PROCEQ DYNA Z, E Pull-off Tester Bedienungsanleitung
Below you will find brief information for Pull-off Tester DYNA Z Z16E, Pull-off Tester DYNA Z Z6E, Pull-off Tester DYNA Z Z16, Pull-off Tester DYNA Z Z6. These testers measure adhesive strength of various materials like coatings, concrete, and plaster. The DYNA Z...E models offer digital displays, data storage, and PC connectivity for advanced analysis. The DYNA Z models use a digital manometer for simpler readings.
Advertisement
Advertisement
Haftprüfgerät DYNA Z / DYNA Z...E
Pull-off Tester
Bedienungsanleitung
Operating Instructions
PROCEQ SA
Riesbachstrasse 57
Postfach 936
CH-8034 Zürich
Switzerland
Tel.:
Fax:
+41 (0)1 389 98 00
+41 (0)1 389 98 12
E-Mail: [email protected]
Internet: www.proceq.com
Änderungen vorbehalten
Subject to change
Copyright © 2002 by PROCEQ SA ZURICH 2002 03 077 D/E
Contents
1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
General Information . . . . . . . . . . . . . . . . . . . . . . . 2
Liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . 2
Designated Use . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Standards and Regulations Applied . . . . . . . . . . . 3
2 Product Description . . . . . . . . . . . . . . . . . . . . . 4
Field of Application . . . . . . . . . . . . . . . . . . . . . . . 4
DYNA Z Pull-Off Tester with Digital Manometer . . . . . . . . . . . . . . . . . . . . 4
DYNA Z...E Pull-Off Tester with Display Unit, DYNAMETER . . . . . . . . . . . . . 5
3 Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
DYNA Z Pull-Off Tester with Digital Manometer . . . . . . . . . . . . . . . . . . . . 6
DYNA Z...E Pull-Off Tester with Display Unit, DYNAMETER . . . . . . . . . . . . . 7
4 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Menu Options in the Main Menu . . . . . . . . . . . . . 9
Functions of the Menu Options . . . . . . . . . . . . . . 9
5 Measuring Process . . . . . . . . . . . . . . . . . . . . . 19
Preparing the Measuring Procedure . . . . . . . . . . 19
Measuring with the Digital Manometer . . . . . . . . 20
Measuring with the DYNAMETER . . . . . . . . . . . 21
6 Maintenance, Storage and Care . . . . . . . . . . . 24
Cleaning After Each Use . . . . . . . . . . . . . . . . . . 24
Performance Check Before Each Use . . . . . . . . 24
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7 Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Form of Delivery . . . . . . . . . . . . . . . . . . . . . . . . 26
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8 Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
© 2002 PROCEQ SA Contents 1
1 Safety
1.1
General Information
Please read through these operating instructions carefully before initial startup. They contain important information about safety, use and maintenance of the pull-off tester and its components.
1.2
Liability
Our "General Terms and Conditions of Sale and
Delivery" apply in all cases. Warranty and liability claims arising from personal injury and damage to property cannot be upheld if they are due to one or more of the following causes:
- Failure to use the pull-off tester and its components in accordance with its designated use
- Incorrect performance check, operation and maintenance of the pull-off tester and its components
- Failure to adhere to the sections of the operating instructions dealing with the performance check, operation and maintenance of the pull-off tester and its components
- Unauthorized structural modifications to the pull-off tester and its components
- Serious damage resulting from the effects of foreign bodies, accidents, vandalism and force majeure
1.3
Safety Regulations
1.3.1
General Information
" Perform the prescribed maintenance work on schedule.
" Carry out a performance check once the maintenance work has been completed.
" Handle and dispose of lubricants and cleaning agents responsibly.
1.3.2
Unauthorized Operators
The pull-off tester and its components are not allowed to be operated by children and anyone under the influence of alcohol, drugs or medication.
Anyone who is not familiar with the operating instructions must not use the pull-off tester and its components or must be supervised when using it.
2 Safety © 2002 PROCEQ SA
1.3.3
Safety Icons
The following icons are used in conjunction with all important safety notes in these operating instructions.
Danger!
This note indicates a risk of serious or fatal injury should certain rules of behavior be disregarded.
Warning!
This note warns you about the risk of material damage, financial loss and legal penalties (e.g.
loss of warranty rights, liability cases, etc.).
This denotes important information.
1.4
Designated Use
The pull-off tester is a mechanical device which serves to check the quality of adhesion and adhesive strength of different materials (concrete, plaster work, coatings, color, etc.) depending on customer specifications.
Warning!
The device is to be used solely on the surfaces to be tested (see chapter 2.1).
