Short Manual / Kurzanleitung
Thermal cyclers / Thermocycler
peqSTAR 2X double block thermocycler
peqSTAR 96X, 384X and in situ X thermocycler /
peqSTAR 2X Doppelblock-Thermocycler
peqSTAR 96X, 384X und in situ X Thermocycler
1212E+D
Creating the future together.
Short Manual peqSTAR thermal cyclers
CONTENTS
1 Safety instructions
2 License Notice
3 System Overview
4 Installation
4.1
Content of delivery
4.2
Installing the thermal cycler
4.3
Operation of the thermal cycler
5 Cycler Operation
5.1
Start screen = 'Run'
5.2
Running stored programs ('Run'):
5.3
Incubation mode ('Run')
5.4
Create or edit PCR programs ('Programs')
5.5
Further dialogues: Diagnostics, GLPs, System
5.6
Further questions…?
6 Technical Service and Ordering Information
7 Technical data
8 Maintenance and repair of the Thermal cycler
8.1
Cleaning, maintenance and repair of the thermal cycler
8.2
Over temp test
8.3
Temperature check-up
8.4
Calibration
9 Warranty
2
2
3
3
3
3
3
4
4
4
5
5
9
10
11
12
14
14
14
14
14
15
A more detailed instruction manual in English or German language is available
for download on our website www.peqlab.com.
PEQLAB_v1212D
1
Short Manual peqSTAR thermal cyclers
1
SAFETY INSTRUCTIONS
Before the first use of the thermal cycler please read this instruction manual entirely.
Special notice should be taken of the following
Caution of dangerous voltage
Please ensure the voltage indicated on the device exactly matches your local electrical supply
Caution of dangerous explosive material
Explosive or reactive material mustn’t be heated in the thermal cycler.
Caution - liquids
Ensure that no liquids can enter the device.
Caution - hot surfaces
The thermal plate and the heated lid quickly reach temperatures above 50 °C. There is danger
of burning! Keep the lid closed until the temperatures reaches 30 °C or less.
Only use materials (tubes, tube caps, plates and sealing films) which are heat resistant to
120 °C
Caution – environment
The ventilation of the device must not be covered.
2
LICENSE NOTICE
This instrument is licensed for research and development and for uses other than human in vitro diagnostics under
one or more of the following patens of Applera Corporation: U.S. Patent Nos. 5,656,493, 5,038,852,
5,333,675, 5,475,610 (claims 1-159 and 164-166), 6,703,236 (claims 7-10) and 7,238,517 or corresponding
claims in their non-U.S. counterparts. No right is conveyed expressly, by implication or by estoppels under any
other patent of Applera, including but not limited to U.S. Patent No. 6,814,934 and its non-U.S. counterparts,
which describe and claim thermal cyclers capable or real-time detection.
PEQLAB_v1212D
2
Short Manual peqSTAR thermal cyclers
3
SYSTEM OVERVIEW
The peqSTAR thermal cyclers are the ideal PCR devices for all molecular biological and biochemical laboratories
working in the field of basic research or routine diagnostic. Eight or sixteen 250 °C HTR (High Temperature
Range) Peltier elements allow quick changes in temperature up to 5 °C/sec (max). The individual monitoring and
controlling of each of the eight Peltier elements equipped with 'Long Life Technology' ensure the tightest correlation
of target and actual temperature across the entire thermal plate leading to highest reproducibility of the results.
Each block is supplied with a lid heating to avoid condensation (except for InSitu blocks).
The peqSTAR thermal cycler is also equipped with a touch-sensitive, graphic colored TFT display of the newest
generation, allowing operation of the device by simply touching symbols and graphic elements on the screen. This
leads to easy and intuitive handling and programming of the device. Additionally a mouse can be connected via
USB. Further-more an Ethernet port is available to connect the device with the Microsoft Windows® or Linux
network. It is therefore possible to use an external server for archiving programs and GLP reports which will be
available globally in the network (e.g. for other thermal cyclers).
It is also possible to control the thermal cycler by a PC integrated to the network or wireless ZigBee connection. The
provided software offers the same functions as the internal thermal cycler software, except for some system
settings. Transfer of data can be done by using commercially available USB memory sticks enabling a nearly
unlimited saving space for programs and GLP reports, which can be printed using a printer connected to the
thermal cycler via the network.
Notice:
4
4.1
The USB ports support only standard cable mice and standard USB sticks.
The maximum acceptable length of the connecting cable of the port is 3 m!
INSTALLATION
Content of delivery
1 Thermal cycler
1 Power cord
1 Short manual
(A detailed instruction manual in English or German is available for download on our website www.peqlab.com.)
1 CD PC-software 'peqSTAR Manager'
Please check delivery for completeness and transport damage upon arrival. If any transport damage is noticed,
contact your distributor or manufacturer immediately.
NOTICE: IF ANY DAMAGE OF THE DEVICE SHOULD BE FOUND DO NOT USE THE DEVICE!
4.2
Installing the thermal cycler
Check the packaging for any transport damage. Remove the packaging and place the thermal cycler onto a solid
surface. The device should not be exposed to direct sunlight. There should be enough space to make sure that the
ventilator grill on the front and back are not covered and sufficient ventilation is provided. There should be a
distance of at least 25 cm to the wall or the next device. Two thermal cyclers should not be placed back-to-back.
Room temperature between 4 °C und 25 °C is advised.
All original packaging should be stored in case a need arises to return the instrument to the supplier.
4.3
Operation of the thermal cycler
Before starting up the thermal cycler for the first time make sure that the voltage range of the device is suitable to
the electricity supply. Plug in the mains plug of the thermal cycler to the electricity supply. The outlet and the power
switch are located at the back of the device. The serial number is written onto a typed label at the bottom of the
device.
The device should only be used by authorized staff. To ensure this, a user administration with appointing rights for
single users is available in the device.
PEQLAB_v1212E
3
Short Manual peqSTAR thermal cyclers
5
CYCLER OPERATION
The operation of the thermal cycler can be run by the internal touch screen, either using the touch sensitive surface
of the display or by using a mouse connected to a free USB port.
If the user has to enter numbers or text the corresponding field has to be touched on the touch screen or clicked
with the mouse. An appropriate keyboard will appear on the screen according to the expected entry.
5.1
Start screen = 'Run'
The last ran PCR program is displayed in the 'Run' dialogue and can directly be restarted with the button 'Start
PCR'.
Note: At peqSTAR 2X double block thermal cyclers the desired block can be chosen in the dialogues 'L' (left) and
'R' (right). For peqSTAR 96X, 384X and InSitu systems this option is not available.
5.2
Running stored programs ('Run'):
1. Open the selection dialogue in the 'Run' dialogue with the button
.
2. Select the desired program in the selection dialogue and load to the 'Run' dialogue with the button
'Open'.
3. The loaded program is started by using the
PEQLAB_v1212E
4
button.
Short Manual peqSTAR thermal cyclers
5.3
Incubation mode ('Run')
Use the button
in the 'Run' dialogue to set the device into the incubation mode. The temperature for the
thermal plate and lid can be set manually. The clock shows the time passed by in the incubation mode.
5.4
Create or edit PCR programs ('Programs')
1. To create or edit PCR programs go to the 'Programs' dialogue.
2. By touching the 'Programs' dialogue the first time during the session the selection dialogue shows up
where a new program can be set over 'New' or an existing program can be opened over 'Open' to edit.
3. On the left side of the program editor all available commands are displayed, on the right side of the
editor the list of commands (Steps) the program is composed of is shown. After selecting a program
command out of the list in the left window it can be inserted into the desired place via the
4. Via double click or the 'Edit' button
parameters of single steps can be changed later.
5. Selected program steps can be deleted from the program by using the 'Delete' button
button.
.
6. Use the 'Save' button
to save the program.
7. Save and start programs in one step by simply using the 'Save&Run' button
.
