GE Medical Systems LOGIQ 200 PRO Series Ultrasound Scanning System Service Manual
Below you will find brief information for Ultrasound Scanning System LOGIQ 200 PRO Series. The LOGIQ 200 PRO Series is a versatile ultrasound system that is designed for a wide range of applications, including abdominal, cardiac, and vascular imaging. It features a variety of advanced technologies that provide high-quality images and improved diagnostic accuracy. The system is also easy to use, with a user-friendly interface that makes it easy for clinicians to perform exams efficiently.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Technical
Publications
2235374
Revision 6
LOGIQt200 PRO Series
Service Manual
GE Medical Systems
Copyright
E
2000, 2001 By General Electric Co.
Operating Documentation
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, write “damage in shipment” on all copies of the freight or express bill before delivery is accepted or “signed for” by a General Electric representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if the inspection is not requested within this 14 day period.
GE Medical Systems -- Americas
USA: (1) 800--437--1171
Canada: (1) 800--668--0732
Latin & South Am: (1) 305--735--2304
GE Medical Systems -- Asia
Japan:
Korea
(81) 426--48--2594
(82)342--740--6000
GE Ultrasound -- Europe
Germany (49) 212--2802--207
GE MEDICAL SYSTEMS
REV 6
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
LIST OF EFFECTIVE PAGES
REV DATE
0 August 1. 1999
1 February 10. 2000
2
3
May 10. 2000
July 20. 2000
4 November 22. 2000
5 January 24. 2001
6 August 14. 2001
PRIMARY REASON FOR CHANGE
Inital release
System Version 3.10 release
System Version 3.15 release
System Version 3.16 release
System Version 3.16b release
Manufacturer Name Change
System Version 3.17a release
PAGE REV
Title Page . . . . . . . . . .
GE Logo Page . . . . . .
6
0
2184996
2184997
2184998
. . . . . . . . . . .
. . . . . . . . . . .
. . . . . . . . . . .
0
1
0
A . . . . . . . . . . . . . . . . . .
i to vi . . . . . . . . . . . . . .
6
0
Chapter 1
1--1 to 1--7
1--8 to 1--12
. . . . . . . . . .
. . . . . . . . .
1--13 to 1--19
1
5
. . . . . . .
1
PAGE REV PAGE
1--20 . . . . . . . . . . . . . . .
5
REV
Chapter 4
4--1 to 4--12 . . . . . . . . .
0
Chapter 2
2--1 to 2--16 . . . . . . . . .
0
Chapter 3
3--1 to 3--9
3--10
3--15
. . . . . . . . . .
. . . . . . . . . . . . . . .
3--11 to 3--14 . . . . . . . .
. . . . . . . . . . . . . . .
0
4
0
6
3--16 . . . . . . . . . . . . . . .
0
Chapter 5
5--1 to 5--8 . . . . . . . . . .
0
Chapter 6
6--1 to 6--2 . . . . . . . . . .
0
6--3
6--4
. . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . .
2
6
6--5 to 6--6
6--7
. . . . . . . . . .
. . . . . . . . . . . . . . . .
0
6
6--8 . . . . . . . . . . . . . . . .
0
PAGE REV
6--9 . . . . . . . . . . . . . . . .
6--10 . . . . . . . . . . . . . . .
6--11 . . . . . . . . . . . . . . .
6
0
6
6--12 . . . . . . . . . . . . . . .
6--13 to 6--14 . . . . . . .
0
3
6--15 to 6--16 . . . . . . .
6--17 . . . . . . . . . . . . . . .
0
3
6--18 to 6--19 . . . . . . .
6--20 . . . . . . . . . . . . . . .
6
3
6--21 to 6--44
6--45
. . . . . . .
0
. . . . . . . . . . . . . . .
6--46 to 6--100 . . . . . .
2
0
6--101 . . . . . . . . . . . . . .
6
PAGE REV
6--102 . . . . . . . . . . . . . .
6--103 . . . . . . . . . . . . . .
6--104 to 6--108 . . . . .
0
6
0
Chapter 7
7--1 to 7--14 . . . . . . . . .
0
Chapter 8
8--1 to 8--3 . . . . . . . . . .
8--4 . . . . . . . . . . . . . . . .
8--5 to 8--12 . . . . . . . . .
0
0
0
A
GE MEDICAL SYSTEMS
REV 0
SECTION
1--1
1--2
1--3
1--4
2--1
2--2
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
TABLE OF CONTENTS
TITLE PAGE
CHAPTER 1 -- INTRODUCTION
SERVICE MANUAL CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3
SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--2--1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--2--2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--4
1--4
1--17
EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--1 EMC Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--2 Notice upon Installation of Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--3 General Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--4 Countermeasures against EMC--related Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--5 Notice on Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--18
1--18
1--18
1--19
1--19
1--19
ADDRESS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--20
CHAPTER 2 -- INSTALLATION
PREINSTALLATION
2--1--1 Introduction
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--1--2 Power Line Requirements
2--1--3 Physical Specifications
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--1--4 Recommended Ultrasound Room Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--3
2--3
2--3
2--4
2--5
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--2 Average Installation Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--3 Installation Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--4 Checking the Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--5 Unpacking LOGIQ 200 PRO Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--6 MTZ Probe Holder Installation (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--7 Transducer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--8 Moving into Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--9 Adjusting System Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--10 Product Locator Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--8
2--8
2--8
2--8
2--8
2--9
2--11
2--14
2--14
2--14
2--15 i
TABLE OF CONTENTS
GE MEDICAL SYSTEMS
REV 0
SECTION
3--1
3--2
3--3
3--4
3--5
3--6
TITLE
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
PAGE
CHAPTER 3 -- SYSTEM CONFIGURATION
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--3
DIMENSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--3
ELECTRICAL SPECIFICATIONS
3--3--1 Power Supply
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--3--2 Facility Power Receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--4
3--4
3--4
STORAGE AND OPERATION REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--4
OPTIONAL PERIPHERALS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--5--1 Peripherals/Accessories Connector Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--5--2 List of Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--5--3 Power Consumption of Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--5
3--5
3--9
3--11
TEST POINT, LED, DIP SWITCH AND RESET SWITCH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--6--1 Test Point List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--6--2 LED List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--6--3 DIP Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--6--4 Reset Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--6--5 Video Signal Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--12
3--12
3--14
3--15
3--15
3--16 ii
TABLE OF CONTENTS
GE MEDICAL SYSTEMS
REV 0
SECTION
4--1
4--2
4--3
5--1
5--2
5--3
5--4
5--5
TITLE
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
PAGE
CHAPTER 4 -- FUNCTIONAL CHECKS
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--1--1 Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--3
4--3
FUNCTIONAL CHECK PROCEDURES
4--2--1 Basic Controls
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--2--2 M--Mode Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--3
4--3
4--5
SMPS ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--3--1 SMPS Assy Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--3--2 SMPS Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--7
4--7
4--10
CHAPTER 5 -- DIAGRAMS
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--3
LOGIQ 200 PRO Series SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--3
BLOCK DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--4
WIRING DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--5
CIRCUIT BOARD DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--7 iii
TABLE OF CONTENTS
GE MEDICAL SYSTEMS
REV 0
SECTION TITLE
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
PAGE
6--1
6--2
CHAPTER 6 -- RENEWAL PARTS
RENEWAL PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--3
DISASSEMBLY/RE--ASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--1 Monitor Assy (FRU No. 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--2 Pot Set Assy (FRU No. 101)
6--2--3 CRT Filter(FRU No. 102)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--4 Monitor Cover Set (FRU No. 103)
6--2--5 Tilt Assy(FRU No. 104)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--6 Filter Clamp Set (FRU No. 105) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--7 Probe Holder (FRU No. 201) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--8 Holder Bracket Assy (FRU No. 202) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--9 Cover Set (FRU No. 203) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--10 Keyboard Assy (FRU No. 204) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--11 Trackball Assy (FRU No. 205) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--12 TGC Assy (FRU No. 206) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--13 Encoder Set (FRU No. 207) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--14 Freeze/Record Key Assy (FRU No. 208)
6--2--15 TGC Knob Set (FRU No. 209)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--16 Rubber Key Set (FRU No. 210) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--17 P.C.Board(s) (FRU No.301 through 304)
6--2--18 BPHV Assy (FRU No. 305)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--19 SCAN Assy (FRU No. 306) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--20 Shield Panel (FRU No. 307) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--21 Mother 3 Assy (FRU No. 308) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--22 Nest Box (FRU No. 309) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--23 DC Fan Assy (FRU No. 310) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--24 Swing Arm Assy (FRU No. 401) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--25 Pipe Assy (FRU No. 402) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--26 Pole Cover (FRU No. 403) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--27 Left Cover (FRU No. 404) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--28 Right Cover (FRU No. 405) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--29 Front Base Cover (FRU No. 406) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--30 Front Cover (FRU No. 407) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--31 Rear Cover (FRU No. 408) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--32 Top Cover (FRU No. 409) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--33 Printer Cover (FRU No. 410) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--34 Printer Bracket Assy(FRU No. 411)
6--2--35 Rear Handle (FRU No. 412)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--36 Neck Frame (FRU No. 413) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--37 Rear Panel Assy (FRU No. 414)
6--2--38 Power S/W Assy (FRU No. 415)
6--2--39 Front Caster (FRU No. 416)
6--2--40 Rear Caster (FRU No. 417)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--64
6--65
6--66
6--68
6--70
6--72
6--74
6--76
6--78
6--80
6--44
6--46
6--48
6--50
6--52
6--54
6--56
6--58
6--60
6--62
6--82
6--84
6--86
6--87
6--29
6--30
6--32
6--34
6--36
6--37
6--38
6--39
6--40
6--42
6--21
6--22
6--24
6--25
6--26
6--27
6--28 iv
TABLE OF CONTENTS
GE MEDICAL SYSTEMS
REV 0
SECTION
6--3
6--4
7--1
7--2
7--3
TITLE
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
PAGE
CHAPTER 6 -- RENEWAL PARTS (continued)
6--2--41 Bumper Set (FRU No. 418) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--42 EMI Cover L (FRU No. 419) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--43 PCB Guide Assy (FRU No. 420) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--44 Terminal Block 12 Assy (FRU No. 501) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--45 Terminal Block 6 Assy (FRU No. 502) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--46 SMPS Assy (FRU No. 503) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--47 Power Trans Assy (FRU No. 504)
6--2--48 Circuit Breaker Set (FRU No. 505)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--88
6--89
6--90
6--92
6--94
6--96
6--98
6--100
FUSE REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--3--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--3--2 Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--101
6--101
6--101
SMPS Assy REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--4--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--4--2 Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--104
6--104
6--104
CHAPTER 7 -- PERIODIC MAINTENANCE
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--1--1 Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3
7--3
PERIODIC MAINTENANCE PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--2--1 Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--2--2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--2--3 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--2--4 Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3
7--3
7--4
7--4
7--4
ELECTRICAL SAFETY TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--1 Outlet Test Wiring Arrangement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--2 Grounding Continuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--3 Chassis Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--4 Probe Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--5 When There’s Too Much Leakage Current...
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--5
7--5
7--6
7--7
7--10
7--14 v
TABLE OF CONTENTS
GE MEDICAL SYSTEMS
REV 0
SECTION
8--1
8--2
TITLE
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
PAGE
CHAPTER 8 -- INSTALLATION FOR OPTIONS
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--3
MOD ASSY INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--2 TOOLS REQUIRED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--3 TIME REQUIRED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--4 PARTS REQUIRED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--5 Functional Check--out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--6 Disassembling Dummy Bracket Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--7 Installing MOD Top Bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--8 Assembling MOD Assy
8--2--9 Jumper Setting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--10 Installing MOD Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--11 Operational Check--out
8--2--12 Final Procedure
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--13 Renewal Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--5
8--6
8--7
8--8
8--8
8--10
8--11
8--12
8--4
8--4
8--4
8--4
8--4
8--4 vi
TABLE OF CONTENTS
GE MEDICAL SYSTEMS
REV 1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CHAPTER 1 -- INTRODUCTION
TABLE OF CONTENTS
TITLE SECTION
1--1
1--2
1--3
1--4
PAGE
SERVICE MANUAL CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3
SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--2--1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--2--2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--4
1--4
1--17
EMC (Electromagnetic Compatibility) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--1 EMC Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--2 Notice upon Installation of Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--3 General Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--3--4 Countermeasures against EMC--related Issues
1--3--5 Notice on Service
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--18
1--18
1--18
1--19
1--19
1--19
ADDRESS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--20
1--1
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
1--2
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1--1 SERVICE MANUAL CONTENTS
This manual provides service information on the LOGIQ 200 PRO Series Ultrasound Scanning System. It contains the following chapters:
1.
Chapter 1, Introduction: Contains a content summary and warnings.
2.
Chapter 2, Installation: Contains physical and electrical requirements that must be considered prior to installation and a complete LOGIQ 200 PRO Series installation procedure with installation checklist.
3.
Chapter 3, System Configuration: Contains system configuration and specifications.
4.
Chapter 4, Functional Checks: Contains functional checks that must be performed as part of the installation, or as required during servicing and periodic maintenance.
5.
Chapter 5, Diagrams: Contains block diagrams and functional explanations of the LOGIQ 200 PRO Series electronics.
6.
Chapter 6, Renewal Parts: Contains a complete list of replacement parts for the LOGIQ 200 PRO Series and disassembly procedures for all changeable FRU.
7.
Chapter 7, Periodic Maintenance: Provides periodic maintenance procedures for the LOGIQ 200 PRO
Series.
8.
Chapter 8, Installation for Options: is provided to keep the option installation instructions supplied with each option.
1--3
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
1--2 SAFETY
1--2--1 Warnings
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
WARNING!
CAREFULLY READ ALL THE WARNINGS LISTED BELOW!
1.
The operator manual should be fully read and understood before operating the LOGIQ 200 PRO Series and kept nearby for quick reference.
2.
Although the ultrasound energy transmitted from the LOGIQ 200 PRO Series transducer is within
AIUM/NEMA standards, unnecessary exposure should be avoided. Only trained personnel should operate the
LOGIQ 200 PRO Series.
3.
To prevent electrical shock, the LOGIQ 200 PRO Series should be connected to a properly grounded power receptacle. Do not use a three to two prong adapter. This defeats safety grounding.
4.
Do not use with Defibrillator when LOGIQ 200 PRO Series is being operated .
5.
Probes are fragile, please handle with care.
6.
Concerning Outside Markings, refer to Illustration 1--1, 1--2, 1--3, 1--4, 1--5, 1--6, 1--7, 1--8, and 1--9.
7.
For the cleaning, disinfection, and sterilization, refer to Probe section in LOGIQ 200 PRO Series Advanced
Reference Manual (2233222--100) and Caution Sheet supplied with each probe.
NOTICE
This medical equipment is approved, in terms of the prevention of radio wave interference, to be used in hospitals, clinics and other institutions which are environmentally qualified. The use of this equipment in an inappropriate environment may cause some electronic interference to radios and televisions around the equipment. Proper handling of this equipment is required in order to avoid such trouble according to the operator and service manuals.
This equipment can be used in residential areas only under the supervision of physicians or qualified technicians.
1--4
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR ALL UNITS)
ILLUSTRATION 1--1
1--5
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
OUTSIDE MARKINGS OFLOGIQ 200 PRO Series (FOR ALL UNITS)
ILLUSTRATION 1--2
Note
For further details regarding the cautions above, refer to NO TAG MOVING INTO POSITION in
Chapter 2.
1--6
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Defibrillator
Caution
Label
OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR ALL UNITS)
ILLUSTRATION 1--3
1--7
INTRODUCTION
GE MEDICAL SYSTEMS
REV 5
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
MODEL NUMBER : 2270968
Class1/Classe1
MANUFACTURED
LOCATION:
GE MEDICAL SYSTEMS KOREA
S.N.
VOLT
:
: 220--240Vac ~
AMP. LONG TERM : 2.1A
KVA : 0.5kVA
PHASE
DESC.
: 1
: LOGIQ 200PRO Series
CISPR 11 / EN 55011
CLASS : A GROUP : 1
CLASSE : A GROUPE : 1
OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR EUROPE )
ILLUSTRATION 1--4
Note
For the symbols shown in the illustration above, refer to latter pages in this chapter.
1--8
INTRODUCTION
GE MEDICAL SYSTEMS
REV 5
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
For USA
Canada,
Mexico
MADE FOR GE MEDICAL SYSTEMS
MILWAKEE, WISCONSIN BY
GE MEDICAL SYSTEMS KOREA
CLASS 1
MODEL 2270969
SERIAL.
MANUFACTURED
VOLTS : 120Vac ~ 1 PHASE
POWER
FREQUENCY
DESC,
: 500VA
: 60HZ
: LOGIQ 200PRO Series
For Brazil
MADE FOR GE MEDICAL SYSTEMS
MILWAKEE, WISCONSIN BY
GE MEDICAL SYSTEMS KOREA
MODEL
SERIAL.
CLASS 1
2270970
MANUFACTURED
VOLTS : 220--240Vac ~ 1 PHASE
POWER : 500VA
FREQUENCY
DESC,
: 50HZ
: LOGIQ 200PRO Series
OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR USA CANADA MAXICO BRAZIL)
ILLUSTRATION 1--5
1--9
INTRODUCTION
GE MEDICAL SYSTEMS
REV 5
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
MADE FOR GE MEDICAL SYSTEMS
MILWAKEE, WISCONSIN BY
GE MEDICAL SYSTEMS KOREA
MODEL
SERIAL.
MANUFACTURED
CLASS 1
2270967
VOLTS
POWER
:220--240Vac ~ 1 PHASE
: 500VA
FREQUENCY
DESC,
: 50HZ
: LOGIQ 200PRO Series
For SEA, China, Hongkong, India, Australia, New Zealand
MADE FOR GE MEDICAL SYSTEMS
MILWAKEE, WISCONSIN BY
GE MEDICAL SYSTEMS KOREA
CLASS 1
MODEL 2270966
SERIAL.
MANUFACTURED
VOLTS : 120Vac ~ 1 PHASE
POWER
FREQUENCY
DESC,
: 500VA
: 60HZ
: LOGIQ 200PRO Series
For Taiwan
OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR SEA CHINA HONGKONG INDIA AUSTRALLIA NEWZEALAND TAIWAN
PHILLIPPINES)
ILLUSTRATION 1--6
1--10
INTRODUCTION
GE MEDICAL SYSTEMS
REV 5
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
MADE FOR GE MEDICAL SYSTEMS
MILWAKEE, WISCONSIN BY
GE MEDICAL SYSTEMS KOREA
CLASS 1
MODEL 2270971
SERIAL.