1.5
Standards and Regulations Applied
- ISO 4624
- BS 1881 part 207
- DIN 1048 part 2
- ZTV SIB 90
- SIA 2002
- NF P 18-853
- ASTM D 4541-85
- ACI 503-30
- EN 1015-12
- EN 1348
International
Great Britain
Germany
Germany
Switzerland
France
USA
USA
Europe (for mortan)
Europe
© 2002 PROCEQ SA Safety 3
2 Product Description
2.1
Field of Application
The DYNA pull-off tester is a mobile and versatile device with the following fields of application:
Measuring the adhesive strength of applied coatings:
This includes, for example, plastic coatings, concrete coats, mortars and plasters, bituminous coats as well as paint finishes and coatings of metals.
Determining the surface strength of concrete and other materials:
The surface strength is tested directly on the component.
Any measuring point can be used and there is no need to embed the sample in the concrete first.
The legs can be moved and/or extended on all models and the measuring procedure can be perfectly adapted to the test situation.
All measurements can be carried out using test discs with different diameters.
The standard diameter of a test disc is Ø 50 mm.
Test discs with other diameters can be ordered directly from PROCEQ or authorized agencies (see chapter 7.2,
Accessories).
All pictures of the pull-off tester (1) are refering to the model Z16(E).
2.2
DYNA Z Pull-Off Tester, with Digital Manometer
2
4
5
1
3
7
6
1 Pull-off tester
2 Digital manometer
3 Hydraulic pump
4 Piston
5 Device plate
6 Draw bolts
7 Test disc
Fig. 2.1 DYNA Z Pull-off tester, with digital manometer
The DYNA Z pull-off tester (1) is fitted with a digital manometer (2) to indicate the tension (or, as an option, the force).
The DYNA Z model can perform the following functions:
Measurement and display of the current value
Measurement and display of the peak value
4 Product Description © 2002 PROCEQ SA
2.3
DYNA Z...E Pull-Off Tester, with Display Unit, DYNAMETER
9 11
1 Pull-off tester
3 Hydraulic pump
5 Device plate
6 Test disc
7 Draw bolt
10 8 5 1 3 6 7
8 Electronic measuring cell
9 Display unit DYNAMETER
10 Keyboard
11 Display
Fig. 2.2
DYNA Z...E Pull-off tester, with display unit,
DYNAMETER
The DYNA Z...E pull-off tester, (1) is equipped with an electronic measuring cell (8) and a display unit,
DYNAMETER (9) with keyboard (10) and display (11).
The DYNA Z...E model has the following functions:
- Measurement of the adhesive strength and adhesion
- Display of the peak value and current value:
The values can either be indicated as force or tension
(surface load) in different units depending on individual specifications
- Input and indication of the sample number and sample dimension
- Specification of the speed of load application
- Storage of up to 1000 measured values
- Transferring the measured data directly to a connected printer or PC
- Online display possible on a PC or laptop with special software
© 2002 PROCEQ SA Product Description 5
3 Startup
3.1
DYNA Z Pull-Off Tester, with Digital Manometer
3.1.1
Starting up the Digital Manometer stops operating, the warning symbol " " appears in the left of the display to indicate that the battery capacity is too low.
3.1.2
Replacing the Battery
Display with current value
Display peak value
Switching the digital manometer on and off
Fig. 3.1
Display of the digital manometer
" Press the "ON" key to switch on the digital manometer
(2).
"0.00" is displayed (see Fig. 3.1).
If no display appears, replace the batteries.
" Press the "ON" key again to switch off the digital manometer (2).
The digital manometer switches off automatically after approx. 10 minutes to save power.
The battery capacity is sufficient for approx.
120 operating hours.
Two hours before the battery is drained and
6 Startup
1 Pull-off tester
2 Digital manometer
Fig. 3.2
Replacing the battery
Warning!
When opening the cover, ensure that the wires between the cover and the pressure sensor in the housing are not torn off.
" Turn the cover of the digital manometer (2) counterclockwise and carefully pull the cover upwards at the same time (see Fig. 3.2).
" Loosen the screw of the battery fastening and replace the battery.
© 2002 PROCEQ SA
Only use 9 volt monobloc batteries
(AM6/6LR61).
" Install the battery in reverse order.
" Then turn the cover clockwise until it reaches its initial position (see Fig. 3.2).
3.2
DYNA Z...E Pull-Off Tester, with Display Unit, DYNAMETER
3.2.1
Connecting Devices
21 22 23 24
21 Input socket for power 22 Input socket for PC/printer supply unit/external battery 23 Input socket for pull-off tester
24 not active
Fig. 3.3
Input sockets of the DYNAMETER
" Connect the connection cable of the pull-off tester to the INPUT A (23) socket of the DYNAMETER.
" If you want to connect the DYNAMETER to a printer or
PC, connect the transmission cable to the INTERFACE
RS 232 (22) socket.