8. For even faster programming the peqSTAR thermocycler comes with a selection of template programs,
which could easily be edited and saved under a new program name.
List of available commands
Adding and removing
program steps
PEQLAB_v1212E
5
Program name
Program window
Short Manual peqSTAR thermal cyclers
Available commands:
Lock Lid
Cycle
Locks the lid avoid accidental opening.
Combines Open Loop, Close Loop,
Temperature, Ramp and Gradient.
Unlock Lid
Pause
Unlocks the lid in case that the lid has been
locked in a previous program step.
Interrupts the program temporarily at the
definable temperature.
Heat Lid
Store
Turns the lid heating on/off.
Holds the definable temperature, turns off
the heated lid automatically.
Temperature
Start Cycle
Temperature step for definable time,
Temperature and Time Incr./Decr. function.
Manual loop beginning.
Gradient
Close Cycle
Realizes different temperatures at the same
time for protocol optimization
(see 'FlexGradient' below for more
information).
Manual loop ending.
Sample program
Create a new program with 'New'. Enter a program name and confirm with OK to access the programming
functions.
01
Heated Lid on, temperature 110 °C
Heat Lid
A 'Heat Lid' step is inserted in each program by default as first step. This step can be edited
or deleted later. The standard settings could be changed under 'System' => 'Firmware' =>
'Settings'.
For 'Store' or 'Temp' steps < 30 °C the lid heating is automatically switched off.
To enter another program step select the desired command out of the list of the available command and add it to
the program window with the '+' arrow. Thereupon the corresponding parameters will be asked.
02
Denaturation at 95 °C for 2 minutes
Temp
03
07
Cycle
04
Open the 'Cycle' function
→ Enter number of cycles – 30 cycles
→ 3 x 'Temp' step
→ Terminate 'Cycle' function with 'Close Cycle'
Denaturation at 95 °C for 30 seconds
Temp
PEQLAB_v1212E
6
Short Manual peqSTAR thermal cyclers
Annealing at 65°C in the 1st cycle for 30 seconds
05
Temp
'Touchdown'-PCR via Temperature-'Decrement'. The annealing temperature of 65 °C in the
1st cycle is reduced by 0.4 °C each cycle. This leads to high specifity in the first cycles and
high amplification efficiency in the later cycles.
The reduction of the cooling rate at this step of the program is not necessary, but should
only be an example. It might make sense e.g. for quantitative annealing of complementary
oligo nucleotide or at annealing steps in the first cycles, when primer annealing should be
very efficient because of low template amount (e.g. after cDNA synthesis)
Elongation at 72 °C for 1 minute in the 1st cycle
06
Temp
07
Via Time -'Increment' the elongation time is prolonged by 5 s each cycle. The incremental
prolongation of the elongation time is useful with high numbers of cycles (>30), when the
Polymerase activity declines.
Terminate 'Cycle' function with 'Close Cycle'.
Alternative programming of steps 03 – 07:
03
Open program cycle – 30 cycles.
Start Cycle
04
Denaturation at 95 °C for 30 seconds
Temp
Annealing at 65°C in the 1st cycle for 30 seconds
05
Temp
'Touchdown'-PCR via Temperature-'Decrement'. The annealing temperature of 65 °C in the
1st cycle is reduced by 0.4 °C each cycle. This leads to high specifity in the first cycles and
high amplification efficiency in the later cycles.
The reduction of the cooling rate at this step of the program is not necessary, but should
only be an example. It might make sense e.g. for quantitative annealing of complementary
oligo nucleotide or at annealing steps in the first cycles, when primer annealing should be
very efficient because of low template amount (e.g. after cDNA synthesis)
Elongation at 72 °C for 1 minute in the 1st cycle
06
Temp
07
Via Time -'Increment' the elongation time is prolonged by 5 s each cycle. The incremental
prolongation of the elongation time is useful with high numbers of cycles (>30), when the
Polymerase activity declines.
Terminate program loop with 'Close Cycle.
Close Cycle
08
Final elongation at 72 °C for 5 minutes
The final elongation assures the complete synthesis of all amplificats.
Temp
09
8 °C forever, lid heat is turned off
Store
PEQLAB_v1212E
By the 'Store' step the lid heat is turned off automatically. This avoids that lid heat and
Peltier elements of the block work against each other. Especially at longer incubation times
at low temperature this increases the life time of the Peltier elements and reduces the power
consumption.
7
Short Manual peqSTAR thermal cyclers
FlexGradient Technology™:
The temperature of each of the eight rows of the block is individually controlled! For the first time you can choose
either the perfectly linear temperature gradient – ideally suited for PCR optimizations, or independent lane control
– ideally suited for the use of different primer pairs in one PCR run.
First, the temperature range is defined by the buttons
'Center' and 'Gradient ±'. The standard gradient is
always realized from the top to the button over the 8
rows of the block. Moreover with the peqSTAR 96X you
have the opportunity to set a 16-gradient. In this case
different temperatures are realized in the left and right
part of the thermoblock.
As in this case you will also have a gradient in the
lanes from left to right, the temperatures displayed in
the software are only realized in the rows 3 and 10.
By selecting one of the temperature buttons
different options will be available: When the checkbox
was deactivated before, the single
temperatures of each row can be adjusted within a
special range using the button 'Adjust Temperature'.
Moreover the selected temperature can be used as new
center value or the gradient step can be changed in a
temperature step using the selected temperature.
Program options:
When setting up a new program, when saving or during programming (via the Info button) the program options
can be set. The following options are available for the user.
1. Power Fail Denaturation:
If a power failure had happened while a program was running the thermal cycler restarts the program at the point it was
stopped as soon as the power returns.
If the option 'Power Fail Denaturation' is activated the thermal block will heat to the stated temperature for the set time before
continuing with the program.
2. Global Program Ramp:
Using this function a global ramp rate in °C/s can be programmed. This global ramp rate will be used in all temperature and
gradient steps of the PCR program, but could be changed in a single temperature or gradient step manually by entering the
desired ramp value at the step.
3. Gradient Control
Using this option you can decide whether in a gradient step the plateaus (Plateaus synchronized) or the Ramps (Ramps
synchronized) should be synchronized for all rows.
4. Tube Control
Via tube control you have the possibility to adjust temperature regulation for a temperature step according to sample volume. If
necessary, the temperature regulation will perform short term deliberate overshoots and undershoots depending on the entered
sample volume during PCR run.
Especially if combined with the emulation mode this function can be helpful if the emulated cycler used the tube control function
before, too.
PEQLAB_v1212E
8
Short Manual peqSTAR thermal cyclers
5. Emulation
Using the emulation function the thermal cycler will perform PCR like the chosen cycler, emulating its performance. This function
could be useful if a PCR program should be transferred to the peqSTAR 2X from another type and model of thermal cycler.
5.5
Further dialogues: Diagnostics, GLPs, System
Diagnostics: Monitoring the actual temperatures of each of the six Peltier elements during a run. In suspicion case
of an instrument malfunction have a look at the Peltiers’ temperature profile.
GLPs: Displaying and organizing GLP (Good Laboratory Practice) reports created by the thermal cycler. This
option is mainly for labs who need to save GLP reports for documentation purpose.
System: Configuration of the system and changing of the general settings of the device.
PEQLAB_v1212E
9
Short Manual peqSTAR thermal cyclers
Elements of the system dialogue:
Time
Hide/Show Curser
Setting up date and time.
Hide or show the cursor on the screen.
User Call
User Management
Setting of acoustic signals or email
notification (Pause, Store, Program End)
Define user and user groups (and their
access rights).
Network settings
Network settings for network integration of
the peqSTAR.
5.6
The default password for the user 'Administrator' is: 'admin'.
'admin' Via
'Logout' button the current user can be logged off. Hereby the user
'Guest' is automatically logged in. This user only has minimal user
rights. These settings can be changed by the administrator if desired.
Hardware
Firmware
Display hardware versions of Control Unit,
Power Unit and Thermo block.