MANUFACTURED
:
VOLTS : 220--240Vac~ 1 PHASE
POWER
FREQUENCY
DESC
: 500VA
: 60HZ
: LOGIQ 200PRO Series
OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR CHILE)
ILLUSTRATION 1--7
1--11
INTRODUCTION
GE MEDICAL SYSTEMS
REV 5
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR KOREA)
ILLUSTRATION 1--8
1--12
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Do not use a Defibrillator simultaneously with the ECG, as its excessive voltage will damage the signal input block of the ECG unit.
OUTSIDE MARKINGS OF LOGIQ 200 PRO Series (FOR UNITS WITH ECG)
ILLUSTRATION 1--9
1--13
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
1--2--1 Warnings (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
NOTE: Two labels are attached on rear of the SMPS assy box inside the rear cover and front of BPHV Board Assy inside the front base cover.
MARKINGS OF LOGIQ 200PRO Series (INSIDE COVERS)
ILLUSTRATION 1--10
1--14
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1--2--1 Warnings (Continued)
The following table describes the purpose and location of safety labels and other important information provided on the equipment.
Label/Icon
Identification and Rating Plate
Type/Class Label
IP Code
(IPX1)
Purpose/Meaning
Manufacturer’s name and address
Date of manufacture
Model and serial numbers
Electrical ratings (Volts, Amps, phase, and frequency)
Used to indicate the degree of safety or protection.
Location
Rear of console near power inlet
Indicates the degree of protection provided by the enclosure per IEC 529. IPX1 indicates drip proof.
Foot Switch
Equipment Type BF (man in the box symbol)
IEC 878-02-03 indicates B Type equipment having a floating applied part.
Probe connectors
Equipment Type CF (heart in the box symbol) IEC 878-02-05 indicate equipment having a floating applied part having a degree of protection suitable for direct cardiac contact.
ECG connector and surgical probes
Device Listing/
Certification Labels
Laboratory logo or labels denoting conformance with industry safety standards such as UL or IEC.
Rear of console
“DANGER -- Risk of explosion used in...”
The system is not designed for use with flammable anesthetic gases.
Rear of console
“CAUTION” The equilateral triangle is usually used in combination with other symbols to advise or warn the user.
Various
“ATTENTION -- Consult accompanying documents” is intended to alert the user to refer to the operator manual or other instructions when complete information cannot be provided on the label.
Various
1--15
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
1--2--1 Warnings (Continued)
Label/Icon
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Purpose/Meaning
“CAUTION -- Dangerous voltage” (the lightning flash with arrowhead in equilateral triangle) is used to indicate electric shock hazards.
Various
Location
“Mains OFF” Indicates the power off position of the mains power switch.
Front of system,
Main power switch
“Mains ON” Indicates the power on position of the mains power switch.
Front of system,
Main power switch
“Equipotentiality” Indicates the terminal to be used for connecting equipotential conductors when interconnecting (grounding) with other equipment.
CAUTION
This is only for ”FUNCTIONAL
GROUNDING”, NOT ”PROTECTIVE
EARTH”.
Rear of console
Indicates Main protective earth terminal
Various
1--16
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1--2--2 Specifications
Type of protection against electric shock: Class I EQUIPMENT (*1)
Degree of protection against electric shock: Type BF EQUIPMENT (*2) (Except ECG)
Type CF EQUIPMENT (*3) (ECG Only)
Ordinary Equipment
Continuous Operation
*1. Class I EQUIPMENT
EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but which includes an additional safety precaution in that means are provided for the connection of the
EQUIPMENT to the protective earth conductor in the fixed wiring of the installation in such a way that
ACCESSIBLE METAL PARTS cannot become LIVE in the event of a failure of the BASIC INSULATION.
*2. Type BF EQUIPMENT
TYPE B EQUIPMENT with an F--TYPE APPLIED PART
TYPE B EQUIPMENT: EQUIPMENT providing a particular degree of protection against electric shock, particularly regarding:
-- allowable LEAKAGE CURRENT;
Normal mode
Patient leakage current Less than 100
µ
A
Single failure mode
Less than 500
µ
A
*3. Type CF EQUIPMENT
EQUIPMENT providing a particular degree of protection higher than that for TYPE OF BF EQUIPMENT against electronic shock particularly regarding allowable LEAKAGE CURRENT, and having an F--TYPE
APPLIED PART.
-- allowable LEAKAGE CURRENT;
Normal mode
Patient leakage current Less than 10
µ
A
Single failure mode
Less than 50
µ
A
1--17
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1--3 EMC (Electromagnetic Compatibility)
1--3--1 EMC Performance
All types of electronic equipment may characteristically cause electromagnetic interference with other equipment, either through air or connecting cables. The term EMC (Electromagnetic Compatibility) indicates capability of the equipment, which curbs electromagnetic influence from other equipment and at the same time does not affect other equipment with similar electromagnetic radiation from itself.
This product is designated to fully comply with EN60601--1--2 (IEC 601--1--2), In Medical electronic equipment
EMC regulations.
Proper installation following this service manual is required in order to achieve the full EMC performance of the product.
The product must be installed as stipulated in 1--3--2, Notice upon Installation of Product.
In case of issues related to EMC, please follow procedures stated in 1--3--4, Countermeasures against
EMC--related Issues.
1--3--2 Notice upon Installation of Product
1) Use either power supply cords provided or designated by GEMS or SGMS. Products equipped with power source plug should be plugged into the fixed power socket which has the protective grounding conductor.
Connect a three--pole plug to a three--pole socket without using a three--pole--to--two--pole converter.
2) Locate the equipment as far as possible from other electronic equipment.
3) Be sure to use cables provided by GEMS and SGMS. Wire these cables following these installation procedures.
(Example) Wire power cables separately from signal cables.
4) Lay out the main equipment and other peripherals following the installation procedures described in
Chapter2, INSTALLATION.
1--18
INTRODUCTION
GE MEDICAL SYSTEMS
REV 1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1--3--3 General Notice
1) Designation of Peripheral Equipment Connectable to This Product
Equipment which conforms to EN60601--1--2 (IEC601--1--2), can be hooked up to the product without compromising its EMC performance.
Avoid using other equipment. Failure to comply with this instruction may result in poor EMC performance of the product.
2) Notice against User Modification
Never modify this product. Unilateral user modification may cause degradation in EMC performance.
Modification of the product includes: a) Changes in cables (length, material, wiring etc.) b) Changes in system installation/layout c) Changes in system configuration/components d) Changes in means of fixing system/parts (cover open/close, screwing cover)
3) Operate the system with all covers closed. If you open any cover for some reason, be sure to close it before starting/resuming operation.
Operating the system with any cover open may affect EMC performance.
1--3--4 Countermeasures against EMC--related Issues
Generally it is very difficult to handle issues related to EMC. It may be time consuming and costly.
General countermeasures
Electromagnetic interference with other equipment
1) Electromagnetic interference may be alleviated by positioning other equipment far from the system.
2) Electromagnetic interference may be mitigated by changing the relative location (installation angle) between the system and other equipment.
3) Electromagnetic interference may be eased by changing wiring locations of power/signal cables of other equipment.
4) Electromagnetic influence may be reduced by altering the path of power supply for other equipment.
1--3--5 Notice on Service
1) Ensure all screws are tight after servicing. Loose screws may cause degradation in EMC performance.
2) In case the high frequency gasket of this system is broken, replace it with a new one immediately.
1--19
INTRODUCTION
GE MEDICAL SYSTEMS
REV 5
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1--4 ADDRESS
This system is not repairable by the customer. If this equipment does not work as indicated in the Operator
Manual, please contact your service support center. If the service engineer needs additional information to repair this equipment, please contact the following address (The necessary information will be provided to the Service
Engineer as needed):
GE Medical Systems
Ultrasound Business Group
4855 W. Electric Ave., Milwaukee, WI 53219
USA
TEL: (1) 800--437--1171
FAX: (1) 414--647--4090
CANADA
TEL: (1) 800--668--0732
LATIN & SOUTH AMERICA
TEL: (1) 305--735--2304
GE Ultrasound Europe
GE Ultraschall GmbH & Co.KG
Beethovenstra β e 239 Postfach 11 05 60
D--42655 Solingen GERMANY
TEL: 0130--81--6370 (OLC--Europe Toll Free Number)
(49) 212--2802--207 (English/German Hotline)
(49) 212--2802--208 (English/German/French Hotline)
FAX: (49) 212--2802--431
GE Yokogawa Medical Systems
On--Line Center (OLC), Asia
Ultrasound Group
67--4, Takakura--cho, Hachioji--shi, Tokyo, 192--0033
JAPAN
TEL: (81) 426--48--2940
FAX: (81) 426--48--2950
GE MEDICAL SYSTEMS KOREA
65--1, Sangdaewon--dong, Chungwon--Ku,
Sungnam--Si, Kyunggi--do,
KOREA
TEL: (82) 31--740--6000
FAX: (82) 31--740--6436
1--20
INTRODUCTION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CHAPTER 2 -- INSTALLATION
TABLE OF CONTENTS
TITLE SECTION
2--1
2--2
PAGE
PREINSTALLATION
2--1--1 Introduction
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--1--2 Power Line Requirements
2--1--3 Physical Specifications
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--1--4 Recommended Ultrasound Room Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--3
2--3
2--3
2--4
2--5
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--2 Average Installation Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--3 Installation Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--4 Checking the Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--5 Unpacking LOGIQ 200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--6 MTZ Probe Holder Installation (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--7 Transducer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--8 Moving into Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--9 Adjusting System Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2--10 Product Locator Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--8
2--8
2--8
2--8
2--8
2--9
2--11
2--14
2--14
2--14
2--15
2--1
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
2--2
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2--1 PREINSTALLATION
2--1--1 Introduction
This section describes various general electrical, operational, and environmental considerations that must be considered before installing the LOGIQ 200 PRO Series Ultrasound unit.
2--1--2 Power Line Requirements
The following power line parameters should be monitored for one week before installation. We recommend that you use an analyzer Dranetz Model 606--3 or Dranetz Model 626:
PARAMETER
Voltage Range
Voltage Range
Line Frequency
Power Transients
:
: Korea
: USA
: Europe, LA
: Japan.
: Asia
LIMITS
: 220 VAC
±
10% (198 -- 242 VAC)
: 115 VAC
±
10% (103 -- 127 VAC)
: 220 -- 240 VAC ¦ 10%(198 --264 VAC)
: 100 VAC
±
10% (90 -- 110 VAC)
: 115 VAC
±
10% (103 -- 127 VAC)
: 220 -- 240 VAC
¦
10%(198 --264 VAC)
: All applications : MAX. 500 VA
: All applications : 50/60Hz (
±
2Hz)
: Less than 25 % of nominal peak voltage for less than 1 millisecond for any type of transient, including line frequency, synchronous, asynchronous, or aperiodic transients.
Decaying Oscillation : Less than 15 % of peak voltage for less than 1 millisecond.
2--3
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2--1--3 Physical Specifications
The LOGIQ 200 PRO Series (excluding accessories) weighs 76 Kg (168 lbs). See Chapter 3, ILLUSTRATION
3--1 for dimensions.
Operating Conditions
The LOGIQ 200 PRO Series is designed to operate within a temperature range of 10
_
C to 40
_
C (50
_
F to 104
_ F), and a relative humidity range of 5 % to 90 % (Non--condensing).
Patient Comfort
Concerning permissible operating temperature and humidity tolerances, we recommend that ambient room temperature should be maintained between 20 _ C to 26 _ C (68 _ F to 79 _ F), Humidity should be maintained between 50 % and 70 % for patient comfort during ultrasound scanning.
Electromagnetic Interference (EMI)
Ultrasound machines are susceptible to interference from the radio frequencies, magnetic fields, and transients in the air or power leads. Possible EMI sources should be identified. Electrical and electronic equipment may produce EMI unintentionally or as a result of a malfunction. These sources include medical lasers, cauterizing guns, computers, monitors, fans, gel warmers, microwave ovens, and cellular phones. The presence of broadcast station or van may also cause interference.
Carefully read the following precautions before installing the unit.
1.
Connect the power plug for any other equipment into the fixed outlet with ground wire.
2.
Securely connect any equipment with permanent ground connection to the earth ground furnished in the building.
3.
Install the unit as far from any electrical or electronic equipment as possible.
If any EMI troubles are known or suspected to be present, try to deal with the equipment suspected to have influence on the Ultrasound machine as follows:
1.
Move the ultrasound machine as far from the equipment as possible.
2.
Change the arrangement of the equipment in the room.
3.
Plug the equipment into other outlet.
4.
Move the power cable or signal cable connected with the equipment.
Securely re--tighten drive any screws for the Ultrasound machine after re--assembling for service operation.
2--4
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
2--1--4 Recommended Ultrasound Room Layout
TABLE 2-- 1 provides the requirements for an ultrasound room:
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
POWER SOURCE
CURRENT RATING
RADIATION SHIELDING
TEMPERATURE
HUMIDITY
HEAT DISSIPATION
FLOOR LOADING
FLOOR CONDITION
LOGIQ 200 PRO Series Weight
TABLE 2-- 1
ULTRASOUND ROOM REQUIREMENTS
230VAC, 50Hz, SINGLE PHASE
115V, 60Hz, SINGLE PHASE
For Europe Version
For USA Version
10A (115V, 100V) ; 5A (220--240V) CIRCUIT BREAKER
NONE REQUIRED for ULTRASOUND ENERGY
20--26
_
C (68--79
_
F) for PATIENT COMFORT
50% to 70% for PATIENT COMFORT
2000 BTU/Hr for LOGIQ 200 PRO SeriesMD
Approximately 240 -- 300 kg/m
2 without Accessories
Gradient : WITHIN 5 degrees
76 kg (168lbs) without Accessories
2--5
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
2--1--4 Recommended Ultrasound Room Layout (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Optional Desirable Ultrasound Room Facilities
These facilities are:
1.
Lab sink with hot and cold water.
2.
Emergency oxygen supply.
3.
Dimmer control for overhead lights.
4.
Film viewer.
5.
Film and linen storage.
6.
Medical equipment storage.
7.
Hospital grade equipment electrical outlet.
8.
Document storage area for operating and service manuals.
9.
Nearby waiting room, dressing room, lavatory.
10. Trash bin.
Recommended and Alternate Ultrasound Console Floor Plans
ILLUSTRATION 2--1 provides a recommended standard floor plan and a minimal floor plan for ultrasound equipment
2--6
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
2--1--4 Recommended Ultrasound Room Layout (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Dedicated power outlet
GE cabinet for software and manuals
Film
Processing
Room
Film Viewer
Counter
Top
LOGIQ
200
Stool
Suction
Line
Emergency
Oxygen
61 cm
Desk
Patient
Toilet
Facility
Scale: each square equals one square foot
Film Viewer
Film Supplies
Sink
107 cm
42 in
A 14 by 17 Recommended Floor Plan
Linens
Probes/supplies
Stool
LOGIQ
200
Examination Table
193 cm
76 cm
30 in
An 8 by 10 Minimal Floor Plan
Dedicated power outlet
GE cabinet for software and manuals
RECOMMENDED ULTRASOUND FLOOR PLAN
ILLUSTRATION 2--1
2--7
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2--2 INSTALLATION
2--2--1 Introduction
This section contains many of the procedures required to install the LOGIQ 200 PRO Series console.
2--2--2 Average Installation Time
The LOGIQ 200 PRO Series has been designed to be installed and checked out by an experienced service technician in approximately three hours. LOGIQ 200 PRO Series consoles with optional equipment may take slightly longer.
2--2--3 Installation Warnings
1.
Since the LOGIQ 200 PRO Series weighs approximately 76 kg (168 lbs) without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items.
2.
There are no operator serviceable components. To prevent shock, do not remove any covers or panels.
Should problems or malfunctions occur, unplug the power cord. Only qualified service personnel should carry out servicing and troubleshooting.
Note
For information regarding packing labels, refer to ILLUSTRATION 2--3, LABELS ON PACKAGE.
3.
After being transported, the unit may be very cold or hot. If this is the case, allow the unit to acclimate before you turn it on. It requires one hour for each 2.5
°
C increment if it’s temperature is below 10
°
C or above 40
°
C.
CAUTION
Equipment damage possibility. Turning the system on without acclimation after arriving at site may cause the system to be damaged.
TABLE 2-- 2
TIME FOR SETTLEMENT
°
C 60 55 50 45 40 35 30 25 20 15 10 5 0 --5 --10 --15 --20 --25 --30 --35 --40
°
F 140 131 122 113 104 95 86 77 68 59 50 41 32 23 14 5 --4 --13 --22 --31 --40 hrs 8 6 4 2 0 0 0 0 0 0 0 2 4 6 8 10 12 14 16 18 20
2--2--4 Checking the Components
When a new system arrives, check that any components are not damaged and or missing. If shipping damage or shortages occur, contact the address shown in Chapter 1.
2--8
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
2--2--5 Unpacking LOGIQ 200 PRO Series
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Do not lift the unit by the Keyboard or Monitor arm. Equipment damage may result.
CAUTION
The unit weighs approximately 76 kg (168 lbs). Be prepared for a sudden shift of weight as the unit is removed from its base (pallet).
Refer to ILLUSTRATION 2--2 while performing the following procedure.
1.
Cut the two BANDs.
2.
Lift the CAP up and off.
3.
Lift the TOP PAD up and off.
4.
Remove the OPTION BOX.
5.
Remove the VINYL COVER.
6.
Remove the MIDDLE PLATE ASSY.
7.
Remove the MONITOR COVER.
8.
Lift the PACKING CASE up and off.
9.
Lift the Monitor up by pressing the <UP/DOWN> button located on the Monitor Arm.
10. Remove the MONITOR PAD.
11. Return the Monitor arm to its lowest position.
12. Carefully roll the LOGIQ 200 PRO Series from the PALLET.
13. Remove the Caution Label attached on the CRT Filter and clean the CRT Filter.
Note
Check the shipping container for special instructions. Verify that the container is intact. In some cases a secondary container may be used. If so, ask the carrier for unpacking instructions.
2--9
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
2--2--5 Unpacking LOGIQ 200 PRO Series (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
MIDDLE PLATE
CAP
TOP PAD
PACKING
PLATE
OPTION BOX
PALLET
BAND
UNPACKING LOGIQ 200 PRO Series
ILLUSTRATION 2--2
MONITOR PAD
2--10
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
2--2--5 Unpacking LOGIQ 200 PRO Series (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
60
d
C
Air Pressure
700 - -1060hPa
Humidity 30- -95%
EXcluding Condensation
- -10
d
C
LABELS ON PACKAGE
ILLUSTRATION 2--3
2--2--6 MTZ Probe Holder Installation (Option)
One MTZ probe holder and two brackets are supplied with the LOGIQ 200 PRO Series console as shown in
ILLUSTRATION 2--4.
Note
Two sets of screw holes are provided at the bottom of standard probe holder and left side of keyboard to install the MTZ probe holder. You can choose the most convenient position for your customer between the two sets of screw holes.
