Gossen MetraWatt METRALINE EXM25 Owner Manual
Gossen MetraWatt METRALINE EXM25 is an intrinsically safe multimeter for Ex areas according to ATEX (EN 60079-0 and EN 60079-11) and EN/IEC 61010 for voltage, resistance, current and frequency measurements as well as with accessories for temperature and high current measurements.
It features an automatic voltage range that sets itself to the used voltage measuring tip EXM 72010. The measuring range serves for fast detection of measured values from 0 - 1000 V without showing decimal places. A resistance measurement function is available for determining impedances from 0.1 Ω - 20 MΩ.
Advertisement
Advertisement
GOSSEN METRAWATT
Bedienungsanleitung 3-349-729-15
Operating Instructions 1/4.14
METRALINE EXM25
GMC-I Messtechnik GmbH
Südwestpark 15
D-90449 Nürnberg
Telefon +49 911 8602-111
Telefax +49 911 8602-777
E-Mail: [email protected]
www.gossenmetrawatt.com
10
11
6
5
4
3
7 8 9
1
2
12
1 Prüfelektroden der Spannungs-Messspitzen
(rote Leitung + /schwarze Leitung -)
2 DATA-HOLD-Taster
3 LED Ω (grün): 0 ... 10 kΩ
4 LED Volt (rot): 12 ... 1000 V
5 LCD-Anzeige
6 -Buchse, schwarze Normbuchse
7 7-polige Universalbuchse für Spannungs-
Messspitzen, Temperatur-Messzusatz,
Stromzange und Steckernetzgerät
8 Ω-Buchse, rote Normbuchse für Widerstandsmessungen
9 A-Buchse, blaue Normbuchse für Strommessungen
10 Ein/Aus-Taste
11 Funktionstaste
12 Handhabe
2
Symbole auf dem Gerät
Achtung! Bedienungsanleitung beachten!
Ex-Kennzeichnung:
Zugelassen für Ex-Bereiche nach ATEX
(EN 60079-0 und EN 60079-11) siehe Kapitel 5.1
EG-Konformitätskennzeichnung
1. Anwendung
METRALINE EXM25 ist ein eigensicheres Multimeter für Ex-Bereiche nach ATEX (EN 60079-0 und
EN 60079-11) und EN/IEC 61010 für Spannungs-,
Widerstands-, Strom- und Frequenzmessungen sowie mit Zubehör für Temperatur- und Hochstrommessungen.
Ausführung M210A mit Akku und M210B mit
Batteriebetrieb.
1.1 Bestimmungsgemäße Verwendung
Das Gerät ist nur für die in der Bedienungsanleitung beschriebenen Anwendungen bestimmt. Hierzu sind besonders die Sicherheitshinweise und die technischen Daten mit den Umgebungsbedingungen zu beachten.
Eine andere Verwendung ist unzulässig und kann zu
Unfällen oder Zerstörung des Gerätes führen.
Diese führen zu einem sofortigen Erlöschen jeglicher Garantie- und Gewährleistungsansprüche.
2. Sicherheitshinweise
Bei bestimmungsgemäßer Verwendung gewährleistet METRALINE EXM25 sowohl die Sicherheit der bedienenden Person als auch die des Gerätes.
Die Spannungs-Messspitzen enthalten in beiden
Prüfspitzen vergossene Vorwiderstände. Sie sind extrem sicher und entsprechen der Überspannungskategorie CAT IV.
Um den sicherheitstechnisch einwandfreien
Zustand zu erhalten und die gefahrlose Verwendung sicherzustellen, ist es unerlässlich, dass Sie vor dem
Einsatz Ihres Gerätes diese Bedienungsanleitung sorgfältig und vollständig lesen und diese in allen
Punkten befolgen.
Das Gerät darf nur innerhalb der gekennzeich-
neten Ex-Zonen (siehe Abschnitt 5.1) und
innerhalb der technischen Grenzwerte (siehe
Abschnitt 5.2) betrieben werden.
Vor Widerstandsmessungen ist die Spannungs-
freiheit des Messobjektes sicherzustellen.
Spannungsmessungen dürfen bis zu
folgenden Grenzwerten durchgeführt werden:
Mit der Messspitze EXM 72010 bis 1000 V.
Eine einwandfreie Anzeige ist nur im Tempera-
turbereich von - 10° C ... + 40° C sichergestellt.
nur an den Handhaben anfassen und die
Prüfelektroden nicht berühren.
3
Nur sachkundige Personen dürfen Arbeiten mit
diesem Produkt duchführen. Der Benutzer muss
mit den Gefahren der Messung und den Einhal-
tungen der Vorsichtsmassnahmen und dem
ordnungsgemäßen Gebrauch des Multimeters
vertraut sein.
Arbeiten dürfen nur mit entsprechender Schutz-
ausrüstung durchgeführt werden.
Beachten Sie die Mindestabstände zu anderen
unter Spannung stehenden oder geerdeten
Anlagenteilen und verwenden Sie persönliche
Schutzausrüstung gemäß den landesspezifischen
Unfallverhütungsvorschriften
(in Deutschland: BGV A3 oder VDE 0105-100).
Das Multimeter kurz vor und nach Möglichkeit
auch nach dem Einsatz auf Funktion geprüft
werden. Führen Sie den Funktionstest durch und
überprüfen Sie das Gerät an einer bekannten
Spannungsquelle (AC und DC). Fällt hierbei die
Anzeige eines oder mehrerer Systeme aus, darf
das Gerät nicht mehr verwendet werden.
Unbefugte dürfen das Messgerät nicht
zerlegen.
Eine Instandsetzung ist nur durch den Hersteller
oder durch vom Hersteller ausdrücklich ermäch-
tigte Werkstätten zulässig (siehe Abschnitt 7).
Gehäuse und die Messleitung auf Beschädi-
gungen zu überprüfen. Wenn Beschädigungen
zu erkennen sind, darf das Multimeter nicht mehr
verwendet werden. Bei starker Verschmutzung
ist das Multimeter vor der Verwendung zu
säubern.
und sauberer Umgebung erfolgen.
4
3. Inbetriebnahme
3.1 Batterie
In Ihr Gerät haben wir bereits einen Energieblock eingesetzt. Es dürfen ausschließlich folgende
Energieblocks verwendet werden:
NiMH-Akku für Gruppe I und II T4:
Typ Z209A EXM-AK9
Lithium-Batterie für Gruppe I und II T6:
Typ Z209B EXM-LB3
Achtung!
Beachten Sie vor der ersten Inbetriebnahme oder nach Lagerung Ihres Gerätes unbedingt den
Abschnitt 6.
3.2 Prüfen von Anzeige und Funktion
METRALINE EXM25 ist an jedem Tag auf einwandfreie Funktion zu überprüfen.