Display, administrate and update software
(Control/Power Unit, Thermo block)
Printer
Display Settings
Set up a printer path for an existing network
printer.
Change settings of the screen display.
Protocol
Service
Internal communication of the thermal cycler
for failure diagnosis
Service function
(only for authorised technical Service)
LAN Control
USB Recovery
Functional button for PC and Master/Slave
control
Manual mounting of USB sticks that are not
automatically detected.
Further questions…?
:
Direct help
Use this button to activate (deactivate) the direct help mode. If the mouse pointer is visible a question mark appears
alongside. By touching a function button or other dialogue element the corresponding help window will appear.
A more detailed instruction manual in English or German is available for download on our website
www.peqlab.com.
PEQLAB_v1212E
10
Short Manual peqSTAR thermal cyclers
6
TECHNICAL SERVICE AND ORDERING INFORMATION
Should you have any questions about the peqSTAR thermocycler, please visit www.peqlab.com to find the
respective contact person.
This short manual is delivered with the following products:
Product
peqSTAR 2X Universal
peqSTAR 2X Universal
Gradient
peqSTAR 96X Universal
peqSTAR 96X Universal
Gradient
peqSTAR 96X HPL
peqSTAR 96X HPL
Gradient
peqSTAR 384X HPL
peqSTAR in situ X
PEQLAB_v1212E
Cat. No.
with 2 x 48 well Universal blocks with standard lids
for 48 x 0.2 ml tubes, 48 well PCR plates or
24 x 0.5 ml tubes with flat caps each
with 2 x 48 well Universal Gradient blocks and standard lids
for 48 x 0.2 ml tubes, 48 well PCR plates or
24 x 0.5 ml tubes with flat caps each
with 96 well Universal block and standard lid
for 96 x 0.2 ml tubes, 96 well PCR plates or
48 x 0.5 ml tubes with flat caps
with 96 well Universal Gradient block and standard lid
for 96 x 0.2 ml tubes, 96 well PCR plates or
48 x 0.5 ml tubes with flat caps
with 96 well block and High Pressure Lid
for 96 x 0.2 ml tubes or 96 well PCR plates
with 96 well Gradient block and High Pressure Lid
for 96 x 0.2 ml tubes or 96 well PCR plates
with 384 well block and High Pressure Lid
for 384 well PCR plates
with in situ block for up to four microscope slides
and integrated buffer reservoir
11
95-07002
95-08002
95-05002
95-06002
95-05002-HPL
95-06002-HPL
95-050384-HPL
95-050FLAT
Short Manual peqSTAR thermal cyclers
7
TECHNICAL DATA
General characteristics
2 x 8 or 16 Peltier elements with Long-Life-technology and 2 x 8 or 16 control circuits with Pt 1000
temperature sensors, respectively
Maximal heating and cooling rate: 5 °C/s
Block uniformity (at 72 °C): ± 0.20 °C
Temperature range thermoblock: 4 to 105 °C
Control accuracy thermoblock: ± 0.1 °C
Variable ramping: 0.1 to 3.0 °C/s
Increment/Decrement time: 0:01 to 9:59 minutes
Increment/Decrement temperature: 0.1 to 9.9 °C
Lid heating temperature range: 40 to 120 °C
Programmable Lid locking mechanism
4 x USB, 1 x Ethernet interfaces (MS Windows® or Linux), maximal length of power cord 3 m!
Note: USB interfaces only support standard USB sticks, cord mouse and keyboards!
Dimensions (B x H x D): 30 x 28 x 38 cm
Weight incl. block: 13.3 kg
Electrical power supply: 90 – 240 V AC, 50 – 60 Hz, 850 VA
Fuse: 10 A time lag
Degree of pollution: 2
Ambient temperature: 10 °C to 30 °C
Maximum relative humidity: 70 %
Maximum altitude for the use of this system: up to 2000 m above sea level
For indoor use only!
Noise level: < 37 dB (A)
Gradient feature (optional)
Maximal gradient range over 8 or 16 columns, respectively: 30 °C (± 15 °C)
Temperature range Gradient: 35 to 105 °C
Gradient accuracy: ± 0.1 °C
Gradient function can be upgraded even after purchase and on-site
FlexGradient™ technology: ideal, linear Gradient or independent lane control
'Gradient Control': synchronized plateau times or ramp rates
User interface/functions
Touch sensitive TFT-Display (VGA, Graphic, 65535 colours), operation via USB mouse possible
Direct help function, Tm- and elongation time calculator
Internal flash memory for 500000 typical PCR programs in free configurable folders/subfolders
Unlimited number of programs via network PC or USB memory stick
Free computer Software for the generation of PCR programs on the computer
Unlimited number of steps/program
Password protected user accounts with variable access rights
GLP reports for complete recording of all runs
Quickstart function of the last used program
Automatic restart after power failure, Power-Fail-Denaturation, Instant Incubation
Remote control of the system by PC software
Free choice of mp3 signal sounds
User notification by email
Master/Slave control (wireless option)
PEQLAB_v1212E
12
Short Manual peqSTAR thermal cyclers
Block options (partly available as interchangeable blocks)
peqSTAR 2X/2X Gradient:
Two 48 well Universal blocks with: lid heating and automatic height adaption for 48 x 0.2 ml tubes, 48 well PCR
plates or 24 x 0.5 ml tubes with flat caps.
peqSTAR 96X/96X Gradient:
96 well Universal block with: lid heating and automatic height adaption for 96 x 0.2 ml tubes, 96 well PCR plates
or 48 x 0.5 ml tubes with flat caps.
peqSTAR 96X HPL/96X HPL Gradient:
96 well block with High Pressure lid (HPL, 100 – 250 N) for the safe sealing of 96 well PCR plates
peqSTAR 384X HPL:
384 well block with High Pressure lid (HPL, 100 – 250 N) for the safe sealing of 384 Well PCR plates
peqSTAR In situ X:
In situ block with integrated buffer reservoir, for up to 4 microscope slides
PEQLAB_v1212E
13
Short Manual peqSTAR thermal cyclers
8
8.1
MAINTENANCE AND REPAIR OF THE THERMAL CYCLER
Cleaning, maintenance and repair of the thermal cycler
For safety reasons the device must be switched off and the mains plug pulled before cleaning is carried out!
The surface of the thermal cycler can be wiped off with a damp cloth or ethanol. Do not use aggressive or scouring
cleaners or organic solvents for cleaning. The device should be protected from aggressive chemicals. Make sure
that no liquid penetrates into the interior of the device.
If sample liquids penetrates into the drill holes of the thermal cycler clean them immediately with mild soap
followed by distilled water. You can use mild disinfectants for decontamination.
Maintenance (Overtemp test and temperature check-up, see 8.2 and 8.3) should be carried out regularly to detect
possible technical failures.
The electric fuses are placed at the back of the device between the power switch and the mains plug.
After pulling off the main plugs and pushing upwards a safety catch the fuses can be changed. Use only fuses with
correct values (indications about the fuse type are placed at the back of the device).
The device may only be opened by a qualified specialist. Unauthorized work on the device voids the warranty. All
kinds of repairs may only be carried out by authorized persons using original replacement parts. The replacement
of single components of the device should not be done by the user but exclusively by authorized specialists.
Therefore the device must be sent to the manufacturer. The surface of the device should be decontaminated with a
lint free cloth soaked with 70 % ethanol before sending. The drill holes of the thermoblock shall be cleaned with a
1.5 % hypochlorite solution with the help of a cotton bud. Please fill in and sign the decontamination certificate
and send it to the manufacturer together with the device.
8.2
Over temp test
A periodical test of the overtemperature protection (→ System → Hardware → Thermo Block → Overtemperature
Test) is recommended.
