Creative Medical PC-900 User and installer manual

Creative Medical PC-900 User and installer manual

Creative Medical PC-900 is a powerful and easy-to-use vital signs monitor that provides accurate and reliable measurements of vital parameters. It is ideal for use in a variety of clinical settings, including hospitals, clinics, and home healthcare. The PC-900 features a large, easy-to-read display, multiple monitoring modes, and a variety of alarms and alerts to ensure patient safety. With its advanced features and user-friendly design, the PC-900 is the perfect choice for monitoring vital signs in any clinical setting.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Creative Medical PC-900 User and installer manual | Manualzz

Vital Signs Monitor

PC-900

User Manual

User Manual for Vital Signs Monitor

General requirements for safety)and MDD 93/42/EEC. It complies with both international and enterprise standards and is also approved by State Technological Supervision Bureau. The Manual is written for the current PC-900 Vital Signs Monitor.

The Manual describes, in accordance with the Vital Signs Monitor’s features and requirements, main structure, functions, specifications, correct methods for transportation, installation, usage, operation, repair, maintenance and storage, etc. as well as the safety procedures to protect both the user and equipment. Refer to the respective chapters for details.

The Manual is published in English and we have the ultimate right to explain the Manual. No part of this manual may be photocopied, reproduced or translated into another language without the prior written consent.

We reserve the right to improve and amend it at any time without prior notice. Amendments will however be published in a new edition of this manual.

Version of This Manual: V1.5

September 13, 2012

All rights reserved.

Marks in the Manual:

Warning: must be followed to avoid endangering the operator and the patient.

Attention: must be followed to avoid causing damage to the monitor.

Note: some important information and tips about operations and application.

3502-2530003

I

User Manual for Vital Signs Monitor

Instructions to User

Dear Users,

Thank you very much for purchasing our product. Please read the following information very carefully before using this device.

Read these instructions carefully before using this monitor. These instructions describe the operating procedures to be followed strictly. Failure to follow these instructions can cause monitoring abnormity, equipment damage and personal injury. The manufacturer is NOT responsible for the safety, reliability and performance issues and any monitoring abnormality, personal injury and equipment damage due to user’s negligence of the operation instructions. The manufacturer’s warranty service does not cover such faults.

WARNING-PACEMAKER PATIENTS. This monitor may continue to count the

pacemaker rate during occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon this monitor ALARMS. Keep pacemaker patients under close surveillance.

Monitoring a single person at a time.

The monitor is defibrillator proof. Verify that the accessories can function

safely and normally and the monitor is grounded properly before conducting defibrillation.

Disconnect the monitor and sensors before MRI scanning. Use during MRI

could cause burn or adversely affect the MRI image or the monitor’s accuracy.

If you have any doubt to the grounding layout and its performance, you must

use the built-in battery to power the monitor.

All combinations of equipment must be in compliance with standard of IEC

60601-1-1 for medical electric system requirements.

Check SpO

2

probe application site periodically (every 30 minutes) to determine circulation, positioning and skin sensitivity.

The SpO

2

measurement of this monitor may not work for all testees. If stable readings can not be obtained at any time, discontinue to use.

Do not immerse the monitor or its accessories in liquid to clean.

Do not use accessories other than those provided/recommended by the

manufacturer.

Each time the monitor is used, check the alarm limits to ensure that they are

appropriate for the patient being monitored.

The monitor is intended only as an adjunct in patient assessment. It must be

used in conjunction with clinical signs and symptoms.

When taking the measure of an infant or neonate’s (less than 10 years old) blood pressure,

do NOT operate in the adult mode. The high inflation pressure may cause lesion or even body putrescence.

II

User Manual for Vital Signs Monitor

The monitor is prohibited from applying to those who have severe hemorrhagic tendency or

who are with sickle cell disease for they may develop partial bleeding when this monitor is used to take the blood pressure measurement.

DO NOT take blood pressure measurement from a limb receiving ongoing transfusion or

intubations or skin lesion area, otherwise, damages may be caused to the limb.

Continuous use of SpO

2

sensor may result in discomfort or pain, especially for those with microcirculatory problem. It is recommended that the sensor should NOT be applied to the same place for over two hours, change the measuring site periodically if necessary.

SpO

2 measuring position must be examined more carefully for some special patient. Do

NOT install the SpO

2

sensor on the finger with edema or vulnerable tissue.

To prevent the risk of the short circuit and to ensure the ECG signal quality, the

equipment must be properly grounded.

Although biocompatibility tests have been performed on all the applied parts, some

exceptional allergic patients may still have anaphylaxis. Do NOT apply to those who have anaphylaxis.

All the connecting cables and rubber tubes of the applying parts should be kept away from

the patient’ s cervix to prevent any possible suffocation of the patient.

All the parts of the monitor should NOT be replaced at will. If necessary, please use the

components provided by the manufacturer or those that are of the same model and standards as the accessories along with the monitor which are provided by the same factory, otherwise, negative effects concerning safety and biocompatibility etc. may be caused.

DO NOT stare at the infrared light of SpO

2 sensor when switch it on, for the infrared may do harm to the eye.

If the monitor falls off accidentally, please do NOT operate it before its safety and

technical indexes have been tested minutely and positive testing results obtained.

It is recommended to take the blood pressure measurement manually. The automatic or

continuous mode should be used at the presence of a doctor/nurse.

Please peruse the relative content about the clinical restrictions and contraindication.

When disposing of the monitor and its accessories, the local law should be followed.

III

User Manual for Vital Signs Monitor

Table of Contents

CHAPTER 1 OVERVIEW ............................................................................................................ 1

1.1

F

EATURES

..................................................................................................................................................... 1

1.2

P

RODUCT

N

AME AND

M

ODEL

...................................................................................................................... 1

1.3

I

NTENDED

U

SE

............................................................................................................................................. 1

1.4

C

ONFORMATION

........................................................................................................................................... 1

1.5

S

YMBOLS ON THE

M

ONITOR

......................................................................................................................... 2

1.6

S

AFETY

......................................................................................................................................................... 2

CHAPTER 2 OPERATING PRINCIPLE ................................................................................... 3

2.1

O

VERALL

S

TRUCTURE

.................................................................................................................................. 3

2.2

C

ONFORMATION

........................................................................................................................................... 3

CHAPTER 3 INSTALLATION AND CONNECTION .............................................................. 4

3.1

A

PPEARANCE

................................................................................................................................................ 4

3.1.1 Front Panel............................................................................................................................................ 4

3.1.2 Side Panel ............................................................................................................................................. 6

3.1.3 Rear Panel ............................................................................................................................................. 7

3.2

I

NSTALLATION

.............................................................................................................................................. 8

3.2.1 Opening the Package and Check .......................................................................................................... 8

3.2.2 Connecting the Power Supply .............................................................................................................. 8

3.2.3 Starting the Monitor .............................................................................................................................. 8

3.3

C

ONNECTION

................................................................................................................................................ 9

3.3.1 ECG Cable Connection ........................................................................................................................ 9

3.3.2 Blood Pressure Cuff Connection ........................................................................................................ 11

3.3.3 SpO

2

Sensor Connection .................................................................................................................... 14

3.3.4 TEMP Transducer Connection ............................................................................................................ 16

3.3.5 Loading printing paper ....................................................................................................................... 16

3.3.6 Battery Installation ............................................................................................................................. 19

CHAPTER 4 OPERATIONS ...................................................................................................... 20

4.1

I

NITIAL

M

ONITORING

S

CREEN

................................................................................................................... 20

4.1.1 Default Display Screen Description ................................................................................................... 21

4.1.2 Operation Instructions ........................................................................................................................ 21

4.2

S

P

O

2

M

ONITORING

S

CREEN

....................................................................................................................... 23

4.2.1 Screen Description .............................................................................................................................. 23

4.2.2 Operation Instructions ........................................................................................................................ 24

4.3

T

REND

G

RAPH

D

ISPLAY

............................................................................................................................. 24

4.3.1 Screen Description .............................................................................................................................. 24

4.3.2 Operation Instructions ........................................................................................................................ 25

IV

User Manual for Vital Signs Monitor

4.4

NIBP L

IST

S

CREEN

..................................................................................................................................... 25

4.4.1 Operation Instructions ........................................................................................................................ 26

4.5

S

ETUP

M

ENU

S

CREEN

................................................................................................................................ 26

4.5.1 ECG and Temperature Setup .............................................................................................................. 27

4.5.2 SpO

2

Setup.......................................................................................................................................... 28

4.5.3 NIBP Setup ......................................................................................................................................... 29

4.5.4 Nurse Call ........................................................................................................................................... 31

4.5.5 System Setup ...................................................................................................................................... 32

4.5.6 Patient Info ......................................................................................................................................... 33

4.5.7 Date/Time ........................................................................................................................................... 33

4.5.8 Recover Default Settings .................................................................................................................... 33

4.6

P

OWER

S

AVING

M

ODE

............................................................................................................................... 33

CHAPTER 5 ALARM ................................................................................................................. 35

5.1

A

LARM

P

RIORITY

....................................................................................................................................... 35

5.2

A

LARM MODES

........................................................................................................................................... 35

5.3

A

LARM

S

ILENCE

......................................................................................................................................... 36

5.4

A

LARM

S

ETTING

........................................................................................................................................ 36

5.5

V

ERIFY

A

DJUSTABLE

A

LARM

F

UNCTION

................................................................................................... 36

CHAPTER 6 TECHNICAL SPECIFICATIONS ..................................................................... 38

6.1

ECG M

ONITORING

..................................................................................................................................... 38

6.2

TEMP M

ONITORING

................................................................................................................................... 39

6.3

NIBP M

ONITORING

.................................................................................................................................... 39

6.4

S

P

O

2

M

ONITORING

..................................................................................................................................... 40

6.5

P

ULSE

M

ONITORING

................................................................................................................................... 40

6.6

D

ATA

R

ECORDING

...................................................................................................................................... 40

6.7

O

THER

T

ECHNICAL

S

PECIFICATIONS

.......................................................................................................... 40

6.8

O

PERATING

E

NVIRONMENT

........................................................................................................................ 41

6.9

C

LASSIFICATION

......................................................................................................................................... 41

6.10

G

UIDANCE AND

M

ANUFACTURER

S

D

ECLARATION

-E

LECTROMAGNETIC

C

OMPATIBILITY

...................... 42

CHAPTER 7 PACKAGING AND ACCESSORIES ................................................................. 47

7.1

P

ACKAGING

................................................................................................................................................ 47

7.2

A

CCESSORIES

............................................................................................................................................. 47

CHAPTER 8 MONITORING PARAMETER .......................................................................... 48

8.1

ECG M

ONITORING

..................................................................................................................................... 48

8.1.1 How to Obtain High Quality ECG and Accurate Heart Rate Value .................................................... 48

8.1.2 Factors affecting ECG signal .............................................................................................................. 49

8.2

NIBP M

ONITORING

.................................................................................................................................... 49

8.2.1 Measuring Principle ............................................................................................................................ 49

8.2.2 Factors affecting NIBP measuring ...................................................................................................... 50

V

User Manual for Vital Signs Monitor

8.2.3 Clinical Limitations ............................................................................................................................ 51

8.3

S

P

O

2

M

ONITORING

..................................................................................................................................... 51

8.3.1 Measuring Principle ............................................................................................................................ 51

8.3.2 SpO

2

Measurement Restrictions (interference reason) ....................................................................... 52

8.3.3 Low SpO

2

measuring value caused by pathology reason ................................................................... 52

8.3.4 Clinical Limitations ............................................................................................................................ 52

8.3.5 Points to be noted in SpO

2

and Pulse Measuring................................................................................ 53

8.4

T

EMPERATURE

M

ONITORING

..................................................................................................................... 53

CHAPTER 9 TROUBLESHOOTING ....................................................................................... 55

9.1

N

O

D

ISPLAY ON THE

S

CREEN

..................................................................................................................... 55

9.2

E

XCESSIVE

ECG S

IGNAL

I

NTERFERENCE OR

T

OO

T

HICK

B

ASELINE

.......................................................... 55

9.3

N

O

B

LOOD

P

RESSURE AND

P

ULSE

O

XYGEN

M

EASURES

............................................................................ 55

9.4

B

LANK

P

RINTING

P

APER

............................................................................................................................ 55

9.5

S

YSTEM

A

LARM

......................................................................................................................................... 55

CHAPTER 10 MAINTENANCE................................................................................................ 56

10.1

S

ERVICE AND

E

XAMINATION

.................................................................................................................... 56

10.1.1 Daily Examination ............................................................................................................................ 56

10.1.2 Routine Maintenance ........................................................................................................................ 56

10.1.3 Battery Maintenance ......................................................................................................................... 56

10.1.4 Service .............................................................................................................................................. 57

10.2

C

LEANING

, S

TERILIZATION AND

D

ISINFECTION

........................................................................................ 57

10.3

C

LEANING

, S

TERILIZATION AND

D

ISINFECTION OF

A

CCESSORIES

............................................................ 58

10.4

S

TORAGE

.................................................................................................................................................. 58

10.5

T

RANSPORTATION

..................................................................................................................................... 58

CHAPTER 11 APPENDIX .......................................................................................................... 59

11.1

P

ROMPT INFORMATION EXPLANATIONS

.................................................................................................... 59

11.2

D

EFAULT

A

LARMING

V

ALUES AND

S

ETUP

R

ANGE

.................................................................................... 60

11.3

A

BBREVIATION OF ARRHYTHMIA

.............................................................................................................. 61

11.4

A

CCESSORIES

L

IST

................................................................................................................................... 62

11.5

I

NSTRUCTIONS FOR

S

P

O

2

P

ROBE

.............................................................................................................. 63

VI

Chapter 1 Overview

1.1 Features

 Blood Pressure, SpO

2 and Pulse Rate are displayed by big, bright digital LEDs;

 ECG waveform, SpO

2 plethysmogram and system parameters are displayed on dot matrix LCD screen;

 Accurate NIBP measurement

with hardware and software over-pressure protection;

 Unique SpO

2 measuring technique ensures sensitive and accurate SpO

2

, Pulse Rate and

Perfusion Index measurement;

 HR and SpO

2 trend curve display for last 12, 24 or 96 hours;

 Up to 12000 groups of NIBP measurements can be stored and reviewed by list;

 Audible & visible alarm with 3 levels of alarm events;

 Nurse call output is available;

 With tourniquet function;

 NIBP measurement is applicable to adult, pediatric and neonate by patient selection;

 Built-in printer is optional to print out waveforms, and text information.

