PW IP67 Water Proof Weighing Indicator User Manual

PW IP67 Water Proof Weighing Indicator User Manual
User Manual
Indicator
PWH3
© Excell Precision Limited 2016. All rights reserved Worldwide.
The information contained herein is the property of Excell Precision Limited and
is supplied without liability for errors or omissions. No part may be reproduced or
used except as authorised by contract or other written permission. The copyright
and the foregoing restriction on reproduction and use extend to all media in
which the information may be embodied.
EXCELL PRECISION CO., LTD
Table of Contents
Special Notice .................................................................................................................................. 2
Before Using the Scale................................................................................................................... 3
Instructions for Use ......................................................................................................................... 3
Preparing to Use the Scale ............................................................................................................ 3
Chapter 1 Introduction .................................................................................................................... 5
1-1 Product Features .................................................................................................................. 5
1-2 Specifications ........................................................................................................................ 5
1-3 Product Appearance ............................................................................................................ 5
1-4 Display Description .............................................................................................................. 6
1-5 Power Supply ........................................................................................................................ 6
1-6 Keypad Function ................................................................................................................... 7
1-7 Self-Test Mode ...................................................................................................................... 9
1-8 Error Messages .................................................................................................................. 10
1-9 Weight Units ........................................................................................................................ 10
Chapter 2 Configurations ............................................................................................................. 11
2-1 General Function Setting
.................................................................... 12
2-1-1 Auto Backlight Function Setting
.................................................. 13
2-1-2 Auto Power-off Timer Setting
...................................................... 14
2-1-3 HI/LO/OK Settings
......................................................................... 15
2-1-4 Hold Function Settings
................................................................. 16
2-1-5 Auto Unit Weight Averaging Setting
2-1-6 Foot Switch Settings
2-1-7
........................................... 19
(Option) ..................................................... 19
Key Function Settings
...................................................... 20
2-2 RS232 Serial Interface Settings
.......................................................... 21
2-2-1 Baud Rate Settings
..................................................................... 22
2-2-2 Communication Protocol Settings
.......................................... 23
2-2-3 Output Format Settings
............................................................... 24
2-2-4 Continuous Transmission Settings
............................................ 26
2-2-5 Selection of the Continuous Transmission Rate
2-2-6 Auto Transmission at Zero
........................... 27
.......................................................... 28
2-2-7 Reset Auto Transmission
............................................................... 29
2-2-8 Output Condition Settings
............................................................. 30
2-2-9 RS232 6/7 Digits Setting
............................................................... 31
2-2-10 RTC Adjustment
............................................................. 32
2-2-11 Y/M/D Print Layout Selection
........................................................ 32
2-2-12 RS485 ID Input (Option)
............................................................... 33
2-2-13 Line Feed Input
............................................................................ 33
Appendix 1 ASCII Code Table .................................................................................................... 38
Appendix 2 7-Segment Display Characters ............................................................................. 39
Appendix 3 Wiring Instructions .................................................................................................. 40
02005159
1
ZSME300000288
EXCELL PRECISION CO., LTD
Special Notice
While installing the load cell, power cord hookup or replacing a new rechargeable
battery, the indicator housing must be opened. It must be done by a technician assigned
by your electronic indicator provider to avoid affecting the waterproof ability of this indicator.
Before opening the housing, make sure the indicator is dry, if there is any liquid on it,
please wipe it with a clean cloth.
HOW TO INSTALL THE HOUSING
After installing the load cell, power cord hookup or replacing a new rechargeable battery,
the housing must be screwed by the assigned order as shown below. Screw lightly first,
then screw them tight using a 12 kgf-cm torsion.
P.s. Please use a torsion-adjustable screw driver.
Screwing order:
3
5
02005159
1
6
4
2
2
ZSME300000288
EXCELL PRECISION CO., LTD
Thank you for purchasing EXCELL WEIGHING INDICATOR,
to help use the product properly, operate smoothly, and
extend its life cycle, please read this manual carefully.
Before Using the Scale
In order to use this scale correctly, we suggest that you read this manual carefully.
Instructions for Use
1. The load placed on the weigh pan must NOT exceed the maximum weighing capacity
of the scale.
2. Protect the scale from high temperatures.
3. Avoid objects impacting with the scale. Do not drop loads onto the scale or subject the
weigh pan to any strong shock loads.
4. Please operate or charge the scale in an open area. Do not squeeze the power cord to
avoid wire on fire.
Preparing to Use the Scale
1. Locate the scale on a firm level surface free from vibrations for accurate weight
readings.
2. Adjust the four levelling feet (if fitted) to set the scale pan level.
3. Avoid operating the scale in direct sunlight or drafts of any kind.
4. If possible avoid connecting the scale to ac power outlet sockets which are adjacent to
other appliances to minimise the possibility of interference affecting the performance of
the scale.
5. Remove any weight that might be on the weigh pan before the scale is switched on and
avoid leaving weight on the pan for long periods of time.
6. All goods weighed should be placed in the centre of the weigh pan for accurate
weighing. The overall dimensions of the goods being weighed should not exceed the
dimension of the weigh pan.
7. Once the scale has been powered on, it will go through an LCD display test and it is
ready for use when the display shows zero.