8.3
Temperature check-up
Maintenance of the temperature at regular intervals (e.g. annually) is recommended. The thermometer for use
should be equipped with wired sensing devices (e.g. Pt1000) and should have an accuracy of measurements of
1/10 class B (∆T = ±0.03 °C) in the range of 0 to 100 °C. While processing the temperature control the room
temperature should be 25 °C at maximum.
8.4
Calibration
If there is a reasonable suspicion that the device has deviated from the calibration range, because the variations of
the temperature check-up are considerably out of the acceptable tolerance, a new calibration must be carried out
by the manufacturer. Please contact your local distributor.
PEQLAB_v1212E
14
Short Manual peqSTAR thermal cyclers
9
WARRANTY
The period of warranty for the product at hand is 24 month from date of purchase. The warranty does not apply
for defects caused by incorrect, improper or inappropriate use. The customer has to check up on obvious defects
promptly upon receipt of the device and send a written report. Otherwise the claim for warranty expires. The
manufacturer reserves the right to carry out changes to the thermal cycler or to the technical documentation of the
thermal cycler on the basis of product development or improvement without prior notice.
10 DISPOSAL INSTRUCTION
This product is subject to the WEEE disposal regulations and cannot be disposed of with regular
waste.
Please contact PEQLAB Biotechnologie GmbH or your local distributor for disposal of peqSTAR
thermal cyclers. Waste electronic equipment is not useless waste. Environmentally friendly
disposal can help to retrieve valuable resources. Thereby you contribute to the protection of the
environment and human health.
NOTES
PEQLAB_v1212E
15
Kurzanleitung peqSTAR Thermocycler
INHALTSVERZEICHNIS
1 Sicherheitshinweise
2 Lizenzhinweis
3 Systemüberblick
4 Installation
4.1
Lieferumfang
4.2
Aufstellung des Thermocyclers
4.3
Inbetriebnahme des Thermocyclers
5 Bedienung
5.1
Startbildschirm = 'Run'
5.2
Starten eines gespeicherten Programms ('Run'):
5.3
Inkubationsmodus ('Run')
5.4
Erstellen oder Editieren von PCR-Programmen ('Programs')
5.5
Weitere Reiter: Diagnostics, GLPs, System
5.6
Noch Fragen…?
6 Technischer Service und Bestellinformationen
7 Technische Daten
8 Wartung & Reparatur des Thermocyclers
8.1
Reinigung, Wartung und Reparatur des Thermocyclers
8.2
Prüfung des Übertemperaturschutzes
8.3
Temperaturüberprüfung
8.4
Kalibrierung
9 Gewährleistung
10 Entsorgungshinweis
17
17
17
18
18
18
18
19
19
19
20
20
24
25
26
27
29
29
29
29
29
30
30
Ein ausführlicheres Manual ist auf unserer Homepage www.peqlab.com
in Deutsch und Englisch zum Download verfügbar.
PEQLAB_v1212D
16
Kurzanleitung peqSTAR Thermocycler Serie
1
SICHERHEITSHINWEISE
Bitte lesen Sie vor Gebrauch des Thermocycler diese Bedienungsanleitung vollständig durch. Die folgenden
Bestimmungen sind unbedingt zu beachten:
Warnung vor gefährlicher elektrischer Spannung
Vor der Inbetriebnahme ist die Spannungsangabe auf dem Typenschild mit den
Netzbedingungen abzugleichen.
Warnung vor explosionsgefährlichen Stoffen
Explosive oder reaktive Stoffe dürfen im Thermocycler nicht temperiert werden.
Warnung - Flüssigkeit
Es ist sicherzustellen, dass keine Flüssigkeit in das Gerät gelangen kann.
Warnung vor Verbrennungsgefahr
Thermoblock, Heizdeckelinnenseite und Reaktionsgefäße erreichen sehr schnell Temperaturen
oberhalb 50 °C. Es besteht Verbrennungsgefahr! Heizdeckel geschlossen hallten, bis
Temperaturen um oder unter 30 °C erreicht sind. Es dürfen keine Materialien (Gefäße,
Deckelverschlüsse) verwendet werden, die nicht ausreichend temperaturbeständig (bis 120 °C)
sind.
Warnung – Umgebung
Die Lüftungsöffnungen des Geräts dürfen nicht abgedeckt werden.
2
LIZENZHINWEIS
This instrument is licensed for research and development and for uses other than human in vitro diagnostics under
one or more of the following patens of Applera Corporation: U.S. Patent Nos. 5,656,493, 5,038,852,
5,333,675, 5,475,610 (claims 1-159 and 164-166), 6,703,236 (claims 7-10) and 7,238,517 or corresponding
claims in their non-U.S. counterparts. No right is conveyed expressly, by implication or by estoppels under any
other patent of Applera, including but not limited to U.S. Patent No. 6,814,934 and its non-U.S. counterparts,
which describe and claim thermal cyclers capable or real-time detection.
3
SYSTEMÜBERBLICK
Die peqSTAR Thermocycler sind die idealen PCR-Geräte für alle molekularbiologischen und biochemischen
Labore, die im Bereich der Grundlagenforschung oder der Routinediagnostik tätig sind. Die sechs 250 °C HTR
(High Temperature Range) Peltier-Technologie gewährleistet schnelle Temperaturänderungen von bis zu 5 °C/sec
(max). Die individuelle Überwachung und Steuerung jedes einzelnen der 8 bzw. 16 mit 'Long-Life-Technologie'
ausgestatteten Peltierelemente führt zu einer überragenden Temperaturhomogenität innerhalb des Blocks und somit
zu höchster Reproduzierbarkeit der Ergebnisse. Zu Verhinderung von Kondensation ist jeder Block mit einer
Deckelheizung ausgerüstet (ausgenommen peqSTAR in situ X).
Die peqSTAR Thermocycler sind mit einem touch-sensitiven, grafischen Farb-TFT-Display der neuesten Generation
ausgestattet, welches die Bedienung des Geräts durch einfache Berührung von Symbolen und grafischen
Elementen auf dem Bildschirm erlaubt. Hierdurch gestalten sich die Handhabung und die Programmierung des
Geräts einfach und intuitiv. Zusätzlich kann über einen USB-Anschluss eine Maus angeschlossen werden.
Das Gerät verfügt weiterhin über eine Ethernet-Schnittstelle, welche die Einbindung des Geräts in ein MSWindows®- oder Linux-Netzwerk erlaubt. Somit bietet sich die Möglichkeit, einen externen Server als zusätzliches
PEQLAB_v1212D
17
Kurzanleitung peqSTAR Thermocycler Serie
Laufwerk des Thermocyclers zu verwenden, so dass Programme und GLP-Reports automatisch archiviert und
gesichert sind und global im Netzwerk (z.B. für weitere Thermocycler) zur Verfügung stehen.
Weiterhin ist es möglich, den Thermocycler über einen mittels Netzwerk oder drahtloser ZigBee-Verbindung
(Option) angeschlossenen PC zu bedienen. Die zu diesem Zweck mit dem Gerät zusätzlich mitgelieferte
Bediensoftware bietet am PC mit Ausnahme einiger Systemeinstellungen dieselbe Funktionalität wie die interne
Bediensoftware des Gerätes.
Die Verwendung eines handelsüblichen USB-Memory-Sticks ist ebenfalls möglich, um Daten zu transferieren. Somit
bietet sich ein nahezu unbegrenzter Speicherplatz für Programme und GLP-Reports.
Der Versuchsablauf kann mit Hilfe eines über das Netzwerk angeschlossenen Druckers dokumentiert werden.
Hinweis:
4
4.1
Die USB-Schnittstellen unterstützen nur Standard-Kabelmäuse und Standard-USB-Sticks.
Die maximal zulässige Länge der Verbindungskabel an den Schnittstellen ist jeweils 3 m!
INSTALLATION
Lieferumfang
1 Thermocycler
1 Netzkabel
1 Kurzanleitung
(Eine ausführliche Bedienungsanleitung ist auf unserer Homepage www.peqlab.com in Deutsch und Englisch zum Download verfügbar.)
