ENDATM
ENVIRONMENTAL CONDITIONS
Ambient / Storage Temperature 0 ... +50°C/-25 ... 70°C (without icing)
Relative Humidity
Relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C.
According to EN60529;
Front panel : IP65
Protection class
Rear Panel : IP20
Height
Max. 2000m
Please read this document carefully before using this product. The guarantee will be invalidated if the device is
damaged by not following instructions detailed in the manual. The company shall not be responsible for any damage or
losses however caused, which may be experienced as a result of the installation or use of this product.
ENDA ET2413K DIGITAL THERMOSTAT
Do not use the device in locations subject to corrosive and flammable gasses.
Thank you for choosing ENDA ET2413K incubator temperature controller devices.
35x77mm sized
On-Off control.
Relay output for heating operation
Single NTC probe input
Offset value can be adjusted for NTC input
Upper and Lower setpoint value limits can be adjusted.
Rotation duration and intervals can be adjusted
Manuel process feature
Humidity duration and intervals can be adjusted
Temperature unit can be selected °C or °F.
Transfer device parameter settings with ENDAKEY
- No power-up required
RS485 ModBus protocol communication feature (optional)
CE marked according to European Norms.
RH
°F
°C
SET
SET
ENDA
ET2413K
Order Code : ET2413K
1
ELECTRICAL CHARACTERISTICS
Supply Voltage
230V AC +%10 -%20, 50/60Hz ; 12V AC/DC ± %10 or 24V AC/DC ±%10
Power Consumption
Max. 5VA
2.5mm² screw-terminal connections
Connection
-60.0 ... +150.0°C (-76.0 ... +302.0°F)
Scale
Sensitivity
0.1°C (Can be selected as 0.1ºC or 1ºC.)
Accuracy
±1°C
±1%
Time Accuracy
4 digits, 12.5mm, 7 segment LED
Display
EN 61326-1: 2013
EMC
EN 61010-1: 2010 (Pollution degree 2, overvoltage category II)
Safety Requirements
OUTPUTS
NO+NC 250V AC, 8A (for resistive load), 1/2hp, 0.37kW 240V AC (for inductive load)
Heat Relay Output
2
1 - Supply Voltage
230........230V AC
24 ........24V AC/DC
12 ........12V AC/DC
Rotation Relay Output
Humidification Relay Output
4
6
7
Without load 30.000.000 switching; 250V AC, 8A (resistive load) 100.000 switching.
CONTROL
Control Type
Control Algorithm
Hysteresis
Single set-point control
On-Off control
Adjustable between 1 ... 20.0°C.
8
9
10
1
While cleaning the device, solvents (thinner, gasoline, acid etc.) or corrosive materials must not be used.
DIMENSIONS
3
4
5
6
7
8
61mm
RH
°F
35mm
ENDAKEY
SET
SET
ENDA
COM B
- Push the flush-mounting
clamp in direction 1 as shown
in the figure below.Then,pull
out the clamp in direction 2 .
°C
ET2413K
A
10 11 12 13
2
Flush mounting
clamp
1
28,5mm
Panel cut-out
71,5mm
Flush mounting
Rubber
clamp
Panel packing
Note:
1) Panel thickness should
be maximum 7mm.
2) If there is no 60mm free
space at the back side of
the device,it would be
difficult to remove it from
the panel.
3
4
5
ET2413K-24
DIGITAL DEFROST
THERMOSTAT
6
7
9 10 11 12
8
Made in Turkey
8 680407 711291
ENDA INDUSTRIAL ELECTRONICS
SN: XXXXXXXXX
TM
ENDA
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Barbaros Cad. No:18 Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
2
Neutral
1) Mains supply cords shall meet the requirements of
IEC 60227 or IEC 60245.
230V AC 2) In accordance with the safety regulations, the power
Supply
supply switch shall bring the identification of the
Fuse should
relevant instrument and it should be easily
be connected Cable size: 1,5mm²
accessible by the operator.
ROTATE
250V AC 8A
7
HEAT
250V AC 8A
RESISTIVE
LOAD
Line
Switch
HUMIDIFIER
250V AC 8A
6
24V AC/DC ±10%
50/60Hz 5VA
184-253V AC
50/60Hz 4VA
Note:
NTC
GND
SUPPLY:
Fuse
F 100 mA
250V AC
ENDAKEY
NOTE:
1
Holding screw
0.4-0.5Nm.