1.2 Product Name and Model

Name: Vital Signs Monitor

Model: PC-900

1.3 Intended Use

This Monitor is a multi-functional instrument designed for monitoring the vital physiological signs of adult and pediatric patients. With the functions of real-time recording and displaying parameters, such as ECG, heart rate, non-invasive blood pressure, functional oxygen saturation, body temperature, and so on, it allows comprehensive analysis of patient’s physiological conditions.

This instrument is applicable for use in hospitals and clinical institutions. The operation should be performed by qualified professionals only.

1.4 Conformation

PC-900

Vital Signs Monitor is module designed product, it consists of

the main control unit,

ECG/TEMP module (optional), NIBP module (optional), SpO

2 measuring

module (optional), printer module

(optional), display panel, and power supply block etc.

and the related accessories for ECG, NIBP, SpO

2

and

Temperature measurement.

☞ According to different needs, you can customize the module configuration by choosing necessary modules. Therefore

, your monitor may not have all the monitoring functions and accessories.

User Manual for Vital Signs Monitor

1.5 Symbols on the Monitor

Adult Patient

Pediatric Patient

Waveform Freeze

Pulse sync indicator

Neonatal Patient Setup Menu

NIBP Start/Cancel

AC Power

Alarm Silence DC Power

Up

Print Type BF applied part

Type CF applied part with defibrillator protection

Warning, refer to User Manual.

OK

Down

ECG Lead Selection

Equal potential terminal

Nurse call output

1.6 Safety

a) This device conforms to IEC60601-1, electric safety classification: Class I, with Type BF and CF applied parts. b) This device can resist against the discharge of defibrillator and the interference of eletro-surgical unit. c) This device can monitor the patients with pace-maker. d) DO NOT use this device while the patient is under MRI scanning.

2

User Manual for Vital Signs Monitor

Chapter 2 Operating Principle

2.1 Overall Structure

The overall structure of the monitor is shown in Fig. 2.1.

Figure 2.1

2.2 Conformation

PC-900 Vital Signs Monitor is module designed product; it consists of ECG/TEMP module (optional), NIBP module (optional), SpO

2

module (optional), main control unit, printer module (optional), display panel, and power supply block etc. and the related accessories for ECG, NIBP, SpO

2

and Temperature measurement.

☞ According to different needs, you can customize the module configuration by choosing necessary modules. Therefore, your monitor may not have all the monitoring functions and accessories.

1. ECG/TEMP module measures ECG signal and detects heart rate with ECG lead wires and electrodes, it also measures temperature with temperature probe.

2. The SpO

2

module detects and calculates pulse rate and oxygen saturation (SpO

2

), and provides plethysmogram and perfusion index as well.

3. The NIBP module performs the measurement of blood pressure by non-invasive way of oscillometric technology, including the diastolic, systolic and mean arterial pressure. The cuffs are designed for adult, pediatric and neonate respectively.

4. The main control unit is in charge of LED and LCD display, keyboard input, data storage, printing, and networking function. Vital Signs Monitor is module designed product; it consists of ECG/TEMP module (optional), NIBP module (optional), SpO

2

module (optional), main control unit, printer

3

User Manual for Vital Signs Monitor module (optional), display panel, and power supply block etc.

Chapter 3 Installation and Connection

3.1 Appearance

3.1.1 Front Panel

Figure 3.1 Front panel illustration

Description:

1

Alarm indicator

Indicator

Red flashing

Alarm Level

High priority alarm

Yellow flashing Medium priority alarm

Alarm Event

Exceeding the limits, low battery voltage

Leads or probe off

4

User Manual for Vital Signs Monitor

Green light Normal

2

SYS: display systolic pressure value

3

DIA: display diastolic pressure value.

4

MAP: Display mean arterial pressure or measuring time of the latest group of NIBP measurement; they will be displayed alternately. The format of NIBP measuring time is “hh:mm”. If the tourniquet is in use, the cuff pressure will be displayed here.

Note: two formats to display NIBP value: “×××mmHg” and “××.×kPa”. Refer to section “4.4.2 NIBP

Setup” to set the unit of NIBP value; the conversion relation between “mmHg” and “kPa”:

1mmHg=0.133kPa.

5

HR(priority indicator): if HR indicator is on, it indicates that the numerical value beside is HR measuring value;

6

Display HR or PR value: when the set of “Setup Menu→ System→priority” is “HR”, it shows HR value here preferentially; if the set is “PR”, PR value will be shown preferentially.

7

PR (priority indicator): if PR indicator is on, it indicates that the numerical value beside is pulse rate value; Unit: “bpm (beats per minute)”.

8 SpO

2

: Display SpO

2

value; Unit: “%”

9 “

”: Bar-graph of pulse intensity

10 LCD panel

11 Pulse sync indicator patient category indicator: “ ” for adult; “ ” for pediatric; “ neonate; Patient category is selected under sub-menu “Patient Info” within the setup menu.

” for

12 Pulse sync indicator: Cardio-pulse/pulse sync indicator. When HR priority indicator is on, its flashing is synchronized with heart beat; When PR priority indicator is on, its flashing is synchronized with pulse.

13

NIBP: start/cancel NIBP measurement

14

Alarm silence key: Enable/disable alarm silence function. When the alarm silence indicator on the left of keys is on, it means the system is in alarm silence status and it lasts this status for 2 minutes. When finishing counting down, the system will resume normal alarm status automatically, if alarm event occurs at this time the alarm sound will be effective again.

DO NOT silence the audible alarm or decrease its volume or patient safety

could be compromised.

15 Alarm silence indicator: When it is on, it indicates that the monitor stays in alarm silence status.

16

Print: the internal printer is optional, press this key to print the current measuring data;

17  Up: shift cursor forward/upward

5

User Manual for Vital Signs Monitor

18 ▇ OK: to confirm selection or modification

19  Down: shift cursor backward/downward

20

Display: shift LCD display modes

21

Waveform Freeze: freeze the current displayed waveform.

22

ECG Lead Selection: select ECG leads among I, II, III, aVR, aVL, aVF and V.

23 : AC Power indicator

24

: DC Power indicator

Status

AC Power indicator DC Power indicator

ON (green)

ON (green)

OFF

OFF

ON (green)

Descriptions

ON (green) this device is using mains power supply

ON (orange) this device is using mains power supply and the battery is being recharged. the battery is being used

ON (green)

ON (orange, blinking)

OFF the battery is being used, but battery voltage is low, the beeper also gives warning. the battery is being recharged while the device is off

25

: Power button: Press power button for 3 seconds to start the monitor or shut off the monitor.

Note: Short time pressing power button for entering the Power Saving Mode screen, then according to your need to make the device stay in the power saving mode or exit from power saving mode (this function is optional and needs hardware support).

26 SpO

2

: SpO

2

sensor connector

27 NIBP: NIBP hose connector

28 TEMP: TEMP probe connector

29 ECG: ECG cable connector

3.1.2 Side Panel

The built-in thermal printer is in the left panel. It is easy for user to print waveform and data.

6

3.1.3 Rear Panel

User Manual for Vital Signs Monitor

Introduction to the rear panel:

1 Handle

2 Fan

3 Nameplate

CE mark

Figure 3.2 Rear Panel

Serial number

Date of manufacture

Authorised representative in the European community

Manufacturer (including address and date)

Disposal of this device according to WEEE regulations

4 “FUSE T3.15 A”: Fuse holder. Fuse specification: T3.15AL/250V Φ520mm.

5 “AC100~240V” : AC power supply socket

6 Loudspeaker

7 Mounting hole for hanging the monitor

8 NET: serial communication port which is used to network with central monitoring system (optional);

9 Nurse-call connector

10 : Equipotential ground terminal

7

User Manual for Vital Signs Monitor

3.2 Installation

3.2.1 Opening the Package and Check

1. Open the package, take out the monitor accessories from the box carefully and place it in a safe stable and easy to watch position.

2. Open the accompanying document to sort the accessories according to the packing list.

 Inspect the monitor for any mechanical damages

 Check all the accessories for any scratch or deformity, especially on connector, wire and probe parts

☞ You can customize the module configuration by choosing necessary modules to meet your own needs. Therefore, your monitor may not have all the monitoring functions and accessories.

If in doubt, please contact the local dealer or our company in case of any problems. We are to offer you the best solution for your satisfaction.

3.2.2 Connecting the Power Supply

1. When powered by AC mains power supply:

 Make sure that the AC power supply is 100-240VAC, 50/60Hz.

 Use the power cable prepared by the manufacturer. Insert one end of it to the power port of the monitor and the other end to the grounded three-phase power jack.

 To eliminate potential differences, the monitor has a separate connection to the equipotential grounding system. Connect one end of the provided ground cable to equipotential grounding port on the rear of the monitor, and connect the other end to one point of the equipotential grounding system.

Caution: ensure that the monitor is grounded correctly.

 After the supply mains has been interrupted when power switch remains in the “on” position and is restored after a period of time that is longer than 30 seconds, the monitor will run by the last settings when restarting the monitor.

2. When powered by built-in battery

 Caution: it’s better to recharge the battery after it is used up, the charging time should be

13~15 hours long.

 The provided battery of the monitor must be recharged after transportation or storage. So if the monitor is switch on without being connected to the AC power socket, it may not work properly due to insufficient power supply.

3.2.3 Starting the Monitor

The system performs self-test and enters initial display after switching on the monitor, and the yellow alarm indicator blinks to inform that the user can begin operating it.

 Check all the applicable functions to make sure that the monitor works normally.

 If the battery is applied please recharge it after using the monitor to ensure sufficient power storage. It will take minimal 8 hours to charge battery from depletion to 90% charge.

8

User Manual for Vital Signs Monitor

 Do not use the device to monitor the patient if there are indications of damage or reminders of error. Please contact the local dealer or our company.

 It’s recommended to delay 1 minute to start it again.

3.3 Connection

3.3.1 ECG Cable Connection

ECG measurement is to collect the ECG signal via the ECG electrodes. Electrode connects the patient and the lead. The lead connects the monitor. The locations of the electrodes are very important for obtaining accurate

ECG signals.

1. Connect the cable to the right-panel connector marked with the ECG icon.

2. Select electrodes to be used. Use only one type of electrode on the same patient to avoid variations in electrical resistance. For ECG monitoring, it is strongly recommended to use silver/silver chloride electrodes. When dissimilar metals are used for different electrodes, the electrodes may be subject to large offset potentials due to polarization. Using dissimilar metals may also increase recovery time after defibrillation.

3. Prepare the electrode sites according to the electrode manufacturer’s instructions.

4. Skin clean

 Clean and dry-abrade skin to ensure low sensor impedance. Mild soap and Water is recommended as a skin cleanser.

Note: Alcohol is not recommended as a skin cleanser; it leaves a film layer that may cause high sensor impedance. If alcohol is used, ensure 30-second dry time.

 Dry-abrading the skin gently with a dry wash cloth, gauze, or skin preparation product is helpful to remove the non-conductive skin layer.

The symbol indicates that the cable and accessories are designed to have special protection against electric shocks, and is defibrillator proof.

The locations of the electrode are in the following Figure:

Figure 3.3 Electrode Location

Note: If skin rash or other unusual symptoms develop, remove electrodes from patient.

5. After starting the monitor, if the electrodes become loose or disconnected during monitoring, the system will display “LEAD OFF” on the screen to alarm the operator.

It might not display ECG wave with 3 leads. The 5 leads should be used to have ECG

9

User Manual for Vital Signs Monitor

wave.

6.The ECG leads and their corresponding locations are as follows:

Symbol

C (V)

Chest electrode

RA

LA

LL

RL

C1(V1)

C2(V2)

C3 (V3)

C4 (V4)

C5 (V5)

C6 (V6)

Position

The intersection between the centerline of the right clavicle and Rib 2

The intersection between the centerline of the left clavicle and Rib 2

Left part of the upper abdomen

Right part of the upper abdomen

The electrodes are placed in different places, the different lead forms will display.

Table 3-1

Safety Instructions for ECG Monitoring

 Use the same type electrode on a patient. If skin rash or other unusual symptom occurs, remove electrodes from patient. Do not attach electrodes on the patient with an inflammation of the skin or scores on skin.