1 CD PC-Software 'peqSTAR Manager'
Diese Lieferung ist nach Erhalt auf Vollständigkeit und auf eventuelle Transportschäden hin zu überprüfen.
Liegen Transportschäden vor, sind der Vertrieb oder der Hersteller unverzüglich zu informieren.
HINWEIS: FALLS SIE TRANSPORTSCHÄDEN AM GERÄT ENTDECKEN, DÜRFEN SIE DIESES KEINESFALLS AN
DAS STROMNETZ ANSCHLIESSEN UND IN BETRIEB NEHMEN!
4.2
Aufstellung des Thermocyclers
Prüfen Sie die Verpackung auf eventuelle Transportschäden. Entfernen Sie die Verpackung und stellen Sie den
Thermocycler auf eine feste Unterlage. Das Gerät sollte keinem direkten Sonnenlicht ausgesetzt sein. Es muss
ausreichend Platz vorhanden sein, so dass die Lüftungsschlitze an der Vorder- und Rückseite des Geräts nicht
verdeckt werden und für eine ausreichende Ventilation gesorgt ist. Das Gerät sollte mindestens 25 cm Abstand zur
Wand oder zum nächsten Gerät haben. Zwei Thermocycler sollten nicht direkt hintereinander stehen. Die
Raumtemperatur sollte zwischen 4 °C und 25 °C liegen.
Der Lieferkarton kann bei einem eventuellen Reparaturfall zum Einsenden des Geräts verwendet werden und sollte
daher aufbewahrt werden.
4.3
Inbetriebnahme des Thermocyclers
Überprüfen Sie vor der ersten Inbetriebnahme des Thermocyclers, ob der für das Gerät geeignete
Spannungsbereich zu Ihrer Stromversorgung passt. Schließen Sie den Thermocycler mit dem mitgelieferten
Netzkabel an das Stromnetz an. Die Gerätesteckdose und der Netzschalter befinden sich auf der Rückseite des
Geräts. Die Seriennummer befindet sich auf einem Typenschild auf der Unterseite des Gerätes.
Das Gerät darf nur durch autorisiertes Personal bedient werden. Um dies sicherzustellen, steht Ihnen hierfür im
Gerät eine Benutzerverwaltung mit Rechtevergabe für einzelne Benutzer zur Verfügung.
Näheres hierzu sowie die detaillierte Bedienungsanleitung finden Sie in den folgenden Kapiteln.
PEQLAB_v1212D
18
Kurzanleitung peqSTAR Thermocycler Serie
5
BEDIENUNG
Der Thermocycler kann wahlweise über den eingebauten Touchscreen, einer berührungsempfindlichen Oberfläche
auf dem Display, oder über eine an einem freien USB-Port angeschlossene PC-Maus bedient werden.
Werden vom Bediener Eingaben in einem Zahlen- oder Textfeld erwartet, so muss das entsprechende Feld über
den Touchscreen berührt oder mit der Maus angeklickt werden. Es erscheint je nach erwarteter Eingabe eine
geeignete Tastatur auf dem Bildschirm.
5.1
Startbildschirm = 'Run'
Das zuletzt gelaufene PCR-Programm wird im 'Run'-Reiter angezeigt und kann über den Button 'Start PCR' sofort
wieder gestartet werden.
Hinweis: Bei peqSTAR 2X Doppelblock-Thermocyclern kann über die Reiter 'L' (links) und 'R' (rechts) der
gewünschte Block ausgewählt werden. Bei den peqSTAR 96X, 384X und in situ X Geräten steht diese Auswahl
nicht zur Verfügung.
5.2
Starten eines gespeicherten Programms ('Run'):
1. Im 'Run'-Reiter den Auswahl-Dialog über den Button
öffnen.
2. Im Auswahldialog das gewünschte Programm markieren und über den Button 'Open' in den 'Run'-Reiter
laden.
3. Das Programm über den Button
PEQLAB_v1212D
starten.
19
Kurzanleitung peqSTAR Thermocycler Serie
5.3
Inkubationsmodus ('Run')
Über den Button
im 'Run'-Reiter kann das Gerät in den Inkubationsmodus versetzt werden. In diesem
Modus können die Temperaturen für Thermoplatte und Deckelheizung manuell eingestellt werden. Die Uhr zeigt
an, wie lange sich der Thermocycler bereits im Inkubationsmodus befindet.
5.4
Erstellen oder Editieren von PCR-Programmen ('Programs')
1. Zum Erstellen oder Editieren von PCR-Programmen wird der 'Programs'–Reiter verwendet.
2. Beim erstmaligen Anwählen des 'Programs'-Reiters erscheint automatisch der Auswahl-Dialog. Hier kann
entweder über 'New' ein neues Programm erstellt werden oder über 'Open' ein bereits vorhandenes
Programm zur Bearbeitung im Programm-Editor geöffnet werden.
3. Im Programm-Editor werden in der linken Spalte alle verfügbaren Befehle angezeigt, im rechten Fenster
können aus
werden die einzelnen Schritte des aktuellen Programms dargestellt. Mit dem Plus-Button
der Liste der verfügbaren Befehle neue Schritte in das Programm eingefügt werden.
können Parameter eines Befehls nachträglich editiert
4. Per Doppelklick bzw. über den Edit-Button
werden.
5. Markierte Programm-Schritte können über den Delete-Button
aus dem Programm entfernt werden.
6. Zum Speichern des Programms wird der Save-Button
benutzt.
7. Über den Save&Run-Button
können Programme in einem Schritt gespeichert und gestartet werden.
8. Um Programme noch schneller zu erstellen, befinden sich auf den peqSTAR Thermocyclern bereits
Musterprogramme, die angepasst und unter neuem Namen abgespeichert werden können.
Liste der verfügbaren Befehle
Einfügen und Entfernen von
Programmschritten
PEQLAB_v1212D
20
Programmname
Programm-Fenster
Kurzanleitung peqSTAR Thermocycler Serie
Verfügbare Befehle:
Lock Lid
Cycle
Verriegelt den Deckel, um zu vermeiden,
dass er versehentlich geöffnet wird.
Kombiniert Start Cycle, Close Cycle,
Temperature und Gradient.
Unlock Lid
Pause
Entriegelt den Deckel, sofern er vorher im
Programmablauf verriegelt wurde.
Hält Programm bei definierbarer
Temperatur vorübergehend an.
Heat Lid
Store
Schaltet die Deckelheizung an bzw. aus.
Hält definierbare Temperatur, schaltet
automatisch die Deckelheizung ab.
Temperature
Start Cycle
Temperaturschritt für definierbare Zeit;
Ramp-Rate, Temperatur- und ZeitIncr./Decr.-Funktion
Manueller Schleifenanfang
Gradient
Close Cycle
Gradientenschritt realisiert unterschiedliche
Temperaturen zur Protokolloptimierung
(siehe weitere Infos unten bei FlexGradient).
Manuelles Schleifenende
Programmbeispiel
Legen Sie mit 'New' ein neues Programm an. Geben Sie dem neuen Programmfile einen Namen, bestätigen Sie
mit OK und gelangen Sie in die Programmierebene.
01
Deckelheizung an, Temperatur 110 °C
Heat Lid
Der 'Heat Lid' Schritt ist standardmäßig als erster Programmschritt eingefügt. Der Schritt
kann nachträglich editiert bzw. gelöscht werden. Die Standardeinstellungen können unter
'System' => 'Firmware' => 'Settings' geändert werden.
Bei 'Store'-Schritten oder 'Temp'-Schritten < 30 °C wird die Deckelheizung automatisch
abgeschaltet.
Zum Einfügen neuer Programmschritte wird der gewünschte Befehl aus der Liste der verfügbaren Befehle markiert
und mit dem '+'-Button in das Programmfenster eingefügt. Die zugehören Parameter werden automatisch
abgefragt.