Bracket1 (RIGHT)
MTZ Probe Holder
Bracket 2 (LEFT)
MTZ PROBE HOLDER SET
ILLUSTRATION 2--4
2--11
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2--2--6 MTZ Probe Holder Installation (Option) (Continued)
1.
Assemble the MTZ probe holder at the bottom of standard probe holder by screwing four (1 -- 4) screws as shown in ILLUSTRATION 2--5.
MTZ PROBE HOLDER INSTALLATION (1)
ILLUSTRATION 2--5
1
2
3
4
2--12
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2--2--6 MTZ Probe Holder Installation (Option) (Continued)
2.
After removing the Keyboard Assy (Refer to NO TAG, 6--2 DISASSEMBLY/RE--ASSEMBLY), Assemble the
MTZ probe holder at the left side of keyboard by screwing four (1 -- 4) screws as shown in
ILLUSTRATION 2--6.
1
2
3
4
MTZ PROBE HOLDER INSTALLATION (2)
ILLUSTRATION 2--6
2--13
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2--2--7 Transducer Connection
1.
Connect a transducer to the left side connector slot #1, transducer receptacle as follows: a. Ensure that the transducer twist lock lever points towards the 12 o’clock position.
b. Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating surface.
c. Twist the transducer twist lock lever to the 4 o’clock position to lock it in place. Twist the lever to the 12 o’clock position to disconnect the transducer.
Note
It is not necessary to turn OFF power to connect or disconnect a transducer.
2.
Connect the main power cable to a hospital grade power receptacle with the proper rated voltage checked during preinstallation. Never use a three--to--two prong adapter; this defeats the safety ground.
2--2--8 Moving into Position
CAUTION
Do not lift the unit by the Keyboard.
Do not tilt the unit more than 5 degrees to avoid tipping it over.
To avoid injury by tipping over. Set the monitor to the lowest position before moving.
CAUTION
Equipment Damage Possibility. Lifting the console by holding covers may damage the covers. Do not lift the console by holding any covers.
In general, a single adult can move the LOGIQ 200 PRO Series. (It is better to pull from the rear rather than push from the front of the unit). Before moving, store all loose parts in the unit. Wrap transducers in soft cloth or foam to prevent damage.
2--2--9 Adjusting System Clock
Set system clock for the LOGIQ 200 PRO Seriesto the local time. For procedure of adjusting the System clock, refer to “Customizing Your System” in the Users manual.
2--14
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2--2--10 Product Locator Installation Card
Fill out proper customer Information the Product Locator Installation Card. Refer to ILLUSTRATION 2--7. Mail this Installation Card “Product Locator” to the address corresponding to your pole.
Note
The Product Locator Installation Card shown in ILLUSTRATION 2--7 may not be same as the provided Product Locator card.
PRODUCT LOCATOR INSTALLATION CARD
ILLUSTRATION 2--7
2--15
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
2--16
INSTALLATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CHAPTER 3 -- SYSTEM CONFIGURATION
TABLE OF CONTENTS
TITLE SECTION
3--1
3--2
3--3
3--4
3--5
3--6
PAGE
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--3
DIMENSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--3
ELECTRICAL SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--3--1 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--3--2 Facility Power Receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--4
3--4
3--4
STORAGE AND OPERATION REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--4
OPTIONAL PERIPHERALS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--5--1 Peripherals/Accessories Connector Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--5--2 List of Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--5--3 Power Consumption of Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--5
3--5
3--9
3--11
TEST POINT, LED, DIP SWITCH AND RESET SWITCH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--6--1 Test Point List
3--6--2 LED List
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--6--3 DIP Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--6--4 Reset Switch
3--6--5 Jumper
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--12
3--12
3--14
3--15
3--15
3--15
3--1
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
3--2
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
3--1 INTRODUCTION
This chapter describes system configuration and specifications.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--2 DIMENSIONS
Regarding LOGIQ 200 PRO Series dimensions, Refer to ILLUSTRATION 3--1 for planning the position of your
LOGIQ 200 PRO Series.
WEIGHT: 76 kg
168 lbs
NOTE
LENGTH : mm (inches)
ABERRATION :
±
5%
400 (15.7)
350 (13.8)
663 (26.1)
250 (9.8)
335 (13.2)
245 (9.6)
192 (7.6)
1200 (47.2)
800 (31.5)
162 (6.4)
OVERALL DIMENSIONS
ILLUSTRATION 3--1
3--3
PRO Series
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--3 ELECTRICAL SPECIFICATIONS
Electrical conduit, junction boxes, outlets, circuit breakers, and switches should be in place before installing the
LOGIQ 200 PRO Series console.
3--3--1 Power Supply
Voltage setup is performed in the factory. Different power cables and circuit breakers are used for the 100 (115)
Vac and 220 (240) Vac versions.
3--3--2 Facility Power Receptacle
A separate power outlet with a 10 amp circuit breaker for 100 (115) Vac units, or a 5 amp circuit breaker for 220
(240) Vac units, is recommended. The specific power receptacle used depends on your country’s power line standards.
The receptacle should have International Electrotechnical Commission (IEC) approval, or equivalent.
3--4 STORAGE AND OPERATION REQUIREMENTS
Probes and peripherals are shipped in separate containers. Shipping weight is approximately 210 lbs (96 kg).
The size of the container is L82 cm x W53 cm x H140 cm (32 in. x 21 in. x 55 in). Table 3--1 provides a summary of temperature, atmospheric pressure, and humidity tolerances for shipping, installation, and operation.
TABLE 3-- 1
STORAGE AND OPERATION REQUIREMENTS
PARAMETER STORAGE OPERATION
TEMPERATURE (
_
C)
( _ F)
--10 to 60
14 to 140
10 to 40
50 to 104
ATMOSPHERIC PRESSURE
(hPa)
HUMIDITY (%)
(Non--condensing)
700 to 1060
5 to 90
700 to 1060
5 to 90
3--4
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
3--5 OPTIONAL PERIPHERALS
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--5--1 Peripherals/Accessories Connector Panel
LOGIQ 200 PRO Series peripherals and accessories can be properly connected using the rear connector panel.
Located on the panel are video input and output connectors, camera expose connector, foot switch connector, power connector and service ports.
This section indicates the pin assignment for each connector (1 -- 4 in ILLUSTRATION 3--2) at pages 3--6 through 3--7.
4
5
1
6
3
2
115V 1.7A Max including front panel
100V 2.0A Max including front panel
220--240V 0.8A Max including front panel
REAR CONNECTOR PANEL
ILLUSTRATION 3--2
3--5
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--5--1 Peripherals/Accessories Connector Panel (Continued)
Note
Each outer (case) ground line of peripheral/accessory connectors are protectively grounded.
Signal ground lines are Not Isolated.
8
9
10
11
12
13
Pin No.
1
2
3
6
7
4
5
1. Pin Assignment of RS232C for Service
Connector : Female, D--SUB, 25--pin
Signal
Frame GND
TXD
RXD
RTS
CTS
DSR
Signal GND
DCD
+ 5 V
*1
(300mA Max)
Pin No.
14
15
16
17
18
19
20
21
22
23
24
25
Signal
DTR
Ring Indicator
*1
: This voltage shall be apply to this Pin in case of MODEM usage.
Note
Output level of RS232C signals :
High
Low
+3V to +15V
--15V to 0V
14
25
1
13
3--6
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--5--1 Peripherals/Accessories Connector Panel (Continued)
2. Pin Assignment of Foot Switch
Connector : Round 5--pin connector
Pin No.
1
2
3
Output Signal
FOOT SW
FOOT SW_G
Pin No.
4
5
Output Signal
Frame GND
4
5
1
2
3
3. Pin Assignment of Mini Jack for Controlling B/W Printer
Connector :Stereo Mini Jack
Pin No.
1
Output Signal
*1
Pin No.
2
Output Signal
Signal GND
*1
: Printer starts printing by receiving the LOW pulse more than 100ms.
Note
Output level of control signals indicated in the above tables are TTL level.
4. Pin Assignment of ECG
Connector :
Pin No.
1
2
3
Output Signal
RA
LF
RF
Pin No.
4
5
Output Signal
Signal GND
Signal GND
1
2
4
5
3
3--7
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
3--5--1 Peripherals/Accessories Connector Panel (Continued)
5. Pin Assignment of Parallel Printer for Printer--A
Connector :
8
9
10
11
12
13
Pin No.
1
2
3
6
7
4
5
Signal
STB (L)
PD0
PD1
PD2
PD3
PD4
PD5
PD6
PD7
ACK (L)
BUSY
PE
SLCT
Pin No.
14
15
16
17
18
19
20
21
22
23
24
25
Signal
AFD(L)
ERR(L)
INIT(L)
SLIN(L)
GND
GND
GND
GND
GND
GND
GND
GND
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
14
25
1
13
6. Pin Assignment of for Printer--B
Connector :
Pin No.
1
2
3
4
5
Signal
DCD
RXD
TXD
DTR
Signal GND
Pin No.
6
7
8
9
Signal
DSR
RTS
CTS
Ring Indicator
6
9
1
5
3--8
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--5--2 List of Optional Peripherals
The tables below shows the suggested optional peripherals for LOGIQ 200 PRO Series.
1. RECORDING DEVICES
TABLE 3-- 2
LIST OF RECORDING DEVICES
DEVICE
Video Cassette Recorder
Video Graphic Printer
Multi Image Camera
Magneto Optic Disk Driver
MANUFACTURER
SONY
SONY
SONY
MITSUBISHI
International Imaging
Electronics
FUJITSU
MODEL
SVO--9500MD
SVO--9500MDP
UP890
P90
IIE460
CATALOG No.
VIDEO SIGNAL
Local
Local
Local
Local
Local
MDD3064AP Local
Note
See each option installation instructions for installation and connection procedures.
NTSC
PAL
3--9
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 6
3--5--2 List of Optional Peripherals (Continued)
2. TRANSDUCER (PROBE)
LOGIQ 200 PRO Series SERVICE MANUAL
TABLE 3-- 3
LIST OF TRANSDUCERS
PROBE
NAME
CBF
CAE
MTZ
CZB
CS
ATR
LH
LE
LI
LT
LB
LD
ERB
10Lb
SY
3C
S317
MATERIAL OF
HEADSHELL
PES
PES
PES
NORYL
PES
PES
PES
PES
PES
PES
PES
NORYL
PES
AREA OF USING
Abdom.
Abdom.
Intercavity
Neonatal
Abdom.
Urology
Superficial
OB/Gyn.
Intraoperative
Intraoperative
OB/Gyn.
Intraoperative
Urology
PES
PES
PES
NORYL
Superficial
Abdom.
OB/Gyn
Cardiology
TYPE CATALOG
NO.
REQUIRED
ADAPTER
Convex H46022CB Not Required
Convex H46022CA Not Required
Convex H46022MT Not Required
Convex H45202CZ Not Required
Convex H45222CS Not Required
Convex H4061PR Not Required
Linear H46022LH Not Required
Linear H46022LE Not Required
Linear H46022LI Not Required
Linear H46022LT Not Required
Linear H46022LB Not Required
Linear H45202LD Not Required
Linear,
Convex
H45202ER Not Required
Linear H45202LM Not Required
Sector H46022SY Not Required
Linear H45202WB Not Required
Sector H45202SD Not Required
PART NO.
FOR JAPAN
P9603AA
P9603AB
P9603AL
2152402
2202315
2201223
P9601AC
P9601AB
P9601AG
P9601AJ
P9601AA
P9601AD
--
--
--
--
2144268--2
2235374
PART NO.
P9603AD
P9603AE
P9603AU
2152422
2202320
2201222
P9601AS
P9601AR
P9601AW
P9601AX
P9601AQ
2124317
--
--
--
--
--
CAUTION
Equipment damage possibility. Be sure to use the following recommended connecting cables to connect recording devices with LOGIQ 200 PRO Series console.
3. CONNECTING CABLES
NAME
Printer Install Kit
MOD Install Kit
PART NO.
2176459
2235371
TABLE 3-- 4
LIST OF CONNECTING CABLES
FIGURE
Printer Cable Assy, BNC cable,
Shutter cable
MOD Driver, Top Bracket,
Bottom Bracket, Screw
NOTE
For B/W Printer
For MOD Driver
3--10
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--5--3 Power Consumption of Optional Peripherals
The table below shows the power consumption of each optional peripheral for LOGIQ 200 PRO Series.
TABLE 3-- 5
POWER CONSUMPTION OF OPTIONAL RECORDING DEVICES
DEVICE
Video Cassette Recorder
Video Graphic Printer
Multi Image Camera
MOD Driver
MODEL
SVO--9500MD
SVO--9500MDP
UP890
IIE460
MDD3064AP
POWER CONSUMPTION
64 W
110 W
60 W
5.3 W
3--11
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--6 TEST POINT, LED, DIP SWITCH AND RESET SWITCH
3--6--1 Test Point List
The table below shows The Test Point list for LOGIQ 200 PRO Series.
LOCATION
MVP ASSY
MSTE ASSY
TBF ASSY
NAME
TP1
TP2
TP3
TP4
TP5
TP6
TP7
TP8
GND
VCC1
--5V
TP1
TP2
TP3
DGND2
AGND1
AGND2
TABLE 3-- 6
TEST POINT
DESCRIPTION
Non--interlaced HSYNC
Non--interlaced VSYNC
M--mode data clock
M_EN
SLN7
B mapping clock
B DTC
CPU clock
Ground
+5V for analog/digital
--5V for analog
+5V for digital
GND
Foot switch status
Ground for Digital
Ground for Analog
Ground for Analog
POSITION
Edge of board
Edge of board
--
--
--
--
--
--
Edge of board
Edge of board
--
Edge of board
Edge of board
--
--
--
--
3--12
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--6--1 Test Point List (continued)
The table below shows The Test Point list for LOGIQ 200 PRO Series.
LOCATION
BPHV ASSY
RBF ASSY
SCAN ASSY
MOTHER 3 ASSY
TABLE 3-- 7
TEST POINT
NAME
TP
TP
¦
¦
THV
SHV
TP HV_REF
V5AP
V5AN
V5BG
V5AD
VCC1
AGND1
AGND2
GND1
GND2
GND3
GND4
GND5
+5V
¦ SHV
MON_12P
ECG_12P
ECG_12M
GND1
GND2
GND3
TRIG
DESCRIPTION
High Voltage for transmitting
High Voltage for High Voltage switch on CONN Assy
D/A output for HV control
+5V for Analog
--5V for analog
+5V for Gain Control Block
+5V for ADC Block
+5V for Digital
Ground for Analog
Ground for Analog
Ground for Digital
Ground for Digital
Ground for Digital
Ground for Digital
Ground for Digital
+5V for digital
¦
SHV for high voltage switch
+12V for Monitor
+12V for ECG Module
--12V for ECG Module
Ground
Ground
Ground
TRIG Signal
POSITION
--
--
--
Edge of board
Edge of board
--
--
--
--
--
--
--
--
--
--
Edge of board
Edge of board
--
--
--
--
--
--
--
3--13
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
3--6--2 LED List
The table below shows the LED list for LOGIQ 200 PRO Series.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
TABLE 3--7
LED
LOCATION
MSTE ASSY
MVP ASSY
NAME
DS9
DS10
DS1~DS8
DS11
DS1
DS2
DS3
DS4
DS5
DS1, DS2
DESCRIPTION
+5V for digital
Watchdog timer out
(CPU FAULT)
Reserved
MOD Active
+5V for digital
Scan Frame Clock
POSITION
Edge of board
NORMAL
ON
OFF
--
--
ON
BLINKING
BPHV ASSY
SMPS ASSY GREEN
LED
DSP running
BPHV ASSY status
(ON:normal)
SMPS ASSY status
(ON:normal)
SMPS ASSY
ON
ON
ON
Note
Three LED (DS2~DS4) of MVP assy should be blinked when the system operated with the
Probe.
ABNORMAL
OFF
ON
--
--
OFF
FIXED
OFF
OFF
OFF
3--14
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 4
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--6--3 DIP Switch
The table below shows the DIP Switch list for LOGIQ 200 PRO Series.
LOCATION NAME Switch No.
TABLE 3--8
DIP SWITCH Setting
DESCRIPTION
ON
STATUS
OFF
-enable
-enable
-disable
-disable
MSTE ASSY S1
S2
1 ~6
7
4
5
2
3
8
1
6
7
8
Assy Version Read, I/O check
System Cold Reset
Reserved
Initialize Setup data to factory default
Reserved
Reserved
A--Mode
Emergency mode for download by MOD or PL
Reserved
Flash memory/SRAM Test & Clear
NTSC/PAL Selector
Note
To activate the dip switch S2--7 fuction, S2--1 should be set to ON.
Note
To activate the A--Mode fuction, S2--4 should be set to ON.
LOCATION NAME Switch No.
MVP ASSY S1 1 ~8
TBF ASSY
RBF ASSY
S1
S1
1 ~8
1 ~7
DESCRIPTION
Assy Version Read, I/O Check
Assy Version Read, I/O check
Assy Version Read, I/O check
PAL NTSC
POSITION
--
Edge of board
DIN connector side
3--6--4 Reset Switch
Reset switch (S3) on edge of MSTE Assy used for resetting the system.
3--15
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3--6--5 Video Signal Specification
The table below shows the video signal specification for LOGIQ 200 PRO Series.
TABLE 3--9
VIDEO SIGNAL SPECIFICATIONS
NTSC
VCR
OUT
Monitor
OUT
VCR
OUT
Total Line per Frame
Vertical Field Frequency
[Line]
[Hz]
525
60
525
60
625
50
Horizontal Scanning Frequency [KHz]
15.733
31.466
15.625
Total Horizontal Line Time
Horizontal Display
Horizontal Porch Width
Horizontal Pulse Width
Horizontal Porch Width
Total Horizontal Blanking
Vertical Front Porch Width
[
µ
s]
[
µ
s]
[
µ
s]
[
µ
s]
[
µ
s]
[
µ
s]
[H]
63.56
49.54
2.76
4.73
6.53
14.02
6.5
31.78
27.3
1.18
2.81
0.49
4.48
0
64.00
48.81
3.09
4.68
7.42
15.19
9.0
Vertical Sync Width
Vertical Back Porch Width
Total Blanking Interval
Pixel Clock
H
Display Area Size
[H]
[H]
[H]
3
22
31.50
[MHZ]
12.27
[
µ
s]
63.56
[mm]
--
4
0
0
24.54
31.87
207X157
2.5
27
38.50
14.75
64
--
PAL
Monitor
OUT
0.3
4.48
0
4
0
0
29.5
32
625
50
31.25
32.00
27.52
1.42
2.76
207X157
3--16
SYSTEM CONFIGURATION
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CHAPTER 4 -- FUNCTIONAL CHECKS
TABLE OF CONTENTS
TITLE SECTION
4--1
4--2
4--3
PAGE
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--1--1 Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--3
4--3
FUNCTIONAL CHECK PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--2--1 Basic Controls
4--2--2 M--Mode Check
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--3
4--3
4--5
SMPS ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--3--1 SMPS Assy Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--3--2 SMPS Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4--7
4--7
4--9
4--1
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
4--2
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
4--1 INTRODUCTION
This chapter provides procedures for quickly checking major functions of the LOGIQ 200 PRO Series console, and SMPS adjustments.