 PC-900 Vital Signs Monitor can only be equipped with ECG leads provided by our company; using

ECG leads supplied by other companies may cause improper performance or poor protection while using defibrillator.

 Electric parts of electrodes, leads and cable are forbidden to contact any other conductive parts

(including ground).

 PC-900 Vital Signs Monitor can resist against defibrillator and electrosurgical unit. Readings may be inaccurate for a short time after or during using defibrillator or electrosurgical unit.

 Transient caused by cable circuitry blocks while monitoring may be similar to the real heartbeat waveform, as a result resistance heart rate alarm rings. If you put the electrodes and cable in proper places according to this manual’s instructions and the instructions for using electrode, the chance of this transient occurring will be decreased.

 Besides the improper connection with electrosurgical unit may cause burns, the monitor may be damaged or arouse deviations of measurement. You can take some steps to avoid this situation, such as do NOT use small ECG electrodes, choosing the position which is far away from the estimated Hertzian waves route, using larger electrosurgical return electrodes and connecting with the patient properly.

10

User Manual for Vital Signs Monitor

 ECG leads may be damaged while using defibrillator. If the leads are used again, please do the functional check first.

When removing the ECG cable, hold the head of the connector and pull it out.

When the monitor is inoperable due to an overload or saturation of any part of the amplifier, it will prompt “Lead off” to remind operator.

No predictable hazard will be caused by the summation of leakage currents when several item of monitor are interconnected.

3.3.2 Blood Pressure Cuff Connection

1. Connect the cable to the right-panel connector marked with the NIBP icon.

2. Unveil and wrap the cuff around patient’s upper arm.

Requirements of the cuff:

1) Appropriate cuff should be selected according to the age of the subject. Its width should be 2/3 of the length of the upper arm. The cuff inflation part should be long enough to permit wrapping 50-80% of the limb concerned. See the table below for the dimensions:

Note: The size of the cuff selected should suit the subjects while measuring.

Cuff Model

Small-sized Pediatric Cuff

Arm Circumference

6cm11cm

Middle-sized Pediatric

10cm19cm

Large-sized Pediatric Cuff

Adult Cuff

18cm26cm

25cm35cm

Table 3-2

Cuff Width

4.5cm

8cm

10.6cm

14cm

When putting on the cuff, unveil and wrap it around the upper arm evenly to appropriate tightness.

2) Remember to empty the residual air in the cuff before the measurement is commenced.

3)

Locate the cuff in such a way that the “φ” mark is at a location where the clearest pulsation of brachial artery is observed.

4) The cuff should be tightened to a degree where insertion of one finger is allowed.

5) The lower end of the cuff should be 2cm above the elbow joint.

11

User Manual for Vital Signs Monitor

Figure 3.4 Cuff Placement

Pressure Accuracy Verification

Pressure Accuracy Verification is a function to inspect the accuracy of pressure measurement by the NIBP module inside the device. Technician or equipment manager should do pressure accuracy verification every half year or year in order to check if the pressure measurement still conforms to the requirement of product performance. If the deviation is beyond the declared specification, it is permitted to return it to factory for repair or calibration.

Before verification, please connect the monitor to a standard pressure meter as the reference equipment like a mercury pressure meter

Figure 3.5 Connection of Pressure calibration fixture

Mode 1: The inflation can be activated by Monitor so the pressure will increase automatically until it exceeds the limit value specified in table A. This pressure limit value depends on the patient type selection as shown in table A:

Adult

Child

240mmHg

200mmHg

Neonate 120mmHg

Table A

During the inflation, the Monitor will close the deflating valve, and the pressure value will be shown during the process. If there is no manual deflation operation, the pressure will persist until deflation by manual operation, so it is necessary to use a manual valve for doing adequate deflation in several steps to verify the pressure accuracy in the full scale of measurement range.

Mode 2: No automatic inflation by Monitor during the pressure accuracy verification.

Increase the pressure manually by the pumping balloon, and the verification can be done by applying different pressure value manually. If the increased pressure exceeds the given limit as shown in table

B, the Monitor will deflate automatically because of over-pressure protection.

12

User Manual for Vital Signs Monitor

Adult 300mmHg

Child 240mmHg

Neonate 140mmHg

Table B

After the verification, do press the button again to return to normal working mode, then

continue other operation, or the NIBP key will be invalid.

Pressure accuracy verification must be operated by technician or equipment manager. Doctor

or nurse is not allowed to do the verification, it is very dangerous especially when the pressure cuff is still on patients.

Air Leakage Check

In order to avoid significant error of blood pressure measurement or even no measurement result caused by air leakage in the pneumatic system including the cuff during measuring, it is recommended to check if there is leak in the pneumatic system as well.

Please remove the cuff from patient while performing the leakage check.

Safety Instructions for NIBP Monitoring

 When taking the measurement of a pediatric or an infant or neonate’s (less than 10 years old) blood pressure, do NOT operate in the adult mode. The high inflation pressure may cause lesion or even body putrescence

 It is recommended to take the blood pressure measurement manually. Automatic measurement should be used at the presence of a doctor/nurse.

 NIBP monitoring is prohibited to those who have severe hemorrhagic tendency or with sickle cell disease, or partial bleeding will appear.

 Pay attention to the color and sensitivity of the limb when measuring NIBP; make sure the blood circulation is not blocked. If blocked, the limb will discolor, please stop measuring or remove the cuff to other positions. Doctor should examine this timely.

 Confirm your patient category (adult or pediatric) before measurement.

 Do NOT bind NIBP cuff on limbs with transfusion tube or intubations or skin lesion area, otherwise, damages may be caused to the limbs.

 The time of the automatic pattern noninvasive blood pressure measurement pull too long, then the body connected with the cuff possibly have the purpura, lack the blood and the neuralgia. When guarding patient, must inspect the luster, the warmth and the sensitivity of the body far-end frequently. Once observes any exception, please immediately stop the blood pressure measurement.

The subject should lie on the back so that the cuff and the heart are in a horizontal position and the most accurate measure is taken. Other postures may lead to inaccurate measurement.

Do not speak or move before or during the measurement. Care should be taken so that the cuff will not be hit or touched by other objects.

13

User Manual for Vital Signs Monitor

The measures should be taken at appropriate intervals. Continuous measurement at too short intervals may lead to pressed arm, reduced blood flow and lower blood pressure, and resulting inaccurate measure of blood pressure. It is recommended the measure be taken at intervals of more than two minutes.

When an adult subject is monitored, the machine may fail in giving the blood pressure measure if the pediatric mode is selected.

Prior to use of the cuff, empty the cuff until there is no residual air inside it to ensure accurate measurement.

Do NOT twist the cuff tube or put heavy things on it.

When unplugging the cuff, hold the head of the connector and pull it out.

The symbol indicates that the cable and accessories are designed to have special protection against electric shocks, and is defibrillator proof.

3.3.3 SpO

2

Sensor Connection

SpO

2

sensor is a very delicate part. Please follow the steps and procedures in operating it. Failure to operate it correctly can cause damage to the SpO

2

sensor.

Operation procedure:

1. Connect the SpO

2

sensor to the connector labeled “SpO

2

”. When unplugging the probe, be

sure to hold the head of the connector and pull it out.

2. If the finger clip SpO

2

sensor is used, insert one finger into the sensor (index finger, middle finger or ring finger with short nail length) as shown in the figure below.

Figure 3.6 Finger clip SpO

2

sensor placement

When selecting a sensor, consider the patient’s category, adequacy of perfusion, availability of probe site and anticipated monitoring duration. Use only SpO

2

probes provided by our company with this monitor. Read the following table for SpO

2

probe information. Refer to Chapter 11.5 for the detailed instructions of each SpO

2 probe.

SpO

2

Probe

SpO

2

Finger clip Sensor (reusable)

Patient Category

Pediatric

14

User Manual for Vital Signs Monitor

SpO

2

Finger rubber Sensor(reusable)

SpO

2

Finger clip Sensor(reusable)

Adult

Adult

3. If the neonate SpO

2

sensor is used, please follow Figure 3.7 to connect.

Sensor adapter: wrapper

Y type sensor

Figure 3.7 Neonate SpO

2

sensor placement

High ambient light sources such as surgical lights (especially those with a xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight can interfere with the performance of an

SpO

2

sensor. To prevent interference from ambient light, ensure that the sensor is properly applied, and cover the sensor site with opaque material.

Failure to take this action in high ambient light conditions may result in inaccurate measurements.

If patient movement presents a problem, verify that the sensor is properly and securely applied; move the sensor to a less active site; use an adhesive sensor that tolerates some patient motion; or use a new sensor with fresh adhesive backing.

For reusable sensors, follow the sensor directions for use for cleaning and reuse. For single-patient use sensors, use a new sensor for each patient. Do not sterilize any sensor by irradiation, steam, or ethylene oxide.

Safety Introductions for SpO

2

Monitoring

15

User Manual for Vital Signs Monitor

 Continuous use of SpO

2

sensor may result in discomfort or pain, especially for those with microcirculatory problem. It is recommended that the sensor should NOT be applied to the same place for over two hours, change the measuring site periodically if necessary.

 SpO

2

measuring position must be examined more carefully for some special patient. Do NOT place the SpO

2

sensor on the finger with edema or fragile tissue.

 If sterile packaging of SpO

2

sensor is damaged, do not use it any more.

 Check the SpO

2

sensor and cable before use. Do NOT use the damaged SpO

2

sensor.

 When the temperature of SpO

2

sensor is abnormal, do not use it any more.

Please do not allow the cable to be twisted or bended.

Do NOT put the SpO

2

sensor and pressure cuff on the same limb, otherwise the NIBP measuring will affect SpO

2

measuring and cause the alarm error.

Using nail polisher or other cosmetic product on the nail may affect the accuracy of measurement.

The fingernail should be of normal length.

The SpO

2

sensor can not be immerged into water, liquor or cleanser completely, because the sensor has no capability of waterproofness.

3.3.4 TEMP Transducer Connection

Connecting methods:

1. Attach the transducers to the patient firmly;

2. Connect the cable to TEMP probe connector in the front panel.

Note: When unplugging the probe, be sure to hold the head of the connector and pull it out.

3.3.5 Loading printing paper

Operation procedures for loading printing paper:

1. Press both “OPEN” notches with force on printer shield with two thumbs to open it.

2. Move the tab of rubber roller lock at the left 90°upwards to unlock it, refer to the following figure with mark ①.

3. Cut one end of the paper into triangle, and load the paper from the underside of the rubber roller.

4. Turn the roller clockwise to get the paper rolled, and put the paper roll into the compartment.

5. Pull the paper out of paper slot on the shield.

6. Move the tab of the rubber roller lock 90° downwards to lock it.

7. Put the shield back in position and secure it.

Operation procedures for taking out printing paper roll:

1~2 steps are the same with the 1~2 steps mentioned above for loading printing paper.

3. Roll the loading roller anti-clockwise and pull the paper out.

4~5 steps are the same with the 6~7 steps mentioned above for loading printing paper.

16

User Manual for Vital Signs Monitor

Figure 3.8 Loading and taking out printing paper

P8 printer may be used due to the different configuration.

P8 printer operation instruction:

off.

Power indicator: green light shows the power is on, while the monitor is out of power, the green light is

Error indicator: red light is constant which shows the printer is out of paper, or the printing paper does not install well. When the printer installs normally, the red light is off.

Power Indicator

Error Indicator

Open button

Paper cartridge

Figure 3.9 P8 Printer

Loading printing paper:

Step 1: press and hold down the cartridge button to open the paper cartridge;

Step 2: Install the paper to the printer properly, pull the paper out of the printer for 2 cm, as shown in figure 3.10.

Step 3: Close the printer cover along the direction of arrow, as shown in figure 3.10.

17

User Manual for Vital Signs Monitor

Figure 3.10 printing paper

18

User Manual for Vital Signs Monitor

3.3.6 Battery Installation

1. Ensure that the monitor is not connected to AC power supply and the monitor is turned off.

2. Open the battery cover and place the battery in the direction as shown in Fig. 3.11 to insert the battery into any one of battery compartments. Do not insert battery with their polarities reversed.

3. Move the battery baffle to secure battery.

4. Close the battery cover.

Underside view

Figure 3.11 Battery Installation

Note:

Do not insert battery terminal with its polarities reversed, or the monitor can not be started.

Please take out the battery before transport or storage.

19

User Manual for Vital Signs Monitor

Chapter 4 Operations

4.1 Initial Monitoring Screen

When the parameter configuration of monitor is “ECG+SpO

2

+NIBP”, insert the ECG cable into the socket labeled “ECG” and attach the ECG leads to the electrodes placed on human body, once powered up, the LCD will display the initial monitoring screen, this is the default display screen as well.

When the parameter configuration of monitor is “SpO

2

+NIBP”, the screen will not display ECG waveform and data, but the operation is similar.

Figure 4.1 Default Display Screen

If the ECG cable is disconnected from the monitor or leads off from the patient, the ECG waveform will become a base line in main display area and “Lead off” will appear at the left side of prompt info area (as shown in Figure 4.2).