02
Denaturierung bei 95 °C für 2 Minuten
Temp
03
07
Cycle
04
Öffnen der Cycle-Funktion
→ Anzahl der Zyklen – 30 Zyklen
→ 3 x 'Temp'-Schritt
→ Cycle-Funktion beenden mit 'Close Cycle'
Denaturierung bei 95 °C für 30 Sekunden
Temp
PEQLAB_v1212D
21
Kurzanleitung peqSTAR Thermocycler Serie
Annealing-Temp. Von 65°C im 1. Zyklus für 30 Sekunden
05
Temp
'Touchdown'-PCR, durch Eingabe eines Temperatur-'Decrements'. Die Annealing-Temp. von
65 °C im 1. Zyklus verringert sich um 0.4 °C pro Zyklus. Dies bewirkt eine hohe Spezifität
in den anfänglichen Zyklen und eine hohe Amplifikationseffizienz in den späteren Zyklen.
Die Verringerung der Kühlrate (Ramp Rate) kann sowohl in Temperatur- also auch in
Gradienten-Schritten eingestellt werden. Sinnvoll ist sie z. B. für das quantitatives Annealing
komplementärer Oligonukleotide oder bei den Annealing-Schritten der anfänglichen Zyklen,
wenn aufgrund geringer Template-Mengen (z. B. nach cDNA-Synthese) die Primer-Bindung
besonders effizient ablaufen soll.
06
Elongation bei 72 °C für 1 Minute im 1. Zyklus
Temp
Durch Eingabe eines Zeit-'Increments' verlängert sich die Elongationszeit um 5 s pro Zyklus.
Die schrittweise Verlängerung der Elongationszeit ist bei hohen Zyklenzahlen (>30) sinnvoll,
wenn die Aktivität der Polymerase nachlässt.
Cycle-Funktion beenden mit 'Close Cycle'.
07
Alternative Eingabe der Schritte 03 – 07:
03
Öffnen einer Programmschleife – 30 Zyklen
Start Cycle
04
Denaturierung bei 95 °C für 30 Sekunden
Temp
05
Annealing-Temp. Von 65°C im 1. Zyklus für 30 Sekunden
Temp
'Touchdown'-PCR, durch Eingabe eines Temperatur-'Decrements'. Die Annealing-Temp. von
65 °C im 1. Zyklus verringert sich um 0.4 °C pro Zyklus. Dies bewirkt eine hohe Spezifität
in den anfänglichen Zyklen und eine hohe Amplifikationseffizienz in den späteren Zyklen.
Die Verringerung der Kühlrate (Ramp Rate) kann sowohl in Temperatur- also auch in
Gradienten-Schritten eingestellt werden. Sinnvoll ist sie z. B. für das quantitatives Annealing
komplementärer Oligonukleotide oder bei den Annealing-Schritten der anfänglichen Zyklen,
wenn aufgrund geringer Template-Mengen (z. B. nach cDNA-Synthese) die Primer-Bindung
besonders effizient ablaufen soll.
06
Elongation bei 72 °C für 1 Minute im 1. Zyklus
Temp
07
Durch Eingabe eines Zeit-'Increments' verlängert sich die Elongationszeit um 5 s pro Zyklus.
Die schrittweise Verlängerung der Elongationszeit ist bei hohen Zyklenzahlen (>30) sinnvoll,
wenn die Aktivität der Polymerase nachlässt.
Schließen der Programmschleife mit 'Close Loop'.
Close Loop
08
Finale Elongation bei 72 °C für 5 Minuten
Die finale Elongation stellt die vollständige Synthese aller Amplifikate sicher.
Temp
09
8 °C bei abgeschalteter Deckelheizung für unbegrenzte Zeit
Store
PEQLAB_v1212D
Im 'Store'-Schritt wird die Deckelheizung automatisch abgeschaltet. Dies verhindert ein
Gegeneinanderarbeiten von Deckelheizung und den Peltierelementen des Blocks, was vor
allem bei längeren Inkubationen bei niedrigen Temperaturen die Lebensdauer der
Peltierelemente erhöht und den Energieverbrauch verringert.
22
Kurzanleitung peqSTAR Thermocycler Serie
FlexGradient Technologie™:
Die Temperatur jeder der 8 Reihen des Blocks wird individuell geregelt! Erstmals können Sie wählen: Zwischen
dem perfekt linearen Temperaturgradienten − ideal für PCR-Optimierungen, oder angepassten Temperaturzonen −
ideal bei Verwendung unterschiedlicher Primer-Paare pro Lauf.
Der Temperaturbereich wird zunächst über die Felder
'Center' und 'Gradient ±' definiert. Der
Standardgradient wird über die 8 Cap-Reihen von
oben nach unten realisiert. Beim peqSTAR 96X gibt es
darüber hinaus die Möglichkeit einen 16er-Gradienten
zu realisieren, bei dem auf der linken und rechten
Blockhälfte unterschiedliche Gradiententemperaturen
realisiert werden. In diesem Fall werden die in der
Software abgebildeten Temperaturen nur in den Reihen
3 bzw. 10 ausgebildet.
Tippt man auf einen Temperaturbutton
, stehen
verschiedene Optionen zur Verfügung: Wurde vorher
über
der lineare Gradient inaktiviert, können
über 'Adjust Temperature' in einem vorgegebenen
Rahmen die Temperaturen der Reihen angepasst
werden.
Unabhängig von der Linearitätseinstellung kann man
die gewählte Temperatur als neue Centertemperatur
verwenden oder den Gradientenschritt in einen
Temperaturschritt mit dem angegebenen Wert
umwandeln.
Programm-Optionen:
Beim Erstellen eines neuen Programms, beim Speichern oder während der Programmierung über den Info-Button
können die Programm-Optionen eingestellt werden. Dem Anwender stehen folgende Optionen zur Verfügung:
1. Power Fail Denaturation:
Ereignet sich während eines Programmlaufs ein Stromausfall, so arbeitet der Thermocycler nach dem Wiedereinschalten des
Stroms immer an der Stelle weiter, an welcher er unterbrochen wurde. Ist die Option ‚Power Fail Denaturation’ aktiviert, so
heizt der Thermoblock vor dem Fortsetzen des Programmablaufs für die eingestellte Zeit auf die angegebene Temperatur.
2. Global Program Ramp:
Hier kann eine globale Heiz- und Kühlrate in °C/s eingestellt werden. Diese wird bei allen Temperatur- oder
Gradientenschritten des Programms verwendet, kann aber über die optionale Angabe eines Ramp-Wertes in einem
Programmschritt nachträglich überschrieben werden.
3. Gradient Control
Hier kann eingestellt werden, ob bei einem Gradientenschritt die Plateauzeiten (Plateaus synchronized) oder die Ramp-Raten
(Ramps synchronized) für jede Reihe identisch sein sollen.
4. Tube Control
Hier kann eingestellt werden, ob die Temperaturregelung in einem Temperaturschritt unter Beachtung der Probenmenge
erfolgen soll. Ggf. fügt die Temperaturregelung bewusst ein von der Probenmenge abhängiges kurzzeitiges Über- oder
Unterschwingen der Temperatur in den Temperaturverlauf ein. Speziell in Kombination mit dem Emulationsmodus kann dies
erforderlich sein, wenn bei dem zu emulierenden Cycler vorher auch mit der Tube Control Option gearbeitet wurde.
5. Emulation
Der Thermocycler kann über diese Funktion veranlasst werden, einen Thermocycler anderen Typs zu emulieren. Dies kann
nützlich sein, wenn ein Programm von einem Fremdfabrikat auf einen peqSTAR übertragen werden soll.
PEQLAB_v1212D
23
Kurzanleitung peqSTAR Thermocycler Serie
5.5
Weitere Reiter: Diagnostics, GLPs, System
Diagnostics: Überwachen der Temperaturprofile der Thermoplatte während eines Laufs. Bei Verdacht auf eine
Gerätefehlfunktion können Sie hier den Temperaturverlauf der Peltierelemente beobachten.