© 2002 PROCEQ SA
3.2.2
Starting up the DYNAMETER
Switching on/off
Starts the selected setting/Reset function
Arrow keys for menu selection and value input
Goes to the main menu
Goes to the measuring display
Stores the values
Fig. 3.4
Keyboard (10) of the DYNAMETER
" Press the "ON/OFF" key for one second.
If nothing appears on the display, replace the batteries.
The following information is displayed briefly:
Serial number of the DYNAMETER
Warning!
Check to see if the serial number of the
DYNAMETER matches the serial number of the pull-off tester (located on the side of the device plate, 5).
You will only receive correct measuring results if the two serial numbers are identical.
Startup 7
- Installed software version
- Confirmation of the automatic self-test ("Test o.k.")
- Expected remaining battery life
Then the display automatically switches to the following measuring display (without the values shown in the example):
Probendimension Current sample numberr
Type of k-Factor
Flow bar
Load pacer
Peak value memory active
Speed of load application
Peak value
Current value
Preselected unit
Fig. 3.5
Measuring display in the display (11) of the
DYNAMETER
The settings of the last measurement are maintained even after the device is switched off.
They are indicated when the device is switched
on again (see Fig. 3.5). The values of the last measurement are no longer displayed.
The display unit is now ready for operation.
8 Startup
All parameters can be changed with the menu options in the main menu (see chapter 4).
" Press the "ON/OFF" key again to switch off the
DYNAMETER.
3.2.3
Replacing the Batteries
" Open the cover at the bottom of the DYNAMETER.
" Disconnect the electric connection and replace the batteries (6 round cells, type 1.5 volt LR6)
Wrap adhesive tape around the battery holder and the batteries before installing them so they cannot become jammed in the battery compartment when they are replaced.
" Close the cover.
© 2002 PROCEQ SA
4 Settings
4.1
Menu Options in the Main Menu
The DYNAMETER has a menu with user navigation.
Please follow the instructions in the displayed menu.
Warning!
The settings may vary depending on the measurement. Check the settings of the DYNAME-
TER prior to every new measurement!
" After starting up the DYNAMETER (see chapter 3.2.2), press the "MENU" key on the keyboard.
The following main menu appears in the display (11):
Fig. 4.1
Main menu
Menu options
Select menu options
Call selected option
Return to measuring display with the selected settings
The menu options of the main menu point to submenus, or directly to certain setting options.
The individual menu options are always selected and set using the arrow keys on the keyboard:
" Select the corresponding menu option in the main menu with the arrow keys #$ .
" Press the "START/RESET" key.
Then the respective submenu appears in the display.
" Make the desired settings (positions and values) using the arrow keys #$ %& .
Press the "MENU" key to return to the main menu (see Fig. 4.1).
Press the "END" key to return to the measuring display (see Fig. 3.5).
4.2
Functions of the Menu Options
In this chapter you will become familiar with the functions of the individual menu options.
The titles of the chapters correspond to the menu options of the main menu.
4.2.1
Data Output
With this menu option you can select the type of data output or delete data after the measurement is finished.
© 2002 PROCEQ SA Settings 9
" Select the menu option "Data Output".
The following submenu appears in the display (11):
Fig. 4.2
Data Output
Display Data
" Select the menu option "Display Data".
The data of the last measurement are displayed.
" Press the
#$ arrow keys to scroll to previous/following data.
If you keep one of the two arrow keys #$ pressed you will automatically scroll in the desired direction.
Print Data
Requirements for data transmission:
- Printer connection: serial interface
- Printer:
EPSON compatible, recommendation:
Martel MCP 9800
- Printer cable:
Art. no. 330 00 460
- Data format of the printer:
9600 baud; 8 data bit, 1 stop bit; no parity
" Prepare the printer for operation in accordance with the manufacturer´s operating instructions.
" Connect the printer cable to the Interface RS 232 socket (see chapter 3.2.1).
" Select the menu option "Print Data".
" Press the "START/RESET" key to start data transmission.
All data in the memory are now transmitted to the printer.
Selective data transmission to the printer is not possible.
Direct data transmission during measuring is not possible.
10 Settings © 2002 PROCEQ SA
Data to the PC
A transfer cable (Art. no. 330 00 456) is necessary to connect the display unit to a PC (IBM-PC or compatible PC).
" Connect the transfer cable to the Interface RS 232 socket (see chapter 3.2.1).
" Prepare the PC for operation in accordance with the manufacturer´s operating instructions.
Please refer to the enclosed information leaflet "Data
Transfer to the PC" for further information on data transmission.
" Select the menu option "PC Data".
" Press the "START/RESET" key to start data transmission.