Figure 4.2 Lead off

The LCD screen will display the information by different display views, short pressing “ ” key to shift screen display among 4 display views: ECG waveform screen (default screen), SpO

2

plethysmogram screen,

SpO

2

trend graph screen, HR trend graph screen and NIBP list screen. Long pressing “ ” key will enter the setup menu screen. For every display view, the display area is divided into 3 parts: title area, main display

20

User Manual for Vital Signs Monitor area, and prompt info area (see Figure 4.1). The prompt info area contains 3 segment of information: status or event indication at the left, patient ID number in the middle, real time clock at the right (also see Figure 4.1).

4.1.1 Default Display Screen Description

Title area:

 Ⅱ ×1:ECG lead status and ECG waveform scale.

“ECG”: indicate the current monitoring parameter is ECG.

“36.8”: temperature numerical value

Main display area:

 When ECG leads is attached on the patient and connected to the monitor well, ECG waveform will be displayed in the main display area.

Prompt Info:

 Status or event indication segment:

This segment will display the ECG leads status, probe status, alarm silence counting-down timer, automatic NIBP measurement counting-down timer, over limit warning and other error messages for technical warning. If more than one event occurs or more status appears, the indication message will be displayed alternately at this segment.

“NIBP C-D: XXX”: the counting-down timer of NIBP measurement is XXX seconds. This prompt message appears only when the NIBP measuring mode is set as “AUTO X”.

“mute C-D: XXX”: the counting-down timer of alarm silence is XXX seconds. This prompt message appears only when the alarm silence is enabled.

 Patient ID segment:

“03”: Patient ID number.

 Real time clock segment:

“11:30:25”: the current time.

4.1.2 Operation Instructions

” key: select ECG lead. When ECG is monitored, press this key to switch the ECG lead among Ⅰ,Ⅱ, Ⅲ, aVR, aVL, aVF and V.

” key: freeze ECG waveform or Plethysmogram on the screen.

” key: shift display mode.

” key: print ECG waveform. Press it again to stop printing.

” key: start/cancel NIBP measurement.

21

User Manual for Vital Signs Monitor

”/ “”key:change ECG waveform scale.

” key: short press this key (about 1 second) to turn on or turn off alarm sound temporarily; longtime press it to enter into the alarm setup shortcut menu as shown in figure 4.3. If not turn off

“ECG Lead off” and “SpO

2

Probe off” manually after alarm lasts for 5 minutes, system will resume alarm silence status.

Figure 4.3 Alarm setup shortcut screen

Alarm setup operation description

1. Press “

” key or “

” key to move cursor to select parameter.

2. Press “■” key to confirm and enter into corresponding alarm parameter setup screen; Press “

” key or “

” key again to turn off corresponding lead off alarm.

3. Press “ ” to exist from Setup Menu Screen.

22

User Manual for Vital Signs Monitor

4.2 SpO

2

Monitoring Screen

Short time press “ Display” key to shift the screen view to SpO

2

monitoring screen, as shown in Figure

4.4.

Figure 4.4 SpO

2

Monitoring Screen

Note: if you need to store the measuring data, please set the option of “store” as “on” on ECG TEMP setting screen.

4.2.1 Screen Description

Title area:

“PI: 3‰”: the perfusion index is 3‰; it displays only when “Setup Menu→SpO

2

→PI Display” is set as “ON”.

Note: PI display function is optional and it needs hardware support.

“PLETH”: Mark of SpO

2

plethysmogram,when “PLETH” displays in title area, the main display area will be SpO

2

plethysmogram, and this display screen is the default screen.

“36.8”: temperature numerical value

Main display area:

When SpO

2

sensor is placed on the patient and connected to the monitor well, a trace of sweeping waveform

(plethysmogram) will be displayed in the main display area (as shown in Figure 4.4).

If the SpO

2

sensor is disconnected from the monitor or off from the patient, the plethysmogram will become a base line in main display area and “Probe off” will appear at the left side of prompt info area (as shown in

Figure 4.5). But when probe is disconected during monitoring patient, there will be visible and audible alarm simultaneously and after 5 minutes, visible and audible alarm will turn off automatically, only Probe Off show in the screen.

23

User Manual for Vital Signs Monitor

Figure 4.5 Probe Off

4.2.2 Operation Instructions

” key: press this key to shift to next display view (SpO

2

trend graph).

” key: Press it to print a trace of SpO

2

plethysmogram, press it again to stop printing.

” key: start/cancel NIBP measurement.

” key: Alarm silence switch, press it to enable/disable alarm silence.

4.3 Trend Graph Display

Short pressing “ Display” key to shift the screen view to trend graph display screen, as shown in Figure

4.6.

Trend graph

Trend length

Figure 4.6 Trend Graph

4.3.1 Screen Description

“12 hours”: the trend length of trend graph; three options: 12”, “24” or “96” hours; when the selection is 12 hours, the upper trend graph will display SpO

2

trend curve for last 12 hours.

“cursor on”: enable the display of cursor on trend graph, i.e. the vertical cursor line displayed in trend graph, so the user can move the cursor to inspect the SpO

2

value at the given time.

“SpO

2

”: indicate that the trend graph beside it is SpO

2

trend. Let the cursor stay here and press“

24

User Manual for Vital Signs Monitor key to confirm, then press “

” key or “

” key again to select trend graph type:

“SpO

2

”: SpO

2

trend graph

“HR” HR trend graph

4.3.2 Operation Instructions

1. Press “

” key or “

” key to highlight “trend length” or “cursor on” selection.

2. Press “

” key to confirm.

3. Press “

” key or “

” key again to select value of trend length (12/24/96 hours) if the selecting box stays in “trend length” option, or to move the cursor if the selecting box stays in “cursor on” option.

Instructions for viewing the trend curve:

 Select “cursor on” and press “

” key to confirm, and “cursor on” becomes “cursor off” , then you can press “▲” key or “▼” key to move the vertical cursor, the list box below will display

SpO

2

/HR value and the time value at the point where the cursor stays. Move cursor back and forth this way, you can view the SpO

2

/HR trend (12/24/96 hours long). Press “

” key again to exit trend viewing.

When pressing “▲” key or “▼” key to move cursor, the moving step is variable. The rule is that the in itial step is 1 point, after pressing “▲” or “▼” key towards the same direction for 5 times, the step becomes 5 points, and with 5 more pressing the step becomes 10, then 20. No matter what step is, as long as you press “▲” or “▼” key towards the other direction, the step becomes 1 and towards the other direction.

4. press:

” key: press this key to shift to next display view.

” key: Press it to print the current displayed trend graph.

” key: start/cancel NIBP measurement

” key: alarm silence switch; press it to enable/disable alarm silence.

4.4 NIBP List Screen

Short pressing “ Display” key to shift the screen to NIBP List screen, as shown in Figure 4.7.

25

User Manual for Vital Signs Monitor

Figure 4.7 NIBP List

The first column is the date, the second column is NIBP measuring time, the third column is NIBP value, and the fourth column is pulse rate (measured by NIBP module). “SYS/DIA/MAP” indicates the value of

“systolic pressure/diastolic pressure/mean arterial pressure”.

4.4.1 Operation Instructions

On NIBP List screen, if NIB

P measurement is more than 6 groups, press “▲” key or “▼” key to scroll up or down through all the measurement values. If NIBP measurement is not more than 6 groups, the keys “▲”or

“▼” are not effective.

” key: press this key to shift to next display view.

” key: print NIBP list.

” key: start/cancel measuring NIBP.

” key: alarm silence switch; press it to enable/disable alarm silence.

4.5 Setup Menu Screen

At any display view screen, long time press “ Display” key to shift the screen to Setup Menu screen, as shown in Figure 4.8. All the functional parameters of the system can be set through Setup Menu.

26

User Manual for Vital Signs Monitor

Figure 4.8 Setup Menu Screen

There are 8 functional groups for setting parameters: “ECG TEMP, SpO

2

, NIBP, Nurse Call, System, Patient

Info, Date/Time and Default” on the Setup Menu Screen.

1. Press“

” key or “

” key to shift cursor to corresponding functional group setting.

2. Pres “

” key to confirm and enter into corresponding functional parameter setup screen.

3. Pres “ ” key under the setup menu will print ECG waveform.

4. Press “ ” to exist from Setup Menu Screen.

At Setup Menu Screen or its submenu screen, when pressing “ ” key, the default display screen will be printed.

The following will cover each functional parameter’s setting up.

Note: If you disabled Hi and Lo limit alarm function of parameter monitoring, all the alarms related to its parameter monitoring will be disabled as well.

4.5.1 ECG and Temperature Setup

Figure 4.9 ECG/TEMP Setup Screen

Screen Description:

“Lead”: ECG lead selection: Ⅰ,Ⅱ,Ⅲ, AVR, AVL, AVF or V;

“Gain”: ECG waveform scale:

“×1/2”- waveform reduced to half of nominal scale

“×1”- nominal waveform scale “×2”- waveform with doubled scale

27

User Manual for Vital Signs Monitor

“1mV”: generating internal 1mV calibration signal. This signal is used to test the function of the machine. It is not used during normal operation. The default set is off.

“HR ”: HR alarm switch; “” indicates HR alarm is on; “ ” indicates HR alarm is off.

“HR Hi/Lo”: high/low limit of HR alarm;

“TEMP ”: temperature alarm

switch; “” indicates temperature alarm is on; “ ” indicates temperature alarm is off.

“TEMP Hi/Lo”: high/low limit of temperature alarm;

“Unit”: body temperature unit. Two options: “℃” or “℉”. Conversion relation: 1℉= (℃X1.8) +32.

“T probe”: the type of temperature probe “KRK”;

4.5.2 SpO

2

Setup

Figure 4.10 SpO

2

Setup Screen

Screen Description:

“SpO

2

”: SpO

2 alarm switch; “” indicates SpO

2 alarm is on; “ ” indicates SpO

2 alarm is off.

“SpO

2

Hi”: high limit of SpO

2 alarm; range: “1~100”.

“SpO

2

Lo”: low limit of SpO

2 alarm; range: “0~99”.

“PR ”: pulse rate alarm switch; “” indicates PR alarm is on; “

” indicates PR alarm is off.

“PR Hi”: high limit of PR alarm; range: “22~250”.

“PR Lo”: low limit of SpO

2 alarm; range: “0~248”.

“PI display”: “on” means PI display is enabled; “off” means PI display is disabled.

Operation Instructions

1. Press “

” key or “

” key to move cursor to select parameter.

2. Press “

” key to confirm and active this parameter setting.

3. Press “

” key or “

” again to adjust or modify parameter value.

4. Press “

”key again to confirm and save the setting.

5. Press “ ” key to return to upper level screen.

28

4.5.3 NIBP Setup

User Manual for Vital Signs Monitor

Figure 4.11A NIBP Setup Figure 4.11B Tourniquet Setup

Figure 4.11C Adjust Setup

NIBP Setup Screen Description:

“SYS ”: systolic pressure alarm switch; “” indicates systolic pressure alarm is on; “

” indicates systolic pressure alarm is off.

“SYS Hi”: high limit of systolic pressure alarm; range: “32~250” mmHg.

“SYS Lo”: low limit of systolic pressure alarm; range: “30~248” mmHg.

“DIA ”: diastolic pressure alarm switch; “” indicates diastolic pressure alarm is on; “

” indicates systolic pressure alarm is off.

“DIA Hi”: high limit of diastolic pressure alarm; range: “22~230” mmHg.

“DIA Lo”: low limit of diastolic pressure alarm; range: “20~228” mmHg.

“MAP ”: mean arterial pressure alarm switch; “” indicates mean arterial pressure alarm is on;

“ ” indicates mean arterial pressure alarm is off.

“MAP Hi”: high limit of mean arterial pressure alarm; range: “28~242” mmHg.

“MAP Lo”: low limit of mean arterial pressure alarm; range: “26~240” mmHg.

“Mode”: NIBP measuring mode, “manual”, “AUTO 1”, “AUTO 2”, …“AUTO 240” and “STAT”

29

User Manual for Vital Signs Monitor etc. options. “AUTO 1” means NIBP measurement takes once every one minute automatically;

“AUTO 60” means NIBP measurement takes once every 60 minutes automatically; In AUTO mode, the counting-down timer is displayed in the “Prompt Info” area, as shown in Figure 4.1.

“unit”: unit of the blood pressure value;

“mmHg” or “kPa” can be selected. Conversion: 1kPa=7.5mmHg.

“>>”: Page down icon. When cursor stays in the “unit” filed, press “▼” key to enter Tourniquet

Setup.

Tourniquet Setup Screen Description:

“Pressure”: when you use Tourniquet function, you need to preset a cuff pressure for hemostasia.

The pressure is adjustable, and its adjusting limit is different for different patient category: for neonates: preset range: 70~100mmHg, default value: “90” mmHg for pediatric: preset range: 80~130 mmHg, default value: “110” mmHg; for adults: preset range: 80~180mmHg, default value: “140” mmHg.

If the pressure drops down slowly under 10mmHg compared with the preset value due to little air leakage in the pneumatic system when time passes by, the monitor will re-inflate to maintain the cuff pressure close to the preset pressure value.

Note: the unit of cuff pressure is the same with the NIBP unit in NIBP Setup.

“Duration”: After presetting the cuff pressure, you need to set the time period for maintaining the preset pressure after inflation. “5, 6, 7,…120” minutes adjustable. The default value is “40” minutes.

If the set value is “xx” minutes, the monitor will count down from “xx” minutes automatically when starting cuff inflation. When time is up, it will deflate automatically.