Eigentest 1 (Anzeigetest):
Drücken Sie die Taste „EIN/AUS“ und halten diese gedrückt. Auf dem LC-Display leuchten alle
Segmente, zusätzlich leuchten die V – LED und die
Ω - LED. Wenn Sie die Taste „EIN/AUS“ loslassen, wird der Wert 000 ... 001 V angezeigt.
Eigentest 2 (Spannungsmessleitung):
Spannungs-Messspitzen EXM 72010 an die
7-polige Universalbuchse anschließen und verriegeln. Gerät einschalten.
Anzeige: 000..001 V, in der oberen Zeile 1000 V.
Beide Prüfelektroden nacheinander etwa 5 mm schräg in die Ω-Buchse stecken.
Display zeigt dabei jeweils Test und rdy an, und die grüne LED leuchtet.
Hinweis: Wenn das Display weiterhin 000 V anzeigt, ist die Messspitze defekt, bitte austauschen.
Eigentest 3 (Spannungsanzeige):
Überprüfen Sie die Funktion an einer bekannten
Spannungsquelle, z.B. 230 V Steckdose.
Eigentest 4 (Widerstandsmessung):
Norm-Prüfleitungen in die Ω- und - Buchse stecken. Gerät auf Ω einstellen, Anzeige: OL kΩ,
Prüfspitzen zusammenhalten, Anzeige muss etwa
0.0 Ω anzeigen und die grüne LED muss leuchten.
Achtung!
Fällt bei der Eigenüberprüfung eine Anzeige auch nur teilweise aus oder wird keine Funktionsbereitschaft angezeigt, darf METRALINE EXM25 nicht mehr verwendet werden.
5
4. Messen und Prüfen
4.1. Allgemeine Hinweise
Batteriestatus (Kapitel 6)
Vorzeichen
Obere Zeile zeigt aktuell gewählten Messbereich.
Blinkt diese Anzeige muss ein spezieller Zusatz z.B. TMZ oder MZ angeschlossen werden.
Untere Zeile zeigt den aktuellen Messwert
Um die Batterie zu schonen, schaltet sich das Gerät etwa 60 s nach der letzten Messung automatisch aus. Im Display erscheint die Meldung „off“
(Ausnahme: siehe Frequenzmessung Abschnitt 4.6).
Die Hintergrund-Beleuchtung schaltet sich ab, falls kein Messwert anliegt oder die Batterie zu schwach ist.
Mit der Ein/Aus Taste können Sie
METRALINE EXM25 auf Spannung einschalten, den Anzeigetest aufrufen und ausschalten.
►
Mit der Funktions-Taste können Sie den gewünschten Messbereich wie folgt durchschalten.
Funktionen Anzeige obere Zeile
Spannung mit Spitze EXM 72010
Spannung mV und V Automatik
1000 V
AuTo
Widerstand Ω
Strom I
Strom mit Messzange I blinkt*
Frequenz FrE
Temperatur mit TMZ T blinkt*
* Messbereiche die durch Blinken signalisiert werden, benötigen Sonderzubehör wie z.B. die
Strommesszange oder den Temperatur-Messzusatz.
6
4.2 Spannung und Polarität prüfen
Achtung!
Spannungs-Messspitzen immer im ausgeschalteten
Zustand an die 7-polige Universalbuchse anschlie-
ßen und verriegeln. Danach das Gerät einschalten und ggf. Bereich wählen.
Setzen Sie die beiden Prüfspitzen mit sicherem
Kontakt auf die Messstellen auf. Der Spannungswert wird in der unteren Displayanzeige angezeigt.
Bei einer Spannung über 12 V blinkt die rote LED und die Tasterfunktionen sind gesperrt.
Hinweise:
- Sollte dauerhaft Hold angezeigt werden, wurden
die Messspitzen im eingeschaltetem Zustand
aufgesteckt. Gerät aus- und einschalten.
- Das Gerät zeigt im Leerlauf wegen der hohen
integrierten Vorwiderstände einige mV an, die das
Messergebnis nicht beeinflussen.
- Das EXM25-Gehäuse besteht aus einem leitfähigem
Werkstoff. Für genaue Spannungsmessungen
>100 Hz ist ein erdfreier Betrieb zu empfehlen.
Der Kontakt mit geerdeten Gegenständen, z.B.
Schaltschränken ist zu vermeiden. Hierfür ist die
Verwendung in einer Ledertasche zu bevorzugen.
1000 V Spannungsbereich (1000 V)
Der Spannungsbereich stellt sich automatisch beim
Einschalten des EXM25 auf die verwendete
Spannungs-Messspitze EXM 72010 ein.
Der Messbereich dient zur schnellen Erfassung von
Messwerten von 0 – 1000 V ohne dabei Komma- stellen anzuzeigen.
Der Anzeige-Bereich des METRALINE EXM25 geht bis 1160 V AC / 1610 V DC mit der Messspitze
EXM 72010.
Beachten Sie die Sicherheitshinweise für Ex-Bereiche
(siehe Abschnitt 2).
Auto-Spannungsbereich (AuTo)
Der Spannungsbereich stellt sich automatisch auf die verwendete Spannungs-Messspitze EXM 72010 ein.
Der Messbereich dient zur genauen Erfassung von
Messwerten von 1 mV – 1000 V wobei der optimale
Messbereich automatisch gewählt wird.
Polarität anzeigen
Die Art der Spannung wird durch die Symbole ~ und – dargestellt. Bei Gleichspannung: Liegt Minus an der Prüfspitze mit Hold-Taster an, so erscheint das
„-“ Vorzeichen. Liegt dort Plus an, so erscheint kein
Vorzeichen vor dem angezeigten Wert.
7
HOLD Funktion
Durch Betätigen der Taste „HOLD“ wird der maximale Spannungswert auf dem Display „gespeichert“.
Der Wert wird für ca. 30 s oder bis Sie die Taste
„HOLD“ erneut drücken angezeigt. Durch erneutes
Anlegen an eine Spannung wird die Hold-Funktion automatisch beendet.
Hinweis: Wenn sich der Messwert für ca. 2 s nicht mehr ändert wurde der Maximalwert erfasst.
4.3 Widerstandsmessung (Ω)
Vor Widerstandsmessungen grundsätzlich prüfen, ob das Messobjekt spannungsfrei ist.
4 mm-Norm-Prüfleitungen in die Ω- und in die
-Buchse stecken. Gerät auf Ω einstellen. In der oberen Displayzeile wird Ω angezeigt.
Hinweis: Der Pfeil meldet „außerhalb des
Messbereiches“. Die Messbereiche dienen zur Bestimmung von Impedanzen von 0,1 Ω – 20 MΩ. Die
Wahl zwischen Ω-, kΩ- und MΩ-Bereichen erfolgt nach Anlegen an die Impedanz automatisch. Die grüne LED signalisiert Widerstandswerte kleiner
10 kΩ.
4.4 Strommessung (I)
Achtung! In Ex-Bereichen dürfen Strommessungen nur in Messkreisen mit einem Scheitelwert von max.
50 V durchgeführt werden.
Es dürfen Ströme bis max. 2 A gemessen werden.
4 mm-Norm-Prüfleitungen in die A- und in die
-Buchse stecken. Der Messbereich dient zum Messen von AC/DC-Strömen im Bereich von 1 mA – 2 A.
Die Wahl zwischen mA und A-Bereich erfolgt nach
Anlegen der Prüfspitzen automatisch.
Achtung!
Bei Strömen über 2 A spricht die vergossene
Sicherung an.
Diese kann nur durch den Hersteller ersetzt werden
(siehe Abschnitt 7).
4.5 Strommessung mit Strommesszange
MZ 1005 (I blinkt)
Achtung!
Messungen mit der MZ 1005 sind nur in
Ex I-Bereichen zulässig.
Bitte beachten Sie die separate Bedienungsanleitung der Messzange MZ 1005.
Schließen Sie die MZ 1005 an die 7-polige Universalbuchse an. In dem Strommesszangenbereich können Sie AC/DC-Ströme von 0,1 – 1000 A messen.
Die Wahl des optimalen Messbereiches erfolgt automatisch.
8
4.6 Frequenzmessung (FrE)
Schließen Sie die Spannungs-Messspitzen
EXM 72010 an die 7-polige Universalbuchse an.
Der Messbereich dient zur Messung von Frequenzen im Bereich von 0,1 Hz – 10 kHz bei Spannungen
> 5 V. Die Wahl zwischen Hz und kHz-Bereich erfolgt nach Anlegen der Prüfspitzen automatisch.
Hinweis:
Im Leerlauf kann die Anzeige 0,00 Hz/kHz ±1 Digit anzeigen.
Bei niedrigen Frequenzen muss das Messsignal einige Sekunden anliegen, um einen zuverlässigen
Wert anzuzeigen.
Die automatische Abschaltung kann durch
Störsignale, z.B. durch Anzeige niedriger Frequenzen bei angeschlossenen Messspitzen, deaktiviert werden.
4.7 Temperaturmessung mit TMZ 25 (T blinkt)
Messungen mit dem eigensicheren TMZ 25 und einem Fe-CuNi-Fühler sind in Ex I- und Ex II
Bereichen zulässig.
Schließen Sie den TMZ 25 an die 7-polige Universalbuchse an und stecken Sie den Temperatur-Fühler auf den TMZ bis er einrastet.
Mit dem TMZ 25 können Sie Temperaturen von
– 80 bis + 600 °C messen. Die Wahl des optimalen
Messbereiches erfolgt automatisch.
Hinweis:
Mit dem Universalfühler ist die Anzeige erst nach etwa 30 s abzulesen, Oberflächenfühler zeigen nach ca. 10 s zuverlässig an. Spezialfühler können geliefert werden.
9
5. Technische Daten METRALINE EXM25
5.1 Kennzeichnung / Ex-Zonen
EG-Baumusterprüfbescheinigung
BVS 13 ATEX E 089
II 2G Ex ib IIC T4/T6 Gb
I M2 Ex ib I Mb
5.2 Sicherheitstechnische Grenzwerte
Eigensichere Stromkreise
Strom (bis U ss
= 50 V Spitze) 2 A innere Induktivität Li < 5 μH
Spannungsfreiheit prüfen mit Messspitze:
Typ EXM 72010
Gruppe IIC AC/DC ≤ 690 V mit Spitze EXM 72010
Gruppe IIB AC/DC ≤ 690 V mit Spitze EXM 72010
Gruppe I AC/DC ≤ 1000 V mit Spitze EXM 72010
Widerstandsbereiche Max.-Werte im Fehlerfall:
Spannung 6 V
Strom 2 mA
Höchstzulässige externe Kapazität/Induktivität:
Ex Gruppe IIC 40 µF/1000mH
Ex Gruppe IIB 1000 µF/1000mH
Ex Gruppe I 3000 µF/1000mH
Frequenz 2 bis 10 kHz
Strom (bis 50 V U ss
) 2 A
Innere Induktivität Li < 5 μH
Energiequelle (vergossene Ausführungen):
NiMH-Akku
Typ Z209A/T4
Nennspannung DC 8,4 V
Maximalspannung U
0
11,2 V
Temperaturklasse T4
Li-Batterie
Typ Z209B Li/T6
Nennspannung DC 3,6 V
Maximalspannung U
0
3,9 V
Temperaturklasse T6
Temperatur mit TMZ 25:
-80°C <TA< +600°C
Strom mit Stromzange Typ MZ 1005:
(nur für Gruppe I )
AC/DC 1000 A
Frequenz bis 500 Hz
10
5.3 Weitere technische Daten METRALINE EXM25
Normen
EN 60079-0 und EN 60079-11 sowie
EN 61010-1 und EN 61010-031
EG-Baumusterprüfbescheinigung
BVS 13 ATEX E 089
II 2G Ex ib T6/T4 Gb und I M2 Ex ib I Mb
Nennspannungsbereiche verwendbar in Stromnetzen für Gruppe
Ex I bis 1 kV, Ex IIB und Ex IIC bis 690 V
Gleichspannung 1 kV mit EXM 72010 rote LED ab 12 V, LCD 3 1/2 stellig:
1, 10, 100, 1000 V (1610 V),
± 1,5% + 3 Digit
Auflösung 0.001 ... 1 V
Wechselspannung 1 kV mit EXM 72010 rote LED ab 12 V, LCD 3 1/2 stellig, TRMS;
10, 100, 1000 V (1160 V)
± 1,5% + 3 Digit bei 0 ... 100 Hz
± 5% + 5 Digit bei 101 ... 500 Hz
Auflösung 0,01 ... 1 V
Eingangswiderstand mit EXM 72010
2 MΩ aufgeteilt auf 4 Widerstände, fest in DATA-HOLD-Messspitzen vergossen
Durchgang/Widerstand grüne LED LCD 3 1/2 stellig
200, 2000 Ω
20, 200, 2000 kΩ, 2 MΩ, 20 MΩ ± 1 % + 5 Digit,
Auflösung 0,1Ω ... 10 kΩ
Strom
AC/DC 1000 mA, 2 A ± 1 % + 2 Digit
Auflösung 1 ... 10 mA
Strom mit Messzange
AC/DC 100, 1000 A, ± 1,5 % + 2/3 Digit
Auflösung 0,1 ... 1 A
Frequenz
200, 2000 Hz ± 1 % + 2 Digit, Auflösung 0,1 ... 1 Hz
10 kHz ± 3 % + 2 Digit, Auflösung 0,01 kHz
Temperatur mit Messzusatz
-80°C... +150°C ± 1,5 % + 3 Digit, Auflösung 0,1°C
-80°C... +600°C ± 1,5 % + 2 Digit, Auflösung 1°C
11
Weitere Funktionen
Automatische Messbereichsschaltung,
Funktionsanzeige, Eigentest, Abschaltautomatik
Stromversorgung eigensicherer Energieblock leicht austauschbar
Li-Batterie (T6) oder NiMH-Akku (T4)
Wechsel im Ex-Bereich zugelassen
Betriebstemperatur
- 10°C ... + 40°C
Elektromagnetische Verträglichkeit
EMV-Anforderungen DIN-EN 61326
Gehäuse
PA schlagzäh, antistatisch, LCD- Abdeckung
PC unzerbrechlich
Anschlüsse
3 Normbuchsen
7-polige Universalbuchse
Schutzart
IP 54, bei Niederschlag verwendbar
Maße / Gewicht
85 x 180 x 38 mm / 335 g
6. Batterieanzeige
Der aktuelle Zustand der Batterie / des Akkus wird
über die dreistufige Batterieanzeige im Display angezeigt.
= Batterie voll
= Batterie halb voll
Es sind noch viele
Messungen möglich.
Achtung!
= Batterie leer
Die Hintergrundbeleuchtung wird
automatisch deaktiviert.
Wenn das leere Batteriesymbol blinkt, sind keine
Messungen mehr möglich, die Batterie muss dann sofort gewechselt oder der Akku geladen werden.
6.1 Batterie austauschen
Der Batteriewechsel ist in Ex-Bereichen möglich.
Es dürfen ausschließlich folgende Energieblocks verwendet werden:
Lithium-Batterie für Gruppe I und II T6:
Typ Z209B EXM-LB3
NiMH-Akku für Gruppe I und II T4:
Typ Z209A EXM-AK9
12
6.2 Akku laden
Achtung!
Akkus dürfen nur außerhalb von Ex-Bereichen geladen werden.
6.2.1 Sicherheitshinweise zum Laden
- Das Gerät NG4 dient nur zum Laden von NiMH-
Akkus der Firma GOSSEN METRAWATT
(Art. Nr. Z209A) in Verbindung mit einem
METRALINE EXM25.
- Lithium Batterien (Art. Nr. Z209B) dürfen nicht
wieder aufgeladen werden. Der Versuch kann
Sach- und Personenschäden nach sich ziehen.
- Der Ladevorgang muss spätestens nach
24 Stunden manuell beendet werden.
- Das Ladegerät darf nur an 230 V AC angeschlossen
und betrieben werden.
- Der Betrieb ist nur im trockenen Innenbereich
zulässig.
- Wärmestau vermeiden
- Das Ladegerät und der angeschlossene
METRALINE EXM25 dürfen nicht unbeaufsichtigt
betrieben werden.
- Es ist darauf zu achten, dass die Isolierung des
Gehäuses bzw. der Anschlussleitung weder
beschädigt noch zerstört ist.
- Das Ladegerät darf nicht mehr verwendet werden,
wenn das Gerät sichtbare Beschädigungen
aufweist oder das Gerät nicht mehr arbeitet.
- Batterien oder Akkus dürfen nicht geöffnet werden.
- Sind die Batterien leer, dürfen sie nicht in den
Hausmüll. Bitte geben Sie die Batterien im Handel
oder an den Recyclinghöfen der Kommunen ab.
Die Rückgabe ist unentgeltlich und gesetzlich
vorgeschrieben.
Technische Daten NG4:
Betriebsspannung: 230 V AC 50 Hz
Leistungsaufnahme: ca. 5 W
Ladestrom: ca. 30 mA
Betriebsbedingung: 0 ... + 40° C relative Luftfeuchttigkeit < 75 % nicht kondensierend
6.2.2 Laden
Vergewissern Sie sich, dass in Ihrem METRALINE
EXM25 ein ladefähiger Akku (Art.Nr. Z209A) und keine Batterie eingesetzt ist. Das Laden erfolgt mit dem
Steckernetzgerät NG4 ohne Entnahme des Akkus aus dem METRALINE EXM25: Das NG4 wird mit der
7-poligen Universalbuchse des EXM25 verbunden und in eine Steckdose 230 V/50 Hz gesteckt. Der
Ladevorgang dauert ca. 14 Stunden.
13
Hinweis:
Akkus sind werkseitig nur vorgeladen. Vor der ersten
Inbetriebnahme sollte der Akku 14 Stunden geladen werden. Die volle Kapazität von NiMH Akkus wird erst nach 2 - 3 Lade-/Entladezyklen erreicht.
(BAT-Anzeige = max. halber Kasten)
Häufiges Überladen schadet dem Akku.
7. Wartung / Lagerung
7.1 Allgemeine Informationen
Der METRALINE EXM25 ist völlig wartungsfrei.
Dennoch ist folgendes für den sicheren Betrieb zu beachten:
Bewahren Sie das Messgerät stets in trockenem und sauberem Zustand auf. Das Gehäuse können Sie mit einem mit Isopropanol (Alkohol) oder Seifenwasser befeuchteten Tuch reinigen.
7.2 Wiederholungsprüfung
Nach EN 61243-3 wird eine Wiederholungsprüfung empfohlen.
Sie soll die Frist von 6 Jahre nicht überschreiten.
Je nach Einsatzbedingungen und Häufigkeit der
Benutzung kann eine frühere Prüfung vom Anwender festgelegt werden.
Die Seriennummer mit Herstelldatum
(WWJJNN=Woche Jahr Nummer) ist auf der Rückseite des Geräts eingeprägt. Reparaturen oder Wiederholungsprüfungen werden durch eine Prüfplakette gekennzeichnet.
7.3 Rücknahme und umweltverträgliche Entsorgung
Bei dem Gerät handelt es sich um ein Produkt der
Kategorie 9 nach ElektroG (Überwachungs- und
Kontrollinstrumente). Dieses Gerät fällt nicht unter die RoHS-Richtlinie. Nach WEEE 2002/96/EG und
ElektroG kennzeichnen wir unsere Elektro- und Elektronikgeräte (ab 8/2005) mit dem nebenstehenden
Symbol nach DIN EN 50419. Diese Geräte dürfen nicht mit dem Hausmüll entsorgt werden. Bezüglich der Altgeräte-Rücknahme wenden Sie sich bitte an unseren Service, Anschrift siehe Kap. 8.
14
8. Reparatur- und Ersatzteil-Service
Kalibrierzentrum und Mietgeräteservice
Eine Reparatur ist nur durch den Hersteller oder durch vom Hersteller ausdrücklich ermächtigte
Werkstätten zulässig.
Bei Beschädigung des Gerätes, Ausfall des Funktionstests nach Abschnitt 3.2 oder zur detaillierten
Überprüfung/Kalibrierung wenden Sie sich bitte an:
www.gmci-service.com oder senden Sie das Gerät mit Fehlerbeschreibung an die folgende Adresse:
GMC-I Service GmbH
Service-Center
Thomas-Mann-Straße 20
D-90471 Nürnberg
Telefon +49 911 817718-0
Telefax +49 911 817718-253
E-Mail [email protected]
www.gmci-service.com
Diese Anschrift gilt nur für Deutschland.
Im Ausland stehen Ihnen unsere jeweiligen Vertretungen oder Niederlassungen zur Verfügung.
9. Produktsupport
Bitte wenden Sie sich im Bedarfsfall an:
GMC-I Messtechnik GmbH
Hotline Produktsupport
Telefon D 0900 1 8602-00
A/CH +49 911 8602-0
Telefax +49 911 8602-709
E-Mail [email protected]
10. Eingeschränkte Garantie und
Haftungsbeschränkung
Durch ständige Qualitätskontrollen, modernste
Elektronik und hochwertige Werkstoffe gewährleisten wir, dass dieses Multimeter für die Dauer von
2 Jahren frei von Material- und Fertigungsdefekten bleibt.
Diese Gewährleistung gilt nicht für Batterien, unsachgemäße Handhabung, nicht bestimmungsgemäße Verwendung, öffnen des Gehäuses, falsche
Lagerung oder Schäden durch Unfälle.
Es werden keine weiteren Garantien wie die Eignung für bestimmte Anwendungen abgegeben.
Wir übernehmen keine Haftung für Begleit- oder
Folgeschäden oder Verluste, gleich welche Ursache zugrunde liegt.
15
11. Zubehör/Ersatzteile
Ledertasche EXM-LED
Art. Nr. Z209L
Die Ledertasche für EXM und Prüfleitungen ist so gestaltet, dass das Gerät bei Messungen nicht entnommen werden muss.
Die zusätzlichen Laschen und Druckknöpfe ermöglichen, das Gerät in bequemer Ableseposition umzuhängen.
Strommesszange MZ 1005
Art. Nr. Z209C für Strommessungen von 0,1 – 1000 A nur für Ex I-Bereiche.
Steckernetzgerät NG4
Art. Nr. Z209P zum Laden des Akkus
NiMH-Akku EXM -AK9/T4
Art. Nr. Z209A
Lithium-Batterie EXM -LB3 Li/T6
Art. Nr. Z209B
Temperaturmesszusatz TMZ 25
Art. Nr. Z209T
– 80 bis + 600 °C für Ex I- und Ex II-Bereiche
Universalfühler EXM-TFU Art. Nr. Z209U
Oberflächenfühler EXM-TFO Art. Nr. Z209S
Spannungs-Messspitzen EXM 72010
Art. Nr. Z209V bis 690 V Ex II und 1000 V Ex I: mit DATA-HOLD-Taster
16
Kurzanleitung METRALINE EXM25
1
Diese Kurzanleitung hilft Ihnen für den Schnell- einstieg. Bitte beachten Sie zu Ihrer Sicherheit, für sicherheitstechnische Ex-Kennwerte und für weitere
Informationen die ausführliche Bedienungsanleitung.
Spitzen oder Zubehör stecken
Spannung
Ex-Gruppe
IIC ≤ 690 V +10 %
IIB ≤ 690 V +10 %
I ≤ 1000 V
Temperatur mit TMZ
Strom mit MZ
2
Widerstand
0,1Ω - 20 MΩ
Achtung: Vorher
Spannungsfreiheit prüfen!
Einschalten und Eigentest
Strom
1 mA - 2 A
Achtung: Nur bis max. 50 V!
Ein/Aus/Test
3
Bereich/Funktion wählen
Funktionen
Spannung mit Spitze EXM 72010
Funktionen durchschalten
Anzeige obere Zeile
1000 V
Spannung mV und V automatisch AuTo
Widerstand Ω
Strom I
Strom mit Messzange I blinkt*
Frequenz FrE
Temperatur mit TMZ T blinkt*
Batteriestatus
Obere Zeile zeigt aktuell gewählte
Funktion.
Untere Zeile zeigt den aktuellen Messwert
Vorzeichen
* Messbereiche die durch Blinken signalisiert werden, benötigen Sonderzubehör wie z.B. Strommesszange oder Temperaturmesszusatz.
17
18
GOSSEN METRAWATT
3-349-729-15
Operating Instructions 1/4.14
METRALINE EXM25
GMC-I Messtechnik GmbH
Südwestpark 15
D-90449 Nürnberg
Phone +49 911 8602-111
Fax +49 911 8602-777
E-Mail: [email protected]
www.gossenmetrawatt.com
10
11
6
5
4
3
7 8 9
1
2
12
1 Test electrodes of voltage measuring line
(red conductor + /black conductor -)
2 Button DATA-HOLD
3 LED Ω (green): 0 ... 10 kΩ
4 LED Volt (red): 12 ... 1000 V
5 LCD display
6 -socket, black standard socket
7 7-pin universal connector for voltage measuring lines, temperature transmitter, clamp-on ammeter and power supply unit
8 Ω-socket, red standard socket for resistance measurements
9 A-socket, blue standard socket fuse socket for current measurements
10 ON/OFF button
11 Function button
12 Handgear
Symbols on the instrument
Attention! Observe user instructions!
Ex marking:
Approved for potentially explosive atmospheres in accordance with ATEX (EN 60079-0 and EN
60079-11) see sectionl 5.1
EC-conformity
20
1. Application
The METRALINE EXM25 is an intrinsically safe multimeter which can be used in areas with a potentionally explosive atmosphere in accordance with ATEX
(EN 60079-0 and EN 60079-11) and EN/IEC 61010 for voltage, resistance, current, frequency measurements with measurement accessories for measurement of temperature and high current.
Model M210A with rechargeable battery and model
M210B with Lithium battery.
1.1 Intended use
This device is intended for use in applications as described in the operating instructions only.
Thus, it is imperative to observe the notes on safety and the technical data in conjunction with the ambient conditions.
Any other form of usage is not permitted and can lead to accidents or destruction of the unit.
Any misuse will result in the expiry of all guarantee and warrantly claims.
2. Safety Precautions
When used for its intended purpose, safety of the operator, as well as that of the instrument, is assured.
The voltage measurement tips contain moulded multipliers within both test prods. They are extremely safe and comply with overvoltage category CAT IV.
In order to maintain flawless technical safety conditions, and to assure safe use, it is imperative that you read these operating instructions thoroughly and carefully before placing your instrument into service, and that you follow all instructions contained therein.
The device may only be used in the designated
Ex zones (see section 5.1) and inside of the safety-
related limit values (see section 5.2).
Before starting resistance measurements make
sure that the test object is at zero-potential.
Perform voltage measurements most up to the
following limits:
With probes EXM 72010 up to 1000 V
Faultless indication of display values is only
guaranteed between -10°C ... +40°C.
avoid covering the display or touching the test
electrodes.
21
Only qualified persons may carry out work with
these device. The user needs to be farmiliar with
the risks for measuring and compliance
with safety regulations and the proper use of the
tester.
Workings may only be performed with appropri-
ate personal protective equipment.
Observe the minimum object distance to other
plant components that are energized or earthed
and use personal protective equipment as speci-
fied by national accident prevention regulations
(in Germany: BGV A3 or VDE 0105-100).
The function of the measuring device must be
checked briefly before and whenever possible
after the use. Carry out the function test and
check the instrument at a known voltage source
(AC and DC). If the indication of one or several
systems fails in the course of checking, the
instrument must not be used again.
The tester may only be dismantled by authorised
personnal.
Maintainance is only allowed by the manufac-
turer or explicitly authorised repair shops
(see section 7).
Before using the device check the housing and
connecting line for visible damage. If damages
are visible the voltage tester may not be placed into
operation. In case of strong dirt contamination, the
tester must be cleaned before use.
The device has to be stored in a clean and dry
environment.
22
3. Putting into operation
3.1 Battery
Your instrument is already supplied with an energieblock. Solely use the following energyblocks:
NiMH accumulator for class I and II T4: type Z209A EXM-AK9
Lithium battery for class I and II T6: type Z209B EXM-LB3
Attention!
Please observe section 6 before initial startup or after your device has been in storage for a long period of time.
3.2 Testing display and function
At every day of use the METRALINE EXM25 has to be checked to ensure that it works properly and faultlessly.
Self test 1 (Display test):
Press and hold button „ON/OFF“.All display segments light up on the display, additionally the V-LED and the Ω-LED light up. When you release button
„ON/OFF“, the value 000 ... 001 V is indicated on the display.
Self test 2 (Voltage measuring line):
Connect and lock the voltage measurement tips
EXM 72010 to the 7-pin universal connector. Switchon the device. Display: 000..001 V, in the upper display line 1000 V. Put the test electrodes, one after another, inclined 5 mm into the W-socket. Display shows ‘Test’ and ‘rdy’ and green LED lights up.
Note:
In case the display continues showing 000 V the measurement tip is damaged, please exchange.
Self test 3 (Voltage indication):
Check functions at a known voltage source - e.g. a
230 V socket.
Self test 4 (Resistance measurement):
Put the standard test lines into the Ω- and -socket.
Adjust the device to Ω, display: OL kΩ, hold test prods together, display indication needs to be about
0.0 Ω and the green LED needs to light up.
Attention!
If indication of one or several systems fail in the course of performing the self test, if function standby is not indicated or if the device is damaged, the
METRALINE EXM25 must not be used again.
23
4. Measuring and testing
4.1. General information battery status (see section 6) leading sign
The upper display line shows currently selected measurement range.
When this indication is flashing you have to connect special accessories e.g. TMZ or MZ
The lower display line shows the actual metered value.
The instrument is switched off automatically approximately 60 seconds after the last measurement in order to extend battery life. In the display the indication „off“ appears (exception: see section 4.6 frequency measurements). The backlight turns off when no measurement result is applied or when the battery is low.
With the ON/OFF button engaging the
METRALINE EXM25 to voltage, call the display test and switch-off the device.
►
Function
With the function button you can select the designated measurement range: upper display line
Voltage with tip EXM 72010 1000 V
Voltage, automatic range mV, V AuTo resistance Ω current I current with clamp-on ammeter I flashes* frequency FrE temperature with TMZ T flashes*
* When the measurement ranges are indicated by a flashing symbol the connection of special accessories is necessary, e.g. a clamp-on ammeter or the temperature transmitter
24
4.2 Testing voltage and polarity
Attention!
Connect the voltage measurement tips to the
7-pin universal connector only when the device is switched-off. Switch-on the device not until they are locked and select rang if necessary.
Put both test prods with safe contact onto the metering point and the voltage value is indicated in the lower display line.
At a voltage of 12 V flashes the red LED and button functions are locked.
Notes:
- If Hold appears permanently, probably the tip
has been attached when the device was switched-
on. Switch METRALINE EXM25 off and on.
- Because of integrated multipliers the device indi-
cates few mV while in open-circuit operation, this
has no influence on the result of measurement.
- The METRALINE EXM25 casing is made of
conductive material. We recommend unearthed
operation for precise voltage measurments
>100 Hz. Avoid contact with grounded parts, e.g.
control cabinet. For this purpose, the use of a
leather bag should be preferred.
1000 V voltage range (1000 V)
The voltage range adjusts automatically to the used DATA-HOLD measurement tip EXM 72010.
The measurement range serves for a fast detection of measurement values between 0 to 1000 V without decimal places.
The display range of the METRALINE EXM25 extends to 1160 V AC / 1610 V DC with EXM 72010.
Observe the safety precautions (see section 2) for
Ex-areas.
Automatic voltage range (AuTo)
The voltage range adjusts automatically to the used voltage measurement tip EXM 72010.
The measurement range serves for a fast detection of measurement values between 1 mV – 1000 V, the optimal measurement range is selected automatically.
Indication of polarity
Type of voltage is indicated by the symbols
~ and – . Direct voltage: is minus applied to the test prod with Hold button, the leading sign „-“ appears, is plus applied, no leading sign appears in front of the indicated value.
25
HOLD function
The maximum voltage value can be „stored“ on the display when activated the button „HOLD“. The value is recorded for approx. 30 seconds or until you press button „HOLD“ again. The Hold-function is stopped when again a voltage is impressed.
Note: When the measured value does not vary for
2 sec, then the maximum value is recorded.
4.3 Resistance measurements (Ω)
Before taking any resistance measurements always check that the test object is at zero-potential.
Connect 4 mm standard test lines to the Ω- and
-socket. Adjust the device to Ω. Ohm is indicated in the upper display line.
Note:
The arrow signalizes „out of measurement range“.
The measurement range serves for a definite determination of impedances of 0,1 Ω – 20 MΩ.
A selection between Ω-, kΩ- and MΩ ranges occurs automatically after applying to the impedance.
The green LED signalises resistance values that a lower than 10 kΩ.
4.4 Current measurement (I)
Attention!
You may perform current measurement in Ex-areas only in measuring circuits with peak values of maximum 50 V.
You may only measure currents up to maximum 2 A.
Connect 4 mm standard test lines to the A- and
-socket.
The measurement range serves for measuring
AC/DC currents in the range of 1 mA – 2 A.
A selection between mA and A occurs automatically after attaching the test prods.
Attention!
The moulded fuse activates with currents of more than 2 A.
The fuse can only be changed by the manufacturer
(see section 7).
4.5 Current measurement with clamp-on ammeter
MZ 1005 (I flashes)
Attention!
Measurements with MZ 1005 are only permissible in Ex I-areas.
Please observe the separate user instructions of the clamp-on ammeter MZ 1005. Connect the MZ 1005 to the 7-pin universal connector.
Within the clamp-on ammeter range you can measure AC/DC currents between 0,1 and 1000 A, the optimal measurement range is selected automatically.
26
4.6 Frequency measurements (FrE)
Connect the voltage measurement tip EXM 72010 to the 7-pin universal connector.
The measurement range serves for measurements of frequencies in the range of 0,1 Hz – 10 kHz with voltages > 5 V. A selection between Hz and kHz occurs automatically after attaching the test prods.
Note:
While the device is in open-circuit operation the display may show 0,00 Hz/kHz ±1 digit.
With low frequencies the measurement signal needs to be placed for a few seconds before it indicates a reliable value.
The automatic shout-down can be deactivated by interfering signals, e.g. because of low frequencies with attached measurement tips.
4.7 Temperature measurements with TMZ 25
(T flashes)
Measurements with the intrinsically safe TMZ 25 and a Fe-CuNi-sensor are only permissible in Ex I- and Ex
II-areas. Connect the TMZ 25 to the 7-pin universal connector and plug temperature sensor on the TMZ until it’s catched.
With the TMZ 25 you can measure temperatures between – 80 and + 600 °C. The optimal measurement range is selected automatically.
Note:
With the universal sensor indication on the display is first readable after approx. 30 s, with surface sensors after approx. 10 s. Special sensors can be delivered.
27
5. Technical data METRALINE EXM25
5.1 Identification marking / Ex zones
EC-Type Examination Certificate
BVS 13 ATEX E 089
II 2G Ex ib IIC T4/T6 Gb
I M2 Ex ib I Mb
5.2 Safety-related limit values
Intrinsically electric circuits
Voltage (by U pp
= 50 V tip) 2 A
Internal inductance Li < 5 μH
Voltage with measurement tips:
Type EXM 72010
Class IIC AC/DC ≤ 690 V with tip EXM 72010
Class IIB AC/DC ≤ 690 V with tip EXM 72010
Class I AC/DC ≤ 1000 V with tip EXM 72010
Resistance ranges maximum values in case of fault:
Voltage 6 V
Current 2 mA
Maximum permissible external capacitance/inductance:
Ex group IIC
Ex group IIB
Ex group I
40 µF/1000mH
1000 µF/1000mH
3000 µF/1000mH
Frequency 2 up to 10 kHz
Current (up to 50 V U ss
) 2 A internal inductance Li < 5 μH
Energie source (moulded design):
NiMH accumulator
Type Typ Z209A/T4
Nominal voltage DC 8,4 V
Maximum voltage U
0
11,2 V
Temperature range T4
Lithium battery
Type
Nominal voltage DC 3,6 V
Maximum voltage U
0
3,9 V
Temperature range T6
Temperature with TMZ 25:
-80°C <TA< +600°C
Curent with clamp-on ammeter type MZ 1005:
(solely for class I)
AC/DC 1000 A
Frequency up to 500 Hz
28
5.3 Technical data METRALINE EXM25
Norm
EN 60079-0 and EN 60079-11 also
EN 61010-1 and 61010-031
EC type examination certificate
BVS 13 ATEX E 089
II 2G Ex ib T6/T4 Gb and I M2 Ex ib I Mb
Nominal measurement ranges usable in electricity networks group Ex I up to
1 kV, Ex IIB and Ex IIC up to 690 V
Direct current 1 kV with EXM 72010 red LED up to 12 V, LCD 3 1/2-digit:
1, 10, 100, 1000 V (1610 V),
± 1,5% + 3 digit, resolution 0,001 ... 1 V
Alternating voltage 1 kV with EXM 72010 red LED up to 12 V, LCD 3 1/2-digit, TRMS;
10, 100, 1000 V (1160 V)
± 1,5% + 3 digit up to 0 ... 100 Hz
± 5% + 5 digit up to 101 ... 500 Hz resolution 0,01 ... 1 V
Input resistance with EXM 72010
2 MΩ distributed to 4 resistances, moulded into DATA-HOLD test prods
Continuity/resistance green LED LCD 3 1/2-digit
200, 2000 Ω
20, 200, 2000 kΩ, 2 MΩ, 20 MΩ ± 1 % + 5 digit, resolution 0,1 Ω ... 10 kΩ
Current
AC/DC 1000 mA, 2 A ± 1 % + 2 digit, resolution 1 ... 10 mA
Current with clamp
AC/DC 100, 1000 A, ± 1,5 % + 2/3 digit, resolution 0,1 ... 1 A
Frequency
200, 2000 Hz ± 1 % + 2 digit, resolution 0,1 ... 1 Hz
10 kHz ± 3 % + 2 digit, resolution 0,01 kHz
Temperature with measurement accessory
-80 ... +150° C ± 1,5 % + 3 digit, resolution 0,1° C
-80 ... +600° C ± 1,5 % + 2 digit, resolution 1° C
Further functions automatic selection of measurement range, indication of function, self-test, automatic switch-off
29
Power supply intrinsically safe energy source, exchangeable
Li-battery (T6) or NiMH accu (T4) exchange in Ex-areas admitted
Operating temperature
- 10°C ... + 40° C
Electromagnetic compatibility
EMV requirements DIN-EN 61326
Casing
PA impact resistant, antistatic, LCD cover
PC unbreakable
Circuit points
3 standard jacks
7-pin universal connector
Protection category
IP 54, device can be used in moist environments
Dimensions/weight
85 x 180 x 38 mm / 335 g
6. Maintainance
The METRALINE EXM25 is completely maintainancefree except for ist own energy source (see section 6.1).
Nevertheless, for safe operation observe the following information:
The METRALINE EXM25 is to be stored in a dry place.
The plastic housing can be cleaned with a cloth dampened with alcohol (isopropyl) or soapy water.
6.1 Battery status
The latest status of the battery or the accumulator is symbolised by a 3 stage battery symbol in the display.
=
=
Many measurements still
can be performed.
=
The backlight deactivates
automatically.
Attention!
When the empty battery symbol flashes, taking measurements is not possible anylonger. The battery needs to be exchanged or the accumulator recharged immediately.
6.2 Change battery
The change of battery is possible in Ex-areas.
Solely use the following energyblocks:
Lithium battery for class I and II T6: type Z209B EXM-LB3
NiMH accumulator for class I and II T4: type Z209A EXM-AK9
30
6.3 Load accu
Attention!
Recharge accumulators outside of Ex-areas.
6.3.1 Safety Precautions for loading
- The device NG4 is used to charge NiMH
accumulators from GOSSEN METRAWATT
(Art. no. Z209A) installed in the METRALINE EXM25.
- Lithium batteries (Art. no. Z209B) are not to be
re-charged.
cause risk to personal safety and damage to the
equipment.
- The charging has to be stopped manually no later
than 24 hours.
- The charger can only be connected to and
operated with 230 V AC.
- The product is only to be used in dry indoor
locations.
- avoid possible damage
- The charger and the connected METRALINE EXM25
should not be operated unattended.
- Ensure that the insulation of the housing or the
main cable is neither damaged or destroyed.
- The charger may not be placed into operation,
if the device exhibits visible damage or the device
does not operate any longer.
- The batteries may not to be opened.
- Depleted batteries must not be disposed with the
domestic waste. Please, return batteries at a local
retailer or municipal recycling depot.
Return is free of charge and required by law.
Technical data NG4:
Operating voltage:
Power input:
Charging current:
Operating instructions: 0 ... + 40° C
230 V AC 50 Hz ca. 5 W ca. 30 mA rel. humidity < 75 % not condensing
6.3.2 Recharge
Make sure a rechargeable accumulator (Art. no. Z209A) and no battery is installed in the METRALINE EXM25.
Recharging occurs with the power supply unit NG4 without removing the accumulator from the
METRALINE EXM25:
The NG4 is connected with the 7-pin universal connector of the EXM25 and put into the 230 V/50
Hz socket. Charging takes about 14 hours.
31
Note:
Accumulators were only shortly charged by the manufacturer. Before the first initial startup the akku should be charged for 14 hours . Full capacitance of
NiMH accumulators ist first reached after 2 - 3 cycles of charging and discharging.
(BAT symbol = maximum half box)
The accumulator suffers damages with frequent supercharging.
7. Maintenance
7.1 General information
The METRALINE EXM25 is absolutely mainteinancefree. Nevertheless, observe the following information in order to maintain safe operation: Always keep the voltage tester dry and clean. The housing can be cleaned with a cloth dampened with isopropyl (alcohol) or soapy water.
7.2 Repeated inspection
According to EN 61243-3 it is recommended to carry out repeated examinations.
It should not exceed the time-limit of 6 years.
Depending on operation conditions and frequency, a previous inspection may be recommendable.
The serial number with the date of manufacturing
(WWYYNN=Week Year Number) is imprinted on the backside of the device. Repeated inspections are offered by the manufacturer and indicated by the inspection plate.
7.3 Device Return and Environmentally
Compatible Disposal
The instrument is a category 9 product (monitoring and control instrument) in accordance with ElektroG
(German Electrical and Electronic Device Law).
This device is not subject to the RoHS directive.
We identify our electrical and electronic devices
(as of August 2005) in accordance with WEEE
2002/96/EG and ElektroG with the symbol shown to the right per DIN EN 50419 .
These devices may not be disposed of with the trash. Please contact our service department regarding the return of old devices (address see chapter 8).
32
8. Repair and Replacement Parts Service
Calibration Center and Rental Instrument Service
Repair is only allowed by the manufacturer or explicitly authorized repair shops.
In case of damages on the device or failure of the function test according to section 3.2 or for detailed inspection/calibration, please contact:
www.gmci-service.com or send the device and a description of failure back to the following adress:
GMC-I Service GmbH
Service Center
Thomas-Mann-Straße 20
90471 Nürnberg, Germany
Phone +49 911 817718-0
Fax +49 911 817718-253
E-Mail [email protected]
www.gmci-service.com
This address is only valid in Germany.
Please contact our representatives or subsidiaries for service in other countries.
9. Product Support
If required, please contact:
GMC-I Messtechnik GmbH
Hotline Produktsupport
Phone +49 911 8602-0
Fax +49 911 8602-709
E-Mail [email protected]
10. Limited warranty and limitation of liability
By continuous quality checks and production controls, most modern electronics and high quality materials we guarantee that the device will be free from defects in material and workmanship for two years.
This warranty does not cover batteries, improper handling, not intended purpose, opening the housing, improper storage or damages from accidents.
No other warranties such as fitness for a particular purpose will be given.
We are not liable for any indirect, incidental or consequential damages or losses arising from any cause or theory.
33
10. Accessories / Spares
Leather bag EXM-LED
Art.no. Z209L
The case for the METRALINE EXM25 and the test cables is designed in such a way that the unit does not have to be taken out for measurements.
Using the additional lugs and push-buttons on the shoulder strap and case the unit can be secured in a comfortable position to facilitate readings.
Clamp-on ammeter MZ 1005
Art. no. Z209C for current measurements of 0,1 – 1000 A only in Ex – I areas.
Power supply unit NG64
Art. no. Z209P for charging the accumulator
NiMH accumulator EXM-AK9/T4
Art. no. Z209A
Lithium battery EXM-LB3 Li/T6
Art. no. Z209B
Temperature transmitter TMZ 25
Art. no. Z209T between – 80 and + 600 °C in Ex-I and Ex–II areas universal sensor EXM-TFU Art. no. Z209U surface sensor EXM-TFO Art. no. Z209S
Voltage measurement tips EXM 72010
Art. no. Z209V up to 690 V Ex II and 1000 V Ex I: with DATA-HOLD button
34
Quick user guide METRALINE EXM25
This quick user guide serves für a quick start. In regard of your own safety please observe safety-related Excharacteristic values and for further information the detailed user instructions.
1
Attach tips or accessories
Voltage
Ex-Group
IIC ≤ 690 V +10 %
IIB ≤ 690 V +10 %
Resistance
0,1Ω - 20 MΩ
I ≤ 1000 V
Temperature with TMZ parts!
Current with MZ
Attention: At first check voltage-free
2
Switching-on and self-test
Current
1 mA - 2 A
Attention: Up to max. 50 V only!
On/Off/test
3
Select range/function
Function
Voltage with tip EXM 72010
Select function upper display line
1000 V
Voltage, automatic range mV / V AuTo resistance Ω current I current with clamp-on ammeter I flashes* frequency FrE temperature with TMZ T flashes* battery status leading sign
The upper display
line shows currently selected measurement range.
The lower display
line shows the actual metered value.
* When the measurement ranges are indicated by a flashing symbol the connection of special accessories is necessary, e.g. clamp-on ammeter or temperature transmitter.