MS2010B
AC Leakage Current Clamp Meter
Instruction Manual
40
400
4
40
MS2010B
REL
Hz/%
/
1. Safety information
The meter is a handheld and battery operated Digital AC Clamp
Meter with multi function. This clamp meter has been designed to meet IEC61010-1 and CAT III over Voltage category.
Warning
To avoid possible electric shock or personal injury and possible damage to the meter or the equipment under test, you are must be adhere to the following rules.
1-1.Safety cautions
•
User must be read the operating instructions thoroughly and
completely before operating your meter. Pay particular attention
to WARNING, which will inform you of potentially dangerous
procedures.
•
Do not apply more than the rated voltage ,of marked on the meter,
between the “INPUT” terminal and “COM” terminal.
•
Do not expose the instrument to direct sunlight, extreme
temperature and moisture or dew full.
•
You always are careful when working with voltage above 60V DC
or 30V AC rms.
•
Keep fingers behind the clamp barrier while measuring current..
•
Always inspect the test lead for damaged insulation or exposed
metal.
1-2. Symbols
Note international Electrical Symbol.
Dangerous Voltage
AC
(Alternating current)
DC (Direct Current)
Ground
Warning see explain in manual
Double insulation
01
2. Meter illustration
40
400
4
40
MS2010B
REL
Hz/%
/
2-1 Panel description
1. Clamp jaw ; it is as current
transformer(CT) when measuring
current flowing through the
conductor.
2. Trigger
3. Clamp Barrier:
4. LCD display
5. “H/*” button : Hold & Back Light
When push this button, the
display will keep the last reading.
Once push again, the Meter will
return the normal mode.
6. Knob: it is rotary switch for
select function.
7. “Hz/%” button: Frequency &
Duty cycle mode
8. “RANGE” button: Change Auto
or Manual mode
9. “FUNC” button : Change function
mode
10. “LPF” button: Used as Low
Pass Filter ON/OFF switch.
11. “INPUT” terminal
12. “COM” terminal
13. “REL” push button:
Relative measurement mode
02
2-2. LCD display
AC
DC
AUTO
REL
LPF
LPF
HOLD
Alternating current/ voltage
Directive current/voltage
% mV, V
Diode
Buzzer
mA, A
C° nF, µF
Auto ranging mode
Ω, kΩ, MΩ
Relative measurement
Hz, kHz
Use LPF
Duty cycle
Voltage unites
Current unites
Temperature unite
Capacitance unite
Resistance unites
Frequency unites
No use LPF
Data hold
Low battery indicator
3. Specification
3-1. General feature
•
Auto ranging Digital Clamp Meter. But the meter may select
manual Function mode by “RANGE” button.
•
3 1/2 digit (200 count) LCD display
•
Over load indication: “OL” symbol will displayed on the LCD.
•
Jaw opening capability: 32mm
03
•
Low battery indication: battery symbol is appears on the LCD.
•
Auto power OFF: If the meter is idle for more than 15 minutes,
the meter automatically turns the power off.
•
Sampling rate: 2 times/s
•
Power supply: 1.5V battery (AAA type) x 3pcs.
•
Operating temperature & Humidity: 0°C to 40°C; <80%
•
Storage temperature & Humidity: -10°C to 50°C; <70%RH
•
Dimension (L x W x H) & Wight : 260 x 92 x 55mm; Approx. 400g
3-2. Electrical Specification
[1] AC Current
Range
40mA
400mA
4A
40A
400A
600A
Resolution
0.01mA
0.1mA
1mA
10mA
100mA
1A
[2] AC Voltage
Range Resolution
400mV
4V
0.1mV
1mV
40V
400V
600V
10mV
0.1V
1V
[3] DC Voltage
Range Resolution
400mV
4V
0.1mV
1mV
40V
400V
600V
10mV
0.1V
1V
Accuracy
(50 ~ 60Hz) (30~50Hz,60~10KHz)
±(1%rdg+8dgt)
±(1.5%rdg+3dgt)
Accuracy
(50 ~ 60Hz) (30~50Hz,60~10KHz)
±(0.8%rdg+5dgt)
Accuracy
±(0.7%rdg+3dgt)
±(0.8%rdg+5dgt)
No available
±(2%rdg+10dgt)
Input impedance
10MΩ
[4] Frequency & Duty cycle measurement
The meter is may be measuring the frequency and duty cycle in the
AC current range or AC voltage range by press the 'Hz/%' push key.
Frequency range is 1Hz to 2KHz in the current range, and then
1Hz to 100KHz in the voltage range.
(4-1) Current measurement range
Range
10Hz
100Hz
1000Hz
Resolution
0.001Hz
0.01Hz
0.1Hz
Accuracy Over load protection
±(2%rdg+5dgt) 250V DC or AC rms
±(1.5%rdg+5dgt)
(by PTC protection circuit) 2kHz
>2kHz
1Hz
No available
1 to 99% 0.1%
***Sensitivity is >AC 10mA(rms)
±3%
(4-2) Voltage measurement range
Range Resolution Accuracy
10Hz 0.001Hz
±(2%rdg+5dgt)
100Hz
1000Hz
10kHz
100kHz
0.01Hz
0.1Hz
1Hz
10Hz
±(1.5%rdg+5dgt)
±(2%rdg+5dgt)
>100kHz
1 to 99% 0.1%
No available
±3%
***Sensitivity is >AC 0.1V (rms)
[5] Resistance
Range
400Ω
4KΩ
40KΩ
400KΩ
4MΩ
40MΩ
Resolution
0.1Ω
1Ω
10Ω
0.1KΩ
1KΩ
10KΩ
Accuracy
±(1%rdg+2dgt)
±(2%rdg+5dgt)
Over load protection
250V DC or AC rms
(by PTC protection circuit)
Over load protection
250V DC or AC rms
(by PTC protection circuit)
04 05
[6] Capacitance measurement
Range
40nF
400nF
4µF
40µF
100µF
Resolution
0.01nF
0.1nF
1nF
0.01µF
0.1µF
Accuracy
±(4%rdg+5dgt)
Over load protection
250V DC or AC rms
(by PTC protection circuit)
[7] Temperature
Range
-20°C to
1000°C
Resolution
1°C -20°C to 0°C
0°C to 400°C
Accuracy
400°C to 1000°C
: ±(5.0%rdg+5dgt)
: ±(1.0%rdg+3dgt)
: ±(2.0%rdg+3dgt)
[8] Diode check
Test current : 1mA
[9] Continuity
If the resistance under testing circuit less than 70Ω, buzzer will sound.
Open circuit voltage: 1.2V.
4. Operating instruction
4-1 LPF usage
The meter have Low- Pass-Filter to reduce the influence of high frequency noise above 1KHz(3db). This LPF feature is available in all AC Voltage or AC Current measurement mode.
The “LPF” button is act as ON/OFF switch for LPF circuit connection. When LPF is activated, Most of noise above 1KHz will decay greatly. So we can obtain more stable and accurate readings on low frequency response.
4-2 Data HOLD
The push key is used to maintain the measurement data unchanging.
Press “HOLD” push key to inter and exit the hold mode in any mode. That act with trigger. The meter will resume the normal measurement mode by pressing the key again.
06
4-3 Range switching
“RANGE” button is the auto/manual measurement push key that act with trigger. The default is auto measurement when power on.
Press this key once, will switch to manual measurement mode, and then press once again, will switch to auto measurement mode.
4-4 Function switching
“FUNC” key is used as the function selection key that acts with trigger. Use the key as switch of DC/AC, Diode/ Continuity and
°C/°F.
4-5 Maximum value hold
“MAX” key is act with trigger. Press this key once, the maximum value is holding(will Displays “MAX” symbol on the LCD), and the press once again, will switch to normal Measurement mode.
After pressing the key, A/D will keep working, and the display value are always up dated and keep the maximum value.
4-6 Back Light
“ ” Push key is used control Black Light. When press this key and held more than 2sec, will enable Back Light for 15secs.
Press the key again within 15secs, back light will disable.
4-7 Sleep mode (Auto power OFF)
If the meter is idle for more than 15minutes, the meter automatically turns the power off.
In this sleep mode, the meter is save battery energy.
In the sleep mode, the meter may be turns normal operating mode by “RANGE”, “MAX”, “ ” Key and rotary switch.
There are two ways to disable the sleep mode as following:
(1) In the auto power off state , you are pushing “HOLD” key.
(2) Push and held “HOLD” key, turn on power by rotary switch
at the same time.
07
5. Measurement operating
5-1 AC Current measurement
1. Set the rotary switch to the desired “A~” position.
2. Press the clamp trigger to open jaw and to clamp one conductor
only, making Sure that the jaw is firmly closed around the
conductor.
3. Set the conductor in that flow the measured current to centre
position of jaw as far as Possible. In this time, measurement
accuracy is best.
4. Read current value on the LCD display.
Not: (1) Use the Low Pass Filter by pushing “LPF” button if
necessary.
(2) When measurement the current of transmission line, must
be clamp one line only between the two line .
5-2 Leakage current measurement
This meter can measure the leakage current on the one-phase or three-phase circuit equipment as well as the conductor in that flow the leakage current.
1. Set the rotary switch to the 20mA~ range.
2. Clamp the conductor in that flow the leakage current, and the
read the leakage 0current value on the LCD display.
5-3 DC/AC Voltage measurement
1. Set the rotary switch to the “V” position. The DC mode is default
mode. To switch to the AC measure mode must be press once
time the “FUNC” button.
2. Insert the red test lead in to the “INPUT” terminal and the black
test lead into the “COM” terminal.
3. Connect the test lead across with the object being measured.
The measured value will be show on the LCD display.
Note:
When DC or AC voltage measurement has been completed,
disconnect the connection between the test lead and circuit
under test.
08
5-4 Frequency & Duty cycle measurement
To measure the frequency or Duty cycle of induced current from the clamp jaw, done according to following procedure.
1. Set rotary switch to the 40/400mA, 4/.40A or 400A position.
2. Press the clamp trigger to open jaw and clamp a conductor in
that flowed Measured current, making sure that the jaw is firmly
closed around the conductor.
3. Press the “Hz/%” push key to into the Frequency mode or Duty
cycle mode.
Measured value will be displayed on the LCD.
To measure the frequency or duty cycle of AC voltage that measured in the “ V” range, one as following:
1.Set rotary switch to the “V” position.
2.Done section 5-4 (2) ~(3) procedure.
5-5 Resistance measurement
1. Set the rotary switch to the “
Ω
” position.
2. Insert the red test lead into the “INPUT” terminal and the black
test lead into the “COM” Terminal.
3. Connect the test lead across with the object being measured.
The measured value will be show on LCD display.
Note:
•
The test lead can add 0.1Ω to 0.2Ω of error to resistance
measurement. To obtain precision reading in low-resistance
measurement, that is the range of 200Ω, short the input terminal
with the test lead probe, and then read out the contact resistance
of test lead. After measuring you can subtract the readied contact
resistance value from the reading value.
•
For high resistance measurement (>10MΩ), it is normal taking
several seconds to obtain stable reading.
•
If The LCD display “OL” symbol, it is indicating open circuit for the
tested resistor or the resistance value of resistor is higher than
the maximum range of the meter.
5-6 Diode check
The diode check function used to check diode, transistor and other semiconductor device.
In the diode testing, the meter sends a current through the semiconductor junction, and then measure the voltage drop across the junction. A good silicon junction will drop between 0.5V to 0.8V.
1. Set the rotary switch to the “
Ω
” position and then
press the “FUNC” push key to into diode check mode.
2. Insert the red test lead into the “INPUT” terminal and the black
test lead into the “COM“ terminal.
3. For forward voltage drop reading on any semiconductor
component, place the red test lead on the component anode and
place the black test lead on the component cathode.
The measured forward drop voltage will show on the LCD display.
4. Reverse the test lead and measure the voltage across the diode again.
•
If diode is good, the display shows “OL”.
•
If diode is shorted, the display shows “0” (zero) in both direction.
•
If display shows “OL” in direction, the diode is open.
5-7 Continuity check
Set the rotary switch to the “
Ω
” position.
1.Press the “FUNC” push button to switch into continuity check mode.
2. Insert the red test lead into the “INPUT” terminal and the black
test lead into the “COM” Terminal.
3. Connect the test lead across with the object being measured.
If the resistance of a circuit under test is less than 50Ω,
the buzzer will sound.
5-8 Capacitance measurement
Warning
To avoid damage to the meter or to the equipment under test,
disconnect power and discharge all high-voltage capacitors
before measuring capacitance.
You can use the DC voltage function to confirm that the
capacitor is discharged.
10
Capacitance ranges are 50.00nF, 500.0nF, 5.000uF, 50.00uF,
200.0uF.
To measure capacitance, connect the Meter as follows:
1. Set the rotary switch to “ ” position.
2. Insert the Capacitor under test to the “INPUT” terminal and
“COM” terminal.
The measured value will be displayed on the LCD disp
5-9 Temperature measurement
Note:
•
MS3202 Multi-Function Socket is used as temperature adapter.
•
Use K-type thermocouple probe.
1. Set the rotary switch to the “TEMP” position.
2. In this time, LCD will display the environment temperature.
3. Insert the MS3202 temperature adapter to the “INPUT” and
“COM” terminal , and then Insert the K-type thermocouple probe
to the MS3203 temperature adapter according to its polarity.
4. To change the temperature physical unite(°C or °F) , you can
press the “FUNC” push button.
6. Maintenance
6-1 Replacing the Battery
When meter display low battery indicator(symbol) on the LCD, you must be replace the battery to maintain normal operation.
(1) Disconnect and remove all test probes from any live source and
meter.
(2) Open the battery cover on the bottom case by screwdriver.
(3) Remove old battery and snap new one into the battery holder.
6-2 Fuse replacement
Replacing the defective fuse should done according to 6-1section procedure.
6-3 Cleaning and Decontamination
The meter can be cleaned with soft clean cloth to remove any oil, grease or grim.
Do not use the liquid solvent or detergent.
11
HYS006738