“Alert T”: the alert time for reminding user that the operation of tourniquet is going to be end after this time period. 1 to 60 minutes adjusting range with 1 minute step, the default value is “5” minutes.

If the set value is “xx” minutes, the monitor will produce alarm sound until ending deflation when counting down time reaches to “xx” minutes. The alarm type is high priority alarm. (For example: the duration is 40 minutes, the alert time is 5 minutes, the alarm will ring for prompt when the duration counting down to 5 minutes. The Prompt Info area starts to prompt: TOUR C-D 300 seconds. )

“Start”: shift cursor to “Start” and press “■” key, “Start” becomes “Stop” and meanwhile the blood cuff starts being inflated; Pressing “Stop” button can stop using this function. After deflation, it will change to “Start” again.

NIBP calibration setup Descriptions:

NIBP Cali Mode 1: Inflating the Pump. Move the cursor to NIBP Cali Mode 1”Start” button, click the OK button to begin the NIBP calibration. (Meanwhile, the “Start” shifts to “Stop”, after the calibration the “Stop” shifts to “Start”)

NIBP Cali Mode 2:Receiving the exterior pressure. The exterior pressure source pressurize to the module to proceed the pressure calibration. Move the cursor to NIBP calibration mode 2”Start” button, click the OK button to begin the NIBP calibration.(Meanwhile, the “Start” shifts to “Stop”, after the

30

User Manual for Vital Signs Monitor calibration the “Stop” shifts to “ Start”)

Gas leak: Move the cursor to Gas leak ”Start” button, click the OK button, the pump inflates to certain pressure and then the valve will be closed for leak detection for ten seconds, then the blood pressure module will deflate automatically and the screen displays measurements.

The NIBP calibration and Gas leak detection can only be carried on when the NIBP measurement is set to mode “Manual”.

Other buttons are disabled except

“■” OK button and “

” Power button during NIBP calibration and Gas leak detection.

Make sure the

“■” OK button is off after the test, or the user could not do other operations.

NIBP Mode Setup Shortcut Screen Descriptions:

In waveform display screen or trend graphic screen or NIBP list screen longtime press “ ”key about

3 seconds can enter into the screen shown in Figure 4.11D. Please refer to “NIBP Setup Screen

Description” for more detailed information.

Figure 4.11D NIBP Mode Setup Shortcut Screen

4.5.4 Nurse Call

Screen Description:

Figure 4.12 Nurse Call Setup Screen

31

User Manual for Vital Signs Monitor

“Output level”: two options “low” or “high” output levels are available.

When the calling system in hospital works in “Normal Open” mode, “low level” should be selected.

When the calling system in hospital works in “Normal Close” mode, “high level” should be selected

“Source”: three kinds of alarm sources can trig the nurse call: high level alarm, medium level alarm and low level alarm (multi-optional). If you don’t make choice, nurse call signal will not be sent out.

“Duration”: two options “pulse” or “continuous” output modes are available;

“continuous”: the continuous mode of output means the nurse call signal will keep until the selected alarm source(es) disappear, i.e. the signal will last from starting alarm to stopping alarm.

“pulse”: the output nurse call signal is pulse signal which lasts for 1 second. When several alarms occur at the same time, only one pulse signal will be sent out.

Note:

Nurse Call function can not be regarded as main alarm notice method, please do not entirely relay on it. You should combine parameter values with alarm level and patient’s clinical behavior and symptom to determine patient’s status.

4.5.5 System Setup

Figure 4.13 System Setup Screen

Screen Description:

“Vol”: set beeper volume, “1~7” level adjustable.

“key beep”: to turn on/off key beep;

“LANG”: language selection. “ENG” for English.

“priority”: priority of “PR” value or “HR” value display. The default set is “HR”.

“backlite”: turn on/off backlight;

“contrast”: adjust LCD display contrast, “0~31” level adjustable;

“care mode”: “Demo” shows the demo waveforms and data. In the demo state, all the signals and data are generated from the monitor for demonstration and testing purpose. “Real” shows the real time waveform, i.e. normal monitoring status;

BT SD: turn on/off the pulse beeping sound.

32

4.5.6 Patient Info

User Manual for Vital Signs Monitor

Figure 4.14 Patient Info Screen

Screen Description:

“ID”: change or set current patient’s ID number, 0~100 adjustable;

“category”: change or set the category of current patient; three options “adult” ,“pediatric” and

“neonate”, the default is “adult”

Note: If the patient ID is changed, all the history data will be cleared, that means SpO

2

trend graph, HR trend graph and NIBP list will become empty.

4.5.7 Date/Time

Figure 4.15 Data/Time Setup Screen

Screen Description:

“yy 07 mm 09 dd 21”: date setting, “07-09-21” shows the date is September 21 st

, 2007.

“hh 10 mm 15 ss 20”: time setting, “09:20:21” shows the time is10:15:20.

4.5.8 Recover Default Settings

On Setup Menu screen, press “

” button or “

” button to shift cursor to “Default”, and then press “

” button, all the setting parameters will be reset to factory default setting value.

4.6 Power Saving Mode

On the initial display screen, you can make the monitor stay in power saving mode for power saving.

33

User Manual for Vital Signs Monitor

Short time press power button to shift screen to “Power Saving Mode” display screen, as shown in Figure

4.16.

Figure 4.16 Power Saving Mode

Press “▲” button or “▼” button to shift cursor to “yes” or “no” and press “

” button to confirm. If your selection is “yes”, all the numerical values displayed on digital LEDs display become darker and the monitor stays in power saving mode.

Short time press power button again to shift screen to “Power Saving Mode” display screen for exiting the sleeping mode.

34

User Manual for Vital Signs Monitor

Chapter 5 Alarm

5.1 Alarm Priority

High Priority:

TOUR C-D:XXX seconds

PR Over limit

SpO

2

over limit

SYS over limit

DIA Over limit

MAP Over limit

NIBP error 1#

NIBP error 2#

NIBP error 3#

NIBP error 4#

NIBP error 5#

Air leak

Cuff error

NIBP over range

Over motion

Over pressure

NIBP timeout

Medium Priority:

Probe Off

5.2 Alarm modes

When an alarm occurs, the monitor responds with visual alarm indications (which are shown by two ways: alarm indicator and alarm message description) and audible alarm indications.

Visual Alarm Indicators

The flashing rates for the three categories of alarms are shown in the table below.

Indicator Color

Alarm Category Flashing Rate

Red flashing

Yellow flashing

Yellow light

High priority alarm

Medium priority alarm

Low priority alarm

2 Hz

0.5 Hz

Constant(on)(non-flashing)

Table 5.1

Refer to Chapter 11.2 Alarm Information for detailed alarm message descriptions.

35

User Manual for Vital Signs Monitor

Audible Alarm Indications

The audible alarm has different tone pitch and on-off beep patterns for each priority category. These are summarized in the Table below.

Alarm Category

High priority alarm

Medium priority alarm

Low priority alarm

Tone Pitch

~400Hz

~500Hz

~500Hz

Beep Chain

10 beeps pause 3 sec.

3 beeps pause 5 sec.

Single beep

Table 5.2

Note: Visual alarm indicators can not be suspended or removed. Audible alarms may be decreased in volume or silenced as described in Chapter 5.

5.3 Alarm Silence

Press key to set or activate the system alarm. In the monitoring screen, press “Alarm” to set the alarm timer. There are four options of alarm silent time: 2 minutes, 5 minutes, 10 minutes and 20 minutes. The time shows up on the upper left corner of the screen. When the alarm timer is activated, the system begins to count down. If alarm occurs during that period, the system alarm will be activated automatically and the monitor will give alarm. If there is no alarm during that period, when the set time has passed the system alarm will be activated as well.

When the monitor alarms, press key to suspend the alarm and set the alarm silence time.

DO NOT silence the audible alarm or decrease its volume if patient safety could

be compromised.

5.4 Alarm Setting

In the Mode Selection screen, move the cursor to the “SETUP”, and press it to enter system setup screen.

Limits setup: Move the gray cursor to the High or Low limits of the alarm settings, and press the

“Alarm” key to turn ON or OFF the alarm for the setting. Yellow color shows ON status, and gray color shows the OFF status.

Refer to Chapter 11.2 for detailed Default Alarming Values of All Parameters and Setup Range.

Whenever the monitor is used, check the alarm limits to ensure that they are

appropriate for the patient being monitored.

5.5 Verify Adjustable Alarm Function

To verify adjustable alarm function, select “Demo” for the item of Mode in system parameter settings menu and adjust alarm limits or change alarm setting, then pay a close attention to the alarm. If the alarm is sent out

36

User Manual for Vital Signs Monitor according to your setting, it means the alarm function is effective.

37

User Manual for Vital Signs Monitor

Chapter 6 Technical Specifications

6.1 ECG Monitoring

1. Input signals range in amplitude: ± (0.5 mVp ~ 5 mVp)

2. Heart rate display range: 15 bpm ~ 350 bpm

3. Heart rate display accuracy: ± 1% or ± 2 bpm, whichever is greater.

4. Heart rate averaging: Averages the recent eight beats having RR intervals falling within the acceptable limits.

5.

Heart rate alarm delay time: ≤ 10s

6. Response time to change in heart rate:

Change from 80 bpm to 120 bpm: < 8 sec

Change from 80 bpm to 40 bpm: < 8 sec

7. Tall T-wave rejection:

Rejects all T-wave less than or equal to 120% of 1mV QRS.

8.

Pacemaker pulse rejection:

Rejects

all pulses of amplitude ± 2mV to ± 700mV and duration

0.1 to 2 ms without overshoot;

Rejects

all pulses of amplitude ± 2mV to ± 400mV and duration

0.1 to 2 ms with overshoot.

9.Sensitivity selection:

×1/2, 5mm/mV tolerance: ± 5%

×1, 10mm/mV tolerance: ± 5%

×2, 20mm/mV tolerance: ± 5%

10. Sweeping speed: 25mm/s tolerance: ±10%

11.

ECG noise level: ≤ 30μV

P-P

.

12. ECG input loop current: ≤ 0.1μA

13. Differential input impedance: ≥ 5MΩ

14. Common-mode rejection ratio (CMRR):

≥ 105dB

15. Time constant:

Monitoring mode: ≥ 0.3s

16. Frequency response: 0.67 Hz~40 Hz(

+0.4dB

,-3.0dB)

38

User Manual for Vital Signs Monitor

Additional declarations to conform the particular standard of IEC 60601-2-27 “Medical electrical equipment – Part 2-27: Particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment”

Applied current less than 0.1 microamperes.

Direct current for respiration, leads-off sensing, and active noise suppression

Response to irregular rhythm

Time to ALARM for tachycardia

A1 Ventricular bigeminy-80BPM

A2 Slow alternating ventricular bigeminy-60BPM

A3 Rapid alternating ventricular bigeminy-120BPM

A4 Bidirectional systoles-90BPM

Waveform B1, Amplitude

0.5 mV

1 mV

2mV

Average Time to Alarm

<8 sec

<8 sec

<8 sec

Waveform B2, Amplitude

1mV

2mV

4mV

Average Time to Alarm

<8 sec

<8 sec

<8 sec

6.2 TEMP Monitoring

1. TEMP measuring range: 25.0

2. TEMP measuring accuracy: ±0.2 ℃

3. TEMP responding time: ≤150s

℃~ 45.0℃

6.3 NIBP Monitoring

1. Measuring method: Oscillometric Technique

2. Atmospheric pressure measuring range: 0 mmHg~300mmHg

3. Accuracy of pressure measurement: ±3 mmHg

4. Cuff inflation time: <10 seconds (typical adult cuff)

5. Measurement time on the average: < 90 seconds

6. Air release time while the measurement is canceled: <2 seconds (typical adult cuff)

7. Initial cuff inflation pressure

Adult: 175 mmHg Infant: 135 mmHg Neonate: 65 mmHg

8. Overpressure protection limit

Adult:

≤ 300 mmHg Infant: ≤ 240mmHg Neonate: ≤ 150 mm

9. NIBP measurement range: press(unit)

SYS mmHg

MAP

DIA mmHg mmHg

Adult

40~275

20~230

10~210

Infant

40~200

20~165

10~150

Neonate

40~135

20~110

10~95

39

User Manual for Vital Signs Monitor

10. NIBP accuracy:

Maximal mean difference: ±5 mmHg

Maximal standard deviation: 8 mmHg

11. Measurement mode: Manual, Auto, STAT

6.4 SpO

2

Monitoring

1. Transducer: dual-wavelength LED

Wavelength: Red light: 660 nm, Infrared light: 905 nm.

Maximal optical output power: less than 2mW maximum average

2. SpO

2

measuring range: 35%~100%

3. SpO

2 measuring accuracy: not greater than 3% for SpO

2

range from 70% to 100%

*NOTE: accuracy defined as root-mean-square value of deviation according to ISO 9919

4. Low perfusion performance: the declared accuracy is attained when the pulse amplitude modulation ratio is as low as 0.4%

6.5 Pulse Monitoring

1. Pulse rate measuring range: 30bpm~240bpm

2. Pulse rate measuring accuracy: ±2bpm or ±2%, whichever is greater.

6.6 Data Recording

1. Sensitivity selection tolerance: ±5%

2. Recording speed: 25mm/s

3. Recording speed accuracy: ±10%

4. Hysteresis: ≤0.5mm

5. Frequency response:

Monitoring mode: 0.5~40Hz

6. Time constant:

Monitoring mode: ≥0.3s

6.7 Other Technical Specifications

1. AC power supply voltage: 100~240VAC

2. AC power frequency: 50/60 Hz

3. Fuse specification: T3

.15AL/250V Φ520mm.