8. The scale requires 15~20 minutes warm up before operation to ensure best accuracy.
9. Please note when the
be charged.
symbol keeps flashing on the screen, the batteries need to
10. Introduction of Storage Battery
02005159
3
ZSME300000288
EXCELL PRECISION CO., LTD
Due to the storage battery adopt the advanced free -maintaining technique, customers
need not to replenish electrolyte.
The scale should be recharged every 3 months to prevent failure of the internal
rechargeable battery.
1. The battery should be charged for 8~10 hours.
2. The temperature of battery should below 45℃.
Maintaining
1. Please do not discharge with over-current when using the battery. Please charge the
battery after discharging current.
2. Please take down the battery when the scale is not used for a long time or break the
connection of cathode.
3. Do not short the battery terminals to check whether there is current. Please check
whether the connection point is firm to guarantee good connection.
4. The battery should be replaced by specialized person. No reverse-battery or the
product will be damaged.
a) Anode of battery should be connected with Anode of product battery(usually red
cable)
b) Cathode of battery should be connected with Cathode of product battery (usually
brown cable or black cable)
c) See the picture following
Brown cable(or black cable)
connected with Anode of
battery
Red cable connected with
Cathode of battery
Safety warnings
1. The electrolyte of battery is caustic which causes metal, cotton, etc to corrode.
2. The hydrogen will be resolved when using or charging the battery and it will cause
explosion when approaches fire.
No burning
02005159
Caution Corrosion
Warning explosion
4
Children faraway
ZSME300000288
EXCELL PRECISION CO., LTD
Chapter 1 Introduction
1-1 Product Features








Sealed to IP68, (Only use cables of 3mm~5mm diameter to ensure
correct sealing of the cable glands)
Kilogram (kg) and pound (lb) weighing modes
Full range tare; Pre-tare; Auto zero tracking; Sampling counting; Gross/Net indication
Adjustable gravity value
Low power indication and auto power off
AC/DC power in and rechargeable battery.
Built-in RS-232
Options: pressure release valve, foot switch
1-2 Specifications

Analogue Input: Input Sensitivity 0.3 V/d (Min.)

Input Signal Range: -1mV~+14mV

Input Zero Range: -1mV~+5mV

Load Cell Excitation: 5V DC

Load Cell Drive Capacity: Up to 8 x 350 load cells

Non-linearity: 0.01% of full scale

A/D Resolution: 500,000 counts (Maximum)

Operating temperature: -10℃ ~ 40℃
1-3 Product Appearance
256 mm
90 mm
170 mm
02005159
5
ZSME300000288
EXCELL PRECISION CO., LTD
(RS-232 and load cell cables size
3mm~5.5mm diameter to ensure
correct sealing of the cable glands)
RS-232 cable entry
Load cell cable entry
Power supply connector
1-4 Display Description
HI
OK
LO
:
:
:
Upper limit
Value between HI and LO
Lower limit
kg
:
Unit “kilogram”
lb
t
Pcs
 0
Net
:
:
:
:
:
:
Unit “pound”
hk tael
Counting mode indication
Zero point indication
Net weight indication
Low Power indication
“” indications description
1 : (STABLE) stable indication
2 : (GROSS) gross weight indication
3 : (
) unit weight insufficient indication
4 : ( PT ) pre-tare indication
5 : Range 2
6 : Range 1
7 : M+
Pcs
1-5 Power Supply
Power Supply Selection
➀ 6V / 4.5Ah rechargeable battery
➁ Input 110VAC or 230VAC ±15% ; Output 10VDC/1A adaptor
Power Consumption
Approximately DC 14 mA (Indicator)
Approximately DC 24 mA (Indicator + Display backlight)
Charging Voltage
DC 10V/1A output adaptor
02005159
6
ZSME300000288
EXCELL PRECISION CO., LTD
Low Battery Warning
Please note when the (
) symbol keeps flashing on the display, the internal battery
should be recharged.
 The scale will turn off automatically after a few hours when the low battery warning
symbol shows up. The scale must be fully charged, before operating again.
 When the battery status indication is full, the power is about 6.1V; Each block is about
0.2V; When the battery warning symbol shows up, the power is about 5.4V.
1-6 Keypad Function
OFF KEY
Press the OFF key to switch the indicator off.
1
ON|ZERO KEY
 When the power is off, press ON|ZERO key to switch the indicator on.
 When the power is on, press ON|ZERO key to return to zero.
 Zero Range : OIML&NTEP is ±2% F.S., and Sri Lanka is ±3% F.S.
1
TARE|PT KEY: Tare / Pre-tare key
To TARE:
Place the container onto the scale, until the weight value is stable, press TARE|PT key for
zero return and the NET indication is shown on the display. Place the object onto the
container and the display shows the net weight value of the object. Remove both object
and container, and negative value of the container will show on the display. Press TARE|PT
key again to clear “tare value”. The scale returns to zero and NET indication goes off.
 Tare can be continuously done until tare value=full load capacity
 Continuous Tare  Press TARE|PT key for continuous weight increase/decrease on
platter.
 If there is Tare, the pre-tare cannot be done. If there is pre-tare first, and the tare weight
more than pre-tare weight, Tare can be done.
 No Tare can be done under gross weight display mode.
To PRE-TARE:
Press TARE|PT key, use keypad to input pre-tare weight. When the cursor flashes on the
rightmost digit, press TARE|PT key again to complete.