4--1--1 Required Equipment
To perform these tests, you’ll need a linear, or a convex transducer.
4--2 FUNCTIONAL CHECK PROCEDURES
4--2--1 Basic Controls
Step Check
1
2
Connect the convex transducer to ”Probe 1” connector.
Power On
3
4
5
6
7
8
Expected Result
Rotate B/M Gain knob
After few seconds, the B mode screen should appears as shown in ILLUSTRATION 4--1.
Image gets lighter with CW rotation and darker with
CCW.
The gray scale adjusts to each new Map selected.
Press Map key to select another gray scale Map.
Press Dyn Range Arrow up or down key.
At lower Dynamic Range settings, image speckle fades and prominent objects in the display are more pronounced from the background image.
Rotate Depth knob.
The depth of image should be increased /decreased.
Slide TGC potentiometers (pots) Image grows darker or brighter at depth equivalent to pot’s location.
Press Zoom. key.
The image should increase to X2 size.
9
Press it again to exit.
Press Frame Avg key.
10
11
12
Press Edge Enhc key.
Press Reverse key.
Press Reverse key again.
Image speckle fades and probe or wire movement is smeared.
The edges inside the focal area(s) should become lighter when you increase and darker as you decrease its value.
The image reverses the left/right orientation.
The image reverses again.
4--3
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
4--2--1 Basic Controls (Continued)
Map1
B--Mode
Cine Gauge
B--MODE DISPLAY SCREEN
ILLUSTRATION 4--1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
GE
4--4
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
4--2--2 M--Mode Check
Step Check
13
Press M key.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Roll trackball, position cursor over area you want to see in motion.
Press M key again.
14
Rotate B/M Gain knob
15
Press Dyn Range Arrow up or down key.
16
Press Sweep Speed key
Press it again to exit.
17
Press Freeze key.
Press it again to exit.
18
Press Edge Enhc key.
19
Press B key.
Expected Result
The M mode timeline should appear next to the B image as shown in ILLUSTRATION 4--2. Whether it takes half the screen or two--thirds depends on the preset.
The Mode cursor should follow trackball movement and the timeline should update for new location of focus.
The full M--mode should appear on the CRT monitor.
The M timeline should get brighter with CW rotation and darker with CCW.
Dynamic Range affects grays and the last added scan mode; to adjust the basic B, M must be off. Turn
Dynamic Range down to increase contrast, turn up to soften.
The timeline speed should increase to 4 second sweeps and decrease to 16 second sweeps.
Fast=4 Medium=8 Slow=16
The image should freeze.
The image revives acquisition.
Changes the M image
The M Mode timeline should disappear and the
B--mode image should appear as shown in
ILLUSTRATION 4--1.
4--5
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
4--2--2 M--Mode Check (Continued)
Map1
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
GE
M--Mode
MuFq
Cine Gauge
M--MODE DISPLAY SCREEN
ILLUSTRATION 4--2
Note
You can select several types of display formats by using the Setup Menu. For the Preset Menu, refer to Customizing Your System in the LOGIQ 200 PRO Series User Manual.
4--6
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
4--3 SMPS ADJUSTMENTS
This section provides SMPS adjustment procedures for the LOGIQ 200 PRO Series. Adjustments should be only made when necessary. SMPS adjustments should be made in accordance with the schedule for periodic maintenance in Chapter 7 of this manual.
Before beginning the SMPS adjustments procedure, make sure the power outlet conforms to the proper power line standards. Refer to Chapter 2, Installation.
Note
If the adjustment pot is turned to far clockwise, the SMPS output shuts down to protect the circuits against the over--voltage. In that case, power the LOGIQ 200 PRO Series OFF and turn the pot counterclockwise all the way. Wait for 5 minutes for recovering of SMPS protection circuit. Then power it ON and try to adjust the SMPS again.
The SMPS Assy is in the bottom of the LOGIQ 200 PRO Series as shown in ILLUSTRATION 4--7.
4--3--1 SMPS Assy Access
1.
Remove the four screw caps and unscrew the screws to remove the Left Cover as shown in the
ILLUSTRATION 4--3.
419 EMI Cover
404 Left Cover
1
2
3
4
LEFT COVER REMOVAL
ILLUSTRATION 4--3
4--7
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
4--3--1 SMPS Assy Access (Continued)
2.
Remove the EMI Cover L as shown in ILLUSTRATION 4--4.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
419 EMI Cover L
3
1
2
5
4
7
6
8
EMI COVER REMOVAL
ILLUSTRATION 4--4
3.
Remove the four screw caps and unscrew the screws to remove the Right Cover as shown in the
ILLUSTRATION 4--5.
RIGHT COVER REMOVAL
ILLUSTRATION 4--5
4--8
405 Right Cover
1
2
3
4
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
4--3--1 SMPS Assy Access (Continued)
4.
Remove the four screw caps and unscrew the screws to remove the Rear Cover as shown in the
ILLUSTRATION 4--6.
408 Rear Cover
1
2
3
4
REAR COVER REMOVAL
ILLUSTRATION 4--6
4--9
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
4--3--2 SMPS Adjustment Procedure
1.
Power LOGIQ 200 PRO Series ON. Wait for about 30 seconds to warm up the console.
2.
Connect a DVM to the appropriate place shown in Table 4--1.
3.
Verify that the voltages are as shown in Table 4--2.
SMPS
±
12V for ECG
+12 for Monitor
+5V for Digital
TABLE 4-- 1
SMPS MEASUREMENT LOCATION
MEASURE AT
ECG_12P, ECG_12M on
MOTHER3 BD ASSY
RETURN AT
AGND3 (Ground) on
MOTHER3 BD ASSY
MON_12P on MOTHER3 BD
ASSY
AGND3 on MOTHER3 BD
ASSY
VCC1 (+5V) on MVP BD ASSY GND1 (GND) on MVP BD
ASSY
ADJUST AT
1 on SMPS ASSY
See ILLUSTRATION 4--7
3 on SMPS ASSY
See ILLUSTRATION 4--7
2 on SMPS ASSY
See ILLUSTRATION 4--7
SMPS
--5V,
±
12V for Analog
+12 for Monitor
+5V for Digital and Analog
TABLE 4--2
SMPS MEASUREMENT TOLERANCES
MIN
--4.8 V
--12.6V
+12.6 V
+11.8 V
+4.8 V
MAX
--5.2 V
--11.4 V
+11.4 V
+12.2 V
+5.2 V
4--10
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
4--3--2 SMPS Adjustment Procedure (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1
2
3
1. Adjustment Point for --5V,
±
12V SMPS
2. Adjustment Point for +5V SMPS
3. Adjustment Point for +12V SMPS
SMPS ASSY
ILLUSTRATION 4--7
4--11
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
4--12
FUNCTIONAL CHECKS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CHAPTER 5 -- DIAGRAMS
TABLE OF CONTENTS
TITLE SECTION
5--1
5--2
5--3
5--4
5--5
PAGE
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--3
LOGIQ 200 SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--3
BLOCK DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--4
WIRING DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--5
CIRCUIT BOARD DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5--7
5--1 DIAGRAMS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
5--2 DIAGRAMS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
5--1 INTRODUCTION
LOGIQ 200 PRO Series is a compact ultrasound scanner supporting a wide range of probes. This gives the system added benefits and flexibility to meet diverse applications.
5--2 LOGIQ 200 PRO Series SYSTEM
The LOGIQ 200 PRO Series has a 48 channel beamformer that digitizes the RF signal. It uses digital delay to focus the acoustic beam. It offers parallel receive beam formation which can increase frame rate by a factor of two in time critical applications. This system also features many advanced image processing controls.
User surveys and the latest technology were used to increase console ease--of--use.
System Features
The key design goals of this system are:
•
High Image Quality
•
Increased User Productivity
•
Multiple Clinical Applications
•
Planned Upgradeability
•
High Mobility
Types of Applications
The system supports many clinical uses. Scan and display parameters may be user selected to default to desired values for each application. The system presets many parameters to clinically determined, optimal values.
•
RAD/ABDOMEN
•
OB/GYN
•
UROLOGY
•
SMALL PARTS
•
UROLOGY
•
CARDIOLOGY
See Illustration 5--1, the LOGIQ 200 PRO Series system can be divided into an beamforming section (from TBF to
RBF assy) and an image reconstruction processing section (from MVP to MSTE). The beamforming section is to focus a RF signal in transmiting and receiving. Image reconstrcution section converts the receiving scan data to some kinds of images that are B mode convex image, B--mode linear image, M--mode image. Another section has the microprocessor driven system control section, which controls the system based on operator commands and system status information.
5--3 DIAGRAMS
GE MEDICAL SYSTEMS
REV 0
5--3 BLOCK DIAGRAM
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Power
Supply
To Mother board
To Mother board
PIF
External Video In
External Video Out
ECG
Shutter Control
Foot Switch
Service
Printer B (RS232C)
Printer A (Parallel)
MSTE
MOD
Keyboard
SHV
BPHV
THV
TBF
RF Data
RBF
BM Data
MVP
B/W Monitor
ECHO SIGNAL
SCAN
NOVA BUS
USC BUS
LOGIQ 200 PRO Series SYSTEM BLOCK DIAGRAM
ILLUSTRATION 5--1
5--4 DIAGRAMS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
5--4 WIRING DIAGRAM
PLUG
L(Brown)
N(Light Blue)
G(Green/Yellow)
TB2
TB1
1
2
3
4
1
NFB1
3
2
4
10A/115V
5A/220V
N(Light Blue)
L(Brown)
ISOLATION TRANSFORMER
PRI
:
SEC
0
100
7
4
115
Core
9
8
7
8
5
0
100
12
0
9
6
115
10
1
100
115
0
11
10
POWER SW
TB2
1
2
3
4
5
6
L
N
AC OUTLET B/W PRT
L G
Orange
Gray
NFB3
NFB4
2A/115V
1A/220V
N
TB3
1
2
N G L
CON1
11
12
2
3
100
115
15 17
100
0
16
NFB2
3A/115V
3
4
5
6
5
6
2
3
7
8
1
4
J2
J3
J1
8
2
4
2
3
5
6
2
1
BPHV ASSY
GND
AC110 V
+SHV
7
+THV
GND
HV--REF
--5V
+5V
-- SHV
--THV
1
2
1
2
4
3
5
MON ASSY
+12V
GND
FG
MON_VID
RTN
REAR PANEL
F SW
SHTR
EXT_VID_IN
EXT_VID_OUT
PRINTER_A
SERVICE
PRINTER_B
ECG
PIOP ASSY
DC FAN SET
MOD
KEY B’D
L N
FG
HK100A--5/A
+5V G
L N FG
HKT--50--522/A
--12V +12V G1 +5V G2
L N FG
HK50A--12/A
+12V G
CON2
B/W PRINTER
PRT_VID
SHTR
1 2
+
5
VA
G
N
D
J25
1 2 3 4 5
+
5
VD
+
5
VD
+
5
VD
G
N
D
G
N
D
J12
1
--
12
VA
+
12
VA
2
J19
3 4 5
--
5
V
G
N
D
G
N
D
1 2
1
+
12
V
J6
G
N
D
+
5V
J7
G
N
D
2
+
12
V
1 2
G
N
+
12
V
G
N
+
12
V
D
G
N
D
D
J3 J23 J24
J1 J29 J4 J16 J17 J18
J5
J11
MOTHER B’D
LOGIQ 200 PRO Series SYSTEM WIRING DIAGRAM
ILLUSTRATION 5--2
5--5
J5 J6 J9 J8 J1 J2 J3 J500
J10
J4
PIF B’D
DIAGRAMS
GE MEDICAL SYSTEMS
REV 0
5--4 WIRING DIAGRAM (Continued)
Brown
Brown
Light Blue
Light Blue
Orange
Gray
Orange
Gray
6
5
4
3
2
1
10
9
8
7
TB2
12
11
115V
100V
100V
0V
115V
100V
0V
0V
115V
100V
0V
115V
Brown
Brown
Light Blue
Orange
Orange
Gray
(100 V)
TB2
12
11
10
9
8
7
6
5
4
3
2
1
115V
100V
0V
115V
100V
0V
115V
100V
0V
115V
100V
0V
AC IN
Peripheral
Out
AC IN
Peripheral
Out
LOGIQ 200 PRO Series SERVICE MANUAL
Brown
Brown
Light Blue
Light Blue
Orange
Orange
Gray
Gray
Brown
Brown
Light Blue
Orange
Orange
Gray
Gray
(115 V)
TB2
12
11
6
5
4
3
2
1
10
9
8
7
115V
100V
0V
115V
100V
0V
115V
100V
0V
115V
100V
0V
6
5
4
3
2
1
10
9
8
7
TB2
12
11
115V
100V
100V
0V
115V
100V
0V
0V
115V
100V
0V
115V
(220 -- 240 V)
(INPUT: 220 -- 240 V
OUTPUT: 115 V)
LOGIQ 200 PRO Series SYSTEM WIRING DIAGRAM (CONTINUED)
ILLUSTRATION 5--3
AC IN
2235374
Peripheral
Out
AC IN
Peripheral
Out
5--6 DIAGRAMS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
5--5 CIRCUIT BOARD DESCRIPTION
The following table lists circuit boards and their respective card cage slot assignments on the mother board of the
LOGIQ 200 PRO Series system.
CARD CAGE SLOT
3
4
1
2
BOARD NAME
TBF BD ASSY
RBF BD ASSY
MVP BD ASSY
MSTE BD ASSY
SCAN BD ASSY
BPHV BD ASSY
MOTHER 3 ASSY
TABLE 5-- 1
CIRCUIT BOARD DESCRIPTION
DESCRIPTION
Transmit Beam Former Board Assembly
Receive Beam Former Board Assembly
Multi--Function Video Processor Board Assembly
Master Enhanced Board Assembly
Scan Board Assembly
Bipolar High Voltage Board Assembly
Motherboard Assembly
NOTE
5--7 DIAGRAMS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
5--8 DIAGRAMS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CHAPTER 6 -- RENEWAL PARTS
TABLE OF CONTENTS
TITLE SECTION
6--1
6--2
PAGE
RENEWAL PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--3
DISASSEMBLY/RE--ASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--1 Monitor Assy (FRU No. 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--2 Pot Set Assy (FRU No. 101)
6--2--3 CRT Filter(FRU No. 102)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--4 Monitor Cover Set (FRU No. 103)
6--2--5 Tilt Assy(FRU No. 104)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--6 Filter Clamp Set (FRU No. 105) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--7 Probe Holder (FRU No. 201) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--8 Holder Bracket Assy (FRU No. 202) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--9 Cover Set (FRU No. 203) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--10 Keyboard Assy (FRU No. 204) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--11 Trackball Assy (FRU No. 205) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--12 TGC Assy (FRU No. 206) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--13 Encoder Set (FRU No. 207) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--14 Freeze/Record Key Assy (FRU No. 208)
6--2--15 TGC Knob Set (FRU No. 209)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--16 Rubber Key Set (FRU No. 210) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--17 P.C.Board(s) (FRU No.301 through 304)
6--2--18 BPHV Assy (FRU No. 305)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--19 SCAN Assy (FRU No. 306) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--20 Shield Panel (FRU No. 307) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--21 Mother 3 Assy (FRU No. 308) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--22 Nest Box (FRU No. 309) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--23 DC Fan Assy (FRU No. 310) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--24 Swing Arm Assy (FRU No. 401) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--25 Pipe Assy (FRU No. 402) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--26 Pole Cover (FRU No. 403) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--27 Left Cover (FRU No. 404) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--28 Right Cover (FRU No. 405) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--29 Front Base Cover (FRU No. 406) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--30 Front Cover (FRU No. 407) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--44
6--46
6--48
6--50
6--52
6--54
6--56
6--58
6--60
6--62
6--64
6--65
6--66
6--68
6--29
6--30
6--32
6--34
6--36
6--37
6--38
6--39
6--40
6--42
6--21
6--22
6--24
6--25
6--26
6--27
6--28
6--1 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
TABLE OF CONTENTS (continued)
SECTION TITLE PAGE
6--2--31 Rear Cover (FRU No. 408)
6--2--32 Top Cover (FRU No. 409)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--33 Printer Cover (FRU No. 410) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--34 Printer Bracket Assy(FRU No. 411) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--35 Rear Handle (FRU No. 412) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--36 Neck Frame (FRU No. 413) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--37 Rear Panel Assy (FRU No. 414) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--38 Power S/W Assy (FRU No. 415)
6--2--39 Front Caster (FRU No. 416)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--40 Rear Caster (FRU No. 417) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--41 Bumper Set (FRU No. 418) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--42 EMI Cover L (FRU No. 419) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--43 PCB Guide Assy (FRU No. 420) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--44 Terminal Block 12 Assy (FRU No. 501) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--45 Terminal Block 6 Assy (FRU No. 502) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--46 SMPS Assy (FRU No. 503) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--47 Power Trans Assy (FRU No. 504) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--2--48 Circuit Breaker Set (FRU No. 505) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--88
6--89
6--90
6--92
6--94
6--96
6--98
6--100
6--70
6--72
6--74
6--76
6--78
6--80
6--82
6--84
6--86
6--87
6--3
6--4
FUSE REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--3--1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--3--2 Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--101
6--101
6--101
SMPS Assy REPLACEMENT
6--4--1 Introduction
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--4--2 Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6--104
6--104
6--104
6--2 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 2
6--1 RENEWAL PARTS
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
LOGIQ 200 PRO Series
MATERIAL LIST (1)
PART NAME
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
PART NO.