4. Internal power supply: 12VDC (rechargeable)

5. Battery specification: 12V 2.3AH (sealed lead-acid battery)

40

User Manual for Vital Signs Monitor

6.8 Operating Environment

Working Environment

Ambient temperature range: 5°C ~ 40°C

Relative humidity: 30 ~ 80%

Atmospheric pressure: 70kPa ~106kPa

Transport and Storage Environment

Ambient temperature range: -20°C ~ 60°C

Relative humidity: 10 ~ 95%

Atmospheric pressure: 50.0kPa ~107.4kPa

6.9 Classification

Safety standard

The type of protection against electric shock

The degree of protection against electric shock

Electro-Magnetic Compatibility:

IEC 60601-1

ClassⅠequipment.

Type BF, CF applied parts

Group I, Class A

41

User Manual for Vital Signs Monitor

6.10 Guidance and Manufacturer’s Declaration-Electromagnetic

Compatibility

Table 1

Guidance and manufacturer’s declaration-electromagnetic emission-

PC-900 Vital Signs Monitor is intended for use in the electromagnetic environment specified below.

The customer or the user of the equipment or system should assure that it is used in such an environment.

Emissions test

RF emissions

CISPR 11

RF emissions

CISPR 11

Harmonic emissions

IEC61000-3-2

Voltage fluctuations/flicker emissions

IEC61000-3-3

Compliance

Group 1

Class A

Class A

Complies

Electromagnetic environment-guidance

PC-900 Vital Signs Monitor uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

PC-900 Vital Signs Monitor is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

for all EQUIPMENT AND SYSTEMS

42

User Manual for Vital Signs Monitor

Table 2

Guidance and manufacturer’s declaration-electromagnetic immunity for all EQUIPMENT AND SYSTEMS

PC-900 Vital Signs Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment or system should assure that it is used in such an environment.

Electromagnetic environment

Immunity test IEC60601 test level Compliance level

Electrostatic discharge(ESD)

IEC61000-4-2

Electrical fast transient/burst

IEC61000-4-4

±6 kV contact

±8kV air

±2kV for power

Supply lines

±1 kV for input/output lines

±6 kV contact

±8kV air

±2kV for power

Supply lines

±1 kV for input/output lines

-guidance

Floors should be wood, concrete or ceramic tile. if floors are covered with synthetic material, the relative humidity should be at least 30%

Mains power quality should be that of a typical commercial or hospital environment.

Surge

IEC 61000-4-5

±1kV line (s) to line(s)

±2kV line(s) to earth

±1kV differential mode

±2kV common mode

Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruptions and voltage variations on power supply input lines

IEC61000-4-11

Power frequency(50Hz/60Hz) magnetic field

IEC61000-4-8

3A/m 3A/m

Mains power quality should be that of a typical commercial or hospital environment. If the user of the equipment or system requires continued operation during power mains interruptions, it is recommended that the equipment or system be powered from an uninterruptible power supply or a battery.

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE U

T is the a.c. mains voltage prior to application of the test level.

43

User Manual for Vital Signs Monitor

Table 3

Guidance and manufacturer’s declaration – electromagnetic immunity-for

EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING

PC-900 Vital Signs Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of PC-900 Vital Signs Monitor should assure that it is used in such an electromagnetic environment.

IMMUNITY test IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Portable and mobile RF communications equipment should be used no closer to any part of

PC-900 Vital Signs Monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

Conducted RF 3 Vrms 3V

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 V/m

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).

b

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey , a should be less than the compliance level in each frequency range .

b

Interference may occur in the vicinity of equipment marked with the following symbol.

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a: Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic site survey should be considered. If the measured field strength in the location in which PC-900

Vital Signs Monitor is used exceeds the applicable RF compliance level above, PC-900 Vital Signs Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may

44

User Manual for Vital Signs Monitor be necessary, such as re-orienting or relocating PC-900 Vital Signs Monitor. b: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Table 4

Recommended separation distances between portable and mobile RF communications equipment and The equipment or system- for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING

PC-900 Vital Signs Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the equipment or system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the equipment or system as recommended below, according to the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter

Rated maximum m output power of

150kHz to 80MHz 80MHz to 800MHz 80MHz to 2,5GHz

transmitter

W

0,01 0.12 0.12 0.23

0,1

1

10

0.38

1.2

3.8

0.38

1.2

3.8

0.73

2.3

7.3

100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

45

User Manual for Vital Signs Monitor

46

User Manual for Vital Signs Monitor

Chapter 7 Packaging and Accessories

7.1 Packaging

The product is packed in high quality corrugated cartons with foam inside to protect the equipment against damage in the shipping and handling process.

Weight: Details see the indication on the outer package.

Dimension: 360(L)×320(W)×410(H) (mm)

7.2 Accessories

(1) ECG cable with lead wire One set

(2) ECG electrodes Ten pieces

(3) NIBP cuff One piece

(4) SpO

2

probe One piece

(5) Temperature probe One piece

(6) Power cord One piece

(7) Grounding wire One piece

(8) User manual One copy

(9) Quality Certificate One copy

(10) Warranty Two copies

(11) Packing list Two copies

Note: The accessories are subject to change. Detailed items and quantity see the Packing List.

47

User Manual for Vital Signs Monitor

Chapter 8 Monitoring Parameter

8.1 ECG Monitoring

8.1.1 How to Obtain High Quality ECG and Accurate Heart Rate Value

The electrocardiogram (ECG or EKG) is primarily a tool for evaluating the electrical events within the heart.

The action potentials of cardiac-muscle cells can be viewed as batteries that cause charge to move throughout the body fluids. These currents represent the sum of the action potentials occurring simultaneously in many individual cells and can be detected by recording electrodes at the surface of the skin. The figure below shows the system of the heart.

First of all, the hospital should be equipped with a 100~240V power supply system with a typical grounding wire. If big interference in ECG continues, connect one end of the grounding wire provided with this equipment to the grounding wire on the back panel of this monitor, and the other end to the special grounding wire, water pipe or radiator.

A common ECG plate electrode used together with this monitor has short shelf life. Generally, the shelf life is only one month after the package is opened. When outdated plate electrode is used, due to skin’s contact impedance and big electrode potential, the chance of interference will be increased, and the ECG baseline will have an unstable inclination. Therefore, always use valid plate electrodes.

48

User Manual for Vital Signs Monitor

8.1.2 Factors affecting ECG signal

 Interference from Electrosurgical Unit;

 Doesn’t filter the interference waveform;

 Poor grounding;

 Electrodes are not placed properly;

 Use expired electrode or use disposable electrode repeatly;

 The skin placed electrode is unclean or poor contract caused by scurf and hair;

 Electrode long-time used.

8.2 NIBP Monitoring

8.2.1 Measuring Principle

Blood pressure may be measured in an invasive way (whereby the sensor will be inserted into blood vessel directly) or a non-invasive way. The non-invasive way includes several methodologies, such as the Korotkoff

Sound Method and oscillating method. The Korotkoff Sound Method is used as a conventional way, whereby stethoscope is used to measure the blood pressure. By the oscillating method, an inflation pump will fill the air, and release it slowly. A computer will record change of the cuff pressure when the air is released. With this record, the blood pressure value will be determined. First of all, make sure the signal quality judgment by computer meets the requirements of accurate calculation (such as sudden limb movement or cuff being hit during the measurement). If the answer is negative, give up the calculation. If the answer is positive, proceed with calculation of the blood pressure value.

As change of the blood pressure is recorded by electric sensor, which sensitivity is much higher than that of human ears, the oscillating method uses different definitions for measurement of diastolic pressure, mean arterial pressure and systolic pressure from the Korotkoff Sound Method. When the oscillating method is used, the circuit in the measuring apparatus will separate the amplitude of the cuff pressure from its change with pulsation. With the oscillating method, the blood pressure at the maximum amplitude of cuff pressure is defined as the mean arterial pressure. The blood pressure at amplitude of cuff pressure forward reduced according to proper proportion is defined as systolic pressure, while the blood pressure at amplitude of cuff pressure backward reduced according to proper proportion is defined as diastolic pressure. The maximum change of pulse pressure occurs at these two points. They are equivalent to the point with pulse sound and the point without pulse sound respectively in the Korotkoff Sound Method.

When the risk of invasive monitoring method outweighs its advantage of accuracy, non-invasive monitoring method shall be used.

Comparison between blood pressure measuring methods

To overcome the effect of human hearing variation and air release speed on measurement accuracy when the conventional Korotkoff Sound Method is used to take measure of blood pressure, people have been dedicated to study of automatic measurement of blood pressure. By now, automatic blood pressure

49

User Manual for Vital Signs Monitor measuring system based on the principle of oscillating method is mature. In practice, however, various problems are encountered, such as why the measures taken by the oscillating method is lower or higher than those taken by Korotkoff Sound Method? Why the measures are inclined to decline? Why, in some cases, no result is obtained in spite of the inflation actions? Why the measure values have big discreteness and even abnormal data in some cases? Why the SpO

2

waveforms may disappear suddenly? ...and so on. The following explanations are devised to give the answers.

The Oscillating method vs. the Korotkoff Sound Method

Blood pressure measurement by the oscillating method and Korotkoff Sound Method has good correlation with the invasive measurement. Notwithstanding, any of the non-invasive blood pressure measurements has its one-sidedness when it is compared to the invasive measurement. The oscillating method has its advantages over the Korotkoff Sound Method in less error, higher reliability and stability. Their differences may be reflected in the following aspects.

1. The measures by the Korotkoff Sound Method are liable to effect of human factors. For example, different people may have different sound judging ability, or different reactivity when listening to heart sound and reading mercury meter. The air release speed and subjectivity may also affect the judgment.

By the oscillating method, the computation is accomplished by the computer, thus relieving the possibility of effect due to human factor.

2. With the Korotkoff Sound Method, the measure is taken on the basis of appearance and disappearance of heart sound. The air release speed and heart rate may have direct effect on the measurement accuracy.

It also has the disadvantages of rapid air release and poor accuracy. In the contrast, with the oscillating method, the determination is calculated on the basis of cuff pressure oscillatory waveform envelope, and the air release speed and heart rate has little effect on the measurement accuracy.

3. Statistics show that, when measuring the hypertension, the measure taken by the oscillating method is likely to be lower than that taken by the Korotkoff Sound Method. When measuring the hypotension, the measure taken by the oscillating method is likely to be higher than that by the Korotkoff Sound

Method. But, it doesn’t mean the advantages or disadvantages between the oscillating method and the

Korotkoff Sound Method. Comparison with the results taken by more accurate method, let’s say comparison of the invasive pressure result with the output value by the blood pressure measuring simulator, will show which method has more accurate results. In addition, higher or lower value should be a statistical concept. It is recommended those used to adopt the Korotkoff Sound Method use different physiological calibration for values determined by the oscillating method.

4. The studies have shown that the Korotkoff Sound Method has the worst accuracy when it comes to measurement of hypotension, while the oscillating method has worse accuracy when it comes to measurement of controlled hypertension relief.

8.2.2 Factors affecting NIBP measuring

 Select a cuff of appropriate size according to the age of the subject.

 Its width should be 2/3 of the length of the upper arm. The cuff inflation part should be long enough to permit wrapping 50-80% of the limb concerned.

Prior to use of the cuff, empty the cuff until there is no residual air inside it to ensure accurate measurement.

50

User Manual for Vital Signs Monitor

Make the cuff mark φ in the position where artery pulsates obviously, the effect will be best.

The lower part of cuff shall 2cm above the elbow joint.

 Do not wrap the cuff on too thick clothes(especially for cotton-padded clothes and sweater ) to take measurement;

 The testee shall lie in bed or sit in chair, make the cuff and heart at the same level, the result will be most accurate, other postures may have inaccurate result;

 During measuring, do not move your arm or the cuff;

 The measuring interval shall longer than 2 minutes, in continuous measurement, too short interval may cause arm extrusion, blood quantity increases, then cause blood pressure increases.

 Keep the patient still and stop talking before and during measuring;

 The patient’s mood also can affect the measuring result, when exciting, the blood pressure goes up.

 The measuring result also affected by time, lower in the morning and higher in the evening;

8.2.3 Clinical Limitations

1. Serious angiospasm, vasoconstriction, or too weak pulse.

2. When extremely low or high heart rate or serious arrhythmia of the subject occurs. Especially auricular fibrillation will lead to unreliable or impossible measurement.

3. Do not take the measurement when the subject is connected with an artificial heart-lung machine.

4. Do not take the measurement when the subject uses diuresis or vasodilator.

5. When the subject is suffering from major hemorrhage, hypovolemic shock and other conditions with rapid blood pressure change or when the subject has too low body temperature, the reading will not be reliable, for reduced peripheral blood flow will lead to reduced arterial pulsation.

6. Subject with hyperadiposis;

In addition, statistics show that 37% people report blood pressure difference of no less than

0.80kPa(6mmHg) between the left and right arms, and 13% people report difference of no less than

1.47kPa (11mmHg).

