove the four PCB Guide Assy (FRU 420). Refer to 6--2--43.

9.

Remove the four P.C.Boards (FRU 301 to 304). Refer to 6--2--17.

10.

Remove the Shield Panel (FRU 307). Refer to 6--2--20.

6--54 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--2--22 Nest Box (FRU No. 309) (Continued)

11.

Remove the SCAN Assy (FRU 306). Refer to 6--2--19.

12.

Remove the Mother3 Assy (FRU 308). Refer to 6--2--21.

13.

Remove the eight Screws (1 -- 8) on both side of the Nest Box. Refer to ILLUSTRATION 6--28.

14.

Remove the Nest Box (FRU 309).

1

2

5 6

3

4

7

8

309 Nest Box

NEST BOX DISASSEMBLY

ILLUSTRATION 6--28

6--55 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--23 DC Fan Assy (FRU No. 310)

Time Required

45 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--29.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Top Cover (FRU 409). Refer to 6--2--32.

5.

Disconnect two connectors (2 x CN01).

6.

Unscrew eight screws (1-- 8) on the fan.

7.

Pull out the Fan Assy outwards.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--56 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--23 DC Fan Assy (FRU No. 310) (Continued)

1

2

3

4

5

6

7

8

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

310 DC Fan Assy

CN01

AC FAN ASSY DISASSEMBLY

ILLUSTRATION 6--29

6--57 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--24 Swing Arm Assy (FRU No. 401)

Time Required

0.6 Hour

Tool Required

Screwdriver

Hex Wrench

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Procedure

Refer to ILLUSTRATION 6--30.

1.

Turn OFF the system and unplug the power cord from outlet.

2.

Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6--2--1.

3.

Remove the Monitor Space Plate.

4.

Unscrew four screws (1 --4) and remove the Cover located at the bottom of the Swing Arm Assy.

5.

Unscrew one screw to remove Ground Terminal ( ).

6.

Unscrew two supporters ( 5 and 6 ) and two screws ( 7 and 8 ).

7.

Remove the Guide.

8.

Unscrew one screws (9).

9.

Remove the Swing Arm Assy (FRU No.401).

6--58 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--24 Swing Arm Assy (FRU No. 401) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

8

7

5

6

4

3

2

1

9

401 Swing Arm Assy

NECK ASSY DISASSEMBLY

ILLUSTRATION 6--30

6--59 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--25 Pipe Assy (FRU No. 402)

Time Required

0.4 Hour

Tool Required

Screwdriver

Hex Wrench

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Procedure

Refer to ILLUSTRATION 6--31.

1.

Turn OFF the system and unplug the power cord from outlet.

2.

Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6--2--1.

3.

Remove the Swing Arm Assy (FRU 401). Refer to 6--2--24.

4.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

5.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

6.

Remove the Rear Cover (FRU 408). Refer to 6--2--31.

7.

Remove the Pole Cover (FRU 403) Refer to 6--2--26.

8.

Remove eleven hexagonal bolts (1 -- 11).

9.

Disconnect two connectors (CN02,J3).

10.

Remove the Pipe Assy.

NOTE

When assembling the Tilt assy, fix two screw(1,7) with one drop of Loctite262.

6--60 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--25 Pipe Assy (FRU No. 402)) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

402 Pipe Assy

Two Connectors

CN02

J3

NECK ASSY DISASSEMBLY

ILLUSTRATION 6--31

4

5

6

1

2

3

9

10

7

8

11

6--61 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--26 Pole Cover (FRU No. 403)

Time Required

20 Minutes

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--32.

1.

Turn OFF the system.

2.

Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6--2--1.

3.

Remove the Swing Arm Assy (FRU 401). Refer to 6--2--24.

4.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

5.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

6.

Remove the Rear Cover (FRU 408). Refer to 6--2--31.

7.

Unscrew two screws (1and 2).

8.

Pull upward the Pole Cover.

6--62 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--26 Pole Cover (FRU No. 403)) (Continued)

403 Pole Cover

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1

2

POLE COVER DISASSEMBLY

ILLUSTRATION 6--32

6--63 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--27 Left Cover (FRU No. 404)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--33.

1.

Turn OFF the system.

2.

Remove four screw caps and unscrew four screws (1 -- 4).

3.

Remove the Left Cover (FRU 404).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

404 Left Cover

1

2

3

4

LEFT COVER DISASSEMBLY

ILLUSTRATION 6--33

6--64 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--28 Right Cover (FRU No. 405)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--34.

1.

Turn OFF the system.

2.

Remove the four screw caps and unscrew four screws (1 -- 4).

3.

Remove the Right Cover (FRU 405).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

405 Right Cover

1

2

RIGHT COVER DISASSEMBLY

ILLUSTRATION 6--34

3

6--65

4

RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--29 Front Base Cover (FRU No. 406)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--35.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Unscrew two screws (1 and 2).

5.

Remove the Front Base Cover (FRU 406).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1

406 Front Base Cover

2

FRONT BASE COVER DISASSEMBLY

ILLUSTRATION 6--35

6--66 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--67 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--30 Front Cover (FRU No. 407)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--36.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

5.

Remove the fore screw caps and unscrew fore screws (1 -- 4).

6.

Remove the Front Cover (FRU 407).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--68 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--30 Front Cover (FRU No. 407) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

407 Front Cover

2

3

1

4

FRONT COVER DISASSEMBLY

ILLUSTRATION 6--36

6--69 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--31 Rear Cover (FRU No. 408)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--37.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the four screw caps and unscrew four screws (1 -- 4).

5.

Remove the Rear Cover(FRU No. 408).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1

408 Rear Cover

2

3

4

REAR COVER DISASSEMBLY

ILLUSTRATION 6--37

6--70 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--71 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--32 Top Cover (FRU No. 409)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--38.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Rear Cover (FRU 408). Refer to 6--2--31.