QUANTITY
2233819 1
2242419
2233820
2233821
2233822
2233823
2233824
2233825
2233826
1
1
1
1 1
1 1
1
1
220V, 4SW
220V, 4SW
100V, 4SW
115V, 4SW
220V, 4SW
220V, 4SW
115V, 2SW
220V, 2SW
1 220V, 2SW
DESCRIPTION
VER. 3.0
VER. 3.0
VER. 3.0
VER. 3.0
VER. 3.0
VER. 3.0
VER. 3.0
VER. 3.0
VER. 3.0
PROBE CABLE ARM 2215992 1 1 1 1 1 1 1 1 1 1 1
MOD INSTALL KIT 2235371 1 1 1 Include MOD
RIGHT & LEFT MTZ INSTALL KIT
FOOT SWITCH ASSY
CINE S/W
VINYL COVER
INSITE KIT
2242590
2162242
2242585 1
2235373
2242586
1 1 1 1
1 1 1
1 1 1
PRINTER INSTALL KIT 2242589 1 1 1 1 1 1 1 1 1 1 1
ATO S/W 2251544 1
1
ADVANCED REF. MANUAL
2233222--100
1 1 1 1 1
SERVICE MANUAL 2235374
QUICK START GUIDE
2234974--100
SONY B/W PRINTER E8310KA
1 1 1 English
1 1 1 1 1 English
1 1 1 English
1 1 1
ECG INSTALL KIT
L500 ECG CABLES
ECG CABLE SET
ECG CABLE ASSY
ECG CABLE ASSY
2242584
P9509KH
2137161
P9509KG
2137160
Clip type
Rod type
Japan only, Clip type
Japan only, Rod type
KOREA (H43132LA)
PHILIPPINES (H43132LK)
JAPAN (H43132LB)
TAIWAN (H43132LC)
CHINA, HONGKONG, INDIA (H43132LD)
SOUTHEAST ASIA, AUSTRALIA, NJ (H43132LD)
CHILE (H43132LH)
BRAZIL (H43132LG)
USA (H43132LF)
EUROPE (H43132LL)
EUROPE (H43132LE)
6--3 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 6
6--1 RENEWAL PARTS (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
PART NAME
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
CONSOLE ASSY
LOGIQ 200 PRO Series
MATERIAL LIST (2)
PART NO.
QUANTITY
2270964 1
2270972
2270965
2270966
2271424
2270967
2270968
2270969
2270970
2270971
1
1
1
1
1
1 1
1
1
220V, 4SW
220V, 4SW
100V, 4SW
115V, 4SW
220V, 4SW
220V, 4SW
220V, 4SW
115V, 2SW
220V, 2SW
1 220V, 2SW
DESCRIPTION
VER. 3.1
VER. 3.1
VER. 3.1
VER. 3.1
VER. 3.1
VER. 3.1
VER. 3.1
VER. 3.1
VER. 3.1
VER. 3.1
PROBE CABLE ARM 2215992 1 1 1 1 1 1 1 1 1 1 1
MOD INSTALL KIT 2235371 1 1 1 1 Include MOD
MTZ INSTALL KIT 2242590 RIGHT
LEFT MTZ INSTALL KIT
FOOT SWITCH ASSY
CINE S/W
2242591
2162242
2242585 1 1 1 1 1
1 1 1
1 1 1
PRINTER INSTALL KIT 2242589 1 1 1 1 1 1 1 1 1 1 1
ATO S/W
VINYL COVER
INSITE KIT
2251544
2235373
2242586
1
1
ADVANCED REF. MANUAL
SERVICE MANUAL
2233222--100
2235374
QUICK START GUIDE
2234974--100
SONY B/W PRINTER E8310KA
ECG INSTALL KIT 2242584
1 1 1 1 1 1 1 1 English
1 1 1 1 1 English
1 1 1 English
1 1 1
KOREA (H43142LA)
PHILIPPINES (H43142LK)
JAPAN (H43142LB)
TAIWAN (H43142LC)
INDIA (H43142LM)
SEA, AUS, NJ, CHINA, HONGKONG (H43142LD)
6--4
CHILE (H43142LH)
BRAZIL (H43142LG)
USA (H43142LF)
EUROPE (H43142LL)
EUROPE (H43142LE)
RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
OPERATOR CONSOLE ASSY
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
MONITOR BLOCK
FRU 100 -- 105
See ILLUSTRATION 6--2
KEYBOARD BLOCK
(FRU 201 -- 210)
See ILLUSTRATION 6--3
PC BOARDS(2)
See ILLUSTRATION 6--4
BODY BLOCK
See ILLUSTRATION 6--5
PC BOARDS(1)
See ILLUSTRATION 6--4
POWER SUPPLY BLOCK
See ILLUSTRATION 6--6
ILLUSTRATION 6--1
6--5 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
OPERATOR CONSOLE ASSY 1/7
101
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
102
105
104
103
ILLUSTRATION 6--2
6--6 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 6
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
MATERIAL LIST (1/8)
FRU
NO.
PART NAME
100 MONITOR ASSY
MONITOR ASSY
MONITOR ASSY
101 POT SET ASSY
102 CRT FILTER (PLASTIC)
CRT FILTER (GLASS)
103 MONITOR COVER SET
MONITOR COVER SET
MONITOR COVER SET
104 TILT ASSY
105 FILTER CLAMP SET
: V3.0
: V3.1
PART NO QTY
.
2235375--3 1
FRU
2235376--3
2249964--3
2148193
2251535
1
1
1
1
DESCRIPTION
1 Including Caution Label,
Glass Filter
1 Including Caution Label
, Plastic Filter
1 Including Caution Label,
For JAPAN only
2 Bright and Contrast,
Include Knob
2 Include 2 brackets
SECTION FOR
REFERENCE
6--2--1
6--2--1
6--2--1
6--2--2
6--2--3
2175438
2235385
2249415
2249966
2249414
2214397
2214398
1
1
1
1
2 Include 2 brackets
For USA, Brazil, Chile,
2 For Plastic Filter
For Glass Filter
For Japan only
2
2 For Plastic Filter
For Glass Filter
6--2--3
6--2--4
6--2--4
6--2--4
6--2--5
6--2--6
6--7 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
OPERATOR CONSOLE ASSY 2/7
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
206, 209
205
208
204
207
210
203
201
202
PG Rivet
ILLUSTRATION 6--3
6--8 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 6
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
FRU
NO.
PART NAME
201 PROBE HOLDER
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
202 HOLDER BRACKET ASSY
203 COVER SET
204 KEY BOARD ASSY
205 TRACKBALL ASSY
206 TGC ASSY
207 ENCODER SET
208 FREEZE/RECORD KEY ASSY
209 TGC KNOB SET
210 RUBBER KEY SET
: V3.0
: V3.1
MATERIAL LIST (2/8)
PART NO QTY.
FRU
2175439
2148199
2235384
2235386--2
2235387
2235388--2
2235390
2235391
2214594
2249182
1
1
1
1
1
1
1
1
1
1
2
2
2
1
2
2
2
2
2
1
DESCRIPTION
Include Probe Cup
Holder and Probe Sheet
Include PG Rivet
SECTION
FOR
REFERENCE
6--2--7
Include FRU 205--210
3 Knobs
8 Knobs
6--2--8
6--2--9
6--2--10
6--2--11
6--2--12
6--2--13
6--2--14
6--2--15
6--2--16
6--9 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
OPERATOR CONSOLE ASSY 3/7
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
308
310
309
301--304
306
305
307
ILLUSTRATION 6--4
6--10 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 6
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
MATERIAL LIST (3/8)
FRU
NO.
PART NAME PART NO QTY FRU DESCRIPTION
301 TBF ASSY
302 RBF ASSY
303 MVP ASSY
304 MSTE ASSY
305 BPHV ASSY
306 SCAN ASSY
307 SHIELD PANEL
308 MOTHER3 ASSY
309 NEST BOX
310 DC FAN ASSY
2216833--2
2216835--2
2233428--3
2266029--3
2216832--9
2266030--10 1
2240652 1
1
1
1
1
1
2217956
2240645
2220786--2
2235392
2244961
1
1
1
1
1
1 software ver 3.02 or later
1
1 software ver 3.05 or later
1 software ver 3.15 or later
1 software ver 3.06b
1 software ver 3.17a
1 Include Bracket
1
2
1 Include Bracket Probe
2
1 2 fans
6--2--18
6--2--19
6--2--20
6--2--21
6--2--22
6--2--23
: V3.0
: V3.1
SECTION
FOR
REFERENCE
6--2--17
6--11 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
OPERATOR CONSOLE ASSY 4/7
LOGIQ 200 PRO Series SERVICE MANUAL
412
408
402
401
403
409
405
420
413
419
414
415
404
417
407
406
416
418
411
410
2235374
ILLUSTRATION 6--5
6--12 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 3
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
FRU
NO.
PART NAME
401 SWING ARM ASSY
402 PIPE ASSY
403 POLE COVER
404 LEFT COVER
405 RIGHT COVER
406 FRONT BASE COVER
407 FRONT COVER
408 REAR COVER
409 TOP COVER
410 PRINTER COVER SET
411 PRINTER BRACKET ASSY
PRINTER BRACKET ASSY
PRINTER BRACKET ASSY
MATERIAL LIST (4/8)
PART NO
QTY.
FRU
2148212
2256761
2148214
1
1
1
1
1
2
DESCRIPTION
Include Cable Assy
Include Cover Pole,
Curtain Pole
2235396
2235397
2148217
2148218--2
2175441
2158086
2235395
2242780
1
1
1
1
1
1
1
1
1
2
1
2
2
1
1
1
Include Caution Label
2242781
2242782
1
1
2
2
For Korea, Taiwan,
Phillippines, USA, Chile
For Japan
For China, Hongkong,
Southeast Asia, India,
Australia, Europe, Brazil
: V3.0
: V3.1
SECTION
FOR
REFERENCE
6--2--24
6--2--25
6--2--26
6--2--27
6--2--28
6--2--29
6--2--30
6--2--31
6--2--32
6--2--33
6--2--34
6--13 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 3
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
FRU
NO.
PART NAME
412 REAR HANDLE
413 NECK FRAME
414 REAR PANEL ASSY
REAR PANEL ASSY
REAR PANEL ASSY
415 POWER S/W ASSY
416 FRONT CASTER
417 REAR CASTER
REAR CASTER
418 BUMPER SET
419 EMI COVER L
420 PCB GUIDE ASSY
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
: V3.0
: V3.1
MATERIAL LIST (4/8)(Continued)
PART NO
QTY.
FRU DESCRIPTION
2148222
2148223--2
2240733
1
1
1
2
2
1
2240734
2240735
1
1
1
1
For Korea, Taiwan,
Phillippines, USA, Chile
For Japan
For China, Hongkong,
Southeast Asia, India,
Australia, Europe, Brazil
2148226
2192111
2192112
1
2
2
1
1
1
2192113 2 2
Swivel
For Japan, Taiwan,
Phillippines, China, Hongkong,
Southeast Asia, India,
Australia, Europe
Fixed
For USA, Brazil, Chile
2148230
2169025
2169263
1
1
1
2
2
2
SECTION
FOR
REFERENCE
6--2--35
6--2--36
6--2--37
6--2--38
6--2--39
6--2--40
6--2--41
6--2--42
6--2--43
6--14 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--15 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
OPERATOR CONSOLE ASSY 5/7
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
504
503
501
505
502
ILLUSTRATION 6--6
6--16 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 3
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
MATERIAL LIST (5/8)
FRU
NO.
PART NAME
501 TERMINAL BLOCK 12 ASSY
502 TERMINAL BLOCK 6 ASSY
503 SMPS ASSY
504 POWER TRANS ASSY
(INPUT:220V--OUTPUT:115V)
POWER TRANS ASSY
(INPUT:100V--OUTPUT:100V)
POWER TRANS ASSY
(INPUT:115V--OUTPUT:115V)
POWER TRANS ASSY
(INPUT:220V--OUTPUT:220V)
505 CIRCUIT BREAKER SET
CIRCUIT BREAKER SET
CIRCUIT BREAKER SET
PART NO QTY.
FRU
2148231 1 2
2248909
2235398
2243660
1
1
1
2
1
1
2243661 1 1
DESCRIPTION
12 pins
6 pins
For Korea, Chile,
Phillippines
For Japan
2243662
2243664
2174991
2174992
2174993
1
1
1
1
1
1
1
1
1
1
For Taiwan, USA
For China, Hongkong,
India, Southeast Asia,
Australlia, Europe,
Brazil
2A(2),3A, 5A ,
For Korea, Chile,
Phillippines
2A(2), 3A, 10A
For Japan, Taiwan,
USA
1A(2), 3A, 5A
For China, Hongkong,
India, Southeast Asia,
Australia, Europe, Brazil
: V3.0
: V3.1
SECTION FOR
REFERENCE
6--2--44
6--2--45
6--2--46
6--2--47
6--2--48
6--17 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 6
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
MATERIAL LIST (6/8)
FRU
NO.
PART NAME PART NO QTY.
FRU DESCRIPTION
601 CABLES SET
602 HARDWARE SET
2235401
2235402
1
1
1
2
: V3.0
: V3.1
CABLES for Boards,
Power Switch and
Cable Tie.
Screws, Washers, Nuts,
Cable Tie, Screw Cap...
Include Washer
SECTION
FOR
REFERENCE
603 CABLE HOOK ARM SET
604 TOP COVER SHEET
605 FUSE SET
2170597
2158087
2249717--2
1
1
1
2
2
1 BPHV--1A250V,
0.5A250V
Monitor--3A 250V
6--
3
6--18 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 6
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
FRU
NO.
PART NAME
701 CBF PROBE
702 CAE PROBE
703 MTZ PROBE
704 CZB PROBE
705 LH PROBE
706 LE PROBE
707 LI PROBE
708 LT PROBE
709 LB PROBE
710 LD PROBE
711 CS PROBE
712 10Lb PROBE
713 ERB PROBE
714 3C PROBE
715 S317 PROBE
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
: V3.0
: V3.1
2152422
2152402
P9601AS
P9601AC
P9601AR
P9601AB
P9601AW
P9601AG
P9601AX
P9601AJ
P9601AQ
P9601AA
2124317
P9601AD
2202320
2202315
H45202LM
H45202ER
H45202WB
H45202SD
MATERIAL LIST (7/8)
PART NO
P9603AD
P9603AA
P9603AE
P9603AB
P9603AU
P9603AL
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
QTY.
FRU
1 1
1 1
1
1
1
1
1
1
1 1
DESCRIPTION
For Japan Only
For Japan Only
Not Include MTZ HOLDER
ASSY
Not Include MTZ HOLDER
ASSY
For Japan Only
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1 For Japan Only
For Japan Only
For Japan Only
For Japan Only
For Japan Only
For Japan Only
For Japan Only
For Japan Only
SECTION FOR
REFERENCE
6--19 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 3
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2233819 -- 2233826
2270964 -- 2270972, 2271424
LOGIQ 200 PRO Series
OPERATOR CONSOLE ASSY
MATERIAL LIST (8/8)
FRU
NO.
PART NAME
717 FOOT SWITCH ASSY
PART NO QTY.
FRU
2162242 1 2
718 RIGHT MTZ OPTION INSTALL KIT 2242590
719 B/W PRINTER INSTALL KIT 2242589
720 ECG INSTALL KIT 2242584
1
1
1
2
2
2
721 CINE S/W
722 INSITE INSTALL KIT
2242585
2242586
723 PROBE CABLE ARM INSTALL KIT 2215992
724 MOD INSTALL KIT 2235371
725 ATO S/W
726 PIOP2 ASSY
2251544
2181871
727 LEFT MTZ OPTION INSTALL KIT 2242591
728 MOD ASSY 2242409
729 VER 3.16 UPGRADE KIT 2273388
1
1
1
1
1
1
1
1
1
2
2
2
2
2
2
2
2
2
DESCRIPTION
FOR B/W PRINTER
Version between 3.10 to 3.15
: V3.0
: V3.1
REFEREN
CE
8--1
8--1
6--20 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2 DISASSEMBLY/RE--ASSEMBLY
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
WARNING!
ONLY QUALIFIED SERVICE PERSONNEL SHOULD REMOVE ANY COVERS OR PANELS.
ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE. BECOME THOROUGHLY
FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT LEVELS TO AVOID
ACCIDENTAL CONTACT
CAUTION
Do not wear the ESD wrist strap when you remove the SMPS Assy. Turn OFF power and unplug the power cord before removing any part of SMPS Assy. However be sure to turn off power and wear the strap before you remove a circuit boards.
CAUTION
Do NOT unplug the power cord before turning OFF the power switch.
6--21 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--1 Monitor Assy (FRU No. 100)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
WARNING!
PERSONAL INJURY HAZARD. VIDEO MONITOR SUPPORT ARM IS SPRING LOADED.
RELEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD
MOVEMENT. KEEP YOUR HEAD AND BODY AWAY FROM ABOVE VIDEO MONITOR
SUPPORT ARM.
Refer to ILLUSTRATION 6--7.
1.
Turn OFF the system.
2.
Lift and set the Monitor to the highest position by pushing down the Up/down Release Button located on the
Swing Arm Assy.
3.
Remove the Monitor Bottom Cover. Refer to 6--2--4.
4.
Disconnect the connector.
5.
Unscrew one screw (1).
6.
Pull the Monitor Assy upwards.
6--22 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--1 Monitor Assy (FRU No. 100) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
100 Monitor Assy connector
1
MONITOR ASSY DISASSEMBLY
ILLUSTRATION 6--7
6--23 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--2 Pot Set Assy (FRU No. 101)
Time Required
6 Minute
Tool Required
Screwdriver
Soldering Iron
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Procedure
Refer to ILLUSTRATION 6--8.
1.
Turn OFF the system.
2.
Remove the Monitor Cover Set. Refer to 6--2--4.
3.
Remove two Potentiometer Knob and unscrew two hexagonal nuts on each Pot.
4.
Unscrew four screws (1 -- 4) and remove the EMI Cover and PCB.
5.
Remove the wire from PCB with Soldering Iron.
6.
Remove the Pot Set Assy.
101 Pot Set Assy
Hexagonal Nut
Potentiometer Knob
POT SET DISASSEMBLY
ILLUSTRATION 6--8
6--24 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--3 CRT Filter(FRU No. 102)
Time Required
2 Minutes
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Not necessary
Procedure
Refer to ILLUSTRATION 6--9.
1.
Turn OFF the system.
2.
Push the two Filter Clamp Assy inward (1) as shown in ILLUSTRATION 6--9 and remove Filter Clamp Assy.
3.
Pull forward CRT Filter.
1
1
102 CRT Filter
105 Clamp Set
CRT FILTER DISASSEMBLY
ILLUSTRATION 6--9
6--25 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--4 Monitor Cover Set (FRU No. 103)
Time Required
4 Minutes
Tool Required
Screwdriver
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Procedure
Refer to ILLUSTRATION 6--10.
1.
Turn OFF the system.
2.
Lift and set the Monitor to the highest position by pushing down the Up/down Release Button located on the
Swing Arm Assy.
3.
Unscrew one screws (1) and remove the Monitor Bottom Cover.
4.
Unscrew four screws (2 -- 5) and remove the Monitor Rear Cover.
5.
Unscrew four screws (6 -- 9) and remove the Monitor Front Cover.
Monitor Rear Cover
2
4
3
5
8
Monitor Bottom Cover
9
Monitor Front Cover
6
1
7
MONITOR COVER ASSY DISASSEMBLY
ILLUSTRATION 6--10
103 Monitor Cover Set
6--26 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--5 Tilt Assy(FRU No. 104)
Time Required
8 Minutes
Tool Required
Screwdriver
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Procedure
Refer to ILLUSTRATION 6--11.
1.
Turn OFF the system.
2.
Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6--2--1.
3.
Remove the Monitor Rear Cover. Refer to 6--2--4.
4.
Unscrew four screws (1 -- 4).
5.
Remove the Tilt Assy.
104 Tilt Assy
3
4
1
2
TILT ASSY DISASSEMBLY
ILLUSTRATION 6--11
6--27 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--6 Filter Clamp Set (FRU No. 105)
Time Required
6 Minutes
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--9.
1.
Turn OFF the system.
2.
Push the two Filter Clamp Assy inward (1) as shown in ILLUSTRATION 6--12 and remove Filter Clamp Assy.
1
1
105 Filter Clamp Set
FILTER CLAMP SET DISASSEMBLY
ILLUSTRATION 6--12
6--28 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--7 Probe Holder (FRU No. 201)
Time Required
2 Minutes
Tool Required
Not necessary
Procedure
Refer to ILLUSTRATION 6--13.
1.
Turn OFF the system.
2.
Pull the PG rivet located at bottom of the Probe Holder.
3.
Pull the Probe Holder out as shown in ILLUSTRATION 6--13.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
201 Probe Holder
PG Rivet
PROBE HOLDER DISASSEMBLY
ILLUSTRATION 6--13
6--29 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--8 Holder Bracket Assy (FRU No. 202)
Time Required
3 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--14.
1.
Turn OFF the system.
2.
Remove the Probe Holder (FRU No. 201). Refer to 6--2--7.
3.
Unscrew three screws (1 -- 3).
4.
Remove the Holder Bracket Assy.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
202 Holder Bracket Assy
1
2
3
HOLDER BRACKET ASSY DISASSEMBLY
ILLUSTRATION 6--14
6--30 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--31 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--9 Cover Set (FRU No. 203)
Time Required
11 Minutes
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--15.
1.
Turn OFF the system.
2.
Remove the Probe Holder (FRU No. 201). Refer to 6--2--7.
3.
Remove the Holder Bracket Assy (FRU No.202) Refer to 6--2--8.
4.
Remove MOD Assy. (FRU No.723). Refer to Chapter 8.
5.
Unscrew the two screw(1 and 2) on the front bottom of Keyboard .
6.
Unscrew the ground screw ( ) and remove the Ground Terminal.
7.
Disconnect the connector and cut the tie wraps off.
8.
Remove the Keyboard Assy (FRU No. 204).
9.
Unscrew nine screws (3 -- 11) and remove the Keyboard assy.
6--32 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--9 Cover Set (FRU No. 203) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Connector
Cover Set
Ground Terminal
1
2
3
4
5
6
11
10
9
7
8
COVER SET DISASSEMBLY
ILLUSTRATION 6--15
6--33 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--10 Keyboard Assy (FRU No. 204)
Time Required
6 Minutes
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--16.
1.
Turn OFF the system.
2.
Remove MOD Assy (FRU.723). Refer to Chapter 8.
3.
Unscrew two screws (1 and 2) on the front bottom of Keyboard.
4.
Unscrew the ground screw ( ) and remove the Ground Terminal.
5.
Disconnect the connector and cut the tie wraps off.
6.
Pull the Keyboard Assy out.
6--34 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--10 Keyboard Assy (FRU No. 204) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
204 Keyboard Assy
Connector
Ground Terminal
1
2
KEYBOARD ASSY DISASSEMBLY
ILLUSTRATION 6--16
6--35 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--11 Trackball Assy (FRU No. 205)
Time Required
8 Minutes
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--17.
1.
Turn OFF the system.
2.
Pull the Keyboard Assy out.
Refer to 6--2--10.
3.
Unscrew three screws (1 -- 3) and disconnect one connector (CN5) on Trackball Assy.
4.
Remove Trackball Assy.
3
1
2
Connector
CN5
TRACKBALL ASSY DISASSEMBLY
ILLUSTRATION 6--17
6--36 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--12 TGC Assy (FRU No. 206)
Time Required
8 Minutes
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--18.
1.
Turn OFF the system.
2.
Pull the Keyboard Assy out . Refer to 6--2--10.
3.
Pull out the TGC Knobs (5 -- 12) on Keyboard Assy.
4.
Unscrew four screws (1 -- 4) and disconnect one connector (CN12).
5.
Remove TGC Assy.
1
2
3
4
Connector
CN12
5
6
7
8
9
11
12
10
TGC Knob
TGC ASSY DISASSEMBLY
ILLUSTRATION 6--18
6--37 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--13 Encoder Set (FRU No. 207)
Time Required
14 Minutes
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Screwdriver, Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6--19
1.
Turn OFF the system.
2.
Pull the Keyboard Assy out. Refer to 6--2--10
3.
Unscrew hexagonal screws (1) on each Encoder Knob and remove three knobs.
4.
Unscrew two screws on each Encoder Bracket and disconnect three connectors (CN1 --CN3).
5.
Remove Encoder Set.
Encoder Bracket
1
CN3
CN2
CN1
ENCODER SET DISASSEMBLY
ILLUSTRATION 6--19
6--38 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--14 Freeze/Record Key Assy (FRU No. 208)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--20.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1.
Turn OFF the system.
2.
Pull the Keyboard Assy out. Refer to 6--2--10.
3.
Unscrew three screws (1 -- 3) and disconnect the connector (CN11).
4.
Remove Freeze/Record Key Assy.
1
2
3
1
2
Connector
Connector
3
CN11
FREEZE/RECORD KEY ASSY DISASSEMBLY
ILLUSTRATION 6--20
6--39 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--15 TGC Knob Set (FRU No. 209)
Time Required
5 Minutes
Tool Required
None
Procedure
Refer to ILLUSTRATION 6--21.
1.
Turn OFF the system.
2.
Pull out the TGC Knobs (1 -- 8).
1 2
3 4
5 6
7 8
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
TGC KNOB SET DISASSEMBLY
ILLUSTRATION 6--21
6--40 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--41 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--16 Rubber Key Set (FRU No. 210)
Time Required
20 Minutes
Tool Required
Screw Driver.
Procedure
Refer to ILLUSTRATION 6--22.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1.
Turn OFF the system.
2.
Remove the Keyboard assy. Refer to 6--2--10.
3.
Pull out the TGC knobs. Refer to 6--2--15.
4.
Using the hexagonial wrench, remove the three Encoder knobs. Refer to 13.
5.
Remove the Freeze/Record key. Refer to 6--2--14.
6.
Disconnect five connectors (CN1, CN2, CN3, CN5, CN11, CN12).
7.
Unscrew 20 screws (1--20) and remove the board from the Keyboard assy. Refer to
ILLUSTRATION 6--22.
8.
Carefully remove 2 Rubber key pieces (1--2) from the Keyboard assy. Refer to
ILLUSTRATION 6--22.
6--42 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--16 Rubber Key Set (FRU No. 210) (continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
2
3
4 5
1
Board
6
CN12
17
7
9
8
CN3
10
CN11
20
CN2
CN1
13
Track
Ball
19
CN5
14
15
16
11
12
18
TGC
Board
Rubber Key 1
FRU 208.
RUBBER KEY SET DISASSEMBLY
ILLUSTRATION 6--22
6--43
Rubber Key 2
Track
Ball
TGC
RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--17 P.C.Board(s) (FRU No.301 through 304)
Time Required
7 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--23.
Refer to Proprietary Service Manual (2242594) for initialization.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
An electrostatic discharge may damage a component. Turn OFF power and wear the wrist strap before you remove a circuit boards. Do not unplug the power cord to keep ground continuity.
Do not bend or flex the boards when mounting/dismounting each board. Surface mount IC boards are very susceptible to damage from flex/torque.
Make sure the Dip switch position is set correctly before changing the MSTE Assy. Refer to
.
chapter 3, section 3--6--3 DIP Switch for correct setting.
Refer to ILLUSTRATION 6--23.
1.
Turn OFF the system.
2.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
3.
Remove the PCB Guide Assy (FRU 420). Refer to 6--2--43.
4.
Unscrew four screws (1 -- 4) and Remove the Bracket Nest.
5.
Disconnect the Cable Assy(5) from Board Assy.
6.
Pull the upper and lower ejector in the direction indicated by the arrows(a). Refer to ILLUSTRATION 6--23.
7.
Pull out the board, do not bend it (b).
6--44 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 2
6--2--17 P.C.Board(s) (FRU No.301 through 304) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374 a a b
Cable Assy(5)
Bracket Nest
1
2
3
4
Card Cage No.
3
4
1
2
Board
TBF
RBF
MVP
MSTE
P.C.BOARD(S) REMOVAL
ILLUSTRATION 6--23
CAUTION
In the case of the system for India, MSTE Dip switch SW2 No.4 should be on for A--Mode
Operation. Make sure the Dip switch position is set correctly before changing the MSTE
Assy. Refer to chapter 3, section 3--6--3 DIP Switch for correct setting.
6--45 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--18 BPHV Assy (FRU No. 305)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--24.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover (FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
5.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
6.
Unscrew four screws (1 -- 4), and remove top cover.
7.
Unscrew four screws (5 -- 8) being used fixing TR.
8.
Disconnect three connectors (J1 -- J3) on the BPHV Assy.
9.
Unscrew and remove four supporters(S1 -- S4).
10.
Remove the BPHV Assy.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--46 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--18 BPHV Assy (FRU No. 305) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
J2
1
J1
S1
S3
3
S2
J3
S4
2
4
BPHV ASSY DISASSEMBLY
ILLUSTRATION 6--24
5
6
7
8
6--47 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--19 SCAN Assy (FRU No. 306)
Time Required
13 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--25.
1.
Turn OFF the system and plug out.
2.
Remove the Shield Panel (FRU 307). Refer to 6--2--20.
3.
Unscrew four screws (1 -- 5).
4.
Disconnect two connector.
5.
Pull out the Probe SCAN Assy.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
306 SCAN Assy
Two Connectors
3
SCAN ASSY DISASSEMBLY
ILLUSTRATION 6--25
1
2
4
5
6--48 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--49 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--20 Shield Panel (FRU No. 307)
Time Required
11 Minutes
Tool Required
Screwdriver
Procedure
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6--26.
1.
Turn OFF the system & plug out
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover (FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
5.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
6.
Remove the Power S/W Assy (FRU 415). Refer to 6--2--38.
7.
Unscrew eleven screws (1 --11).
8.
Remove the Shield Panel (FRU 307).
6--50 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--20 Shield Panel (FRU No. 307) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
307 Shield Panel
9
10
8
7
11
6
5
4
1
2
3
SHIELD PANEL DISASSEMBLY
ILLUSTRATION 6--26
6--51 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--21 Mother 3 Assy (FRU No. 308)
Time Required
0.5 Hours
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
An electrostatic discharge may damage a component. Turn OFF power and wear the wrist strap before you remove a circuit boards. Do not unplug the power cord to keep ground continuity. Do not bend or flex the boards when mounting/dismounting each board.
Surface mount IC boards are very susceptible to damage from flex/torque.
Refer to ILLUSTRATION 6--27.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29
5.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
6.
Remove the EMI Cover L (FRU 419). Refer to 6--2--42.
7.
Remove the four PCB Guide Assy (FRU 420). Refer to 6--2--43.
8.
Remove the four P.C.Boards (FRU 301 to 304). Refer to 6--2--17.
9.
Remove the Shield Panel (FRU 307). Refer to 6--2--20.
10. Remove the Scan Assy (FRU306). Refer to 6--2--19.
11.
Disconnect 29 connectors connected with the mother board. (J1 -- J29)
12.
Unscrew two screws (a and b), and remove the Bracket Harness. If necessary, cut the tie wrap.
13.
Unscrew 25 screws (1 -- 25).
14.
Remove the Mother3 Assy (FRU No. 308).
6--52 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--21 Mother 3 Assy (FRU No. 308) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
308 Mother Assy
J4
1
J7
J10
8
J13
J16
J17
9 10
J20
J23
J26
19
15
J24
20
J18
2
J1
J2 J5
3
J8
J11
J14
4
11
J21
16
J27
21
22
J30
J31
5
J9
J15
12
J22
17
J28
23 a b
Bracket Harness
J3 J6
6
7
J12 J19
13
18
14
J25
24
J29
25
MOTHER3 ASSY DISASSEMBLY
ILLUSTRATION 6--27
CAUTION
Before connect the cables, Check and distinguish the labels especially for J12, J19, J3, J6.
6--53 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--22 Nest Box (FRU No. 309)
Time Required
0.6 Hour
Tool Required
Screwdriver
Hexagonal Wrench
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
An electrostatic discharge may damage a component. Turn OFF power and wear the wrist strap before you remove a circuit boards. Do not unplug the power cord to keep ground continuity.
Do not bend or flex the boards when mounting/dismounting each board. Surface mount IC boards are very susceptible to damage from flex/torque.
Procedure
Refer to ILLUSTRATION 6--28.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29
5.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
6.
Remove the Rear Cover (FRU 408). Refer to 6--2--31.
7.
Remove the EMI Cover L (FRU 419). Refer to 6--2--42.
8.
Remove the four PCB Guide Assy (FRU 420). Refer to 6--2--43.
9.
Remove the four P.C.Boards (FRU 301 to 304). Refer to 6--2--17.
10.
Remove the Shield Panel (FRU 307). Refer to 6--2--20.
6--54 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--2--22 Nest Box (FRU No. 309) (Continued)
11.
Remove the SCAN Assy (FRU 306). Refer to 6--2--19.
12.
Remove the Mother3 Assy (FRU 308). Refer to 6--2--21.
13.
Remove the eight Screws (1 -- 8) on both side of the Nest Box. Refer to ILLUSTRATION 6--28.
14.
Remove the Nest Box (FRU 309).
1
2
5 6
3
4
7
8
309 Nest Box
NEST BOX DISASSEMBLY
ILLUSTRATION 6--28
6--55 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--23 DC Fan Assy (FRU No. 310)
Time Required
45 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--29.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Top Cover (FRU 409). Refer to 6--2--32.
5.
Disconnect two connectors (2 x CN01).
6.
Unscrew eight screws (1-- 8) on the fan.
7.
Pull out the Fan Assy outwards.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--56 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--23 DC Fan Assy (FRU No. 310) (Continued)
1
2
3
4
5
6
7
8
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
310 DC Fan Assy
CN01
AC FAN ASSY DISASSEMBLY
ILLUSTRATION 6--29
6--57 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--24 Swing Arm Assy (FRU No. 401)
Time Required
0.6 Hour
Tool Required
Screwdriver
Hex Wrench
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Procedure
Refer to ILLUSTRATION 6--30.
1.
Turn OFF the system and unplug the power cord from outlet.
2.
Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6--2--1.
3.
Remove the Monitor Space Plate.
4.
Unscrew four screws (1 --4) and remove the Cover located at the bottom of the Swing Arm Assy.
5.
Unscrew one screw to remove Ground Terminal ( ).
6.
Unscrew two supporters ( 5 and 6 ) and two screws ( 7 and 8 ).
7.
Remove the Guide.
8.
Unscrew one screws (9).
9.
Remove the Swing Arm Assy (FRU No.401).
6--58 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--24 Swing Arm Assy (FRU No. 401) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
8
7
5
6
4
3
2
1
9
401 Swing Arm Assy
NECK ASSY DISASSEMBLY
ILLUSTRATION 6--30
6--59 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--25 Pipe Assy (FRU No. 402)
Time Required
0.4 Hour
Tool Required
Screwdriver
Hex Wrench
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Procedure
Refer to ILLUSTRATION 6--31.
1.
Turn OFF the system and unplug the power cord from outlet.
2.
Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6--2--1.
3.
Remove the Swing Arm Assy (FRU 401). Refer to 6--2--24.
4.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
5.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
6.
Remove the Rear Cover (FRU 408). Refer to 6--2--31.
7.
Remove the Pole Cover (FRU 403) Refer to 6--2--26.
8.
Remove eleven hexagonal bolts (1 -- 11).
9.
Disconnect two connectors (CN02,J3).
10.
Remove the Pipe Assy.
NOTE
When assembling the Tilt assy, fix two screw(1,7) with one drop of Loctite262.
6--60 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--25 Pipe Assy (FRU No. 402)) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
402 Pipe Assy
Two Connectors
CN02
J3
NECK ASSY DISASSEMBLY
ILLUSTRATION 6--31
4
5
6
1
2
3
9
10
7
8
11
6--61 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--26 Pole Cover (FRU No. 403)
Time Required
20 Minutes
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--32.
1.
Turn OFF the system.
2.
Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6--2--1.
3.
Remove the Swing Arm Assy (FRU 401). Refer to 6--2--24.
4.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
5.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
6.
Remove the Rear Cover (FRU 408). Refer to 6--2--31.
7.
Unscrew two screws (1and 2).
8.
Pull upward the Pole Cover.
6--62 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--26 Pole Cover (FRU No. 403)) (Continued)
403 Pole Cover
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1
2
POLE COVER DISASSEMBLY
ILLUSTRATION 6--32
6--63 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--27 Left Cover (FRU No. 404)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--33.
1.
Turn OFF the system.
2.
Remove four screw caps and unscrew four screws (1 -- 4).
3.
Remove the Left Cover (FRU 404).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
404 Left Cover
1
2
3
4
LEFT COVER DISASSEMBLY
ILLUSTRATION 6--33
6--64 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--28 Right Cover (FRU No. 405)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--34.
1.
Turn OFF the system.
2.
Remove the four screw caps and unscrew four screws (1 -- 4).
3.
Remove the Right Cover (FRU 405).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
405 Right Cover
1
2
RIGHT COVER DISASSEMBLY
ILLUSTRATION 6--34
3
6--65
4
RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--29 Front Base Cover (FRU No. 406)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--35.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Unscrew two screws (1 and 2).
5.
Remove the Front Base Cover (FRU 406).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1
406 Front Base Cover
2
FRONT BASE COVER DISASSEMBLY
ILLUSTRATION 6--35
6--66 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--67 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--30 Front Cover (FRU No. 407)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--36.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
5.
Remove the fore screw caps and unscrew fore screws (1 -- 4).
6.
Remove the Front Cover (FRU 407).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--68 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--30 Front Cover (FRU No. 407) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
407 Front Cover
2
3
1
4
FRONT COVER DISASSEMBLY
ILLUSTRATION 6--36
6--69 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--31 Rear Cover (FRU No. 408)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--37.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the four screw caps and unscrew four screws (1 -- 4).