GLPs: Verwalten und Anzeigen der vom Thermocycler erzeugten GLP (Good Laboratory Practice)-Reports. Diese
Option ist im Wesentlichen für Labore, die zu Dokumentationszwecken GLP-Reports speichern müssen.
System: Konfiguration der Gerätes und Einstellen allgemeiner Geräteparameter
PEQLAB_v1212D
24
Kurzanleitung peqSTAR Thermocycler Serie
Elemente des System-Reiters:
Clock
Hide/Show Curser
Einstellen des aktuellen Datums und Uhrzeit
Aus-/Einblenden des Maus-Cursors
User Call
User Management
Einstellen der akustischen Meldungen oder
Email-Benachrichtigung (Pause, Store,
Program End)
Definieren von Benutzern und Benutzergruppen (und deren Zugriffsrechte)
Network Settings
Netzwerkeinstellungen zur Netzwerkintegration des peqSTARS
5.6
Das werkseitige Passwort für den Benutzer 'Administrator' ist:
'admin'. Über den Button 'Logout' kann sich der aktuelle Benutzer
'admin'
abmelden. Dadurch wird automatischer der Benutzer 'Guest'
angemeldet, der werkseitig nur über minimale Benutzerrechte
verfügt. Der Administrator kann diese Einstellungen ändern.
Hardware
Firmware
Anzeige der Hardware-Versionen für
Control Unit, Power Unit und Thermoblock
Anzeige, Verwaltung und Update der Software (Control Unit & Power Unit)
Printer
Display Settings
Eintragen eines Druckerpfads für einen
vorhandenen Netzwerkdrucker.
Einstellungen der Bildschirmanzeige
verändern.
Protocol
Service
Ausgabe der interne Kommunikation des
Cyclers zur Fehlerdiagnose
Service-Funktionen (nur für den
authorisierten, technischen Service)
LAN Control
USB Recovery
Funktionsbutton für die PC- bzw.
Master/Slave-Steuerung
Manuelles Mounten von nicht automatisch
erkannten USB-Sticks
Noch Fragen…?
:
Hilfemodus
Wird dieser Button gedrückt, schaltet das Gerät in den Hilfemodus (der Button erscheint dann gedrückt). Durch
Berühren eines Funktionsbuttons oder eines Dialogelements wird der dazugehörige Hilfedialog angezeigt.
Ein ausführlicheres Manual ist auf unserer Homepage www.peqlab.com in Deutsch und Englisch zum Download
verfügbar.
PEQLAB_v1212D
25
Kurzanleitung peqSTAR Thermocycler Serie
6
TECHNISCHER SERVICE UND BESTELLINFORMATIONEN
Bei technischen Fragen kontaktieren Sie uns bitte unter +49 (0)9131 610 7020 oder per E-Mail an
[email protected] Ausführliche Informationen zu unseren Produkten finden Sie in unserem aktuellen Produktkatalog,
den wir Ihnen auf Wunsch gerne zusenden, oder unter www.peqlab.com.
Diese Kurzanleitung erhalten Sie mit folgenden Produkten:
Produkt
peqSTAR 2X Universal
peqSTAR 2X Universal
Gradient
peqSTAR 96X Universal
peqSTAR 96X Universal
Gradient
peqSTAR 96X HPL
peqSTAR 96X HPL
Gradient
peqSTAR 384X HPL
peqSTAR in situ X
PEQLAB_v1212D
Cat. No.
Mit 2 x 48 Well Universalblöcken mit Standarddeckeln
für je 48 x 0.2 ml Tubes, 48 Well PCR-Platten oder
24 x 0.5 ml Tubes mit flachem Deckel
Mit 2 x 48 Well Universal-Gradientenblöcken und
Standarddeckeln für je 48 x 0.2 ml Tubes, 48 Well PCRPlatten oder 24 x 0.5 ml Tubes mit flachem Deckel
mit 96 Well Universalblock und Standarddeckel
für 96 x 0.2 ml Tubes, 96 Well PCR-Platten oder
48 x 0.5 ml Tubes mit flachem Deckel
mit 96 Well Universal-Gradientenblock und Standarddeckel
für 96 x 0.2 ml Tubes, 96 Well PCR-Platten oder
48 x 0.5 ml Tubes mit flachem Deckel
mit 96 Well Block und High Pressure Lid
für 96 x 0.2 ml Tubes oder 96 Well PCR-Platten
mit 96 Well Gradientenblock und High Pressure Lid
für 96 x 0.2 ml Tubes oder 96 Well PCR-Platten
mit 384 Well Block und High Pressure Lid
für 384 Well PCR-Platten
mit in situ Block für bis zu vier Objektträger
und integriertem Pufferreservoir
26
95-07002
95-08002
95-05002
95-06002
95-05002-HPL
95-06002-HPL
95-050384-HPL
95-050FLAT
Kurzanleitung peqSTAR Thermocycler Serie
7
TECHNISCHE DATEN
Allgemeine Merkmale
2 x 8 bzw. 16 Peltier-Elemente mit Long-Life-Technologie und 2 x 8 bzw. 16 Regelkreise mit Pt 1000Temperatursensoren
Maximale Heiz- und Kühlrate: 5 °C/s
Blockuniformität (bei 72 °C): ± 0,20 °C
Temperaturbereich Thermoblock: 4 bis 105 °C
Regelgenauigkeit Thermoblock: ± 0,1 °C
Variables Ramping: 0.1 bis 3.0 °C/s
Increment/Decrement Zeit: 0:01 bis 9:59 Minuten
Increment/Decrement Temperatur: 0.1 bis 9.9 °C
Temperaturbereich Deckelheizung: 40 bis 120 °C
Programmierbare Deckelverriegelung
4 x USB, 1 x Netzwerkanschluss (MS Windows® oder Linux), maximale Länge der Verbindungskabel jeweils
3 m! Hinweis: Die USB-Schnittstellen unterstützen nur Standard-USB-Sticks, -Kabelmäuse und -Tastaturen!
Maße (B x H x T): 30 x 28 x 38 cm
Gewicht inkl. Block: 13.3 kg
Stromversorgung: 90 – 240 V AC, 50 / 60 Hz, 850 VA
Feinsicherung: 10 A träge
Verschmutzungsgrad: 2
Umgebungstemperatur: 10 °C bis 30 °C
Max. relative Luftfeuchtigkeit: 70 %
Max. Höhe für den Einsatz des Gerätes: Bis 2000 m ü. NN
Verwendung nur in Innenräumen!
Geräuschpegel: < 37 dB (A)
Optionale Gradientenfunktion
Maximaler Gradient über 8 bzw. 16 Reihen: 30 °C (± 15 °C)
Temperaturbereich Gradient: 35 bis 105 °C
Gradientengenauigkeit: ± 0,1 °C
Gradienten-Upgrade auch nachträglich und vor Ort möglich
FlexGradient™-Technologie: Idealer, linearer Gradient oder angepasste Temperaturzonen
'Gradient Control': synchronisierte Plateauzeiten oder Ramp-Raten
Bedienung/Funktionalität
Touchsensitives TFT-Display (VGA, Grafik, 65535 Farben), Bedienung über USB-Mouse möglich
Direkthilfe-Funktion, Tm- und Elongationszeit-Kalkulator
Interner Speicher für 500.000 typische PCR-Programme in frei erstellbaren Ordnern/Unterordnern
Beliebig viele Programme über Netzwerk-Rechner oder USB-Memory-Stick
PC-Software für die Fernsteuerung/Überwachung von Läufen und das Erstellen von PCR-Programmen am PC
Unbegrenzte Anzahl von Schritten/Programm
Passwortgeschützte User Accounts mit variablen Zugriffsrechten
GLP-Reports für lückenlose Aufzeichnung aller Läufe
Quickstart-Funktion des zuletzt gelaufenen Programmes
Auto-Restart nach Stromausfall, Power-Fail-Denaturation, Instant Incubation
Fernsteuerung und Überwachung von Geräten mittels PC-Software
Frei wählbare MP3-Signaltöne
Benutzerrufe per E-Mail
Master/Slave-Steuerung (optional auch kabellos)
PEQLAB_v1212D
27
Kurzanleitung peqSTAR Thermocycler Serie
Blockvarianten (z.T. als Wechselblöcke verfügbar)
peqSTAR 2X/2X Gradient:
Zwei 48 Well Universalblöcke mit jeweils: Heizdeckel mit automatischer Höhenanpassung für 48 x 0.2 ml Tubes,
48-Well PCR-Platten oder 24 x 0.5 ml Tubes mit flachem Deckel
peqSTAR 96X/96X Gradient:
96 Well Universalblock mit Heizdeckel mit automatischer Höhenanpassung für 96 x 0.2 ml Tubes, 96 Well PCRPlatten oder 48 x 0.5 ml Tubes mit flachem Deckel.