The data are transmitted in the following order (the examples below correspond to how they are displayed on the monitor of your PC):
Example with force measurement:
Name of
DYNAMETER wd
Sample number
500 kN
4.4
8.1
12.3
14.5
Unit
Intermediate values
(when pressing the
PRINT/STORE key)
Peak value
Examples with surface load:
Sample number Sample dimension
(Standard dimension)
501 d 50 mm N/mm 2
5.8
6.5
7.7
Unit
Intermediate values
(when pressing the
PRINT/STORE key)
Peak value
Sample number
502 A 2500.00 mm 2
5.8
Test surface
(Diameter 50x50; see also chapter 4.2.3)
N/mm 2 Unit
Peak value
© 2002 PROCEQ SA Settings 11
Clear Data
" Select the menu option "Clear Data".
The message "Clear Data?" appears on the display.
Warning!
Data cannot be cleared individually. Clearing data always means that the entire memory is deleted.
" Press the "START/RESET" key to delete the entire memory.
4.2.2
Sample No.
A sample number is assigned to every new measuring procedure; the number is shown in the measuring display (Fig. 3.5).
With this menu option you can assign individual consecutive sample numbers for different measurement series.
For example, measurement series A could start with
"001000" whereas measurement series B starts with
"002000". In this way you can organize your measuring results systematically and adapt them to different measured objects.
" Select the menu option "Sample No.".
The following submenu appears in the display (11)
(without the values shown in the example):
Sample number (example)
Fig. 4.3
Sample No.
" Set the desired sample number using the arrow keys #$ %&
4.2.3
Sample Dimension
Measuring procedures can be carried out using test discs with different diameters. In order for the measuring results to be calculated accordingly, the surface of the test disc must be entered via this menu option.
Settings in this menu option depend on the setting "Unit" (see chapter 4.2.7)
No settings are required for measurements with a value for the tensile force (kN, kg, 10 3 kg or kip).
12 Settings © 2002 PROCEQ SA
The following settings must be performed for measurements with a value for the surface load (N/mm 2 , MPa, kg/cm 2 or psi ):
" Select the menu option "Sample Dimension".
The following submenu appears in the display (11):
Diameter Ø 50 mm
Size of sample surface
Fig. 4.4
Sample Dimension
" If you use a test disc with the diameter Ø 50 mm, select "Test Disc 50mm".
The tensile force is converted into the tension with a surface area of 1963 mm2 and shown in the measuring display.
" If you use another test disc, select "A=".
" Set the required surface area in mm 2 using the arrow keys %& .
The sample dimension can only be entered in mm 2 . If you use another unit of measurement
(e.g. inch), you have to convert the surface area into mm 2 accordingly.
4.2.4
k-Factor
The k-Factor serves for converting the electric signal emitted by the measuring cell (8, see Fig. 2.2) of the pull-off tester into the unit of the indicated measured value (force or tension).
To achieve maximum precision of the measured values, the pull-off tester with the DYNAMETER display unit is calibrated with ten load levels at our factory. These values are indicated in the menu option, "Calibration F10"
(see chapter 4.2.5, Fig. 4.6).
" Check to see if the serial number of the DYNAMETER matches the serial number of the pull-off tester (see chapter 3.2.2).
" Select the menu option "k-Factor".
© 2002 PROCEQ SA Settings 13
The following submenu appears in the display (11):
Factory calibration
Manual input
Fig. 4.5
k-Factor
" Select position C.
The pull-off tester uses the k-Factor values of the factory calibration for operation (see chapter 4.2.5).
" If you want to use the display unit with a different pulloff tester, select position m.
" Set the k-Factor of the pull-off tester using the arrow keys #$ %& .
Specifications for the k-Factor can be found in the calibration certificate or on the type plate of the pull-off tester, depending on the model.
The measured values are calculated over the entire force area with the same k-Factor. This means that the displayed values deviate slightly more (approx.1.5%) from the actual value (due to the non-linearity of the sensor).
14 Settings
Nominal tensile force
Only use the pull-off tester with factory calibration for tests, in accordance with the relevant standards.
For this purpose always select position C
(F10 calibration, see also chapter 4.2.5).
4.2.5
Calibration F10
The k-Factors determined during the factory setting are displayed for your information in the "Calibration F10" menu option (see Fig 4.6). Calibration is done in steps of 10% over the entire load area (10% - 100% of the nominal load).
" Select the menu option "Calibration F10".
The following submenu appears in the display (11):
Preselected unit
Load levels k-Factors of the individual load levels
Fig. 4.6
Calibration F10
© 2002 PROCEQ SA
Warning!
The pull-off testers are available as a calibrated unit (pull-off tester - DYNAMETER display unit). The nominal tensile force in the display
(see Fig. 4.6) must be equivalent to the nomi-
nal value of the pull-off tester. The nominal
load and displayed k-Factor values of the
individual load levels must not be changed by the user. The values are only displayed for information purposes.