5.
Remove the Rear Cover(FRU No. 408).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1
408 Rear Cover
2
3
4
REAR COVER DISASSEMBLY
ILLUSTRATION 6--37
6--70 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--71 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--32 Top Cover (FRU No. 409)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--38.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Rear Cover (FRU 408). Refer to 6--2--31.
5.
Pull up the Pole Cover (FRU 403). Refer to 6--2--26.
6.
Remove the Cable Hook Arm (1)
7.
Unscrew two screws (2 -- 3) and four bolts (4 --7).
8.
Remove the Top Cover.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--72 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--32 Top Cover (FRU No. 409) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1
2
3
4
5
6
7
409 Top Cover
REAR COVER DISASSEMBLY
ILLUSTRATION 6--38
6--73 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--33 Printer Cover (FRU No. 410)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--39.
1.
Turn OFF the system.
2.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
3.
Remove the Top Cover (FRU 409). Refer to 6--2--32.
4.
Unscrew the six screw (1 -- 6).
5.
Remove the Printer Cover (FRU No. 410).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1
2
3
4
5 6
410 Printer Cover
PRINTER COVER DISASSEMBLY
ILLUSTRATION 6--39
6--74 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--75 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--34 Printer Bracket Assy(FRU No. 411)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Procedure
Refer to ILLUSTRATION 6--40.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
5.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
6.
Remove the Top Cover (FRU 409). Refer to 6--2--32.
7.
Remove the Printer Cover (FRU 410). Refer to 6--2--33.
8.
Disconnect three cables connected to Back Bracket Assy as listed below. Cut the tie wraps off if necessary.
a. Unscrew and Disconnect two ring terminals (1 and 2) from the Terminal Block 6 Assy. Refer to 6--2--45,
ILLUSTRATION 6--51.
b. Unscrew and disconnect one ring terminals(Protective Earth) from the chassis.
c. Disconnect two connectors (J5 and J9) from the Mother Assy. Refer to 6--2--21, ILLUSTRATION 6--27.
9.
Remove four hexagonal bolts (1 -- 4) and pull out Printer Bracket Assy with Back Bracket Assy.
10.
Unscrew the four screw (5 -- 8).
11.
Disassemble Printer Bracket Assy with Back Bracket Assy (FRU411).
6--76 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--34 Printer Bracket Assy(FRU No. 411) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
5
6
Back Bracket Assy
411 Printer Bracket Assy
7
8
1
2
3
4
PRINTER BRACKET ASSY AND BACK BRACKET ASSY DISASSEMBLY
ILLUSTRATION 6--40
6--77 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--35 Rear Handle (FRU No. 412)
Time Required
1 Hours
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6--41.
1.
Turn OFF the system.
2.
Remove the Keyboard Assy (FRU 204). Refer to 6--2--10.
3.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
4.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
5.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
6.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
7.
Remove the Top Cover (FRU 409). Refer to 6--2--32.
8.
Remove the Printer Cover (FRU 410). Refer to 6--2--33.
9.
Remove the Neck Frame (FRU 413). Refer to 6--2--36.
10.
Remove four hexagonal bolts (1 -- 4).
11.
Remove the Rear Handle.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--78 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--35 Rear Handle (FRU No. 412) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
412 Rear Handle
1
3
2
4
REAR HANDLE DISASSEMBLY
ILLUSTRATION 6--41
6--79 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--36 Neck Frame (FRU No. 413)
Time Required
0.5 Hours
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6--42.
1.
Turn OFF the system.
2.
Remove the Keyboard Assy (FRU No. 204). Refer to 6--2--10.
3.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
4.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
5.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
6.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
7.
Remove the Top Cover (FRU 409). Refer to 6--2--32.
8.
Remove the Printer Cover (FRU 410). Refer to 6--2--33.
9.
Remove four hexagonal bolts (1 -- 4).
10.
Remove the Neck Frame (FRU No. 413).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--80 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--36 Neck Frame (FRU No. 413) (Continued)
1
2
3
4
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
413 Neck Frame
NECK FRAME DISASSEMBLY
ILLUSTRATION 6--42
6--81 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--37 Rear Panel Assy (FRU No. 414)
Time Required
0.5 Hour
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--43.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Rear Cover (FRU 408). Refer to 6--2--31.
5.
Remove the EMI Cover (FRU 419). Refer to 6--2--42.
6.
Disconnect all connectors out of the Rear Panel assy. If necessary, cut tie wraps. Connectors are listed below.
a. Unscrew and disconnect three ring terminals (1A, 2A and 5A) from the Terminal Block 6 Assy. Refer to
6--2--45, ILLUSTRATION 6--51.
b. Disconnect four connectors (J2,J3,J6 and J8) from the Mother Assy. For each cable, refer to 6--2--21,
ILLUSTRATION 6--27.
7.
Unscrew four screws (1 -- 4) and pull the Rear Panel Assy out.
8.
Remove the Rear Panel Assy.
6--82 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--37 Rear Panel Assy (FRU No. 414) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
414 Rear Panel Assy
1
2
3
4
REAR PANEL ASSY DISASSEMBLY
ILLUSTRATION 6--43
6--83 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--38 Power S/W Assy (FRU No. 415)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6--44.
1.
Turn OFF the system and unplug the power cord from the outlet.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
5.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
6.
Unscrew four screws (1 -- 4).
7.
Disconnect fore Terminals from the Power S/W Assy.
8.
Remove the Power S/W Assy.
6--84 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--38 Power S/W Assy (FRU No. 415) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1
3
2
4
415 Power S/W Assy
POWER S/W ASSY DISASSEMBLY
ILLUSTRATION 6--44
6--85 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--39 Front Caster (FRU No. 416)
Time Required
15 Minutes
Tool Required
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6--45.
1.
Turn OFF the system.
2.
Put a box below the Power Trans box.
3.
Unscrew four hexagonal bolts (1 -- 4) on each Front Caster.
4.
Remove the Front Caster.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
416 Front Caster
1
2
3
4
FRONT CASTER DISASSEMBLY
ILLUSTRATION 6--45
6--86 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--40 Rear Caster (FRU No. 417)
Time Required
15 Minutes
Tool Required
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6--46.
1.
Turn OFF the system.
2.
Put a box below the Power Trans box.
3.
Unscrew four hexagonal bolts (1 -- 4) on each Rear Caster.
4.
Remove the Rear Caster.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
417 Rear Caster
1
2
3
4
REAR CASTER DISASSEMBLY
ILLUSTRATION 6--46
6--87 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--41 Bumper Set (FRU No. 418)
Time Required
10 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6--47.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
5.
Unscrew three hexagonal bolts (1 -- 3) on each Bumper Set
6.
Remove the Bumper Set (FRU 418).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
418 Bumper Set
3
2
1
1 2
3
BUMPER SET DISASSEMBLY
ILLUSTRATION 6--47
6--88 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--42 EMI Cover L (FRU No. 419)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--48.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Unscrew eight screws (1 -- 8).
4.
Remove the EMI Cover L (FRU 419).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
419 EMI Cover L
3
1
2
5
4
7
8
6
EMI COVER L DISASSEMBLY
ILLUSTRATION 6--48
6--89 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--43 PCB Guide Assy (FRU No. 420)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6--34.
1.
Turn OFF the system.
2.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
3.
Unscrew four screws (1 -- 4).
4.
Remove the PCB Guide Assy (FRU 420).
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
420 PCB Guide Assy
PCB GUIDE DISASSEMBLY
ILLUSTRATION 6--49
1
4
2
3
6--90 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--91 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--44 Terminal Block 12 Assy (FRU No. 501)
Time Required
10 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6--50.
1.
Turn OFF the system and unplug the power cord from the outlet.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Unscrew the two screws (1 and 2) and pull out the Terminal Block 12 Assy.
4.
Unscrew and Disconnect twelve ring terminals (TB2--1 ~ TB2--12). Refer to NO TAG,Chapter 5. Cut the tie wrap, if necessary.
5.
Unscrew and Disconnect twelve ring terminals (TB2--1 ~ TB2--12).
6.
Remove the Terminal Block 12 Assy.
6--92 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--44 Terminal Block 12 Assy (FRU No. 501) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
501 Terminal Block 12 Assy
TB2--1
TERMINAL BLOCK 12 ASSY DISASSEMBLY
ILLUSTRATION 6--50
1
TB2--12
2
6--93 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--45 Terminal Block 6 Assy (FRU No. 502)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6--51.
1.
Turn OFF the system and unplug the power cord from the outlet.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Rear Cover (FRU 408). Refer to 6--2--31.
5.
Unscrew and Disconnect six ring terminals (TB3--1 ~TB3--6).
6.
Unscrew and Disconnect six ring terminals (TB3--1 ~TB3--6).
7.
Unscrew the two screws (1and 2) and pull out the Terminal Block 6 Assy.
8.
Remove the Terminal Block 6 Assy.
6--94 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--45 Terminal Block 6 Assy (FRU No. 502) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
502 Terminal Block 6 Assy
TB3--1
TB3--6
1
TB3--1
2
TB3--6
TERMINAL BLOCK 6 ASSY DISASSEMBLY
ILLUSTRATION 6--51
6--95 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--46 SMPS Assy (FRU No. 503)
Time Required
10 Minutes
Tool Required
Screwdriver
Hexagonal Wrench (Key Size: 4 mm)
Procedure
Refer to 6--4, SMPS Assy REPLACEMENT.
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--96 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--97 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--47 Power Trans Assy (FRU No. 504)
Time Required
20 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6--52.
1.
Turn OFF the system and unplug the power cord from the outlet.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
5.
Remove the Rear Cover (FRU 408). Refer to 6--2--31.
6.
Remove the SMPS Assy (FRU 503). Refer to Refer to 6--4, SMPS Assy REPLACEMENT..
7.
Disconnect sixteen cables listed below. Cut the tie wraps off if necessary. For each cable, refer to
ILLUSTRATION 6--52.
a. Unscrew and Disconnect twelve ring terminals (TB2--1 ~ TB2--12) from the Terminal Block 12 Assy. Refer to 6--2--44.
b. Unscrew and Disconnect two ring terminals. Refer to 6--2--45.
c. Disconnect two terminals from the Power S/W Assy.
8.
Unscrew two screws ( 1 and 2) of front side and two screws (3 and 4) of rear side of the system.
9.
Pull out the Power Trans Assy to the backward of the system.
6--98 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--47 Power Trans Assy (FRU No. 504) (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
TB2--12
TB2--1
504 Power Trans Assy
1
2
3
4
POWER TRANS ASSY DISASSEMBLY
ILLUSTRATION 6--52
6--99 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--2--48 Circuit Breaker Set (FRU No. 505)
Time Required
30 Minutes
Tool Required
Screwdriver
Procedure
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6--53.
1.
Turn OFF the system & unplug the power cord from the outlet.
2.
Remove the Power Trans Assy (FRU 504). Refer to 6--2--47.
3.
Unscrew four screws (1 -- 4)on the front of Power Trans Assy. Refer to ILLUSTRATION 6--53.
4.
Remove the defective Circuit Breaker.
1
3
4
2
CIRCUIT BREAKER DISASSEMBLY
ILLUSTRATION 6--53
6--100 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 6
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--3 FUSE REPLACEMENT
6--3--1 Introduction
LOGIQ 200 PRO Series has four replaceable fuses. The table below states the items of those fuses.
LOCATION
BPHV Assy
BPHV Assy
BPHV Assy
Monitor
TABLE 6-- 1
REPLACEABLE FUSES
Ref. No.
F1
F2
F3
F1
RATING
1 A, SB
500 mA, SB
500 mA, SB
3 A, SB
250 V
250 V
250 V
250 V
REMARKS
(SB ; Slow Blow type)
(SB ; Slow Blow type)
(SB ; Slow Blow type)
(SB ; Slow Blow type)
6--3--2 Replacement Procedures
Concerning replacing each fuse of LOGIQ 200 PRO Series, refer to the following steps;
Note
This system contains no operator serviceable parts inside. Please refer servicing to qualified personnel from GE. (See Address of Chapter 1.)
Parts Required
FRU
605
PART NAME
FUSE SET
Tools Required
Screwdriver
Hexagonal Wrench (Key Size: 3mm)
QTY
1
PART #
2249717--2
REMARKS
6--101 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
6--3--2 Replacement Procedures (Continued)
(A) F1, F2 and F3 on BPHV Assy
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Make sure that only fuse with the required rated current and of the same type is used for replacement.
1. Turn the system power OFF.and unplug the power cord from the outlet. Follow the OSHA Lockout/Tagout
Requirements (29 CFR 1910.147, Direction 46--015902).
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6--2--29.
5.
Remove the Front Cover (FRU 407). Refer to 6--2--30.
6.
Remove the BPHV top cover. Refer to 6--2--18.
7.
Replace the defective fuse(s) with new one(s). For the location of fuses, refer to ILLUSTRATION 6--54.
8.
Re--install the Covers.
F1 1A 250V
BPHV A’SSY
F3 0.5A 250V
F2 0.5A 250V
BPHV ASSY
ILLUSTRATION 6--54
6--102
Fuse Cap
F3
Fuse Socket
RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 6
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--3--2 Replacement Procedures (Continued)
(B) F1 on Monitor Assy
1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction
46--015902).
2.
Remove the Monitor Bottom Cover and Monitor Rear Cover. Refer to 6--2--4.
3.
Replace the defective fuse with new one. For the location of fuses, refer to ILLUSTRATION 6--55. Make sure that only fuse with the required rated current and of same type is used for replacement.
4.
Re--install the Monitor Bottom Cover and Monitor Rear Cover.
F1 3A 250V
MONITOR ASSY
ILLUSTRATION 6--55
6--103 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--4 SMPS Assy REPLACEMENT
6--4--1 Introduction
This section describes the procedures for replacing the SMPS Assy unit of LOGIQ 200 PRO Series.
6--4--2 Replacement Procedure
READ and UNDERSTAND these instructions thoroughly before proceeding with the replacement. Perform each step in sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer.
CAUTION
Possible operational damage. Failure to strictly follow ESD (Electrostatic Discharge) precautions during this upgrade may cause constant or intermittent operational abnormalities. Strictly follow all precautions.
Check each step as it is completed.
Parts Required
FRU
503
PART NAME
SMPS ASSY
QTY
1
PART #
2235398
REMARKS
Time Required
Approximately one (1) hours
Tools Required
Standard field Service Tool kit
(A) Removing Covers
1.
Turn the system power OFF and unplug the unit.
2.
Remove the Left Cover (FRU 404). Refer to 6--2--27.
3.
Remove the Right Cover(FRU 405). Refer to 6--2--28.
4.
Remove the Rear Cover (FRU 408). Refer to 6--2--31.
5.
Remove the EMI Cover L (FRU 419). Refer to 6--2--42.
6--104 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--4--2 Replacement Procedure (Continued)
(B) Disconnecting Cables
1.
Disconnect seven cables listed below. Cut the tie wraps off if necessary. For each cable, refer to
ILLUSTRATION 6--56.
a. Disconnect four connectors (J6, J12, J19, J25) from the Mother Assy.
b. Unscrew and Disconnect two ring terminals (TB3--3 and TB3--5) from the Terminal Block 6 Assy.
c. Unscrew one ground screw (1).
TB3--3
TB3--5
J12
J19
1
J25
J6
LEFT SIDE VIEW
ILLUSTRATION 6--56
6--105 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--4--2 Replacement Procedure (Continued)
(C) Replacing SMPS Assy
1.
Unscrew two hexagonal bolts (1 and 2). Two bolts are located at the front of SMPS Assy.
2.
Pull out the SMPS Assy forewards. Refer to ILLUSTRATION 6--57.
WARNING!
PERSONAL INJURY HAZARD! SMPS ASSY UNIT WEIGHTS APPROXIMATELY 3.5
KILOGRAMS (7.8 POUNDS). PAY ATTENTION NOT TO BE INJURED.
3.
Remove the SMPS Assy and place new one on the pallet.
4.
Slide and move the pallet to the position just under the console. Make sure that the SMPS Assy unit is placed in the correct direction.
5.
Screw two hexagonal screws to assemble new SMPS Assy unit to the console.
1
2
SMPS ASSY UNIT REMOVAL
ILLUSTRATION 6--57
6--106 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
6--4--2 Replacement Procedure (Continued)
(D) Final Procedure
1.
Connect four cables with the appropriate connectors. For the details of cables, refer to 6--4--2, (B),
Disconnecting Cables.
2.
Re--assemble the removal parts.
3.
Plug the system and power it ON.
4.
Perform all SMPS Assy checks and leakage current test.
5.
Perform functional checks to verify that the system is functioning properly. Refer to Chapter 4, FUNCTIONAL
CHECKS, in this manual.
6.
This completes the replacement of the SMPS Assy Unit.
CAUTION
Before connect the cables, Check and distinguish the labels especially for J12, J19, J3, J6.
6--107 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
6--108 RENEWAL PARTS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CHAPTER 7 -- PERIODIC MAINTENANCE
TABLE OF CONTENTS
TITLE SECTION
7--1
7--2
7--3
PAGE
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--1--1 Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3
7--3
PERIODIC MAINTENANCE PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--2--1 Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--2--2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--2--3 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--2--4 Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3
7--3
7--4
7--4
7--4
ELECTRICAL SAFETY TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--1 Outlet Test Wiring Arrangement
7--3--2 Grounding Continuity
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--3 Chassis Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--4 Probe Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--3--5 When There’s Too Much Leakage Current...
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7--5
7--5
7--6
7--7
7--10
7--14
7--1 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
7--2 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--1 INTRODUCTION
This chapter provides Periodic Maintenance procedures. It is an effective and efficient method of reducing equipment down time and service cost.
7--1--1 Periodic Maintenance
Periodic Maintenance consists of visual inspections, system performance checks, and periodic cleaning of certain critical parts. Perform this procedure at one year intervals. This schedule for maintenance is based on an estimated equipment usage of 40 hours per week.
7--2 PERIODIC MAINTENANCE PROCEDURE
7--2--1 Visual Inspection
1.
Check all cables for cracks nicks and abrasions in their insulation.
2.
Check that cable sheaths are securely clamped in cable ground/connector clamps where applicable.
3.
Check that cables are correctly routed and secured.
4.
Check that there are no missing screws or fasteners and replace if necessary.
5.
Check that all detachable cable connectors are properly seated in their sockets.
6.
Check that all equipment covers and fixtures are in place and undamaged.
7.
Check that all plug--in boards are properly seated in cardcage connector.
8.
Check that the keyboard keys, controls and Trackball function properly and that no mechanical damage is apparent.
7--3 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--2--2 Cleaning
1.
Clean CRT filter.
2.
Clean out the interior of the Console by using a vacuum cleaner.
7--2--3 Measurement
1.
