Monitex WhitenMAX BR800 teeth whitening unit User manual

Monitex WhitenMAX BR800 teeth whitening unit User manual

The WhitenMAX BR800 is a professional teeth whitening unit that uses LED light to accelerate the bleaching process. It features two flux modes, a timer with adjustable settings, and a blue and red LED option. The unit is easy to use and designed for efficient and effective teeth whitening.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Monitex WhitenMAX BR800 User Manual | Manualzz

www.monitex.com.tw

Professional teeth whitening unit

User Manual

Table of Contents

Preface ------------------------------------------------------------ 1

Disclaimer -------------------------------------------------------- 1

Safety Precautions --------------------------------------------- 2

Product Parts --------------------------------------------------- 3

Assembly --------------------------------------------------------- 3

Operation -------------------------------------------------------- 10

Trouble Shooting Tips ---------------------------------------- 13

Specification ---------------------------------------------------- 17

Preface:

1. Thank you for choosing Monitex Product for your advance integration.

Prior to using Monitex WhitenMAX BR800 Professional Whitening

Unit, please read this User Manual carefully and keep this manual

well preserved for future reference.

2. Please contact your local dealer if any damage or problem occurs.

Attempt to try and repair this unit by unauthorized technician personnel

WILL automatically lead to the forfeit and negation of the warranty

agreement.

3. The specification are subject to change and revision without prior

notice.

Disclaimer:

1. Monitex shall not assume any responsibility if any malfunction or

damage was caused by improper operation, removal, modification

or maintenance.

2. We shall not assume any responsibility if any malfunction or damage

was caused by operating under unsuited environment.

3. We shall not assume any responsibility if any malfunction or damage

was caused by forceful majeure.

4. We reserve the right to explain for the given cause of the Defective

products which might have been caused by natural disasters or

abnormal temperature, humidity that is inconcurrent to the guidance

instructed in this manual.

1

Table of Contents

Preface ------------------------------------------------------------ 1

Disclaimer -------------------------------------------------------- 1

Safety Precautions --------------------------------------------- 2

Product Parts --------------------------------------------------- 3

Assembly --------------------------------------------------------- 3

Operation -------------------------------------------------------- 10

Trouble Shooting Tips ---------------------------------------- 13

Specification ---------------------------------------------------- 17

Preface:

1. Thank you for choosing Monitex Product for your advance integration.

Prior to using Monitex WhitenMAX BR800 Professional Whitening

Unit, please read this User Manual carefully and keep this manual

well preserved for future reference.

2. Please contact your local dealer if any damage or problem occurs.

Attempt to try and repair this unit by unauthorized technician personnel

WILL automatically lead to the forfeit and negation of the warranty

agreement.

3. The specification are subject to change and revision without prior

notice.

Disclaimer:

1. Monitex shall not assume any responsibility if any malfunction or

damage was caused by improper operation, removal, modification

or maintenance.

2. We shall not assume any responsibility if any malfunction or damage

was caused by operating under unsuited environment.

3. We shall not assume any responsibility if any malfunction or damage

was caused by forceful majeure.

4. We reserve the right to explain for the given cause of the Defective

products which might have been caused by natural disasters or

abnormal temperature, humidity that is inconcurrent to the guidance

instructed in this manual.

1

Safety Precaution:

Please follow instruction steps in this manual for correct and safe operations of the device. Pay more attention to the following four warning signs beside the operation descriptions where applicable:

Danger-This message appears where the operation may potentially

cause severe injury if instruction not followed correctly.

Warning-This message appears where the operation may cause

serious damage or defect to other objects.

Caution-This message appears where the operation may cause slight

damage to other objects or slight bodily harm if operation

not properly conducted.

Important-This message appears where the operation may involve

risk of defecting the device itself if not properly conducted.

Transport and storage Environment conditions:

Temperature: -20to 65

(-4

to 149

)

Humidity: 15-90%, non-condensing

Maintenance:

If there is any dirt or stains on the surface of control panel, metal parts, or mouth piece, please use dried cloth with Medical-use Alcohol to gently wipe clean. If there is a hard stain then please use a neutral cleaning agent for removal.

Important: Never use a solvent or volatile oil for cleaning. This

can lead to fading in the paint, or display board.

Special Symbols:

Symbol Explanation

Type B applied part

Attention, consult

ACCOMPANYING DOCUMENTS

Signifies the device has met all essential requirements of

European Medical Device Directive 93/42/EEC.

Symbol used to indicate the polarity of the DC power input.

Direct Current

Recycle

2

Product Parts:

Packaging:

B

F

A

C

D

A - Pentagular Mobile Cart

B - N shape Iron Stand

C - Operating Panel

D - Rotating Arm

E - Light Unit and Mouth Piece

F - Localized Plug Adaptor (Europe

Plug built-in and us plug on option,

please see page 9)

G - Iron Rim

H - Plastic Washer Rim

I - Protective Glasses

I

G H

E

Assembly & Dis-assembly:

Caution: Please always space the site for your precaution before executing your installation as some of the iron-made parts can be damaged of its paint,or damaging to other equipment if it’s handled improperly.

A. Pentagular Mobile Cart:

Place Mobile Cart in an open space for preparation of assembly.

3

Safety Precaution:

Please follow instruction steps in this manual for correct and safe operations of the device. Pay more attention to the following four warning signs beside the operation descriptions where applicable:

Danger-This message appears where the operation may potentially

cause severe injury if instruction not followed correctly.

Warning-This message appears where the operation may cause

serious damage or defect to other objects.

Caution-This message appears where the operation may cause slight

damage to other objects or slight bodily harm if operation

not properly conducted.

Important-This message appears where the operation may involve

risk of defecting the device itself if not properly conducted.

Transport and storage Environment conditions:

Temperature: -20to 65

(-4

to 149

)

Humidity: 15-90%, non-condensing

Maintenance:

If there is any dirt or stains on the surface of control panel, metal parts, or mouth piece, please use dried cloth with Medical-use Alcohol to gently wipe clean. If there is a hard stain then please use a neutral cleaning agent for removal.

Important: Never use a solvent or volatile oil for cleaning. This

can lead to fading in the paint, or display board.

Special Symbols:

Symbol Explanation

Type B applied part

Attention, consult

ACCOMPANYING DOCUMENTS

Signifies the device has met all essential requirements of

European Medical Device Directive 93/42/EEC.

Symbol used to indicate the polarity of the DC power input.

Direct Current

Recycle

2

Product Parts:

Packaging:

B

F

A

C

D

A - Pentagular Mobile Cart

B - N shape Iron Stand

C - Operating Panel

D - Rotating Arm

E - Light Unit and Mouth Piece

F - Localized Plug Adaptor (Europe

Plug built-in and us plug on option,

please see page 9)

G - Iron Rim

H - Plastic Washer Rim

I - Protective Glasses

I

G H

E

Assembly & Dis-assembly:

Caution: Please always space the site for your precaution before executing your installation as some of the iron-made parts can be damaged of its paint,or damaging to other equipment if it’s handled improperly.

A. Pentagular Mobile Cart:

Place Mobile Cart in an open space for preparation of assembly.

3

B. Assembly of N-shape Iron Stand & Pentagular Mobile Cart:

onto the Pentagular Mobile Cart.

C. Assembly of N-Shape Iron Stand & the Rotating Arm:

> Place the Rotating Arm in the open area for preparing installation.

> Place the Iron Rim onto the bottom of the N-Shape Iron Stand.

> Place the Plastic Washer Rim inside the rim for secure purpose.

> Please repeat the procedure stated previously by inserting the

Silver Iron Rim and the Plastic Washer Rim onto the top of the

N-Shape Iron Stand to receive clockwise turning for secure

screw-lock.

> Pull out the plug from inside

the N- shape Iron Stand and

connect the plug onto the jack

pre-installed inside the

Pentagular Mobile Cart.

4 5

B. Assembly of N-shape Iron Stand & Pentagular Mobile Cart:

onto the Pentagular Mobile Cart.

C. Assembly of N-Shape Iron Stand & the Rotating Arm:

> Place the Rotating Arm in the open area for preparing installation.

> Place the Iron Rim onto the bottom of the N-Shape Iron Stand.

> Place the Plastic Washer Rim inside the rim for secure purpose.

> Please repeat the procedure stated previously by inserting the

Silver Iron Rim and the Plastic Washer Rim onto the top of the

N-Shape Iron Stand to receive clockwise turning for secure

screw-lock.

> Pull out the plug from inside

the N- shape Iron Stand and

connect the plug onto the jack

pre-installed inside the

Pentagular Mobile Cart.

4 5

>Plug-in the Power Cable and the Connecting cable between Light Unit

and the Display Panel check and make sure all connection fixed well!

Plug in the Power Cable to the Display Panel.

D. Assembly of the Rotating Arm & the Light Unit:

> Please prepare the Light

Unit with protect Mouth

Piece for installation.

Plug in Connecting Cable between Light Unit & the

Display Panel.

> Plug-in the extended Connecting Cable from the Rotating Arm to

the extended plug located in the top of the N-Shape Iron Stand to

complete the connection between the Display Panel & the Light Unit.

> Make sure the Mouth Piece

is properly placed onto the

Light Unit.

Caution: Please make sure always to disinfect the Light Unit

and Mouth Piece with medical-use alcohol before

carrying any operation.

> Follow the instruction illustrated below by placing the Iron Rim

and the Plastic Washer Rim to secure the light unit onto the

Rotating Arm.

6 7

>Plug-in the Power Cable and the Connecting cable between Light Unit

and the Display Panel check and make sure all connection fixed well!

Plug in the Power Cable to the Display Panel.

D. Assembly of the Rotating Arm & the Light Unit:

> Please prepare the Light

Unit with protect Mouth

Piece for installation.

Plug in Connecting Cable between Light Unit & the

Display Panel.

> Plug-in the extended Connecting Cable from the Rotating Arm to

the extended plug located in the top of the N-Shape Iron Stand to

complete the connection between the Display Panel & the Light Unit.

> Make sure the Mouth Piece

is properly placed onto the

Light Unit.

Caution: Please make sure always to disinfect the Light Unit

and Mouth Piece with medical-use alcohol before

carrying any operation.

> Follow the instruction illustrated below by placing the Iron Rim

and the Plastic Washer Rim to secure the light unit onto the

Rotating Arm.

6 7

> Turn the Iron Rim by clockwise to fix the support of Light Unit.

Important: Please make sure to position the Light Unit tightly

& securely to prevent Light Unit from falling off or

cause unnecessary damage.

Plug-in:

Make sure to plug in the Light Unit from the cable extended from the Rotating Arm:

Wheels:

Congratulation! You have successfully assembled your BR-800

WhitenMAX LED Bleaching Light Unit.

Important:

There are 3 handle locks located in 3 of the 5 wheels under the Mobile Cart.

After safely placing BR-800 to your appointed location or chair-side , please press down the handle locks to secure your unit and prevent BR-800 from mobility.

To set it free for mobile, simply unlock it by lifting the handle lock in accordance to the following picture!

Power Plug:

Each BR-800 is built-in with a 100 ~ 240V Power adaptor inside of the Pentagular Mobile Cart. The preset extension cord is European Plug.

WARNING : Beware of high temperature and strong light. Do not look directly into the light source if without using any eye protection.

WARNING : Light unit is a heat generating source. The temperature

º might reach to 65 C for long term usage. Beware of high temperature.

8

Additional power plug adaptor is available on request as optional accessories.

Please store the packaging materials safely for future use!

Perchance if the unit is required to be disassembled for transportation or sent for repair, and it’s requested to reverse carefully the above procedures and store the parts in the original packaging material for delivery and/or storage.

9

> Turn the Iron Rim by clockwise to fix the support of Light Unit.

Important: Please make sure to position the Light Unit tightly

& securely to prevent Light Unit from falling off or

cause unnecessary damage.