For new calibration procedures, PROCEQ offers separate calibration instructions to the testing authority.
4.2.6
Peak Value
With the menu option "Peak Value" you can set how values are to be displayed and stored.
" Select the menu option "Peak Value".
The following submenu appears in the display (11):
" Select the desired position using the arrow keys
#$ .
DYNAMETER offers two basic possibilities for the dis-
play and storage of values:
If you select "Memory on", the DYNAMETER automatically stores the peak value after the force has significantly decreased by approx. 1% of the nominal value during a measuring procedure (if the sample breaks or in the case of manual relief).
If you select "Memory off", you can perform measuring procedures with loads up to the specified test force. In this case the data of the current measurement must be stored manually.
To document that the test force has been reached proceed as follows:
" When the specified test force has been reached press the "PRINT/STORE" key on the keyboard to store the current value.
Fig. 4.7
Peak Value
© 2002 PROCEQ SA Settings 15
4.2.7
Unit
With the menu option "Unit" you can select the units below, depending on the specifications:
- the unit for the force: kN, kg, 10 3 kg, kip or
- the unit for the tension (force per surface unit):
N/mm 2 , MPa, kg/cm 2 , psi to be displayed in the measuring display.
" Select the menu option "Unit".
The following submenu appears in the display (11):
4.2.8
Language
" Select the menu option "Language".
The following submenu appears in the display (11):
Fig. 4.8
Unit
" Select the desired unit.
The measured values are converted into the corresponding unit and shown in the measuring display.
When you select a unit for the tension (N/mm 2 , MPa, kg/cm 2 , psi), you need to ascertain and enter the value for the test surface (see chapter 4.2.3).
16 Settings
Fig. 4.9
Language
" Select the desired dialog language.
4.2.9
Flow Bar
The flow bar indicates the tendency of the load increase or decrease which is influenced with the crank of the pull-off tester.
With the menu option "Flow Bar" you can control the
display speed of the bar movement. This setting also influences the display speed of the bar movement of the load pacer (see chapter 4.2.10).
© 2002 PROCEQ SA
This will not influence the speed of load application and the duration of the measuring procedure, which is determined by means of the load pacer (see chapter 4.2.10).
" Select the menu option "Flow Bar".
The following submenu appears in the display (11):
With these values set, the measuring procedure takes approx. 40 s.
" Select approx. 2/3 of the maximum speed for the flow bar.
If this speed is set, the flow bar and the load pacer bar move across the measuring display once from left to right.
During longer measuring procedures, the flow bar and the load pacer bar might move across the measuring display from left to right several times.
This avoids the display speed being too low.
You can change the display speed according to your requirements. The display speed is also influenced inversely proportional to the specified speed of load application (see chapter 4.2.10).
Fig. 4.10
Flow Bar
" Use the arrow keys
#$ to set the display speed.
Example of possible settings:
- k-Factor: Position C (see chapter 4.2.4)
- Specified speed of load application: 0.1kN/sec
(see chapter 4.2.10)
- Test disc: Ø 50 mm (see chapter 4.2.3)
- Sample breaks at: 2N/mm 2 (= 3.9kN)
© 2002 PROCEQ SA Settings 17
4.2.10 Load Pacer
With the menu option "Load Pacer" you can specify the speed of load application prescribed by the relevant standards.
While the measuring procedure is performed, the bar movement of the load pacer determines the flow bar´s speed of movement (see chapter 4.2.9).
The unit for the preselection must always be "kN / s" even if another unit had been selected for the display.
" Select the menu option "Load Pacer".
The following submenu appears in the display (11):
Conversion example:
The speed of the load increase (in accordance with EN
1015-12, chapter 8) is between 0.003 and 0.100 N/
(mm 2 x s).
For a medium speed of 0.050 N/(mm 2 xs) and a test disc with a diameter of Ø 50 mm, the speed is as follows in kN/s:
0.050 x 1963 x 0.001 ~ 0.10 kN/s
Mean velocity Surface of test disc
(Ø 50 mm) in mm 2
Conversion from N to kN
Fig. 4.11
Load Pacer
" Set the required speed of load application in "kN / s" using the arrow keys #$ %& .
" If necessary, convert the existing data into kN/s.
18 Settings © 2002 PROCEQ SA
5
5.1
Measuring Process
Warning!
Before performing a measuring procedure, read through the relevant chapters for the required settings (see chapter 4). Incorrect settings will result in wrong measuring results!
Warning!
Observe the relevant standards and regulations when performing preparatory work or settings for a measuring procedure (e.g., making cores).