Pre-tare mode, keypad function as followed:
ON|ZERO  upward key (0~9 digit entry)
NET|B/G  move cursor leftward
UNIT
 downward key (0~9 digit entry)
TARE|PT
 move cursor rightward
UNIT KEY
Press the UNIT key to switch weight units; the display icons will indicate the active units.
NET|B/G KEY
In the Tare mode, the screen displays the “TARE” icon; press the NET|B/G key to switch
between the “Net value” and the “Gross value”. When the ”GROSS” arrow  shows up,
the Gross Weight in the screen = Tare weight + Net weight, meanwhile, keys except the
1NET|B/G key, have no response.
M+|PRINT KEY
Totalization function. M+/Print function is available when RS232 is on keypad transmission
mode. (
output).
1
02005159
7
ZSME300000288
EXCELL PRECISION CO., LTD
This key is a composite key, while totalization is shown and weight returns to net zero,
press M+ key to erase memory. RS232 will output MC print format (
output).
 If there is new weight added on platter(it is less than 20 divisions in Brazil version, no
accumulation ), a new item will be added to totalization. If this weight is not taken off,
nothing can be added to totalization.Display will show the totalization numbers for one
second, then show net weight for one second, then the scale returns to the current
weight, and prints out the last item for totalization.
 To clear totalization data, press M+ key to let display shows up totalization numbers,
then press the M+|PRINT key again, to clear totalization data. RS232 will print out
totalization numbers, total weight, etc.
P.s. Weight must return to net zero if to perform clear function.
KEY
Function key (In
, select
key as “MC” or “HR”)
Foot Switch Mode
This function is optional. Use
to select TARE|PT key as “ZERO” or “PRINT” key
 If “PRINT” key is set (
= 10/11), all totalization data will be printed out, and
totalization will be cleared.
 If it is Brazil version and foot switch is set as Print function, it has totalization function
and print function.
Simple Counting Mode
Use UNIT key to switch unit to Pcs, to go into simple counting mode.
1. Use NET|B/G key to select a sample number from “10, 20, 50, 100, or 200”. Display will
show
,
,
,
,
in an sequential order by pressing
NET|GROSS key.
2. Select a sampling number, and put appropriate weight on platter, and press UNIT key.
Display will show “
”. The scale will go into counting mode after weight is
stable, and display will show the sample number.
 Unit weight insufficient (
): Sample unit weight is less than 0.2d or total sample
weight is less than 20d (d=division)
 While sampling, if there is insufficient sample or unit weight insufficient indication “”,
the scale is still usable, but there may be slight inaccuracy.
 After Power-off, the scale automatically memorizes the sampling number, and it is
available when “Pcs” unit is selected next time.
 If the setting is “automatic average unit weight “, if the object on platter > the previous
sampling number more than 5 pcs, and also < less than 100% the previous sampling
number, the scale will execute unit weight calibration automatically.
Pcs
02005159
8
ZSME300000288
EXCELL PRECISION CO., LTD
1-7 Self-Test Mode
When power is off, hold NET|B/G key and press ON|ZERO key, Wait till display shows
to enter “Self-Test Mode”.
Press UNIT key
Press ON|ZERO key
Press UNIT key
Press ON|ZERO key
Press UNIT key
Press ON|ZERO key
Press UNIT key
Press ON|ZERO key
INTERNAL VALUE MODE (must hook up full-bridge Load Cell to test)
 Press TARE|PT key to enter, and the display shows internal value
 Please check the internal value is within normal range is 0 ~ 400000 (no load)
 Check whether the backlight is on
 Press ON|ZERO key to back to the last screen, the display shows
KEYPAD TEST MODE
 Press TARE|PT key to enter, display shows
Keypad’s internal code: TARE|PT key = 06, UNIT key= 05,
key = 02
1NET|B/G key=04
 Press ON|ZERO key to back to the last screen , the display shows
FIRMWARE VERSION DISPLAY MODE
 Press TARE|PT key to enter , display shows the firmware version
 Press TARE|PT key again, the display shows maintenance number
 Press ON|ZERO key to back to the last screen, display shows
,
for 2 seconds
BACK TO THE LAST SCREEN
Press TARE|PT key to exit self-test mode, the scale will re-power on automatically.
02005159
9
ZSME300000288
EXCELL PRECISION CO., LTD
1-8 Error Messages
 The EEPROM is not working correctly.
The EEPROM is not set yet, or the circuit on PCB is broken.
 Zero is higher than the zero range when switching the indicator on.
 Zero is lower than the zero range when switching the indicator on.
 A/D value is unstable.
 The weight of the object is over 9 divisions of the maximum capacity .
 The weight of the object is under -1/6 maximum capacity.
 ADIC value is over the maximum range.
 The scale is not in level status.
(Only available with level detector equipped)
1-9 Weight Units
kg
g
lb
lb/oz
oz
GN
dwt
ct
hk.tael
viss
02005159
1g =
1g =
1g =
1g =
1g =
1g =
1g =
1g =
1g =
1kg =
0.001 kg
1g
0.002204623 lb
0.03527396 oz
0.03527396 oz
15.432358 GN
0.6430149 dwt
5 ct
0.02645546 Hk.catty
0.612245 viss
10
ZSME300000288
EXCELL PRECISION CO., LTD
Chapter 2 Configurations
Configuration Workflow
In the weighing mode, press NET|B/G key and ON|ZERO key s at the same time to enter
the configuration mode.