Note:

Some practitioners may report big discreteness or abnormal value of the blood pressure measures when the oscillating method is used. As a matter of fact, the so-called “big discreteness” must be a term in the sense of statistical significance of mass data. Abnormal data may be observed in some individual cases. It is normal in the scientific experiments. It may be caused by an apparent reason, or by an unknown factor in some cases. Such individual doubtful experimental data may be identified and eliminated using the special statistical technique. It is not a part of this manual. The practitioner may eliminate the apparently unreasonable data according to the experience.

8.3 SpO

2

Monitoring

8.3.1 Measuring Principle

51

User Manual for Vital Signs Monitor

Based on Lamber-Beer law, the light absorbance of a given substance is directly proportional with its density or concentration. When the light with certain wavelength emits on human tissue, the measured intensity of light after absorption, reflecting and attenuation in tissue can reflect the structure character of the tissue by which the light passes. Due to that oxygenated hemoglobin (HbO

2

) and deoxygenated hemoglobin (Hb) have different absorption character in the spectrum range from red to infrared light (600nm~1000nm wavelength), by using these characteristics, SpO

2

can be determined. SpO

2

measured by this monitor is the functional oxygen saturation -- a percentage of the hemoglobin that can transport oxygen. In contrast, hemoximeters report fractional oxygen saturation – a percentage of all measured hemoglobin, including dysfunctional hemoglobin, such as carboxyhemoglobin or metahemoglobin.

8.3.2 SpO

2

Measurement Restrictions (interference reason)

 Intravascular dyes such as indocyanine green or methylene blue

 Exposure to excessive illumination, such as surgical lamps, bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight.

 Vascular dyes or external used color-up product such as nail enamel or color skin care

 Excessive patient movement

 Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line

 Exposure to the chamber with high pressure oxygen

 There is an arterial occlusion proximal to the sensor

 Blood vessel contraction caused by peripheral vessel hyperkinesias or body temperature decreasing

8.3.3 Low SpO

2

measuring value caused by pathology reason

 Hypoxemia disease, functional lack of HbO

2

 Pigmentation or abnormal oxyhemoglobin level

 Abnormal oxyhemoglobin variation

 Methemoglobin disease

 Sulfhemoglobinemia or arterial occlusion exists near sensor

 Obvious venous pulsations

 Peripheral arterial pulsation becomes weak

 Peripheral blood supply is not enough

8.3.4 Clinical Limitations

 As the measure is taken on the basis of arteriole pulse, substantial pulsating blood stream of subject is required. For a subject with weak pulse due to shock, low ambient/body temperature, major bleeding, or use of vascular contracting drug, the SpO

2 waveform (PLETH) will decrease. In this case, the measurement will be more sensitive to interference.

52

User Manual for Vital Signs Monitor

 For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with icterus problem, the SpO

2

determination by this monitor may be inaccurate.

 The drugs such as dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor blamed for serious error of SpO

2

measurements.

 As the SpO

2

value serves as a reference value for judgment of anemic anoxia and toxic anoxia, the measurement result of some patients with serious anemia may also present as good SpO

2 value.

8.3.5 Points to be noted in SpO

2

and Pulse Measuring

 The finger should be properly placed (see the attached illustration of this instruction manual), or else it may cause inaccurate measurement result.

 Make sure that capillary arterial vessel beneath the finger is penetrated through by red and infrared lights.

 The SpO

2

sensor should not be used at a location or limb tied with arterial or blood pressure cuff or receiving intravenous injection.

 Do not fix the SpO

2

sensor with adhesive tape, or else it may result in venous pulsation and consequential inaccurate measurement result of SpO

2

.

 Make sure the optical path is free from any optical obstacles like adhesive tape.

 Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby light, infrared heater, and direct sunlight etc.

 Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy.

 Please do not use the SpO

2

sensor when having the MRI, or burn may be caused by faradism.

 Always observe the plethysmogram (waveform), which is auto-scaled within the range of 100. The

SpO

2

reading may be unlikely true when the waveform is not smooth or irregular. If in doubt, rely on your clinical judgment, rather than the monitor readout.

 A functional tester can not be used to assess the accuracy of the pulse oximeter monitor or a SpO

2 sensor. However, a functional tester, such as SpO

2

simulator can be used to check how accurately a particular pulse oximeter is reproducing the given calibration curve. Before testing the oximeter by a functional tester, please firstly ask the manufacturer which calibration curve is used, if necessary, request the manufacturer for its dedicated calibration curve and download it into the tester.

8.4 Temperature Monitoring

The sensor is thermo-resistor type (25℃

5kΩ) and is supplied with constant micro current. Calculating the temperature of measured part through measuring the voltage. There is a period responding time, so the accurate temperature value display after a while. The temperature monitoring can be divided into two measuring method: measure through body surface temperature and through the temperature inside the body cavity (placed in mouth or anus).

Normal value: body surface: 36.5℃~37℃; inside body cavity: 36.5℃~37.7℃

53

User Manual for Vital Signs Monitor

Notes:

 Attach the TEMP transducer to the patient; generally if the TEMP transducer and skin doesn’t contact closely, the measured value becomes lower, so for those who have requirement for temperature, add a proper martial to transducer and fix it with adhesive tape to make them contact firmly.

 Especially for pediatric patient, they like sports, pay more attention to the transducer fixing.

54

User Manual for Vital Signs Monitor

Chapter 9 Troubleshooting

9.1 No Display on the Screen

Shut down the machine and unplug the power. Use a universal meter to check if the outlet has proper voltage, if the power cable is in good condition, and if the power cable is properly connected with this apparatus or outlet. Remove the fuse from the back cover of this machine, and make sure it is in good condition.

9.2 Excessive ECG Signal Interference or Too Thick Baseline

1. Check if the plate electrodes are properly located, and if valid plate electrodes are used.

2. Check whether the lead wires are inserted properly. If no ECG curve is displayed, check if the ECG lead wires are broken.

3. Make sure the mains outlet has standard grounding wire.

4. Check if the grounding wire of the apparatus is properly grounded.

9.3 No Blood Pressure and Pulse Oxygen Measures

1. Check if the blood pressure cuff is properly wrapped around the arm according to the operating instructions, if the cuff leaks, and if the inlet is closely connected with the NIBP jack on the side panel. Check if the indicator of the pulse oxygen probe flashes and if the pulse oxygen probe is properly connected to the SpO

2

jack on the side panel.

2. If the problems still exist, please contact the local dealer.

9.4 Blank Printing Paper

1. Check whether the printing paper is installed with its face reversed. Please reinstall it and let the sensitive page face upward.

2. If the problems still exist, please contact the local dealer.

9.5 System Alarm

1. When the parameter value is higher or lower than the alarm limits, the alarm will ring. Please check whether the alarm limit value is proper or the condition of the patient.

2. Probe off. Please check the connection of the probes.

Note: In case of trouble of this machine in the service, follow the instructions below to eliminate the problem first. If the attempt fails, contact the dealer in your local area or the manufacturer. Do not open the cabinet without permission.

55

User Manual for Vital Signs Monitor

Chapter 10 Maintenance

10.1 Service and Examination

10.1.1 Daily Examination

Before using the monitor, the checks below should be carried out:

Check the monitor for any mechanical damage;

Inspect the exposed parts and the inserted parts of all the leads, and the accessories;

Examine all the functions of the monitor that are likely to be used for patient monitoring, and ensure that it is in good working condition;

Make sure that the monitor is grounded properly.

Pay close attention to the fluctuation of the local power supply voltage. A manostat is recommended when necessary.

In case any indication of damage about the function of the monitor is detected and proven, it is not allowed to apply it to the patient for any monitoring.

10.1.2 Routine Maintenance

After each maintenance or the yearly maintenance, the monitor can be thoroughly inspected by qualified personnel, including function and safety examinations. The designed life of this monitor is 5 years. In order to ensure its long service life, please pay attention to the maintenance.

If the hospital fails to carry out a satisfactory maintenance program about the monitor, it may

get disabled and harm the patient’s safety and health.

In case of ECG leads damage or aging, please replace the lead.

If there is any indication of cable and transducer damage or they deteriorate, they are

prohibited from any further use.

The adjustable units in the monitor such as potentiometer are not allowed to adjust without

permission to avoid unnecessary failures that affect normal application.

10.1.3 Battery Maintenance

Please pay attention to the polarity of battery, do NOT insert it into battery compartment with

reversed polarities;

Do NOT use the batteries manufactured by other companies, if being inserted, the device will

may be damaged;

In order to avoid damaging the battery, do NOT use other power supply device to charge the

battery;

56

User Manual for Vital Signs Monitor

After battery ageing phenomenon occurring, to avoid explosion risk do NOT throw the

battery into fire.

Do not hit or strike it with force;

Do not use this battery on other devices;

Do not use this battery below -10 or above 40;

Dispose of the battery, the local law should be followed.

In order to maintain battery supply time and prolong battery lifetime, please charge the

battery every one or two months if don’t use battery for a long time. And do charge battery at least 12-15 hours every time. Before connect to AC, do start monitor with battery’s power supply, until battery power is used up and monitor turn off automatically, then connect monitor to AC and have it charged for 12-15 hours continuously. The speed of charge will be the same no matter whether the monitor is working or not. The reason why discharge the battery before charge is to avoid the decrease of capacity caused by battery’s memory effect.

If the monitor won’t be used for a long time, do have it charged fully before conservation.

When starting the monitor by battery power only which is short of supply, monitor will turn

off automatically. In order to avoid the damage to battery caused by excessive discharge, please pay attention to following. After monitor turns off automatically, there is still small drain current inside battery, so it is suggested that user should press the power button again to cut off the power supply. If battery keeps in a state of small drain current, battery will be damaged and can’t be repaired because of excessive discharged. It is recommended to use the battery once a month to ensure its strong power supply capacity and long service life, and recharge it after running out of the power.

If battery is damaged, please replace with same type and specification battery marked by

“CCC” or “CE” in time, or contact the company directly

10.1.4 Service

If the monitor has functional malfunction or is not working, please contact the local dealer or our company, and we are to offer the best solution as soon as possible for your satisfaction. Only qualified service engineer specified by the manufacture can perform the service. Users are not permitted to repair it by themselves.

10.2 Cleaning, Sterilization and Disinfection

 Kept the monitor from dust.

 It is recommended to clean the outer shell and screen of the monitor to keep it clean. Only non-corrosive cleanser such as clear water is permitted.

 Use the cloth with alcohol to wipe the surface of the monitor and transducers, and dry it with dry and clean cloth or simply air-dry.

 The monitor can be sterilized and disinfected, please clean it first.

Switch off the monitor and disconnect the power cable before cleaning.

Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage.

57

User Manual for Vital Signs Monitor

Clean the exterior of the connector only.

Dilute the cleanser.

Do not let any liquid flow into the shell or any parts of the monitor.

Do not let the cleanser and disinfectant stay on its surface.

Do not perform high pressure sterilization to the monitor.

Do not put any parts of the monitor or its accessories in the liquid.

If the monitor is accidentally wetted it should be thoroughly dried before use. The rear cover

can be removed by qualified service technician to verify absence of water.

Do not pour the disinfector on its surface while sterilization.

10.3 Cleaning, Sterilization and Disinfection of Accessories

It is recommended to clean the accessories (including sensor, leads and plugs) with a piece of gauze which has been soaked in 75% Alcohol or 70% Ispropanol before use.

Do not use damaged accessories.

Accessories can not be entirely immerged into water, liquor or cleanser.

Do not use radial, steam or epoxyethane to disinfect accessories.

Do wipe off the remained alcohol or ispropanol on the accessories after disinfection, for good

maintenance can extend the life of accessories.

10.4 Storage

If the equipment will not be used for long period of time, wipe it clean and keep it in the packaging, which shall be kept in a dry and good ventilation place free from dust and corrosive gases.

Storage environment: ambient temperature: -20~60°C relative humidity: 10%~95% atmosphere: 50kPa~107.4kPa

10.5 Transportation

This monitor should be transported by land (vehicle or railway) or air in accordance with the contractual terms. Do not hit or drop it with force.

58

User Manual for Vital Signs Monitor

Chapter 11 Appendix

11.1 Prompt information explanations

Mute C-D: XXX seconds Alarm silence count down: XXX seconds

NIBP C-D: XXX seconds NIBP auto measuring cycle count down: XXX seconds

TOUR C-D: XXX seconds

Probe off

PR over limit

SpO

2

over limit

Tourniquet alert count down: XXX seconds

SpO

2

PR value exceeds the high/low alarm limit

SpO

2

probe fells off

value exceeds the high/low alarm limit

SYS over limit

DIA over limit

Systolic pressure value exceeds the high/low alarm limit

Diastolic pressure value exceeds the high/low alarm limit

MAP over limit

NIBP error 1#

NIBP error 2#

NIBP error 3#

NIBP error 4#

Cuff error

NIBP error 5#

Air leak

NIBP over range

Over motion

Over pressure

NIBP timeout

MAP value exceeds the high/low alarm limit

Sensor or other hardware error

Very weak signal because of the cuff, or the patient has very weak pulse

Blood pressure amplifier overflow due to excessive movement

Leaking during the pneumatic device testing

Cuff is not wrapped correctly, or is not connected

Abnormal condition of CPU, such as register overflow, divided by zero

Air moving part, tube or the cuff leak air

The measurement range exceeds 255mmHg (for neonates: over 135

The repeated measurement due to moving, excessive noise during the stepping inflation and measuring pressure and pulse, e.g. during patient shaking motion

Cuff press exceeds the safety limit value of software. Limit value for adult:

290mmHg; Limit value for pediatric: 145mmHg;

Or caused by cuff extrusion or flapping cuff with force.