Plug-in:

Make sure to plug in the Light Unit from the cable extended from the Rotating Arm:

Wheels:

Congratulation! You have successfully assembled your BR-800

WhitenMAX LED Bleaching Light Unit.

Important:

There are 3 handle locks located in 3 of the 5 wheels under the Mobile Cart.

After safely placing BR-800 to your appointed location or chair-side , please press down the handle locks to secure your unit and prevent BR-800 from mobility.

To set it free for mobile, simply unlock it by lifting the handle lock in accordance to the following picture!

Power Plug:

Each BR-800 is built-in with a 100 ~ 240V Power adaptor inside of the Pentagular Mobile Cart. The preset extension cord is European Plug.

WARNING : Beware of high temperature and strong light. Do not look directly into the light source if without using any eye protection.

WARNING : Light unit is a heat generating source. The temperature

º might reach to 65 C for long term usage. Beware of high temperature.

8

Additional power plug adaptor is available on request as optional accessories.

Please store the packaging materials safely for future use!

Perchance if the unit is required to be disassembled for transportation or sent for repair, and it’s requested to reverse carefully the above procedures and store the parts in the original packaging material for delivery and/or storage.

9

Operation:

Functional Introduction:

Please see below diagram and description for instruction of use.

High

Min Sec

BR-800

Low

Monitor

Control Panel

Danger:

Please do not look directly into the lights, nor acquiring the emitting light directly onto patient. Such action can cause a danger or harm to patients’ eyes

10

Monitor:

Selection of

Light Source

Min Sec

BR-800

Timer

The timer first starts count down upon push “Start/Stop” button.

The “Min” indicate the number of minutes currently remain of.

The “Sec” indicate the last minute of count-down.

Pre-Set Time is 10 Minutes

Indication of your

Mode Selection:

High: Mode 1

Low: Mode 2

High

BR-800

Low

Over Heat Indicator:

When the temperature reaches 40ºC, it automatically activate fan motor for heat dissipating .

Important:

When the temperature reaches 70ºC, as our protection mechanism, the build-in Thermometers will start to blink on the drop while the unit ceases functioning till the temperature climb down.

11

Operation:

Functional Introduction:

Please see below diagram and description for instruction of use.

High

Min Sec

BR-800

Low

Monitor

Control Panel

Danger:

Please do not look directly into the lights, nor acquiring the emitting light directly onto patient. Such action can cause a danger or harm to patients’ eyes

10

Monitor:

Selection of

Light Source

Min Sec

BR-800

Timer

The timer first starts count down upon push “Start/Stop” button.

The “Min” indicate the number of minutes currently remain of.

The “Sec” indicate the last minute of count-down.

Pre-Set Time is 10 Minutes

Indication of your

Mode Selection:

High: Mode 1

Low: Mode 2

High

BR-800

Low

Over Heat Indicator:

When the temperature reaches 40ºC, it automatically activate fan motor for heat dissipating .

Important:

When the temperature reaches 70ºC, as our protection mechanism, the build-in Thermometers will start to blink on the drop while the unit ceases functioning till the temperature climb down.

11

Control Panel:

Power Button

LED Dual Color Display:

Green: Standby Mode

Orange: Push the power button to enter

Operation mode.

Mode 1 / Mode 2 :

Two Flux Modes

Illumination

Mode 1 (High):

100% Full Power Output

Mode 2 (Low):

70% Power Output

Ajustable Whitening

Timer:

TIME Adjustable: up to 1~20 Minutes.

Adjusting interval:

1 minute per push.

Light Source Selection:

Blue emits Medical use LED Cool Blue light.

Red emits Red light to

Accelerate the bleaching process.

B&R signifies Blue &

Red Light turns on

Simultaneously.

P.S. There are a total of

4 x 3W Blue Led bulb &

1 x 3W Red Led bulb built-in.

Start & Stop Button

One press START/STOP initiates the operation.The following press terminates the operation.

P.S. BR800 enters Stand by Mode when in idled for more than 1 minute.

12

Trouble Shooting & FAQ:

1. LED malfunctions:

What happens if the Blue or Red LED does not work?

Possible Cause: BR800 Blue LED are connected in serial manner.

Blue LEDs are connected together, and Red LED is connected individually. If one of the Blue LED does not work, all four will cease function, while the Red LED remain functioning individually.

Solution: Please send BR800 back to the supplier for repair.

What happens if none of the LED work?

Possible Cause 1:

The connection between the light unit and the display panel might be off. Please check for their connection status as ref to pictures below .

Possible Cause 2:

The connection between the light unit and the pentagular Pentagular Mobile Cart cart might be off.

Please check their connection status as shown on the right picture.

13

Control Panel:

Power Button

LED Dual Color Display:

Green: Standby Mode

Orange: Push the power button to enter

Operation mode.

Mode 1 / Mode 2 :

Two Flux Modes

Illumination

Mode 1 (High):

100% Full Power Output

Mode 2 (Low):

70% Power Output

Ajustable Whitening

Timer:

TIME Adjustable: up to 1~20 Minutes.

Adjusting interval:

1 minute per push.

Light Source Selection:

Blue emits Medical use LED Cool Blue light.

Red emits Red light to

Accelerate the bleaching process.

B&R signifies Blue &

Red Light turns on

Simultaneously.

P.S. There are a total of

4 x 3W Blue Led bulb &

1 x 3W Red Led bulb built-in.

Start & Stop Button

One press START/STOP initiates the operation.The following press terminates the operation.

P.S. BR800 enters Stand by Mode when in idled for more than 1 minute.

12

Trouble Shooting & FAQ:

1. LED malfunctions:

What happens if the Blue or Red LED does not work?

Possible Cause: BR800 Blue LED are connected in serial manner.

Blue LEDs are connected together, and Red LED is connected individually. If one of the Blue LED does not work, all four will cease function, while the Red LED remain functioning individually.

Solution: Please send BR800 back to the supplier for repair.

What happens if none of the LED work?

Possible Cause 1:

The connection between the light unit and the display panel might be off. Please check for their connection status as ref to pictures below .

Possible Cause 2:

The connection between the light unit and the pentagular Pentagular Mobile Cart cart might be off.

Please check their connection status as shown on the right picture.

13

Possible Cause 3 :

The power is off, please connect the plug back to the wall power socket, or make sure that the electricity is coming out of the socket before plugging in.

Possible Cause 4:

If the cause 3 does not fit to the scenario, a doubt of the built-in adaptor is broken. Please send the unit back to your local dealer for repair.

2. Why can I not fix the adjusted angle of the light unit?

Important:

The active angle suitable for the weight of our light unit is 60º vertical angle

(see right pic.).

Please maintain the working angle within this range.

Possible Cause 1:

There are two knobs on each side of the lamp unit. Only 1 of either knob is needed to be secured to enable the locking mechanism to work. If both knobs are secured, the locking mechanism will fail.

Possible Cause 2:

If the root cause of above can not be determined please send the unit back to your nearest local dealer for re-adjustment. Thank you.

14

3. Why can I not operate the lamp when pressing the

power button?

Possible Cause : If the power cord has plugged into the Wall Plug, but the Power button does still not turn to green indicator (stand by mode) please check for the power of the wall plug and replug the Display Panel.

If the power plug in the wall is normal, please send the unit back to your nearest dealer for examination.

4. Why is that my Display Panel or the buttons do not work?

Possible Cause:

If by checking after all connection, and plugging in to the Wall Plug, the display panel still does not function, please check for the power of the wall plug and replug the Display Panel.

If the wall plug is normal, please send the unit back to your nearest dealer for examination.

5. Unable to ultilize Pentagular Push Cart freely !

Please check if the Lock

Switch are all released & the handle lifting upwards!

15

Possible Cause 3 :

The power is off, please connect the plug back to the wall power socket, or make sure that the electricity is coming out of the socket before plugging in.

Possible Cause 4:

If the cause 3 does not fit to the scenario, a doubt of the built-in adaptor is broken. Please send the unit back to your local dealer for repair.