The LCD shows
.
Overall workflow of the Advanced Function setting mode:
Press UNIT key
Press ON|ZERO key
Press UNIT key
Press ON|ZERO key
Press UNIT key
Press ON|ZERO key
Press UNIT key
Press ON|ZERO key
 General Function Setting Mode
 External Weight Calibration
 RS232 Bi-direction Function Setting
 Exit the Advanced Function Setting Mode
Refer to the following sections for the detailed operation
procedures of each function setting.
02005159
11
ZSME300000288
EXCELL PRECISION CO., LTD
2-1 General Function Setting
Workflow of the General Function setting:
Press TA RE|PT key
key in the parameter
“00 ~ 13”
ON|ZERO key  Upward key (0~ 9 digit ent ry)
Key in “00” and press
1 TARE |PT key to return to
the Advanced function
setting menu
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
 Return to the Advanced Function Setting Mode Menu
 Automatic Backlight Function Settings
 Automatic Power-off Timer Settings
 Hi/Lo/OK Settings
 Reserved
 Reserved
 Hold Function Settings
 Auto Unit Weight Averaging Setting
 Reserved
 Reserved
 Reserved
 Foot switch on/off settings (Option)

key function settings
 Reserved
02005159
12
ZSME300000288
EXCELL PRECISION CO., LTD
2-1-1 Auto Backlight Function Setting
Select
in the general function setting mode
backlight function setting.
to change the
Press TARE |PT key
LCD displays the last status
 Default setting = off
Use ON|ZERO or UNIT key
to select function “on” or “off
On  backlight function is on
Off  backlight function is off
Press TARE|PT key to go
back to the advanced
function setting menu
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
Auto backlight function
When the weight is over 10d, the display backlight will be on. After the weight
is stable for 10 seconds or when the scale returns to zero, the display
backlight switches off.
02005159
13
ZSME300000288
EXCELL PRECISION CO., LTD
2-1-2 Auto Power-off Timer Setting
Select
in the general function setting mode
automatic power-off timer setting.
to change the
Press TARE|PT key
LCD displays the last status
Default setting = 0 (No auto power-off)
Use ON|ZERO or UNIT key
to key in parameter
Automatic power-off timer setting
Use ON|ZERO or UNIT key to key in
parameter
Press TARE|PT key to
go back to the
Advanced function
setting menu
0  No auto power-off
1  When the scale is idle for 1
minute, the scale
automatically switches off
2  When the scale is idle for 2
minutes, the scale
automatically switches off
9  When the scale is idle for 9
minutes, the scale
automatically switches off
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
Auto power-off function
When the weight on weigh pan is less than 10d or keeps idle for the set time, the
scale will automatically switch off.
02005159
14
ZSME300000288
EXCELL PRECISION CO., LTD
2-1-3 HI/LO/OK Settings
Select
in the general function setting mode
to set the HI/LO/OK
function.
 When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is disabled.
Press TARE|PT key
lb
.
Key in the desired high
limit value
LCD displays the last status
lb
.
Press TARE|PT key
lb
.
LCD displays the last status
Key in the desired low limit
value
lb
.
LCD displays the last status
Press TARE|PT key
Alarm setting
(a)(b)(c)
The definition of a, b, c positions:
Key in the parameter
setting for alarm
(a)  1 = beep sound on
0 = beep sound off
(b)  1 = beep sound on when stable
0 = beep sound on when unstable
Press TARE|PT key
(c)  1 = beep sound on when the weight is
between high and low limits
0 = beep sound on when the weight is
not between high and low limits and
higher than 10d
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
02005159
15
ZSME300000288
EXCELL PRECISION CO., LTD
2-1-4 Hold Function Settings
 For OIML or NTEP approval models,
hold=0
setting cannot be changed, and
Hold function setting
LCD displays the last parameter setting
Hold function setting
Use ON|ZERO or UNIT key to key in the
parameters
 Default setting = 0
0  Hold function disabled
1  “Peak hold” mode
2  “Stable hold 1” mode
3  “Stable hold 2” mode
4  “Animal scale hold 1” mode
5  “Animal scale hold 2” mode
Press TARE|PT key
Use ON|ZERO or UNIT key
to key in the parameters
Press TARE|PT key
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
= Hold function disabled
= “Peak hold” mode
The scale keeps displaying the maximum weight when the weight is continually
changing. Press any key to exit this mode.
= “Stable hold 1” mode
When the weight is stable, the LCD shows the current weight value. Press any
key to exit this mode.
= “Stable hold 2” mode
When the weight is stable, the LCD shows the current weight value. When the
weight returns to zero (<10d), the hold mode is cancelled automatically.
= “Animal scale hold 1” mode
When no load, the display shows “- - - - - -”. After the animal or object is on the
platter and the weight is stable, the display shows the weight value and hold .
When the animal or object is off the platter, the display shows “ - - - - - - ” and the
hold function is off. If the weight is hardly stable, display shows the average
weight in 10 seconds and holds the status, until the weight is less than 10e and
display shows “- - - - - -” or press any key to calculate a new weight.