Equipment is protected throughout by
DOUBLE INSULATION
5mm
For removing mounting clamps:
GND
9
Depth
77mm
Made in Turkey
NTC
HUMIDFIER
250V AC 8A
2
HEAT
250V AC 8A
RESISTIVE
LOAD
ENDAKEY
11 12
Suitable for flush -panel mounting
W77xH35xD61mm
Approx. 190g (After packing)
Self extinguishing plastics.
Housing Type
Dimensions
Weight
Enclosure Material
8 680407 711321
24V AC/DC ±10%
50/60Hz 5VA
NTC
GND
Made in Turkey
230V AC +10% -20%
50/60Hz 5VA
5
Without load 30.000.000 switching; 250V AC, 8A (resistive load) 100.000 switching.
Humidification Relay
ENDA INDUSTRIAL ELECTRONICS
ET2413K-24-RS
DIGITAL DEFROST
THERMOSTAT
8 680407 711284
HEAT
250V AC 8A
RESISTIVE
LOAD
ROTATE
250V AC 8A
HUMIDIFIER
250V AC 8A
3
SN: XXXXXXXXX
ROTATE
250V AC 8A
ENDA INDUSTRIAL ELECTRONICS
2
Rotation Relay
HOUSING
ENDA ET2413K is intended for installation in control panels. Make sure that the device is used
only for intended purpose. The electrical connections must be carried out by a qualified staff
and must be according to the relevant locally applicable regulations. During an installation, all of
the cables that are connected to the device must be free of electrical power.
1
NO 250V AC, 8A (for resistive load), 1/2hp, 0.37kW 240V AC (for inductive load)
LIFE EXPECTANCY FOR RELAYS
Heat Relay
Without load 30.000.000 switching; 250V AC, 8A (resistive load) 100.000 switching.
2 - ModBus (Optional)
RS.........ModBus
Blank.....N/A
ET2413K-230
DIGITAL DEFROST
THERMOSTAT
NO 250V AC, 8A (for resistive load), 1/2hp, 0.37kW 240V AC (for inductive load)
1/3
ET2413K-EN-01-170327
°F.
FAHRENHEIT LED : In parameter value or the measured temperature value “°F” unit while this LED
lights up. In the hidden menu at the same time the user menu parameter is shown the LED lights up.
RH HUMIDIFIER LED : Lights up while humidifying process.
ROTATION LED : Lights up while rotating process.
HEAT LED : Lights up while heating (heat output is active).
SET
RH
°F
°C
While in "Programming Mode", provides the transition to the next parameter. If parameter is being
adjusted, it increases parameter’s value. Constantly holding this key, the parameter value rapidly
increases.
While in "Programming Mode", provides the transiton to the previous parameter. If parameter is
being adjusted, it decreases parameter’s value. Constantly holding this key,the parameter value
rapidly decreases.
SET
SET
ENDA
In "Running Mode", indicates the set value.
In “Programming Mode”, indicates the selected parameter value.
ET2413K
FRONT PANEL COMMANDS
1. Viewing and Changing The Set Value
-24.0
1000
SET
Measurement
value
-30.0
-29.9
-30.0
SET
While in "Running Mode", if
key is pressed set value is displayed for 3 seconds. While in this case, the set value is changed with
keys.
2. Locking and Unlocking Keys
-24.0
1000
Measurement
value
SET
Loc
unl
Keys are locked.
Keys are unlocked.
SET
While in "Running Mode", if
keys are pressed together for 2 seconds, Loc message is displayed and the keys are locked.
If the keys are locked
SET
SET
keys are pressed for 2 seconds again, unL message is displayed and keys are unlocked. While keys are locked and if
key is pressed, set value can be
SET
displayed but can not be changed. While the keys are locked and if any key is pressed ( except
key ), Loc message appears.
3. Manuel Rotating Process
While in the "Running Mode", if
key is pressed for 2 seconds, defrost process starts manually. If r.dur parameter is 0 , manual defrost will be inactive.
4. Activating / Inactivating The Control Outputs
-24.0
1000
Measurement
value
When in the "Running mode", if
C.dI5
The control ouput becomes inactive.