Make sure that electrical outlet of hospital should provide the power line standard.
Refer to Chapter 2, Installation.
2.
Ensure all SMPS voltages are within tolerance.
Refer to section 4--3, SMPS ADJUSTMENT in Chapter 4, FUNCTIONAL CHECKS, for adjustment procedure.
3.
Make sure that leakage currents are within tolerance.
Refer to section 2--3, ELECTRICAL SAFETY TESTS in Chapter 2, INSTALLATION, for measurement procedures.
7--2--4 Note
Visual inspection of probes (“head”, cable and connector)
7--4 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--3 ELECTRICAL SAFETY TESTS
The following tests are performed at the factory and should be performed at the site. These tests are: grounding continuity, chassis leakage current, probe leakage current, and ECG leakage current. All measurements are made with an Electrical Safety Analyzer Model Dale 600/600E built by Dale Technology Corporation or equivalent device.
Recommended Tool
PART NUMBER
46--285652G1
46--328406G2
2113015
NAME DESCRIPTION
Electric Safety Analyzer DALE 600 For 110/115V Units
Electric Safety Analyzer DALE 600E For 220/240V Units
Leakage Current Ultrasound Kit For 100/115V and 220/240V Units
7--3--1 Outlet Test Wiring Arrangement
Test all outlets in the area for proper wiring arrangement by plugging in the Dale 600/600E and noting the combination of lights that are illuminated (Refer to ILLUSTRATION 7--1).
AC outlet
Combination of
Light
Condition CORRECT
WIRING
REVERSE
POLARITY
OPEN
GROUND
OUTLET TEST
ILLUSTRATION 7--1
OPEN
HOT
HOT/GROUND
REVERSED
Note
No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the
Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected.
7--5 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
7--3--2 Grounding Continuity
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CAUTION
Electric Shock Hazard. The patient must not be contacted to the equipment during this test.
Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case
(ILLUSTRATION 7--2). The ground wire resistance should be less than 0.2 ohms.
CONSOLE
GROUND PIN
ACCESSIBLE METAL PARTS
1. Monitor Housing
2. Rear Panel Connector
3. Any Caster/Wheel Support
OHMMETER
GROUND CONTINUITY TEST
ILLUSTRATION 7--2
Meter Procedure
Follow these steps to test the ground wire resistance.
1.
Turn the LOGIQ 200 PRO Series unit OFF.
2.
Plug the unit into the meter, and the meter into the tested AC wall outlet (Refer to ILLUSTRATION 7--2).
3.
Plug the black chassis cable into the meter’s ”CHASSIS” connector and attach the chassis cable clamp to an exposed metal part of the LOGIQ 200 PRO Series unit (Refer to ILLUSTRATION 7--2).
4.
Set the meter’s ”FUNCTION” switch to the RESISTANCE position.
5.
Set the meter’s ”POLARITY” switch to the OFF (center) position.
6.
Measure the ground wire resistance and keep a record of the results with other hard copies of PM data kept on site.
7--6 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--3--3 Chassis Leakage Current Test
Definition
Leakage current is the electrical current that could flow ground wire break. The unit, the probes, the ECG module, and all external peripherals must be tested.
Generic Procedure
The test verifies the isolation of the power line from the chassis. The testing meter is connected from accessible metal parts of the case to ground. Measurements should be made with the unit ON and OFF, with the power line polarity Normal and Reversed. Record the highest reading of current.
POWER
OUTLET
H (BLACK)
N (WHITE)
G (GREEN)
POLARITY REVERSING
SWITCH
MOMENTARY SWITCH
SETUP FOR CHASSIS SOURCE LEAKAGE CURRENT
ILLUSTRATION 7--3
LEAKAGE
TEST
METER
CONSOLE
ACCESSIBLE
METAL
PARTS
7--7 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--3--3 Chassis Leakage Current Test (Continued)
Meter Procedure
Follow these steps to test the unit for leakage current.
1.
Turn the LOGIQ 200 PRO Series unit OFF.
2.
Plug the unit into the meter, and the meter into the tested AC wall outlet (Refer to ILLUSTRATION 7--4).
Outlet Test
Lamps
Functions
Selector
Neutral
Open /closed
Normal
Reverse
Polarity
Power
On/Off
Chassis
Cable
Clamp
AC outlet
Power
Cable
GROUND AND CHASSIS LEAKAGE CURRENT TEST
ILLUSTRATION 7--4
3.
Plug the black chassis cable into the meter’s ”CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the LOGIQ 200 PRO Series unit (Refer to ILLUSTRATION 7--4).
4.
Set the tester’s ”FUNCTION” switch to CHASSIS position.
5.
Follow the test conditions described in the following table for every test point on the unit.
TABLE 7-- 1
CHASSIS LEAKAGE CURRENT TEST CONDITIONS
TEST
1
2
3
4
5
CONDITION
Mounting screw for transducer receptacle
Wheel support
Mounting screw for CRT housing
Mounting screw for peripheral plugged into unit
Mounting screw for other peripheral powered by unit
6.
Keep a record of the results with other hard copies of PM data kept on site.
7--8 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--3--3 Chassis Leakage Current Test (Continued)
Data Sheet for Chassis Source Leakage Current
Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the table below (TABLE 7-- 2).
TABLE 7-- 2
MAXIMUM ALLOWANCE LIMIT FOR CHASSIS SOURCE LEAKAGE CURRENT
COUNTRY
USA
OTHERS
NORMAL
CONDITION
N/A
0.1mA
OPEN
GROUND
0.3mA
0.5mA
REVERSE
POLARITY
0.3mA
0.5mA
OPEN
NEUTRAL
N/A
0.5mA
CAUTION
Electric Shock Hazard. When the meter’s ground switch is OPEN, don’t touch the unit!
Unit
Power
Tester Tester
POLARITY NEUTRAL
Switch
Test1
Probe or GROUND connector
Switch
Enter name of tested peripheral here:
ON
ON
ON
ON
OFF
OFF
OFF
OFF
NORM
NORM
REV
REV
NORM
NORM
REV
REV
TABLE 7-- 3
DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT
Test2 wheel
Test3
CRT
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
Test4 Test5
7--9 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--3--4 Probe Leakage Current Test
Definition
This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
Generic Procedure
Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst case condition.
POWER
OUTLET
H (BLACK) POLARITY REVERSING SWITCH
N (WHITE)
G (GREEN)
MOMENTARY
SWITCH
CONSOLE
LEAKAGE
TEST
METER
PROBE
SETUP FOR PROBE LEAKAGE CURRENT
ILLUSTRATION 7--5
Note
Each probe will have some amount of leakage current, dependent on its design. Small variations in probe leakage currents are normal from probe to probe. Other variations will result from differences in line voltage and test lead placement.
7--10 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
7--3--4 Probe Leakage Current Test (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Meter Procedure Using Probe Adapter
The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins.
The probe’s imaging area is immersed in a saline solution along with a grounding probe from the meter to complete the current path. Saline solution is a mixture of water and salt. The salt adds a free ion to the water, making it conductive. Normal saline solution is 0.9% salt or 1/2 gram salt per 1 liter of water. If saline is not available, a mixture of 1 quart water with one or more grams of table salt, mixed thoroughly, will substitute.
CAUTION
To avoid probe damage and possible electric shock, do not immerse probes into any liquid beyond the level indicated in the probe users manual. Do not touch the probe, conductive liquid or any part of the unit under test while the ISO TEST switch is depressed.
Follow these steps to test each transducer for leakage current.
1.
Turn the LOGIQ 200 PRO Series unit OFF.
2.
Plug the unit into the test meter, and the meter into the tested AC wall outlet.
3.
Connect the probe for test with the meter’s appropriate adapter (Refer to ILLUSTRATION 7--6).
4.
Plug the saline probe into the meter’s ”CHASSIS” connector (Refer to ILLUSTRATION 7--6).
Ultrasound
Probe
Saline
Probe
Probe Adapter with probe connected
Saline Solution
Nonconductive
Container
AC outlet
TRANSDUCER SOURCE LEAKAGE CURRENT TEST
ILLUSTRATION 7--6
7--11 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--3--4 Probe Leakage Current Test (Continued)
5.
Plug the probe adapter into the meter’s connector marked ”EXTERNAL” (Refer to ILLUSTRATION 7--6 ).
6.
Set the meter’s ”FUNCTION” switch to EXTERNAL position.
7.
Add the saline probe and the imaging area of the probe into the saline bath.
8.
Have unit power ON for the first part; turn it OFF for the second half.
9.
Depress the ISO TEST rocker switch and record the highest current reading.
10. Follow the test conditions described in the following table (TABLE 7-- 5) for every transducer.
11. Keep a record of the results with other hand copies of PM data.
Meter Procedure
Follow these steps to test each transducer for leakage current.
1.
Turn the LOGIQ 200 PRO Series unit OFF.
2.
Plug the unit into the test meter, and the meter into the tested AC wall outlet.
3.
Plug the external probe into the meter’s ”EXTERNAL” connector.
4.
Set the meter’s ”FUNCTION” switch to EXTERNAL position.
5.
Connect the probe for test with the connector of the console.
6.
Apply a liberal amount of gel to the transducer’s lens.
7.
Wrap the transducer in aluminum foil, making sure there is good contact with its acoustic aperture.
8.
Either suspend the transducer by its cable or place it on an insulated surface.
9.
Have unit power ON for the first part; turn it OFF for the second half.
10. Hold the external probe against the aluminum foil on the transducer when the unit is ready to scan.
11. Depress the ISO TEST rocker switch and record the highest current reading.
12. Follow the test conditions described in the following table (TABLE 7-- 5) for every transducer.
13. Keep a record of the results with other hand copies of PM data.
7--12 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
7--3--4 Probe Leakage Current Test (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
Data Sheet for Transducer Source Leakage Current
Follow the foregoing test procedure for every transducer. The test passes when all readings measure less than the value shown in the table below (TABLE 7-- 4).
TABLE 7-- 4
MAXIMUM ALLOWANCE LIMIT FOR PROBE LEAKAGE CURRENT
1.
Type BF Applied Part Leakage Current Limits
[Non--conductive (floating) Surface and Cavity Transducers]
COUNTRY
USA
OTHERS
NORMAL
CONDITION
0.05mA
0.1mA
OPEN
GROUND
0.05mA
0.5mA
REVERSE
POLARITY
0.05mA
0.5mA
OPEN
NEUTRAL
0.05mA
0.5mA
2.
Type CF Applied Part Leakage Current Limits
[Surgical Transducer and ECG Connections]
COUNTRY
USA
OTHERS
NORMAL
CONDITION
0.01mA
0.01mA
OPEN
GROUND
0.01mA
0.05mA
REVERSE
POLARITY
0.01mA
0.05mA
OPEN
NEUTRAL
0.01mA
0.05mA
Transducer tested:
TABLE 7-- 5
DATA SHEET FOR TRANSDUCER SOURCE LEAKAGE CURRENT
LOGIQ 200 PRO Series
Unit Power
ON
Tester Power
Polarity Switch
NORM
ON
ON
ON
NORM
REV
REV
OFF
OFF
OFF
OFF
NORM
NORM
REV
REV
Tester Ground or Neutral Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
Measurement
7--13 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
7--3--5 When There’s Too Much Leakage Current...
Chassis Fails
Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent. Replace any defective part.
Tighten all grounds. Ensure star washers are under all ground studs.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead.
Note
No outlet tester can detect the condition where the white neutral wire and the green grounding wire are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected.
Probe Fails
Test the probe in another transducer interface connector.
If excessive leakage current is slot dependent, inspect the Transducer Interface board for bent pins, poor connections, and ground continuity.
If the problem still has stayed with the probe, replace the probe.
Peripheral Fails
Tighten all grounds. Ensure star washers are under all ground studs.
Inspect wiring for bad crimps, poor connections, or damage.
Still Fails
If all else fails, begin isolation by removing the probes, external peripherals, then the on board ones, one at a time while monitoring the leakage current measurement.
New Unit
If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety
Failure Report to document the system problem. Remove unit from operation.
7--14 PERIODIC MAINTENANCE
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
CHAPTER 8 -- INSTALLATION FOR OPTIONS
TABLE OF CONTENTS
TITLE SECTION
8--1
8--2
PAGE
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--3
MOD ASSY INSTALLATION
8--2--1 INTRODUCTION
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--2 TOOLS REQUIRED
8--2--3 TIME REQUIRED
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--4 PARTS REQUIRED
8--2--5 Functional Check--out
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--6 Disassembling Dummy Bracket Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--7 Installing MOD Top Bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--8 Assembling MOD Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--9 Jumper Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--10 Installing MOD Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--11 Operational Check--out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--12 Final Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--2--13 Renewal Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8--6
8--7
8--8
8--8
8--10
8--11
8--12
8--4
8--4
8--4
8--4
8--4
8--4
8--5
8--1 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
This page is left blank intentionally.
8--2 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
8--1 INTRODUCTION
This chapter provides the installation procedures and renewal parts for optional devices. The installation procedures for the following optional devices are included in this chapter.
MOD Assy : FUJITSU MDD3064AP
Keep the option installation instructions on this chapter after completing installation for other options.
8--3 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
8--2 MOD ASSY INSTALLATION
LOGIQ 200 PRO Series SERVICE MANUAL
8--2--1 INTRODUCTION
These Section describes the installation of the optional MOD FUJITSU MDD3064AP for LOGIQ 200.
2235374
8--2--2 TOOLS REQUIRED
Standard field Service Tool kit
8--2--3 TIME REQUIRED
Approx. 0.5 Hour
8--2--4 PARTS REQUIRED
3.
4.
5
ITEM
1.
2.
PART NAME
MOD ASSY
MOD Top Bracket
MOD Bottom Bracket
SCREW M3X6
SCREW M4X10
1
4
4
QTY
1
1
PART #
2242409
2240642
2240643
2159624
2159632
REMARKS
8--2--5 Functional Check--out
READ and UNDERSTAND these instructions thoroughly before proceeding with the installation. Perform each step in sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer.
CAUTION
Possible operational damage. Failure to strictly follow ESD (Electrostatic Discharge) precautions during this upgrade may cause constant or intermittent operational abnormalities. Strictly follow all precautions.
Check each step as it is completed.
Perform functional checks to verify that the system is functioning properly. Refer to Functional Checks in Chapter
4.
8--4 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
8--2--6 Disassembling Dummy Bracket Assy
1.
Turn the system power OFF and unplug the unit.
2.
Unscrew four screws (1--4) and remove MOD dummy bracket. Refer to ILLUSTRATION 1--1.
Note
Keep removed screws (1--4) because they are required to mount the MOD Top Bracket.
1
MOD Dummy Bracket
2
3
4
MOD DUMMY BRACKET DISASSEMBLY
ILLUSTRATION 1--1
8--5 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
8--2--7 Installing MOD Top Bracket
1.
Screw 4 screws (1--4) and install MOD Top Bracket. Refer to ILLUSTRATION 1--2.
MOD TOP BRACKET ASSEMBLY
ILLUSTRATION 1--2
MOD Top Bracket
8--6 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
8--2--8 Assembling MOD Assy
1.
Assemble MOD Assy and MOD Bottom Bracket with four screws (1--4). Refer to ILLUSTRATION 1--3.
1
4
2
3
MOD ASSEMBLY
ILLUSTRATION 1--3
8--7 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
8--2--9 Jumper Setting
1.
Confirm that Jumper Setting is as illustrated in ILLUSTRATION 1--4.
Rear View
3
4
1
2
Default setting
JUMPER SETTING
ILLUSTRATION 1--4
8--2--10 Installing MOD Assy
1.
Insert MOD Assy with Bottom Bracket in to the MOD Top Bracket and screw 2 screws.
Refer to ILLUSTRATION 1--5.
2.
Connect two connectors (CN1, CN2) from the console to MOD Assy.
8--8 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
8--2--10 Installing MOD Assy (Continued)
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
1
CN1
CN2
1
2
INSTALLING MOD ASSY
ILLUSTRATION 1--5
2
8--9 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
8--2--11 Operational Check--out
1.
Turn the system power ON.
2.
Enter the service software after completing the power--up sequence. For the service software, refer to LO--
GIQ 200 PRO Series Proprietary Manual, 2242594. The Service Utility Menu:Top Menu will be displayed as shown in
ILLUSTRATION 4
. The password is required to enter. For the password, refer to the service note.
Service Utility MENU : Top Menu
Current Software Version is 3.0
99/03/12 09:16:37
System Diagnosis
Board/Program Version
System Reset
Key Operation History
* Option Set
Initialize :1 (1:America 2:Europe 3:Asia 4:Japan)
System Clock Adjust
Software/Probe data upgrade
Exit
SERVICE UTILITY MENU:TOP MENU
ILLUSTRATION 4
3.
Move the cursor to the item “Option Set” and press Set key. Option Set Menu will be displayed.
8--10 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 0
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
8--2--11 Operational Check--out (continued)
4.
Set among the optional items, MOD Driver to the “:Installed”. Refer to ILLUSTRATION 5.
Service UtilitY MENU : Option Set
Current Software Version is 3.0
99/03/12 09:16:37
CINE
ATO
* MOD Driver
Biopsy Guideline
:1 (1:Not Install 2:Install)
:1 (1:Not Install 2:Install)
:2 (1:Not Install 2:Install)
:1 (1:Not Install 2:Install)
MGOB Calculation :1 (1:Not Install 2:Install)
Cardiac Calculation :1 (1:Not Install 2:Install)
Urology Calculation :l (1:Not Install 2:Install)
Serial Printer
Paralled Printer
InSite
Exit
:l (1:Not Install 2:Install)
:1 (1:Not Install 2:Install)
Data Management Center:l (1:Not Install 2:Install)
:1 (1:Not Install 2:Install)
OPTION SET MENU
ILLUSTRATION 5
5.
6.
Note
To activate the options, move the cursor to the item, type 2, and press the Set key.
Exit the Option Set Menu and Power OFF the system.
Turn the system power ON.
8--2--12 Final Procedure
1.
Properly dispose of excess material.
2.
This completes the installation of the LOGIQ 200 PRO Series MOD Option.
8--11 INSTALLATION FOR OPTIONS
GE MEDICAL SYSTEMS
REV 6
8--2--13 Renewal Parts
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
3.
4.
5
ITEM
1.
2.
PART NAME
MOD ASSY
MOD Top Bracket
MOD Bottom Bracket
SCREW M3X6
SCREW M4X10
MATERIAL LIST (1/1)
1
4
4
QTY
1
1
PART #
2295442
2240642
2240643
2299256
2159632
REMARKS
1. MOD ASSY
2. MOD Top Bracket
3. MOD Bottom Bracket
RENEWAL PARTS
ILLUSTRATION 1--6
8--12 INSTALLATION FOR OPTIONS
advertisement