9113A
– Input for RTD, TC and mA
– Active / passive mA output via the same two terminals
– 1 or 2 channels
– Can be supplied separately or installed on power rail, PR type 9400
– SIL 2-certified via Full Assessment
Advanced features
• Configuration and monitoring by way of detachable display
front (PR 4511/4501); process calibration and signal
simulation.
• Copying of the configuration from one device to others of the
same type via the display front.
• TC inputs can use either the internal CJC or a terminal with a
built-in Pt100 sensor (PR 5910, channel 1 / PR 5913, channel
2) for higher accuracy.
• Advanced monitoring of internal communication and stored
data.
• SIL 2 functionality is optional and must be activated in a menu
point.
Applications
Application
• The device can be mounted in and receive signals from nonclassified area and zone 2.
• Conversion and scaling of temperature (Pt, Ni and TC) and
active current signals.
• 9113A has been designed, developed and certified for use in
SIL 2 applications according to the requirements of IEC
61508.
Technical characteristics
• 1 green and 2 red front LEDs indicate operation status and
malfunction.
• 2.6 kVAC galvanic isolation between input, output and supply.
Mounting
• The devices can be mounted vertically or horizontally without
distance between neighbouring units.
This page is automatically generated on the basis of information provided on www.prelectronics.com and affiliated websites. It is provided to you as a service and for information
purpose only. While we have attempted to maintain the information as accurately as possible, the page may contain errors or omissions for which we disclaim any and all liability
Sensor error detection................................... Programmable ON / OFF
Environmental Conditions
Operating temperature...................................
Storage temperature......................................
Calibration temperature.................................
Relative humidity............................................
Protection degree...........................................
Installation in..................................................
-20°C to +60°C
-20°C to +85°C
20...28°C
< 95% RH (non-cond.)
IP20
Pollution degree 2 &
measurement / overvoltage
cat. II
Mechanical specifications
Dimensions (HxWxD).....................................
Dimensions (HxWxD) w/ 4501
/ 4511.............................................................
Weight approx................................................
Weight incl. 4501 / 4511 (approx.).................
DIN rail type...................................................
Wire size........................................................
Screw terminal torque....................................
Vibration.........................................................
2...13.2 Hz......................................................
13.2...100 Hz..................................................
109 x 23.5 x 104 mm
109 x 23.5 x 116 / 131 mm
250 g
265 g / 350 g
DIN EN 60715/35 mm
0.13...2.08 mm2 AWG 26...14
stranded wire
0.5 Nm
IEC 60068-2-6
±1 mm
±0.7 g
Common specifications
Supply
Supply voltage............................................... 19.2...31.2 VDC
Max. required power...................................... ≤ 0.8 W / ≤ 1.4 W (1 ch. / 2
ch.)
Max. power dissipation, 1
/ 2 ch.............................................................. ≤ 0.8 W / ≤ 1.4 W
Fuse............................................................... 400 mA SB / 250 VAC
Isolation voltage
Test /working: Input to any............................. 2.6 kVAC / 300 VAC
reinforced isolation
Analog output to supply................................. 2.6 kVAC / 300 VAC
reinforced isolation
Status relay to supply..................................... 1.5 kVAC / 150 VAC
reinforced isolation
TC input
Thermocouple type........................................ B, E, J, K, L, N, R, S, T, U,
W3, W5, LR
Cold junction compensation
(CJC) via ext. sensor in
connector 5910.............................................. 20...28°C ≤ ±1°C, -20...20°C /
28...70°C ≤ 2°C
CJC via internally mounted
sensor............................................................ ±(2.0°C + 0.4°C * Δt)
Δt =................................................................. Internal temp.-ambient temp.
Sensor error detection................................... Programmable ON or OFF
(only wire breakage)
Sensor error current: When
detecting / else............................................... Nom. 2 μA / 0 μA
Current input
Measurement range.......................................
Programmable measurement ranges............
Input resistance..............................................
Sensor error detection...................................
0...23 mA
0...20 and 4...20 mA
Nom. 20 Ω + PTC 50 Ω
Programmable ON / OFF
Output specifications
Current output
Signal range...................................................
Programmable signal ranges.........................
Load (@ current output).................................
Load stability..................................................
Sensor error indication...................................
NAMUR NE43 Upscale/Downscale...............
Output limitation, on 4...20
and 20...4 mA signals....................................
Output limitation, on 0...20
and 20...0 mA signals....................................
Current limit....................................................
0...23 mA
0...20/4...20/20...0/20...4 mA
≤ 600 Ω
≤ 0.01% of span / 100 Ω
0 / 3.5 / 23 mA / none
23 mA / 3.5 mA
3.8...20.5 mA
0...20.5 mA
≤ 28 mA
Passive 2-wire mA output
Max. external 2-wire supply........................... 26 VDC
Max. load resistance [Ω]................................ (Vsupply-3.5)/0.023 A
Effect of external 2-wire
supply voltage variation................................. < 0.005% of span / V
Response time
Temperature input, programmable
(0...90%, 100...10%)...................................... 1...60 s
mA / V input (programmable)......................... 0.4...60 s
Status relay
Max. voltage................................................... 110 VDC / 125 VAC
Max. current................................................... 0.3 ADC / 0.5 AAC
Max. AC power.............................................. 62.5 VA / 32 W
Programming................................................. Communication enabler 4511
/ Programming front 4501
Signal / noise ratio......................................... Min. 60 dB (0...100 kHz)
Signal dynamics, input................................... 24 bit
Signal dynamics, output................................. 16 bit
Accuracy........................................................ Better than 0.1% of selected
range
EMC immunity influence................................ < ±0.5% of span
Extended EMC immunity: NAMUR
NE21, A criterion, burst.................................. < ±1% of span
*of span.......................................................... = of the currently selected
measurement range
Input specifications
EMC............................................................... 2014/30/EU
LVD................................................................ 2014/35/EU
RoHS.............................................................. 2011/65/EU
EAC................................................................ TR-CU 020/2011
Approvals
UL................................................................... UL 61010-1
DNV-GL Marine............................................. Stand. f. Certific. No. 2.4
SIL.................................................................. SIL 2 certified & fully assessed
acc. to IEC 61508
9113A-061617
RTD input
RTD type........................................................ Pt10/20/50/100/200/250/300/P
t400/500/1000;
Ni50/100/120/1000
Cable resistance per wire
(max.)............................................................. 50 Ω
Sensor current............................................... Nom. 0.2 mA
Effect of sensor cable resistance
(3-/4-wire)....................................................... < 0.002 Ω / Ω