peqSTAR 96X HPL/96X HPL Gradient:
96 Well Block mit High Pressure Lid (HPL, 100 – 250 N) für das sichere Verschließen von 96 Well PCR-Platten
peqSTAR 384X HPL:
384 Well Block mit High Pressure Lid (HPL, 100 – 250 N) für das sichere Verschließen von 384 Well PCR-Platten
peqSTAR In situ X:
In situ Block mit integriertem Pufferreservoir für bis zu 4 Objektträger
PEQLAB_v1212D
28
Kurzanleitung peqSTAR Thermocycler Serie
8
8.1
WARTUNG & REPARATUR DES THERMOCYCLERS
Reinigung, Wartung und Reparatur des Thermocyclers
Aus Sicherheitsgründen ist vor der Reinigung des Thermocyclers unbedingt das Gerät auszuschalten und der
Netzstecker zu ziehen!
Das Gehäuse des Gerätes kann mit einem leicht feuchten Tuch oder Ethanol gereinigt werden. Zur Reinigung keine
aggressiven oder scheuernde Reinigungsmittel oder organische Lösungsmittel verwenden. Das Gerät muss vor
aggressiven Chemikalien geschützt werden. Es ist darauf zu achten, dass keine Flüssigkeiten in das Innere des
Gerätes laufen. Sollten Probenflüssigkeit in die Bohrungen des Thermocyclers gelangen, sollten diese sofort mit
milder Seifenlauge und anschließend mit destilliertem bzw. demineralisiertem Wasser gereinigt werden. Milde
Desinfektionsmittel können zur Dekontamination verwendet werden.
Die Wartung (Prüfung des Übertemperaturschutzes und externe Temperaturüberprüfung, siehe 8.2 und 8.3) sollte
zur Detektion eventueller Defekte regelmäßig durchgeführt werden.
Die elektrischen Schmelzsicherungen befinden sich zwischen Netzschalter und Netzstecker auf der
Geräterückseite. Durch Abziehen des Netzsteckers und anschließendem Aufwärtsschieben einer Sperrklinke
können sie entnommen und gewechselt werden. Es dürfen nur Sicherungen mit korrekten Stromwerten verwendet
werden (ein Schild mit Angaben über den Sicherungstyp befindet sich auf der Geräterückseite).
Das Gerät darf nur von einem qualifizierten Fachmann geöffnet werden. Die Gerätegarantie erlischt mit einem
unautorisierten Eingriff in das Gerät. Sämtliche Reparaturen dürfen nur von autorisiertem Personal durchgeführt
werden, es dürfen hierzu nur Original-Ersatzteile verwendet werden.
Der Austausch einzelner Gerätekomponenten soll ebenfalls nicht durch den Anwender, sondern ausschließlich
durch autorisiertes Fachpersonal erfolgen. Dazu ist eine Einsendung des Gerätes an den Hersteller notwendig.
Wird ein Gerät zur Reparatur an den Hersteller geschickt, so sollte es vor dem Versand dekontaminiert werden.
Dazu sollte das Gehäuse mit einem mit 70 %igem Ethanol versetzten fusselfreien Tuch abgewischt werden. Die
Bohrungen des PCR-Blocks sollten mit Hilfe eines Wattestäbchens mit einer 1,5 %igen Hypochloritlösung gereinigt
werden. Dem Gerät ist beim Versand an den Hersteller die ausgefüllte und unterschriebene
Dekontaminationsbescheinigung beizulegen.
8.2
Prüfung des Übertemperaturschutzes
Eine regelmäßige Überprüfung des Übertemperaturschutzes (→ System → Hardware → Thermo Block →
Overtemperature Test) wird empfohlen.
8.3
Temperaturüberprüfung
Eine regelmäßige Überprüfung (z. B. jährlich) der Temperatur wird empfohlen. Das dazu verwendete Thermometer
sollte über einen Draht-Einstechfühler (z.B. Pt100) verfügen und im Bereich von 0 bis 100 °C eine Messgenauigkeit
von 1/10 Klasse B (∆T = ± 0,03 °C) aufweisen. Bei der Durchführung der Temperaturkontrolle darf die
Raumtemperatur maximal 25 °C betragen.
8.4
Kalibrierung
Besteht der begründete Verdacht, dass das Gerät dekalibriert ist, weil z.B. die Abweichungen bei der
Temperaturüberprüfung deutlich außerhalb der zulässigen Toleranzen liegen, muss eine erneute Kalibrierung
durch den Hersteller erfolgen. Hierzu ist der Vertriebspartner zu kontaktieren.
PEQLAB_v1212D
29
Kurzanleitung peqSTAR Thermocycler Serie
9
GEWÄHRLEISTUNG
Die Gewährleistungsfrist für das vorliegende Produkt beträgt 24 Monate. Die Frist beginnt mit dem Kaufdatum. Die
Gewährleistung gilt nicht für Mängel, die durch nicht bestimmungsgemäße, missbräuchliche oder unsachgemäße
Verwendung entstanden sind. Der Kunde hat das Gerät unverzüglich nach Erhalt auf augenscheinliche Mängel zu
überprüfen und diese schriftlich anzuzeigen. Anderenfalls besteht kein Anspruch auf Gewährleistung. Der
Hersteller behält sich vor, an den Thermocyclern oder der technischen Dokumentation der Thermocycler im
Rahmen der kontinuierlichen Produktentwicklung und -verbesserung ohne vorherige Ankündigung Änderungen
vorzunehmen.
10 ENTSORGUNGSHINWEIS
Elektrische Geräte, die mit dem Symbol der durchgekreuzten Mülltonne gekennzeichnet sind,
dürfen EU-weit nicht mit dem normalen, unsortierten Hausmüll entsorgt werden.
Bitte kontaktieren Sie die PEQLAB Biotechnologie GmbH, wenn Sie einen entsprechend
gekennzeichneten peqSTAR-Thermocycler entsorgen wollen.
Altgeräte sind kein wertloser Abfall. Durch umweltgerechte Entsorgung können wertvolle
Rohstoffe wiedergewonnen werden. Sie leisten dadurch einen Beitrag zum Schutze unserer
Umwelt und der menschlichen Gesundheit.
NOTIZEN
PEQLAB_v1212D
30
D
AT
UK
USA
PEQLAB Biotechnologie GmbH, 91052 Erlangen, Freecall (D): 0800 100 20 16, [email protected], www.peqlab.de
PEQLAB Biotechnologie GmbH, 6404 Polling, Tel: +43 (0) 5238 84 169, [email protected], www.peqlab.at
PEQLAB Ltd., Southampton SO31 7ZN, Freephone (UK): 0808 202 1302, [email protected], www.peqlab.co.uk
PEQLAB LLC, Wilmington, DE 19810, Toll-Free (US): 877 737 5220, [email protected], www.peqlab.us
Creating the future together.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement