Surgimedics Turbo 230V 50Hz PN 906750 Smoke Evacuator Owner's Manual

Below you will find brief information for Smoke Evacuator Turbo 230V 50Hz PN 906750. This document describes a smoke evacuation system designed to remove plume/smoke created during surgical procedures such as electrosurgery, laser surgery, and laparoscopic or power tool surgical procedures. The system features a ULPA filter, which captures particles as small as 0.1 microns with 99.9995% efficiency, and a high-grade activated carbon for gas and odor absorption. The Turbo 230V 50Hz PN 906750 allows for adjustable suction power, ensuring safe and effective removal of smoke throughout your surgical procedures.

PDF Download
Document
Surgimedics Smoke Evacuator Turbo 230V 50Hz PN 906750 Owner's Manual | Manualzz

Advertisement

Advertisement

Advertisement

Key features

  • ULPA filter
  • Activated carbon filter
  • Adjustable suction power
  • Footswitch control
  • Filter test function
  • Easy to clean
  • Safe and effective smoke evacuation
  • Pre-filter for larger particles

Frequently asked questions

The SurgiFresh® Turbo Smoke Evacuation System is intended to be used to evacuate plume/smoke created during electrosurgery, laser surgery, and laproscopic or power tool surgical procedures.

The ULPA filter should be replaced every 90 days or sooner if it fails the filter test or produces an abnormal odor.

First, review the troubleshooting section of this manual. If the problem persists, contact Surgimedics Service.

After each use, the SurgiFresh® Turbo Smoke Evacuation System should be cleaned and prepared for future use. Turn off the power switch and unplug the unit from the wall receptacle. Remove the unit from the immediate procedure area. Wipe the footswitch and footswitch tubing with an appropriate hospital disinfectant. Coil the footswitch tubing around the cord wrap at the back of the unit. Wipe the power cord with an appropriate hospital disinfectant. Coil the power cord around the cord wrap. Remove the tubing from the pre-filter or suction canister. If a suction canister was used, securely close lids/caps and properly dispose canister. Remove the pre-filter. Dispose of all operative waste materials according to your institution’s procedures for biohazardous materials. Check the ULPA filter to be sure no fluids are on top of the filter. Replace if fluids are present. Install a new pre-filter. Thoroughly wipe all external surfaces of the SurgiFresh® Turbo with an appropriate hospital disinfectant. Follow the procedures approved by your institution.

The maximum static suction pressure of the SurgiFresh® Turbo is 92 inches H20 (172mm Hg).