CENB
The control output becomes active.
key is pressed for 2 seconds, C.dI5 message is displayed and control outputs becomes to the inactive position, the device
works as the indicator. When the control outputs are disabled; if
key is pressed for 2 seconds C.EnB is disabled and the device continues to control function.
5. Changing Parameter Values
If
keys are pressed together for 2 seconds “User Menu” is entered, then first parameter’s name is displayed in the user menu.
SET
While a parameter is selected, by pressing
key, parameter value can be displayed. This parameter can be changed with
keys. If no operation
SET
performed for 3 seconds or during this time,
key is pressed while parameter value displayed, parameter name will be displayed again. While parameter name
displayed, if by pressing together
keys, "Running Mode" is entered.
ERROR MESSAGES
Pfa
P5C
NTC probe is broken or not connected.
NCT probe is short circuit.
Temperature value is lower than the scale.
Temperature value is higher than the scale.
FACTORY SETTINGS
If
key is held down while the device is powered up, d.PAr message appears and factory parameters restored.
TM
TM
ENDA
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Barbaros Cad. No:18 Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
2/3
ET2413K-EN-01-170327
OUTPUT GRAPHICS
Power
HEATING OUTPUT GRAPHICS
SET+ r.HYS
SET
SET- r.HYS
Read
button
HE
(heating)
HEAT
(Temperature)
output
control
TRANSFERRING THE PARAMETERS FROM ENDAKEY TO DEVICE
ROTATION (CYCLE) OUTPUT GRAPHICS
While in "Running Mode", if
key on device or “Read” button on “ENDAKEY”
is pressed, “ dL ” message appears on display. While " dL " message
displayed on device and if
key is pressed, parameter values from
"ENDAKEY” transferred to the device. If the parameter transfer is successful,
the " rEf " message appears and the device begins to work with the loaded
parameter values. If the parameters are wrong, incorrect or " ENDAKEY " is
failure, " Err " message appears and parameters will not be changed on
device.
r.int
The time between two consecutive rotating operations
r.dur
Rotation duration (If r.dur=0 is selecled, automatic and manual rotating is disabled.)
ROTATION
INDICATOR
LED
Rotation
mode
TRANSFERRING THE PARAMETERS FROM DEVICE TO ENDAKEY
While in "Running Mode",
if key is pressed on device, “ uL ” message
appears on display and
key is pressed again, parameters are read and
transferred to the device. If process succes, “ 5uc ” message appears. In case
of failure, " Err " message displayed and parameters will not be changed on
device.
NOTE 1 : No power-up required for transfering the parameter by using
"ENDAKEY". For long battery life, "ENDAKEY" must be disconnected from
device after the transferring process.
NOTE 2 : Please specify at order “ENDAKEY” if required.
Rotation
mode
ROTATION
OUTPUT
HUMIDIFICATION OUTPUT GRAPHICS
h.int
Time between two consecutive humidifications
h.dur
Humidification duration (If h.dur=0 is selected, humidification is disabled.)
HUMIDIFICATION
INDICATOR
LED
Humidification
Mode
HUMIDIFICATION
OUTPUT
CONTROL PARAMETERS
uPL
LoL
HY5
oFF
Upper limit for setpoint
Lower limit for setpoint
Heating differential (hysteresis)
Offset value for heating
Min.
Max.
LoL
-60.0
0.1
-20.0
150.0
uPL
20.0
20.0
UNIT
DEFAULT
VALUE
°C
°C
°C
°C
150
-60
2
0
CONFIGURATION PARAMETERS
Unit
Dpnt
Temperature unit
Decimal point (no= decimal point isn’t shown 22°C,YE5=decimal point is shown 22.3°C.)
°C
°F
°C
No
YE5
YE5
ROTATION CONTROL PARAMETERS
r.dur
Rotating duration ( If r.dur = 0 selected, automatic and manual rotation is disabled ).
0:00
99:00
r.int
Time between 2 consecutive rotation.
0:00
99:00
min:sec
hr:min
1:00
min:sec
hr:min
1:00
1:00
HUMIDIFICATION CONTROL PARAMETERS
h.dur
Humidification duration ( If h.dur = 0 selected, automatic and manual humidification is disabled ).
0:00
99:00
h.int
Time between 2 consecutive humidification.
0:00
99:00
TM
ENDA
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Barbaros Cad. No:18 Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
1:00
3/3
ET2413K-EN-01-170327