5.

Pull up the Pole Cover (FRU 403). Refer to 6--2--26.

6.

Remove the Cable Hook Arm (1)

7.

Unscrew two screws (2 -- 3) and four bolts (4 --7).

8.

Remove the Top Cover.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--72 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--32 Top Cover (FRU No. 409) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1

2

3

4

5

6

7

409 Top Cover

REAR COVER DISASSEMBLY

ILLUSTRATION 6--38

6--73 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--33 Printer Cover (FRU No. 410)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--39.

1.

Turn OFF the system.

2.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

3.

Remove the Top Cover (FRU 409). Refer to 6--2--32.

4.

Unscrew the six screw (1 -- 6).

5.

Remove the Printer Cover (FRU No. 410).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1

2

3

4

5 6

410 Printer Cover

PRINTER COVER DISASSEMBLY

ILLUSTRATION 6--39

6--74 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--75 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--34 Printer Bracket Assy(FRU No. 411)

Time Required

30 Minutes

Tool Required

Screwdriver

Hexagonal Wrench

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Procedure

Refer to ILLUSTRATION 6--40.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

5.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

6.

Remove the Top Cover (FRU 409). Refer to 6--2--32.

7.

Remove the Printer Cover (FRU 410). Refer to 6--2--33.

8.

Disconnect three cables connected to Back Bracket Assy as listed below. Cut the tie wraps off if necessary.

a. Unscrew and Disconnect two ring terminals (1 and 2) from the Terminal Block 6 Assy. Refer to 6--2--45,

ILLUSTRATION 6--51.

b. Unscrew and disconnect one ring terminals(Protective Earth) from the chassis.

c. Disconnect two connectors (J5 and J9) from the Mother Assy. Refer to 6--2--21, ILLUSTRATION 6--27.

9.

Remove four hexagonal bolts (1 -- 4) and pull out Printer Bracket Assy with Back Bracket Assy.

10.

Unscrew the four screw (5 -- 8).

11.

Disassemble Printer Bracket Assy with Back Bracket Assy (FRU411).

6--76 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--34 Printer Bracket Assy(FRU No. 411) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

5

6

Back Bracket Assy

411 Printer Bracket Assy

7

8

1

2

3

4

PRINTER BRACKET ASSY AND BACK BRACKET ASSY DISASSEMBLY

ILLUSTRATION 6--40

6--77 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--35 Rear Handle (FRU No. 412)

Time Required

1 Hours

Tool Required

Screwdriver

Hexagonal Wrench

Procedure

Refer to ILLUSTRATION 6--41.

1.

Turn OFF the system.

2.

Remove the Keyboard Assy (FRU 204). Refer to 6--2--10.

3.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

4.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

5.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

6.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

7.

Remove the Top Cover (FRU 409). Refer to 6--2--32.

8.

Remove the Printer Cover (FRU 410). Refer to 6--2--33.

9.

Remove the Neck Frame (FRU 413). Refer to 6--2--36.

10.

Remove four hexagonal bolts (1 -- 4).

11.

Remove the Rear Handle.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--78 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--35 Rear Handle (FRU No. 412) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

412 Rear Handle

1

3

2

4

REAR HANDLE DISASSEMBLY

ILLUSTRATION 6--41

6--79 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--36 Neck Frame (FRU No. 413)

Time Required

0.5 Hours

Tool Required

Screwdriver

Hexagonal Wrench

Procedure

Refer to ILLUSTRATION 6--42.

1.

Turn OFF the system.

2.

Remove the Keyboard Assy (FRU No. 204). Refer to 6--2--10.

3.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

4.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

5.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

6.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

7.

Remove the Top Cover (FRU 409). Refer to 6--2--32.

8.

Remove the Printer Cover (FRU 410). Refer to 6--2--33.

9.

Remove four hexagonal bolts (1 -- 4).

10.

Remove the Neck Frame (FRU No. 413).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--80 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--36 Neck Frame (FRU No. 413) (Continued)

1

2

3

4

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

413 Neck Frame

NECK FRAME DISASSEMBLY

ILLUSTRATION 6--42

6--81 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--37 Rear Panel Assy (FRU No. 414)

Time Required

0.5 Hour

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--43.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Rear Cover (FRU 408). Refer to 6--2--31.

5.

Remove the EMI Cover (FRU 419). Refer to 6--2--42.

6.

Disconnect all connectors out of the Rear Panel assy. If necessary, cut tie wraps. Connectors are listed below.

a. Unscrew and disconnect three ring terminals (1A, 2A and 5A) from the Terminal Block 6 Assy. Refer to

6--2--45, ILLUSTRATION 6--51.

b. Disconnect four connectors (J2,J3,J6 and J8) from the Mother Assy. For each cable, refer to 6--2--21,

ILLUSTRATION 6--27.

7.

Unscrew four screws (1 -- 4) and pull the Rear Panel Assy out.

8.

Remove the Rear Panel Assy.

6--82 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--37 Rear Panel Assy (FRU No. 414) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

414 Rear Panel Assy

1

2

3

4

REAR PANEL ASSY DISASSEMBLY

ILLUSTRATION 6--43

6--83 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--38 Power S/W Assy (FRU No. 415)

Time Required

10 Minutes

Tool Required

Screwdriver

Procedure

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Do NOT unplug the power cord before OFF the power switch.

Refer to ILLUSTRATION 6--44.

1.

Turn OFF the system and unplug the power cord from the outlet.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

5.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

6.

Unscrew four screws (1 -- 4).

7.

Disconnect fore Terminals from the Power S/W Assy.

8.

Remove the Power S/W Assy.

6--84 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--38 Power S/W Assy (FRU No. 415) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1

3

2

4

415 Power S/W Assy

POWER S/W ASSY DISASSEMBLY

ILLUSTRATION 6--44

6--85 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--39 Front Caster (FRU No. 416)

Time Required

15 Minutes

Tool Required

Hexagonal Wrench

Procedure

Refer to ILLUSTRATION 6--45.

1.

Turn OFF the system.

2.

Put a box below the Power Trans box.

3.

Unscrew four hexagonal bolts (1 -- 4) on each Front Caster.

4.

Remove the Front Caster.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

416 Front Caster

1

2

3

4

FRONT CASTER DISASSEMBLY

ILLUSTRATION 6--45

6--86 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--40 Rear Caster (FRU No. 417)

Time Required

15 Minutes

Tool Required

Hexagonal Wrench

Procedure

Refer to ILLUSTRATION 6--46.

1.

Turn OFF the system.

2.

Put a box below the Power Trans box.

3.

Unscrew four hexagonal bolts (1 -- 4) on each Rear Caster.

4.

Remove the Rear Caster.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

417 Rear Caster

1

2

3

4

REAR CASTER DISASSEMBLY

ILLUSTRATION 6--46

6--87 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--41 Bumper Set (FRU No. 418)

Time Required

10 Minutes

Tool Required

Screwdriver

Hexagonal Wrench

Procedure

Refer to ILLUSTRATION 6--47.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

5.

Unscrew three hexagonal bolts (1 -- 3) on each Bumper Set

6.

Remove the Bumper Set (FRU 418).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

418 Bumper Set

3

2

1

1 2

3

BUMPER SET DISASSEMBLY

ILLUSTRATION 6--47

6--88 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--42 EMI Cover L (FRU No. 419)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--48.

1.

Turn OFF the system.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Unscrew eight screws (1 -- 8).

4.

Remove the EMI Cover L (FRU 419).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

419 EMI Cover L

3

1

2

5

4

7

8

6

EMI COVER L DISASSEMBLY

ILLUSTRATION 6--48

6--89 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--43 PCB Guide Assy (FRU No. 420)

Time Required

5 Minutes

Tool Required

Screwdriver

Procedure

Refer to ILLUSTRATION 6--34.

1.

Turn OFF the system.

2.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

3.

Unscrew four screws (1 -- 4).

4.

Remove the PCB Guide Assy (FRU 420).

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

420 PCB Guide Assy

PCB GUIDE DISASSEMBLY

ILLUSTRATION 6--49

1

4

2

3

6--90 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--91 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--44 Terminal Block 12 Assy (FRU No. 501)

Time Required

10 Minutes

Tool Required

Screwdriver

Hexagonal Wrench

Procedure

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Do NOT unplug the power cord before OFF the power switch.

Refer to ILLUSTRATION 6--50.

1.

Turn OFF the system and unplug the power cord from the outlet.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Unscrew the two screws (1 and 2) and pull out the Terminal Block 12 Assy.

4.

Unscrew and Disconnect twelve ring terminals (TB2--1 ~ TB2--12). Refer to NO TAG,Chapter 5. Cut the tie wrap, if necessary.

5.

Unscrew and Disconnect twelve ring terminals (TB2--1 ~ TB2--12).

6.

Remove the Terminal Block 12 Assy.

6--92 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--44 Terminal Block 12 Assy (FRU No. 501) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

501 Terminal Block 12 Assy

TB2--1

TERMINAL BLOCK 12 ASSY DISASSEMBLY

ILLUSTRATION 6--50

1

TB2--12

2

6--93 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--45 Terminal Block 6 Assy (FRU No. 502)

Time Required

10 Minutes

Tool Required

Screwdriver

Procedure

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Do NOT unplug the power cord before OFF the power switch.

Refer to ILLUSTRATION 6--51.

1.

Turn OFF the system and unplug the power cord from the outlet.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Rear Cover (FRU 408). Refer to 6--2--31.

5.

Unscrew and Disconnect six ring terminals (TB3--1 ~TB3--6).

6.

Unscrew and Disconnect six ring terminals (TB3--1 ~TB3--6).

7.

Unscrew the two screws (1and 2) and pull out the Terminal Block 6 Assy.

8.

Remove the Terminal Block 6 Assy.

6--94 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--45 Terminal Block 6 Assy (FRU No. 502) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

502 Terminal Block 6 Assy

TB3--1

TB3--6

1

TB3--1

2

TB3--6

TERMINAL BLOCK 6 ASSY DISASSEMBLY

ILLUSTRATION 6--51

6--95 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--46 SMPS Assy (FRU No. 503)

Time Required

10 Minutes

Tool Required

Screwdriver

Hexagonal Wrench (Key Size: 4 mm)

Procedure

Refer to 6--4, SMPS Assy REPLACEMENT.

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--96 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--97 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--47 Power Trans Assy (FRU No. 504)

Time Required

20 Minutes

Tool Required

Screwdriver

Hexagonal Wrench

Procedure

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Do NOT unplug the power cord before OFF the power switch.

Refer to ILLUSTRATION 6--52.

1.

Turn OFF the system and unplug the power cord from the outlet.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

5.

Remove the Rear Cover (FRU 408). Refer to 6--2--31.

6.

Remove the SMPS Assy (FRU 503). Refer to Refer to 6--4, SMPS Assy REPLACEMENT..

7.

Disconnect sixteen cables listed below. Cut the tie wraps off if necessary. For each cable, refer to

ILLUSTRATION 6--52.

a. Unscrew and Disconnect twelve ring terminals (TB2--1 ~ TB2--12) from the Terminal Block 12 Assy. Refer to 6--2--44.

b. Unscrew and Disconnect two ring terminals. Refer to 6--2--45.

c. Disconnect two terminals from the Power S/W Assy.

8.

Unscrew two screws ( 1 and 2) of front side and two screws (3 and 4) of rear side of the system.

9.

Pull out the Power Trans Assy to the backward of the system.

6--98 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--47 Power Trans Assy (FRU No. 504) (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

TB2--12

TB2--1

504 Power Trans Assy

1

2

3

4

POWER TRANS ASSY DISASSEMBLY

ILLUSTRATION 6--52

6--99 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--2--48 Circuit Breaker Set (FRU No. 505)

Time Required

30 Minutes

Tool Required

Screwdriver

Procedure

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Do NOT unplug the power cord before OFF the power switch.

Refer to ILLUSTRATION 6--53.

1.

Turn OFF the system & unplug the power cord from the outlet.

2.

Remove the Power Trans Assy (FRU 504). Refer to 6--2--47.

3.

Unscrew four screws (1 -- 4)on the front of Power Trans Assy. Refer to ILLUSTRATION 6--53.

4.

Remove the defective Circuit Breaker.

1

3

4

2

CIRCUIT BREAKER DISASSEMBLY

ILLUSTRATION 6--53

6--100 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 6

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--3 FUSE REPLACEMENT

6--3--1 Introduction

LOGIQ 200 PRO Series has four replaceable fuses. The table below states the items of those fuses.

LOCATION

BPHV Assy

BPHV Assy

BPHV Assy

Monitor

TABLE 6-- 1

REPLACEABLE FUSES

Ref. No.

F1

F2

F3

F1

RATING

1 A, SB

500 mA, SB

500 mA, SB

3 A, SB

250 V

250 V

250 V

250 V

REMARKS

(SB ; Slow Blow type)

(SB ; Slow Blow type)

(SB ; Slow Blow type)

(SB ; Slow Blow type)

6--3--2 Replacement Procedures

Concerning replacing each fuse of LOGIQ 200 PRO Series, refer to the following steps;

Note

This system contains no operator serviceable parts inside. Please refer servicing to qualified personnel from GE. (See Address of Chapter 1.)

Parts Required

FRU

605

PART NAME

FUSE SET

Tools Required

Screwdriver

Hexagonal Wrench (Key Size: 3mm)

QTY

1

PART #

2249717--2

REMARKS

6--101 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

6--3--2 Replacement Procedures (Continued)

(A) F1, F2 and F3 on BPHV Assy

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Make sure that only fuse with the required rated current and of the same type is used for replacement.

1. Turn the system power OFF.and unplug the power cord from the outlet. Follow the OSHA Lockout/Tagout

Requirements (29 CFR 1910.147, Direction 46--015902).

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Front Base Cover (FRU 406). Refer to 6--2--29.

5.

Remove the Front Cover (FRU 407). Refer to 6--2--30.

6.

Remove the BPHV top cover. Refer to 6--2--18.

7.

Replace the defective fuse(s) with new one(s). For the location of fuses, refer to ILLUSTRATION 6--54.

8.

Re--install the Covers.

F1 1A 250V

BPHV A’SSY

F3 0.5A 250V

F2 0.5A 250V

BPHV ASSY

ILLUSTRATION 6--54

6--102

Fuse Cap

F3

Fuse Socket

RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 6

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--3--2 Replacement Procedures (Continued)

(B) F1 on Monitor Assy

1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction

46--015902).

2.

Remove the Monitor Bottom Cover and Monitor Rear Cover. Refer to 6--2--4.

3.

Replace the defective fuse with new one. For the location of fuses, refer to ILLUSTRATION 6--55. Make sure that only fuse with the required rated current and of same type is used for replacement.

4.

Re--install the Monitor Bottom Cover and Monitor Rear Cover.

F1 3A 250V

MONITOR ASSY

ILLUSTRATION 6--55

6--103 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--4 SMPS Assy REPLACEMENT

6--4--1 Introduction

This section describes the procedures for replacing the SMPS Assy unit of LOGIQ 200 PRO Series.

6--4--2 Replacement Procedure

READ and UNDERSTAND these instructions thoroughly before proceeding with the replacement. Perform each step in sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer.

CAUTION

Possible operational damage. Failure to strictly follow ESD (Electrostatic Discharge) precautions during this upgrade may cause constant or intermittent operational abnormalities. Strictly follow all precautions.

Check each step as it is completed.

Parts Required

FRU

503

PART NAME

SMPS ASSY

QTY

1

PART #

2235398

REMARKS

Time Required

Approximately one (1) hours

Tools Required

Standard field Service Tool kit

(A) Removing Covers

1.

Turn the system power OFF and unplug the unit.

2.

Remove the Left Cover (FRU 404). Refer to 6--2--27.

3.

Remove the Right Cover(FRU 405). Refer to 6--2--28.

4.

Remove the Rear Cover (FRU 408). Refer to 6--2--31.

5.

Remove the EMI Cover L (FRU 419). Refer to 6--2--42.

6--104 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--4--2 Replacement Procedure (Continued)

(B) Disconnecting Cables

1.

Disconnect seven cables listed below. Cut the tie wraps off if necessary. For each cable, refer to

ILLUSTRATION 6--56.

a. Disconnect four connectors (J6, J12, J19, J25) from the Mother Assy.

b. Unscrew and Disconnect two ring terminals (TB3--3 and TB3--5) from the Terminal Block 6 Assy.

c. Unscrew one ground screw (1).

TB3--3

TB3--5

J12

J19

1

J25

J6

LEFT SIDE VIEW

ILLUSTRATION 6--56

6--105 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--4--2 Replacement Procedure (Continued)

(C) Replacing SMPS Assy

1.

Unscrew two hexagonal bolts (1 and 2). Two bolts are located at the front of SMPS Assy.

2.

Pull out the SMPS Assy forewards. Refer to ILLUSTRATION 6--57.

WARNING!

PERSONAL INJURY HAZARD! SMPS ASSY UNIT WEIGHTS APPROXIMATELY 3.5

KILOGRAMS (7.8 POUNDS). PAY ATTENTION NOT TO BE INJURED.

3.

Remove the SMPS Assy and place new one on the pallet.

4.

Slide and move the pallet to the position just under the console. Make sure that the SMPS Assy unit is placed in the correct direction.

5.

Screw two hexagonal screws to assemble new SMPS Assy unit to the console.

1

2

SMPS ASSY UNIT REMOVAL

ILLUSTRATION 6--57

6--106 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

6--4--2 Replacement Procedure (Continued)

(D) Final Procedure

1.

Connect four cables with the appropriate connectors. For the details of cables, refer to 6--4--2, (B),

Disconnecting Cables.

2.

Re--assemble the removal parts.

3.

Plug the system and power it ON.

4.

Perform all SMPS Assy checks and leakage current test.

5.

Perform functional checks to verify that the system is functioning properly. Refer to Chapter 4, FUNCTIONAL

CHECKS, in this manual.

6.

This completes the replacement of the SMPS Assy Unit.

CAUTION

Before connect the cables, Check and distinguish the labels especially for J12, J19, J3, J6.

6--107 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

6--108 RENEWAL PARTS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CHAPTER 7 -- PERIODIC MAINTENANCE

TABLE OF CONTENTS

TITLE SECTION

7--1

7--2

7--3

PAGE

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--1--1 Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3

7--3

PERIODIC MAINTENANCE PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--2--1 Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--2--2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--2--3 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--2--4 Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3

7--3

7--4

7--4

7--4

ELECTRICAL SAFETY TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--1 Outlet Test Wiring Arrangement

7--3--2 Grounding Continuity

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--3 Chassis Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--4 Probe Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--3--5 When There’s Too Much Leakage Current...

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7--5

7--5

7--6

7--7

7--10

7--14

7--1 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

7--2 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--1 INTRODUCTION

This chapter provides Periodic Maintenance procedures. It is an effective and efficient method of reducing equipment down time and service cost.

7--1--1 Periodic Maintenance

Periodic Maintenance consists of visual inspections, system performance checks, and periodic cleaning of certain critical parts. Perform this procedure at one year intervals. This schedule for maintenance is based on an estimated equipment usage of 40 hours per week.

7--2 PERIODIC MAINTENANCE PROCEDURE

7--2--1 Visual Inspection

1.

Check all cables for cracks nicks and abrasions in their insulation.

2.

Check that cable sheaths are securely clamped in cable ground/connector clamps where applicable.

3.

Check that cables are correctly routed and secured.

4.

Check that there are no missing screws or fasteners and replace if necessary.

5.

Check that all detachable cable connectors are properly seated in their sockets.

6.

Check that all equipment covers and fixtures are in place and undamaged.

7.

Check that all plug--in boards are properly seated in cardcage connector.

8.

Check that the keyboard keys, controls and Trackball function properly and that no mechanical damage is apparent.

7--3 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--2--2 Cleaning

1.

Clean CRT filter.

2.

Clean out the interior of the Console by using a vacuum cleaner.

7--2--3 Measurement

1.

Make sure that electrical outlet of hospital should provide the power line standard.

Refer to Chapter 2, Installation.

2.

Ensure all SMPS voltages are within tolerance.

Refer to section 4--3, SMPS ADJUSTMENT in Chapter 4, FUNCTIONAL CHECKS, for adjustment procedure.

3.

Make sure that leakage currents are within tolerance.

Refer to section 2--3, ELECTRICAL SAFETY TESTS in Chapter 2, INSTALLATION, for measurement procedures.

7--2--4 Note

Visual inspection of probes (“head”, cable and connector)

7--4 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--3 ELECTRICAL SAFETY TESTS

The following tests are performed at the factory and should be performed at the site. These tests are: grounding continuity, chassis leakage current, probe leakage current, and ECG leakage current. All measurements are made with an Electrical Safety Analyzer Model Dale 600/600E built by Dale Technology Corporation or equivalent device.

Recommended Tool

PART NUMBER

46--285652G1

46--328406G2

2113015

NAME DESCRIPTION

Electric Safety Analyzer DALE 600 For 110/115V Units

Electric Safety Analyzer DALE 600E For 220/240V Units

Leakage Current Ultrasound Kit For 100/115V and 220/240V Units

7--3--1 Outlet Test Wiring Arrangement

Test all outlets in the area for proper wiring arrangement by plugging in the Dale 600/600E and noting the combination of lights that are illuminated (Refer to ILLUSTRATION 7--1).

AC outlet

Combination of

Light

Condition CORRECT

WIRING

REVERSE

POLARITY

OPEN

GROUND

OUTLET TEST

ILLUSTRATION 7--1

OPEN

HOT

HOT/GROUND

REVERSED

Note

No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the

Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected.

7--5 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

7--3--2 Grounding Continuity

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CAUTION

Electric Shock Hazard. The patient must not be contacted to the equipment during this test.

Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case

(ILLUSTRATION 7--2). The ground wire resistance should be less than 0.2 ohms.

CONSOLE

GROUND PIN

ACCESSIBLE METAL PARTS

1. Monitor Housing

2. Rear Panel Connector

3. Any Caster/Wheel Support

OHMMETER

GROUND CONTINUITY TEST

ILLUSTRATION 7--2

Meter Procedure

Follow these steps to test the ground wire resistance.

1.

Turn the LOGIQ 200 PRO Series unit OFF.

2.

Plug the unit into the meter, and the meter into the tested AC wall outlet (Refer to ILLUSTRATION 7--2).

3.

Plug the black chassis cable into the meter’s ”CHASSIS” connector and attach the chassis cable clamp to an exposed metal part of the LOGIQ 200 PRO Series unit (Refer to ILLUSTRATION 7--2).

4.

Set the meter’s ”FUNCTION” switch to the RESISTANCE position.

5.

Set the meter’s ”POLARITY” switch to the OFF (center) position.

6.

Measure the ground wire resistance and keep a record of the results with other hard copies of PM data kept on site.

7--6 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--3--3 Chassis Leakage Current Test

Definition

Leakage current is the electrical current that could flow ground wire break. The unit, the probes, the ECG module, and all external peripherals must be tested.

Generic Procedure

The test verifies the isolation of the power line from the chassis. The testing meter is connected from accessible metal parts of the case to ground. Measurements should be made with the unit ON and OFF, with the power line polarity Normal and Reversed. Record the highest reading of current.

POWER

OUTLET

H (BLACK)

N (WHITE)

G (GREEN)

POLARITY REVERSING

SWITCH

MOMENTARY SWITCH

SETUP FOR CHASSIS SOURCE LEAKAGE CURRENT

ILLUSTRATION 7--3

LEAKAGE

TEST

METER

CONSOLE

ACCESSIBLE

METAL

PARTS

7--7 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--3--3 Chassis Leakage Current Test (Continued)

Meter Procedure

Follow these steps to test the unit for leakage current.

1.

Turn the LOGIQ 200 PRO Series unit OFF.

2.

Plug the unit into the meter, and the meter into the tested AC wall outlet (Refer to ILLUSTRATION 7--4).

Outlet Test

Lamps

Functions

Selector

Neutral

Open /closed

Normal

Reverse

Polarity

Power

On/Off

Chassis

Cable

Clamp

AC outlet

Power

Cable

GROUND AND CHASSIS LEAKAGE CURRENT TEST

ILLUSTRATION 7--4

3.

Plug the black chassis cable into the meter’s ”CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the LOGIQ 200 PRO Series unit (Refer to ILLUSTRATION 7--4).

4.

Set the tester’s ”FUNCTION” switch to CHASSIS position.

5.

Follow the test conditions described in the following table for every test point on the unit.

TABLE 7-- 1

CHASSIS LEAKAGE CURRENT TEST CONDITIONS

TEST

1

2

3

4

5

CONDITION

Mounting screw for transducer receptacle

Wheel support

Mounting screw for CRT housing

Mounting screw for peripheral plugged into unit

Mounting screw for other peripheral powered by unit

6.

Keep a record of the results with other hard copies of PM data kept on site.

7--8 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--3--3 Chassis Leakage Current Test (Continued)

Data Sheet for Chassis Source Leakage Current

Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the table below (TABLE 7-- 2).

TABLE 7-- 2

MAXIMUM ALLOWANCE LIMIT FOR CHASSIS SOURCE LEAKAGE CURRENT

COUNTRY

USA

OTHERS

NORMAL

CONDITION

N/A

0.1mA

OPEN

GROUND

0.3mA

0.5mA

REVERSE

POLARITY

0.3mA

0.5mA

OPEN

NEUTRAL

N/A

0.5mA

CAUTION

Electric Shock Hazard. When the meter’s ground switch is OPEN, don’t touch the unit!

Unit

Power

Tester Tester

POLARITY NEUTRAL

Switch

Test1

Probe or GROUND connector

Switch

Enter name of tested peripheral here:

ON

ON

ON

ON

OFF

OFF

OFF

OFF

NORM

NORM

REV

REV

NORM

NORM

REV

REV

TABLE 7-- 3

DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT

Test2 wheel

Test3

CRT

OPEN

CLOSED

OPEN

CLOSED

OPEN

CLOSED

OPEN

CLOSED

Test4 Test5

7--9 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--3--4 Probe Leakage Current Test

Definition

This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.

Generic Procedure

Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst case condition.

POWER

OUTLET

H (BLACK) POLARITY REVERSING SWITCH

N (WHITE)

G (GREEN)

MOMENTARY

SWITCH

CONSOLE

LEAKAGE

TEST

METER

PROBE

SETUP FOR PROBE LEAKAGE CURRENT

ILLUSTRATION 7--5

Note

Each probe will have some amount of leakage current, dependent on its design. Small variations in probe leakage currents are normal from probe to probe. Other variations will result from differences in line voltage and test lead placement.

7--10 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

7--3--4 Probe Leakage Current Test (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Meter Procedure Using Probe Adapter

The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins.

The probe’s imaging area is immersed in a saline solution along with a grounding probe from the meter to complete the current path. Saline solution is a mixture of water and salt. The salt adds a free ion to the water, making it conductive. Normal saline solution is 0.9% salt or 1/2 gram salt per 1 liter of water. If saline is not available, a mixture of 1 quart water with one or more grams of table salt, mixed thoroughly, will substitute.

CAUTION

To avoid probe damage and possible electric shock, do not immerse probes into any liquid beyond the level indicated in the probe users manual. Do not touch the probe, conductive liquid or any part of the unit under test while the ISO TEST switch is depressed.

Follow these steps to test each transducer for leakage current.

1.

Turn the LOGIQ 200 PRO Series unit OFF.

2.

Plug the unit into the test meter, and the meter into the tested AC wall outlet.

3.

Connect the probe for test with the meter’s appropriate adapter (Refer to ILLUSTRATION 7--6).

4.

Plug the saline probe into the meter’s ”CHASSIS” connector (Refer to ILLUSTRATION 7--6).

Ultrasound

Probe

Saline

Probe

Probe Adapter with probe connected

Saline Solution

Nonconductive

Container

AC outlet

TRANSDUCER SOURCE LEAKAGE CURRENT TEST

ILLUSTRATION 7--6

7--11 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--3--4 Probe Leakage Current Test (Continued)

5.

Plug the probe adapter into the meter’s connector marked ”EXTERNAL” (Refer to ILLUSTRATION 7--6 ).

6.

Set the meter’s ”FUNCTION” switch to EXTERNAL position.

7.

Add the saline probe and the imaging area of the probe into the saline bath.

8.

Have unit power ON for the first part; turn it OFF for the second half.

9.

Depress the ISO TEST rocker switch and record the highest current reading.

10. Follow the test conditions described in the following table (TABLE 7-- 5) for every transducer.

11. Keep a record of the results with other hand copies of PM data.

Meter Procedure

Follow these steps to test each transducer for leakage current.

1.

Turn the LOGIQ 200 PRO Series unit OFF.

2.

Plug the unit into the test meter, and the meter into the tested AC wall outlet.

3.

Plug the external probe into the meter’s ”EXTERNAL” connector.

4.

Set the meter’s ”FUNCTION” switch to EXTERNAL position.

5.

Connect the probe for test with the connector of the console.

6.

Apply a liberal amount of gel to the transducer’s lens.

7.

Wrap the transducer in aluminum foil, making sure there is good contact with its acoustic aperture.

8.

Either suspend the transducer by its cable or place it on an insulated surface.

9.

Have unit power ON for the first part; turn it OFF for the second half.

10. Hold the external probe against the aluminum foil on the transducer when the unit is ready to scan.

11. Depress the ISO TEST rocker switch and record the highest current reading.

12. Follow the test conditions described in the following table (TABLE 7-- 5) for every transducer.

13. Keep a record of the results with other hand copies of PM data.

7--12 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

7--3--4 Probe Leakage Current Test (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

Data Sheet for Transducer Source Leakage Current

Follow the foregoing test procedure for every transducer. The test passes when all readings measure less than the value shown in the table below (TABLE 7-- 4).

TABLE 7-- 4

MAXIMUM ALLOWANCE LIMIT FOR PROBE LEAKAGE CURRENT

1.

Type BF Applied Part Leakage Current Limits

[Non--conductive (floating) Surface and Cavity Transducers]

COUNTRY

USA

OTHERS

NORMAL

CONDITION

0.05mA

0.1mA

OPEN

GROUND

0.05mA

0.5mA

REVERSE

POLARITY

0.05mA

0.5mA

OPEN

NEUTRAL

0.05mA

0.5mA

2.

Type CF Applied Part Leakage Current Limits

[Surgical Transducer and ECG Connections]

COUNTRY

USA

OTHERS

NORMAL

CONDITION

0.01mA

0.01mA

OPEN

GROUND

0.01mA

0.05mA

REVERSE

POLARITY

0.01mA

0.05mA

OPEN

NEUTRAL

0.01mA

0.05mA

Transducer tested:

TABLE 7-- 5

DATA SHEET FOR TRANSDUCER SOURCE LEAKAGE CURRENT

LOGIQ 200 PRO Series

Unit Power

ON

Tester Power

Polarity Switch

NORM

ON

ON

ON

NORM

REV

REV

OFF

OFF

OFF

OFF

NORM

NORM

REV

REV

Tester Ground or Neutral Switch

OPEN

CLOSED

OPEN

CLOSED

OPEN

CLOSED

OPEN

CLOSED

Measurement

7--13 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

7--3--5 When There’s Too Much Leakage Current...

Chassis Fails

Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent. Replace any defective part.

Tighten all grounds. Ensure star washers are under all ground studs.

Inspect wiring for bad crimps, poor connections, or damage.

Test the wall outlet; verify it is grounded. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead.

Note

No outlet tester can detect the condition where the white neutral wire and the green grounding wire are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected.

Probe Fails

Test the probe in another transducer interface connector.

If excessive leakage current is slot dependent, inspect the Transducer Interface board for bent pins, poor connections, and ground continuity.

If the problem still has stayed with the probe, replace the probe.

Peripheral Fails

Tighten all grounds. Ensure star washers are under all ground studs.

Inspect wiring for bad crimps, poor connections, or damage.

Still Fails

If all else fails, begin isolation by removing the probes, external peripherals, then the on board ones, one at a time while monitoring the leakage current measurement.

New Unit

If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety

Failure Report to document the system problem. Remove unit from operation.

7--14 PERIODIC MAINTENANCE

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

CHAPTER 8 -- INSTALLATION FOR OPTIONS

TABLE OF CONTENTS

TITLE SECTION

8--1

8--2

PAGE

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--3

MOD ASSY INSTALLATION

8--2--1 INTRODUCTION

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--2 TOOLS REQUIRED

8--2--3 TIME REQUIRED

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--4 PARTS REQUIRED

8--2--5 Functional Check--out

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--6 Disassembling Dummy Bracket Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--7 Installing MOD Top Bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--8 Assembling MOD Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--9 Jumper Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--10 Installing MOD Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--11 Operational Check--out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--12 Final Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--2--13 Renewal Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8--6

8--7

8--8

8--8

8--10

8--11

8--12

8--4

8--4

8--4

8--4

8--4

8--4

8--5

8--1 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

This page is left blank intentionally.

8--2 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

8--1 INTRODUCTION

This chapter provides the installation procedures and renewal parts for optional devices. The installation procedures for the following optional devices are included in this chapter.

MOD Assy : FUJITSU MDD3064AP

Keep the option installation instructions on this chapter after completing installation for other options.

8--3 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

8--2 MOD ASSY INSTALLATION

LOGIQ 200 PRO Series SERVICE MANUAL

8--2--1 INTRODUCTION

These Section describes the installation of the optional MOD FUJITSU MDD3064AP for LOGIQ 200.

2235374

8--2--2 TOOLS REQUIRED

Standard field Service Tool kit

8--2--3 TIME REQUIRED

Approx. 0.5 Hour

8--2--4 PARTS REQUIRED

3.

4.

5

ITEM

1.

2.

PART NAME

MOD ASSY

MOD Top Bracket

MOD Bottom Bracket

SCREW M3X6

SCREW M4X10

1

4

4

QTY

1

1

PART #

2242409

2240642

2240643

2159624

2159632

REMARKS

8--2--5 Functional Check--out

READ and UNDERSTAND these instructions thoroughly before proceeding with the installation. Perform each step in sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer.

CAUTION

Possible operational damage. Failure to strictly follow ESD (Electrostatic Discharge) precautions during this upgrade may cause constant or intermittent operational abnormalities. Strictly follow all precautions.

Check each step as it is completed.

Perform functional checks to verify that the system is functioning properly. Refer to Functional Checks in Chapter

4.

8--4 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

8--2--6 Disassembling Dummy Bracket Assy

1.

Turn the system power OFF and unplug the unit.

2.

Unscrew four screws (1--4) and remove MOD dummy bracket. Refer to ILLUSTRATION 1--1.

Note

Keep removed screws (1--4) because they are required to mount the MOD Top Bracket.

1

MOD Dummy Bracket

2

3

4

MOD DUMMY BRACKET DISASSEMBLY

ILLUSTRATION 1--1

8--5 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

8--2--7 Installing MOD Top Bracket

1.

Screw 4 screws (1--4) and install MOD Top Bracket. Refer to ILLUSTRATION 1--2.

MOD TOP BRACKET ASSEMBLY

ILLUSTRATION 1--2

MOD Top Bracket

8--6 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

8--2--8 Assembling MOD Assy

1.

Assemble MOD Assy and MOD Bottom Bracket with four screws (1--4). Refer to ILLUSTRATION 1--3.

1

4

2

3

MOD ASSEMBLY

ILLUSTRATION 1--3

8--7 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

8--2--9 Jumper Setting

1.

Confirm that Jumper Setting is as illustrated in ILLUSTRATION 1--4.

Rear View

3

4

1

2

Default setting

JUMPER SETTING

ILLUSTRATION 1--4

8--2--10 Installing MOD Assy

1.

Insert MOD Assy with Bottom Bracket in to the MOD Top Bracket and screw 2 screws.

Refer to ILLUSTRATION 1--5.

2.

Connect two connectors (CN1, CN2) from the console to MOD Assy.

8--8 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

8--2--10 Installing MOD Assy (Continued)

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

1

CN1

CN2

1

2

INSTALLING MOD ASSY

ILLUSTRATION 1--5

2

8--9 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

8--2--11 Operational Check--out

1.

Turn the system power ON.

2.

Enter the service software after completing the power--up sequence. For the service software, refer to LO--

GIQ 200 PRO Series Proprietary Manual, 2242594. The Service Utility Menu:Top Menu will be displayed as shown in

ILLUSTRATION 4

. The password is required to enter. For the password, refer to the service note.

Service Utility MENU : Top Menu

Current Software Version is 3.0

99/03/12 09:16:37

System Diagnosis

Board/Program Version

System Reset

Key Operation History

* Option Set

Initialize :1 (1:America 2:Europe 3:Asia 4:Japan)

System Clock Adjust

Software/Probe data upgrade

Exit

SERVICE UTILITY MENU:TOP MENU

ILLUSTRATION 4

3.

Move the cursor to the item “Option Set” and press Set key. Option Set Menu will be displayed.

8--10 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 0

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

8--2--11 Operational Check--out (continued)

4.

Set among the optional items, MOD Driver to the “:Installed”. Refer to ILLUSTRATION 5.

Service UtilitY MENU : Option Set

Current Software Version is 3.0

99/03/12 09:16:37

CINE

ATO

* MOD Driver

Biopsy Guideline

:1 (1:Not Install 2:Install)

:1 (1:Not Install 2:Install)

:2 (1:Not Install 2:Install)

:1 (1:Not Install 2:Install)

MGOB Calculation :1 (1:Not Install 2:Install)

Cardiac Calculation :1 (1:Not Install 2:Install)

Urology Calculation :l (1:Not Install 2:Install)

Serial Printer

Paralled Printer

InSite

Exit

:l (1:Not Install 2:Install)

:1 (1:Not Install 2:Install)

Data Management Center:l (1:Not Install 2:Install)

:1 (1:Not Install 2:Install)

OPTION SET MENU

ILLUSTRATION 5

5.

6.

Note

To activate the options, move the cursor to the item, type 2, and press the Set key.

Exit the Option Set Menu and Power OFF the system.

Turn the system power ON.

8--2--12 Final Procedure

1.

Properly dispose of excess material.

2.

This completes the installation of the LOGIQ 200 PRO Series MOD Option.

8--11 INSTALLATION FOR OPTIONS

GE MEDICAL SYSTEMS

REV 6

8--2--13 Renewal Parts

LOGIQ 200 PRO Series SERVICE MANUAL

2235374

3.

4.

5

ITEM

1.

2.

PART NAME

MOD ASSY

MOD Top Bracket

MOD Bottom Bracket

SCREW M3X6

SCREW M4X10

MATERIAL LIST (1/1)

1

4

4

QTY

1

1

PART #

2295442

2240642

2240643

2299256

2159632

REMARKS

1. MOD ASSY

2. MOD Top Bracket

3. MOD Bottom Bracket

RENEWAL PARTS

ILLUSTRATION 1--6

8--12 INSTALLATION FOR OPTIONS

advertisement

Key Features

  • Versatile ultrasound system
  • High-quality images
  • Advanced technologies
  • User-friendly interface
  • Improved diagnostic accuracy

Frequently Answers and Questions

What is the LOGIQ 200 PRO Series?
The LOGIQ 200 PRO Series is a versatile ultrasound system that is designed for a wide range of applications, including abdominal, cardiac, and vascular imaging.
What are the key features of the LOGIQ 200 PRO Series?
The LOGIQ 200 PRO Series features a variety of advanced technologies that provide high-quality images and improved diagnostic accuracy. It also has a user-friendly interface.
What kind of applications is the LOGIQ 200 PRO Series suitable for?
The LOGIQ 200 PRO Series is suitable for abdominal, cardiac, and vascular imaging.

Related manuals

Download PDF

advertisement