Preparing the Measuring Procedure
Core
Fig. 5.1 Preparing the test surface
Test disc 6
Fig. 5.2
Test surface with stuck on test disc
" Cut into the test surface with a core drill (if necessary) so that the break can occur over the defined circular surface (see Fig. 5.1).
" Attach the test disc (6) to the test surface using an adequate adhesive (see chapter 8, Appendix).
" Wait until the required adhesive strength is reached.
© 2002 PROCEQ SA
12
3
4
14
13
15
16
7
6
3 Hydraulic pump
4 Piston
6 Test disc
7 Draw bolt
12 Crank
13 Wheel
14 Draw spindle
15 Coupling
16 Legs
Fig. 5.3
Preparing the pull-off tester
If the pull-off tester is to be stored for a longer period of time (2-3 months), move the piston (4) up and down 2-3 times by turning the crank (12) of the hydraulic pump (3) in a clockwise and counterclockwise direction. This reduces friction losses caused by the piston seals.
Warning!
Avoid hasty, jerky movements, particularly when the piston is in its initial and end
position, because otherwise the piston might jam (see also chapter 6.2.1, page 24).
Measuring Process 19
" Turn the crank (12) back to its initial position in a counterclockwise direction until slight resistance is encountered (see Fig. 5.3).
" Then turn the crank (12) once in a clockwise direction
(to relieve the hydraulic system).
This ensures that the full piston stroke is available during the measuring procedure.
" Connect the coupling (15) of the draw spindle (14) to the draw bolt of the test disc (6).
" Turn the wheel (13) clockwise until slight resistance is encountered.
" Position the pull-off tester in such a way (1) that the tensile force is applied perpendicular to the test surface:
To achieve this, adjust the legs (16) of the pull-off tester (1) until no "pulling at a slant" can occur.
"Pulling at a slant" during measuring will falsify the measuring result and cause increased wear of the pull-off tester. Premature repair or new calibration might become necessary!
" After alignment, slightly release the draw spindle (14) with the wheel.
5.2
Measuring with the Digital Manometer
Fig. 5.4
Display of the digital manometer
" Ensure that the pull-off tester (1) is prepared in accordance with chapter 5.1.
" Ensure that the digital manometer (2) is switched off or switch it off.
" Then press the "ON" key to switch on the digital manometer (2).
The pull-off tester is now ready for measuring. The current value is displayed.
" Press the "PEAK" key if the digital manometer (2) is to indicate the peak value.
If the digital manometer is not under load and switched on in PEAK mode, the peak value may increase slowly to approx. 20 digits. This will not influence the measuring result.
20 Measuring Process © 2002 PROCEQ SA
12
14
13
7
6
6 Test disc
7 Draw bolt
12 Crank
13 Wheel
14 Draw spindle
Fig. 5.5
Measuring Procedure
" Turn the crank (12) steadily clockwise until the specified test force is reached, or until the test surface breaks.
" Always press the "ON" key after the measurement is complete to switch off the device or perform a new measuring procedure.
" After each measurement, turn the crank (12) back counterclockwise until slight resistance is encountered.
" Then turn the crank (12) maximum one turn in a clockwise direction.
The pull-off tester is now ready for a new measuring procedure.
© 2002 PROCEQ SA
5.3
Measuring with the DYNAMETER
5.3.1
Preparing the DYNAMETER
" Check all settings in accordance with chapter 4.2.
" After checking the settings and if the pull-off tester is not under load, press the "END" key to return to the measuring display.
" Start the measuring procedure.
5.3.2
Carrying out the Measuring Procedure
" Ensure that the pull-off tester (1) is prepared in accordance with chapter 5.1.
" Press the "START/RESET" key.
To avoid mistakes, press the "START/RESET" key before each new measuring procedure.
Measuring Process 21
12 Crank
6 Test disc
9 DYNAMETER
Fig. 5.6
Measuring procedure
" Turn the crank (12) steadily clockwise to increase the force on the test surface.
The movement of the crank directly influences the speed of the flow bar in the measuring display (see chapter 4.2.9).
" Turn the crank (12) just enough for the flow bar to follow the speed of the load pacer flow bar.
22
Depending on the display speed (see chapter
4.2.9) and the value of the specified test force, it might be necessary to move the flow bar 2 or 3 times from left to right before the specified peak value is reached or the sample breaks.
Measuring Process
Two types of data storage are now possible depending on the option you have selected (see chapter 4.2.6):
Measuring with the
"Memory on" setting
After the force of the pulloff tester has dropped by approx. 1% of the nominal load, "Peak" flashes in the measuring display and the peak value is displayed and stored.
The DYNAMETER stores the data of
- Peak value
- Sample no.
- Test surface (if preselected, see chapter 4.2.3).
Measuring with the
"Memory off" setting
" When the specified value has been reached press the "PRINT/
STORE" key:
The DYNAMETER stores the data of
- Current value
- Sample no.