= “Animal scale hold 2” mode
When no load,the display shows “0.000”. After the animal or object is on the
platter and the weight is stable, the display shows the weight value and hold.
When the weight added or removed on the platter is larger than the weight set in
hold 5, the hold function release and to calculate a new weight. If the weight is
hardly stable, the display shows the average weight in 10 seconds and holds the
status. In this mode, ON|ZERO and TARE|PT keys are useless.
02005159
16
ZSME300000288
EXCELL PRECISION CO., LTD
Hold Function Setting (Animal scale)
Hold function setting
LCD displays the last parameter setting
Press TARE|PT key
Hold function setting
Use ON|ZERO or UNIT key to key in 4 and
pPress TARE|PT key to enter the hold
function setting mode.
Use ON|ZERO or UNIT key
to key in the parameters
Press TARE|PT key
%
Use ON|ZERO or UNIT key
to key in the parameters
Error range setting
Use ON|ZERO or UNIT key to key in the
parameters ( Among 1%~100%)
Default setting = 10%
%
Press TARE|PT key
Use ON|ZERO or UNIT key
to key in the parameters
Counts averaging setting
Use ON|ZERO or UNIT key to key in the
parameters (1,2,4,8,16,32,64)
Default setting = 8
Press TARE|PT key
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
17
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
Hold Function Setting (Animal scale)
Press TARE|PT key
Hold function setting
LCD displays the last parameter setting
Use ON|ZERO or UNIT key
to key in the parameters
Hold function setting
Use ON|ZERO or UNIT keys to key in 5
Release Hold range setting
LCD displays the last parameter setting
Press TARE|PT key
.
kg
Use ON|ZERO or UNIT key
to key in the parameters
.
Release Hold range setting
Use ON|ZERO or UNIT keys to key in the
parameters (0 ~ full scale )
Default setting = full scale
kg
Press TARE|PT key
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
18
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-1-5 Auto Unit Weight Averaging Setting
Auto unit weight averaging setting
LCD displays the last parameter setting
Press TARE|PT key
Auto unit weight averaging setting
Use ON|ZERO or UNIT key to key in the
parameters
 Default setting = 0
0  Auto unit weight averaging
1  Non auto unit weight averaging
Use ON|ZERO or UNIT key
to key in the parameters
ON|ZERO key  Upward key (0~ 9 digit ent ry)
Press TARE|PT key
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
2-1-6 Foot Switch Settings
(Option)
Foot switch on “ZERO” function
Display shows parameter from
the previous setup
Press TARE|PT key
Foot switch on “TARE” function
Use ON|ZERO or UNIT key
to select parameter
Foot switch on “PRINT” function
RS232 print out data sent from
pressing M+ key
If it is Brazil version and
=
_
, the format is
format and it has totalization function.
Use ON|ZERO or UNIT key
to select parameter
Press TARE|PT key
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
19
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-1-7
Key Function Settings
key function represents “MC”
function
Display shows parameter from
the previous setup
Press TARE|PT key
key function represents
“HR” function
Use ON|ZERO or UNIT key
To select parameter
key function represents
tare or pre-tare value
ON|ZERO key  Upward key (0~ 9 digit ent ry)
Press TARE|PT key
02005159
20
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-2 RS232 Serial Interface Settings
Press TARE|PT key
Key in the parameter
00 ~ 08
ON|ZERO key  Upward key (0~ 9 digit ent ry)
Key in the parameter
“00”, then press
1 TARE|PT key to exit
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
 Exit the RS232 Serial Interface Setting Mode
 Baud Rate Settings
 Communication Protocol Settings
 Output Format Settings
 Transmission method setup
 The Selection of Continuous Transmission Rate
 Auto Transmission at Zero
 Reset of Auto Transmission
 Output Condition Settings
 RS232 6/7 digits setting
 RTC adjustment
 Y/M/D Print Format Selection
 RS485 ID Input (Option)
 Line Feed Input
02005159
21
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-1 Baud Rate Settings
Select
Baud Rate.
in the RS232 serial interface setting mode
to set the
Press TARE|PT key
Baud rate setting
LCD displays the last value
 Default value: 9600 (bits/sec)
Use ON|ZERO or UNIT key
to select the desired
Baud rate
Baud rate setting
Use ON|ZERO or UNIT key to select the
desired Baud rate 600、1200、2400、
4800、9600、19200 (bits/sec)
Press TARE|PT key
Key in the parameter “00”
Press TARE|PT key to exit
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
22
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-2 Communication Protocol Settings
Select
in the RS232 serial interface setting mode
Communication Protocol.
to set the
Press TARE|PT key
Communication protocol setting
LCD shows the last setting value.

Default setting:
Use ON|ZERO or UNIT key
select the protocol setting
Communication protocol setting
Use ON|ZERO or UNIT key to select the
protocol setting: n 8 1, E 7 1, O 7 1
Press TARE|PT key
Key in the parameter “00”
Press TARE|PT key to exit
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
23
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-3 Output Format Settings
Select
Output Format.
in the RS232 serial interface setting mode
to set the
RS232 output format setting
LCD shows the last setting.
 Default setting:
Press TARE|PT key
RS232 output format setting
Use ON|ZERO or UNIT key to select the
output format:
Use ON|ZERO or UNIT key
to select the output format
= Same data as the scale
= Gross weight
= Net weight
= Same data as the scale in
simple format
= Same gross data as the scale in
simple format
= Same net data as the scale in
simple format
= Hi/Lo/OK status + Same data as
the scale in simple format
= Hi/Lo/OK status + Simple gross
weight
= Hi/Lo/OK status + Simple net
weight
= Tare value
= M+ Transmission 1 note 1
= M+ Transmission 2 note 2
= Brazil printing format note 3
Press TARE|PT key
Key in the parameter “00”
Press TARE|PT key to exit
 Format 12 are only for Brazil
Format 10.11 are not available on
counting mode
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
02005159
24
ZSME300000288
EXCELL PRECISION CO., LTD
Note1:
Ticket No.
Date year/month/day & day/month/year (choose 1 out of 2 date formats)
Time
G
T
N
Total Net (This line will only print when data is erased, showing net weight total of every count)
Note2:
Ticket No.
Date year/month/day & day/month/year (choose 1 out of 2 date formats)
Time
G
T
N
Total Weight (This line will only print when data is erased, showing gross weight total of every count)
Note3:
If it is not Brazil version, it can only print under continuous transmition.
If it is Brazil version, under decimal units mode,
=
and
= _
, it can print out. It is not available on counting mode.
For example:
70.15Kg print as: =51.07000=51.07000=51.07000=51.07000=51.07000
negative 70.15Kg print as: =51.0700-=51.0700-=51.0700-=51.0700-=51.0700-
02005159
25
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-4 Continuous Transmission Settings
Select
in the RS232 serial interface setting mode
Continuous Transmission status.
to set the
Press TARE|PT key
Transmission setting
LCD shows the last setting
 Default setting:
Use ON|ZERO or UNIT key
to select the transmission
type
Transmission setting
Use ON|ZERO or UNIT key to select the
transmission type:
Press TARE|PT key
_
Key in the parameter “00”
= Command mode
= Continuous transmission
= Constantly auto
transmission
= Shut down RS232
= M+ Mode
Press TARE|PT key to exit
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
26
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-5 Selection of the Continuous Transmission Rate
Select
in the RS232 serial interface setting mode
Continuous Transmission Rate.
to set the
Press TARE|PT key
Continuous transmission rate setting
The LCD shows the last setting.
 Default setting:
Use ON|ZERO or UNIT key
to select the continuous
transmission rate.
Use ON|ZERO or UNIT key to select the
continuous transmission rate:
= 1 time/sec
= 2 times/sec
= 4 times/sec
= 8 times/sec
= 16 times/sec
= More than16 times/sec
Note: Actual output times vary by
hardware implements
Press TARE|PT key
Key in the parameter “00”
Press TARE|PT key to exit
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
27
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-6 Auto Transmission at Zero
Select
in the RS232 serial interface setting mode
Auto Transmission at Zero.
to set the
Press TARE|PT key
Auto transmi ssion at zero
LCD shows the last setting
 Default setting:
Use ON|ZERO or UNIT key
to key in the desired
parameters
Auto transmi ssion at zero
Use ON|ZERO or UNIT key
to key in the
desired parameters:
= 00 division of weight value
Press TARE|PT key
= 01 division of weight value
= 99 divisions of weight value
Key in the parameter “00”
Press TARE|PT key to exit
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
 When the value of “Auto transmission at zero” set as
, the “Auto
transmission at zero” function is not available. Because when the scale is at zero and
keeping stable, the data are keeping transmitted, and that would be the same as
“Continuous Transmission”.
02005159
28
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-7 Reset Auto Transmission
Select
Transmission.
in the RS232 serial interface setting mode
Press TARE|PT key
to Reset Auto
Resetting auto transmission
The LCD shows the last setting
 Default setting:
Use ON|ZERO or UNIT key
to key in the desired
parameters
Resetting auto transmi ssion
Use ON|ZERO or UNIT key
to key in the
desired parameters:
= 00 division of weight value
= 01 division of weight value
Press TARE|PT key
= 99 divisions of weight value
Key in the parameter “00”
Press TARE|PT key to exit
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
 When the value of “Auto transmission at zero” set as
, the “Auto
transmission at zero” function is not available. Because when the scale is at zero and
keeping stable, the data are keeping transmitted, and that would be the same as
“Continuous Transmission”.
02005159
29
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-8 Output Condition Settings
Select
in the RS232 serial interface setting mode
Auto Transmission.
to Reset of
Press TARE|PT key
Output condition settings
The LCD shows the last setting.

Use ON|ZERO or UNIT key
to select the desired setting
Default setting:
Output condition settings
Use ON|ZERO or UNIT key to select the
desired setting:
= All output
Press TARE|PT key
=
Stable output
(No output when OL or unstable)
=
Stable output
(OL included)
Key in the parameter “00”
Press TARE|PT key to exit
RS232 2-way serial
interface settings
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
30
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-9 RS232 6/7 Digits Setting
Select
in the RS232 serial interface setting mode
Auto Transmission.
to Reset of
Press TARE|PT key
Output Condition Restriction Settings
Display shows the previous setup
 Factory Default is
(6 places)
Use ON|ZERO or UNIT key
to select the desired
setting
Output Condition Restriction Settings
Use ON|ZERO or UNIT key to select desired
setting.
Output condition:
= 6 digits
= 7 digits
Press TARE|PT key
Key in the parameter “00”
Press TARE|PT key to exit
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
31
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-10 RTC Adjustment
Select
Transmission.
in the RS232 serial interface setting mode
to reset of Auto
Enter the date in YY/MM/DD
Use upward and downward key to
select, and use TARE|PT key to confirm
Press TARE|PT key
Move cursor to the right end and
Press TARE|PT key
Press TARE|PT key
Enter time in HH/MM/SS
Use upward and downward key to
select, and use TARE|PT key to
confirm, and move rightward.
Press TARE|PT key
Move cursor to the right end and
Press TARE|PT key to confirm.
RTC adjustment is complete
Press TARE|PT key
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
2-2-11 Y/M/D Print Layout Selection
Select
in the RS232 serial interface setting mode
Auto Transmission.
to reset of
Display shows the previous setup
Use upward and downward key to select,
and use TARE|PT key to confirm
Press TARE|PT key
If y_m_d, printed as Year/Month/Day
If d_m_y, printed as Day/Month/Year
Use ON|ZERO or UNIT key
to select the desired
setting
ON|ZERO key  Upward key (0~ 9 digit ent ry)
Press TARE|PT key
02005159
32
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
2-2-12 RS485 ID Input (Option)
Select
in the RS232 serial interface setting mode
Auto Transmission.
to reset of
Press TARE|PT key
Display shows the previous setup
Use upward and downward key to
select, and use TARE|PT key to confirm
Press TARE|PT key
ON|ZERO key  Upward key (0~ 9 digit ent ry)
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
2-2-13 Line Feed Input
Line feed would be available only when
was set as
or
.
Factory default: 2
Display shows the previous setup
Use upward and downward key to
select, and use TARE|PT key to
confirm
Press TARE|PT key
Press TARE|PT key
ON|ZERO key  Upward key (0~ 9 digit ent ry)
02005159
33
1
UNIT key  Downward key (0~9 digit entry)
1
TARE|PT key  Move cursor rightward
1
NET|B/G key  Move cursor leftward
ZSME300000288
EXCELL PRECISION CO., LTD
 Command Mode
Command Format A
Host
Slave
Command
Command
MZ
Zero
SO
Command mode
MT
Tare
UA
Switch to the first unit
MG
Gross weight
UB
Switch to the second unit
MN
Net weight
UC
Switch to the third unit
CT
Clear TARE value
UD
Switch to the forth unit
SC
Continuous transmission UE
Switch to the fifth unit
SA
Auto transmission
UF
Switch to the sixth unit
%
Stop continuous transmission and enter the command mode
Note: UA ~ UF settings are depended on the model of the scale
Command Format B
Host
Slave
RW
RG
RN
RT
RB
Note:
Command
Data
Read current weight
Read Gross weight
RH
RI
Read Gross (simple)
Read Net (simple)
Read comparison situation +
Read Net weight
RJ
current display of weight
(simple)
Read comparison situation +
Read TARE
RK
Gross (simple)
Read current display of
Read comparison situation +
RL
weight (simple)
Net (simple)
a. add % before the command to read continuously
b. add # before the command to transmit a stable value
RJ,RK,RL Command Description
If RL command entered, and
if weight is higher than HI (FNC 03 setting), and if the current weight is 10 kg, the following
will be displayed: “100+ 10.000”
If weight is higher than HI (FNC 03 setting), and if the current weight is 0.5 kg, the
following will be displayed: “001+ 0.500”
If weight is between HI and LOW, and if the current weight is 1 kg, the following will be
displayed: 010+ 1.000
02005159
34
ZSME300000288
EXCELL PRECISION CO., LTD
Read weight comparison setting value RS
: Groups(00 ~ 09) : Setting Items
Show “HI” presetting value
Show “LO” presetting value
HI
LO
Note : (Group) is various depended on different models
00  The first group
01  The second group
02  The third group
EX:
ANS:
RS02LO<CR><LF>
Show “LO” presetting value
RS02LOXXXXXX<CR><LF>
Command Format C
Host
Command+ Data
Slave
Command+ Data
Write weight comparison setting value WSXXXXXX
: Groups(00 ~ 09) : Setting Items XXXXXX: Setting Value
HI
Write “HI” setting value
LO
Write “LO” setting value
Note :  (Group) is various depended on different models
00  The first group
01  The second group
02  The third group
EX:
ANS:
Write “HI” setting value
WS00HI001000<CR><LF>
WS00HI001000<CR><LF>
Command Format D
Host
Slave
1
Data
Value (e.g. Price)
2
3
4
5
Position of decimal point
6
1
CR
LF
When the Slave receives this data format, it will transfer the data and display it on its LCD.
˙
 The function is effective when the weight value is over 0.
 The above 4 (ABCD) command formats are RS232 bi-directional. There are the
following error messages received by Slave terminal (scale).
Error messages:
E1: Wrong command
E2: Command format error (Wrong parameters)
E3: Command not recognised
02005159
35
ZSME300000288
EXCELL PRECISION CO., LTD
 Output data format
1. 7 places (first decimal place not included)
Weight format
Gross
S
Net
S
Tare
S
Plus OL
O
Minus OL O
Unstable U
T
T
T
L
L
S
Simple format
G/N
+ 1
G/N
+ 0
G/N
+ 0
Plus OL
+ SP
Minus OL - SP
,
,
,
,
,
,
G
N
T
G
G
G
S
T
R
S
S
S
,
,
,
,
,
,
.
1
1
SP
SP
2
2
2
SP
SP
3
3
.
SP
SP
.
4
3
SP
SP
+ 0 1 2 3 4 5
+ 1 . 2 3 . 4
+ 0 1 2 . 3 4
+ SP SP SP SP SP SP
- SP SP SP SP SP SP
+ 0 1 2 3 4 .
4
5
4
SP
SP
5
.
5
SP
SP
Comparison status + Simple format
Byte0 Byte1 Byte2 +/1
.
2
6 7 SP
5 6 t
5 6 SP
SP SP SP
SP SP SP
5 6 SP
SP o z
l . g
SP k g
CR LF
SP SP SP
SP SP SP
SP l b
6
6
6 CR LF
SP
SP
3
.
4
5
6
CR
7 SP
6 t
6 SP
SP SP
SP SP
6 SP
SP o z
l . g
SP k g
CR LF
SP SP SP
SP SP SP
SP l b
LF
Byte0 : HI 30H/31H
Byte1 : OK 30H/31H
Byte2 : LO 30H/31H
2. 6 places ( first decimal place not included)
Weight format
Gross
S
Net
S
Tare
S
Plus OL O
Minus OL O
Unstable U
Simple format
G/N
+
G/N
+
G/N
+
Plus OL
+
Minus OL -
T
T
T
L
L
S
.
1
1
SP
SP
,
,
,
,
,
,
G
N
T
G
G
G
2
2
2
SP
SP
S
T
R
S
S
S
3
3
.
SP
SP
,
,
,
,
,
,
+ 1 2 3 4 5 6
+ . 2 3 . 4 5
+ 1 2 . 3 4 5
+ SP SP SP SP SP SP
- SP SP SP SP SP SP
+ 1 2 3 4 . 5
.
4 5
4 5
.
3 4 5
SP SP SP
SP SP SP
Comparison status + simple format
Byte0 Byte1 Byte2 +/.
2
Byte0 : HI 30H/31H
Byte1 : OK 30H/31H
Byte2 : LO 30H/31H
02005159
6
6
6 CR LF
SP
SP
3
.
36
4
5
6
CR
LF
ZSME300000288
EXCELL PRECISION CO., LTD
 Serial Data Transfer/Receive Format
LSB
MSB
n,8 ,1
S
8-bit data
STOP
e,7 ,1
S
7-bit data
P STOP
o,7 ,1
S
7-bit data
P STOP
1
0
1
0
1
0
Note:
S
: Start bit
STOP: Stop bit
P
: Parity bit
02005159
37
ZSME300000288
EXCELL PRECISION CO., LTD
Appendix 1 ASCII Code Table
Symbol
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Q
R
S
T
U
V
W
X
Y
Z
02005159
ASC II
Code
41H
42H
43H
44H
45H
46H
47H
48H
49H
4AH
4BH
4CH
4DH
4EH
4FH
50H
51H
52H
53H
54H
55H
56H
57H
58H
59H
5AH
ASC II
Code
61H
62H
63H
64H
65H
66H
67H
68H
69H
6AH
6BH
6CH
6DH
6EH
6FH
70H
71H
72H
73H
74H
75H
76H
77H
78H
79H
7AH
Symbol
a
b
c
d
e
f
g
h
i
j
k
l
m
n
o
p
q
r
s
t
u
v
w
x
y
z
38
Symbol
0
1
2
3
4
5
6
7
8
9

ASC II
Code
30H
31H
32H
33H
34H
35H
36H
37H
38H
39H
0DH
ZSME300000288
EXCELL PRECISION CO., LTD
Appendix 2 7-Segment Display Characters
Digit
7 segment letter
Alphabet
7 segment letter
Alphabet
0
A
N
1
B
O
2
C
P
3
D
Q
4
E
R
5
F
S
6
G
T
7
H
U
8
I
V
9
J
W
K
X
L
Y
M
Z
02005159
39
7 segment letter
ZSME300000288
EXCELL PRECISION CO., LTD
Appendix 3 Wiring Instructions
Load cell wiring method
(1) As shown below, when a load cell is connected with a 4PIN cable,
SEN+ and SEN- can be unconnected. J11 and J12 on PCB must be
tin-soldered in short circuit.
(2) If Load Cell is 6PIN, please wire as the following diagram:
02005159
40
ZSME300000288
EXCELL PRECISION CO., LTD
RS232 wiring instruction
To connect RS232, please open the housing, RS232’s PIN connectors locate on the
bottom-right corner of the main board. The most common connection method is using
9PIN and 25PIN, as shown below:
PC
Printer
PIN
Function
2
Transmit Data
Female 9 PINS
Function
SG
3
Receive Data
RxD
5
Signal Ground
TxD
PIN
Function
PWH3
Male 25 PINS
Function
2
Receive Data
SG
3
Transmit Data
RxD
7
Signal Ground
TxD
PWH3
To use other connection methods, please identify the signal and following the
above principles. After it is finished, please install the housing by the
instruction in the SPECIAL NOTICE.
02005159
41
ZSME300000288
Was this manual useful for you? yes no