2. Why can I not fix the adjusted angle of the light unit?

Important:

The active angle suitable for the weight of our light unit is 60º vertical angle

(see right pic.).

Please maintain the working angle within this range.

Possible Cause 1:

There are two knobs on each side of the lamp unit. Only 1 of either knob is needed to be secured to enable the locking mechanism to work. If both knobs are secured, the locking mechanism will fail.

Possible Cause 2:

If the root cause of above can not be determined please send the unit back to your nearest local dealer for re-adjustment. Thank you.

14

3. Why can I not operate the lamp when pressing the

power button?

Possible Cause : If the power cord has plugged into the Wall Plug, but the Power button does still not turn to green indicator (stand by mode) please check for the power of the wall plug and replug the Display Panel.

If the power plug in the wall is normal, please send the unit back to your nearest dealer for examination.

4. Why is that my Display Panel or the buttons do not work?

Possible Cause:

If by checking after all connection, and plugging in to the Wall Plug, the display panel still does not function, please check for the power of the wall plug and replug the Display Panel.

If the wall plug is normal, please send the unit back to your nearest dealer for examination.

5. Unable to ultilize Pentagular Push Cart freely !

Please check if the Lock

Switch are all released & the handle lifting upwards!

15

6. Abnormal Burning smell occurs.

Check from where the Burning smell.

If the smell is coming from the machine. Please send back to your nearest dealer for examination.

Thank you.

7. Why was my BR800 overheated within operation of less

than 20 minutes?

Possible Cause :

Listen to the sound of the built-in motor fan. This fan should be normally activated when the temperature reaches 40ºC.

If the fan motor does not function, the thermo sensor will indicate

Overheating sign to shut off

BR800 for protection. Most likely to be taken within 20 minutes.

Please send the unit back to the dealer for further checking and repair.

16

Specification:

Output Spectrum : 450 ~ 470 nm & 620 ~ 630 nm

Voltage : AC 100~240V 50/60HZ

Power : 19 W

Weight : 9.4 kgs net.

Dimensions on :

Total Arm Length : 960 mm

Total Height : 1245 mm

17

6. Abnormal Burning smell occurs.

Check from where the Burning smell.

If the smell is coming from the machine. Please send back to your nearest dealer for examination.

Thank you.

7. Why was my BR800 overheated within operation of less

than 20 minutes?

Possible Cause :

Listen to the sound of the built-in motor fan. This fan should be normally activated when the temperature reaches 40ºC.

If the fan motor does not function, the thermo sensor will indicate

Overheating sign to shut off

BR800 for protection. Most likely to be taken within 20 minutes.

Please send the unit back to the dealer for further checking and repair.

16

Specification:

Output Spectrum : 450 ~ 470 nm & 620 ~ 630 nm

Voltage : AC 100~240V 50/60HZ

Power : 19 W

Weight : 9.4 kgs net.

Dimensions on :

Total Arm Length : 960 mm

Total Height : 1245 mm

17

INDUSTRIAL CO., LTD

6F, 70, SEC. 1, ,

CITY, 24158, TAIPEI COUNTY, TAIWAN

TEL : 886-2-29993869 FAX: 886-2-29993647

E-mail : [email protected]

http://www.monitex.com.tw

advertisement

Key Features

  • Two flux modes (High & Low)
  • Timer with adjustable settings (1-20 minutes)
  • Blue and red LED option
  • Easy to use
  • Mobile cart design

Frequently Answers and Questions

What happens if the Blue or Red LED does not work?
If one of the Blue LEDs does not work, all four will cease function, while the Red LED remain functioning individually. Please send BR800 back to the supplier for repair.
Why can I not fix the adjusted angle of the light unit?
There are two knobs on each side of the lamp unit. Only 1 of either knob is needed to be secured to enable the locking mechanism to work. If both knobs are secured, the locking mechanism will fail.
Why can I not operate the lamp when pressing the power button?
Please check if the connection between the light unit and the display panel is properly connected and if the power cord is plugged into a functioning wall outlet. If these checks are successful, please send the unit back to your nearest dealer for examination.

Related manuals

Download PDF

advertisement