- Test surface (if preselected, see chapter 4.2.3).
The "PRINT/STORE" key can also be pressed several times during a measuring procedure (if sample changes, e.g. a fracture, occur). The values will then be stored in this order.
" After every measurement, turn the crank (12) back counterclockwise until slight resistance is encountered.
© 2002 PROCEQ SA
" Then turn the crank (12) one turn in a clockwise direction.
If you want to carry out a new measuring procedure:
" Prepare the pull-off tester (1) according to the instructions (see chapter 5.1).
" Check all settings (see chapter 4.2).
" Press the "START/RESET" key.
The peak value in the measuring display is deleted and the sample number is increased by 1.
The pull-off tester (1) is now ready for a new measuring procedure.
5.3.3
Display of the Measured Values
After every measuring procedure the stored values can be displayed directly on the DYNAMETER (9):
" Select the menu option "Data Output" in the main menu (see chapter 4.2.1).
" Select "Display Data" in the submenu.
" Select the desired sample number using the arrow keys #$ .
5.3.4
Storage Capacity of the DYNAMETER
The memory capacity is 1000 groups. When the memory is full, no further values will be accepted.
For this reason it is important that you transmit the data to a printer or PC in time (see chapter 4.2.1).
Afterwards you can delete the memory (see chapter
4.2.1).
© 2002 PROCEQ SA Measuring Process 23
6 Maintenance, Storage and Care
6.1
Cleaning After Each Use
Warning!
Never immerse the DYNAMETER and pull-off tester in water or wash it under a running tap!
Use neither abrasives nor solvents for cleaning!
6.1.1
Pull-Off Tester
" Clean the pull-off tester with a clean, dry cloth.
" If necessary, clean the gaps with a pneumatic pump
(no compressed air!).
" Clean the digital manometer or DYNAMETER with a clean, dry cloth after use.
" Clean dirty input sockets and connectors of the DYNA-
METER with a clean, dry brush.
6.2
Performance Check Before Each Use
6.2.1
Pull-Off Tester
" Ensure that the pull-off tester is complete.
" Ensure that the coupling (15, see Fig. 5.3) is screwed in correctly.
" Fix the legs (16) in their basic position.
" Check to see if there are traces of oil on the pull-off tester.
Warning!
Oil is an indication of leaking seals. Check the oil level immediately!
Checking the oil level
H
A
: Height of initial position
H
E
: Height of end position
H
E
H
A
Fig. 6.1
Measuring the difference of level
24 Maintenance © 2002 PROCEQ SA
" Turn the crank (12) back counterclockwise until slight resistance is encountered (initial position).
" Measure the distance H
A
(see Fig. 6.1) between the highest point of the device plate (5) and the outer top edge of the piston (4) in this position.
" Now turn the crank (12) clockwise until slight resistance is encountered (end position).
" Measure the distance H
E tion.
(see Fig. 6.1) in this posi-
The difference of level ∆H (H
E
- H
A
) between the initial and end position should be 3 - 4 mm.
" Turn the crank (12) back counterclockwise until it reaches its initial position.
" Then turn the crank (12) one turn in a clockwise direction.
Warning!
If the difference of level is too small, the oil volume will not be sufficient. Immediately send the device to an authorized service center.
" Check the cables for damage.
" Check the battery capacity (see chapter 3.2.2).
If a battery life of 4 to 6 hours is indicated on the DYNAMETER when it is started up, you should take a new set of batteries to the measuring site.
6.3 Storage
" Store the DYNAMETER and pull-off tester in the original case in a clean, dust-free room.
" If the device is not used for a long period of time, remove the batteries from the DYNAMETER or digital manometer.
6.4
Maintenance
6.4.1
Pull-Off Tester and Display Units
We recommend performing a general check and new calibration of the pull-off tester and DYNAMETER every
two years.
Warning!
Any warranty or liability claims will expire if maintenance work is carried out improperly.
For this reason, always send the pull-off tester and its components to a service center authorized by PROCEQ.
© 2002 PROCEQ SA Maintenance 25
7 Data
7.1
Form of Delivery
DYNA Pull-off tester Z16 / Z6 with digital manometer
DYNA Pull-off tester Z16E / Z6E with DYNAMETER
Model Z with digital manometer
- Test disc Ø 50 mm
- Draw bolt with connection thread M8
- 1 battery for digital manometer 9V AM6/6LR61 (120h)
- Carrying case 300 x 280 x 240 mm
- Calibration Certificate
- Operating instructions
Weight of pull-off tester: 3.5 kg
Total weight: 6 kg
26 Data
Model Z...E with display unit DYNAMETER
DYNA Pull-off tester:
- Test disc Ø 50 mm
- Draw bolt with connection thread M8
- Carrying case
300 x 280 x 240 mm
Total weight: 6 kg
Display unit DYNAMETER
- Display unit
- Carrying strap
- 6 batteries LR, 1.5V (60h)
- Carrying case
320 x 295 x 105 mm
- Calibration Certificate
- Operating instructions
Total weight: 1.8 kg
© 2002 PROCEQ SA
7.2
Accessories
Test discs
Article no.
Material Dimensions
Test discs Standard versions
345 08 001
345 08 002
Aluminum
Aluminum
Quantity/ Thread Used with devices
Dia. 50 x 25 mm 10 pcs.
Dia. 50 x 25 mm 10 pcs.
M 8
M12
Z 6
'
Z 16
'
Z 25
'
Test discs Special versions
345 00 782 Steel Dia. 50 x 25 mm 10 pcs.
345 08 051
345 08 053
Aluminum
Aluminum
50 x 50 x 25 mm 10 pcs.
Dia. 100 x 25 mm 10 pcs.
345 08 009
345 08 008
Aluminum Dia. 20 x 20 mm 1 pcs.
Aluminum Dia. 100 x 25 mm 1 pcs.
M 8
M 8
M 8
M 8
M12
'
'
'
'
'
'
'
'
'
Draw bolt
345 09 002
345 10 002
Steel
Steel
M 8
M12
' '
'
Z 50
'
'
'
Draw bolts
345 09 007
345 12 007
Electric drive, compl. WAF 12
Electric drive, compl. WAF 14
'
'
'
'
Electric drive
© 2002 PROCEQ SA Data 27
7.3
Technical Data
7.2.1
Pull-Off Tester Z with Digital Manometer
Z6 Z16
Max. tensile force: 6 kN
Thread of draw spindle:
M20 x 1
16 kN
M20 x 1
Thread of draw bolt: M8
Permissible stroke 4 mm
(with pump):
Accuracy:
Resolution:
< 2 %
0.01 N/mm 2
M8
3.5 mm
< 2 %
0.01 N/mm 2
7.3.2
Pull-Off Tester Z...E with DYNAMETER
Z6E Z16E Z25E Z50E
Max. tensile force: 6 kN 16 kN 25 kN 50 kN
Thread of draw spindle: M20 x 1 M20 x 1 M22 x 2 M24 x 2
Thread of draw bolt: M8 M8 M12 M12
Permissible stroke 4 mm 3.5 mm 5 mm 6 mm
(with pump):
Accuracy:
Resolution:
< 1 % < 1 % < 1 % < 1 %
0.001kN 0.01kN
0.01kN 0.01kN
28 Data
7.3.3
Display Unit DYNAMETER
- Non-volatile memory for 1000 measured values
- Display on 128 x 128 graphics LCD
- Interface RS 232 C
- Integrated software for the transmission of measured values to the printer and/or PC
- Temperature range: -10°C to +60°C
- Battery operation with 6 batteries for approx. 60 hours
© 2002 PROCEQ SA
Rapid adhesives for Pull-off Tester
3M Scotch-Weld DP 100/DP 110
Araldit Rapide
Araldit 2012
CHEMO-Resin EP-KL-11
MG Spezialkleber
Sikadur 31 Normal/Rapid
Silikal R I/21
Technovit 2190
Tepicolle KB 60 G
X60
3M (Schweiz) AG
Ciba Spezialitätenchemie AG
Ciba Spezialitätenchemie AG
Krämer Chemie GmbH
Hannelore Moser
Sika AG
Karl Ulrich & Co AG
Kulzer & Co. GmbH
Alfred Klotz AG
Karl Bubenhofer AG
HBM Darmstadt
Schenk AG for use on wet surfaces
Araldit AV 144 / 2-HV 997
Araldit AV 2101-HV 2957 for high environment temperature
Araldit AV 138M-Hv 998
© 2002 PROCEQ SA
Ciba Spezialitätenchemie AG
Ciba Spezialitätenchemie AG
Ciba Spezialitätenchemie AG
CH-8803 Rüschlikon
CH-4057 Basel
CH-4057 Basel
D-66130 Saarbrücken-Güdingen
D-79183 Waldkirch 3
CH-8048 Zürich
D-63533 Mainhausen 1
D-61350 Bad Homburg
CH-8040 Zürich
CH-9200 Gossau
CH-8606 Nänikon
CH-4057 Basel
CH-4057 Basel
CH-4057 Basel
Appendix 29

Download
Advertisement
Key features
- Digital manometer or DYNAMETER display
- Measurement of adhesive strength and adhesion
- Storage of up to 1000 measured values
- Data transfer to printer or PC
- Online display on a PC or laptop
- Adjustable legs for different test situations
- Various test disc diameters available
- Calibration certificate included