Adult measurement is more than 120 seconds, neonate pediatric measurement is more

59

User Manual for Vital Signs Monitor

11.2 Default Alarming Values and Setup Range

Mode

Parameter

High

HR

Low

High

SYS

Low

High

DIA

Low

High

MAP

Low

High

SpO

2

Low

High

Pulse rate

Low

High

TEMP

Low

Adult

180bpm

40bpm

180mmHg

60mmHg

120mmHg

50mmHg

160mmHg

50mmHg

100%

90%

180bpm

40bpm

39.0℃

35.0℃

Default

Pediatric

200bpm

50bpm

130mmHg

50mmHg

90mmHg

40mmHg

110mmHg

40mmHg

100%

85%

200bpm

50bpm

39.0℃

35.0℃

60

Neonate

220bpm

50bpm

110mmHg

50mmHg

90mmHg

30mmHg

100mmHg

30mmHg

100%

85%

220bpm

50bpm

39.0℃

35.0℃

User Manual for Vital Signs Monitor

11.3 Abbreviation of arrhythmia

1. ECG TACHY

2. ECG BRADY

3.

ECG VPCEST

4. MISS BEAT

5.

VE EARLY

6. SVE EARLY

7. VE COUPLET

8.

SVE COUPLET

9. VE RUN

10.

SVE RUN

11.

VE SHORT RUN

12. SVE SHORT RUN

13.

VE BIGEMINY

14.

SVE BIGEMINY

15. VE TRIGEMINY

16.

SVE TRIGEMINY

17. VE INSERT

18. SVE INSERT

19.

VE RONT

20. SVE RONT

61

User Manual for Vital Signs Monitor

11.4 Accessories List

Part No.

15010513

5101-0101310

15044051

15044061

15044041

15044063

15024402

15021402

15022402

15023402

15020400

15084120

5101-5236310

2903-0000000

2911-0003032

900093

Part Name

ECG cable

ECG electrode

Adult SpO

2

Finger clip sensor

Adult SpO

2

Finger rubber sensor

Pediatric SpO

2

Finger clip sensor

Neonate SpO

2

Y-type sensor

Adult NIBP cuff(25~35cm)

Small-sized Pediatric NIBP cuff

Middle-sized Pediatric NIBP cuff

Large-sized Pediatric NIBP cuff

Neonate NIBP cuff(5.4*9.1cm)

Skin TEMP probe

Thermal printer paper

Power cord

Grounding wire

Net wire

Remark

Optional

Optional

Optional

Optional

Optional

Optional

Optional

Optional

Optional

For more information regarding the accessories, please contact your local sales representative or the manufacturer.

Note: Part no. is subject to change without prior notice, please refer to the label of parts or packlist.

62

User Manual for Vital Signs Monitor

11.5 Instructions for SpO

2

Probe

Instructions for Neonate SpO

2

Y-type Sensor

Intended Use

When used with a compatible patient monitor or a pulse oximeter device, this sensor is intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO

2

) and pulse rate monitoring for neonates (1-3 kg).

Contraindications

This sensor is contraindicated for use on active patients or for prolonged use.

Instructions for Use

1) Insert the two sensor tips into the slots on the rubber wrap (A); place the sensor on the neonate’s foot (B), wrap the rubber belt around the foot and tighten accordingly (C).

2) Plug the sensor into the oximeter and verify proper operation as described in the oximeter operator’s manual.

3) Inspect the monitoring site every 4 hours for skin integrity.

Cleaning & Disinfection

Unplug the sensor before cleaning or disinfecting. Surface-clean sensor and cable with a soft gauze pad by saturating it with a solution such as 70% isopropyl alcohol. If low-level disinfection is required, use a

1:10 bleach solution.

Warnings

1) Some factors may affect the accuracy of saturation measurements. Such factors include: excessive

2) patient motion, fingernail polish, use of intravascular dyes, excessive light, poorly perfused finger, extreme finger sizes or improper placement of the sensor.

3) The sensor must be checked for skin integrity at least every 4 hours. Because individual skin condition affects the ability of the skin to tolerate sensor placement, it may be necessary to change the sensor site.

4) Do not use NIBP or other constructing instruments on same appendage as sensor for blood flow interrupted by NIBP cuff or circulatory patient condition will result in no pulse found or loss of pulse.

5) Do not use the sensor during MRI scanning. Carefully route cables to reduce the possibility of patient entanglement or strangulation.

6) Do not alter or modify the sensor. Alterations or modifications may affect performance or accuracy.

63

User Manual for Vital Signs Monitor

7) Do not use the sensor if the sensor or the sensor cable appears damaged.

Caution: Do not sterilize by irradiation steam, or ethylene oxide

Instructions for Pediatric SpO

2

Finger Clip Sensor

Intended Use

When used with a compatible patient monitor or a pulse oximeter device, the sensor is intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO

2

) and pulse rate monitoring for pediatric patients weighing between 10~40kg.

Contraindications

This sensor is contraindicated for use on active patients or for prolonged use.

Instructions for Use

1) With the upper and lower jaws open, place an index finger evenly on the base of the clip. Push the finger tip against the stop so that it is over the sensor window (A). If an index finger cannot be positioned correctly, or is not available, other fingers can be used.

2) Note: When selecting a sensor site, priority should be given to an extremity free of an arterial catheter, blood pressure cuff, or intravascular infusion line.

3) Spread open the rear tabs of the sensor to provide even force over the length of the pads (B).

4) The sensor should be oriented in such a way that the cable is positioned along the top of the hand

(C).

5) Plug the sensor into the oximeter and verify proper operation as described in the oximeter operator’s manual.

6) Inspect the monitoring site every 4 hours for skin integrity.

7) Before each use, surface-clean sensor and cable with a soft gauze pad by saturating it with a solution such as 70% isopropyl alcohol. If low-level disinfection is required, use a 1:10 bleach solution.

Caution: Do not sterilize by irradiation steam, or ethylene oxide.

Warnings

8) Some factors may affect the accuracy of saturation measurements. Such factors include: excessive patient motion, fingernail polish, use of intravascular dyes, excessive light, poor blood perfusion in the finger, extreme finger sizes or improper placement of the sensor.

9) Using the sensor in the presence of bright lights may result in inaccurate measurements. In such cases, cover the sensor site with an opaque material.

10) The sensor must be moved to a new site at least every 4 hours. Because individual skin condition affects the ability of the skin to tolerate sensor placement, it may be necessary to change the sensor site more frequently with some patients. If skin integrity changes, move the sensor to another site.

11) Do not apply tape to secure the sensor in place or to tape it shut; venous pulsation may lead to inaccurate saturation measurements.

12) Do not immerse sensor as it causes short.

13) Do not use NIBP or other constructing instruments on same appendage as sensor for blood flow interrupted by NIBP cuff or circulatory patient condition will result in no pulse found or loss of pulse.

64

User Manual for Vital Signs Monitor

14) Do not use the sensor or other oximetry sensors during MRI scanning.

15) Carefully route cables to reduce the possibility of patient entanglement or strangulation.

16) Do not alter or modify the sensor. Alterations or modifications may affect performance or accuracy.

17) Do not use the sensor if the sensor or the sensor cable appears damaged.

Instructions for Adult SpO

2

Finger Rubber Sensor

Intended Use

When used with a compatible patient monitor or a pulse oximeter device, this SpO

2

sensor is intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO

2

) and pulse rate monitoring for patients weighing greater than 50kg.

Contraindications

This sensor is contraindicated for use on active patients or for prolonged use.

Instructions for Use

1) Hold the sensor with its opening towards the patient’s index finger (A). The sensor should be oriented in such a way that the sensor side with a finger tip sign is positioned on the top.

2) Insert the patient’s index finger into the sensor until the fingernail tip rests against the stop at the end of the sensor. Adjust the finger to be placed evenly on the middle base of the sensor. Direct the cable along the top of the patient’s hand. Apply adhesive tape to secure the cable (B). If an index finger cannot be positioned correctly, or is not available, other fingers can be used.

3) Plug the sensor into the oximeter and verify proper operation as described in the oximeter operator’s manual.

4) Inspect the monitoring site every 4 hours for skin integrity.

Cleaning & Disinfection

Unplug the sensor before cleaning or disinfecting. Surface-clean sensor and cable with a soft gauze pad by saturating it with a solution such as 70% isopropyl alcohol. If low-level disinfection is required, use a

1:10 bleach solution.

Caution: Do not sterilize by irradiation steam, or ethylene oxide.

Warnings

1) This sensor is for use only with compatible patient monitors or pulse oximeter devices. Use of this sensor with instruments other than compatibles may result in improper performance.

2) Some factors may affect the accuracy of saturation measurements. Such factors include: excessive patient motion, fingernail polish, use of intravascular dyes, excessive light, poorly perfused finger, extreme finger sizes or improper placement of the sensor.

3) The sensor site must be checked for skin integrity at least every 4 hours. Because individual skin condition affects the ability of the skin to tolerate sensor placement, it may be necessary to change the sensor to another finger.

65

User Manual for Vital Signs Monitor

4) Do not use NIBP or other constructing instruments on same appendage as sensor for blood flow interrupted by NIBP cuff or circulatory patient condition will result in no pulse found or loss of pulse.

Do not use the sensor during MRI scanning.

5) Carefully route cables to reduce the possibility of patient entanglement or strangulation.

6) Do not alter or modify the sensor. Alterations or modifications may affect performance or accuracy.

7) Do not use the sensor if the sensor or the sensor cable appears damaged.

Instructions for Adult SpO

2

Finger Clip Sensor

Intended Use

When used with a compatible patient monitor or a pulse oximeter device, the sensor is intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO

2

) and pulse rate monitoring for patients weighing greater than 40kg.

Contraindications

This sensor is contraindicated for use on active patients or for prolonged use.

Instructions for Use

1) With the upper and lower jaws open, place an index finger evenly on the base of the clip. Push the finger tip against the stop so that it is over the sensor window (A). If an index finger cannot be positioned correctly, or is not available, other fingers can be used.

2) Note: When selecting a sensor site, priority should be given to an extremity free of an arterial catheter, blood pressure cuff, or intravascular infusion line.

3) Spread open the rear tabs of the sensor to provide even force over the length of the pads (B).

4) The sensor should be oriented in such a way that the cable is positioned along the top of the hand (C).

5) Plug the sensor into the oximeter and verify proper operation as described in the oximeter operator’s manual.

6) Inspect the monitoring site every 4 hours for skin integrity.

7) Before each use, surface-clean sensor and cable with a soft gauze pad by saturating it with a solution such as 70% isopropyl alcohol. If low-level disinfection is required, use a 1:10 bleach solution.

Caution: Do not sterilize by irradiation steam, or ethylene oxide.

Warnings

1) Some factors may affect the accuracy of saturation measurements. Such factors include: excessive patient motion, fingernail polish, use of intravascular dyes, excessive light, poorly perfused finger, extreme finger sizes or improper placement of the sensor.

2) Using the sensor in the presence of bright lights may result in inaccurate measurements. In such cases, cover the sensor site with an opaque material.

3) The sensor must be moved to a new site at least every 4 hours. Because individual skin condition affects the ability of the skin to tolerate sensor placement, it may be necessary to change the sensor site more frequently with some patients. If skin integrity changes, move the sensor to another site.

4) Do not apply tape to secure the sensor in place or to tape it shut; venous pulsation may lead to inaccurate saturation measurements.

5) Do not immerse sensor as it causes short.

66

User Manual for Vital Signs Monitor

6) Do not use NIBP or other constructing instruments on same appendage as sensor for blood flow interrupted by NIBP cuff or circulatory patient condition will result in no pulse found or loss of pulse.

7) Do not use the sensor or other oximetry sensors during MRI scanning.

8) Carefully route cables to reduce the possibility of patient entanglement or strangulation.

9) Do not alter or modify the sensor. Alterations or modifications may affect performance or accuracy.

10) Do not use the sensor if the sensor or the sensor cable appears damaged.

Creative offers a 6-month warranty against manufacturing defects for the SpO

2

sensors mentioned above in its undamaged condition.

If you have any question regarding any of SpO

2

sensor instructions, please contact info@ creative-sz.com, or your local dealer.

67

advertisement

Key Features

  • Measures SpO2, NIBP, ECG, PR, and temperature
  • Large, easy-to-read display
  • Multiple monitoring modes
  • Variety of alarms and alerts
  • Compact and portable
  • Battery-operated for portability
  • Defibrillator-proof

Frequently Answers and Questions

What parameters does the PC-900 measure?
The PC-900 measures SpO2, NIBP, ECG, PR, and temperature.
Is the PC-900 easy to use?
Yes, the PC-900 is designed to be user-friendly with a large, easy-to-read display and intuitive controls.
Is the PC-900 portable?
Yes, the PC-900 is compact and portable and can be easily transported between different locations.
Is the PC-900 battery-operated?
Yes, the PC-900 can be battery-operated for portability, making it ideal for use in remote or mobile settings.
Is the PC-900 defibrillator-proof?
Yes, the PC-900 is defibrillator-proof, ensuring patient safety during defibrillation procedures.

Related manuals

Download PDF

advertisement